Vaisala OMT364 Technical data

OMT364 Oxygen Transmitter for Hazardous Areas
www.vaisala.com
The Vaisala SPECTRACAP® Oxygen Transmitter OMT364 for hazardous areas.
Features/Benefi ts
Minimizes need for sample
conditioning equipment
In-situ probe or sampling cell
Tolerates aggressive chemicals
Tolerates excessive amounts of
moisture even in liquid form
Low maintenance
Calibration interval 12 months
Software and keypad user
interfaces and an LCD display
Example applications:
fermentation, composting process monitoring, inerting processes and fl are gas recovery
The Vaisala SPECTR ACAP® Oxygen Transmitter OMT364 is designed for oxygen monitoring in moist and aggressive process gases.
Laser based measurement
The OMT364 incorporates a compact tunable diode laser (TDL) gas spectrometer built in the measurement probe. This optical measurement technology is well known for unmatched stability and robustness. The patented SPECTRACAP TDL technology in a compact probe suitable for  eld use.
®
sensor offers uniquely
FM Approved OMT364 for Hazardous Areas
The non-incendive OMT364 transmitter is permitted for use in Class I, Division 2 of the NEC area classi cation according to FM Approvals SM . The probe of the
®
500
instrument can be installed in areas that are classi ed as Class I, Division 1 or 2.
Direct installation
In many applications, the OMT364 can be  ange-mounted directly into a process. No sampling or sample conditioning equipment is needed. This feature provides real time measurement data without sampling or sample switching delays.
Sampling cell installation
An optional sampling cell is available for processes with high temperatures, elevated pressures or extremely dif cult mechanical conditions. Due to the robustness of the SPECTRACAP sensitivity to gas  ow and pressure variations, the sampling system is simple and it can be installed near the sampling point.
®
sensor and its low
Technical data
Performance
Measurement range 0 ... 100 % O Accuracy (including noise, linearity and repeatability) ±0.2 % O Temperature range over T range ±2 % of reading Stability ±1 % of reading/yr Zero drift ±0.1 % O
2/
Response time of measurement 3 s Diffusion limited response in still air T63/T
without fi lters 10 s / 20 s with stainless steel mesh fi lter 10 s / 25 s with stainless steel mesh fi lter and
PTFE fi lters 30 s / 70 s Pressure dependence without pressure compensation
0.8 ... 1.2 bar -2 % of reading
1.2 ... 1.4 bar -5 % of reading Accuracy of pressure compensation ±0.25 % of reading Background gas effects for CO
and H2O uncompensated
2
< 1% of reading for <6 vol -% CO
< 1% of reading for gas dew point < 30 ˚C
Accuracy of background gas compensation
0 ... 50 vol-% CO
2
±-0.5 % of reading
0 ... 300 g/m3 H2O (Td=80 ˚C) ±1 % of reading
Operating environment
Operating temperature range for
probe (in-line installation) -20 ... +80 ˚C electronics (housing) (Kalrez® seals) transmitter (ambient gas
measurement) -20 ... +60 ˚C Storage temperature range -55 ... +80 ˚C Operating pressure range 0.8 ... 1.4 bar Maximum pressure range for probe up to 10 bar Compliance
IEC(EN)-61326 Electrical equipment for measurement,
control and laboratory use. EMC requirements. Laser safety Class 1 laser product
-20 ... +60 ˚C
Classifi cation
OMT364 CLASSIFICATIONS FOR THE TRANSMITTER
2
USA and Canada FM, Non-incendive for Class I,
2
OMT364 CLASSIFICATIONS FOR THE PROBE
yr
USA and Canada FM, Suitable for Class I, Division 1,
Division 2, Groups A, B, C, and D.
T
a
Groups A, B, C, and D.
T
90
a
Inputs and outputs
Power supply
input range 11 ... 36 VDC
Power consumption
maximum 6W typical 3W
Analog output 0/4 ... 20 mA sourcing
maximum load 500 Ω Serial output (2-wire, non-isolated) RS-485 Alarm/control relay 30 VAC/60 VDC
2
Mechanics
Housing material G-AlSI10 Mg (DIN 1725) Probe material AISI 316 Housing classifi cation IP66 Weight, transmitter and probe 2.2 kg Weight, sample cell 2.6 kg Mounting fl ance diameter 97 mm Cable bushing Cable gland M20 x 1.5 Stainless steel mesh fi lter openings ø 310 μm Wetted materials
®
AISI 316, Kalrez
SiN, MgF
, PTFE (optional),
2
Options and accessories
Hydrophobic PTFE fi lter, pore size 8 μm with
Kalrez® O-rings 217056 1/2” NPT conduit fi tting 217917 Sample cell with wall-mounting bracket
gas fi ttings for O.D. 6 mm tube
volume 260 m
T63 response time with 1 l/min
sample fl ow and mesh fi lter, 18 s
with Kalrez
3
®
O-rings 216619
= -20 ˚C to +60 ˚C
= -20 ˚C to +80 ˚C
For more information, visit www.vaisala.com or contact us at sales@vaisala.com
Ref. B210823EN-A ©Vaisala 2009
This material is subject to copyright protection, with all copyrights retained by Vaisala and its individual partners. All rights reserved. Any logos and/or product names are trademarks of Vaisala or its individual partners. The reproduction, transfer, distribution or storage of information contained in this brochure in any form without the prior written consent of Vaisala is strictly prohibited. All specifi cations — technical included — are subject to change without notice.