Calibrating the Resistor Transducer Adapter
The Vanguard Resistor Transducer Adapter must be calibrated each time a different resistor transducer is connected to the adapter. Follow the steps below to calibrate the adapter:
a. Connect the resistor transducer to the adapter. b. Turn on the circuit breaker timer. The following “RUNNING” screen will be displayed on
the resistor transducer adapter’s LCD screen:
RUNNING
Res Enc Rev: 1.00
Press either the [] or [] key to access the Main Menu.
c. The Main Menu will be displayed:
CALIBRATE ENCODER
ADJUST CONTRAST
<DIAGNOSTIC>
Press the [] key to select the CALIBRATE ENCODER option:
<CALIBRATE ENCODER>
ADJUST CONTRAST
DIAGNOSTIC
Press the [ENTER] key.
d. The following screen will be displayed:
SET ENCODER TO POSITION 1
“ENTER” WHEN DONE
Move the resistor transducer to one end of its travel and then press the [ENTER] key.
e. The following screen will be displayed:
SET ENCODER TO POSITION 2
“ENTER” WHEN DONE
Move the resistor transducer to the opposite end of its travel and then press the [ENTER] key.
1
REV 2 CT-7500 S2 USER’S MANUAL
f. The following screen will be displayed:
/ TO SET DISTANCE
5.00 IN 12.70 CM
“ENTER” WHEN DONE
Press the [] or [] key to set the value to the known traveled distance (distance between the endpoints of the transducer) and then press the [ENTER] key. The following screen will be displayed momentarily:
CALIBRATION SAVED
The following screen will then be displayed:
RUNNING
Res Enc Rev: 1.00
2