VEGA BAR87 User Manual

Page 1
Operating Instructions
Pressure transmitter with metallic measuring cell
VEGABAR 87
Slave sensor for electronic differential pres­sure
Document ID: 45053
Page 2
Contents
1 About this document
1.1 Function ........................................................................................................................... 4
1.2 Target group ..................................................................................................................... 4
1.3 Symbolism used ............................................................................................................... 4
2 For your safety
2.1 Authorised personnel ....................................................................................................... 5
2.2 Appropriate use ................................................................................................................ 5
2.3 Warning about incorrect use ............................................................................................. 5
2.4 General safety instructions ............................................................................................... 5
2.5 CE conformity ................................................................................................................... 5
2.6 Measuring range - permissible process pressure ............................................................. 6
2.7 Environmental instructions ............................................................................................... 6
3 Product description
3.1 Conguration .................................................................................................................... 7
3.2 Principle of operation........................................................................................................ 8
3.3 Packaging, transport and storage ................................................................................... 11
3.4 Accessories and replacement parts ............................................................................... 12
4 Mounting
4.1 General instructions to use the instrument ..................................................................... 13
4.2 Ventilation and pressure compensation .......................................................................... 14
4.3 Combination Master - Slave ........................................................................................... 15
4.4 Dierential pressure measurement ................................................................................ 16
4.5 Interface measurement .................................................................................................. 16
4.6 Density measurement .................................................................................................... 17
4.7 External housing ............................................................................................................ 19
5 Connecting to power supply
5.1 Preparing the connection ............................................................................................... 20
5.2 Connecting ..................................................................................................................... 20
5.3 Single chamber housing ................................................................................................. 21
5.4 External housing with version IP 68 (25 bar)................................................................... 22
5.5 Connection example ...................................................................................................... 24
6 Set up with the display and adjustment module
6.1 Parameter adjustment - Quick setup .............................................................................. 25
6.2 Parameter adjustment - Extended adjustment................................................................ 28
7 Diagnosis, asset management and service
7.1 Maintenance .................................................................................................................. 38
7.2 Rectify faults ................................................................................................................... 38
7.3 Exchanging the electronics module ................................................................................ 38
7.4 Exchange process assembly with version IP 68 (25 bar) ................................................ 38
7.5 How to proceed in case of repair .................................................................................... 40
8 Dismounting
8.1 Dismounting steps.......................................................................................................... 41
8.2 Disposal ......................................................................................................................... 41
9 Supplement
9.1 Technical data ................................................................................................................ 42
45053-EN-131011
2
VEGABAR 87 • Slave sensor for electronic dierential pressure
Page 3
9.2 Dimensions .................................................................................................................... 47
Safety instructions for Ex areas
Please note the Ex-specic safety information for installation and op­eration in Ex areas. These safety instructions are part of the operating instructions manual and come with the Ex-approved instruments.
45053-EN-131011
Editing status: 2013-10-09
VEGABAR 87 • Slave sensor for electronic dierential pressure
3
Page 4

1 About this document

1 About this document

1.1 Function

This operating instructions manual provides all the information you need for mounting, connection and setup as well as important instruc­tions for maintenance and fault rectication. Please read this informa­tion before putting the instrument into operation and keep this manual
accessible in the immediate vicinity of the device.

1.2 Target group

This operating instructions manual is directed to trained specialist personnel. The contents of this manual should be made available to these personnel and put into practice by them.

1.3 Symbolism used

Information, tip, note
This symbol indicates helpful additional information.
Caution: If this warning is ignored, faults or malfunctions can result.
Warning: If this warning is ignored, injury to persons and/or serious
damage to the instrument can result.
Danger: If this warning is ignored, serious injury to persons and/or
destruction of the instrument can result.
Ex applications
This symbol indicates special instructions for Ex applications.
List
The dot set in front indicates a list with no implied sequence.
Action
→
This arrow indicates a single action.
1 Sequence of actions
Numbers set in front indicate successive steps in a procedure.
Battery disposal
This symbol indicates special information about the disposal of bat­teries and accumulators.
45053-EN-131011
4
VEGABAR 87 • Slave sensor for electronic dierential pressure
Page 5

2 For your safety

2 For your safety

2.1 Authorised personnel

All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator.
During work on and with the device the required personal protective equipment must always be worn.

2.2 Appropriate use

VEGABAR 87 is a slave sensor for electronic dierential pressure
measurement.
You can nd detailed information on the application range in chapter
"Product description".
Operational reliability is ensured only if the instrument is properly used according to the specications in the operating instructions manual as well as possible supplementary instructions.

2.3 Warning about incorrect use

Inappropriate or incorrect use of the instrument can give rise to application-specic hazards, e.g. vessel overll or damage to system components through incorrect mounting or adjustment.

2.4 General safety instructions

This is a high-tech instrument requiring the strict observance of stand­ard regulations and guidelines. The user must take note of the safety
instructions in this operating instructions manual, the country-specic installation standards as well as all prevailing safety regulations and accident prevention rules.
The instrument must only be operated in a technically awless and reliable condition. The operator is responsible for trouble-free opera-
tion of the instrument. During the entire duration of use, the user is obliged to determine the
compliance of the necessary occupational safety measures with the
current valid rules and regulations and also take note of new regula­tions.

2.5 CE conformity

The device fullls the legal requirements of the applicable EC guide­lines. By axing the CE marking, we conrm successful testing of the product.
You can nd the CE Certicate of Conformity in the download section of our homepage.
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
5
Page 6
2 For your safety

2.6 Measuring range - permissible process pressure

The permissible process pressure is specied on the type label with "process pressure", see chapter "Conguration". For safety reasons, this range must not be exceeded. This applies also if a measuring cell with higher measuring range (order-related) than the permissible pressure range of the process tting is installed.

2.7 Environmental instructions

Protection of the environment is one of our most important duties.
That is why we have introduced an environment management system
with the goal of continuously improving company environmental pro­tection. The environment management system is certied according
to DIN EN ISO 14001.
Please help us fulll this obligation by observing the environmental
instructions in this manual:
Chapter "Packaging, transport and storage"
Chapter "Disposal"
45053-EN-131011
6
VEGABAR 87 • Slave sensor for electronic dierential pressure
Page 7
Type plate

3 Product description

3 Product description
3.1 Conguration
The nameplate contains the most important data for identication and
use of the instrument:
1
2
3
4 5
6
7 8 9
10
Fig. 1: Layout of the type label (example)
1 Instrument type
2 Product code
3 Field for approvals 4 Power supply and signal output, electronics 5 Protection rating 6 Measuring range 7 Permissible process pressure 8 Material, wetted parts 9 Hardware and software version 10 Order number 11 Serial number of the instrument 12 Symbol of the device protection class 13 ID numbers, instrument documentation 14 Reminder to observe the instrument documentation 15 Notied authority for CE marking 16 Approval directive
16
15
14
13
12
11
Serial number
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
The type label contains the serial number of the instrument. With it you can nd the following data on our homepage:
Product code of the instrument (HTML)
Delivery date (HTML)
Order-specic instrument features (HTML)
Operating instructions at the time of shipment (PDF)
Order-specic sensor data for an electronics exchange (XML)
Test certicate pressure transmitters (PDF)
Go to www.vega.com, "VEGA Tools" and "Serial number search".
As an alternative, you can nd the data via your Smartphone:
Download the smartphone app "VEGA Tools" from the "Apple App
Store" or the "Google Play Store"
Scan the Data Matrix code on the type label of the instrument or
Enter the serial number manually in the app
7
Page 8
3 Product description
Scope of this operating instructions manual
Scope of delivery
Electronicdierential
pressure measurement
Measured variables
This operating instructions manual applies to the following instrument
versions:
Hardware from 1.0.0
Software version from 1.0.0
The scope of delivery encompasses:
VEGABAR 87 pressure transmitter
Connection cable to the master sensor
Cable gland for the master sensor
Documentation
this operating instructions manualTest certicate measuring accuracy (optional)Supplementary instructions manual "Plug connector for con-
tinuously measuring sensors" (optional)
Ex-specic "Safety instructions" (with Ex versions) – if necessary, further certicates

