Waygate Technologies USM Go+ Operating Manual

Page 1
USM Go USM Go+
Technical Reference and Operating Manual Id. No. 49 155
Page 2
This issue 6 (02/2013) applies to the following software versions:
USM Go: 2.08 (March 6, 2013)
USM Go+: 2.08 (May 28, 2013)
You will find the software version and the serial number of your instrument on the second operating level (CONFIG2 - ABOUT)
Page 3

First operating level (Base)

To change between the first and the second operating level:
USM Go: Press the joystick for 2 seconds.
USM Go+: Press the center key of the keypad for 2 seconds.
USM Go Issue 6 (02/2013) 0-3
Page 4

First operating level (Options)

To change between the first and the second operating level:
USM Go: Press the joystick for 2 seconds.
USM Go+: Press the center key of the keypad for 2 seconds.
0-4 Issue 6 (02/2013) USM Go
Page 5

Second operating level

USM Go Issue 6 (02/2013) 0-5
Page 6

Second operating level (continued)

0-6 Issue 6 (02/2013) USM Go
Page 7

Status display icons

Icon Meaning Icon Meaning
SD memory card is inserted, flashes when the SD card is accessed
Angle-beam probe 30° … 90°, flat surface, Reflection from the backwall
Freeze active (Freeze), Display is „frozen“.
Angle-beam probe 30°, curved surface, Reflection from the inner surface of tube
Magnify gate is active Angle-beam probe 80°, curved surface,
Reflection from the outer surface of tube
Pulser-receiver separation is turned off Angle-beam probe 90°, surface wave
Pulser-receiver separation is turned on DAC mode = TCG is active
Pulser-receiver separation is turned on and set
DGS reference echo has been recorded
to through-transmission mode
Reject function is active DGS reference echo has been recorded,
transfer loss > 0
AGT is active dB REF is active
Calibration reminder
USM Go Issue 6 (02/2013) 0-7
Page 8

Power level indicators Keypad functions

3
4
2
2
3
4
5
5
USM Go USM Go+
Icon Meaning
Battery charge level, remaining operating time in hours (approximate value)
Charger/power adapter is connected, percentage of battery charge level (approximate value)
Warning: Low battery charge level, remaining operating time in minutes (approximate value)
1
1
1 Increasing the gain level in increments
2 Decreasing the gain level in increments
3 Navigation operating levels and function groups
4 Function key 1, individually assignable
5 Function key 2, individually assignable
0-8 Issue 6 (02/2013) USM Go
Page 9

Navigation using joystick (USM Go) or keypad (USM Go+)

USM Go USM Go+ Function
Navigation between function groups, adjusting values
Navigation within a function group, adjusting values
Changing between operating levels (press for 2 seconds)
USM Go Issue 6 (02/2013) 0-9
Page 10
0-10 Issue 6 (02/2013) USM Go
Page 11
Contents
0 Übersichten
First operating level (Base). . . . . . . . . . . . 0-3
First operating level (Options). . . . . . . . . . 0-4
Second operating level . . . . . . . . . . . . . . . 0-5
Second operating level (continued) . . . . . 0-6
Status display icons . . . . . . . . . . . . . . . . . 0-7
Power level indicators. . . . . . . . . . . . . . . . 0-8
Keypad functions . . . . . . . . . . . . . . . . . . . 0-8
Navigation using joystick (USM Go)
or keypad (USM Go+). . . . . . . . . . . . . . . . 0-9

1 Introduction

1.1 Safety information . . . . . . . . . . . . . . . . . 1-2
Battery operation . . . . . . . . . . . . . . . . . . . 1-2
Software . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Defects/errors and exceptional stresses. . 1-3
1.2 Important information on
ultrasonic testing . . . . . . . . . . . . . . . . . . 1-3
Prerequisites for testing with
ultrasonic test equipment . . . . . . . . . . . . . 1-3
Operator training . . . . . . . . . . . . . . . . . . . . 1-4
Technical test requirements . . . . . . . . . . . 1-4
Limits of testing . . . . . . . . . . . . . . . . . . . . . 1-5
Ultrasonic wall thickness measurement . . 1-5
Effect of the test object material . . . . . . . . 1-5
Effect of temperature variations . . . . . . . . 1-6
Measurement of remaining wall thickness 1-6
Ultrasonic evaluation of flaws . . . . . . . . . . 1-6
Flaw boundary method . . . . . . . . . . . . . . . 1-6
Echo display comparison method . . . . . . . 1-7
1.3 The USM Go. . . . . . . . . . . . . . . . . . . . . . . 1-8
Instrument versions USM Go and USM Go+1-9
Options . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
Special features of the USM Go . . . . . . . 1-11
1.4 The DMS Go. . . . . . . . . . . . . . . . . . . . . . 1-12
1.5 How to use this manual . . . . . . . . . . . . 1-12
USM Go Issue 6 (02/2013) 0-1
Page 12
Contents
1.6 Layout and presentation in
this manual . . . . . . . . . . . . . . . . . . . . . . 1-13
Attention and note symbols. . . . . . . . . . . 1-13
Listings . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13
Operating steps. . . . . . . . . . . . . . . . . . . . 1-13

2 Standard package and accessories

2.1 Standard package . . . . . . . . . . . . . . . . . . 2-2
2.2 Add-on functions . . . . . . . . . . . . . . . . . . 2-3
2.3 Preconfigured function packages. . . . . 2-4
2.4 Recommended accessories. . . . . . . . . . 2-5

3 Initial start-up

3.1 Instrument positioning . . . . . . . . . . . . . . 3-2
3.2 Power supply. . . . . . . . . . . . . . . . . . . . . . 3-2
Operation with charger/power adapter . . . 3-2
Operation using batteries . . . . . . . . . . . . . 3-4
Charging the batteries. . . . . . . . . . . . . . . . 3-8
3.3 Connecting a probe . . . . . . . . . . . . . . . . 3-9
3.4 Inserting the SD memory card. . . . . . . 3-10
3.5 Starting the USM Go . . . . . . . . . . . . . . . 3-11
Powering On . . . . . . . . . . . . . . . . . . . . . . 3-11
Powering Off . . . . . . . . . . . . . . . . . . . . . . 3-12
Factory default setting (Reset) . . . . . . . . 3-12

4 Principles of operation

4.1 Overview of operator's controls . . . . . . 4-2
4.2 Display screen. . . . . . . . . . . . . . . . . . . . . 4-3
A-scan representation . . . . . . . . . . . . . . . . 4-3
Functions on the display screen . . . . . . . . 4-4
Gain. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Measurement line . . . . . . . . . . . . . . . . . . . 4-5
Status display icons . . . . . . . . . . . . . . . . . 4-6
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
0-2 Issue 6 (02/2013) USM Go
Page 13
Contents
4.3 Navigation and function keys . . . . . . . . 4-7
Navigation. . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Function keys . . . . . . . . . . . . . . . . . . . . . . 4-7
Key combinations . . . . . . . . . . . . . . . . . . . 4-8
Power key . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
4.4 Operational concept . . . . . . . . . . . . . . . . 4-9
Operating levels . . . . . . . . . . . . . . . . . . . . 4-9
Selecting and setting functions . . . . . . . . . 4-9
Function HOME . . . . . . . . . . . . . . . . . . . 4-11
Activating functions. . . . . . . . . . . . . . . . . 4-12
4.5 Important default settings . . . . . . . . . . 4-13
Language setting . . . . . . . . . . . . . . . . . . 4-13
Units setting . . . . . . . . . . . . . . . . . . . . . . 4-14
Decimal separator. . . . . . . . . . . . . . . . . . 4-14
Date format, Date, and Time . . . . . . . . . 4-15
Selecting the instrument orientation . . . . 4-16
4.6 Default settings of the display. . . . . . . 4-16
Selecting the color scheme. . . . . . . . . . . 4-16
Selecting the A-scan color . . . . . . . . . . . 4-17
Selecting the grid . . . . . . . . . . . . . . . . . . 4-18
Setting the brightness . . . . . . . . . . . . . . . 4-18
4.7 Saving the settings. . . . . . . . . . . . . . . . 4-19
Recalling settings . . . . . . . . . . . . . . . . . . 4-21
Displaying dataset name. . . . . . . . . . . . . 4-22

