Welch Allyn assumes no responsibility for any injury to anyone, or for any illegal or improper use of the product, that may
result from failure to use this product in accordance with the instructions, cautions, warnings, or statement of intended use
published in this manual.
Welch Allyn, Propaq, and Smartcuf are registered trademarks of Welch Allyn. ParamSet is a trademark of Welch Allyn.
Masimo, SET, LNOP, and LNCS are registered trademarks, and FastSAT and APOD are trademarks, of Masimo Corporation.
Possession or purchase of a Masimo SpO
-equipped monitor does not convey any express or implied license to use the
2
device with unauthorized sensors or cables which would, alone or in combination with this device, fall within the scope of
one or more of the patents relating to this device.
Nellcor and Oximax are registered trademarks of Nellcor Puritan Bennett.
Software in this product is copyright by Welch Allyn or its vendors. All rights are reserved. The software is protected by
United States of America copyright laws and international treaty provisions applicable worldwide. Under such laws, the
licensee is entitled to use the copy of the software incorporated with this instrument as intended in the operation of the
product in which it is embedded. The software may not be copied, decompiled, reverse-engineered, disassembled or
otherwise reduced to human-perceivable form. This is not a sale of the software or any copy of the software; all right, title
and ownership of the software remain with Welch Allyn or its vendors.
For information about any Welch Allyn product, contact your local Welch Allyn representative: www.welchallyn.com/about/
company/locations.htm.
DIR 80020372 Ver B
Revision date: 2016-07
Welch Allyn, Inc.
4341 State Street Road
Skaneateles Falls, NY 13153 USA
www.welchallyn.com
Regulatory Affairs Representative
Welch Allyn LimitedNavan Business Park
Dublin Road
Navan, County Meath
Republic of Ireland
Index ...................................................175
Page 5
1
1
Introduction
Intended use
The Propaq LT Series (802LT0N, and 802LT0S) monitors are portable devices intended to
be used by clinicians and medically qualified personnel for single- or multiparameter
vital-signs monitoring of ambulatory and nonambulatory patients, including neonate,
pediatric and adult patients. These devices are indicated for ECG, noninvasive blood
pressure (NIBP), respiration and SpO2. The most likely locations for patients to be
monitored by these devices are hospital general medical-surgical, telemetry, and
intermediate care floors, hospital emergency departments, transport, emergency medical
services, and other healthcare applications.
Federal USA law restricts this device to sale, distribution, or use by or on the order of a
licensed healthcare professional.
Even though this manual describes some monitoring techniques, the monitor is intended
for use only by trained and experienced clinicians who know how to measure and
interpret vital signs.
Contraindications
Symbols
There are no known contraindications for this monitor.
Table 1. Directions for use
WARNING The warning statements in this manual identify conditions or practices that
could lead to illness, injury, or death. Warning statements appear with a yellow background in a black and
white document.
CAUTION The caution statements in this manual identify conditions or practices that could
result in damage to the equipment or other property, or loss of data. This definition applies
to both yellow and black and white symbols.
Follow instructions/directions for use (DFU) -- mandatory action.
A copy of the DFU is available on this website.
A printed copy of the DFU can be ordered from Welch Allyn for delivery within 7 calendar
days.
Power - Turn the monitor power on or off.Down - Move the cursor down to the next display
Action - Act based on what is highlighted.
(See “Using the action button” on page 25.)
Left - Move the cursor left to the next display
item; decrease the parameter value.
Right - Move the cursor right to the next
display item; increase the parameter value.
Up - Move the cursor up to the next display
item.
Cradle
Monitor Release - Press and then hold while removing the monitor from the cradle.
item.
Alarm reset - Silence the current alarm tone for
90 seconds or reset a silenced alarm tone.
Display - Cycle to the next configured display
format, or cancel the current control, setup, or
pop-up menu.
Snapshot - Recorda 21
and waveform data.
Start/Stop NIBP - Start or stop an NIBP
measurement.
-second period ofnumeric
Table 3. Status indicators
Monitor
(green flashing) Monitoring normally (no active alarms).
Low priority alarm
(yellow) At least one alarm is disabled.
Medium priority alarm
(yellow flashing) Technical alarm.
High priority alarm
(red flashing) Patient alarm.
Upper and lower alarm limits for this parameter are on.
Upper and lower alarm limits for this parameter are disabled.
High priority alarm active.
High priority alarm active, Paused.
Medium or low priority alarm active.
Medium or low priority alarm active, paused.
A snapshot exists for this period.
The snapshot for this period has been replaced with a more recent snapshot.
Page 7
Directions for UseIntroduction3
Table 3. Status indicators (continued)
(green) The battery is fully charged.
The battery is partially full.
The battery is partially full and is charging.
(yellow) The battery is low.
The battery is low and is charging.
(red)The battery is near failure; the monitor will shut down soon. If this indicator appears
while the monitor is in the cradle, the battery cannot be charged and must be replaced.
The battery is near failure and is charging. The monitor will shut down if removed from the cradle.
(green) Cradle is powered.
(green) Monitor battery is charging.
NOTE: When the battery is fully charged, this indicator is not lit.
(yellow) Cradle fault or battery fault.
Table 4. Labels
Proceed with caution. If in doubt, refer to the
accompanying documentation.
The monitor or accessory meets all essential
requirements of the European Medical Device
Directive 93/42/EEC for a Class II-b product.
Professional use onlyType CF patient connections, isolated for
See the accompanying manual.Hazard Class 9, IATA/ICAO (International Air
Reorder numberProduct identifier
Recycle the monitor, cradle, and battery
separately from other waste. (Refer to
www.welchallyn.com/weee for collection-
point and additional information.)
High voltage. Do not touch during
defibrillation.
Recycle the battery separately from other
disposables.
Li ++
Lithium
Rechargeable batteryNIBP/PSNI Non Invasive Blood Pressure
Monitor
Enclosure protection: Drip
per EN60529:1991.
Intertek ETL Listed Mark
indicating successful
independent safety testing.
direct cardiac application and protected
against defibrillation.
Transport Association/International Civil
Aviation Organization).
Direct current.
Battery replacement specification.
This monitor is approved for use during all
phases of flight aboard U.S. Army aircraft.
AWR: MIL STD: 461E. Army: CE101, CS101,
T, CS115, S116, RE102, RE103.
The monitor is safe for patients and clinicians when used in accordance with the
instructions and with the warning and caution statements presented in this manual.
All personnel must read and understand all warning and caution statements
presented in this manual before using the monitor.
•Failure to understand and observe any warning statement in this manual could
lead to patient injury, illness, or death.
•Failure to understand and observe any caution statement in this manual could
lead to equipment damage or loss of patient data.
Operator Position
The operator of this device should be positioned within 1 meter of the front panel at an
angle not greater than 20 degrees.
Page 9
Directions for UseIntroduction5
Disconnect AC power
WARNING Never move the monitor, cradle or mobile stand by pulling on any of
the cords. This may cause the monitor to tip over or damage the cord. Never pull
on the power cord when disconnecting the cord from the mains outlet. When
disconnecting the power cord, always grasp the attachment plug. Keep the cord
away from liquids, heat, and sharp edges. Replace the power cord if the strain
relief or cord insulation is damaged or begins to separate from the attachment
plug.
1. Grasp the attachment plug.
2. Pull the attachment plug from the mains outlet.
Note
The monitor automatically switches to battery power when AC mains power is
interrupted.
These statements apply to all aspects of patient monitoring. Statements which apply
specifically to one aspect of monitoring, such as NIBP or SpO2monitoring, are presented
in the corresponding sections of the manual.
WARNING Many environmental variables, including patient physiology and
clinical application, can affect the accuracy and performance of the monitor. The
clinician must verify all vital-signs information prior to patient intervention.
WARNING Always check the patient mode (adult, pediatric, or neonate) when
monitoring a new patient. The patient mode determines default alarm limits and
internal algorithm settings. Make sure the monitor has settings that are
appropriate before monitoring the patient.
WARNING The power cord is the disconnect device to isolate this equipment
from supply mains. Position the equipment so that it is not difficult to reach or
disconnect the cord.
WARNING The monitor might not meet its performance specifications if stored
or used outside the specified temperature and humidity ranges.
WARNING Use of respiration monitoring by impedance pneumography can
affect the operation of some pacemakers. If pacemaker operation is affected, turn
off respiration pneumography. (See Figure 52 on page 53.)
WARNING Do not connect more than one patient to a monitor.
WARNING Do not connect more than one monitor to a patient.
WARNING Electric shock hazard. Do not allow the conductive parts of the
patient cable, electrodes, or associated connections of defibrillation-proof applied
parts, including the neutral conductor of the patient cable and electrodes, to
come into contact with other conductive parts, including earth ground. An
electrical short might result, risking electric shock to patients and damage to the
device.
WARNING Do not operate this product in the presence of flammable
anaesthetics or other flammable substances in combination with air or
oxygen-enriched environments. Failure to observe this warning can result in an
explosion.
WARNING Do not use the monitor in a Magnetic Resonance Imaging (MRI)
suite or a hyperbaric chamber. Such use can cause fire or explosion resulting in
patient injury and monitor damage.
WARNING Do not operate this monitor near equipment that emits strong
electromagnetic or radio-frequency signals. Electronic equipment of this type can
cause electrical interference with monitor operation, which can distort the ECG
signal and prevent accurate rhythm analysis.
Page 11
Directions for UseIntroduction7
WARNINGWARNING Pacemaker signals can differ from one pacemaker to the next. The
Association for Advancement of Medical Instrumentation (AAMI) cautions that “in
some devices, rate meters may continue to count the pacemaker rate during
occurrences of cardiac arrest or some arrhythmias. Do not rely entirely upon rate
meter alarms. All pacemaker patients should be kept under close or constant
observation.” See “Pacer pulse rejection” on page 138 for disclosure of the
pacemaker pulse rejection capability of this instrument.
WARNING Use only accessories approved by Welch Allyn. The use of any other
accessories can result in inaccurate patient data, can damage the equipment, and
can void your product warranty. Refer to the accessory list or
www.welchallyn.com.
WARNING Patient injury risk. Use only accessories approved by Welch Allyn,
including electrodes, lead wires, and patient cables. These approved accessories
are required for electrical protection of the patient during cardiac defibrillation.
WARNING Always use accessories according to the standards of your facility
and according to the manufacturer's directions.
WARNING Use only ECG cables supplied or specified by Welch Allyn. Use of
any other ECG cables can negate defibrillator protection and can create a risk of
patient injury due to shock.
WARNING Frequently inspect—electrically and visually—all cables, sensors, and
electrode wires. Replace any damaged cables, sensors or wires. Failure to
properly inspect and keep in excellent working order all cables, sensors, and
electrode wires can result in hazards to patients and to equipment failure and
damage.
WARNING Always properly connect the electrosurgery return circuit. Improper
circuit connection can cause current to return through monitor electrodes and
probes, creating a burn hazard for patients.
WARNING Always keep patient motion to a minimum. Motion artifact can cause
inaccurate measurement of patient vital signs.
WARNING Carefully route and secure patient cabling, using the supplied
garment clips. Improperly routed and secured cabling can cause the patient to
become entangled in the cables, creating a strangulation hazard.
WARNING When the patient is wearing the monitor or being transported by
stretcher with the monitor connected, always take care to position the monitor
carrying straps on the patient. Be certain that the straps do not and cannot cross
the neck or throat and cause choking, and the straps do not restrict movement of
the patient’s arms or legs.
WARNING Never use a monitor that is not working properly. If the monitor is not
working properly, patient waveforms might be inaccurate or might not be
displayed.
WARNING If the monitor is damaged, or if you see any indication that the
monitor is not operating properly, disconnect it from the patient. Do not return it
to service until it has been inspected and, if necessary, repaired by qualified
service personnel.
WARNING High-power radars are allocated as primary users of the bandwidth
between 5.25 GHz and 5.35 GHz and between 5.65 GHz and 5.85 GHz. These
radars can cause interference with this device and can damage this device.
WARNING Do not use the pulse oximeter as a replacement or substitute for
ECG-based arrhythmia analysis.
WARNING The bedside patient monitor is the primary alarming source for the
patient.
WARNING The leading cause of patient death or serious injury reported with the
use of patient monitoring equipment is failure to respond to alarms notifying the
user of an adverse change in patient condition. If you are relying on visual alarm
notifications, maintain a clear line of sight and remain within 4 meters of the
monitor. If you are relying on audio alarm notifications, make sure that you can
hear audio alarms from where you are. Set the volume as needed considering the
environment and ambient noise levels. Verify that the alarm is audible to a
clinician working at the maximum distance from the monitor.
WARNING False alarms may occur in some situations. You must understand
and address the cause of the false alarms whenever possible to eliminate the
possibility of repeated false alarms and alarm fatigue, which might result in a
failure to respond to an actual alarm situation.
WARNING Electric shock hazard. Do not allow patient to make contact with
system input ports (input connectors, such as SpO
, ECG, etc.) and system
2
output connectors (output port, such as USB). Additionally, you should not make
contact with the patient and the system input or system output connectors at the
same time.
WARNING Electric shock hazard. All system input and output connectors are
intended for connecting devices that comply with IEC 60601-1 or other IEC
standards (for example, IEC 60950), as appropriate to the device. Connecting
additional devices to the electrocardiograph might increase chassis or patient
leakage currents.
WARNING Defective batteries can damage the monitor. Visually inspect the
battery at least monthly, if the battery shows any signs of damage or cracking, it
must be replaced immediately and only with a battery approved by Welch Allyn.
WARNING Patient injury risk. Keep the monitor, reusable electrodes, and the
patient cable clean. Patient contact with contaminated equipment can spread
infection.
WARNING Patient injury risk. The monitor has not been designed for use with
high-frequency (HF) surgical equipment and does not protect against burns to the
patient.
WARNING Patient injury risk. The data captured from this monitor should not be
used as a sole means for determining a patient’s diagnosis or prescribing
treatment.
WARNING To avoid serious injury or death, take precautions consistent with
good clinical practice during patient defibrillation:
•Avoid contact with the electrocardiograph, patient cable, and patient.
•Place defibrillator paddles properly in relation to electrodes.
Page 13
Directions for UseIntroduction9
WARNING To avoid serious injury or death, take precautions consistent with
good clinical practice during patient defibrillation:
•Before defibrillation, verify that patient leads are properly connected.
•After defibrillation, pull each patient lead out of the patient cable and inspect
the tips for charring (black carbon marks). If there is any charring, the patient
cable and individual leads must be replaced. If there is no charring, fully
reinsert the leads into the patient cable. (Charring can occur only if a lead is
not fully inserted into the patient cable before defibrillation.)
