VEGA BAR17 User Manual

Operating Instruction
VEGABAR 17
s
Document ID:
27636
Process pressur
e

Contents

Contents
1.1 Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.2 Target group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.3 Symbolism used . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2 For your safety
2.1 Authorised personnel . . . . . . . . . . . . . . . . . . . . . . . .
2.2 Appropriate use . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.3 Warning about incorrect use . . . . . . . . . . . . . . . . . . .
2.4 General safety instructions . . . . . . . . . . . . . . . . . . . .
2.5 Safety label on the instrument . . . . . . . . . . . . . . . . . .
2.6 CE conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.7 Measuring range - permissible process pressure . . . .
2.8 Safety instructions for Ex areas . . . . . . . . . . . . . . . . .
2.9 Environmental instructions. . . . . . . . . . . . . . . . . . . . .
3 Product description
3.1 Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.2 Principle of operation . . . . . . . . . . . . . . . . . . . . . . . .
3.3 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.4 Packaging, transport and storage . . . . . . . . . . . . . . .
4 Mounting
4.1 General instructions . . . . . . . . . . . . . . . . . . . . . . . . .
4.2 Instructions for installation . . . . . . . . . . . . . . . . . . . . .
4.3 Mounting steps. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5 Connecting to power supply
5.1 Preparing the connection . . . . . . . . . . . . . . . . . . . . .
5.2 Connection procedure. . . . . . . . . . . . . . . . . . . . . . . .
5.3 Wiring plan. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
document
4 4 4
5 5 5 5 6 6 6 6 6
7 8 8 8
10 10 10
12 13 16
6 Setup
6.1 Setup steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.2 Recalibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7 Maintenance and fault rectication
7.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7.2 Remove interferences . . . . . . . . . . . . . . . . . . . . . . . .
7.3 Instrument repair . . . . . . . . . . . . . . . . . . . . . . . . . . .
8 Dismounting
8.1 Dismounting steps . . . . . . . . . . . . . . . . . . . . . . . . . .
8.2 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
9 Supplement
9.1 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
9.2 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2 VEGABAR 17
18 18
21 21 22
23 23
24 30
27636-EN-120531
Contents
Supplementary documentation
Information:
Suppleme with the delivery. You can nd them listed in chapter "Product
description".
Editing status: 2012-05-29
ntary documents appropriate to the ordered version come
27636-EN-120531
VEGABAR 17 3

1 About this document

1 About this document

1.1 Function

is operating instructions manual provides all the information you
Th need for mounting, connection and setup as well as important instructions for maintenance and fault rectication. Please read this information before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device.

1.2 Target group

This operating instructions manual is directed to trained qualied personnel. The contents of this manual should be made available to these personnel and put into practice by them.

1.3 Symbolism used

Information, tip, note
This
symbol indicates helpful additional information.
Caution: If this result. Warning: If this warning is ignored, injury to persons and/or serious damage to the instrument can result. Danger: If this warning is ignored, serious injury to persons and/or destruction of the instrument can result.
applications
Ex
Th
symbol indicates special instructions for Ex applications.
is
l List
The dot set in front indicates a list with no implied sequence.
warning is ignored, faults or malfunctions can
à Action
This a
rrow indicates a single action.
1 Sequence
Numbers set in front indicate successive steps in a procedure.
Battery disposal
symbol indicates special information about the disposal of
This batteries and accumulators.
27636-EN-120531
4 VEGABAR 17

2 For your safety

2 For your safety
2.1 Auth
All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator.
During work on and with the device the required personal protective equipment must always be worn.
orised personnel

2.2 Appropriate use

VEGABAR 17 is a pressure transmitter for measurement of gauge pressure, absolute pressure and vacuum.
You can nd detailed information on the application range in chapter
"Product description".
Operational reliability is ensured only if the instrument is properly used according to the specications in the operating instructions manual as well as possible supplementary instructions.
For safety and warranty reasons, any invasive work on the device beyond that described in the operating instructions manual may be carried out only by personnel authorised by the manufacturer. Arbitrary conversions or modications are explicitly forbidden.

2.3 Warning about incorrect use

Inappropriate or incorrect use of the instrument can give rise to application-specic hazards, e.g. vessel overll or damage to system components through incorrect mounting or adjustment.

2.4 General safety instructions

This is a high-tech instrument requiring the strict observance of standard regulations and guidelines. The user must take note of the safety instructions in this operating instructions manual, the country­specic installation standards as well as all prevailing safety regulations and accident prevention rules.
The instrument must only be operated in a technically awless and reliable condition. The operator is responsible for trouble-free operation of the instrument.
During the entire duration of use, the user is obliged to determine the compliance of the necessary occupational safety measures with the current valid rules and regulations and also take note of new regulations.
27636-EN-120531
VEGABAR 17 5
2 For your safety
2.5 Safety label
The safety approval markings and safety tips on the device must be observed.
on the instrument

2.6 CE conformity

The device fullls the legal requirements of the applicable EC guidelines. By axing the CE marking, we conrm successful testing of the product.
You can nd the conformity ce rticate in the download section of our homepage.
2.7 Measuring range - permissible process pres-
sure
Due to the application, a measuring cell with a measuring range higher than the permissible pressure range of the process tting may have been integrated. The permissible process pressure is stated with
"Process pressure" on the type label, see chapter 3.1 "Conguration". For safety reasons, this range must not be exceeded.