5 Operation

5.1 Overview of the functions . . . . . . . . . . . 5-2
Function groups first operating level . . . . . 5-3
Function groups second operating level . . 5-4
5.2 Setting the gain. . . . . . . . . . . . . . . . . . . . 5-6
Setting the dB increment of gain. . . . . . . . 5-6
5.3 Assignment of function keys. . . . . . . . . 5-8
5.4 Setting the display range
(function group RANGE) . . . . . . . . . . . 5-10
RANGE . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
PROBE DELAY. . . . . . . . . . . . . . . . . . . . 5-11
VELOCITY . . . . . . . . . . . . . . . . . . . . . . . 5-12
DISPLAY DELAY . . . . . . . . . . . . . . . . . . 5-13
USM Go Issue 6 (02/2013) 0-3
Page 14
Contents
5.5 Setting the pulser
(function group PULSER). . . . . . . . . . . 5-14
VOLTAGE (pulser voltage) . . . . . . . . . . . 5-14
ENERGY. . . . . . . . . . . . . . . . . . . . . . . . . 5-15
WIDTH . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16
DAMPING . . . . . . . . . . . . . . . . . . . . . . . . 5-17
PRF MODE (pulse repetition frequency) . .5-17
5.6 Setting the receiver
(function group RECEIVER). . . . . . . . . 5-19
FREQUENCY . . . . . . . . . . . . . . . . . . . . . 5-19
RECTIFY. . . . . . . . . . . . . . . . . . . . . . . . . 5-20
DUAL (pulser-receiver separation) . . . . . 5-20
REJECT . . . . . . . . . . . . . . . . . . . . . . . . . 5-21
5.7 Setting the gates (function groups
GATE A and GATE B) . . . . . . . . . . . . . . 5-22
Tasks of the gates. . . . . . . . . . . . . . . . . . 5-22
A-START/B-START
(starting point of the gate) . . . . . . . . . . . . 5-23
A-WIDTH/B-WIDTH
(width of the gates) . . . . . . . . . . . . . . . . . 5-23
A-THRESHOLD/B-THRESHOLD (response and measurement
threshold of the gate) . . . . . . . . . . . . . . . 5-24
TOF MODE . . . . . . . . . . . . . . . . . . . . . . . 5-25
Starting point of gate B . . . . . . . . . . . . . . 5-27
Automatic gate height . . . . . . . . . . . . . . . 5-28
5.8 Calibrating the USM Go . . . . . . . . . . . . 5-29
Calibrating the display range. . . . . . . . . . 5-29
Choice of the measuring point . . . . . . . . 5-29
Calibration with straight-beam and
angle-beam probes . . . . . . . . . . . . . . . . . 5-30
Calibration using dual-element probes . . 5-34
5.9 Making measurements . . . . . . . . . . . . . 5-37
General notes . . . . . . . . . . . . . . . . . . . . . 5-37
5.10 dB-difference measurement
(function group dB REF) . . . . . . . . . . . 5-38
Recording a reference echo . . . . . . . . . . 5-39
Deleting a reference echo . . . . . . . . . . . . 5-39
Echo height comparison . . . . . . . . . . . . . 5-40
0-4 Issue 6 (02/2013) USM Go
Page 15
Contents
5.11 Rating of welds
(function group AWS D1.1) . . . . . . . . . 5-41
Rating of welds according to AWS D1.1. 5-42
5.12 Flaw position calculation with
angle-beam probes. . . . . . . . . . . . . . . . 5-45
PROBE ANGLE . . . . . . . . . . . . . . . . . . . 5-46
THICKNESS . . . . . . . . . . . . . . . . . . . . . . 5-47
X VALUE. . . . . . . . . . . . . . . . . . . . . . . . . 5-48
O-DIAMETER . . . . . . . . . . . . . . . . . . . . . 5-48
COLOR LEG. . . . . . . . . . . . . . . . . . . . . . 5-49
5.13 Defining the probe angle . . . . . . . . . . . 5-50
BLOCK . . . . . . . . . . . . . . . . . . . . . . . . . . 5-51
5.14 Enabling options (Upgrade). . . . . . . . . 5-52
5.15 Configuring the USM Go for
test tasks . . . . . . . . . . . . . . . . . . . . . . . . 5-53
TOF MODE. . . . . . . . . . . . . . . . . . . . . . . 5-53
Phantom echo detector. . . . . . . . . . . . . . 5-58
Configuring the measurement line . . . . . 5-59
Enlarged display of reading . . . . . . . . . . 5-62
LARGE (alarm signal). . . . . . . . . . . . . . . 5-64
MAGNIFY GATE (spanning the gate). . . 5-65
Activating the magnify gate function . . . . 5-66
Automatic A-scan freeze (Freeze) . . . . . 5-68
5.16 Setting the display . . . . . . . . . . . . . . . . 5-70
ASCAN FILL . . . . . . . . . . . . . . . . . . . . . . 5-71
Working with Echo Max . . . . . . . . . . . . . 5-72
5.17 General setup . . . . . . . . . . . . . . . . . . . . 5-73
EVAL MODE . . . . . . . . . . . . . . . . . . . . . . 5-73
Gate logic . . . . . . . . . . . . . . . . . . . . . . . . 5-74
Selecting the pulser type. . . . . . . . . . . . . 5-75
Locking the joystick. . . . . . . . . . . . . . . . . 5-76
Configuring the alarm output. . . . . . . . . . 5-77
Power saving mode . . . . . . . . . . . . . . . . 5-79
TOF in LAYER . . . . . . . . . . . . . . . . . . . . 5-80
Backwall echo attenuation (BEA) . . . . . . 5-82
Displaying the envelope curve
(ENVELOPE) . . . . . . . . . . . . . . . . . . . . . 5-83
Automatic gain control
(Auto Gain Control) . . . . . . . . . . . . . . . . . 5-84
USM Go Issue 6 (02/2013) 0-5
Page 16
Contents
Calibration reminder . . . . . . . . . . . . . . . . 5-86
Password protection . . . . . . . . . . . . . . . . 5-87
5.18 Distance-amplitude correction (DAC) . 5-91
Recording a DAC curve . . . . . . . . . . . . . 5-92
Setting up the DAC . . . . . . . . . . . . . . . . . 5-94
Turning the DAC evaluation off. . . . . . . . 5-95
Deleting the DAC curve . . . . . . . . . . . . . 5-96
Editing DAC points . . . . . . . . . . . . . . . . . 5-96
Adding DAC points . . . . . . . . . . . . . . . . . 5-97
Multiple DAC curves . . . . . . . . . . . . . . . . 5-97
AWS D1.1 in DAC/TCG . . . . . . . . . . . . . 5-99
Sensitivity correction . . . . . . . . . . . . . . . . 5-99
Echo evaluation using DAC/TCG . . . . . 5-100
5.19 Distance-amplitude curve according to
JIS Z3060-2002 (JISDAC) . . . . . . . . . . 5-102
Activating JISDAC
(DAC according to JIS) . . . . . . . . . . . . 5-102
Recording a DAC curve . . . . . . . . . . . . 5-102
Setting up JISDAC . . . . . . . . . . . . . . . . 5-105
Sensitivity correction . . . . . . . . . . . . . . . 5-106
Turning the JISDAC evaluation off . . . . 5-106
Deleting the DAC curve. . . . . . . . . . . . . 5-107
Echo evaluation using DAC . . . . . . . . . 5-107
5.20 Distance-amplitude curve according to JB/T4730 and
GB 11345 (CNDAC) . . . . . . . . . . . . . . . 5-109
Evaluations according to CNDAC . . . . . 5-109
Standards and reference blocks . . . . . . 5-110
Activating the CNDAC. . . . . . . . . . . . . . 5-111
Recording a DAC curve . . . . . . . . . . . . 5-111
Setting up CNDAC . . . . . . . . . . . . . . . . 5-114
Sensitivity correction . . . . . . . . . . . . . . . 5-115
Adjusting reference lines. . . . . . . . . . . . 5-116
Turning the CNDAC evaluation off . . . . 5-116
Deleting the DAC curve. . . . . . . . . . . . . 5-117
Echo evaluation using DAC . . . . . . . . . 5-117
5.21 Evaluation according to the
DGS method . . . . . . . . . . . . . . . . . . . . 5-119
Using the DGS for measurements . . . . 5-119
Validity of the DGS method. . . . . . . . . . 5-121
0-6 Issue 6 (02/2013) USM Go
Page 17
Contents
Starting the echo height evaluation
according to DGS . . . . . . . . . . . . . . . . . 5-123
Basic settings for the
DGS measurement . . . . . . . . . . . . . . . . 5-123
Recording a reference echo and
turning the DGS curve on . . . . . . . . . . . 5-125
Locks, error messages . . . . . . . . . . . . . 5-127
Sound attenuation and
transfer correction. . . . . . . . . . . . . . . . . 5-128
Using multiple DGS curves. . . . . . . . . . 5-128
Turning the DGS evaluation off . . . . . . 5-129
Deleting a DGS reference echo . . . . . . 5-129
Probe data . . . . . . . . . . . . . . . . . . . . . . 5-130
trueDGS angle-beam probes . . . . . . . . 5-133

6 Documentation

6.1 Test reports. . . . . . . . . . . . . . . . . . . . . . . 6-2
Storing test reports . . . . . . . . . . . . . . . . . . 6-2
Displaying test reports . . . . . . . . . . . . . . . 6-4
Printing test reports. . . . . . . . . . . . . . . . . . 6-6
Deleting test reports . . . . . . . . . . . . . . . . . 6-7
Storing the A-scan and parameters
in the test report . . . . . . . . . . . . . . . . . . . . 6-9
6.2 Storing memos . . . . . . . . . . . . . . . . . . . 6-10
Creating a new memo file . . . . . . . . . . . . 6-10
Editing a memo file . . . . . . . . . . . . . . . . . 6-11
Attaching a memo file to test report . . . . 6-12
6.3 Storing a report header . . . . . . . . . . . . 6-13
Creating a new header file . . . . . . . . . . . 6-13
Editing a header file . . . . . . . . . . . . . . . . 6-14
Including a header file in the test report . 6-15
6.4 Viewing and storing parameters . . . . . 6-16
6.5 Videos . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
Recording a video . . . . . . . . . . . . . . . . . . 6-17
Viewing a video. . . . . . . . . . . . . . . . . . . . 6-19
6.6 Documentation using UltraMATE . . . . 6-21
6.7 Data Recorder (option). . . . . . . . . . . . . 6-22
Creating a Data Recorder file . . . . . . . . . 6-23
Activating a Data Recorder file . . . . . . . . 6-25
Storing readings in the grid matrix . . . . . 6-26
USM Go Issue 6 (02/2013) 0-7
Page 18
Deleting readings . . . . . . . . . . . . . . . . . . 6-27
A-scan preview . . . . . . . . . . . . . . . . . . . . 6-28
Viewing Data Recorder files . . . . . . . . . . 6-28
Turning the grid matrix on/off . . . . . . . . . 6-29
Service interface (Mini RS232-C) . . . . . . . 8-3
8.2 Peripherals. . . . . . . . . . . . . . . . . . . . . . . . 8-4

9 Appendix

7 Maintenance and care

7.1 Instrument care. . . . . . . . . . . . . . . . . . . . 7-2
7.2 Battery care . . . . . . . . . . . . . . . . . . . . . . . 7-2
Battery care. . . . . . . . . . . . . . . . . . . . . . . . 7-2
Charging the batteries. . . . . . . . . . . . . . . . 7-3
7.3 Maintenance . . . . . . . . . . . . . . . . . . . . . . 7-3
7.4 Software updates . . . . . . . . . . . . . . . . . . 7-4
Download of update files. . . . . . . . . . . . . . 7-4
Installing an update. . . . . . . . . . . . . . . . . . 7-5

8 Interfaces and Peripherals

8.1 Interfaces. . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
USB interface . . . . . . . . . . . . . . . . . . . . . . 8-3
0-8 Issue 6 (02/2013) USM Go
9.1 Function directory. . . . . . . . . . . . . . . . . . 9-2
9.2 EU Declaration of Conformity . . . . . . . 9-11
9.3 Manufacturer/Service addresses. . . . . 9-11
9.4 Environmental protection regulations 9-13
WEEE directive (Waste Electrical and
Electronic Equipment) . . . . . . . . . . . . . . . 9-13
Disposal of batteries . . . . . . . . . . . . . . . . 9-14
9.5 Recycling directives . . . . . . . . . . . . . . . 9-16
Overview . . . . . . . . . . . . . . . . . . . . . . . . . 9-16
Materials to be disposed of separately . . 9-19
Other materials and components . . . . . . 9-22
Recycling data of the USM Go . . . . . . . . 9-29
Page 19

