VEGA BAR12 User Manual

Ope
rating Instructions
VEGABAR 12
Document ID:
35544
Proc
ess pressure

Contents

Conten
1 About this document
2 For your safety
3 Product description
4 Mounting
5 Connecting to power supply
1.1 Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.2 Target group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.3 Symbolism used . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1 Authorised personnel . . . . . . . . . . . . . . . . . . . . . . . .
2.2 Appropriate use . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.3 Warning about misuse . . . . . . . . . . . . . . . . . . . . . . .
2.4 General safety instructions . . . . . . . . . . . . . . . . . . . .
2.5 Safety label on the instrument . . . . . . . . . . . . . . . . . .
2.6 CE conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.7 Measuring range - permissible process pressure . . . .
2.8 Environmental instructions. . . . . . . . . . . . . . . . . . . . .
3.1 Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.2 Principle of operation . . . . . . . . . . . . . . . . . . . . . . . .
3.3 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.4 Packaging, transport and storage . . . . . . . . . . . . . . .
4.1 General instructions . . . . . . . . . . . . . . . . . . . . . . . . .
4.2 Mounting instructions . . . . . . . . . . . . . . . . . . . . . . . .
4.3 Mounting steps. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.1 Preparing the connection . . . . . . . . . . . . . . . . . . . . .
5.2 Wiring plan. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.3 Switch on phase. . . . . . . . . . . . . . . . . . . . . . . . . . . .
4 4 4
5 5 5 5 6 6 6 6
7 8 8 8
10 10 10
11 12 12
6 Set up
6.1 Setup steps without adjustment . . . . . . . . . . . . . . . . .
6.2 Setup steps with PC operation . . . . . . . . . . . . . . . . .
7 Maintenance and fault rectication
7.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7.2 Fault rectication . . . . . . . . . . . . . . . . . . . . . . . . . . .
7.3 Instrument repair . . . . . . . . . . . . . . . . . . . . . . . . . . .
8 Dismounting
8.1 Dismounting steps . . . . . . . . . . . . . . . . . . . . . . . . . .
8.2 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
9 Supplement
9.1 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
9.2 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2 VEG
13 13
14 14 15
16 16
35544-EN-110616
17 20
ABAR 12
Contents
ng status: 2011-05-27
Editi
Supplementary documentation
Informa
tion:
Supplementary documents appropriate to the ordered version come with the delivery. You can nd them listed in chapter "Product
description".
35544-EN-110616
VEGABAR 12 3
1 Abou
t this document
1 Abou
t this document

1.1 Function

This operating instructions manual provides all the information you need for mounting, connection and setup as well as important instructions for maintenance and fault rectication. Please read this information before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device.
1.2 Target group
This operating instructions manual is directed to trained qualied personnel. The contents of this manual should be made available to these personnel and put into practice by them.
1.3 Symbolism used
Inform
ation, tip, note
This symbol indicates helpful additional information.
Cauti
on: If this warning is ignored, faults or malfunctions can
result. Warning: If this warning is ignored, injury to persons and/or serious damage to the instrument can result. Danger: If this warning is ignored, serious injury to persons and/or destruction of the instrument can result.
applications
Ex
This symbol indicates special instructions for Ex applications.
l List
The dot set in front indicates a list with no implied sequence.
à Action
Th
is arrow indicates a single action.
1 Sequence
Numbers set in front indicate successive steps in a procedure.
4 VEG
35544-EN-110616
ABAR 12
your safety
2 For
or your safety
2 F

2.1 Authorised personnel

All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator.
During work on and with the device the required personal protective equipment must always be worn.

2.2 Appropriate use

VEGABAR 12 is a pressure transmitter for the bubble-through method with neutral gases.
You can nd detailed information on the application range in chapter
"Product description".
Operational reliability is ensured only if the instrument is properly used according to the specications in the operating instructions manual as well as possible supplementary instructions.
For safety and warranty reasons, any invasive work on the device beyond that described in the operating instructions manual may be carried out only by personnel authorised by the manufacturer. Arbitrary conversions or modications are explicitly forbidden.

2.3 Warning about misuse

Inappropriate or incorrect use of the instrument can give rise to application-specic hazards, e.g. vessel overll or damage to system components through incorrect mounting or adjustment.

