Weinmann Accuvac Basic Aspirator Service manual

ACCUVAC Basic
Aspirator
WM 10700
Contents
Introduction
1.
Overview
2.
Description
2.1
2.2
3.
Operation
4.
Operating and Display Elements
5.
Maintenance
5.1
6.
Functional checks
6.1
6.2
7.
Troubleshooting
. . . . . . . . . . . . . . . . . . . . . . . . . 3
. . . . . . . . . . . . . . . . . . . . . . . . . . . 4
. . . . . . . . . . . . . . . . . . . . . . . . . . 5
Purpose Function
Disposal
Intervals Performing the functional check
. . . . . . . . . . . . . . . . . . . . . . 5
. . . . . . . . . . . . . . . . . . . . . . 5
. . . . . . . . . . . . . . . . . . . . . . . . . . 7
. . . . . . . . . . . . . . . . . . . . . . . . . 7
. . . . . . . . . . . . . . . . . . . . . 7
. . . . . . . . . . . . . . . . . . . . . 8
. . . . . . . . . . . . . . . . . . . . . . 8
. . . . . . . . . . . . . . . . . . . . . 10
. . . . . . . . . . 7
. . . . . . . 8
8.
Repairs: Information and Instructions
8.1
General
8.2
Opening the device
8.3
Closing the device
8.4
Replacing release catch
8.5
Replacing membrane keyboard
8.6
Changing power pack
8.7
Changing fuse F1 or F2
8.8
Check connector between circuit board
and power pack . . . . . . . . . . . . . . .
8.9
Fitting new circuit board WM 10780.
8.10
Fitting new pressure gage
8.11
Fitting new vacuum regulator
8.12
Fitting new pump
8.13
Changing fuse in vehicle plug
8.14
Changing the muffler
8.15
Electrical circuit diagram
9.
Spare Parts
9.1
Spare parts list
10.
Tools and Test Equipment
11.
Technical Changes
12.
Technical Data
12.1
Safe distances
13.
Repair and Test Report
. . . . . . . . . . . . . . . . . . . . . 12
. . . . . . . . . . . . . 12
. . . . . . . . . . . . . . 13
. . . . . . . . . . . . . . . 21
. . . . . . . . . . . . 22
. . . . . . . . . . . . . . . . . . . . . . . . . 24
. . . . . . . . . . . . . . . . 24
. . . . . . . . . . . . . . 26
. . . . . . . . . . . . . . . . . . . 26
. . . . . . . . . . . . . . . . . . . . . . 27
. . . . . . . . . . . . . . . . . 28
. . . . . . . . . . . . . . . . 29
. . . . . . 12
. . . . . . . . . . 13
. . . . . 14
. . . . . . . . . . . 16
. . . . . . . . . . 17
18 18
. . . . . . . . . 20
. . . . . . . 20
. . . . . . 22
. . . . . . . . . . 23
© Copyright The content and presentation are copyright protected and may only be used by authorised the course of their service operations. The content must not be reproduced or passed on to third parties. The complete documents must be returned on termination of the cooperation with
2
WEINMANN GmbH & Co. KG.
WEINMANN Service Partners in
WEINMANN.

Introduction

For decades manufacturing and marketing appliances for emer gency medical treatment, inhalation In 1986 ACCUVAC
The improved vides users with an appliance that is an indispen sable aid in many emergency health care situa tions. The for oral or transtracheal aspiration of secretion. Aspiration is indicated in particular for tracheoto my patients undergoing respiration.
The indications include:
Neuromuscular disorders, e.g. paraplegia, amyotrophic lateral sclerosis (ALS)
Late symptoms of polio
Muscular dystrophies
Cystic fibrosis
Kyphoskoliosis and posttubercular syndrome
Chronic obstructive lung diseases, e.g. severe chronic bronchitis
Pulmonary fibrosis, e.g. silicosis
Other forms of alveolar hypoventilation
Using the ACCUVAC Basic aspirator gives the user more time to look after the patient and perform other measures.
WEINMANN has been developing,
oxygen therapy and
therapy.
WEINMANN launched the first
aspirator on the market.
ACCUVAC Basic
ACCUVAC Basic
aspirator pro
is primarily intended
-
-
The purpose of these Service and Repair Instruc-
-
-
-
tions is to make sure you as an expert specialist are familiar with the ACCUVAC Basic aspirator: its functionality and technology and how to repair it. Combined with training you have already re­ceived from WEINMANN, this makes you a mem­ber of the “authorized expert personnel” category, which means you can give your customers proper instruction, remedy faults independently, perform the final checks specified in the Operating Instruc­tions, and carry out any repairs in accordance with these Service and Repair Instructions.
In the event of a warranty claim the ACCUVAC Basic aspirator is to be sent to WEINMANN.
So that we can process warranty claims or re­quests for generous treatment of complaints, please send the purchase receipt (invoice) with the appliance.
Repairs may only be carried out by WEINMANN or by expert personnel.
You are responsible for any repairs you make and for the relevant warranty!
Only original WEINMANN spare parts must be used for repairs.
Please remember:
Your customer trusts you and relies on your quality, just as you rely on WEINMANN.
Note:
For the following information, please consult the ACCUVAC Basic Operating Instructions:
Safety information
Fitting accessories
Operation
Hygienic preparation
Warranty
Introduction
3

