VEGA BAR21 User Manual

Operating Instructions
VEGABAR 21
87654321
+ -
-
DISPLAY
4 ¼ 20 mA
+
VEGADIS 10
Level and Pressure
+ ­4 20 mA
12 36 V DC E12
­+
DISPLAY
VEGADIS 10
OK
87654321
p

Safety information

Safety information
The described module must only be installed and operated as described in these operating instructions. Please note that other action can cause damage for which VEGA does not take responsibility.
2 VEGABAR 21

Note Ex-area

Please note the approval documents attached (yellow binder), and especially the included safety data sheet.
Contents
Contents
Safety information ........................................................................ 2
Note Ex-area ................................................................................ 2
1 Product description
1.1 Function and configuration .................................................. 4
1.2 Self-monitoring ..................................................................... 4
1.3 Technical data ....................................................................... 5
1.4 Approvals and certificates ................................................ 10
1.5 Dimensions ......................................................................... 10
2 Mounting................................................................................... 13
3 Electrical connection
3.1 Connection information ...................................................... 13
3.2 Connection plan ................................................................. 14
3.3 Connection examples ........................................................ 15
4 Set-up
4.1 Adjustment of the basic functions .................................... 16
4.2 Menu-guided adjustment with additional functions ......... 17
4.5 Indicating module ............................................................... 22
5 Diagnosis
5.1 Maintenance ....................................................................... 23
5.2 Failure removal ................................................................... 23
6 Instrument modification
6.1 Interchanging adjustment modules .................................. 25
7 Measurement with isolating systems ................................. 26
VEGABAR 21 3

1 Product description

Product description

1.1 Function and configuration

VEGABAR 21 is an efficient pressure trans­mitter with isolating system for process pres­sure measurement and hydrostatic level measurement. The hydrostatic pressure of the medium acts on the separating dia­phragm of the isolating system and is trans­ferred via the isolating liquid to a ceramic­capacitive measuring cell CERTEC®. The pressure on this measuring cell effects a capacitance change within the measuring cell. This capacitance change is detected by an ASIC (Application specific integrated circuit) and converted by an integral oscilla­tor with microcomputer into a pressure pro­portional signal. The exact, digital processing of measured data with highest resolution ensures perfect technical data.
Isolating systems are used to protect the measuring cell against aggressive products and high temperatures. Due to the suitable selection of isolating liquid and diaphragm material, systems with a temperature resist­ance up to 400°C are possible.
The oscillator is powered by a separate VEGA-signal conditioning instrument, a stabi­lised power supply unit or a DCS (active input). After the adjustment, a standardised current signal 4 … 20 mA is available which can be displayed (e.g. in DCS-systems) or further processed.

1.2 Self-monitoring

Important electronics components are checked on their function and internal param­eters such as sensor value, temperature and operating voltage are monitored to increase the reliability.
VEGABAR 21 with ceramic CERTEC®-meas­uring cell offers the advantage of a continu­ous self-monitoring. Measuring and reference capacitance of the measuring cell are in a defined relation over the complete measuring range. Each deviation from these data is a reliable indicator for a malfunction of the measuring cell.
If during these routines failures or malfunc­tions are detected, a fault signal is triggered via the 4 … 20 mA-output (current jump to 3,6 mA or 22 mA).
Two versions are available for adjustment:
- adjustment module directly on VEGABAR
- adjustment module in external housing (VEGADIS 10)
4 VEGABAR 21
Product description

