Velleman HHPS5 Service manual

Page 1
Page 2
Velleman
Legen Heirweg 33
9890 Gavere
Belgium
Internet Site: http://www.velleman.be
PersonalScope
TM
2
HHPS5 - 1999 - ED1
Page 3
PersonalScope
TM

CONTENTS

CONTENTS
FCC information for the USA ............................... 4
Important............................................................... 5
GENERAL............................................................. 6
Features................................................................. 5
Options.................................................................. 5
Technical specifications......................................... 6
SAFETY and WARNINGS.................................... 7
Symbols displayed on the unit................................ 7
USE......................................................................... 8
Survey of the connections and controls .................. 8
Power supply ......................................................... 9
Survey of the indications on the screen .................10
OPERATION........................................................11
Turning on/off the Personal Scope........................11
Resetting the Personal Scope.................................11
Adjusting the contrast...........................................11
Choice of input.....................................................11
Switching from one readout to another..................12
Measuring probe adjustment.................................13
Auto-setup function ..............................................13
Signal screen ........................................................14
Screen memory.....................................................15
Set-up function keys..............................................15
The markers .........................................................15
Setting up the voltage per division.......................16
Setting up the time base........................................15
Moving the signal in Y-position............................16
Trigger functions..................................................17
Setting up a X10 measuring probe ........................18
MAINTENANCE..................................................18
TROUBLESHOOTING .......................................19
The screen remains blank or there is no signal......19
RMS readout is incorrect ......................................19
No signal on the oscilloscope display....................19
Incorrect frequency readout...................................19
Voltage readout is not correct ...............................19
WARRANTY........................................................19
DIAGRAM............................................................20
3
Page 4
PersonalScope
TM

FCC information for the USA

This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a res­idential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harm­ful interference to radio communications. However, there is no guarantee that interference will not occur in a partic­ular installation. If this equipment does cause harmful in­terference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and re-
ceiver.
Connect the equipment into an outlet on a circuit dif-
ferent from that to which the receiver is connected.
Consult the dealer or an experienced radio/TV techni-
cian for help.
Important
This equipment was tested for FCC compliance under con­ditions that include the use of shielded test leads between it and the peripherals. It is important that you use shielded cables and connectors to reduce the possibility of causing Radio and Television interference. Shielded probes, suit-
able for the HPS5 oscilloscope can be obtained from the authorized Velleman dealer.
If the user modifies the HPS5 oscilloscope or its connec­tions in any way, and these modifications are not approved by Velleman, the FCC may withdraw the user’s right to operate the equipment.
The following booklet prepared by the Federal Communi­cations Commission may be of help: “How to identify and Resolve Radio-TV Interference problems”. This booklet is available from the US Government Printing Office, Wash­ington, DC20402 Stock No. 044-000-00345-4.
4
Page 5
PersonalScope
TM

GENERAL

GENERAL
The Velleman PersonalScope is not a graphical multimeter but a complete portable oscilloscope at the size and the cost of a good multimeter. Its high sensitivity - down to 5mV/ div - and extended scope functions make this unit ideal for hobby, service, automotive and development purposes. Because of its extreme value for money, the PersonalScope is well-suited for educational use in schools and colleges. Suitable for measurements on audio equipment, mains voltage applications, digital signals, all kind of sensors, signal analysis in automotive applications, car stereo etc… It’s ultra fast full auto setup function takes care of all the necessary settings and makes measuring waveforms very easy, a feature that will be appreciated both by experts and novices.

