Waygate Technologies DM5E Operating Manual

GE
Sensing & Inspection Technologies Ultrasonic
DM5E
Operating Manual
P/N 1259551 Rev. 1
February 2009
DM5E
Corrosion Thickness Gauge
Operating Manual
P/N 1259551 Rev. 1 February 2009
GESensingInspection.com
©2009 General Electric Company. All rights reserved.
Contents
Chapter 1. General Information
1.1 Supplying Power to the DM5E. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Powering the Instrument ON and OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3 Key Features of the DM5E . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3.1 DM5E Corrosion Thickness Measurement Base Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.4 Contents of this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Chapter 2. Understanding the Keypad, Menu System & Displays
2.1 Keypad Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.2 Interpreting the Display Screens. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.3 Working with the Configuration Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
DM5E Operating Manual iii
Contents
Chapter 3. Setting Up the DM5E
3.1 Connecting a Probe and Loading a Setup File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.2 Configuring the Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.2.1 Setting the Instrument Gain. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.2.2 Setting the Update Rate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.2.3 Specifying the Nominal Thickness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.3 Instrument Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.4 Setting the Maximum and Minimum Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.5 B-SCAN Display Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.6 Locking and Unlocking Instrument the Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Chapter 4. Measuring Thickness
4.1 Selecting the Displayed View Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.2 Normal Measurement View Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.3 Min Scan and Max Scan Measurement View Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
4.4 Differential/Rate-of-Reduction Measurement View Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
4.5 B-SCAN Measurement View Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
4.6 Dual Multi Measurement View Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
iv DM5E Operating Manual
Contents
Chapter 5. Using the Optional Data Recorder
5.1 Creating a New Data Recorder File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
5.2 Recalling and Erasing Stored Data Recorder Files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5.2.1 Recalling a Data Recorder File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5.2.2 Erasing a Data Recorder File. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
5.3 Recording Thickness Measurements in Data Recorder Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
5.3.1 Storing Minimum of B-Scan in Data Recorder Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
5.3.2 Navigating Through Data Recorder Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Chapter 6. I/O Features
6.1 Connecting the DM5E to a PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Chapter 7. Specifications
7.1 Instrument Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
7.2 Data Recorder Option Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
7.3 DM5E Probe/Transducer Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
DM5E Operating Manual v
Contents
Chapter 8. Maintenance
8.1 Care of the instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
8.2 Proper Cable Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
8.3 Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Appendix A. Supplemental Information
A.1 Resetting the Operating Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
A.2 Firmware Updates. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
A.3 Web Site Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
A.4 EMC Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
A.5 Manufacturing & Customer Service Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Appendix B. Regulatory Compliance
B.1 Waste Electrical and Electronic Equipment (WEEE) Directive. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
B.2 Battery Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
B.2.1 What do the Markings Mean? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
B.2.2 The Risks and Your Role in Reducing Them . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
B.3 EN15317 Technical Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
vi DM5E Operating Manual
Preface
Important Notice
The following information must be read and understood by any user of a GE Sensing & Inspection Technologies ultrasonic thickness gauge. Failure to follow these instructions can lead to errors in thickness measurements or other test results. Decisions based on erroneous results can, in turn, lead to property damage, personal injury or death.
General Warnings
Proper use of ultrasonic test equipment requires three essential elements:
Selection of the correct test equipment
Knowledge of the specific “test application requirements”
Training on the part of the instrument operator
This operating manual provides instruction in the basic setup and operation of the thickness gauge. There are, however, additional factors which affect the use of ultrasonic test equipment. Specific information regarding these additional factors is beyond the scope of this manual. The operator should refer to textbooks on the subject of ultrasonic testing for more detailed information.
DM5E Operating Manual vii
Preface
Operator Training
Operators must receive adequate training before using ultrasonic test equipment. Operators must be trained in general ultrasonic testing procedures and in the setup and performance required by a particular test. Operators must understand:
Sound wave propagation theory
Effects of the velocity of sound of the test material
Behavior of the sound wave where two different materials are in contact
Areas covered by the sound beam
More specific information about operator training, qualification, certification, and test specifications is available from various technical societies, industry groups, and government agencies.
viii DM5E Operating Manual
Preface
Testing Limitations
In ultrasonic testing, information is obtained only from within the limits of the sound beam. Operators must exercise great caution in making inferences about the test material outside the limits of the sound beam. For example, when testing large materials it may be impossible or impractical to inspect the entire test piece.
