Vaisala DMP248 OPERATING MANUAL

DMP248 Dewpoint
Transmitter
OPERATING
MANUAL
M210492EN-A
JULY 2003
PUBLISHED BY Vaisala Oyj Phone (int.): (+358 9) 894 91
Visit our Internet pages at http://www.vaisala.com/
© Vaisala 2003 No part of this manual may be reproduced in any form or by any
means, electronic or mechanical (including photocopying), nor may its contents be communicated to a third party without prior written permission of the copyright holder.
The contents are subject to change without prior notice.
_________________________________________________________________________CONTENTS
Table of contents
CHAPTER 1 GENERAL INFORMATION.......................................................................................................1
SAFETY................................................................................................................................................................1
WARRANTY .........................................................................................................................................................1
CHAPTER 2 PRODUCT DESCRIPTION.........................................................................................................3
GENERAL CHARACTERISTICS...............................................................................................................................3
THE OPERATING PRINCIPLE OF THE DMP248.......................................................................................................4
Use in high pressure.......................................................................................................................................6
CHAPTER 3 INSTALLATION...........................................................................................................................7
SELECTING THE PLACE OF INSTALLATION............................................................................................................7
MOUNTING THE TRANSMITTER ............................................................................................................................7
Mounting; overview.......................................................................................................................................9
Mounting the probe directly to the process..................................................................................................10
Installing the probe through the ball valve assembly...................................................................................11
Mounting the probe with a quick connect....................................................................................................17
Sample cell...................................................................................................................................................18
Grounding....................................................................................................................................................20
Connections..................................................................................................................................................22
Connection to an AC supply........................................................................................................................23
CHAPTER 4 COMMISSIONING.....................................................................................................................25
SECURITY LOCK JUMPER....................................................................................................................................25
SELECTING THE ANALOGUE OUTPUTS................................................................................................................26
CONNECTING THE RS 232C SERIAL BUS............................................................................................................28
CHAPTER 5 COMMANDS...............................................................................................................................33
COMMANDS AND SECURITY LOCK JUMPERS.......................................................................................................33
LED COMMANDS...............................................................................................................................................34
DISPLAY/KEYPAD COMMANDS...........................................................................................................................35
CHAPTER 6 CALIBRATION AND ADJUSTMENT.....................................................................................55
HUMIDITY CALIBRATION AND ADJUSTMENT......................................................................................................55
Two-point calibration and adjustment adjustment procedure ......................................................................56
Using serial commands.............................................................................................................................................56
Using display/keypad commands..............................................................................................................................57
Using LED commands..............................................................................................................................................58
Humidity calibration table............................................................................................................................59
TEMPERATURE CALIBRATION ............................................................................................................................59
One point offset correction...........................................................................................................................59
Using serial commands.............................................................................................................................................59
Using display/keypad commands..............................................................................................................................60
Using LED commands..............................................................................................................................................60
Two-point temperature calibration and adjustement....................................................................................61
Using serial commands.............................................................................................................................................61
Using display/keypad commands..............................................................................................................................62
Using LED commands..............................................................................................................................................62
