Caution: Federal US law restricts sale of the device identified in this manual to, or on the order of, a licensed
physician.
Welch Allyn assumes no responsibility for any injury, or for any illegal or improper use of the product, that may result
from failure to use this product in accordance with the instructions, cautions, warnings, or indications for use
published in this manual.
Welch Allyn is a registered trademark of Welch Allyn, Inc. CP 150, and CardioPerfect are trademarks of Welch Allyn,
Inc.
Patent information
For patent information, please visit www.welchallyn.com/patents.
PDI and Sani-Cloth are registered trademarks of PDI, Inc. CaviWipes is a trademark of Metrex Research Corporation.
Software in this product is copyright Welch Allyn or its vendors. All rights are reserved. The software is protected by
United States of America copyright laws and international treaty provisions applicable worldwide. Under such laws,
the licensee is entitled to use the copy of the software incorporated within this instrument as intended in the
operation of the product in which it is embedded. The software may not be copied, decompiled, reverse-engineered,
disassembled or otherwise reduced to human-perceivable form. This is not a sale of the software or any copy of the
software; all right, title and ownership of the software remains with Welch Allyn or its vendors.
This document is written for clinical professionals with a working knowledge of medical
procedures and terminology required for monitoring cardiac patients.
Before using the electrocardiograph for clinical applications — or before setting up,
configuring, troubleshooting, or servicing the electrocardiograph — you must read and
understand this document and all other information that accompanies the
electrocardiograph and related options or accessories.
1
Intended use
The Welch Allyn electrocardiography products (subject devices) are intended for use by
trained operators in health facilities. The subject devices provide the following diagnostic
functions:
•
Acquiring, viewing, storing, and printing ECG waveforms using ECG front-end
modules (patient cables) and associated accessories that provide signal acquisition
for up to twelve (12) leads of patient ECG waveforms through surface electrodes
adhered to the body.
•Using optional algorithms to generate measurements, data presentations, graphical
presentations, and interpretative statements on an advisory basis. These are
presented for review and interpretation by the clinician based upon knowledge of the
patient, the result of physical examination, the ECG tracings, and other clinical
findings.
Indications for use
The electrocardiograph is one of the tools that clinicians use to evaluate, diagnose, and
monitor patient cardiac function.
The 12-lead ECG interpretive algorithm provides a computer-generated analysis of
potential patient cardiac abnormalities which must be confirmed by a physician with
other relevant clinical information.
Contraindications
The electrocardiograph has no known contraindications.
•The electrocardiograph allows users to perform 12-lead ECG measurements and
analysis.
•The electrocardiograph supports STAT, Auto, and Rhythm test types.
•The electrocardiograph provides the ability to print test records on an internal printer
or external network printer.
•The electrocardiograph allows storage of test records in device memory, external
storage media, and external software applications.
Features
Pacemaker detection
The software detects the possible presence of a pacemaker. If you confirm that the
patient has a pacemaker, the ECG report includes no interpretation, and it indicates that
a pacemaker was detected.
Automatic ECG interpretation (optional)
The optional MEANS interpretation algorithm, developed by the University of Rotterdam
in the Netherlands, provides automatic analysis of ECG tests. For more information, see
the MEANS Physicians’ Manual or the PEDMEANS Physicians’ Manual on the CD that
came with your electrocardiograph. The MEANS algorithm is used for adult patients 18
years and older. The PEDMEANS algorithm is used for pediatric patients from 1 day
through 17 years old.
CAUTION Check for the presence of a pacemaker before using ECG with
interpretation.
Directions for use Introduction 3
Controls, indicators, and connectors
No. FeatureDescription
1LCD screen800 x 480 pixels color touchscreen provides a graphical user
interface.
2Power switch and LEDPower-on/Standby switch.
The LED indicates the charging status when connected to AC
power:
•Green: The battery is charged.
•Amber: The battery is charging.
3Patient cable connectorProvides connection for patient cable.
4PrinterPrinter provides a printout of patient Auto ECG, Stat ECG, or
1Ethernet connectorProvides a hardwired connection to the computer network.
The LEDs indicate active network status when the ethernet cable
is connected to a network.
2Clients USBUSB, type "mini B." Provides connection to an enabled host.
3Host USBUSB, type "A." Provides four host USB connections for optional
accessories.
4Power connectionProvides an external AC power connection.
5AC fuseProvides access to AC fuse.
6Ground lug (equipotential terminal)Provided for electrical safety testing and as a means for
connection of a potential equalization conductor.
7Battery compartment (behind cover)Houses the Li-ion battery.
