Diese Bedienungsanleitung ist e ine Publikation von Voltcraft®, Lindenweg 15, D-92242 Hirschau,
Tel.-Nr. 0180/586 582 7 ( www.voltcraft.de).
Alle Rechte einschließlich Übersetzung vorbehalten. Reproduktionen jeder Art, z . B. Fotokopie, Mikroverfilmung, oder die
Erfassung in elektronischen Datenverarbeitungsanlagen, bedürfen der schriftlichen Genehmigung des Herausgebers.
Nachdruck, auch auszugsweise, verboten.
Diese Bedienungsanleitung entspricht dem technischen Stand bei Drucklegung. Ä nderung in Technik und A usstattung
vorbehalten.
These operating instructions are a publication by Voltcraft®, Lindenweg 15, D-92242 Hirschau/Germany,
Phone +49 180/586 582 7 (www.voltcraft.de).
All rights including translation reserved. Reproduction by any method, e.g. photocopy, microfilming, or the capture in electronic
data processing systems require the prior written approval by the editor. Reprinting, also in part, is prohibited.
These operating instructions represent the technical s tatus at the time of printing. Changes in technology and equipment
reserved.
Ce mode d'emploi est une publication de la société Voltcraft®, Lindenweg 15, D-92242 Hirschau/Allemagne,
Tél. +49 180/586 582 7 (www.voltcraft.de).
Tous droits réservés, y compris de traduction. Toute reproduction, quelle qu'elle soit (p. ex. photocopie, microfilm, saisie dans
des installations de traitement de données) nécessite une autorisation écrite de l'éditeur. Il est interdit de le réimprimer, même
par extraits.
Ce mode d'emploi correspond au niveau t echnique du moment de la mise sous presse. Sous réserve de modifications
techniques et de l'équipement.
Deze gebruiksaanwijzing is een publicatie van de firma Voltcraft®, Lindenweg 15, D-92242 Hirschau/Duitsland,
Tel. +49 180/586 582 7 (www.voltcraft.de).
Alle rechten, vertaling inbegrepen, voorbehouden. Reproducties van welke aard dan ook, bijvoorbeeld fotokopie,
microverfilming of de registratie in elektronische gegevensverwerkingsapparatuur, vereisen de schriftelijke toestemming van
de uitgever. Nadruk, ook van uittreksels, verboden.
Deze gebruiksaanwijzing voldoet aan de technische stand bij het in druk bezorgen. Wijziging van techniek en uitrusting
voorbehouden.
Utility key 4/4 35
Default Settings 36
Built-in Help 37
5. MEASUREMENT 38
Basic Measurements 38
Activating a channel 38
Using Autoset 39
Running and stopping the trigger 40
Changing the horizontal position and scale 41
Changing the vertical position and scale 42
Using the probe compensation signal 43
Automatic Measurements 45
Measurement items 45
Automatically measuring the input signals 47
Cursor Measurements 50
Using the horizontal cursors 50
Using the vertical cursors 51
Math Operations 53
Overview 53
Adding, subtracting or multiplying signals 54
Using the FFT function 55
6. CONFIGURATION 56
Acquisition 56
Selecting the acquisition mode 56
Selecting Delay mode 57
Real time vs Equivalent time sampling mode 59
Display 59
Selecting vector or dot drawing 59
Accumulating the waveform 59
Adjusting the display contrast 60
3
Selecting the display grid 60
Horizontal View 61
Moving the waveform position horizontally 61
Selecting the horizontal scale 62
Selecting the waveform update mode 62
Zooming the waveform horizontally 63
Viewing waveforms in the X-Y mode 64
Vertical View (Channel) 65
Moving the waveform position vertically 65
Selecting the vertical scale 65
Selecting the coupling mode 65
Expand Vertical Scale Centre / Ground 66
Inverting the waveform vertically 67
Limiting the waveform bandwidth 68
Selecting the probe attenuation level 68
Trigger 70
Trigger type 70
Trigger parameter 70
Confi guring the edge trigger 73
Confi guring the video trigger 74
Confi guring the pulse width trigger 74
Manually triggering the signal 76
Remote Control Interface 77
System Settings 78
Viewing the system information 78
Selecting the language 78
7. SAVE/RECALL 79
File Structures 79
Display image fi le format 79
Waveform fi le format 79
Setup fi le format 81
Using the SD card fi le utilities 81
Quick Save (HardCopy) 83
Save 85
Saving all (panel settings, display image, waveform) 89
Recall 90
File type/source/destination 90
Recalling the default panel settings 91
Recalling a reference waveform to the display 92
Recalling panel settings 92
Recalling a waveform 93
The input signal does not appear in the display. 98
I want to remove some contents from the display. 98
The waveform does not update (frozen). 98
The probe waveform is distorted. 98
Autoset does not catch the signal well. 99
I want to clean up the cluttered panel settings. 99
The saved display image is too dark on the background. 99
The accuracy does not match the specifi cations. 99
The SD card slot does not accept my card. 99
The oscilloscope will not allow a 2M waveform to be saved. 99
Model-specifi c specifi cations 101
Common specifi cations 102
Probe Specifi cations 104
GDS-1062A / 1152A Probe 104
11. DISPOSAL 105
12. INDEX 106
5
INTRODUCTION1.
Dear Customer,
In purchasing this Voltcraft® product, you have made a very good decision for which we would like
to thank you.
Voltcraft® - In the fi eld of measuring, charging and network technology, this name stands for
high-quality products which perform superbly and which are created by experts whose concern is
continuous innovation.
From the ambitious hobby electronics enthusiast to the professional user, products from the
Voltcraft® brand family provide the optimum solution even for the most demanding tasks. And the
remarkable feature is: we offer you the mature technology and reliable quality of our Voltcraft®
products at an almost unbeatable price-performance ratio. In this way, we aim to establish a long,
fruitful and successful co-operation with our customers.
We wish you a great deal of enjoyment with your new Voltcraft® product!
All names of companies and products are trademarks of the respective owner. All rights
reserved.
Front panel diagram .....................................12
Frequency measure ...................................... 46
Frequency rejection ...................................... 72
SAFETY INSTRUCTIONS2.
This chapter contains important safety instructions that should be followed when operating and
storing the oscilloscope. Read the following before any operation to ensure your safety and to keep
the oscilloscope in the best condition.
Safety Symbols
These safety symbols may appear in this manual or on the oscilloscope.
Warning: Identifi es conditions or practices that could result in injury or loss of life.
WARNING
Caution: Identifi es conditions or practices that could result in damage to the
oscilloscope or to other objects or property.
CAUTION
DANGER High Voltage
Attention: Refer to the Manual
Protective Conductor Terminal
Earth (Ground) Terminal
106
7
Safety Guidelines
General
Guideline
(Measurement categories) EN 61010-1:2001 specifi es the measurement categories
and their requirements as follows. The GDS-1000A falls under category II.
Make sure the BNC input voltage does not exceed 300V peak.•
Never connect a hazardous live voltage to the ground side of the BNC •
connectors. It might lead to fi re and electric shock.
Do not place heavy objects on the oscilloscope.•
Avoid severe impact or rough handling that may damage the oscilloscope.•
Avoid discharges of static electricity on or near the oscilloscope.•
Use only mating connectors, not bare wires, for the terminals.•
Do not block the cooling fan vent.•
Do not perform measurements at power sources and building installation sites •
(Note below).
The oscilloscope should only be disassembled by a qualifi ed technician.•
Measurement category IV is for measurement performed at the source of a •
low-voltage installation.
Measurement category III is for measurement performed in a building •
installation.
Measurement category II is for measurement performed on circuits directly •
connected to a low voltage installation.
Measurement category I is for measurements performed on circuits not directly •
connected to Mains.
The power supply voltage should not fl uctuate more than 10%.•
Connect the protective grounding conductor of the AC power cord to an earth •
ground.
Applicable model & probeGDS-1152A
Position x 10 Attenuation Ratio10:1
Bandwidth DC ~ 150MHz
Input Resistance 10MΩ when used with 1MΩ
Input Capacitance 17pF approx.
Maximum Input
Voltage
Position x 1 Attenuation Ratio1:1
Bandwidth DC ~ 6MHz
Input Resistance 1MΩ when used with 1MΩ
Input Capacitance 47pF approx.
Maximum Input
Voltage
Operating Cond. Temperature –10°C ~ 55°C
Relative Humidity ≤85% @35°C
Safety Standard EN 61010-031 CAT II
* Note: GW Instek reserves the right to change the probe model type at anytime without notice for
probe model types of similar specifi cation.
input
500V CAT I, 300V CAT II
(DC+Peak AC)
Derating with frequency
input
300V CAT I, 150V CAT II
(DC+Peak AC)
Derating with frequency
DISPOSAL11.
In order to preserve, protect and improve the quality of environment, protect human
health and utilise natural resources prudently and rationally, the user should return
unserviceable product to relevant facilities in accordance with statutory regulations.
The crossed-out wheeled bin indicates the product needs to be disposed separately and
not as municipal waste.
8
105
Probe Specifi cations
GDS-1062A/1152A Probe
Applicable model & probeGDS-1062A
Position x 10 Attenuation Ratio10:1
Bandwidth DC ~ 60MHz
Input Resistance10M when used with 1M input
Input Capacitance 23pF approx.
Maximum Input
Voltage
Position x 1 Attenuation Ratio1:1
Bandwidth DC ~ 6MHz
Input Resistance 1MΩ when used with 1MΩ
Input Capacitance 128pF approx.