1. Overview

-1,0
-0,8
-0,6
-0,4
-0,2
0
22 Vacuum
Fingertip
19 Aspiration tube with
Disposable collection canister
21 Disposable bag
20 Intermediate tube
tube
24 Collection canister
23 T-piece
11 Adapter/charger plug
25 Holder
Mains / charger unit
FW7405M/14,
complete WM 2610
Accessory bag
WM 10655
Wall bracket
WM 15208
Connection cable
WM 10650
15 Fuse F100 16 Fuse F101 (hidden)
ACCUVAC Basic interior
7 Disposable
6 Release catch
5 Membrane keyboard
canister
switch
4 On/Off
27 Vent tab
28 Filter cover
31 Bracing
29 Bacteria filter
30 Locking tab
32 Secretion cover
18 Plug X100 17 Power pack
Accessories
26 Re-usable collection canister WM 15269
8 Motor unit
9 Tube holder plate
33 Ball (overfill guard)
34 Sealing ring
35 Fingertip
10 Loops for
37 Collection canister
36 Aspiration tube
accessory bag
38 Holder set WM 15271
3 Voltage indicator
2 Vacuum control
(hidden)
13 Muffler
ACCUVAC Basic from front with disposable collection canister
1 Pressure gauge
ACCUVAC Basic from rear without collection canister
14 Suction port
12 Power socket
plug
(hidden)
11 Adapter/charger

2. Description

-1,0
-0,8
-0,6
-0,4
-0,2
0

2.1 Purpose

ACCUVAC Basic is a mobile and portable electri­cally powered aspirator (suction pump).
It is used for:
aspirating accumulations of blood, secretions and food from the oral cavity, the nose and throat region and the bronchial system;
deflating vacuum mattresses and inflatable splints.
ACCUVAC Basic:
can when used by a skilled operator eliminate obstruction of the respiratory tract and hence the risk of respiratory failure;

2.2 Function

An electrically operated diaphragm pump generates the vacuum necessary for aspiration.
Use the vacuum control 2 to set the desired vacuum, which is continuously variable between –0.05 bar and –0.8 bar. You can read off the set value on the pressure gage 1.
cuts energy consumption by reducing power output on reaching the necessary vacuum;
can optionally be powered by a rechargeable internal power pack;
or by an external DC source supplying 12.0 -
13.8 V;
is also suitable for use in wards.
ACCUVAC Basic must not be used:
in medical rooms where potential equalization is necessary (e.g. heart surgery);
in explosion-risk areas.
The aspirated material passes via the aspiration tube
36 into the disposable collection canister 7.
Disposable collection canister
The disposable collection canister 7 consists of the collection canister 24 itself, the disposable collection bag 21 with intermediate tube 20, the aspiration tube with fingertip 19, the T-piece 23 and the vacuum tube
22.
The disposable collection canister 7 is inserted in the holder 25 screwed on the side of the motor unit 8.
The vacuum tube 22 of the disposable canister 24 is pushed onto the suction port 14 of the motor unit 8.
Description 5
The aspirated material passes through the aspira­tion tube 19 into the disposable collection bag 21. The disposable collection bag is intended for once-only use. When it is full, you can remove it from the collection canister 24 and dispose of it complete with contents.
An overflow valve filter is integrated in the disposable collection bag 21. This prevents secre­tions and liquid from finding their way into the motor unit 8 and passing into the environment via the muffler 13.
Power supply
Power for operating the appliance and charging the power pack can be drawn from:
the built-in power pack 17.
a 12-volt electrical cable system, using the connecting cable WM 10650 which is available as an accessory.
the mains and charger unit WM 2645 which is available as an accessory.
Important!
The AC adapter/charger plug 11 supplied with the appliance is only designed for charging the power pack with the appliance switched off. It must not be used to operate the appliance itself, as this will over­load it and cause it to overheat.
The voltage indicator 3 shows the charge status of the power pack.
Charging of the power pack starts automatically as soon as the appliance is switched off and con­nected to an external power supply (see ”12. Technical Data“ on page 27).
6 Description

3. Operation

ACCUVAC Basic may only be used by trained staff in­structed in aspiration techniques. Incorrect use can cause serious bodily harm.
Operation is described in the operating instructions.

