Remote lead socket
(15 pin D type)
Also used for
downloading data
Rubber boot
to PC
4 line display
Battery
Compartment
(back)
Commonly used keys
ON/OFF UP
ON
OFF
ENTER & SET/CAL
SET/CAL
ENTER
Accepts parameters
Enters values
ON/OFF Key
UP key
DOWN key
ENTER/SETCAL
keys
Pages up through screens
Scrolls up through options, ie Fuel
DOWN
Pages down through screens
Scrolls down through options
KM9106 Operators Manual
1
1.2 Analyser Layout
Oxygen SensorFilter Bridge
Connection lead
for Oxygen
sensor
or Sulphur Filter
Particle Filter
Battery
enclosure
Printer unit
Handset
(stored in
pocket)
Dual
pressure ports
Water trap
connection
Water trap
located on
side of instrument
Accessory
storage
space
(leads/water
trap etc.)
Instrument case
2
KM9106 Operators Manual
1.3 Standard Probe Configuration
KM9106 Operators Manual
3
1.4 Analyser Connections
4
KM9106 Operators Manual
2. SAFETY WARNING
This analyser extracts combustion gases that may be toxic in relatively low
concentrations. These gases are exhausted from the bottom of the instrument. This
instrument must only be used in well ventilated locations. It must only be used by
trained and competent persons after due consideration of all the potential hazards.
3. FIRST TIME USE
Charge the battery for 12 hours, following this an overnight charge should be
sufficient for an average 8 hour day.
Whilst charging the green LED will be illuminated, the LED will flash when the
battery is fully charged.
Check that you have all the items you have ordered.
We offer a wide choice of probes which are not supplied as standard and must be
ordered as a separate item.
Take time to read this manual fully.
TIP: Take a look at the Spare Parts list and order some replacement filters
and paper rolls now.
When using the analyser for the first time you have the following under your control
:-
PARAMETER SECTION
Display Contrast 5.4
Backlight 5.4
Language 6.4
Line Rejection for mains frequency 6.4
Gas Measurement Scale 6.4
Temperature scale 6.4
Pressure scale 6.4
Reference oxygen 6.4
Time and date 6.1
KM9106 Operators Manual
5
Printed header name and telephone number 7.1
6
KM9106 Operators Manual
4. NORMAL START UP SEQUENCE
4.1 Every Time You Use The Analyser
BEFORE SWITCH-ON CHECK THAT:
the Oxygen sensor is connected
the particle filter is not dirty
the sulphur filter is fitted for heavy oil or coal
the water trap and probe line are empty of water
all hose connections, etc, are properly made
the paper roll is fitted
the probe is sampling ambient FRESH air
the water trap is vertical
the flue temperature is connected
the instrument is placed on a clean, flat, level surface
Switch ON the instrument by pressing
4.2 Automatic Calibration
ON
During this sequence the analyser pumps fresh air into the sensors to allow toxic
sensors to be set to zero and the Oxygen sensor to be set to 20.9 %.
After switch-on the analyser will
briefly display the Kane logo and
telephone numbers:-
777 KANE-MAY 777
TEL +44 (0) 1707 375550
FAX +44 (0) 1707 393277
Followed by the logger menu screen :-
9 9 SELECT FUNCTION 9 9
1
. . LOGGER CONTROL
2 . . QUINTOX CONTROL
Use the and keys to
position the cursor over 2 . .Quintox Control.
Press to access Quintox
Control
ENTER
99 SELECT FUNCTION 99
1 . . LOGGER CONTROL
2
. . QUINTOX CONTROL
KM9106 Operators Manual
7
And show the countdown screen :-
-- -- -- -- -- -- -- -- -- -- -- -- --
9 CALIBRATING 9 9
300
-- -- -- -- -- -- -- -- -- -- -- -- --
The display will countdown from 300 to zero in one second steps. If the analyser has
been used recently it may complete automatic calibration in less than 300 seconds
otherwise it will count to zero.