WARNING Patient injury risk. Ensure that each test identifies the patient to
avoid the risk of associating reports with the wrong patients. If any report does
not identify the patient, either write the patient identification information on the
report immediately following the ECG test or enter identifying information before
sending test results.
WARNING To maintain diagnostic accuracy and to comply with IEC60601-02-51
and IEC 60601-02-25, do not scale (resize) when sending a saved ECG to an
external printer.
WARNING Patient injury risk. To prevent cross-contamination and the spread of
infection:
General cautions
Note
Note
•Dispose of single-patient use components (for example, electrodes) after use.
•Regularly clean and disinfect all components that come in contact with
patients.
•Reprocess the device accessories (for example, patient cable, leads, and
reusable electrodes) between patients.
Caution Do not autoclave the monitor.
Caution Autoclave accessories only if the manufacturer's instructions clearly
direct you to do so. Many accessories can be damaged by autoclaving.
Caution Changes or modifications not expressly approved by Welch Allyn could
void the purchaser’s authority to operate the equipment.
Caution This product contains no user-serviceable components. Any
unauthorized changes to the product invalidate Welch Allyn’s warranty and also
invalidate all applicable regulatory certifications and approvals.
This monitor is suitable for use in the presence of electrosurgery.
•Monitoring of neonate, pediatric and adult patients
•Display of ECG, SpO2, and Resp waveform traces
•Accurate reading of NIBP in the presence of motion artifact, using Welch Allyn’s
patented Smartcuf®motion-tolerant technology
•SpO2monitoring with advanced technology for accuracy under conditions of low
perfusion
•Configurable adjustments to alarm limits with ParamSet™technology
•Standalone operation with local physiological and technical alarms
•Color LCD for display of numerics and waveform data
•Configurable display formats and monitoring capabilities
•Rechargeable lithium-ion battery
•Weight of approximately 2 pounds (0.9 kg)
•Durability
•Tolerance of brief exposure to water
•HIPAA support
•Error detection
Models
The monitor is available in two models.
FeatureModel
3-lead and 5-lead ECGxx
Respiration rate (Resp)xx
Masimo SpO
Nellcor
Noninvasive blood pressure (NIBP)xx
Cradle to recharge the monitor batteryxx
USBOptionOption
Upload patient data from the monitor to a PC
and download custom monitor configurations
from a PC to the monitor
USB option
The monitor can be purchased with the optional USB data transfer capability, which
enables communication between the monitor and a PC. (See “About the USB data
transfer option” on page 17.)
802LT0N
2
®
SpO
2
OptionOption
Model
802LT0S
x
x
Page 17
Directions for UseIntroduction13
HIPAA considerations
Each medical facility is responsible for creating and enforcing policies and procedures to
guarantee compliance with the regulations defined in 45 CFR 160-164 of the Health
Insurance Portability and Accountability Act (HIPAA) of 1996.
The Propaq LT Monitor, the Propaq LT Monitor Configuration Utility, and the Propaq LT
Monitor AutoPrint Utility incorporate security features that support your implementation
of the HIPAA requirements for ensuring that patient information is kept private and
confidential.
Monitor
•Clinicians can lock the monitor display to prevent the display of patient vital signs. For
patient protection when the monitor display is locked, the display is restored instantly
in the event of a button press (unless button-pad lock-out is enabled), or an alarm.
•Clinicians can lock the monitor buttons to prevent any unauthorized access to the
monitor controls. To protect the patient when the buttons are locked, access to the
buttons is restored instantly in the event of an alarm.
Configuration utility
The Configuration Utility never contains patient data.
AutoPrint utility
•All patient data is stored in a Welch Allyn proprietary data format which is readable
only by machine.
•All patient data is deleted from the PC when it is sent to the printer.
The monitor runs through an operational self-test each time it is powered on. Always
verify that it follows the power-on sequence shown below. When Step 3 completes as
described, the monitor has verified that the visual and audible alarm indicators are
working properly. If Step 4 does not complete as described, remove the monitor from
service and have it examined by a qualified service person.
To turn on the monitor
1. Press and holduntil a tone sounds.
2. A tone sounds as the green, yellow, and red lights turn on and off. This occurs twice.
3. The splash screen appears, identifying the product and displaying the message
“Diagnostics in progress”. The green light turns on and off quickly while the low tone
sounds, the yellow light turns on and off quickly while the medium tone sounds, and
the red light turns on and off quickly while the loud tone sounds.
WARNING To ensure patient safety, listen for audible indicators and watch for
visual alerts at power-up. Correct any system errors before using the monitor. In
addition to the audible indicators, the screen Status area displays color coding,
icons, and messages that help you to distinguish clinical priority and actions, if
needed.
Figure 6. Splash screen
Propaq
LT
Diagnostics in progress
Application Version 1.00.00
Bootcore Version 1.00.00
c
Welch Allyn 2005
4. The splash screen is replaced by one of the power-on screens shown here:
Page 20
16Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
Select ( ) to Start a New Patient.
No data saved .
Start New Patient
Info
Demo
Portl and Wests ide Hospital
Emerg ency Department
Select ( ) to delete data
and start a new patient.
Portl and Wests ide Hospital
Emerg ency Department
Note!
There is pati ent data stored for
Hall, Robert E. ID: 3456187
Start New Patient
Continue Patient
Info
Demo
English
Deutsch
Français
Español
Italiano
Polski
Nederl
Svenska
Português
Japanese
Figure 7. Power-up screens
Portland Westside Hospital
Emergency Department
No data saved.
Select ( ) to Start a New Patient.
Start New Patient
Info
To start monitoring from this point, see “Standalone monitoring” on page 43. To practice
using the monitor, see “Using demo mode” on page 35.
Selecting a language
To change the language of the monitor interface
1. If the monitor power is on, pressto turn it off.
2. Pressto turn the monitor power on.
3. Immediately after pressing, and before any screen appears, simultaneously press
and. Keep the buttons pressed until the language selection screen appears.
Demo
Portland Westside Hospital
Emergency Department
There is patient data stored for
Hall, Robert E. ID: 3456187
Select ( ) to delete data
Start New Patient
and start a new patient.
Continue Patient
Note!
Info
Demo
Figure 8. Language selection screen
English
Deutsch
Français
Español
Italiano
Polski
Nederl
Svenska
Português
Japanese
4. Press,, , andto highlight the language you wish to use.
5. Press.
Power-on continues in the selected language.
Note
If you select French, HR/PR alarm limits cannot be turned off unless you go
through the service menus to change this setting. If you do change this setting,
the change stays in effect through power cycles.
For information about using the service menus, refer to the Propaq LT MonitorService Manual (810-2712-XX).
The monitor always powers on in this language until the setting is changed again. The
setting can be changed again only if one of the following occurs:
•This procedure is repeated.
Page 21
Directions for UseOverview of monitor operation17
•A new configuration is downloaded from a PC. (See “Using the Configuration
Utility to configure the monitor” on page 104).
About the charging/communications cradle
When the charging/communication cradle (the cradle) is attached to AC power (or to
vehicle DC power) and the monitor is seated in the cradle, the cradle does the following:
•Recharges the monitor battery, whether patient monitoring is occurring or not.
•Powers the monitor, conserving the charge on the monitor battery.
•(If the cradle has the USB data transfer option) enables data transfer between a
PC and a monitor. See “About the USB data transfer option” on page 17.
WARNING Never download a configuration file to the monitor while you are
monitoring a patient with the monitor in the cradle.
•Changing the monitor configuration shuts down the monitor and deletes all
patient data from the monitor.
Figure 9. Monitor and cradle
The cradle is intended to hold the monitor when the patient is in bed or is otherwise
stationary. The cradle can sit on a flat surface near the patient or attach to the bed rail
(using the bed-rail hook).
If the monitor is in the cradle and the AC power adapter is connected to the cradle, the
monitor runs on AC power rather than battery power. This keeps the monitor battery at full
charge so that the monitor can then run on battery power when it is removed from the
cradle to accompany the patient away from the bed.
The cradle can be connected to AC power at all times, whether the monitor is present
or not.
About the USB data transfer option
Monitor
Cradle
A cradle configured with the optional USB communication port and connected to a PC
with the Propaq LT Configuration Utility and the Propaq LT AutoPrint Utility can be used
for the following:
•Creating custom monitor configurations on the PC and downloading them to any
number of monitors. (See “Monitor configuration” on page 91.)
•Uploading the configuration file from the monitor to the PC.
Page 22
18Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
•Uploading patient data from the monitor to the PC for printing. (See “Printing
patient data” on page 111.)
Setting up the cradle
1. Place the cradle on a table or shelf, or hang it on a bed rail.
WARNING Place the cradle so that it cannot fall on the patient.
Caution Locate the cradle near the patient but not so close that it interferes
with patient care.
2. Connect the AC adapter to an AC power outlet and to the cradle. The indicator
(green) on the front of the cradle indicates that the AC power adapter is connected.
Figure 10. Cradle status indicators
Green Monitor battery charging
Yellow Battery or charger failed
(Indicates either that the temperature of
the battery pack is out of range or that
service is required.)
Green AC power connected
3. Insert the monitor into the cradle. When the monitor is properly seated, it clicks into
place.(green) on the cradle indicates that the monitor battery is charging. (See
“Recharging the battery” on page 133.)
Note
The monitor can be inserted into the cradle when power is on or off. If monitor
power is on, inserting the monitor in the cradle or removing the monitor from the
cradle does not interrupt patient monitoring.
Page 23
Directions for UseOverview of monitor operation19
Removing the monitor from the cradle
To remove the monitor, depress the release button on the front of the cradle. With the
button depressed, hold the cradle securely with one hand, grasp the monitor firmly with
the other hand, and lift the monitor out.
Figure 11. Monitor release button
Monitor release button
Page 24
20Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
Displaying data
Patient vital signs appear on the 3.5-inch (diagonal measurement) monitor display.
Figure 12. Components of the vital-signs display (A)
1
234
5
6
7
0
Patient name (if available)
1
Waveform source
2
Waveform display scale
3
Patient ID (if available)
4
Time of day
5
Patient room number
6
Patient mode
7
Figure 13. Components of the vital-signs display (B)
8
9
14131211
10
Page 25
Directions for UseOverview of monitor operation21
Battery status (See Table 3, “Status indicators” on page 2.)
8
greenpartially to fully charged
yellownearly discharged
reddischarged; the monitor will shut down soon
Color-coded waveforms, ECG Lead identifiers, and vital-signs numerics
9
Green
Cyan
Purple
Yellow
SpO2pulse amplitude
10
SpO2numeric data
11
Respiration rate numeric data
12
NIBP numeric data
13
HR (heart rate) is displayed if ECG is active.
14
PR (pulse rate) is displayed if ECG is not active and SpO
The monitor indicates an HR/PR measurement outside the measurable range as follows:
---out
+++ out
???undetermined
ECG and HR/PR
NIBP
Resp
SpO
2
-of-range low
-of-range high
or NIBP is active.
2
Page 26
22Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
140/78
80
12
97
%
HAL L, ROBERT E.
345 6187
3:0 0:06P
Ad ult
Rm 23 9
HR/min
NIBP mmHg (102)
@2:47P Manual
Resp/min
SpO2
SpO22x2x
II 1mV/cm
About display formats
The monitor can be configured to display any of these formats. By default, the Large
Numerics display format cannot be removed from the configuration:
Table 5. Display formats
Large numerics
Blood pressure, heart rate, respiration rate, and
SpO2measurements displayed in large numerics.
Bell icons indicating...
Lower alarm limit
Upper alarm limit
Single waveform
•3 seconds of one ECG or SpO2signal or ...
•12 seconds of the Resp signal.
Vital signs displayed in medium-sized numerics.
HALL, ROBERT E.
II 1mV/cm
SpO2
140/78
80
HR/min
NIBP mmHg (102)
@2:47P Manual
Adult
3456187
Rm 239
12
Resp/min
3:00:06P
97
SpO2
Tabular trends
Current vital signs displayed above the table in small
numerics.
Historical vital signs displayed in a table.
Dual waveform
•6 seconds of an ECG or SpO2signal or 24
seconds of the Resp signal or...
•Any two of the following: 3 seconds of an ECG
or SpO2signal and 12 seconds of the Resp
signal.
Other vital signs displayed numerically below the
waveforms.
Tabular trends with single waveform
•3 seconds of an ECG or SpO2signal or...
•12 seconds of the Resp signal.
Other vital signs displayed numerically below the
waveforms.
Historical vital signs displayed in a table below the
waveform.
Page 27
Directions for UseOverview of monitor operation23
The monitor can be configured to cycle quickly through three of the five available display
formats when you press. For example, if the configuration specifies three formats—
Large Numerics, Single Waveform, and Tabular Trends—you can pressrepeatedly
to cycle through those formats.
Note
Figure 14. Cycling through the configured display formats
The Large Numerics display cannot be removed from the three screen display
cycle.
Single waveform
Large numerics
Tabular trends
Certain properties of each display type can also be configured. See “Display Format” on
page 123.
To view a tabular display if no tabular trends format is specified
1. Access the Setup menu. (See “To access the setup menus” on page 31.)
2. Highlight Trends and press.
Note
All valid display configurations include at least one waveform format.
Timing out the display and the back light
To conserve battery run life, the monitor display and the back light can be configured to
turn off if no operator activity (that is, a button press) is detected for a specified number of
minutes.
•When the display is off, no patient data is visible.
•When the back light is off, patient data is visible only under direct light.
Note
If the display and the back light are turned off due to a time out, they turn on again
immediately when an alarm occurs or, if the buttons lock-out is not enabled, when
you press any monitor button.
The monitor configuration determines whether the time-out feature is enabled or
disabled, and defines the default time-out period if it is enabled. If the feature is enabled in
Page 28
24Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
the configuration, you can temporarily change the time-out period or disable the time out
through the Setup -> Timings menu.
Locking out the display, back light, and buttons
If lock-outs are enabled, you can lock out one or more of the following elements:
ButtonsTo prevent unauthorized use
Back Light To extend the battery run life
DisplayTo prevent unauthorized viewing of patient information
The monitor configuration defines which of these, if any, can be locked out. For
information on enabling or disabling lock-outs for these elements, see “Monitor
configuration” on page 91.
To lock out the configured elements
Hold down,, andsimultaneously for 5 seconds.
To unlock the configured elements
Hold down,, andsimultaneously for 5 seconds.
Note
Alarms immediately unlock any locked elements.
Page 29
Directions for UseOverview of monitor operation25
About navigation
You navigate the monitor screens using,,and (arrow buttons),(action button),
and(display button).
Using the arrow buttons
Use,,andto do the following:
•Highlight an item on the display. (See “Using the highlights” on page 25.)
•Select options from a control menu.
•Useandto select options from a pop-up menu.