4. Operating and Display Elements

-0,4
-0,6
-0,8
-1,0
-0,2
0
2 Vacuum control 3 Voltage indicator 4 On/Off switch1 Pressure gauge

5. Maintenance

The ACCUVAC Basic needs no maintenance, but please be sure to observe the intervals specified for regular final checks (see ”6.1 Intervals“ on page 8).
In the interests of maintaining power pack opera­tion and life, we recommend you to calibrate the power pack every 8 weeks as described in

5.1 Disposal

Do not dispose of the unit with domestic waste. For proper waste disposal of the equipment, please contact an approved and certified waste disposal site for electronic goods. Ask your Environmental Officer or town council for the address.
Section 8.1.2 of the instructions for use. This in­cludes the required discharging and recharging of the power pack.
We recommend that you have any servicing, such as inspections and repair work, carried out by the manufacturer – WEINMANN – or by expert per­sonnel.
Operation 7

6. Functional checks

If the final check reveals defects or deviations from the specified values, ACCUVAC Basic must not be used again until the faults have been rectified.
We therefore recommend that you always keep a stock of the following:
Disposable collection bag 21 WM 10732
Aspiration tube with fingertip 19 WM 10733

6.1 Intervals

To ensure that a properly functioning ACCUVAC Basic is always available, it is essential to observe the fol­lowing intervals.
Before every use
Perform a functional check (see ”6.2 Perform­ing the functional check“ on page 8).
After every use
Clean, disinfect and/or sterilize the unit and its parts (see operating instructions ”5. Cleaning and disinfecting“);
Perform a functional check (see ”6.2 Perform­ing the functional check“ on page 8).
Every 3 months
Check the charge level of the power pack 17 by switching on ACCUVAC Basic and read­ing the voltage indicator 3. If the top LED of the voltage indicator does not light up, the power pack should be recharged (see operating in­structions ”4.4 Charging ACCUVAC Basic“).
At least every 6 months
Perform a functional check (see ”6.2 Perform­ing the functional check“ on page 8).
Make a visual inspection of the muffler for clogging. If it is clogged, fit a new muffler (see ”8.14 Changing the muffler“ on page 22).
After all repairs
Clean, disinfect and/or sterilize the unit and its parts (see operating instructions ”5. Cleaning and disinfecting“);
Perform a functional check (see ”6.2 Perform­ing the functional check“ on page 8).

6.2 Performing the functional check

1. Assemble ACCUVAC Basic ready for use.
2. Check that all tubes and the individual parts of the disposable collection canister 7 are in per­fect condition. Any damaged or worn parts must be replaced.
3. Check that all tubes are securely connected and the disposable collection bag 21 is firmly installed.
4. Switch on ACCUVAC Basic.
8 Functional checks
5. Check the charge level on the voltage indica­tor 3. Always recharge the power pack with­out delay if the yellow LED on the voltage indicator lights up (see operating instructions ”4.4 Charging ACCUVAC Basic“).
6. Battery test A battery test should always be performed when there are doubts about the performance of the rechargeable battery, however at the latest two years after the battery was last changed.
Procedure
Charge the ACCUVAC Basic for 8 hours us­ing the WM 2610 mains / charger unit FW7405M/14, complete or for 14 hours us­ing the WM 10750 plug-in power supply unit. Set a short interval timer to 20 minutes and start the ACCUVAC. After an operating time of 20 minutes the red LED should not be lit up and the ACCUVAC should be in opera­tion.
If the red LED lights up after 20 minutes, or the ACCUVAC is no longer working, the battery is spent and must be replaced. In this case please replace the battery as described in sec­tion 8.6.
7. Insert the stopper in the fingertip 19.
Stopper
8. Use your thumb to hold the suction nozzle closed.
9. Switch on the aspirator and select the maxi­mum vacuum of –0.8 bar by turning vacuum control 2 fully to the right. ACCUVAC Basic must reach this vacuum in not more than 20 seconds.
If the aspirator takes longer than 20 seconds to reach this vacuum, this means its suction capacity is reduced. Check for possible faults (see ”7. Troubleshooting“ on page 10).
10. Attach the test pressure gauge 0 to –1 bar WM 15294 to the fingertip.
11. Check the maximum vacuum for accuracy with test pressure gauge 0 to –1 bar. The vacuum must be –0.8 bar, tolerance +/– 0.04 bar.
12. Remove the test pressure gauge from the fingertip.
13. Switch ACCUVAC Basic off again.
Suction tube
Fingertip
-0,4
-0,6
-0,2
-0,8 bar
0
O
2
-1,0
Functional checks 9

