canister
Fitting an accessory bag
Fitting a rinsing glass
. . . . . . . . . . . . . . . . . . .
Preparing for aspiration
Aspiration
After aspiration
Charging
. . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . .
. . . . . . . . . . . . . .
. . .
re-usable
. . . . . . . . . . . . . . .
collection
. . . .
. . . . . .
. . . . .
. . . . . . . . . . . . . .
. . . . . . . . . .
ACCUVAC
Basic
. .
10
11
11
13
13
14
16
16
2
5.
Hygienic preparation
2
3
5
5
5
8
8
5.1
Preparations
5.2
Cleaning, disinfecting and
sterilizing
5.3
6.
7.
8.
9.
10.
11.
Reassembling the disposable
collection canister
Functional check
6.1
Intervals
6.2
Performing the functional check
Troubleshooting
7.1
Power pack
7.2
Changing fuses
7.3
Changing the muffler
Maintenance
8.1
Intervals
8.2
Disposal
Product and accessories
9.1
Standard product
9.2
Accessories
9.3
Spare parts
Technical Data
10.1
Safe distances
Warranty
. . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . . . .
. . . . . . . . .
. . . . . . . . . . . . . . .
. . . . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . .
. . . . . . .
. . . . . . . . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . .
. . . . . . . . . . . . . . . .
. . . . . . . . . . .
19
19
20
22
23
23
25
26
27
29
31
32
32
32
33
33
34
36
37
38
39
-1,0
-0,8
-0,6
-0,4
-0,2
0
ACCUVAC Basic
from front with disposable collection canister
5 Membrane keyboard
6 Release catch
7
Disposable
canister
2 Vacuum control
4 On/Off
switch
1 Pressure gage
3 Voltage indicator
9 Tube holder plate
ACCUVAC Basic from rear without collection canister
14 Suction port
13 Muffler
(hidden)
12 Power socket
(hidden)
11 Adapter/charger
plug
8 Motor unit
10 Loops for
accessory bag
ACCUVAC Basic interior
Disposable collection canister
15 Fuse F116 Fuse F2 (hidden)
18 Plug X117 Power pack
Accessories
26 Re-usable collection canister WM 15269
27 Vent tab
28 Filter cover
29 Bacteria filter
30 Locking tab
31 Bracing
32 Secretion cover
33 Ball (overfill guard)
34 Sealing ring
35 Fingertip
25 Holder
Accessory bag
WM 10655
19 Aspiration tube
with fingertip
20 Intermediate tube
21 Disposable bag
22 Vacuum
tube
23 T-piece
24 Collection canister
11 Adapter/charger plug
Wall bracket
WM 15208
36 Aspiration tube
37 Collection canister
38 Holder set WM 15271
13,8 V
Mains / charger unit
WM 2645
Connection cable
WM 10650
1. Description
1.1 Purpose
ACCUVAC Basic is a mobile and portable electrically
powered aspirator (suction pump).
It is used for:
•aspirating accumulations of blood, secretions and
food from the oral cavity, the nose and throat
region and the bronchial system;
•deflating vacuum matresses and inflatable splints.
ACCUVAC Basic:
•can when used by a skilled operator eliminate
obstruction of the respiratory tract and hence the
risk of respiratory failure;
•cuts energy consumption by reducing power
output on reaching the necessary vacuum;
•can optionally be powered by a rechargeable
internal power pack;
•or by an external DC source supplying 12.0 -
13.8 V;
•is also suitable for use in wards.
ACCUVAC Basic must not be used:
2Description
•in medical rooms where potential equalization is
necessary (e.g. heart surgery);
•in explosion-risk areas.
1.2 Function
-1,0
-0,8
-0,6
-0,4
-0,2
0
An electrically operated diaphragm pump generates
the vacuum necessary for aspiration.
Use the vacuum control 2 to set the desired vacuum,
which is continuously variable between –0.05 bar
and –0.8 bar. You can read off the set value on the
pressure gage 1.
The aspirated material passes via the aspiration tube
19 into the disposable collection canister 7.
Disposable collection canister
The disposable collection canister 7 consists of the collection canister 24 itself, the disposable collection bag
21, the aspiration tube with fingertip 19, the interme-
diate tube 20, the T-piece 23 and the vacuum tube 22.
The disposable collection canister 7 is inserted in the
holder 25 screwed on the side of the motor unit 8.
The vacuum tube 22 of the disposable canister 24 is
pushed onto the suction port 14 of the motor unit 8.
