Caution: Federal US law restricts sale of the device identified in this manual to, or on the order of, a licensed physician.
Welch Allyn assumes no responsibility for any injury, or for any illegal or improper use of the product, that may result from failure to
use this product in accordance with the instructions, cautions, warnings, or indications for use published in this manual.
Welch Allyn is a registered trademark of Welch Allyn, Inc. CP 50, CP 50 Plus, and CardioPerfect are trademarks of Welch Allyn, Inc.
PDI and Sani-Cloth are registered trademarks of PDI, Inc. CaviWipes is a trademark of Metrex Research Corporation.
Software in this product is copyright Welch Allyn or its vendors. All rights are reserved. The software is protected by United States of
America copyright laws and international treaty provisions applicable worldwide. Under such laws, the licensee is entitled to use the
copy of the software incorporated within this instrument as intended in the operation of the product in which it is embedded. The
software may not be copied, decompiled, reverse-engineered, disassembled or otherwise reduced to human-perceivable form. This
is not a sale of the software or any copy of the software; all right, title and ownership of the software remains with Welch Allyn or its
vendors.
Welch Allyn Technical Support
European Customer Care + 35 3 46 906 7790France + 33 155 69 58 49
Germany + 49 6950 985 132China + 86 21 6327 9631
Latin America (Miami office) + 1 305 669 9003Netherlands + 31 202 061 360
This document is written for clinical professionals with a working knowledge of medical
procedures and terminology required for monitoring cardiac patients.
Before using the electrocardiograph for clinical applications — or before setting up,
configuring, troubleshooting, or servicing the electrocardiograph — you must read and
understand this document and all other information that accompanies the
electrocardiograph and related options or accessories.
Intended use
The Welch Allyn electrocardiography products (subject devices) are intended for use by
trained operators in health facilities. The subject devices provide the following diagnostic
functions:
•Acquiring, viewing, storing, and printing ECG waveforms using ECG front-end
modules (patient cables) and associated accessories that provide signal acquisition for
up to twelve (12) leads of patient ECG waveforms through surface electrodes adhered
to the body.
•Using optional algorithms to generate measurements, data presentations, graphical
presentations, and interpretative statements on an advisory basis. These are
presented for review and interpretation by the clinician based upon knowledge of the
patient, the result of physical examination, the ECG tracings, and other clinical
findings.
Indications for use
The electrocardiograph is one of the tools that clinicians use to evaluate, diagnose, and
monitor patient cardiac function.
The 12-lead ECG interpretive algorithm provides a computer-generated analysis of
potential patient cardiac abnormalities which must be confirmed by a physician with other
relevant clinical information.
Auto ECG A report typically showing a 10-second acquisition of 12 leads of ECG
information combined with patient data, measurements, and optional interpretation.
Auto ECGs can be saved to the electrocardiograph’s test directory or to a USB massstorage device. In CP 50 Plus models, auto ECGs can also be sent to a Welch Allyn
CardioPerfect workstation.
Stat ECG An auto ECG that starts instantly without waiting for you to enter patient data
or adjust the waveforms.
Rhythm ECG A continuous, real-time printout of rhythm strips with a user-defined lead
configuration. Rhythm ECGs are printouts only. They cannot be saved.
Auto ECG report example
A: 3 rows, 4 columns. B: Patient data. C: ECG measurements. D: Automatic ECG interpretation (optional). E: Report status
label. F: Rhythm leads. G: Gain, speed, AC filter, frequency range, date and time, software version, device ID, practice ID.
H: Calibration pulse.
Rhythm ECG report example
Directions for useIntroduction3
Features
Pacemaker detection
The software detects the possible presence of a pacemaker. If you confirm that the
patient has a pacemaker, the ECG report includes no interpretation, and it indicates that a
pacemaker was detected.
Automatic ECG interpretation (optional)
The optional MEANS interpretation algorithm, developed by the University of Rotterdam
in the Netherlands, provides automatic analysis of ECG tests. For more information, see
the MEANS Physicians' Manual or the PEDMEANS Physicians’ Manual on the CD that
came with your electrocardiograph. The MEANS algorithm is used for adult patients 18
years and older. The PEDMEANS algorithm is used for pediatric patients from 1 day
through 17 years old.
Separate the battery from other
disposables for recycling.
Separate the device from other
disposables for recycling.
See www.welchallyn.com/weee.
Miscellaneous symbols
Heartbeats per minuteHot surface. Do not touch.
ManufacturerDate of manufacture
Defibrillation-proof Type CF applied partsDefibrillation-proof Type BF applied parts
Reorder numberSerial number
Atmospheric air-pressure limits
Recycle.
China RoHS (restriction of hazardous
substances) symbols for control of pollution
caused by electronic information products.
5-year environment-friendly use period (EFUP)
for batteries. 10-year EFUP for the device.
For details, see the accompanying
documentation.
Do not reuse.Batch code
Meets essential requirements of European
European community representative
Medical Device Directive 93/42/EEC.
Software activation code
Directions for useIntroduction7
General warnings
Warnings indicate conditions or practices that could lead to illness, injury, or death.
Warnings related to the environment
WARNING To avoid a possible explosion, do not use the electrocardiograph in
the presence of flammable anesthetics: mixtures containing air, oxygen, or
nitrous oxide.
WARNING When transporting the electrocardiograph on a cart, tuck the patient
cable away from the wheels so that it does not present a hazard.
