Vaisala HMM100 Technical data

www.vaisala.com
HMM100 Humidity Module for Environmental Chambers
The Vaisala HUMICAP® Humidity Module HMM100.
Features/Benefits
Full temperature compensation
over the operating temperature range of -70˚C ... +180˚C
High temperature tolerance,
also suitable for heat­sterilization
Excellent measurement
accuracy with Vaisala HUMICAP
Durable
Easy field calibration by
trimmers
Maintenance-free
Easy to install
Applications: test chambers,
incubators
®
180R sensor
The HMM100
The Vaisala HUMICAP® Humidity Module HMM100 is an open frame module for integration into environmental chambers. The modules provide a single analog output channel for relative humidity (RH) or dewpoint (T
Two probes are available, one made of plastics, the other of stainless steel. Several cable lengths up to 3 meters are available. Both the probes have the Vaisala HUMICAP 180R sensor which ensures excellent measurement accuracy.
Robust and reliable
The HMM100 probe works in freezing conditions (-70 °C) and also in temperatures up to +180 °C. The
HMM100 is easy to install and the probe can be freely placed in a test chamber as the speed of airow does not affect the measurement.
Maintenance-free
Compared to psychrometers, the
).
d
®
HMM100 is practically maintenance­free. There is no wick that needs changing and there is no need for a water tank or water pump. Thus, environmental stress screening can be done reliably.
Accessories
The accessories available are a component board mounting bracket with a lid, probe clamp, USB-cable for service use, a module housing and a probe mounting ange.
Technical data
54
2.12
30.7
1.21
12
0.47
See Table
46.5
1.83
36.5
1.44
55
2.17
49
1.93
3.5
0.14
AW 7 [0.28]
Performance
RELATIVE HUMIDITY
Measurement range 0 ... 100 %RH Accuracy (incl. non-linearity, hysteresis and repeatability)
temperature range -20 ... +40 °C
0 ... 90 %RH ±2 %RH 90 ... 100 %RH ±3 %RH
temperature range -40 ... -20 °C, +40 ... +180 °C
0 ... 90 %RH ±2.5 %RH 90 ... 100 %RH ±3.5 %RH
Factory calibration uncertainty (+20 °C) ±1.5 %RH
®
Humidity sensor Vaisala HUMICAP
DEW POINT TEMPERATURE
180R
Measurement range -20 ... +100 °C (-4 ... +212 °F)T Accuracy (incl. non-linearity, hysteresis and repeatability when dew point depression <20 °C ±2 °C T (Ambient temperature - dew point)
Operating environment
Operating temperature range
component board -5 ... +55 °C (+23 ... +131 °F) stainless steel and plastic probe -70 ... +180 °C (-94 ... +356 °F) porous PTFE filter stainless steel, sintered filter -70 ... +180 °C (-94 ... +356 °F) plastic grid, membrane filter -20 ... +80 °C (-4 ... +176 °F)
Electromagnetic compatibility Complies with EMC standard
EN61326-1, for use in light industrial environments
Mechanics
Service cable connector M8 series 4-pin (male) Probe diameter 12 mm Probe cable lengths 0.6/1.55/2.9 m Probe material
plastics PPS
stainless steel AISI316/PPS Probe mounting clamp AISI316 Mounting bracket material
lid ABS/PC
bottom plate Al Module housing material ABS/PC (cover)
Options and accessories
d
Humidity sensor HUMICAP180R Membrane filter 10159HM
d
Plastic grid filter 6221 Porous PTFE filter 219452SP Stainless steel sintered filter HM47280SP Mounting bracket with lid 225979 Module housing (IP65) 226060 Probe mounting flange 226061 Probe mounting clamp set (10 pcs) 226067 USB cable 226068
Dimensions
Input and outputs
Operating voltage
2-wire model 24 VDC 3-wire model 10 ... 35 VDC or 24 VAC
15 ... 35 VDC or 24 VAC when 0 ... 10 V
output is used Power consumption 6 mA Analog output types (1 output selectable)
2-wire model 4 ... 20 mA (loop-powered)
3-wire model 0 ... 20 mA, 0 ... 1 V/5 V/10 V Max. wire size 0.5 ... 1. 5 mm Service port M8 connector for USB cable
For more information, visit www.vaisala.com or contact us at sales@vaisala.com
2
(AWG)
Ref. B210813EN-A ©Vaisala 2010
This material is subject to copyright protection, with all copyrights retained by Vaisala and its individual partners. All rights reserved. Any logos and/or product names are trademarks of Vaisala or its individual partners. The reproduction, transfer, distribution or storage of information contained in this brochure in any form without the prior written consent of Vaisala is strictly prohibited. All specifications — technical included — are subject to change without notice.