UNI-T UTD2062C, UTD2042C, UTD3025C, UTD3062C, UTD2082C Operating Manual

...
Page 1
Page 2
Introductio
n
Dear users
:
.
Afer reading it, you are suggested to keep the manual properly. Please keep it with the instrument together or place it in an accessible location for future use
.
Page 3
General Safety Overview
The instrument is designed and manufactured strictl
y
according to “Safety Requirements of Electronic Measuring Instruments” (GB4793) and IEC61010-1 Safety Standard . It conforms to “Insulation Overvoltage Standard” (CAT 600V) and Safety Standard wit
h
Pollution Level You are expected to know the followin
g
safety precaution measures to avoid personal injury or product damage or any other product connected to it
.
Please use the product according to the regulations so as to avoid the possible damage
.
Only personnel with professional training can execute maintenance procedure to avoid fire or personal injury. Appropriate power wires should be used and special power wires must be approved by the state
.
Reliable grounding of products: The product is grounded by ground conductor of power supply
.
Ground conductor must be connected to earth to avoid electric shock. Please guarantee correct
grounding before connecting input/output terminal of
the product
.
Correct connection of oscillograph probe: Earth wire
of probe is the same to earth potential. Ple ase do not
connect earth wire to high volt age
.
Review of rated values of all terminals: Please refer to all rated values and marks on the product surface to avoid fire or excessive current shock. Please refer to
product manual before connecting products so as to
know details of rated value s.D o not open cover plate
of machine box: It is not allowed to start or operate the product when opening external cover or panel
.
Ċ
Ċ
Correct plug-in/out: Please do not plug in/out whe
n
connecting probe or testing conduits to voltage source
.
Operating Manual for UTD2000/3000
Page 4
Operating Manual for UTD2000/3000
Usage of appropriate fuse: It is only allowed to use fuse type and rated indicator designated by the product
.
To avoid exposed circuits: It is not allowed to touc
h
exposed connector or element after power supply
.
Operation is not allowed if there is any doubt for product trouble: Please consult qualified maintenance personnel to inspect the product if there is any trouble doubt
.
It has to maintain appropriate ventilation. It is not allowed to operate in wet locations. It is not allowed to operate in combustible and explosive environment. It has to keep product surface clea
n
and dry
.
Page 5
Terms in the manual: The following terms may be seen in the manual
:
Terms for products: The following terms may be Danger: It refers to direct injury near the mark
.
Warning: It refers to potential injury near the mark
.
Note: It refers to potential danger for the product and other properties
.
Warning: Warning s tatement means the
conditions and behaviors that may endanger lives.
Note: Notice statement means the conditions and behaviors that may cause other property damage
Protective
ground terminal
Ground terminal
for chassis
Ground terminal
for testin
g
Caution
!
Refer to manual
High voltage
Safety Ter ms and Symbols
Product symbol: Following symbols may be show
n
on the product surface
.
Operating Manual for UTD2000/3000
Page 6
Operating Manual for UTD2000/3000
The manual is about operation information of OPERATING MANUAL FOR UTD2000/3000 digital storage oscillograph. The manual includes the following chapters Chapter 1: User Guideline: It is to introduce functions of digital storage oscillograph easily and provide installation guidelines
.
Chapter 2: Instrument Setup:Operation methods of OPERATING MANUAL FOR UTD2000/3000 digital storage oscillograph
.
Chapter 3: Application example: Examples to solve various measurement problems Chapter 4: System Prompt and Troubleshootin
g
Chapter 5: Appendix: Technical Indicators
A
ppendix B: Fittings of OPERATING MANUAL FOR UTD2000/3000 series digital storage oscillograph
A
ppendix C: Maintenance and Cleaning Maintenance
Preface
Page 7
The manual includes 3 series of digital storage oscillograph as follows. Detailed models are shown in the following form
:
Operating Manual for UTD2000/3000
UTD2 C/UTD3025
C
02
5
UTD2 C/UTD3042
C
04
2
UTD2 C/UTD3062
C
06
2
UTD2 C/UTD3082
C
08
2
UTD2 C/UTD3102
C
10
2
UTD2 C/UTD3152
C
15
2
UTD2 C/UTD3202
C
20
2
UTD2 CE/UTD3042CE04
2
UTD2 CE/UTD3062CE06
2
UTD2 CE/UTD3082CE08
2
UTD2 CE/UTD3102CE10
2
UTD2 CE/UTD3152CE15
2
UTD2202CE/UTD3202CE UTD2025C
L
UTD2052C
L
UTD2052CE
X
UTD2102CE
X
25MH
z
40MH
z
60MH
z
80MH
z
100MH
z
150MH
z
200MH
z
40MH
z
60MH
z
80MH
z
100MH
z
150MH
z
200MH
z
25MH
z
50MH
z
50MH
z
100MH
z
250MS/s 500MS/s 500MS/s 500MS/s 500MS/s 500MS/s 500MS/s
1GS/s 1GS/s 1GS/s 1GS/s 1GS/s 1GS/s
1GS/s 1GS/s
250MS/s 500MS/s
UTD2000C/3000
C
UTD2000E/3000E
UTD2000
L
UTD2000E
X
UTD2000/300
0
Series Model
Band width
Real-time
Page 8
Operatin g Manual for UTD2000/3000
UTD2000/3000 series digital storage oscillograp
h
provides simple front panel with clear functions to users for all basic operations. Scale and position of various channels can provide intuitional operatio
n
to meet usage habits of traditional instruments
.
Users can use it skillfully without learning and being familiar with operation for a long time. Users can also press AUTO key directly to accelerate adjustment and facilitate measurement
.
A
ppropriate waveform and gear setup can be
shown on the instrument panel
.
UTD2000/3000 series digital storage oscillograp
h
also has high-performance indicators and powerful functions to fulfill measurement tasks quickl
y
besides easy usage. Quicker signals can be observed by UTD2000/3000 series digital storage oscillograph by real-time sampling of 500MS/s (or 1GS/s) and equivalent sampling of 25GS/s (or 50GS/s). It can capture and analyze waveform
easily by powerful trigger and analysis capacity
.
Users can also observe and analyze signal problems quickly and clearly by clear LCD displa
y
and mathematical calculation function. We can see that such series of digital storage oscillograph ca
n
meet your measurement requirements by following
performance characteristics
:
Double analog channels
High-resolution colorful LCD display system wit
h
resolution of 320 240 or 800 480
To support plug and play USB storage instrument
to communicate with computer
A
utomatic waveform and status setup
Waveform, setup, bitmap storage, waveform and
setup reoccurrence
Ɣ
Ɣ
Ɣ
Ɣ
Ɣ
h h
Page 9
Ɣ
Ɣ
Ɣ
Ɣ
Ɣ
Ɣ
Ɣ
Ɣ
Ɣ
Ɣ
Ɣ
Ɣ
hr

Āā
Āā
Elaborate view window extension function for
delicate analysis of w aveform details and profile
A
utomatic measurement of 28 kinds of
waveform parameters
A
utomatic cursor tracing and measurement
functio
n
Unique waveform recording and playback
functions
Embedded FFT Multiple mathematical calculation functions
(including +, -, for waveform
Edge, video, pulse width and ALT TRIG
functions
Multi-language menu displa
y
Fittings of UTD2000/3000 series digital storage oscillograp
h
2 probes of 1.2m 1:1/10:1 please refer to manual of probe Fittings of probe details so as to meet EN61010-031:2008 standar
d
1 power wire to meet national standards 1 Operating Manual 1 Product Warrant
y
USB connecting wire: UT-D06 or UT-D0
5
Communication control software for UTD2000/3000 series oscillograp
h
Operating Manual for UTD2000/3000
Page 10
Item
Elementary understanding of Trigger System..........................................
.
Page
..........................................................................................
.
.........................................................................................
.
......................................................................................
.
............................................................
.
Elementary understanding of Vertical System...............................................
.
Elementary understanding of Horizontal System..........................................
.
General Safet y Overvi ew Preface
Setup of Vertical System........................................................................
.
Setup of Horizontal System.....................................................................
.
Setup of Trigger System..........................................................................
.
General Inspectio
n
Function Inspectio
n
Probe Compensatio
n
A
utomaticSetup of Waveform Displa
y
Chapter 1: User Guideline ..................................................................................
.
5 5
8
9 9
11
1
4151627
31
Operatin g Manual for UTD2000/3000
Content
s
Chapter 2: Instr ument Setup.......................................................................
1
Page 11
Page
...................................................................................
.
..........................................................................................
.
Setup of Auxiliary Functions .............................................................................
.
A
utomatic Measurement...................................................................................
.
Cursor Measurement ........................................................................................
.
....................................................................................................
....................................................
.
.............
.
..................................................................
.
Setup of Display System
Storage and Callout
Operation Ke
y
Chapter 3: Application Examples
Example 1: Measurement of Simple Signals
Example 2: Delay Obser vati on of Sin e Sig na l s by Circ ui t.. ... .... ... ... ... ... .... ... .
.
Example 3: Capture of Sing le S igna l .... ... ... ... .... ... ... ... .... ... ... ... ... .... ... ... ... .... .
.
Example 4: Reduction of Ran do m Noi se for S igna ls. ... ....................... ... ... ....
.
Example 5: A p plic a tion Cu r sor Me asu r em en t.. ... .... ... ... ... .... ... ... ... .... ... ... ... ...
.
Example 6: Application of X-Y Functio
n
Example 7: Trigger of Video Signals
................. ... .... ... ... ... .... ... ... ....... ... ...
.
.................. ... .... ... ... ... ... .... ... ...
.
4
8505560687072727375767879
81
4
4
Operating Manual for UTD2000/3000
Item
Setup of Sampling System................................................................
.
Page 12
Item Page
.......................................................
.
............................................................................
.
....................
.
.....................................................
.
...............................................................................................
.
.........................................
Examp
Examp : Usage of Upgrading Procedure of U disk
System Prompt Informatio
n
Troubleshootin
g
A
ppendix A: Technical Indicators......................................................
.
A
ppendix B: Fittings of UTD2000/3000 Series Digital Storage Oscillograph..
.
A
ppendix C: Maintenan ce and Clea ning Mai nte nan ce... ..... .... .... .... ..... .. ....
.
Example 10: Usage of Storage Functio
n
Chapter 4: S ystem Prompt and Troubleshooting
Chapter 5: Technical Indicators............................................................
Example 8: Pass/Fail Detectio
n
Example
9
8
889
91
91
10
4
10
583848588
Operating Manual for UTD2000/3000
Page 13
Operating Manual for UTD2000/3000
Chapter 1: User Guideline
Ƹ
Ƹ Ƹ
Function inspectio
n
Elementary understanding of vertical system Elementary understanding of horizontal system
UTD2000/3000 series digital storage oscillograph is a small and handy desktop product. Basic test can be done by providing convenient front panel to facilitate operation for users
.
UTD2000/3000 series are to provide simple front panel with clear functions to users for basic operation. Panel also includes knob and function key
.
Knob functions are similar to those of other digital storage oscillograph
.
The chapter is to show execution of following steps
General inspectio
n
5 keys in a row in the right side of display panel are
menu operation keys. (They are defined as F1 to F
5
keys from top to bottom.) You can also set different options for current menu by them. Other keys are function keys. You can enter into different functio
n
menus or gain specific function application directly
by them
.
Elementary understanding of trigger system
Please know front operation panel firstly after getting UTD2000/3000 series digital storage
oscillograph. The chapter is to introduce operation
and functions of front panel for UTD2000/3000products so as to be familiar wit
h
usage within the shortest period
.