2.8 Safety instructions for Ex areas

Please note the Ex-specic safety information for installation and operation in Ex areas. These safety instructions are part of the operating instructions manual and come with the Ex-approved instruments.

2.9 Environmental instructions

Protection of the environment is one of our most important duties. That is why we have introduced an environment management system with the goal of continuously improving company environmental protection.
The environment management system is certied according to DIN EN ISO 14001.
Please help us full this obligation by observing the environmental
instructions in this manual:
l Chapter "Packaging, transport and storage" l Chapter "Disposal"
6 VEGABAR 17
27636-EN-120531

3 Product description

4
3
2
1

3.1 Structure

3 Product description
pe of delivery
Sco
Structure
e scope of delivery encompasses:
Th
l VEGABAR 17 process pressure transmitter l Depending on the version, with plug connector, connection cable
or terminal housing
l Documentation
- this operating instructions manual
- Ex-specic "Safety instructions" (with Ex versions)
- if necessary, further certicates
Fig. 1: VEGABAR 17 with
1 Process tting 2 Housing with electronics 3 Pressure compensation (beneath the knurled nut) 4 Plug connector
plug
connector
according to ISO 4400
Type label
The type label contains the most important data for identication and use of the instrument:
l Article number l Serial number l Technical data l Article numbers, documentation
With the serial number, you can access the delivery data of the instrument via
www.vega.com, "VEGA T
ools" and "serial number
search". In addition to the type label outside, you can also nd the
serial number on the inside of the instrument.
27636-EN-120531
VEGABAR 17 7
3 Product description
cation area
Appli
Functional principle
Voltage supply
Packaging
3.2 Principle of
operation
VEGABAR 17 is a pressure transmitter for measurement of gauge pressure, absolute pressure or vacuum. Measured products are gases, vapours and liquids. The front ush versions are also suitable for use in viscous or contaminated products.
The process pressure acts on the sensor element via the stainless steel diaphragm. The process pressure causes a resistance change which is converted into a corresponding output signal and outputted as measured value.
1)
4 20 mA two-wire electronics for voltage supply and measured
value transmission on the same cable.

3.3 Operation

The VEGABAR 17 has no adjustment options.
However, there are two built-in potentiometers for recalibration of zero
and span.

3.4 Packaging, transport and storage

Your instrument was protected by packaging during transport. Its capacity to handle normal loads during transport is assured by a test according to DIN EN 24180.
The packaging of standard instruments consists of environment­friendly, recyclable cardboard. For special versions, PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies.
Transport
Transport must be carried out under consideration of the notes on the transport packaging. Nonobservance of these instructions can cause damage to the device.
Transport inspection
The delivery must be checked for completeness and possible transit damage immediately at receipt. Ascertained transit damage or concealed defects must be appropriately dealt with.
Storage
Up to the time of installation, the packages must be left closed and stored according to the orientation and storage markings on the outside.
Unless otherwise indicated, the packages must be stored only under the following conditions:
l Not in the open
1)
Measuring transmission liquid. Measuring ranges up to 25 bar: strain gauge (DMS) sensor element on the rear of the stainless steel diaphragm (dry).
ranges up to 16 bar: piezoresistive sensor element with internal
8 VEGABAR 17
27636-EN-120531
3 Product description
Stora
ge and transport
temperature
l Dry and dust free l Not expose l Protected against solar radiation l Avoiding mechanical shock and vibration
l Storage and transport temperature see chapter "Supplement -
d to corrosive media
Technical data - Ambient conditions"
l Relative humidity 20 85 %
27636-EN-120531
VEGABAR 17 9
4 Mounting

4 Mounting

bility for the pro-
Suita cess condition
Diaphragm protection
Checking the diaphragm
Mounting position
s
4.1 General ins
Make sure that all parts of the instrument exposed to the process, in particular the sensor element, process seal and process tting, are suitable for the existing process conditions. These include above all the process pressure, process temperature as well as the chemical properties of the medium.
You can nd the specications in chapter "Technical data" and on the type label.
To protect the diaphragm, the process tting is covered by a protective cap.
Remove the protective cap just before installation so that the diaphragm will not get damaged. It is recommended to keep the cap and use it again later for storage or transport.
tructions

4.2 Instructions for installation

Please check the diaphragm visually for damage and leaking uid before mounting and setting up the instrument. Make sure that the diaphragm doesn't get damaged during installation.
Caution:
The instr condition and has an undamaged diaphragm.
VEGABAR 17 functions in any installation position. It is mounted according to the same directives as a manometer (DIN EN 839-2).
ument may only be used if it is in a technically awless
Information:
We recommen siphons from our line of accessories.
d using lock ttings, measuring instrument holders and