•Useandto change the values of numeric parameters.
Using the action button
Useto do the following:
•Display the control menu for a blue-highlighted item.
•Return from a control menu to the primary display.
•Access the Setup menu when Setup is highlighted.
•Display tabular and graphical trends when Trends is highlighted.
•Display snapshots when Snapshot is highlighted.
•Turn on the display or the back light if either has been turned off by a time-out.
•Display a pop-up menu.
Using the display button
Useto do the following:
•Cycle through the configured display formats.
•Return from a control menu to the primary display.
•Close a pop-up menu.
Using the highlights
Every screen contains a single element—the current context—highlighted by a blue field.
Some screens also contain elements—parameter values—highlighted by a green field.
About blue highlights
A blue highlight identifies the current context. For example, Figure 15 illustrates a
highlighted row in a trends display and a highlighted setting in the Waveform Size menu.
Page 30
26Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
Waveform Size
0.2 mV/cm
0.5 mV/cm
1 mV/cm
2 mV/cm
4 mV/cm
8 mV/cm
140/78
80
12
97
%
HAL L, ROBERT E.
345 6187
3:0 0:06P
Ad ult
Rm 23 9
HR/min
NIBP mmHg (102)
@2:47P Manual
Resp/min
SpO2
SpO22x2x
II 1mV/cm
Figure 15. Examples of highlighted elements
Drop-down menu
Highlights
Waveform Size
0.2 mV/cm
0.5 mV/cm
1 mV/cm
2 mV/cm
4 mV/cm
8 mV/cm
Vital-signs display
In a display screen (see “About display formats” on page 22), pressingcauses the
monitor to replace the current screen with another screen related to the current context.
For example, if SpO2 is highlighted in the Two waveforms display and you press...
Figure 16. Using the action button ()
HALL, ROBERT E.
II 1mV/cm
SpO2
140/78
80
HR/min
NIBP mmHg (102)
@2:47P Manual
Adult
3456187
Rm 239
12
Resp/min
3:00:06P
97
SpO2
(Two waveforms display)
With SpO2highlighted, press
...the monitor presents the SpO
About green highlights
Green highlights identify the current values of parameters within a given context. For
example, in the control menu shown in Figure 17, the current settings of the SpO
parameters are highlighted in green.
control menu (Figure 17).
2
2
Page 31
Directions for UseOverview of monitor operation27
Menus
Using control menus
Figure 17. SpO2control menu (example)
Control context
Parameters
Current settings
Control menu
A control menu includes a topic name for the current context (for example, SpO2); a
column of parameters with one highlighted (for example, SpO2 Monitoring); and a
column of options, with one item in each set of options highlighted (for example, On, On,
100, On, 90, Low).
•The blue highlight indicates the parameter currently enabled for modification.
•The green highlights indicate the current settings for all parameters in the menu.
At the bottom of the screen for all control menus are links to Exit, Trends, Snapshots,
and Setup.
ExitReturn to the vital-signs display.
TrendsView a tabular history.
SnapshotsView a series of 21-second waveform snapshots of the current patient’s
vital signs.
SetupAccess the setup menu. (See “To access the setup menus” on page 31.)
Example: Using a control menu
Using the example (Figure 17), you would do the following to raise the SpO2lower alarm
limit to 95 (Step 1) and shut off the HR/PR tone (Step 2):
1. With SpO2 Monitoring highlighted, scroll (using) to highlight Lower Limit, and
pressas many times as needed to raise this alarm limit to 95.
Note
If you decrease an upper alarm limit to a value almost as low as the lower limit,
the lower limit decreases so that it is always lower than the upper limit.
If you increase a lower alarm limit to a value almost as high as the upper limit, the
upper limit increases so that it is always higher than the lower limit.
Page 32
28Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
2. Scroll (using) to HR/PR Tone, and press eitheroras many times as needed to
highlight Off.
Note
When you change a setting (for example, by turning off an alarm limit or by
increasing or decreasing an alarm limit), the change takes effect immediately.
3. Pressorto exit the control screen and return to the vital-signs display.
Note
When you exit a control menu, the values displayed at the time you exit are the
values in effect for the monitor. If you change a parameter setting and then decide
before exiting the control menu to keep the previous setting values, you must
return the parameters to the original values before you exit the control menu.
30Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
About the Resp control menu
ParameterOptionsParameterOptions
Resp MonitoringOff OnLower AlarmOff On
Limit AlarmOff OnLower Limit
Upper LimitAdult2 - 148
Adult4 - 150Pediatric2 - 148
Pediatric4 - 150Neonate3 - 148
Neonate5 - 150Resp LeadLd1 (RA-LA) Ld2 (RA
-LL)
Using setup menus
Use the Setup menus—Alarms, ECG, NIBP, SpO2, and Timings—to define settings for
monitor behavior.
SettingOptions
Alarms
Suspend Audible AlarmsOff, On (with a time value)
Alarm ToneLow, Med, High
ECG
Resp MonitoringOff, On
ECG BandwidthMonitor, Extended
Power Source Filter60 Hz, 50 Hz, Off
Pacer IndicatorOff, On
NIBP
NIBP FormatSD, SD(m), sd(M)
NIBP UnitsmmHg, kPa
SmartcufOff, On
(Masimo only)
SpO
2
SensitivityNormal, Maximum, APOD
FastSAT
Averaging Time2, 4, 8, 10, 12, 14, 16
Timings
Back Light Time Out (Min)2, 5, 10, 15, 30, On, Off
Off, On
Display Time Out (Min)2, 5, 10, 15, 30, On
Demo ModeDisabled, Low, High
Page 35
Directions for UseOverview of monitor operation31
Use the Setup -> Configuration menu to change the monitor default configuration
without using the Configuration Utility. Any changes to the configuration settings requires
a password. Contact your service administrator.
OptionNote
Configuration
Select ( ) to save current settings.Replace the power-up defaults with the current
Select ( ) to use factory settings.Replace the power-up defaults with the factory
Select ( ) to use file: [file name]Replace the power-up defaults with the last
parameter settings.
settings.
downloaded configuration settings.
Figure 19. Confirm Factory Settings
To access the setup menus
From any main display screen (such as Large Numerics, Dual Waveform...):
1. Highlight, HR/PR, SpO2, NIBP, Resp,or.
2. Press.
3. Highlight Setup (at the bottom of the screen) and press.
Figure 20. Setup menus
Setup
Page 36
32Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
Note
If you change parameter settings and then change the patient mode (from adult to
pediatric, for example):
•All parameters are reset to the configuration default values for the new
patient mode.
•All stored patient data is lost.
WARNING The Setup menus are also used to access the Service menu. Do not
enter the Service menu unless you are a qualified service person.
Page 37
Directions for UseOverview of monitor operation33
Welch Allyn Propaq 802LT0N
Serial# AB72383-1 V1.00.00
Portland Westside
Emergency Department
PtldWstsdED10Jun05.mnt
Ann Jones, MD
503-530-0101 x9999
Patient Mode
Wireless Communications
Start New Patient
Info
Demo
Adult
Snapshots
14 of 20 saved
Disabled
Continue Patient
About monitor information screens
Both the start-up information screen and the monitoring information screen provide
information about the monitor.
To view the start-up information screen
From the power-on screen (Figure 7 on page 16), highlight Info.
Figure 21. Start-Up Information
Welch Allyn Propaq 802LT0N
Serial# AB72383-1 V1.00.00
Portland Westside
Emergency Department
PtldWstsdED10Jun05.mnt
Ann Jones, MD
503-530-0101 x9999
Patient Mode
Wireless Communications
Adult
Snapshots
Disabled
14 of 20 saved
Start New Patient
Continue Patient
Info
Demo
The start-up information screen provides the following:
•Monitor type, serial #, and software version # (V X.XX.XX)
•Medical facility name and department
•Configuration file name
•Contact person’s name and telephone number
•Current patient mode (Adult, Pediatric, Neonate)
•Number of data snapshots saved (0 - 20)
•Network communications status (Disabled)
To exit the start-up information screen, do one of the following:
•Highlight Continue Patient or Start New Patient to start monitoring.
•Highlight Demo to enter Demo mode.
•Pressto turn off the monitor.
Page 38
34Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
503-530-0101 x9999
Emergency Department
PtldWstsdED10Jun05.mnt*
Ann Jones, MD
Portland Westside
To view the monitoring information screen
Highlight(in the upper right corner of the vital-signs display) and press.
Figure 22. Monitoring information
The monitoring information screen provides the following information:
•Continuous numeric and waveform display of patient vital signs
•Medical facility name and unit
•Configuration file name
•Contact person’s name and telephone number
Note
If a monitor configuration is saved after a configuration file is downloaded, the file
name shown in the information screen is followed by an asterisk (*). This is true
even if the saved configuration is identical to the downloaded configuration.
Portland Westside
Emergency Department
PtldWstsdED10Jun05.mnt*
Ann Jones, MD
503-530-0101 x9999
For more information about changing the monitor configuration, see “Changing
the default settings” on page 67.
To exit the monitoring information screen, do one of the following:
•To return to the primary display, press, or highlight Exit and press.
•To see a tabular display of vital signs, highlight Trends and press.
•To view saved snapshots of vital signs, highlight Snapshots and press.
•To view the Setup menu, highlight Setup and press.
•To turn off the monitor, press.
Page 39
Directions for UseOverview of monitor operation35
Select ( ) to enter Demo Mode
No data saved.
Start New Patient
Info
Demo
(unplug patient cables)
Using demo mode
In Demo mode, the monitor displays simulated patient data for all vital signs. You can use
Demo mode to familiarize yourself with the following:
•Viewing vital-signs displays
•Modifying alarm limits and other settings
•Cycling through display formats
•Responding to alarm conditions
To start Demo mode, all of the following must be true:
•The power-on screen is displayed.
•The monitor is not in NIBP Auto mode.
•Patient data was deleted when the monitor was last shut down; that is, the
monitor contains no stored patient data.
•The SpO2and ECG cables are not connected to the monitor.
To enter demo mode low
1. Verify that the SpO2and ECG cables are not connected to the monitor.
2. Cycle the monitor off (deleting any saved data) and then on again.
3. When the main screen appears, highlight Demo and press.
Note
If you enter Demo mode with an SpO2or ECG cable connected, the monitor
enters Demo mode for only a second or two before shutting down and powering
up in monitor mode.
If you are in Demo mode and you connect an SpO2or ECG cable or select NIBP
auto mode, the monitor shuts down and powers up to the start-up screen.
Figure 23. Power-on screen: Demo highlighted
No data saved.
Select ( ) to enter Demo Mode
(unplug patient cables)
Start New Patient
Info
Demo
Demo mode is indicated by the message ‘SIMULATION’ in the upper left corner of
the screen.
Page 40
36Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
Figure 24. Demo mode: initial display
When you enter Demo mode, the monitor is in ‘Demo mode low’. In this mode, the
simulated vital signs of the patient are steady and do not cause any alarms at the default
alarm limit settings. You can explore the monitor displays and menus, and you can change
the same settings and values in Demo mode that you can change in normal mode.
In Demo mode low, if you adjust the alarm limits to put the simulated patient’s vital signs
out of limits, the monitor simulates an alarm condition. Another way to simulate an alarm
condition is to switch the monitor to Demo mode high, which uses higher numeric values.
To switch to demo mode high
1. Access the Setup menu. (“To access the setup menus” on page 31.)
2. Pressto highlight Timings.
Figure 25. Setup: timings
Timings
Default demo mode (Low)
Page 41
Directions for UseOverview of monitor operation37
Off
II 1mV/cm
Exit
Trends
Snapshots
Setup
80
HR/min
NIBP mmHg
Resp/min
SpO2
Demo Mode
Backlight Timeout (min)
Setup
12
97
%
ID: 01018VDO9PBH
3:00:06P
Ad ult
Rm 23 9
Alarms
ECG
NIBP
Timings
Service
Display Timeout (min)2
Low
5
Off2 5
SIMULATION
High
Configuration
The Timings menu specifies the Demo mode—Demo mode low, which simulates
normal vital signs, and Demo mode high, which simulates a higher heart rate (HR),
higher respiration rate (Resp), and lower oxygen saturation (SpO2).
3. Pressto highlight Demo mode, and pressorto highlight High.
The monitor generates an alarm within seconds of reading this new set of simulated
vital signs. The red alarm indicator (Figure 28) illuminates and flashes, the numerics
for all violating vital signs—HR, SpO2, and Resp—turn red and flash, and the highlight
moves to the most recent alarming numeric.
Page 42
38Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
With the monitor simulating an alarm, you can practice responding to alarms. (See
“Responding to a high-priority alarm” on page 77.)
To exit demo mode
Press. The Demo Mode Power Off screen appears.
Figure 29. Demo mode power off
•To shut down the monitor, press. (No data can be saved from Demo mode.)
•To resume Demo mode, highlight Cancel and press.
•To access the Setup menu, highlight Setup and press.
Page 43
Directions for UseOverview of monitor operation39
Power saving
To maximize battery life, the monitor display shuts off when the following conditions are
all true:
•No button press for a period of n seconds. (n is configurable.) See “Timing out the
display and the back light” on page 23.
•No active alarms
•No Patient ID entry window
Turning off the monitor
To turn off the monitor
Press.
The Power-off screen appears:
Figure 30. Power-off screen
Highlight the desired action and press.
Note
If you pressto power down before leaving the power-on screen (Figure 7 on
page 16), the monitor shuts down without presenting the screen shown above.
About error detection
The monitor can detect conditions that prevent it from operating properly. If this occurs, it
displays an error message and an error number. Follow the directions displayed on the
screen.
Transporting the monitor with the patient
An ambulatory patient can wear or carry the monitor using the wearable strap or the
patient carry strap (optional accessories).
Page 44
40Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
WARNING When the patient is wearing or carrying the monitor, carefully route
any patient cabling to reduce the possibility of patient entanglement or
strangulation. Use the supplied garment clips to secure the cable properly.
WARNING When positioning straps on the patient, make sure the straps do not
entangle the patient’s neck or cause choking.
WARNING Make sure the straps do not restrict the movement of the patient’s
limbs or create a hazard for the patient when the patient is walking or moving.
WARNING Never use a strap to carry or pick up both the monitor and the cradle.
The straps are not intended to support, and cannot support, the combined weight
of the monitor and the cradle.
To attach the wearable strap
WARNING Do not put the wearable strap on the patient while the patient is inbed. The intended use of the wearable strap is to keep the monitor—without the
cradle—with the patient when the patient is ambulatory.
Figure 31. Wearable strap
1. Place the wearable strap on the sitting or standing patient and adjust all components
for a comfortable, secure fit (Figure 31 left).