General inspectio
n
Ƹ
Ƹ
Ƹ
A
utomatic setup of waveform displa
y
Probe compensatio
n
Ƹ
1
Page 14
Operating Manual for UTD2000/3000
2
5.7 inch front panel of UNI-T
probe compensatio
n
signal output
vertical control
horizontal control
Operatio
n
controls
multi-purpose
knob controller
commo
n
men
u
trigger control
EXT TRIG INPUT
analog signal input
interface
Page 15
Operating Man ual for UT D2000/30 00
3
probe compensatio
n
signal output
vertical control
horizontal control
Operatio
n
controls
multi-purpose
knob controller
commo
n
men
u
trigger control
EXT TRIG INPUT
analog signal input
interface
7inch front panel of UNI-T
Page 16
Operatin g Manual for UTD2000/3000
4
display of horizontal trigger positio
n
time display of central scale line
vertical scale coefficient of display channel
display setup of
main time base
waveform display window
Menu varies wit
h
different functio
n
keys
.
mark of channel
2
display of trigger status
mark of channel 1
Interface display diagram of digital storage oscillograph (i.e. UTD2000L series
)
Page 17
Operating Manual for UTD2000/3000
5
General inspectio
n
Function inspectio
n
It is suggested to inspect the instrument b
y
following steps when getting a new UTD2000/3000 series digital storage oscillograph
.
1. To check if there is damage due to transportatio
n
or not
.
Please replace carton or foaming plastic protectio
n
cushion with severe damage instantly
.
2. Please inspect fitting details and shown it i
n
“Fittings of UTD2000/3000 series digital Storage Oscillograph” shown in the manual. You can check if there are deficient fittings or not by referring to it
.
Please contact UNI-T distributor or local agent if there are deficient or damaged fittings
.
3. Please inspect the entire machine and contact UNI-T distributor or local agent for damaged appearance
,
abnormal operation or failing to pass performance test
.
Please pay attention to packages and report to
transportation department and UNI-T distributor if machine damage is caused by transportation
.
1. Connect the power suppl
y
You can supply power to the machine and voltage is A
C
100V to AC 240V. Frequency is 45Hz to 440Hz. It has to press [UTILITY] menu and [F1] for self-correctio
n
after power supply 30 minutes after operation of hot machine so that digital storage oscillograph operates under the
best status. As shown in Figure 1 4, please
enter into next page and press [F1] to call out ex-factor
y
setup
.
UNI-T will arrange maintenance or replacement
.
Please carry out once fast function inspection to check if the machine operates normally or not
.
Please operate according to following steps
:
Page 18
Warning: Please confirm safety grounding of digital storage oscillograph to avoid dangers
.
2. Signal access of digital storage oscillograph is double-channel input for UTD2000/3000 series digital storage oscillograph. There is also 1 EXT TRIG INPUT channel. Please connect signals according to following steps
:
(As shown in Figure 1-5), connect probe of digital storage oscillograph to CH1 input terminal and set attenuation ratio switch on probe to 10×
Á
Figure 1 5 Setup of attenuation ratio switch o
n
probe
ķ
ˊ
Operating Manual for UTD2000/3000
6
Figure 1-
4
power switc
h
function butto
n
F
1
MEASURE
CURSO
R
DISPLAY
A
CQUIRE STORAGE
UTILITY
MULTIPURPOSE
SELECT
AUTO
CH
1CH2
F
2F3
F
4
F
5
POSITION
SCALE SCALE
VOLTS DIV
VVm
POSITION
SECD IV
SS
LEVEL
MATH
FORCE
HELP
MENU
TRI
G
MENU
HORI
VERTICA
L
CH
1
TRIGGE
R
HORIZONTA
L
CH2
SET
Operating Man ual for UT D2000/30 00
Page 19
Operating Man ual for UT D2000/30 00
7
ĸIn digital storage oscilloscope to set the probe
attenuation factor. The attenuation coefficient of vertical gear ratio change instruments, whic
h
makes measurement resultsSetup method of attenuation coefficient of pr obe is as follows to show amplitude of measured s ignals correctly
.
Please press
Figure 1-6 Setup of deflection coefficient of probe on digital storage oscillograp
h
F4 key for menu display of 10
×
Connect probe and grounding clip to
corresponding connection ends then press [AUTO]
button. As shown in Figure 1-7, square wave (1kHz
,
about 3V and peak value) can be displayed withi
n
several seconds. It has to inspect CH2 by the same
method then press [OFF] function button so as to power off CH1. It has to press [CH2] function butto
n
to power on CH2 and repeat the second and third
steps.
Ĺ
Proberatio
n
Figure 17㧙
Page 20
Operating Manual for UTD2000/3000
Probe compensatio
n
It needs to adjust it when connecting probe to an
y
input channel for the first time so that probe can be appropriate to input channel. Probe without compensation correction can also bring i
n
measurement tolerance or error. Please operate according to following steps for adjustment of probe compensation
:
1. To set probe menu attenuation coefficient to 10× and place probe switch to 10× then connect digital storage oscillograph probe to CH1. Please guarantee reliable contact to probe. It has to connect probe end to signal output connector2of
probe compensator then connect grounding clip to earth wire connector of probe compensator. It also has to power on CH1 and press [AUTO]
.
3. Adjust variable capacitance on the probe b
y
screwdriver of non-metallic handle for waveform display of “insufficient compensation” or “excessive compensation” shown in above figure until “correct compensation” for waveform displa
y
on the screen as above
.
Warning: Please guarantee perfect insulatio
n
conduits of probe and do not contact metallic part of probe when connecting to HV source to avoid electric shock of probe during HV measurement
.
Observation of displayed waveform
8
Exceedin
g
compensatio
n
Correct
compensation
Insufficient
compensation Figure 1
8
Compensation correction of probe
Page 21
Operatin g Manual for UTD2000/3000
9
Automatic setup of waveform displa
y
UTD2000/3000 series digital storage oscillograp
h
has automatic setup functions. It can adjust vertical deflection coefficient, scanning time base and trigger mode automatically according to input signal for the most appropriate waveform display
.
Frequency of measured signals shall be more tha
n
or equal to 50Hz and duty cycle shall exceed 1% according to automatic setup requirements
.
1. Connect measured signals to signal input channel
.
2. Press [AUTO] button so that digital storage oscillograph can set ver tical deflection coefficient
,
scanning time base and trigger mode automatically. They can be adjusted again after careful observation and automatic setup so that waveform display can realize the best effect
.
A
s shown in the following figure, there are a series of keys and knobs in vertical control a rea. Please introduce usage of vertical setup gradually as follows
.
Automatic setup
Elementary understanding of vertical system
5.7-inch vertical
control panel of scree
n
7-inch vertical control panel of scree
n
Figure 1 9 Vertical control area on the panel
Page 22
Operating Manual for UTD2000/3000
10
1. Use vertical position knob so that middle displa
y
of waveform signal in the window. Vertical displa
y
position of signal can be controlled by vertical
position knob
.
Ground mark moves vertically with waveform whe
n
rotating vertical position knob
.
Measurement knack You can observe difference between waveform and signal ground to measure DC component
of signal quickly if adopting DC channel coupling mode. DC components in the signal will be filtered if adopting AC coupling mode . You can also display AC component of signal
by higher sensitivity conveniently
.
Vertical position of double analog channels shall be
restored to short-cut key of zero pointnamely
,
SET TO ZERO
.
The key is to return vertical displacement/horizontal displacement/hold off position to zero point
(or middle point)
.
Of w hich: UTD2000L /2000EX se ries will trigger electric level Return to 50% of position
.
2 Alter vertical setup and observe alternation of status information. You can determine variation of an
y
vertical gear by s tatus bar below waveform window
.
Vertical gear of “volt/grid” can be altered by rotatin
g
vertical scale knob to show corresponding variation for
channel gear display which corresponds to status bar
.
User can press [CH1], [CH2] , [MATH] and [REF] to display operation menu, mark, and waveform and
gear status information of corresponding channels
.
Page 23
Operating Manual for UTD2000/3000
1
1
Users can also press [OFF] key to c lose currentl
y
selected channels for UTD2000C/3000C and UTD2000E/3000E oscillograph. They can also click on CH1, CH2 and MATH keys twice to close corresponding channels for UTD2000L/2000EX oscillograph
.
A
s shown in following figure, there is 1 key and
2
knobs in horizontal control area. Please be familiar
with setup of horizontal time base gradually as follows
:
Elementary understanding of horizontal system
Horizontal cont rol panel of 5.7-inc
h
Horizontal cont rol panel of 7-inc
h
Figure 1 10 Horizontal control area on the panel
Page 24
Operating Manual for UTD2000/3000
1. Use horizontal SCALE knob to alter horizontal time base gear then observe alternation of status information. To rotate horizontal SCALE knob to alter “s/div” time base gear so as to show corresponding variation of time base gear displa
y
for corresponding channels of status bar
.
Horizontal scanning speed steps from 2ns/di
v
s/div by way of 1 mode
.
Note: Horizontal scanning time base gear varies for UTD2000/3000 series digital storage oscillograph due to different models
.
γ
㧙㧙50 2
5
2. Use horizontal POSITION knob to adjust horizontal position of signals in waveform window
.
To use horizontal POSITION knob to control trigger displacement of signals. Horizontal movement of waveform with knob can be observed when rotatin
g
horizontal POSITION knob for application for trigger displacement
.
3. Please press [MENU] button for 5.7-inc
h
oscillograph (and press [HORI MENU] button for 7­inch oscillograph) to display Zoom menu. Please press [F3] to open view window extension under this menu then press [F1] to close view window extension and return to main time base. Hold off time can be set under this menu
.
1
2
Page 25
Operating Manual for UTD2000/3000
1
3
Trigger point can be r estored to vertical m iddle point quickly by short-cut key of [SET TO ZERO] so as to restoring displacement of trigger point to horizontal zero point short-cut key. Users can also adjust horizontal position of signal in waveform window by rotating horizontal knob
.
Term explanatio
n
Trigger point: It refers to posi tion of actual trigger point by comparing with middle point of storage instrument. Users can move trigger horizontally b
y
rotating horizontal POSITION knob
.
Hold off: It refers to interval to start up trigger circuit a gain
.
Hold off time can be set by rotatingmulti-purpose knob controller
.
A
s shown in Figure 1 there is 1 knob and
3
(or 2) keys in control area of trigger menu. Pleas e be familiar with setup of trigger system gradually as follows
:
POSITIO
N
Elementary understanding of trigger system
㧙11
5.7-inch screen 7-inch scree
n
Figure 1-1
2
Trigger men
u
Figure 1 Trigger menu on the panel㧙11
Page 26
1. To user trigger level knob to alter trigger level and instruct trigger level line by seeing trigger mark on the screen so that it moves vertical ly with knob rotation. Variation of tr igger level in lower part of screen can be observed when moving trigger level at the same time
.
2. (As shown in Figure 1-12,) TRIGGER MENU ca
n
be used to alter trigger setup
.
User can press [F1] key to select “edge” trigger
.
n trigger signal amplitude
.
i
4.To press FORCE button: To produce 1 trigger
signal forcefully for mainly application for normal and single trigger mode
.
Operating Manual for UTD2000/3000
14
They can press [F2]key to select “trigger source
to CH1
.
They can press [F3] key to set “gradient” of edge
type to ascending
.
They can press [F4 ]key to set “trigger mode” to
automatic. They can also press F5 key to set
“trigger coupling” to AC
.
They can press [50%] button for 5.7-inch oscillograp
h
and press [SET TO ZERO] key for 7-inch oscill ograph
.
It also has to set vertical middle point of trigger level
2
.