The aspirated material passes through the aspiration
tube 19 into the disposable collection bag 21. The disposable collection bag is intended for once-only use.
When it is full, you can remove it from the collection
canister 24 and dispose of it complete with contents.
An overflow valve filter is integrated in the disposable
collection bag 21. This prevents secretions and liquid
from finding their way into the motor unit 8 and passing
into the environment via the muffler 13.
Description3
Power supply
Power for operating the appliance and charging the
power pack can be drawn:
•from the built-in power pack 17.
•from a 12-volt electrical cable system, using the
connecting cable WM 10650 which is available
as an accessory.
•from the mains and charger unit WM 2645
which is available as an accessory.
ImportantThe AC adapter/charger plug 11 supplied with the
appliance is only designed for charging the power
pack with the appliance switched off. It must not be
used to operate the appliance itself, as this will overload it and cause it to overheat.
The voltage indicator 3 shows the charge status of the
power pack.
Charging of the power pack starts automatically as
soon as the appliance is switched off and connected
to an external power supply (see “10. Technical Data”
on page 37).
4Description
2. Safety
2.1 Special symbols on the appliance
Rating plate ACCUVAC Basic
Do not dispose of the unit in domestic
waste!
Protection class BF
Protection class II, safety insulation
Year of manufacture
2.2 Safety rules
For your own safety and the safety of your patients,
and in accordance with the requirements of the
93/42/EEC, we would draw your attention to the
following points:
•Please read these operating instructions carefully.
They belong to the appliance and must be kept
available at all times.
•Before working with ACCUVAC Basic, you must
understand how to use it.
•Be sure to observe these operating instructions
when working with ACCUVAC Basic.
Safety5
•To avoid infection or contamination with bacteria
please read section “5. Hygienic preparation” on
page 19.
•ACCUVAC Basic may only be used by trained
staff instructed in aspiration techniques. Incorrect
use can cause serious bodily harm.
•ACCUVAC Basic must be used for the designated
purpose only (see “1.1 Purpose” on page 2).
•Maintenance operations such as inspection and
repair work must be performed by expert personnel or by the manufacturer Weinmann.
•Malfunctions and a lack of biocompatibility may
result if third-party articles are used. Please bear in
mind that in these cases any warranty entitlement
and liability shall lapse where the accessories recommended in the instructions for use or original
spare parts are not utilised.
•During the aspiration process, take particular care
not to cause injuries to the patient’s mouth or
throat, e.g. to mucous membrane.
The suction can be interrupted by briefly opening
the fingertip, for example if the nozzle becomes
firmly attached to the skin.
•Ensure sterile conditions during bronchial aspiration, and always use sterile catheters.
6Safety
Important!•The disposal of liquids such as blood and secre-
tions and any items contaminated with such liquids
should be carried out according to the national
guidelines for hygiene.
•The ACCUVAC Basic should only be operated in
the upright position, in order to prevent liquid from
wetting the overflow valve filter integrated in the
disposable collection bag 21. If this happens, the
overflow valve filter becomes impermeable to air
and the disposable collection bag 21 has to be
changed.
•The AC adapter/charger plug supplied is only for
recharging the power pack when the appliance is
switched off. It must not be used for operating the
appliance itself, as this will overload it and cause
it to overheat.
•To extend its service life, the power pack 17 must
not be allowed to become fully discharged. If the
red charge control LED 3 lights up, it is time to recharge immediately.
•To eliminate the risk of fully discharging the power
pack, ACCUVAC Basic must never be stored in
an uncharged condition. Always recharge the
power pack first as described in “4.4 Charging
ACCUVAC Basic” on page 16.
•Lead cell power packs discharge even during nonuse. The rate of auto-discharge increases with rising temperature and amounts to about 20 % after
three months’ non-use at 20 °C. For this reason be
sure to carry out functional checks at the specified
intervals (see “6. Functional check” on page 23).
•Never sterilize the motor unit 8 in an autoclave.
•Never dip the motor unit 8 in disinfectant solution.
•We recommend keeping alternative means of
aspiration ready in case of appliance failure.
•Do not use mobile telefones right beside
ACCUVAC Basic.
ACCUVAC Basic can however be operated
without any problem in the back of an ambulance
even if a mobile telephone is being used in the
cab.
Safety7
3. Assembly
ACCUVAC Basic is supplied ready for use.
ImportantBefore using ACCUVAC Basic for the first time, fully
charge the power pack (see “4.4 Charging
ACCUVAC Basic” on page 16).
3.1 Assembly with wall bracket
To mount ACCUVAC Basic on a wall, use wall
bracket WM 15208, available as an accessory.