CALIBRATION OF THE ANALOGUE OUTPUTS.......................................................................................................63
Using serial commands................................................................................................................................63
Using display/keypad commands.................................................................................................................63
VAISALA __________________________________________________________________________ I
OPERATING MANUAL_______________________________________________________________
Using LED commands .................................................................................................................................64
CHAPTER 7 MAINTENANCE.........................................................................................................................67
REFERENCE MEASUREMENTS.............................................................................................................................67
SELF-DIAGNOSTICS............................................................................................................................................67
TEMPERATURE CHANNEL ADJUSTMENT WITH PT 100 SIMULATORS ...................................................................68
MEASUREMENT OF OUTPUT CURRENTS USING TEST POINTS ...............................................................................70
ADJUSTING THE CONTRAST OF THE DISPLAY......................................................................................................71
VAISALA SERVICE CENTERS ..............................................................................................................................71
CHAPTER 8 TECHNICAL DATA ...................................................................................................................72
MEASURED VARIABLES......................................................................................................................................72
OUTPUTS............................................................................................................................................................73
GENERAL...........................................................................................................................................................74
ELECTRONICS ....................................................................................................................................................75
SERIAL INTERFACE MODULES.............................................................................................................................75
ELECTROMAGNETIC COMPATIBILITY .................................................................................................................76
CHAPTER 9 OPTIONS......................................................................................................................................76
CHAPTER 10 SPARE PARTS AND ACCESSORIES ....................................................................................77
APPENDIX 1 SERIAL COMMANDS...............................................................................................................79
APPENDIX 2 POWER SUPPLY MODULE ..................................................................................................101
APPENDIX 3 INSTALLING AND USING THE RS 485/422 SERIAL PORT MODULE.........................105
APPENDIX 4 INSTALLING AND USING THE CURRENT LOOP MODULE........................................115
APPENDIX 5 ERROR MESSAGES................................................................................................................125
APPENDIX 6 WIRING DIAGRAM................................................................................................................129
APPENDIX 7 ALARM OUTPUT UNIT .........................................................................................................131
APPENDIX 8 PRESSURE CONVERSION CHART.....................................................................................137
PRESSURE CONVERSION CHART........................................................................................................................137
Multiplication factors .............................................................................................................................................137
II______________________________________________________________________M210492EN-A
CHAPTER 1_______________________________________________________ GENERAL INFORMATION
CHAPTER 1 GENERAL INFORMATION
Safety
Throughout the manual important instructions regarding the safety considerations are focused as follows.
WARNING
CAUTION
NOTE
Warranty
Warning denotes a hazard. It calls attention to a procedure, practice, condition or the like, which, if not correctly performed or adhered to, could result in injury to or death of personnel.
Caution denotes a hazard. It calls attention to a procedure, practice, condition or the like, which, if not correctly performed or adhered to, could result in damage to or destruction of part or all of the product.
Note highlights important information. It calls attention to an essential procedure, practice, condition or the like.
Vaisala issues a guarantee for the material and workmanship of this product under normal operating conditions for one (1) year from the date of delivery. Exceptional operating conditions, damage due to careless handling and misapplication will void the guarantee.
VAISALA __________________________________________________________________________1
CHAPTER 2_______________________________________________________ PRODUCT DESCRIPTION
CHAPTER 2 PRODUCT DESCRIPTION
General characteristics
The DMP248 transmitter is a microprocessor-based instrument for the measurement of dewpoint temperature in low humidities. The transmitter measures other quantities as well: relative humidity, temperature and ppm concentration (dry). When the dewpoint temperature is below 0 °C, the transmitter calculates the frostpoint instead of the dewpoint. The dewpoint output can be scaled freely, for example, dewpoint -40...+20 °C can be set to correspond to 0...1 V. The DMP248 transmitter has two analogue outputs and can be connected to a serial bus via the RS 232C interface or optionally through an RS 485/422 serial module or a current loop module.
The transmitter can be configured in many ways. It can have either a blank cover or a cover with a local display and keypad with which the user can operate the transmitter. The power supply voltage can be se­lected from three alternatives (24 VDC/VAC, 115 VAC, 230 VAC). Two analogue output signals are selected from the measured quantities; the signals can be scaled. The transmitter can be supplied with two, five or ten metre sensor head cable. The alarm output option enables two separate alarms that can be freely set by user.
