The Pen-Ray Power Supplies are made specifically for the Pen-Ray Lamps. These power supplies provide proper voltage and current for starting and operat-
DESIGN
Each power supply is designed to be maintenance free. There is an internal circuit protection fuse which is designed to prevent mains circuit shut down in the unlikely case of a catastrophic internal failure. The connector must be used with the mating connector supplied with the Pen-Ray lamps. The connector is certified
to UL and CE compliance for use with this power supply. The power supply carries the CE, ETL and ETLc marks.
WARNINGS
Never handle the connector when the unit is in position as a hazardous shock condition will exist.
Do not attempt to operate lamps with operating voltages greater than 300 volts continuously. Starting voltages do not apply.
If the equipment is used in a manner not specified by UVP, LLC, the protection provided by the equipment may be impaired.
Unit may become HOT during continuous use.
If internal circuit protection fuse blows, the unit should be returned to the factory for service. The internal protection fuse is designed to trip if the primary coil short circuits for any reason, protecting the user and the circuit it is plugged into. The unit may be cleaned with any mild cleaning agent. Although the powder coated finish is extremely durable, avoid acids and caustic substances. Do not use solvent cleaners on plastic end caps.
ENVIRONMENT AL CONDITIONS
For indoor use only
Temperature: 5 to 40OC
Maximum or Relative Humidity: 80% for temperatures up to 31OC decreasing linearly to 50% relative humidity at 40OC
Mains supply voltage fluctuations not to exceed +10% of the nominal voltage
Installation is category II
Pollution degree is 2
OPERATING INSTRUCTIONS
The operation of the PS series power supply is as follows:
1. Make sure the switch is in the off position.
2. Connect the Pen-Ray lamp to the connector making sure the connector is completely inserted into its mating part.
3. Plug the unit into proper mains power supply.
4. Turn on the unit and the lamp should immediately light up.
5. Always unplug the unit when disconnecting or connecting the lamp.
MAINTENANCE
Keep the connector clean and free from debris, oil, fluids, etc. If the cord should become frayed or cracked, it should be replaced immediately. Call UVP for replacement service.
If connector arcing should occur due to dirt in the connector or attempting to operate a Pen-Ray lamp which operates in excess of 300volts, do not continue operation. The unit should be returned to the factory for repair. (Cleaning the carbonized area between the connector pins may restore function to the power supply).
SPECIFICA TIONS
Primary Model Primary Current Starting Max. Lamp Lamp No. Voltage AMPS Voltage Voltage Current Dimensions
PS-1 115Vac/60Hz 0.40 2300Vac 300Vac 18mA 6.3x3.7x2.3" PS-1 230Vac/50Hz 0.21 2300Vac 300Vac 18mA (160x94x53mm)
PS-2 115Vac/60Hz 0.40 2000Vac 300Vac 6mA 6.3x3.7x2.3" PS-2 230Vac/50Hz 0.21 2000Vac 300Vac 6mA (160x94x53mm)
PS-3 115Vac/60Hz 0.40 2100Vac 300Vac 10mA 6.3x3.7x2.3" PS-3 230Vac/50Hz 0.21 2100Vac 300Vac 10mA (160x94x53mm)
PS-4 115Vac/60Hz 1.00 2800Vac 300Vac 26mA 4.6 x 5.8 x 3.8” PS-4 230Vac/50Hz 0.55 2800Vac 300Vac 26mA (117x147x97mm)
PS-9 115Vac/60Hz 0.90 2400Vac 300Vac 48mA 5.8 x 4.6 x 3.6” PS-9 230Vac/50Hz 0.50 2400Vac 300Vac 48mA (147x117x92mm)
PS-11 115Vac/60Hz 0.40 2300Vac 300Vac 18mA 5.2x2.2x3.5" PS-11 230Vac/50Hz 0.21 2300Vac 300Vac 18mA (132x56x89mm)
NOTE: 115V ETL LISTED US/CANADA; 230V CE COMPLIANCE
81-0154-01 Rev F
Web Site: www.uvp.com
UVP, LLC 2066 W. 11th Street, Upland, CA 91786 • Tel: (909) 946-3197 or (800) 452-6788
Ultra-Violet Products Ltd, Unit 1, Trinity Hall Farm Estate, Nuffield Road,
Cambridge, CB4 1TG UK • Tel: +44(0)1223 420022
Cambridge, CB4 1TG UK • Tel: +44(0)1223 420022
Web Site: www.uvp.com
81-0154-01 Rev F
Primary
Ultra-Violet Products Ltd, Unit 1, Trinity Hall Farm Estate, Nuffield Road,