1
15 mm
1
10
mm
˘ 3 mm
1. Find a suitable mounting location on an even, ver-
tical surface. The outline of the ACCUVAC on the
drilling template shows the space required.
2. Hold up the drilling template and level it with a
spirit level.
3. Mark the required screw holes on the mounting
location.
4. Drill the screw holes with a Ø 3 mm drill.
5. Screw the wall bracket on with the screws supplied.
6. Push the appliance plug11of the connection cable or of the mains/charger unit into the guide rail
of the wall bracket until it engages with the
tongue.
Important: the mounting substrate must be suffi-
ciently strong to hold the wall bracket and the
ACCUVAC securely in accordance with
CEN 1789 Medical Vehicles and their Equipment - Road Ambulances.
8Assembly
As an alternative to the AC adapter/charger plug
supplied with the appliance, you can also use the
connecting cable WM 10650 available as an
accessory.
7. Insert the AC adapter/charger plug in a 230 V/
50 Hz power socket, or if using the connecting
cable, insert the vehicle plug into a 12 volt DC
power source.
The vehicle plug is supplied complete with a red
adapter ring. This is needed if you want to plug
the vehicle plug into a cigarette lighter socket. To
connect to the mains/charger unit WM 2645
(accessory) or a 12 volt vehicle socket, remove the
red adapter ring.
8. Slide ACCUVAC Basic downwards into the wall
bracket.
If ACCUVAC Basic appliance is switched off, it
will automatically be charged up from the DC
power source.
9. To remove the ACCUVAC Basic from the bracket
ready for use, press the release catch 6 and lift
ACCUVAC Basic out of the wall bracket.
Assembly9
3.2 Fitting a re-usable collection canister
Instead of the disposable collection canister 7 supplied
as standard, you can also fit a sterilizable re-usable
collection canister 26 with holder 38, available as an
accessory. This conversion kit is available under order
number WM 15261.
1. Remove the disposable collection canister 7 from
its holder 25.
2. Unscrew the holder 25 from the motor unit 8.
3. In the same place, screw on the holder 38 for the
re-usable collection canister 26.
10Assembly
4. Push the re-usable collection canister 26 into the
holder 38. Since the collection canister is directly
connected to the suction port 14 of the motor unit
8, there is no need for an intermediate tube.
3.3 Fitting an accessory bag
The accessory bag WM 10655 is available as an
optional extra. It is used to hold aspiration catheters
and other small parts. The accessory bag cannot be
fitted at the same time as a rinsing glass (see “3.4 Fitting a rinsing glass” on page 11).
1. Use the velcro-type strips to attach the accessory
bag to the loops on the motor unit.
3.4 Fitting a rinsing glass
On the left side of the appliance it is possible to fit an
additional collection canister in the form of a rinsing
glass for holding a rinsing liquid, e.g. water.
The rinsing glass kit WM 15229 consists of:
•Collection canister 37 WM 10631
•Holder set 38 WM 15271
1. Use a pointed object to pierce the two closed drill
holes on the left of the appliance case.
Assembly11
2. Fix the holder 38 to the left side of the appliance.
3. Push the rinsing glass into the holder.
12Assembly
4. Operation
-1,0
-0,8
-0,6
-0,4
-0,2
0
4.1 Preparing for aspiration
1. Unwind the aspiration tube 19 from the tube holder plate 9.
Fingertip
Catheter
2. If necessary, adapt an aspiration catheter of
suitable size for tracheal or nasopharyngeal aspiration to fit the nozzle.
Aspirator tube
Nozzle
3. Press the button marked O/I. The lower green LED
for the power indicator and the green LED for the
O/I button lights up.
Do not switch on ACCUVAC Basic while there is
still a vacuum. This would cause the pump motor
to draw too much current when switching on,
which could actuate fuse F2 16.
4. Set the desired vacuum at the vacuum control 2.
For this purpose, close the fingertip with the stopper and the suction opening with your thumb.
Wait until the pressure gage 1 shows a constant
vacuum. Now use the vacuum control 2 to set the
desired vacuum.
ACCUVAC Basic is now ready for operation, and you
can start aspiration.
Operation13
4.2 Aspiration
ImportantMake sure that ACCUVAC Basic is always upright
During the aspiration process, take care not to cause
any injury to the patient’s mouth and throat region, and
especially to mucous membrane.
•The fingertip can be opened to interrupt suction
briefly, e.g. if the nozzle is clinging to the skin.
