Waygate Technologies MIC 20 Operating Manual

Krautkramer MIC 20
Technical Reference and Operating Manual
Ident No. 28 702
This Issue 02, 04/2005 applies to the software version V 01.00
The version number of the software is displayed in the menu bar line on the screen or in the menu ConfigInfo (ref. chapter 4.7).
Subject to change without notice.
0-2 Issue 02, 04/2005 Krautkramer MIC 20
1 Introduction ....................................... 1-1
1.1 Safety information ................................... 1-2
Power supply ............................................. 1-2
Software .................................................... 1-3
1.2 About this manual ................................... 1-3
Important information ................................. 1-3
1.3 Layout and presentation in this manual . 1-4
Attention and note symbols ....................... 1-4
Listings ...................................................... 1-4
Operating steps ......................................... 1-4
1.4 Prerequisites for hardness testing .......... 1-5
Operator training ........................................ 1-5
Technical test requirements ....................... 1-6
Choice of the appropriate test equipment ... 1-6
1.5 Important information about hardness
testing using the Krautkramer MIC 20 .... 1-7
Test material .............................................. 1-7
Test method ............................................... 1-7
Conversion of hardness values .................. 1-9
1.6 The Krautkramer MIC 20 ......................... 1-10
Special features of the MIC 20.................. 1-11
Large variety of application possibilities ....1-12
2 Standard package and accessories 2-1
2.1 Standard package .................................... 2-3
2.2 Required accessories .............................. 2-5
2.3 Recommended accessories
(general) .................................................... 2-7
2.4 Recommended accessories
(UCI method) ............................................ 2-8
2.5 Recommended accessories
(rebound method) ...................................2-11
2.6 Spare parts requirements
(UCI method) ...........................................2-12
2.7 Spare parts requirements
(rebound method) ...................................2-13
Krautkramer MIC 20 Issue 02, 04/2005 0-1
ContentsContents
3 Initial start-up ..................................... 3-1
3.1 Power supply ........................................... 3-2
Operation using the power supply unit ........ 3-2
Operation using the battery pack
MIC 20-BAT ............................................... 3-3
3.2 Connecting a probe or
an impact device ...................................... 3-4
Connecting a UCI probe ............................. 3-5
Connecting an impact device ..................... 3-6
Disconnecting the connecting cable ........... 3-7
Exchanging the test attachment ................ 3-8
3.3 Turning on/off .......................................... 3-8
Turning on .................................................. 3-8
Turning off .................................................. 3-8
Emergency-stop ......................................... 3-9
Cold start ................................................... 3-9
4 Operation ........................................... 4-1
4.1 Fundamental principles of operation ..... 4-2
Operator’s controls and displays ................ 4-3
Display screen ........................................... 4-3
User interface ............................................ 4-4
Virtual keyboard......................................... 4-5
Operation using the keypad or
touch screen .............................................. 4-6
Installing the instrument ............................. 4-6
4.2 Probe and impact device ......................... 4-7
Guiding the probe ....................................... 4-7
Operating the impact device ...................... 4-8
4.3 Hardness testing .....................................4-10
Starting and closing a test series .............. 4-10
Viewing and evaluating test results ...........4-11
Conversion into other hardness scales ...... 4-14
4.4 Calibration ...............................................4-15
Carrying out the calibration ....................... 4-16
Saving calibration data ..............................4-18
0-2 Issue 02, 04/2005 Krautkramer MIC 20
Deleting calibration data ............................4-19
Restoring the standard calibration ............. 4-20
4.5 Saving the test data ................................4-21
Saving files ............................................... 4-22
Deleting files or directories ........................4-22
Opening and closing directories ................4-23
Creating new directories ............................4-23
Opening files .............................................4-24
Editing saved files .................................... 4-24
File management using instrument keys ...4-25
4.6 Printing test reports ................................ 4-25
Prerequisites for printing ...........................4-26
Selecting and printing a test series ........... 4-27
4.7 Instrument configuration ........................4-30
Evaluation parameters ..............................4-30
System settings........................................ 4-35
System information ...................................4-39
4.8 Saving and loading
instrument parameters ...........................4-41
Saving instrument parameters .................. 4-41
Loading and deleting
instrument parameters ..............................4-41
4.9 Functional tests ......................................4-42
Functional test for UCI method .................4-42
Functional test for rebound method ........... 4-43
4.10 Calibrating the touch screen ..................4-44
4.11 Troubleshooting ......................................4-45
Error messages ........................................ 4-45
Errors ........................................................ 4-45
5 Care and maintenance ...................... 5-1
5.1 Care .......................................................... 5-2
Care of instrument ..................................... 5-2
Care of batteries ........................................ 5-2
Charging the batteries ................................ 5-3
Krautkramer MIC 20 Issue 02, 04/2005 0-3
5.2 Maintenance ............................................. 5-4
Cleaning the probe ..................................... 5-4
Cleaning the impact device ........................ 5-4
6 Interfaces and data transfer.............. 6-1
6.1 Interfaces .................................................. 6-2
Serial interface RS232 ............................... 6-3
Ethernet interface ...................................... 6-3
6.2 Data transfer to a printer .......................... 6-4
Requirements for printing ........................... 6-4
6.3 Data transfer to a computer ..................... 6-5
7 Hardness testing method ................. 7-1
7.1 The UCI method ....................................... 7-2
The treatment of test material .................... 7-3
7.2 The rebound method ............................... 7-5
The treatment of test material .................... 7-6
7.3 Conversion of hardness values ............... 7-8
7.4 Information on the
statistical evaluation ............................... 7-12
Statistical average .................................... 7-12
Relative range of spread ...........................7-13
Relative standard deviation ....................... 7-13
Calculation of process capability ...............7-13
8 Specifications .................................... 8-1
Basic instrument ........................................ 8-2
Connections and interfaces ........................ 8-2
Power supply and operating time ................ 8-3
Ambient conditions .................................... 8-3
Evaluation .................................................. 8-4
9 Annex ................................................. 9-1
9.1 EC Declaration of Conformity ................. 9-2
9.2 Manufacturer/Service addresses ............. 9-2
10 Index ................................................. 10-1
0-4 Issue 02, 04/2005 Krautkramer MIC 20
Introduction 1
Krautkramer MIC 20 Issue 02, 04/2005 1-1
IntroductionIntroduction Safety information
1.1 Safety information
The Krautkramer MIC 20 is designed and tested ac­cording to DIN EN 61 010 Part 1, March 1994, “Safety requirements for electric measuring, control, and labora­tory devices”, and was technically in a perfectly safe and faultless condition when leaving the manufacturing works.
In order to maintain this condition and to ensure a safe operation, it is absolutely necessary that you read the following safety information before putting the instru­ment into operation.
A Attention:
The Krautkramer MIC 20 is an instrument meant for materials testing. Any use in medical or other appli­cations is not allowed!
The instrument may only be used in industrial envi­ronments.
Power supply
Battery pack and power supply unit
The Krautkramer MIC 20 can be operated using the battery pack MIC 20-BAT, or a power supply unit. You can charge the battery pack MIC 20-BAT in the instru­ment itself while the instrument is supplied with power via the power supply unit and turned off. As soon as you turn the instrument on, the power supply to the battery compartment is interrupted.
NiCd or NiMH single cells
The operation using NiCd or NiMH single cells is like­wise possible but is not recommended because of the clearly reduced operating time.
A Attention:
You cannot charge NiCd or NiMH single cells in the instrument itself but only by means of an external bat­tery charger approved for this purpose.
If you are only using the power supply unit for a longer period of time, you should remove the batteries from the instrument.
1-2 Issue 02, 04/2005 Krautkramer MIC 20
IntroductionIntroductionSafety information
H Note:
Alkaline cells are not suitable due to their high internal impedance.
Software
According to the current state of the art, software is never completely free from errors or defects.
For this reason, before using any software-controlled test equipment, it should be ensured that the required functions operate perfectly in the intended combination.
If you have any questions regarding the use of the in­strument, please contact your nearest GE Inspection Technologies representative.
1.2 About this manual
This manual describes the operation of the hardness tester Krautkramer MIC 20.
Please read carefully through this manual in order to be able to operate all functions of your instrument quickly and reliably. You’ll be able to use the complete range of instrument functions and, at the same time, to avoid faults and operating errors which may lead to incorrect test results.
Important information
Even if you might be familiar with hardness testing methods, please always observe the information in chapters 1.4 and 1.5. In chapter 1.4, you will find impor­tant limitations and prerequisites for hardness testing in general (training, knowledge of the specific technical test requirements and limits of testing, choice of the appropriate test device).
In chapter 1.5, you will find concrete information about hardness testing using the MIC 20 that you always have to follow in order to ensure correct test results.
Krautkramer MIC 20 Issue 02, 04/2005 1-3
IntroductionIntroduction Layout and presentation in this manual
1.3 Layout and presentation in this manual
To make it easier for you to use the manual, the operat­ing steps, notes, etc. are always presented in the same way. This will help you find individual pieces of informa­tion quickly.
Attention and note symbols
A Attention:
You will find the Attention symbol in the case of any peculiarities or special aspects in the operation which could affect the correctness of the results.
H Note:
At Note, you will find e.g. references to other chapters or special recommendations for a function.
Listings
Listings are presented in the following form:
Variant A
Variant B
❚ ...
Operating steps
Operating steps appear as shown in the following ex­ample:
– Place the handheld probe vertically to the surface to
be tested.
– Hold the probe tight with one hand so that the foot
stays perpendicular to the surface during the mea­surement.
1-4 Issue 02, 04/2005 Krautkramer MIC 20
IntroductionIntroductionPrerequisites for hardness testing
1.4 Prerequisites for hardness testing
In this operating manual you will find all essential infor­mation on how to operate the Krautkramer MIC 20. In addition, there are a number of factors which affect the test results. As a description of these factors would go beyond the scope of an operating manual, only the three most important conditions are therefore described here:
operator training
knowledge of special technical test requirements and
limits
choice of the appropriate test equipment
A Attention:
Lack of knowledge of the above-mentioned subjects may lead to incorrect test results with unforeseeable consequences.