Options Alarm output 2 relays 8A/230V SPCO relays Power supply 24 VDC (VAC) (standard), 115/230 VAC Serial interface RS 232C (standard), RS 485/422, current loop Display cover cover with or without local display & keypad Cable length 0.56, 2, 5 or 10 metres
VAISALA __________________________________________________________________________3
OPERATING MANUAL_______________________________________________________________
Pw
The operating principle of the DMP248
The DMP248 transmitter incorporates the DRYCAP® sensor which is optimized to be used in low humidities but has also an excellent tolerance against condensation. The DRYCAP® sensor uses an operating principle based on changes in capacitance as its thin polymer film absorbs water molecules together with a combined temperature measurement with a Pt 100 resistive temperature sensor. The capacitance of the thin polymer film has a direct response proportional to RH, but combined with the temperature signal the response of the DRYCAP® sensor is rather proportional to Pw (water
vapour pressure) or to the dewpoint. While frostpoints (dewpoints below 0 °C) in principle can be
determined by using traditional RH transmitters, it is very difficult considering the required accuracy at the dry end calibration. As relative humidity levels approach zero the accuracy rapidly decreases and the offset soon becomes the largest source of errors when monitoring frostpoints. Therefore, the focus is to minimize the offset (error at 0% RH) when monitoring the process gas.
For example to monitor a process with a frostpoint of -40 C and a temperature of +20 C translates to a relative humidity of 0.55%. An offset error of -0.2 %RH, which is well within specifications for a normal RH-transmitter, brings the measured RH down to 0.35%. This would bring the calculated frostpoint down to -44 C. Thus seemingly minor offset errors caused by drift or bad calibration translate into unacceptable frostpoint errors when the RH is low.
To solve the accuracy problem the DMP248 transmitter utilizes a patented method that automatically adjusts the dry end measurement in frostpoints. The offset calibration algorithm incorporated into the DMP248 transmitter uses the fact that the capacitance of a thin film polymer sensor is proportional to RH as seen in formula 1.
RH RH Gain
= +
out
0
*
Pws T
( )
(2-1)
where: RH0=output in completely dry state
Pw=water vapor pressure Pws(T)=temperature dependent water vapor saturation pressure
4______________________________________________________________________M210492EN-A
CHAPTER 2_______________________________________________________ PRODUCT DESCRIPTION
0.4
The function Pws(T) is well known from literature. Thus, it is possible to determine the offset (RH0) if measurements are made at
two or (preferably) more temperatures assuming a constant Pw during the process.
0.3
20°C
0.2
0.1
0
RHout(%)
-0.1
-0.2
-0.3 0 0.005 0.01 0.015 0.02 0.025 0.03 0.035 0.04 0.045
FIGURE 2-1 Device output during hypothetical offset calibration
30°C
1/Pws(T)
cycle.
For example, for a transmitter with a -0.2%RH offset error is made 11 measurements at temperatures 20...30 °C. The frostpoint is assumed to stay constant at -40 °C and that the 'Gain' is nominal 100%.
As the temperature increases the RH drops. Since the response is linear, a straight line is fitted through the data points. The line
intercepts the y-axis at RH0 as shown in FIGURE 2-1. Now the RH0 is known and it can be subtracted from the indicated value 0.35% RH to get the correct value of 0.55% RH.
FIGURE 2-2 shows the DRYCAP
sensor as mounted on a DMP248 probehead. Through a combination of the polymer sensor and Pt 100 sensor, the DRYCAP will accurately measure the water vapor pressure used in determining low dewpoints. During auto-calibration the Pt 100 element is used to first heat and then measure the temperature of the sensor while cooling back to ambient temperature. A complete cycle of auto-calibration takes 60...70 seconds. When the auto-calibration cycle is active, the transmitter locks the output values to those measured prior to auto-calibration.
VAISALA __________________________________________________________________________5
OPERATING MANUAL_______________________________________________________________
FIGURE 2-2 The DRYCAP
Note that the auto-calibration takes place only if the DMP248 is used in ambient humidities below 10 %RH (dewpoint below -12 °C at 20 ° C) and at ambient temperature 0...+80 °C.
Use in high pressure
If the process pressure differs from the normal ambient pressure, the value has to be entered in the transmitter memory to ensure the best possible measurement accuracy. The pressure setting is used for pressure compensation of the DMP248 transmitter. Note that although dewpoint is a pressure dependent parameter, this setting cannot be used for calculating dewpoints in different pressures. The probe should be installed to a place with pressure equal to that of the process in order to ensure the most reliable measurement. For converting pressure units, see Appendix 8.
