WAGNER L622 Owner's Manual

Page 1
WAGNER MODEL L622
"DIGITAL RECORDING"
HAND-HELD MOISTURE METER
OWNER'S MANUAL
DOCUMENT NO: 500-62201-001 REV: M
Page 2
500-62201-001 Rev. M Page 2 of 33
Copyright Notice
Wagner Model L622 “Digital Recording” Moisture Meter Owner’s Manual
© Copyright 1996-2011 Wagner Meters. All Rights Reserved Worldwide. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language in any form by any means without the express written permission of Wagner Meters.
Technical Support
Should you require support, please consult your printed documentation to resolve your problem. If you are still experiencing difficulty, you may contact a Wagner Technical Service representative during normal business hours— 7:30 a.m. to 4:00 p.m. Pacific Standard Time, Monday through Friday. Voice: (541) 582-0541 Fax: (541) 582-4138 E-Mail: support@wagnermeters.com Mail: Technical Services Department Wagner Meters 326 Pine Grove Road Rogue River, OR 97537
Page 3
500-62201-001 Rev. M Page 3 of 33
Table of Contents
FCC Compliance Statement..............................................4
L622 Digital Recording Hand-Held Moisture Meter...........5
L622 Specifications...........................................................5
Getting Started..................................................................6
Turning on the L622..........................................................7
Readings Mode.................................................................7
Storing Readings...............................................................7
Changing Groups..............................................................8
Activating Main Menu........................................................8
Locale Menu......................................................................9
Changing Language..........................................................9
Changing Date Format......................................................10
Changing Decimal Format.................................................11
Changing Species.............................................................12
Editing Specific Gravity .....................................................12
Editing Species Name.......................................................12
Mean/Std Statistics............................................................13
High/Low Statistics............................................................13
Clear Menu........................................................................14
Clear Group.......................................................................15
Clear Readings..................................................................15
Clear All Groups................................................................15
Printing Reports ................................................................16
Group Report ....................................................................17
Group Summary Report....................................................17
Readings Report................................................................17
Sample Group Report........................................................18
Sample Group Summary Report .......................................19
Sample Readings Report ..................................................20
Calibration.........................................................................21
Checking Calibration .........................................................21
Backlight............................................................................22
Battery...............................................................................22
Restore Factory................................................................. 22
Technical Notes.................................................................23
Determining Specific Gravity .............................................24
Factory Species List..........................................................25
Commentary On Specific Gravity ......................................26
Connecting to a Serial Printer............................................27
DB25P to RJ11 Adapter Pin Assignments.........................27
DB9S to RJ11 Adapter Pin Assignments...........................27
Question and Answers ......................................................28
Warranty............................................................................32
Repair Service................................................................... 33
Page 4
500-62201-001 Rev. M Page 4 of 33
FCC Compliance Statement
This equipment has been tested and found to comply within the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in residential installations.
This equipment generates, uses and can radiate radio frequency energy, and if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation.
If this equipment does cause interference to radio or television equipment reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna
• Move the equipment away from the receiver
• Plug the equipment into an outlet on a circuit different from that
to which the receiver powered
• If necessary, the user should consult the dealer or an
experienced radio/television technician for additional suggestions.
CAUTION: Only equipment certified to comply with Class B (computer input/output devices, terminals, printers, etc.) should be attached to this equipment.
Finally, any changes or modifications to the equipment by the user not expressly approved by the grantee or manufacturer could void the users authority to operate such equipment.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and (2) This device must accept any interference received, including
interference that may cause undesired operation.
Canadian Department of Communications
Compliance Statement
This Class B digital apparatus meets all requirements of the Canadian Interference Causing Equipment Regulations.
Avis de conformité aux normes du ministère des
Communications du Canada
Cet appareil numerique de la classe B respecte toutes les exigences du Reglement sur le material broilleur du Canada.
Page 5
500-62201-001 Rev. M Page 5 of 33
L622 Digital Recording Moisture Meter
Wagner introduces the L622 Digital Recording Hand-Held Moisture Meter. This new model replaces the proven L612 model instrument, providing improved moisture measurement accuracy in the 15%+ moisture content range. As with the L612 model, the L622 utilizes Wagner’s Intellisense™ technology to truly read the moisture IN the wood, not ON the wood.
• Stores and displays up to 33 user-selected species
adjustments by name.
• Stores up to 5000 separate moisture readings.
• Stores readings in up to 200 groups.
• Built-in real time clock/calendar.
• Send reports directly to a printer or data to a
computer for advanced analysis using supplied Stat­Pak software.
Using non-volatile memory, stored readings are retained even when the batteries are removed. A backlit LCD display features a 2-line, 32-character display. Function buttons and scroll arrows operate a menu system that guides the user step by step through the selecting, storing and reporting process. The meter has auto-shut­off to extend battery life.
Like all Wagner Hand-Held Moisture Meters, moisture readings are virtually unaffected by wood temperature in the normal operating range outside of a kiln environment. Refer to the stack probe manual for more information about measuring wood in kilns. Contact Wagner technical support for guidelines when wood is below freezing.