2. Connect the wearable strap securely to the monitor strap mounts (Figure 31 right).
3. Carefully arrange the strap and the monitor on the patient to avoid bruising or other
skin injuries.
Page 45
Directions for UseOverview of monitor operation41
To use the patient carry strap
WARNING Do not use the patient carry strap to lift or carry both the monitor and
the cradle. The patient carry strap is not intended to support (and cannot support)
the weight of both the monitor and the cradle. Attempting to carry both the
monitor and the cradle with a patient carry strap could lead to patient injury and to
damage to the monitor and the cradle.
1. Remove the monitor from the cradle.
2. Detach the monitor cables from any accessories.
3. Verify that all cables are disentangled from the bed and any bedside tables.
4. Connect the ends of the carry strap to the strap mounts on the monitor.
Note
Be sure that each end of the carry strap snaps into place.
Figure 32. Patient carry strap
To use the transport stretcher carry strap
1. With the monitor facing away from the stretcher and the patient, attach one end of
the strap to a monitor strap mount.
2. Run the monitor strap under the stretcher restraint straps, near the patient’s waist.
3. Attach the other end of the strap to the other strap mount.
Figure 33. Monitor secured to the patient during stretcher transport
4. Before transporting the patient, verify that all monitor cables are clear.
Page 46
42Overview of monitor operationWelch Allyn Propaq LT Vital Signs Monitor
Page 47
43
Select ( ) to delete data
and start a new patient.
Portland Westside Hospital
Emergency Department
Note!
There is patient data stored for
Hall, Robert E. ID: 3456187
Start New Patient
Continue Patient
Info
Demo
3
Standalone monitoring
Overview
The monitor is available in standalone models (802LT0N and 802LT0S).
This chapter describes the operation of the standalone monitor.
About the Model 802LT0N and 802LT0S (standalone) monitors
The model 802LT0N and 802LT0S monitors operate in standalone mode. They measure
and display vital signs, store patient data, and locally indicate alarm conditions.
Preparing for a new patient
To begin monitoring a new patient
1. If the monitor is on, pressto turn it off.
If prompted to save or delete existing data, highlight Delete & Shut Down and
press. The monitor deletes any saved data and temporary settings, and then shuts
down.
2. Pressto turn on the monitor.
•If the monitor holds stored data from the previously monitored patient, it displays
the “data saved” start-up screen:
Figure 34. Start-up with saved data
Portland Westside Hospital
Emergency Department
Note!
There is patient data stored for
Hall, Robert E. ID: 3456187
Select ( ) to delete data
and start a new patient.
Continue Patient
“no data saved” start-up screen appears:
Start New Patient
•If the data from the previously monitored patient was deleted on shut-down, the
A B C D E F G H I J K L M N O P Q R S T U V
W X Y Z 1 2 3 4 5 6 7 8 9 0 -
Select ( ) to enter highlighted character
Confirm
Cancel
Backspace
Figure 35. Start-up with no saved data
Portland Westside Hospital
Emergency Department
No data saved.
Select ( ) to Start a New Patient.
Start New Patient
Info
Demo
3. Highlight Start New Patient and press. The first configured data display appears.
In place of a patient name, ID: is followed by a string that is generated by the monitor
when you select ‘Start New Patient’. This AutoID string, unique to each new patient,
identifies the patient until you supply appropriate patient identification data.
Note
To bypass patient ID entry, press.
Figure 36. Data display with AutoID
Start-up AutoID
4. In the primary data display, highlight ID: (upper left) and press.
The Patient Information Entry screen appears:
Figure 37. Patient information entry
Last
Middle
First
Select ( ) to enter highlighted character
Confirm
ID#
Backspace
Cancel
Rm#
Entry fields
Character field
5. Enter the ID and name of the new patient.
a.Pressorto highlight (blue) a character in the character field (Figure 38).
b. Pressto copy it to the green-highlighted location in the entry fields (Figure 37).
c.Repeat from step a until all characters are entered into the field.
d. Pressuntil the insertion point in the entry field changes from green to blue.
e. Pressorto move the insertion point to another entry field.
f.Pressto return to the character field.
Page 49
Directions for UseStandalone monitoring45
Last
Middle
First
ID#
Rm#
A B C D E F G H I J K L M N O P Q R S T U V
W X Y Z 1 2 3 4 5 6 7 8 9 0 -
Select ( ) to enter highlighted character
Confirm
Cancel
HALLERORO
Backspace
3456187
Patient Mode
Adult
Pediatric
Neonate
Patient Mode
Adult
Pediatric
Neonate
g. Repeat from step a until all fields are complete.
Note
To correct an error in an entry field: Place the cursor (step d and step e)tothe
right of the error location, highlight and enter Backspace in the character field to
delete the erroneous character, and then enter the correct character.
Figure 38. Patient information entry (continued)
Last
Middle
HALL
Select ( ) to enter highlighted character
Confirm
First
ID#
3456187
Backspace
Cancel
Rm#
Next letter highlighted
6. Highlight Confirm and press.
\
Note
Name alone is not sufficient to confirm a patient ID; thus, you can confirm the
patient name only after you have entered the patient ID.
7. Verify that the patient mode (adult, pediatric, or neonate) is set correctly for this
patient.
NeonatalTerm birth through 28 days, or up to 44 gestational weeks.
PediatricBetween 29 days and 12 years.
Adult13 years and older.
If the current patient mode setting is not correct:
a.Highlight the current patient mode (Adult, Pediatric,orNeonate) and press.
b. From the Patient Mode selection menu, highlight the appropriate patient mode
and press.
Figure 39. Changing patient mode
Patient Mode
Adult
Pediatric
Neonate
Patient Mode
Adult
Pediatric
Neonate
c.A confirmation screen appears:
Figure 40. Confirming a change of patient mode
d. To confirm the change, highlight Confirm and press.
When you change the patient mode and confirm the change:
•All vital-signs data for the patient is lost.
•All monitor settings revert to the defaults for the new patient mode.
8. Verify that the displayed time and date are correct.
•The monitor displays time in the 12-hour (AM/PM) or 24-hour format, and displays
the date in the mm/dd/yy, dd.mm.yy, or yy/mm/dd format.
•The date does not appear on the primary display screens. It appears on the
snapshot list and on the snapshot display.
a.Highlight the time display (in the upper right corner of the screen) and press.
The Time/Date screen appears.
Figure 41. Time/date screen: 24-hour format
HALL, ROBERT E.
II 1mV/cm
HR/min
Time Format
Hour
Minute
Month
Day
Yea r
Date Format
NIBP mmHg
Set
12 Hour
14
45
17
2005
mm/dd/yy
Adult
3456187
Resp/min
24 Hour
dd.mm.yy
Rm 239
Cancel
14:49:37
SpO2
24-hour format
yy/mm/dd
b. If the displayed time or date is not correct, pressorto move the highlight
from one parameter to another, and pressorto change the value of the
highlighted parameter. For example, to change the time display format from
24-hour to 12-hour, highlight Time Format and press eitheroronce.
Figure 42. Time/date screen: 12-hour format
HALL, ROBERT E.
II 1mV/cm
HR/min
Time Format
Hour
Minute
Month
Day
Yea r
Date Format
NIBP mmHg
Set
12 Hour
2005
mm/dd/yy
3456187
Adult
2P
45
17
Resp/min
24 Hour
dd.mm.yy
Rm 239
Cancel
yy/mm/dd
c.When the time and date are correct and formatted appropriately, pressto
accept the changes and return to the vital-signs display.
Note
If you change the time or date settings and then decide not to accept the
changes, pressto cancel the changes and return to the vital-signs display.
14:49:54
SpO2
12-hour format
Page 51
Directions for UseStandalone monitoring47
Select ( ) to delete data
and start a new patient.
Portland Westside Hospital
Emergency Department
Note!
There is patient data stored for
Hall, Robert E. ID: 3456187
Start New Patient
Continue Patient
Info
Demo
Select ( ) to continue
monitoring this patient.
Portland Westside Hospital
Emergency Department
Note!
There is patient data stored for
Hall, Robert E. ID: 3456187
Start New Patient
Continue Patient
Info
Demo
Continuing to monitor a patient on power-up
If patient data was saved when the monitor was last turned off (see “To turn off the
monitor” on page 39), you can resume monitoring that patient when the monitor is turned
on again. (When patient data is saved, the monitor settings are also saved.)
To resume monitoring the same patient:
1. Turn on the monitor. The “patient data stored” screen appears:
Figure 43. Start-up with saved data
Portland Westside Hospital
Emergency Department
Note!
There is patient data stored for
Hall, Robert E. ID: 3456187
Select ( ) to delete data
Start New Patient
and start a new patient.
Continue Patient
Info
Demo
2. Verify that the displayed name and ID match that of the current patient.
3. Highlight Continue Patient and press.
Figure 44. Start-up with saved data
Portland Westside Hospital
Emergency Department
Note!
There is patient data stored for
Hall, Robert E. ID: 3456187
Select ( ) to continue
Start New Patient
monitoring this patient.
Continue Patient
Monitoring ECG and Resp
Overview
Note
In this manual, lead refers to an ECG electrode or lead wire, and Lead refers to a
waveform source.
Info
Demo
You can monitor heart signs (ECG) and respiration rate (Resp) using either a 3-lead or a
5-lead ECG cable. Using a 3-lead cable, you can display one signal waveform for Lead I, II,
or III. Using a 5-lead cable, you can display either one or two signal waveforms (I, II, III, V;
and if enabled in the configuration, aVR,aVL,oraVF). You can also display the SpO2or
Resp waveform in place of an ECG waveform.
WARNING Always monitor and set alarms for SpO2when using impedance
pneumography to monitor respiratory function.
WARNING When monitoring respiration via impedance pneumography, always
select the ECG Lead with the most prominent QRS complex. The monitor rejects
cardiovascular artifact, but this function depends upon accurate ECG R-wave
detection.
WARNING Do not place the monitor near another respiration monitor. Resp
measurement frequencies can cause mutual interference.
WARNING Do not perform impedance pneumography on paced patients.
Pacemaker pulses can sometimes be falsely counted as breaths.
WARNING Always keep patient motion to a minimum. Motion artifact can cause
incorrect readings of breath rate or heart rate.
WARNING If a disconnected lead is too close to other electrical devices, it can
cause a false heart rate, a false respiration rate, or a failure to display a “Lead Fail”
message.
WARNING The monitor displays + + + for HR numerics between 301-350 beats
per minute. For heart rates above 350 beats per minute, it might display
incorrectly low heart readings, due to intermittent picking of R-waves.
WARNING The monitor does not provide internal arrhythmia analysis; therefore,
arrhythmias can cause the monitor to display inaccurate heart rates.
WARNING Make sure the patient mode is correct. Incorrect patient mode can
result in inaccurate heart rate readings and inappropriate alarm settings.
WARNING Electric shock hazard. Do not allow the conductive parts of the
patient cable, electrodes, or associated connections of defibrillation-proof applied
parts, including the neutral conductor of the patient cable and electrodes, to
come into contact with other conductive parts, including earth ground. An
electrical short might result, risking electric shock to patients and damage to the
device.
Page 53
Directions for UseStandalone monitoring49
WARNING (1) During a surgical procedure, do not use small ECG electrodes.
(2) Select ECG electrode attachment points remote from the surgical site and
remote from the electrosurgical return electrode. (3) Use electrosurgical return
electrodes with the largest practical contact area. (4) Assure proper application of
the electrosurgical return electrode to the patient.
High-intensity radio-frequency (RF) energy from external sources, such as an
improperly connected electrosurgical unit, can induce heat into electrodes
and cables, which can cause burns on the patient and can lead to
measurement errors.
WARNING Always use the provided garment clips to route ECG cables away
from the patient’s head.
WARNING Use only accessories approved by Welch Allyn. The use of any other
accessories can result in inaccurate patient data, can damage the equipment, and
can void your product warranty. Refer to the accessory list or
www.welchallyn.com.
WARNING Always use accessories according to the standards of your facility
and according to the manufacturer's directions.
Note
WARNING Never use ECG cables with loose or faulty detachable lead wires.
These can cause erratic behavior of the ECG and respiration waveforms due to
intermittent ECG lead wire connections.
WARNING Resp is derived from the same leads as the ECG channel, so the
monitor determines which signals are cardiovascular artifact and which signals are
a result of respiratory effort. If the breath rate is within five per cent of the heart
rate or is a multiple or submultiple of the heart rate, the monitor might ignore
breaths and trigger a respiration alarm.
Caution Never use an ECG cable longer than 10 feet (3 meters) including
extensions. If you use an ECG extension cable with an ECG cable longer than
4 feet, the monitor acts as though no ECG cable is connected.
Caution To protect the monitor from damage during defibrillation or
electrosurgery, for accurate ECG information, and for protection against noise
and other interference, use only ECG electrodes and cables specified or supplied
by Welch Allyn (these cables have the required current-limiting resistors). Follow
recommended application procedures.
The monitor contains type CF fully isolated patient-connected circuitry, but it is
not intended for direct application on a patient’s heart.
Severe artifact and interference (such as defibrillation interference) can cause the
waveform to move off of the display for a few seconds before it is restored.
Impedance pneumography (Resp) is not recommended for use with
high-frequency ventilation.
The monitor counts as breaths any respiratory efforts larger than twice the
background cardiovascular artifact.
Note
Use only silver/silver chloride electrodes. Other electrodes, such as stainless
steel electrodes, squeeze-bulb electrodes, or electrodes with dissimilar metals,
are subject to large offset potentials due to polarization. Other electrodes can also
have slower recovery time after the application of defibrillator pulses.
1. Inspect the ECG cable. Replace it if it shows any signs of wear, breakage, or fraying.
2. Plug the cable into the monitor.
3. Select electrode sites on the patient, choosing flat areas and avoiding fatty or bony
areas and major muscles.
Figure 45. ECG leads - actual placement
The monitor display is not capable of resolving a 0.1mV RTI signal at a gain setting
of 8mV/cm. Adjust the gain setting of the monitor if necessary.
V1
RA
LL
RL
5-lead, adult and pediatric
LA
V6
Six possible V lead electrode
placement sites for the C lead.
RA
LL
3-lead, adult and pediatric
RA
LA
3-lead, neonatal
LA
LL
4. Shave or clip the hair from the electrode sites.
5. Thoroughly clean the skin, using soap and water, isopropyl alcohol, or skin preparation
pads, and lightly rub it dry.
WARNING For safe and effective use of electrodes, follow the electrode
manufacturer’s directions.
6. If you are using gelled electrodes, verify that the electrode expiration date has not
passed and that the gel is intact and not dried out.
If you are not using gelled electrodes, apply a mound of gel (1/4-inch to 1/2-inch, or
0.6-cm to 1.3-cm) to each electrode contact area.