1.3 Technical data

Mechanical data
Materials, wetted parts
Process connection stainless steel 1.4571 Diaphragm stainless steel 1.4571, Hastelloy C276 or B2,
Extension stainless steel 1.4571
Materials, non-wetted parts
Housing high resistance plastic PBT (Polyester) Earth terminal stainless steel 1.4305 Window of the indicating module Lexan
Weight
VEGABAR approx. 0,8 … 8 kg
Adjustment and indicating elements
Adjustment of the basic functions 2 keys, 1 rotating switch Menu-guided adjustment with additional function
- adjustment elements 4 keys
- indicating elements DOT-Matrix display, 3 lines with 7 figures each Indicating module LC-display with
Monel 400, Tantalum, PTFE-foil on 1.4571
(dependent on the isolating system)
- bargraph (20 segments)
- digital value (4-digit)
- tendency indicators for raising or falling values
Electrical data
Connection
Cable entry M20 x 1,5 (for cable-ø 5 … 9 mm) Screw terminals for cross-section area of conductor up to
Protective measures
Protection
1)
Protection class III Overvoltage category III
1)
For maintaining the housing protection, the use of the suitable seal for the cable in the Pg is necessary. If the used seal does not fit with the cable, it should be replaced by a suitable one.
VEGABAR 21 5
2,5 mm
IP 65
2
Product description
Supply and signal circuit
Supply voltage 12 … 36 V DC Permissible residual ripple USS £ 1 V Output signal
- range 3,8 … 20,5 mA
- resolution 6 µA Current limitation 22 mA Fault signal 22 mA (3,6 mA) Integration time 0 … 10 s adjustable Average delay time 70 ms Connection line 2-wire Max. permissible load dependent on the supply voltage
(see load diagram)
1000
750
in Ohm
total
500 250
Load RL
0
12
18 24 30 36
Voltage of the external energy UH in Volts
Indicating and adjustment circuit
For connection to VEGADIS 10 and/or the indicating module Data transmission digitally Connection line 4-wire Max. line length 25 m
Adjustment ranges
Zero -20 % … +95 % adjustable of nominal range Span 3,3 % … 120 % adjustable of nominal range Recommended turn down:
- accuracy class 0,25 up to 1 : 5
- accuracy class 0,1 up to 1 : 10
6 VEGABAR 21
Product description
Measuring ranges
Nominal measuring Gauge pressure Low pressure range resistance
1)
resistance
Relative pressure 0 … 0,4 bar 20 bar -0,4 bar
0 … 1,0 bar 25 bar -1,0 bar
0 … 2,5 bar 35 bar -1,0 bar
0 … 5,0 bar 45 bar -1,0 bar
0 … 10,0 bar 60 bar -1,0 bar
0 … 20,0 bar 90 bar -1,0 bar
0 … 40,0 bar 140 bar -1,0 bar
0 … 60,0 bar 200 bar -1,0 bar
-0,05 … 0,05 bar 10 bar -0,1 bar
-0,1 … +0,1 bar 15 bar -0,2 bar
-0,2 … +0,2 bar 20 bar -0,4 bar
-0,5 … +0,5 bar 25 bar -1,0 bar
-1,0 … +0,0 bar 25 bar -1,0 bar
-1,0 … +1,5 bar 35 bar -1,0 bar
-1,0 … +4,0 bar 45 bar -1,0 bar
-1,0 … +10,0 bar 60 bar -1,0 bar
-1,0 … +20,0 bar 90 bar -1,0 bar
-1,0 … +40,0 bar 140 bar -1,0 bar
-1,0 … +60,0 bar 200 bar -1,0 bar
Absolute pressure 0 … 1,0 bar 25 bar
0 … 2,5 bar 35 bar
0 … 5,0 bar 45 bar
0 … 10,0 bar 60 bar
0 … 20,0 bar 90 bar
0 … 40,0 bar 140 bar
0 … 60,0 bar 200 bar
1)