L622 Specifications
Physical Size (in): L 8.5 x W 4 x H 3.75 (mm) L 216 x W 102 x H 95
Scanning Area (in): 2.5 x 2.5 (mm) 63.5 x 63.5
Scanning Depth (in): 1.0 in. (25.4 mm) Weight: 18 oz. (510 g) Power: 4 AA Alkaline Batteries Battery Life: ~8 hours of continual use*. Real Time Clock: Records both date and time Auto Power Shut Down: After 1 minute of non-use. Moisture Measurement Range: 5 to 30% (dependent on
species setting) Specify Gravity Range: 0.30 to 0.75 Meter Operating Temperature: 30° to 120°F Comm Port: RS-232 Serial Interface Probe Port: L712/L722 Sensor Probe
Interface Patent Numbers: 5,486,815 and 5,621,391
*Battery life is significantly reduced when using the display backlight. It is recommended to use rechargeable Nickel-Cadmium or Nickel­Metal Hydride batteries when extended backlight usage is planned. The Ni-Cd or Ni-MH batteries and charger may usually be purchased at most electronics supply stores. Be sure to use a battery charger that has a battery conditioner and an auto shut-off feature to avoid over charging of the batteries.
Page 6
500-62201-001 Rev. M Page 6 of 33
Getting Started
This sections informs you about
Turning on the L622
Moisture readings
Storing a reading
What is a group
Viewing statistics
The menu system
Changing species
Before you begin, make sure your L622 has a fresh set of batteries.
Turning on the L622
Turn on the L622 by pressing the STORE button for one second. The display will momentarily show the model, software version, and revision numbers. Your L622 is now ready to take moisture readings.
Moisture readings
Take moisture readings by pressing the bottom of the L622 to the wood surface. The first line of the display shows the species for which the L622 is calibrated. The second line shows the reading in percent moisture content.
Storing a reading
While pressing the bottom of the L622 to the wood surface at the location of interest, press the STORE button to save that reading in a group.
What is a group
A group is a set of readings that belong together. You determine which readings belong together. Switch to another group by first pressing the GROUP button. Press the ARROW buttons to select a different number. Press the STORE button to activate the selection. The active group number is shown in the bottom right corner of the display.
Viewing statistics
Statistical values such as mean, standard deviation, highest reading, and lowest reading are calculated on individual groups. They may be accessed for viewing by using the menu system.
The menu system
For example, to view the mean and standard deviation values for the active group, press the MENU button to activate the main menu. Press the ARROW button to scroll through the menu until the menu item Mean/Std is displayed. Press the STORE button to activate the menu item.
Changing species
If the L622 is not set to the species you are measuring, change it using the menu system. See the section “Factory Species List” for a complete list of available species. If your species is not listed, you may customize a species as outlined in the section “Changing Species.”
Page 7
500-62201-001 Rev. M Page 7 of 33
Turning on the L622
Readings Mode
Storing Readings
Operation
To turn on the L622, press the
STORE
button for one second. The display will briefly show the Model, Version, and Revision numbers.
The L622 will turn off after 1 minute of inactivity. Inactivity means no change in moisture readings or buttons pressed.
The L622 takes continuous percent moisture content (%MC) readings and puts them on the display. When a reading is below 5%, the first line shows the current species and group field label. The second line shows the current reading and the active group number. When a reading is above 5%, the first line becomes a bar graph representing the current reading.
When storing a valid reading, the second line will display the message Storing along with the sample number.
When attempting to store an invalid reading, the second line will display message Too Low along with the last sample number.
Display
WAGNER L622 V xx.xx Rxx.xx
Douglas Fir GRP .0 % 0
Storing 14 or Too Low 13
STORE
To turn the meter on, press the
STORE
button for one second.
Press the
STORE button to store the
current reading into the active group.
GROUP
Press the
GROUP button to activate
the change group function.
MENU
Press the
MENU button to activate the
main menu.
ARROWS
Refer to the Factory Default section.
Page 8
500-62201-001 Rev. M Page 8 of 33
Changing Groups
Activating Main Menu
Main Menu Selection
Operation
Press the
GROUP button to activate
the change group function. Use the
ARROW buttons to select the
desired group number. Press the
STORE button to activate the
change. The group number range is from 0 to
199.
Press the
MENU button to activate the
main menu. Press the
ARROW buttons to scroll
through the menu. Press the
STORE button to activate the
selected menu item.
Locale Species Mean/Std High/Low Clear Print Calibration Backlight Battery
Display
Group 1
Menu Language
STORE
Press the
STORE button to save the
change and return to the readings mode
Press the
STORE button to activate the
selected menu item.
GROUP
Press the
GROUP button to cancel the
change and return to the readings mode
Press the
GROUP button to return to
the readings mode
MENU
Press the
MENU button to cancel the
change and return to the readings mode
Press the
MENU button to activate the
selected menu item.
ARROWS
Press the
ARROW buttons to
increment or decrement the group number.
Press the
ARROW buttons to scroll
through the menu.
Page 9
500-62201-001 Rev. M Page 9 of 33
Locale Menu
Locale Menu Selection
Changing Language
Operation
Enter the main menu and activate the Locale item.
Press the
ARROW buttons to scroll
through the Locale menu. Press the
STORE button to activate the
selected menu item.
Language Date Decimal
Enter the Locale menu and activate the Language item.
Press the
ARROW buttons to scroll
through the Language menu. The languages available are dependent on your locale.
Press the
STORE button to activate the
language for the meter text.
Display
Locale Language
Language English
STORE
Press the
STORE button to activate the
selected menu item.
Press the
STORE button to activate the
language for the meter text.
GROUP
Press the
GROUP button to return to
the readings mode.
Press the
GROUP button to return to
the readings mode.
MENU
Press the
MENU button to activate the
selected menu item.
Press the
MENU button to activate the
language for the meter text.
ARROWS
Press the
ARROW buttons to scroll
through the menu.
