UVP SI-950 User Manual

SI-950 Ultraviolet Incubator
User Instructions
Web Site: www.uvp.com

81-0221-01 Rev K

UV Incubator Page 1
Temperature
Range
Ambient to 68 OC
Sensor
LM345 Integrated Temperature Sensor
Controller
PI
Setpoint
Digital
Display
Digital LED
Accuracy
+0.5OC
Uniformity
+0.1OC at 37OC
Relative Humidity
~ 80%
Unit Heat Load
Typical operation
Incubation
Heating Element
1250 Watts, 4,265,1777 BTUXH
Dimensions
Exterior W x H x D
17.5 x 17 x 14” (445 x 432 x 356mm)
Interior W x H x D
14 x 10.7 x 10.9” (356 x 272 x 277mm)
Shelves (2 standard)
Dimensions (W x D)
13.3 x 8.5” (338 x 216mm)
Construction
Formed stainless steel
Surface area
113.05 in2
Construction
Interior volume cu ft
0.95 cubic feet (26.9 liters)
Interior
Stainless steel
Exterior
Aluminum powder coated
Door
Polycarbonate
Electrical
115V
60 Hz
230V
50 Hz
100V
50/60 Hz

Introduction

This single chamber unit is designed for the variable-temperature inc ubation and humidification envelopment of microbiology organisms, biology assays and any protocol requiring heat incubation in the range of ambient +10°C to 68°C.
The Incubator is constructed with built-in overhead shortwave 254nm UV tubes for decontamination of the chamber between experiments.

Safety Information

The UV Incubator is designed with function, reliability and safety in mind. It is the user’s responsibility to install the unit in conformance with local electrical codes.

Specifications

UV Incubator Page 2
Additional Specifications
Pollution degree
2%
Installation Category
II
Altitude
2000m
Incubator is designed for operation inside a building only
Main supply
Not to exceed 10% of the nominal supply voltage.
Indoor Use

Installing the Incubator

1. Remove the incubator from its packaging.
2. Place the unit on a level working surface and provide adequate room in front of the system to allow the door to be opened easily.
3. Plug the female end of the power cord into the back of the unit.
4. Plug the male end of power cord into a surge-protected electrical outlet.
5. Press the I/0 toggle switch located in front of unit to turn the incub ator on. The touch pad should illuminate.
6. Select the desired temperature by pressing the touchpad up and down arrows. The internal temperature will be displayed on the touchpad.

Operating the Incubator

Setting the Temperature

When the power is turned on, the LED display flashes the software version number for several seconds. The display will then show the internal ambient temperature for several seconds and adjust to the set temperature (displayed in degrees Celsius). This setting is adjustable from the ambient temperature to 68
The current setpoint value can be altered by pressing the “UP” and “DOWN” buttons on the touchpad while the value is being displayed on the LED display. To change the setpoint:
1. Momentarily press then release either the “UP” or the “DOWN” button. The LED intensity will
flash to indicate that the displayed value is the current setpoint.
2. Increase or decrease the setpoint value by pressing the “UP” or “DOWN” buttons respectively.
If either button is held down for more than several seconds, setpoint value will increase or decrease continuously.
3. When the desired setpoint is reached, wait approximately five seconds without pressing either
button. The display will revert to normal mode showing the actual temperature.
The new setpoint becomes effective and is stored in the non-volatile memory when the display reverts to the normal mode.

Calibrating the Temperature Reading

The incubator is calibrated at the UVP factory at 37 oC. UVP recommends that temperature recalibration be performed at the UVP factory as recalibration by the user may void the warranty. Obtain an RGA (Returned Goods Authorization) number from UVP prior to returning a unit to the factory. Call UVP’s customer service department for information.
o
C.
UV Incubator Page 3
during incubation.

Setting the Decontamination Cycle

The SI-950 UV Incubator provides an automated process for eliminating contamination by utilizing the germicidal properties of ultraviolet light. The UV light is typically operated when no samples are inside the chamber. The entire chamber is decontaminated, eliminating viable fungi, bacteria and yeast. Visible light decontamination makes the SI -950 ideal for cGMP sterility testing and BL3 laboratories.
To operate the decontamination function, turn the master power on. Then, turn the UV key lock to the horizontal position to operate the germicidal lamp housed within the chamber.
Note: Do not attempt to incubate cultures with the germicidal lamp on. The germicidal lamp is used to decontaminate the internal chamber of the incubator between experiments.
Insert the UV key to operate the shortwave UV. Remove the key
after using the UV to avoid
accidental exposure of samples
UV key in the OFF position.
Warning: Exposure to UV light is dangerous. Keep the door closed.
The incubator door is UV blocking and will not allow UV radiation to pass through. The germicidal lamp will shut off if the door is opened.
A Word of Caution: The UV Incubator contains a powerful source of UV radiation that will cause damage to unprotected eyes and skin. Before operating the unit, be sure all personnel in the area are properly protected. It is preferable that the unit be installed and operated in an area where access and exposure are limited while the unit is in operation. Even though the unit shuts the UV off when the door is open, UV Blocking Eyewear should be worn. UVP carries a complete line of UV Blocking Eyewear, including Spectacles, Goggles and Faceshield. See “Replacement Parts” for more information.
Extended periods of time can be used to decontaminate the interior of the incubator chamber by the destruction of organisms such as:
-Bacillus anthracis -Proteus vulgaris
-Bacillus subtilus -Salmonella
-Corynebacterium -Streptococcus hemolyticus
-Escherichlia Coli -Asperg ill us nig er
-Micrococcus candidus -Saccharoyces ellipoideus
-Mycobacterium tuberculosis -Influenza
-Chlorella vulgaris -And others