7. Attach lead wires to the electrodes before applying them to the patient.
Page 55
Directions for UseStandalone monitoring51
140/78
80
12
97
%
HAL L, ROBERT E.
345 6187
3:0 0:06P
Ad ult
Rm 23 9
HR/min
NIBP mmHg (102)
@2:47P Manual
Resp/min
SpO2
SpO22x2x
II 1mV/cm
8. Apply the electrodes to the patient in the proper locations (Figure 45).
Note
At least three appropriate electrode connections are required for ECG/Resp
monitoring.
Figure 46. ECG leads - placement displays, 5-lead and 3-lead
The locations of the circles displayed on the monitor (Figure 46) for each lead are
fixed, and do not indicate the exact placement of the electrodes on the patient.
If the monitor detects that some lead wires are not connected, it displays an “ECG
Fault” technical alarm and a chest diagram indicating the location of the disconnected
lead or leads. If the disconnected lead(s) invalidate the Lead used for HR
determination, then the monitor reassigns, if possible, the Lead used for HR. If the
reassignment succeeds, the monitor then displays another technical alarm with the
message “ECG Lead changed”.
Figure 47. ECG leads - disconnected leads
Disconnected leads
X
X
X
9. When all leads are properly connected, confirm that the monitor displays the ECG
waveform, heart rate, and other patient data.
Lead I
Lead II 6 sec
Lead III
aVR
aVL
aVF
V
Resp
SpO2
Add 2nd Wave
%
STE WART, ANN
776 2940
7:0 5P
Ad ult
Rm 26 3
HR/min
Resp/min
SpO2
Resp 4x
60
20
97
130/74
NIBP mmHg (88)
@7:05P Manual
Waveform Source
Lead I
Lead II
Lead III
aVR
aVL
aVF
V
Resp 24 sec
SpO2
Add 2nd Wave
%
HR/min
NIBP mmHg
Resp/min
SpO2
II 1mV/cm
I 4mV/cm
STE WART, ANN
776 2940
12: 41:32
Ad ult
Rm 26 3
80
12
97
Waveform Size
0.2 mV/cm
0.5 mV/cm
1 mV/cm
2 mV/cm
4 mV/cm
8 mV/cm
To change the waveform selection
1. Highlight the current waveform source selection (Lead II, for example) and press.
Figure 49. Waveform source: II
STEWART, ANN
II 1mV/cm
Waveform Source
Lead I
Lead II 6 sec
Lead III
aVR
aVL
aVF
V
Resp
SpO2
80
Add 2nd Wave
HR/min
NIBP mmHg
7762940
Adult
12
Resp/min
Rm 263
15:14:28
97
%
SpO2
2. Highlight your waveform source choice and pressor.
Figure 50. Waveform source: Resp
STEWART, ANN
Resp 4x
Waveform Source
Lead I
Lead II
Lead III
aVR
aVL
aVF
V
Resp 24 sec
SpO2
60
Add 2nd Wave
HR/min
NIBP mmHg (88)
@7:05P Manual
130/74
7762940
Adult
20
Resp/min
Rm 263
7:05P
97
%
SpO2
Note
If you scroll to the bottom selection of the Waveform Source menu—either Add
2nd Wave or Remove 2nd wave—the selection takes effect immediately and the
monitor returns to the primary data display.
Note
In the Waveform Source menu, either the source Lead or the waveform period
can be highlighted. If the waveform period is highlighted, a second trace of the
same source will be cascaded to double the period obtained from a single trace.
To change the waveform size
1. Highlight the current waveform scale (1mV/cm, for example) and press.
Figure 51. Waveform size popup menu
STEWART, ANN
II 1mV/cm
Waveform Size
0.2 mV/cm
0.5 mV/cm
1 mV/cm
I 4mV/cm
2 mV/cm
4 mV/cm
8 mV/cm
80
HR/min
NIBP mmHg
Adult
7762940
Rm 263
12
97
Resp/min
12:41:32
%
SpO2
Page 57
Directions for UseStandalone monitoring53
II 1mV/cm
Exit
Trends
Snapshots
Setup
60Hz
50Hz
80
HR/min
140/78
NIBP mmHg (102)
Resp/min
SpO2
OffOnOn
Power source filter
Resp Monitoring
Setup
12
97
%
HAL L, ROBERT E.
345 6187
3:0 0:06P
Ad ult
Rm 23 9
Alarms
ECG
NIBP
Timings
Off
Configuration
ECG Bandwidth
Pacer indicator
Monitor
OffOnOn
Extended
Service
2. Highlight the desired scaling factor and press. (Waveform size does not affect
QRS-detector sensitivity.)
About pacemakers and ECG monitoring
If the patient being monitored has a pacemaker, the monitor detects and can indicate the
occurrence of pacemaker signals. If the Pacer Indicator setting is ON, the monitor displays
and prints vertical dashed lines to indicate detected pacemaker signals. If Pacer Indicator
is OFF, the monitor continues to detect the pacemaker signals but does not display or
print the pacer markers.
WARNING Signals differ between pacemakers. The Association for
Advancement of Medical Instrumentation (AAMI) cautions that “in some devices,
rate meters may continue to count the pacemaker rate during occurrences of
cardiac arrest or some arrhythmias. Do not rely entirely upon rate meter alarms.
All pacemaker patients should be kept under close or constant observation.”
WARNING Use of respiration monitoring by impedance pneumography can
affect the operation of some pacemakers. If pacemaker operation is affected, turn
off respiration pneumography. (See Figure 52 on page 53.)
WARNING Always use fresh ECG electrodes and make sure the ECG cable lead
wires make good connections. The presence of pacer-like noise can cause the
displayed heart rate to be erratic even though the ECG trace might look
undistorted with the pacer indicator off.
To enable or disable the display of pacer indicators
1. Access the Setup menu. (See “To access the setup menus” on page 31.)
2. Highlight ECG.
Figure 52. Turning the pacer indicator off in the ECG setup menu
HALL, ROBERT E.
II 1mV/cm
80
HR/min
Setup
Resp Monitoring
ECG Bandwidth
Power source filter
Pacer indicator
Exit
Alarms
Service
140/78
NIBP mmHg (102)
ECG
Trends
Adult
NIBP
3456187
Rm 239
12
Resp/min
Timings
Off
Monitor
60Hz
50Hz
Off
Snapshots
3:00:06P
97
SpO2
Configuration
Extended
Off
Setup
%
Pacer indicator Off
3. Highlight Pacer indicator and pressorto highlight Off or On.
4. To exit the Setup menu, pressor, or highlight Exit and press.
If the pacemaker signal is strong enough, the monitor displays it as a waveform
spike. This is true with Pacer indicator ON or OFF.
In accordance with the Pacer Pulse Rejection specification (“Pacer pulse
rejection” on page 138.), pacemaker pulses are not counted as heartbeats
whether Pacer Indicator is On or Off.
Noise on the ECG signal might be detected as pacer signals, causing the pacer
indicator to appear on the display. If you do not need to indicate pacemaker
signals, turn off the pacemaker indicator for a better ECG waveform display.
Improving the waveform display
If the power source filter is off, noise from the power source can cause an unclear or
noisy waveform.
To change the power source filter to reduce noise
1. Access the Setup menu. (See “To access the setup menus” on page 31.)
2. Highlight ECG.
Verify that the setting for Power source filter is correct for the power source in your
facility. If you do not know what this setting should be, consult a qualified service
person.
3. Highlight Power source filter and pressoras needed to select 60 Hz, 50 Hz,
or Off.
Figure 53. Turning on the 60-Hz power source filter in the ECG setup menu
HALL, ROBERT E.
II 1mV/cm
80
HR/min
Setup
Alarms
Service
Resp Monitoring
ECG Bandwidth
Power source filter
Pacer indicator
Exit
4. Exit the Setup menu by pressingor.
Monitoring respiration
Resp is based on impedance pneumography, where respirations are sensed from the
ECG electrodes.
All ECG cables listed for the Propaq LT monitor in the accessory list or at
www.welchallyn.com permit respiration monitoring and electrosurgical
interference suppression.
To measure Resp with Lead II selected, the LL lead must be attached to the
patient.
Note
Note
Adult
140/78
NIBP mmHg (102)
ECG
NIBP
Trends
3456187
Rm 239
12
Resp/min
Timings
Off
Monitor
60Hz
50Hz
Off
Snapshots
3:00:06P
97
SpO2
Configuration
Extended
Off
Setup
%
Power source filter 60 Hz
Page 59
Directions for UseStandalone monitoring55
%
HAL L, R OBERT E.
345 6187
3:0 8:16
Ad ult
Rm 23 9
HR/min
Resp/min
SpO2
Resp 2x2x
80
12
97
NIBP mmHg
%
HAL L, R OBERT E.
345 6187
3:0 8:16
Ad ult
Rm 23 9
HR/min
Resp/min
SpO2
Resp 8x8x
80
12
97
NIBP mmHg
Resp is part of ECG monitoring. The Resp numeric is displayed (in purple) in the lower
right corner. To view the Resp waveform:
Change the waveform source to Resp. (See “To change the waveform selection” on
page 52.)
WARNING Use of respiration monitoring by impedance pneumography can
affect the operation of some pacemakers. If pacemaker operation is affected, turn
off respiration pneumography. (See Figure 52 on page 53.)
Figure 54. Resp waveform, default size (2x)
HALL, ROBERT E.
Resp
80
HR/min
For more detail, change the waveform size to 8x. (See “To change the waveform size”
on page 52.)
Note
Figure 55. Resp waveform enlarged for detail (8x)
Waveform size does not affect breath-detector sensitivity.
The monitor can be purchased with the Nellcor or the Masimo SpO2option. Most of this
chapter pertains to either configuration. Monitors with the Masimo technology, however,
can provide additional SpO2data. (“About the Masimo SpO2display” on page 60.)
WARNING Accurate measurements of oxygen saturation, when derived from
pulse oximetry, depend to a great extent on patient condition and proper sensor
placement. Patient conditions such as shivering and smoke inhalation can result in
erroneous readings. If you believe a measurement might be inaccurate, verify it
using another clinically accepted measurement method, such as arterial blood gas
measurements using a co-oximeter.
WARNING Use only SpO
www.welchallyn.com. Inspect sensors and cables, and discard any that are
damaged. Do not use a sensor with exposed optical components.
WARNING If you need to increase the length of the sensor cable, use only one
extension. Use of multiple extensions can adversely affect performance. Do not
attach any cable that is intended for computer use to the SpO2connector at the
monitor.
WARNING Tissue damage and erroneous measurements can be caused by
incorrect application or use of a sensor. (Examples of bad practices: wrapping the
sensor too tightly, applying supplemental tape, failing to periodically inspect the
sensor site, leaving a sensor on too long in one place.) Refer to the
manufacturer’s directions for specific instructions on application and use, and for
description, warnings, cautions, and specifications.
2
accessories listed in the accessory list or at
2
WARNING Do not modify the sensor.
WARNING Do not wet the sensor or immerse it in fluid. Do not attempt to
sterilize a sensor.
WARNING Sensors exposed to ambient light while not applied to a patient can
exhibit seminormal saturation readings. Be sure the sensor is securely placed on
the patient and check its application often to ensure accurate readings.
WARNING Inaccurate measurements might be caused by venous pulsations.
WARNING The pulse oximeter can be used during defibrillation, but the
readings might be inaccurate for a short time.
WARNING Do not use the pulse oximeter as an apnea monitor.
WARNING During SpO2monitoring, a very sudden and substantial change in
pulse rate can result in erroneous pulse rate readings. Always validate the patient
data and patient condition before effecting an intervention or a change in patient
care.
Page 61
Directions for UseStandalone monitoring57
WARNING Interfering substances: Carboxyhemoglobin can erroneously
increase readings; the level of increase is approximately equal to the amount of
carboxyhemoglobin present. Methemoglobin and other dysfunctional
hemoglobins can also cause erroneous readings. Further assessment beyond
pulse oximetry is recommended. Intravascular dyes, or any substances
containing dyes, that change usual arterial pigmentation can cause erroneous
readings. Darkly pigmented skin can adversely affect SpO2readings.
WARNING For a premature infant, high oxygen levels might predispose the
infant to develop retinopathy. Therefore, the upper alarm limit for oxygen
saturation must be carefully selected in accord with accepted clinical standards
and considering the accuracy range of the monitor.
WARNING Functional testers cannot be used to assess the accuracy of a pulse
oximeter monitor.
Caution If liquid gets into the SpO2connector cavity, discontinue SpO
2
monitoring until the liquid is removed and the cavity is dry.
Note
This monitor does not normalize the SpO2waveform.
1. Inspect the SpO2cable. Replace it if it shows any signs of wear, breakage, or fraying.
2. Plug the cable into the sensor and the monitor.
Each SpO2sensor is intended for application to a specific site and site size on the
patient. To obtain optimal performance, use the right sensor and apply it as instructed
by the sensor manufacturer.
3. Clean the application site. Remove anything, such as nail polish, that could interfere
with the operation of the sensor.
4. Attach the SpO2sensor to the patient according to the manufacturer’s directions for
use, observing all warnings and cautions.
5. Confirm that the monitor displays SpO2data within thirty seconds of being connected
to the patient.
•If ambient light is too bright, shield the sensor site with opaque material. Failure
to do so can result in inaccurate measurements. Light sources that can affect
performance include the following:
-surgical lights (especially those with a xenon light source)
-bilirubin lamps
-fluorescent lights
-infrared heating lamps
-direct sunlight.
•To help reduce unnecessary SpO2alarms when NIBP and SpO2are monitored
simultaneously, place the NIBP cuff and the SpO2sensor on different limbs.
•Do not attach the SpO2sensor on the same limb as an arterial catheter or
intravascular line.
•The pulse signal can disappear if any of the following conditions exists:
-an NIBP cuff is inflated on the same limb as the sensor
-arterial occlusion occurs near the sensor
-the patient is in cardiac arrest or shock
-the patient has hypotension, severe vasoconstriction, severe anemia, or
hypothermia
•If poor perfusion affects performance for an adult, consider using the adult nasal
sensor.
•If a sensor is connected and the sensor light does not come on within 3 seconds:
-Verify that SpO
is turned on (Figure 56 on page 58)
2
-Replace the sensor
•If excessive or prolonged patient movement interferes with measurements,
consider the following possible solutions:
-be sure the sensor is secure and properly applied
-use a new sensor with fresh adhesive backing
-select a different type of sensor
-move the sensor to a less active site
6. Periodically verify that the sensor remains properly positioned on the patient.
About SpO2spot check
Note
When SpO2is turned on, the monitor generates an alarm condition whenever SpO
readings are interrupted, such as when the sensor is disconnected from the patient after
the monitor begins taking SpO2readings.
