Vanguard Accu-Trans User Manual

Page 1
OPERATING INSTRUCTIONS
for the
ACCU-TRANS™
TRANSFORMER-WINDING
RESISTANCE METER
Series II
Part Number VIC-8000-S2
1710 Grevillea Court Ontario, California 91761
TEL: (909) 923-9390 July 2001 FAX: (909) 923-9391 Rev 0
Page 2
Page 3
Accu-Trans Operating Procedures
SAFETY SUMMARY
The following safety precautions must be observed during all phases of test set-up, test hookups, testing, and test-lead disconnects.
Do Not Service or Test Alone
Do not perform test procedures or service unless another person is also present who is capable of rendering aid and resuscitation.
Avoid Contact with High Voltage
Because electric utility station environments contain high voltages and currents, there’s always the possibility of personal contact with an unexpected lethal voltages generated by magnetically induced and/or electrostatic leakage from nearby live circuitry. When test units are connected to de-energized ("dead") lines, regardless of how short these lines may be, always discharge the lines before attaching test lead. Because of the possibly deadly consequences of physical contact with such lines, engineers and technicians must always treat electrical equipment and hookups as though a lethal condition will eventually occur. Therefore, no matter how unlikely it may be, never assume anything about the safety of any test setup. Ensure the safety of personal by checking
first-hand to eliminate every possible hazard!
Avoid Contact With High-Voltage “Kicks”
Any transformer with many windings turns and an iron core supports induced magnetic fields, such that even low-level test currents produce a magnetic field that, if interrupted,
can produce a high-voltage spike capable of causing severe injury or death. Accordingly, its important for personnel safety that test leads be securely attached to the transformer terminals to prevent their being accidentally disconnected during testing. All personnel are warned to stand clear of the transformer during all testing. Test technicians should always ensure that all non-technical personnel (who may not understand the hazards) be required to stand clear of all transformers and equipment during all phases of transformer testing.
Do Not Modify Test Equipment
Because of the added risk of introducing additional or unknown hazards, do not install substitute parts or perform any unauthorized modification to the Accu-Trans. To ensure all designed safety features are maintained, it is recommended that all Accu-Trans repairs be performed at Vanguard Instruments Co. or by an authorized repair-service. Unauthorized equipment modifications can create safety hazards and will void the Accu-Trans warranty.
Follow Exact Operating Procedures
Any deviation from the operating procedures described in this manual may create one or more safety hazards, damage the unit or cause test result errors; Vanguard Instruments Co. assumes no liability for unsafe or improper use of the Accu-Trans.
i
Page 4
Page 5
Accu-Trans Operating Procedures
1
Table of Contents
SAFETY SUMMARY................................................................................................................. I
1.0 INTRODUCTION...................................................................................................................3
1.1 APPLICABILITY................................................................................................................3
1.2 SUPERSEDURE NOTICE..................................................................................................3
1.3 GENERAL DESCRIPTION................................................................................................3
1.4 FUNCTIONAL DES CRIPTION.........................................................................................3
1.5 FURNISHED ACCESSORIES ...........................................................................................3
2.0 ACCU-TRANS SPECIFICATIONS.......................................................................................4
3.0 CONTROLS and INDICATORS ............................................................................................5
4.0 PRETEST SETUP ...................................................................................................................7
4.1 OPERATING VOLTAGES.................................................................................................7
4.2 ACCU-TRANS LCD CONTRAST CONTROL.................................................................8
4.3 ACCU-TRANS PRINTER PAPER CONTROL.................................................................8
5.0 ACCU-TRANS PRINTER PAPER.........................................................................................8
6.0 OPERATING PROCEDURES................................................................................................8
6.1 GENERAL PROCEDURES................................................................................................8
6.2 STEP-BY-STEP PROCEDURES AT START-UP MENU..............................................10
6.2.1 START-UP MENU:.......................................................................................................10
6.2.2 TEST TRANSFORM ER PROCEDURE.......................................................................10
6.3 SETUP MENU:..................................................................................................................11
6.3.1 TRANSFORMER ID ENTRIES:..................................................................................12
6.3.2 REVIEW TEST RECORDS PROCEDURE .................................................................13
6.3.3 RESTORE RECORDING PROCEDURES ..................................................................14
6.3.4 PRINTING TEST-RECORD DIRECTORY.................................................................15
6.3.5. ERASE RECORDING PROCEDURE..........................................................................17
6.3.6 ENABLING COMPUTER INTERFACE.....................................................................18
6.3.7 SETTING DATE AND TIME.......................................................................................18
6.3.8 ENABLE PASSIVE TEST MODE ...............................................................................19
6.3.9 VOLTAGE REGULATOR TEST.................................................................................19
Page 6
Accu-Trans Operating Procedures
2
Table of Figure
Figure 1.0 Accu-Trans Controls and Indicators............................................................................5
Figure 2.0 90 to 130Vac Jumper Setting.......................................................................................7
Figure 3.0 210 to 240Vac Jumper Setting.....................................................................................7
Figure 4.0 Summary of Step-by-Step Accu-Trans Operat ion.......................................................9
Figure 5.0 Typical Test Report Printout......................................................................................12
Figure 6.0 Typical Record Directory Printout .............................................................................17
Page 7
Accu-Trans Operating Procedures
3
1.0 INTRODUCTION
1.1 Applicability
This manual applies to a second-generation Accu-Trans™ (Series II, henceforth, called the Accu-Trans), part number VIC-80000­S2 manufactured by the Vanguard Instruments Co. A prior manual supports a first generation Accu-Trans (i.e., a Series I configuration).