Directions for use Introduction 5
Symbols
Documentation symbols
Power symbols
WARNING The warning statements in
this manual identify conditions or
practices that could lead to illness, injury,
or death.
Consult operating instructions.
Power on/standby
(LED off) No AC power is presentThe battery is absent or faulty
(LED green) AC power is present, and
battery is charged. (LED amber) AC power
is present, and battery is charging.
Caution The caution statements in this
manual identify conditions or practices
that could result in damage to the
equipment or other property, or loss of
data. This definition applies to both
yellow and black and white symbols.
Separate the battery from other
disposables for recycling
Separate the device from other
disposables for recycling.
See www.welchallyn.com/weee.
Lithium ion battery
Miscellaneous symbols
Manufacturer
Reference/Model numberSerial number
Recycle
China RoHS (restriction of hazardous
substances) symbols for control of
pollution caused by electronic
information products. 5-year
environment-friendly use period (EFUP)
for batteries. 10-year EFUP for the device.
For details, see the accompanying
documentation.
Defibrillation-proof Type CF applied parts
Reorder/Catalog number
Batch code
Directions for use Introduction 7
Intertek ETL listedDo not reuse
Electromagnetic compatibility (EMC) information
For information about electromagnetic compatibility (EMC), see the Welch Allyn website.
Warnings indicate conditions or practices that could lead to illness, injury, or death.
Warnings related to the environment
WARNING The power cord is considered the disconnect device for
isolating this equipment from supply mains. Do not position the equipment
so that it is difficult to reach or disconnect.
WARNING To avoid a possible explosion, do not use the
electrocardiograph in the presence of flammable anesthetics: mixtures
containing air, oxygen, or nitrous oxide.
WARNING When transporting the electrocardiograph on a cart, tuck the
patient cable away to keep them clear of the wheels and to minimize trip
hazards.
Warnings related to accessories and other equipment
WARNING To avoid the risk of electric shock, this equipment must only
be connected to a supply mains with protective earth.
WARNING For operator and patient safety, peripheral equipment and
accessories that can come in direct patient contact must be in compliance
with all appropriate safety, EMC, and regulatory requirements.
WARNING All signal input and output (I/O) connectors are intended for
connection of only devices complying with IEC 60601-1, or other IEC
standards (for example, IEC 60950), as appropriate to the device.
Connecting additional devices to the electrocardiograph might increase
chassis or patient leakage currents.
WARNING The electrocardiograph has not been designed for use with
high-frequency (HF) surgical equipment and does not protect against
hazards to the patient.
WARNING Defective batteries can damage the electrocardiograph.
Visually inspect the battery at least monthly, if the battery shows any signs
of damage or cracking, it must be replaced immediately and only with a
battery approved by Welch Allyn.
WARNING Improper disposal of batteries may create an explosion or
contamination hazard. Never dispose of batteries in refuse containers.
Always recycle batteries according to local regulations.
Warnings related to using the electrocardiograph
WARNING No modification of this equipment is allowed.
WARNING This device captures and presents data reflecting a patient’s
physiological condition. When reviewed by a trained physician or clinician,
this data can be useful in determining a diagnosis. However, the data
should not be used as a sole means for determining a patient’s diagnosis or
prescribing treatment.
WARNING To provide CF protection use only accessories approved by
Welch Allyn. Visit www.welchallyn.com. The use of any other accessories
can result in inaccurate patient data, can damage the equipment, and can
void your product warranty.
WARNING To avoid serious injury or death, take these precautions during
patient defibrillation:
•Avoid contact with the electrocardiograph, patient cable, and patient.
•Verify that the patient leads are properly connected.
•Place defibrillator paddles properly in relation to electrodes.
•After defibrillation, pull each patient lead out of the patient cable and inspect
the tips for charring (black carbon marks). If there is any charring, the patient
cable and individual leads must be replaced. If there is no charring, fully
reinsert the leads into the patient cable. (Charring can occur only if a lead is
not fully inserted into the patient cable before defibrillation.)
WARNING To prevent the spread of infection, take these precautions:
•Dispose of single-use components (for example, electrodes) after using
them once.
•Regularly clean all components that come in contact with patients.
•Avoid ECG testing for patients with open, infectious sores.
WARNING Avoid positioning any leads or cables so that they could easily
trip someone or become wrapped around a patient’s neck.
WARNING To ensure safe use of the device, follow the documented
maintenance procedures.
WARNING Only qualified service personnel should attempt to repair the
electrocardiograph. In case of a malfunction, call Technical Support.
WARNING Do not perform ST segment analysis on the ECG screen
display, since these ECG representations are scaled. Make manual
measurements of ECG intervals and magnitudes on printed ECG reports
only.