Features

High contrast LCD with wide viewing angle
Full auto set up for volt/div and time/div
Trigger mode: run - normal - once - slope -/+
True RMS and dB measurements
Readout: Vdc-Vpp-Vrms-dBm
x1 and x10 probe calculation option
Display mode: none - markers - grid - crosshair
Marker readout: dt - 1/dt (frequency) - dV
AC/DC input coupling selection
Automatic zero reference for DC measurements
Sceen hold fuction
Auto power off or continuous on
Low battery detection
Includes battery charge circuit

Options

Insulated measurement probe x1 / x10: PROBE60S
Soft carry case: CHPS5
Adapter: type PS905 for 230V / PS905USA for 110V
5
Page 6
GENERAL
PersonalScope

Technical specifications

Maximum sample rate 5MHz for repetitive signals (0.5MHz for single shot events) Input amplifier bandwidth 1MHz ( -3dB at 1V/div setting) Input impedance 1Mohm // 20pF (standard oscilloscope probe) Maximum input voltage 100Vpeak (AC + DC) Input coupling DC, AC and GND (GND for auto zero reference) Vertical resolution 8 bit (6 bit on LCD) Linearity
± 1bit
LCD Graphics 64 x 128 pixels, 64 x 96 pixels for signal dBm measurement (0 dB= 0.775V)
From -73dB tot +40dB (up to 60dB with X10 probe) ±0.5dB accuracy
True-rms AC measurement From 0.1mV to 80V (up to 400Vrms with X10 probe) 2.5% accuracy Peak peak and DC range 0.1mV to 180V (1mV to 600V met x10 probe) 2% accuracy Timebase range 20s, 10s, 5s, 2s, 1s, 0.5s, 0.2s, 0.1s, 50ms, 20ms, 10ms, 5ms, 2ms, 1ms, 0.5ms,
0.2ms, 0.1ms, 50us, 20us, 10us, 4us, 2us / div. Input sensitivity range (at x1) 5mV, 10mV, 20mV, 50mV, 0.1V, 0.2V, 0.4V, 1V, 2V, 4V, 8V, 20V /div. Probe calibration output Approx. 1KHz / 5Vtt Supply voltage 9VDC/ min 300mA adapter (unregulated) 12VDC if regulated Batteries (option) Alkaline type AA or Ni Cd / NiMH rechargeable ( 5 pcs required)
Charge current for rechargeable batteries 90mA Battery life Up to 20h with Alkaline batteries Operating temperature
0 to 50°C (32 to 122°F)
Dimensions 105x220x35mm (4.13x7.95x1.38”) Weight 395g (14oz.) ex. Batteries
TM
6
Page 7
PersonalScope
TM

SAFETY and WARNINGS

SAFETY and WARNINGS
Symbols displayed on the unit
!
Important safety information, see user manual.
Ground level
The PersonalScope is ideally suited for measurements of category II installations with pollution degree 1 and using a
maximum of 600V, in accordance with the IEC1010-1 norm. The enclosure is UL 94V0 listed
Consequently, all measurements should be avoided in case of polluted or very humid air. One should also refrain from
measuring conductors or installations that use voltages that exceed 600Vrms above ground level. CAT II indicates conformity for measurements of domestic installations.
The maximum input voltage for the connections of the unit stands at 100Vp (AC+DC)Do NOT open the enclosure while performing measurements.Remove all test leads before opening the enclosure in order to avoid electrical shock.Use a measuring probe with an insulated connector ( e.g. type PROBE60S) when measuring voltages exceeding 30V.
!
Remove all non-rechargeable batteries when connecting an AC/DC adapter !
7
Page 8
USE
USE
PersonalScope
TM

Survey of the connections and controls

1 - BNC input connector (max. input 100Vp AC+DC) 2 - Adapter connection (observe the polarity!) 3 - Battery compartment 4 - X10 probe testing signal behind battery cover 5 - Serial number 6 - Display 7 - Battery charge indication LED 8 - Reset push button 9 - Keyboard
1
5
123456
!
I/O
probe
test
3
2
9V DC
x10
4
8
Display
Display
mark 1-2
mark 1-2
Reset
Reset
Hold
Hold
P E R S O N A L
P E R S O N A L
6
trig
trig
Trigger
Trigger
mode
mode
AC/DC
AC/DC
S C O P E
S C O P E
Slope
Slope
Auto
Auto
t-V/div
t-V/div
9
Y-pos
Y-pos
probe
probe
x1/x10
x1/x1
Charge
Charge
HPS5
HPS5
7
8
Page 9
PersonalScope
TM
USE