3.2 Principle of operation

The VEGABAR 87 slave sensor is combined with a sensor from the
VEGABAR 80 series for electronic dierential pressure measurement.
The sensors are connected via a screened four-wire cable. The measured value from the slave sensor is read in and factored into the calculations. Parameter adjustment is carried out through the master sensor.
The dierential pressure measurement is suitable for the measure­ment of the following process variables:
Level
Flow
Dierential pressure
Density
Interface
45053-EN-131011
8
VEGABAR 87 • Slave sensor for electronic dierential pressure
Page 9
3 Product description
Fig. 2: Electronic level dierence measurement through Master/Slave combina­tion
Application area
The VEGABAR 87 is a pressure transmitter for pressure and level measurements of liquids with higher temperatures in the chemical, food processing and pharmaceutical industry.
Measured products
Measured products are liquids.
Depending on the instrument version and the measurement setup, the measured products can be also viscous.
Measuring system
The process pressure acts on the sensor element via the stainless steel diaphragm and an internal transmission liquid. The process pressure causes a resistance change which is converted into a cor­responding output signal and outputted as measured value.
The METEC® measuring cell is the measuring unit. It consists of
the ceramic-capacitive CERTEC® measuring cell and a special, temperature-compensated isolating system.
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
9
Page 10
3 Product description
4321
Fig. 3: Conguration of the METEC® measuring cell with VEGABAR 87
1 Process diaphragm 2 Isolating liquid 3 FeNi adapter 4 CERTEC® measuring cell
Pressure types
Measuring system
The measuring cell design depends on the selected pressure type.
Relative pressure: the measuring cell is open to atmosphere. The ambient pressure is detected in the measuring cell and compensated. It thus has no inuence on the measured value.
Absolute pressure: the measuring cell is evacuated and encapsu­lated. The ambient pressure is not compensated and does hence inuence the measured value.
Relative pressure, climate-compensated: the measuring cell is
evacuated and encapsulated. The ambient pressure is detected through a reference sensor in the electronics and compensated. It thus has no inuence on the measured value.
The process pressure acts on the sensor element via the stainless steel diaphragm and an internal transmission liquid. The process pressure causes a resistance change which is converted into a cor­responding output signal and outputted as measured value.
The METEC® measuring cell is the measuring unit. It consists of
the ceramic-capacitive CERTEC® measuring cell and a special, temperature-compensated isolating system.
45053-EN-131011
10
VEGABAR 87 • Slave sensor for electronic dierential pressure
Page 11
Packaging
3 Product description
4321
Fig. 4: Conguration of the METEC® measuring cell with VEGABAR 87
1 Process diaphragm 2 Isolating liquid 3 FeNi adapter 4 CERTEC® measuring cell

3.3 Packaging, transport and storage

Your instrument was protected by packaging during transport. Its
capacity to handle normal loads during transport is assured by a test
based on ISO 4180.
The packaging of standard instruments consists of environment­friendly, recyclable cardboard. For special versions, PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies.
Transport
Transport inspection
Storage
Storage and transport temperature
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
Transport must be carried out in due consideration of the notes on the
transport packaging. Nonobservance of these instructions can cause
damage to the device.
The delivery must be checked for completeness and possible transit
damage immediately at receipt. Ascertained transit damage or con­cealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and
stored according to the orientation and storage markings on the outside.
Unless otherwise indicated, the packages must be stored only under
the following conditions:
Not in the open
Dry and dust free
Not exposed to corrosive media
Protected against solar radiation
Avoiding mechanical shock and vibration
Storage and transport temperature see chapter "Supplement -
Technical data - Ambient conditions" Relative humidity 20 … 85 %
11
Page 12
3 Product description
Protective cap

3.4 Accessories and replacement parts

The protective cover protects the sensor housing against soiling and
intense heat from solar radiation.
You will nd additional information in the supplementary instructions
manual "Protective cover" (Document-ID 34296).
Flanges
Welded socket
Electronics module
Screwed anges are available in dierent versions according to the
following standards: DIN 2501, EN 1092-1, BS 10, ANSI B 16.5, JIS B 2210-1984, GOST 12821-80.
You can nd additional information in the supplementary instructions
manual "Flanges according to DIN-EN-ASME-JIS" (Document-ID
31088).
Welded sockets are used to connect the sensors to the process.
You can nd further information in the supplementary instructions
"Welded socket VEGABAR series 80" (Document-ID 45082).
The electronics module VEGABAR series 80 is a replacement part for pressure transmitters of VEGABAR series 80. There is a dierent version available for each type of signal output.
You can nd further information in the operating instructions "Elec-
tronics module VEGABAR series 80" (Document-ID 45054).
12
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
Page 13
Suitability for the process conditions

4 Mounting

4 Mounting

4.1 General instructions to use the instrument

Make sure that all parts of the instrument exposed to the process are suitable for the existing process conditions.
These are mainly:
Active measuring component
Process tting
Process seal
Process conditions are particularly:
Process pressure
Process temperature
Chemical properties of the medium
Abrasion and mechanical inuences
You can nd the specications of the process conditions in chapter "Technical data" as well as on the nameplate.
Protection against mois­ture
Screwing in
Vibrations
Temperature limits
Protect your instrument further through the following measures
against moisture penetration:
Use the recommended cable (see chapter "Connecting to power
supply") Tighten the cable gland
Loop the connection cable downward in front of the cable gland
This applies particularly to:
Outdoor mounting
Installations in areas where high humidity is expected (e.g. through
cleaning processes)
Installations on cooled or heated vessels
On instruments with process tting thread, the hexagon must be tight­ened with a suitable screwdriver. Wrench size see chapter "Dimen-
sions".
Warning:
The housing must not be used to screw the instrument in! Applying tightening force can damage internal parts of the housing.
In case of strong vibrations at the mounting location, the instrument
version with external housing should be used. See chapter "External housing".
Higher process temperatures often mean also higher ambient temperatures. Make sure that the upper temperature limits stated in chapter "Technical data" for the environment of the electronics hous-
ing and connection cable are not exceeded.
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
13
Page 14
4 Mounting
Filter elements
2
1
Fig. 5: Temperature ranges
1 Process temperature 2 Ambient temperature

4.2 Ventilation and pressure compensation

Ventilation and pressure compensation are carried out with VE­GABAR 87 via a lter element. It is air permeable and moisture-
blocking.
Caution:
The lter element causes a time-delayed pressure compensation. When quickly opening/closing the housing cover, the measured value can change for approx. 5 s by up to 15 mbar.
For eective ventilation, the lter element must always be free of buildup.
Caution:
Do not use a high-pressure cleaner. The lter element could be dam­aged, which would allow moisture into the housing.
The following paragraphs describe how the lter element is arranged in the dierent instrument versions.
Instruments in non-Ex and Ex-ia version
14
The lter element is mounted into the electronics housing. It has the
following functions:
Ventilation electronics housing
Atmospheric pressure compensation (with relative pressure meas-
uring ranges)
VEGABAR 87 • Slave sensor for electronic dierential pressure
45053-EN-131011
Page 15
4 Mounting
4 4 4
1 2 3
Fig. 6: Position of the lter element - non-Ex, Ex-ia version
1 Housing plastic, stainless steel precision casting 2 Housing aluminium 3 Housing stainless steel, electropolished 4 Filter element
With the following instruments a blind plug is installed instead of the lter element:
Instruments in protection IP 66/IP 68 (1 bar) - ventilation via capil-
laries in x connected cable Instruments with absolut pressure
Instruments in IP 69K version
The lter element is mounted into the electronics housing. It has the
following functions:
Ventilation electronics housing
Atmospheric pressure compensation (with relative pressure meas-
uring ranges)
1
Fig. 7: Position of the lter element - IP 69K version
1 Filter element
Instruments with absolute pressure have a blind plug mounted instead of the lter element.