Once the calibration sequence is
complete an audible beep will be heard
and will show the selected fuel on the
following display:-
Press
This zeros the toxic sensor and sets Oxygen to 20.9%. The next screen is the MAIN
DISPLAY of the analyser:-
ENTER
-- -- -- -- -- -- -- -- -- -- -- -- -NATURAL GAS
PRESS ENTER KEY
-- -- -- -- -- -- -- -- -- -- -- -- --
DATE . . . 07-08-96
TIME . . . 12:31:35
INSTABILITY 0
BATTERY % . . 54
MAIN DISPLAY
NOTE : It is advisable to repeat Oxygen Calibration every 2 hours for maximum
All parameters are detailed in Appendix A - MAIN DISPLAY PARAMETERS.
8
KM9106 Operators Manual
4.3 Main Displays
The main display can be changed to show the following :-
• Page Mode displays 4 lines of data in set format, each page is predefined.
• Line scroll mode allows you to customise the display with the data you require.
To change between the different modes :-
Press followed by
SET/CAL
PAGE MODE
SET
The and keys change
between Page and Line Scroll Modes
Press to select
ENTER
4.3.1 Page Mode
LINE SCROLL MODE
SET
Use the and keys to change the information displayed on the
screen. The following are a number of the pages available. Other parameters on other
screens are detailed in Appendix A.
CO2 % . . . 0.0
FLUE C . . . 0.0
INLT NOT FITTED
AMBIENT C . . . 21
CO/CO2 R . . 0.0001
P INDEX % . . . 0.01
XAIR % . . . 0.0
Prs mbar 0.00
This screen only shows readings if optional
sensors are fitted. In this instance the SO2
sensor is NOT FITTED.
NO ppm . . 0000
NO2 ppm . . 0000
NOx ppm . . 0000
SO2 NOT FITTED
KM9106 Operators Manual
9
4.3.2 Line Scroll Mode
Line scroll mode allows you to customise the display.
Use the key to change the bottom line of the display. Once the correct line is
displayed press to confirm and move the line up. Select the next parameter and
repeat until all lines display the desired parameters.
Change bottom line using
NETT C . . . 0.0
O2 % . . . 20.9
CO ppm . . 0000
CO2 % . . . 0.0
to select and move
parameter up
O2 % . . . 20.9
CO ppm . . 0000
CO2 % . . . 0.0
CO2 % . . . 0.0
Select next parameter.
Repeat above until
display reads desired data
O2 % . . . 20.9
CO ppm . . 0000
CO2 % . . . 0.0
CO/CO2 R . . 0.0001
10
KM9106 Operators Manual
4.4 Sampling the Flue Gas
Once the automatic calibration procedure has been completed and the specific fuel has
been selected the probe can be inserted into the desired sampling point.
It is recommended that the sampling point be located at least two flue diameters
downstream of any bend and that the probe tip is in the centre of the flue (this is
normally the point of the hottest temperature). With balanced flues and other domestic
units the probe should be positioned far enough into the flue so that no air can ‘back
flush’ into the probe.
The probe depth stop cone provided with the instrument allows the probe to be used
in holes whose diameters range from 8 mm to 21 mm (5/16 to 4/5 inch).
The standard probe is rated at 650°C/1202°F. Temperatures of up to 1100°C/2012°F
can be accommodated using an optional high temperature probe.
TIP:To conserve battery power, switch off the pump when you are not taking a
measurement. Use the key to turn the pump ON and OFF.
PUMP
KM9106 Operators Manual
11
4.5 Taking a Pressure Reading
With the optional pressure module fitted a flue draught measurement can be made at
any time.
Connect the standard probe to the appropriate pressure sensor inlet and place the
probe in the flue.
The pressure reading will be displayed :-
CO/CO2 R . . 0.0001
P INDEX % . . . 0.01
XAIR % . . . 0.0
Prs mbar 0.00
NOTE: Care must be taken to connect the probe to the correct port as the pressure
will be displayed in reverse i.e. a positive pressure rather than negative
draught.
To perform a combustion test and display draught pressure at the same time a special
probe is required. Contact Kane International or Authorised Distributor for details.