Power supply

The HPS5 PersonalScope can be powered by means of an adapter, ordinary batteries or rechargeable ones. Rechargeable batteries are advised for intensive use. The greater the capacity of the rechargeable batteries, the greater the autonomy of the unit. Rechargeable batteries of the 500mA/h type will yield ap­prox. 5 hours of autonomy, while 800mA/h-types provide up to 8 hours of autonomy.
Attention:
Use a non-regulated adapter of 9VDC that supplies
a min. of 300mA. Observe the polarity and consult your retailer if necessary.
When using a regulated adapter, the user should
verify whether it can supply 12VDC.
Batteries (option):
The unit can be used with either ordinary alkaline bat-
teries or rechargeable ones e.g. our HR6 type (5 x).
Open the battery cover by loosening the screw:
Insert the batteries in the holders. Mind the polarity !
probe
test
9
Page 10
USE
PersonalScope
TM
Important: Do NOT perform measurements unless the
!
battery compartment is closed. The battery cover should only be removed for the calibration of the probe 's X10-position.
F When first using the rechargeable batteries or if the
batteries are completely discharged, the user should charge them for at least 12 hours before using the unit.
F When the unit is OFF, the charging time for
800mA/h type batteries will be approx. 12 hours.
F The “Charge” indication LED on the front panel
will light up when the batteries are being charged.
ATTENTION: Remove all non-rechargeable batteries when connecting the AC/DC adapter !
The message “Lo . bat” will flash in the bottom right cor­ner of your display when the batteries need replacement (or charging). Insufficient battery voltage may entail erro-
neous measuring results.
Survey of the indications on the screen :
Dot indicates the vertical position of the signal on the screen
Trigger opening indication (fixed in the middle of the screen)
Signal window with (possibly) the markers , cross or dot to indicate the various divisions.
Time per division or time between markers (when used)
Display of converted/calculated frequency 1/dt between markers (if present)
Trigger information or screen hold indica­tion.
Input-coupling indication : ac/dc/gnd and X10 probe setup
Measurement readouts, Vp (V peak-to-peak)
; Vr (Volt rms) ; dB (Decibels) ; V= (DC volt.)
Readout of the voltage between the mark­ers (when present)
Selected voltage per division Indication of the selected function (blue
keys) or battery-low indication
10
Page 11
PersonalScope
TM

OPERATION

OPERATION

Turning on/off the Personal Scope

Press the “ON/OFF” switch briefly to (de)activate the unit. The screen will briefly display the logo above the message “timer on”, which means that the unit will deacti­vate after eight minutes if no other key has been pressed during this interval. In order to bypass automatic deactivation : Keep the ON/OFF switch pressed down while activating the unit. The screen will briefly display the message “timer off”.
F When the unit is deactivated, the last setup will be
saved as long as the power is not cut and as long as the unit is not “reset” (see below).

Resetting the Personal Scope.

In case of unusual behaviour of the keypad or the display, it is possible to reactivate the unit by using the "Reset" function. The unit will subsequently return to the manufacturer-programmed setup. Press the sunk “Reset” push button for at least 10 seconds.
F Do not use a sharp tool !
Reset

Adjusting the contrast

Press the “contrast” key until the desired contrast is displayed. The user can either keep the key pressed down or press it repeatedly. When pressing this key, the contrast will always increase to maximum level before starting again at the lowest available level and working its way up again
AC/DC
the input can be connected to the signal through a direct link or by using a decoupling capacitor. Select DC for measuring DC voltage. Press the “AC/DC” key to select either AC or DC input coupling (see indications on the screen).