Press the
ARROW buttons to scroll
through the menu.
Page 10
500-62201-001 Rev. M Page 10 of 33
Changing Date Format
Date Menu Selection
Operation
Enter the Locale menu and activate the Date item.
Press the
ARROW buttons to scroll
through the Date menu. Press the
STORE button to activate the
selected menu item.
MM/DD/YY DD/MM/YY
Selecting the MM/DD/YY item will format the date in the month/day/year order. Selecting the DD/MM/YY item will format the date in the day/month/year order.
Display
Date MM/DD/YY
STORE
Press the
STORE button to activate the
selected menu item.
GROUP
Press the
GROUP button to return to
the readings mode.
MENU
Press the
MENU button to activate the
selected menu item.
ARROWS
Press the
ARROW buttons to scroll
through the menu.
Page 11
500-62201-001 Rev. M Page 11 of 33
Changing Decimal Format
Decimal Menu Selection
Operation
Enter the Locale menu and activate the Decimal item.
Press the
ARROW buttons to scroll
through the Numbers menu. Press the
STORE button to activate the
selected menu item.
Period Comma
Selecting the
Period format will use
the period ‘.’ character as the decimal separator. Selecting the
Comma format will use
the comma ‘,’ character as the decimal separator.
Display
Decimal Period
STORE
Press the
STORE button to activate the
selected menu item.
GROUP
Press the
GROUP button to return to
the readings mode.
MENU
Press the
MENU button to activate the
selected menu item.
ARROWS
Press the
ARROW buttons to scroll
through the menu.
Page 12
500-62201-001 Rev. M Page 12 of 33
Changing Species
Editing Specific Gravity
Editing Species Name
Operation
Enter the main menu and activate the Species item.
Use the
ARROW buttons to scroll
through the species list (See the section “Factory Species List”).
Press the
STORE button to activate the
selected species.
Enter the Edit SG mode through the Species item.
Use the
ARROW buttons to change the
SG number. Press the
STORE button to save and
activate changes.
Enter the Edit Name mode through the Species item and Edit SG mode.
Use the
ARROW buttons to change the
alpha character. Press the
MENU button to move the
edit cursor to the right. Press the
STORE button to save and
activate changes.
Display
Species SG Walnut,Black.41
Edit SG Walnut,Black.45
Edit Name Walnut,Black.45
STORE
Press the
STORE button to activate the
selected species.
Press the
STORE button to save and
activate changes.
Press the
STORE button to save and
activate changes.
GROUP
Press the
GROUP button to return to
the readings mode.
Press the
GROUP button to cancel
changes and return to the readings mode.
Press the
GROUP button to cancel
changes and return to the readings mode.
MENU
Press the
MENU button to activate the
Edit SG mode.
Press the
MENU button to activate the
Edit Name mode.
Press the
MENU button to move the
edit cursor one character to the right. That character will be cleared.
ARROWS
Press the
ARROW buttons to scroll
through the species list.
Press the
ARROW buttons to
increment or decrement the SG number.
Press the
ARROW buttons to cycle
through the alphabet.
Page 13
500-62201-001 Rev. M Page 13 of 33
Mean/Std Statistics
High/Low Statistics
Operation
Enter the main menu and activate the Mean/Std item.
Use the
ARROW buttons to scroll
through the group statistics. Only groups that contain readings will be shown.
Press the
STORE button to return to
the readings mode.
Enter the main menu and activate the High/Low item.
Use the
ARROW buttons to scroll
through the group statistics. Only groups that contain readings will be shown.
Press the
STORE button to return to
the readings mode.
Display
Mean Std GRP
8.3 4.7 0
High Low GRP
11.0 6.1 1
STORE
Press the
STORE button to return to
the readings mode.
Press the
STORE button to return to
the readings mode.
GROUP
Press the
GROUP button to advance to
the next valid group.
Press the
GROUP button to advance to
the next valid group.
MENU
Press the
MENU button to return to the
readings mode.
Press the
MENU button to return to the
readings mode.
ARROWS
Press the
ARROW buttons to
increment or decrement the group number.
Press the
ARROW buttons to
increment or decrement the group number.
Page 14
500-62201-001 Rev. M Page 14 of 33
Clear Menu
Clear Menu Selection
Operation
Enter the main menu and activate the Clear menu.
Press the
ARROW buttons to scroll
through the Clear menu. Press the
STORE button to activate the
selected item.
Group Readings All Groups
Display
Clear Group
STORE
Press the
STORE button to activate the
selected item.
GROUP
Press the
GROUP button to return to
the readings mode.
MENU
Press the
MENU button to activate the
selected item.
ARROWS
Press the
ARROW buttons to scroll
through the menu.
Page 15
500-62201-001 Rev. M Page 15 of 33
Clear Group
Clear Readings
Clear All Groups
Operation
Enter the Clear menu and activate the Group item.
Press the
ARROW buttons to select the
group to clear. Only groups that contain readings will be shown.
Press the
STORE button to clear the
selected group.
Enter the Clear menu and activate the Readings item.
Press the
ARROW buttons to select the
reading to clear. Press the
STORE button to clear the
selected reading.
Enter the Clear menu and activate the All Groups item.
Press the
STORE button to clear all
groups. Press any other button to cancel
operation and return to the readings mode.
Display
Group 0
Reading GRP
17.6 % 10 21
All Groups Yes = STORE
STORE
Press the
STORE button to clear the
selected group.
Press the
STORE button to clear the
selected reading.