GE Inspection Technologies organizes training courses in the field of hardness testing. You will receive informa­tion on the scheduled dates on request.
Operator training
The reliable and safe operation of a hardness testing device requires a proper training in materials testing.
A proper training comprises for example adequate knowledge of:
hardness testing on metallic materials
effects due to material properties, especially due to
the microstructure, on hardness testing and on the corresponding choice of the appropriate hardness tester
problems relating to the comparability of different
hardness values, such as Vickers, Rockwell, and Brinell
effects due to surface finish on the hardness value
effects of the test load on the determined hardness
value
Krautkramer MIC 20 Issue 02, 04/2005 1-5
IntroductionIntroduction Prerequisites for hardness testing
Technical test requirements
Every hardness test is subject to specific technical test requirements. The most important ones are:
definition of the scope of testing
choice of the appropriate test method
consideration of material properties
determination of limits for evaluation
Choice of the appropriate test equipment
It is the task of those responsible for the test to fully inform the operator about the technical test require­ments. Moreover, a clear and thorough interpretation of the corresponding test specifications is absolutely nec­essary.
Information about test methods and test specifications may be obtained, for example, from various institutions, industrial companies, and authorities.
1-6 Issue 02, 04/2005 Krautkramer MIC 20
IntroductionIntroductionImportant information about hardness testing using the Krautkramer MIC 20
1.5 Important information about hardness testing using the Krautkramer MIC 20
Please find in the following a summary of the most important technical test requirements that you always have to observe in order to obtain reliable and reproduc­ible test results.
A Attention:
Do not carry out the hardness test twice at the same test position, otherwise measuring errors may occur due to the hardness increase of the surface. The dis­tance between the test positions should be at least 3 mm.
Krautkramer MIC 20 Issue 02, 04/2005 1-7
Test material
The surfaces must be free from any impurities (oil, dust, etc.) and rust. The peak-to-valley height should not exceed approx. 30 % of the penetration depth. Rough surfaces lead to a higher variation range of the single readings. If necessary, polish rougher surfaces, for example using our grinding set MIC 1060 (ref. chap­ter 2).
Test method
The MIC 20 supports both quasi-static hardness testing according to the UCI method and dynamic hardness testing according to the rebound method.
All Krautkramer UCI probes and the rebound impact devices D, G, and E are supported in these processes.
The UCI method
The UCI method is a comparative method (contribution of the Young’s modulus to the measurement) with a very high reproducibility of measurements. The UCI method does not replace the classical Vickers hard­ness testing method but constitutes a fast and reliable addition to it.
IntroductionIntroduction Important information about hardness testing
The direct comparison with the results of the standard­ized Vickers measurement according to the material samples is therefore indispensable for the assessment of the measuring accuracy of the UCI method. This means:
The test forces (probes) must match the surface quality of the material:
Smooth, homogeneous surfaces require low test
forces.
Rougher, coarse-grained surfaces require as high
test forces as possible.
A Attention:
It is absolutely necessary that you calibrate your MIC 20 to the material to be tested.
The calibration is required only once for this since you can save calibrations and recall them again as required without any problem (ref. chapter 4.4).
The calibration for low-alloy or unalloyed steel is already predefined in the instrument. You should check them from time to time (ref. chapter 4.9 Functional tests).
Please also read the notes on the UCI method in chap­ter 7.
The Rebound method
The rebound method is a dynamic method with a very high reproducibility of measurements. The rebound method does not replace the classical Brinell hardness testing method or other standardized methods but con­stitutes a fast and reliable addition to them.
The still existing energy of an impact body after the rebound from the material surface is measured. It should be kept in mind in this regard that the loss of energy suffered in this process also depends on the mechanical properties of the material, i.e. mainly on its Young’s modulus or modulus of elasticity.
The direct comparison with the results of the standard­ized Vickers measurement according to the material samples is therefore indispensable for the assessment of the measuring accuracy of the method. This means:
1-8 Issue 02, 04/2005 Krautkramer MIC 20
IntroductionIntroductionImportant information about hardness testing
A Attention:
It is absolutely necessary that you set your MIC 20 to the suitable material group before carrying out the test and that you additionally calibrate it if necessary.
The calibration is necessary only once in this process since you can save the calibrations and recall them again as required without any problem (ref. chapter 4.4).
The material group for low-alloy or unalloyed steel is predefined in the instrument as a default setting.
You should check this from time to time (ref. chapter
4.9 Functional tests).
Test objects having a mass of at least 5 kg can be tested without any additional supports; more leight­weight objects require a support to which they have to be rigidly coupled. You should use large metal supports that do not give way for this purpose.
The test objects should have a minimum wall thickness of 20 mm. We recommend the UCI method, or the use of a suitable UCI probe in combination with the MIC 20 for test objects having smaller wall thicknesses.
A Attention:
Should the test objects give way or spring, the result may be measuring errors!
Please read also chapter 7 on the rebound method.
Conversion of hardness values
The conversion of hardness values into other hardness scales is only possibly with restrictions.
Hardness values determined according to different methods cannot be converted into each other by means of generally applicable relations. The restrictions regarding the conversion, stated in the DIN 50 150, ASTM E140 specifications, must therefore always be taken into account!
The indentation behavior of the material is determined by its stress-strain behavior. Shape and material of the indenter, the size of indentation, and consequently the measured area vary, depending on the test method used.
The conversion of hardness values both into one an­other and into tensile strength values can be inaccurate
Krautkramer MIC 20 Issue 02, 04/2005 1-9
Introduction Important information about hardness testing
or unacceptable depending on the material, treatment condition, and surface quality.
A Attention:
Any illegal or unacceptable conversions may lead to serious errors in the interpretation of test results.
1.6 The Krautkramer MIC 20
The Krautkramer MIC 20 is a mobile and easy-to-use hardness tester which can be equally used for tests according to the UCI method and for tests according to the rebound method.
After connecting the probe, the instrument is automati­cally set to the corresponding test method; tests can be carried out quickly and without any problem.
The fact that both methods are supported leads to a large variety of application possibilities with only one single instrument. Additional probes extend the range of applications if required.
The operation of the Krautkramer MIC 20 (calibration, setup, evaluation, data memory) is carried out by means of the graphic user interface presented on the display and adapted to the known Windows standard. The mouse is replaced by the touch screen for this purpose.
As an alternative, conventional buttons are available for most operating functions.
1-10 Issue 02, 04/2005 Krautkramer MIC 20
IntroductionThe Krautkramer MIC 20
Special features of the MIC 20
Measurements on castings and forgings, hardened
surfaces and welds
Automatic adaptation of the test method and of the
test parameters to the connected probe
Color LCD 5.7" for displays and operating functions
(TFT or CSTN)
Support of all Krautkramer UCI probes
Support of the Krautkramer rebound
impact devices D, G, and E
Non-directional measurement (patented signal
processing with the rebound method)
Storage of measurement data and clear representa-
tion of measurement series as a diagram and histo­gram
Simple and fast calibration, saving and recalling of
calibration data at the press of a button
Output of measurement data directly to a printer or
transfer to a computer
Krautkramer MIC 20 Issue 02, 04/2005 1-11
Introduction The Krautkramer MIC 20
Operation using the battery pack or power supply
unit
Large variety of application possibilities
You can use the Krautkramer MIC 20 to measure any­where and in any direction; the direction must not be set beforehand.
The MIC 20 is mainly suitable
for measuring hardness of low-alloy or unalloyed
steels,
for measuring hardness of high-alloy steels,
for measuring hardness of nonferrous metals.
The MIC 20 is available in two versions:
Basic version MIC 20 with color CSTN screen
MIC 20 TFT with color TFT display screen
1-12 Issue 02, 04/2005 Krautkramer MIC 20
Standard package and accessories 2
Krautkramer MIC 20 Issue 02, 04/2005 2-1
Standard package and accessories Standard package
This chapter informs you about the standard package and the accessories available for the Krautkramer MIC 20.
It describes
component parts of the product
required and recommended accessories
spare parts requirements
2-2 Issue 02, 04/2005 Krautkramer MIC 20
2.1 Standard package
Product code Description Order no.
MIC 20 Portable Hardness Tester according to the quasi-static 35 468
Standard package and accessoriesStandard package
UCI principle and the dynamic Rebound principle. Display of the hardness values in HV, HB, HRC, HRB, N/mm in HL and HS.
including:
Instrument MIC 20 35 493
Mains adapter 101 075
Transport case 101 554
Operating manual German 28 701
or Operating manual English 28 702
2
and additional for the Rebound principle
Krautkramer MIC 20 Issue 02, 04/2005 2-3
Standard package and accessories Standard package
Product code Description Order no.
MIC 20 TFT Portable Hardness Tester according to the quasi-static 35 479
UCI principle and the dynamic Rebound principle. Display of the hardness values in HV, HB, HRC, HRB,
2
N/mm
and additional for the Rebound principle
in HL and HS.
including:
Instrument MIC 20 TFT 35 492
Mains adapter 101 075
Transport case 101 554
Operating manual German 28 701
or Operating manual English 28 702
2-4 Issue 02, 04/2005 Krautkramer MIC 20
2.2 Required accessories
Product code Description Order no.
MIC 201-A Handheld probe 10 N 34 104
MIC 205-A Handheld probe 50 N 34 105
MIC 2010-A Handheld probe 98 N 34 106
MIC 201-AL Extended handheld probe 10 N 34 392
MIC 205-AL Extended handheld probe 50 N 34 282
MIC 201-AS Short handheld probe 10 N 34 711
MIC 205-AS Short handheld probe 50 N 34 712
Standard package and accessoriesRequired accessories
Handheld Probes (UCI method)
(each probe complete with cable)
Krautkramer MIC 20 Issue 02, 04/2005 2-5
Standard package and accessories Required accessories
Product code Description Order no.
Motorized Probes (UCI method)
(each probe complete with cable)
MIC 211-A Motor probe 8.6 N 34 381
MIC 2103-A Motor probe 3 N 34 382
MIC 2101-A Motor probe 1 N 35 577
Impact Devices (rebound method)
(each without connecting cable for impact device)
Dyna D Impact device with 3 mm tungsten-carbide 34 248
Dyna G Impact device with 5 mm tungsten-carbide 34 549
Dyna E Impact device with diamond-tipped body 34 588
Dyna 50 Connecting cable for impact device 34 329
2-6 Issue 02, 04/2005 Krautkramer MIC 20
Recommended accessories (general)
2.3 Recommended accessories (general)
Product code Description Order no.