7. Troubleshooting

… fit new pump (8.12, page 21)
… fit new circuit board (8.9, page 18)
Perform several charge/discharge cycles. If
unsuccessful, fit new power pack (8.6,
page 16)
Correct polarity and if necessary replace fuse
F1 (8.7, page 17)
Make sure connection snaps in
Check that all tubes are firmly connected and
disposable bag is securely installed
… charge power pack (operating instructions
4.4, page 16)
… fit new pump (8.12, page 21)
Connect pump to functioning circuit board and
switch on. If pump does not run, …
Connect circuit board to functioning pump and
switch on. If pump does not run, …
Faulty pump
Circuit board defective
Fuse F100 or F101 in appliance defective Fit new fuse (8.7, page 17)
Fuse in vehicle plug defective Fit new fuse (8.13, page 22)
Power pack fully discharged
Incorrect polarity of vehicle power socket
Snap-in connection between circuit board and
Disconnect secretion container from pump unit.
Switch on pump and run at a setting of 0.8 bar.
Hold thumb over suction port. The maximum
vacuum is reached within 5 sec. and the pump
stops. If the pumps starts up again within 10 to
20 sec., the leak is in the pump unit.
Connect mains / charger unit FW7405M/14,
complete WM 2610. If the pump now runs
much more “powerfully” than with the power
pack, you must …
Connect pump to functioning circuit board and
power pack not properly engaged
Leak in suction side of appliance
Power pack not sufficiently charged
Fault in pump
switch on. If pump does not run, …
Muffler clogged Fit new muffler (8.14, page 22)
Vacuum regulator defective Fit new vacuum regulator (8.11, page 20)
Fault Cause Localize fault Remedy
Appliance does not start, O/I indicator and
voltage indicator show ready for operation
Appliance does not start, O/I indicator does
not show ready for operation
Appliance does not reach maximum vacuum of
–0.8 bar in 20 seconds, but voltage indicator
shows ready for operation
No vacuum is generated
No reading on pressure gage
External power supply must be between 12.0 V
and 13.8 V
Make sure connection snaps in (8.8, page 18)
Use AC adapter/charger plug WM 10750 or
charger WM 2610 (accessory)
Check vehicle electrical system
Fit new power pack (8.6, page 16)
Check internal connection between printed
circuit board (X100) and power pack (8.8,
page 18)
Do not switch on pump under vacuum
Inadequate external power supply
Fault Cause Localize fault Remedy
Fuse F100 or F101 defective Fit new fuse (8.7, page 17)
Fuse in vehicle plug defective Fit new fuse (8.13, page 22)
Snap-in connection between circuit board and
power pack not properly engaged
Battery does not recharge
Charger does not meet specifications
Vehicle electrical system is not supplying 12.0 –
13.8 V
Power pack damaged by being fully
discharged
Power pack at end of service life
Internal connection faulty
Fuse F101 defective Fit new fuse (8.7, page 17)
Power pack is not fully charged up despite
charging for 14 hours (top LED in voltage
indicator does not light up)
Top LED of voltage indicator lights up after
charging, but pump does not run when
powered by power pack
Pump is often switched on under vacuum
conditions
Pump defective fit new pump (8.12, page 21)
Fuse F2 frequently blows
Troubleshooting 11

8. Repairs: Information and Instructions

8.1 General

An ESD workplace is essential for making repairs to the ACCUVAC Basic.
No work should be performed on the appliance without a detailed knowledge of the Operating In­structions and the Service and Repair Instructions, which must always be complied with.
ACCUVAC Basic is only intended for the purpose described (see ”2.1 Purpose“ on page 5).

8.2 Opening the device

Tools and equipment required:
Phillips screwdriver size 2
1. Switch off ACCUVAC Basic.
2. Disconnect the aspirator from the external power supply.
3. Remove collection canister 7 and any accessories.
4. Unscrew the holder 25 for the disposable col­lection canister 7.
When replacing components or individual parts, be sure to use only original WEINMANN parts.
A functional check (see ”6.2 Performing the func­tional check“ on page 8) must be performed after every repair.
When ordering the rear part of the case, please state the appliance type, appliance number and year of manufacture.
5. Open the case by unscrewing the 6 cross­head screws 52. When opening the case, be careful not to damage the silicone sealing cord.
6. Carefully pull the front and rear case elements apart.
12 Repairs: Information and Instructions
52

8.3 Closing the device

Tools and equipment required:
Phillips screwdriver size 2
1. Carefully put the front and rear case elements together again.
2. Screw the case together again, making sure the silicone sealing cord is correctly inserted and is not jammed or otherwise damaged.
3. Perform a functional check (see ”6.2 Perform­ing the functional check“ on page 8).

8.4 Replacing release catch

52
Tools and equipment required:
Phillips screwdriver size 2
Screwdriver, size 1 ;
Flat or pointed pliers.
1. Open the device (see ”8.2 Opening the de­vice“ on page 12).
2. Place the ACCUVAC Basic on its front.
3. Push out retaining pins 64 from release catch
63 and remove them.
4. Remove the old or defective release catch 63. To do so, use a flat / blunt object to press down snap lock 66, which is located below release catch 6.
5. Take the new release catch 63 and insert it in the rear wall of the case.
6. Take O-ring 65 and locate it in the rear wall of the case between the loops of the release catch and the rear wall attachment point.
7. Take the retaining pins 64 and insert them in the bushing from outside to inside until you hear them click into place.
8. Close the device (see ”8.3 Closing the de­vice“ on page 13).
9. Perform a functional check (see ”6.2 Perform­ing the functional check“ on page 8).
63
64
65
66
Repairs: Information and Instructions 13