Page 27
Operating Manual for UTD2000/3000
1
5
Chapter 2: Instrument Setup
By now, you have been familiar with operation of vertical control area, horizontal control area and trigger system menu for UTD2000/3000 series digital storage oscillograph preliminarily. You should be familiar with setup of digital storage oscillograph b
y
menu operation by introduction shown in previous chapter. It is suggested to read Chapter 1 if you have not been familiar with above operation and methods
.
The chapter is mainly to introduce following themes
Set trigger systems of [TRIGGER LEVEL]
,
[MENU], ]50%] and] FORCE]
;
Set sampling mode of [ACQUIRE]
;
It is suggested to read the chapter carefully to know many measurement functions for UTD2000/3000 and system operation method
.
Ƶ
Ƶ
Ƶ
Ƶ Ƶ Ƶ Ƶ Ƶ Ƶ
Ƶ Set vertical systems of [CH1], ]CH2],] MATH]
,
]REF] , ]OFF] ]VERTICAL POSITION] and ]VERTICAL SCALE]
;
Set horizontal systems of ]MENU] or ]HO
R
MENU], ]HORIZONTAL POSITION] or [HORIZONTAL SCALE]
;
Set display mode of [DISPLAY]
;
Store and call-out for [STORAGE]
;
Set auxiliary system of [UTILITY]
;
A
utomatic measurement of [MEASURE]
;
Cursor measurement of [CURSOR]
;
Use execution buttons of [AUTO] and
[RUN/STOP]
;
Page 28
Operating Manual for UTD2000/3000
Setup of vertical system
CH1 and CH2 channels and setup Every channel has its own vertical menu. Ever
y
item can be set individually acc ording to different channels. As shown in Form 2-1, the system ca
n
display operation menu of CH1 or CH2 channel b
y
pressing CH1 or CH2 function key
.
1
6
1 1
0100
100
0
h
h
h
h
Setup
Descriptio
n
Function menu
Couplin
g
ACDC
To obstruct DC component of input signal
.
To cut off input signals by A
C
and DC components of input signals
.
Groundin
g
Bandwidt
h
limit
O
N
OFF
To limit band width to 20MH
z
so as to reduce display noise.
Full band widt
h
Volt /g ri
d
Coarse tunin
g
To set vertical deflectio
n
coefficient for coarse tunin
g
by 1-2-5 system. To refine
within setup scope of coarse
tuningforfinetuningsoasto
improve vertical resolution
.
Fine tunin
g
Probe
To select one value according
to probe attenuatio
n
coefficient to keep correct vertical deflection coefficient
.
There are totally 4 kinds
,
namely,1 ,10 ,10
0
and 1000
.
hh h
h
Reverse phase
O
N
OFF
To realize reverse directio
n
function for waveform
;
To display waveform normally.
Form 2-1
Page 29
1 Measured signal is a sine signal with DC component by taking exerting signal to CH1 channel as a
n
example
.
To press F1 to select AC and AC coupling mode. D
C
component contained by measured signal will be obstructed. Waveform display is shown in followin
g
figure
.
Press F1 to DC to pass through DC and A
C
components of measured signals of CH1 channel
.
Waveform display is also shown in following figure
:
Figure 2 2 S ynchronous dis play o f DC and A
C
components of signal Press F1 to select grounding and set channels to grounding mode. DC and AC components contained by measured signals will be obstructed. Waveform display is shown in following figure
:
displayed on the scr een
.
Setup of channel coupling㧦
1
7
Operatin g Manual for UTD2000/3000
Figure 2 Jammed DC component of signal㧙1
Setup of
A
C couplin
g
DC couplin
g
setup
Page 30
Operating Manual for UTD2000/3000
(Note: Input signal still shall be connected to channel circuit under such mode althoug
h
waveform is not
Figure 2 3 Synchronous obstruction of DC and
A
C components of signal
18
Ground couplin
g
setup
2.
Setup of cha nnel bandwidth limit
Input one sine signal of about 40MHz for CH1 as an
example. It has to press [CH1] to open CH1 channel then press [F2] to set bandwidth limit to OFF
.
Channel band width is full band width to pass
through high-frequency components contained by
measured signal. Waveform display is also shown
in following figure
.
Figure 2 Waveform display when closin
g
bandwidth limit
4
Bandwidt
h
suppressio
n
20MH
z
Bandwidth suppression icon
Page 31
Operating Man ual for UT D2000/30 00
1
9
Please press [F2] to set band width limitation to ON for great attenuation of high-frequency component and noise of more than 20MHz contained b
y
measured signals. Waveform display is also show
n
in following figure
.
Figure 2 5Waveform display when openin
g
bandwidth limit
It has to set probe attenuation coefficient i
n
operation menu of channel to coordinate wit
h
attenuation coefficient setup of probe. Probe coefficient in channel menu shall be set to 10× if probe attenuation coefficient is 10:1. It has to analogize to guarantee correct voltage
.
Volt/grid gears of vertical deflection coefficient can
be divided into coarse tuning and fine tuning
.
Please refer to following figure for setup and
vertical gear when using 10:1 probe
.
3. Setup of multiplying factor for probe
Close
bandwidt
h
limit
Probe
attenuatio
n
coefficient
Figure 2 Probe attenuation coefficient shown
in channel men
u
6
Page 32
Operating Manual for UTD2000/3000
4.
2
025
Regulation setup of vertical volt/grid Setup of reverse phase of waveform
Reverse phase of waveform:
㨪
㧙㧙
Volt/grid scope is 2mV/di
v
V/div (or 10V/div), or 1mV/div V/div with step by way of 1 Deflection coefficient shall be altered within current vertical gear scope by less step during fine tuning so as to realize uninterrupted and continuous adjustment of vertical deflection coefficient within all vertical gears
.
5
20
Setup of fine tunin
g
Figure 2 Coarse tuning and fine tuning of vertical deflection coefficient
7
It is to show overturning degree of phase location of displayed signals. Please refer to Figure 2 for waveform without reverse phase. Please refer to
Figure 2 for waveform with reverse phase
.
㧙㧙8
9
Without reverse phase
Figure 2 Setup of reverse phase of vertical
channel (or without reverse phase
)
8
Page 33
Operating Manual for UTD2000/3000
2
1
Figure 2 9 Setup of reverse phase of vertical channel
I. Realization of mathematical calculation
functio
n
Mathematical calculation functions are to show +, -
,
, / as well as FFT calculation results for CH1 and
CH2 channel waveform. Menu is shown as follows:
h
Reverse phase of waveform
Mathematical calculation
Figure 2 Mathematical calculatio
n
1
0
Page 34
Operatin g Manual for UTD2000/3000
Analysis of FFT frequency spectrum
Time domain signal can be converted to frequency
domain signal when using mathematical calculatio
n
of fast Fourier transtable (FFT). Following types of
signals can be observed conveniently by usin
g
FFT
Measure harmonic content and distortion withi
n
measurement system
;
Ɣ
ƔƔShow noise characteristics in DC power
;
A
nalyze vibration
;
2
2
Function Menu
Setup Descriptio
n
Type
Mat
h
,.---hr
Informatio
n
source
CH1
CH
2
To set information source 1 as Ch1 channel waveform
.
To set information source 1 as Ch2 channel waveform
.
,
.
Information source 1 Information source
2
Information source 1 Information source
2
Information source 1 Information source
2
Information source 1 Information source
2
,
.
Ø
r
Operator
Informatio
n
source
2
CH1
CH
2
To set information source
2
as CH1 channel waveform
.
To set information source
2
as CH2 channel waveform
.
Form 2-2: Introduction to mathematical calculatio
n
Ø
r
Page 35
Form 2-3: FFT menu Descriptio
n
FFT operation knack:
Signals with DC component or deviation will brin
g
in error or deviation of FFT waveform component
.
A
C coupling mode can be selected to reduce D
C
component.
A
verage obtainment mode of digital storage oscillograph can be set to reduce random noise and aliasing frequency component of repeated or single pulse event
.
Operating Manual for UTD2000/3000
2
3
Functio
n
Setup
Descriptio
n
Typ
e
FFT
FFT mathematical calculation
Informatio
n
source
CH1 CH
2
To set CH1 and CH2 as calculation waveform
.
Window
Hammin
g
Blackma
n
Rectangle
Set Hanning window functio
n
Set Hamming window functio
n
Set Blackman window functio
n
Set Rectangle window functio
n
Hannin
g
Vertical unit
Vrm s dBVrm
To set vertical unit to Vrms or dBVrms
.
Page 36
Operating Manual for UTD2000/3000
Selection of FFT window
Mathematical storage oscillograph will carry out FFT conversion for time record with limited length when assuming continuous repetition of YT waveform. YT waveform will have the same amplitude in beginning and ending locations for no interrupation. However, high-frequency transient interruption will be caused in connector due to different waveform amplitudes for non-integral cycle of YT waveform. Such effect shall be called as leakage in frequency domain. Therefore, value in forced beginning and ending locations is 0 b
y
multiplying original waveform by a window function to avoid leakage. Please refer to following form for application of window function
.
Form 2-
4
24
FFT window
Characteristic
s
The most appropriate measurement content
Rectangle
The best frequency and the poorest amplitude resolution to be basicall
y
similar to circumstances without window
;
Transient or short impulse with basically equivalent signal level; Sine with equivalent amplitude wit
h
similar frequency; Wide band random noise of wave
spectrum with slow variation
;
Hannin
g
With excellent frequency resolution and poor amplitude resolution by comparin
g
with rectangular window
;
Sine, cycle and narrow band random noise
Hammin
g
Frequency resolution of Hamming window is better than that of Hanning window slightly
.
Transient or s hort impulse with large difference for signal level
;
Blackma
n
The best amplitude resolution and the poorest frequency resolutio
n
Mainly used for single-frequency signal to search
for higher order of harmonic wave
.
Page 37
Operating Manual for UTD2000/3000
2
5
Term expl an atio n: FFT resolution:
Nyquist frequency:
Reference waveform
It is defined as factor of samplin
g
and calculation point. FFT resolution increases when sampling rate decreases for fixed calculatio
n
point quantity
.
Original waveform can not be reset until using sampling rate of at least 2f for waveform with the maximum frequency of “f”. It is also called as Nyquist guideline. “f” refers to Nyquist frequency and “2f” refers to Nyquist sampling rate
.
Please refer to waveform stored in nonvolatile memory (NVM) of digital storage oscillograph or peripheral U disk with name of RefA and RefB as follows. Please operate acc ording to followin
g
steps to call out or close reference waveform
.
5. Please press REF menu key on the front panel for 5.7-screen oscillograp
h
Please press [STORAGE]key and enter into the
second page to select call-out for 7-inc
h
oscillograph (namely, UTD2000L/2000EX series).
2. Please press [RefA] (RefA reference option) and
select information source then select its position b
y
rotating multi-purpose knob in upper part of front panel. 1 10 (or 1 20) can be selected for the position. Press “call-out” to call out waveform which is stored inthe position originally after selecting a certain position such as “1”. Pleas e plug in U disk if waveform is stored in the U d isk
.
There are 2 options for magnetic disk as follows
:
DSO/USB; Please press [F2] key and select USB
.
(The menu can not be activated until plugging in U
disk.) Called out waveform is displayed on the screen for waveform called out and stored from U disk. Please press cancel key (F5) Return to previous menu after waveform call-out
.
3. To press [RefB] (RefB reference option) and
select

Page 38
information source 2 which participates i
n
calculation and method is the same to the secondstep. Please measure and observe related waveform by UTD2000/3000 series digital storage oscillograph during actual application. Please compare current waveform with reference waveform for analysis. Please press REF key to display reference waveform menu. Setup is also shown in following form
.