You can leave the fingertip open all the time and
keep your thumb over it. To release the suction all
you have to do is lift your thumb.
during operation so that the overflow valve filter integrated in the disposable collection bag 21 is not wetted by secretion. If the overflow valve filter is wetted
with secretion, it becomes impermeable and the disposable collection bag 21 has to be replaced.
During aspiration, keep an eye on the fluid level in the
disposable collection bag 21 and replace it with a
new one in good time.
14Operation
ImportantBe sure to check the power pack charge level at regu-
lar intervals during aspiration. If the red LED on the
voltage indicator 3 lights up, aspiration should be continued with a second appliance. Using the appliance
beyond this point can damage the power pack by
discharging it completely, with the result that its full
capacity will no longer be available.
Disposal of disposable collection bag
1. Pull the end of the aspiration tube with fingertip 19
off the port.
2. Detach the intermediate tube 20 from the T-piece
23 of the collection canister 24.
3. Push the free end of the intermediate tube 20 onto
the connection for the aspiration tube. This closes
the disposable collection bag 21 and reduces the
risk of fluid escaping.
4. Remove the disposable collection bag 21 and dispose of it together with the aspiration tube 19.
ImportantBe sure to observe the relevant rules for disposal
(see “2. Safety” on page 5).
Operation15
4.3 After aspiration
After aspiration:
1. Switch off ACCUVAC Basic by pressing the
button marked O/I.
2. Fit a new disposable collection bag 21
(see “ Disposal of disposable collection bag” on
page 15).
3. Clean ACCUVAC Basic (see “5. Hygienic preparation” on page 19).
4.4 Charging ACCUVAC Basic
16Operation
We recommend you to charge ACCUVAC Basic as
soon as the yellow voltage indicator LED lights up. By
doing so you will ensure that it operates for an
adequate period next time it is used.
The power supply for charging must deliver 12.0 V –
13.8 V DC and at least 0.3 ampere. A full charge
takes a maximum of 14 hours.
The following sources can be used for charging:
•The AC adapter/charger plug 11 supplied.
•The 12-volt vehicle electrical system with the aid of
the accessory connecting cable WM 10650.
•The mains/charger unit WM 2645 available as
an accessory.
–+
A red adapter ring is fitted to the vehicle connector. This is needed if you want to insert the vehicle
connector in a cigarette lighter socket. To connect
to the mains/charger unit WM 2645 (accessory)
or a 12-volt vehicle socket, you must remove the
red adapter ring.
Caution: Check the vehicle plug for correct
polarity. Reversed polarity can damage
ACCUVAC Basic.
Charging without wall bracket
1. Switch off ACCUVAC Basic.
2. Place the plug 11 at the appliance end of the connecting cable in the guide slot at the rear of the appliance and push it up behind the cover plate.
3. Switch on the power supply.
4. Charging starts as soon as you have switched on
the power supply. When the top LED lights up, the
power pack is approx. 75 % charged. Please
charge the battery for another two 2 hours at least.
Charging with wall bracket
1. Switch off ACCUVAC Basic.
2. Insert ACCUVAC Basic in the wall bracket.
Operation17
3. Charging starts as soon as you switch on the current. When the top LED lights up, the power pack
is approx. 75 % charged. Please charge the battery for another two 2 hours at least.
Refresher charging
Regular refresher charging extends the service life of
the power pack. Leave ACCUVAC Basic connected
to the external power supply even after the power
pack is fully charged. This cannot overcharge the power pack provided the charging voltage is kept within
the range 12.0 – 13.8 V.
18Operation
5. Hygienic preparation
Hygienic preparation of ACCUVAC Basic and the accessories used must be carried out daily during use
and before every change of patient. Observe the instructions for the disinfectant used. We recommend
GIGASEPT FF for immersion disinfection and TERRALIN
for wipe disinfection.
Never immerse the ACCUVAC Basic motor unit in disinfectant or other liquids. Always disinfect simply by
wiping with disinfectant. Otherwise you may damage
the device and thereby endanger users and patients
(see “5.2 Cleaning, disinfecting and sterilizing” on
page 20).
Be sure to carry out a functional check after every hygienic preparation (see “6.2 Performing the functional
check” on page 25).
5.1 Preparations
1. Pull off the aspiration tube 19.
2. Remove the intermediate tube 20 from the T-piece
23 of the collection canister 24.
Hygienic preparation19
3. Push the free end of the intermediate tube 20 onto
the aspiration tube port 19. This closes the disposable collection bag 21 and thereby reduces
the risk of secretion escaping.