4.3 Combination Master - Slave

In principle, any sensor combination within the VEGABAR 80 series is allowed. The following requirements must be fullled:
Conguration, Master sensor suitable for electronic dierential
pressure Pressure type is identical for both sensors, i.e. relative pressure/
relative pressure or absolute pressure/absolute pressure The master sensor must always measure the higher pressure
Measurement setup as shown in the following chapters
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
15
Page 16
4 Mounting
Measurement setup
4.4 Dierentialpressuremeasurement
With the Master/Slave combination, a gauge dierent measurement can be realized.
Measurement setup
1
Fig. 8: Measurement setup with a level dierence measurement
2
3

4.5 Interface measurement

With the Master/Slave combination also an interface measurement
can be realised.
Requirements for a functioning measurement are:
Vessel with changing level
Products with steady densities
Interface always between the measurement points
Total level always above the upper measurement point
16
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
Page 17
4 Mounting
Measurement setup
1
Fig. 9: Measurement setup with interface measurement, h = distance between the two measuring points
1 VEGABAR 87 2 VEGABAR 87 - Slave sensor
The interface measurement is possible in open as well as in closed
vessels.
2
0,8
1,0
h

4.6 Density measurement

With the Master/Slave combination also a density measurement can
be realised.
Requirements for a functioning measurement are:
Vessel with changing level
Product with consistent density
Distance between the measurement points as large as possible
Level always above the upper measuring point
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
17
Page 18
4 Mounting
1
Fig. 10: Measurement setup with density measurement, h = distance between the two measuring points
1 VEGABAR 87 2 VEGABAR 87 - Slave sensor
2
h
Density measurement is possible in open as well as in closed ves­sels. Small changes in the density cause only small changes in the
measured dierential pressure. So a suitable measuring range has to
be selected.
18
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
Page 19
Conguration
Mounting
4 Mounting

4.7 External housing

1
2
3
4
Fig. 11: Arrangement measurement loop, external housing
1 Vessel 2 Sensor 3 Connection cable sensor - external housing 4 External housing 5 Signal cable
1. Mark the holes according to the following drilling template
2. Fasten wall mounting plate with 4 screws
90 mm (3.54")
70 mm (2.76")
3mm
(0.12")
8 mm
(0.32")
(3.66")
(4.33")
93 mm
110 mm
Fig. 12: Drilling template - wall mounting plate
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
R 3,5 mm
(0.14")
19
Page 20

5 Connecting to power supply

Safety instructions
5 Connecting to power supply

5.1 Preparing the connection

Always keep in mind the following safety instructions:
Connect only in the complete absence of line voltage
If overvoltage surges are expected, overvoltage arresters should
be installed
Voltage supply
Cable gland ½ NPT
Connection technology
Connection procedure
The voltage supply and the signal transmission are carried out via the
four-wire, screened connection cable from the master sensor.
You can nd the data for this signal circuit in chapter "Technical data".
With plastic housing, the NPT cable gland or the Conduit steel tube
must be screwed without grease into the threaded insert. Max. torque for all housings see chapter "Technical data".

5.2 Connecting

The voltage supply and signal output are connected via the spring-
loaded terminals in the housing.
The connection to the display and adjustment module or to the inter­face adapter is carried out via contact pins in the housing.
Information:
The terminal block is pluggable and can be removed from the
electronics. To do this, lift the terminal block with a small screwdriver
and pull it out. When reinserting the terminal block, you should hear it snap in.
Proceed as follows:
1. Unscrew the housing cover
2. If a display and adjustment module is installed, remove it by turn­ing it slightly to the left.
3. Loosen compression nut of the cable entry
4. Remove approx. 10 cm (4 in) of the cable mantle, strip approx.
1 cm (0.4 in) of insulation from the ends of the individual wires
5. Insert the cable into the sensor through the cable entry
20
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
Page 21
5 Connecting to power supply
Fig. 13: Connection steps 5 and 6
6. Insert the wire ends into the terminals according to the wiring plan
Information:
Solid cores as well as exible cores with wire end sleeves are insert­ed directly into the terminal openings. In case of exible cores without end sleeves, press the terminal from above with a small screwdriver; the terminal opening is freed. When the screwdriver is released, the
terminal closes again.
You can nd further information on the max. wire cross-section under
"Technical data/Electromechanical data"
7. Check the hold of the wires in the terminals by lightly pulling on
them
8. Connect the screen to the internal ground terminal, connect the
outer ground terminal to potential equalisation
9. Tighten the compression nut of the cable entry. The seal ring must completely encircle the cable
10. Reinsert the display and adjustment module, if one was installed
11. Screw the housing cover back on
The electrical connection is hence nished.

5.3 Single chamber housing

The following illustration applies to the non-Ex, Ex-ia and Ex-d ver-
sion.
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
21
Page 22
5 Connecting to power supply
Electronics and connec­tion compartment
Master
Overview
5678
4
1
2
Fig. 14: Wiring plan VEGABAR 87 Slave sensor
1 To the sensor 2 Ground terminal for connection of the cable screen
1)

5.4 External housing with version IP 68 (25 bar)

22
Fig. 15: VEGABAR 87 in IP 68 version 25 bar, non-Ex and axial cable outlet, external housing
1)
Connect screen here. Connect ground terminal on the outside of the housing
to ground as prescribed. The two terminals are galvanically connected.
VEGABAR 87 • Slave sensor for electronic dierential pressure
45053-EN-131011
Page 23
5
2
( )
+
(-)
6
8
A
Electronics and connec­tion compartment for power supply
Terminal compartment, housing socket
5 Connecting to power supply
1
4...20m
1
2
3
Fig. 16: Electronics and connection compartment
1 Electronics module 2 Cable gland for voltage supply 3 Cable gland for connection cable, transmitter
45053-EN-131011
1234
5
Fig. 17: Connection of the sensor in the housing base
1 Yellow 2 White 3 Red 4 Black 5 Shielding 6 Breather capillaries
VEGABAR 87 • Slave sensor for electronic dierential pressure
2
341
6
23
Page 24
5 Connecting to power supply
Electronics and connec­tion compartment
Master
Connection example, slave sensor
5678
4
1
2
Fig. 18: Wiring plan VEGABAR 87 Slave sensor
1 To the sensor 2 Ground terminal for connection of the cable screen
2)

5.5 Connection example

4...20mA
( )
+
(-)
1
2
5
678
3
4
1
Fig. 19: Connection example VEGABAR 87 slave sensor
1 VEGABAR 87 2 VEGABAR 87 slave sensor 3 Connection cable 4 Supply and signal circuit VEGABAR 87
The connection between VEGABAR 87 and the VEGABAR 87 slave sensor is made with a standard cable according to the chart:
VEGABAR 87 Slave sensor
Terminal 5 Terminal 5
Terminal 6 Terminal 6
Terminal 7 Terminal 7
Terminal 8 Terminal 8
5678
4
2
24
2)
Connect screen here. Connect ground terminal on the outside of the housing
to ground as prescribed. The two terminals are galvanically connected.
VEGABAR 87 • Slave sensor for electronic dierential pressure
45053-EN-131011
Page 25
Presettings