4.3 Mounting steps

Welding the socket
Sealing/Screwing in
10 VEGABAR 17
For mounting VEGABAR 17, a welded socket is required. You can nd
these components in the supplementary instructions manual "Welded socket and seals".
Use the seal tting to the instrument, or in case of NPT connections, a high-resistance sealing material.
Screw VEGABAR 17 into the welded socket with a wrench applied to the hexagon of the process tting. For the proper torque see chapter "Technical data", for spanner size see chapter "Dimensions".
27636-EN-120531
Fig. 2: Mounting of VEGABAR 17
4 Mounting
27636-EN-120531
VEGABAR 17 11

5 Connecting to power supply

5 Connecting to power supply
Note safety instructions
Take note of sa-
fety instructions for Ex applica­tions
Select power supply
Select connection cable
5.1 Preparin
Always keep in mind the following safety instructions:
l Connect only in the complete absence of line voltage l If overvoltage surges are expected, overvoltage arresters should
be installed
Tip:
We recomm
In hazardous areas you must take note of the respective regulations,
conformit supply units.
The supply voltage and the current signal are carried on the same two­wire connection cable.
Provide a reliable separation between the supply circuit and the mains circuits according to DIN VDE 0106 part 101.
VEGA power supply units VEGATRENN 149AEx, VEGASTAB 690, VEGADIS 371 as well as all VEGAMETs meet this requirement. When using one of these instruments, protection class III is ensured for VEGABAR 17.
Keep in mind the following additional factors that inuence the
operating voltage:
l Output voltage of the power supply unit can be lower under
l Inuence of additional instruments in the circuit (see load values in
The instrument is connected with standard two-wire cable without screen. If electromagnetic interference is expected which is above the test values of EN 61326 for industrial areas, screened cable should be used.
Use cable with round cross-section. A cable outer diameter of 5 … 9 mm (0.2 0.35 in) ensures the seal eect of the cable gland. If you are
using cable with a dierent diameter or cross-section, exchange the seal or use a suitable cable gland.
For the version with round plug connector M12 x 1, a suitable, ready- made connection cable (article no. ASL.1S.) in 5 m, 10 m or 25 m lengths is available from the line of VEGA accessories.
y and type approval certicates of the sensors and power
nominal load (with a sensor current of 20.5 mA or 22 mA in case of fault message)
chapter "Technical data")
g the connection
end VEGA overvoltage arrester ÜSB 62-36G.X.
27636-EN-120531
12 VEGABAR 17
5 Connecting to power supply
screening and
Cable grounding
Selec
t conne tion cable applications
c-
for Ex
Connect the cable screen on both ends to ground potential.
If potenti
processing side must be made via a ceramic capacitor (e. g. 1 nF, 1500 V). The low frequency potential equalisation currents are thus
suppressed, but the protective eect against high frequency interfer­ence signals remains.
Warning:
Considerab as vessels with cathodic corrosion protection. Very large equalisation currents can ow through the cable screen when the screen is grounded on both ends. To avoid this, the cable screen must be connected to ground potential only on one end (inside the switching cabinet) in such applications. The cable screen must not be connected to the internal ground terminal in the sensor and the outer ground terminal on the housing not to potential equalisation!
Information:
The tube, etc.) are conductively connected with the inner and outer ground terminal on the housing. This connection exists either as a direct metallic contact or via the shielding of the special connection cable on instruments with external electronics. You can nd specications on the potential connections within the instrument in chapter "Technical
data".
Ta appl
al equalisation currents are expected, the connection on the
le potential dierences exist inside galvanic plants as well
metallic parts of the instrument (antenna, transmitter, concentric
ke note of the corresponding installation regulations for E
ications.
x

5.2 Connection procedure

Connection via angle
plug connec
27636-EN-120531
VEGABAR 17 13
tor
Proceed as follows:
1 Loosen the screw on the rear of the plug connector 2 Remove the plug connector and seal from VEGABAR 17 3 Remove the plug insert from the plug housing
1
2
3
2
3
1
4
5 Connecting to power supply
Fig. 3: Loosen the plug
1 Cable gland 2 Plug insert 3 Plug housing
insert
4 Remove approx. 5 cm of the cable mantle, strip approx. 1 cm
insulation from the individual wires
5 Lead the cable through the cable gland into the plug housing 6 Connect the wire ends to the screw terminals according to the
wiring plan
Fig. 4: Connection to the
1 Cable gland 2 Plug housing 3 Plug insert 4 Plug seal
screw terminals
7 Snap the plug insert into the plug housing and insert the sensor
seal
8 Plug the plug insert with seal to VEGABAR 17 and tighten the
screw
The electrical connection is nished.
Connection via angle plug connector with hin­ged cover
Proceed as follows:
14 VEGABAR 17
27636-EN-120531
4
32
1
3
2
1
1
5
2
3
4
5 Connecting to power supply
1 Loosen the screw in the cover of the plug connector 2 Open the
cover and remove it
3 Press the plug insert downwards 4 Loosen the screws of the strain relief and cable entry
Fig. 5: Loosen the plug
1 Plug insert 2 Strain relief 3 Cable gland 4 Plug housing
insert
5 Remove approx. 5 cm of the cable mantle, strip approx. 1 cm
insulation from the individual wires
6 Lead the cable through the cable gland into the plug housing 7 Connect the wire ends to the screw terminals according to the
wiring plan
Fig. 6: Connection to the
1 Cable gland 2 Cover 3 Plug housing 4 Plug insert 5 Plug seal
screw terminals
27636-EN-120531
VEGABAR 17 15
1
2
3
+
-
1
3
+
-
1
324
1
5 Connecting to power supply
d plug connector
Angle according to ISO 440
0
8 Snap the plug insert into the plug housing and insert the sensor
seal
Information:
Note the correct arrangement, see illustration
9 Tighten the screws on the strain relief and cable entry 10 Hook in the cover and push onto the plug connection, tighten
cover screw
11 Plug the plug insert with seal to VEGABAR 17 and tighten the
screw
The electrical connection is nished.