4. Remove the disposable collection bag 21 and dispose of it together with the aspiration tube 19.
ImportantBe sure to observe the relevant rules when dispos-
ing of the collection bag (see “2. Safety” on
page 5).
5. Remove the vacuum tube 22.
6. Lift the disposable canister 7 from its holder 25.
7. Remove the tube holder plate 9 by pushing it to the
left until it disengages and then pulling it towards
you.
5.2 Cleaning, disinfecting and sterilizing
Hygienic preparation of the ACCUVAC Basic and the
accessories used should be performed as described in
the following table.
Observe the instructions regarding use of disinfectant. For
immersion disinfection we recommend GIGASEPT FF
and TERRALIN for wipe disinfection.
20Hygienic preparation
You are recommended to wear suitable gloves (e.g.
household or disposable gloves) during disinfection
procedures.
Rinse in
PartCleaningDisinfecting
Collection canister
Vacuum tubeNot permitted
Disposable
Aspiration tube
collection canister
with fingertip
Collection canister
Secretion cover
Overfill ball
Filter cover
Aspiration tube
(1)
Filter
Re-usable collection canister
Nozzle with
fingertip
Motor unit
Tube holder plateIn warm water
Accessory bag
In warm water
with mild house-
hold cleaner
Immerse in dilute
(3)
solution
Disposable item, re-use not permitted. Use new part for every patient
In warm water
with mild house-
hold cleaner
In clear warm
(2)
water
Immerse in dilute
(3)
solution
Not permitted
Disposable item, re-use not permitted. Use new part for every patient.
Wipe with damp
cloth
with mild house-
hold cleaner
Disinfectant wipeNot permitted
Immerse in dilute
(3)
solution
washing
machine
Rinse at up to
95 °C
Rinse at up to
95 °C
(2)
Rinse at up to
40 °C
Sterilization
Hot steam
sterilization
up to 121 °C
(4)
Hot steam
sterilization
up to 134 °C
(5)
Hot steam
sterilization
up to 134 °C
(5)
Not permitted
(1) Always dry the filter before using it again. A damp filter reduces
the pump’s suction capacity.
(2) Do not use any surfactants or alcohol for cleaning, as they can
adversely affect the hydrophobic properties of the filter.
(3) After disinfection, rinse the parts thoroughly with distilled water
and let them dry.
(4) Sterilization with superheated steam at 121 °C in units
complying with EN 285, residence time 20 minutes.
(5) Sterilization with superheated steam at 134 °C in units
complying with EN 285, residence time 5 minutes.
Hygienic preparation21
5.3 Reassembling the disposable collection canister
After cleaning, disinfection or sterilization, reassemble
the parts as follows:
1. Refit tube holder plate 9 by sliding it to the right.
2. Insert the collection canister 7 in its holder 25.
3. Use the vacuum tube 22 to connect the T-piece of
the collection canister 24 to the suction port 14 on
the motor unit 8.
4. Insert a new disposable collection bag 21.
5. Push the intermediate tube 20 onto the T-piece 23.
22Hygienic preparation
6. Push a new aspiration tube 19 onto the connection
port of the disposable collection bag 21.
6. Functional check
If the functional check reveals defects or deviations
from the specified values, ACCUVAC Basic must not
be used again until the faults have been rectified.
We therefore recommend that you always keep a
stock of the following:
•Disposable collection bag 21WM 10732
•Aspiration tube with fingertip 19 WM 10733
6.1 Intervals
To ensure that a properly functioning
ACCUVAC Basic is always available, it is essential to
observe the following intervals.
Before every use
•Perform a functional check (see “6.2 Performing
the functional check” on page 25).
After every use
•Clean, disinfect and/or sterilize the unit and its
parts (see “5. Hygienic preparation” on
page 19);
•Perform a functional check (see “6.2 Performing
the functional check” on page 25).
Functional check23
Every 3 months
•Check the charge level of the power pack 17 by
switching on ACCUVAC Basic and reading the
voltage indicator 3. If the top LED of the voltage indicator does not light up, the power pack should
be recharged (see “4.4 Charging
ACCUVAC Basic” on page 16).
At least every 6 months
•Perform a functional check (see “6.2 Performing
the functional check” on page 25).
•Make a visual inspection of the muffler for clogging. If it is clogged, fit a new muffler (see “7.3
Changing the muffler” on page 31).
After all repairs
•Clean, disinfect and/or sterilize the unit and its
parts (see “5. Hygienic preparation” on
page 19);
•Perform a functional check (see “6.2 Performing
the functional check” on page 25).