Note the nominal pressure of the process connection!
VEGABAR 21 7
Accuracy (similar to DIN 16 086, DIN V 19 259 - 1 and IEC 770)
Influence of the ambient temperature
Average temperature coefficient of the zero signal
- accuracy class 0,25 < 0,15 %/10 K
- accuracy class 0,1 < 0,05 %/10 K
1) 2)
Product description
Long-term stability
Long-term drift of the zero signal
1) 3)
< 0,1 % (meas. cell, without considering the isolating system)
Deviation
Reference conditions (acc. to IEC 770)
- temperature 15°C … 35°C
- rel. humidity 45 % … 75 %
- air pressure 860 mbar … 1060 mbar Determination of characteristics limit point adjustment acc. to DIN 16 086 Characteristics linear Deviation in characteristics
1)
- accuracy class 0,25 < 0,25 %
- accuracy class 0,1 < 0,1 % Hysteresis Repeatability
1)
1)
< 0,02 % < 0,02 %
Other actuating variables
Calibration position standing, diaphragm points downwards Vibration resistance mechanical vibrations with 4 g and
5 … 100 Hz, tested acc. to the regulations of German Lloyd GL-characteristics 2
Operating conditions
Ambient conditions
Ambient temperature -40°C … +85°C
- indication module -20°C … +70°C Storage and transport temperature -40°C … +85°C Medium temperature -40°C … +100°C Medium temperature, abhängig von Druckmittlerflüssigkeit Temperature
- silicone oil -40°C … +130°C / -40°C … +130°C
- silicone oil and cooling element -40°C … +200°C / -40°C … +130°C
- silicone oil and capillaries 1 m -40°C … +200°C / -40°C … +130°C
- vegetable oil -10°C … +130°C / -10°C … +130°C
- vegetalble oil and cooling element -10°C … +250°C / -10°C … +130°C
- high temperature oil a. cooling element -10°C … +400°C / -10°C … +200°C
- high temperature oil and capillaries 1 m -10°C … +400°C / -10°C … +200°C
1)
Relating to the nominal range.
2)
In the compensated temperature range of 0°C … +80°C, reference temperature 20°C. Without isolating diaphragm, capillaries, cooling element (see influence of the ambient temperature).
3)
Acc. to IEC 770, item 6.1.2 relating to the nominal range.
4)
At pabs < 1 bar (vacuum range ) the isolating liquid boils already at low temperatures.
8 VEGABAR 21
p
> 1 bar / p
abs
4) < 1 bar
abs
Product description
Zusätzlicher Temperatureinfluß durch Druckmittler
Process connection Temperature coefficient
[mbar/10 K]
FA Flange DN 25 PN 40 acc. to DIN 2501,
seal surface acc. to DIN 2526 Form D 1,8
FB Flange DN 40 PN 40 acc. to DIN 2501,
seal surface acc. to DIN 2526 Form D 1,0
FC Flange DN 50 PN 40 acc. to DIN 2501,
seal surface acc. to DIN 2526 Form D 2,0
FE Flange DN 50 PN 40 with ext. 100 mm 1,9
FG Flange DN 50 PN 40 with ext. 200 mm 2,1
FH Flange DN 80 PN 40 acc. to DIN 2501,
seal surface acc. to DIN 2526 Form D 0,4
FK Flange DN 80 PN 40 with ext. 100 mm 0,5
FN Flange DN 1" 300 lbs acc. to ANSI B 16.5,
seal surface RF 1,8
FO Flange DN 2" 600 lbs acc. to ANSI B 16.5,
seal surface RF 2,0
- by cooling elements, dependent on diaphragm-ø 0,1 … 1,5 mbar/10 K
- by 1 m capillary line dependent on diaphragm-ø 0,1 … 15 mbar/10 K
VEGABAR 21 9
Product description