When a less-than-complete inspection is to be performed, the operator must be shown the specific areas to inspect. Inferences about the condition of areas not inspected, based on data from the evaluated areas, should only be attempted by personnel fully trained in applicable statistical and probability techniques. In particular, materials subject to erosion or corrosion, in which conditions can vary significantly in any given area, should only be evaluated by fully trained and experienced operators.
Sound beams reflect from the first interior surface encountered. Because of part geometry and overlapped flaws or overlapped surfaces, thickness gauges may measure the distance to an internal flaw rather than to the back wall of the material. Operators must take steps to ensure that the entire thickness of the test material is being examined.
DM5E Operating Manual ix
Preface
Ultrasonic Thickness Measurement Critical Operating Procedures
The following operating procedures must be observed by all users of ultrasonic thickness gauges in order to minimize errors in test results.
1. Calibration of Sound Velocity
The principle of operation of an ultrasonic thickness gauge is that the instrument measures the time of flight of an ultrasonic pulse through the test piece and multiplies this time by the velocity of sound in the material. Thickness measuring error is minimized by ensuring that the sound velocity to which the instrument is calibrated is the sound velocity of the material being tested. Actual sound velocities in materials often vary significantly from the values found in published tables. In all cases, best results are obtained if the instrument is calibrated on a velocity reference block made from the same material as the test piece; this block should be flat and smooth and as thick as the maximum thickness of the test piece.
Operators should also be aware that the sound velocity may not be constant in the material being tested; heat treating, for example, can cause significant changes in sound velocity. This must be considered when evaluating the accuracy of the thickness provided by this instrument. Instruments should always be calibrated before testing, and the calibration should be checked after testing, to minimize testing errors.
2. Probe Zero Procedure
Before performing calibration with a contact probe, the probe zero procedure must be performed first. If the zero procedure is not performed or if it is performed improperly, inaccurate thickness readings will result.
x DM5E Operating Manual
Preface
Ultrasonic Thickness Measurement Critical Operating Procedures (cont.)
3. Effects of Temperature on Calibration
Temperature variations change the sound velocity of materials and transducer delay lines and, therefore, calibrations. All calibrations should be performed on-site and with test blocks at the same temperature as the test piece, to minimize errors due to temperature variations.
4. Transducer Selection
The transducer used in testing must be in good condition, without noticeable wear of the front surface. Badly worn transducers will have a reduced effective measuring range. The specified range of the transducer must include the complete range of thicknesses to be tested. The temperature of the material to be tested must be within the transducer's temperature range.
5. Use of Couplants
Operators must be familiar with the use of ultrasonic couplants. Testing skills must be developed so that couplant is used and applied in a consistent manner to minimize variations in couplant layer thickness and errors in test results. Calibration and actual testing should be performed under similar coupling conditions, using a minimum amount of couplant and applying consistent pressure on the transducer.
DM5E Operating Manual xi
Preface
Ultrasonic Thickness Measurement Critical Operating Procedures (cont.)
6. Doubling
Ultrasonic thickness gauges will, under certain conditions, display readings which are twice (or in some cases, three times) the actual material thickness being measured. This effect, commonly known as “doubling,” can occur below the minimum specified range of the transducer. If the transducer being used is worn, doubling is possible at a thickness greater than the minimum of the specified range.