Maximum Input
Voltage
Operating Cond. Temperature –10°C ~ 55°C
Relative Humidity ≤85% @35°C
Safety Standard EN 61010-031 CAT II
500V CAT I, 300V CAT II
(DC+Peak AC)
Derating with frequency
input
300V CAT I, 150V CAT II
(DC+Peak AC)
Derating with frequency
FuseFuse type: T1A/250V•
Cleaning the
oscillo-scope
Operation
Environment
To ensure fi re protection, replace the fuse only with the specifi ed type and •
rating.
Disconnect the power cord before replacing the fuse.•
Make sure the cause of fuse blowout is fi xed before replacing the fuse.•
Disconnect the power cord before cleaning the oscilloscope.•
Use a soft cloth dampened in a solution of mild detergent and water. Do not •
spray any liquid into the oscilloscope.
Do not use chemicals containing harsh products such as benzene, toluene, •
xylene, and acetone.
Location: Indoor, no direct sunlight, dust free, almost non-conductive pollution •
(Note below)
Relative Humidity: < 80%•
Altitude: < 2000m•
Temperature: 0°C to 50°C•
(Pollution Degree) EN 61010-1:2001 specifi es pollution degrees and their
requirements as follows. The oscilloscope falls under degree 2.
Pollution refers to “addition of foreign matter, solid, liquid, or gaseous (ionized
gases), that may produce a reduction of dielectric strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive pollution occurs. •
The pollution has no infl uence.
Pollution degree 2: Normally only non-conductive pollution occurs. •
Occasionally, however, a temporary conductivity caused by condensation must
be expected.
Pollution degree 3: Conductive pollution occurs, or dry, nonconductive pollution •
occurs which becomes conductive due to condensation which is expected. In
such conditions, equipment is normally protected against exposure to direct
sunlight, precipitation, and full wind pressure, but neither temperature nor
humidity is controlled.
104
Storage
environment
Location: Indoor•
Relative Humidity: < 85%•
Temperature: -10°C to 60°C•
9
Power cord for the United Kingdom
When using the oscilloscope in the United Kingdom, make sure the power cord meets the following
safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED
IMPORTANT: The wires in this lead are coloured in accordance with the following code:\
Green / Yellow: Earth
Blue: Neutral
Brown: Live (Phase)
As the colours of the wires in main leads may not correspond with the coloured marking identifi ed in
your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth terminal marked with
either the letter E, the earth symbol or coloured Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is marked with the letter
N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal marked with the letter L or P
or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact the supplier.
This cable/appliance should be protected by a suitably rated and approved HBC mains fuse: refer
to the rating information on the equipment and/or user instructions for details. As a guide, a cable of
0.75mm2 should be protected by a 3A or 5A fuse. Larger conductors would normally require
13A types, depending on the connection method used.
Any exposed wiring from a cable, plug or connection that is engaged in a live socket is extremely
hazardous. If a cable or plug is deemed hazardous, turn off the mains power and remove the cable,
any fuses and fuse assemblies. All hazardous wiring must be immediately destroyed and replaced
in accordance to the above standard.
Cursors and
Measurement
Control Panel
Function
Display LCD 5.6 inch, TFT, brightness adjustable
Interface USB SlaveUSB1.1 & 2.0 full speed compatible
Probe
Compensation
Signal
Power Source Line Voltage 100V~240V AC, 47Hz~63Hz
OperationAmbient temperature 0 ~ 50°C
EnvironmentRelative humidity ≤ 80% @35°C
StorageAmbient temperature –10°C to 60°C
EnvironmentRelative humidity ≤ 80% @60°C
Dimensions 341.5(W) x 162.3 (H) x 159 (D) mm
Weight Approx. 2.5kg
Average 2, 4, 8, 16, 32, 64, 128, 256
Voltage Vpp, Vamp, Vavg, Vrms, Vhi, Vlo, Vmax,
Vmin, Rise Preshoot/ Overshoot, Fall
Preshoot/ Overshoot
Time Freq, Period, Rise Time, Fall Time, +
Width, – Width, Duty Cycle
Delay FRR, FRF, FFR, FFF, LRR, LRF, LFR, LFF
Cursors Voltage difference (∆ V) and
Time difference ( ∆T) between cursors
Auto Counter Resolution: 6 digits, Accuracy: ±2%
Signal source: All available trigger source
except the Video trigger
Autoset Automatically adjust Vertical Volt/div,
Horizontal Time/div, and Trigger level
Save/Recall Up to 15 sets of measurement conditions
and waveforms
Resolution (dots) 234 (Vertical) x 320 (Horizontal)
Graticule 8 x 10 divisions
Display Contrast Adjustable
Connector(printers and fl ash disk not supported)
SD Card Slot Image (BMP) and waveform data (CSV)
Frequency range 1kHz ~ 100kHz adjustable, 1kHz step
1.5div or 15mV (25MHz~100MHz)
External Trigger~ 50mV (DC~25MHz)
Sensitivity~ 100mV (25MHz~100MHz)
Rise Time < 2.3ns approx.
Cursor keyCursor
12
Runs cursor measurements (page 22).
101
APPENDIX10.
Fuse Replacement
ProcedureRemove the power cord and remove the fuse socket using a minus driver.1.
Utility keyUtilityConfi gures the Hardcopy function (page 23), shows the
Help keyHelp
Autoset key
Autoset
system status (page 34), selects the menu language
(page 34), runs the self calibration (page 34), and
confi gures the probe compensation signal (page 35).
Shows the Help contents on the display (page 37).
Automatically confi gures the horizontal, vertical, and
trigger settings according to the input signal (page 23).
Replace the fuse in the holder.2.
RatingsT1A, 250V
100
Measure keyMeasure
Save/Recall keySave/Recall
Hardcopy keyHardcopy
Run/Stop keyRun/Stop
Trigger level
knob
Trigger menu
key
Single trigger
key
Trigger force
key
Horizontal menu
key
Horizontal
position knob
TRIGGER
LEVEL
MENU
SINGLE
FORCE
MENU
Confi gures and runs automatic measurements (page
26).
Saves and recalls images, waveforms, or panel settings
(page 27).
Stores images, waveforms, or panel settings to an SD
card (page 83).
Runs or stops triggering (page 41).
Sets the trigger level (page 72).
Confi gures the trigger settings (page 31).
Selects the single triggering mode (page 76).
Acquires the input signal once regardless of the trigger
condition at the time (page 76).
Confi gures the horizontal view (page 61).
Moves the waveform horizontally (page 61).
13
TIME/DIV knobTIME/DIVSelects the horizontal scale (page 42).
Vertical position
knob
CH1/CH2 keyCH 1
Moves the waveform vertically (page 42).
Confi gures the vertical scale and coupling mode for
each channel (page 21).
Autoset does not catch the signal well.
The Autoset function cannot catch signals under 30mV or 30Hz.
Please operate the oscilloscope manually. See page 76 for details.
I want to clean up the cluttered panel settings.
Recall the default settings by pressing the Save/Recall key—Default Setting.
For default setting contents, see page 36.
VOLTS/DIV
knob
Input terminalCH1
Ground terminal
MATH keyMATH
SD card
connector
Probe
compensation
output
External trigger
input
Power switch
14
VOLTS/DIV
EXT TRIGAccepts an external trigger signal (page 70).
Outputs a 2Vp-p, square signal for compensating the
probe (page 96) or demonstration.
Powers the oscilloscope on or off.
The saved display image is too dark on the background.
Use the Inksaver function which reverses the background color. For details, see page 35.
The accuracy does not match the specifi cations.
Make sure the device is powered on for at least 30 minutes, within +20°C~+30°C. This is necessary
to stabilize the unit to match the specifi cation.
The SD card slot does not accept my card.
Make sure is: 1. a standard SD card (MMC and SDHC is not supported), 2. 2GB or less, and 3.
formatted as FAT or FAT32.
The oscilloscope will not allow a 2M waveform to be saved.
Make sure that only 1 channel is active. Make sure that the signal has been triggered and that the
STOP or Single key has been pressed.
Ensure the time base is slower than 10 ns/div. See page 80.
For more information, contact your local dealer or GWInstek at www.gwinstek.com.tw / marketing@
goodwill.com.tw.
99
FAQ9.
The input signal does not appear in the display.•
I want to remove some contents from the display.•
The waveform does not update (frozen).•
The probe waveform is distorted.•
Autoset does not catch the signal well.•
I want to clean up the cluttered panel settings.•
The accuracy does not match the specifi cations.•
The SD card slot does not accept my card.•
The oscilloscope will not allow a 2M waveform to be saved.•
The input signal does not appear in the display.
Make sure you have activated the channel by pressing the CH key.
I want to remove some contents from the display.
To clear the math result, press the Math key again.
To clear the cursor, press the Cursor key again.
To clear the Help contents, press the Help key again.
Rear Panel
Power cord
socket
Fuse socket
Power cord socket accepts the AC mains, 100 ~ 240V,
50/60Hz.
The fuse socket holds the AC main fuse, T1A/250V.
For the fuse replacement procedure, see page100.
The waveform does not update (frozen).
Press the Run/Stop key to unfreeze the waveform. See page 26 for details. For trigger setting details,
see page 70.
If this does not help, press the CH key. If the signal still does not appear, press the Autoset key.
The probe waveform is distorted.
You might need to compensate the probe. For details, see page 96.