1.2 Supersedure Notice
This Operator’s manual is the basic issue for the Accu-Trans and does not supersede any previously published manual.
1.3 General Description
Vanguard Instruments Company produces the Accu-Trans, a microprocessor-controlled winding resistance meter that measures low resistances of a transformer’s inductive windings, in a range from 1 micro-ohm to 200 ohms with very high accuracy. This instrument can also measure any low milliohm resistance (e.g., the contact resistance of a circuit breaker). This Accu-Trans is simple-to-use; it features keypad controls and an alpha/numeric display of the measured resistance, test current, and operator-entered test parameters and identifying data. The Accu-Trans uses a 48-volt dc power supply and a special current-regulator to quickly measure transformer-winding resistances. A typical single- phase 400MVA
transformer winding resistance can be measured in less than 3 minutes. To ensure
operator safety, the Accu-Trans automatically discharges the energy stored in the transformer at the end of each test. A built-in thermal printer prints test results on 2.5-inch-wide thermal paper. Resistance and test current display on a 4-line by 20­character, back lighted LCD. Up to 63 records (96 readings each) can be stored in the Accu-Trans FLASH EEPROM. The user can recall stored reports for printing. A RS-232C serial interface port is provided for
diagnostic testing and down loading test records. The Accu-Trans is field-portable, rugged; it features simple, easy operation by first-time users with little training.
1.4 Functional Description
The Accu-Trans’s operation is based on the electrical relationships described by Ohm’s law: R=V/I, where I is a known current and V is the dc voltage measured across the unknown resistance (typically, a circuit­breaker’s contacts). The value of the unknown resistance under test is the direct function of the measured voltage divided by the current and being calculated by the microprocessor. Resistance reading is then displayed on a back-lighted 4 line by 20 character LCD. The Accu-Trans’ 48V dc power supply applies up to 2 amperes test current to the device under test. Accu-Trans voltmeter test leads are run separately from the current­bearing test leads to the resistive load; thus, voltages are measured at the terminals of the resistance being measured, eliminating error from the I•R voltage drop in the test current cables. The Accu-Trans makes precise micro-ohm measurements possible without calculating compensation for test-current­lead resistance errors. A built-in discharge circuit automatically discharges the energy stored in the transformer at the end of each test.
1.5 Furnished Accessories
The Accu-Trans is shipped with two 45-foot test cables with “quick disconnect” type test plugs on the unit end and battery-type clamps at the test load end. One power cord, one ground cable, and a cable-carrying bag.