NOTE
The probe can be installed in the process through the ball valve assembly provided that the process pressure is less than 10 bars. This way, the process does not have to be shut down when installing or removing the probe. However, if the probe is not removed from the process as such (e.g. the process is shut down first), the process pressure can be max. 20 bars.
sensor mounted on a DMP248 probe.
6______________________________________________________________________M210492EN-A
CHAPTER 3_______________________________________________________________INSTALLATION
CHAPTER 3 INSTALLATION
Selecting the place of installation
Select a place which gives a true picture of the environment or process; also select a place that is as clean as possible. Air should circulate freely around the sensor.
It is recommended that the sensor head is installed directly in the process through the ball valve assembly. When the ball valve assembly is used, the chamber or the duct does not have to be emptied or shut down for installation or removal of the probe. Install the sensor head transversely against the direction of the process flow.
If the probe head has to be installed aside of the process gas flow or the process is very hot or particularly dirty, the probe can be installed in a “leak-through” position. In this installation, the probe is mounted behind the ball valve assembly and if necessary, a cooling coil and/or a filter can be mounted in between. The flow passes through the sensor head and leaks out through a vent hole in the fitting body enabling a reasonable response time. In hot and dirty processes, a sample system can also be used.
Mounting the transmitter
In FIGURE 3-1 and FIGURE 3-2, you can see the dimensions of the DMP248 transmitter:
VAISALA __________________________________________________________________________7
OPERATING MANUAL _______________________________________________________________
PROBE UP
PROBE PUSHED DOWN
cable length 2, 5 or 10 m
ø5.5
178
31
ø13.5
adjustment
range120 mm
29
R1/2 ISO 7/1
non leaking screw (A) (factory setting) or leak screw (B) (included in the package)
149
clasp nut
fitting body
NOTE
104
CL
ENT
120
ø6.5
65
133
145
FIGURE 3-1 Dimensions of the DMP248 electronics housing (in
mm).
1. Always mount the transmitter housing with the cable bushings
pointing downwards to ensure IP65 (NEMA4) rating
2. Make sure that the connection cable has the right thickness
(7...10 mm) and that the cable bushing is carefully tightened.
3. Pay always special attention to closing the transmitter cover
carefully and remember to tighten all four screws.
FIGURE 3-2 Probe dimensions (in mm).
8 _____________________________________________________________________M210492EN-A
CHAPTER 3_______________________________________________________________INSTALLATION
NOTE
Take care not to damage the pipe of the probe. If the pipe is damaged, the probe head is less tight and it will not go through the clasp nut.
Mounting; overview
sealing with:
1. LOCTITE® No 542 + activ. No 7649 (t=-55...+150 °C)
2. MEGA-PIPE EXTRA No 7188 (t=-55...+170 °C)
3. PTFE tape (t=-60...+210 °C) NOTE: the tape does not lock the parts together. Therefore, use two fork spanners (hex 24 and 27 mm) for tightening and opening the clasp nut of the probe
FIGURE 3-3 Sealing and thread cutting for the fitting body.
The fitting body can be installed e.g. on standard pipe fittings (G 1/2 ISO 228/1) or on a thread in the process wall. If the wall thickness is less than 10.5 mm, it is recommended to use a welded sleeve (see
FIGURE 3-3). Note that the minimum recommended distance of the
fitting body and probe head is 40 mm (see FIGURE 3-3).
fitting body hex = 24mm tapered thread R1/2 ISO 7/1
parallel thread G1/2 ISO 228/1 (BS 2779, JIS B0202)
ø19mm drilling
>10.5mm
>40mm
Process or pipe wall
Adjust the probe to a suitable distance according to the type of installation, and tighten the clasp nut first manually. Then, mark the fitting body and the clasp nut and tighten the nut a further 50...60° with a fork spanner (see FIGURE 3-4).
probe
a pen
fitting body
FIGURE 3-4 Tightening the clasp nut.
clasp nut
60°
max.