5.3 Wiring plan

Fig. 7: Wiring VEGABAR 17
1 Voltage supply and signal output
Round plug connector M12 x 1
Fig. 8: Wiring plan, round plug
1 Voltage supply and signal output
Connection via connection cable with 4-pole socket M12 x 1 (accessory)
16 VEGABAR 17
plan, angle
connector according to ISO 4400, top view to
plug
connector M12 x 1, top view to VEGABAR 17
27636-EN-120531
1
2
3
321
+
– +
4 5
1
2
Cable
5 Connecting to power supply
Wire colour Socket
Brown 1 White 2 Blue 3 Black 4
outlet
Terminal housing
Fig. 9: Wiring plan cable
1 brown (+) power supply and signal output 2 green (-) power supply and signal output 3 blue = cable screen
Fig. 10: Wiring plan, terminal housing
power supply or the processing system
1 To 2 Control instrument (4 … 20 mA measurement)
outlet
27636-EN-120531
VEGABAR 17 17

6 Setup

6 Setup

6.1 Setup steps

mounting and electrical connection, VEGABAR 17 is ready for
After
operation.
VEGABAR 17 delivers a current of 4 20 mA corresponding to the actual process pressure.
Further settings are not necessary.

6.2 Recalibration

With both instruments with thread ring or eld housing, zero and span can be adjusted via integrated potentiometers. Adjustment range:
l Zero ± 5 % l Span ± 5 %
This allows, for example, the consideration of an installation position dierent from the reference installation position.
A shifting of zero shifts span also respectively.
Note:
The potenti calibration equipment (at least 3 times more precise than the deviation of VEGABAR 17).
Recommended recalibration cycle: 1 year.
ometer for span should only be used if you have adequate
Instrum
ents with plug connector or let
18 VEGABAR 17
cable out-
Proceed as follows:
1 Loosen the plug connector and screw the screwed ring in
connected status
2 Place the plug connector onto the instrument place and pull both
carefully out of the instrument
27636-EN-120531
1
2
3
4
5
6
2
1
6 Setup
Fig. 11: Open the instrument
1 Plug 2 Plug seal 3 Screwed ring 4 Instrument plug 5 Plug seal 6 Housing
connector
3 Set zero in unpressurized status, check 4 mA signal in the circuit 4 Set span with exact reference pressure 5 Check zero
Fig. 12: Adjustment of zero
1 zero (Z) 2 span (S)
and span
6 Assemble the instrument and connect it.
27636-EN-120531
VEGABAR 17 19
4
5
6 Setup
Instruments with termi­nal housin
g
Proceed as follows:
1 Screw on
Fig. 13: Adjustment of zero
1 zero (Z) 2 span (S)
the housing cover in connected status
and span
2 Set zero in unpressurized status, check 4 mA signal in the circuit 3 Set a span with a suciently precise reference pressure 4 Check zero 5 Screw the housing cover back on
20 VEGABAR 17
27636-EN-120531
7 Maintenance and fault recticati
7 Maintenance and fault rectication
on
Reaction when malfunc­tions occur
reasons
Failure
Fault rectication
24 hour ser
Check the 4 20 mA signal
vice hotline
7.1 Maintena
If the instrument is used properly, no special maintenance is required in normal operation.
nce