24Functional check
6.2 Performing the functional check
1. Check that all tubes and the individual parts of the
disposable collection canister 7 are in perfect condition. Any damaged or worn parts must be replaced.
2. Check that all tubes are securely connected and
the disposable collection bag 21 is firmly installed.
3. Switch on ACCUVAC Basic.
4. Check the charge level on the voltage indicator 3.
Always recharge the power pack without delay if
the yellow LED on the voltage indicator lights up
(see “4.4 Charging ACCUVAC Basic” on
page 16).
5. Insert the stopper in the fingertip.
6. Use your thumb to hold the suction nozzle closed.
7. Switch on the aspirator and select the maximum
vacuum of –0.8 bar by turning vacuum control 2
fully to the right. ACCUVAC Basic must reach this
vacuum in not more than 20 seconds.
If the aspirator takes more than 20 seconds to
reach this vacuum, its suction capacity is reduced.
Check for possible faults (see “7. Troubleshooting”
on page 26).
8. Switch ACCUVAC Basic off again.
Functional check25
7. Troubleshooting
FaultCauseRemedy
Appliance does not start,
O/I indicator and voltage
indicator show ready for
operation
Appliance does not start,
O/I indicator does not
show ready for operation
Appliance does not switch
on or off
Appliance does not reach
maximum vacuum of
–0.8 bar in 20 seconds,
but voltage indicator
shows ready for operation
Not charging
Faulty pump
Fuse F1 or F2 in appliance defective Fit new fuse (7.2, page 29)
Fuse in vehicle plug defectiveFit new fuse (7.2, page 29)
Power pack fully discharged
Incorrect polarity of vehicle power
socket
Snap-in connection between circuit
board and power pack not properly
engaged
Electronic faultHave it repaired
Leak in suction side of appliance
Power pack not sufficiently chargedCharge power pack (4.4, page 16)
Fault in pump
AC adapter/charger plug defective
Inadequate external power supply
Fuse F1 or F2 defectiveFit new fuse (7.2, page 29)
Fuse in vehicle plug defectiveFit new fuse (7.2, page 29)
Snap-in connection between circuit
board and power pack not properly
engaged
Have repairs carried out by factory or
by expert personnel.
Perform several charge/discharge
cycles. If unsuccessful, fit new power
pack (7.1, page 27)
Correct polarity and if necessary
replace fuse F1 (7.2, page 29)
Make sure connection snaps in
Check that all tubes are firmly
connected and disposable bag is
securely installed
Have repairs carried out at factory or by
expert personnel.
Have repairs carried out at factory or
by expert personnel
External power supply must be
between 12.0 and 13.8 volt.
Make sure connection snaps in
26Troubleshooting
FaultCauseRemedy
Charger does not meet specifications
Power pack is not fully
charged up despite
charging for 14 hours (top
LED in voltage indicator
does not light up)
Top LED of voltage
indicator lights up after
charging, but pump does
not run when powered by
power pack
Fuse F2 frequently blows
Vehicle electrical system is not
supplying 12.0 – 13.8 V
Power pack damaged by being fully
discharged
Power pack at end of service life
Internal connection faulty
Fuse F2 defectiveFit new fuse (7.2, page 29)
Pump is often switched on under
vacuum conditions
Pump defective
7.1 Power pack
Use AC adapter/charger plug
WM 10750 or charger WM 2645
(accessory)
Check vehicle electrical system
Fit new power pack (7.1, page 27)
Check internal connection between
printed circuit board (X1) and power
pack
Do not switch on pump under vacuum
Have repair carried out at factory or
by expert personnel
ACCUVAC Basic is fitted with a high-grade lead cell
power pack.
Changing power pack
ImportantNever touch the circuit board, as this can damage the
electronic system.
1. Switch off ACCUVAC Basic.
2. Disconnect the aspirator from the external power
supply.
3. Remove collection canister 7 and any accessories.
Troubleshooting27
4. Unscrew the holder 25 for the disposable collection canister 7.
5. Open the case by unscrewing the 6 cross-head
screws. When opening the case, be careful not to
damage the silicone sealing cord.
6. Carefully disconnect the power pack connector 18
(X1) from the circuit board.
7. Unscrew the power pack holder (4 cross-head
screws).
28Troubleshooting
8. Detach both leads from the power pack 17.
9. Remove the faulty power pack.
Help protect the environment!
Don’t throw the old power pack in the garbage
can – take it to a local collection point for environment-friendly disposal.