Press the
STORE button to clear all
groups.
GROUP
Press the
GROUP button to advance to
the next valid group.
Press the
GROUP button to advance to
the next valid group.
Press the
GROUP button to return to
the readings mode.
MENU
Press the
MENU button to cancel and
to return to the readings mode.
Press the
MENU button to return to the
readings mode.
Press the
MENU button to return to the
readings mode.
ARROWS
Press the
ARROW buttons to
increment or decrement the group number.
Press the
ARROW buttons to
increment or decrement reading number.
Press the
ARROW button to return to
the readings mode.
Page 16
500-62201-001 Rev. M Page 16 of 33
Printing Reports
Print Menu Selection
Operation
Enter the main menu and activate the Print menu.
Press the
ARROW buttons to scroll
through the Print menu. Press the
STORE button to activate the
menu item.
Group Summary Readings
Display
Print Group
STORE
Press the
STORE button to activate the
selected menu item.
GROUP
Press the
GROUP button to return to
the readings mode.
MENU
Press the
MENU button to activate the
selected menu item.
ARROWS
Press the
ARROW buttons to
increment or decrement the menu.
Page 17
500-62201-001 Rev. M Page 17 of 33
Group Report
Group Summary Report
Readings Report
Operation
Enter the Print menu and activate the Group item.
Press the
ARROW buttons to select the
group to print. Press the
STORE button to send the
group report to the printer.
Enter the Print menu and activate the Summary item.
Press the
STORE button to send the
report to the printer.
Enter the Print menu and activate the Readings item.
Press the
ARROW buttons to select the
group to print. Press the
STORE button to send the
readings report to the printer.
Display
Group 0
Print Summary
Group 0
STORE
Press the
STORE button to send the
report to the printer.
Press the
STORE button to send the
report to the printer.
Press the
STORE button to send the
report to the printer.
GROUP
Press the
GROUP button to advance to
the next valid group.
Press the
GROUP button to return to
the readings mode.
Press the
GROUP button to advance to
the next valid group.
MENU
Press the
MENU button to cancel and
return to the readings mode.
Press the
MENU button to send the
report to the printer.
Press the
MENU button to cancel and
return to the readings mode.
ARROWS
Press the
ARROW buttons to
increment or decrement the group number.
Press the
ARROW buttons to
increment or decrement the menu.
Press the
ARROW buttons to
increment or decrement the group number.
Page 18
500-62201-001 Rev. M Page 18 of 33
Sample Group Report
L622 Stat-Pak 15:06:08 07/14/10 Group Report
Company Name
Meter #: 1 First: 10:10:07 07/14/10 Last: 10:18:14 07/14/10
Group Readings Mean Std High Low Species SG 0 32 11.9 2.3 15.1 5.6 Douglas Fir .50
% Readings %Total Histogram < 0 .0 6 2 6.2 ******** 8 1 3.1 **** 10 6 18.7 ************************ 12 11 34.3 ********************************************* 14 11 34.3 ********************************************* 16 1 3.1 **** 18 0 .0 20 0 .0 22 0 .0 24 0 .0 26 0 .0 > 0 .0
Page 19
500-62201-001 Rev. M Page 19 of 33
Sample Group Summary Report
L622 Stat-Pak 15:07:39 07/14/10 Summary Report
Company Name
Meter #: 1 First: 10:10:07 07/14/10 Last: 10:23:43 07/14/10
Group Readings Mean Std High Low Species SG 0 32 11.9 2.3 15.1 5.6 Douglas Fir .50 7 36 14.3 4.0 24.0 6.8 Douglas Fir .50 11 35 13.2 3.0 19.2 7.5 Douglas Fir .50 199 31 12.8 3.4 22.3 7.9 Douglas Fir .50
---------------------------------------------------------------------------­ 4 134 13.1 3.3 24.0 5.6 Summary
% Readings %Total Histogram < 0 .0 6 3 2.2 *** 8 9 6.7 *********** 10 21 15.6 ************************** 12 36 26.8 ********************************************* 14 31 23.1 ************************************** 16 16 11.9 ******************** 18 10 7.4 ************ 20 4 2.9 ***** 22 3 2.2 *** 24 1 .7 * 26 0 .0 > 0 .0
Sample Readings Report
Page 20
500-62201-001 Rev. M Page 20 of 33
L622 Stat-Pak 15:09:32 07/14/10 Readings Report
Company Name
Meter #: 1 First: 10:10:07 07/14/10 Last: 10:18:14 07/14/10
Group Readings Mean Std High Low Species SG 0 32 11.9 2.3 15.1 5.6 Douglas Fir .50
# % # % # % # % # % 1 12.3 2 12.3 3 12.6 4 13.9 5 15.1 6 14.8 7 14.4 8 13.0 9 12.4 10 12.2 11 11.3 12 10.6 13 10.3 14 9.7 15 9.4 16 10.1 17 10.9 18 11.2 19 11.6 20 11.9 21 12.4 22 13.1 23 14.2 24 14.1 25 12.3 26 13.9 27 14.1 28 13.7 29 14.2 30 8.4 31 5.6 32 6.3
Page 21
500-62201-001 Rev. M Page 21 of 33
Calibration
Operation
Enter the main menu and activate the Calibration item.
Display
Calibration xx.x %
STORE
Press the
STORE button to return to
the readings mode.
GROUP
Press the
GROUP button to return to
the readings mode.
MENU
Press the
MENU button to return to the
readings mode.
ARROWS
Press the
ARROW button to return to
the readings mode.