8.5 Replacing membrane keyboard

Tools and equipment required:
Phillips screwdriver size 1;
7mm open-end wrench.
1. Open the device (see ”8.2 Opening the de­vice“ on page 12).
2. Carefully disconnect the power pack connec­tor 18 (X100) from the circuit board.
3. Unscrew the power pack holder (4 cross-head screws 41).
4. Detach both leads from the power pack 17.
5. Remove the power pack.
X100
6. Carefully remove cap 45 of knob 44, p.ex. à using a razor blade (for example).
7. Slightly loosen the nut on the adjuster knob and pull the adjuster knob off the regulator unit.
8. Carefully pull tube 58 off regulator unit 46.
9. Use a 7mm open-end wrench to unscrew the tube connection from the pressure gauge.
10. Cut open cable tie 43 (locking strip).
45
44
59 586061
41
17
46
50
14 Repairs: Information and Instructions
43
11. Carefully push pressure gauge 1 upward out of resilient ring 42.
12. Remove resilient ring 42.
13. Disconnect the electrical power pack connec­tors X100, X101 and X103 from the board.
14. Detach the front keyboard from inside through the front of the case, by pressing the mem­brane keyboard out upwards and carefully pulling it off.
15. Clean the old adhesive area until no adhesive residues are left.
16. Take the new membrane keyboard 5 and re­move the protective layer from the adhesive surface. Run the ribbon cable through the opening in the case front.
Take care when inserting the ribbon cable through the opening in the housing; it must be properly routed without any kinks.
17. Stick the new membrane keyboard 5 in the correct position on the case.
18. Pull the protective film off the new membrane keyboard.
19. Insert resilient ring 42 in the front of the case again.
20. Using a little methylated spirits, carefully insert the new pressure gauge 1 downwards into the resilient ring 42 (making sure to align it correct­ly). Roll the rubber cuff onto the pressure gauge with your fingers.
42
1
42
1
21. Secure the pressure gage by fitting cable tie
43 around the resilient ring.
22. Insert the board again and screw it up firmly.
23. Carefully restore the electrical connections
X100, X101 and X103.
24. Carefully push tube 58 onto the regulator unit.
25. Attach the tube connector to the pressure gage using a 7mm open-end wrench.
26. Insert the power pack again and screw it firmly in place.
43
Repairs: Information and Instructions 15
27. Make the electrical connections to the power pack.
28. .When replacing the knob, ensure that the gap between knob and keyboard membrane (with the spindle screwed in) is approx.
1.0 mm.
1mm
29. Use a 10 mm tubular wrench to tighten the nut securing knob 44.
30. Insert cap 45 (slit in cap must match ridge in knob).
31. Close the device (see ”8.3 Closing the de­vice“ on page 13).
32. Perform a functional check (see ”6.2 Perform­ing the functional check“ on page 8).
45
44

8.6 Changing power pack

ACCUVAC Basic is fitted with a high-grade lead cell power pack.
Tools and equipment required:
Phillips screwdriver size 2.
1. Open the device (see ”8.2 Opening the de­vice“ on page 12).
2. Carefully disconnect the power pack connec­tor 18 (X100) from the circuit board.
X100
3. Unscrew the power pack holder (4 cross-head screws 41).
4. Detach both leads from the power pack 17.
5. Remove the faulty power pack.
Help protect the environment!
Don’t throw the old power pack in the gar­bage can – take it to a local collection point for environment-friendly disposal.
16 Repairs: Information and Instructions
41
17
6. Attach the two cables to the connections on the new power pack: red to plus; black to
minus.
If you accidentally reverse the connections, the electronic system will be protected by fuse 16 (F2), which will then have to be replaced.
7. Fit the new power pack with its holder, making sure you do not jam any cables.
8. Carefully push power pack connector 18 (X100) onto the circuit board until it snaps into place.
9. Close the device (see ”8.3 Closing the de­vice“ on page 13).
10. Perform a functional check (see ”6.2 Perform­ing the functional check“ on page 8).

8.7 Changing fuse F1 or F2

X100
Important! Avoid touching the electronic components on the cir­cuit board, as this could damage them (electrostatic charge).
Tools and equipment required:
Phillips screwdriver size 2.
1. Open the device (see ”8.2 Opening the de­vice“ on page 12).
2. Remove the faulty fuse 15/16. The fuses are identified on the circuit board.
3. Insert a new fuse. Always use approved fuses (see ”12. Technical Data“ on page 27).
4. Close the device (see ”8.3 Closing the de­vice“ on page 13).
5. Perform a functional check (see ”6.2 Perform­ing the functional check“ on page 8).
Repairs: Information and Instructions 17

8.8 Check connector between circuit board and power pack

Tools and equipment required:
Phillips screwdriver size 2.
1. Open the device (see ”8.2 Opening the de­vice“ on page 12).
2. Check connector X100.
3. Close the device (see ”8.3 Closing the de­vice“ on page 13).
4. Perform a functional check (see ”6.2 Perform­ing the functional check“ on page 8).
X100