10 Specifications

10.1 Specifications of
USM Go and USM Go+ . . . . . . . . . . . . . 10-2
Display screen . . . . . . . . . . . . . . . . . . . . 10-2
Display . . . . . . . . . . . . . . . . . . . . . . . . . . 10-3
Connectors . . . . . . . . . . . . . . . . . . . . . . . 10-3
Memory. . . . . . . . . . . . . . . . . . . . . . . . . . 10-4
Pulser . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-4
Receiver . . . . . . . . . . . . . . . . . . . . . . . . . 10-5
Gates . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-6
Memory. . . . . . . . . . . . . . . . . . . . . . . . . . 10-6
Environment . . . . . . . . . . . . . . . . . . . . . . 10-7
Protection . . . . . . . . . . . . . . . . . . . . . . . . 10-8
Options . . . . . . . . . . . . . . . . . . . . . . . . . . 10-9
10.2 Specifications according to
EN 12668 . . . . . . . . . . . . . . . . . . . . . . . 10-11
Contents

11 Index

USM Go Issue 6 (02/2013) 0-9
Page 20
0-10 Issue 6 (02/2013) USM Go
Page 21

Introduction 1

USM Go Issue 6 (02/2013) 1-1
Page 22
1 Introduction Safety information
ATTENTION

1.1 Safety information

The USM Go has been designed and tested according to DIN
EN 61010-1: 2011-07, Safety requirements for electrical equipment for measurement, control and lab oratory use, and was technically in perfectly safe and faultless condition when leaving the manufacturing works.
In order to maintain this condition and to ensure a safe operation, you should always read the following safety information carefully before putting the instrument into operation.
The USM Go is an instrument for materi­als testing. Any use for medical or any other applications is not permitted!
The instrument may only be used in industri­al environments.
The USM Go is waterproof according to IP67. It can be operated either with the corresponding lithium-ion bat teries or with the charger/power adapter. The charger/ power adapter meets the requirements of electrical safety class II.
-

Battery operation

For the battery operation of the USM Go, we recom­mend the corresponding lithium-ion battery. You should only use this battery for the battery operation.
­You can charge the lithium-ion battery either within the instrument itself or in an external charger. If a lithium-ion battery is inserted, charging starts automatically as soon as you connect the charger/power adapter to the USM Go and to the mains power supply.
For power supply, please also see Chapter 3.2 Power supply, page 3-2. For the use of batteries, please also see Chapter 7.2 Battery care, page 7-2.

Software

According to the current state of the art, software is nev­er completely free from errors. Before using any soft­ware-controlled test equipment, it is therefore neces­sary to make sure that the required functions operate perfectly in the intended combination.
If you have any questions about the use of your test equipment, please contact your nearest representative of GE Sensing & Inspection Technologies.
1-2 Issue 6 (02/2013) USM Go
Page 23
Important information on ultrasonic testing 1 Introduction

Defects/errors and exceptional stresses

If you have reason to believe that a safe operation of your USM Go is no longer possible, you have to discon nect the instrument and secure it against unintentional re-connection. Remove the lithium-ion battery.
A safe operation is no longer possible for example
if the instrument shows visible damages,
if the instrument no longer operates perfectly,
after prolonged storage under adverse conditions
(e.g. exceptional temperatures or especially high air humidity, or corrosive environmental conditions),
after being subjected to heavy stresses during trans­portation.

1.2 Important information on ultrasonic testing

­Please read the following information before using your
USM Go. It is important that you understand and ob serve this information to avoid any operator errors that might lead to false test results. Such false test results could result in personal injuries or property damages.

Prerequisites for testing with ultrasonic test equipment

This operating manual contains essential information on how to operate your test equipment. In addition, there are a number of factors that affect the test results, but a description of all these factors goes beyond the scope of this operating manual. The three most important pre requisites for a safe and reliable ultrasonic inspection are:
Operator training
Knowledge of special technical test requirements and
limits
-
-
Choice of appropriate test equipment
USM Go Issue 6 (02/2013) 1-3
Page 24
1 Introduction Important information on ultrasonic testing

Operator training

The operation of an ultrasonic test device requires prop­er training in ultrasonic test methods.
Proper training comprises for example adequate knowl­edge of:
the theory of sound propagation,
the effects of sound velocity in the test material,
the behavior of the sound wave at interfaces between
different materials,
the propagation of the sound beam,
the influence of sound attenuation in the test object
and the influence of surface quality of the test object.
Lack of such knowledge could lead to false test results with unforeseeable consequences. You can contact for example NDT societies or organizations in your country (DGZfP in Germany; ASNT in the USA), or also GE Sensing & Inspection Technologies, for information on the existing opportunities for training of ultrasonic in spectors as well as on the qualifications and certificates that can finally be obtained.
-

Technical test requirements

Every ultrasonic test is subject to specific technical test requirements. The most important ones are:
the definition of the scope of inspection
the choice of the appropriate test method
the consideration of material properties
the determination of limits for recording and evalua-
tion.
It is the task of the those with overall responsibility for testing to ensure that the inspector is fully informed about these requirements. The best basis for such infor mation is experience with identical test objects. It is also essential that the relevant test specifications be clearly and completely understood by the inspector.
GE Sensing & Inspection Technologies regularly holds specialized training courses in the field of ultrasonic testing. The scheduled dates for these courses will be given to you on request.
-
1-4 Issue 6 (02/2013) USM Go
Page 25
Important information on ultrasonic testing 1 Introduction

Limits of testing

The information obtained from ultrasonic tests only re­fers to those parts of the test object which are covered by the sound beam of the probe used.
Any conclusions from the tested parts to be applied to the untested parts of the test object should be made with extreme caution.
Such conclusions are generally only possible in cases where extensive experience and proven methods of sta tistical data acquisition are available.
The sound beam can be completely reflected from boundary surfaces within the test object so that flaws and reflection points lying deeper remain undetected. It is therefore important to make sure that all areas to be tested in the test object are covered by the sound beam.

Ultrasonic wall thickness measurement

All ultrasonic wall thickness measurements are based on a time-of-flight measurement. Accurate measure ment results require a constant sound velocity in the test object. In test objects made of steel, even with varying alloying constituents, this condition is mostly fulfilled. The variation of sound velocity is so slight that it is only
-
of importance for high-precision measurements. In oth er materials, e.g. nonferrous metals or plastics, the sound velocity variations may be even larger and thus affect the measuring accuracy.

Effect of the test object material

If the material of the test object is not homogeneous, the sound waves may propagate at different velocities in dif ferent parts of the test object. An average sound velocity
­should then be taken into account for the range calibra
tion. This is achieved by using a reference block with a sound velocity equal to the average sound velocity of the test object.
If substantial sound velocity variations are expected, then the instrument calibration should be adjusted to the actual sound velocity values at shorter time intervals. Failure to do so may lead to false thickness readings.
-
-
-
USM Go Issue 6 (02/2013) 1-5
Page 26
1 Introduction Important information on ultrasonic testing

Effect of temperature variations

The sound velocity within the test object also varies as a function of the material's temperature. This can cause appreciable errors in measurements if the instrument has been calibrated on a cold reference block, whereas the measurement is carried out on a warm test object. Such measurement errors can be avoided either by ad justing the temperature of the reference block used for calibration or by taking the temperature effect into con sideration on the basis of a correction factor obtained from published tables.

Measurement of remaining wall thickness

The measurement of the remaining wall thickness on plant components, e.g. pipes, tanks, and reaction ves sels of all types which are corroded or eroded from the inside, requires a perfectly suitable gauge and special care in handling the probe.
The inspectors should always be informed about the corresponding nominal wall thicknesses and the likely amount of wall thickness losses.
-

Ultrasonic evaluation of flaws

In present-day test practice, there are basically two dif­ferent methods of flaw evaluation:
If the diameter of the sound beam is smaller than the ex­tent of the flaw, then the sound beam can be used to ex­plore the boundaries of the flaw and thus determine its
­area.
-
If, however, the diameter of the sound beam is larger that the extent of the flaw, the maximum echo indication from the flaw must be compared with the maximum echo indication from an artificial flaw provided for com parison purposes.

Flaw boundary method

The smaller the diameter of the probe's sound beam, the more accurately the boundaries, i.e. the actual flaw area, can be determined by the flaw boundary method. If, however, the sound beam is relatively broad, the flaw area determined can substantially differ from the actual flaw area. Care should therefore be taken to select a probe which will give a sufficiently narrow sound beam at the position of the flaw.
-
1-6 Issue 6 (02/2013) USM Go
Page 27
Important information on ultrasonic testing 1 Introduction

Echo display comparison method

The echo from a small, natural flaw is usually smaller than the echo from an artificial comparison flaw, e.g. cir cular disc flaw of the same size. This is due, for in­stance, to the roughness of the surface of a natural flaw, or to the fact that the sound beam does not impinge on it at right angles.
If this fact is not taken into account when evaluating nat­ural flaws, there is a risk of false evaluation.
In the case of very jagged or fissured flaws, e.g. shrink holes in castings, it may be that the sound scattering oc curring at the boundary surface of the flaw is so strong that no echo at all is produced. In such cases, a different evaluation method should be chosen, e.g. use of the backwall echo attenuation in the evaluation.
The distance sensitivity of the flaw echo plays an impor­tant part when testing large components. Attention should be paid here to choosing artificial comparison flaws which are as far as possible governed by the same "distance laws" as the natural flaws to be evaluated.
The ultrasonic wave is attenuated in any material. This sound attenuation is very low, e.g. in parts made of fine­grained steel, likewise in many small parts made of oth
-
er materials. However, if the sound wave travels larger distances through the material, a high cumulative sound attenuation can result, even with small attenuation coef ficients. There is then a danger that echoes from natural flaws appear too small. For this reason, an estimate must always be made of the effects of attenuation on the evaluation result and taken into account if applicable.
If the test object has a rough surface, part of the incident sound energy will be scattered at its surface and is not
­available for the test. The larger this initial scattering, the
smaller the flaw echoes appear, and the more errors oc cur in the evaluation result.
It is therefore important to take the effect of the test ob­ject's surfaces on the height of the echo into account (transfer correction).
-
-
-
USM Go Issue 6 (02/2013) 1-7
Page 28
1 Introduction The USM Go

1.3 The USM Go

The USM Go is a lightweight and compact ultrasonic flaw detector which is especially suitable
for locating and evaluating material flaws,
for measuring wall thicknesses,
for saving and documenting test results.
Due to its design, the USM Go can be used in almost all flaw detection applications in a wide range of industries, including aerospace, power generation, automotive, as well as oil and gas. These include:
Weld inspection
Trigonometric projections
AWS
DAC
DGS
Inspection of forgings and castings
Manual PRF adjustment
Rail inspection
High PRF (up to 2000 Hz)
Lightweight: 850 g (1.87 lb)
Small and ergonomic
Inspection of composites
RF display
2 gates A and B
Gate B is triggered by the event in gate A
For even more demanding applications
Narrow band-pass filters
Low-noise digital amplifiers
Optional square pulsers
DAC (TCG) with 120 dB/µs slope
Backwall echo attenuation (BEA)
Phantom echo detector
DGS
1-8 Issue 6 (02/2013) USM Go
Page 29
The USM Go 1 Introduction

