NOTICE
page 4 of 20
+ WARNING indicates a hazardous situation which, if not avoided, could
result in death or serious injury.
NOTICE is used to address practices not related to personal injury.
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 11/08/2015 // 999281
Page 5
Safety information!
page 5 of 20
IMPORTANT
In case of devices which have been in contact in contact with biological substances
of risk level 2 contact the VACUUBRAND service absolutely before shipping the
device. These devices have to be completely disassembled and decontaminated by
the user prior to shipment.
Do not return devices which have been in contact with biological substances
of risk level 3 or 4. These devices cannot be checked, maintained or repaired. The
devices must not returned to VACUUBRAND due to a residual risk in spite of decontamination.
The same conditions apply to on-site work.
This manual describes the decontamination of the aspiration systems BVC and the
handle VHC
The applicability of the decontamination method and the selection of the appropriate disinfectants have to be ensured by the user.
Ensure that work is done only by an appropriately trained and supervised personal.
Obey local and national safety requirements. Ensure that the personal is familiar
with the safety procedures which relate to the products processed by the pumping
system. Ensure that the device is decontaminated before repair and that you take
adequate precautions to protect people from the effects of hazardous substances if
contamination has occurred.
pro
.
NOTICE
+ Take adequate precautions to protect people from the effects of dangerous sub-
stances. Wear appropriate personal protective equipment (e. g., protective clothing,
safety glasses, protective gloves) and perform work in a hood.
Prior to decontamination isolate equipment from mains. Allow sufcient cooling.
Ensure that the device cannot be operated accidentally.
Wait 5 seconds before starting work to allow the capacitors to discharge.
Before starting check that the required parts (e. g. tools, disinfectants) are available
and of the correct type.
Check the operating sequence mentally on feasibility, safety requirements and consequences on safety and function of the equipment.
Obey regulations when disposing of solvents and chemicals.
In order to comply with law (occupational, health and safety regulations, safety at
work law and regulations for environmental protection) vacuum pumps, components
and measuring instruments returned to the manufacturer can be repaired only when
certain procedures (see section ”Repair - Maintenance - Return - Calibration”)
are followed.
Scrapping and waste disposal
Dispose of the pump and any components removed from it safely in accordance with
all local and national safety and environmental requirements.
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 11/08/2015 // 999281
Page 6
page 6 of 20
Materials
To facilitate the selection of a compatible disinfectant the materials of the devices
NOTICE
SurfacesMaterial*
BVC basic / G
Supportcoated stainless steel
Valve block
FeetNBR
BVC control / G // BVC professional / G
Support, cover pumpPBT
are listed.
For information about the compatibility with the materials, ask the manufacturer of
the disinfectant.
PA, PTFE, PP, PPS, POM, ETFE,
aluminium alloy
Pump housing, base platealuminium alloy
On / off switchPC, EPTR
IEC plugPET
FeetPU
Screwsstainless steel
Touchpanelglass
Liquid level sensor foil (only BVC professional / G)PET
We reserve the right for technical modications without prior notice!
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 11/08/2015 // 999281
We reserve the right for technical modications without prior notice!
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 11/08/2015 // 999281
Page 8
Decontamination of the vacuum pump
IMPORTANT
page 8 of 20
Important information
Devices may be contaminated due to operation.
The vacuum pump of the BVC may contaminated as well. The protective lter with
a pore size of 0.2 µm does not avoid safely a contamination of the vacuum pump in
all cases.
Also in case of a defect of the lter or operation of the BVC without lter pathogenic
germs (e. g., bacteria, viruses, prions, fungi) may enter the pump.
Prior to decontamination the devices have to be disassembled.
When disassembling the devices avoid contact of contaminated parts with not contaminated parts.
Adopt suitable measures to prevent the release of dangerous, toxic, explosive, cor-
rosive, noxious or polluting uids, vapors and gases and a possible contamination
of personal and environment.
Prior to decontamination clean with a damp cloth if necessary. Use water or a mild
detergent as cleaning agent.
Decontaminate all disassembled parts and the used tools using a appropriate
degassing method. Comply with exposure time.
The applicability of the decontamination method and the selection of the appropriate
disinfectant have to be ensured by the user.
After decontamination perform an appropriate verication procedure, e. g., use
contact plates in order to control disinfecting measures.