Using the SpO2Spot Check feature, however, you can take any number of spot SpO
readings at random intervals, attaching and detaching the sensor repeatedly without
generating alarms.
To prepare to take a spot check reading
1. Highlight SpO2 and press.
2. Highlight SpO2 Monitoring, pressto highlight Off, and pressto return to the
Figure 56. Turning off SpO2monitoring
The Spot Check feature is available only if it is enabled in the monitor
configuration. Refer to “Using the Configuration Utility to configure the monitor”
on page 104.
main screen.
2
2
Pressto move the
highlight to Off.
‘SpO2’ has changed to ‘SpO2@ (time)’. Spot checks are now enabled.
Page 63
Directions for UseStandalone monitoring59
SpO2
On
Off
Spot Check
Figure 57. SpO2monitoring turned off
SpO2Monitoring is Off
To take an SpO
1. Enable spot checks. (See “To prepare to take a spot check reading” on page 58.)
2. Attach the sensor to the monitor and the patient.
3. Highlight SpO2 @ XX:XX and press. The SpO
Note
Figure 58. SpO2drop-down menu
The SpO2drop-down menu can be accessed only when SpO2is set to Off.
spot check reading
2
drop-down menu appears.
2
SpO2
On
Off
Spot Check
4. Pressto highlight Spot Check.
•The drop-down menu disappears. SEARCH appears above SpO2Spot Check (with
pulse rate if SpO2is the source of pulse rate).
Figure 59. SpO2spot check: waiting for an SpO2signal
6. To take another spot check later, repeat from step 2 (above).
To return to continuous SpO2measurements
1. Highlight SpO2 @ XX:XX and press.
2. Pressto turn automatic SpO2On. The pop-up menu disappears.
To adjust the SpO2and ECG pulse tone volume
1. Highlight SpO2 and press.
2. Highlight HR/PR Tone.
The pulse amplitude bar can help determine a correct SpO2sensor placement.
This bar indicates physiological pulse activity, but is not directly related to pulse
volume.
•The spot check ends, and SpO
•The SpO2text on the display screen now includes the time of the most recent
SpO2measurement. For example: SpO2 @ 3:05P.
Spot-check readings are included in trend displays.
monitoring is again turned off.
2
3. Highlight the desired volume level (Off, Low, Med,orHigh) and press.
About the Masimo SpO2display
The Masimo SpO2option provides the following additional features and indicators.
Note
The monitor (model 802LT0S) must be configured specifically to display the
Masimo parameters.
Page 65
Directions for UseStandalone monitoring61
Signal IQ (SIQ)(Signal Identification and Quality indicator) A horizontal line
below the SpO2waveform, with vertical spikes indicating
the relative reliability of the oxygen saturation readings.
(Higher spikes indicate greater reliability.)
Perfusion Index (PI) An assessment, expressed as a percentage between 0.02
and 20, of the relative pulse strength at the monitoring
site.
FastSATThe tracking of rapid changes in arterial oxygen saturation.
SensitivityNormal, APOD, and Maximum.
NormalFor typical monitoring.
APODFor monitoring when the probe is likely to detach from the
patient because of wet skin, motion, or other unusual
circumstances.
MaximumFor monitoring patients with extremely low perfusion.
Averaging TimeThe period (in seconds) over which multiple SpO2
WARNING Always use a properly fitting cuff, placing it carefully on the patient
according to the instructions presented below. Failure to fit and locate the cuff
correctly can lead to inaccurate pressure readings.
WARNING During monitoring, periodically observe the patient’s limb to make
sure that the circulation is not impaired for a prolonged period. Prolonged
impairment of circulation or improper cuff placement can cause bruising.
WARNING Do not use the monitor to simultaneously measure NIBP on one
patient and monitor ECG on another patient.
WARNING If an NIBP measurement is suspect, repeat the measurement. If you
are still uncertain about the reading, verify it using another method.
WARNING Do not take NIBP measurements during cardiopulmonary bypass.
WARNING When monitoring NIBP, match the monitor patient mode to the NIBP
cuff. For neonates, set the monitor to Neonatal Mode unless the circumference
of the limb is too large for the cuff. In that case, use the Pediatric Mode.Be
aware, however, that the maximum cuff inflation limits are based on the patient
mode, not the cuff; the maximum cuff inflation limits for Pediatric Mode are
greater than for Neonate Mode. (See “NIBP” on page 142 for values.)
WARNING The decision to use the device on pregnant or pre-eclamptic patients
is at the discretion of the trained clinician using the equipment.
WARNING Do not apply the NIBP cuff over a wound.
WARNING Patient injury risk. Inaccurate measurement risk. Do not place the
cuff where it can disturb proper circulation. Do not place the cuff on any area
where circulation is compromised or on any extremity used for intravenous
infusions.
WARNING Patient injury risk. Do not place the cuff on the arm on the same side
of a mastectomy. If necessary, use the femoral artery in the thigh to take a
measurement.
Caution Pulse-rate measurements generated through the blood pressure cuff
or through SpO2are subject to artifact and might not be as accurate as heart-rate
measurements generated through ECG or through manual palpation.
Caution Do not allow the blood pressure cuff hose to be restricted by
compression.
When the monitor is powered on, the default cuff inflation pressure is based on the
patient mode. (See “Default inflation pressure” on page 142.) After an NIBP
measurement occurs, the monitor adjusts the inflation pressure to optimize subsequent
NIBP measurements.
Page 67
Directions for UseStandalone monitoring63
Note
Always cycle the monitor power before you begin to monitor another patient.
Normal physiological pressure variations affect NIBP measurements from reading
to reading.
If the monitor is in Adult mode and a neonate cuff is connected to the monitor,
the monitor generates a technical alarm.
If the battery charge is low and the monitor is not in the cradle, the battery icon
indicates low battery and NIBP monitoring is disabled.
To monitor blood pressure
1. Select cuff size based on limb circumference. Use only hoses and cuffs listed in the
accessory list or at www.welchallyn.com.
2. Squeeze all the air from the cuff before placing the cuff on the patient.
3. Place the cuff on the limb, as near heart level as possible.
Figure 61. Cuff placement
Cuff applied evenly and snugly. The
center of the cuff is at heart level,
and the bottom edge is one inch (2.5
cm) above the antecubital fossa.
Possible cuff placements for neonates
•The cuff must fit snugly without being uncomfortably tight.
•The hose must be free of kinks and not pinched.
4. Align the point where the tubing connects to the cuff over the brachial or other
appropriate artery.
Note
If you are simultaneously monitoring blood pressure and SpO2, you can reduce or
eliminate unnecessary SpO2alarms by placing the cuff and the SpO2sensor on
different limbs.
5. Screw the hose connector onto the NIBP air connector on the top of the monitor (see
Figure 4 on page 11).
6. Pressto start a reading.
Within a few seconds, the pump starts, the cuff begins to inflate, and the manometer
screen appears.
7. When the cuff is fully inflated, the manometer bar dynamically displays the pressure
reading.
Figure 63. NIBP: reading in progress
STEWART, ANN
NIBP in progress.
II 1mV/cm
HR/min
NIBP
NIBP Mode:
NIBP mmHg
Manometer
Manual
Adult
Systolic
7762940
Rm 263
Resp/min
Diastolic
3:00:06P
%
SpO2
Mean
Manometer pressure indicator bar
145
50
100
150
200
250
300
Exit
Trends
Snapshots
Setup
When the cycle completes, control returns to the previous direct-access display. To
view the final NIBP readings on the manometer scale, select Manometer from the
Setup NIBP menu.
Figure 64. NIBP: reading complete
STEWART, ANN
II 1mV/cm
79
PR/min
Manometer
NIBP
NIBP Mode:
2 2 50
Exit
132/83
NIBP mmHg
Manual
100
Trends
Adult
(98)
Systolic
150
7762940
Rm 263
Resp/min
Diastolic
200
Snapshots
3:00:06P
SpO2
Mean
250
Setup
Systolic, diastolic, and MAP
(numeric)
%
Systolic, diastolic, and MAP
(graphical)
300
•If you need to stop the reading and vent the cuff at any time during the
measurement, press.
•If the monitor cannot get a valid NIBP reading, it displays the message
‘NIBP retry in progress’ in the upper left corner of the screen. Then, depending on
settings and patient mode, the monitor attempts one or two more
Page 69
Directions for UseStandalone monitoring65
measurements. If the retries fail to produce a valid reading, the monitor displays
an NIBP error message.
Note
For every inch the cuff is placed above the heart, add 1.9 mmHg (0.253 kPa) to the
displayed NIBP reading. For every inch below the heart, subtract 1.9 mmHg
(0.253 kPa).
8. If motion artifact such as shivering or coughing interferes with NIBP readings, do the
following:
•Position the patient’s limb away from the body so that the applied cuff is not in
contact with the patient’s body or any other object (such as a bed rail). Keep the
cuff as close to heart level as possible.
•Verify that the Smartcuf filter is ON. (See “Improving NIBP accuracy with
Smartcuf” on page 65.)
•Verify that the ECG leads are properly connected to the patient, and monitor ECG
during NIBP. (ECG monitoring is required for Smartcuf.)
Note
The message ???/???/(???) in an NIBP TREND display or printout indicates that
the monitor could not complete an NIBP measurement during that period.
Note
On-demand NIBP readings (manual or turbo) are delayed by 8 seconds if they are
started while the monitor is in a power-saving mode such as display time-out.
Improving NIBP accuracy with Smartcuf
Many factors can adversely affect an NIBP measurement: cardiac arrhythmias, sudden
changes in blood pressure, patient motion such as convulsions or shivering, sudden cuff
movement, vibration, vehicle motion, or a weak pulse. The Smartcuf feature increases
NIBP measurement accuracy in the presence of moderate motion artifact or diminished
pulses.
s
Note
Smartcuf can function only when ECG is being monitored.
To enable Smartcuf
1. Simultaneously monitor ECG and NIBP.
2. Access the Setup menu. (See “To access the setup menus” on page 31.)
3. Highlight NIBP, highlight Smartcuf, and enable the Smartcuf filter.
If Smartcuf is enabled and motion artifact is so severe that it still affects measurement
accuracy, the measurement is marked with the symbolon the display. During certain
types of arrhythmias and other situations where a good ECG signal cannot be obtained,
consider disabling Smartcuf, as follows:
To disable Smartcuf
1. Access the Setup menu.
2. Highlight NIBP, highlight Smartcuf, and disable the Smartcuf filter.
In the Auto NIBP mode, for intervals shorter than 5 minutes, the monitor immediately
begins taking NIBP readings at the specified interval. For intervals of 5 minutes or longer,
the readings begin when the time of day is a multiple of the interval. (If the interval is 15
minutes, for example, then the readings begin at 00, 15, 30, or 45 minutes after the hour.)
To start automatic NIBP readings
1. Select and apply the appropriate NIBP cuff and hose. (“Monitoring blood pressure
(NIBP)” on page 62.)
2. Highlight NIBP and press.
3. Highlight NIBP Mode and select Auto.
4. Highlight Auto Interval (min) and select an interval.
Note
To stop automatic NIBP readings
1. Highlight NIBP and press.
2. Highlight NIBP Mode and select Manual.
After you invoke an automatic NIBP, expect a delay before the monitor starts the
first measurement. The delay can be as long as the interval selected.
Taking NIBP readings using Turbo mode
WARNING Patient injury risk. Bruising of the patient’s limb can result from the
NIBP cuff when the device is used in Turbo mode.
In Turbo mode, the monitor starts an NIBP reading and then takes as many more readings
as possible within five minutes.
To use the Turbo mode
1. Select and apply the NIBP cuff and hose. (“Monitoring blood pressure (NIBP)” on
page 62.)
2. Highlight NIBP and press.
3. Highlight NIBP Mode and select Turbo.
Note
To end the Turbo mode
If you cycle the monitor power, NIBP returns to auto mode.
Pressor select NIBP Mode Manual. The monitor returns to Auto NIBP
measurement mode.
Page 71
Directions for UseStandalone monitoring67
II1mV/cm
Exit
Trends
Snapshots
Setup
80
HR/min
NIBP mmHg
Resp/min
SpO2
Select ( ) to save current settings.
12
97
%
ID:01018VDO9PBH
3:00:06P
AdultRm239
Select ( ) to use factory settings.
Select ( ) to use file: n/a
Setup
Alarms
ECG
NIBP
Timings
Service
Configuration
NIBP measurements in power-saving mode
When a manual or turbo NIBP activity awakens the monitor from power-saving mode
(“Power saving” on page 39), cuff inflation pressure is reset to default levels and cuff
inflation is delayed for up to 8 seconds.
NIBP disabled when the battery is low
If the battery is low and the monitor is operating on battery power, NIBP functions are
disabled and the monitor displays the message “NIBP off. Low battery.”
If you attempt to start an NIBP measurement during a low-battery condition, the monitor
displays an technical alarm with the message “Low battery. NIBP disabled.”
Note
Inserting a monitor into a powered cradle during a low-battery condition
immediately enables NIBP monitoring.
Changing the default settings
Note
You can change the default settings (the monitor configuration) using the following
methods:
To save the current settings as the new default settings
1. Access the Configuration Setup menu.
Changing the default settings is password protected. Contact your service
administrator for details on accessing this area.
•Save the current settings as the new default settings.
•Restore the last downloaded configuration.
•Restore the factory settings.
•Download another configuration from a PC. (See “Using the Configuration Utility
to configure the monitor” on page 104.)
80
HR/min
Setup
Select ( ) to save current settings.
Select ( ) to use factory settings.
Select ( ) to use file: n/a
Exit
Alarms
Service
NIBP mmHg
ECG
NIBP
Trends
12
Resp/min
Timings
Snapshots
97
SpO2
Configuration
Setup
%
2. Highlight Select () to save current settings and press.
3. In the confirmation screen, highlight Confirm and press.
1. Access the Configuration Setup menu, highlight Select ( ) to use file:
configuration_file_name, and press.
2. In the confirmation screen, highlight Confirm and press.
To restore the factory configuration
1. Access the Configuration Setup menu, highlight Select () to use factory settings,
and press.
2. In the confirmation screen, highlight Confirm and press.
Page 73
Directions for UseStandalone monitoring69
Select ( ) to delete patient data
You have pressed the Power Off button.
Delete & Shut Down
Cancel
pSetup
and shut down.
15: 01:24
Ad ult
Rm 23 9
There is patient data stored for
Hall, Robert E. ID: 3456187
Save & Shut Down
HAL L, ROBERT E.
345 6187
To discontinue monitoring
1. Press. The Power Off screen appears.
Figure 65. Power-off
Cancel
15:01:24
Setu
HALL, ROBERT E.
You have pressed the Power Off button.
