QUICK REFERENCE GUIDE
Vaisala HUMICAP®
Humidity Module HMM100
Download
the complete User's Guide from:
www.vaisala.com/HMM100
© Vaisala 2009. All rights reserved.
*M211002EN*
GENERAL
The HMM100 series humidity measurement modules are open
frame modules intended for integration into environmental
chambers. The modules provide a single analog output channel
for relative humidity (RH) or dewpoint (T
Technical Data for more information.
NOTE All HMM100 series modules include
automatic temperature compensation across
the operating temperature range. No external
temperature compensation should be applied
to the output.
NOTE The measurement probe with cable is
attached to the component board at Vaisala.
Do not disconnect and reconnect the cable.
). See section
d
INSTALLATION
When selecting the mounting location for the probe, note the
following:
- The probe should be entirely in the measurement
environment, with sufficient distance to the outer walls of the
chamber or duct. Air should flow freely around the probe.
- The probe location should represent well the environmental or
process conditions, and it should be as clean as possible.
Attach the module securely using the mounting holes on
1.
the corners of the component board. Use the optional
mounting bracket if needed. Be sure to provide an earth
connection from the plated mounting hole. See Figure 1.
Refer to the User's Guide for Vaisala order codes of
accessories.
Install the probe in the location. Use the optional probe
2.
mounting flange or mounting clamp if needed. Insert
enough cable to the same space with the probe in order to
prevent heat conduction, and to allow the probe to be
moved when calibrating.
Connect the wires to the screw terminals:
3.
- 2-wire output: see Figure 2. Provide a 24 VDC supply
voltage.
- 3-wire output: see Figure 3. Provide the required
supply voltage according to the output type:
10 ... 35 VDC or 24 VAC for 0 ... 20 mA, 1 / 5 V
13 ... 35 VDC or 24 VAC for 0 ... 10 V
Verify the installation: power up the module and check
4.
the output reading.
Figure 1 Component Board (3-wire version)
Figure 2 Wiring for 2-wire Output
Figure 3 Wiring for 3-wire Output
When using the 3-wire output module, you may also connect a
fourth wire to the “Out -“ terminal. It is internally connected to
the “Vin -" terminal. This will help reduce electrical
interference, especially when using the 0 ... 1V output.
ADJUSTMENT USING TRIMMERS
HMM100 series modules have two trimmers mounted on the
component board. These trimmers allow for easy adjustment of
the offset and gain of the humidity measurement. The modules
are shipped with the trimmers centered.
Turning the trimmer counter-clockwise will decrease the offset
or gain, and turning clockwise will increase it. The adjustment
range is ± 5% for both offset and gain.
Offset correction requires a < 50 %RH reference. Offset and gain
correction requires additionally a > 50 %RH reference, with over
30 %RH difference between the two points.
NOTE The trimmers will only turn 135 degrees
each way, less than half a rotation. Do not
force the trimmer past the stopping point.
Figure 4 Trimmers for Offset and Gain
The procedure below provides an example adjustment of offset
and gain using the saturated salt calibrator HMK15. It is
assumed that the module is powered and you can measure the
output reading.
Note the current position of the trimmers.
1.
Insert the probe in the LiCl salt chamber of the humidity
2.
calibrator (11 %RH), and wait for 20 – 40 minutes for
the reading to stabilize.
If necessary, adjust the offset trimmer using a small
3.
Phillips-head screwdriver to achieve the output reading
that corresponds to 11 %RH.
For 4 ... 20 mA output, the reading should be 5.76 mA.
For 0 ... 5 V output, the reading should be 0.55 V.
Insert the probe in the NaCl salt chamber (75 %RH) and
4.
wait for the reading to stabilize.
If necessary, adjust the gain trimmer to achieve the
5.
output reading that corresponds to 75 %RH.
For 4 ... 20 mA output, the reading should be 16 mA.
For 0 ... 5 V output, the reading should be 3.75 V.
Optional: repeat steps 2 – 4 to check the adjustment result.
6.
If you want to undo the adjustment, return the trimmers to
their original position.
Visit our Internet pages at www.vaisala.com
TECHNICAL DATA
Property Description / Value
Measured parameters
Relative humidity
Dewpoint
Accuracy
-20 ... +40 °C
-40 ... -20 °C, +40 ... +180 °C
Sensor Vaisala HUMICAP® 180R
Outputs
2-wire
3-wire
Supply voltage
2-wire
3-wire (0 ... 20 mA, 1 / 5 V)
3-wire (0 ... 10 V)
Input current < 4 mA
Connections Screw terminals,
Probe types and cable lengths Plastic, 0.6 m cable
Sensor protection options Plastic grid filter
Operating temperature range
Component board
Probe (both types)
Plastic grid and membrane filter
PTFE and stainless steel filter
Ref. M211002EN-A
0 ... 100 %RH
-20 ... +100 °C
(-4 ... +212 °F) Td
± 2 %R
H (0 ... 90 %RH)
± 3 %RH (0 ... 100 %RH)
± 2.5 %RH (0 ... 90 %RH)
± 3.5 %RH (0 ... 100 %RH)
4 ... 20 mA (loop powered)
0 ... 20 mA, 1 / 5 / 10 V
24 VDC
10 ... 35 VDC or 24 VAC
13 ... 35 VDC or 24 VAC
0.5 ... 1.5 mm2
4-pin M8 connector for
service use
Plastic, 1.55 m cable
Plastic, 2.9 m cable
Stainless steel, 1.55 m cable
Stainless steel, 2.9 m cable
Membrane filter
Porous PTFE filter
Stainless steel sintered filter
-5 ...+55 °C (+23 ... +131 °F)
-70 ...+180 °C (-94...+356°F)
-20 ... +80 °C (-4 ... +176 °F)
-70 ...+180 °C (-94...+356°F)
Figure 5 Dimensions in mm