When using a new transducer, any reading which is less than twice the minimum specified range of the transducer may be a “doubled” reading, and the thickness of the material being tested should be verified by the use of other methods. If the transducer shows any sign of wear, doubling may occur at a second echo, or other echo signal combinations may produce a readable signal. The instrument reading and apparent thickness are up to about twice the actual value, resulting in a thickness greater than twice the minimum of the specified range. This thickness should be determined by calibrating the instrument/transducer combination on reference blocks that represent the complete range of possible thicknesses that may be encountered in testing. This is particularly important when a test piece is being ultrasonically measured for the first time or in any case where the history of thickness of the test specimen is unknown.
xii DM5E Operating Manual
Preface
Safety Information
IMPORTANT: The DM5E is an instrument for materials testing. Any use for medical applications or other purposes is not allowed. The
DM5E may only be used in industrial environments.
IMPORTANT: For battery operation of the DM5E, two “AA” size batteries are needed. You should only use the products recommended by
us for battery operation. Use of well-known battery brands is highly recommended.
IMPORTANT: Disposal of all components of the DM5E and its batteries must be done in compliance with all applicable regulations. See
Appendix B, Regulatory Compliance, for detailed information.
Software
According to the current state of the art, software is never completely free from errors. Before using any software-controlled test equipment, please make sure that the required functions operate perfectly in the intended combination.
DM5E Operating Manual xiii
Preface
Defects/Errors and Exceptional Stresses
If you have reason to believe that safe operation of your DM5E is no longer possible, you must disconnect the instrument and secure it against unintentional reconnection. Remove the batteries if necessary.
Safe operation is no longer possible if conditions such as the following exist:
The instrument shows visible damages
The instrument no longer operates perfectly
The instrument has been subjected to prolonged storage under adverse conditions such as exceptional temperatures and/or especially
high air humidity, or corrosive environmental conditions
The instrument has been subjected to heavy stresses during transportation
Service
Every effort has been made to provide you with a reliable product. However, should service become necessary, GE Sensing & Inspection Technologies, has established a number of Factory Trained Service Centers. For the location of the nearest facility, see the back cover of this manual.
xiv DM5E Operating Manual
Chapter 1. General Information
Chapter 1. General Information
The DM5E is an ultrasonic corrosion thickness measurement device. The instrument is available in three versions:
DM5E Basic
DM5E with Dual Multi functionality
DM5E DL with Dual Multi and Data Recorder
This chapter introduces the key features of the DM5E, and the next chapter explains the keypad functions and display-menu contents. Closely reviewing the material in these two chapters will help you make better use of the more detailed information found in subsequent chapters of this manual.
1.1 Supplying Power to the DM5E
The instrument is powered by two “AA” size batteries, providing up to approximately 100 hours of service life. To install the batteries, remove the battery-compartment cover as shown in Figure 1 on page 2.
IMPORTANT: You cannot charge or power the DM5E through its USB port.
DM5E Operating Manual 1
Chapter 1. General Information
1.1 Supplying Power to the DM5E (cont.)
Rear of Instrument
Install 2 “AA” batteries in compartment as shown
Figure 1: DM5E Connections and Batteries (shown with battery cover removed)
2 DM5E Operating Manual
Top of Instrument
Probe Cable Connection
I/O Device Connection
(mini USB)
Chapter 1. General Information
1.1 Supplying Power to the DM5E (cont.)
When the battery indicator is in the last quarter, as indicated by the display-screen symbol , replace the batteries as soon as possible. The DM5E automatically shuts OFF when the batteries are too weak for reliable operation. However, your settings are saved and restored when the instrument is turned ON again. When testing in remote locations, always carry spare batteries.
1.2 Powering the Instrument ON and OFF
Power the instrument ON by pressing and holding until the instrument display activates. Press and hold while the instrument is ON to turn it OFF.
1.3 Key Features of the DM5E
Large hollow/filled thickness digits
Alarm bar graph
Standard and custom parameter setups
Supports new set of DA5xx probes
Alphanumeric naming of data files (DM5E DL)