Note that the frequency accuracy and duty factor are not specifi ed for probe compensation waveforms
and therefore it should not be used for other reference purposes.
98
USB slave port
Calibration
output
Accepts a type B (slave) male USB connector for
remote controlling the oscilloscope (page 77).
Outputs the calibration signal used in vertical scale
accuracy calibration (page 95).
15
Display
Press Wavetype repeatedly to 4.
select the standard square wave.
Wave Type
WaveformsChannel 1: YellowChannel 2: Blue
Trigger statusTrig’dA signal is being triggered
Trig?Waiting for a trigger condition
Input signal
frequency
Trigger
confi guration
Horizontal
status
Vertical status
AutoUpdating the input signal regardless of trigger condi-
STOPTriggering is stopped
For trigger setting details, see page 70.
Updates the input signal frequency (the trigger source signal) in real-time.
“< 20Hz” Indicates that the signal frequency is less than the lower frequency limit
(20Hz) and thus not accurate.
Shows the trigger source, type, and slope. In case of the Video trigger, shows the
trigger source and polarity.
Shows the channel confi gurations: coupling mode, vertical scale, and horizontal
scale.
tions
Press the Autoset key. The 5.
compensation signal will appear in
the display.
Press the Display key, then 6. Type
to select the vector waveform.
Turn the adjustment point on the probe until the signal edge becomes sharp.7.
Over
Compensation
Autoset
Dispaly
NormalUnder
Type
Vectors
Compensation
16
97
When fi nished, connect the 8.
calibration signal to the Channel 2
input and repeat the procedure.
The calibration is completed and the display goes back to the previous state.9.
Probe Compensation
Done!!
Setting up the Oscilloscope
BackgroundThis section describes how to set up the oscilloscope properly including adjusting
the handle, connecting a signal, adjusting the scale, and compensating the
probe. Before operating the oscilloscope in a new environment, run these steps
to make sure the oscilloscope is functionally stable.
ProcedurePull both bases of the handle out 1.
slightly.
ProcedureConnect the probe between Channel1 input and the probe compensation 1.
96
output (2Vp-p, 1kHz square wave) on the front panel. Set the probe attenuation
to x10.
x1
x10
Press the Utility key.2.
Press 3. ProbeComp.
Utility
ProbeComp
Menu
Turn to one of the three preset 2.
positions.
Connect the power cord.3.
Press the power switch. The display 4.
will become active in approximately
10 seconds.
Reset the system by recalling the 5.
factory settings. Press the Save/
Recall key, then Default Setup.
For details regarding the factory
settings, see page 91.
Save/Recall
Default
Setup
17
Connect the probe between the Channel1 input terminal and probe 6.
compensation signal output (2Vp-p, 1kHz square wave).
Set the probe attenuation to x10.7.
MAINTENANCE8.
Two types of maintenance operations are available: calibrating the vertical resolution, and
compensating the probe. Run these operations when using the oscilloscope in a new environment.
Vertical Resolution Calibration
Press the Autoset key. A square 8.
waveform will appear in the center of
the display. For details on Autoset, see
page 23.
Press the Display key, then 9. Type and
select the vector waveform type.
Turn the adjustment point on the probe to fl atten the square waveform edge.10.
Over Under
Compensation Normal Compensation
Display
Autoset
Type
Vectors
ProcedurePress the Utility key.1.
Press 2. More.
Press3. Self Cal Menu.
Press 4. Vertical. The message
“Set CAL to CH1, then press
F5” appears at the bottom of the
display.
Connect the calibration signal 5.
between the rear panel CAL out
terminal and the Channel1 input.
Press F5. The calibration automatically starts.6.
The Channel1 calibration will 7.
complete in less than 5 minutes.
Utility
More
Self CAL
Menu
Vertical
CALCH1
Ch1 calibration 1/3
18
95
Press Recall Waveform. The 3.
display shows the available source
and destination options.
Recall
Waveform
Press Source repeatedly to select 4.
the fi le source, internal memory or
external SD card. Use the Variable
knob to change the memory
location (W1 ~ W15).
Memory Internal memory, W1 ~ W15
SD card External fl ash drive, no practical limitation on the
amount of fi les. The waveform fi le must be placed in the root directory to be recognized.
Press Destination. Use the 5.
Variable knob to select the memory
location.
RefA, B Internally stored reference waveforms A, B
Press 6. Recall to confi rm recalling.
When completed, a message
appears at the bottom of the
display.
Note
File utilitiesTo edit the SD card contents (create/
delete/ rename fi les and folders),
press File Utilities. For details, see
page 81.
The fi le will not be saved if the power is turned Off or the SD
card is disconnected before completion.
Source
Memory
VARIABLE
Destination
VARIABLE
Recall
File
Utilities
Setting up the oscilloscope is completed. You may continue with the other 11.
operations.
Measurement: page 38 Confi guration: page 56
94
19
QUICK REFERENCE4.
This chapter lists the oscilloscope menu tree, operation shortcuts, built-in help coverage, and default
factory settings. Use this chapter as a handy reference to access the oscilloscope functionalities.
Menu Tree and Shortcuts
ConvenionsExamples
Normal= Press the functional key for “Normal”
Press 3. Recall Setup.
Press 4. Source repeatedly to select
the fi le source, internal or external
memory. Use the Variable knob to
change the memory.
Display
Refs.
Source
Memory
VARIABLE
Average
Normal ~
Average
Normal
VAR
20
= Repeatedly press the functional key for “Average”
= Select a menu from “Normal” to “Average” and press its functionality key
= Press the functionality key for “Normal”, and then use the Variable knob
Memory Internal memory, S1 ~ S15
SD card External card, no practical
limitation on the amount of fi le. The setup fi le must be placed in the root directory to be recognized.
Press 5. Recall to confi rm recalling.
When completed, a message
appears at the bottom of the
display.
Note
File utilitiesTo edit SD card contents (create/
delete/ rename fi les and folders),
press File Utilities. For details, see
page 81.
The fi le will not be saved if the power is turned Off or the SD
card is disconnected before completion.
Recalling a waveform
Procedure(For recalling from an external SD 1.
card) Insert the card into the slot.
Press the Save/Recall key.2.
Recall
File
Utilities
Save/Recall
93
Recalling a reference waveform to the display
ProcedureThe reference waveform must be stored in advance. 1.
Press the Save/Recall key. Save/2.
Recall
Press 3. Display Refs. The reference
waveform display menu appears.
Select the reference waveform, 4.
Ref A or Ref B, and press it. The
waveform appears on the display
and the period and amplitude
of the waveform appears in the
menu.
To clear the waveform from the 5.
display, press RefA/B again.
Recalling panel settings
Procedure(For recalling from an external SD 1.
card) Insert the card into the slot.
Press the Save/Recall key.2.
Save/Recall
Display
Refs.
Ref.A Off
Ref.A Off
200mV
250uS
Ref.A Off
Save/Recall
Acquire
Normal
Average
Peak
Detect
Delay
On
Sample Rate
500MS/s
2/ 4/ 8/ 16/ 32/
64/ 128/ 256
On/ Off
Select acquisition mode
Normal ~ Peak-Detect
Select average number
Average
Turn Delay on/off
Delay On
CH1/2 key
CH 1
Coupling
Invert
Off
BW Limit
Off
Probe
x1
Expand
Center
On/ Off
On/ Off
x1/ x10/ x100
Center/Ground
Turn channel on/off
CH 1/2
Select coupling mode
Coupling
Invert waveform
Invert
Turn bandwidth limit on/off
BW Limit
Select probe attenuation
Probe(1x~100x)
Expand type
Expand
92
21
Cursor key 1/2
Recalling the default panel settings
Cursor
Source
CH1
X1
-5.000uS
0.000uV
X2
5.000uS
0.000uV
X1X2
10.00uS
100.0kHz
0.000uV
XY
Cursor key 2/2
Cursor
Source
CH1
Y1
123.4mV
Y2
22.9mV
Y1Y2
16.0V
X Y
CH1/ 2/ MATH
CH1/ 2/ MATH
Turn cursor on/off
Cursor
Move X1 cursor
X1 VAR
Move X2 cursor
X2 VAR
Move both X1 and X2 cursor
X1X2 VAR
Switch to Y cursor
X Y
Turn cursor on/off
Cursor
Move Y1 cursor
Y1 VAR
Move Y2 cursor
Y2 VAR
Move both Y1 and Y2 cursor
Y1Y2 VAR
Switch to X cursor
X Y
ProcedurePress the Save/Recall key.1.
Default panel
setup
Setting contents The following is the default panel setting contents.
Acquisition Mode: Normal
Channel Coupling: DC Invert: Off
Cursor Source: CH1 Horizontal: None
Display Type: Vectors Accumulate: Off
Horizontal Scale: 2.5us/Div Mode: Main Timebase
Math Type: + (Add) Channel: CH1+CH2
Measure Item: Vpp, Vavg, Frequency, Duty cycle, Rise Time
Trigger Type: Edge Source: Channel1
Utility SaveImage, InkSaver Off
Press 2. Default Setup. The factory
installed setting will be recalled.
BW limit: Off voltage: x1
Vertical: None
Graticule:
Position: 0.00 Div Unit/Div: 2V
Mode: Auto Slope:
Coupling: DC Rejection: Off
Noise Rejection: Off
Save/Recall
Default
Setup
22
91
Press Save to confi rm saving. When 6.
completed, a message appears at the
bottom of the display.
Note
Together with the current setup/waveform/ image, the last saved waveform fi le 7.