Enclose an evidence of verication of the decontamination stating the detection method with the device.
After decontamination return the device disassembled for repair.
Ensure that all parts are packed safely.
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 11/08/2015 // 999281
Page 9
page 9 of 20
Disassembling the BVC control / professional
IMPORTANT
Connect person to ground before coming in contact with electronic
parts.
Tools required (metric):
- Diaphragm key SW 46 (order no.: 637777)
- Screw driver Torx TX20
- Open-ended wrench SW 19
- Tweezers
➨Unscrew elbow tting with silencer at the outlet of the
pump.
➨ Depending on BVC version, remove connection tubing at
the hose nozzle or detach quick coupling.
➨ Remove bottle from support.
➨ Dismantled bottle into individual components.
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 11/08/2015 // 999281
Page 10
page 10 of 20
➨ Unscrew hose nozzle, or hose nozzle with coupling, using
an open-ended wrench.
➨ Unscrew cover, paying attention to washers.
➨ Remove the cover of the housing cover from the housing.
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 11/08/2015 // 999281
Page 11
page 11 of 20
View of the disassembled pump head parts ME 1C
7
8
12
1110
9
1: Cover of housing cover
2: Housing cover
3: Housing cover insert
4: Valves
5: Head cover
6: Guiding pin
6
345
1
2
7: Diaphragm clamping disc with square head
screw
8: Diaphragm
9: Diaphragm support disc
10: Washers
11: Rod
12: Housing
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 11/08/2015 // 999281
Page 12
page 12 of 20
➨ Unscrew four screws at the pump head, paying attention
to washers.
Remove housing cover.
➨ Unscrew union nut and remove hose from elbow tting.
➨ Remove housing cover insert with elbow tting.
➨ Note position of valves.
➨ Remove head cover and valves.
+ Never remove parts by using a pointed or sharp-edged
tool (e.g., screw driver), we recommend to use a rubber
mallet or compressed air (to be blown carefully into port).
➨ Lift diaphragm carefully sidewise.
+ Never use a pointed or sharp-edged tool to lift the dia-
phragm.
➨ Use the diaphragm key to grip the diaphragm support disc
below the diaphragm.
➨ Press down the diaphragm clamping disc to bring the
diaphragm to its lowest oscillating position. Unscrew diaphragm support disc with diaphragm and diaphragm
clamping disc.
+ If the diaphragm is difcult to separate from the diaphragm
support disc, immerse assembly in naphtha or petroleum
ether. Do not inhale vapors!
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 11/08/2015 // 999281
Page 13
page 13 of 20
➨ Lay down the BVC as shown in the gure.
➨ Unscrew screws at the base plate.
➨ Slide the base plate to the side.
IMPORTANT
Connect person to ground before coming in contact with electronic
parts.
➨ Remove earth cables.
➨ Remove hose from pressure sensor using tweezers.
➨ Remove sensor hose and dismantled into individual com-
ponents.
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 11/08/2015 // 999281
Page 14
page 14 of 20
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 11/08/2015 // 999281
Page 15
page 15 of 20
Disassembling the BVC basic
Disassemble the BVC basic into individual components as shown in the gure.
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 11/08/2015 // 999281
Page 16
page 16 of 20
Disassembling the VHC
➨ Select continuous aspiration.
➨ Turn cap (1) and remove (2).
1
2
pro
➨ Pull adapter with hose out of the VHC
➨ Remove adapter from hose.
➨ Pull aspiration hose out of the handle.
pro
.
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 11/08/2015 // 999281
Page 17
page 17 of 20
NOTICE
It is not possible to disassemble the VHC
pro
IMPORTANT
Do not return the aspiration hose.
completely.
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 11/08/2015 // 999281
Page 18
Repair - Maintenance - Return - Calibration
IMPORTANT
page 18 of 20
Every employer (user) is held responsible for the health and safety of his employees. This also applies to service personnel performing repair, maintenance, return
or calibration.
The health and safety clearance form informs the contractor about any possible
contamination of the device and forms the basis for the risk assessment.
In case of devices which have been in contact with biological substances of
risk level 2 contact the VACUUBRAND service absolutely before dispatching
the device. These devices have to be completely disassembled and decontaminated by the user prior to shipment. Do not return devices which have
been in contact biological substances of risk level 3 or 4. These devices cannot
be checked, maintained or repaired. Also decontaminated devices must not returned
to VACUUBRAND due to a residual risk.