UVP, LLC 2066 W. 11th Street, Upland, CA 91786 • Tel: (909) 946-3197 or (800) 452-6788
NOTE: 115V ETL LISTED US/CANADA; 230V CE COMPLIANCE
PS-11 230Vac/50Hz 0.21 2300Vac 300Vac 18mA (132x56x89mm) PS-11 115Vac/60Hz 0.40 2300Vac 300Vac 18mA 5.2x2.2x3.5"
PS-9 230Vac/50Hz 0.50 2400Vac 300Vac 48mA (147x117x92mm) PS-9 115Vac/60Hz 0.90 2400Vac 300Vac 48mA 5.8 x 4.6 x 3.6”
PS-4 230Vac/50Hz 0.55 2800Vac 300Vac 26mA (117x147x97mm) PS-4 115Vac/60Hz 1.00 2800Vac 300Vac 26mA 4.6 x 5.8 x 3.8”
PS-3 230Vac/50Hz 0.21 2100Vac 300Vac 10mA (160x94x53mm) PS-3 115Vac/60Hz 0.40 2100Vac 300Vac 10mA 6.3x3.7x2.3"
PS-2 230Vac/50Hz 0.21 2000Vac 300Vac 6mA (160x94x53mm) PS-2 115Vac/60Hz 0.40 2000Vac 300Vac 6mA 6.3x3.7x2.3"
PS-1 230Vac/50Hz 0.21 2300Vac 300Vac 18mA (160x94x53mm) PS-1 115Vac/60Hz 0.40 2300Vac 300Vac 18mA 6.3x3.7x2.3"
No. Voltage AMPS Voltage Voltage Current Dimensions Model Primary Current Starting Max. Lamp Lamp
SPECIFICA TIONS
supply). carbonized area between the connector pins may restore function to the power operation. The unit should be returned to the factory for repair. (Cleaning the operate a Pen-Ray lamp which operates in excess of 300volts, do not continue If connector arcing should occur due to dirt in the connector or attempting to
replacement service. become frayed or cracked, it should be replaced immediately. Call UVP for Keep the connector clean and free from debris, oil, fluids, etc. If the cord should
MAINTENANCE
5. Always unplug the unit when disconnecting or connecting the lamp.
4. Turn on the unit and the lamp should immediately light up.
3. Plug the unit into proper mains power supply.
completely inserted into its mating part.
2. Connect the Pen-Ray lamp to the connector making sure the connector is
1. Make sure the switch is in the off position. The operation of the PS series power supply is as follows:
OPERATING INSTRUCTIONS
C decreasing
O
Pollution degree is 2
Installation is category II
Mains supply voltage fluctuations not to exceed +10% of the nominal voltage
C
O
C
O
linearly to 50% relative humidity at 40
Maximum or Relative Humidity: 80% for temperatures up to 31
Temperature: 5 to 40
For indoor use only
ENVIRONMENT AL CONDITIONS
on plastic end caps. extremely durable, avoid acids and caustic substances. Do not use solvent cleaners may be cleaned with any mild cleaning agent. Although the powder coated finish is circuits for any reason, protecting the user and the circuit it is plugged into. The unit service. The internal protection fuse is designed to trip if the primary coil short If internal circuit protection fuse blows, the unit should be returned to the factory for
Please read operating instructions before turning unit on.
the CE, ETL and ETLc marks. to UL and CE compliance for use with this power supply. The power supply carries
the mating connector supplied with the Pen-Ray lamps. The connector is certified unlikely case of a catastrophic internal failure. The connector must be used with circuit protection fuse which is designed to prevent mains circuit shut down in the Each power supply is designed to be maintenance free. There is an internal
DESIGN
output from and maximum life span of the Pen-Ray lamps. ing the lamps. Mating recommendations are available to provide optimum energy These power supplies provide proper voltage and current for starting and operat­The Pen-Ray Power Supplies are made specifically for the Pen-Ray Lamps.
INTRODUCTION
WARNING
Unit may become HOT during continuous use.
provided by the equipment may be impaired.
If the equipment is used in a manner not specified by UVP, LLC, the protection
continuously. Starting voltages do not apply.
Do not attempt to operate lamps with operating voltages greater than 300 volts
condition will exist.
Never handle the connector when the unit is in position as a hazardous shock
WARNINGS
Operating Instructions
PS-1, PS-2, PS-3, PS-4, PS-9 and PS-11
Pen-Ray Power Supplies
Loading...