Instrument versions USM Go and USM Go+

A joystick is used in the USM Go for navigation, for changing settings, and for selecting adjustment values.
USM Go
These functions are carried out by means of five keys in the keypad of the USM
Go+. The arrow keys in the key-
pad correspond to the movement of the joystick in the corresponding direction, and pressing the center key corresponds to pressing the joystick.
USM Go+
USM Go Issue 6 (02/2013) 1-9
Page 30
1 Introduction The USM Go

Options

Various options extend the basic functions of the USM Go and can be enabled by a code in each case.
USM Go Base
Basic version, for universal ultrasonic test jobs.
USM Go AWS
Amplitude evaluation according to AWS D1.1 for the weld inspection.
USM Go DAC
DAC amplitude evaluation using up to 16 points ac­cording to EN 1712, EN 1713, EN 1714, ASME, and ASME
III, in conformity with JIS Z3060
DAC (TCG) 110 dB dynamic
DAC (BEA) 120 dB/µs slope
USM Go DGS
DGS amplitude evaluation according to EN 1712
USM Go with an on-board data logger
Recording and documentation of measurement val­ues in linear and grid file mode.
USM Go with square pulser
Enables the fine adjustment of initial pulse parame­ters
Voltage setting 120 … 300 V in increments of 10 V
Pulse duration setting 30 … 500 ns in increments of
10
ns
PRF
3G (gate C)
1-10 Issue 6 (02/2013) USM Go
Page 31
The USM Go 1 Introduction

Special features of the USM Go

small, lightweight, rugged
dust-tight and waterproof housing according to IP67
long operating time (six hours) due to lithium-ion bat-
tery with internal and external charging possibility
one-handed operation is possible
keys for direct adjustment of gain
two independent gates for accurate wall thickness
measurements from the material surface up to the first echo or between two backwall echoes, including a measurement on coated materials with a resolution of 0.01 mm (up to 100 mm), referred to steel
Magnify gate: spreading of the gate range over the entire display range
high-resolution color display (800 × 480 Pixel) for the representation of digitized signals
color display of gates and the corresponding readings for easy distinction
easily perceptible reflection geometry when using an­gle-beam probes due to changing A-scan or back­ground color at every reflection point
high memory capacity with 2 GB SD card. SD cards with a memory capacity up to 16
GB can be used.
increased calibration range: up to 9999 mm (steel), depending on the frequency range
semiautomatic two-point calibration
pulse repetition frequency adjustable in 3 steps
(AUTO LOW, MED, HIGH) or MANUALLY in steps of 5
Hz.
optional connection of phantom-echo detector
choice of the frequency range for the connected
probe
Signal display: full-wave rectification, positive or neg­ative half rectification, and radio frequency
display of seven user-selectable measured readings at the top of the A-scan, one of which is displayed in enlarged mode, or four user-selectable measured readings displayed in enlarged mode
USM Go Issue 6 (02/2013) 1-11
Page 32
1 Introduction The DMS Go

1.4 The DMS Go

The USM Go uses the same operating principle as the portable thickness gauge DMS Go.
You can extend your USM Go by a DMS Go quickly and easily by means of a software upgrade. You will then have two instruments available in one housing. When you power the instrument on, you can choose the instru ment that you want to use (see Chapter 3.5 Starting the USM Go, page 3-11).
A separate operating manual is available for the DMS Go. The functions of the DMS Go are therefore not de scribed in the operating manual of the USM Go.

1.5 How to use this manual

This operating manual applies to all instrument versions of the USM Go. Any differences in the functions or ad justment values are marked in each case.
Before operating the instrument for the first time, it is ab­solutely necessary that you read the chapters 1, 3, and
4. They will inform you about the necessary prepara
­tions of the instrument, give you a description of all keys and displays, and explain the operating principle.
In doing this, you will avoid any errors or failures of the instrument and be able to use the full range of instru
­ment functions.
You will find the specifications of the instrument in Chapter 10 Specifications.
-
-
-
1-12 Issue 6 (02/2013) USM Go
Page 33
Layout and presentation in this manual 1 Introduction
ATTENTION
Note

1.6 Layout and presentation in this manual

To make it easier for you to use this manual, all operat­ing steps, listings, and special notes are always pre­sented in the same way. This will help you find individual pieces of information quickly.

Attention and note symbols

The ATTENTION symbol indicates peculiar­ities and special aspects in the operation which could affect the accuracy of the re sults.
Note contains e.g. references to other chap­ters or special recommendations for a func­tion.
-

Listings

Listings are presented in the following form:
Variant A
Variant B
● ...

Operating steps

Operating steps appear as shown in the following exam­ple:
– Loosen the two screws at the bottom.
– Remove the cover.
–…
USM Go Issue 6 (02/2013) 1-13
Page 34
1-14 Issue 6 (02/2013) USM Go
Page 35

Standard package and accessories 2

USM Go Issue 6 (02/2013) 2-1
Page 36
2 Standard package and accessories Standard package

2.1 Standard package

Product code Description Order number
Ultrasonic flaw detector USM Go or USM Go+
TC-096 Transport case
LI-138 Lithium-ion battery, 7.4 V, 3.9 Ah, rechargeable
LiBC-139 AC power adapter/charger, 100V … 260V AC
SD memory card 2 GB
Display screen protector foils (10 pieces)
WS-342 Safety hand strap
Brief operating manual
Operating manual on CD
Manufacturer's certificate
2-2 Issue 6 (02/2013) USM Go
Page 37
Add-on functions 2 Standard package and accessories

2.2 Add-on functions

Product code Description Order number
DAC/TCG Echo evaluation method DAC, JISDAC, CNDAC, TCG
DGS Echo evaluation method DGS
AWS Echo evaluation method AWS D1.1
SWP Square-wave pulser
PPRF Phantom echo detector
3Gate Third gate C
DL Wall thickness data logger
USM Go Issue 6 (02/2013) 2-3
Page 38
2 Standard package and accessories Preconfigured function packages

2.3 Preconfigured function packages

Product code Description Order number
Basic Ultrasonic flaw detector USM Go or USM Go+
DAC Basic with DAC/TCG, AWS, SWP
DGS Basic with DGS, AWS, SWP
Advanced Basic with DAC, DGS, AWS, SWP, PPRF
2-4 Issue 6 (02/2013) USM Go
Page 39
Recommended accessories 2 Standard package and accessories

2.4 Recommended accessories

Product code Description Order number
LI-138 Lithium-ion battery, 7.4 V, 3.9 Ah, rechargeable
LiBC-139 AC power adapter/charger, 100V … 260V AC
CA-040 Battery adapter for external charging of battery
TC-096 Transport case
CH-097 Shoulder strap
WH-098 Shoulder bag for instrument and couplant
WS-342 Safety hand strap
EK-492 Ergonomic set (CH-097, WH-098, WS-342)
CBL-604 Probe cable: Lemo 00-90° - Microdot
CBL-819 Probe cable: Lemo 00-90° - Lemo 00
USM Go Issue 6 (02/2013) 2-5
Page 40
2 Standard package and accessories Recommended accessories
Product code Description Order number
CBL-820 Probe cable: Lemo 00-90° - Lemo 01
CBL-821 Probe cable: Lemo 00-90° - KBA 533
CBL-822 Probe cable: Lemo 00-90° - BNC
EN-499 Certificate EN 12668-1
2-6 Issue 6 (02/2013) USM Go
Page 41

Initial start-up 3

USM Go Issue 6 (02/2013) 3-1
Page 42
3 Initial start-up Instrument positioning

3.1 Instrument positioning

Fold out the prop-up stand on the rear side of the USM Go and position the instrument on a flat base so that you can easily read the display.
If the instrument has been brought from a cold room into a warmer one, wait until it has adapted to the room tem perature before you power it on (to avoid condensation).
If (in rare cases) condensation has developed inside the instrument, the cover may mist up from the inside. In this case, open the cover until the damp has dried up. You should not power the instrument on until this has hap pened.
-

3.2 Power supply

The USM Go can be operated either with an external charger/power adapter or with the corresponding lithi um-ion battery.
You can also connect the USM Go to the mains power supply if the battery is in the instrument. A discharged
­battery is charged in this case, during the instrument op eration.

Operation with charger/power adapter

Connection to power supply
For the operation using a charger/power adapter, you should only use the charger/power adapter included in the standard package.
The charger/power adapter is automatically adjusted to every AC voltage between 90 voltage).
V and 240 V (nominal
-
-
3-2 Issue 6 (02/2013) USM Go
Page 43
Power supply 3 Initial start-up
ATTENTION
1
2
Connecting the instrument
Connect the USM Go to the mains socket-outlet by means of the corresponding charger/power adapter. The socket-contact for connecting the charger/power adapter is located on the side of the USM Go.
– Align the Lemo plug of the charger/power adapter
with the red mark on the socket (1).
– Push the plug into the socket until it locks into place
with a clearly audible click.
– When removing the Lemo plug, pull the metal sleeve
on the plug back first in order to open the lock.
In order to power the instrument off correctly, press the power On/Off key (2) on the side of the instrument. If the power supply is inter
-
rupted (removing the battery, disconnecting the power plug), the operation does not end correctly.
USM Go Issue 6 (02/2013) 3-3
Page 44
3 Initial start-up Power supply
21