1.4 Approvals and certificates

Approvals
Ex Zone 2 StEx Zone 10 Ship approval CENELEC EEx ia IIC ATEX II 1G EEx ia IIC and WHG ATEX II 1G EEX ia IIC and ship approval
If for certain applications the use of approved instruments is required, the appropriate legal documents (test reports, test certificates and conformity certificates) have to be noted. These are supplied with the appropriate instrument.
CE-conformity
VEGABAR 21 corresponds to the requirements of EMC (89/336/EWG) and NSR (73/23/EWG). The conformity has been judged acc. to the following standards: EMC Emission EN 50 081 - 1: 1992
Susceptibility EN 50 082 - 2: 1995
NSR EN 61 010 - 1: 1993
NAMUR-regulations
The NAMUR-regulations NE 21, May 1993 are met.

1.5 Dimensions

Housings
without indication module with indication module
85
~76
72
M20x1,5
Earth connection
10 VEGABAR 21
82
Breather facility
Product description
Process connections
Capillaries 1 m (TK = 15)
Cooling element (TK = 4,5)
Standard
100
d2
51
FA, FB, FC FH FN, FO
d4
k
D
D = outer flange diameter b = flange thickness k = diameter of hole circle d2= diameter of holes d4= seal ledge diameter d5= extension diameter f = seal ledge strength RL= extension length dM= diaphragm diameter
Standard
83
51
f
b
L
R
d5
FE, FG, FK
Flange connection acc. to DIN 2501, seal ledge acc. to DIN 2526 form D
Order Flange Holes Seal ledge Extension code Size D b k No. d2d4fRLd
d
5
M
FA DN 25 / PN 40 115 22 85 4 14 68 2 –– –– –– FB DN 40 / PN 40 150 18 110 4 18 88 3 –– –– –– FC DN 50 / PN 40 165 20 125 4 18 102 3 –– –– –– FE 100 48,3 47 FG 200 48,3 47 FH DN 80 / PN 40 200 24 160 8 18 138 3 –– –– –– FK 100 76 72
Flange connection acc. to ANSI B 16.5, seal ledge RF
FN DN 1" / 300 lbs 125 22 89 4 20 51 2 FO DN 2" / 600 lbs 165 32 127 8 20 92 6,4
VEGABAR 21 11
Product description
Tube isolating diaphragms for mounting between flanges with connection acc. to DIN 2501
Order DN PN D Mb code
RC 50 6 400 95 54,5
Tube isolating diaphragms with connection acc. to DIN 11 851
Order DN PN G1 L2 D2 d6 g m code
RF 50 25 Rd78x1/6 156 70 68,5 11 18
Tube isolating diaphragms with Clamp­connection
Order DN PN L D d D1 D2 Mb code
RH 2" 40 156 64 56,3 56 75 48
60
D
Mb
g
d6
d
D
Mb
L2
D2
L
D2
D1
12 VEGABAR 21
Mounting, electrical connection