Form 2-5 Selection of storage positio
n
It is allowed to select from 1 to 20 for internal storage position. Please plug in U disk and press F2 to select USB magnetic disk for external storage instrument. Please refer to STORAGE menu for waveform storage
.
Operating Manual for UTD2000/3000
2
6
Function menu
Setup
Descriptio
n
Storage positio
n
120a
12
0
a refer to positions of
20 groups of waveform
respectively
;
There are 200 groups of waveform positions for storage to USB
;
(Storage position and quantity
vary withmachine model.
)
Magnetic disk
DSO USB
To select internal storage position;
To select external storagepositio
n
(until plugging inU disk)
;
Close
To close called out waveform
;
Callout
To call out selected wavefo rm;
Cancel
Return to previous menu;
Page 39
Operating Man ual for UT D2000/30 00
2
7
Setup of horizontal system
Horizontal control knob Control knob on horizontal surface can be used to alter horizontal scale (or time base) to trigger horizontal position (or trigger position) shown i
n
RAM. Vertical middle point toward horizontal direction on the screen is time reference point for waveform. Horizontal scale can be altered for waveform extension or shrinkage by comparin
g
with screen center. Horizontal position can also alter by comparing with position of waveform trigger point
.
Horizontal position: To adjust horizontal position of channel waveform (including mathematical calculation). Resolution of such control key varies according to time base
.
Horizontal scale: To adjust main time base, namely
,
s/div. Window width can vary with delay scannin
g
time base by rotating horizontal scale knob whe
n
opening extension time base. Please refer to introduction to extension time base for details. (As shown in the following form), horizontal menu ca
n
be displayed for horizontal control key menu
.
Form 2 -
6
Function menu
Setup
Descriptio
n
Main time base
1
.2.
To open main time base
;
To close view window
extensionaccordingtomai
n
time base after opening
;
View window extensio
n
To open extension time base
;
Hold off
To regulate inhibition period
;
üü
üü
üü
üü
üü
Page 40
Mark Description:
Trigger horizontal distance of position b
y
comparing with middle point of view window
.
ķ
ˊĸˊĹˊ
Ļ
ˊ
Mark represents position of current waveform
view window in RAM
Mark position of trigger point in RAM
.
Mark position of trigger point in current
waveform view window
.
Display horizontal time base (or main time
base), namely, s/div
.
ĺ
ˊ
Term explanation:
Y
-T mode:
X-Y mode:
Y axle refers to voltage and X axle
refers to time quantum under this mode
.
X axle refers to CH1 voltage and Y axle
refers to CH2 voltage under this mode
.
Slow scanning mode: Instrument will enter into slow scanning sampling mode when settin
g
horizontal time base control to 100ms/div or less. It
is suggested to set channel coupling to DC whe
n
observing low-frequency signal by applying for slow scanning mode
.
s-div: Time base control can expand or compress
waveform for horizontal scale (or time base) unit if
stopping waveform sampling (by using RUN/STOP
key).
Operatin g Manual for UTD2000/3000
2
8
Figure 2-11 Interface of horizontal system
Page 41
Operating Manual for UTD2000/3000
2
9
View window extension
View window extension is to argument a section of waveform to review image details conveniently
.
Setup of view window extension can not be slower than that of main time base
.
A
s shown in above figure), it can be divided into
2
display areas under extension time base. Original waveform is displaye d for upper part. Sel ected ar ea can move horizontally by rotating horizontal
POSITION knob or increase or decrease by rotating
horizontal SCAL E knob
.
(
Waveform with horizontal extension in originall
y
selected waveform area is shown in lower part. (As shown in above figure,) please pay attention that resolution increases for extension time base b
y
comparing with main time base. Waveform displayed
by the entire lower party corresponds to area selected for upper part so users can rotate horizontal SCALE knob to reduce selected area and increase extension time base, namely, horizontal extensio
n
multiple of waveform
.
ThemodemustbeusedforCH1andCH2atthe same time
.
CH1 voltage will be displayed on X axle and CH
2
voltage
Extensio
n
time base
Horizontal extensio
n
of the waveform
Horizontally extended part of the waveform
Figure 2 1 Screen display under view window extensio
n
2
Page 42
Operatin g Manual for UTD2000/3000
will be displayed on Y axle after selecting X-
Y
display mode
.
Note: Digital storage oscillograph can capture waveform by any sampling speed under normal X-
Y
mode. Similarly, it is allowed to adjust samplin
g
rate, vertical gear of channels and time base gear under X-Y mode. Default sampling rate is 100MS/s under X-Y mode. In general, sampling rate ca
n
reduce to gain excellent Lissajou’s figure. Followin
g
functions do not effect during X-Y display
.
Automatic testing mode
Cursor measurement mode Reference or mathematical calculation waveform View window extension functio
n
Trigger cont
rol
Ƶ Ƶ Ƶ Ƶ Ƶ
30
Figure 2 Waveform display under X-Y mode㧙1
3
Page 43
Operatin g Manual for UTD2000/3000
3
1
Setup of trigger system
Trigger can determine data acquisition time and waveform display for digital storage oscillograph. It can convert instable display to meaningful waveform once correct setup. Digital storage oscillograph will collect sufficient data to map waveform in the left side of trigger point firstly. It will acquire data continuously and synchronousl
y
when waiting for trigger conditions. It also will acquire sufficient data continuously to map waveform in the right side of trigger point. Trigger control area of operation panel of digital storage oscillograph includes trigger level adjustment knob and trigger menu key of [TRIG MENU]
;
5.7-inch oscillograph will set 50% of trigger level i
n
vertical middle point of signal (and realize UTD2000L /2000EX by press [SET TO ZERO] key)
.
It has to press forced trigger key of [FORCE]
.
Trigger level: Trigger level will set signal voltage which corresponds to trigger point
.
50%: To set trigger level to vertical point of trigger signal amplitude
.
FORCE: To produc e 1 trigger signal forcefully for mainly application for normal and single modes shown in trigger modes
.
TRIG MENU: To trigger and set menu key
.
Page 44
Operating Manual for UTD2000/3000
3
2
Trigger control
Trigger mode: Edge, pulse width, video and ALT TRIG
;
Edge trigger: Trigger when edge of trigger signal reaches a certain given level
.
Pulse width trigger: Trigger when pulse width of trigger signal reaches certain set trigger conditions Video trigger: Carry out field or line trigger for standard video signals. (UTD2025CL/UTD2000EXseriesis no such function.
)
A
LT TRIG: It is appropriate to trigger signal without frequency relevance. Various kinds of trigger menus
are also shown as follows
:
Edge trigger
:
Edge trigger mode is to trigger threshold value of input signal edge. That is, to trigger in ascending
/
descending edge of input signal when selecting “edge trigger”
.
Form 2-
7
Page 45
Operating Manual for UTD2000/3000
3
3
Function menu
Setup
Descriptio
n
Type
Edge
Selection of information source
CH1 CH
2
EXT EXT/
5
Municipal power
A
lteratio
n
Set CH1 as trigger signal of info r ma ti on s ou rce
.
et CH2 as trigger signal of inf or m atio n sou r ce
.
Set EXT TRIG INPUT channel as trigge r sig na l of information sourc e
.
Divide outer trigge r so ur c e by 5 to ex pa nd outer trigger level sco pe
.
Set municipal powe r tri gg er
A
LT TR IG of CH1 an d CH2 si gn als ;
.
Gradient
A
scen
d
Descen
d
A
scend/descend
Set ascending edge trigger for signal
;
Set descending edge trigger for signal
;
Set ascending/descending edge trigger of signal
;
Trigger mode
A
utomatic Normal Single
Set waveform acquisition without detecting trigger conditions
;
Set waveform acquisition after meeting trigger conditions
;
Set sampling of a waveform and shutdown after detecting once trigger
;
Trigger couplin
g
ACDC
High-frequenc
y
inhibitio
n
Low-frequenc
y
inhibitio
n
Obstruct DC component of input signal
;
Pass through AC & DC components of input signal
;
Inhibit high-frequency components of more than 80kHz for signal
;
Inhibit low-frequency components of less than 80kHz for signal
Page 46
Operating Manual for UTD2000/3000
Pulse width trigge
r
Trigger time s hall be determined for pulse width trigger according to pulse width. You can capture abnormal pulse by setting pulse width conditions
.
Form 2-8 (the first page
)
34
Function menu
Setup
Descriptio
n
Type
Pulse widt
h
Selection of informatio
n
source
CH1 CH
2
EXT EXT/
5
Municipal power
A
lteratio
n
To set CH1 as trigger signal of information source
.
To set CH2 as trigger signal of information source
.
To set EXT TRIG INPUT channel as trigger signal of information source.
To divide outer trigger source by 5 to expand outer trigger level scope.
To set municipal power for tri gger
.
A
LT TRIG of CH1 and CH2 signals
Pulse widt
h
conditions
More tha
n
Less tha
n
Equal t
o
To trigger when pulse width exceeds set value
;
To trigger when pulse width is less than set value
;
To trigger when pulse width is equivale
Pulse widt
h
setup
To set pulse wi dth of 2 0ns to 1 0s then regulate by multi -purpose knob in
upper part of front panel
.
Next page 1/
2
Enter into next page
.
Page 47
Operating Manual for UTD2000/3000
3
5
Form 2-9 (the second page
)
Function menu
Setup
Descriptio
n
Type
Pulse widt
h
Trigger polarit
y
Positive pulse widt
h
Negative pulse widt
h
Trigger mode
To set positive pulse width as trigger signal
;
To set negative pulse width as trigger signal
;
A
utomatic Normal Single
The system will acquire waveform data automaticall
y
and display scanning base line on the screen without trigger signal input. It will convert to trigger scannin
g
automatically when producing trigger signal
.
To produce trigger scanning if any trigger signal is produced
.
To produce once trigger then stop during trigger signal input
.
Trigger couplin
g
D
C
AC
High-frequenc
y
inhibitio
n
High-frequency inhibitionTo pass through AC & D
C
components of trigger signal
.
To obstruct DC components of trigger signal
.
To obstruct pass through of high-frequency c omponents of signal and only low-frequency components can pass through
.
Previous page 2/
2
Return to previous page
.
Page 48
Operatin g Manual for UTD2000/3000
Video touc
h
To trigger in field or row with NTSC or PAL standard video signals after selecting video trigger for UTD2025CL oscillograph shown in 5.7-inch or 7- inch screen. Trigger coupling is pre-set to DC. P lease refer to following form for trigger menu
.
Form 2-10 Setup of video trigger
3
6
Function menu
Setup
Descriptio
n
Type
Selection of
Pulse widt
h
information source
CH1 CH
2
EXT
To set CH1 as trigger signal of information source
.
To set CH2 as trigger signal of information source
.
To set EXT TRIG INPUT channel as trigger signal of information source
.
To set 5 times of attenuation for outer trigger signal as trigger signal
.
A
LT TRI G of CH1 and CH2 signals
EXT/
5
A
lteratio
n
Standar
d
PA
L
NTS
C
Video signal to be appropriate to PAL system
;
Video signal to be appropriate to NTSC system
;
Synchronizatio
n
A
ll rows
Designated row
Odd f iel
d
Even fiel
d
To set synchronous trigger of video row
.
To set synchronous trigger of designated video row for adjustment by multifunctional knob in upper part of front panel
.
To set synchronous trigger for video odd field
.
To set synchronous trigger for video even field
.
Page 49
Operating Man ual for UT D2000/30 00
3
7
Please refer to Figure 2-14 for screen display whe
n
selecting standard system of PA L and row synchronization as synchronization mode. Please refer to Figure 2-15 for field synchronization of synchronization mode
.