DM5E Operating Manual 3
Chapter 1. General Information
1.3 Key Features of the DM5E (cont.)
Lockout feature with password access
Large LCD display with backlight and adjustable contrast
Importing of data recorder files directly into Microsoft Excel (DM5E DL)
Approximately 100 hours of operation with 2 “AA alkaline batteries
Easy-to-use, single-level menu system
Multiple display screen languages
User-selectable measurement resolution to 0.001 inch (0.01 mm)
Normal (thickness), MinScan, MaxScan, and Differential/Rate of Reduction Measurement modes are standard
Timed B-Scan
Optional Dual Multi function (DM5E and DM5E DL)
Optional data recorder (DM5E DL) stores up to 50,000 thickness measurements in data recorder file
Field upgrades available
Compatible with UltraMATE
©
and UltraMATE LITE© software programs
4 DM5E Operating Manual
Chapter 1. General Information
1.3.1 DM5E Corrosion Thickness Measurement Base Instrument
1.3.1a Contents of the DM5E Base Instrument
DM5E instrument
2 “AA” batteries
Plastic carrying case
Couplant sample
Upgrade Utility Program on CD-ROM (requires mini-USB to PC cable)
Operating manual in CD-ROM
Operating instruction card
Certificate of conformity
Note: A mini-USB to PC cable is only included with the DM5EDL version. This cable can be purchased separately for use with the
DM5E Basic or DM5E version.
DM5E Operating Manual 5
Chapter 1. General Information
1.3.1b Instrument Versions
DM5E Basic
DM5E with Dual Multi functionality
DM5E DL with Dual Multi and Data Recorder
1.4 Contents of this Manual
This DM5E Operating Manual is divided into eight chapters and one appendix. All chapters apply to all instrument versions, except for two functions:
Data Recorder, which is described in Chapter 5, applies only to the DM5E DL version
Dual Multi mode applies only to the DM5E with Dual Multi functionality and to the DM5E DL
Note: Any DM5E can be retrofitted to incorporate the above functions.
6 DM5E Operating Manual
Chapter 2. Understanding the Keypad, Menu System & Displays
Chapter 2. Understanding the Keypad, Menu System & Displays
The DM5E's display, keypad, and functional commands are easy to interpret and use. In this chapter, you'll find a brief explanation of all display and keypad features. You'll also find references to manual sections where more detailed information is available.
Note: The DM5E's display screen contents varies depending on the optional features installed and the operational settings selected. The
display screens shown in this chapter are typical of those found in all the instrument configurations.
2.1 Keypad Features
The instrument's keypad includes a combination of dedicated keys and virtual function keys. The key functions, as illustrated in Figure 2 on page 8, are as follows:
Arrow Keys - Use to adjust the value of the selected parameter and for on-screen navigation.
Function Keys - Use to activate any functional control shown in the display directly above the key.
Cal/On Key - Press once to launch the calibration process or to make on screen selections. Press and hold to power the instrument
ON and OFF.
Mode Key - Press to activate the standard display modes and any optional display modes that are installed.
DM5E Operating Manual 7
Chapter 2. Understanding the Keypad, Menu System & Displays
2.1 Keypad Features (cont.)
Select a probe setup
Select, adjust and confirm
operating settings
Thickness measurement display
Figure 2: DM5E Keypad and Key Functions
Press to activate standard display modes (shown here) and the optional display modes (if installed). Optional display modes include Data Recorder Contents.
Press to activate the functional control (when present) above the key.
Press to launch the calibration process or make on-screen selections. Press and hold to power the instrument ON or OFF.
Press to adjust the value of the selected parameter and for on-screen navigation.
8 DM5E Operating Manual
Chapter 2. Understanding the Keypad, Menu System & Displays
2.2 Interpreting the Display Screens
This section describes the layout of the DM5E's primary display modes including:
Measurement Display Mode - Indicates measured thickness, contains status icons, displays data recorder location and file name (when
this option is installed and activated). Figure 3 on page 10 shows four representative display screens. The display's appearance varies based on the installed instrument options as well as instrument display settings. Additional measurement view modes are available when the Dual Multi or Data Recorder options are installed.
Probe Setup Display Mode - Allows the user to select a standard preloaded instrument setup, each of which is matched to a specific
probe model (see Figure 4 on page 11).
File Display Mode - Allows the user to create and store thickness readings in data recorder files. This feature is only available when the
data recorder option is installed and activated (see Figure 5 on page 12). When installed and activated, the data recorder option allows users to create data recorder files, store thickness readings, and navigate through file contents.