Choice of input

Depending on (part of) the signal to be measured,
F When measuring the ”ripple” of a DC voltage : put
the input on AC to limit the measurement to the AC component of the signal.
11
Page 12
OPERATION
PersonalScope
TM

Switching from one readout to another

The Personal Scope offers many voltage measur­ing possibilities. Press the “readout” key repeatedly to select one of four possible readouts :
Peak voltage (Vp). The signal's peak-to-peak voltage
(difference between highest and lowest value) is displayed.
True RMS readout (Vr). The area of the AC wave is calculated and converted to voltage. Please remember that a correct readout invariably consists of at least one or two visible periods (one can also use the auto-setup function).
dB measurement (dB). The measured signal is converted to dBm (0dB=0.775V). Once more, the user should make sure that at least one or two periods are displayed or select the auto-setup function).
DC voltage measurement (V=). This function enables the user to measure DC voltages (only for DC input coupling)
F Useful tip for measuring DC voltages:
The readout can be set at zero (reference) for any position on the screen by keeping the AC/DC key pressed down. Always use the "run" trigger mode for DC voltage measurement.
NOTE: If the signal goes off-screen or when the dB-
signal is too small for measurement, the
display will show.....
12
Page 13
PersonalScope
TM
OPERATION
probe x1/x10

Measuring probe adjustment

We advise the user to use a measuring probe equipped with a X10 setting when high or unknown voltages or high impedances need to be measured. When using the measuring probe in the X10 position, the measuring impedance will increase to 10Mohm, thereby
reducing the mea­suring instrument's charge on the cou-
x1 x10
pling. Press the “x1/x10
x1 x10
key in order to ad­just the display to the setup of the measuring probe. Upon choosing X10, “X10” will be
indicated on the screen and all measuring values will be adjusted accord­ingly.
IMPORTANT : X10 measuring probes should be cali­brated. Read “Setting up a X10 measuring probe".
Auto

Auto-setup function

The auto-setup function is ideally suited for quick measurements as no manual setup has to be made and everything is automatic.
F Use the auto-setup function when the screen no
longer displays a signal after the manual setup.
Connect the signal to be checked to the BNC input by
means of a special measuring lead or measuring probe.
Activate the unit.
Press “Auto” in order to (de)activate the auto-setup
function. When “auto setup” is engaged, the Volt/div and Time/div readouts are highlighted. The auto setup function will provide correct measurements for repetitive and stable signals of max. 400KHz.
The display may read as follows :
13
Page 14
OPERATION
IMPORTANT: Put the appropriate measuring probe in the X10 position for measuring high
!
voltages (>100Vp + DC).
NOTE: The Y-position does not work in auto-setup, which
is why the trigger setup remains in run or normal mode.
The auto-setup function will deactivate when using
the Trigger , Volt and time per division and Y position functions. (see further).
GRID
PersonalScope
TM
Display
divisions on your screen or in positioning movable markers on the screen. Press the “Display” key repeatedly in order to engage the various aids:
A grid divides the screen into reference points.
A cross inserts a co-ordinate system into the screen.
Markers: Moveable markers in order to measure the
signal (see below for use).
Empty : no aids.
14

Signal screen

The signal screen is an aid in displaying the
CROSSHAIR
MARKERS
Page 15
PersonalScope
TM
OPERATION
Hold

Screen memory

It may be useful to "freeze" certain signals on your screen. This will allow the user to study the signal or perform measurements with the markers.
Press ‘Hold’ to freeze the screen.
F The screen is also saved when the unit is deacti-
vated.
F The memory will be erased if “reset” is pressed or if
the battery voltage is too low.