7.2 Remove interferences

The operator of the system is responsible for taking suitable measures to rectify faults.
VEGABAR 17 oers maximum reliability. Nevertheless, faults can occur during operation. These may be caused by the following, e.g.:
l Sensor l Process l Voltage supply l Signal processing
The rst measure to be taken is to check the output signal. In many cases, the causes can be determined this way and the faults rectied.
Should these measures not be successful, please call in urgent cases the VEGA service hotline under the phone no. +49 1805 858550.
The hotline is available to you 7 days a week round-the-clock. Since we oer this service world-wide, the support is only available in the English language. The service is free of charge, only the standard telephone costs will be charged.
? No 4 20 mA signal
l Connection to voltage supply wrong
à Check connection according to chapter "Connection steps"
and if necessary, correct according to chapter "Wiring plan"
l No voltage supply
à Check cables for breaks; repair if necessary
l Operating voltage too low or load resistance too high
à Check, adapt if necessary
? Steady output signal with pressure change
l electronics module or measuring cell defective
à Exchange the instrument or send it in for repair
In Ex applications, the regu
circuits must be observed.
27636-EN-120531
VEGABAR 17 21
lations for the wiring of intrinsically safe
7 Maintenance and fault recticati
on
Reaction after fault rec­tication
Depending on the reason for the fault and the measures taken, the
described in chapter "Set up" may have to be carried out again.
steps

7.3 Instrument repair

If a repair is necessary, please proceed as follows:
You can download a return form (23 KB) from our Internet homepage
www.vega.com under: "Dow form".
By doing this you help us carry out the repair quickly and without having to call back for needed information.
l Print and ll out one form per instrument l Clean the instrument and pack it damage-proof
l Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
l Please ask the agency serving you for the address of your return
shipment. You can nd the respective contact data on our website www.vega.com under: "Compa
nloads - Forms and certicates - Repair
ny - VEGA worldwide"
22 VEGABAR 17
27636-EN-120531

8 Dismounting

8 Dismounting
8.1 Dism
Warning:
Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel, high temperatures, corrosive or toxic products etc.
Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order.
ounting steps

8.2 Disposal

The instrument consists of materials which can be recycled by specialised recycling companies. We use recyclable materials and have designed the electronics to be easily separable.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/EG and the respective national laws. Pass the instrument directly on to a specialised recycling company and do not use the municipal collecting points. These may be used only for privately used products according to the WEEE directive.
Correct disposal avoids negative eects on humans and the environ­ment and ensures recycling of useful raw materials.
Materials: see chapter "Technical data"
If you have no way to dispose of the old instrument properly, please
contact us concerning return and disposal.
27636-EN-120531
VEGABAR 17 23