8.9 Fitting new circuit board WM 10780

Tools and equipment required:
Phillips screwdriver size 2.
1. Open the device (see ”8.2 Opening the de­vice“ on page 12).
2. Carefully remove the electrical power pack connectors X100 and X101 of wiring harness
53 from the circuit board.
53
3. Carefully open up cable grip of ribbon cable connector X103. Then carefully remove ribbon cable (do not touch the ribbon cable contacts with your fingers, as this can cause oxidation).
X103
X101 X100
X101 X100
18 Repairs: Information and Instructions
4. Unscrew circuit board (3 screws
52
).
5. Insert the new circuit board in reverse order.
Important!
The lengths of tubes WM 10766 and WM 10765 are not compatible with all circuit boards. The over­view below shows you the cases in which you will have to adapt the length of the tubes:
Old circuit board New circuit board Action
WM 10780 Index
WM 10780 Index ≤ C WM 10780 Index > C
WM 10780 Index > C WM 10780 Index > C No modification required
≤ C WM 10780 Index ≤ C No modification required
Use tubes with new lengths: – WM 10766 Index > a (42 mm instead of 50 mm) – WM 10765 Index > a (208 mm instead of
200 mm)
52
Carefully connect electronic power pack con-
X101
nector
. Then route connecting cable so
it does not come into contact with the pump.
Important:
Do not connect electrical power pack connector
Wait half a minute to allow the capacitors on the circuit board to discharge.
6. Carefully connect the electrical power pack connector
7. Close the device (see ”8.3 Closing the de­vice“ on page 13).
8. Perform a functional check (see ”6.2 Perform­ing the functional check“ on page 8).
X100
X100
yet.
.
Repairs: Information and Instructions 19

8.10 Fitting new pressure gage

Tools and equipment required:
Phillips screwdriver size 2.
1. Open the device (see ”8.2 Opening the de­vice“ on page 12).
2. Use a 7mm open-end wrench to unscrew the tube connection from the pressure gauge.
3. Cut open cable tie 43.
4. Carefully push faulty pressure gauge upward out of resilient ring 42.
5. Using a little methylated spirits, carefully insert the new pressure gauge 1 downwards into the resilient ring 42 (making sure to align it correct­ly). Roll the rubber cuff onto the pressure gauge with your fingers.
6. Secure the pressure gage by fitting cable tie
43 around the resilient ring.
7. Attach the tube connector to the pressure gauge using a 7mm open-end wrench.
8. Close the device (see ”8.3 Closing the de­vice“ on page 13).
9. Perform a functional check (see ”6.2 Perform­ing the functional check“ on page 8).
43
42143

8.11 Fitting new vacuum regulator

Tools and equipment required:
Phillips screwdriver size 2,
Tubular wrench, 10 mm.
1. Open the device (see ”8.2 Opening the de­vice“ on page 12).
2. Carefully remove cap 45 of knob 44, using a razor blade (for example).
3. Use a 10 mm tubular wrench to unscrew the knob.
45
44
20 Repairs: Information and Instructions
4. Remove tube 58 from vacuum regulator.
5. Unscrew the 2 cylinder-head screws 47 and remove the faulty vacuum regulator.
6. Fit the new regulator unit 46 in reverse order.
7. When replacing the knob, ensure that the gap between knob and keyboard membrane (with the spindle screwed in) is approx. 1.0 mm.
8. Use a 10 mm tubular wrench to tighten the nut securing knob 44.
9. Insert cap 45 (slit in cap must match ridge in knob).
10. Close the device (see ”8.3 Closing the de­vice“ on page 13).
11. Perform a functional check (see ”6.2 Perform­ing the functional check“ on page 8).
58
46
47
1mm

8.12 Fitting new pump

The pump is only available as a complete replacement unit.
Tools and equipment required:
Phillips screwdriver size 2.
1. Open the device (see ”8.2 Opening the de­vice“ on page 12).
2. Detach silicone fabric tube 56 from inlet port of pump.
3. Detach rolled-up silicone tube 57 from muffler connection.
4. Unscrew 4 screws 41 from support.
5. Detach tube 57 from faulty vacuum pump, at­tach it to outlet of new pump 44 and wind it round motor.
6. Finish fitting the new vacuum pump in reverse order.
57
44
56
7. Close the device (see ”8.3 Closing the de­vice“ on page 13).
8. Perform a functional check (see ”6.2 Perform­ing the functional check“ on page 8).
41
Repairs: Information and Instructions 21

8.13 Changing fuse in vehicle plug

Tools and equipment required:
Phillips screwdriver size 2.
1. Use a screwdriver to open the vehicle plug.
Note:
The central contact of the plug is the plus pole. The plus lead of the cable has either a square cross-section or colored markings. The outer contact of the plug is the minus pole. The minus lead of the cable is round and black.
Caution:
Check the vehicle plug for correct polarity. Re­versed polarity can damage ACCUVAC Basic.
2. Change the faulty fuse 62. Use only approved fuses (see ”12. Technical Data“ on page 27).
3. Screw the vehicle plug together again.
4. Perform a functional check (see ”6.2 Perform­ing the functional check“ on page 8).
62

8.14 Changing the muffler

Tools and equipment required:
Phillips screwdriver size 2.
1. Use a screwdriver to unscrew the cover plate 48 (2 cross-head screws 49).
2. Remove the old muffler 13.
49
48
13
3. Insert a new muffler 13.
22 Repairs: Information and Instructions
4. Refit the cover plate 48.
Note that there is a projecting lug on the back of the cover plate. Be sure to fit the cover plate so that this lug locates the muffler in position.
5. Perform a functional check (see ”6.2 Perform­ing the functional check“ on page 8).