The same conditions apply to on-site work.
No repair, maintenance, return or calibration is possible unless the correctly
completed health and safety clearance form is returned. Devices sent are rejected if applicable. Send a completed copy of the health and safety clearance
form to us in advance. The declaration must arrive before the equipment. Enclose a
second completed copy with the product.
Remove all components from the device that are no original VACUUBRAND components. VACUUBRAND will not be responsible for lost or damaged components that
are no original components.
Drain the device completely of uids and residues. Decontaminate the device.
Close all openings airtight especially if using substances hazardous to health.
To expedite repair and to reduce costs, please enclose a detailed description of the
problem and the product’s operating conditions with every product returned.
If you do not wish a repair on the basis of our quotation, the device may be returned
to you disassembled and at your expense.
In many cases, the components must be cleaned in the factory prior to repair.
For cleaning we use an environmentally friendly water based process. Unfortunately
the combined attack of elevated temperature, cleaning agent, ultrasonic treatment
and mechanical stress (from pressurised water) may result in damage to the paint.
Please mark in the health and safety clearance form if you wish a repaint at your
expense just in case such a damage should occur. We will also replace parts for cosmetic reasons at your request and at your expense.
Before returning the device
Pack the device properly, if necessary, please order original packaging materials at
your costs.
Mark the package completely
Enclose the completed health and safety clearance form.
Notify the carrier of any possible contamination if required.
Scrapping and waste disposal
Dispose of the equipment and any components removed from it safely in accordance
with all local and national safety and environmental requirements. Particular care
must be taken with components and waste oil which have been contaminated with
dangerous substances from your processes. Do not incinerate uoroelastomer seals
and O-rings. You may authorize us to dispose of the equipment at your expense.
Otherwise we return the device at your expense.
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 11/08/2015 // 999281
2. Serial no.: ...........................................................................................................................................................
3. Reason for return / malfunction: .........................................................................................................................
* Contact the VACUUBRAND service absolutely before dispatching the device.
** Devices which have been in contact with biological substances of risk level 3 or 4 cannot be checked, main-
tained or repaired. Also decontaminated devices must not returned to VACUUBRAND due to a residual risk.
☐ yes ☐ no
7. Radioactive contamination:
☐ yes ☐ no
8. Prior to return to the factory the device has been decontaminated: ☐ yes ☐ no
11. If the paint is damaged, we wish a repaint or a replacement of parts for reason of appearance
(repaint and replacement at customer’s expense): ☐ yes ☐ no
12. Legally binding declaration
We assure for the returned device that all substances, which have been in contact with the device are listed in
section 5 and that the information is complete and that we have not withheld any information. We declare that all
measures - where applicable - have been taken listed in section “Repair - Maintenance - Return - Calibration”.
By our signature below, we acknowledge that we accept liability for any damage caused by providing incomplete or incorrect information and that we shall indemnify VACUUBRAND from any claims as regards damages
from third parties. We are aware that as expressed in § 823 BGB (Public Law Code of Germany) we are directly
liable for injuries or damages suffered by third parties, particularly VACUUBRAND employees occupied with
handling/repairing the product. Shipping of the device must take place according to regulations.
Release for repair grant by VACUUBRAND (date / signature): ...................................................................................................................
VACUUBRAND GMBH + CO KG
Alfred-Zippe-Straße 4
97877 Wertheim - Germany
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 11/08/2015 // 999281
T +49 9342 808-5660 F +49 9342 808-5666
E-Mail: service@vacuubrand.com
www.vacuubrand.com
Page 20
page 20 of 20
Disclaimer: Our technical literature is only intended to inform our customer. The applicability of general
empirical values and results obtained under lab conditions to your specic operations depends on a
number of factors beyond our control. It is, therefore, strictly the users’ responsibility to very carefully
check the application of these data to their specic requirements. No claims arising from the information
provided in this literature will, consequently, be entertained.
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. 11/08/2015 // 999281
Alfred-Zippe-Str. 4 · 97877 Wertheim / Germany
T +49 9342 808-0 · F +49 9342 808-5555
info@vacuubrand.com · www.vacuubrand.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.