(one from W1 ~ W15) and setup fi le (one from S1 ~ S15) are also included
in the folder.
capture or it can be used as a reference waveform.
Insert the card into the slot.
Press the Save/Recall key twice to 2.
access the Save menu.
Press 3. Save Image.
Press4. Ink Saver repeatedly to invert the
background color (on) or not (off).
Press 5. Destination.
SD card External card, no practical limitation on the amount of fi les.When saved, the image fi le will be placed in the root directory.
Press 6. Save to confi rm saving. When
completed, a message appears at the
bottom of the display.
Note
rename fi les and folders), press File
Utilities. For details, see page 81.
The fi le will not be saved if the power is turned off or the
SD card is disconnected before completion.
Save/Recall Save/Recall
Save
Image
Ink Saver
Off
Destination
SD Card
Save
File
Utilities
MATH
Operation
FFT
Source
CH1
Window
Hanning
Position
0.00 Div
Unit/Div
1dB
CH1+CH2
CH1-CH2
CH1xCH2
FFT
FFT rms
CH1/2
Flattop/
Rectangular/
Blackman/
Hanning
-12div ~ +12div
20/10/5/2/1 dB
Math on/off
Math
Select math operation type (+/–/x/FFT/
FFT rms)
Operation
Select FFT source channel
Source
Select FFT window
Window
Select FFT result position
Position VAR
Select vertical scale
Unit/Div
88
25
Measure key
Measure
Measure
Press 4. Source. Use the Variable knob to
select the source signal.
Source
VARIABLE
Vpp
1:204mV
2: 300mV
Vavg
1: 1.93mV
2: 28.0mV
Frequency
1: 1.000kHz
2: 3.003kHz
Duty Cycle
1: 50.01%
2: 49.88%
Rise Time
1. 76.20us
2. 70.11us
Source 1
CH1
Source 2
CH2
Voltage
Vpp
Previous
Menu
View all/Select
Measurements
Time/Delay)
Turn on/off measurement Measure
Select measurement type Voltage/Time/Delay
Select measurement item VAR or Icon(F3) / VAR
Go back to previous menu Previous Menu
Run/Stop key
Run/StopFreeze/unfreeze waveform or trigger
Run/Stop
(Voltage/
CH1 ~ CH2 Channel 1 ~ 2 signal
Math Math operation result
RefA, B Internally stored reference waveforms A, B
Press 1. Destination repeatedly to select
the fi le destination. Use the Variable
knob to select the memory location.
Memory Internal memory, W1 ~ W15
SD Normal Save to the SD card with a 4k waveform memory length.
SD 1M Save to the SD card with a 1M waveform memory length.
For 2 channel operation only.
SD 2M Save to the SD card with a 2M waveform memory length.
For single channel operation only.
Ref Internal reference waveform, A/B
Press 2. Save to confi rm saving. When
completed, a message appears at the
bottom of the display.
Note
The fi le will not be saved if the power is turned off or the
SD card is disconnected before completion.
rename fi les and folders), press File
Utilities. For details, see page 81.
Destination
Memory
VARIABLE
Save
File
Utilities
26
87
Press 4. Destination repeatedly to select
the saved location. Use the Variable
knob to change the internal memory
location (S1 ~ S15).
Destination
Memory
VARIABLE
Save/Recall key 1/9
Save/Recall
Save/Recall
Memory Internal memory, S1 ~ S15
SD card External card, no practical
limitation for the amount of fi le.When saved, the setup fi le will beplaced in the root directory.
Press 5. Save to confi rm saving. When
completed, a message appears at the
bottom of the display.
Note
The fi le will not be saved if the power is turned off or the
SD card is disconnected before completion.
Press the Save/Recall key twice to 2.
access the Save menu.
Press 3. Save Setup.
Save/Recall Save/Recall
Save
Setup
28
85
Press the Utility key.2.
Press 3. Hardcopy Menu.
Press 4. Function repeatedly to select Save
Image or Save All.
To invert the color in the display image, press 5.
Ink Saver. This turns Ink Saver on or off.
Press 6. Mem Leng repeatedly to select SD Normal or SD 1M/2M. SD Normal and
SD 1M/2M sets the waveforms to a 4k
and 1M/2M memory length when saving,
respectively.
1M memory length is available when both CH1 and CH2 are active; 2M
memory length is available when a single channel is active only.
Press the Hardcopy key. The fi le or folder 7.
will be saved to the root directory of the SD
card.
Utility
Hardcopy
Menu
Function
Save All
Ink Saver
Off
Mem Leng
SD Normal
Hardcopy
Save/Recall key 5/9
Save Setup
Save
Setup
Destination
Memory
Save
File
Utilities
Save/Recall key 6/9
Save Waveform
Save
Waveform
Source
Destination
Memory
Save
File
Utilities
Memory/SD card
(SD Card only)
To File Utilities
CH1/2Math
Ref A/B
Memory
SD Normal/SD 1M
SD 2M/Refs.
(SD Card only)
To File Utilities
Select other menu
Save Setup
Select destination
Destination
Save setup
Save
Go to SD card fi le utilities
File Utilities
Select other menu
Save Waveform
Select source
Source
Select destination
Destination
Save waveform
Save
Go to SD card fi le utilities
VAR
VAR
VAR
File Utilities
84
29
Save/Recall key 7/9
Save Setup
Save
Setup
Destination
Memory
Save
File
Utilities
Memory/SD card
(SD Card only)
To File Utilities
Select other menu
Save Image
Turn on/off ink saver
Ink Saver
Save image
Save
Go to SD card fi le utilities
File Utilities
When editing is completed, press Save. 3.
The fi le/folder creation or rename will be
completed.
Deleting a folder
or fi le
Move the cursor to the folder or fi le location 1.
and press Delete. The message “Press F4 again to confi rm this process” appears at
the bottom of the display.
If the fi le/folder still needs to be deleted, 2.
press Delete again to complete the deletion.
To cancel the deletion, press any other key.
Quick Save (HardCopy)
BackgroundThe Hardcopy key works as a shortcut for saving display
images, waveform data, and panel settings into an SD card.
Save
Delete
Delete
Hardcopy
Save/Recall key 8/9
Save All
Save
Image
Ink Saver
Off
Destination
SD Normal
Save
File
Utilities
30
On / Off
SD Normal / SD 1M
(SD Card only)
To File Utilities
Select other menu
Save All
Turn on/off ink saver
Ink Saver
Select destination
Destination
Save all
Save
Go to SD card fi le utilities
File Utilities
VAR
Hardcopy key can be confi gured into two types of operations: save image and
save all (image, waveform, setup).
Using the Save/Recall key can also save fi les with more
options. For details, see page 85.
Functionalities Save image
SD Card
restriction
ProcedureInsert an SD card to the slot.1.
(*.bmp)
Save all
The GDS-1000A series accepts the following SD card.
Size: 2GB or less Format: FAT or FAT32
Save/Recall
Saves the current display image into an SD card.
Saves the following items into an SD card.
Current display image (*.bmp)•
Current system settings (*.set)•
Current waveform data (*.csv)•
Last stored system settings (*.set)•
Last stored waveform data (*.csv)•
83
SD card
indicator
Save
Image
Destination
SD Card
Press 3. File Utilities. The display shows the
SD card contents.
Use the Variable knob to move the cursor. 4.
Press Select to go into the folder or go back
to the previous directory level.
When an SD card is inserted into the oscilloscope, an indicator appears at the
right bottom corner of the display. (The SD card shouldn’t be removed when a fi le
is saved or retrieved from the SD card).
SD card
File
Utilities
VARIABLE
Select
Save/Recall key 9/9
File Utilities
Select
New
Folder
Rename
Delete
Previous
Menu
Trigger key 1/6
Entcr
Character
Back
Space
Save
Previous
Menu
Select fi le / folder
Select
VAR
Create or rename folder/fi le
New Folder/Rename
Enter character / Backspace /
VAR
Save / Previous menu
Delete folder/fi le
Delete
Go to previous menu
Previous menu
Creating a
new folder /
renaming a fi le
or folder
82
Move the cursor to the fi le or folder location 1.
and press New Folder or Rename. The fi le/
folder name and the character map will
appear on the display.
Use the Variable knob to move the pointer 2.
to the characters. Press Enter Character to
add a character or Back Space to delete a
character.
New
Folder
Rename
VARIABLE
Enter
Character
Back
Space
Trigger TypeTrigger Holdoff
MENUMENU
Type
Edge
Source
CH1
Slope /
Coupling
Mode
Auto
Holdoff
40.0ns
Set to
Minimum
Select Trigger type or Trigger
Holdoff menu
Type
31
Trigger key 2/6
Setup fi le format
Video Trigger
Save
Image
Source
CH1
Standard
NTSC
Polarity
Line
Trigger key 3/6
Edge Trigger
Type
Edge
Source
CH1
Slope /
Coupling
Mode
Auto
CH1/2
NTSC/SECAM/PAL
/
Field1/Field2/Line
CH1/2Ext/Line
To Slope / Coupling
Auto / Normal
Select video trigger type
Type
Select trigger source
Source
Select video standard
Standard
Select video polarity
Polarity
Select video fi eld/line
Line
Select edge trigger type
Edge
Select trigger source
Source
Go to slope/coupling menu (page 73)
Slope/Coupling
Select trigger mode
Mode
VAR
Formatxxxx.set (proprietary format)
ContentsAcquiremode•
A setup fi le saves or recalls the following settings.
source channel•
video standard•
video line•
slope/coupling•
vertical position•
invert on/off•
probe•
source channel•
unit/div•
Using the SD card fi le utilities
BackgroundWhen an SD card is inserted into the oscilloscope, fi les utilities (fi le deletion,
SD Card
restriction
ProcedureInsert an SD card into the card slot.1.
folder creation and fi le/folder renaming) are available from the front panel.
