Vaisala MMT330 Technical data

MMT330 Moisture and Temperature Transmitter Series for Oil
The display shows measurement trends, real time data and history.
www.vaisala.com
The MMT330 transmitter family offers a range of solutions for demanding moisture in oil measurements.
Features/Benefits
Continuous on-line
measurement of moisture in oil
Ball valve installation – no need
to shut down the process
Incorporates Vaisala HUMICAP®
Sensor – more than 30 years of field performance
Ten years of experience in
measuring moisture in oil
Excellent long-term stability
Easy to calibrate and maintain
in the field - Compatible with Vaisala HUMICAP® Hand-Held Moisture for Oil Meter MM70
NIST traceable calibration
(certificate included)
Analog outputs, WLAN/LAN
The Vaisala HUMICAP® Moisture and Temperature Transmitter Series for Oil MMT330 enables fast and reliable detection of moisture in oil. The MMT330 can be used in on-line moisture monitoring and as a control device, allowing separators and oil driers to be started only when needed.
Proper monitoring saves both oil and the environment. With the MMT330 it is easy and economical to monitor the changes of moisture in oil.
Reliable Vaisala HUMICAP® technology
The MMT330 incorporates the latest generation of the Vaisala HUMICAP Sensor, which is the result of ten years of eld experience. It was developed for demanding moisture measurement in liquid hydrocarbons.
The sensor's excellent chemical tolerance provides accurate and reliable measurement over a wide measurement range.
For diverse applications and demanding conditions
Because of the variety of probes, the transmitter can be used in lubrication systems, hydraulic systems, and transformers.
Indicates the margin to water saturation
The MMT330 measures moisture in
®
oil in terms of the water activity (aw) and temperature (T). Water activity indicates directly whether there is a risk of free water formation. The measurement is also independent of oil type and age.
Water content as ppm conversion
In addition to water activity, the MMT330 can output ppm, the average mass concentration of water in oil. Vaisala has this conversion readily availabe for mineral transformer oil.
For other oils, the oil specic conversion coefcients can be programmed to the transmitter if the water solubility of the oil is known.
Graphical measurement trend and historical display
The MMT330 can be ordered with a large numerical and graphical display with a multilingual menu. It allows the user to monitor operational data, measurement trends and up to 1-year measurement history. The optional data logger with real-time clock makes it possible to generate over four years of measured history, and
zoom in on any desired time or time frame.
The display alarm allows tracking of any measured parameter, with a freely congurable low and high limit.
Data collection and (wireless) transfer to PC
The recorded measurement data can be viewed on the display or transferred to a PC with Microsoft Windows can also be connected to a network with an optional (W)LAN interface, which enables a (wireless) Ethernet connection.
®
software. The transmitter
Versatile outputs and easy installation
The MMT330 provides up to three analog outputs. Galvanic isolation of supply power and analog outputs is also available. For serial interface the USB connection, RS232 and RS485
The Vaisala HUMICAP® Hand-Held Moisture for Oil Meter MM70 is designed for field checking MMT330 transmitters.
can be used. In addition, alarm relay option is available.
The MMT330 has several options for transmitter mounting. Transmitters are delivered pre-congured with all settings installation ready.
The MMT332 probe is installed
97 (3.82)
65.5 (2.58) 7 (0.28)
35 (1.38)
Ø 12 (0.47)
9 (0.35)
M6
45°
Ø 36 (1.42)
16 (0.63)
77 (3.03)
33 (1.30)
NPT1/2"
ISO 3/8" or
Ø 12 (0.47)
Hex 22m (0.87)
226/448 (8.90/17.6)
188/410 (7.40/16.1)
35 (1.37)
14 (0.55)
Ø 13.5 (0.53)
35 - 157/379(1.37 - 6.2/14.9)
Ø 12 (0.47)
R1/2" ISO7/1
using a flange. It is for high pressure applications.
Installation options
MMT332 For High Pressure Installations
Pressure range 0 ... 250 bar /
0 ... 3625 psia
Probe diameter 12 mm / 0.5 inch
Installation
Flange 36 mm / 1.4 inch
Temperature
Measurement range -40 ... +180 °C
(-40 ... 356 °F)
Dimensions
Dimensions in mm (inches)
The MMT337 probe, with optional Swagelok connector, is ideal for tight spaces with a thread connection. The small probe is designed for integrating into confined spaces with small diameter lines.
The MMT338 is ideal for installations in pressurized processes where the probe needs to be removed while the process is running. The probe depth is adjustable.