Warnings related to accessories and other equipment
WARNING For operator and patient safety, peripheral equipment and
accessories that can come in direct patient contact must be in compliance with all
appropriate safety, EMC, and regulatory requirements.
WARNING All signal input and output (I/O) connectors are intended for
connection of only devices complying with IEC 60601-1, or other IEC standards
(for example, IEC 60950), as appropriate to the device. Connecting additional
devices to the electrocardiograph might increase chassis or patient leakage
currents. To maintain operator and patient safety, consider the requirements of
IEC 60601-1-1. Measure the leakage currents to confirm that no electric shock
hazard exists. In the case of a USB printer, the printer (non-medical electrical
equipment) shall be situated outside the patient environment (reference IEC
60601-1-1). The printer used should be approved to the appropriate safety
standard for non-medical electrical equipment (IEC 60950, or its national variants),
and use of an isolation transformer is recommended. If there is a requirement for
the printer to be situated within the patient environment it is the responsibility of
the user to ensure that the system provides a level of safety in compliance with
IEC 60601-1 and 60601-1-1.
WARNING The electrocardiograph has not been designed for use with highfrequency (HF) surgical equipment and does not protect against hazards to the
patient.
WARNING This device captures and presents data reflecting a patient’s
physiological condition. When reviewed by a trained physician or clinician, this
data can be useful in determining a diagnosis. However, the data should not be
used as a sole means for determining a patient’s diagnosis or prescribing
treatment.
WARNING To avoid serious injury or death, take these precautions during
patient defibrillation:
•Avoid contact with the electrocardiograph, patient cable, and patient.
•Verify that the patient leads are properly connected.
•Place defibrillator paddles properly in relation to electrodes.
•After defibrillation, pull each patient lead out of the patient cable and inspect
the tips for charring (black carbon marks). If there is any charring, the patient
cable and individual leads must be replaced. If there is no charring, fully
reinsert the leads into the patient cable. (Charring can occur only if a lead is
not fully inserted into the patient cable before defibrillation.)
WARNING To prevent the spread of infection, take these precautions:
•Dispose of single-use components (for example, electrodes) after using them
once.
•Regularly clean all components that come in contact with patients.
•Avoid ECG testing for patients with open, infectious sores.
WARNING Avoid positioning any leads or cables so that they could easily trip
someone or become wrapped around a patient’s neck.
WARNING To ensure safe use of the device, follow the documented
maintenance procedures.
WARNING Only qualified service personnel should attempt to repair the
electrocardiograph. In case of a malfunction, call Technical Support.
WARNING Do not perform ST segment analysis on the ECG screen display,
since these ECG representations are scaled. Make manual measurements of
ECG intervals and magnitudes on printed ECG reports only.
WARNING To avoid injury, do not touch the print head immediately after printing.
It might be hot.
Directions for useIntroduction9
General cautions
Cautions indicate conditions or practices that could damage the equipment or other
property.
Caution When removing the electrocardiograph from storage, allow it to
thermally stabilize to surrounding environmental conditions before using it.
Caution To prevent possible damage, do not use sharp or hard objects to press
the touch screen or the buttons. Only use fingertips.
Caution Do not expose the patient cable to strong ultraviolet radiation.
Caution Do not pull or stretch the patient cable. Doing so could result in
mechanical or electrical failures. Form the patient cable into a loose loop before
storing.
Caution Avoid positioning the patient cable where it might get pinched,
stretched, or stepped on. Otherwise, measurements might no longer be
accurate, and repair might be necessary.
Caution Using the equipotential terminal for anything but grounding purposes
may contribute to damage of the device.
Caution Use only parts and accessories, including thermal paper, that were
supplied with the device and available through Welch Allyn. The use of
accessories other than those specified may result in degraded performance or
unsafe use of this device.
Caution Portable and mobile RF communications equipment can affect the
performance of the electrocardiograph.
Caution The electrocardiograph meets the Class A requirements of
IEC 60601-1-2 regarding incidental emission of radio frequency interference. As
such it is suitable for use in commercial grade electrical environments. If the
electrocardiograph is used in residential grade electrical environments and you
experience incidental interference with other equipment that uses radio
frequency signals to operate, minimize the interference.
Caution Other medical equipment—including but not limited to defibrillators,
ultrasound machines, pacemakers, and other stimulators—may be used
simultaneously with the electrocardiograph. However, such devices may disturb
the electrocardiograph signal.
Caution The power cord must be disconnected from AC power before
cleaning, maintaining, transporting, or servicing.
Caution The requirements of AAMI EC11, Section 3.2.7.2, Frequency and
Impulse Response, for an impulse triangle waveform may be impacted by up to
5 milliseconds of small amplitude dampened ringing immediately after the
impulse when the muscle filter (35 Hz) is turned on or a small amplitude offset
when the baseline filter (0.5 Hz) is turned on. These requirements are unaffected
by any other combination of filters turned on or off. Measurements performed by
the optional interpretation algorithm are unaffected by any filter selections.
11
2
Setup
Connecting the patient cable
WARNING Do not allow the conductive parts of the patient cable, electrodes, or
associated connections of defibrillation-proof applied parts, including the neutral
conductor of the patient cable and electrodes, to come into contact with other
conductive parts, including earth ground. Otherwise, an electrical short might
result, risking electric shock to patients and damage to the device.
WARNING To avoid injury to the patient or damage to the device, never plug
patient leads into any other device or wall outlet.
Caution Always connect the patient cable and the leads properly during
defibrillation. Otherwise, the connected leads could be damaged.
Loading...
+ 35 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.