There is patient data stored for
Hall, Robert E. ID: 3456187
Select ( ) to delete patient data
and shut down.
Delete & Shut Down
3456187
Adult
Save & Shut Down
Rm 239
•If you intend to continue monitoring the same patient when the monitor is turned
on again, and if you want to save the stored vital-signs data and monitor settings
(to print them at a PC), highlight Save & Shut Down and then press. The
monitor saves the patient data and the monitor settings, and then turns off.
•If you do not intend to continue monitoring the same patient when the monitor is
turned on again, highlight Delete & Shut Down and then press. The monitor
turns off without saving the data and the settings.
2. Disconnect the leads and sensors from the patient.
If you pressand then decide that instead of turning off the monitor you want to resume
monitoring the same patient, do one of the following:
•Highlight Cancel and press.
•Wait for 30 seconds.
•Press.
Note
When you power down from Demo mode, you cannot save settings and
patient data.
A Physiological (high-priority) alarm warns of a patient condition, such as a vital-sign
reading that is outside of acceptable limits. When an alarm occurs, the red light on the
monitor flashes and the numerics of the violating alarm limits on the display turn red.
A Technical alarm warns of an equipment condition, such as a low battery or a detached
lead. When a technical alarm occurs, the yellow light on the monitor flashes and a
message describing the error condition appears on the display. If tones are not
suspended, the technical alarm tone sounds.
WARNING Patient injury risk. Do not set the alarm parameters to extreme
levels. Setting extreme parameters could render the alarm system useless,
causing the potential for patient injury.
Page 76
72Physiological and technical alarmsWelch Allyn Propaq LT Vital Signs Monitor
Table 6. Alarm priorities
Physiological alarms
Alarm messageDescriptionPriority
DIA HighThe diastolic pressure exceeded
the high alarm limit or is over
range.
DIA LowThe diastolic pressure exceeded
the low alarm limit or is under
range.
MAP HighThe mean pressure exceeded the
high alarm limit or is over range.
MAP LowThe mean pressure exceeded the
low alarm limit or is under range.
SYS HighThe systolic pressure exceeded the
high alarm limit or is over range.
SYS LowThe systolic pressure exceeded the
low alarm limit or is under range.
NIBP UnknownIndicates that one or more of
Systolic, Diastolic, Map parameter
values are unknown.
NIBP PR UnknownIndicates that NIBP PR parameter
value is unknown.
PR HighThe pulse value exceeded the high
alarm limit or is over range.
PR LowThe pulse value exceeded the low
alarm limit or is under range.
SpO2 HighThe SpO2 value exceeded the high
alarm limit or is over range.
SpO2 LowThe SpO2 value exceeded the low
alarm limit or is under range.
SpO2 Sat UnknownIndicates that SpO2 Sat parameter
value is unknown.
SpO2 PR UnknownIndicates that SpO2 PR parameter
value is unknown.
RR HighThe Resp Rate value exceeded the
high alarm limit or is over range.
RR LowThe Resp Rate value exceeded the
low alarm limit or is under range.
Technical Alarms
Alarm MessageDescriptionPriority
No SensorNo sensor is attached/detected by
the device while SpO2 monitoring
was active.
Defective SensorA defective sensor is detected
while SpO2 monitoring was active.
Sensor OffThe sensor is not properly attached
to the patient
Weak Pulse, Can't Find SYS/DIANot enough pulses to determine
the systolic or diastolic pressures.
No Pulse DetectedThe cuff might not be properly
applied to the patient, or the
patient might not have detectable
pulses due to shock or arrhythmias.
5 Minute WarningThe monitor battery charge is very
low; the monitor will shut down in
5 minutes or less.
High
High
High
High
High
High
High
High
High
High
High
High
High
High
High
High
High
High
High
High
High
High
Page 77
Directions for UsePhysiological and technical alarms73
Excess OffsetAt least onechannel has excessive
offset. At least one electrode is
old, contaminated, or defective.
Cable DisconnectedThe ECG cable is disconnected.Medium
Lead FailureA lead failure occurs for one or
more ECG leads.
Lead FailureOne or more electrodes have very
poor or no contact.
Noisy SignalElectrodes have poor contact and
might be dried out.
Defective SensorA defective sensor is detected.Medium
Air Leak, Check HoseThe monitor could not properly
inflate the cuff.
Kinked Hose, Check HoseThe monitor could not properly
inflate the cuff.
Over Pressure ConditionThe pressure in the cuff exceeded
the acceptable limits for the
current patient mode.
Artifact, Can't Find SYS/DIAThe systolic or diastolic pressures
are unreliable due to artifact.
Usually caused by patient motion.
Valid BP Not FoundThe patient mode setting is
incorrect or the wrong hose or cuff
is being used for the current
patient mode.
Low Battery, NIBP DisabledThe battery is too far discharged to
operate the NIBP channel.
Cuff Too Large for Patient ModeMedium
Kinked or Neonate HoseA hose is kinked or a neonate hose
is detected in the adult patient
mode.
Artifact Present, Minimize MotionThe monitor has detected toomuch
artifact to allow accurate readings.
Wrong CableThe ECG cable does not contain
1k current-limiting resistors,
which are required for Resp
operation and to protect the
monitor from damage during
defibrillation.
No SensorNo sensor is attached/detected by
the device.
Unknown SensorLow
Interference DetectedAn outside signal or outsideenergy
is preventing the reading.
Too Much Ambient LightToo much light on the sensor site. Low
Use ECG to Reduce ArtifactNIBP artifact prevents a valid
reading.
Calibrating, Please WaitNIBP Calibration is in progress.Low
Service Required, NIBP DisabledLow
Calibrating, Minimize MotionMotion is detected during a
periodic NIBP calibration.
Dock Temperature FaultThe battery is too cold or too hot to
charge.
Dock Timeout Faultthe charger has timed outLow
Dock Battery Too Lowthe battery is discharged too far to
be charged
Battery Too HighLow
Medium
Medium
Medium
Medium
Medium
Medium
Medium
Medium
Medium
Medium
Medium
Medium
Low
Low
Low
Low
Low
Low
Low
Page 78
74Physiological and technical alarmsWelch Allyn Propaq LT Vital Signs Monitor
Warning Message
Silencing an alarm
A tone sounds whenever the monitor detects an alarm condition.
To silence the currently sounding tone for 90 seconds
1. Press.
Note
Silencing the tone does not affect the other alarm indicators.
30 Minute WarningThe monitor battery charge is low,
and the monitor will shut down in
30 minutes or less.
Battery ExhaustedThe monitor battery charge is too
low to support monitor function.
Dock Overcharge Faulta cell in the battery pack is
overcharged due to cell imbalance
Dock Not Powered FaultLow
USB Lost ConnectionThe monitor detects a problem in
communication with the cradle.
Warning MessageDescriptionPriority
Some alarms off.At least one alarm is turned off.Low
Low
Low
Low
Low
•The red light (high-priority alarm) flashes or the yellow light (meduim-priority
alarm) flashes or remains on (low-priority alarm).
•After 90 seconds, if the condition is not corrected, the tone starts again.
•If the condition is corrected within 90 seconds of silencing the tone, the monitor
resets the tones for the next alarm.
If a new alarm condition occurs while an earlier alarm is silenced, the tone sounds
again.
Figure 66. Sample high-priority alarm screen
Page 79
Directions for UsePhysiological and technical alarms75
Figure 67. Sample medium-priority alarm screen
2. Check the patient and provide appropriate care.
3. After caring for the patient, verify that alarm limits are enabled and correctly set.
WARNING If you turn off or modify any alarm limits while responding to an
alarm, restore those alarm limits before you resume monitoring.
Suspending the alarm tone
If this feature is enabled in the monitor configuration (See “Monitor configuration” on
page 91), you can suspend all alarm tones for all parameters—preventing the alarm tone
from sounding if an alarm condition occurs—while monitoring a patient. If an alarm
condition occurs while the alarm tones are suspended, the monitor presents visual alarm
indicators but does not sound the tone.
In the monitor configuration, the alarm tone suspension period can be set to Disabled,to
Always On, or to a period: 90 sec or 2, 3, 4, 5, 10, 15, 30,or60 minutes.
Note
Note
The factory default suspension period is 4 minutes.
•If it is configured to Disabled, you cannot suspend the alarm tone at the monitor.
•If it is configured to Always On and you set Suspend Audible Alarms to On,
then the alarm tone remains suspended until:
•you set Suspend Audible Alarms to Off or
•monitor power is turned off and then turned on again or
•the monitor is reconfigured
Suspend Audible Alarms does not affect the behavior of the alarm reset
feature (). Pressingalways either silences a sounding alarm tone for 90
seconds or resets the audible alarm if it was already silenced.
To suspend the alarm tone
1. Access the Setup menu. (See “To access the setup menus” on page 31.)
2. Highlight Suspend Audible Alarms).
Page 80
76Physiological and technical alarmsWelch Allyn Propaq LT Vital Signs Monitor
II 1mV/cm
Exit
Trends
Snapshots
pSetup
Low
Med
80
HR/min
NIBP mmHg
Resp/min
SpO2
OffOnOn
Alarm Tone
Suspend Audible Alarms
Setup
(90 sec)
12
97
%
876 5432
3:0 0:06 P
Ad ult
Rm 23 9
Alarms
ECG
NIBP
Timings
High
Service
0:01:30 remaining
Au dio al arm s s usp end ed
ID: 010 18VDO9PBH
Configuration
Figure 68. Suspend Audible Alarms: Off
The configured suspension period—90 seconds in this example—is displayed to the
right of the line.
3. Highlight On.
Note
If audible alarms are suspended, pressingcancels the suspension.
Figure 69. Suspend Audible Alarms: On
ID: 01018VDO9PBH
Audio alarms suspended
II 1mV/cm
80
HR/min
Setup
Suspend Audible Alarms
Alarm Tone
Exit
Alarms
Service
NIBP mmHg
ECG
Trends
8765432
Adult
Rm 239
Resp/min
NIBP
Timings
Off
0:01:30 remaining
Low
Snapshots
12
Configuration
(90 sec)
Med
3:00:06P
97
SpO2
High
Setu
‘Tone suspended’ indicator
%
Configured suspension period
Suspension time remaining
•The alarm tone is suspended immediately.
If an alarm condition occurs during the suspension period, the alarm tone does
not sound.
•A countdown timer appears below the line to indicate the
time remaining in the
suspension period.
•‘Audio alarms suspended’ appears in yellow in the upper left corner of the screen.
•When the suspension period elapses, the alarm tone is again enabled.
Page 81
Directions for UsePhysiological and technical alarms77
Changing alarm limits
At the monitor
Typically, each institution determines the appropriate alarm limits for adult, pediatric, and
neonatal patients and then configures the monitor with those alarm limits before putting
the monitor into service. These become the default alarm limits for the monitor—the
alarm limits used each time the monitor is powered on.
The configured alarm limits remain in effect until the monitor configuration is changed.
However, you can temporarily change the current patient’s alarm limits.
Note
To temporarily change alarm limits for the current patient
1. Highlight the vital sign for which you want to set custom limits.
2. Press.
3. Highlight the limit you want to change.
4. Set a new alarm limit.
5. Repeat from step 3 for other alarm limits you want to change for the same vital sign.
6. Press.
When the monitor is turned off, you will have a choice of saving the current settings and
stored patient data or deleting the current settings and patient data. Neither choice affects
the default settings for the monitor.
The default alarm limits are set by your facility. Contact your facility administrator
for more information.
About ParamSet
Using ParamSet, you can quickly widen the alarm limits by a configured percentage
(relative to the patient’s alarming reading) for any vital sign.
For information about ParamSet, see “ParamSet Settings” on page 127.
Responding to a high-priority alarm
A high-priority alarm condition is indicated on the monitor in the following ways:
•The RED indicator (rightmost of three) above the display screen flashes.
•The alarm banner is displayed in the status panel.
•The numerics for the vital sign in alarm are displayed in RED.
•If alarm tones are not suspended, the alarm-tone sequence sounds repeatedly—3
short tones, a short pause, 2 short tones, and a long pause.
Page 82
78Physiological and technical alarmsWelch Allyn Propaq LT Vital Signs Monitor
Responding to a medium-priority alarm
A medium-priority alarm condition (“Overview” on page 71) is indicated on the monitor in
the following ways:
•A flashing yellow light above the monitor display.
•A yellow message on the monitor display (for example, NO ECG CABLE DETECTED).
•Repeated sounding of the medium-priority alarm tone sequence (if tones are not
suspended): 3 long tones and a pause. For a low-priority alarm, a single beep occurs.
Figure 70. Example medium-priority screen
The yellow light flashes during an alarm
and while the alarm is silenced.
To respond at the monitor to a technical alarm
1. Pressto silence the tone for 90 seconds; pressto clear the alarm.
2. Troubleshoot and correct the problem. (See “Technical alarm messages and status
messages” on page 80.)
About battery charge status
Table 7. Battery status indicators
DisplayStatus/alarm
text
Green
Green
Green
Text on a yellow field identifies the
alarm condition.
BatteryMonitorRecommended
action
Fully chargedOkayNone
Partially fullOkayNone
Partially full; charging OkayNone
Yellow
TECHNICAL ALARM
Low Battery.
Charge battery soon.
LowCan function for up to
30 minutes, but NIBP
is disabled.
Prepare to discontinue
monitoring. If possible,
insert the monitor intoa
cradle.
Page 83
Directions for UsePhysiological and technical alarms79
Table 7. Battery status indicators
DisplayStatus/alarm
Yellow
Red
Red
text
TECHNICAL ALARM
Battery Too Low.
Shutting down.
BatteryMonitorRecommended
Low; chargingNormal function.Do not remove the
Almost completely
discharged
Almost completely
discharged; charging
Shutting down soon.
Can function for up to
5 minutes, but NIBP is
disabled.
Normal function.Do not remove the
action
monitor from the cradle.
Prepare to discontinue
monitoring. If possible,
insert the monitor intoa
cradle.
If this indicator appears
when the monitor isin a
powered cradle, then
the battery is damaged
and must be replaced.
In this case, all stored
patient data will be
deleted when the
monitor is removed
from the cradle.
monitor from the cradle.
Page 84
80Physiological and technical alarmsWelch Allyn Propaq LT Vital Signs Monitor
Technical alarm messages and status messages
Table 8. Alarm messages
Alarm typeMessagePossible cause and suggested response
ECGECG Fault.
XX lead failed.
ECG Fault.
XX, XX leads failed.
ECG Fault.
Multiple lead fail.
ECG Fault.
Excessive offset.
ECG Fault.
Cable disconnected.
NIBPNIBP Fault.
Air leak.
Check hose.
NIBP Fault.
Kinked hose.
Check hose.