The GDS-1000A series accepts the following SD card.
Size: 2GB or less
Format: FAT or FAT32
Press the Save/Recall key. Select any save 2.
or recall function. For example SD card
destination in the Save image function.
32
Save/Recall
(Example)
81
Waveform data
format
Waveform
Memory Depth
Waveform fi le
contents: other
data
One division includes 25 points of
horizontal and vertical data. The
vertical point starts from the center
line. The horizontal point starts from
the leftmost waveform.
25
0
255075
–25
The time or amplitude represented by each data point depends on the vertical
and horizontal scale. For example:
Vertical scale: 10mV/div ( 4mV per point)
Horizontal scale: 100us/div (4us per point)
The memory depth is limited to 1 M points when both channels are activated
or 2M points when only a single channel is activated. However the full memory
depth is only available after a signal is triggered and the STOP or Single key has
been pressed.
There are a number of conditions when all of the available memory is not utilized
due to a limited number of different sample rates. This can be caused by an
un-triggered signal, or a time/div setting that is too fast to display all the points
on screen.
Note: 2M point memory lengths are only available for time bases slower than
10ns/div on a single channel, and 1 M point memory lengths are only available
for time bases slower than 25ns/div on two channels.
When the scope is running the number of points displayed on screen is limited
to 4000. Equivalent time sampling is always limited to 4000 points on screen at
any one time.
A waveform fi le also includes the following information.
• Memory length
• source channel
• vertical offset
• vertical scale
• coupling mode
• waveform last dot address
• date and time
• trigger level
• vertical position
• time base
• probe attenuation
• horizontal view
• horizontal scale
• sampling period
• sampling mode
Trigger key 4/6
Pulse Trigger
Type
Pules
Source
CH1
When <
20.0ns
Slope /
Coupling
Mode
Auto
CH1/2/Ext/Line
>/</=/≠
20ns~10s
To Slope / Coupling
Auto / Normal
Select pulse trigger type
Type
Select trigger source
Source
Select pulse trigger condition and pulse width
When VAR
Go to slope/coupling menu (page 73)
Slope/Coupling
Select trigger mode
Mode
Trigger key 5/6
Coupling/Slope
Slope
Coupling
AC
Rejection
Off
Noise Rej
Off
Previous
Menu
/
AC/DC
LF/HF/Off
On/Off
Select trigger slope type
Slope
Select trigger coupling mode
Coupling
Select frequency rejection
Rejection
Turn noise rejection on/off
Noise Rej
Go back to previous menu
Previous Menu
80
33
Trigger key 6/6
Trigger Holdoff
Holdoff
40.0ns
Set to
Minimum
40ns~2.5s
Select Holdoff time
VAR
Set to minimum Holdoff time
Set to Minimum
SAVE/RECALL7.
The save function allows saving display images, waveform data, and panel settings into the
oscilloscope’s internal memory or an external SD card. The recall function allows recalling the default
factory settings, waveform data, and panel settings from the oscilloscope’s internal memory or an
external SD card.
File Structures
Utility key 1/4
Utility
Hardcopy
Menu
ProbeComp
Menu
Language
English
System
Info.
More
Utility key 2/4
Calibration
Self CAL
Menu
Previous
Menu
To Hardcopy
menu
To Probe
Comp menu
English/
Chinese(T) etc
To Self Cal
menu
Vertical
Go to hardcopy menu
Hardcopy
Go to probe compensation menu
ProbeComp
Select language
Language
Show system information
System Info.
Go to self calibration menu
More
Enter self calibration
Self CAL
Go to previous menu
Previous Menu
Three types of fi le are available: display image, waveform fi le, and panel settings.
Display image fi le format
Formatxxxx.bmp (Windows bitmap format)
ContentsThe current display image in 234 x 320 pixels, color mode. The background color
can be inverted (Ink saver function).
Waveform fi le format
Formatxxxx.csv (Comma-separated values format which can be opened in spreadsheet
Waveform type CH1, 2
Storage location Internal
applications such as Microsoft Excel)
Math
memory
External SD
card
Ref A, BTwo reference waveforms are used as a buffer to recall
Input channel signal
Math operation result (page 85)
The oscilloscope’s internal memory,
which can hold 15 waveforms.
An SD card (SD 2G or less, FAT or
FAT32 format) can hold practically an unlimited number
of waveforms.
a waveform in the display. You have to save a waveform into internal memory or an SD card, then copy the
waveform into the reference waveform slot (A or B),
and then recall the reference waveform into the display.
34
79
System Settings
Utility key 3/4
The system settings show the oscilloscope’s system information and allow changing the language.
Viewing the system information
ProcedurePress the Utility key.1. Utility
In the Single
trigger mode
Press System Info. The upper half of the display 2.
shows the following information.
• Manufacturer • Model
• Serial number • Firmware version
Press any other key to go back to the waveform 3.
display mode.
System
Info.
More
Selecting the language
ParameterLanguage selection differs according to the region to
ProcedurePress the Utility key.1. Utility
which the oscilloscope is shipped.
• English • Chinese (traditional)
• Chinese (simplifi ed) • Others
Press 2. Language repeatedly to select the
language.
Language
English
Hardcopy
Function
Save All
Ink Saver
Off
Mem Leng
SD Normal
Previous
Menu
Utility key 4/4
Probe compensation
Wave Type
Frequency
1K
Duty Cycle
50%
Default
1kHz
Previous
Menu
Save Image
SaveAll
On / Off
SD Normal
SD 1M
(
1k ~ 100k
(
5% ~ 95%
/
only)
only)
Select Hardcopy function
Function
Turn on/off Inksaver
Ink Saver
Set the memory length
Mem Leng
Go to previous menu
Previous Menu
Select probe compensation signal
Wave Type
Set frequency for square wave
Frequency VAR
Set duty cycle for square wave
Duty Cycle VAR
Go to previous menu
Previous Menu
78
35
Default Settings
Remote Control Interface
Here are the factory installed panel settings which appear when
pressing the Save/Recall key Default Setup.
AcquisitionMode: Normal
ChannelScale: 2V/DivInvert: Off
Coupling: DCProbe attenuation voltage: x1
BW limit: OffChannel 1 & 2: On
CursorSource: CH1Cursor: Off
DisplayType: VectorsAccumulate: Off
Grid:
HorizontalScale: 2.5us/DivMode: Main Timebase
MathType: + (Add)Position: 0.00 Div
MeasureItem: Vpp, Vavg, Frequency, Duty Cycle, Rise Time
The Remote control interface section describes how to set up the USB interface for PC connection.
The details of remote control commands are described in the GDS-1000A Programming Manual.
USB connection PC sideType A, host
GDS-1000A sideType B, slave
Speed1.1/2.0 (full speed)
ProcedureConnect the USB cable to the USB slave port.1.
When the PC asks for the USB driver, select dso_cdc_1000.inf (Windows 2.
XP) or dso_vista_cdc.inf (Vista 32bit) which are downloadable from the GW
website, www.gwinstek.com.tw, GDS-1000A product corner.
On the PC, activate a terminal application such as MTTTY (Multi-Threaded 3.
TTY). To check the COM port No., see the Device Manager in the PC. For
WindowsXP, select Control panel System Hardware tab.
Run this query command via the terminal application.4.
*idn?
This command should return the manufacturer, model number, serial number,
and fi rmware version in the following format.
GW, GDS-1152A, XXXXXXX, V1.00
Confi guring the command interface is completed. Refer to the programming 5.
manual for the remote commands and other details.
36
77
Manually triggering the signal
Built-in Help
NoteThis section describes how to manually trigger the input signals when the oscil-
To acquire the
signal regarless
of trigger
conditions
In the Single
trigger mode
loscope does not capture them. This section applies to the Normal and Single
trigger mode, since in the Auto trigger mode, the oscilloscope keeps updating the
input signal regardless of the trigger conditions.
To acquire the input signal regardless of the trigger
condition, press the Force key. The oscilloscope
captures the signals once.
Press the Single key to start waiting for the trigger
condition. To break out of the Single mode, press
the Run/Stop key. The trigger mode changes to the
Normal mode.
FORCE
SINGLE
Run/Stop
The Help key shows the contents of the built-in help support.
When you press a function key, its descriptions appear in the display.
Applicable keys Acquire
Cursor
(Vertical) (Horizontal) (Trigger)
CH 1MATHMENUMENU
ProcedurePress the Help key. The display 1.
Display
Measure
changes to the Help mode.
Press a functional key to access its 2.
help contents. (example: Acquire key)
Use the Variable knob to scroll the Help 3.
contents up and down.
Utility
Save/Recall
CH 2
Help
HardcopyRun/Stop
Help
Autoset
SINGLE
FORCE
Help
Acquire
VARIABLE
Press the Help key again to exit the 4.
Help mode.
76
Help
37
MEASUREMENT5.
The Measurement chapter describes how to properly observe a signal using the oscilloscope’s basic
functions, and how to observe a signal in a detailed manner using some of the advanced functions
such as :
Automatic measurements, cursor measurements, and math operations.