2 Mounting

VEGABAR can be mounted in any position. The cable entries have to point downwards to avoid humidity ingress. For this purpose the housing can be rotated by 330° against the mounting part.
For mounting, a seal must be used corre­sponding to the connection. This seal is either supplied with VEGABAR or must be provided by the customer.
Compensation of the atmospheric pressure
The atmospheric pressure of instruments used for overpressure measurements is compensated via a breather facility inte­grated in the housing.

3 Electrical connection

3.1 Connection information

The electronics in VEGABAR 21 requires a power supply of 12 … 36 V DC. It is pro- vided in two-wire technology, i.e. supply voltage and current signal are led to the terminals via the same two-wire connection cable.
This external energy is provided via a sepa­rate power supply unit:
- power supply unit, e.g. VEGASTAB 690
- processing unit with integral DC-current source (e.g. active DCS-input)
- VEGAMET or VEGADIS 371
Observe that the external energy source is reliably separated from the mains circuits acc. to DIN VDE 0106, part 101. The above mentioned VEGA-instruments meet this re­quirement and protection class III is therefore ensured.
The external energy source must supply a terminal voltage of at least 12 V to the trans­mitter. The actual terminal voltage on the transmitter depends on the following factors:
- output voltage of the external energy source under nominal load.
- Electrical resistors of the connected instru­ments in the circuit (see connected instru­ments, load resistor).
Note the following instructions for electrical connection:
- The connection must be made acc. to the specific installation standards (e.g. in Ger­many according to the VDE-regulations).
- The terminal voltage must not exceed 36 V to avoid damage of the electronics.
- The electrical connection is provided with a reverse polarity protection.
- The wiring between VEGABAR and power supply can be made with standard two­wire cable.
VEGABAR 21 13
Electrical connection
- If strong electromagnetic interferences have to be expected, screened cable is recommended. The screening should be earthed at one side of the sensor.
- If overvoltages are expected, we recom­mend the installation of VEGA-overvoltage arresters.
- In the PG a seal must be used which fits to the cable.

3.2 Connection plan

Note:
A current meter for local control of the output current can be connected to terminals 1 and 3. This measurement can be made during operation without interrupting the supply line.
VEGABAR
VEGABAR (with or without indication)
Current meter for local control
1)
Earth
1
876532
rd wt vio or
+
Supply voltage 12 36 V DC 4 20 mA
1)
If screened cable is used, the screen must be connected in the terminal box on one side to earth and the earth terminal must be earthed outside on the housing according to the prescription. Both terminals are connected in the housing.
14 VEGABAR 21
Electrical connection

3.3 Connection examples

Supply via power supply unit
The processing is made by an indicating instrument.
Current meter for local control
+ – –
1 2 3
Terminals VEGABAR
Supply via a DCS with active input circuit
The processing is made via a DCS with active input circuit.
Current meter for local control
analogue / digital indicat­ing instrument
Power supply unit
~
+
+ – –
1 2 3
Terminals VEGABAR
DCS
+
DCS
VEGABAR 21 15