Figure 2 14 Video trigger: Row synchronizatio
n
Figure 2 15 Video trigger: Field synchronizatio
n
A
LT TRIG mode
A
LT TRIGTrigger signal comes from 2 vertical channels durin g ALT TRIG. Suc h trigge r mode can be adopted for synchronous observation of 2 signals without regard to signal frequency. Please refer to following figure for trigger alternation waveform display
.
Page 50
Setup of trigger coupling mode
Please enter into trigger setup menu for setup of trigger coupling mode and gain the most sform synchronization. Please refer to following form for menu of trigger coupling mode
.
Operating Manual for UTD2000/3000
Form 2-11 Trigger alternation menu
Function menu
Setup Descriptio
n
Type
Edge
Set trigger mode to edge
Trigger source
A
lternatio
n
A
LT TRIG of CH1 and
CH
2
Gradient
A
scendin
g
Set trigger gradient to ascending edge
.
Trigger mode
A
utomatic
Set trigger mode as automatic one
Trigger couplin
g
AC
Set trigger couplin
g
modeto AC mode
Figure 2 Observation of 2 signals wit
h
different frequency by
.
16
3
8
Page 51
Form 2-1
2
Operating Manual for UTD2000/3000
3
9
Function menu
Setup
Descriptio
n
Type
Trigger source
Gradient
Trigger mode
Coupling
ACDC
High-frequency inhibitio
n
Low-frequencyinhibitio
n
Set pass-throu gh obstr ucti on of DC compone nts
.
Set pass-through of all components
.
Obstruct pass-through of high-frequency part of signal. Only low­frequency components can pass through
.
Obstruct pass-through of low-frequency part of signal. Only high­frequency components can pass through
.
üü
üü üü
üü
Page 52
Adjustme nt of hold off time
A
djustment of hold off time can be used to observe complex waveform (such as pulse s tring series)
.
Inhibition time refers to waiting period of digital storage oscillograph to start up trigger circuit again
.
Digital storage oscillograph will not trigger until inhibition time ends during inhibition period. For instance
,
as shown in Figure 2-17, the first pulse among a group of pulse series shall be triggered to set inhibition time to width of pulse string
.
Please refer to following form for hold off menu
.
Form 2-1
3
Operating Manual for UTD2000/3000
40
Function menu
Setup
Description
Main time base
1. To open main time base
;
2. To close view window extension according to main t ime base after openin
g
view window extension
.
View window extensio
n
To open extension time base
.
Hold off
To adjust inhibition time
.
üü
üü
üü
üü
Page 53
Operating Manual for UTD2000/3000
4
1
Figure 2—17 Hold off used for synchronization of complex waveform
Operation Description
1. To select edge, trigger source and gradient show
n
in TRIG MENU according to synchronization mode of normal signal then adjust trigger level for stable waveform display as practicable as possible
.
2. To press HORI MENU key to display horizontal menu
.
3. To adjust multi-purpose knob in upper part of panel to alter inhibition time for stable waveform
display as practicable as possible
.
Term explanations
1. Trigger source: To gain trigger from many kinds of
information sources, namely, input channel (CH1
&
CH2), external trigger (EXT & EXT/5) and municipal
power
.
Input channel The most common trigger information source is optional input channel Selected one is used as channel of trigger
information source for normal operation regardless
of input displa
y
External trigger: Such trigger information source
can be used for trigger on the third informatio
n
channel during synchronous data acquisition o
n
channels For instance signals f rom external clock
or circuit t
o
meas
ure can be used as trigger information source
.
EXT and EXT/5 trigger sources shall use external trigger signals
Ƶ

Ƶ

Trigger
Holdoff
Page 54
connected to EXT TRIG interface. Signal can be used for EXT and EXT can be used within signal trigger level scope of -3V to +3V
.
A
utomatic trigger: The system will acquire
waveform data automatically so as to displa
y
scanning base line on the screen without trigger signal input. It will convert to trigger scannin
g
automatically for signal synchronization whe
n
producing trigger signal
.
It is allowed to select EXT/5 for oscillograph of 5.7­inch screen. Outer tri gger shall be divided by 5 so that trigger scope can expand to -15V to +15V for trigger of digital storage oscillograph under large signal
.
Municipal power: Such trigger mode can be used
to observe signal with regard to municipal power
.
For instance, it can provide relationship betwee
n
lighting equipment and power supply equipment so as to gain stable synchronization
.
2 Trigger mode: It is to determine behavior mode of digital storage oscillograph without trigger event
.
This digital storage oscillograph can also provide
3
triggers, namely, automatic, normal and single trigger modes
.
Ƶ
Ƶ
Operating Manual for UTD2000/3000
4
2
Note: Without trigger signal is allowed under
automatic” mode when setting scanning waveform to 5-ms/div or less time base
.
Normal trigger: Digital storage oscillograph ca
n
not acquire waveform until meeting trigger conditions under ordinary trigger mode. Trigger scanning can be produced when producing trigger signal
.
Single trigger: User can press “operation” button
once under single trigger mode for entry and trigger
waiting for digital storage oscillograph. It will sample and display all acquired waveform then stop after digital storage oscillograph detects once trigger
.
Ƶ
Ƶ
Page 55
Operatin g Manual for UTD2000/3000
4
3
3
.
4. Pre-trigger/delay trigger: To trigger data acquired in ahead/after event. Trigger position is often set i
n
the horizontal center on the screen. You ca
n
observe pre-trigger and delay information of 5di
v
(or 6div). You can also rotate horizontal position to regulate horizontal displacement of waveform and
review more pre-trigger details. You can observe waveform in ahead of trigger by observing pre­trigger data. For instance, burring reason can be shown by observation and analysis on pre-trigger data after capturing burring during circuit startup
.
Trigger coupling: Trigger coupling can determ ine category of signal component conveyed to trigger circuit. Coupling types include DC, AC, low­frequency inhibition and high-frequency inhibition
.
Pass through all components of signals under
DC mode
.
Ƶ
Ƶ
Ƶ
Ƶ
Obstruct DC components with signal attenuatio
n
to be less than 400Hz under AC mode
.
Obstruct DC components with attenuation of low-frequency component of less than 80kHz under “low-frequency inhibition” mode
.
A
ttenuate high-frequency components of more
than 80kHz under “high-frequency inhibition” mode
.
Page 56
Operating Manual for UTD2000/3000
Setup of sampling system
As shown in following figure, ACQ UIRE in control area is function key of sampling system
.
Figure 2 18 Function key of sampling system User can use ACQUIRE key to pop out samplin
g
setup menu so as to adjust sampling mode b
y
control button shown in the menu
.
44
Page 57
Operating Manual for UTD2000/3000
4
5
Form 2-14 Sampling men
u
Function menu
Setup
Descriptio
n
Obtainment mode
Samplin
g
Peak-value detectio
n
A
verage
A
ctivate ordinary sampling mode
.
A
ctivate peak-value detection mode
.
Set average sampling mode and display average times
.
A
verage times
2~25
6
Set average times for step by multiples of 2, namely, 2, 4, 8, 16
,
32, 64, 128 and 256. As shown in Figure 2-18, average times ca
n
alter by pressing left multi-purpose knob
.
Sampling mode
Real-time Equivalent
Set sampling mode to real-time sampling mode
.
Set sampling mode to equivalent sampling mode
.
Fast acquisitio
n
O
N
A
cquire by higher screen refreshing rate to reflect dynamic
effect of waveform excell ently
.
Close fast acquisition
.
Page 58
It has to observe waveform display variation b
y
altering obtainment mode. Please refer to Figure 2­19 and 2-20 for sampled waveform display if signal contains large noise when adopting 32 times of average mode or without adopting average mode
.
Operating Manual for UTD2000/3000
4
6
Figure 2-20 Waveform by adopting 32 times of average mode
Figure 2 Waveform without adoptin
g
1
9
Note
Please select real-time sampling mode for observation
of single signal
.
2. Please select equivalent sampling mode for observatio
n
of high-frequency periodic signal
.
3. Please select peak-value detection mode to avoid mixed signal envelope. Please select average sampling mode wit
h
step of multiples of 2 for average value to reduce random noise of displayed signal. Average times can be selected
form 2 to 256.
1
Page 59
Term explanatio
n
Real-time sampling: It refers to data required for once acquisition
.
Equivalent sampling: It refers to repeated sampling mode. It can facilitate careful observation of repeated periodic signals. It can be adopted to gain horizontal resolution which is higher than real-time sampling resolution greatly
.
Sampling mode: It can be classified into equivalent and real-time modes
.
Peak-value detection mode
:
Digital storage oscillograph will find out the maximum and minimum input signals within ev ery sampling interval for waveform display by using such values under such obtainment mode. It can also obtain and display narrow impulse or they may be omitted for larger noise under such mode. Average mode is as follows: Digital storage oscillograph may obtain several waveform and calculate average value then display final waveform under this obtainment mode. Such mode can be adopted to reduce random noise
.
Operating Manual for UTD2000/3000
4
7
Page 60
Operating Manual for UTD2000/3000
Setup of display system
A
s shown in following figure, DISPLAY in control area is function k ey of display sy stem
Figure 2 Function key of sampl ing system Please press DISPLAY button to pop out setup menu shown in following figure. Display mode can be adjusted by pressing menu control button
.
㧙
4
8
Page 61
Operatin g Manual for UTD2000/3000
4
9
Key points
:
Display type: To fill in blank between adjacent sampling points during vector display. Only sampling points will be displayed for point display
.
Function menu Setup
Description
s
Display type
Vector point
Only to display sampling points by wiring
.
Format
YT X
Y
Operation mode of digital storage oscillograp
h
X-Y display mode; X input for CH1 and Y input for CH2
;
Continuous
Close 1s 2s 5s Infinite
Real-time updating of screen waveform
;
To update after keeping waveform data shown in the screen for 1s
;
To update after keeping waveform data shown in the screen for 2s
;
To update after keeping waveform data shown in the screen for 5s
;
To always keep display of original waveform data shown in the screen until this function is inactivated if new data is added into displa
y
continuously
;
Waveform brightness
1% - 100%
To set waveform brightness (without this function for UTD2000C/3000C
)
Form 2-15 Display menu
Page 62
Storage and callout
A
s shown in following figure, STORAGE in MENU
control area is f unction k ey of storage system
.
Operation steps Please press STORAGE Enter into type menus
.
There are 3 types, namely, waveform, setup and bitmap
.
Please refer to Figure 2-16 for selection of type
waveform and entry into following waveform storage menu. Please call out REF (reference
waveform shown in the second page) from Section
2 of the chapter after saving waveform
.
Form 2-16 Waveform storage menu (the first page)
2.
Operating Manual for UTD2000/3000
50
Figure 2 22 Function key (storage) for sampling system STORAGE key can be used to display storage setup menu. You can save oscillograph waveform or setup status to internal storage area or U disk to call out saved waveform by Ref
A
(or RefB) or call out setup status by STORAGE key. Waveform display area of oscillograph can be saved to DSO contents in U disk in a form of bitmap. Saved bitmap can be read by PC computer
.
Page 63
Operatin g Manual for UTD2000/3000
5
1
Form 2-16
Function menu
Setup
Descriptio
n
Type
Wav e form
Select waveform storage and callout menu
Information source
CH1 CH
2
Select waveform from CH1 channel
.
Select waveform from CH2 channel
.
Storage positio
n
120a
12
0
a refer to positions to store 20 groups of waveform
.
There are 200 groups of waveform positions for storage to USB
.
(Storage position and quantity vary with machine model.
)
Save
Save wavef orm
Next page 1/
2
Enter into next page
.