Basic Measurements
Press 3. Source repeatedly to select the trigger
source.
Range Channel 1, 2, Ext
Press 4. Mode repeatedly to select the trigger mode,
Auto or Normal. To select the Single trigger mode,
press the Single key.
Range Auto, Normal
Source
CH1
Mode
Auto
SINGLE
This section describes the basic operations required in capturing and viewing an input signal. For
more detailed operations, see the following chapters.
Measurements : from page 38•
Confi gurations: from page 56•
Activating a channel
Activating a
channel
To activate an input channel, press the Channel
key, CH1 or CH2. The channel indicator
appears at the left side of the display and the
channel icon changes accordingly.
CH 1CH 2
Press 5. When repeatedly to select the pulse
condition. Then use the Variable knob to set the
pulse width.
Condition > , < , = , ≠
Width 20ns ~ 10s
Press 6. Slope/Coupling to set trigger slope and
coupling.
Press 7. Slope repeatedly to select the trigger slope,
which also appears at the bottom of the display.
Range Rising edge, falling edge
Press Coupling repeatedly to select the trigger 8.
coupling.
Range DC, AC
Press Rejection to select the frequency rejection 9.
mode.
Range LF, HF, Off
Press Noise Rej to turn the noise rejection on or off.10.
Range On, Off
Press Previous menu to go back to the previous 11.
menu.
When <
20.0ns
VARIABLE
Slope /
Coupling
Slope
Coupling
AC
Rejection
Off
Noise Rej
Off
Previous
Menu
38
75
Confi guring the video trigger
ProcedurePress the Trigger menu key.1. MENU
Press 2. Type repeatedly to select video trigger. The
video trigger indicator appears at the bottom of the
display.
Press 3. Source repeatedly to select the trigger
source channel.
Range Channel 1, 2,
Press 4. Standard repeatedly to select the video
standard.
Range NTSC, PAL, SECAM
Press 5. Polarity repeatedly to select the video signal
polarity.
Range positive, negative
Press 6. Line repeatedly to select the video fi eld line.
Use the Variable knob to select the video line.
Field 1, 2
Video line NTSC: 1 ~ 262 (Even), 1 ~ 263 (Odd)
PAL/SECAM: 1 ~ 312 (Even),
1 ~ 313 (Odd)
Type
Video
Source
CH1
Standard
NTSC
Polarity
Line
VARIABLE
Confi guring the pulse width trigger
ProcedurePress the Trigger menu key.1. MENU
Channel 1 off Channel 1 on
Channel
indicator
Channel icon
De-activating a
channel
To de-activate the channel, press the Channel key twice (once if the channel menu
is already selected).
Using Autoset
BackgroundThe Autoset function automatically confi gures the panel settings to the best
ProcedureConnect the input signal to the oscilloscope and press 1.
viewing conditions, in the following way.
Selecting the horizontal scale•
Positioning the waveform horizontally•
Selecting the vertical scale•
Positioning the waveform vertically•
Selecting the trigger source channel•
Activating the channels•
Autoset
the Autoset key.
The waveform appears in the center of the display.2.
Press 2. Type repeatedly to select pulse width trigger.
The pulse width trigger indicator appears at the
bottom of the display.
74
Type
Pulse
39
Before Autoset After Auoset
Confi guring the edge trigger
ProcedurePress the Trigger menu key.1. MENU
Undo
Autoset
Undo option
Undoing the
Autoset
Adjusting the
trigger level
LimitationAutoset does not work in the following situation.
To undo the Autoset, press Undo (available
for a few seconds).
If the waveform is still unstable, try
adjusting the trigger level up or down by
using the Trigger Level knob.
• Input signal frequency less than 20Hz
• Input signal amplitude less than 30mV
Undo
LEVEL
Running and stopping the trigger
BackgroundIn the trigger Run mode, the oscilloscope constantly searches for a trigger condition
and updates the signal into the display when the condition is met.
In the trigger Stop mode, the oscilloscope stops triggering and thus the last
acquired waveforms stay in the display. The trigger icon at the top of the display
changes into Stop mode.
When the scope is running the number of points displayed on screen is limited
to 4000. Equivalent time sampling is always limited to 4000 points on screen at
any one time.
Press 2. Type repeatedly to select edge trigger.
Press 3. Source repeatedly to select the trigger
source.
Range Channel 1, 2, Line, Ext
Press 4. Mode repeatedly to select the Auto or
Normal trigger mode. To select the single trigger
mode, press the Single key.
Range Auto, Normal
Press 5. Slope/coupling to enter into the trigger
slope and coupling selection menu.
Press 6. Slope repeatedly to select the trigger slope,
rising or falling edge.
Range Rising edge, falling edge
Press 7. Coupling repeatedly to select the trigger
coupling, DC or AC.
Range DC, AC
Press 8. Rejection to select thefrequency rejection
mode.
Range LF, HF, Off
Press 9. Noise Rej to turn the noise rejection on
or off.
Range On, Off
Press 10. Previous menu to go back to the previous
menu.
Type
Edge
Source
CH1
Mode
Auto
SINGLE
Slope /
Coupling
Slope
Coupling
AC
Rejection
Off
Noise Rej
Off
Previous
Menu
40
73
Trigger slopeTriggers on the rising edge.
Triggers on the falling edge.
Pressing the Trigger Run/Stop key switches between the
Run and Stop mode.
Run/Stop
Trigger coupling AC
Frequency
rejection
Noise rejection Rejects noise signals.
Trigger level
Confi guring Holdoff
BackgroundHoldoff function defi nes the waiting period before GDS-1000A starts triggering
Panel operationPress the Trigger menu key twice.1. MENUMENU
DC
LF
HF
LEVEL
again after a trigger point. The holdoff function is especially useful for waveforms
with two or more repetitive frequencies or periods that can be triggered.
To set the Holdoff time, use the Variable knob. 2.
The resolution depends on the horizontal scale.
Range 40ns~2.5s
Pressing Set to Minimum sets the Holdoff time to the
minimum, 40ns.
Note: The holdoff function is automatically disabled when the waveform update
mode is in Roll mode.
Triggers only on AC component.
Triggers on AC+DC component.
Puts a high-pass fi lter and rejects the frequency below
50kHz.
Puts a low-pass fi lter and rejects the frequency above 50kHz.
Using the trigger level knob moves the trigger point up or
down.
Holdoff
40.0ns
Trigger Run mode Trigger Stop mode
Waveform
operation
Auto or
Trig icon
Waveforms can be moved or scaled in both the Run and Stop mode. For details,
see page 61 (Horizontal position/scale) and page 65 (Vertical position/scale).
Stop icon
Changing the horizontal position and scale
For more detailed confi gurations, see page 61.
Setting the
horizontal
position
The horizontal position knob moves the waveform
left or right.
The position indicator moves along with the waveform and the distance from the
center point is displayed as the offset in the upper side of the display.
72
41
Horizontal offset
Position indicator
The Auto trigger status appears in the upper right corner of
the display.
Selecting the
horizontal
scale
Range 1ns/Div ~ 10s/Div, 1-2.5-5 increment
To select the timebase (scale), turn the TIME/DIV
knob; left (slow) or right (fast).
To move the waveform up or down, turn the vertical
position knob for each channel.
SingleThe oscilloscope acquires the
NormalThe oscilloscope acquires and updates the input signals only
HoldoffThe holdoff function defi nes the waiting period before GDS-1000A starts trigger-
Video standard
(video trigger)
Sync polarity
(video trigger)
Video line
(video trigger)
Pulse condition
(pulse trigger)
ing again after a trigger point. The Holdoff function ensures a stable display.
NTSCNational Television System Committee
PALPhase Alternative by Line
SECAMSEquential Couleur A Mémoire
line1~263 for NTSC, 1~313 for PAL/SECAM
Sets the pulse width (20ns ~ 10s) and the triggering condition.
> Longer than = Equal to
< Shorter than ≠ Not equal to
input signals once when a trigger
event occurs, then stops acquiring.
Pressing the Single key triggers on the input signals again.
The Single trigger status appears in the upper right corner
of the display.
(Searching) (Triggered)
when a trigger event occurs.
The Normal trigger status appears in the upper right corner
of the display.
(Searching) (Triggered)
Positive polarity
Negative polarity
Selects the trigger point in the video signal.
fi eld 1 or 2
SINGLE
42
71
Trigger
The Trigger function confi gures the conditions by which the oscilloscope captures the incoming
signals.
Trigger type
EdgeTriggers when the signal crosses an amplitude threshold in either positive or
negative slope.
VideoExtracts a sync pulse from a video format signal and triggers on a specifi c line
or fi eld.
PulseTriggers when the pulse width of the signal matches the trigger settings.
Select vertical
scale
As the waveform moves, the vertical position of the cursor appears at the bottom
left corner of the display.
Run/Stop modeThe waveform can be moved vertically in
To change the vertical scale, turn
the VOLTS/DIV knob; left (down)
or right (up).
Range 2mV/Div ~ 10V/Div, 1-2-5 increments
The vertical scale indicator for each channel on the bottom left of the display
changes accordingly.
Stop mode In Stop mode, the vertical scale
setting can be changed but the
waveform shape stays the same.
Trigger mode AutoThe oscilloscope updates the input signal regardless of the
trigger conditions (if there is no trigger event, the oscilloscope generates an internal trigger). Select this mode especially when viewing rolling waveforms at a slow timebase.