9 Supplement

9 Supplement
9.1 Technical
data
General data
Parameter, pressure Gauge pressure, absolute pressure, vacuum Measuring principle Piezoresistive/Thin lm DMS Communication interface None
Materials and weights
Materials, wetted parts
- Process tting 316Ti
- Diaphragm 316Ti
- Diaphragm with front ush version 316Ti, Hastelloy C4
- Seal, O-ring FPM, FKM, EPDM, NBR
Materials, non-wetted parts
- Internal transmission liquid Synthetic oil, Halocarbon oil
2)3)
- Housing 316Ti
- Terminal housing 316Ti
- Ground terminal 316Ti
- Plug PA
- Cable gland PA, 316Ti
- Plug seal Silicone
- Connection cable PUR
Available cable length max. 40 m Weight approx.
- Version with plug connector, cable outlet 0.2 kg (0.441 lbs)
- Version with terminal housing 0.35 kg (0.772 lbs)
Torque max. 50 Nm (36.88 lbft)
Output variable
Output signal 4 20 mA Zero and span adjustable via potentiometer ±5 % Dead time 1 ms Step response time (10 90 %)
- Standard version ≤ 1 ms
2)
Synthetic processing industry. For measuring ranges up to 25 bar not available.
3)
Halocarbon ranges, not with absolute measuring ranges < 1 bar
oil: For measuring ranges
oil: With version oil
and greasefree, not with vacuum measuring
up to 16 bar, FDA listed for the food
.
abs
24 VEGABAR 17
27636-EN-120531
9 Supplement
- Version for medium temperature < -30 °C
10 ms
(-22 °F)
- Vers
ion for measuring ranges > 25 bar 10 ms
- Version with front-ush diaphragm 10 ms
Reference conditions and actuating variables (according to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1
- Temperature +15 … +25 °C (+59 +77 °F)
- Relative humidity 45 75 %
- Air pressure 860 … 1060 mbar/86 106 kPa (12.5 15.4 psi)
Determination of characteristics Limit point adjustment according to IEC 61298-2 Reference installation position upright, diaphragm points downward Inuence of the installation position depending on the chemical seal version
Input variable
The availability of the respective measuring range depends on the corresponding process tting. The specications on overload capacity are only an overview and refer to the measuring cell.
Limitations due to the material and form of the process tting are possible. The specications on
the type label always apply.
Nominal range Overload capacity, max.
pressure
Gauge pressure
-0.1 0 bar/-10 0 kPa 1 bar/100 kPa -1 bar/-100 kPa
-0.16 0 bar/-16 0 kPa 1.5 bar/150 kPa -1 bar/-100 kPa
-0.25 0 bar/-25 0 kPa 2 bar/200 kPa -1 bar/-100 kPa
-0.4 0 bar/-40 0 kPa 2 bar/200 kPa -1 bar/-100 kPa
-0.6 0 bar/-60 0 kPa 4 bar/400 kPa -1 bar/-100 kPa
-1 0 bar/-100 0 kPa 5 bar/500 kPa -1 bar/-100 kPa
-1 3 bar/-100 300 kPa 10 bar/1000 kPa -1 bar/-100 kPa
0 … 0.1 bar/0 … 10 kPa 1 bar/100 kPa -1 bar/-100 kPa 0 … 0.16 bar/0 … 16 kPa 1.5 bar/150 kPa -1 bar/-100 kPa 0 … 0.25 bar/0 … 25 kPa 2 bar/200 kPa -1 bar/-100 kPa 0 … 0.4 bar/0 … 40 kPa 2 bar/200 kPa -1 bar/-100 kPa 0 … 0.6 bar/0 … 60 kPa 4 bar/400 kPa -1 bar/-100 kPa 0 … 1 bar/0 … 100 kPa 5 bar/500 kPa -1 bar/-100 kPa 0 … 1.6 bar/0 … 160 kPa 10 bar/1000 kPa -1 bar/-100 kPa 0 … 2.5 bar/0 … 250 kPa 10 bar/1000 kPa -1 bar/-100 kPa 0 … 4 bar/0 … 40 kPa 17 bar/1700 kPa -1 bar/-100 kPa 0 … 6 bar/0 … 600 kPa 35 bar/3500 kPa -1 bar/-100 kPa 0 … 10 bar/0 … 1000 kPa 35 bar/3500 kPa -1 bar/-100 kPa 0 … 16 bar/0 … 1600 kPa 80 bar/8000 kPa -1 bar/-100 kPa
Overload capacity, min. pressure
27636-EN-120531
VEGABAR 17 25
9 Supplement
Nominal range Overload capacity, max.
0 … 25 bar/0 … 2500 kPa 50 bar/5000 kPa -1 bar/-100 kPa 0 … 40 bar/0 … 4000 kPa 80 bar/8000 kPa -1 bar/-100 kPa 0 … 60 bar/0 … 6000 kPa 120 bar/12 MPa -1 bar/-100 kPa 0 … 100 bar/0 … 10 MPa 200 bar/20 MPa -1 bar/-100 kPa 0 … 160 bar/0 … 16 MPa 320 bar/32 MPa -1 bar/-100 kPa 0 … 250 bar/0 … 25 MPa 500 bar/50 MPa -1 bar/-100 kPa 0 … 400 bar/0 … 40 MPa 800 bar/80 MPa -1 bar/-100 kPa 0 … 600 bar/0 … 60 MPa 1200 bar/120 MPa -1 bar/-100 kPa 0 … 1000 bar/0 … 100 MPa 1500 bar/150 MPa -1 bar/-100 kPa
Absolute pressure
0 … 0.25 bar/0 … 25 kPa 2 bar/200 kPa 0 … 0.4 bar/0 … 40 kPa 2 bar/200 kPa 0 … 0.6 bar/0 … 60 kPa 4 bar/400 kPa 0 … 1 bar/0 … 100 kPa 5 bar/500 kPa 0 … 1.6 bar/0 … 160 kPa 10 bar/1000 kPa 0 … 2.5 bar/0 … 250 kPa 10 bar/1000 kPa 0 … 4 bar/0 … 400 kPa 17 bar/1700 kPa 0 … 6 bar/0 … 600 kPa 35 bar/3500 kPa 0 … 10 bar/0 … 1000 kPa 35 bar/3500 kPa 0 … 16 bar/0 … 1600 kPa 80 bar/8 MPa
Deviation
4)
pressure
Overload pressure
capacity, min.
Deviation 0.5 %
Inuence of the product or ambient temperature
5)