Two pressure ports are provided on the instrument for use with a Pitot tube. Contact
Kane International Ltd. for details of this probe and its availability.
4.6 Regular Checks During Sampling
Care must be taken at all times not to exceed the analyser’s operating specification. In
particular ensure the following :-
• Do not exceed the maximum temperature of the flue probe.
• The analyser’s internal temperature does not exceed normal operating range,
typically 0-40°C.
• DO NOT PLACE THE ANALYSER ON A HOT SURFACE.
• The water trap is vertical at all times. Water condenses in the probe line and can
quickly fill the water trap when the probe is moved. Take care and watch the water
trap closely.
• The in-line particle filter is clean and does not become blocked.
12
KM9106 Operators Manual
4.7 Normal Shutdown Sequence
DO THIS EVERY TIME YOU USE THE ANALYSER.
Remove the probe from the flue - TAKE CARE ! THE PROBE WILL BE HOT -
and allow it to cool naturally. Do not immerse the probe in water as this will be drawn
into the analyser and damage the pump and sensors.
Once the probe is removed from the flue, the oxygen reading 20.9% and the CO
reading is zero press and the analyser will switch off.
OFF
Check the water trap and probe tubes for water before packing away.
4.8 Electromagnetic Compatibility
The European Council Directive 89/336/EEC requires that electronic equipment does
not generate electromagnetic disturbances that exceed defined levels and has an
adequate level of immunity to enable it to be operated as intended. The specific
standards applicable to this product are detailed in the appendices.
Since there are many electrical products in use that pre-date this Directive and may
emit electromagnetic radiation in excess of the standards defined in the Directive
there may be occasions where it would be appropriate to check the analyser prior to
use. The following procedure should be adopted:
Go through the normal start up sequence in the location where the equipment is to be
used.
Switch on all localised electrical equipment that might be capable of causing
interference.
Check that all readings are as expected. (A level of disturbance in the readings is
acceptable). If not adjust the position of the instrument to minimise interference or
switch off, if possible, the offending equipment for the duration of the test.
N.B. Maximum cable lengths must be less than 3 metres.
At the time of writing this manual (January 1997) Kane International Ltd is not aware
of any field based situation where such interference has ever occurred and this advice
is only given to satisfy the requirements of the Directive.
4.9 HC104 Module
With this module fitted and working the CO2 value is always the measured value not
the calculated value. The handset display and print out indicate that the CO2 value is
a measured value by showing CO2m. If the module is not fitted, the KM9106
automatically defaults to calculated CO2. If during start up calibration a fault is
detected in the CO2 module, the instrument will default to calculated. If the
instrument displays CO2m FAULT during operation, by switching OFF and then ON,
the instrument will default to calculated CO2.
KM9106 Operators Manual
13
5. USING THE KEYPAD
5.1 Basic Operation
Basic operation of the keypad to change the display in PAGE and LINE SCROLL
mode is detailed section 4.3. These modes give you the facility to perform the
following :-
• Page Mode displays 4 lines of data in set format, each page is predefined.
• Line scroll mode allows you to customise the display with the data you require.
5.2 QUICK key operation
To allow parameters to be viewed quickly Quintox has a number of QUICK keys.
Many of these keys have two functions.
To select an LOWER PARAMETER
parameter simply press the key i.e.
to display CO2 and three other
CO
2
values
To select a UPPER PARAMETER
parameter press the UPPER FUNCTION
key
Followed by the UPPER PARAMETER
you require parameter key i.e.
NETT to display Nett temperature plus O
FLUE to display Flue temperature plus Ambient, O2 and Prs.
INLET to display Inlet temperature plus Ambient, O2 and Flue
O
2
to display Oxygen reading plus CO, CO2 and Nett
CO2 to display Carbon Dioxide calculation or reading CO, Nett and XAIR
CO to display Carbon Monoxide plus Nett, O2 and CO2
AUX to display AUX1 and CxHy readings plus Nett and CO
FUEL to display chosen fuel and its parameters
EFF to display Gross efficiency plus O2 and CO
14
KM9106 Operators Manual
, CO and Eff.