Page 8
Accu-Trans Operating Procedures
4
2.0 ACCU-TRANS SPECIFICATIONS
Table 1. ACCU-TRANS SPECIFICATIONS
TYPE Special-purpose test equipment, portable, micro-ohmmeter SIZE (inches) 16.8 wide by 12.6 high by 10.6 deep WEIGHT Less than 21 pounds RESISTANCE 1 micro-ohm to 200 ohms TEST CURRENT 2 amperes max DISPLAY LCD, back-lighted, 4-line by 20-character PRINTER 2.5-inch wide thermal printer ACCURACY ± 1 % reading, ± 1 count POWER 2 amps, 90-120 / 200-240 V ac (selectable), 50/60 Hz SERIAL INTERFACE RS-232C connector port ENVIRONMENT Operating: 0°C to 55°C; Storage: -40°C to 65°C WARRANTY One-year parts & labor (post warranty service is available)
NOTE: THE ABOVE SPECIFICATIONS ARE VALID AT NOMINAL OPERATING VOLTAGE AND AT A TEMPERATURE OF 25 °C (77°F)
?ACCU-TRANS SPECIFICATIONS MAY BE UPGRADED AND CHANGED WITHOUT PRIOR NOTICE.
Page 9
Accu-Trans Operating Procedures
5
3.0 CONTROLS AND INDICATORS
(See Figure 1 and refer to Table 1.) Accu-Trans controls and indicators are shown in Figure 1.0. A leader line with an index number points to each control and indicator, which is cross-referenced to a functional description in Table 2. This table describes the function of each item on the
control-panel. The purpose of the controls and indicators may seem obvious, but users should become familiar with them before using the Accu-Trans. Accidental misuse of the controls will usually cause no serious harm. Users should also be familiar with the Safety Summary found on the front page of this operator’s manual.
Figure 1.0 Accu-Trans Controls and Indicators
Page 10
Accu-Trans Operating Procedures
6
Table 1.0 Functional Descriptions of Accu-Trans Controls and Indicators
Fig. 1
Index Panel Markings Functional Descriptions
1
RS-232C
2 (Not titled; 16-key pad) Pushbutton operating controls; 10 key pad, plus
3 (Not titled; LCD) Liquid-Crystal Display, 4-line by 20 character;
4 120/240 2A, 50-60 Hz
Fuse: 250 Vac, 3A Slow Blow
5 GROUND Stud, 5/16-18 thread, with hand-turned wing
6 (Not titled; thermal printer) Built in thermal printer; prints test results in the
7
HIGH CURRENT
PRESENT
8 (Not titled; test lead jacks) Test connector jacks, female; for connecting
Computer interface port; 9-pin connector; female DB type.The data are set to 19,200 baud, 1 start bit, 2 stop bits, 8 data bits, and no parity bit.
PIN SIGNAL 2 Rx 3 Tx 5 Signal Gnd
START, STOP, ENTER, CLEAR, two paper­drive buttons; number keys 2 thru 9 have dual functions as letter entries (a la telephone dialer key pad).
back lighted and readable in sunlight; Displays menus, user selections, status readouts, and test results. Input power connector with third-wire safety ground, ON (1)/OFF (0) rocker toggle switch; has built-in fuse protection.
nut; Safety Ground; This must be connected to station ground before connecting cables to transformer and begin testing.
field on 2.5-inch-wide paper. NOTE: For best high-contrast print quality, it is recommended that only VIC thermal paper be used. LED indicator light, red; When lighted, this indicator warns operators that there is a possibility that lethal high voltage will exist if any test lead is disconnected while current is running through an inductive load. Failure to
heed this warning can result in shock and/or fatal injury to personnel.
heavy gauge current leads and voltage-sensing test leads. Plug-in voltage plugs are not polarity sensitive, but must be plugged into the jacks adjacent to their companion current jacks.
Page 11
Accu-Trans Operating Procedures
7
4.0 PRETEST SETUP
4.1 Operating Vo ltages
The Accu-Trans operating voltages are selectable between 90-130Vac, 50/60Hz or 200-240, 50/60Hz. Voltage selection is set
Table 2.0 Voltage Selection
VOLTAGE SELECTION TERMINAL BLOCK JUMPERS
90-130Vac Brown to Blue & Yellow to Green
200-240Vac Blue to Yellow
by the placement of jumpers on the power
terminal block as listed in Table 2.0 below
and Figures 2.0 and 3.0.
120Vac Jumper Installation
Figure 2.0 90 to 130Vac Jumper Setting
Figure 3.0 210 to 240Vac Jumper Setting
220Vac Jumper Installation
Page 12
Accu-Trans Operating Procedures
8
4.2 Accu-Trans LCD Contrast Control
To darken the LCD display, press and hold the “Paper Contrast” switch for two
seconds; to lighten the LCD display, press and hold the “Paper Contrast” switch for
two seconds.