VAISALA __________________________________________________________________________9
OPERATING MANUAL_______________________________________________________________
NOTE
Be careful not to tighten the clasp nut more than 60° as this may result in difficulties when trying to open it.
The probe is delivered with non-leaking screw A mounted. For by­pass measurements, this screw is removed and replaced with leaking screw B (included) and an O-ring is placed on the groove of the sintered filter prior to installation. Make sure to tighten the screw carefully.
Screw B has a small (0.08 mm) laser-made hole in the middle; the gas or air to be measured passes through the sintered filter and by the sensor, and leaks out through the screw.
Mounting the probe directly to the process
Select a point, which gives a true picture of the process. The transmitter can be installed directly in the process wall, especially if the pressure of the process is 1 bar (atmospheric processes).
process wall
recommended adjustment range
25 mm
FIGURE 3-5 Installing the probe in an atmospheric process.
25...135 mm
Leave at least 25 mm (1 ") of probe head free to enable a faster response time in unpressurized processes
welded sleeve (G1/2, Ø 40 mm)
non leaking screw (screw A, factory setting)
If the probe is installed in process pipes where the water is likely to collect at the measurement point, take care to install the sensor head so that it will not be immersed in water.
When the probe is installed directly on the process wall or pipe, note that a closing valve may be needed on both sides of the installed probe so that the sensor head can be removed from the process for calibration and maintenance.
10_____________________________________________________________________M210492EN-A
CHAPTER 3_______________________________________________________________INSTALLATION
If the sensor head is installed in a pressurized chamber, always make sure that the pressure of the chamber is equalized with the ambient pressure prior to removing the probe.
capped nut DIN 917-M22x1.5
when the probe is pulled out for maintenance, cap the hole with a capped nut; this way, the process can be open although the probe is not in place
sealing
welded sleeve
(G1/2)
process pipe
FIGURE 3-6 Installing the sensor head directly on the process pipe.
Non leaking screw (screw A)
closing valve (ball valve)
Installing the probe through the ball valve assembly
The best way to install the sensor head is through the ball valve assembly. Use a 1/2” ball valve assembly with a ball hole of 14 mm or more. In this kind of installation, it is not necessary to empty or shut down the process for installing or removing the sensor head. If the sensor head is installed in a process pipe, please note that the nominal size of the pipe must be at least 1 inch. See FIGURE 3-7 -
FIGURE 3-8 for detailed instructions.
VAISALA _________________________________________________________________________11
OPERATING MANUAL_______________________________________________________________
probe
non-leaking screw A
handle
>30 mm
ball of the ball valve (hole diameter at least 14 mm)
NOTE
process pipe / chamber
FIGURE 3-7 Installing the sensor head through the DMP248BVS
ball valve assembly.
The probe can be installed in the process through the ball valve assembly provided that the process pressure is less than 10 bars. This way, the process does not have to be shut down when installing or removing the probe. However, if the process is shut down before removing the probe, the process pressure can be max. 20 bars.
See FIGURE 3-8- FIGURE 3-11 for detailed description of installation through the ball valve assembly. This installation is possible provided that the process pressure is less than 10 bars. Note also that if the sensor head is installed in a process pipe, the nominal size of the pipe must be at least 1 inch.
12_____________________________________________________________________M210492EN-A
CHAPTER 3_______________________________________________________________INSTALLATION
STEP 1: mount the probe with the ball valve assembly closed;
tighten the clasp nut manually.
bushing R1/2 cone/G1/2(40 bar) e.g. Camozzi 2520-1/2-1/2
(the bushing serves for moving the probe (sinter) to such a distance from the ball valve that the valve can be closed)
clasp nut
fitting body R1/2 cone, sealed
ball valve 1/2" (40 bar) e.g. Atlas Copco:BAL-1A 15 (G1/2)
FIGURE 3-8 Installing the probe through the ball valve assembly;
>30 mm
bushing R1/2 cone sealed
nipple R1/2 cone sealed
step 1.
