Vaisala PTB330 User Manual

ELECTRICAL CONNECTIONS

2008-08-04

VAISALA BAROCAP® DIGITAL BAROMETER PTB330

 

 

Operating voltage

 

10 ... 35 VDC

 

 

Typical power consumption at +20 °C

 

 

 

 

(Uin 24 VDC, one pressure sensor)

 

 

 

 

RS-232

 

25 mA

 

 

RS-485

 

40 mA

 

 

Uout

 

25 mA

 

 

Iout

 

40 mA

 

 

display and backlight

 

+ 20 mA

 

 

Analog outputs

 

 

 

 

 

current output

 

0 ... 20 mA, 4 ... 20 mA

 

 

voltage output

 

0 ... 1 V, 0 ... 5 V, 0 ... 10 V

 

 

External loads

 

 

 

 

 

current ouputs

 

RL < 500 ohm

 

 

0 ... 1V output

 

RL > 2 kohm

 

 

0 ... 5V and 0 ... 10V outputs

RL > 10 kohm

 

 

Max wire size

 

1.5 mm2 (AWG 16)

 

Electrical Connections

 

 

 

 

 

 

 

 

 

WARNING

Make sure that you connect only de-energized wires.

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 1 Signal and Power

Supply Screw Terminals

Basic Wiring Using Screw Terminals

1.Open the four cover screws and open the barometer cover.

2.Insert the power supply wires and signal wires through the cable bushing in the bottom of the barometer. Ground the screen of the electrical cable.

3.Connect the RS-232 user port cables to terminals TxD, GND and RxD on the screw terminal block. See Figure 1 on the left.

4.If you have the optional analog output module, connect the cables to the terminals Ch+ and Ch– on the analog output module. The terminals are shown in Figure 2.

5.When wiring the optional AC power supply, RS-485, or relay module, refer to the PTB330 User’s Guide.

6.In case you need to ground the barometer housing, the grounding connector is found inside the housing. Make sure that different groundings are made to the same potential. Otherwise harmful ground currents may be generated.

7.Connect the power supply wires to (+) and (-) terminals marked

POWER 10 ... 36V=.

8.Close the cover and tighten the cover screws.

Figure 2 Analog output module

www.vaisala.com

Ref. M210958EN-A

Vaisala PTB330 User Manual

ELECTRICAL CONNECTIONS

2008-08-04

Wiring with M12 (8-Pin) Connector

Figure 3 Optional 8-Pin Connector

Table 1 Pin Assignments of the 8-Pin Connector

PIN/Terminal

Wire

Serial Signal

 

Analog Signal

 

 

RS-232 (EIA-232)

RS-485 (EIA-485)

 

1

White

Data out TX

A –

-

2

Brown

(serial GND)

(serial GND)

Signal GND (for both channels)

3 (optional)

Green

External power control

(serial GND)

Signal GND (for both channels)

4

Yellow

-

-

Analog output

5

Grey

Supply –

Supply –

Supply –

6

Pink

Supply +

Supply +

Supply +

7

Blue

Data in RX

B –

-

8

Shield/Red

Cable shield

Cable shield

Cable shield

Wiring with Optional D-9 Connector

Figure 3 Optional D-9 Connector

Table 2 Pin Assignments of the D-9 Connector

PIN/Terminal

Wire

Serial Signal

 

Analog Signal +

 

 

 

 

RS232

 

 

RS-232 (EIA-232)

RS-485 (EIA-485)

 

1

Red

 

 

 

2

White

TX

 

TX

3

Black

RX

 

RX

4 (optional)

Yellow

External power control

External power control

External power control

5

Brown

Ground

 

Ground

6

Green

 

LO

Aout

7

Blue

Ground for supply

Ground for supply

GND supply

 

 

voltage

voltage

 

8

Grey

 

HI

AGND

9

Orange

Supply voltage (10...30

Supply voltage (10...30

Supply

 

 

VDC)

VDC)

 

www.vaisala.com

Ref. M210958EN-A