4 Set-up

Set-up
The electrical connection as well as the set­up can be facilitated by mounting the cover or indicating module short-term laterally or displaced to the bottom on VEGABAR.
VEGABAR
+ ­4 ¼ 20 mA
12 ¼ 36 V DC E12
OK
87654321
-
DISPLAY
+
VEGADIS 10
Tighten screws moderately
Indicating module
4.1 Adjustment of the basic
functions
Adjustment elements
Rotating switch: Choose the requested function
Key +: Change
OPERATE t
i
ZERO SPAN
value (raising)
Adjustment
For adjustment of zero and span, a current meter must be connected to terminals 1 and
3. The measured value is identical with the output current.
1 Adjustment of zero
(e.g. process pressure zero or empty vessel)
Set rotating switch to zero
By pushing the + and “–“-keys simultane-
ously, the current jumps directly to 4 mA or adjusting by pushing the “+“ and “–“-keys a current of 4 mA
Adjustment range of zero:
-20 % +95 % of the nominal measuring range
2 Adjustment of span
(e.g. process pressure or vessel level maximum)
Set rotating switch to span
By pushing the + and “–“-keys simultane-
ously, the current jumps directly to 20 mA or adjusting by pushing the “+“ and “–“- keys a current of 20 mA
Adjustment range of span: 3,3 % 120 % of the nominal measuring range
Key –: Change value (falling)
Note:
- A modification of zero does not influence the span.
Adjustment system
Select the requested function with the rotat­ing switch.
Change the value with keys +- and “–“.
Reset rotating switch to OPERATE, the
- It is also possible to adjust currents for part fillings or part pressures, e.g. 8 mA for 25 % and 16 mA for 75 %. VEGABAR cal­culates automatically the current values for 0 % or 100 % (only possible with ³ 3,3 %).
adjusted values are transferred to the EEPROM-memory. They remain there even in case of voltage loss.
16 VEGABAR 21
Set-up
Integration time
An integration time ti of 0 10 s can be ad­justed for damping of pressure shocks.
Procedure
Set rotating switch to t
By pushing the “-“-key 10-times, ensure
i
first of all that the integration time is set to 0 s
For every 1 s requested integration time, push the “+“-key once
The integration time is the time required by the current output signal to reach 90 % of the actual level after a step change of the proc­ess pressure.
4.2 Menu-guided adjustment with
additional functions
The adjustment with the multiple function module is menu orientated and is made via the four keys in conjunction with the clear text indication. The jump from the measured value indication to the main menu is made with the OK-key. Use the ®-key to change from one menu point to the other within the menu.
Branching
A branching is indicated with the symbol and enables a jump to the below menu with the OK-key. In this menu you can find pa­rameters which are linked (possibly in other submenus).
Parameters have no or symbol. The value of the parameters can be modified with the “+“ and “–“-keys or chosen from the list. When these buttons are pushed once, the value flashes first. When pushing again, the value is modified. The modified value can be saved with the OK-key. Push the ®-key to interrupt the input (without storing the modifi­cation).
Certain parameters can only be indicated and their values cannot be modified.
Reset
A reset is marked with the symbol and enables with the OK-key a jump to the above menu.
Automatic reset to measured value indication is triggered 60 minutes after having pushed a key for the last time.
Adjustment elements
Indication of:
- measured value
- functions in menu
Jump to the below menu
with OK-key
D Reset to the above menu
with OK-key
VEGABAR 21 17
Dependent on the param­eter, change value (raising) or choose from list
+
OPERATE
3.00
mbar
Dependent on the param­eter, change value (falling) or choose from list
OK
Choose menu point or inter­rupt input
Acc. to the menu point, save the adjusted value or change the menu stage
Menu plan
Set-up
Operate
0,2
mbar
Ad-
just-
ment
Adjust
with
press.
Measured value indication
Calcu-
lation
ti
1
s
Adust
without
press
Unit
0,0
mbar
psi kPa
Output
Unit
0,0
mbar
psi kPa % Scal. mA
Current
output
4-20mA
20-4mA
Scaling
indica-
tion
0%=
0
Fault
signal
22mA
Escape
3,6mA
Escape
The scaling can only be carried out, when in the menu point "Unit the option scal had been chosen before.
100%=
1000
Decimal
point
888,8
Escape
Escape
For pressure transmitters with a measuring range ³ 1 bar the unit bar can be selected instead of mbar.
Offset
correc-
tion
Offset
corr. OK?
Zero
0,0
mbar
Span
100,0
mbar
Only with pressure transmitters for relative pressure measurement
Escape
Simula-
tion
Sim.
xx,x
mbar
Sensor unpres-
surized?
Zero
4,000
mA
Span
4,000
mA
Escape
18 VEGABAR 21
Set-up
Addi-
tional
functiion
Sensor
data
Escape
Lan-
guage
Lan-
guage
English
Reset
Reset OK ?
Escape
Deutsch Franc. Ital. Español
Herst.
Date
49.98
1)
By pushing the “+“ and “–“-keys simultaneously,
P min
1)
-0,3
mbar
Reset
Now!
OK ?
P max
150
mbar
1) 1) 1)
Temp.
30,7
°C
the min. and max. values can be reset to the actual value.
T min
23,5
°C
T max
36,2
°C
Diagno-
stic no.
- - -
Escape
In white letters you see the parameters which can be modified with the “+“ or “–“-key and saved with the OK-key.
The sensor and measured value information which cannot be modified in these positions are in itallic print.
List These options can be chosen with the
+ or “–“-key and saved with theOK-key.
Adjust
with
press.
Escape
Use the arrow key, to move in the menu plan to the right.
In menu points with these symbols you can move with the OK-key to the bottom or top.
Span
100,0
mbar
Lan-
guage
English
Deutsch Franc. Ital. Español
VEGABAR 21 19
Set-up
Adjustment by considering the actual pressure (life-adjustment)
The life-adjustment comprises two steps: 1 Adjustment of zero 2 Adjustment of span