VAISALA _________________________________________________________________________13
OPERATING MANUAL_______________________________________________________________
STEP 2: open the ball valve assembly.
manual press tool
120mm
probe pipe
148 mm
ø5.5
handle
ball of the ball valve
adjustment range
61
> 14
=
ø14
marking groove
leak screw (B) (hex. 1.5 mm)
(40)
29
15
R1/2 ISO 7/1
ø13.5
fitting ferrule
clasp nut (hex 27 mm)
fitting body (hex. 24 mm)
O-ring
DRYCAP® sensor
filter
FIGURE 3-9 Installing the probe through the ball valve assembly;
step 2 (measures in mm).
14_____________________________________________________________________M210492EN-A
CHAPTER 3_______________________________________________________________INSTALLATION
STEP 3: push the probe head through the ball valve assembly into the process. If the pressure is high, use a manual press tool. Note that the sensor head must be pushed so deep that the filter is completely inside the process flow.
MANUAL PRESS TOOL
VALVE OPEN
VALVE CLOSED
FILTER
FIGURE 3-10 Installing the probe through the ball valve assembly;
step 3.
For by-pass measurements, the probe is mounted behind the ball valve assembly and non-leaking screw A on the fitting body is replaced with leaking screw B and O-ring is placed on the groove of the sintered filter. Screw B has a small (0.08 mm) laser-made hole in the middle; the gas or air to be measured passes through the sintered filter and by the sensor, and leaks out through the screw. The process pressure reduces in the hole of the screw B. This installation is recommended if the process flow rate is >20 m/s and there is over-pressure in the process.
VAISALA _________________________________________________________________________15
OPERATING MANUAL _______________________________________________________________
gas escape channel: use this position for by-pass measurements
leak screw (B)
sintered filter
O-ring
DRYCAP® sensor
process pipe or chamber
NOTE keep the marking groove in sight when using leak screw (B)
FIGURE 3-11 Installing the sensor head for by-pass measurements.
When pushing the probe head through the ball valve assembly, be careful not to break the sintered filter. Open and close the ball valve assembly with the marking groove always in sight. In by-pass measurements, the clasp nut is tightened manually prior to pressing the probe through the valve. When the probe has been pressed through and the valve is open, the nut is tightened 50...60° with a fork spanner (hexagon 27 mm).
16 ____________________________________________________________________M210492EN-A
CHAPTER 3_______________________________________________________________INSTALLATION
Mounting the probe with a quick connect
The probe can also be installed with a quick-connect, which acts as a closing valve between the process and the probe. The installation can be easily done even in small ducts with standard parts, and the probe is easily removed when necessary. It is necessary for the probe to be installed in the leak-through position for a reasonable response time.
FIGURE 3-12 illustrates an example of using a quick connect with the
DMP248 probe. The chamber can be made of stainless steel AISI 316.
keep the marking groove in sight when using leak screw
leak screw
leak screw
15 mm
30 mm
chamber made of hexagonal bar (hex=27 mm)
sealing
G1/2
50 mm
G1/8 or G1/4
bayonet socket
fitting part
of the quick-
connect
process pipe
FIGURE 3-12 Installing the probe with a quick-connect.
VAISALA _________________________________________________________________________17
OPERATING MANUAL_______________________________________________________________
Sample cell
It may be necessary to use the sample cell if the process (e.g. a pipe) is too small for the DMP248 sensor head. Furthermore, if the process is very hot (>80 °C) or particularly dirty, the probe is installed in a sample cell behind a cooling coil and/or filter. In this case, the ambient temperature must be at least 10 °C warmer than the process dewpoint in order to avoid condensation in the sample tubing.