4.3 Accu-Trans Printer Paper Control
To advance the Accu-Trans printer paper, press and release the “Paper Contrast”
switch. To retract the Accu-Trans printer paper, press and release the “Paper Contrast” switch.
5.0 ACCU-TRANS PRINTER PAPER
The Accu-Trans built-in thermal printer uses
2.5-inch wide thermal paper for printing test results. To maintain the highest quality test­result printing and to avoid paper jamming, the use of paper supplied by our factory is highly recommended. Additional paper can be ordered from either of the two sources listed below:
Vanguard Instruments Co, Inc.
1710 Grevillea Court
Ontario, CA 91761 Tel: 909-923-9390
Fax: 909-923-9391
Part Number: TP-3 Paper
OR
BG Instrument Co.
13607 E. Trent Avenue
Spokane, WA 99216
Tel: 509-893-9881
Fax: 509-893-9803
Part Number: TP-3 paper
6.0 OPERATING PROCEDURES
Before using the Accu-Trans for measuring any resistance, operators should review the
Summary of Step-by-Step Operation of the Accu-Trans found in Figure 4.0. Accu-
Trans operations are simple, requiring the selection of choices from the display menus and responding to the displayed prompts. However, first-time operators should review
Figure 1.0 to become familiar with the total
Accu-Trans operation and the logical
branching for various test options. (More
experienced operators may use this figure as
a handy help and reference guide.
WARNING
Do not touch or disconnect any test lead that is connected to a transformer terminal while high current is being conducted during any test. Failure to
heed this warning can result in lethal electrical shock to personnel and/or damage to equipment.
6.1 General Procedures
a) Ground Accu-Trans to substation ground
(item 5 in Figure 1.0).
b) Plug Accu-Trans power cable into a
power outlet.
c) Insert current-cable plugs and voltage-
sensing cable plugs into respective control-
panel jacks (index 8 of Figure 1.0). Ensure
that voltage plugs are plugged into jacks that
are adjacent to companion current jacks.
d) Attach test-cable clamps to transformer
terminals of winding that is to be measured
(or if load is not an inductance, simply
attach test leads to opposite terminals of the
resistive load).
e) Turn on Accu-Trans power (item 4 in
Figure 1) by pressing “I” on rocker switch.
f) Observe that after configuration data is
displayed briefly, then the start-up menu
displays (with selection options for “1. RUN
TEST,” “2. SETUP”).
Page 13
Accu-Trans Operating Procedures
9
Figure 4.0 Summary of Step-by-Step Accu-Trans Operation
Page 14
Accu-Trans Operating Procedures
10
TEST IN PROGRESS
MEASURED RESISTANCE
1. TEST
XFMR
04/15/01
6.2 Step-by-Step Procedures at Start­Up Menu
6.2.1 START-UP Menu:
Observe that the Start-Up menu displays as shown below:
2. SETUP 09:28:03
a) Press key #1 to start test. Press #2 to select other options (i.e. enter record­identification data, recall test results, etc.)
6.2.2 Test Transformer Procedure
The following procedure shows the steps to get a resistance reading from a device under test. a) To start a test, press key #1 (TEST XFMR) from main menu. When the test mode is selected, a “WARNING!” display is shown below:
This display doesn’t prompt any action; but
simply reminds the user that a test is in
progress. The time for this display depends
on size of the winding’s inductance and the
test mode selected.
The Accu-Trans determines when the
resistance reading is stable and shows
resistance value on LCD as follows:
“STOP” TO ACCEPT
120.0 MICRO-OHM
The Accu-Trans continues to apply the test
voltage on the transformer’s winding and
updates the resistance value on the display.
d) The user presses STOP key to retain this
data and go to the next step. (If the displayed
data is not O.K., press CLEAR to return to
the Start-Up Menu.)
**********WARNING!********** DANGEROUS FLASH-OVER MAY OCCUR IF CABLES
ARE DISCONNECTED!
This warning reminds operators that the next sequence of test steps will run current through the load, which, if inductive, will cause lethal voltages to be generated if the cables are disconnected before the current is ramped down to zero.
b) Press the ENTER key and proceed to the next step. The start test display appears:
c) Press the START key to Start test. The test-in progress display appears below:
PLEASE WAIT………
MEASURED RESISTANCE FINAL RESISTANCE
120.0 MICRO-OHM
e) Press ENTER key and go to the next
step. The print-option menu is now
displayed.