WARNING To maintain diagnostic accuracy and to comply with IEC
60601-02-51 and IEC 60601-02-25, do not scale (re size) when sending a
saved ECG to an external printer.
Directions for use Introduction 9
WARNING To avoid injury, do not touch the print head immediately after
printing. It might be hot.
General cautions
Cautions indicate conditions or practices that could damage the equipment or other
property.
CAUTION Federal US law restricts sale of the device identified in this
manual to, or on the order of, a licensed physician.
CAUTION When removing the electrocardiograph from storage, allow it to
thermally stabilize to surrounding environmental conditions before using it.
CAUTION To prevent possible damage, do not use sharp or hard objects
to press the touch screen or the buttons. Only use fingertips.
CAUTION Do not expose the patient cable to strong ultraviolet radiation.
CAUTION Do not pull or stretch the patient cable. Doing so could result in
mechanical or electrical failures. Form the patient cable into a loose loop
before storing.
CAUTION Avoid positioning the patient cable where it might get pinched,
stretched, or stepped on. Otherwise, measurements might no longer be
accurate, and repair might be necessary.
CAUTION Using the equipotential terminal for anything but grounding
purposes may contribute to damage of the device.
CAUTION Use only parts and accessories, including thermal paper, that
were supplied with the device and available through Welch Allyn. The use
of accessories other than those specified may result in degraded
performance or unsafe use of this device.
CAUTION Portable and mobile RF communications equipment can affect
the performance of the electrocardiograph.
CAUTION The electrocardiograph meets the Class A requirements of IEC
60601-1-2 regarding incidental emission of radio frequency interference. As
such it is suitable for use in commercial grade electrical environments. If
the electrocardiograph is used in residential grade electrical environments
and you experience incidental interference with other equipment that uses
radio frequency signals to operate, minimize the interference.
CAUTION Other medical equipment—including but not limited to
defibrillators, ultrasound machines, pacemakers, and other stimulators—
may be used simultaneously with the electrocardiograph. However, such
devices may disturb the electrocardiograph signal.
CAUTION The power cord must be disconnected from AC power before
cleaning, maintaining, transporting, or servicing.
CAUTION The requirements of AAMI EC11, Section 3.2.7.2, Frequency
and Impulse Response, for an impulse triangle waveform may be impacted
by up to 5 milliseconds of small amplitude dampened ringing immediately
after the impulse when the muscle filter (35 Hz) is turned on or a small
amplitude offset when the baseline filter (0.5 Hz) is turned on. These filters,
in any combination turned on or off, meet the AAMI requirements.
Measurements performed by the optional interpretation algorithm are
unaffected by any filter selections.
NoteThe entire patient cable, up to and including the electrodes are considered
to be an Applied Part.
Setup
Connecting the patient cable
11
WARNING Do not allow the conductive parts of the patient cable,
electrodes, or associated connections of defibrillation-proof applied parts,
including the neutral conductor of the patient cable and electrodes, to
come into contact with other conductive parts, including earth ground.
Otherwise, an electrical short might result, risking electric shock to patients
and damage to the device.
WARNING To avoid injury to the patient or damage to the device, never
plug patient leads into any other device or wall outlet.
WARNING To provide CF protection use only accessories approved by
Welch Allyn. Visit www.welchallyn.com. The use of any other accessories
can result in inaccurate patient data, can damage the equipment, and can
void your product warranty.
CAUTION Always connect the patient cable and the leads properly during
defibrillation. Otherwise, the connected leads could be damaged.
Loading the thermal paper
The electrocardiograph prints on z-fold thermal paper.
Store the paper in a cool, dry, dark place.
•
•Do not expose it to bright light or UV sources.
•Do not expose it to solvents, adhesives, or cleaning fluids.
•Do not store it with vinyls, plastics, or shrink wraps.
SetupCP 150 12-lead resting electrocardiograph
12
Powering the electrocardiograph
The electrocardiograph runs on AC or battery power. Plug the electrocardiograph into AC
power as often as possible so that the built-in charger can keep the battery charged.
Regardless of the battery condition, you can use the electrocardiograph whenever it is
plugged in.
WARNING When you use AC power, always plug the electrocardiograph
into a hospital-grade outlet to avoid the risk of shock.
WARNING If the integrity of the building’s safety ground is in doubt,
operate this device on battery power to avoid the risk of shock.
To turn on or turn off
Press .
Verifying proper operation
To ensure the accuracy of test data, you should verify proper operation of the
electrocardiograph before it is used for the first time with patients. You should reverify
proper operation once a year.
Loading...
+ 36 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.