10. Attach the two cables to the connections on the
new power pack: red to plus; black to minus.
7.2 Changing fuses
If you accidentally reverse the connections, the
electronic system will be protected by fuse 16 (F2),
which will then have to be replaced.
11. Fit the new power pack with its holder, making
sure you do not jam any cables.
12. Carefully push power pack connector X1 onto the
circuit board until it snaps into place.
13. Screw the case together again, making sure that
the silicone sealing cord is correctly inserted and
is not jammed or otherwise damaged.
14. Perform a functional check (see “6.2 Performing
the functional check” on page 25).
Changing fuse F1 or F2
ImportantNever touch the circuit board, as this can damage the
electronic system.
1. Unscrew the case of ACCUVAC Basic
(see “ Changing power pack” on page 27).
2. Remove the faulty fuse 15/16. The fuses are identified on the circuit board.
3. Insert a new fuse. Always use approved fuses (see
“10. Technical Data” on page 37).
Troubleshooting29
Fuse
4. Screw the case together again, making sure that
the silicone sealing cord is correctly inserted and
is not jammed or otherwise damaged.
5. Perform a functional check (see “6.2 Performing
the functional check” on page 25).
Changing fuse in vehicle plug
1. Use a screwdriver to open the vehicle plug.
Note:
The central contact of the plug is the plus pole. The
plus lead of the cable has either a square cross-
section or colored markings.
The outer contact of the plug is the minus pole. The
minus lead of the cable is round and black.
Caution: Check the vehicle plug for correct
polarity. Reversed polarity can damage
ACCUVAC Basic.
2. Change the faulty fuse. Use only approved fuses
(see “10. Technical Data” on page 37).
3. Screw the vehicle plug together again.
4. Perform a functional check (see “6.2 Performing
the functional check” on page 25).
30Troubleshooting
7.3 Changing the muffler
1. Use a screwdriver to unscrew the cover plate
(2 cross-head screws).
2. Remove the old muffler 13.
3. Insert a new muffler.
4. Refit the cover plate. Make sure that the pad stuck
to the cover plate is pressing against the muffler.
5. Perform a functional check (see “6.2 Performing
the functional check” on page 25).
Pad
Troubleshooting31
8. Maintenance
8.1 Intervals
The ACCUVAC Basic requires no maintenance, but
please observe the intervals specified for regular functional checks (see “6.1 Intervals” on page 23).
We recommend that you have any servicing, such as
inspections and repair work, carried out by the manufacturer – Weinmann – or by expert personnel.
8.2 Disposal
Do not dispose of the unit with domestic waste. To dispose of the unit properly, please contact a licensed,
certified electronic scrap disposal merchant. This
address is available from your Environment Officer or
from your local authority.
32Maintenance
9. Product and accessories
9.1 Standard product
1. ACCUVAC Basic with disposable canister
ACCUVAC BasicWM 10700
consisting of:
–Aspirator unit
(disposable canister)WM 10710
–AC adapter/charger plug,
completeWM 10750
2. ACCUVAC Basic with re-usable canister
ACCUVAC BasicWM 10709
consisting of:
–Aspirator unit
(re-usable canister)WM 10726
–AC adapter/charger plug,
completeWM 10750
–Bacteria filterWM 10675
Product and accessories
33
9.2 Accessories
The following accessories are not part of the standard
product and must be ordered separately:
1. Mains/charger unitWM 2645
2. Connection cableWM 10650
3. Accessory bagWM 10655
4. Wall bracket
incl. fixing kitWM 15208
5. Conversion kit to re-usable
collection canisterWM 15261
2. Holder set for
disposable collection canisterWM 15172
consisting of:
–HolderWM 10735
–Fixing elements
3. Reset button set, greenWM 15397
4. Tube holder plate, greenWM 10723
5. Lead cell power packWM 10747
6. MufflerWM 10665
7. Instructions for useWM 16138
8. Fuse for vehicle plugWM 10673
9. Fuse for external power
supply (F1) WM 2692
10. Fuse for power pack (F2)WM 2692
11. Set I, disposable collection bagsWM 15226
(10 x WM 10732)
12. Set II, disposable collection bagsWM 15227
(20 x WM 10732)
13. Set III, disposable collection bagsWM 15228
(50 x WM 10732)
14. Set I, aspiration tube with fingertip WM 15327
(10 x WM 10733)
15. Set II, aspiration tube with fingertip WM 15328
(20 x WM 10733)
16. Set III, aspiration tube with fingertip WM 15329