6 Set up with the display and adjustment module

6 Set up with the display and adjustment
module

6.1 Parameter adjustment - Quick setup

To quickly and easily adapt the sensor to the application, select
the menu item "Quick setup" in the start graphic on the display and adjustment module.
Carry out the following steps in the sequence specied below. The presettings apply to all applications.
You can nd "Extended adjustment" in the next sub-chapter.
1. Measurement loop name
In the rst menu item you assign a suitable measurement loop name.
Permitted are names with max. 19 characters.
2. Application
In this menu item you activate/deactivate the slave sensor for elec­tronic dierential pressure and select the application.
Note:
It is absolutely necessary to activate the slave sensor in advance to
have the applications displayed in the electronic dierential pressure
measurement menus.
3. Units
In this menu item you determine the adjustment and temperature units of the instrument. Depending on the selected application in the menu item "Application", dierent adjustment units are available.
Quick setup - Level meas­urement
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
4. Unit, static pressure
In this menu item, you determine the unit of the static, i.e. superim­posed pressure.
5. Position correction
In this menu item you compensate the inuence of the installation position of the instrument (oset) to the measured value.
6. Min. adjustment
In this menu item you carry out the min. adjustment for level.
Enter the percentage value and the corresponding pressure value for
the min. level.
7. Max. adjustment
In this menu item you carry out the max. adjustment for level.
25
Page 26
6 Set up with the display and adjustment module
Enter the percentage value and the corresponding pressure value for
the max. level.
The quick setup for level measurement is nished.
Quick setup - Flow meas­urement
4. Position correction
In this menu item you compensate the inuence of the installation position of the instrument (oset) to the measured value.
5. Min. adjustment
In this menu item you carry out the min. adjustment for ow.
Enter the percentage value and the corresponding pressure value for the min. ow.
6. Max. adjustment
In this menu item you carry out the max. adjustment for ow.
Enter the percentage value and the corresponding pressure value for the max. ow.
7. Linearization
In this menu item, you select the characterics of the output signal.
Quicksetup-Dierential
pressure measurement
26
The quick setup for ow measurement is nished.
4. Unit, static pressure
In this menu item, you determine the unit of the static, i.e. superim­posed pressure.
5. Position correction
In this menu item you compensate the inuence of the installation position of the instrument (oset) to the measured value.
6. zero adjustment
In this menu item you carry out the zero adjustment for the dierential pressure.
Enter the respective pressure value for 0 %.
7. span adjustment
In this menu item you carry out the span adjustment for the dierential pressure
Enter the respective pressure value for 100 %.
VEGABAR 87 • Slave sensor for electronic dierential pressure
45053-EN-131011
Page 27
Quick setup - Density measurement
Quick setup - Interface measurement
6 Set up with the display and adjustment module
The quick setup for dierential pressure measurement is nished.
4. Unit, static pressure
In this menu item, you determine the unit of the static, i.e. superim­posed pressure.
5. Position correction
In this menu item you compensate the inuence of the installation position of the instrument (oset) to the measured value.
6. Distance
In this menu item, you enter the installation distance between master and slave sensor.
7. Min. adjustment
In this menu item you carry out the min. adjustment for density.
Enter the percentage value and the corresponding density value for
the min. density.
8. Max. adjustment
In this menu item you carry out the max. adjustment for density.
Enter the percentage value and the corresponding density value for
the max. density.
The quick setup for density measurement is nished.
4. Unit, static pressure
In this menu item, you determine the unit of the static, i.e. superim­posed pressure.
5. Position correction
In this menu item you compensate the inuence of the installation position of the instrument (oset) to the measured value.
6. Distance
In this menu item, you enter the installation distance between master and slave sensor.
7. Min. adjustment
In this menu item, you carry out the adjustment for the min. height of the interface.
Enter the percentage value and the corresponding height of the
interface.
8. Max. adjustment
In this menu item, you carry out the adjustment for the max. height of the interface.
Enter the percentage value and the corresponding height of the
interface.
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
27
Page 28
6 Set up with the display and adjustment module
The quick setup for interface measurement is nished.