Operation using batteries

You should only use the corresponding lithium-ion bat­tery for the battery operation.
Inserting batteries
The battery compartment is located on the rear of the in­strument. The cover is fastened with two attachment screws.
– Turn the two attachment screws (1) of the battery
compartment counterclockwise by one quarter of a turn each in order to loosen them.
– Lift the cover off upward. In the open battery compart-
ment, you will see several connector pins (2) on one side.
3-4 Issue 6 (02/2013) USM Go
Page 45
Power supply 3 Initial start-up
1 2
– Place the battery in the battery compartment so that
the marking faces upwards and the contacts are pushed against the connector pins (1).
– Insert the cover of the battery compartment with the
side opposite to the screws at first, and push the lugs
(3) into the housing recesses.
– Press the cover firmly down on the side of the screws
and turn the two screws
(2) clockwise by one quarter
of a turn each in order to lock the cover.
3
USM Go Issue 6 (02/2013) 3-5
Page 46
3 Initial start-up Power supply
1 2
Checking the charge level of the lithium-ion battery
The lithium-ion battery is provided with a battery charge level indicator. Five light-emitting diodes
(1) indicate the level of battery charge. Check the battery charge level before inserting it into the instrument.
The number of diodes that are lit up has the following meaning:
5 LEDs: Battery charge level 100 … 80 %
4 LEDs: Battery charge level 80 … 60 %
3 LEDs: Battery charge level 60 … 40 %
2 LEDs: Battery charge level 40 … 20 %
1 LED: Battery charge level 20 … 10 %
1 LED is flashing: Battery charge level <10 %
– Press the key (2) next to the LEDs. The LEDs indi-
cate the battery charge level.
3-6 Issue 6 (02/2013) USM Go
Page 47
Power supply 3 Initial start-up
Note
Power level indicator
The USM Go is equipped with a power level indicator that allows to estimate the remaining operating time of the instrument. A battery icon with the corresponding charge level is displayed in the top right corner on top of the A-scan. The charge level is indicated in percent by means of the battery icon.
Icon Meaning
Battery charge level, remaining operating time in hours (approximate value)
Charger/power adapter is connected, percentage of battery charge level (approximate value)
Warning: Low battery charge level, remaining operating time in minutes (approximate value)
The USM Go is automatically powered off if the opera­tion is no longer ensured. All settings are retained during battery exchange and are immediately available again afterwards.
If the battery charge level is low, it is abso­lutely necessary that you finish your test job, power off the instrument, and replace the battery. You should carry a second battery along with you if you cannot use mains pow er supply to operate the instrument.
-
USM Go Issue 6 (02/2013) 3-7
Page 48
3 Initial start-up Power supply

Charging the batteries

You can charge the lithium-ion batteries either directly within the instrument or in an external charger.
Internal charging
If a lithium-ion battery is inserted, charging starts auto­matically as soon as you connect the charger/power adapter to the USM Go and to the mains power supply. You can carry out ultrasonic tests and charge the batter ies at the same time.
The charging time is approx. ten hours with simultane­ous ultrasonic testing. If the instrument is not used for ul­trasonic testing, the charging time is approx. eight hours. This charging time applies to ambient tempera tures of 25 … 30 °C.
-
Charging status
The LED on the charger/power adapter indicates the status of charging.
off: Charger/power adapter is not con-
nected to the power supply
yellow steady light: Charger/power adapter is not con-
nected to the instrument or no bat­teries are inserted into the instru-
-
ment
flashing green light: Charging
green steady light: Charging is completed,
batteries are charged
External charging
You can charge lithium-ion batteries with an external charger of the USM Go. Do not use any other chargers for charging the lithium-ion batteries for the USM Go.
3-8 Issue 6 (02/2013) USM Go
Page 49
Connecting a probe 3 Initial start-up
ATTENTION

3.3 Connecting a probe

To prepare the USM Go for operation, you have to con­nect a probe to it. Any GEIT probe can be used for the USM
Go, provided the appropriate cable is available and the operating frequency is within an adequate range.
If a probe is connected incorrectly, the con­sequence would be a mismatching which may lead to considerable power losses or even to echo waveform distortions.
The probe is connected to the sockets on the side of the instrument casing.
Both connector sockets are equally suitable (connected in parallel) for connecting probes equipped with only one ultrasonic element (ultrasonic transducer) so that it does not matter which one of the two sockets is used.
When connecting a dual-element (TR) probe (having one transmitter or pulser element and one receiver ele ment) or two probes (of which one is transmitting and the other one receiving), attention should be paid to the correct allocation of connecting cables (please see sym bols on the instrument):
Icon Meaning
Pulser connection
Receiver connection
-
-
USM Go Issue 6 (02/2013) 3-9
Page 50
3 Initial start-up Inserting the SD memory card
2

3.4 Inserting the SD memory card

You can use any SD memory card in the USM Go. To insert and to remove the memory card, you have to open the watertight cover at the top of the instrument.
– Push the lock of the hinged cover (1) in the direction
of the arrow in order to open the cover.
– Insert the SD memory card into the card slot so that
the contacts (2) of the card face the instrument front side.
– Press the card down into the card slot until it locks
into place.
– Close the cover and make sure that it is locked tightly.
If necessary, push the lock up to the limit stop in the opposite direction of the arrow in order to close the cover watertight again.
– To remove the SD card, open the cover and shortly
press down the card to unlock it.
1
3-10 Issue 6 (02/2013) USM Go
Page 51
Starting the USM Go 3 Initial start-up
1

3.5 Starting the USM Go

Powering On

To start the USM Go, shortly press the Power key (1) on the side of the instrument casing.
The software is initialized. During this, the display will re­main blank for about 3 seconds. If the license for the DMS Go is also installed, the display will show the page for selecting the required instrument. Select the re quired instrument by means of the joystick (USM Go) or the arrow keys of the keypad (USM
After that, the start display showing the name of the in­strument and information on the software, serial num­ber, and the installed options will appear.
The instrument carries out a self-check and then switch­es over to stand-by mode.
The settings of all function values and the default set­tings (language and units) are the same as before pow­ering the instrument off.
Go+).
-
USM Go Issue 6 (02/2013) 3-11
Page 52
3 Initial start-up Starting the USM Go

Powering Off

To power the USM Go off, shortly press the Power key
(1) on the side of the instrument casing.
The settings of all function values and the default set­tings (language and units) are retained after powering off.

Factory default setting (Reset)

If you can no longer use the functions of your instrument or if the instrument no longer reacts as expected, you can reset it to the factory default settings. Any data saved to the SD card will be retained, all other individual settings, e.g. language and units, will be reset to the fac tory default settings.
– Power the instrument off.
– Press the further end of the Gain key (1), the further
end of the Function key multaneously, and hold all three keys pressed until the start display or the page of the instrument selec tion appears.
(2), and the Power key (3) si-
The instrument starts with the factory default settings (for language selection, page 4-13).
see Section Language setting,
1
-
-
2
3
3-12 Issue 6 (02/2013) USM Go
Page 53

Principles of operation 4

USM Go Issue 6 (02/2013) 4-1
Page 54
4 Principles of operation Overview of operator's controls
1
3
4
1
2
2
3
4
5
6 7
5
USM Go USM Go+

4.1 Overview of operator's controls

1 Increasing the gain level in increments
2 Decreasing the gain level in increments
3 Operating levels and function groups navigation
4 Function key 1, individually assignable
4-2 Issue 6 (02/2013) USM Go
5 Function key 2, individually assignable
6 Display for representation of A-scan and functions
7 Power key for powering on and off
Page 55
Display screen 4 Principles of operation

4.2 Display screen

A-scan representation

The USM Go has a high-resolution display screen for the display of the A-scan.
A-scan display in the normal mode
A-scan display in the zoom mode
The gain and the adjusted dB step value are shown in the top left corner of the display screen. The access to all other instrument functions is disabled in the zoom mode of the A-scan display.
Toggling the A-scan display mode
To toggle between the normal and the zoomed A-scan display mode, shortly press the joystick (USM the center key of the keypad (USM
Go+) once.
Go) or
USM Go Issue 6 (02/2013) 4-3
Page 56
4 Principles of operation Display screen

Functions on the display screen

Function groups
The names of the seven function groups are shown at the bottom of the display screen. The currently selected function group is highlighted.
First operating level (A-scan):
Second operating level (settings):
Functions
On the first operating level, the functions of the currently selected function group are shown on the left of the dis play screen, next to the A-scan.
The functions are hidden in the zoomed A-scan display mode; no operation is possible in this case.
-
4-4 Issue 6 (02/2013) USM Go
Page 57
Display screen 4 Principles of operation
Note

Gain

The current gain value and the adjusted dB step value are always displayed in the top left corner of the screen.

Measurement line

The measurement line on top of the A-scan shows sev­en different measurement readings. One reading can be displayed in enlarged mode in a box on the far right. The readings for the individual boxes are user selectable (
see Section Configuring the measurement line,
page 5-59).
One or four measurement readings can be displayed in enlarged mode. The number of the other boxes of the measurement line is then reduced ( larged display of reading, page 5-62).
see Section En-
In addition to the measurement reading, the measuring point (peak or flank) is displayed with a symbol in sound path measurements:
^ = measuring point Peak
/ = measuring point Flank
Examples:
SA^ = sound path within gate A, at the measuring point Peak
SA/ = sound path within gate A, at the measuring point Flank
The measuring point for amplitude is marked for the corresponding gate bar at the top edge of the display by a triangle pointing up wards in the color of the gate, the measuring point for distance is marked by a triangle pointing downwards.
-
USM Go Issue 6 (02/2013) 4-5
Page 58
4 Principles of operation Display screen

Status display icons

To the left of the A-scan, below the measurement line, there is an area for various status displays. The status display icons inform about active functions and certain settings ( the beginning of the present operating manual).
see Section Status display icons, page 0-7 at

Alarms

You can display an alarm signal in the form of a virtual LED in the far right box on top of the A-scan ( tion LARGE (alarm signal), page 5-64).
When an alarm is triggered, the color of the alarm signal changes from green to red.
see Sec-
4-6 Issue 6 (02/2013) USM Go
Page 59
Navigation and function keys 4 Principles of operation
Note

4.3 Navigation and function keys

Navigation

A joystick is used in the USM Go for navigation, for changing settings, and for selecting adjustment values.
These functions are carried out by means of five keys in the keypad of the USM pad correspond to the movement of the joystick in the corresponding direction, and pressing the center key corresponds to pressing the joystick.
The navigation is used
for toggling between the operating levels,
for selecting function groups,
for selecting and setting functions,
for selecting the zoom mode of A-scan display
Go+. The arrow keys in the key-

Function keys

Two key groups consisting of two function keys each are arranged next to the display screen.
The two upper function keys are used for changing the gain and turning the function AUTO Key combinations, page 4-8).
The two lower function keys are used
for changing values or settings,
for selecting options,
for triggering user-assigned functions (see Chapter
5.3 Assignment of function keys, page 5-8).
You can change many values either by means of the navigation (small increments) or by means of the lower function keys (large increments).
80 on (see Section
USM Go Issue 6 (02/2013) 4-7
Page 60
4 Principles of operation Navigation and function keys

Key combinations

You can carry out some functions by means of key com­binations. To achieve this, you have to press several keys at the same time (
ator's controls, page 4-2).
Function Keys
HOME
Function key 1 + function key 2
see Section Overview of oper-
AUTO 80
UPDATE
RESET
Increase gain + decrease gain
Decrease gain + function key 2 + Power key
Increase gain + function key 2 + Power key

Power key

The key for powering the instrument on and off is locat­ed on the instrument side, next to the probe connectors.
4-8 Issue 6 (02/2013) USM Go
Page 61
Operational concept 4 Principles of operation