70
Using the probe compensation signal
BackgroundThis section introduces how to use
Waveform type
View the probe
compensation
the probe compensation signal for
general usage, in case the DUT
signal is not available or to get a
second signal for comparison. For
probe compensation details, see
page 96.
Note that the frequency accuracy and duty factor are not guaranteed. Therefore
the signal should not be used for reference purpose.
Connect the probe between the compensation signal output and Channel 1.
input.
Square waveform used for probe compensation. 1k ~ 100kHz, 5% ~ 95%.
Demonstration signal for showing the effects
of peak detection.
43
Waveform
Press the Utility key.2. Utility
Use the variable knob to edit the voltage or current 3.
attenuation.
The voltage scale in the channel indicator changes accordingly. There is no 4.
change in the waveform shape.
Rangex1, x10, x100
NoteThe attenuation factor adds no infl uence on the real signal; it only changes the
voltage scale on the display.
VARIABLE
Probe
compensation
44
Press 3. ProbeComp.
Press Wave type repeatedly to 4.
select the wave type.
(For 5. only) To change
the frequency, press Frequency
and use the Variable knob.
Range 1kHz ~ 100kHzVARIABLE
(For 6. only) To change
the duty cycle, press Duty Cycle
and use the Variable knob.
Range 5% ~ 95%VARIABLE
For probe compensation details, see page 96.
ProbeComp
Menu
Wave Type
Frequency
1 K
Duty Cycle
50%
69
Limiting the waveform bandwidth
Automatic Measurements
BackgroundBandwidth limitation puts the input signal into a 20MHz (3dB) low-pass fi lter. This
ProcedurePress the Channel key.1. CH1
ExampleBW Limit Off BW Limit On
function is useful for cutting off high frequency noise to see the clear waveform
shape.
Press 2. BW Limit to turn on or off the limitation.
When turned on, the BW indicator appears next to
the Channel indicator in the display.
BW Limit
Off
Selecting the probe attenuation level
BackgroundA signal probe has an attenuation switch to lower the original DUT signal level
to the oscilloscope input range, if necessary. The probe attenuation selection
adjusts the vertical scale so that the voltage level on the display refl ects the real
value, not the attenuated level.
ProcedurePress the Channel key.1. CH1
Automatic measurement function measures input signal attributes and updates them in the display.
Up to 5 automatic measurement items can be updated at any one time on the side menus. All
automatic measurement types can be displayed on screen if necessary.
Difference between global
high and global low voltage
(=Vhi Vlo)
FRR
FRF
FFR
FFF
LRR
LRF
LFR
LFF
Press the Probe repeatedly to select the 2.
attenuation level.
68
Probe
x1
45
VloGlobal low voltage.
Press the Channel key.1. CH1
Time
measurement
items
Delay
measurement
items
Vavg
Vrms
ROVShoot
FOVShoot
RPREShoot
FPREShoot
Freq
Period
Risetime
Falltime
+Width
–Width
Duty CycleRatio of signal pulse
FRR
Averaged voltage of the fi rst cycle.
RMS (root mean square) voltage.
Rise overshoot voltage.
Fall overshoot voltage.
Rise preshoot voltage.
Fall preshoot voltage.
Frequency of the waveform.
Waveform cycle time
(=1/Freq).
Rising time of the pulse
(~90%).
Falling time of the pulse
(~10%).
Positive pulse width.
Negative pulse width.
compared with whole cycle
=100x (Pulse Width/Cycle)
Time between:
Source 1 fi rst rising edge and
Source 2 fi rst rising edge
Press F5 to toggle between Expand Center and 2.
Expand Ground.
To change the vertical scale, turn the VOLTS/DIV 3.
knob; left (down) or right (up).
The vertical scale indicator on the bottom left of the
display changes accordingly.
CH1
Expand
Center
VOLTS/DIV
100mV
Inverting the waveform vertically
ProcedurePress the Channel key.1. CH1
Press 2. Invert. The waveform becomes inverted
(upside down) and the Channel indicator in the
display shows a down arrow.
Original Inverted
Invert
Off
46
67
Press 2. Coupling repeatedly to select the
coupling mode.
Range
Coupling
DC coupling mode. The whole
portion (AC and DC) of the
signal appears on the display.
Ground coupling mode. The
display shows only the zero
voltage level as a horizontal
line. This mode is useful for
measuring the signal amplitude
with respect to the ground
level.
AC coupling mode. Only the AC
portion of the signal appears on
the display. This mode is useful
for observing AC waveforms
mixed with DC signal.
Expand Vertical Scale Centre / Ground
BackgroundNormally when the vertical scale is increased, the scaled image is centered from
ground. However a signal with a voltage bias could be obscured when the vertical scale is increased. The Expand Center function expands the image from the
center of the signal, rather than ground.
FRF
FFR
FFF
LRR
LRF
LFR
LFF
Time between:
Source 1 fi rst rising edge and
Source 2 fi rst falling edge
Time between:
Source 1 fi rst falling edge and
Source 2 fi rst rising edge
Time between:
Source 1 fi rst falling edge and
Source 2 fi rst falling edge
Time between:
Source 1 fi rst rising edge and
Source 2 last rising edge
Time between:
Source 1 fi rst rising edge and
Source 2 last falling edge
Time between:
Source 1 fi rst falling edge and
Source 2 last rising edge
Time between:
Source 1 fi rst falling edge and
Source 2 last falling edge
Automatically measuring the input signals
Viewing the
measurement
result
66
Press the Measure key.1. Measure
The measurement results appear on the menu bar, constantly updated. 5 2.
measurement slots (F1 to F5) can be customized.
47
Example
Main modeXY mode
Editing a
measurement
item
Change
measurement
item
Change
measurement
source
48
Press the corresponding menu key (3. F1~F5) to
select the measurement slot to be edited.
The editing menu appears.4.
Use the Variable knob to select a different 5.
measurement item.
Press 6. F1 repeatedly to change Source1 from
CH1 to CH2 or MATH.
Range CH1, 2, Math
Press 7. F2 repeatedly to change the channel
for Source2.
Range CH1, 2, Math
Voltage
Vpp
VARIABLE
Source 1
CH1
Source 2
CH2
XY mode
Vertical View (Channel)
The Vertical view section describes how to set the vertical scale, position, bandwidth limitation,
coupling mode, and attenuation.
Moving the waveform position vertically
ProcedureTo move the waveform up or down, turn the
Selecting the vertical scale
ProcedureTo change the vertical scale, turn the VOLTS/DIV
Range2mV/Div ~ 10V/Div, 1-2-5 increments
Selecting the coupling mode
ProcedurePress the Channel key.1. CH 1
vertical position knob for each channel.
VOLTS/DIV
knob; left (down) or right (up).
65
Example
Setting the zoom width Zooming in
Zooming width
Viewing waveforms in the X-Y mode
View all
measurements
Press 8. F3 to view all measurement items.
All the measurements appear in the center of 9.
the screen.
Voltage
Vpp
BackgroundThe X-Y mode compares the voltage of Channel 1 and Channel 2 waveforms in a
ProcedureConnect the signals to Channel 1 (X-axis) 1.
Adjusting the
X-Y mode
waveform
64
single display. This mode is useful for observing the phase relationship between
the two waveforms.
and Channel 2 (Y-axis).
Make sure both Channel 1 and 2 are 2.
activated.
Press the Horizontal key.3. MENU
Press XY. The display shows two waveforms 4.
in XY format; Channel 1 as Xaxis, Channel
2 as Y-axis.
Horizontal position CH1 Position knob
Horizontal scale CH1 Volts/Div knob
Vertical position CH2 Position knob
Vertical scale CH2 Volts/Div knob
CH1CH2
XY
Press 10. F3 to return.
NoteAll the editing operations can still be performed when viewing all the measurement
items.
Press Previous Menu to confi rm the item 11.
selection and to go back to the measurement
results view.
Previous
Menu
49
Cursor Measurements
Cursor lines, horizontal or vertical, show the precise position of the input waveforms or the math
operation results. The horizontal cursors can track time, voltage and frequency, whilst the vertical
cursors can track voltage. All measurements are updated in realtime.
Using the horizontal cursors
ProcedurePress the Cursor key. The cursors appear in 1.
the display.
Cursor
Selecting the
Roll mode
manually
Press the Horizontal menu key.1. MENU
Press 2. Roll. The horizontal scale automatically
becomes 250ms/div and the waveform starts
scrolling from the right side of the display (If the
oscilloscope is already in the Roll mode, there
will be no change).
Zooming the waveform horizontally
Roll
Press X Y to select the horizontal (X1&X2) 2.
cursor.
Press Source repeatedly to select the source 3.
channel.
Range CH1, 2, MATH
The cursor measurement results will appear in the menu, F2 to F4.4.
Parameters X1Time/Voltage position of the left cursor. (relative to zero)
X2Time/Voltage position of the right cursor. (relative to zero)
X1X2The distance between the X1 and X2.
The time difference between X1 and X2.
Moving the
horizontal
cursors
50
us
HzThe time distance converted to frequency.
VThe voltage difference. (X1-X2)
To move the left cursor, press X1 and then use
the Variable knob.
To move the right cursor, press X2 and then
use the Variable knob.
X Y
Source 1
CH1
X1
-5.000us
0.000uV
X1
5.000us
0.000uV
Procedure/
range
Press the Horizontal Menu key.1. MENU
Press 2. Window.
Use the horizontal position knob to move the 3.
zoom range sideways, and TIME/DIV knob to
change the zoom range width.