2.4 General safety instructions

This is a high-tech instrument requiring the strict observance of standard regulations and guidelines. The user must take note of the safety instructions in this operating instructions manual, the country­specic installation standards as well as all prevailing safety regulations and accident prevention rules.
The instrument must only be operated in a technically awless and reliable condition. The operator is responsible for trouble-free operation of the instrument.
During the entire duration of use, the user is obliged to determine the compliance of the necessary occupational safety measures with the current valid rules and regulations and also take note of new regulations.
35544-EN-110616
VEGABAR 12 5
2 For
your safety
2.5 Safety
The safety approval markings and safety tips on the device must be observed.
label on the instrument

2.6 CE conformity

This device fullls the legal requirements of the applicable EC guidelines. By attaching the CE mark, VEGA provides a conrmation of successful testing. You can nd the CE conformity declaration in the download area of
2.7 Measu
www.vega.com.
ring range - permissible process pres-
sure
Due to the application, a measuring cell with higher measuring range than the permissible pressure range of the process pressure can be integrated. The permissible process pressure is stated with "process pressure" on the type label, see chapter 3.1 "Conguration". For safety reasons, this range must not be exceeded.

2.8 Environmental instructions

Protection of the environment is one of our most important duties. That is why we have introduced an environment management system with the goal of continuously improving company environmental protection.
The environment management system is certied according to DIN EN ISO 14001.
Please help us full this obligation by observing the environmental
instructions in this manual:
l Chapter "Packaging, transport and storage" l Chapter "Disposal"
6 VEG
35544-EN-110616
ABAR 12
3
2
1
3 Produc
t description
pe of delivery
Sco
Structure
3 Produc
t description

3.1 Structure

The scope of delivery encompasses:
l VEGABAR 12 process pressure transmitter l Mounting accessories, depending on the version l Documentation
- this operating instructions manual
- Operating instructions 35584 - PC operation VEGABAR 12
(optional)
- if necessary, further certicates
terminal connection
Type
label
Fig. 1: VEGABAR 12 with
1 Process tting 2 Housing with electronics 3 Pressure compensation
The type label contains the most important data for identication and use of the instrument:
l Article number l Serial number l Technical data l Article number, documentation
The serial number allows you to access the delivery data of the instrument via
search".
www.vega.com, "VEG
A Tools" and "serial number
35544-EN-110616
VEGABAR 12 7
3 Produc
t description
Appli
cation area
Functional principle
Voltage supply
Packaging
3.2 Princi
VEGABAR 12 is a pressure transmitter for gauge pressure measure- ment of neutral gases.
The sensor element is the CERTEC ceramic diaphragm. The process pressure causes a capacitance change in the measuring cell via the ceramic diaphragm. This change is converted into an appropriate output signal and outputted as measured value.
Two-wire electronics 4 … 20 mA for power supply and measured value transmission over the same cable.
ple of operation
®
measuring cell with rugged

3.3 Operation

The optional operation of the instrument is carried out via a PC with interface adapter and VEGA adjustment software.

3.4 Packaging, transport and storage

The device was protected by packaging during transport. Its capacity to handle normal loads during transport is assured by a test according to DIN EN 24180.
The packaging of standard instruments consists of environment­friendly, recyclable cardboard. For special versions, PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies.
Transport
Transport inspection
Storage
8 VEG
Transport must be carried out under consideration of the notes on the
transport packaging. Nonobservance of these instructions can cause damage to the device.
The delivery must be checked for completeness and possible transit damage immediately at receipt. Ascertained transit damage or concealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and stored according to the orientation and storage markings on the outside.
Unless otherwise indicated, the packages must be stored only under the following conditions:
l Not in the open l Dry and dust free l Not exposed to corrosive media l Protected against solar radiation l Avoiding mechanical shock and vibration
35544-EN-110616
ABAR 12
3 Produc
t description
Storage and transport temperature
l Storag
e and transport temperature see chapter "Supplement -
Technical data - Ambient conditions"
l Relative humidity 20 85 %
35544-EN-110616
VEGABAR 12 9
4 M
ounting
bility for the pro-
Suita
conditions
cess
Mounting position
4 Moun
ting

4.1 General instructions

Make sure that all parts of the instrument exposed to the process, in particular the sensor element, process seal and process tting, are suitable for the existing process conditions. These include above all the process pressure, process temperature as well as the chemical properties of the medium.
You can nd the specications in chapter "Technical data" or on the type label.