Figure 2-23 Waveform storage
Figure 2-24 Waveform storage by U disk
üü
üü
Page 64
Operatin g Manual for UTD2000/3000
Form 2-17 Storage menu (the second page
)
5
2
Function menu
Setup
Descriptio
n
Magnetic disk
DSO USB
To select internal RAM of digital storage oscillograph the
n
select external U disk. (This function can not be used until plugging in U disk.
)
Storage dept
h
Ordinar
y
Long storage
To set storage depth to ordinary; (Callout only can be done i
n
REF area when saving data to U disk.
)
To set storage depth to long storage. (Note: This function ca
n
not be activated until plugging in U disk. Callout can not be done until using loading function of computer communicatio
n
software or waveform analysis load when saving data to U disk.
)
Previous page 2/
2
Return to previous page
.
üü
Page 65
2
A
s shown in Form 2-18, please select setup and enter into setup storage menu
.
Figure 2-25 Storage setup
Form 2-18 Setup of storage menu
5
3
Operating Manual for UTD2000/3000
Function menu
Setu
p
Description
Setup
To select panel setup menu
.
Setup of (storage position
)
12
0
Operation setup to save 20 groups of panels: To select by multi
-
purpose knob in upper part of front panel. (Position and quantit
y
vary with machine model
Save
Storage setup
Recall
Callout setup
Page 66
54
3
.
A
s shown in Form 2-19, please select bitmap for entry into bitmap storage menu
.
Remark: This function can not be called out until plugging in U disk
.
Form 2-19 Bitmap storage menu
Operating Manual for UTD2000/3000
Function menu
Setup
Descriptio
n
Bitmap
Select bitmap menu
Storage positio
n
12
0

200 bitmap data can be saved by selecting multi-purpose knob in
upper part of front panel. (It is 1 to 10 for UTD2000C series.
)
Save
Save bitmap data
üü
üü
Figure 2-26 Bitmap storage
Page 67
Setup of auxiliary functions
A
s shown in following figure, UTILITY in MENU
control area is function key
.
Figure 2 27 Function key (and function) of samplin
g
User can press UTILITY key to pop out setup menu of auxiliary system function
.
Form 2-20 (the first page
)
5
5
Operating Man ual for UT D2000/30 00
Function menu
Setup
Descriptio
n
Self-correctio
n
Cancel
To carry out self-correction
.
To cancel self-correction and return to previous page
.
Executio
n
By detectio
n
Please refer to Form 2-2
3
To set waveform Pass/Fail
.
(Only UTD2000E/3000E
,
UTD2025C/3025C and UTD2025CL have this function.
)
Waveform recordin
g
Please refer to Form 2-2
2
Set waveform recordin
g
operation
.
Language Multilanguage
Select interface language
.
Page 68
5
6
Operating Manual for UTD2000/3000
Form 2-21-1 (the second page
)
Function menu
Function menu
Setup
Setup
Descriptio
n
Descriptio
n
Fast correctio
n
ON/OFF
A
ctivate/inactivate fast correction function of channels within scope
of 2mV/div to 10 mV/div. (Only UTD2025C/3025C has this function.
)
Interface style
The first style The secondstyle The third style Thefourthstyle
Set 4 interface styles for digital storage oscillograph
.
Gridding brightness
1% - 100%
A
djust gridding brightness of the screen by multi-purpose knob
.
UTD2000C/3000C also has no this function
.
Next page 2/
3
Enter into next page
.
Form 2-21-2 (the third page
)
üü
System informatio
n
Display information of current oscillograph system
.
Frequency meter
A
ctivate/inactivate function of frequency meter
.
The first page 3/
3
Return to the first page
.
Page 69
Form 2-22 Waveform recording menu
5
7
Operatin g Manual for UTD2000/3000
Function menu
Setup Descriptio
n
Information source
CH1 CH
2
CH1+CH
2
Select CH1 as recording signal source
.
Select CH2 as recording signal source
.
Select ct CH1+CH2 as recording signal source
.
Operatio
n
Record by pressing recording key for synchronous display of recorded quantity in lower part of the screen
.
1.Playback key
;
2. Press the key f or playback and display of current playback screen No. in lower right corner of screen. Playback can be stopped by rotating multi-purpose knob controller in upper part of panel. However, waveform of a certain screen can be selected for repeated playback by continuing to rotate it
.
3Press to stop then press for playback of all items
.
ƽ
Ƶ
Ƶ
Stop recordin
g
Save
1~20
0
Save newly recorded waveform for U disk plug-in. Storage position can be selected by multi-purpose knob
.
Recall
1~20
0
Call out recorded wavef orm from U disk then sele ct recor din
g
position by multi- purp ose kno b
.
Retur
n
Return to previous level of menu
.
üü
Page 70
5
8
Operating Manual for UTD2000/3000
Form 2-23 By detectio
n
Function menu
Setup
Descriptio
n
Detection status
Ope
n
Close
Open/close information of waveform pass/fail times
Information source
CH1 CH
2
MAT
H
Ref
A
RefB
Select CH1 as detected signal source
.
Select CH2 as detected signal source
.
Select MATH as detected signal source
.
Select RefA as detected signal source
.
Select RefB as detected signal source
.
Output conditions
Pass Pass/Halt Fail/Halt Fail
Output concurrency prompt when detecting waveform pass
;
Output and pause when detecting waveform pass
;
Output and pause when detecting waveform fail
;
Output concurrency when detecting waveform fail
;
Tem p la te set up
Refer to Form 2-24 for details
Enter into template setup menu
.
Retur
n
Return to previous menu
.
Page 71
5
9
Key note:Key note
:
Self-correction: Self-correction procedure is to correct measurement error caused by digital storage oscillograph due to ambient variation. You can operate the procedure according to demand
.
Please open power of digital storage oscillograp
h
then wait for instrument pre-heating for 30 minutes for more correct calibration. Please press UTILIT
Y
key (as an auxiliary function) for operatio
n
according to prompt shown in the screen
.
Language selection: There are many kinds of languages for UTD2000/3000 series digital storage oscillograph. Please press UTILITY me
n
u butto
n
and select appropriate language
.
Operatin g Manual for UTD2000/3000
Form 2-24 Template s etup
Function menu
Setup
Descriptio
n
Template output
Set horizontal and vertical Pass/Fail tolerance scope accordingtoadjustment
Horizontal 1-200Pixel
Set horizontal tolerance scope
Vertical
1-100Pixel
Setvertical tolerance scope
Retur
n
Return to Pass detectio
n
men
u
Page 72
60
Automatic measurement
A
s shown in following figure, MEASURE is automatic measurement function key. You can be familiar with powerful automatic measurement function of UTD2000/3000 series digital storage oscillograph gradually by following introduction
.
Application example
Measurement menu of such series of oscillograp
h
can be used to measure 28 kinds of waveform parameters. User can press MEASURE key Enter
into parameter measurement display menu firstly.
The menu includes 5 areas to display measurement value at the same time. They correspond to functio
n
keys of F1 to F5 res pectively. Please press corresponding F keys Enter into measurement
category selection menu if measurement category
shall be selected for any area. Measurement category selection menu can be classified into voltage and time kinds Enter into voltage or time measurement category respectively. Please press F1 to F5 to select measurement category the
n
return to parameter measurement display menu. In addition,user ca
n
Operating Manual for UTD2000/3000
Figure 2-28 Function key (automatic measurement
)
of sampling system
Page 73
press F5 to select “All parameters” to dis play all voltage/time measurement parameters. User ca
n
press F2 to select channel to measure. (Channel is ineffective until opening.) User can press F1 Retur
n
to parameter measurement display menu if not altering current measurement category. The first example: Steps are as follows if displayin
g
measured peak value of CH2 channel in F1 area
.
1. To press F1 key Enter into measurement category selection menu
;
2. To press F2 to select CH2
;
3. To press F3 to select voltage
;
4. To press F5 (shown in the second page, 2/4) to see that F3 position is “peak value
5. To press F3 to select “peak value” and return to parameter measurement display menu automatically
;
Peak value is display ed in F1 area on the homepage of measurement menu
.
The secondexample: Setup of delay measurement:
Delay measurement function is used for interval between ascending edges of 2 information sources
,
namely, interval from ascending edge of the first cycle of a certain information source to that of the first cycle of additional information source
.
Measurement steps are as follows
:
1. To refer to above example and select area displayed by delay measurement value after entering into measurement menu (shown in 3/
3
page of time category)
;
2. To select reference information source of CH1 then select delay information source of CH2
;
3. To select reference information source of then select delay information source of CH2
;
4. To press F5 to confirm that delay measurement is displayed in selected area
.
CH
1
6
1
Operatin g Manual for UTD2000/3000
Page 74
Automatic measurement of voltage parameters
Voltage parameters which can be measured b
y
UTD2000/3000 series oscillograph automaticall
y
are as follows
:
Peak value (Vpp): Voltage from the highest waveform point to the lowest waveform point
;
The maximum value (Vmax): Voltage from the highest waveform point to GND (ground)
;
The minimum value (Vmi n): Voltage from the lowest waveform point to GND (ground)
;
Middle value (Vmid): Half of sum of waveform top and bottom voltage
;
A
mplitude (Vamp): Voltage from waveform top to
bottom
;
Middle value (Vmid): Half of sum of waveform top and bottom voltage
;
Top value (Vtop): Voltage from flat waveform top t
o
GND (ground)
;
Bottom value (Vbase): Voltage from waveform
bottom to GND (ground);
Overshoot: Ratio of difference of the maximum waveform value and top value and amplitude
;
Preshoot: Ratio of difference of the minimum
waveform value and bottom value and amplitude;
A
verage value: Average amplitude of signals withi
n
a cycle
;
Root-mean-square (Vrms): It is effective value
.
Root-mean-square refers to DC voltage of converted energy within 1 cycle according to A
C
signal to correspond to equivalent energy
.
Operating Manual for UTD2000/3000
6
2
Page 75
Automatic measurement of time parameters
UTD2000/3000 series oscillograph can measure signal frequency, cycle, ascending time
,
descending time, positive pulse width, negative pulse width, delay (of 9 combinations), positive dut
y
cycle, negative duty cycl e and 10 time parameters automatically. Definitions of such time parameters are as follows
:
RiseTime: It refers to period when waveform amplitude increases from 10% to 90%
;
FallTime: It refers to period when waveform amplitude decreases from 90% to 10%
;
Positive pulse width (+Wi dth): It refers to pulse width of positive pulse under 50% of amplitude
;
Negative pulse width(-Width): It refers to pulse width of negative pulse under 50% of amplitude
;
Delay (ascending edge): It refers to delay from ascending edge to ascending edge
;
Delay (descending edge): It refers to delay from
descending edge to descending edge
;
Positive duty cycle (+Duty): It refers to ratio of positive pulse width and cycle
;
Negative duty cycle (-Duty): It refers to ratio of negative pulse width and cycle
;
Operation Description: Please press MEASURE
key so that the screen can display 5 measurement
values. As shown in Form 2-25, user can press any
key of F1 to F5 so that the screen enters into
me
asurement selection menu
.
Measurement menu
Operatin g Manual for UTD2000/3000
6
3
Page 76
Operating Manual for UTD2000/3000
64
Form 2-2
5
Function menu
Setup
Descriptio
n
Retur
n
Return to parameter measurement display menu
.
Information source
CH1
Select channel of measurement parameters
.
CH
2
Select channel of measurement parameters
.
Voltage kin
d
Enter into voltage parameter menu
.
Time kin
d
Enter into time parameter menu
.
A
ll parameters
Display/close all measurement parameters
.
Form 2-2
6
Function/measurement item
Descriptio
n
Retur
n
Return to menu shown in Form 2-25
.
Preshoot
Return to parameter measurement display menu and substitute original position parameter after selection
.