4.4 Operational concept

Operating levels

The USM Go is an easy-to-use instrument. It has two operating levels between which you can toggle by pressing the joystick (USM (USM
Go+).
The first operating level shows the A-scan and is used during normal operation.
It contains seven function groups for the settings during normal operation.
The second operating level contains all functions for the instrument configuration. This is where you will also find functions for saving and printing, for data management using the data logger, and for special applications, e.g. software updates.
Go) or the center key

Selecting and setting functions

Shown below the A-scan are the seven function groups which you can directly select using the navigation. The name of the currently selected function group is high lighted and the corresponding four functions are dis­played on the left, next to the A-scan.
You can likewise directly select the individual functions using the navigation.
If a function is selected, the name of the function group is marked red. You can then change the value using the navigation or the function keys.
-
USM Go Issue 6 (02/2013) 4-9
Page 62
4 Principles of operation Operational concept
Note
Coarse and fine adjustment of functions
You can choose between coarse and fine adjustment for some functions.
The fine adjustment is made using the navigation. The fastness of change is influenced by the movement of the joystick during this (e.g. for the function VELOCITY).
You can lock the joystick in the USM Go. Any changes to the adjustments using the joy
-
stick are then no longer possible (see Sec­tion Locking the joystick, page 5-76).
Use the lower function keys to make the coarse adjust­ment. The value will then change either in large incre­ments (e.g. for the function RANGE) or you can choose from a series of factory-saved values (e.g. for the func
-
tion VELOCITY)
During the fine adjustment, the name of the function is displayed in lower-case letters (range), whereas it is displayed in capital letters during the coarse adjustment (RANGE).
As long as a function is selected, you can only change the corresponding value but you cannot toggle between the function groups. To toggle between the function groups, you have to first select a function group again using the navigation (the name is highlighted).
4-10 Issue 6 (02/2013) USM Go
Page 63
Operational concept 4 Principles of operation
Note

Function HOME

You can or must select the function HOME in various cases (e.g. to confirm certain settings). To do this, press the function keys at the same time ( view of operator's controls, page 4-2).
You can assign the function HOME to one of the function keys (
of function keys, page 5-8).
Choosing the start value
You can use the function HOME for some functions to quickly choose the start value of the adjustment range. To do this, press the two lower function keys at the same time after selecting the function. In this way, you can e.g. set the value of the function DISPLAY DELAY to 0.000 µs.
see Section Over-
see Section Assignment
This quick adjustment option is available for the follow­ing functions:
Function Function group
PROBE DELAY RANGE
DISPLAY DELAY RANGE
REJECT RECEIVER
USM Go Issue 6 (02/2013) 4-11
Page 64
4 Principles of operation Operational concept
Note

Activating functions

You have usually two options for choosing functions, for triggering actions, or for changing settings on the sec ond operating level:
using the function keys directly after selecting a func­tion,
using the function keys or the navigation after activat­ing a function.
The direct change of settings or functions is disabled for some functions and, in any case, you must activate the function first before you can change any settings. This is the case e.g. for the function DIRECTORY for selecting a directory on the SD memory card.
-
Example
– Switch over to the second operating level.
– In the function group EVAL, select the function MAG-
NIFY GATE.
Option 1:
– Press the function keys to directly choose the gate for
the magnify gate function.
Option 2:
– Press the joystick (USM Go) or the center key of the
keypad (USM
Go+) briefly to activate the function
MAGNIFY GATE. The currently selected setting is highlighted.
– After that, change the setting using either the naviga-
tion or the function keys.
– After changing the setting, press the joystick
(USM
Go) or the center key of the keypad
(USM
Go+) briefly again to deactivate the function. You can then use the navigation again to switch over to another function.
4-12 Issue 6 (02/2013) USM Go
Page 65
Important default settings 4 Principles of operation

4.5 Important default settings

Language setting

Use the function LANGUAGE (function group CONFIG1 on the second operating level) to choose the
language for the displayed texts.
The following languages are available:
Bulgarian Chinese German
English Finnish French
Italian Japanese Dutch
Norwegian Polish Portuguese
Romanian Russian Swedish
Spanish Czech Hungarian
– Switch over to the second operating level.
– In the function group CONFIG1, select the function
LANGUAGE.
– Press the function keys to choose the required lan-
guage. The language changes immediately.
USM Go Issue 6 (02/2013) 4-13
Page 66
4 Principles of operation Important default settings

Units setting

You can use the function UNITS (function group CONFIG1 on the second operating level) to select the
required units (mm, in, or µs). You can change the units any time. All values are adjusted accordingly.
– Switch over to the second operating level.
– In the function group CONFIG1, select the function
UNITS.
Press the function keys to choose the required units.

Decimal separator

You can choose the decimal separator mark. All data are displayed and saved using the selected decimal separator.
– Switch over to the second operating level.
– In the function group CONFIG1, select the function
DECIMAL.
Press the function keys to choose the required setting.
4-14 Issue 6 (02/2013) USM Go
Page 67
Important default settings 4 Principles of operation
ATTENTION

Date format, Date, and Time

The date is saved together with the inspection results. You can set the date format, the date, and the time us ing the corresponding functions of the function group CONFIG1 on the second operating level.
In view of a correct documentation, always make sure that you are using the correct date and the correct time. Remember to turn the clock during the change from winter time to summer time.
– Switch over to the second operating level.
– In the function group CONFIG1, select the function
DATE FORMAT.
– Press the function keys to change the date format.
The time format is changed together with the date for mat.
– Select the function DATE.
– Press the joystick (USM Go) or the center key of the
keypad (USM
Go+) to change the setting. The first
value (day, month, or year) is highlighted.
– Use the navigation to change the highlighted value
and switch over to the next value.
– Change the other values in the same way.
-
– Finally, press the joystick (USM Go) or the center key
of the keypad (USM
Go+) to close the setting. The
new values apply immediately.
– Select the function TIME.
– Change the time in the same way as you did previ-
ously for the date.
– Finally, press the joystick (USM Go) or the center key
of the keypad (USM
Go+) to close the setting. The
new values apply immediately.
-
USM Go Issue 6 (02/2013) 4-15
Page 68
4 Principles of operation Default settings of the display

Selecting the instrument orientation

You can configure the instrument for right-handed or left-handed operation (referred to the probe). The screen display is then rotated through 180°, the func tions of navigation and function keys are adjusted ac­cordingly.
– Switch over to the second operating level.
– In the function group CONFIG1, select the function
ORIENTATION.

4.6 Default settings of the display

The USM Go is equipped with a high-resolution color display. You can optimize the display according to your own viewing habits and to the operating environment.

Selecting the color scheme

-
– Press the function keys to choose the required orien-
tation.
4-16 Issue 6 (02/2013) USM Go
Page 69
Default settings of the display 4 Principles of operation
Note
You can choose one of four color schemes using the function COLOR (function group CONFIG1 on the sec ond operating level). The color scheme determines the color of all displays and of the background. You can set the color of the A-scan separately (please see the sec tion below).
All color schemes are suitable for indoor use. For outdoor use, we recommend Scheme 3 and Scheme 4.
– Switch over to the second operating level.
– In the function group CONFIG1, select the function
COLOR.
– Press the function keys to select the required color
scheme.

Selecting the A-scan color

-
-
You can choose the color of the A-scan using the function A-SCAN COLOR (function group CONFIG1 on the sec ond operating level). The color options depend on the se­lected color scheme (see Section Selecting the color scheme, page 4-16).
– Switch over to the second operating level.
– In the function group CONFIG1, select the function
A-SCAN COLOR.
-
– Press the function keys to choose the required color
for the A-scan.
USM Go Issue 6 (02/2013) 4-17
Page 70
4 Principles of operation Default settings of the display
Note

Selecting the grid

Using the function GRID (function group CONFIG1 on the second operating level), you can choose a grid for the A-scan. Two grids, each with and without a ruler at the bottom edge of the display, are available as options.
– Switch over to the second operating level.
– In the function group CONFIG1, select the function
GRID.
– Press the function keys to choose the required grid
for the A-scan.

Setting the brightness

To set the brightness of the display, use the function BRIGHTNESS (function group CONFIG1 on the second operating level). You can choose a value from 1 to 10.
A high brightness value increases the power consumption, due to which the operating time is reduced in battery operation.
4-18 Issue 6 (02/2013) USM Go
Page 71
Saving the settings 4 Principles of operation
Note
– Switch over to the second operating level.
– In the function group CONFIG1, select the function
BRIGHTNESS.
– Press the function keys to select the required value.
You can extend the operating time by means of the power saver function ( Power saving mode, page 5-79).
see Section

4.7 Saving the settings

You can save the current instrument settings to a file on the SD memory card. The files for the USM filename extension UGO.
Go have the
USM Go Issue 6 (02/2013) 4-19
Page 72
4 Principles of operation Saving the settings
ATTENTION
– Switch over to the function ACTION and press the
The filename can be up to 14 characters long for saving data sets. However, only the first 7 characters are shown in the display on top of the A-scan (
see Section Displaying dataset
name, page 4-22).
Please note this limitation when saving data sets in order to avoid any confusions be
­tween filenames starting with the same char­acter.
function keys to select the function STORE DATA SET.
– Switch over to the function FILENAME and briefly
press the joystick (USM keypad (USM
Go+).
Go) or the center key of the
– Use the navigation to choose the option
<NEW
FILE> and briefly press the joystick (USM Go)
or the center key of the keypad (USM
Go+).
– Use the navigation to choose the first character of the
filename.
-
– Use the navigation to switch over to the next item,
and choose the next character.
– Switch over to the second operating level.
– Press the joystick (USM Go) or the center key of the
– In the function group FILES, select the function
DIRECTORY and briefly press the joystick (USM Go) or the center key of the keypad (USM
Go+). The di-
keypad (USM try.
Go+) briefly to close the filename en-
rectory of the SD memory card appears.
– Press the function keys to select a directory on the
SD card.
– Press the joystick (USM Go) or the center key of the
keypad (USM
Go+) briefly to choose the selected di-
rectory.
4-20 Issue 6 (02/2013) USM Go
Page 73
Saving the settings 4 Principles of operation
Note
Note
– Switch over to the function ENTER and briefly press
the joystick (USM (USM
Go+).
Go) or the center key of the keypad
The current instrument settings are saved to the select­ed directory on the SD memory card under the name en­tered here.
You can create and delete directories by in­serting the SD memory card into the SD card reader of your PC or by connecting the USM
Go to a PC via a USB cable (see Sec-
tion USB interface, page 8-3).