4.2 Mounting instructions

VEGABAR 12 functions in any installation position.
Informa
tion:
We recommend using the measuring instrument holder of the line of mounting accessory.

4.3 Mounting steps

For mounting VEGABAR 12, a connection with G1/8" outer thread is required.
10 VEG
35544-EN-110616
ABAR 12
ecting to power supply
5 Conn
Note safety instructions
Select power supply
Select connection cable
Cable screening and
groun
ding
5 Conn
ecting to power supply

5.1 Preparing the connection

Always keep in mind the following safety instructions:
l Connect only in the complete absence of line voltage l If voltage surges are expected, install overvoltage arresters
The supply voltage and the current signal are carried on the same two­wire connection cable.
Provide a reliable separation between the supply circuit and the mains circuits according to DIN VDE 0106 part 101.
Keep in mind the following additional factors that inuence the operating voltage:
l Output voltage of the power supply unit can be lower under
nominal load (with a sensor current of 3.6 mA in case of fault message)
l Inuence of additional instruments in the circuit (see load values in
chapter "Technical data")
The instrument is connected with standard two-wire cable without screen. If electromagnetic interference is expected which is above the test values of EN 61326 for industrial areas, screened cable should be used.
Connect the cable screen on both ends to ground potential.
If potential equalisation currents are expected, the connection on the
processing side must be made via a ceramic capacitor (e. g. 1 nF, 1500 V). The low frequency potential equalisation currents are thus
suppressed, but the protective eect against high frequency interfer­ence signals remains.
ning:
War
Within galvanic plants as well as vessels with cathodic corrosion protection there are considerable potential dierences. Considerably equalisation currents can be caused via the cable scrren when the screen is earthed on both ends. To avoid this, the cable screen must only connected to ground potential on one side of the switching cabinet in such applications. The cable screen must not be connected to the internal ground terminal in the sensor and the outer ground terminal on the housing not to the potential equalisation!
35544-EN-110616
VEGABAR 12 11
1
2
+
+
-
-
ecting to power supply
5 Conn
tion:
Informa
The metal parts of the instrument (antenna, transmitter, concentric tube, etc.) are conductive connected with the inner and outer ground terminal on the housing. This connection exists either directly metallic or with instruments with external electronics via the screen of the special connection cable. You can nd specications to the potential connections within the instrument in chapter "Technical data".

5.2 Wiring plan

Fig. 2: VEGABAR 12 - wiring
1 Voltage supply and signal output 2 PC adjustment interface
plan

5.3 Switch on phase

After connecting VEGABAR 12 to power supply or after a voltage recurrence, the instrument carries out a self-check:
l Internal check of the electronics l 4 20 mA output jumps to the failure signal 3.6 mA
Then VEGABAR 12 delivers a current of 4 20 mA to the cable. The value corresponds to the actual level as well as to settings already carried out, e.g. the factory setting.
35544-EN-110616
12 VEG
ABAR 12
6 Set
up
6 Set
up

6.1 Setup steps without adjustment

After mounting and electrical connection, VEGABAR 12 is ready for operation.
VEGABAR 12 delivers a current of 4 20 mA according to the actual process pressure. Further settings are not necessary.

6.2 Setup steps with PC operation

The procedure is described in the separate operating instructions "PC
operation for VEGABAR 12".
35544-EN-110616
VEGABAR 12 13
aintenance and fault rectication
7 M
Rea
ction when malfunc-
tions
occur
Causes of faults
Fault recticat
24 hour service hotline
ion
7 Maint
enance and fault rectication