The width of the bar in the middle of the display is the actual zoomed area.
Zoom range 1ns ~ 25s
Press 4. Window Zoom. The specifi ed range gets
zoomed.
Window
TIME/DIV
Window
Zomm
63
Selecting the horizontal scale
Select horizontal
scale
To select the timebase (scale), turn the TIME/DIV knob; left
(slow) or right (fast).
Range 1ns/Div ~ 50s/Div, 1-2.5-5-10 increment The timebase indicator at
the bottom of the display updates the current horizontal scale.
TIME/DIV
Selecting the waveform update mode
BackgroundThe display update mode is switched automatically or manually according to the
horizontal scale.
Main modeUpdates the whole displayed waveform at once.
Roll modeUpdates and moves the waveform gradually from the right side of the display
The main mode is automatically selected when the horizontal scale (timebase)
is fast.
Horizontal scale ≤100ms/div
Trigger All modes available
to the left. The Roll mode is automatically selected when the horizontal scale
(timebase) is 250ms or greater. When in the Roll mode, an indicator appears at
the bottom of the display.
Main mode Roll mode
Timebase ≤250ms/div (V100Sa/s)
Trigger Auto mode only
To move both cursors at once, press X1X2
and then use the Variable knob.
Remove cursors Press Cursor to remove the onscreen
cursors.
X1X2
10.00us
100.0kHz
0.000uV
Cursor
Using the vertical cursors
ProcedurePress the Cursor key.1. Cursor
Press X Y to select the vertical 2.
(Y1&Y2) cursor.
Press Source repeatedly to select the 3.
source channel.
Range CH1, 2, MATH
The cursor measurement results will appear in the menu.4.
Press Source repeatedly to select the 5.
source channel.
Range CH1, 2, MATH
Parameters Y1Voltage level of the upper cursor
Y2Voltage level of the lower cursor
Moving the
vertical cursors
Y1Y2The voltage difference between the upper and lower
To move the upper cursor, press Y1 and then
use the Variable knob.
cursor
X Y
Source
CH1
Source
CH1
Y1
123.4mV
62
51
To move the lower cursor, press Y2 and then
use the Variable knob.
Y2
12.9mV
Press the grid icon repeatedly to select 2.
the grid.
Full
To move both cursors at once, press Y1Y2
and then use the Variable knob.
Remove cursors Press Cursor to remove the onscreen
cursors.
Y1Y2
10.5mV
Cursor
Parameters
Shows the full grid.
Shows the outer frame and X/Y axis.
Shows only the outer frame.
Horizontal View
The Horizontal view section describes how to confi gure the horizontal scale, position, waveform
update mode, window zoom, and X-Y mode.
Moving the waveform position horizontally
ProcedureThe horizontal position knob moves the waveform left or
right. The position indicator at the top of the display shows
the center and current position.
Center position Moving right
52
61
Press 2. Accumulate to turn on the
waveform accumulation.
To clear the accumulation and start it 3.
over (refresh), press Refresh.
ExampleAccumulation off Accumulation on
Accumulate
On
Refresh
Adjusting the display contrast
ProcedurePress the Display key.1. Display
Press 2. Contrast.
Turn the Variable knob left to lower the 3.
contrast (dark display) or right to raise
the contrast (bright display).
Contrast
VARIABLE
Selecting the display grid
ProcedurePress the Display key.1. Display
Math Operations
The Math operations can add, subtract, multiply or perform FFT/FFT RMS on the input waveforms.
The resulted waveform can be measured using the cursors, and saved or recalled just like normal
input signals.
Overview
Addition (+)Adds the amplitude of CH1 & CH2 signals.
Subtraction (–) Extracts the amplitude difference between CH1 & CH2.
Multiplication (×) Multiplies CH1 and CH2.
FFTPerforms a FFT calculation on a signal. Four types of FFT windows are available:
FFT RMSPerforms a FFT RMS calculation on a signal. RMS is similar to FFT, however
Hanning FFT
window
Flattop FFT
window
Rectangular
FFT window
Hanning, Flattop, Rectangular, and Blackman.
the amplitude is calculated as RMS and not dB. Four types of FFT windows are
available: Hanning, Flattop, Rectangular, and Blackman.
Frequency resolutionGood
Amplitude resolution Suitable for....Not good
Frequency resolutionFrequency measurement on
periodic waveforms
Frequency resolutionNot good
Amplitude resolutionGood
Suitable for....Amplitude measurement on
periodic waveforms
Frequency resolutionVery good
Amplitude resolutionBad
Suitable for....Single-shot phenomenon (this
mode is the same as having no
window at all)
60
53
Blackman FFT
window
Frequency resolutionBad
Amplitude resolutionVery good
Suitable for....Amplitude measurement on
periodic waveforms
Adding, subtracting or multiplying signals
ProcedureActivate both CH1 and CH2.1. CH1CH2
Press the Math key.2.
MATH
Real time vs Equivalent time sampling mode
BackgroundThe oscilloscope automatically switches between two sampling modes, Real-time
Real-time
sampling
Equivalent-time
sampling
and Equivalenttime, according to the number of active channels and sampling
rate.
Once sampled data is used to reconstruct a single waveform. Short-time events
might get lost if the sampling rate gets too high. This mode is used when the
sampling rate is relatively low (1GSa/s or lower).
Multiple numbers of sampled data are accumulated to reconstruct a single
waveform. ETS restores more waveform detail but takes longer to update the
waveform. This mode is used when the sampling rate becomes higher than
1GSa/s. The maximum equivalent-time sampling rate is 25GSa/s.
Display
Press Operation repeatedly to 3.
select addition (+), subtraction (–) or
multiplication (×).
Operation
CH1+CH2
The Display section describes how to confi gure the display settings:
drawing type, waveform accumulation, contrast adjustment, and grid settings.
Selecting vector or dot drawing
The math measurement result appears in 4.
the display.
To move the math result vertically, use 5.
the Variable knob. The position will be
displayed in Posistion.
To clear the math result from the display, 6.
press the Math key again.
54
Unit/Div
2V
Position
0.00 Div
VARIABLE
MATH
ProcedurePress the Display key.1. Display
Press 2. Type repeatedly to select the waveform
drawing.
TypesDots Only the sampled dots are displayed.
Vectors The sampled dots are connected by lines.
Type
Vectors
Accumulating the waveform
BackgroundAccumulation preserves the old waveform drawings and overwrites new waveforms
ProcedurePress the Display key.1.
on top of it. It is useful for observing waveform variation.
Display
59
Using the FFT function
ProcedurePress the Math key.1. MATH
Delay offWith delay off, the trigger point is fi xed. When the
ProcedurePress the Acquire key.1. Acquire
Press Delay On/Off to turn delay on of off.2.
Use the Horizontal Position knob to increase or 3.
decrease the delay time when Delay is set to On.
Use the Horizontal Position knob to increase or 4.
decrease the delay time when Delay is set to On.
horizontal scale is increased the screen will be
magnifi ed according to the trigger point.
Delay
On
TIME/DIV
Press 2. Operation repeatedly to select
FFT or FFT RMS
Press 3. Source repeatedly to select the
source channel.
Press 4. Window repeatedly to select the
FFT window type.
The FFT result appears. The horizontal scale changes from time to frequency, 5.
and the vertical scale from voltage to dB or RMS.
To move the FFT waveform vertically, 6.
use the Variable knob. The position will
be updated in Position.
Range –12.00 Div ~ +12.00 Div
To select the vertical scale of FFT 7.
waveform, press Unit/Div(FFT) or Volt/Div(FFT RMS) repeatedly.
Range 1, 2, 5, 10, 20 dB/Div
Voltage Volt/Div
To clear the FFT result from the display, 8.
press the Math key again.
Operation
FFT
Source
CH1
Window
Hanning
VARIABLE
Position
0.00 Div
Unit/Div
1dB
MATH
58
55
CONFIGURATION6.
ExampleThe peak detect mode reveals the occasional glitch.
The Confi guration chapter describes how to confi gure panel settings to make measurements and
observations suited to the application needs.
Acquisition
The acquisition process samples the analog input signals and converts them into digital format for
internal processing. You may select the normal, average, or peak detect acquisition mode.
Selecting the acquisition mode
ProcedurePress the Acquire key.1. Acquire
Select the acquisition mode 2.
between Normal, Average and
Peak Detect.
RangeNormalAll of the acquired data is used to
AverageMultiple data is averaged to form a
Peak detectTo activate the Peak detect mode, press
Normal
Average
Peak
Detect
draw the waveform.
waveform. This mode is useful for
drawing a noise-free waveform. To select
the number, press Average repeatedly.
Average number: 2, 4, 8, 16, 32, 64,
128, 256
Peak-Detect. Only the minimum and
maximum value pairs for each acquisition
interval (bucket) are used. This mode is
useful for catching abnormal glitches in
a signal.
Normal mode Peak detect mode
Selecting Delay mode
BackgroundWhen delay time is ON, the displayed output is delayed for a defi ned amount of
time from the trigger point. Using the delay function is useful for observing an
area of the waveform that occurs some time after the trigger point.
Delay onThe delay point will be located in the center of the
display. When changing the delay time, the screen
will stay centered on the delay point when delay time
is turned ON. Thus the delay point will be fi xed in the
center of the screen.
Delay Time
Trigger Point
Delay Point
When the horizontal scale is changed the delay point
will stay centered on the screen whilst the tripper point
will move (leftwards when magnifying).
56
57
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.