Recalling settings

You can recall and use instrument settings saved to the SD memory card.
You can only recall datasets having the file­name extension UGO. Other files on the SD memory card are not displayed as options.
– Switch over to the second operating level.
– In the function group FILES, select the function
DIRECTORY and briefly press the joystick (USM Go) or the center key of the keypad (USM rectory of the SD memory card appears.
– Press the function keys to select a directory on the
SD card.
– Press the joystick (USM Go) or the center key of the
keypad (USM
Go+) briefly to choose the selected di-
rectory.
Go+). The di-
USM Go Issue 6 (02/2013) 4-21
Page 74
4 Principles of operation Saving the settings
Note
– Switch over to the function ACTION and press the
function keys to select the function RECALL DATA SET.
– Switch over to the function FILENAME and briefly
press the joystick (USM keypad (USM
Go+).
Go) or the center key of the
– Use the navigation to select the name of the required
dataset.
– Press the joystick (USM Go) or the center key of the
keypad (USM
Go+) briefly. The selection is automat-
ically changed to the function ENTER.
– Press the joystick (USM Go) or the center key of the
keypad (USM
Go+) briefly.
The instrument settings of the selected dataset are re­called and they are immediately active after closing the recalling process.

Displaying dataset name

-
You can display the filename of the currently recalled in­strument settings (dataset name) in the measurement line on top of the A-scan.
The dataset name can only be displayed in the box on the far right.
4-22 Issue 6 (02/2013) USM Go
Page 75
Saving the settings 4 Principles of operation
ATTENTION
– Switch over to the second operating level.
– Select the function group EVAL.
– Select the function LARGE.
– Press the function keys to choose the parameter
DSET as value.
– Switch over to the first operating level.
The filename can be up to 14 characters long for saving data sets (on the second operating level). However, only the first 7 characters are shown in the display on top of the A-scan (on the first operating level).
Please note this limitation when saving data
After recalling the instrument settings, the correspond­ing filename (dataset name) is displayed in red letters in
sets in order to avoid any confusions be tween filenames starting with the same char­acter.
-
the measurement line on top of the A-scan in the box on the far right.
USM Go Issue 6 (02/2013) 4-23
Page 76
4 Principles of operation Saving the settings
4-24 Issue 6 (02/2013) USM Go
Page 77

Operation 5

USM Go Issue 6 (02/2013) 5-1
Page 78
5 Operation Overview of the functions
Note
Note

5.1 Overview of the functions

The functions of the USM Go are combined into function groups on two operating levels.
– Press the joystick (USM Go) or the center key of the
keypad (USM tween the two operating levels.
– Use the navigation to select a function or a setting.
– Press the function keys to change a setting.
The gain is always directly available to you by means of the two keys at the top, next to the display.
You will find an overview of the function groups and their corresponding functions on the first pages of this oper ating manual.
Go+) for about 2 seconds to toggle be-
You will find detailed descriptions of how to use the operator's controls in ciples of operation.
Chapter 4 Prin-
The first operating level contains seven function groups in the default view.
When certain functions are activated, other associated function groups become visible by navigating upwards. You can return to the previous view by pressing the center key.
The second operating level contains eight function groups.
First operating level
Second operating level
-
5-2 Issue 6 (02/2013) USM Go
Page 79
Overview of the functions 5 Operation

Function groups first operating level

RANGE This is where you will find functions required for the basic setting of the signal representation on the
display screen.
PULSER This group is a combination of functions used for setting the pulser.
RECEIVER This group is a combination of functions used for setting the receiver.
dB REF This group contains functions for evaluation. The name and functions of this group vary depending
on the selected evaluation method.
AUTOCAL This is where you will find the functions for the semi-automatic calibration of the instrument.
GATE A In this group, you will find all functions required for setting the gate A.
GATE B In this group, you will find all functions required for setting the gate B.
GATES Only if the option 3B is enabled: All functions for setting the gates A, B, and C.
The function groups GATE A and GATE B are then no longer visible.
USM Go Issue 6 (02/2013) 5-3
Page 80
5 Operation Overview of the functions

Function groups second operating level

FILES This is where you will find functions used for the management of files, test reports, and videos.
EVAL This group is a combination of functions for evaluation methods. In addition, you can configure the
measurement line here (
dB REF This group contains functions for the evaluation method dB Ref. The name and functions of this
group vary depending on the selected evaluation method.
DAC/TCG This group contains the functions for the evaluation method DAC/TCG (option).
DGS This group contains the functions for the evaluation method DGS (option).
AWS D1.1 This group contains the functions for the evaluation method AWS D1.1 (option).
JISDAC This group contains the functions for the evaluation method JISDAC (option).
CNDAC This group contains the functions for the evaluation method CNDAC (option).
see Section Measurement line, page 4-5).
5-4 Issue 6 (02/2013) USM Go
Page 81
Overview of the functions 5 Operation
CONFIG1 This is where you will find various default settings, e.g. language, color, and A-scan display options.
CONFIG2 This is where you will find special functions for the instrument setup for test and measurement
applications.
CONFIG3 This is where you will find additional functions for the instrument setup for test and measurement
applications.
CONFIG4 This group contains functions for the automatic gain control.
DR This group contains all functions for the data logger (option).
USM Go Issue 6 (02/2013) 5-5
Page 82
5 Operation Setting the gain

5.2 Setting the gain

To enable you to set the gain quickly and directly, this function is always available to you with the two keys at the top, next to the display screen.
You can use the gain to adjust the sensitivity necessary for making echoes from reflectors to be detected visible on the display screen at the required height.
– Press the upper key to increase the gain. The current
gain is displayed in the top left corner of the display screen.
– Press the lower key to decrease the gain.

Setting the dB increment of gain

By pressing the keys, you will always set the gain by a certain dB increment. You can define this dB increment.
5-6 Issue 6 (02/2013) USM Go
Page 83
Setting the gain 5 Operation
Note
The following settings are possible:
LOCK (locked)
0.2 dB
0.6 dB
1.0 dB
2.0 dB
6.0 dB
0.2 … 60.0 dB
The setting LOCK locks the gain thus pre­venting its adjustment by mistake.
– Switch to the second operating level.
– In the function group CONFIG2, select the function
dB STEP.
– Press the function keys to choose the required dB in-
crement.
You can define the dB increment of the seventh level us­ing the function USER GAIN STEP.
You can set the display range by coarse adjustment steps or by fine adjustment (
see Section Coarse and
fine adjustment of functions, page 4-10).
– Select the function USER GAIN STEP.
– Set the user gain step by means of the function keys
or using the navigation.
USM Go Issue 6 (02/2013) 5-7
Page 84
5 Operation Assignment of function keys

5.3 Assignment of function keys

You can assign various functions to the two function keys at the bottom, next to the display screen, so that you will be able to trigger these functions any time with out switching the operating level.
Each one of the two function keys can have two function assigned to it, one of which is triggered by briefly press ing the key, whereas the other one is triggered by press­ing the key for a longer time (approx. 2 seconds). All in all, this means that you can assign four different func tions to the two function keys.
-
The following settings are possible:
NONE (no function selected)
FREEZE (see page 5-68)
JOYSTICK LOCK (see page 5-76)
COPY (see page 6-2)
AUTO80 (see page 5-92)
MAGNIFY GATE (see page 5-65)
dB STEP (see page 5-6)
RECALL DATASET (see page 4-21)
ORIENTATION (see page 4-16)
ENVELOPE (see page 5-83)
HOME (see page 4-8)
ANGLE+
-
ANGLE-
BW GAIN+ (see page 5-82)
-
BW GAIN- (see page 5-82)
5-8 Issue 6 (02/2013) USM Go
Page 85
Assignment of function keys 5 Operation
– Switch to the second operating level.
– In the function group CONFIG2, select the function
FUNCTION
1 to assign functions to the upper func-
tion key.
– Activate the function (see Section Activating func-
tions, page 4-12). The upper option is selected and you can now choose the function for the brief key press.
– Use the navigation to choose the required function for
the brief key press.
– Press one of the function keys to choose the function
for the long key press. The selection changes to the lower option and the name of the function changes to
(HOLD)func
1.
– Press the function keys to choose the function for the
long key press.
– Finally, deactivate the function (HOLD)func 1.
– Switch to the function FUNCTION 2 to assign func-
tions to the lower function key.
USM Go Issue 6 (02/2013) 5-9
Page 86
5 Operation Setting the display range (function group RANGE)
Note

5.4 Setting the display range (function group RANGE)

For an exact adjustment of the sound veloci­ty and the probe delay, please start by read­ing Chapter 5.8 Calibrating the USM Go, page 5-29.
You need the function group RANGE for the basic set­ting of the display range. The display range must be ad­justed to the sound velocity used (function VELOCITY) and to the probe used (function PROBE DELAY).
– Switch to the first operating level.
– Select the function group RANGE.
5-10 Issue 6 (02/2013) USM Go
Page 87
Setting the display range (function group RANGE) 5 Operation
Note
Note

RANGE

You can use the function RANGE to set the range (the display range) for making measurements.
You can set the display range by coarse adjustment steps or by fine adjustment ( fine adjustment of functions, page 4-10).
The adjustment range is 0.5 … 20000.00 mm.
The adjustment range for display range de­pends on the sound velocity setting and on the frequency range setting (function FRE QUENCY in the function group RECEIVER).
– Select the function RANGE.
– Set the required display range by means of the func-
tion keys or using the navigation.
– Press the two function keys at the same time to quick-
ly adjust an average value (254.00 mm).
see Section Coarse and
-

PROBE DELAY

Every probe is equipped with a delay line between the transducer and the coupling face. The sound pulse must first pass through this delay line before it can enter the test object. You can compensate for this influence of the delay line in the probe in the function PROBE DELAY.
You can set the probe delay by coarse adjustment steps or by fine adjustment ( justment of functions, page 4-10).
The adjustment range is 0 … 1000.000 µs.
If the value for the probe delay is not known, please read USM Go, page 5-29 to determine this value.
– Select the function PROBE DELAY.
– Set the probe delay by means of the function keys or
using the navigation.
– Press the two function keys at the same time to set
the value to zero.
see Section Coarse and fine ad-
Chapter 5.8 Calibrating the
USM Go Issue 6 (02/2013) 5-11
Page 88
5 Operation Setting the display range (function group RANGE)
Note
ATTENTION