7.1 Maintenance

If the instrument is used properly, no special maintenance is required in normal operation.
7.2 Fault rectication
The operator of the system is responsible for taking suitable measures to rectify faults.
VEGABAR 12 oers maximum reliability. Nevertheless, faults can occur during operation. These may be caused by the following, e.g.:
l Sensor l Process l Voltage supply l Signal processing
The rst measure to be taken is to check the output signal. In many cases, the causes can be determined this way and the faults rectied.
Should these measures not be successful, please call in urgent cases the VEGA service hotline under the phone no. +49 1805 858550.
The hotline is available to you 7 days a week round-the-clock. Since we oer this service world-wide, the support is only available in the English language. The service is free of charge, only the standard telephone costs will be charged.
Checking the 4 20 mA sign
al
? 4 20 mA signal not stable
l no atmospheric pressure compensation
à Check pressure compensation
? No 4 20 mA signal
l Connection to voltage supply wrong
à Check connection according to chapter "Connection steps"
and if necessary, correct according to chapter "Wiring plan"
l No operating voltage
à Check cables for breaks; repair if necessary
l Operating voltage too low or load resistance too high
à Check, adapt if necessary
14 VEG
35544-EN-110616
ABAR 12
aintenance and fault rectication
7 M
Reaction after fault rec-
cation
ti
? Current
signal 3.6 mA
l electronics module or measuring cell defective
à Exchange the instrument or send it in for repair
Depending on the failure reason and measures taken, the steps
described in chapter "Set up" must be carried out again, if necessary.

7.3 Instrument repair

If a repair is necessary, please proceed as follows:
You can download a return form (23 KB) from our homepage at
vega.com
under: "Downloads - Forms and certicates - Repair form".
By doing this you help us carry out the repair quickly and without having to call back for needed information.
l Print and ll out one form per instrument l Clean the instrument and pack it damage-proof
l Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
l Please ask the agency serving you for the address of your return
shipment. You can nd the respective agency on our website www.vega.com
under: "Company - VEGA worldwide"
www.
35544-EN-110616
VEGABAR 12 15
8 Dismoun
ting
8 Dismou
nting

8.1 Dismounting steps

Warning:
Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel, high temperatures, corrosive or toxic products etc.
Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order.

8.2 Disposal

The instrument consists of materials which can be recycled by specialised recycling companies. We use recyclable materials and have designed the electronics to be easily separable.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/EG and the respective national laws. Pass the instrument directly on to a specialised recycling company and do not use the municipal collecting points. These may be used only for privately used products according to the WEEE directive.
Correct disposal avoids negative eects on humans and the environ­ment and ensures recycling of useful raw materials.
Materials: see chapter "Technical data"
If you have no way to dispose of the old instrument properly, please
contact us concerning return and disposal.
16 VEG
35544-EN-110616
ABAR 12
9 Sup
plement
9 Supp
lement