MIC 20-BAT NiMH accumulator pack, 4.5 Ah, internal charging 35 452
MIC 1060 Battery grinding set 34 380
MIC 300 Technical book on hardness testing (only in German) 28 837
Printer cable (serial/parallel) for connection to 101 761 Hewlett Packard Printers of the hpdeskjet 9xx series
UDAT UltraDAT data management program for transfer and 35 549
documentation of hardness measurement data
Data transfer cable from the instrument to a PC 101 785
ZG-F Couplant 54 558
Standard package and accessories
Krautkramer MIC 20 Issue 02, 04/2005 2-7
Standard package and accessories
2.4 Recommended accessories (UCI method)
Product code Description Order no.
Guiding devices and test supports
MIC 270 Surface attachment for MIC 201-A, MIC 205-A, 32 084
MIC 2010-A
MIC 271 Prism adapter for MIC 201-A, MIC 205-A 32 993
MIC 220 Test specimen holder for MIC 2101-A, MIC 2103-A 30 766
and MIC 211-A
MIC 221 Universal test support with instrument holder 33 541
MIC 227 Test support for determination of hardness progression 35 264
MIC 222-A Precision test support 35 546
MIC 2221 Magnetic foot for support MIC 222-A 33 909
MIC 2220 Support block for flat, thin parts 33 651
Recommended accessories (UCI method)
MIC 225 Camshaft test stand 34 869
2-8 Issue 02, 04/2005 Krautkramer MIC 20
Recommended accessories (UCI method)
Product code Description Order no.
MIC 100 Probe attachment set 29 929
MIC 100-1 Prism attachment 29 920
MIC 120 Magnetic probe shoe for curved surfaces 31 854
MIC 223 Quick test support with magnetic foot 34 020
MIC 25C Hardness Reference Plate 25 HRC 33 905
MIC 45C Hardness Reference Plate 45 HRC 33 906
MIC 65C Hardness Reference Plate 65 HRC 33 907
MIC 1V100 Hardness Reference Plate 150 HV10 34 279
MIC 2V010 Hardness Reference Plate 240 HV1 33 896
Standard package and accessories
Guiding devices for motor probes
Hardness reference plates with MPA certificate
MIC 2V050 Hardness Reference Plate 240 HV5 33 899
Krautkramer MIC 20 Issue 02, 04/2005 2-9
Standard package and accessories
Product code Description Order no.
MIC 2V100 Hardness Reference Plate 240 HV10 33 902
MIC 5V010 Hardness Reference Plate 540 HV1 33 897
MIC 5V050 Hardness Reference Plate 540 HV5 33 900
MIC 5V100 Hardness Reference Plate 540 HV10 33 903
MIC 8V010 Hardness Reference Plate 840 HV1 33 898
MIC 8V050 Hardness Reference Plate 840 HV5 33 901
MIC 8V100 Hardness Reference Plate 840 HV10 33 904
Recommended accessories (UCI method)
2-10 Issue 02, 04/2005 Krautkramer MIC 20
Recommended accessories (rebound method)
Standard package and accessories
2.5 Recommended accessories (rebound method)
Product code Description Order no.
Hardness reference blocks (rebound method)
MIC D62 Hardness reference block 620 HV100 34 393
MIC D62MPA Hardness reference block 620 HV100, 34 573
certified by MPA, Germany
MIC G38 Hardness reference block 380 HV100 34 631
MIC G38MPA Hardness reference block 380 HV100, 34 657
certified by MPA, Germany
Test attachments for Dyna D and E (rebound method)
Dyna 41 Set (5 pcs.) of test attachments for cylindrical and 34 536
hollow-cylindrical surfaces
Dyna 42 Set (5 pcs.) of test attachments for spherical and 34 539
hollow-spherical surfaces
Krautkramer MIC 20 Issue 02, 04/2005 2-11
Standard package and accessories
2.6 Spare parts requirements (UCI method)
Product code Description Order no.
MIC 1050 Probe cable for handheld probes 34 071
MIC 1051 Probe cable for motor probes 34 378
MIC 1052 Probe cable for shortened handheld probes for 34 713
MIC 201-AS and MIC 205-AS
Ball headed probe grip 33 854
Spare parts requirements (UCI method)
2-12 Issue 02, 04/2005 Krautkramer MIC 20
Spare parts requirements (rebound method)
Standard package and accessories
2.7 Spare parts requirements (rebound method)
Product code Description Order no.
Impact body D, new 34 443
Impact body G, new 34 596
Impact body E, new 34 593
Impact body D, replacement in exchange by 34 572 Krautkramer Service Department
Impact body G, replacement in exchange by 35 265 Krautkramer Service Department
Dyna 50 Connecting cable for impact device 34 329
Cleaning brush for Dyna D or Dyna E 34 420
Cleaning brush for Dyna G 34 618
Standard test attachment for impact device D or E 34 312
Test attachment for impact devices D or E, 34 656 Ø = 13.5 mm
Krautkramer MIC 20 Issue 02, 04/2005 2-13
Standard package and accessories
Product code Description Order no.
Spare parts requirements (Rebound method)
Standard test attachment for impact device G 34 569
Test attachment for impact device G, 34 634 diameter = 50 mm
Replacement grinder for battery grinding set MIC 1060 18 115
2-14 Issue 02, 04/2005 Krautkramer MIC 20
Initial start-up 3
Krautkramer MIC 20 Issue 02, 04/2005 3-1
Initial start-up Power supply
3.1 Power supply
You can operate the Krautkramer MIC 20 either with the battery pack MIC 20-BAT or with the power supply unit. The operation with the power supply unit is possible even if the battery pack MIC 20-BAT is in the instru­ment. The power supply to the battery compartment is then automatically interrupted.
Operation using the power supply unit
The Krautkramer MIC 20 is supplied either with a desk­top power supply unit or with a plug-in power supply unit. The power supply unit is automatically set to any a.c. voltage between 100 and 240 V (nominal voltage).
Operation using the desktop power supply unit
The desktop power supply unit is provided with a cable equipped with a Euro plug connector and inlet connec­tor for non-heating appliances.
Plug the cable’s inlet connector for non-heating appli­ances in the socket for non-heating appliances of the power supply unit.
Operation using the plug-in power supply unit
The plug-in power supply unit is delivered with two dif­ferent socket-outlet adapters – for Euro and U.S. stan­dard. If the adapter plug on your power supply unit does not correspond to your socket-outlet standard, you can exchange it.
Just pull off the attached adapter, and replace it with the required one.
H Note:
You should exchange the socket-outlet adapter only once; the plug-in power supply unit is not meant for frequent exchanges.
3-2 Issue 02, 04/2005 Krautkramer MIC 20
Initial start-upPower supply
Connecting
Use the corresponding power supply unit to connect the Krautkramer MIC 20 to a suitable mains socket-outlet. The socket-contact for the power supply unit cable is located at the back of the instrument.
Operation using the battery pack MIC 20-BAT
You can operate the Krautkramer MIC 20 with the NiMH battery pack MIC 20-BAT (ref. chapter 2).
The operation is likewise possible with 6 NiCd or NiMH single C-cells each, however, it is not recommended because of the clearly reduced operating time.
Inserting batteries
The battery compartment is located in the instrument bottom.
– Move the fastener towards the lid center in order to
open the lid, and remove the lid.
– Insert the plug of the battery pack into the socket-
contact in the battery compartment until it snaps into place. When inserting the battery pack, make sure that no squeezing or kinking of the cables is pos­sible.
– If you use single cells, insert the individual cells into
the battery compartment one by one. Follow the in­structions regarding alignment and polarity in the battery compartment.
– Place the lid back on, move the fastener back and
close the battery compartment so that the fastener locks home perceptibly.
Krautkramer MIC 20 Issue 02, 04/2005 3-3
Initial start-up Power supply
H Notes:
Remove the batteries from the instrument if you’re not going to use it for a longer period of time!
Used or defective batteries are special waste and have to be disposed of as provided by the law!
When using the battery pack MIC 20-BAT (unless NiMH or NiCd single cells are used), the current status of the battery is indicated on the display of the MIC 20 by an icon:
An alarm is output with low voltage. In such a case, exchange the batteries at once. The Krautkramer MIC 20 is automatically turned off if the voltage gets too low in order to ensure a reliable mode of function­ing.
You will find more details on the care and charging of batteries in chapter 5.
3.2 Connecting a probe or an impact device
You can connect probes for the quasi-static hardness testing according to the UCI method as well as impact devices for the dynamic hardness testing according to the rebound method to the MIC 20.
Probes and impact devices are connected to the MIC 20 via corresponding connecting cables. The con­necting cables are equipped with a round plug for the connection to the probe or to the impact device, and with a square plug for the connection to the instrument.
H Note:
You should only change the probe or the impact device when the MIC 20 is switched off.
3-4 Issue 02, 04/2005 Krautkramer MIC 20
Connecting a UCI probe
– Align the round plug of the cable correctly with the
socket of the probe by means of the markings, and push it carefully into the socket until it locks into place.
– Plug the square plug of the connecting cable in the
socket at the back of the instrument.
– Check that both plugs fit tightly.
– Attach the probe handle to the probe if necessary.
– Screw on the conical test attachment for short-time
measurements using the UCI handheld probes MIC 201-A, MIC 205-A, or MIC 2010-A. Use the cy­lindrical test attachment for measurements with de­fined dwell times.
– Switch the MIC 20 on. An icon in the top left corner
of the screen shows that a probe has been correctly connected:
Initial start-upConnecting a probe or an impact device
Conical probe attachment for short-time measurements
Cylindrical probe attachment for measurements with a defined dwell time
Krautkramer MIC 20 Issue 02, 04/2005 3-5
Initial start-up Connecting a probe or an impact device
H Note:
If you want to carry out measurements with defined dwell times, please use the support
MIC 222-A (support with precise probe guidance).
This will help you to avoid measuring inaccuracies to a large degree.