PRINT TEST RESULTS
1. YES
2. NO
f) Press key #1 to print test results. Typical
Accu-Trans test report is shown in figure 5.
g) Select key #2 to bypass printing results.
* XFMR CHARGING*
Page 15
Accu-Trans Operating Procedures
11
RUN ANOTHER TEST?
1. ENTER
XFMR DATA
1. COMPUTER CONTROL
h) When either option is selected, the next window displays with the prompt “KEEP THIS READING? Go to the next step.
KEEP THIS READING
1.YES
2. NO
i) If the KEEP THIS READING prompt is answered with a YES, the following display appears. j) Go to step l. If NO is selected, the next display is skipped and the second display following appears.
TEST SAVED
m) At the “RUN ANOTHER TEST?”
Menu, the operator may choose to run
another test, by selecting “YES”. Return to
step 6.2.2.a to run test.
n) If NO is selected (i.e., don’t run another
test) the program returns to the Start-Up.
Menu (section 6.2.1).
6.3 SETUP Menu:
This menu is displayed by again selecting 2
(SETUP) on the Start-UP MENU
2. REVIEW RECORD
3. RESTORE RECORD
4. NEXT PAGE
The TEST SAVED display advises that the resistance measurement has been saved.
k) Press ENTER and go to the next step. “RUN ANOTHER TEST?” message will appears.
l) Select NO from “KEEP THIS RESULTS” will take the user to the “RUN ANOTHER TEST?” menu.
1. YES
2. NO
a) Select key 1 to 4. When key #4 (NEXT
PAGE) is selected, More choices for setup
menu are shown (below).
2. SET TIME
3. ENABLE TEST MODE
4. VOLTAGE REG. TEST
Page 16
Accu-Trans Operating Procedures
12
COMPANY:
Figure 5.0 Typical Test Report Printout
6.3.1 Transformer ID Entries:
The operator can enter test data using the following procedures.
OPERATOR’S NOTE
Test Identification data are entered with the keypad (item 2 of Figure 1.0). The keypad has 10 keys that are alpha­numerically marked, like pushbuttons on a telephone. Keys are pressed for each character position (marked by the cursor) in the identification area; thus: If, for example, we consider the key marked 2/A B C, the first press of the key selects the number “2,” pressing the key a second time selects the letter “A,” pressing the key a third time selects the letter “B, and pressing the key a fourth time selects the letter “C.” Further pressing of the key returns to the number selection; the cycle repeats
as the key continues to be pressed. The characters selected are entered at the position of the cursor. Pressing the up key (∧ ) advances the cursor by one
space; pressing the down key (∨ ) moves the cursor back one space. If a character is erroneously entered, select the new key to get the correct entry.
The first Identification (ID) display is the
COMPANY screen, which is shown below:
a) Enter the characters of the utility
company’s name. When this is done, press
Page 17
Accu-Trans Operating Procedures
13
STATION:
CIRCUIT:
MANUFACTURER:
SERIAL NUMBER:
KVA RATING:
ENTER key and go to the next step. The STATION screen will then be displayed.
b) Enter characters for naming the station where the transformer to be tested is located. When all the characters are selected or identifying the station, press ENTER key and go to the next step. The CIRCUIT screen displays.
c) Enter all characters needed to identify the circuit name, then press ENTER key and go to the next step. The manufacturer screen displays.
go to the next step. The SERIAL NUMBER
screen displays.
f) Enter the alpha-numeric characters
needed to identify the serial number, then
press ENTER key and go to the next step.
The KVA RATING screen displays.
g) Enter the numbers to identify the KVA
rating for the transformer being tested, then
press ENTER key and go to the next step.
The OPERTOR screen displays.
d) Enter all characters needed to identify the manufacturer of the transformer being tested, then press ENTER key and go to the next step. The MODEL screen displays.
MODEL:
e) Enter the characters of the transformer’s model number, then press ENTER key and
OPERATOR:
h) Enter the characters needed to identify
the operator who is performing the tests.
This completes entries for the transformer
identification data. Press ENTER key to
enter the selected data: the program returns
to the START-UP MENU.
6.3.2 Review Test Records Procedure
This procedure describes steps to review a
test record residing in Accu-Trans
working memory. The user can view the
record on the LCD display or from a thermal
printout.