6
40
28
80
G1/4
ISO
228/1
PUSH PROBE
S
INTER AGAINST
THIS EDGE,
TIGHTEN THE
G1/4
6
S
Sample gas inlet
use connector
R1/4 ISO 7/1
68
SAMPLE CELL
CLASP NUT.
DMP248SC
IN
G1/2
ISO228/1
OUT
view
A - A
Sample gas outlet
use connector R1/4 ISO 7/1
FIGURE 3-13 Dimensions (in mm) of the DMP248SC sample cell.
25
22
80
25
20
for DMP248 probe
A
A
fixing screws
(e.g. M6x60 or 1/4" x 11/4")
TOP VIEW SIDE VIEW
40
FIGURE 3-14 Fastening to a metal plate (top view) and on a concrete
frame (side view)
18_____________________________________________________________________M210492EN-A
CHAPTER 3_______________________________________________________________INSTALLATION
process pipe
28
6
6
S
G1/4
ISO
228/1
SAMPLE CELL
DMP248SC
CLASP NUT.
THIS EDGE,
PUSH PROBE
TIGHTEN THE
S
INTER AGAINST
G1/4
screw
FRONT VIEW TOP VIEW
metal hose clamp
thread M6 or 1/4"-20 UNC
metal
plate
thickness e.g.
3mm (1/8")
AISI 316
40
Dimensions of the metal plate
68
80
FIGURE 3-15 Fastening to a process pipe with the help of a metal
plate
An overpressure in the process is necessary to create a flow through the sample cell. Note that the pressure of the sample cell must not differ from that of the process because dewpoint temperature changes with pressure. In dirty processes, it may be necessary to use a filter between the cooling coil and the sample cell. One more simple way of using the sample cell with user provided accessories is shown in
FIGURE 3-16. The flow through the sample cell is controlled with the
needle valve and the pressure is kept equal to that of the process.
FIGURE 3-16 Installing the probe in high temperatures (an example).
VAISALA _________________________________________________________________________19
OPERATING MANUAL_______________________________________________________________
Grounding
A single electrical cable with a screen and three to ten wires is recommended for power and analogue output/serial bus connections. The cable diameter should be 7...10 mm.
The screen of the electrical cable must be grounded properly to achieve best possible EMC performance. Recommended cable shield is done in the cable gland as shown.
remove the brass disks, rubber ring and nut from the transmitter
housing
strip 165 mm of the cable insulation, but leave 25 mm of the braid
visible
slip the nut and rubber ring over the cable insulation
slip the brass disk that has the bigger hole in it over the braid so
that it rests against the cable insulation
slip the other brass disk over the wires to the middle of the braid
flexible wires 0.5 mm²
(AWG 20), stranded wires
recommended
3
140
165
braid
brass disks
rubber ring
nut
cable
25
D = Ø 7...10 mm
(If the cable diameter is less
than 7mm, use a shrinking
tube or an adhesive tape)
shielding tube
braid
brass disks
push back the braid and press it between the two brass disks to
achieve a full 360° grounding; the fold between the disks should have the same diameter as the brass disks
secure the braid with a shielding tube
20_____________________________________________________________________M210492EN-A
CHAPTER 3_______________________________________________________________INSTALLATION
insert the wires into the transmitter housing through the gland
tighten the nut
connect the wires into the screw terminals and fasten a cable tie
around the wires
cable tie
transmitter housing
gland
brass disks
rubber ring
nut
NOTE
When the cable is grounded as explained, the metallic parts of the sensor head, the screen of its cable, the transmitter housing and the screen of the signal cable to external system are all connected to each other. After this, the whole system can be grounded from one point only. If the grounding is made via several points (sensor head, transmitter housing, signal cable), make sure that the different groundings are made to the same grounding potential. Otherwise, harmful grounding currents may be generated. If you do the grounding via the transmitter housing, use one serrated lock washer between a mounting screw and the housing; the lock washer breaks the paint on the housing.