Checking Calibration
Calibration is factory set. It may be checked by using a
Wagner Calibration Verification Block (CVB) (P/N: 840­60130-003). Use only to verify L622 factory calibration is within tolerance limits. If calibration is out of tolerance, return L622 to factory or repair depot for re-calibration. Never adjust L622 to this device. Protect CVB from direct sunlight. Store the CVB in a cool dry place. Follow these steps for checking calibration.
1. Enter the Calibration verification mode using the
MENU button.
2. Hold the L622 in mid-air so the bottom is several feet
from any object. The correct reading for air is .0 %MC. If the reading is outside the range, -2 to 2 %MC, the L622 is out of calibration.
3. Place the CVB, rubber feet down, on a metal surface:
such as a file cabinet, a piece of sheet metal, etc.
4. Orientate the long axis of the L622 with the long axis
of the CVB. Press the L622 sensor in the center of the CVB and observe the reading. The correct reading is
20.0 %MC. If the reading is outside the range, 18.0 to
22.0 %MC, the L622 is out of calibration.
Warning—There are two (2) hole plugs located on either side of the
L622 handle. Do not remove these plugs because the adjustments below them are factory set and should not be adjusted by the customer. If adjustment is attempted, the unit will not be in calibration and will require factory re-calibration.
Page 22
500-62201-001 Rev. M Page 22 of 33
Backlight
Battery
Restore Factory
Operation
Enter the main menu and activate the Backlight item.
Press the
STORE button to change state of
the backlight to ON or OFF.
Enter the main menu and activate the Battery item.
When a low battery condition occurs, the battery voltage bar graph will be replaced with the message Too Low. Also, when in the readings mode, GRP will be replaced with BAT.
Life-time for alkaline batteries is ~8 hours (continuous use, backlight OFF).
Pressing the
ARROW buttons at the same
time while in the readings mode will clear memory and restore the L622 to factory settings. Warning, you will lose all changes made to the species list.
Press the
STORE button to restore factory
settings. Press the
GROUP or MENU button to
cancel operation and return to the readings mode.
Display
Backlight ON
xxxxxx
4.0 Volts
Factory Yes = STORE
STORE
Press the
STORE button to change state of
the backlight.
Press the
STORE button to return to the
readings mode.
Press the
STORE button to restore factory
settings.
GROUP
Press the
GROUP button to return to the
readings mode.
Press the
GROUP button to return to the
readings mode.
Press the
GROUP button to cancel the
operation and return to the readings mode.
MENU
Press the
MENU button to return to the
readings mode.
Press the
MENU button to return to the
readings mode.
Press the
MENU button to cancel the
operation and return to the readings mode.
ARROWS
Press the
ARROW buttons to return to the
readings mode.
Press the
ARROW buttons to return to
the readings mode.
No function.
Page 23
500-62201-001 Rev. M Page 23 of 33
Technical Notes
Clock/Calendar—The L622 is contains a real time clock
and calendar. The clock/calendar is set through the Stat­Pak program. All reports use the clock/calendar to show the date and time a report is printed, and Group and Reading reports show when the first and last readings where taken.
Company Name—Your company name may appear on the reports. The company name is set through the Stat­Pak program.
Readings and Groups—The L622 can store up to 5000 readings in up to 200 groups. Groups are numbered 0 to 199 and can be selected at random. There is no limit to the number of readings per group as long as the total readings of all groups do not exceed 5000.
Automatic Menu Escape—When in the menu system and 15-seconds of no key presses, the L622 will automatically return to the Readings mode ready to store the next reading in the current group.
Repeat Key—If a key (button) is held down for more than one (1) second, then the key pressed is repeated at a rate of one (1) repeat each second. After 5 repeats the rate doubles and after an additional 5 repeats the rate doubles again. Use this feature to scroll through a long list of species or readings.
Page 24
500-62201-001 Rev. M Page 24 of 33
Determining Specific Gravity
The L622 calibration is based on an equation that relates measurement correction to the specific gravity of a given species. Species of wood in different regions may have a specific gravity that differs from that which is published in wood drying handbooks. If you do not know the species of the wood you are using, or the specific gravity differs from the handbook because of a different growing region, then use the following procedure for determining specific gravity.
1. Select a sample of wood with all edges being true. Carefully measure the dimensions of your sample using a caliper. You will need the length, width, and thickness.
2. Convert these measurements into feet or meters (ft. or m).
3. Carefully measure the weight of your sample.
4. Convert the weight to pounds or kilograms (lb. or kg).
5. Calculate specific gravity (SG).
An English units illustration:
Length: 10 in. converted to feet is 10 in./ 12 = 0.833 ft. Width: 7.5 in. converted to feet is 7.5 in. / 12 = 0.625 ft. Thickness: 1.5 in. converted to feet is 1.5 in./ 12 = 0.125 ft. Weight: 20 oz. converted to lb. is 20 oz. / 16 = 1.25 lb. Volume: L x W x T: 0.833 x 0.625 x 0.125 = 0.065 ft
3
. SG = (Weight / Volume) / Density of Water SG = (1.25 lb. / 0.065 ft
3
) / 62.34 lb./ ft3 = 0.31
A metric units illustration:
Length: 254 mm converted to meters is = 0.254 m. Width: 190 mm converted to meters is = 0.190 m. Thickness: 38 mm converted to meters is = 0.038 m. Mass: 565 converted to kg is = 0.565 kg. Volume: L x W x T: 0. 254 x 0. 190 x 0. 038 = 0.00183 m
3
. SG = (Mass / Volume) / Density of Water SG = (0.565 kg / 0.00183 m
3
) / 1000 kg/m3 = 0.31
In order to ensure that the value obtained for the specific gravity is statistically significant, a number of pieces must be measured and the average determined. Use this value of specific gravity for your species of wood.