Set-up function keys

HPS5
The selected function is dis­played in the bottom right
mark 1-2
Reset
Hold
Display
trig
Trigger
Trigger
mode
mode
AC/DC
P E R S O N A L
S C O P E
t-V/div
Y-pos
probe x1/x10
Charge
Slope
Slope
Auto
hand corner of the screen.
The four blue function keys work in conjunction with the arrow keys.
mark 1-2

The markers

The user can perform measurements on a certain
signal by using the four moveable markers.
This can be useful when measuring the interval between two points or the ampli­tude of any given peak.
The following indications will appear on the screen :
The time interval between two vertical
markers. The calculated frequency (primarily used for
the measurement of periods). The voltage between two horizontal mark-
ers.
The markers can be moved through the arrow keys. Keeping the key pressed down will move the marker quickly, pressing it briefly will move the marker by 1 position. The “mark 1-2” key is used to select the desired marker.
15
Page 16
OPERATION
PersonalScope
TM
F Determining the frequency of a signal requires the
measurement of a period. The easiest way to do this would be by placing the vertical markers either on two consecutive peaks or two identical slopes of a signal.
t-V/div
screen vertically by adjusting the displayed voltage per division. (V/div = voltage per division). The divisions can be made visible through the Display key (see page 16). Press the “t-V/div” key Press the or the key to select the sensitivity : from 5mV to a maximum of 20V per division.
F 50mV to 200V with X10 probe selection
t-V/div
fewer periods of a signal (t/div = time per division). The divisions can be visualised through the Display key (see page 16). Press the “t-V/div” key Press the ⇐ the ⇒ key to set the time base between 20s and 2µS per division.

Setting up the voltage per division

The user can enlarge or reduce the signal on the

Setting up the time base

Adjusting the time base will visualise more or
screen will be adjusted once the trigger level is reached.
NOTE: Use the minimum time base (2 microsec) as a starting point when measuring a signal and select longer time bases until the signal is displayed properly. Other­wise the display may not correctly reflect the signal under measurement due to aliasing.
Y-pos
Press the key or the key in order to move the signal in the direction of the arrows. Prolonged pressing will make the Y- position change faster. A small dot in the left-hand corner of the screen will indicate the direction in which the signal has moved. In this manner, the user will know in which direction the signal was going when it went off­screen.

Moving the signal in Y-position

Press the “Y-pos” key
ATTENTION : the positions 10µs, 4µs and 2µs only operates correctly with periodic signals, for which the unit will automatically select the ‘normal’ trigger mode. The
16
Page 17
PersonalScope
TM
OPERATION
trig

Trigger functions

The Personal Scope is equipped with a number of "trigger" aids that are used when the signal can be displayed on-screen.
Press the “trig” key.
The (trigger mode) key enables the user to select the appropriate triggering method, the (slope) key is used to select the trigger slope (see below).
Press the key repeatedly to choose between :
“run” mode: The signal will always appear on the
screen, regardless of the level involved. This position is used most frequently and should always be used for measuring DC voltages in particular.
“norm” mode: The screen will be adjusted pro­vided that the signal falls within the trigger level (interruption in the vertical line on the left-hand side of your screen). Use this mode when you want to start displaying the signal when it reaches a preset threshold value.
“once” mode: The screen is adjusted one single time, viz. when the trigger level is reached. Use this
mode to detect e.g. a short, once-only voltage peak.
ATTENTION : in the positions 10µs, 4µs and 2µs the unit will automatically select the ‘norm’ trigger mode. The trigger level has to be reached for the screen to be adjusted.
F Both the and the key remain active in order
to move the signal along the Y-axis, which enables the user to programme a trigger reference
Repeatedly press the key to choose between :
Triggering on the rising slope of the signal. The screen
will only display the signal when a positive slope is "detected", viz. the signal has to rise in vertical direc-
tion in order to trigger. Triggering on the falling slope of the signal : The screen
will only display the signal when a negative slope is "detected", viz. the signal has to drop in vertical direc-
tion in order to trigger.
Trigger on rising slope
NOTE : There is a time difference of ±12µs between the trigger point (the interruption in the vertical line on the left-hand side of the screen) and the start of the actual display. This is specifically true on higher frequencies.
Trigger on falling slope
17
Page 18
OPERATION