Page 18
Accu-Trans Operating Procedures
14
1. ENTER
XFMR ID
4. NEXT PAGE
REVIEW RECORD
2 TESTS
TEST NUMBER: 1
press the down () key to roll down, through
NOTE
To review a test record stored in Flash EEPROM, the user must first restore test record from Flash EEPROM to working memory (see paragraph 6.3.3).
These procedural steps begin by selecting SETUP on the Start-Up Menu, which displays the Set-Up Menu, shown below:
2. REVIEW RECORD
3. RESTORE RECORD
a) In the Set-Up Menu (see above), press 2 to select the REVIEW RECORD procedure. It starts with the display shown below. Go to the next step.
the saved test records.
The next menu displays the number of test.
in the record, time and date of the test.
There are 2 tests in the test record below.
06/12/ 01 12:21:34
f) Press the UP (∧ ) key to review the
resistance measured in the first test, which
show in the following display.
ACTIVE TEST
10.88 OHM AT 0.824 Amps
1. SCROLL TEST RECORD
2. PRINT TEST RECORD
b) Press 1 (SCROLL TEST RECORD) to view all the test records in memory (see menu above).
c) Press 2 (PRINT TEST RECORDS) to output test record to the thermal printer. See Fig. 5.0 for typical test record printout.
d) Pressing the 1 key (SCROLL TEST RECORD) produces the RECORD “ID” INFO display. This menu displays the information related to the test.
RECORD “ID” INFO:
g) Press the UP (∧ ) key again to display the
second test on the record as shown below:
TEST NUMBER: 2 ACTIVE TEST
122.8 MILLI-OHMs AT 2.000 Amps
h) When the operator has reviewed the test
measurement for the selected record in the
display above, press STOP to return to the
Start-Up Menu. This ends the steps for
reviewing saved test records.
6.3.3 Restore Recording Procedures
This procedure allows the user to restore a
test record from the Accu-Trans Flash
EEPROM to working memory. The user can
then review the test record using the
REVIEW RECORD command (6.3.2).
e) To scroll through the test records in memory, press the UP (∧ ) key to roll up or
The Restore Recording Procedure steps
begin at the Set-Up Menu, shown next:
Page 19
Accu-Trans Operating Procedures
15
1. RESTORE RECORD
RESTORE RECORD
RESTORE RECORD
RECORD RESTORED!
1. ENTER
XFMR ID
REVIEW RECORD
1. ENTER XFMR ID
2. REVIEW RECORD
3. RESTORE RECORD
4. NEXT PAGE
a) To begin the Restore Recording steps, press key # 3 (RESTORE RECORDING), which shows a menu for three main procedures, as displayed below. Go to next step.
2. DIRECTORY
3. ERASE RECORDS
4. NEXT PAGE
b) Press key # 1 (RESTORE RECORD) to begin the procedure. The next display appears. Go to the next step.
1. ENTER RECORD NUMBER
2. SCROLL TO SELECT
c) Press # 1 (ENTER RECORD NUMBER) to produce the Restore Record display, as shown below. Go to next step.
The above display confirms to the operator
that the selected record has been restored.
e) When the restoration is confirmed, press
ENTER key to select the viewing option as
shown below.
1. SCROLL TEST RECORD
2. PRINT TEST RECORD
f) If the user wants to scroll through the
test record, press key #1 and continue from
step 6.3.2.b.
i) Press key # 2 will produce a test record
print out on the thermal printer. See figure
5.0 for a typical test record printout.
6.3.4 Printing Test-Record Directory
Printing the test record directory begins with
the Set-Up Menu shown below (the same
display from which the Review Records
steps began):
NUMBER:
The above display prompts the operator to enter the record number.
d) When the record number is displayed, press ENTER and go to the next step. The next display will appear.
2. PREVIEW RECORD
3. RESTORE RECORD
4. NEXT PAGE
Page 20
Accu-Trans Operating Procedures
16
1. RESTORE RECORD
PRINT DIRECTORY
a) Press key # 3 to select the Restore Recording step sequence, and go to the next step. The following menu displays:
2. DIRECTORY
3. ERASE RECORD
b) Press key # 2 (DIRECTORY) and go to the next step. The menu below displays:
1. FULL DIRECTORY
2. SHORT DIRECTORY
c) Press key # 1 to print the entire directory
of test records; Press key # 2 to print the
short directory.