Recommendations for UV Decontamination

Typically, the germicidal lamp is used during periods of non-use (overnight and during weekend periods of non-cultivation) to ensure a clean, non-infectious incubatory surface exists prior to sample testing.
UV key in the ON position.
UV Incubator Page 4

Care and Cleaning

The units are built to provide trouble-free oper atio n.
NOTE: The door is made of a polycarbonate plastic material and is subject to scuffing and scratches if improperly cleaned.
To ensure correct operation:
Wipe excess water from inside the unit and outside the unit with an absor b ent soft cl oth or sponge.
Use mild soap and water with a soft cloth or sponge to clean the exterior of the unit.
Use a mild detergent on the interior surfaces.
Warning: Unplug the UV incubator before cleaning around the UV bulb contacts.
Clean the door with a mild detergent; never use organic based compounds or cleansers
containing alcohol or ammonia.
Do not use abrasive pads or cleansers.
A plastic cleaner solution is recommended to clean the door. Such cleaner is available
from local plastic supply distributors.

Replacement Parts and Repl acement Instructions

Replacement parts are listed below for the SI-950 UV Incubator:
Key, replacement 53-0178-03
Tube, 8-watt, 254nm shortwave germicidal 34-0007-01
Tray, stainless steel ventilated 10-0341-08
Spectacles, UV blocking (UVC-303) 98-0002-01
Goggles, UV blocking (UVC-503) 98-0002-02
Face Shield, UV blocking (UVC-803) 98-0002-04

Replacing the UV Shortwave Bulb

1. Turn the unit power off and unplug the unit from the power outlet.
2. Open the incubator door and locate the bulb. Carefully rotate the tube and slide the bulb out of the socket.
3. Using two hands, carefully hold the bulb and fit the bulb end into the socket. Replace with a new tube by sliding the tube into the socket and rotating into place.
UV Incubator Page 5

Maintenance, Repair and Technical Assistance

UVP offers technical support for all of its products. If you have questions about the product’s use, operation or repair, please contact UVP’s offic es at the locations below.

Troubleshooting

Plates Drying Out
Ensure that the door is securely closed to prevent any external air from entering the incubator chamber.

Returned Goods Authorization

Note: A Returned Goods Authorization (RGA) number must be obtained from UVP Customer Service before returning any product.
You may contact UVP’s Customer Service based on your location:
If you are in North America, South
America, East Asia or Australia:
Call (800) 452-6788 or (909) 946-
3197, and ask for Customer Service
during regular business days, between
7:00 am and 5:00 pm, PST.
E-mail your message to:
info@uvp.com
Fax Customer Service, and send it to
(909) 946-3597
Write to: UVP, LLC 2066 W. 11th
Street, Upland, CA 91786 USA
If you are in Europe, Africa, the
Middle East of Western Asia:
Call +44(0) 1223-420022, and ask for
Customer Service during regular
business days between 9:00 am and
5:30 pm.
E-mail your message to: uvp@uvp.co.uk
Fax Customer Service, and send it to:
+44(0) 1223-420561
Write to: Ultra-Violet Products Ltd
Unit 1, Trinity Hall Farm Estate, Nuffield
Road, Cambridge CB4 1TG UK

Warranty

UVP's products are guaranteed to be free of defects in materials, workmanship and manufacture for one (1) year from date of purchase. Consumable and disposable parts including, but not limited to tubes and filters, are guaranteed to be free from defects in manufacture and materials for ninety (90) days from date of purchase. Transilluminators carry a two (2) year warranty from date of purchase. If equipment failure or malfunction occurs during the warranty period, UVP shall examine the inoperative equipment and have the option of repairing or replacing any part(s) whic h, in the judgment of UVP, were originally defective or became so under conditions of normal usage and service.
No warranty shall apply to this instrument, or part thereof, that has been subject to accident, negligence, alteration, abuse or misuse by the end-user. Moreover, UVP makes no warranties whatsoever with respect to parts not supplied by UVP or that have been installed, used and/or serviced other than in strict compliance with instructions appearing in this manual. In no event shall UVP be responsible to the end-user for any incidental or consequential damages, whether foreseeable or not, including but not lim ite d to property damage, inability to use equipment, lost business, lost profits, or inconvenience arising out of or connected with the use of instruments produced by UVP. Nor is UVP liable or responsible for any personal injuries occurring as a result of the use, installation and/or servicing of equipment.