Installation options
MMT337 with Small Sized Probe
Pressure range 0 ... 10 bar / 0 ... 145 psia
Probe diameter 12 mm / 0.5 inch
Installation
Fitting Body R 3/8" ISO
Fitting Body 1/2" ISO
Fitting Body NPT 1/2"
Temperature
Measurement range -40 ... +180 °C
(-40 ... 356 °F)
Installation options
MMT338 with Probe for Pipeline Installations
Pressure range with ball valve
0 ... 40 bar / 0 ... 580 psia
up to 120 °C (248 °F) and 40 bar
Adjustable length 35 ... 157/379 mm /
1.37 ... 6.2 /14.9 inch
Installation
Fitting Body R1/2" ISO
Fitting Body NPT 1/2"
Ball Valve Set BALLVALVE-1
Sampling Cell DMT242SC2
Temperature
Measurement range -40 ... +180 °C
(-40 ... 356 °F)
Dimensions
Dimensions in mm (inches)
Dimensions
Dimensions in mm (inches)
Technical data
Measured values
WATER ACTIVITY
Measurement range a
w
0 ... 1
Accuracy (including nonlinearity, hysteresis and repeatability)
0 ... 0.9 ±0.02
0.9 ... 1.0 ±0.03 Response time (90 %) at +20 °C in still oil (with stainless steel filter) 10 min. Sensor HUMICAP
Performance
TEMPERATURE
Measurement range
MMT332 -40 ... +180 °C (-40 ... +356 °F) MMT337 -40 ... +180 °C (-40 ... +356 °F) MMT338 -40 ... +180 °C (-40 ... +356 °F)
Accuracy at +20 °C (+68 °F) ± 0.2 °C (0.36 °F)
Operating environment
Operating temperature
for probes same as measurement ranges
for transmitter body -40 ... +60 °C (-40 ... +140 °F) with display 0 ... +60 °C (+32 ... +140 °F) Pressure range for probes See probe specifications
Complies with EMC standard EN61326-1, Electrical equipment for measurement, control and laboratory use - EMC requirements; Industrial environment.
Max wire size 0.5 mm2 (AWG 20) stranded wires recommended Digital outputs RS-232, RS-485 (optional) Service connection RS-232, USB Relay outputs 0.5 A, 250 VAC, SPDT, Potential Free (optional) Ethernet interface (optional)
Supported standards 10/100Base-T Connector RJ45 Protocols Telnet
®
WLAN interface (optional)
Supported standards 802.11b Antenna connector type RP-SMA Protocols Telnet Security WEP 64/128,WPA
Authentication / Encryption
Open / no encryption Open / WEP WPA Pre shared key / TKIP WPA Pre shared key / CCMP (a.k.a. WPA2)
Optional data logger with real-time clock
Logged parameters max. three with trend/min/max values Logging interval 10 sec (fixed) Max. logging period 4 years 5 months Logged points 13,7 million points per parameter Battery lifetime min. 5 years
Display LCD with backlight, graphic
trend display of any parameter
Display menu languages English, Chinese, Spanish, German,
French, Japanese, Russian, Swedish,
Finnish
Inputs and outputs
Operating voltage 10 ... 35 VDC, 24 VAC
with optional power supply module 100 ... 240 VAC 50/60 Hz Power consumption @ 20 °C (U
RS-232 max 25 mA
2 x 0...1V / 0...5V / 0...10V max 25 mA
U
out
2 x 0...20 mA max 60 mA
I
out
display and backlight + 20 mA Analog outputs (2 standard, 3rd optional)
current output 0 ... 20 mA, 4 ... 20 mA
voltage output 0 ... 1 V, 0 ... 5 V, 0 ... 10 V Accuracy of analog outputs at 20 °C ± 0.05 % full scale Temperature dependence of the
analog outputs ± 0.005 %/°C full scale External loads
current ouputs R
0 ... 1V output R
0 ... 5V and 0 ... 10V outputs R
24VDC)
in
< 500 ohm
L
> 2 kohm
L
> 10 kohm
L
Mechanics
Cable bushing M20x1.5 for cable diameter 8 ... 11mm/0.31 ... 0.43” Conduit fitting 1/2”NPT Interface cable connector (optional) M12 series 8 pin (male)
option 1 with plug (female) with 5 m / 16.4 ft black cable
option 2 with plug (female) with screw terminals USB-RJ45 Serial Connection Cable (incl. Mi70 Link software) Probe cable diameter 5.5 mm Probe cable lengths 2 m, 5 m or 10 m Housing material G-AlSi 10 Mg (DIN 1725) Housing classification IP 65 (NEMA 4X)
219685
Mounting options
169 (6.65)
183 (7.20)
116 (4.57)
Ø 7 (0.28)
96 (3.78)
77 (3.03)
71 (2.80)
Dimensions
Dimensions in mm (inches)
Mounting with Wall
Mounting with DIN Rail
Mounting Kit
Pole Installation with Installation Kit for Pole or Pipeline
HUMICAP® is a registered trademark of Vaisala.
Installation Kit
Mounting Rain Shield
with Installation Kit
For more information, visit www.vaisala.com or contact us at sales@vaisala.com
TYPE A PPROVED PRODUCT CERTI FICATE NO.: A-11440
Ref. B210953EN-A ©Vaisala 2010
This material is subject to copyright protection, with all copyrights retained by Vaisala and its individual partners. All rights reserved. Any logos and/or product names are trademarks of Vaisala or its individual partners. The reproduction, transfer, distribution or storage of information contained in this brochure in any form without the prior written consent of Vaisala is strictly prohibited. All specifications — technical included — are subject to change without notice.
Loading...