Page 25
500-62201-001 Rev. M Page 25 of 33
Factory Species List
The L622 has a factory default of 32 species. The species name or specific gravity (SG) value may be changed from the Editing Specific Gravity mode or Editing Species Name mode menu item. This allows you to tailor the L622 species setting to a species not listed. The factory species list may also be custom configured through the Stat-Pak program. The values shown are median SG values. Any single wood species will have an SG variation. Again, if necessary, you can edit these SG values if you feel that your particular species has a slightly different SG median value based on your geographic region. Additionally, the following resources are available to obtain specific gravity values for species not found in the L622 and Stat-Pak software.
http://www.wagnersg.com/ http://www.fpl.fs.fed.us/utilities/information.php http://www.woodworkerssource.com/wood_library.php
You can create your own custom species list in the meter as well as in Stat-Pak.
Species SG
Alder,Red .41 Cedar,EastRed.47 Cedar,WestRed.32 Cherry,Black .50 Douglas Fir .48 Fir,Balsam .35
Fir,Cal Red .38 Fir,Subalpine.32 Fir,White .39 Hemlock,East .40 Hemlock,West .45 Hickory,Nutmg.60 Hickory,Pecan.66 Hickory,Mockr.72 Larch,Western.52 Maple,Sugar .63 Oak,North Red.63 Oak,South Red.59 Oak,White .68 Pine,E White .35 Pine,Jack .43 Pine,Loblolly.51 Pine,Ldgepole.41 Pine,Longleaf.59 Pine,Pnderosa.40 Pine,Shrtleaf.51 Pine,Slash .59 Pine,Sugar .36 Poplar,Yellow.42 Redwood, Old .40 SYP .56 Custm Species.50
Page 26
500-62201-001 Rev. M Page 26 of 33
Commentary On Specific Gravity
In 1992, a study was conducted at the Forest Research Laboratory of Oregon State University on species correction for the Wagner Hand-Held Moisture Meters. The species tested were Douglas Fir, Lodgepole Pine, Western Red Cedar, Western Hemlock, White Fir, Western Larch, Engelmann Spruce and White Oak. Three to four 40-piece samples of each species were tested. Specific gravity was found to be the primary factor on species correction. A species correction equation as a function of specific gravity and the meter reading was obtained using multiple regression technique (R-square =
0.95) as follows: CF = 8.8772 + (0.2492 * MM) - (15.8649 * SG) - (0.6204 * SG
* MM) CF = species correction; MM = meter reading; SG = species
average specific gravity in oven-dry weight and 12 %MC volume basis
Wood is not a uniform material. Specific gravity of solid sawn lumber varies within the piece and between pieces. In the OSU study, the average sample specific gravity for each species differed from the individual sample by plus and minus 1% to plus and minus 8%. For general applications, average specific gravity values can be found in the Wood Handbook (USDA Agriculture Handbook '72, 1987). Except for one species whose experimental value is 7% higher, the species' overall average specific gravity values obtained in the OSU study are comparable with those in the Wood Handbook. The exception may be caused by unknown biases in the sampling scheme. The Wood Handbook values are used in the species list.
Species correction can be determined for lumber sorted, or otherwise known, to have specific gravity different from the species' average. One example is lumber graded under the Dense rules. If the specific gravity of a lumber sample is known, species correction for this lumber sample can be obtained by the species correction equation.
The species correction equation provides a way to expand the use of the Wagner Hand-Held Moisture Meters for lumber of any species groups having similar species specific gravity values. One example is Hem-Fir. For a species group, one way to determine the species correction is by the use of a weighted average of the individual species' average specific gravity values. The weighing procedure used in the ASTM D2555 by standing timber volume can be used. Species correction is not recommended for any species group having a broad range of species specific gravity values. There are no recognized limits on species group species correction. Species correction for species groups should be used with knowledge on the variability of specific gravity of all individual species involved and the effect of it on species correction. If the species mix in the lumber production of a species group is controlled or known to have specific gravity different from that used for the species group, a better estimation of species correction can be determined using the known specific gravity in the above species correction equation.
Page 27
500-62201-001 Rev. M Page 27 of 33
Connecting to a Serial Printer
To print a report to a serial printer, the following equipment is needed: L622, coiled modular serial cable, DB25P to RJ11 adapter, and a serial printer.
• The serial printer must be ON, on line and configured for 9600 baud, 8 data bits, no parity, 1 stop bit.
• Plug one end of the modular serial cable into the COMM port of the L622.
• Plug the other end into the DB25P to RJ11 modular adapter.
• Plug the DB25P to RJ11 adapter into the serial printer.
• Follow the Print Report instructions.
DB25P to RJ11 Adapter Pin Assignments
Pin Description
2 Receive 3 Transmit 7 Ground 1 No Connection 4-6 No Connection 8-25 No Connection
DB9S to RJ11 Adapter Pin Assignments
Pin Description
2 Transmit 3 Receive 5 Ground 1 No Connection 4 No Connection 6-9 No Connection
Page 28
500-62201-001 Rev. M Page 28 of 33
Question and Answers
Q: I’m nervous about buying a new technology. What about Wagner Meters? How long have they been building this type of Hand-Held Moisture Meter?