The appropriate output current is displayed via the DOT-matrix.
1 Adjustment of zero
(e.g. process pressure zero or vessel empty)
Set with the “+“ or “–“-key the cur-
Zero
4,000
rent value to 4,000 mA. Then push
mA
the OK-key.
2 Adjustment of span
(e.g. process pressure or vessel level max.)
Set with the “+“ or “–“-key the cur-
Span
20,000
rent value to 20,000 mA. Then push
mA
the OK-key.
Note:
- A modification of zero does not influence the adjusted span.
- For standard adjustment of zero/span “+“ and “–“ are pushed simultaneously. The value jumps directly to 4 mA/20 mA.
- In case of a high turn down generally the + and --key should be pushed simulta­neously due to the resolution.
- When pushing the “+“ or “–“-keys individu­ally, the output current remains at the last value, only after storing with the OK-key it takes the adjusted value.
- It is also possible to adjust currents for part fillings or part pressures, e.g. 8 mA for 25 % and 16 mA for 75 %. Then VEGABAR calculates automatically the current values for 0 % or 100 % (only possible with >3,3 %).
Adjustment without considering the actual pressure (dry adjustment)
The adjustment without pressure comprises four steps: 1 Enquiry of the unit in which the adjustment
should be carried out 2 Offset correction 3 Adjustment of zero 4 Adjustment of span
The offset correction (only with overpressure) defines the reference basis for the measure­ment. It can be carried out:
- before or after the adjustment of zero and
span
- before or after installation of VEGABAR.
VEGABAR must be unpressurised when the offset correction is carried out!
The adjustment without pressure can be carried out assembled or unassembled (e.g. workroom). An actually available pressure has no meaning for the adjustment.
1 Enquiry of the unit
Choose with the “+“ or “–“-key the
Adjust
without
2 Adjustment of zero
mbar
4 Adjustment of span
Span
100,0
mbar
unit. Save the selected unit with the
press.
OK-key.
Unit
0,0
mbar
psi kPa
Adjust zero with the “+“ or “–“-key
Zero
0,0
and save with the OK-key.
Adjust span with the “+“ or “–“-key and save with the OK-key.
20 VEGABAR 21
Set-up
Note:
- A modification of zero does not influence the adjusted span.
- By pushing the “+“ and “–“-keys the output current remains on the last value, after saving with the OK-key, the output cur­rent takes the adjusted value.
Processing
Adjustment of the integration time
ti
For damping of pressure shocks an
1
integration time ti of 0 s … 10 s can
s
With the OK-key the adjusted value can be saved.
Selection of the unit
Scaling
be adjusted with the “+“ or “–“-key.
In the measured value indication
Unit
0,0
the actually measured pressure is
mbar
displayed on the DOT-matrix. The
psi
appropriate unit can be chosen out
kPa
of a list with the “+“ or “–“-key and
%
the OK-key. When choosing the
scal. mA
unit Scal. the following menu points are available.
The bar graph relates to the adjusted meas­uring range and change proportionally to the current output. Digital values < –10 % or > 110 % are displayed as flashing figure.
By scaling user specific figures as 4-digit digital values are coordinated on the indicat­ing module to the values 4 mA or 20 mA of the current output.
Outputs
Current output/Selection of the character­istics
The current output provides the actually measured pressure as analogue current signal 4 20 mA relating to the adjustment.
Current
4-20mA
20-4mA
Fault signal
If during the continuous self-monitoring fail­ures, damages or interferences in the mea­suring cell or the electronics are detected, the fault signal is triggered via the current output.
The characteristics can also be
output
inverted, i.e. adjusted to 20 ... 4 mA with the “+“ or “–“-key and the OK-key.
In operating condition the actually measured pressure is displayed on the indicating mod­ule as 4-digit digital value.
Scaling
indica-
tion
0%=
100%=
0
1000
Decimal
point
888,8
Escape
Simulation
Simula-
With the “+“ or “–“-key and the
Fautl
signal
OK-key you choose if the failure
22mA
current should be 22 mA or
3,6mA
3,6 mA.
To check the outputs of VEGABAR and connected components, you
tion
can adjust with the “+“ or “–“-key an individual pressure or %-value. The
Sim.
adjusted value flashes with acti-
xx,x
vated simulation. With the OK-key
mbar
the simulation can be finished.
VEGABAR 21 21
Set-up
Additional functions
Important sensor data can be indicated via DOT-matrix for information and diagnosis purposes:
- manufacturing date
- pointer function (p
- pointer function (p
- actual temperature value (Temp)
- pointer function (T
- pointer function (T
- diagnosis number
Addi-
tional
function
Sensor
data
Herst.
P min
Date
49.98
1)
The min. and max.-values can be set to the actual value by pushing the “+“ and “–“-keys simultane­ously.
P max
-0,3
mbar
1)
1) 1) 1)
Language
150
mbar
min
max
min
max
Temp.
30,7
)
)
)
)
T max
T min
23,5
°C
°C
36,2
°C
Diagno-
stic no.
- - -
Reset
The reset function deletes the ad-
Reset
justed values and resets the param­eters to the data of the factory set­ting. The adjustment data
Reset
correspond again to the nominal
OK ?
measuring range.
Reset
Now!
OK ?
4.5 Indicating module
Digital value
- 4 digits and decimal point
- with single range and basic function mod­ule fixed indication range in bar
- the multiple function module can be indi­vidually scaled.
The connection is made acc. to the scheme in section Electrical connection“.
Tendency indication
As factory setting VEGABAR is
Lan-
adjusted to the order language.
guage
Bargraph
With the “+“ or “–“-key and the OK-key the languages German, English, Italian or Spanish can be
Lan-
guage
English
Deutsch Franc. Ital. Español
22 VEGABAR 21
chosen.
Digital value
Diagnosis