A
mplitude
he same as above
Overshoot
The same as above
Next page (1/4
)
Page turnin
g
Of which: Menu included by “voltage kind” is shown i n Form 2-26 to 2-29
.
Page 77
6
5
Form 2-27
Function/measurement item
Function/measurement item
Descriptio
n
Descriptio
n
Operating Man ual for UT D2000/30 00
Previous page
Return to previous page
.
A
verage value
Return to parameter measurement display menu and substitute original position parameter after selection
.
Peak value
he same as above
he same as above
Root-mean-square Next page (2/4
)
Page turnin
g
Form 2-2
8
Previous page Top value
Return to previous page
.
Return to parameter measurement display menu and substitute original position parameter after selection
.
Bottom value
The same as above The same as above
Middle value
Next page (3/4
)
Page turnin
g
Page 78
6
6
Operatin g Manual for UTD2000/3000
Function/measurement item
Function/measurement item
Descriptio
n
Descriptio
n
Previous page
Retur
n
Return to previous page
.
Return to menu shown in Form 2-25
.
The maxim um value
Frequenc
y
Return to parameter measurement display menu and substitute original position parameter after selection
.
Return to parameter measurement display menu and substitute original position parameter after selection
.
The minimum value
Cycle
The same as above
The same as above
Next page (4/4
)
Next page (1/3
)
Return to the first page (shown in Form 2-26
)
Page turnin
g
Form 2-2
9
Form 2-3
0
üü üü
Menu included by “time category” is shown in Form 2-30 to Form 2-32
.
The same as above
RiseTime
Page 79
6
7
Operatin g Manual for UTD2000/3000
Function/measurement item
Function/measurement item
Descriptio
n
Descriptio
n
Previous page
Previous page
Return to previous page
.
Return to previous page
.
Frequenc
y
Dela
y
Return to parameter measurement display menu and substitute original position parameter after selection
.
Enter into delay option menu (shown in Form 2-32a) after selection
.
Cycle
Positive duty cycle
The same as above
Return to parameter measurement display menu and substitute original position parameter after selection
.
Next page (2/3
)
The first page (3/3
)
Page turnin
g
Return to the first page (shown in Form 2-30)
.
Form 2-31
Form 2-3
2
The same as above
The same as above
RiseTime
Negative duty cycle
Page 80
Cursor measurement
User can press CURSOR button to displa
y
measurement cursor and cursor menu then use multi-purpose knob controller to alter cursor
position. As shown i n following figure, CURSOR in
MENU control area is cursor measurement functio
n
key
.
Figure 2-29 Function key (or cursor) of samplin
g
system
6
8
Operating Manual for UTD2000/3000
Return to parameter measurement display men
u
and substitute original position parameter after selection
.
Form 2-32a
Functio
n
Setup
Descriptio
n
Channel
Channel
CH1/CH
2
/MAT
H
CH1/CH
2
/MAT
H
Select measured channel
.
üü üü
üü üü
Enter
Page 81
User can move cursor for measurement under CURSOR mode. There are 3 modes, namely, voltage
,
time and tracing. V can be measured by pressin
g
SELECT and COARSE keys on the panel and multi­purpose knob controller to adjust positions of
2
cursors respectively during voltage measurement
.
Similarly, it can b e measured if selecting time. Cursor of digital storage oscillograph can also trace signal variation automatically under tracing mode wit
h
waveform display
.
Remark: For 5.7-inch oscillograp
h
1. Effect of SELECT key is to select the cursor
.
2. Effect of COARSE key is to adjust movement speed of cursor
.
For 7-inch oscillograph
:
User can press multi-function knob key to select the cursor. Movement speed of cursor can also alter according to rotation speed
.
1. Voltage/time measurement mode: The first or
secondcursor will appear at the same time. Cursor position on the screen can be adjusted by multi­purpose knob controller. User can press SELECT ke
y
(or multi-function knob key) to select cursor to adjust
.
Displayed reading is also time or voltage between
2
cursors
.
2. Tracing mode:Horizontal and vertical cursors ca
n
intersect to become a crossing cursor. Crossin
g
cursor can position on the waveform automatically to
adjust horizontal position of crossing cursor on the waveform by rotating multi-purpose knob controller
.
Digital storage oscillograph can also displa
y
coordinate of cursor point at the same time
.
3. Measurement value will be displayed in upper right corner of the screen automatically whe
n
opening cursor function
.
Operating Manual for UTD2000/3000
6
9
Page 82
Usage of operation ke
y
RUN/STOP key is in upper right corner of front panel of digital storage oscillograph. User ca
n
press the key so that green lamp is on to show operation status. Red lamp can be on after pressin
g
the key to show stop status
.
Figure 2-30 Operation/stop ke
y
A
s shown in above figure, automatic setup is to
simplify operation. Digital storage oscillograph ca
n
adjust vertical deflection coefficient and horizontal
time base gear automatically for stable waveform display on the screen according to waveform amplitude and frequency when user presses AUTO key. System setup is shown in following form durin
g
automatic setup
:
Automatic setup
Operatin g Manual for UTD2000/3000
70
Page 83
RUN/STOP: To acquire waveform or stop
acquisition continuously;
The key can be pressed for switchover of waveform
sampling between operation and stop. User ca
n
press the key under operation status so that green
lamp is on. “Auto” can be displayed in upper part of
the screen. Old digital storage oscillograph ca
n
acquire waveform continuously while red lamp is o
n
by pressing the key under stop status. Digital storage oscillograph can stop acquisition whe
n
displaying “Stop” in upper part of the screen
.
Operatin g Manual for UTD2000/3000
7
1
Functio
n
Setup
Obtainment mode Samplin
g
Display format
Set to YT
Horizontal positio
n
A
utomatic adjustment
A
djustment according to signal
frequenc
y
Second/gri
d
Trigger couplin
g
AC
Hold off
The minimum value
Trigger level
Set to 50%
A
utomatic
Trigger mode
Trigger source
Set to CH1; Set to C H2 whe
n
exerting on signal for it if there is no signal for CH1
;
Trigger gradient
A
scendin
g
Trigger type
Edge
A
llVertical band widt
h
Volt /g ri
d
A
djustment according to signal
amplitude
Page 84
Operating Manual for UTD2000/3000
7
2
Chapter 3: Applicatio n Example
The first example: Measurement of simple signal
It has to observe a unknown signal in the circuit for quick display and measurement of signal frequenc
y
and peak value
.
2
1. Please operate according to following steps for quick signal display
:
1. Please operate according to following steps for quick signal display
:
To set probe menu attenuation coefficient to 10*
and set switch on the probe to 10*
.
To connect CH1 probe to measured point of
circuit
.
To press AUTO button
.
Digital storage oscillograph can set automaticall
y
for the best waveform display.You can al so adjust vertical or horizontal gear on above basis so that waveform display can meet related requirements
.
A
utomatic measurement of signal voltage and
time parameters
Digital storage oscillogra
ph can measure most of display signals automatically. Please operate according to following steps for measurement of signal frequency and peak value
.
To press MEASURE key to display automatic
measurement menu
;
To press F1 to enter into measurement menu
category for selectio
n
To press F3 to select voltage categor
y
To press F5 to turn to 2/4 page then press F3 to
select measurement type of peak value
To press F2 to enter into measurement menu category for selection then press F to select time categor
y
ķ
ķ
ĸ
 Ĺ ĺ
Ļ
ļ
To press F2 to select measurement type of
frequency.
Page 85
Operating Manual for UTD2000/3000
7
3
The secondexample: Delay observation of sine
wave signal by circuit
The same as above example, please set
attenuation coefficient of probe and digital storage
oscillograph to 10*. Please connect CH1 channel
of digital storage oscillograph to circuit signal input
terminal and connect CH2 channel to output terminal
.
Operation steps
:
1. Signal display of CH1 and CH2 channels
:
To press AUTO button
To continue to adjust horizontal or vertical gear so that waveform display can meet your test requirements
.
To press CH1 key to select CH1 then rotate vertical position knob to adjust vertical positio
n
of CH waveform
To press CH2 key to select CH2 then adjust vertical position of CH2 waveform to facilitate
ķ ĸ
Ĺ

ĺ
Figure 3-1 Automatic measurement
Page 86
Operating Manual for UTD2000/3000
74
comparison without waveform overlapping for the first and second channels
.
2. To measure delay of sine signal caused b
y
circuit then observe waveform variation.
ķĸDelay of automatic measurement channel
To press MEASURE button to display automatic measurement menu
;
To press F1 key to enter into measurement menu category for selection
;
To press F4 key to enter into time measurement parameter list
;
To press F5 key twice to enter into 3/3 page
;
To press F2 key to select delay measurement
;
To press F1 key to select CH1 then press F2 key to select CH2 then press F5 as enter ke
y
You can see delay value from “CH1-CH2 delay” i
n
F1 area
.
Observation of waveform variation (shown i
n
following figure
)
Figure 3-2: Waveform dela
y
Page 87
Operating Manual for UTD2000/3000
7
5
The third example: Capture of single signal
A
dvantages and characteristics of digital storage oscillograph include convenient capture of aperiodic signals such as pulse and burring, etc. It has to know this signal to set trigger level and trigger edge to capture a signal signal. For instance: Trigger level shall be set to about 2V and set trigger edge to ascending edge trigger if pulse is a logic signal of TTL level. User can observe it b
y
automatic or ordinary trigger mode so as to determine trigger level and trigger edge for unknown signals
.
Operation steps are as follows
:
1. To set attenuation coefficient of probe and CH1 channel like the previous example
;
2. To set triggers
;
To press MENU button in trigger control area to
display trigger setup men
u
To set trigger type to edge, trigger source to CH1
,
gradient to ascend, trigger mode of single mode and trigger coupling to AC by using menu operation keys of F1 to F5 respectively under this menu
.
To adjust horizontal time base and vertical gear to
appropriate scope
To rotate TRIGGER LEVEL knob to adjust
appropriate trigger level
To press RUN/STOP execution button to wait for
signals which meet trigger conditions. Once
sampling can be done for display on the screen if a
certain si
gnal reaches set trigger level. Occidental
events can be captured easily by utilizing t he function. For instance: Sudden burring with large
amplitude Please set trigger level to be higher
than normal signal slightly then press
RUN STOP button for waiting The machine will
trigger automatically and record waveform before and after trigger during burring. Horizontal POSITION knob in horizontal
ķ
ĸ
Ĺ
ĺ
Ļ

Page 88
Operating Manual for UTD2000/3000
7
6
control area on the panel can be rotated to alter horizontal trigger position so as to gain negative delay trigger with different length to facilitate observation of waveform before burring
.
Figure 3-3 Single signal
You can adjust setup of digital storage oscillograph
for overlapping of random noise to tested signal so
as to filter or reduce noise and avoid disturbance to
ontology signal during measurement. (Please refer
to following figure for waveform.
)
Figure 3-4 To reduce random noise of signals Operation steps are as follows
:
1. As shown in previous example, please set attenuation coefficient of probe and CH1 channel
.
The fourth example: To reduce random noise of
the si gnal
Page 89
7
7
2. To connect signals for stable waveform display o
n
digital storage oscillograph
.
Please refer to previous example for operation
.
Please refer to previous chapter for adjustment of horizontal time base and vertical gear
.
3. To improve trigger by setting trigger coupling
.
(1) To press MENU button in trigger area to displa
y
trigger setup menu
.
(2) To set trigger coupling to low-frequenc
y
inhibition or high-frequency inhibition. Low­frequency inhibition is to set a high-pass filter to filter low-frequency signal component of less tha
n
80kHz and allow pass through of high-frequenc
y
signal component. High- frequency inhibition is to set a low-pass filter to filter high-frequency signal component of more than 80kHz and allow pass through of low-frequency signal component. It ca
n
be set to inhibit low-frequency or high-frequenc
y
noise respectively to gain stable trigger
4. To reduce display noise by setting samplin
g
modes
.