6.2 Parameter adjustment - Extended adjustment

For technically demanding measurement loops you can carry out
extended settings in "Extended adjustment".
Main menu
The main menu is divided into ve sections with the following func­tions:
Setup: Settings, e.g., for measurement loop name, application, units, position correction, adjustment, signal output
Display: Settings, e.g., for language, measured value display, lighting Diagnosis: Information, e.g. on instrument status, pointer, measure-
ment reliability, simulation
Additional adjustments: PIN, date/time, reset, copy function Info: Instrument name, hardware and software version, date of manu-
facture, sensor features In the main menu item "Setup", the individual submenu items should
be selected one after the other and provided with the correct param­eter values.
The following submenu points are available:
In the following section, the menu items from the menu "Setup" for
electronic dierential pressure measurement are described in detail. Depending on the selected application, dierent sections are relevant.
Information:
Further menu items of the menu "Setup" as well as the complete menus "Display", "Diagnosis", "Additional adjustments" and "Info"
are described in the operating instructions of the respective master
sensor.
45053-EN-131011
Setup - Application
28
In this menu item you activate/deactivate the slave sensor for elec­tronic dierential pressure and select the application.
VEGABAR 87 • Slave sensor for electronic dierential pressure
Page 29
6 Set up with the display and adjustment module
The following applications are available:
Level
Flow
Dierential pressure
Density
Interface
Note:
It is absolutely necessary to activate the slave sensor in advance to
have the applications displayed in the electronic dierential pressure
measurement menus.
Enter the requested parameters via the appropriate keys, save your
settings with [OK] and jump to the next menu item with the [ESC] and the [->] key.
Setup - Units
Setup - Position correc­tion
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
In this menu item, you determine the units for the "Min. adjustment/ zero" and "Max. adjustment/span" as well as the static pressure.
If the level should be adjusted in a height unit, the density of the me­dium must also be entered later during the adjustment.
In addition, the unit for the measuring cell and electronics temperature
is determined in the menu item "Peak values".
Enter the requested parameters via the appropriate keys, save your
settings with [OK] and jump to the next menu item with the [ESC] and the [->] key.
Particularly with chemical seal systems, the installation position of the instrument can shift the measured value (oset). The position correction compensates this oset. Hence the actual measured value is accepted automatically. With relative pressure measuring cells also a manual oset can be carried out. With a master/slave combination, the position correction is carried out for both sensors.
If the actual measured value should be taken over as correction value
during automatic position correction, this value must not be inu­enced by product coverage or static pressure.
With the manual position correction, the oset value can be deter­mined by the user. Select for this purpose the function "Edit" and enter the requested value.
29
Page 30
6 Set up with the display and adjustment module
Save your settings with [OK] and move with [ESC] and [->] to the next menu item.
After the position correction is carried out, the actual measured value is corrected to 0. The corrective value appears with an inverse sign as oset value in the display.
The position correction can be repeated as often as necessary. How-
ever, if the sum of the corrective values exceeds 20 % of the nominal
measuring range, then no position correction is possible.
Parameterization example
VEGABAR 87 always measures pressure independently of the pro­cess variable selected in the menu item "Application". To output the
selected process variable correctly, an allocation to 0 % and 100 % of the output signal must be carried out (adjustment).
With the application "Interface" the hydrostatic pressure at min. and
max. height of the interface is entered for adjustment. See the follow-
ing example:
100%
2
0,8
1,0
Fig. 20: Parameter adjustment example "Min./max. adjustment, interface measurement"
1 Min. interface = 0 % correspond to 0.0 mbar 2 Max. interface = 100 % correspond to 490.5 mbar 3 VEGABAR 87 4 VEGABAR 87 - Slave sensor
")
2 m
(78.74
h
0%
1
If these values are not known, an adjustment with interface layers of for example 10 % and 90 % is also possible. By means of these set­tings, the layer of the actual interface is then calculated.
The real product level during the adjustment is not important,
because the adjustment is always carried out without changing the
product level. These settings can be made ahead of time without the
instrument having to be installed.
45053-EN-131011
30
VEGABAR 87 • Slave sensor for electronic dierential pressure
Page 31
6 Set up with the display and adjustment module
Note:
If the adjustment ranges are exceeded, the entered value will not be accepted. Editing can be interrupted with [ESC] or corrected to a value within the adjustment ranges.
Setup - Min. adjustment Level
Setup - Max. adjustment Level
Proceed as follows:
1. Select the menu item "Setup" with [->] and conrm with [OK]. Now select with [->] the menu item "Adjustment", then "Min.
adjustment" and conrm with [OK].
2. Edit the percentage value with [OK] and set the cursor to the requested position with [->].
3. Set the requested percentage value (e.g. 10 %) with [+] and save with [OK]. The cursor jumps now to the pressure value.
4. Enter the pressure value corresponding to the min. level (e.g.
0 mbar).
5. Save settings with [OK] and move with [ESC] and [->] to the max. adjustment.
The min. adjustment is nished.
For an adjustment with lling, simply enter the actual measured value indicated at the bottom of the display.
Proceed as follows:
1. Select with [->] the menu item "Max. adjustment" and conrm with [OK].
2. Edit the percentage value with [OK] and set the cursor to the requested position with [->].
3. Set the requested percentage value (e.g. 90 %) with [+] and save with [OK]. The cursor jumps now to the pressure value.
4. Enter the pressure value for the full vessel (e.g. 900 mbar) cor­responding to the percentage value.
5. Save settings with [OK]
The max. adjustment is nished.
For an adjustment with lling, simply enter the actual measured value indicated at the bottom of the display.
Setup - Min. adjustment,
ow
Proceed as follows:
1. Select the menu item "Setup" with [->] and conrm with [OK]. Now select with [->] the menu item "Min. adjustment" and conrm with [OK].
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
31
Page 32
6 Set up with the display and adjustment module
2. Edit the mbar value with [OK] and set the cursor to the requested position with [->].
3. Set the requested mbar value with [+] and store with [OK].
4. Change with [ESC] and [->] to the span adjustment
The min. adjustment is nished.
For an adjustment with pressure, simply enter the actual measured value indicated at the bottom of the display.
Setup - Max. adjustment,
ow
Setup - zero adjustment,
dierentialpressure
Proceed as follows:
1. Select with [->] the menu item "Max. adjustment" and conrm with [OK].
2. Edit the mbar value with [OK] and set the cursor to the requested
position with [->].
3. Set the requested mbar value with [+] and store with [OK].
The max. adjustment is nished.
For an adjustment with pressure, simply enter the actual measured value indicated at the bottom of the display.
Proceed as follows:
1. Select the menu item "Setup" with [->] and conrm with [OK]. Now select with [->] the menu item "zero adjustment" and conrm with [OK].
2. Edit the mbar value with [OK] and set the cursor to the requested
position with [->].
3. Set the requested mbar value with [+] and store with [OK].
4. Change with [ESC] and [->] to the span adjustment
The zero adjustment is nished.
Information:
The zero adjustment shifts the value of the span adjustment. The span, i.e. the dierence between these values, however, remains
unchanged.
For an adjustment with pressure, simply enter the actual measured value indicated at the bottom of the display.
45053-EN-131011
32
VEGABAR 87 • Slave sensor for electronic dierential pressure
Page 33
6 Set up with the display and adjustment module
Setup - span adjustment,
dierentialpressure
Setup - Distance, density
Proceed as follows:
1. Select with [->] the menu item "Span adjustment" and conrm with [OK].
2. Edit the mbar value with [OK] and set the cursor to the requested
position with [->].
3. Set the requested mbar value with [+] and store with [OK].
The span adjustment is nished.
For an adjustment with pressure, simply enter the actual measured value indicated at the bottom of the display.
Proceed as follows:
1. Select the menu item "Setup" with [->] and conrm with [OK]. Now select with [->] the menu item "Distance" and conrm with [OK].
2. Edit the sensor distance with [OK] and set the cursor to the
requested position with [->].
3. Set the requested distance with [+] and store with [OK].
The adjustment of the distance is hence nished.
Setup - Min. adjustment, density
Proceed as follows:
1. Select the menu item "Setup" with [->] and conrm with [OK]. Now select with [->] the menu item "Min. adjustment" and conrm with [OK].
2. Edit the percentage value with [OK] and set the cursor to the requested position with [->].
3. Set the requested percentage value with [+] and save with [OK]. The cursor jumps now to the density value.
4. Enter the min. distance suitable for the percentage value.
5. Save settings with [OK] and move with [ESC] and [->] to the max. adjustment.
The min. adjustment for density is nished.
Setup - Max. adjustment,
Proceed as follows:
density
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
33
Page 34
6 Set up with the display and adjustment module
1. Select the menu item "Setup" with [->] and conrm with [OK]. Now select with [->] the menu item "Max. adjustment" and con-
rm with [OK].
2. Edit the percentage value with [OK] and set the cursor to the requested position with [->].
3. Set the requested percentage value with [+] and save with [OK]. The cursor jumps now to the density value.
4. Enter the max. density value matching the percentage value.
The max. adjustment for density is nished.
Setup - Min. adjustment ­Interface
Setup - Max. adjustment ­Interface
Proceed as follows:
1. Select the menu item "Setup" with [->] and conrm with [OK]. Now select with [->] the menu item "Min. adjustment" and conrm with [OK].
2. Edit the percentage value with [OK] and set the cursor to the requested position with [->].
3. Set the requested percentage value with [+] and save with [OK]. The cursor jumps now to the height value.
4. Enter the min. height of the interface suitable for the percentage
value.
5. Save settings with [OK] and move with [ESC] and [->] to the max. adjustment.
The min. adjustment, interface is hence nished.
Proceed as follows:
1. Select the menu item "Setup" with [->] and conrm with [OK]. Now select with [->] the menu item "Max. adjustment" and con-
rm with [OK].
2. Edit the percentage value with [OK] and set the cursor to the requested position with [->].
3. Set the requested percentage value with [+] and save with [OK]. The cursor jumps now to the height value.
4. Enter the max. height of the interface matching the percentage
value.
The max. adjustment for interface is nished.
45053-EN-131011
34
VEGABAR 87 • Slave sensor for electronic dierential pressure
Page 35
6 Set up with the display and adjustment module
Additional adjustments
- Reset
With a reset, certain parameter adjustments carried out by the user
are reset.
The following reset functions are available:
Delivery status: Restoring the parameter settings at the time of shipment from the factory incl. the order-specic settings. A user-pro­grammable linearization curve as well as the measured value memory
will be deleted.
Basic settings: Resetting the parameter settings incl. special parameters to the default values of the respective instrument. A user programmable linearization curve as well as the measured value
memory is deleted. The following table shows the default values of the instrument. De-
pending on the instrument version or application, all menu items may not be available or some may be dierently assigned:
Reset - Setup
Menu item Parameter Default value
Measurement loop name
Application Slave for elec-
tronic dierential pressure
Application No reset
Units Unit of measure-
ment
Static pressure bar
Position correc­tion
Adjustment Distance (with
density and inter­face)
Zero/Min. adjust­ment
Span/Max. adjust­ment
Damping Integration time 0.0 s
Linearization Linear
Sensor
No reset
mbar (with nominal measuring range
≤400 mbar)
bar (with nominal measuring rang-
es ≥1 bar)
0.00 bar
1.00 m
0.00 bar
0.00 %
Nominal measuring range in bar
100.00 %
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
35
Page 36
6 Set up with the display and adjustment module
Menu item Parameter Default value
Current output Current output -
Lock adjustment Last setting
Reset - Display
Menu item Default value
Menu language Order-specic
Displayed value 1 Current output in %
Displayed value 2 Measuring cell temperature in °C
Backlight Switched o
Reset - Diagnosis
Menu item Parameter Default value
Sensor status -
Peak value Pressure Actual measured value
Simulation -
Mode
Current output ­Min./Max.
Temperature Actual temperature values from meas-
Output characteristics
4 … 20 mA Reaction when malfunction occurs
≤ 3.6 mA
3.8 mA
20.5 mA
uring cell, electronics
36
Reset - Additional settings
Menu item Parameter Default value
PIN 0000
Date/Time Actual date/Actual time
Copy in­strument settings
Special pa­rameters
Scaling Scaling size Volume in l
Scaling format 0 % corresponds to 0 l
Current out­put
HART mode Address 0
Current output - Size Lin. percent - Level
Current output - adjustment 0 … 100 % correspond to
-
No reset
100 % corresponds to 0 l
4 … 20 mA
VEGABAR 87 • Slave sensor for electronic dierential pressure
45053-EN-131011
Page 37
6 Set up with the display and adjustment module
Menu item Parameter Default value
Eective
pressure transmitter
Unit m3/s
Adjustment 0.00 % correspond to 0.00
m3/s
100.00 %, 1 m3/s
Setup - Characteristics
values,eectivepressure
transmitter
In this menu item, the units for the eective pressure transmitter are determined as well as the selection of mass or volume ow.
Furthermore the adjustment for the volume or mass ow rate at 0 % or
100 % is carried out.
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
37
Page 38

7 Diagnosis, asset management and service

7 Diagnosis, asset management and service

7.1 Maintenance

Maintenance
Reaction when malfunc­tions occur
If the instrument is used properly, no special maintenance is required in normal operation.
In some applications, product buildup on the diaphragm can inuence the measuring result. Depending on the sensor and application, take precautions to ensure that heavy buildup, and especially a hardening
thereof, is avoided.