5 Diagnosis

5.1 Maintenance

VEGABAR process pressure transmitters are maintenance free.

5.2 Failure removal

Fault signals
With selftest and continuous self-monitoring VEGABAR offers high reliability. If however interferences occur, the diagnosis of VEGA­BAR distinguishes between atypical process conditions and faults in the VEGABAR.
Atypical process conditions
Exceeding or decreasing of the measuring range limits (fault signal extinguishes when the value is again in the measuring range).
Failure in VEGABAR
Failure in the electronics, interferences or damages in the measuring cell.
The below table assists in the processing of fault signals.
Reason Fault signal via
Clear de- “OPERATE Bargraph 0 % creasing or ???? or 100 % Current exceeding bar Digital value value of the flashes 3,6 mA measuring or range 22 mA
Overload Bargraph 0 % range of the or 100 %, digital measuring value: four flashing cell segments - - - -
Failure in all segments VEGABAR flash
DOT-matrix Bargraph Current
Digital indication output
On instruments with menu-guided adjustment with additional functions, possible reasons are displayed in the menu point Diagnosis no. in case of failure.
Diagnosis no. Meaning
1 No frequency from C/f-converter
2 Frequency signal of the capacitor
outside the limit values
3 Frequency signal of the reference
capacitor outside the limit values
4 Frequency signal temperature outside
the limit values
7 Communication to EEPROM interfered
9 Error in CRC-check sum EEPROM
Fault finding
If the indicated value does not correspond to the level in the vessel or to the process pres­sure, the following measures have to be carried out:
- check the pressure compensation (only with overpressure measuring ranges)
- check the electrical connection.
Checking pressure compensation
Open the housing of VEGABAR. The meas­ured value should not change. However if the indicated value changes, compensation of the atmospheric pressure is not ensured and the measured value will be affected. Check the breather facility on the VEGABAR-hous­ing.
VEGABAR 21 23
Checking the electrical components
Diagnosis
–+
1 2 3
mA
Voltage
V
Current meter for local control
Current
mA
External energy source
~
+
Voltage
- the terminal voltage on VEGABAR must be at least 12 V DC
Current
Current value Condition
3,8 20,5 mA standard range for output current
0 mA signal line open circuit
3,6 mA oscillator or pressure sensor ele-
22 mA short-circuit in the signal line,
ment defect
oscillator or pressure sensor ele­ment defect
24 VEGABAR 21
Instrument modification
6 Instrument modification
6.1 Interchanging adjustment modules
Indication module
Module for basic functions
Module for menu-guided adjustment
VEGABAR
Module without adjustment
Exchange of the adjustment module
Removal of the adjustment module
1 Separate VEGABAR from the power
supply
2 Loosen the screws on the cover of
VEGABAR or VEGADIS 10 3 Loosen connection line from the terminals 4 Loosen screws of the adjustment cover 5 Remove adjustment cover and plug con-
nection
Insertion of an adjustment module
6 Plug the connection of the new adjust-
ment cover to the plug-in socket of the
oscillator 7 Fasten new adjustment cover 8 Fasten connection lines to terminals 9 Close cover of VEGABAR or VEGADIS 10 10 Connect VEGABAR to power supply
Retrofitting of an indicating module
1 Separate VEGABAR from power supply 2 Loosen screws on cover of VEGABAR or
VEGADIS 10 3 Remove cover 4 Connect indicating module via the four
lines to the screw terminals (note 3.2
Connection plan“)
The modular construction of VEGABAR ena­bles to retrofit, interchange or remove adjust­ment modules and the indicating module. A retrofitting can be e.g. necessary when a
Note:
Step 4 is facilitated when the indicating mod­ule is temporarily (carefully) mounted with two screws.
VEGABAR with factory setting should be adapted to modified measurement condi­tions. Already saved data (e.g. adjustment values) are not saved in the adjustment mod­ule but in an EEPROM of the oscillator and a
5 Place indicating module in the correct
position on VEGABAR or VEGADIS 10 6 Tighten screws of the cover 7 Connect VEGABAR to power supply
new adjustment is therefore not necessary.
The connection of the modules is made:
- on the adjustment module via a 4-pole plug connection
- on the indicating module via screw termi­nals.
VEGABAR 21 25

7 Measurement with isolating systems

Measurement with isolating systems
Application limits
Isolating systems extend the temperature and product application range of pressure transmitters considerably. Therefore check in addition that the complete measurement loop suits the temperature. If due to the tempera­ture conditions, a high ambient temperature can be caused on the instrument, the use of a cooling element or a capillary will be neces­sary.
The max. permissible gauge pressure on the separating diaphragm should not exceed a value of -500 mbar. Should higher low pres­sure values be expected, please contact us respectively.
Error treatment
The influence of an isolating system is directly determined by the temperature of the isolat­ing liquid in front of the measuring cell. The temperature coefficient Tk (see 1.5 Technical data) is valid for silicone oil. It reduces with water-glycerine to 50 % of the values.
The total temperature error results in the sum of Tk
Pressure transmitter
Tk
= Tk
tot
Example: You use an isolating system with a Tk of 2,7 mbar/10 K (acc. to 1.5 Technical data).
Your pressure transmitter has a measuring range of 0 … 1 bar and a Tk of 0,05 %/10 K (acc. to technical data). This corresponds to a Tk of 0,5 mbar/10 K.
Tk
=
total
= 2,7 mbar/10 K + 0,5 mbar/10 K = 3,2 mbar/10 K
We recommend the use of a process pres­sure transmitter with the max. available accu­racy level to keep the total error within the limits.
and Tk
Isolating diaphragm
Isolating diaphragm
+ Tk
Pressure transmitter
:
26 VEGABAR 21
Notes
VEGABAR 21 27
VEGA Grieshaber KG Am Hohenstein 113 D-77761 Schiltach Phone (0 78 36) 50 - 0 Fax (0 78 36) 50 - 201 e-mail info@vega-g.de
ISO 9001
The statements on types, application, use and operating conditions of the sensors and processing systems correspond to the actual knowledge at the date of printing.
Technical data subject to alteration.
2.23 421 / Oct. 99
Loading...