A short directory lists the last 20 test records
stored in the Accu-Trans memory.
This ends the print directory sequence.
Page 21
Accu-Trans Operating Procedures
17
1. RESTORE RECORD
ERASE RECORD
Figure 6.0 Typical Record Directory Printout
6.3.5.Erase Recording Procedure
The Erase Recording sequence begins with the Set-Up Menu, shown below. Go to the next step.
1. ENTER XFMR ID
2. PREVIEW RECORD
3. RESTORE RECORD
4. NEXT PAGE
a) Press # 3 (RESTORE RECORD) and go to the next step. The display below appears.
2. DIRECTORY
3. ERASE RECORD
b) Press # 3 (ERASE RECORD) and go to next display.
ERASE RECORD
1. ERASE SINGLE REC.
2. ERASE ALL RECORDS
c) To erase a single a record, press key # 1;
to erase all records press key # 2. If a single
recording is selected go to the next step; The
prompt below displays. To erase all
recordings go to step f.
NUMBER:
Page 22
Accu-Trans Operating Procedures
18
RECORD NUMBER
ERASE ALL RECORDS!
RECORD ERASED!
COMPUTER
ITF MODE
1. ENTER
XFMR ID
d) Enter the number of the record that is to be erased and press ENTER key. Go to the next step. The confirmation display appears as shown below:
ERASED
e) When the erased record number is confirmed, press ENTER key to return to the start-Up Menu ending the sequence for erasing a single record.
f) When key # 2 (ERASE ALL RECORDS) is pressed in the Erase Record display (back in step c, the following prompt displays. Go to the next step.
6.3.6 Enabling Computer Interface
The Computer Interface Mode allows the
user to retrieve test records stored in the
Acccu-Trans memory via the RS-232C port
An IBM PC program is provided with each
Accu-Trans allowing the user using his PC
to retrieve these test records.
a) This procedure begins in the Set-Up
Menu, shown below. Go to the next step:
2. REVIEW RECORD
3. RESTORE RECORD
4. NEXT PAGE
b) Pressing key #4 (NEXT PAGE) and the
menu below appears. Go to the next step.
1. COMPUTER CONTROL
2. SET TIME
3. ENABLE TEST MODES
ARE YOU SURE? “ENTER” TO CONTINUE
g) Press ENTER key to erase all test records. If ERASE ALL RECORDS was mistakenly selected, press “STOP” to abort the erase process and return to the Start-Up menu. When ENTER key is pressed, the following confirmation message is displayed. Press ENTER key to return to the Start-Up Menue:
This ends the Erase Record(s) sequence.
c) Press key # 1 (COMPUTER
CONTROL) and go to the next step. The
display below appears:
j) Pressing the STOP key terminates the
computer control, at which time the Start-Up
Menu displays, ending this step sequence.
6.3.7 Setting Date and Time
The date and time set begins at the Set-Up
Menu, shown below:
1. ENTER XFMR ID
2. REVIEW RECORD
3. RESTORE RECORD
4. NEXT PAGE
Page 23
Accu-Trans Operating Procedures
19
a) Begin the date and time set by pressing
key # 4 (NEXT PAGE) and going to the next step. The menu below appears. Go to the next step.
1. COMPUTER CONTROL
2. SET TIME
3. ENABLE TEST MODE
b) Press key # 2 (SET TIME) and go to the next step. The display below appears:
ENTER
MM-DD-YY HH:MM:SS
c) Enter the date and time, which appear in the dashed-line areas, then press ENTER key to set the calendar and clock and return to the Start-UP Menu. This ends the Set Time step sequence.
6.3.8 Enable Passive Test Mode
Operators have a choice between two test modes: ACTIVE or PASSIVE. Normally, the ACTIVE test mode is best suited for quickly testing transformer resistances. In this mode, the Accu-Trans automatically ramps the test current up to its final test value (when the magnetic field is saturated) in the shortest possible time. However, in a very large inductive load, the ACTIVE test mode reading may appear to be unstable. In this situation, the user may want to try the PASSIVE test mode test. In this mode, the Accu-Trans will take longer to provide a reading, however, the reading is more stable.