The following specications apply to values within the compensated temperature range, i.e. 0 80 °C (176 °F), reference temperature 20 °C (68 °F ).
Average temperature coecient of the zero signal
- Standard < 0.2 %/10 K
- Meas. ranges 0 … 0.1 and 0 … 0.16 bar < 0.4 %/10 K
Average temperature coecient of the span < 0.2 %/10 K
The following specications are valid for values not within the compensated temperature range.
Average temperature coecient of the zero signal
- Standard typ. < 0.2 %/10 K
- Meas. ranges 0 … 0.1 and 0 … 0.16 bar typ. < 0.4 %/10 K
4)
5)
Relating ducibility. Relating
to the adjusted
to the set
span, incl. non-linearity, hysteresis and non-repro-
span, incl. hysteresis and repeatability.
26 VEGABAR 17
27636-EN-120531
Average temperature coecient of the span typ. < 0.2 %/10 K
2
1
55°C
(131°F)
80°C
(176°F)
130°C
(266°F)
150°C
(302°F)
Long-term stability (according to DIN 16086, DINV 19259-1 and IEC 60770-1)
Long-term drift of the zero signal
6)
< 0.2 %/year
Ambient conditions
Ambient temperature (note temperature derating!)
- Cable outlet -20 +80 °C (-4 +176 °F)
- Round plug connector M12 x 1 -25 … +80 °C (-13 +176 °F)
- Angled plug connector according to
-40 +80 °C (-40 +176 °F)
ISO 4400
- Terminal housing -50 +80 °C (-58 +176 °F)
- with cooling element -20 +80 °C (-4 +176 °F)
9 Supplement
Fig. 14: Temperature
1 Ambient 2 Process
derating VEGABAR 17
temperature temperature
Storage and transport temperature
- Standard -30 +100 °C (-22 +212 °F)
- with cooling element -20 +100 °C (-4 +212 °F)
Process conditions
Product temperature
- Standard -30 +100 °C (-22 +212 °F)
- additional -30 +125 °C (-22 +257 °F)
- with cooling element -20 +150 °C (-4 +302 °F)
- Measuring ranges from 400 bar -30 +70 °C (-22 +158 °F)
Shock resistance
- Version with terminal housing 600 g according to IEC 60068-2-27 (mechanical shock)
6)
Under
reference
conditions, relating
to
the adjusted span.
27636-EN-120531
VEGABAR 17 27
9 Supplement
- Version with plug connector or cable outlet
according to IEC 60068-2-27 (mechanical
1000 g shock)
- Version with cooling element 400 g according to IEC 60068-2-27 (mechanical
shock)
Vibration resistance
- Version with terminal housing or cooling element
- Version with plug connector or cable outlet
10 g according to IEC 60068-2-6 (resonance vibration)
20 g according to IEC 60068-2-6 (vibration at resonance)
Electromechanical data
Angled plug connector
- Version 4-pin according to ISO 4400
- Outer cable diameter 6 8 mm
Circular plug connector
- Version 4-pole M12 x 1
Cable outlet
- Diameter 6.8 mm
Terminal housing
- Cable entry for cable outer diameter 6 8 mm
- Spring-loaded terminals for wire cross-
2.5 mm² (AWG 14)
section up to
Voltage supply
Operating voltage
- Version with plug or cable outlet 10 30 V DC
- Version with terminal housing 11 30 V DC
Load
- Version with cable outlet RA ≤ (U-10V)/0,02 A- (length of the cable version in
m x 0.14 Ω)
- Version with plug see diagram
- Version with terminal housing see diagram
Electrical protective measures
Protection rating
7)
- with angled plug connector IP 65
- with round plug connection IP 65
- with cable outlet IP 67, IP 68 (0.5 bar)
- with terminal housing IP 67
Voltage resistance Insulation according to EN 50020, 6.4, 12
7)
According
to EN 60529/IEC 529.
28 VEGABAR 17
27636-EN-120531
9 Supplement
Interference resistance
- HF 10 V/m
- Burst 2 kV
protective measures
Other
- Interpolation protection Available
- Overvoltage protection up to 36 V DC
Approvals
Instruments with approvals can have dierent technical data depending on the version. That's why the associated approval documents have to be noted with these instruments. They are
part of the delivery or can be downloaded under number search" as well as via "Downloads" and "Approvals".
www.vega.com via "VE
GA Tools" and "serial
27636-EN-120531
VEGABAR 17 29
G1/4
~116 mm (4.57 ")
2 mm
(0.08")
13 mm
(0.51")
2 mm
ø 5 mm
(0.20")
ø 9,5 mm
(0.37")
SW27
G1/2
~123 mm (4.84 ")
ø 27 mm
(1.06")
3 mm
(0.12")
3 mm
20 mm
(0.79")
ø 6 mm
(0.24")
ø 17,5 mm
(0.69")
G1/2
14 mm
(0.55")
10 mm
(0.39")
~119,5 mm (4.70")
G1/2
~123 mm (4.84 ")
20 mm
(0.79")
3 mm
(0.12")
ø 17,5 mm
(0.69")
G1/2
10 mm
(0.39")
ø 18 mm
(0.71")
20,5 mm
(0.81")
~126 mm (4.96")
G1
20,5 mm
(0.81")
10 mm
(0.39")
~126 mm (4.96")
ø 30 mm
(1.18")
SW41
G1
28 mm
(1.10")
~135 mm (5.31")
ø 30 mm
(1.18")
1/4NPT
1/2NPT
~116 mm (4.57")
13 mm
(0.51")
~123 mm (4.84")
19 mm
(0.75")
GBX
NBX NDX
84L/84B
GDX
TBX
851/85L/85B861/86L/86B
GTX
9 Supplement