7.2 Rectify faults

The operator of the system is responsible for taking suitable meas­ures to rectify faults.
Procedure for fault recti-
cation
Reaction after fault recti-
cation
24 hour service hotline
The rst measures are:
Evaluation of fault messages, for example via the display and
adjustment module
Checking the output signal
Treatment of measurement errors
Further comprehensive diagnostics options are available with a PC
with PACTware and the suitable DTM. In many cases, the reasons can
be determined in this way and faults rectied.
Depending on the reason for the fault and the measures taken, the steps described in chapter "Setup" must be carried out again or must be checked for plausibility and completeness.
Should these measures not be successful, please call in urgent cases the VEGA service hotline under the phone no. +49 1805 858550.
The hotline is also available outside normal working hours, seven days a week around the clock.
Since we oer this service worldwide, the support is provided in
English. The service itself is free of charge, the only costs involved are the normal call charges.

7.3 Exchanging the electronics module

In case of a defect, the user can replace the electronics module with another one of identical type.
In Ex applications, only instruments and electronics modules with ap­propriate Ex approval may be used.
If there is no electronics module available on site, one can be ordered from the agency serving you.
7.4 Exchange process assembly with version
IP 68 (25 bar)
With version IP 68 (25 bar), the user can exchange the process as­sembly on site. Connection cable and external housing can be kept.
45053-EN-131011
38
VEGABAR 87 • Slave sensor for electronic dierential pressure
Page 39
7 Diagnosis, asset management and service
Required tools:
Hexagon socket wrench, size 2
Caution:
The exchange may only be carried out in the complete absence of
line voltage.
In Ex applications, only a replacement part with appropriate Ex ap­proval may be used.
Proceed as follows when carrying out the exchange:
1. Losen the xing screw with the hexagon socket wrench
2. Carefully detach the cable assembly from the process assembly
53
4
2
1
Fig. 21: VEGABAR 87 in IP 68 version, 25 bar and lateral cable outlet, external housing
1 Process assembly 2 Plug connector 3 Cable assembly 4 Connection cable 5 External housing
3. Loosen the plug connector
4. Mount the new process assembly on the measuring point
5. Plug the connector back in
6. Mount the cable assembly on the process assembly and turn it to the desired position
7. Tighten the xing screw with the hexagon socket wrench
The exchange is nished.
If there is no replacement part available on site, one can be ordered
from the agency serving you.
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
39
Page 40
7 Diagnosis, asset management and service
The necessary serial number can be found on the type label of the
instrument or on the delivery note.

7.5 How to proceed in case of repair

You can nd a repair form as well as detailed information on how to proceed under www.vega.com/downloads and "Forms and certi-
cates".
By doing this you help us carry out the repair quickly and without hav­ing to call back for needed information.
If a repair is necessary, please proceed as follows:
Print and ll out one form per instrument
Clean the instrument and pack it damage-proof
Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
Please contact the agency serving you to get the address for
the return shipment. You can nd the agency on our home page
www.vega.com.
40
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
Page 41

8 Dismounting

8 Dismounting

8.1 Dismounting steps

Warning:
Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel or pipeline, high temperatures, cor­rosive or toxic products etc.
Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order.

8.2 Disposal

The instrument consists of materials which can be recycled by spe­cialised recycling companies. We use recyclable materials and have designed the parts to be easily separable.
Correct disposal avoids negative eects on humans and the environ-
ment and ensures recycling of useful raw materials.
Materials: see chapter "Technical data"
If you have no way to dispose of the old instrument properly, please contact us concerning return and disposal.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/EG and the respective national laws. Pass the instrument directly on to a spe­cialised recycling company and do not use the municipal collecting points. These may be used only for privately used products according
to the WEEE directive.
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
41
Page 42