In addition, other accessories are available to make the measurements easier for you to carry out (ref. chapter 2).
Connecting an impact device
The following impact devices are available for use in combination with the MIC 20:
Dyna D
Standard impact device for all materials that can be tested
Dyna G
Impact device for solid test objects, e.g. castings or forgings
Dyna E
Impact device for the hardness range over 650 HV
A Attention:
The impact device Dyna G must only be used up to a hardness to be tested of max. 650 HB, otherwise the impact body may be destroyed.
H Note:
A larger test attachment (diameter 50 mm) is available for the impact device Dyna G for a more stable mea­surement on large, flat workpieces.
3-6 Issue 02, 04/2005 Krautkramer MIC 20
Connecting a probe or an impact device
Initial start-up
For measurements on test objects having curved sur­faces, you have special test attachments at your dis­posal in order to achieve a better positioning with the impact devices Dyna D and Dyna E (ref. chapter 2).
– Align the round plug of the cable correctly with the
socket of the probe by means of the markings, and push it carefully into the socket until it locks into place.
– Plug the square plug of the connecting cable in the
socket at the back of the instrument.
– Check that both plugs fit tightly.
– Screw on the suitable test attachment for measure-
ments on curved surfaces if necessary.
– Switch the MIC 20 on. An icon in the top left corner
of the screen shows that a probe has been correctly connected:
Disconnecting the connecting cable
The round plug is provided with a bayonet lock, the square plug has two locking clips.
– Carefully pull the outer ring of the round plug on the
probe or on the impact device in order to release the lock, and pull off the plug.
– Press the two lateral clips of the square plug on the
instrument in order to release the lock, and pull off the plug.
Krautkramer MIC 20 Issue 02, 04/2005 3-7
Initial start-up
Connecting a probe or an impact device
Exchanging the test attachment
You can use special test attachments for testing on small parts and on uneven surfaces in order to achieve a better positioning of the probe or of the impact device.
– Slowly unscrew the currently used test attachment
counter-clockwise by hand.
– Place the required test attachment on carefully.
Make sure that the threads are not damaged by tilt­ing.
– Carefully screw on the test attachment clockwise
and by only using manual force all the way to the limit stop.
3.3 Turning on/off
Turning on
– Briefly press the button o to turn the instrument on.
The operating system will start, and some information, e.g. about the software, is briefly displayed on the screen. When the instrument is ready, you will see the graphic user interface.
Turning off
– Briefly press the button o to turn the instrument off.
A Attention:
You should always use the button erly turn the instrument off. In the case of an interrup­tion of the power supply (battery removal, unplugging of power plug), the instrument is not turned off properly, and data may be lost.
o in order to prop-
3-8 Issue 02, 04/2005 Krautkramer MIC 20
Turning on/off
Emergency-stop
If the instrument no longer shows any reaction, you can turn it off without saving the current data. Afterwards, you can turn it back on as usual.
Initial start-up
– To turn off, keep the button
the screen is blanked, and the instrument is turned off.
o pressed down until
Cold start
If the instrument no longer reacts and cannot be started properly, you can reset the settings with a cold start.
A Attention:
In a cold start, the files LastMeasure.mes and last.set are deleted so that the last measurement series and the saved calibration data may be lost.
– To switch on the instrument, briefly press
the button
– As soon as the start screen (blue background) ap-
pears, press the button
Krautkramer MIC 20 Issue 02, 04/2005 3-9
o.
.
3-10 Issue 02, 04/2005 Krautkramer MIC 20
Operation 4
Krautkramer MIC 20 Issue 02, 04/2005 4-1
Operation Fundamental principles of operation
4.1 Fundamental principles of operation
4-2 Issue 02, 04/2005 Krautkramer MIC 20
OperationFundamental principles of operation
Operator’s controls and displays
1 Temperature sensor (no operating function)
2 Select keys to select the main menus
3 Touch-sensitive screen (touch screen), for direct
operation of the graphic interface
4 Key to turn on and off
5 Red LED,
is lit in the case of alarms and if the preset threshold values for hardness measurements are exceeded
6 Green LED,
is lit with an active charging process, flashes quickly with a finished charging process, flashes slowly if the battery voltage or the tempera­ture are out of tolerances
7 Arrow keys for navigation in the menus, and for acti-
vation of functions
8 Select keys for submenus and functions
Display screen
The Krautkramer MIC 20 is equipped with a touch screen enabling a direct operation of the menus appear­ing on the screen.
The graphic user interface is adapted to the known Windows standards. The operation by means of a mouse (selecting and clicking) is in this case replaced with direct touch on the screen. A mouse pointer is therefore not necessary.
To select or to mark an element of the user interface, just touch the corresponding point of the screen briefly with your finger or with the pen provided for the instru­ment.
The pen is located in the holder at the instrument bot­tom.
A Attention:
Do not touch the touch screen with any hard or sharp­edged objects (e.g. ball-point pen or screw driver). The touch-sensitive surface may be severely damaged by them.
Krautkramer MIC 20 Issue 02, 04/2005 4-3
Operation Fundamental principles of operation
Do not apply any high pressure to the screen, the touch screen needs only slight pressure to react.
User interface
The 3 main menus Data, Config, and Measure are dis- played on the screen. The 3 main menus have different user interfaces.
The buttons as well as the option and other boxes shown on the user interface can be operated by directly touching the touch screen, or by pressing the corre­sponding key next to the screen.
H Note:
The contents of the menu differ in some details, de­pending on the fact whether a probe or an impact de­vice is connected to the MIC 20.
Main menu Measure Main menu Config
4-4 Issue 02, 04/2005 Krautkramer MIC 20
Main menu Data
Virtual keyboard
A virtual keyboard is available for text and value inputs and is automatically displayed if the corresponding input fields are activated.
The virtual keyboard can also be turned on and off by hand.
– Touch the button with the keyboard icon on it, next to
the battery icon. The virtual keyboard is displayed.
– Touch the broad bar above the displayed keys, keep
the pen pressed down, and move the keyboard with the pen pressed down.
– Touch the button with the keyboard icon on it once
again. The virtual keyboard is turned off again.
OperationFundamental principles of operation
Krautkramer MIC 20 Issue 02, 04/2005 4-5
Operation Fundamental principles of operation
Operation using the keypad or touch screen
You can use either the keys arranged around the screen to operate all menus and functions or, as an alternative, directly the touch screen.
At the beginning, use both alternatives in order to test the advantages of either one in practice.
In general, this manual only describes the direct opera­tion by means of the touch screen. If the operation al­ternative using the keyboard presents a special advan­tage, it is additionally mentioned at that point.
The keys and menus or functions are arranged in such a way that a clear assignment can always be seen.
Consequently, you’ll always have the following two al­ternatives:
– Touch the button of a function.
or
– Press the key next to (or below) the function.
Installing the instrument
The instrument has a prop-up lever stand at its bottom enabling you to install the instrument at varying angles.
You should always ensure that you have a good view of the screen in order to work as ergonomically as pos­sible and to avoid overstraining yourself unnecessarily.
– Fold out the lever stand, and let it lock into the re-
quired position.
– Install the instrument in such a way that you have a
good and non-dazzling view of it.
4-6 Issue 02, 04/2005 Krautkramer MIC 20
Probe and impact device
Operation
4.2 Probe and impact device
Guiding the probe
A Attention:
Make sure that the probe is connected before switching the MIC 20 on so that the instrument is automatically set to the UCI method.
Reliable measurement results can only be obtained if you know how to handle the probe correctly and safely.
Please always observe the following rules:
Avoid any rough placing of the probe on the test sur-
face in order not to damage the diamond.
Do not move the probe under load on the test material.
Pay attention to guiding the probe smoothly and
steadily using your both hands.
Do not carry out measurements twice at the same
test position. The distance between the test positions should be at least 3 mm.
H Note:
After placing the probe on the test surface, the mea­surement must be carried out within 3 seconds. Other­wise an error message will be given and you have to start the process again.
– Select the main menu Measure.
– If necessary, remove the rubber protection cap from
the probe.
– Screw on the conical test attachment for short-time
measurements.
– Hold the probe with your both hands in order to have
the best possible control during guidance and during impressing in the material surface.
– Place the probe vertically on the surface to be
tested. An icon at the top edge on the screen shows the contact of the probe with the test material:
Krautkramer MIC 20 Issue 02, 04/2005 4-7
Operation
Probe and impact device
– Hold the probe with one hand tightly enough so that it
doesn’t tilt and press it evenly and steadily against the surface with the other hand.
– Increase the pressure steadily until the probe’s spe-
cific load is attained and you hear an acoustic signal. The measured hardness value is displayed on the screen.
– Lift off the probe and place it again at another posi-
tion in order to carry out other measurements.
H Note:
If you have pressed the probe down up to the limit stop without any measured value being determined (no acoustic signal), check the following possible causes:
The load was applied too fast. Try again by increas-
ing the pressure more slowly.
The instrument has not been calibrated correctly
beforehand.
The hardness of the surface tested is beyond the
probe’s range of measurement.
Operating the impact device
A Attention:
Make sure that the impact device is connected before switching the MIC 20 on so that the instrument is auto­matically set to the rebound method.
The impact device Dyna G must only be used up to a hardness to be tested of max. 650 HB, otherwise the impact device may be destroyed.
Reliable measurement results are only obtained if you know how to handle the impact device correctly and safely.
You should always observe the following rules:
Always use an impact device suitable for the appli-
cation and equipped with the appropriate test attach­ment.
Do not carry out measurements twice at the same
test position. The distance between the test positions should be at least 3 mm.
4-8 Issue 02, 04/2005 Krautkramer MIC 20
Probe and impact device
Operation
Never load the impact device at the actual test posi-
tion (surface hardening) but somewhere else (e.g. at a table or another position of the test object).
Keep the following in mind when using the test at-
tachments Dyna 41 and Dyna 42 so that the impact device is correctly loaded: do not tension the impact device by pushing the load tube through „in the air” but by e.g. pressing one finger against it or by plac­ing the impact device on the test surface. After that, put the impact device at the actual test position.
– Select the main menu Measure.