Note
The Accu-Trans always defaults to the ACTIVE test mode when power is first applied to the unit. The user needs to enable the PASSIVE test mode before using it.
a) This procedure begins in the Set-Up
Menu, shown below. Go to the next step:
1. ENTER XFMR ID
2. REVIEW RECORD
3. RESTORE RECORD
4. NEXT PAGE
b) Pressing key #4 (NEXT PAGE) and the
menu below appears:
1. COMPUTER CONTROL
2. SET TIME
3. ENABLE TEST MODES
c) Press key #3 to activate PASSIVE Test
mode.
Note
Once the PASSIVE test mode is selected, the Accu-Trans will display the test mode selection menu (shown below) in step 6.2.2.a after the user presses key #1 (TEST XFMR).
1. ACTIVE TEST
2. PASSIVE TEST
6.3.9 Voltage Regulator Test
Note
This test verifies that a regulating switch in a primary-voltage tap regulator changes contacts through its selection range with no break in the circuit (i.e., it checks the tap switch’s make-before-break). This test is important because, in actual use, any breaks in current in a regulating tap
Page 24
Accu-Trans Operating Procedures
20
1. TEST
XFMR
08/14/01
1. ENTER
XFMR ID
1. COMPUTER CONTROL
ARE DISCONNECTED!
VOLTAGE REG. TEST
switch generates large reactive voltage spikes that are hazardous and exceed the switch’s voltage ratings, thus causing irreparable damage. By connecting the Accu-Trans continuity test input across the switching input and running the regulator switch through all the contacts in it’s range, any breaks in continuity caused by faulty tap switching are detected.
a) To run the Voltage-Regulator test, perform the following step-by-step sequence, starting with the STARTUP MENU (see below):
2. SETUP 12:21:01
b) Press key #2 (SETUP) to produce the SETUP MENU (see below):
e) Press ENTER key for next display (see
below):
f) Press START key to run the voltage
regulator tap-switch test. The test begins as
the magnetic field in the transformer ramps
up to a stable plateau with the advisory that
the transformer is charging (the red High-
voltage warning indicator lights) When the
transformer current stabilizes, the following
prompt displays (see below):
*****WARNING!*****
DANGEROUS FLASH-OVER
MAY OCCUR IF CABLES
VOLTAGE REG. TEST
“START” TO RUN TEST
or
“STOP” TO ABORT
2. REVIEW RECORD
3. RESTORE RECORD
4. NEXT PAGE
c) Press key #4 (NEXT PAGE) to produce the next menu of test options (see next display):
2. SET TIME
3. ENABLE TEST MODES
4. VOLTAGE REG. TEST
d) Connect Accu-Trans input to the regulating tap-switch input. Press key #4 (VOLTAGE REG. TEST) to start the regulator test (high-voltage advisory notice displays—see following):
CURRENT RAMP: xx %
*XFMR CHARGING*
No operator action required; when
transformer current ramps nears a flat
plateau, the run tap switch prompt displays
(see below):
VOLTAGE REG. TEST
RUN TAP CHANGER NOW
“STOP” TO EXIT TEST
XFMR CHARGING
If no transient is detected as taps are
switching, the display above will remain.
Press “STOP” to end the voltage regulator
test.
Page 25
Accu-Trans Operating Procedures
21
If a switch break is detected, an advisory displays (shown below). Press STOP to end the voltage-regulator test.
TRANSIENT DETECTED!
.
Page 26
Accu-Trans Operating Procedures
APPENDIX A
ACCU-TRANS Troubleshooting Guide
Item Symptom Possible Problem Solution
1 Reading is incorrect. 1. Poor connection at the
Clips.
2. Broken sensing lead.
2 No Test Current.
(Current % read zero) and resistance reading=0
3 Have test current but
resistance reading =
0.
1. No test current going through DUT.
2. Drive circuit not working.
1. Broken sense leads.
2. Reverse sense leads.
1.Check connection to make sure sensing and current teeth are touching the device under test (DUT).
2. Inspect sensing cables.
1. Check Connection to DUT.
2. Run Calibration Test.
1. Inspect sense leads.
NOTE: DUT=Device Under Test
Page 27
Page 28
1710 Grevillea Court, Ontario, CA 91761, USA
Phone: 909-923-9390 Fax: 909-923-9391
Website: http//www.vanguard-instruments.com
ACCU-TRANS 09/09/01: OAK
Loading...