9 Supplement

9 Supplement

9.1 Technical data

Materials and weights
Materials, wetted parts
Process tting 316L Transmitter 316L
Suspension cable FEP Seal, suspension cable FKM, FEP
Connection tube 316L
Diaphragm Hastelloy C276 Protective cap PFA
Seal for process tting (in the scope of delivery)
Ʋ Thread G1½ (DIN 3852-A) Klingersil C-4400 Ʋ Threaded tting Klingersil C-4400
Materials, non-wetted parts
Isolating liquid Essomarcal (med. white oil, FDA-approved) Straining clamp 1.4301 Threaded tting 316L
Electronics housing Plastic PBT (polyester), Alu die-casting powder-coated,
External housing plastic PBT (Polyester), 316L
Socket, wall mounting plate external
housing Seal between base and wall mounting
plate Seal, housing cover NBR (stainless steel housing), silicone (Alu/plastic hous-
Inspection window in housing cover for display and adjustment module
Ground terminal 316Ti/316L Connection cable between transmitter
and external electronics housing with IP 68 (25 bar) version
Type label support on connection cable PE hard Connection cable with IP 68 (1 bar)
version Connection cable, master and slave
sensor
Weights
Basic weight 0.7 kg (1.543 lbs)
Suspension cable 0.1 kg/m (0.07 lbs/ft)
Connection tube 1.5 kg/m (1 lbs/ft)
316L
plastic PBT (Polyester), 316L
EPDM (xed connected)
ing) Polycarbonate (UL-746-C listed)
PE, PUR
PE
PE, PUR
45053-EN-131011
42
VEGABAR 87 • Slave sensor for electronic dierential pressure
Page 43
9 Supplement
Straining clamp 0.2 kg (0.441 lbs) Threaded tting 0.4 kg (0.882 lbs)
Input variable
Adjustment
Adjustment range of the min./max. adjustment relating to the nominal measuring range:
Ʋ Percentage value -10 … 110 % Ʋ Pressure value -20 … 120 %
Adjustment range of the zero/span adjustment relating to the nominal measuring range:
Ʋ zero -20 … +95 % Ʋ span -120 … +120 %
3)
Ʋ Dierence between zero and span max. 120 % of the nominal range
Recommended max. turn down 20 : 1 (no limitation)
Nominal measuring ranges and overload capability in bar/kPa
The specications are only an overview and refer to the measuring cell. Limitations due to the material and version of the process tting as well as the selected pressure type are possible. The specications on the nameplate apply.
Nominal range Overload capacity, max.
pressure
Gauge pressure
0 … +0.1 bar/0 … +10 kPa +15 bar/+1500 kPa -0.2 bar/-20 kPa
0 … +0.4 bar/0 … +40 kPa +30 bar/+3000 kPa -0.8 bar/-80 kPa
0 … +1 bar/0 … +100 kPa +35 bar/+3500 kPa -1 bar/-100 kPa
0 … +2.5 bar/0 … +250 kPa +50 bar/+5000 kPa -1 bar/-100 kPa
0 … +10 bar/0 … +1000 kPa +90 bar/+9000 kPa -1 bar/-100 kPa
0 … +25 bar/0 … +2500 kPa +130 bar/+13000 kPa -1 bar/-100 kPa
Absolute pressure
0 … 1 bar/0 … 100 kPa 35 bar/3500 kPa 0 bar abs.
0 … 2.5 bar/0 … 250 kPa 50 bar/5000 kPa 0 bar abs.
0 … 10 bar/0 … 1000 kPa 90 bar/9000 kPa 0 bar abs.
0 … 25 bar/0 … 2500 kPa 130 bar/13000 kPa 0 bar abs.
Overload capacity, min. pres­sure
Nominal measuring ranges and overload capacity in psi
The specications are only an overview and refer to the measuring cell. Limitations due to the material and version of the process tting as well as the selected pressure type are possible. The specications on the nameplate apply.
Nominal range Overload capacity, max.
Gauge pressure
0 … +1.5 psig +225 psig -2.901 psig
3)
Values less than -1 bar cannot be set.
45053-EN-131011
pressure
Overload capacity, min. pres­sure
VEGABAR 87 • Slave sensor for electronic dierential pressure
43
Page 44
9 Supplement
Nominal range Overload capacity, max.
pressure
0 … +5 psig +435 psig -11.60 psig
0 … +15 psig +510 psig -14.51 psig
0 … +30 psig +725 psig -14.51 psig
0 … +150 psig +1300 psig -14.51 psig
0 … +300 psig +1900 psig -14.51 psig
Absolute pressure
0 … 15 psi 510 psi 0 psi
0 … 30 psi 725 psi 0 psi
0 … 150 psi 1300 psi 0 psi
0 … 300 psi 1900 psi 0 psi
Overload capacity, min. pres­sure
Reference conditions and actuating variables (according to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1
Ʋ Temperature +15 … +25 °C (+59 … +77 °F) Ʋ Relative humidity 45 … 75 %
Ʋ Air pressure 860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig) Determination of characteristics Limit point adjustment according to IEC 61298-2 Characterstic curve Linear
Reference installation position upright, diaphragm points downward
Inuence of the installation position < 0.2 mbar/20 Pa (0.003 psig)
Deviation (according to IEC 60770)
Specications refer to the set span. Turn down (TD) is the ratio: nominal measuring range/set span.
Accuracy class Non-linearity, hysteresis and re-
peatability with TD 1 : 1 up to 5 : 1
0.1 % < 0.1 % < 0.02 % x TD
Non-linearity, hysteresis and re­peatability with 5 : 1
Inuenceoftheproductorambienttemperature
Thermal change zero signal and output span
Turn down (TD) is the relation nominal measuring range/adjusted span.
Ceramic/Metal measuring cell - Standard
Medium or ambient temperature T
-40 … 0 °C (-40 … +32 °F) < {0.15 % + 0.015 %/K x (-T)} x {0.5+0.5 x TD}
0 … 40 °C (+32 … +104 °F) < {0.0075 %/K x (T - 20 °C)} x {0.5+0.5 x TD}
40 … 100 °C (+104 … +212 °F) < {0.15 % } x {0.5+0.5 x TD}
100 … 120 °C (+212 … +248 °F) < {0.15 % + 0.005 %/K x (T - 100 °C)} x {0.5+0.5 x TD}
Ceramic/Metal measuring cell - climate-compensated
Compensated temperature range 0 … +100 °C (+32 … +212 °F)
44
VEGABAR 87 • Slave sensor for electronic dierential pressure
45053-EN-131011
Page 45
9 Supplement
Non-compensated temperature range -40 … 0 °C (-40 … +32 °F), -100 … 130 °C
(212 … +276 °F)
Nominal measuring range in bar/ kPa
0 … 10 bar/0 … 1000 kPa 0 … 150 psig
0 … 25 bar/0 … 2500 kPa 0 … 350 psig
0 … 1 bar/0 … 100 kPa 0 … 15 psig
0 … 2.5 bar/0 … 250 kPa 0 … 35 psig
0 … 0.4 bar/0 … 40 kPa 0 … 6 psig < {0.15 % + 0.4 x
Nominal measuring range in psig
In the compensated temperature range
< {0.05 % + 0.1 x
(0.5+0.5 x TD)}
< {0.1 % + 0.2 x
(0.5+0.5 x TD)}
(0.5+0.5 x TD)}
In the non-compen­sated temperature range
< {0.1 % + 0.15 x
(0.5+0.5 x TD)}
< {0.1 % + 0.3 x
(0.5+0.5 x TD)}
< {0.15 % + 0.6 x
(0.5+0.5 x TD)}
Long-term stability (according to DIN 16086 and IEC 60770-1)
Specications refer to the set span. Turn down (TD) is the ratio: nominal measuring range/set span.
Long-term drift of the zero signal
Time period All measuring ranges Measuring range
One year < 0.05 % x TD < 0.1 % x TD
Five years < 0.1 % x TD < 0.2 % x TD
Ten years < 0.2 % x TD < 0.4 % x TD
0 … +0.025 bar/0 … +2.5 kPa
Long-term drift of the zero signal - Version climate-compensated
Nominal measuring range in bar/kPa Nominal measuring range
in psig
0 … 10 bar/0 … 1000 kPa 0 … 150 psig
0 … 25 bar/0 … 2500 kPa 0 … 350 psig
0 … 1 bar/0 … 100 kPa 0 … 15 psig
0 … 2.5 bar/0 … 250 kPa 0 … 35 psig
0 … 0.4 bar/0 … 40 kPa 0 … 6 psig < (1 % x TD)/year
Long-term drift of the ze­ro signal
< (0.1 % x TD)/year
< (0.25 % x TD)/year
Process conditions
Product temperature
Ʋ Suspension cable -12 … +100 °C (+10 … +212 °F)
Ʋ Connection tube -12 … +100 °C (+10 … +212 °F)
Mechanical stress, depending on the instrument version
Vibration resistance
Ʋ Suspension cable 4 g at 5 … 200 Hz according to EN 60068-2-6 (vibration
with resonance)
Ʋ Connection tube 1 g (with lengths > 0.5 m (1.64 ft), the tube must be sup-
ported in addition)
Shock resistance 100 g, 6 ms according to EN 60068-2-27 (mechanical
shock)
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
45
Page 46
9 Supplement
Ambient conditions
Version Ambient temperature Storage and transport temper-
Version with connection tube -40 … +80 °C (-40 … +176 °F) -60 … +80 °C (-76 … +176 °F)
Version with suspension cable FEP -20 … +80 °C (-4 … +176 °F) -20 … +80 °C (-4 … +176 °F)
Version IP 68 (1 bar) with connec­tion cable PE
-20 … +60 °C (-4 … +140 °F) -20 … +60 °C (-4 … +140 °F)
ature
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Options of the cable entry
Ʋ Cable diameter (options) 5 … 9 mm (0.20 … 0.35 in)
6 … 12 mm (0.24 … 0.47 in)
10 … 14 mm (0.39 … 0.55 in) Ʋ Cable entry ½ NPT Ʋ Blind plug M20 x 1.5; ½ NPT Ʋ Closing cap M20 x 1.5; ½ NPT
Plug options
Ʋ Signal circuit Plug M12 x 1, according to ISO 4400, Harting HAN,
7/8" FF Ʋ Indication circuit M12 x 1 plug
Wire cross-section (spring-loaded terminals)
Ʋ Massive wire, cord 0.2 … 2.5 mm² (AWG 24 … 14) Ʋ Stranded wire with end sleeve 0.2 … 1.5 mm² (AWG 24 … 16)
Electromechanical data - version IP 68 (25 bar)
Connection cable between IP 68 instrument and external housing
Ʋ Conguration four wires, one suspension wire, one breather capillary,
screen braiding, metal foil, mantle Ʋ Wire cross-section 0.5 mm² (AWG 20) Ʋ Wire resistance < 0.036 Ω/m (0.011 Ω/ft) Ʋ Standard length 5 m (16.40 ft) Ʋ Max. length 180 m (590.5 ft) Ʋ Min. bending radius at 25 °C/77 °F 25 mm (0.985 in) Ʋ Diameter approx. 8 mm (0.315 in) Ʋ Colour Blue
Cable entry/plug
4)
Ʋ External housing – 1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm), 1 x
blind plug M20 x 1.5
or:
1␣x plug (depending on the version), 1␣x blind stopper
M20␣x␣1.5
4)
Depending on the version M12 x 1, according to ISO 4400, Harting, 7/8" FF.
46
VEGABAR 87 • Slave sensor for electronic dierential pressure
45053-EN-131011
Page 47
9 Supplement
Spring-loaded terminals for wire cross-
2.5 mm² (AWG 14)
section up to
Voltage supply, interface to the master sensor
Voltage supply Through the master sensor Data transmission digital (I²C-Bus) Conguration, connection cable 4-wire, screened Cable length max. 25 m
Potential connections in the instrument
Electronics Not non-oating Ground terminal Galvanically connected with the process tting
Electrical protective measures
Housing material Version IP-protection class NEMA protection
Plastic Single chamber IP 66/IP 67 NEMA 4X/6P
Aluminium Single chamber IP 66/IP 68 (0.2 bar)
IP 68 (1 bar)
Stainless steel, electro-
polished
Stainless steel, investment casting
Stainless steel Transmitter for external
Single chamber Single chamber
Single chamber IP 66/IP 68 (0.2 bar)
housing
IP 66/IP 68 (0.2 bar)
IP 69K
IP 68 (1 bar)
IP 68 (25 bar) -
NEMA 4X/6P
-
NEMA 4X/6P
-
NEMA 4X/6P
NEMA 4X
Overvoltage category III Protection class II
Approvals
Instruments with approvals can have dierent technical data depending on the version.
For that reason the associated approval documents of these instruments must be carefully noted. They are part of the delivery or can be downloaded under www.vega.com and "VEGA Tools" as
well as under "Downloads" and "Approvals".