NIBP Fault.
Overpressure condition.
Lead XX (LA, LL, RA, C, or RL) has very poor contact or no
contact with the patient. Check for proper connection and
replace the electrode if needed.
Leads XX and XX (any two leads on a 5-lead cable) have very
poor contact orno contact with thepatient. Check for proper
connection; replace electrodes if needed.
At least three leads of a 5-lead cable or at least two leads
of a 3-lead cable have very poor contact or no contact with
the patient. Check for proper connection; replace electrodes
if needed.
At least one channel has excessive offset. At least one
electrode is old, contaminated, or defective. Replace the
electrodes.
The ECG cable is unplugged.
The monitor could not properly inflate the cuff. Check the
hose and cuff for leaks.
The monitor could not properly inflate the cuff. Check for a
hose kink between the monitor and the patient.
The pressure in the cuff exceeded the acceptable limits for
the current patient mode. Check the hose and retry the
measurement.
NIBP Fault.
Weak Pulses.
Can’t find Sys/Dia.
NIBP Fault.
Artifact.
Can’t find Sys/Dia.
NIBP Fault.
No pulses detected.
NIBP Fault.
Connect ECG to reduce NIBP artifact.
Not enough pulses to determine the systolic or diastolic
pressures, but a mean pressure is available. Squeeze all air
from the cuff and reapply it.
The systolic or diastolic pressures are unreliable due to
artifact, but a mean pressure is available. Usually caused by
patient motion.
The cuff might not be properly applied to the patient, or the
patient might not have detectable pulses due to shock or
arrhythmias.
WARNING The monitor cannot determine
whether this alarm has a physiologic cause or a
cuff application cause. Always evaluate the
patient for presence of life
when this message occurs.
NIBP artifact prevents a valid reading. Connect ECG
electrodes to improve NIBP measurements. (See “Improving
NIBP accuracy with Smartcuf” on page 65.)
-threatening conditions
Page 85
Directions for UsePhysiological and technical alarms81
Table 8. Alarm messages
Alarm typeMessagePossible cause and suggested response
NIBP Fault.
No valid blood pressure found.
The patient mode setting is incorrect or the wrong hose or
cuff is being used for the current patient mode.
NIBP Fault.
Calibrating.
Please wait.
NIBP Fault.
Calibrating.
Minimize motion.
NIBP Fault.
Low battery.
NIBP disabled.
NIBP Fault.
Service required.
NIBP disabled.
NIBP Fault.
Kinked or neonate hose.
NIBP Fault.
Artifact present.
Minimize motion.
BatteryLow Battery.
Charge battery soon.
The monitor periodically calibrates (zeroes) the NIBP
channel to make sure it can properly make NIBP
measurements. No NIBP monitoring can be done until the
calibration is completed. Other normal monitor operation
continues during NIBP calibration.
Motion is detected during a periodic NIBP calibration.
Minimize patient motion or motion on the cuff, or
disconnect the cuff. Motion-generated noise on the
pressure transducer can cause the calibration to continue
indefinitely.
The battery is too far discharged to operate the NIBP
channel. Insert the monitor into a powered cradle.
Have the monitor serviced.
A hose is kinked or a neonate hose is detected in the adult
patient mode. Check the hose and the patient mode
selection.
The monitor has detected too much artifact to allow
accurate readings. Take steps to reduce artifact. Position
the patient’s limb away from the body so the applied cuff is
not in contact with the patient’s body or any other object
such as a bed rail.
The monitor battery charge is low, and the monitor will shut
down in 30 minutes or less. Insert the monitor into the
cradle. If no cradle is available, find an alternative method
of monitoring the patient before the monitor shuts down.
Very Low Battery.
Charge battery now.
Battery Too Low.
Shutting down.
The monitor batterycharge is very low;the monitor will shut
down in 5 minutes or less. Insert the monitor into a cradle or
find another way to monitor the patient before the monitor
shuts down.
The monitor battery charge is too low to support monitor
function. Monitor operation can continue only after the
battery is recharged or replaced or until the monitor is
inserted in a powered cradle.
Page 86
82Physiological and technical alarmsWelch Allyn Propaq LT Vital Signs Monitor
Table 8. Alarm messages
Alarm typeMessagePossible cause and suggested response
ChargerCharger Fault.
SpO
2
Masimo
only
Service charger.
Charger Disabled.
Battery temperature too high or low.
Battery Fault.
Replace battery.
SpO2Fault.
No sensor detected.
SpO2Fault.
Defective SpO2 sensor.
SpO2Fault.
Defective SpO2 sensor/
No sensor detected.
Ambient light.Too much light on the sensor site. Verify that the sensor
Sensor off.The sensor is not properly attached to the patient; reconnect
Interference.An outside signal or outside energy is preventing the
Service required.
The battery is too cold or too hot to charge. Normalize the
battery temperature before attempting to charge it.
The battery is missing; the battery is discharged too far to
be charged; the charger has timed out; a cell in the battery
pack is overcharged due to cell imbalance; the fuse is
blown. Service required.
An SpO2sensor has been disconnected from the monitor
after being connected for more than a few seconds.
Replace the sensor.
Replace the sensor.
shield is covering the site and, if necessary, reduce the
ambient light, shade the sensor site, or replace the sensor.
it to the monitor and reattach it to the patient. If the sensor
is properly attached, it is defective; replace it.
reading. Remove the interference.
RespResp Fault.
Lead fail.
Resp Fault.
Noisy signal.
Check electrodes.
Resp Fault.
Inappropriate ECG cable.
GeneralMultiple Faults.Multiple technical alarms have been triggered
One or more electrodes have very poor or no contact. Check
for proper connection; replace electrodes if needed.
Electrodes have poor contact and might be dried out.
Replace electrodes.
The ECG cable does not contain 1 k current
resistors, which are required for Resp operation and to
protect the monitor from damage during defibrillation.
Replace the cable with one of the proper type.
simultaneously.
WARNING If you acknowledge this alarm
message before determining which alerts are
triggered, you cannot identify individual alerts.
Table 9. Status messages
MessageNotes
Audio alarms suspended.
Some alarms off.At least one alarm is turned off.
Press Select ( ) for controls.
NIBP off. Low battery.The battery is too low; NIBP monitoring is disabled.
-limiting
Snapshot in progress.
NIBP in progress.
Page 87
Directions for UsePhysiological and technical alarms83
Table 9. Status messages
MessageNotes
NIBP calibrating.The monitor will resume operation when the calibration is complete.
Button pad locked.Buttons (including the on/off button) remain locked until you unlock them or
Check SpO2 sensor.The sensor cable is faulty, incorrectly inserted, or incompatible. This message
Retry in progress.An NIBP automatic retry is in progress.
Low perfusion index.Masimo only.
SpO2 interference.Masimo only.
SpO2 sensor off.Masimo only.
SpO2 ambient light.Masimo only.
Low Signal IQ.Masimo only.
until an alarm sounds.
persists until a valid cable is connected or until monitor power is cycled.
Page 88
84Physiological and technical alarmsWelch Allyn Propaq LT Vital Signs Monitor
Page 89
85
5
Overview
Storing and reviewing patient data
The monitor stores up to 24 hours (at one-minute intervals) of trends information for the
patient being monitored. When data storage is at capacity, the data from each new
reading replaces the data from the oldest.
Table 10. Store and review patient data
All monitors
Type of dataTRENDS
• Tabular or graphical.
• Include maximum of 24 hours of trend data taken at:
1
-minute intervals
NIBP and SpO
Snapshots
SNAPSHOTS
• 21 seconds of numeric and waveform data: 14 seconds
before and 7 seconds after the snapshot request.
• Pressto take snapshots. (Maximum: 5 snapshots
per minute.)
• Maximum of 20 snapshots stored. At capacity, each
new snapshot overwrites the oldest.
Spot Checks
2
Where storedMonitor
Where reviewedMonitor
Printer (with AutoPrint)
Capturing a data snapshot
Pressto capture a 21-second period of numeric and waveform patient data. The
monitor captures the 14 seconds preceding and the 7 seconds following the button press.
The monitor can store 20 snapshots. After 20 snapshots have been taken, each new
snapshot replaces the oldest snapshot in memory.
Snapshots and trend data can be reviewed at the monitor.
Reviewing snapshots
To review snapshots
From any main display screen (such as Large numerics, Dual waveform...):
1. Highlight, HR/PR, SpO2, NIBP, Resp,or.
2. Press.
3. Highlight Snapshots and press. The Snapshots display appears.
Figure 71. Snapshots display
STEWART, ANN
Snapshots
2/21/05
80
HR/min
80
II 1mV/cm
Adult
NIBP mmHg
n/a
#5 of 5
1 Sec after
7762940
Rm 263
12
Resp/min
12
15:14:28
97
%
SpO2
97
14:34:58
4. The monitor first displays Select. With Select highlighted, pressto view the list of
up to 20 snapshots. To view another snapshot, highlight the snapshot menu (#5 of 5
in the example above), press, and highlight any of the available snapshots.
The blue markers at the bottom of the waveform display represent one-second
intervals. The numeric data above the waveform represents the patient’s vital signs
measured at a specific time relative to that indicated in the lower right corner. For
example: In Figure 71, the data was measured at “1 Sec after” 14:34:58.
5. To view the waveform and the numeric data for a specific time within the 21 seconds
of data, highlight 1 Sec after and then use orscroll the display to the desired time.
For example, to see the vital signs taken five seconds after the trigger point of the
snapshot, highlight 1 Sec after and press and holdas needed to increment the
display to 5 Sec after.
Figure 72. Snapshots display: 5 seconds after the button press
STEWART, ANN
Snapshots
2/21/05
80
HR/min
80
II 1mV/cm
Adult
NIBP mmHg
n/a
#5 of 5
5 Sec after
7762940
Rm 263
Resp/min
12
12
15:14:28
97
%
SpO2
97
14:34:58
Page 91
Directions for UseStoring and reviewing patient data87
STE WART, ANN
776 2940
15: 14:28
Ad ult
Rm 26 3
HR/min
NIBP mmHg
Resp/min
SpO2
II1mV/cm
Snapshots
2/21/05
5 Sec after
14:34:58
80
12
97
80
n/a
12
97
%
#5 of 5
Trends
Tabular
Tab/Wave
Graphical
Snapshots
Exit
6. To change the source of the displayed waveform, see “To change the waveform
selection” on page 52.
7. To change the size of the waveform, see “To change the waveform size” on page 52.
8. From this display, you can switch to another type of display or you can exit and return
to the primary display. To do this, highlight Snapshots and press. The Trends menu
appears, from which you can exit or switch to another display—Tabular, Tabular and
Waveform, or Graphical. To switch to the split display, for example, highlight Tab/Wave and press.
Figure 73. Snapshots display: switching to another display
STEWART, ANN
Trends
Snapshots
Tabular
Tab/Wave
Graphical
II 1mV/cm
Snapshots
Exit
2/21/05
Reviewing trends
To review trends
From any main display screen (such as Large Numerics, Dual Waveform...):
1. Highlight, HR/PR, SpO2, NIBP, Resp,or.
2. Press.
3. Highlight Trends and press. The Trends display appears.
Useandto scroll through the list of trends.
80
HR/min
80
Adult
NIBP mmHg
n/a
#5 of 5
5 Sec after
7762940
Rm 263
12
Resp/min
12
15:14:28
97
%
SpO2
97
14:34:58
•Red indicates an alarm.
•indicates that a snapshot exists for this trend. Highlightand pressto
view the snapshot.
•indicates that the snapshot for this trend was overwritten with a more recent
snapshot and is not available for viewing.
•--- indicates a value below the monitor’s measurement range.
•+++ indicates a value above the monitor’s measurement range.
Each facility can define as many default configurations as required to both meet the
needs of individual departments and units and to conform to the facility’s general
methods and standards.
•Each time the patient mode is changed, the monitor settings return to the default
configuration settings for that patient mode.
•Each time the monitor is turned on, the clinician chooses either to retain the most
recently used temporary, patient-specific settings or to reset the monitor to its
default configuration.
Clinicians can use the monitor buttons to customize settings to meet the needs of
individual patients. These custom settings can be
•discarded when the monitor is turned off and then restarted.
•retained when the monitor is turned off and then restarted.
•saved as the new default configuration for the monitor.
Note
If you change alarm limits while monitoring a patient, you can make these
changes either temporary or permanent. See “Changing alarm limits” on
page 77.
About factory configuration
When the monitor arrives from the factory, it is fully configured and ready for operation.
The factory configuration defines every setting and behavior for the monitor. Each
setting and behavior defined in the factory configuration remains in effect until it is
changed by a custom configuration or by a temporary adjustment. Unless the factory
configuration is replaced by a custom configuration, the factory settings and behaviors are
restored to the monitor each time the monitor is powered up and Start New Patient is
selected.
When a custom configuration is downloaded to the monitor, the settings and behaviors
defined in the custom configuration become the power-up defaults for the monitor, and
the factory configuration is no longer in effect. However, the factory configuration
information can always be restored by qualified service personnel and can also be
restored by downloading the configuration file PropaqLTFactoryConfig.mnt from a PC to
the monitor.
The factory settings are shown in the “Parameter configuration matrix” on page 92.
Most facilities, before putting a new monitor into service, reconfigure it to conform to
local protocol and to adapt it to the clinical environment in which it is to be used. This
reconfiguration results in a set of default settings and monitor behaviors.
Creating a custom default configuration requires the following:
•A Propaq LT Monitor Configuration Utility running on a PC, and
•A cradle with the USB data transfer option
For information about the monitor Configuration Utility, refer to “Using the Configuration
Utility to configure the monitor” on page 104.
To learn which parameters can be defined using the Configuration Utility, see the
“Parameter configuration matrix” on page 92.
About temporary configuration
While monitoring a patient, you can use the monitor buttons to temporarily adjust many of
the monitor settings to accommodate the needs of a specific patient. These custom
settings persist through a power cycle if you select to save patient data on shut-down.
They persist until they are changed manually, or until the monitor is powered down and
then powered up to monitor another patient.
Note
To learn which parameters can be temporarily defined using the user interface, see the
“Parameter configuration matrix” on page 92.
If you change alarm limits while monitoring a patient, you can make these
changes either temporary or permanent. For complete information, see
“Changing alarm limits” on page 77.
Parameter configuration matrix
This table lists the factory settings and the possible values for all parameters, and
indicates which parameters and settings can be defined by the Configuration Utility, or by
the clinician through the monitor buttons.
Note
Note
Table 11. Parameter configuration matrix
ParameterFactory settingPossible valuesWhere set
General
This table shows only those parameters that can be changed.
Except for Can Disable HR/PR Alarm Limits, this table does not show
parameters intended for use only by qualified service personnel.