VAISALA _________________________________________________________________________21
OPERATING MANUAL_______________________________________________________________
+
Connections
CH1- and CH2- are connected
CH2
CH1+
CH1 -
CH2 -
together internally
+
+
X2
X1
OPENED COVER OF THE DMP248
CURRENT/VOLTAGE
-
V
mA
OUTPUTS
-
V
mA
POWER SUPPLY (INTERNAL OR EXTERNAL)
Do not use power supply ground (-) as output signal ground
24 V +
FIGURE 3-17 Electrical connections
Power supply 24 VDC
24 VAC (see Chapter Connection to an AC supply)
Output signals 0...20 mA
4...20 mA
0...1 V
0...5 V
0...10 V
Power supply ground (-) is connected to the housing with parallel connection of 15 nF capacitor and 300 k resistor.
22_____________________________________________________________________M210492EN-A
CHAPTER 3_______________________________________________________________INSTALLATION
Connection to an AC supply
The DMP248 transmitter can also be connected to an AC supply without an external rectifier. However, when more than one transmitter is connected for example to one 24 VAC transformer, a common loop is formed and there is an increased risk of a short­circuit. To avoid this, always use separate floating supply for each transmitter (see FIGURE 3-18A). However, if several transmitters have to share one transformer, the phase () must always be connected to + connector in each transmitter (see FIGURE 3-18 B).
A) NO COMMON LOOP FORMED - RECOMMENDED
DMP248 transmitter Controller
24 VAC
24 VAC
DMP248 transmitter
B) COMMON LOOP FORMED -
DMP248 transmitter
24 VAC
supply
voltage
supply
voltage
DMP248 transmitter
signal
supply
supply
output
voltage
signal
output
voltage
NOT RECOMMENDED!
Controller
signal
output
signal
output
shared common
line
FIGURE 3-18 Connecting the transmitter to an AC supply.
VAISALA _________________________________________________________________________23
CHAPTER 4_____________________________________________________________COMMISSIONING
CHAPTER 4 COMMISSIONING
When the DMP248 transmitter leaves the factory, its measurement ranges and output signals have already been selected. The user can subsequently change the measurement units between metric and non­metric and select and scale the output signals with software functions, see Chapter Selecting and scaling the analogue output quantities and Appendix 1.
Security lock jumper
Before the settings can be changed, the user must first remove the security lock jumper in connector X15 (see FIGURE 4-1). The security lock jumper makes it impossible to change the transmitter settings by mistake. The jumper should be removed only for changing the settings and for calibration; the auto-calibration is active only with the jumper connected. When the security lock jumper is connected, some commands cannot be used (see Chapter Commands and security lock jumpers).
CHANGE OF SETTINGS DISABLED
X15
OPENED COVER OF THE DMP248
FIGURE 4-1 Location of the security lock jumper.
VAISALA _________________________________________________________________________25
OPERATING MANUAL_______________________________________________________________
OPENED COVER OF THE DMP248
Selecting the analogue outputs
The DMP248 transmitter can be ordered ready with the current or voltage outputs required. If the outputs need to be changed, move the jumpers in connector X15 into positions as shown in FIGURE 4-2.
CH1
CH2
C H 2
C H 1
CURRENT OUTPUTS 0 ... 20 / 4 ... 20 mA
CH2
CH1
C H 1
C H 2
VOLTAGE OUTPUTS 0 ... 5 V / 0 ... 10 V
CH2
CH1
C H 2
C H 1
VOLTAGE OUTPUTS 0 ... 1 V
CH2
CH1
C H 1
C H 2
CH1 0 ... 1 VOLTAGE OUTPUT CH2 CURRENT OUTPUT
X15
FIGURE 4-2 Selecting the analogue outputs with jumpers.
The software also has to be informed which outputs are in use. This is done either through the serial interface or the menus on local display when in use. The serial command is AMODE and the display/keypad command "Mode ð Analog outputs ð Mode" (see Chapter 5
commands). If the outputs need to be scaled, see serial
26_____________________________________________________________________M210492EN-A
Loading...
+ 111 hidden pages