8.15 Electrical circuit diagram

48
projecting lug
+
Power pack, lead-cell WM 10747
-
black
X2
black
X1
Vacuum pump WM 10674
Membrane keyboard, Basic
red
red
red
black
4 3
X1
2 1
Printed circuit board WM 10780
X2
X3
Repairs: Information and Instructions 23

9. Spare Parts

9.1 Spare parts list

Note:
The item numbers in the following table are identical with the numbers used in these Service and Repair Instructions and the Operating Instructions.
Item No. Description Order No.
1 Pressure gauge WM 10754
5 Membrane keyboard ACCUVAC Basic WM 10745
7
24 21 19
9 Tube holder plate, green WM 10723
11 Adapter/charger plug WM 10750
13 Muffler WM 10665
15 Fuse, external power supply (F1) WM 2692
16 Fuse, power pack (F2) WM 2692
17 Lead-acid power pack WM 10747
Disposable collection canister consisting of: – Collection canister – Disposable collection bag – Aspiration tube with fingertip
WM 10730
WM 10731 WM 10732 WM 10733
22 Vacuum tube WM 10740
23 T-piece WM 10738
25
26
Holder set for disposable collection canister consisting of: – Holder – Fixing elements
Re-usable canister, complete
WM 15172
WM 10735 WM 51091
WM 15269
consisting of:
28 29 31 32 33 34 35 36 37
38 Holder set for re-usable canister
– Filter cover – Bacteria filter – Bracing clip – Secretion cover – Ball (overfill guard) – Sealing ring – Nozzle with fingertip – Aspiration tube – Collection canister
WM 10632 WM 10675 WM 10641 WM 10636 WM 10643 WM 10635 WM 10666 WM 10662 WM 10631
WM 15271 consisting of: – Holder – Fixing elements
39 Front case element, Basic, assembled, reconditioned* WM 10706
WM 10640
WM 53053
24 Spare Parts
40 Rear case element, Basic, assembled, reconditioned* WM 10707
41 Oval head screw KB 40x14 WM 23158
42 Resilient ring WM 10757
43 Cable tie WM 12498
2
Vacuum control consisting of:
44 45 46 47
48 Cover plate, green WM 10725
49 Countersunk screw, KB 40x12 WM 58360
50
51 PCB, Basic WM 10780
52 Oval head screw M3x14 WM 53032
53 Internal wiring harness, Basic WM 10786
54 Cable tie WM 4668
55 Oval head screw for clip, KB 35x8 WM 58350
– Knob – Cap – Regulator unit – Cylinder head screw M3x8, DIN 912
Vacuum pump, complete (new)* Vacuum pump, complete, exchange unit*
Tube system, internal
WM 10772 WM 2635 WM 10770 WM 50595
WM 10694 WM 10605
WM 10760
consisting of:
56
57 Tube, silicone 7x2.5 – 690 mm long WM 10668
58
59 Tube, silicone fabric 6x3.5; 208 mm long WM 10765
60 Tube, silicone fabric 6x3.5; 42 mm long WM 10766
61 T reducer 6-4-6 WM 10764
62 Fuse, vehicle plug WM 10673
6
– T reducer 8-6-8 – Tube, silicone fabric 8x3.5 – 210 mm long – Tube, silicone fabric 8x3.5 – 70 mm long
Pressure gauge tube consisting of: – Tube, silicone fabric 3x3 140mm long – Sintered cylinder 2x
Set release catch, green
WM 10763 WM 10669 WM 10664
WM 10752
WM 10753 WM 10658
WM 15397
consisting of:
63 64 65
66 Snap lock WM 10627
– Release catch, green – Retaining pin – O-ring 2.9 x 1.78
WM 10724 WM 10697 WM 1145/80
Washer B 4.3 DIN 125 WM 50240
Washer B 3.2 DIN 125 WM 50235
Spring washer A3 DIN 127 WM 50350
Round cord 930 mm long WM 10612
Operating instructions WM 16138
* When ordering, please state type, appliance no. and year of manufacture
Spare Parts 25

10. Tools and Test Equipment

Following is a list of all tools and test equipment mentioned in these Service and Repair Instructions.
See the relevant chapter for details of the tools and test equipment needed in each case.
Cross-head screwdriver, size 1
Cross-head screwdriver, size 2
Tubular wrench, 10 mm – for knob
Open-end wrench, 7 mm – for tube connection to pressure gage
Ball-joint socket wrench 2.5 mm – for regulator valve
Wire cutters – for cable tie on pressure gage
Pliers
Vacuum test pressure gage set WM 15294
Special tools can be obtained from the manu­facturer, WEINMANN.