A: Wagner Meters has been building quality moisture
measurement instrumentation for over 40 years. They are a leading supplier of sophisticated moisture measurement systems for the primary forest products industry. The advanced electro-magnetic wave technology used in Wagner Hand-Held Moisture Meters has been closely scrutinized by professional wood grading associations and scientists in the industry. It has proven reliable, consistent and much easier to use than pin-type resistance moisture metering.
Q: How can I take accurate moisture readings without sticking pins into my wood? Why doesn’t my new Wagner Hand-Held Moisture Meter read the same moisture content as my old pin-type meter?
A: Pin-type meters work on a primitive, resistance principle
that basically measures the flow of electricity through a substance. This method is subject to many environmental variables that can dramatically affect moisture readings such as chemicals in the water trapped within the wood and the temperature of the wood. Pin-type meter readings must always be corrected for any difference in temperature above or below 70 degrees F. Wagner Hand-Held Moisture Meters use advanced electro-magnetic wave technology and are sensitive to changes in density and the actual moisture content of the wood.
Q: What is the narrowest piece of lumber I can measure accurately with the Wagner Hand-Held Moisture Meter?
A: Model L622 measures boards as narrow as 2 ½" in width. Q: What thickness boards can I measure? A: You can accurately measure boards as thin as ½" and up
to 2" thick. Note: Make sure there is nothing behind boards less than 1"
thick when you take the measurement because Wagner Hand­Held Moisture Meters will read through the thinner board and measure the moisture in the material behind. Under no circumstances measure a piece of wood that is sitting on a metal surface. The Wagner L622 measures to a depth of 1", which will reach the center of a 2" board)
Q: Can I get accurate results on 4x or thicker lumber?
A: No hand-held moisture meter can accurately read to the center of 4x material unless you are willing and able to drive pins 1" into the lumber all the way up and down its length and breadth. However using a Wagner Hand-Held Moisture Meter, you can quickly and easily scan 4x lumber on both sides (4 x 4’s on all four sides). Then only the center 1
5/8" would be
unmeasured. Note: Most wood grading agencies are generally not
concerned about the moisture content in the center of thicker beams and posts. They consider 1" deep scanning more than adequate.
Q: Where is the reading taken with a pin-type meter? With a Wagner Hand-Held Moisture Meter?
A: Pin-type meters take their measurements at the depth that
you’ve been able to drive the pins. . . and only in a line
Page 29
500-62201-001 Rev. M Page 29 of 33
between the non-insulated portion of the pins (often only the tips). In contrast, Wagner Hand-Held Moisture Meters generate a three-dimensional field that measures from the surface of the wood to a depth of 1" under the entire sensor (approximately 2 1/2" x 2 1/2" on the L622).
Q: How does this difference in measuring techniques affect accuracy?
A: If you’re using a pin-type meter, the moisture content you
are reading is determined by the micro-thin path the electricity takes to travel from one pin to the other. In effect, it measures only the moisture content of that very tiny path. If there is a single wet fiber between the pins, the electric charge will flow easily along that fiber and cause pin-type meters to exaggerate the moisture content in the wood when in fact it is just a very small fiber that is wet. However, if the place you choose to drive the pins into the wood is simply extraordinarily dry and untypical of the rest of the piece you will get an exaggerated dry reading. On the other hand, Wagner Hand­Held Moisture Meters take an average of the moisture content discovered by the full scan of the three-dimensional field so small wet fibers are not read as large wet spots. Plus, it only takes seconds to scan the entire board.
Q: How are moisture meters affected by surface moisture?
A: Most moisture meters can be affected by standing water,
or visible water on the board. You should always wipe off as much excess water as possible. Once the standing water is removed, Pin-type meters with non-insulated pins will register a highly exaggerated moisture reading. Wagner Hand-Held Moisture Meters will read slightly higher than normal (probably less than 3 to 4%). If water is allowed to soak into the wood, it will naturally show a higher moisture content. If a piece of wood is quite rough, it will soak up the water quite readily and readings will be affected.
Q: What are the effects of relative humidity on Wagner Hand-Held Moisture Meter readings?
A: As long as there is not condensation on the bottom surface
of the instrument there is no effect from changes in relative humidity.
Q: Do the meters on the Wagner Hand-Held products ever require readjustment? Does my Wagner Hand-Held Moisture Meter need to be calibrated? If so, how often must it be done?
A: Occasionally Wagner Hand-Held Moisture Meters require
adjustment. However, the process of checking zero points and calibration is very simple.
Note: Wagner Hand-Held Moisture Meters are originally calibrated at the factory. Type and amount of use will determine how long this original calibration will last. A calibration block is available for the customers who must check their calibration often. Anytime that the meter is not reading correctly on that calibration block, it should be sent into the factor for calibration.
Q: I have a very thin veneer over a doorstock and I’m trying to measure the moisture content of the core. Can I measure accurately through laminated materials?
A: If you’re measuring an all-wood door with a very thin
veneer wood laminate, you can probably use a correction factor to determine the moisture content of the core material.
Page 30
500-62201-001 Rev. M Page 30 of 33
Note: If you’re measuring a door that has a plastic laminate or Formica-type laminate, the Formica laminate is going to have its own density, which is going to affect the reading of the meter. You can determine the variance caused by the laminate by first measuring only the core and then the core with the laminate. For example, if the core measures 12% without the laminate and 13.5% with, you will then know to correct your readings of the combined material by 1.5%.