2
UPPER PARAMETER KEYS
ALWAYS Press UPPER FUNCTION first then
SCALE to display Scaling setup parameters
AMBIENT to display ambient temperature plus sensor, Flue and Inlet
SO2 to display Sulphur Dioxide reading plus Nett, O2 and CO
NO2 to display Nitrogen Dioxide reading plus NO, Nett and CO
NO to display Nitric Oxide reading plus NOx, Nett and CO
NO
X
PRESSURE to display the pressure reading plus Flue, Nett and O2
LOSSES to display all four losses
λto display Excess Air plus CO, Eff and CO2.
All measured and calculated values are detailed in Appendix A - Main Display
Parameters.
to display the Oxides of Nitrogen reading plus NO, CO and Nett
5.3 DISPLAY HOLD
The display hold function allows you to freeze values on the instrument allowing
them to be viewed or printed. No measurements are taken once this function has been
activated. This feature gives the following benefits :-
• Data can be viewed at a particular point in the boiler tuning process.
• Multiple printouts may be obtained of the same data.
• Scales can be changed between printouts giving different units. i.e. ppm and mg/m3
In this function only the battery level will be updated and all other parameters are
frozen. This does not apply when AUTOPRINT is ON, the time and date are also
updated in this mode.
KM9106 Operators Manual
15
5.4 DISPLAY BACKLIGHT AND CONTRAST
The display contrast can be adjusted to suit different lighting conditions and the
backlight can be switched on or off.
Hold down either the or key and release once the desired level is
set.
Press to confirm setting.
ENTER
NOTE: The screen may flip from all dark to blank very quickly. Do not panic this
is normal - keep pressing the same key until the desired level is displayed.
TIP: If for any reason the display is not visible at switch on, simply
disconnect the handset for a few seconds and re-connect. Then reset
the display contrast as detailed above.
5.5 PUMP
The Pump can be toggled on or off from the handset.
TO TOGGLE THE PUMP ON /OFF
Press
PUMP
TIP: When the pump is switched off the O
reading will go down as the
2
oxygen sensor consumes the oxygen in its housing!
16
KM9106 Operators Manual
6. USER SELECTABLE SETTING
The following features are under your control at any time and can be changed as
detailed later in this section.
Time and date
Fuel type
Efficiency
Language
Line Rejection
Gas Scale
ppm(n) or mg/m
Compensation
Temperature scale
Pressure scale
Reference oxygen
NOx Calculation
temperature
Inlet
Oxygen calibration
Toxic sensor zero
Day/ Month order is selectable and the real time clock and
calendar are fully adjustable
.
Standard pre-programmed fuels can be selected or users can
define their own fuel characteristics
Efficiency readings can be selected to be based on Gross or Nett
values.
The analyser is programmed with ten languages.
For optimum mains electricity noise rejection a software
.
filter set for either 50 Hz or 60 Hz. Select 50 Hz in Europe
and 60 Hz in the USA. Check your country's power
frequency.
3
(n)
Normalised or un-normalised ppm or mg/m3 scalings can be
selected. Normalised is also known as referenced readings to
Oxygen. See Reference Oxygen below.
Some sensors are cross sensitive to other gases. Where
appropriate sensors are fitted so this cross sensitivity can be
compensated for, improving accuracy. During re-calibration
this compensation must be disabled
The analyser is programmed for both Celsius and Fahrenheit
The analyser is programmed for inWG, mBar, cmWG and
.
kPa.
Toxic gas measurements can be referenced to defined oxygen
levels.
Determines the level for calculating NOx. Depends on the
sensors fitted and local authority preferences.
The flue temperature probe can be used to measure and
programme the air inlet temperature to the boiler
If the analyser is being used for multiple checks over the
working day it is advisable to re-set the oxygen sensor at
regular intervals. This function allows re-set without the need
to repeat the start-up routine
The CO and other optional toxic sensors can be reset to zero if
KM9106 Operators Manual
17
they drift. This may happen if the sensor is taken to very high
R
R
concentrations for long periods of time or over-ranged.