Setting up a X10 measuring probe

When used in the X10 position, a measuring probe should always be callibrated to the measuring instrument being used, in this case the Personal Scope.
Open the battery cover.
Set the probe input to X10
Set the voltage per division to 1V
Set the time per division to 0.2ms
Select AC for the input.
Use the probe to perform measurements at the preselected point under the battery cover
x10
Adjust the trimmer of the measuring probe in order to obtain a square wave signal
probe
test
with a top that is as flat as possible.
TM
NG
PersonalScope
NG
OK

MAINTENANCE

Clean the display with a shammy. NEVER use a dustcloth or paper in order to avoid scratches. The rest of the unit can be cleaned with a soft, dry cloth. NEVER use water to clean the unit.
18
Page 19
PersonalScope
TM

TROUBLESHOOTING

TROUBLESHOOTING
The screen remains blank or there is no signal :
No power supply
Batteries are discharged
Contrast adjustment is incorrect
Press RESET for at least 10 seconds
REMARK: Temporarily remove both batteries and adapter in case RESET is not yet operational.
RMS readout is incorrect :
Make sure that at least 1 and preferably even 2 periods
are displayed.
The batteries are discharged.
No signal on the oscilloscope display :
Time/div setting is in the wrong position. Try 1ms or
choose auto-setup mode.
The unit is in the Hold position
Trigger function is set in the “once” position
The programmed trigger level is not reached (choose
“run” mode)
Y position is wrong
The input signal is too high, change the volt/div. setting
or choose auto-setup.
Incorrect frequency readout:
An incorrect time/div. setting has been chosen. (start at
2µs)
Voltage readout does not correspond with the actual value :
The measuring probe is in the X10 position
The batteries are discharged
The zero refrence is not set correctly for DC measure-
ments.
Take into account that the RMS readout only measures
the AC voltage