Q: Will Wagner Hand-Held Moisture Meters work the same on rough lumber as they do on smooth clean lumber?
A: There are little fibers in very rough material that actually
allow a minute layer of air between the meter and the main body of the wood. However, this should not materially affect the reading, or if it does, only slightly lower.
Note: It’s important on rough material to use some pressure and force the meter down firmly against the wood. Occasionally, the measurement of exceptionally rough material may necessitate adding 1 to 2% to get an accurate reading.
Q: Can you check moisture content of plywood, particle board or wafer board with hand-held meters?
A: Because of the glues and mixed species nature of these
materials, it is very difficult to take reliable moisture readings with pin-type or Wagner Hand-Held Moisture Meters.
Note: However, if you would like to work up your own calibrations for materials you use repetitively, you can contact Wagner for guidelines and suggestions.
Q: Does it matter which way you set a Wagner Hand-Held Moisture Meter on the wood? Cross grain, with the grain, or at an angle?
A: Because Wagner Hand-Held Moisture Meters use
advanced electro-magnetic wave technology, they are completely unaffected by orientation on the wood.
Q: Are the readings that I take with my Wagner Hand-Held Moisture Meter affected by the temperature of the wood like those taken with a pin-type meter? What about frozen wood?
A: Unlike pin-type meters which require corrections for
temperatures above or below 70 degrees F, the readings on the Wagner Hand-Held Moisture Meter are essentially unaffected by the temperature of the wood. Moisture content can accurately be measured as soon as the hot wood is taken out of the kiln. When the same wood is measured hours later, again with the Wagner Hand-Held Moisture Meter, the readings stay consistently the same, unless the wood continues to dry during the cooling process.
For frozen wood, as long as the moisture content of the wood you are measuring is below 15%, you can get reliable readings. When moisture content readings in frozen wood exceed 15%, you will need to make corrections.
Q: How rugged are the Wagner Hand-Held Moisture Meters? Are they too delicate to be used on an abusive production line?
Page 31
500-62201-001 Rev. M Page 31 of 33
A: The Wagner L622 Moisture Meter is a tough production­line model, but can possibly be damaged by being dropped or slammed down hard on wood surfaces.
Q: How accurate are Wagner Hand-Held Moisture Meters? A: The Wagner Hand-Held Moisture Meter is as accurate, or
more accurate than any moisture meter that is on the market. This can be verified by several university studies.
Q: Are Wagner Hand-Held Moisture Meters safe to use? A: Wagner’s Hand-Held Moisture Meters have been tested
and certified to comply with FCC and CE regulations.
Q: What is proper moisture content in wood? What moisture content is considered too high or too low?
A: There is no one
correct answer for this question. As a rule, differences in woods and their uses determine the moisture content desired. For instance, if the wood is to be used in construction as a stud for building, the moisture content requirement could be under 15% to 19%. If the wood is to be glued, and it is too dry it will not bond. If it is too wet, it will not hold. Ideally the moisture content of wood to be used in furniture is between 6 and 8%.
The best way to determine the proper moisture content of the wood you are using is to contact the nearest University Wood Product Representative. You may also call the Forest Products Research Laboratory in Madison, WI (608-231-
9200).
Page 32
500-62201-001 Rev. M Page 32 of 33
Warranty
Wagner Meters warrants this product against defects in material and workmanship for one (1) year from the date of purchase, subject to the following terms and conditions:
Wagner’s liability under this warranty shall be limited, at Wagner’s option, to repair or replacement of this product or any part thereof, which is demonstrated to be defective. To exercise this warranty, customer must telephone, fax or e-mail Wagner’s Customer Service Department for factory instructions for shipment. This limited warranty does not apply if accident, negligent handling, misuse, alteration, damage during shipment, or improper service have damaged the product. Wagner Meters shall, in no event, be liable for incidental or consequential damages for the breach of any express or implied warranty with respect to this product or its calibration.
With proper care and maintenance, as recommended in the manual, the meter should stay in calibration; however, because Wagner Meters has no control over the manner in which the unit will be used, it makes no warranty that the meter will stay in calibration for any specific period of time. Wagner Meters recommends purchasing a calibration standard or return the unit to the factory for diagnostic checkup and re-calibration, on the anniversary date of purchase, each year the meter is in service.
This warranty is in lieu of all other warranties, whether oral or written, express or implied. Any implied
warranties, including implied warranties of merchantability and fitness for a particular purpose, are excluded. Agents and employees of Wagner Meters are not
authorized to make modifications to this warranty or additional warranties binding on Wagner Meters. Accordingly, additional statements, whether oral or written, except written statements from an officer of Wagner Meters do not constitute warranties and should not be relied upon by the customer.
This warranty is personal to the customer purchasing the product from Wagner Meters and is not transferable.
Page 33
500-62201-001 Rev. M Page 33 of 33
Repair Service
All products returned for repair must be accompanied by a Repair Order form found on our website at www.wagnerrepairs.com Repair should be returned to the following address:
Wagner Meters
326 Pine Grove Road
Rogue River, Oregon 97537
Phone (541) 582-0541
Fax (541) 582-4138
Standard repairs: Customer must pay the expense of shipping the product to and from Wagner.
Warranty repairs: Customer must pay the expense of shipping the product to Wagner. Wagner will pay the cost of return shipment by surface carrier within the Continental United States. Customer must pay all extra costs of expedited shipping or shipment to and from locations outside of the Continental United States.
E-mail: support@wagnermeters.com Web: www.wagnermeters.com