9.1 Technical data

General data
Parameter, pressure Gauge pressure
Measuring principle Ceramic-capacitive, dry measuring cell
Communication interface Standard interface
Materials and weights
Materials, wetted parts
- Process tting 316L
- Diaphragm sapphire ceramic
- Measuring cell seal FKM
Materials, non-wetted parts
- Electronics housing 316L
- Connection terminals PA
- Ohmic contact Between ground terminal, housing and process
tting
Torque max. 20 Nm (14.75 lbf ft) Weight approx. 0.1 kg (0.22 lbs)
Output variable
Output signal 4 … 20 mA Fault signal 3,6 mA Max. output current 22 mA Step response time 160 ms (0 63 % )
®
(99.9 % oxide ceramic)
Input variable
Adjustment
Adjustment range of the zero/span adjustment relating to the nominal measuring range:
- zero -20 +20 %
- span 20 +120 %
Max. turn down 5 : 1
Nominal measuring ranges and overload resistance
The overload specications are only an overview and refer to the measuring cell. Limitations by material and process tting version are possible. The specications on the type label are applicable.
35544-EN-110616
VEGABAR 12 17
9 Suppl
ement
range Overload capacity, max.
Nominal
Gauge pressure
0 1 bar/0 100 kPa 35 bar/3500 kPa -1 bar/-100 kPa
0 … 2.5 bar/0 … 250 kPa 50 bar/5000 kPa -1 bar/-100 kPa
0 4 bar/0 400 kPa 65 bar/6500 kPa -1 bar/-100 kPa
Absolute pressure
0 … 1,6 bar/0 … 160 kPa 50 bar/5000 kPa -1 bar/-100 kPa
0 … 2.5 bar/0 … 250 kPa 50 bar/5000 kPa -1 bar/-100 kPa
0 4 bar/0 400 kPa 65 bar/6500 kPa -1 bar/-100 kPa
pressure
Overload capacity, min. pressure
Reference conditions and actuating variables (similar to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1
- Temperature +15 … +25 °C (+59 +77 °F)
- Relative humidity 45 75 %
- Air pressure 860 … 1060 mbar/86 106 kPa (12.5 15.4 psi)
Determination of characteristics Limit point adjustment according to IEC 61298-2
Characterstic curve Linear
Reference installation position upright, diaphragm points downward Inuence of the installation position < 0.2 mbar/20 Pa (0.003 psig)
Deviation determined according to the limit point method according to IEC 60770
1)
Deviation < 0.5 %
Inuence of the ambient temperature
Average temperature coecient of the zero
3)
signal
2)
< 0.15 %/10 K
Long-term stability (similar to DIN 16086, DINV 19259-1 and IEC 60770-1)
Long-term drift of the zero signal
4)
< 0.1 %/2 years
Ambient conditions
Ambient, storage and transport temperature -20 +85 °C (-4 +185 °F)
Process conditions
Product temperature -20 +85 °C (-4 +185 °F)
1)
Relating non-reproducibility.
2)
Relating to
3)
In the reference temperature 20 °C (68 °F).
4)
Relating
the nominal measuring range, incl. non-linearity, hysteresis and
to
the nominal measuring range.
compensated temperature range of 0 … +80 °C (+32 … +176 °F),
to
the nominal measuring range.
18 VEG
35544-EN-110616
ABAR 12
9 Sup
plement
ation resistance mechanical vibrations with 4 g and 5 100 Hz
Vibr
Electromechanical data
Screw terminals for cable cross-section up to 1.5 mm² (AWG 16)
Voltage supply
Operating voltage 8 30 V DC
Permissible residual ripple U
ss
< 1 V
Load see diagram
Electrical protective measures
Protection rating
6)
IP 20
Protection class III
Overvoltage category III
5)
5)
Tested according
6)
According to EN 60529/IEC 529.
to the guidelines of German Lloyd, GL directive 2.
35544-EN-110616
VEGABAR 12 19
63 mm (2.48")
57,5 mm (2.26")
10 mm
(0.39")
max. 6,5 mm
(0.26")
SW/SWS 24
(0.95")
G 1/8
ø 26 mm
(1.02")
9 Suppl
9.2 D
ement
imensions
VEGABAR 12
Fig. 3: Dimensions VEGABAR 12
20 VEG
35544-EN-110616
ABAR 12
9 Sup
plement
9.3 Indus
VEGA product lines are global protected by industrial property rights. Further information see http://www.vega.com. Only in U.S.A.: Further information see patent label at the sensor
housing.
VEGA Produktfamilien sind weltweit geschützt durch gewerbliche Schutzrechte.
Nähere Informationen unter http://www.vega.com.
Les lignes de produits VEGA sont globalement protégées par des droits de propriété intellectuelle.
Pour plus d'informations, on pourra se référer au site http://www.vega.
com.
VEGA lineas de productos están protegidas por los derechos en el campo de la propiedad industrial.
Para mayor información revise la pagina web http://www.vega.com.
Линии продукции фирмы ВЕГА защищаются по всему миру правами на интеллектуальную собственность.
Дальнейшую информацию смотрите на сайте http://www.vega.com.
VEGA 站<http://www.vega.com>
trial property rights

9.4 Trademark

All the brands as well as trade and company names used are property of their lawful proprietor/originator.
35544-EN-110616
VEGABAR 12 21
9 Suppl
ement
22 VEG
35544-EN-110616
ABAR 12
9 Sup
plement
35544-EN-110616
VEGABAR 12 23
VEGA Grieshaber KG
ISO 9001
Am Hohenstein 113 77761 Schiltach
Germany Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail: info@de.vega.com
www.vega.com
Printing date:
statements concerning scope of delivery, application,
All
practical use and operating conditions of the sensors and
processing systems correspond to the information avail-
able at the time of printing.
© VEGA Grieshaber KG, Schiltach/Germany 2011
Subject to change without prior notice 35544-EN-110616