Configuration Display Mode - Instrument controls are accessed through this screen. The controls listed on the configuration display
screen vary depending on the installed options (see Figure 6 on page 13). The contents of the Configuration display depends on the instrument's installed-option configuration.
DM5E Operating Manual 9
Chapter 2. Understanding the Keypad, Menu System & Displays
2.2 Interpreting the Display Screens (cont.)
Figure 3: Measurement Display Mode
10 DM5E Operating Manual
2.2 Interpreting the Display Screens (cont.)
Chapter 2. Understanding the Keypad, Menu System & Displays
Figure 4: Probe Setup Display Mode
DM5E Operating Manual 11
Chapter 2. Understanding the Keypad, Menu System & Displays
2.2 Interpreting the Display Screens (cont.)
Figure 5: File Display Mode
12 DM5E Operating Manual
2.2 Interpreting the Display Screens (cont.)
Chapter 2. Understanding the Keypad, Menu System & Displays
Figure 6: Configuration Display Mode
DM5E Operating Manual 13
Chapter 2. Understanding the Keypad, Menu System & Displays
2.3 Working with the Configuration Display
By repeatedly pressing , you can access the Configuration display and you can adjust the following DM5E settings:
VIEW MODE - In base-model units this parameter offers five settings. See Section 4.1 on page 36 to select the view to be displayed.
DATA RECORDER - When the optional data recorder is installed, setting this parameter to ON activates the file recording capability and
the data recorder file display.
MINIMUM ALARM - Enables and sets the minimum alarm value (see Section 3.4 on page 30).
MAXIMUM ALARM - Enables and sets the maximum alarm value (see Section 3.4 on page 30).
UNITS - Sets the units of measurement to inches or millimeters (see Section 3.2 on page 21).
NOMINAL THICKNESS - Sets the nominal thickness value that is then used to calculate and the differential and rate of reduction
measurement values when selected by the VIEW MODE parameter (see Section 3.2.3 on page 25).
RESOLUTION - Controls the format (number of decimal places) of the thickness measurement that is displayed and stored in a Data
Recorder file.
UPDATE RATE - Sets the rate at which the measurement reading is updated (see Section 3.2.2 on page 24).
14 DM5E Operating Manual
Chapter 2. Understanding the Keypad, Menu System & Displays
2.3 Working with the Configuration Display (cont.)
BACKLIGHT - Turns the display backlight feature to ON, OFF, or AUTO (AUTO switches backlighting ON whenever a key is pressed or
the probe is connected).
CONTRAST - Select this control and then press or to adjust the display contrast level.
LANGUAGE - Sets the instrument display language (see Section 3.2 on page 21).
BSCAN MINIMUM - Sets the minimum value that can be drawn when BSCAN is processing (see Section 3.2.1 on page 23).
BSCAN MAXIMUM - Sets the maximum value that can be drawn when BSCAN is processing (see Section 3.2.1 on page 23).
DUAL MULTI - Used to switch between Dual Multi and IP test modes, when Dual Multi option is allowed (see Section 3.2 on page 21).
READING HOLD - Can be set to hold the last measurement value when the probe is uncoupled in measurement mode (see Section 3.2 on
page 21).
DM5E Operating Manual 15
Chapter 2. Understanding the Keypad, Menu System & Displays
2.3 Working with the Configuration Display (cont.)
TIME - Used to set the current time (see Section 3.2 on page 21).
DATE - Used to set the current date (see Section 3.2 on page 21).
GAIN - Used to select a different level of gain (LOW/HIGH/AUTO/DEFAULT) (see Section 3.2.1 on page 23).
RADIX - Selects a period (.) or comma (,) to be used as a decimal point (see Section 3.2 on page 21).
AUTO POWER DOWN - When set to ON, this control conserves battery life by powering the instrument OFF if no key presses or
measurements occur for the set number of minutes. When set to OFF, the instrument is only powered OFF when is pressed and held.
16 DM5E Operating Manual
Loading...
+ 70 hidden pages