– Tension the impact device by pushing the load tube
all the way through to the limit stop and by moving it slowly back to the initial position.
– Place the impact device vertically on the test sur-
face and press it slightly against it with one hand so that it doesn’t tilt.
– Use the index finger of the other hand to press the
release button. An acoustic signal is heard if the measurement was successful.
– Lift off the impact device. The measured hardness
value is displayed on the screen.
Release button
Load tube
Test attachment
Krautkramer MIC 20 Issue 02, 04/2005 4-9
Operation
Hardness testing
4.3 Hardness testing
A Attention:
In order to obtain correct measurement results, it is absolutely necessary to calibrate the MIC 20 to the test object’s material before starting any measurements.
You have to repeat the calibration process for every new material. You must likewise repeat the calibration whenever you change the test method.
The calibration is described in detail in the following chapter 4.4. As you have to carry out measurements for the calibration, please read the present chapter first in order to learn how to edit measurement series and view measurement results.
H Note:
When the MIC 20 is switched off, the instrument set­tings (including calibration) of the last measurement are saved and will again be available unchanged when the instrument is switched on the next time. Check the current settings prior to every measurement series.
Starting and closing a test series
If you use test series, you can end a test series any time, and start a new one. The statistical evaluations always refer to the currently active test series.
– Select the main menu Measure.
– Select one of the functions Curve, Histogr. or Sta-
tistics at the bottom edge of the display.
– Touch the button New in order to start a new test
series.
– Touch the button Delete in order to delete a test se-
ries that has possibly already been recorded. The readings are deleted, and a new test series is started.
You can now start measuring.
The processes of saving and editing test results are described later in this manual (ref. chapter 4.5).
4-10 Issue 02, 04/2005 Krautkramer MIC 20
Hardness testing
Operation
Viewing and evaluating test results
Various information is shown to you in the main menu Measure, depending on the selected view, e.g. the cur­rent hardness value, the list of previous measured val­ues, or the statistical distribution of measured values. Details depend on the current method of measurement. You can choose between three display view modes for measurements:
You have a choice between three display modes for the measurements:
Curve
Histogram
Statistics
H Note:
The different display modes have no effect on the pro­cess of measurement.
The current reading (s), the current statistical average (x), and the number of measurements (n) can be seen in all four display modes.
– Select the main menu Measure.
– Select one of the functions at the bottom edge of the
display in order to see the corresponding display mode.
Curve function
The measured hardness values are displayed as a curve in a diagram. The shape of the curve helps you with the evaluation of the test results.
Krautkramer MIC 20 Issue 02, 04/2005 4-11
Operation
Hardness testing
The horizontal green lines show the tolerance limit set­tings. Readings within the tolerance range are displayed in green, readings outside in red.
In addition to the diagram, other information is dis­played for every measured value:
Probe Test load of the probe or
type of the impact device
t Dwell time (only with UCI method)
CAL Calibration number (Off = standard material)
Mat. Selected material group
(only with rebound method)
Date Date of measurement
Time Hour of time of measurement
File File name when editing an already saved
measurement series
You can delete individual test results.
– Touch the required reading to select it.
– Touch the button Delete. The selected reading is
immediately deleted.
H Note:
You can also select the reading by means of the arrow keys to the right of the screen display.
4-12 Issue 02, 04/2005 Krautkramer MIC 20
Hardness testing
Operation
Histogr. function
The histogram shows the statistical distribution of the readings of the current test series.
Besides the graphics, additional statistical single val­ues are displayed (ref. chapter 7.4):
S Standard deviation (absolute and percentage)
R Range of spread (absolute and percentage)
cp Process capability index
cpk Critical process capability index
Statistics function
The Statistics function shows the test results of a test series in tabular form.
Krautkramer MIC 20 Issue 02, 04/2005 4-13
Operation
Hardness testing
Besides the table, additional statistical single values are displayed (ref. chapter 7.4):
S Standard deviation (absolute and percentage)
R Range of spread (absolute and percentage)
Min Minimum value
Max Maximum value
Test results situated within the set tolerance limits are shown in blue characters, test results outside the toler­ance are shown in red characters.
You can delete individual test results.
– Touch the required reading in order to select it.
– Touch the button Delete. The selected reading is
immediately deleted.
H Note:
You can also select the reading by means of the arrow keys to the right of the screen.
Conversion into other hardness scales
You can directly convert the test results obtained into other hardness scales.
A Attention:
Any illegal or unacceptable conversions may lead to serious errors in the interpretation of test results.
You should always pay attention to the information given in chapter 7.3 of this manual regarding the con­version of test results.
4-14 Issue 02, 04/2005 Krautkramer MIC 20
Hardness testing
Operation
H Note:
The list of hardness scale options is not automatically adapted to the test method or the impact device used. If your choice is not accepted by the instrument, the selected conversion is not possible. A corresponding note is displayed on the screen.
– Touch the abbreviation of the current hardness scale
(e.g. HV) at the item Average (x) in the right-hand screen section. An options menu is displayed.
– Touch the abbreviation of the required hardness
scale (e.g. HB). All values of the current measurement series are displayed in the new scale.
4.4 Calibration
A Attention:
In order to obtain correct measurement results, it is absolutely necessary to calibrate the MIC 20 to the test object’s material before starting any measurements.
You have to repeat the calibration for every new mate­rial. When changing the test method, you have to like­wise repeat the calibration procedure.
H Note:
If you want to measure on low-alloy or unalloyed steels, there’s no need to set the instrument because this ma­terial group is the default setting of the MIC 20.
When switching the MIC 20 off, the instrument settings (including calibration) of the last measurement are saved and will be available again unchanged when the instrument is switched on the next time. Check the current settings prior to every measurement series.
Krautkramer MIC 20 Issue 02, 04/2005 4-15
Operation
Calibration
Carrying out the calibration
The calibration procedure is basically identical for the UCI method and for the rebound method.
You only have to start by selecting one of the material groups stored in the instrument as an additional action with the rebound method.
You need a reference block made of the same material with known hardness for the calibration. During the cali­bration, you determine the calibration number that you need in other measurements on this material.
The calibration number is an auxiliary value that simpli­fies the subsequent change of the calibration setting. The calibration number has no physical reference what­soever and consequently no name either.
As you can save the calibration in the MIC 20, you only have to carry out the calibration once for a certain ma­terial.
Only with calibration for rebound method:
If you want to carry out measurements on high-alloy steels or on other materials, e.g. nonferrous metals,
you have to set the MIC 20 to the required material. To do this, you start by choosing the suitable material from the 9 existing ones. For precision measurements, you have to carry out a calibration afterwards for the adjustment.
– Touch the button Cal. The calibration menu is dis-
played.
– Touch the list box Material group. The list of the
stored material groups appears.
4-16 Issue 02, 04/2005 Krautkramer MIC 20
Calibration
Operation
– Touch the material group corresponding to the test
object. The list is closed again.
– Finally, touch the button OK to close the menu.
Notes on the material groups:
The MIC 20 makes 9 material groups available. How­ever, not all material groups are suitable for the different impact devices. Please observe the assignment in the table below:
Material group Impact device
ST - steel and cast steel D E G
AST - tool steel D E
SST - stainless steel D
GCI - gray cast iron D G
NCI - nodular cast iron D G
AL - aluminum cast alloys D
BRZ - bronze D
CU - wrought copper alloys D
Calibration for UCI and rebound method:
– Carry out a series of measurements consisting of
about 5 single measurements on your reference block.
– Touch the button Cal. The calibration menu is dis-
played.
The current statistical average of your measurements is displayed at the item Real value. However, this value does not correspond to the known, real hardness value because you have not yet calibrated the instrument.
You will now have to adapt the displayed value to the known hardness value of your reference block.
– Double-touch the item Target value in order to mark
the value displayed there.
BRS - brass D
Krautkramer MIC 20 Issue 02, 04/2005 4-17
Operation
Calibration
– Touch the button Select at the bottom edge of the
screen. The virtual keyboard is displayed, and you can input the required target value.
– Key in the value by touching the corresponding char-
acters on the keyboard.
– Finally, touch the displayed Return (ENTER) key. The
keyboard disappears again.
According to the input hardness value, the MIC 20 will automatically determine the calibration number which is immediately displayed at the item Cal. Number.
Having calibrated the MIC 20, you can carry out hardness tests as described above.
– Touch the button OK in order to close the calibration
menu. You will see the normal display screen again. An icon at the top edge of the screen shows that a calibration is active (calibration number non-0):
A Attention:
The calibration is active when the calibration menu is closed, however, it is not automatically saved. If you want to use a calibration again some time later, you have to save it under a new name (please see the fol­lowing section).
Saving calibration data
If you want to use a calibration several times (e.g. after switching the instrument off and on again, or after ex­changing the probe), you have to save it. To do this, you have to enter a name for the calibration.
H Notes:
You cannot edit or overwrite saved calibrations. In order to save any changes, you have to enter a new name.
Calibrations for the UCI method are independent of the probe, and consequently they apply to all UCI probes. Calibrations for the rebound method are dependent of the
4-18 Issue 02, 04/2005 Krautkramer MIC 20
Calibration
Operation
type of the impact device, and the letter of the impact device is therefore added to their name (D, G, or E).
– Touch the button Cal. The calibration menu is dis-
played and you will see the current calibration.
– Double-touch the item Name in order to mark the
name displayed there.
– Touch the button Select at the bottom edge of the
screen. The virtual keyboard is displayed, and you can key in the required name for the calibration.
– Finally, touch the button OK in order to close the
menu. The current calibration is now saved under the input name, and you can always restore it at a later date.
Deleting calibration data
You can always delete calibrations that are no longer needed.
– Touch the button Cal. The calibration menu is dis-
played.
– Touch the item List containing the saved calibration
data. The list appears.
– Touch the name of the required calibration.
– Touch the button Delete at the bottom edge of the
screen. The current calibration is deleted after con­firming the safety prompt. The following calibration in the list is now active.
– Finally touch the button OK in order to close the
menu.
Krautkramer MIC 20 Issue 02, 04/2005 4-19
Operation
Calibration
Restoring the standard calibration
If you want to reset the calibration to the standard or default setting again after calibrating to a special mate­rial, just choose the option STANDARD from the list of calibration data. This setting is factory-programmed and cannot be deleted.