9.2 Dimensions

BAR 17 - standard housing
VEGA
Fig. 15: VEGABAR 17 standard connection, GTX = G½ A according
25 bar, 851/85L/85B = G1 B front-ush with O-ring up to 1.6 bar, 861/86L/86B = G½ B front-ush with O-ring > 1.6 bar, NBX = ¼ NPT thread, NDX = ½ NPT thread
30 VEGABAR 17
housing, GBX = G¼ B manometer
to DIN 3852-E, TBX = G½ B, inside G¼ B, 84L/84B = G1 B front-ush max.
connection, GDX = G½ B manometer
27636-EN-120531
SW27
G1/4
G1/2
1/4NPT
1/2NPT
G1
10 mm
(0.39")
~136 mm (5.35 ")
~136 mm (5.35")
~143 mm (5.63 ")
ø 27 mm
(1.06")
SW41
G1
28 mm
(1.10")
~155 mm (6.10")
ø 30 mm
(1.18")
13 mm
(0.51")
2 mm
(0.08")
3 mm
(0.12")
3 mm
20 mm
(0.79")
13 mm
(0.51")
2 mm
ø 5 mm
(0.20")
ø 6 mm
(0.24")
ø 9,5 mm
(0.37")
ø 17,5 mm
(0.69")
G1/2
ø 30 mm
(1.18")
G1/2
10 mm
(0.39")
ø 18 mm
(0.71")
G1/2
14 mm
(0.55")
10 mm
(0.39")
~139,5 mm (5.49")
20,5 mm
(0.81")
20,5 mm
(0.81")
20 mm
(0.79")
3 mm
(0.12")
ø 17,5 mm
(0.69")
~146 mm (5.75")
~143 mm (5.63 ")
~146 mm (5.75")
~143 mm (5.63 ")
19 mm
(0.75")
GBX
NBX NDX
84L/84B
GDX
TBX
851/85L/85B861/86L/86B
GTX
9 Supplement
VEGABAR 17 - Standard ho
using (Ex version)
Fig. 16: VEGABAR 17 standard housing, GBX = G¼ B manometer connection, GDX = G½ B manometer connection, GTX = G½ A according
25 bar, 851/85L/85B = G1 B front-ush with O-ring up to 1.6 bar, 861/86L/86B = G½ B front-ush with O-ring > 1.6 bar, NBX = ¼ NPT thread, NDX = ½ NPT thread
27636-EN-120531
VEGABAR 17 31
to DIN 3852-E, TBX = G½ B, inside G¼ B, 84L/84B = G1 B front-ush max.
SW27
G1/4
G1/2
1/4NPT
1/2NPT
G1
20,5 mm
(0.81")
10 mm
(0.39")
~128 mm (5.04 ")
~128 mm (5.04")
~135 mm (5.31 ")
ø 27 mm
(1.06")
SW41
G1
28 mm
(1.10")
~147 mm (5.79")
ø 30 mm
(1.18")
13 mm
(0.51")
2 mm
(0.08")
3 mm
(0.12")
3 mm
20 mm
(0.79")
~135 mm (5.31")
13 mm
(0.51")
2 mm
ø 5 mm
(0.20")
ø 6 mm
(0.24")
ø 9,5 mm
(0.37")
ø 17,5 mm
(0.69")
G1/2
ø 30 mm
(1.18")
G1/2
10 mm
(0.39")
ø 18 mm
(0.71")
G1/2
14 mm
(0.55")
10 mm
(0.39")
~131,5 mm (5.18")
20,5 mm
(0.81")
20 mm
(0.79")
3 mm
(0.12")
ø 17,5 mm
(0.69")
19 mm
(0.75")
ø 61 mm (2.40")
max. 90 mm
(3.54")
~138 mm (5.43")
~138 mm (5.43")
~135 mm (5.31 ")
GBX
NBX NDX
84L/84B
GDX
TBX
851/85L/85B861/86L/86B
GTX
9 Supplement
VEGABAR 17 - terminal housing
Fig. 17: VEGABAR 17 terminal housing, GBX = G¼ B manometer connection, GDX = G½ B manometer connection, GTX = G½ A according
25 bar, 851/85L/85B = G1 B front-ush with O-ring up to 1.6 bar, 861/86L/86B = G½ B front-ush with O-ring > 1.6 bar, NBX = ¼ NPT thread, NDX = ½ NPT thread
32 VEGABAR 17
to DIN 3852-E, TBX = G½ B, inside G¼ B, 84L/84B = G1 B front-ush max.
27636-EN-120531
VEGABAR 17 - Cooling elements, plug, cable outlet
G1/2
G1
~48 mm
(1.89")
~ 17,5 mm
(0.69")
27,5 mm
(1.08")
~ 17,5 mm
(0.69")
13,5 mm
(0.53")
12 mm
(0.47")
1
2
3 4
5
Fig. 18: VEGABAR 17 - Cooling elements, plug, cable outlet
1 Cooling 2 Cooling element G1 B 3 Plug according to ISO 4400 4 Cable outlet 5 M12 x 1 plug
element G½ B
9 Supplement
27636-EN-120531
VEGABAR 17 33
9 Supplement
34 VEGABAR 17
27636-EN-120531
9 Supplement
27636-EN-120531
VEGABAR 17 35
VEGA Grieshaber KG
ISO 9001
Am Hohenstein 113 77761 Schiltach
Germany Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail: info.de@vega.com
www.vega.com
Printing date:
All statements concerning scope of delivery, application,
practical use
and operating conditions of the sensors and
processing systems correspond to the information avail-
able at the time of printing.
© VEGA Grieshaber KG, Schiltach/Germany 2012
Subject to change without prior notice 27636-EN-120531
Loading...