VELOCITY

Please always make sure that the setting of
If µs are selected as units, the function VE- LOCITY is deactivated for safety reasons and it does not appear on the display screen.
You can use the function VELOCITY to set the sound velocity in the test object.
You can choose between different materials and their sound velocities. The additional information TR. and TRANS refer to the sound velocities in transverse wave excitation mode. In longitudinal wave excitation mode, the material name remains without additional informa
-
tion.
If you deviate from a specified material velocity when using the navigation, the material name is automatically replaced by the information CUSTOM. You can set the sound velocity by coarse adjustment steps or by fine ad justment (see Section Coarse and fine adjustment of functions, page 4-10).
– Select the function VELOCITY.
– Set the sound velocity by means of the function keys
or using the navigation.
-
the function VELOCITY is correct. The USM Go calculates all range and distance indica tions on the basis of the value adjusted here.
-
The adjustment range is 250 … 16000 m/s.
5-12 Issue 6 (02/2013) USM Go
Page 89
Setting the display range (function group RANGE) 5 Operation

DISPLAY DELAY

You can use this function to choose whether to display the adjusted display range (for example 250 mm) start ing from the surface of the test object or in a section of the test object starting at a later point. This allows you to shift the complete screen display and consequently also the display zero.
For example, if the display should start from the surface of the test object, you have to adjust the value 0 for DIS PLAY DELAY.
You can set the display delay by coarse adjustment steps or by fine adjustment (
see Section Coarse and
fine adjustment of functions, page 4-10).
The adjustment range is -15.000 … 3500.000 µs.
– Select the function DISPLAY DELAY.
– Set the sound velocity by means of the function keys
or using the navigation.
– Press the two function keys at the same time to set
the value to zero.
-
-
USM Go Issue 6 (02/2013) 5-13
Page 90
5 Operation Setting the pulser (function group PULSER)

5.5 Setting the pulser (function group PULSER)

You will find all functions for setting the pulser in the function group PULSER.
– Switch to the first operating level.
– Select the function group PULSER.

VOLTAGE (pulser voltage)

You can use the function VOLTAGE to set the pulser voltage.
The following settings are possible:
HIGH – high voltage
LOW – low voltage
The setting HIGH is recommended for all tests in which maximum sensitivity is important, e.g. for the detection of small flaws. Choose the setting LOW for broadband probes or if narrow echoes are required (better lateral resolution).
Square-wave pulser
If your USM Go is equipped with a square-wave pulser (option) and if the square-wave pulser is chosen as pulser type ( page 5-75), then you can set the pulser voltage within the range of 120 … 300 V in steps of 10 V. Instead of the function ENERGY, the function WIDTH is then avail able.
see Section Selecting the pulser type,
-
5-14 Issue 6 (02/2013) USM Go
Page 91
Setting the pulser (function group PULSER) 5 Operation
ATTENTION
Note

ENERGY

Use the data sheet for your probe to check which maximum voltage is allowed to be ap plied.
The pulser voltage and the pulse width can be automatically limited, depending on the pulse repetition frequency (
see Section PRF
MODE (pulse repetition frequency), page 5-17) setting. This function helps to avoid heat accumulations in the pulser elec­tronics.
– Select the function VOLTAGE.
– Press the function keys to choose the required pulser
voltage.
You can use the function ENERGY to set the penetra­tion or sound energy.
­The following settings are possible:
HIGH – high energy
LOW – low energy
The setting HIGH is recommended for all tests in which maximum sensitivity is important, e.g. for the detection of small flaws. Choose the setting LOW for broadband probes or if narrow echoes are required (better lateral resolution).
– Select the function ENERGY.
– Press the function keys to choose the required set-
ting.
USM Go Issue 6 (02/2013) 5-15
Page 92
5 Operation Setting the pulser (function group PULSER)
Note

WIDTH

This function is only available if the option square-wave pulser is chosen as pulser type (
see Section Selecting
the pulser type, page 5-75).
You can use the function WIDTH to adjust the pulse width for the square-wave pulser. You can adjust a val ue within the range of 30 … 500 ns (nanoseconds) in steps of 10
ns.
The following equation produces an approximation for the suitable pulse width:
Nominal width in nanoseconds
= 500/probe frequency in MHz
For example, the equation for a 2.25 MHz probe results in:
Nominal width in nanoseconds
= 500/2.25 ns = 222 nanoseconds
The pulser voltage and the pulse width can be automatically limited, depending on the pulse repetition frequency (
see Section PRF
MODE (pulse repetition frequency),
-
page 5-17) setting. This function is used for limiting the signal loss.
– Select the function WIDTH.
– Press the function keys to select the required value.
5-16 Issue 6 (02/2013) USM Go
Page 93
Setting the pulser (function group PULSER) 5 Operation

DAMPING

This function is used for matching the probe. By setting the damping of the probe oscillating circuit you can vary the height, width, and resolution of the echo display.
The following settings are possible:
1000 Ohm
low damping, the echoes become higher and broad­er.
50 Ohm
reduces the echo height but produces narrower echoes with a higher resolution.
– Select the function DAMPING.
– Press the function keys to select the required value.

PRF MODE (pulse repetition frequency)

The pulse repetition frequency indicates the number of times an initial pulse is triggered per second. You can decide whether you need the highest possible PRF val ue or whether you are satisfied with a low value. You have three fixed steps and one user-variable step avail able for the adjustment.
The larger your test object, the smaller PRF values are needed to avoid phantom echoes. However, the A-scan update rate becomes lower in the case of smaller PRF values; high values are therefore required if a test object should be scanned fast.
The best way to determine the suitable PRF value is by experimenting: Start from the highest step and reduce the value until there are no more phantom echoes.
-
-
USM Go Issue 6 (02/2013) 5-17
Page 94
5 Operation Setting the pulser (function group PULSER)
Note
The following settings are possible:
AUTO LOW 400 Hz
AUTO MED 1000 Hz
AUTO HIGH 1500 Hz
MANUAL
You can detect and avoid phantom echoes by means of the optional function phantom PRF (
see Section Phantom echo detector,
page 5-58).
– Select the function PRF MODE.
– Press the function keys to select the required value.
– If you have selected MANUAL, set the required value
using the navigation.
5-18 Issue 6 (02/2013) USM Go
Page 95
Setting the receiver (function group RECEIVER) 5 Operation

5.6 Setting the receiver (function group RECEIVER)

You will find all functions for setting the receiver in the function group RECEIVER.
– Switch to the first operating level.
– Select the function group RECEIVER.

FREQUENCY

In this function, you can set the frequency of the receiver according to the frequency of your probe.
The following settings are possible:
BROADBAND
1 - 5 MHz
2 MHz
2.25 MHz
4 MHz
5 MHz
10 MHz
13 MHz
15 MHz
– Select the function FREQUENCY.
– Press the function keys to select the required value.
USM Go Issue 6 (02/2013) 5-19
Page 96
5 Operation Setting the receiver (function group RECEIVER)

RECTIFY

You can use the function RECTIFY to select the rectifi­cation mode of the echo pulses according to your appli­cation.
The following settings are possible:
FULLWAVE
All half-waves are displayed above the base line on the screen.
POS HALFWAVE
Only positive half-waves are displayed above the base line on the screen.
NEG HALFWAVE
Only negative half-waves are displayed above the base line on the screen.
RF (radio frequency)
There is no rectification. Both positive and negative wave portions are displayed with true amplitude.
– Select the function RECTIFY.
– Press the function keys to choose the required set-
ting.

DUAL (pulser-receiver separation)

You can use the function DUAL to activate the pulser­receiver separation ( probe, page 3-9).
OFF
Single-element operation; the probe connector sock­ets are connected in parallel.
ON
Dual mode for using dual-element probes; one probe connector is connected to the amplifier input, the ini tial pulse is applied to the other probe connector.
THROUGH
Through-transmission mode for using two separate probes; the receiver is connected to one probe con nector and the pulser is connected to the other. As the sound wave passes through the test object only once in through-transmission mode, all range and wall thickness measurement functions are adjusted accordingly.
– Select the function DUAL.
– Press the function keys to choose the required set-
ting.
see Chapter 3.3 Connecting a
-
-
5-20 Issue 6 (02/2013) USM Go
Page 97
Setting the receiver (function group RECEIVER) 5 Operation
ATTENTION

REJECT

You can use the function REJECT to suppress unwant­ed echo indications, e.g. structural noise from the test object.
The height in % indicates the minimum height that the echoes must attain in order for them to be displayed on the screen at all. The reject setting cannot be higher than 80 %.
You should handle this function with great caution as it may of course happen that you suppress echoes from flaws as well. Many test specifications expressly forbid using the reject function.
– Select the function REJECT.
– Press the function keys to select the required value.
USM Go Issue 6 (02/2013) 5-21
Page 98
5 Operation Setting the gates (function groups GATE A and GATE B)

5.7 Setting the gates (function groups GATE A and GATE B)

You will find all functions for setting the gates A and B in the function groups GATE
– Switch to the first operating level.
– Select the function group GATE A or GATE B.
A and GATE B.

Tasks of the gates

The gates monitor the area of the test object where you expect to detect a flaw. If an echo exceeds or falls below the gate, an alarm signal is output ( Alarms, page 4-6).
The gates A and B are independent of each other. Gate
A can also assume the function of an echo start
gate for gate
The gate chooses the echo for the digital time-of­flight and amplitude measurement. The measured value is displayed in the measurement line ( tion Measurement line, page 4-5).
Display of the gates
The gates are displayed in different colors for easier al­location.
B.
see Section
see Sec-
5-22 Issue 6 (02/2013) USM Go
Page 99
Setting the gates (function groups GATE A and GATE B) 5 Operation

A-START/B-START (starting point of the gate)

You can set the starting point of the gates A or B within an adjustment range of 0
You can set the value by coarse adjustment steps or by fine adjustment ( ment of functions, page 4-10).
– Select the function A-START or
B-START.
– Set the required value by means of the function keys
or using the navigation.
see Section Coarse and fine adjust-
… 27940 mm.

A-WIDTH/B-WIDTH (width of the gates)

You can set the width of the gates A or B within an ad­justment range of 1.00 … 27940.00 mm.
You can set the value by coarse adjustment steps or by fine adjustment ( ment of functions, page 4-10).
– Select the function A-WIDTH or
B-WIDTH.
– Set the required value by means of the function keys
or using the navigation.
see Section Coarse and fine adjust-
USM Go Issue 6 (02/2013) 5-23
Page 100
5 Operation Setting the gates (function groups GATE A and GATE B)

A-THRESHOLD/B-THRESHOLD (response and measurement threshold of the gate)

You can define the threshold value of the gates A or B within the range of 5 ing an alarm if this value is exceeded or not reached.
In RF mode, you can additionally set the threshold with­in the range of –5 … –95 %.
– Select the function A-THRESHOLD or
B-THRESHOLD.
– Press the function keys to select the required value.
… 95 % screen height for trigger-
5-24 Issue 6 (02/2013) USM Go
Loading...