– Touch the button Cal. The calibration menu is dis-
played.
– Touch the item List containing the saved calibration
data. The list appears.
– Touch the option STANDARD. The default setting is
restored.
– Finally, touch the button OK in order to close the
menu.
The icon for the active calibration is no longer shown at the top edge of the screen.
H Note:
You can always restore all other saved calibration data as well in exactly the same way by making your choice from the list.
4-20 Issue 02, 04/2005 Krautkramer MIC 20
Saving the test data
Operation
4.5 Saving the test data
A Attention:
Before starting a new test series, you can save the current test series to a file in the memory of the Krautkramer MIC 20. You can restore and view the data saved in this way any time later on.
You can interrupt and save test series any time by starting a new test series.
Depending on the configuration of the instrument, you are automatically prompted to save as soon as the last value that was set for a test series has been recorded.
– Select one of the functions Curve, Histogram or
Statistics at the bottom edge of the display.
– Touch the button New in order to start a new test
series. Three new buttons are displayed at the bot­tom edge of the display.
– If necessary, touch the button Edit if you want to
make some further changes to the current test se­ries (record other readings, delete reading or conver­sions). You will again see the display mode that was last selected.
– Touch the button Delete in order to delete the current
test series.
– Touch the button Save in order to save the current
test series. The window for selecting the directory and file name appears.
You can now select a directory and assign a file name as you’re used to do in common Windows applications.
H Note:
The selection of the directory and the assignment of the file name are not necessary if you activate Auto save to (please see chapter 4.7).
Krautkramer MIC 20 Issue 02, 04/2005 4-21
Operation
Saving the test data
Saving files
– Touch the name of a directory in order to select the
directory.
– Touch the text box at the bottom edge of the display.
The virtual keyboard is displayed, and you can key in the required file name.
– Enter a name by touching the corresponding charac-
ters on the keyboard.
– Finally, touch the displayed ENTER key. The key-
board disappears again.
– Touch the button Save. The test series is saved to
the currently selected directory using the specified name.
Deleting files or directories
A Attention:
When a directory is deleted, all files stored there are likewise deleted. You cannot cancel the deletion.
– Touch the name of a directory or a file in order to
select the directory or file.
– Touch the button Delete. A safety prompt appears.
– Touch the button Yes in order to confirm the deletion
process. The selected directory or the selected file is deleted.
4-22 Issue 02, 04/2005 Krautkramer MIC 20
Saving the test data
Operation
Opening and closing directories
– Double-touch a directory. The directory is opened.
or
– Briefly touch a directory in order to select it, and then
touch the button Open.
– Double-touch an open directory. The directory is
closed.
or
– Briefly touch an open directory in order to select it,
and then touch the button Close.
Creating new directories
– Touch the name of a directory in order to select the
directory.
– Touch the text box at the bottom edge of the display.
The virtual keyboard is displayed, and you can key in a name of a directory.
– Enter a name by touching the corresponding charac-
ters on the keyboard.
– Finally, touch the displayed ENTER key. The key-
board will again disappear.
– Touch the button NewDir. The new directory is cre-
ated in the currently selected directory using the specified name.
Krautkramer MIC 20 Issue 02, 04/2005 4-23
Operation
Saving the test data
Opening files
You can open saved files, and view the readings of a test series stored there.
– Touch the name of a file in order to select the file.
– Touch the button Open. The file is opened.
After opening the file, the main menu Measure will auto­matically appear. You can change between the different display modes, as well as view the saved data.
If you start a new test series, the open file is closed without editing it.
You can also edit the open file. If the edited data should be saved, you have to save the file afterwards using a new name.
Editing saved files
If you want to edit an already saved file, you have to save this file afterwards using a new name.
– Open the required file.
– If necessary, edit the required data in the main menu
Measure (e.g. deleting of single readings).
– Go to the main menu Data.
You can now enter a new file name as described above, and save the edited file.
H Note:
You cannot add any other readings to a saved file. The file LastMeasure.mes, where the current readings that have not yet been saved are automatically filed when the MIC 20 is switched off, is an exception. You can continue with the measurement series after opening this file the next time the instrument is switched on.
4-24 Issue 02, 04/2005 Krautkramer MIC 20
Saving the test data
Operation
File management using instrument keys
All main functions of file management, known to you from the Windows Explorer, are available to you in the main menu Data. Analogously to the operation using the mouse, the operation by means of the touch screen is the most convenient way with the Krautkramer MIC 20.
The operation using the keys arranged around the dis­play screen, however, only makes a limited functional­ity available.
You can always navigate in the file structure, as well as select files and directories using the arrow keys to the right of the display screen.
The keys below the display screen enable you to oper­ate the buttons appearing above each one of them, and carry out the corresponding functions (e.g. Delete, Open, or Close).
The access to the text box for keying in file or directory names, however, is not possible by means of the in­strument keys. A character can only be keyed in via the virtual keyboard.
4.6 Printing test reports
You can print out the results of a test series in the form of a test report. To do this, you have to save the test results to a file beforehand.
You have a choice between two types of reports:
Compact
Complete
The compact-type report contains the following data:
Date and name of the file
Information about the instrument and the probe or
impact device
Name of the operator
Remarks on the test series
All statistical evaluations referring to the test series
Data regarding the tolerance threshold settings
Krautkramer MIC 20 Issue 02, 04/2005 4-25
Operation
Printing test reports
H Note:
The compact-type report contains no single readings.
The complete-type report contains, in addition to the data of the compact-type report, the following:
graphic display of the hardness values in a diagram
(analogously to the display mode Curve)
list of hardness values in tabular form (analogously
to the display mode Statistics)
For both report types you can addionally select the options Color and Draft. The option Draft will accelerate the printing process with a slight loss of quality.
Prerequisites for printing
The following requirements must be met for the printout of test reports.
A PCL-compatible laser or ink-jet printer must be con­nected to the serial interface of the instrument. To con­nect the printer, you need a serial printer cable, or a serial-parallel converter (order no. 101 761).
H Note:
Information about the configuration of your model of printer you find in the documentation of the printer used.
Serial printer
– Connect the printer with the serial interface (9-pin
Sub-D-socket).
– Set the data transmission parameters on your printer
as follows:
1 start bit 1 stop bit 8 data bits no parity check software handshake on
– Ensure that the baud rate set in the printer agrees
with that selected in the MIC 20 (ref. following sec­tion).
4-26 Issue 02, 04/2005 Krautkramer MIC 20
Printing test reports
Operation
Parallel printer
– Connect the printer by the serial-parallel-cable with
the serial interface (9-pin Sub-D-socket).
– Ensure that the baud rate setting of the serial-parallel
converter agrees with that selected in the MIC 20 (ref. following section). The recommended baud rate is 4800.
Selecting and printing a test series
After connecting and configuring a compatible printer, you can select the required measured value file (*.mes), and print out the test report.
H Note:
Due to the baud rate of 4800, the printout is recom­mended in the draft mode.
– Go to the main menu Data.
– Touch the required measured value file in order to
select it.
– Touch the button Print. A dialog box appears.
– Touch the list boxes one after another, and select the
printer, the interface, and the mode of the report.
– Touch the button OK. The printing process starts
immediately.
Krautkramer MIC 20 Issue 02, 04/2005 4-27
Operation
Specimen test report complete
Printing test reports
4-28 Issue 02, 04/2005 Krautkramer MIC 20
Specimen test report compact
OperationPrinting test reports
Krautkramer MIC 20 Issue 02, 04/2005 4-29
Operation
Instrument configuration
4.7 Instrument configuration
In order to use the Krautkramer MIC 20 in an optimum way, you can configure it according to your require­ments. The configuration of the instrument comprises three areas:
Evaluation
Instrument
Info
H Note:
In the main menu Config, you have the additional but­ton Select at your disposal. This button enables you to alternatively carry out most of the settings using the instrument keys. The touch screen or the virtual key­board is only required for the input of a comment text if applicable.
You can use the key below the button Select e.g. to open list boxes in order to make the required choice there.
Evaluation parameters
The following describes the various options available to you for the evaluation and the representation of test results.
H Note:
The contents of the menu differ in some details, de­pending on the fact whether a probe or an impact de­vice is connected to the MIC 20.
– Select the submenu Evaluation in the main menu
Config.
4-30 Issue 02, 04/2005 Krautkramer MIC 20
Instrument configuration
Operation
Scale
The test result is displayed as HV value in the default setting. You can select another hardness scale as re­quired.
H Notes:
The list of the hardness scale options is not automati­cally adapted to the test method or the impact device used. If your choice is not accepted by the instrument, the selected conversion is not possible. A warning mes­sage may be displayed under certain circumstances.
If readings are beyond the range of values when con­verting into another hardness scale, they cannot be converted. These values are reset to zero, and they are not taken into consideration in the statistical evaluation according to the currently active scale.
– Touch the list box, and select the required hardness
scale.
Conversion
You can choose between the conversion tables accord­ing to DIN 50 150, according to ASTM E140, and the DynaMIC-specific conversion table for the conversion of hardness values.
H Note:
The list of the conversion table options is not automati­cally adapted to the test method or the impact device used. If your choice is not accepted by the instrument, the selected conversion is not possible. A warning mes­sage may be displayed under certain circumstances.
– Touch the list box, and select the required standard.
Krautkramer MIC 20 Issue 02, 04/2005 4-31
Operation
Instrument configuration
No of Measurements
You can specify the number of single measurements to be contained in a test series. After recording the last reading, a message is then automatically displayed on the screen prompting you to save the test series.
– Double-touch the text box No of Measurements in
order to select the current value.
– Touch the keyboard icon at the top edge of the dis-
play screen in order to display the virtual keyboard.
– Enter the required number, and touch the ENTER
key in order to make the keyboard disappear again.
– Touch the check box in order to activate the function.
If the function is activated, the box is ticked off.
H Note:
If the function is not activated, test series are not auto­matically closed. However, you can also save test se­ries which are not finished any time (ref. chapter 4.5).
Upper Threshold, Lower Threshold
You can enter a minimum value and a maximum value for measured hardness values. If a hardness value is outside these tolerance thresholds, an alarm is output, and the red LED at the instrument front panel is lit.