6.1 Time and Date (Setting numbers)
This section gives details on setting the time and date and also the general principle of
setting a number from 0-9 used in other functions.
Each number in the date is to be set. The cursor _ under the two in the DAY above
indicates this number can be changed. Set each number in the date until correct using
the method below, this is also the method for SETTING NUMBERS.
Parameter
DATE: Select each number using or
Controls Options
0 - 9
key and
ENTE
The cursor _ moves to each number in turn
until the last number is set.
DATE 26-01-97
SET
18
KM9106 Operators Manual
To Set The Time:
TIME 16-01-12
SET
As with the date each number in the time is to be set.
Parameter Controls Options
TIME: Select each number using or key
and
ENTER
0 - 9
Once the last number has been set the screen will show the main display last shown
before entering the set time routine.
6.2 Changing a fuel
This section gives details on changing a standard fuel and entering a user fuel.
TO SET THE FUEL
Press the
SET/CAL
-- -- -- -- -- -- -- -- -- -- -- -- --
SET
-- -- -- -- -- -- -- -- -- -- -- -- --
Pressing the key displays
FUEL
STANDARD OR USER?
STANDARD FUEL
Parameter
FUEL
STANDARD
Controls Options
Select using or key and
ENTER
STANDARD
FUEL
USER FUEL
KM9106 Operators Manual
19
To Set A Standard Fuel:
There are over 70 standard fuels programmed into Quintox. The fuels are arranged
into tables for each country and the table should be selected for the origin of the fuel
used in your boiler. Each table contains different fuel types, choose the fuel that is
closest to the fuel used in your boiler.
SELECT FUEL TABLE
ENGLISH
Parameter
Controls Options
FUEL TABLE:
Select table for your country using or
key and
ENTER
ENGLISH
FRANCAIS
DEUTSCH
NEDERLANDS
ITALIANO
CASTELLANO
SVENSKA
SUOMI
OESTERREICH
ZEALAND
SELECT FUEL TYPE
NATURAL GAS
Parameter
FUEL TYPE:
The table shown above is for the UK. Fuel lists vary depending on the country.
Controls Options for
English Fuel Table
Select fuel type using or key
and
ENTER
NATURAL GAS
NATURAL GAS 2
TOWN GAS
LIGHT OIL
HEAVY OIL
COAL
ANTHRACITE
COKE
PROPANE
BUTANE
GASCOR
NEW
LPG
20
KM9106 Operators Manual
To Set A User Fuel:
If one of the standard fuels does not approximate to the one you are using in your
boiler then it is possible to set the Quintox up for a USER FUEL. The information
required to be able to set this are the Chemical Breakdown of the fuel and the Gross
and Nett calorific value. Details of the calculation are shown in the Appendix.
Parameter
K1g 0.350
K2 11.89
K4 32
Controls Options
SET
K1n 0.393
K3 9.83
O2r 3.0
USER FUEL:
Set each number as date using or
key and
ENTER
0 - 9
TIP: If you enter this routine in error, ENTER past all of the numbers to
exit.
Check fuel set by pressing
FUEL
NATURAL GAS
K1g 0.350
K2 11.89
K4 32
K1n 0.393
K3 9.83
O2r 3.0
6.3 Gross or Net Efficiency
The Quintox can calculate efficiency in of one of two ways.
Efficiency = 100% - Losses. See the Appendix for the Efficiency calculation.
•Gross efficiency uses the gross calorific value of the fuel and deems that the
latent heat of vapourisation is lost up the flue of the boiler and is taken as a loss.
Gross is used in the UK and USA.
•Net Efficiency uses the net calorific value and assumes the latent heat is not lost
up the flue. For Natural gas this efficiency can be 11% higher than the Gross
figure. Net is used in France and Germany.
NOTE: Latent heat is the heat required to turn water at 100°C into steam at
100°C, i.e. a change of state from liquid to vapour without rise in
temperature has taken place.
KM9106 Operators Manual
21
Loading...
+ 46 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.