WARRANTY

This product carries a one-year warranty as far as the craftsmanship and possible flaws in the materials are concerned. The warranty expires ONE YEAR after the date of purchase. The warranty will only apply if the unit is wrapped in the original packing material and either presented to VELLEMAN-KIT N.V or to an official distributor together with a copy of the original purchasing document. VELLEMAN-KIT N.V. is under the obligation to repair defects and flaws, but is free to either replace or repair defective parts. The warranty does not apply to software, fuses, measuring probes and batteries. VELLEMAN-KIT N.V will not be held responsible for any flaw or defect which the company feels is due to negligence on behalf of the user, to modification or opening of the unit, or to accidents or abnormal use or treatment of the product. VELLEMAN-KIT N.V will not reimburse the transport costs or risks, the costs for removing and replacing the product or any other costs that are directly or indirectly related to the defect. VELLEMAN-KIT N.V. accepts no liability for whatever damages may be caused by a malfunctioning product.
19
Page 20
POWER SUPPLY SECTION
SK1
DC JACK
SK2
+
1 2
­BATTERY
PRLL4001
GND
D7
D13
PRLL4001
T2
BDP31
GND GNDGND
D6
PRLL4001
R1 15
PWR
GND
FUSE 2A
PMBT2222A
R2 330
LD1
LED3RL
R33
22K
PersonalScope
+V1
R17 10K
LOW BAT
ZD1
PMBT2222A
DGND
SPR
TP4
L4 220µ
C25
+
100µ
T6
1
SCREENING
1
GND
C26
+
AGND
100µ
+V1 +V6
+V2
-V1
-V2
C27
+
100µ
AGNDGND
T12
R35 4K7
R16
10K
+
GND
100µ
100µ
L1 220µ
+
C20
100µ
AGNDDGND
C3
+
R32 1K
C4
C5
+
100µ
GND
3
C7
100n
JP1 0
L2
220µL3220µ
+
C24
100µ
AGND
D1
BAT54SD2BAT54S
Personal Scope: Power Supply Section
R34 22K
DGND
+
C23
100µ
AGND
T13 PMBT2907A
PMBZ5240B
JP2 0
D11
BAS85
LM2940S5.0(3) IC1
IN1OUT
GND2GND
GND
T11
PMBT2907A
C2
+
100µ
GND
T4
PMBT2222A
GND
PMBT2907A
4
+V3
T1
BDP32
R5 47K
F1
GND PWM
R60 47K
+V1
GND
D12
BAV99
R3 330
T5
R15 10K
R26
1K
+
C1 100µC6100n
T3
PMBT2222A
R27 1K
TM
20
Page 21
R50 100K
R41 39K-1%
R42 910K-1%
R43
510-1%
R6 47K-1%
R40 100-1%
R44 2K4-1%
T7
+V1
DGND
C38 1n5/200V
TM
RY1
EDR301A05
PMBT2222A
C29
BAV99
D4
100µ
AGND
AGND
AGND
R20 10K
+
12 14 15 11
1 5 2 4
BAV99
D5
AC/DC
+V6
R28 1K
+V4
16
IC3 MC74HC4052
X0
X
X1
VDD
X2
Y
X3
Y0
A Y1 Y2
B Y3
INH
VSS8VEE
6
7
DGND
13 3
10 9
C12
PersonalScope
SK3 BNC
C35
R39 100
47n/200V
AGND
CV1
20p
J1
Transient
C36
330p
CV2
C30
50p
47p
C37
10n
AGND
AGND
CV1: Transient adjustment at 2V/div CV2: Transient adjustment at 1V/div RV1: Offset adjustment (switch between 1 & 2V/div) RV2: Gain adjust at 0.1V/div (first adjust RV3 !)
+V1
D9
EDR301A05
RY2
C39
8 14
RY3
EDR301A05
R49 33K
100n
DGND DGND
ANALOG SECTION
BAV70
RV2
YPOS
OS NULL-
OS NULL+ IN+
IN-
IC4
100
4
-V2
OUT
VCC -
LF357N(8)
R36 4K7-1%
R51 47-1%
R53 180-1%
R55 560-1%