In addition, hardness values outside the tolerance thresholds are marked red in the main menus Measure, viz. in the display modes Curve and Statistics.
H Note:
If the input thresholds are beyond the range of values when changing to another hardness scale, they are automatically adjusted to the permissible maximum or minimum value. When changing back to the previous hardness scale, this may lead to threshold shifts.
– Double-touch the text box Upper Threshold or
Lower Threshold in order to select the current value.
– Touch the keyboard icon at the top edge of display
screen in order to have the virtual keyboard dis­played.
4-32 Issue 02, 04/2005 Krautkramer MIC 20
Instrument configuration
Operation
– Enter the required number, and touch the ENTER
key in order to make the keyboard disappear again.
– Touch the corresponding check box in order to acti-
vate the required function. If the function is acti­vated, the box is ticked off.
Material group (only with rebound method)
For dynamic measurements according to the rebound method, you have to choose one of the material groups stored in the MIC 20 in order to calibrate the instrument to the test material.
H Note:
Not all material groups can be selected for the different impact devices (D, E, or G). A corresponding note may be displayed if applicable. Please observe the assign­ment in the table below:
Material group Impact device
ST - steel and cast steel D E G
AST - tool steel D E
SST - stainless steel D
GCI - gray cast iron D G
NCI - nodular cast iron D G
AL - aluminum cast alloys D
BRS - brass D
BRZ - bronze D
CU - wrought copper alloys D
– Touch the list box and select the required material
group.
Krautkramer MIC 20 Issue 02, 04/2005 4-33
Operation
Instrument configuration
Dwell time (only with UCI method)
You can specify the dwell time of test load in seconds if required. This function is only useful in measurements by means of a support when using a special test at­tachment or motor probes (ref. chapter 2).
– Double-touch the text box Dwell time in order to
select the current value.
– Touch the keyboard icon at the top edge of display
screen in order to have the virtual keyboard dis­played.
– Enter the required number, and touch the ENTER
key in order to make the keyboard disappear again.
Auto save to
You can specify a directory where the measured value files are automatically filed during saving.
If you activate this option, you don’t have to select a directory and a file name by hand during every new saving process.
The name of the directory plus an appended numerical value, which is likewise increased by 1 with every new file, are then automatically assigned to the measured value files.
To start with, you have to select the required directory in the main menu Data. Chapter 4.5 describes how to proceed in the main menu Data.
– Go to the main menu Data.
– If necessary, create a new directory.
– Briefly touch the required directory in order to select it.
– Go to the main menu Config, and select the
submenu Evaluation.
The selected memory directory is displayed at the parameter Auto save to for verification purposes.
– Touch the check box to activate the Auto-save
option.
If the option is activated, the main menu Data is not opened any more during saving. After touching the but­ton Save, you can immediately start a new test series.
4-34 Issue 02, 04/2005 Krautkramer MIC 20
Instrument configuration
Operation
System settings
You can define various default settings for the Krautkramer MIC 20.
– Select the submenu Instrument in the main menu
Config.
Backlight
You can set the backlight of the display screen accord­ing to a scale from 0 to 9, with 9 being the value for the brightest backlight.
– Double-touch the text box in order to select the cur-
rent value.
– Touch the keyboard icon at the top edge of display
screen in order to have the virtual keyboard displayed.
– Enter the required value, and touch the ENTER key
in order to make the keyboard disappear again.
H Note:
To save energy, the display screen is automatically slightly dimmed after a short while if you’re not operat­ing the instrument.
After a further period of time without any operation, the display screen is completely turned off.
As soon as you carry out an action, the display screen switches back to the normal backlight. You cannot influ­ence this energy-saving function.
Krautkramer MIC 20 Issue 02, 04/2005 4-35
Operation
Instrument configuration
Contrast (only with MIC 20 equipped with CSTN screen)
You can adjust the contrast of the display screen on a scale from 0 to 9. The value for the highest contrast is 9.
– Double-touch the text box in order to mark the cur-
rent value.
– Touch the keyboard icon at the top edge of the
screen in order to display the virtual keyboard.
– Key in the required value and touch the Return (EN-
TER) key in order to make the keyboard disappear again.
Automatic turn-off
You can set a time in minutes after which the instru­ment is automatically turned off if there is no operation of the keys, the probe, or of the touch screen.
The current instrument configuration and the current test series are saved, and they are available again when the instrument is started the next time.
After the automatic turn-off, you have to turn the instru­ment on again by means of the On/Off key.
If you enter the value 0 (zero), this function is deacti­vated, and the instrument is no longer automatically turned off.
– Double-touch the text box in order to select the cur-
rent value.
– Touch the keyboard icon at the top edge of display
screen in order to have the virtual keyboard dis­played.
– Enter the required number, and touch the ENTER
key in order to make the keyboard disappear again.
4-36 Issue 02, 04/2005 Krautkramer MIC 20
Instrument configuration
Operation
Beep
You can activate an acoustic signal to acknowledge different actions. The actions acknowledged with an acoustic signal include the successful recording of a reading, or errors during recording of readings.
– Touch the list box, and select the required option.
Language
You can select the dialog language for the operation of the Krautkramer MIC 20. All texts of the user interface are displayed in the selected language. Reports are printed in the selected language as well.
– Touch the list box, and select the required language.
Date
Set the correct date, and make sure that the set date always corresponds to the current date. Date and time are saved together with the readings and used for print­ing test reports.
– Touch the list box containing the currently set date. A
calendar is displayed.
– Use the arrow icons in the header area of the calen-
dar to select the required month.
– Touch the required date. The calendar disappears,
and you will see the new date in the list box.
H Note:
In case of a power interruption (e.g. when exchanging the batteries) longer than 5 minutes the date and time setting will get lost.
Krautkramer MIC 20 Issue 02, 04/2005 4-37
Operation
Instrument configuration
Time
Set the correct hour of time, and make sure that the set time always corresponds to the current time. Date and time are saved together with the readings and used for printing test reports.
The time is indicated in hours, minutes, and seconds.
– Touch the number of hours in the box Time in order
to select it.
– Touch the arrow keys to the right of the box in order
to change the selected value.
– Set the minutes and, if required, the seconds in the
same way.
Network address (TCP/IP)
If required, specify an IP address for the instrument here. During data transfer operations to a computer using the UltraDAT program, this address must agree with the IP address specified in the program.
– Double-touch the text box in order to mark the cur-
rent address.
– Touch the keyboard icon at the top edge of the
screen in order to display the virtual keyboard.
– Key in the required IP address and touch the Return
(ENTER) key in order to make the keyboard disap­pear again.
4-38 Issue 02, 04/2005 Krautkramer MIC 20
Instrument configuration
Operation
System information
The submenu Info provides you with various informa- tion about the instrument and the probe. These data are only meant for information, and they cannot be edited.
In the lower section of the window, you can key in a short comment or information text to be saved together with the current test series, and also to be printed out in the test report.
– Select the submenu Info in the main menu Config.
Probe Type
Type of the currently connected probe (H = handheld probe, M = motor probe) or of the connected impact device (D, G, or E).
Probe Serial Number
Serial number of the currently connected probe.
Load (N) (only with UCI method)
Nominal test load of the currently connected probe in N.
Evaluation Software
Version number of the current evaluation software.
System Software Version
Version number of the current system software.
Krautkramer MIC 20 Issue 02, 04/2005 4-39
Operation
Instrument configuration
System Hardware Version
Version number of the current system hardware.
Total Measurements
Total number of measurements carried out using the currently connected probe.
Last Service
Date of the last instrument check by the authorized service of GE Inspection Technologies.
Info
This is where you can key in information or comments to be saved together with the current test series (e.g. name of operator, name of test object, and data on the test position).
H Note:
These data only apply to the current test series. As soon as a new test series is started, the Info text box is cleared again.
– Double-touch the text box in order to select the cur-
rent text.
– Touch the keyboard icon at the top edge of display
screen in order to have the virtual keyboard dis­played.
– Key in the required text, and touch the ENTER key in
order to make the keyboard disappear again.
4-40 Issue 02, 04/2005 Krautkramer MIC 20
Saving and loading instrument parameters
Operation
4.8 Saving and loading instrument parameters
You can save the current specific parameters of the instrument. The current settings regarding the measure­ments are saved with a test series.
You can load the saved instrument parameters any time in order to quickly configure the instrument.
Saving instrument parameters
– Go to the main menu Config, and select the
submenu Instrument.
– Touch the button Save. A window for the input of the
file name appears.
– Briefly touch the text box File name. The virtual
keyboard is displayed.
– Key in the required file name, and touch the ENTER
key in order to make the keyboard disappear again. The current instrument parameters are saved.
Loading and deleting instrument parameters
– Go to the main menu Config, and select the
submenu Instrument.
– Touch the button Load. A window for the selection of
a parameter file appears.
– Touch the name of the required file in order to select it.
– Touch the button Load. The selected file is loaded
and immediately active.
– Touch the button Delete. The selected file is deleted.
Krautkramer MIC 20 Issue 02, 04/2005 4-41
Operation
Functional tests
4.9 Functional tests
Functional test for UCI method
Checking the Vickers diamond optically
Check the Vickers diamond under a microscope from time to time.
Ensure that the pyramid form of the diamond is not damaged.
Checking the measuring accuracy
All UCI probes are calibrated using hardness reference plates certified for the corresponding test load by the German material testing institute “Materialprüfungs­anstalt MPA NRW”. Every hardness reference plate shows natural variations that also affect the deviation of the value measured by the corresponding probe from the plate’s value.
Under well defined conditions, as e.g. with the measure­ment in the test support MIC 222 on Krautkramer hard­ness reference plates with acoustic coupling (thin oil film or ultrasonic couplant ZG between the reference
plate and the support), a deviation from the plate’s value amounting to ±3.6 % of the average from 5 mea­surements is still permissible (with a range of variation of max. 5 % relative to the average).
Individual deviations from the plate’s value may occur in freehanded measurements, depending on how the probe is handled. However, they should not amount to a deviation of more than 5 % from the plate’s value in 10 measurements.