9.2 Dimensions

The following dimensional drawings represent only an extract of the possible versions. Detailed
dimensional drawings can be downloaded at www.vega.com under "Downloads" and "Drawings".
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
47
Page 48
9 Supplement
Housing
~ 69 mm
M20x1,5/ ½ NPT
~ 69 mm
(2.72")
(2.72")
ø 79 mm
(3.11")
1
ø 79 mm
(3.11")
(4.41")
112 mm
M20x1,5/ ½ NPT
~ 59 mm
~ 116 mm
(4.57")
(2.3")
ø 86 mm
(3.39")
2
ø 80 mm
(3.15")
M20x1,5
(4.57")
116 mm
M20x1,5/ ½ NPT
~ 59 mm
(2.32")
ø 80 mm
(3.15")
3
(4.41")
112 mm
M20x1,5/ ½ NPT
(4.61")
117 mm
M20x1,5/
4
½ NPT
5
(4.1")
104 mm
Fig. 22: Housing versions in protection IP 66/IP 67 and IP 66/IP 68 (0.2 bar)
1 Plastic housing 2 Aluminium housing 3 Stainless steel housing, electropolished 4 Stainless steel housing, precision casting 5 Stainless steel housing, electropolished IP 69K
45053-EN-131011
48
VEGABAR 87 • Slave sensor for electronic dierential pressure
Page 49
External housing with IP 68 version
82 mm (3.23")
41,6 mm
(3.15")
80 mm
(1.64")
9 Supplement
1
(4.25")
108 mm
41,6 mm
(1.64")
2
Fig. 23: VEGABAR 87, IP 68 version with external housing
1 Lateral cable outlet 2 Axial cable outlet 3 Plastic version 4 Stainless steel version 5 Seal 2 mm (0.079 in) - only with 3A approval
3
~ 66 mm
(2.60")
4
110 mm x 90 mm
110 mm x 90 mm
(4.33" x 3.54")
(4.33" x 3.54")
(2.32")
59 mm
(2.01")
51 mm
5
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
49
Page 50
9 Supplement
VEGABAR 87
48-52 mm
(1.89" -
2.05")
175 mm
L
197 mm
52 mm
(6.89")
ø 8 mm
(0.32")
(7.76")
SW 30 mm
(1.18")
22 mm (0.87")
(1.52")
38,5 mm
(2.05")
L
G1 ½ / 1½ NPT
(0.55")
14 mm
(0.79")
20 mm
L
SW 46 mm
(1.81")
G1 ½ / 1½ NPT
G1 ½ / 1½ NPT
(0.87")
22 mm
L
ø 29 mm
(1.14")
ø 40 mm
(1.58")
1 2 3 4
Fig. 24: VEGABAR 87, standard ttings
1 Straining clamp 2 Threaded tting 3 Thread G1½ 4 Thread 1½ NPT 5 Lock tting
50
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
Page 51
VEGABAR87,angeconnection
d2
9 Supplement
(2.01")
51 mm
D
ø 8 mm
L
(6.97")
177 mm
(0.32")
ø 40 mm
(1.58")
Fig. 25: VEGABAR 87, ange connection
1 Flanges according to DIN 2501 2 Flanges according to ANSI B16.5
f
b
d4
k
DN PN Dk
mm
1
. .2540 115 18 85
B3 32 40 140 18 100 4xø18782
A8 40 40 150 18 110
B2 50 40 16520 125
B5 80 40 20024 160 8xø18 1383
CD 100 16 220 20 180 8xø18 158 3
A6 150 16 285 22 240 8xø22 162 3
2
"lbs Db kd2d4f
CA
2" 150
CB
3" 150
BB 4" 300
DN PN Dkb
inch
1
. . 25 40
B3 32 40
A8 40 40
B2 50 40
B5 80 40
A6 150 16
2
"lbs Db kd2d4f
CA
2" 150
CB
3" 150
BB 4" 300
152,5
190.5
254
5.51"
5.91"
6.50"
7.87"
11.22"
6"
7.50"
10"
19
19
30.5
0.71"
0.71"
0.71"
0.79"
0.95"
0.79"
0.87"
0.75
0.75"
1.25"
b
"
d2
4xø14
4xø18
4xø18
4xø16120,5
4xø16152,5 152
200 8xø16
d2 d4 f
4xø 0.55"
3.35"4.53"
4xø 0.71"
3.94"
4.33"
4xø 0.71"
4xø 0.71"
4.92"
8xø 0.71"
6.30"
8xø 0.87"
9.45"
4xø0.63"4.75"
4xø0.63"6" 6"
7.87" 8xø0.63"
d4
68
88
102
92
157
2.68"
3.07"
3.47"
4.02"
5.43"
6.22
6.38"
3.63"
6.19"
"8xø 0.71"7.09"8.66"CD 10016
f
2
3
3
3
3
3
0.08"
0.08"
0.12"
0.12"
0.12"
0.12"
0.12"
0.12"
0.12"
0.12"
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
51
Page 52
9 Supplement
VEGABAR87,hygienictting
(2.01")
51 mm
ø 64 mm
(2.52")
(2.01")
51 mm
ø 78 mm (3.07") ø 92 mm (3.62")
L
(6.97")
177 mm
ø 8 mm
(0.32")
ø 40 mm
(1.58")
1 2
Fig. 26: VEGABAR 87, hygienic ttings
1 Clamp 2" PN16 (ø64 mm) DIN 32676, ISO 2852/316L 2 Bolting DN 50
L
45053-EN-131011
52
VEGABAR 87 • Slave sensor for electronic dierential pressure
Page 53
9 Supplement

9.3 Industrial property rights

VEGA product lines are global protected by industrial property rights. Further information see www.vega.com.
Only in U.S.A.: Further information see patent label at the sensor housing.
VEGA Produktfamilien sind weltweit geschützt durch gewerbliche Schutzrechte.
Nähere Informationen unter www.vega.com.
Les lignes de produits VEGA sont globalement protégées par des droits de propriété intellec­tuelle. Pour plus d'informations, on pourra se référer au site www.vega.com.
VEGA lineas de productos están protegidas por los derechos en el campo de la propiedad indus­trial. Para mayor información revise la pagina web www.vega.com.
Линии продукции фирмы ВЕГА защищаются по всему миру правами на интеллектуальную собственность. Дальнейшую информацию смотрите на сайте www.vega.com.
VEGA系列产品在全球享有知识产权保护。 进一步信息请参见网站<www.vega.com>。

9.4 Trademark

All the brands as well as trade and company names used are property of their lawful proprietor/
originator.
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
53
Page 54
INDEX
INDEX
A
Adjustment
– Density 33 – Dierential pressure, ow 31, 32 – Interface 34 – Level 31 – Menu 28 – Overview 30 – Unit 29
C
Connecting
– Mechanically 13
Connection
Steps 20 – Technology 20
D
Default values 35
E
Eective pressure transmitter
– Characteristics values 37
F
Fault rectication 38 Functional principle 9, 10
M
Maintenance 38
Measurement setup
– Density measurement 17 – Interface measurement 16 – Level dierence measurement 16
Quick setup 25
P
Position correction 29 Pressure compensation 14, 15
R
Repair 40 Reset
– Basic settings 35 – Delivery status 35
S
Service
– Hotline 38
Setup
– Extended adjustment 28
54
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
Page 55
Notes
45053-EN-131011
VEGABAR 87 • Slave sensor for electronic dierential pressure
55
Page 56
Printing date:
All statements concerning scope of delivery, application, practical use and operat­ing conditions of the sensors and processing systems correspond to the information available at the time of printing. Subject to change without prior notice
© VEGA Grieshaber KG, Schiltach/Germany 2013
VEGA Grieshaber KG Am Hohenstein 113 77761 Schiltach Germany
Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail: info.de@vega.com www.vega.com
45053-EN-131011
Loading...