11. Technical Changes

Technical change From Device No. Date
Case reinforced, use of washers discontinued 5372 20.11.01
Stronger release catch 5932 22.02.02
Vacuum pump without side walls 9412 28.10.02
Appliance plate with e 1 mark 10735 28.06.04
Redesigned circuit board 22502 05.03.09
26 Tools and Test Equipment

12. Technical Data

ACCUVAC Basic
Product category according to 93/42/EEC
Dimensions WxHxD in mm
Weight approx. 5.4 kg
Canister volume 900 ml
Suction capacity at 12 V with free flow
Max. vacuum at 12 V 0.8 bar (80 kPa)
Aspiration tube
Overflow valve filter
Motor output 50 W
Rated voltage 12 V
Maximum current consumption
Temperature range Operation Charging Storage
Electromagnetic compatibility: Radio interference suppression Radio interference resistance
Classification according to EN 60601-1: Protection against electric shock: class II Degree of protection against electric shock: BF Degree of protection against water: IPX 1(drip-water)
Classification according to EN ISO 10079-1
II b
385x280x140
> 20 l/min
diam. 5 mm, length 1800 mm
at least 99.8 % for particle size 2.8 µm
3.5 A
–18 °C to +40 °C + 0 °C to +40 °C –40 °C to +70 °C
EN 60601-1-2
EN 55011 IEC 1000-4 Parts 2–5 & 11
high vacuum/high flow
ACCUVAC Basic
EN 60601-1,
Norms complied with
Vehicle plug fuse
Internal fuse F1 external power supply
Internal fuse F2 power pack
Power pack type Lead cell, 3.4 Ah
Charging voltage 12.0 to 13.8 V
Operating time after charging for 14 hours
Operating mode S2 60 min
Service life of power pack
AC adapter/charger
Disposable collection canister: Collection canister: PC Disposable collection bags, tubes: PVC Nozzle with fingertip: PE
Case PC/ABS
Tube holder plate ABS
EN ISO 10079-1, EN 1789
8 A, DIN 72581, identification color white
4 A slow-acting, low breaking capacity G fuse links 5x20 mm, conforms to IEC 127
4 A slow-acting, low breaking capacity G fuse links 5x20 mm, conforms to IEC 127
30 min at maximum suction
400 charge/discharge cycles in approx. 3 years
primary: 230 V/50 Hz secondary: 13.8 V /0.3 A
Materials
From serial no. 10735/2004
(Subject to technical change without notice)
Technical Data 27

12.1 Safe distances

The ACCUVAC Basic is intended for operation in an electromagnetic environment in which high-fre­quency interference variables are controlled. The customer or user of the ACCUVAC Basic can help avoid electromagnetic interference by maintaining
Recommended safe distances between portable or mobile HF telecommunication
devices (e.g. mobile phones) and the ACCUVAC Basic
Rated output of
HF device
in W
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.69 3.69 7.38
100 11.67 11.67 23.33
150 kHz - 80 MHz
d=(3,5/V1) x √√√√P
Safe distance depending on transmission frequency
the minimum safe distance between portable and mobile high-frequency telecommunication devices (transmitters) and the ACCUVAC Basic depending on the rated output of the transmitter as given be­low.
in m
80 MHz - 800 MHz
d=(3,5/V1) x √√√√P
800 MHz – 2,5 GHz
d=(3,5/V1) x √√√√P
28 Technical Data

13. Repair and Test Report

Keep a record of all tests or repairs performed (please copy attached form for use).
Date Signature
Service performed in
accordance with ACCUVAC-
Service and Repair Instructions
Company
___________ ________________
maintenance / repair / comments
Company
Date Signature
___________ ________________
Company
Date Signature
___________ ________________
Company
Date Signature
___________ ________________
WEINMANN GmbH + Co.
22525 Hamburg
Device master data Maintenance and repair work carried out in accordance with service document
WM 10600 ACCUVAC Rescue
WM 10700 ACCUVAC Basic
Manufacturers:
Device mode: ACCUVAC
Device no.: _______________________
Production date: ___________________
Repair and Test Report 29
WEINMANN Geräte für Medizin GmbH+Co. KG
P.O. Box 540268 • D-22502 Hamburg Kronsaalsweg 40 • D-22525 Hamburg T: +49-(0)40-5 47 02-0 F: +49-(0)40-5 47 02-461 E: info@weinmann.de www.weinmann.de
Center for Production, Logistics, Service
WEINMANN Geräte für Medizin GmbH+Co. KG Siebenstücken 14 D-24558 Henstedt-Ulzburg T: +49-(0)4193-88 91-0 F: +49-(0)4193-88 91-450
WM 16291d - 06.09
Loading...