To adopt average sampling mode to eliminate display of random noise to thinner waveform and facilitate observation & measurement if waveform is thickened by overlapping random noise to measured signal. Signal details can be observed
convenientl
y by reduced random noise after gainin
g
average value. Detailed operation is as follows To press ACQUIRE button in panel menu area to display sampling setup menu To press F1 key as menu operation key to set obtainment mode to average status. (As shown in following figure,
)
please adjust average times by pressing F2 key as
menu operation key with step from 2 to 256 b
y
multiples of 2 so that waveform display can meet observation & test equirements
.
ķ
Operating Manual for UTD2000/3000
Page 90
7
8
Figure 3-5 Inhibited signal noise
To reduce display noise by reducing wav eform
brightness
.
Note: Average sampling mode can be used to slow down updating speed for waveform display as normal phenomenon
.
This digital storage oscillograph can measure 2
8
kinds of waveform parameters automatically. All automatic measurement parameters can be measured by cursor. Waveform time and voltage can be measured quickly by using cursor
.
Please operate according to following steps to measure 1 staircase voltage of staircase signal
:
1. To press CURSOR key to display cursor measurement menu
.
2. To press F1 key as menu operation key to set cursor type to voltage
.
3. To rotate multi-purpose knob controller to place the first cursor to 1 staircase of staircase signal
.
4. To press SELECT or PUSHSELECT to select the cursor then rotate multi-purpose knob controller to place the secondcursor to another staircase of staircase signal
.
ĸ
The fifth example: Application curso
r
measurement
To measure 1 staircase voltage of staircase signal
Operating Manual for UTD2000/3000
Page 91
7
9
Ƹ
V namely pressure difference can be show
n
in cursor menu automaticall
y
Please refer to following figure for details
.
Note: Please set cursor type to time if onlyin
g
adopting the secondstep as above for time measurement by cursor
.
Figure 3-6 Voltage difference of signal by cursor measurement
To review phase location difference of signals of
2
channels
.
Example: Phase location variation caused by 1 circuit for tested signals. Please connect digital storage oscillograph to cir cuit to monitor input/output signal of circuits. Please operate according to following steps to review input/output terminal of circuits by way of X-Y coordinate diagram
.
1. To set probe menu attenuation coefficient to 10* and set switch on the probe to 10*
.
2. To connect CH1 probe to network input the
n
connect Ch2 probe to network output
.
3. To press CH1 and CH2 menu keys and open
2
channels if they have not been displayed
.
4. To press AUTO button
.
5. To adjust vertical scale knob for approximate equivalent display amplitude of 2 loops of signals
.
The sixth example: Application of X-Y function
Operating Man ual for UT D2000/30 00
Page 92
80
6. To press DISPLAY menu key to call out displa
y
control menu
.
7. To press F2 to select X-Y. Digital storage oscillograph will display input/output characteristics of circuit under Lissa Jous's figure mode
.
8. To adjust vertical scale and vertical position knob to reach the best effect for waveform
.
9. (As shown in following figure), please observe and calculate phase location difference by elliptic oscillograph figure method
.
VLQș $% RU &' 2I ZKLFK ³ș ´ UHIHUV WR SKDVH
location difference angle between channels. Please
refer to above figure for definition of A, B, C and D.
Therefore,phase location difference angle is calculated as follows =±arcsin A B or
±arcsin C D Calculated phase locatio
n
difference angle shall be within Q uadrant I or IV
(namely, (0~ ) or ( ~ ) if main axle of ellipse
is within Quadrant I or III. Calculated value is within
( ~ ) or ( ~ ) if main axle of ellipse is withi
n
Quadrant II or IV. Additionally, frequency and phase location relationship between 2 signals can be calculated according to figure if frequency or phase location difference of 2 measured signals is integral multiple
.


Operating Manual for UTD2000/3000
Signal must be in the horizontal center
Figure 3-
7
3
2
2
2
2
3
2
Page 93
8
1
10.Form of X Y phase location difference The seventh example: Trigger of video signal
Trigger o f video fiel
d
Video trigger function can be used for all modes except UTD2000L/2000EX series. (Of which
:
UTD2025CL has video trigger function.
)
It has to adopt video trigger to gain stable video output signal display for observation of a certai
n
video circuit.
Please operate according to following steps for trigger in video field
.
1. To press[ MENU ]button in trigger control area to
display trigger menu
.
2. To press[ F1] key as menu operation key to select video type
.
3. To press[ F2] key as menu operation key to set trigger source to CH1
.
4. To press[ F3] key as menu operation key to select video standard of P
A
L
.
Operating Manual for UTD2000/3000
Frequenc
y
ratio of signal
1:1
Phase location difference
4
5
°
9
0
°
18
0
°
27
0
°
36
0
°
Page 94
Operating Manual for UTD2000/3000
8
2
5. To press [F4] key as menu operation key to select synchronization of odd field or even field
.
6. To adjust horizontal time base by rotatin
g
horizontal SCALE knob in horizontal control area so as to gain clear waveform display
.
Figure 3-8 Trigger of video field
Please operate according to following steps for trigger in video row
:
1. To press [MENU] button in trigger control area to
display trigger area
.
2. To press[ F1] key as menu operation key to select
video.
3. To press [F2] key as menu operation key to set trigger source to Ch1
.
4. To press[F3] key as menu operation key to select video standar
d
of PAL
.
5. To press[ F4] key as menu operation key to select synchronization to designated row
.
6. To rotate multi-purpose knob controller for trigger in any row
.
7. To adjust horizontal time base by rotatin
g
horizontal SCALE knob in horizontal control area so as to gain clear waveform display
.
Trigger o f video row
Page 95
Operating Manual for UTD2000/3000
8
3
Figure 3-9 Trigger of video row
It has to check if input signal is within rule scope or not. Fail refers to out of scope and Pass refers to within the scope. Pass/Fail signal can be shown b
y
output port
.
1. To press[UTILITY] button then press[F2]Enter into
Pass/Fail menu
.
2. Setup of information source: Enter into Pass/Fail menu then press[F2] to select information source
.
3. Template setup: To press[F4] Enter into template setup menu (Template). To press[F2] and[F3] show
n
in Template me nu to set horizo ntal and ver tic al tolerance scope by multi-function knob, (namely, 1­200Pixel toward ho
rizontal direction and 1-100Pixel
toward vertical direction.) Please press F4 finall
y
Return to Pass/Fail menu.
4. Setup of output conditions: To press[F3] to set output conditions shown in Pass/Fail menu
.
5. To begin detection: To press[F1] to ope
n
detection status shown in Pass/Fail menu
.
The eighth example: Pass/Fail detection
of the rear panel
Rear panel of Pass / Fail output
Pass FailInterface Schematic
Page 96
Operating Man ual for UT D2000/30 00
84
The ninth example: Usage o f updatin
g
procedure of U-disk
Updating procedure of U-disk can be used more conveniently and flexibly. Please operate according to following step s if using this fun cti on
:
1. To download procedure documents to update from Internet then save them in U-disk. (It is suggested to use U-disk suggested by us.
)
2. To close oscillograph and insert U-disk then ope
n
oscillogr aph
.
3. To turn to confirmation interface of whether to update it or not if there is only 1 procedure document in U-disk. Please press[F5] for exit or press F1 to e xit procedure updating. Document selection interace will appear if there are at least 2 procedure documents i
n
U-disk. Please press[F5] key for confirmation (or press[F1] to exit procedure updating). Please turn to confirmation interface of whether to update it or not
then press[F5] for updating)
.
4. To wait for updating for prompt of successful updating. Please power off oscillograph then open it to fulfill updating pr oce dure
.
Note:
1. Please wait for updating for several minutes
.
2. It is not allowed to power off oscillograph or plug
out U-disk during updating to avoid updating failure or unforeseenable error
.
3. Please power off oscillograph then open it for re­updating after updating failure
.
Page 97
Operating Manual for UTD2000/3000
8
5
The tenth example: Usage of storage function
1. Setup of storage function:
Storage functions of oscillograph include setup storage, waveform storage, bitmap storage and specific screen copy function of UTD2000L/2000EX series of oscillograph
.
It has to select display channel, trigger channel
,
vertical and horizontal gears of oscillograph for setup. Please press[STORAGE] key to sel ect torage type of “setup” by pressing[F1 ]key. As shown in following figure, it has to select position to save by multi-function button then press F3 for saving
.
It can be tested directly during next test after saving
setup to reduce re-setup process. UTD2000C series
of oscillograph can save 10 kinds of setup while UTD2000E and UTD2000L/2000EX series of oscillograph can save 20 kinds of setup
.
Figure 3-10 Saving setup interface
Page 98
Operating Manual for UTD2000/3000
8
6
2. Waveform storage function:
UTD2000 series of oscillograph can provide waveform storage function. User can press[STORAGE]key and [F1] key to select storage type of “waveform”. User ca
n
press [F2]key to select channel to save waveform the
n
press[F5]Ente r in to ne x t page . M ag ne ti c dis k in F1 menu position can designate storage position and default storage position is digital storage oscillograp
h
(DSO) without plugging in U-disk. Stored magnetic disk position can be [DSO] or [USB] after plugging in U-disk
.
Default storage depth shown by F2 is ordinary and saved waveform only can be displayed by oscill ograph
.
Storage depth can be selected to long storage to save “.data” for exporta ti on by wa v ef orm ana l ysi s s oft wa r e when plugging in U-disk and selecting magnetic disk position to [USB]. It is allowed to analyze content data of every point of saved waveform. User can also press F5 Return to previous me nu th en select storage position by multi-fun c ti on bu tto n the n pre ss [F3] fo r
saving
.
Figure 3-12 Waveform storage by U-disk
Figure 3-11 Waveform storage
Page 99
Operating Manual for UTD2000/3000
8
7
A
s shown in following figure, saved waveform ca
n
be recalled by short-cut key of [REF] on the panel for UTD2000C and UTD2000E series of oscillograph. User can press [F3] shown in the second page of STORAGE main menu for recall for UTD2000L/2000EX series of oscillograph
.
Figure 3-13 Vertical control panel of 5.7-inch scree
n
A
s shown in following figure, user can press[ F1] menu key of STORAGE main menu to select storage type of “bitmap” after plugging in U-disk and successful connection of osc illograph. User ca
n
rotate multi-function knob to select saved bitmap No. Saved bitmap in U-disk with “.BMP” format ca
n
be read out by computer directly
.
Figure 3-14 Bitmap storage
3. Bitmap storage function:
Page 100
Operating Man ual for UT D2000/30 00
8
8
4. Pr tSc copy & storage function
Introduction to S ystem Prompt
There is 1 PrtSc key in upper left corner of the scree
n
for U TD2000L/2000EX series of oscillograph. User can press the key for screen copy after plugging in U­disk. Displayed image can be saved in U-disk by wa
y
of “.BMP” format
.
Figure 3-15 Screen cop
y
Regulation to the extreme: It prompts that multi­purpose knob has been regulated to the greatest extent under current status and continued adjustment is not allowed. The prompt will display whe
n
regulating vertical deflection coefficient switch, time base switch, X displacement, vertical displacement and trigger level to the terminal
.
Successful connection of U-disk : The prompt will display on the screen after connecting U-disk to digital storage oscillograph and correct connection
.
Plug-out of U-disk: The prompt will display on the screen when plugging out U-disk from digital storage oscillograph.Saving: The prompt will display on the screen with schedule bar below it when saving waveform
Chapter 4: System Prompt and Troubleshootin
g