(50 x WM 10733)
36Product and accessories
10. Technical Data
ACCUVAC Basic
Appliance class
MPG
Dimensions
WxHxD in mm
Weight approx. 5.4 kg
Canister volume900 ml
Suction capacity at
12 V with free flow
Max. vacuum at
12 V
Aspiration tube
Overflow valve filter
Motor output50 W
Rated voltage12 V
Maximum current
consumption
Temperature range
– Operation
– Charging
– Storage
Electromagnetic
compatibility:
– Radio interference
suppression
– Radio interference
resistance
Classification according to EN 60601-1:
– Protection against electric shock: class II;
with charger: class I
– Degree of protection against electric shock: BF
– Degree of protection against water: IPX 1
(drip-water)
Classification according to EN ISO 10079-1:
– high vacuum/high flow
II b
385x280x140
> 20 l/min
0.8 bar (80 kPa)
diam. 5 mm,
length 1800 mm
at least 99.8 % for
particle size 2.8 µm
3.5 A
–18 °C to +40 °C
+ 0 °C to +40 °C
–40 °C to +70 °C
EN 60601-1-2
EN 55011
IEC 1000-4 Parts 2–5 & 11
Norms complied
with
Vehicle plug fuse
Internal fuse F1
external power
supply
Internal fuse F2
power pack
Power pack typeLead cell, 3.4 Ah
Charging voltage12.0 to 13.8 V
Operating time
after charging for
14 hours
Operating modeS2 60 min
Service life of
power pack
AC adapter/
charger
Disposable collection canister:
– Collection canister: PC
– Disposable collection bags, tubes: PVC
– Nozzle with fingertip: PE
CasePC/ABS
Tube holder plateABS
From serial no. 10735/2004
ACCUVAC Basic
EN 60601-1,
EN ISO 10079-1,
EN 1789
8 A, DIN 72581,
identification color white
4 A slow-acting,
low breaking capacity
G fuse links 5x20 mm,
conforms to IEC 127
4 A slow-acting,
low breaking capacity
G fuse links 5x20 mm,
conforms to IEC 127
30 min at maximum suction
400 charge/discharge
cycles in approx. 3 years
primary: 230 V/50 Hz
secondary: 13.8 V/
0.3 A
Materials
Technical Data37
10.1 Safe distances
The ACCUVAC Basic is intended for operation in an
electromagnetic environment in which high-frequency
interference variables are controlled. The customer or
user of the ACCUVAC Basic can help avoid electromagnetic interference by maintaining the minimum
safe distance between portable and mobile high-frequency telecommunication devices (transmitters) and
the ACCUVAC Basic depending on the rated output
of the transmitter as given below.
Recommended safe distances between portable or mobile HF telecommunication
devices (e.g. mobile phones) and the ACCUVAC Basic
Rated output of
HF device
in W
0.010.120.120.23
0.10.370.370.74
11.171.172.33
103.693.697.38
10011.6711.6723.33
150 kHz - 80 MHz
d=(3,5/V1) x √√√√P
Safe distance depending on transmission frequency
in m
80 MHz - 800 MHz
d=(3,5/V1) x √√√√P
800 MHz – 2,5 GHz
d=(3,5/V1) x √√√√P
38Technical Data
11. Warranty
•Weinmann offers a warranty that the product, when
used in accordance with requirements, will remain free
from defects for a period of two years from date of
purchase. For products whose durability is clearly indicated as less than two years, the warranty expires on the
expiration date indicated on the packaging or in the
user’s manual.
•Claims against the warranty can be made only when
accompanied by the sales receipt, which must show
salesperson and date of purchase.
•We offer no warranty in the case of:
– Disregard of usage instructions
– Operating errors
– Improper use or handling
– Third-party intervention by non-authorized persons for
the purpose of device repair
– Acts of God, e.g., lightning strikes, etc.
– Transport damage as a result of improper packaging
of returned items
– Lack of maintenance
– Operational and normal wear and tear, which inclu-
des, for example, the following components.
– Filter
– Batteries / recheargable batteries
– Articles for one-time usage, etc.
– failure to use original spare parts.
•Weinmann is not liable for consequential harm caused
by a defect if it is not based on intention or gross negligence. Weinmann is also not liable for minor physical
injury to life or limb resulting from negligence.
•Weinmann reserves the right to decide whether to eliminate defects, to deliver a defect-free item or to reduce
the purchase price by a reasonable amount.
•If Weinmann rejects a claim against the warranty, it assumes no expense for transport between customer and
manufacturer.
•Implied warranty claims remain unaffected by these
changes.