R45 2K4-1%
+V2
7
VCC+
8
NC
6
C4022p
R54 27-1%
R52 47-1%
R56 750-1%
R30 1K
AGND
TP3
1
ANALOG OUT
AD IN
C43
68p
AGND
C11
100n
C28
100µ
AGND
12 14 15 11
1 5 2 4
+
+V4
16
IC2 MC14052B
X0
X
X1
VCC
X2
Y
X3
Y0
A Y1 Y2
B Y3
INH
VEE
GND
6
7
8
DGND
ZD2
PLVA2656A
AGND
13 3
AGND
10 9
C8
100nC9100n
R29 1K
R47 33K
DGND DGND
-V2
+V1
R38 4K7
ATT4
R46
ATT5
33K
PMBT2222A
T8
DGND
18p
17
R19 10K
26 13
T9
PMBT2222A
R48 33K
C13
100n
DGND
BAS45AL
D8
BAV99
C10
R7
100n
AGND
+V6
D3
AGND
R18 10K
ATT1
ATT2 ATT3
4K7-1%
47K
RV1 22K
1
5 3
2
R37
Personal Scope: Analog input section
21
Page 22
DIGITAL SECTION
RV3: Callibration at 50mV/Div RV4: Center setting in auto setup mode
+V6
R57
4.7
+
C21
C14
REFTS REFT
REFB REFBS
AN IN
TLC5510INS
+V1
2
IN+ IN- 3OUT
5
BAL
LM311N(8)
8
IC5
VDDA14VDDA15VDDA18VDDD
AGND20DGND
AGND21DGND
AGND DGND
R58 1M
IC6
V+
BAL/STRB
GND
4
1
DGND
100n
4µ7
AGND AGND
16
CALIBRATION
R4 100
C17
100n
AD IN
17
23 22
19
R21
10K
RV3
C16
100n
500
AGND
R22
R23
10K
10K
AGND
TP1
1
CENTER REF
+V1
11
VDDD
V-
13
DGND
OE D0
D1 D2 D3 D4 D5 D6 D7
CLKIN
2
24
6 7
PLVA2656A
ZD4
DGND
CENTER YPOS
1
7
TM
TP2
1
YPOS
PersonalScope
C32
470n
470n
DGND
RV4
R10 47K
C31
AGND
DGND
R13 47K
C33
470n
R31 1K
17 16 15
10 11 12 13 14
LCD1
4 6
7 8 9
5
47K CENTER
Auto center
C19
100n
RST CS2 CS1 D/I
E DB0 DB1 DB2 DB3 DB4 DB5 DB6 DB7 R/W
PG12864
J2
3 2 5 6
LM358M(8)
DGND
DGND
IC8
R14 270K
R11 22K
+V2
8
IN1+
OUT1
V+
IN1­IN2+
OUT2
IN2-
V-
4
-V2
C34
470n
+V1
2
3
VDD
CONTRAST
19
A
20
K
VSS
VEE
1
18
-V1
R9 47K
AGND
DATABUS[DB0...DB7]
+V3
R24
T10
10K
C22
100µ
ZD3
DGND
DGND
PMBTA14
C18
+
100n
DGND
SW1
DGND
Mark
PLVA2656A
Contr.UpRightProbeX1
DisplayInput
BZD27-C6V8
R8
C42
22p
10K
DGND DGND
2 3
AC/DC
4
ATT1
5
ATT2
6
ATT3
7
ATT4
8
ATT5
36 41 38 39 37 42 43 44
ZD5
DGND
1 17 28 40
PIC16C65A20/L(44)
PLVA2656A ZD6
DGND
C15
100n
1
AD OE
3
DB0
4
DB1
5
DB2
6
DB3
7
DB4
8
DB5
9
DB6
10
DB7
12
AD CLK
Ypos
Meter
ZD7
DGND
PLVA2656A
+V1
R59 1K
tVdiv
Trig
Down Left Hold
AutoOnOff
PLVA2656A
+V1
ZD8
IC7
OSC1/CLKIN14OSC2/CLKOUT MCLR/VPP RA0 RA1 RA2 RA3 RA4/T0CKI RA5/SS RB0/INT RB4 RB2 RB3 RB1 RB5 RB6 RB7 NC NC NC NC
DGND
BAS85
35
RC0/T1OSO/T1CKI
VDD12VDD
VSS13VSS
34
DGND
D10
X1
20Mc C41
RC1/T1OSI/CCP2
RC2/CCP1
RC3/SCK/SCL
RC4/SDI/SDA
RC5/SDO RC6/TX/CK RC7/RX/DT
RD0/PSP0 RD1/PSP1 RD2/PSP2 RD3/PSP3 RD4/PSP4 RD5/PSP5 RD6/PSP6 RD7/PSP7
RE0/RD
RE1/WR
RE2/CS
22p
15 16
LOW BAT
18 19 20 25 26 27
PWR
29 21
DB0
22
DB1
23
DB2
24
DB3
30
DB4
31
DB5
32
DB6
33
DB7
9 10
AD OE
11
AD CLK
+V1
R25 10K
SK4
1
PROBE TEST
PWM
R12 47K
Personal Scope: Digital Section
22
Page 23
®
Legen Heirweg 33, 9890 Gavere
Belgium Europe
+32 (0)9 3843611 http://www.velleman.be
HHPS5 - 1999 - ED1
France
VELLEMAN ELECTRONIQUE +33 (0) 320158615
Netherlands
VELLEMAN COMPONENTS +31 (0) 765147563
USA
VELLEMAN INC. +1 (817) 284 7785
Loading...