Make yourself thoroughly familiar with the handling of the probe and practice the handling of the probe using the measurements on the hardness reference plates until you obtain stable measured values.
Any sudden changes indicate a defective diamond or a de-calibration of the probe.
Have the instrument and the probe checked by the GE Inspection Technologies service if any major deviations occur.
– Carry out 3 to 5 measurements on the hardness ref-
erence plate. Ensure that the distance between the measuring positions is at least 3 mm.
4-42 Issue 02, 04/2005 Krautkramer MIC 20
Functional tests
Operation
– Read the average and compare it with the reference
value of the hardness reference plate.
Functional test for rebound method
Check the functioning of impact device and of MIC 20 before using them for the first time, and afterwards from time to time (after 1,000 impacts in each case at the latest) by measuring the hardness value HL on the cor­responding hardness reference block.
Minor deviations from the reference value of ±5 HL can be compensated for by the calibration function. If any major deviations occur, it is necessary to replace the spherical carbide metal tip.
– Carry out 3 to 5 measurements on the hardness ref-
erence plate. Ensure that the distance between the measuring positions is at least 3 mm.
– Read the average, and compare it with the reference
value of the hardness reference plate.
Have the instrument and the impact device checked by the GE Inspection Technologies service if any major deviations occur.
Krautkramer MIC 20 Issue 02, 04/2005 4-43
Operation
Calibrating the touch screen
4.10 Calibrating the touch screen
If the operation of the touch screen frequently fails, you can recalibrate the touch-sensitive screen.
H Note:
The calibration of the touch screen does not affect measurements and test results in any way.
– Go to the main menu Config, and select the
submenu Info.
– Touch the button Cal. Touch. A window for the cali-
bration of the display screen appears.
– Press the center of the displayed crosshairs as ex-
actly as possible. The crosshairs are moved to an­other position.
– Press the center of the crosshairs as exactly as
possible one more time.
– Repeat the action until the calibration is finished, and
a new information message is displayed.
– Briefly touch the touch screen once in order to con-
firm the new calibration, and to transfer it.
If you don’t transfer the calibration, the old setting is used again after 30 seconds.
The calibration is finished, and you will again see the main menu Config after a short while.
4-44 Issue 02, 04/2005 Krautkramer MIC 20
Troubleshooting
Operation
4.11 Troubleshooting
After being turned on, the Krautkramer MIC 20 carries out an automatic system self-check. In addition, the system also monitors itself during the operation.
Error messages
If any system or operating errors occur, they are indi­cated by a corresponding error message on the display screen. Generally, the error message also contains information about the cause of the error, and about the further procedure.
Errors
Error Possible remedy
Instrument does not react when turned on. Check power supply and battery charge status.
Instrument no longer reacts when operated. Keep the On/Off key pressed down long enough
(approx. 5 s) to turn off the instrument. Current data are not saved.
Krautkramer MIC 20 Issue 02, 04/2005 4-45
4-46 Issue 02, 04/2005 Krautkramer MIC 20
Care and maintenance 5
Krautkramer MIC 20 Issue 02, 04/2005 5-1
Care and maintenance Care
5.1 Care
A Attention:
Never use water for cleaning! Water penetrating inside may cause severe damages to the instrument and the probe.
Do not use any solvents! Paint coating and plastic parts may become brittle or be damaged by them.
Care of instrument
Clean the instrument and the connecting cable for the probe or the impact device at regular intervals by only using a dry cloth.
Do not use any sharp-edged objects or other tools for cleaning the display screen. The touch-sensitive foil may be damaged by them.
Care of batteries
Capacity and life of batteries mainly depend on their correct handling. You should therefore pay attention to the tips below. You should charge the batteries in the following cases:
before the initial start-up,
after a storage period of 3 months or longer,
after frequent partial discharge.
H Note:
Used or defective batteries are special waste and must be disposed of as provided by the law!
5-2 Issue 02, 04/2005 Krautkramer MIC 20
Care and maintenanceCare
Charging the batteries
A Attention:
Never try to charge alkaline batteries (explosion hazard)!
You can only charge the NiMH battery pack MIC 20-BAT in the instrument itself.
You can only charge single NiMH or NiCad batteries outside the Krautkramer MIC 20 using a battery charger approved for this task.
If necessary, read the operating manual of the battery charger, and pay special attention to the safety precau­tions recommended there.
The battery pack MIC 20-BAT is always automatically charged in the Krautkramer MIC 20 if the instrument is supplied with power via the power supply unit and turned off at the same time. If the instrument is turned on, the connection to the battery compartment is al­ways interrupted if the instrument is supplied with power via the power supply unit.
The green LED on the instrument front panel is lit dur­ing the charging process. If the green LED flashes quickly, the charging process is finished.
H Note:
Used or defective batteries are special waste, and must be disposed of as provided by the law!
Krautkramer MIC 20 Issue 02, 04/2005 5-3
Care and maintenance Maintenance
5.2 Maintenance
As a general principle, no maintenance work is required on the Krautkramer MIC 20.
A Attention:
Any repair work must only be carried out by the autho­rized service of GE Inspection Technologies.
Cleaning the probe
Clean the probe by just using a dry cloth.
Cleaning the impact device
A Attention:
Never use oil for cleaning the impact device!
The impact device must be cleaned after approx. 1,000 measurements or if any incorrect readings are obtained.
– Unscrew the test attachment and remove the impact
body from the guide tube.
– Clean the test attachment, impact body, and test tip
using a dry cloth.
– Clean the guide tube using a cleaning brush.
5-4 Issue 02, 04/2005 Krautkramer MIC 20
Interfaces and data transfer 6
Krautkramer MIC 20 Issue 02, 04/2005 6-1
Interfaces and data transfer Interfaces
6.1 Interfaces
The Krautkramer MIC 20 has two bidirectional inter­faces that you can use to transfer the data to a printer or to a computer.
RS232 interface Ethernet interface
6-2 Issue 02, 04/2005 Krautkramer MIC 20
Interfaces and data transferInterfaces
Serial interface RS232
Plug Pin Signal Level Direction
1 DCD RS232 Input
2 RXD RS232 Input
3 TXD RS232 Output
4 DTR RS232 Output
Sub-D 5 GND 0 V Input
6 DSR RS232 Input
7 RTS RS232 Output
8 CTS RS232 Input
9––
H Note:
You need a serial printer cable or a serial-parallel con­verter for the connection of a printer (Order number 101 761).
Ethernet interface
Plug Pin Signal Level Direction
1 TXD + 3.3 V Output
2 TXD – 3.3 V Output
3 RXD + 3.3 V Input
4––
RJ45 5
6 RXD – 3.3 V Input
7––
8––
The signals are galvanically isolated.
H Note:
You need a crossed patch cable with RJ45 plugs for the connection to a computer (Order number 101 785).
Krautkramer MIC 20 Issue 02, 04/2005 6-3
Interfaces and data transfer Data transfer to a printer
6.2 Data transfer to a printer
You can print out the results of a test series in the form of a test report. To do this, you have to save the test results to a file beforehand (please see chapter 4 of this manual).
Requirements for printing
The following requirements have to be met for printing test reports.
A PCL-compatible laser or ink-jet printer must be con­nected with the serial interface of the instrument. To connect the printer, you need a serial printer cable or a serial-parallel converter (Order number 101 761).
H Note:
You will find information about the configuration of your printer type in the documentation of the printer used.
Transfer parameters
The transfer parameters must be set as follows for a serial printer:
1 start bit
1 stop bit
8 data bits
no parity check
software handshake on
In the case of a parallel printer using the serial-parallel converter, the baud rate setting on the converter must correspond to the baud rate selected in the MIC 20. A baud rate of 4800 is recommended.
6-4 Issue 02, 04/2005 Krautkramer MIC 20
6.3 Data transfer to a computer
You can transfer the results of saved test series to a computer, and save them to an Excel file there.
You need the UltraDAT program for the data transfer (ref. chapter 2).
UltraDAT enables the direct access to the file structure of the instrument and has a large variety of preview options. Besides the convenient selection of the test series to be exported, additional information can be added to the test series, e.g. test object, operator’s name, or additional remarks.
H Note:
The data transfer to a computer is carried out either via the Ethernet interface. To do this, you need a corre­sponding data cable (ref. section 6.1).
For more details on the data transfer, please read the documentation for the UltraDAT program (quick refer­ence guide and online help).
Interfaces and data transferData transfer to a computer
Krautkramer MIC 20 Issue 02, 04/2005 6-5
6-6 Issue 02, 04/2005 Krautkramer MIC 20
Hardness testing method 7
Krautkramer MIC 20 Issue 02, 04/2005 7-1
The hardness testing methods The UCI method
7.1 The UCI method
The following section gives you helpful information about the UCI hardness testing method of the MIC 20.
Unlike conventional low-load hardness testers, the MIC 20 does not evaluate the Vickers indentation by means of a microscope but by electronic means using the UCI method.
This brings about an advantage with regard to the opti­cal evaluation because a high reproducibility of the results is ensured even with small test indentations.
The Vickers diamond is attached to the tip of a round metal rod. This metal rod is excited to longitudinal oscil­lations in its resonant frequency of approx. 78 kHz.
If the Vickers diamond and the test material come into contact with each other, the resonant frequency changes. This happens as a function of the size of the indentation surface of the Vickers diamond. This is in turn the measure for the hardness of the material tested.
Resonant frequencies can be measured quite accu­rately. The UCI method is therefore especially suitable for a considerable simplification and acceleration of the evaluation of the Vickers indentations, and conse­quently of the complete test process.
As a result of this, there are two more benefits:
The hardness measurement is carried out under load
(no impairment of the measurement by elastic resilience).
Basis for the hardness measurement is the surface
of the indentation and not the length of the indenta­tion diagonals. The measurement is consequently less influenced by the surface roughness. The measurement of burnished or gunmetal finish sur­faces is also possible.
H Note:
In the UCI method, the measured value depends on the elastic properties of the material (Young’s modulus or modulus of elasticity, Poisson’s ratio) so that the instru­ment has to be calibrated for hardness measurements on different materials (ref. chapter 4.4).
7-2 Issue 02, 04/2005 Krautkramer MIC 20
Loading...