UNI-T UT60B, UT60C, UT60E Operating Manual

3
4 4 5 6 7 8 9
11
12 12
14 15
16
17 18 19
20
21 22
22 24 24 25 25 25 25 26
27 27 27 28
Unpacking Inspection Rules For Safe Operation
Page
Table of Contents
1
Model UT60B/C/E: OPERATING MANUAL
D. Continuity Test
International Electrical Symbols
AC Voltage Measurement
Capacitance Measurement
The Use of Relative Value Mode
The Meter Structure
Turning on the Display Backlight General Specifications
Resistance
Overview Safety Information
Rotary Switch Functional Buttons Display Symbols Measurement Ranges
A. B.
Measurement Operation
A. B. C. D. Testing for Continuity E. F. G. H. I. Temperature Measurement
(Model UT60C/UT60E)
J.
Operation of Hold Mode The POWER Button
The BLUE Button Sleep Mode (Model UT60B/UT60C) Accuracy Specifications
A. DC Voltage B. AC Voltage C.
Selecting a Measurement Range
Manual Ranging and Autoranging
DC Voltage Measurement
Measuring Resistance
Frequency Measurement
DC or AC Current Measurement
Testing Diodes
Measuring Duty Cycle
Capacitance
Frequency & Duty Cycle
DC Current
F. G. H. Temperature(Model UT60C/UT60E)
E.
I.
Diode Test
28 28 28 29 29
30 31
31 32 33
35 35
35
Page
2
J. AC Current
Maintenance
A. General Service B. Testing the Fuses C. Replacing the Battery D. Replacing the Fuses
RS232C Serial Port (Model UT60E)
A. RS232C Port Cable B. Setting of RS232C Serial Ports C. System Requirements for
Installing the UT60E Interface Program
Model UT60B/C/E: OPERATING MANUAL
3
Model UT60B/C/E: OPERATING MANUAL
Overview
This Operating Manual covers information on safety and cautions. Please read the relevant information carefully
and observe all the Warnings and Notes strictly.
Warning
T o avoid electric shock or personal injury, read the “Safety Information” and “Rules for Safe Operation”
carefully before using the Meter.
Digital MultimeterModelUT60B, UT60Cand UT60E (hereafter referred to as “the Meter”) has autorange and
manual range options with maximum reading 3999. The enclosure structure design adopted advanced “co-injection” technique in order to provide sufficient insulation.
In addition to the conventional measuring functions, there is a RS232C standard serial port equipped with Model UT60E for easy connection with computer to realize macro recording and monitoring and capture of transient dynamic data, displaying change of waveform during the measurement, providing data and evidence to engineering technicians for scientific research.This is also a highly applied digital multimeter of good performance with full overload protection and backlight function. Model UT60E also has true rms reading for AC voltage and AC current measurements.
Except where noted, the descriptions and instructions in this Operating Manual apply to all Model UT60B/UT60C
/UT60E
.
Unless otherwise identified, all figures show the Model
UT60B.
Qty
1 piece
1 pair 1 pair
1 piece 1 piece 1 piece 1 piece
Item
1 2 3 4
5 6
7
Description
Operating Manual Test Lead Test Clip
Model UT60E:CD-ROM (Installation Guide & Computer Interface Software)
Model UT60C/UT60E:Point Contact Temperature Probe
Model UT60E:RS232C Interface Cable
9V Battery (NEDA1604, 6F22 or 006P) (installed)
Model UT60B/C/E: OPERATING MANUAL
CAT IV: Primary supply level, overhead lines, cable systems etc.
4
Unpacking Inspection
Open the package case and take out the Meter. Check the following items carefully to see any missing or damaged part:
In the event you find any missing or damage, please contact your dealer immediately.
Safety Information
This Meter complies with the standards IEC61010: in pollution degree 2, overvoltage category (CA T. III 1000V , CAT. IV 600V;)and double insulation.
CAT. III: Distribution level, fixed installation, with smaller transient overvoltages than CAT. IV
Use the Meter only as specified in this operating manual, otherwise the protection provided by the Meter may be impaired.
In this manual, a Warning identifies conditions and actions that pose hazards to the user, or may damage the Meter or the equipment under test.
A Note identifies the information that user should pay attention on.
International electrical symbols used on the Meter and in this Operating Manual are explained on page 6.
Rules For Safe Operation
To avoid possible electric shock or personal injury, and to avoid possible damage to the Meter or to the equipment under test, adhere to the following rules:
l Before using the Meter inspect the case. Do not use the
Meter if it is damaged or the case (or part of the case) is removed. Look for cracks or missing plastic. Pay attention to the insulation around the connectors.
l Inspect the test leads for damaged insulation or exposed
metal. Check the test leads for continuity. Replace damaged test leads with identical model number or electrical specifications before using the Meter.
l Do not apply more than the rated voltage, as marked on
the Meter, between the terminals or between any terminal and grounding.
l The rotary switch should be placed in the right position
and no any changeover of range shall be made during measurement is conducted to prevent damage of the Meter.
l When the Meter working at an effective voltage over 60V
inDC or 30V rms in AC, special care should be taken for there is danger of electric shock.
l Use the proper terminals, function, and range for your
measurements.
l Do not use or store the Meter in an environment of high
temperature, humidity, explosive, inflammable and strong magnetic field. The performance of the Meter may deteriorate after dampened.
l When using the test leads, keep your fingers behind the
finger guards.
l Disconnect circuit power and discharge all high-voltage
capacitors before testing resistance, continuity, diodes, current, or capacitance.
l Before measuring current, check the Meter’s fuses and
turn off power to the circuit before connecting the Meter to the circuit.
l
Replace the battery as soon as the battery indicator " " appears. With a low battery, the Meter might produce false readings that can lead to electric shock and personal injury.
l Remove test lead, temperature probe, RS232C interface
Warning
5
Model UT60B/C/E: OPERATING MANUAL
Model UT60B/C/E: OPERATING MANUAL
cable and test clip from the Meter and turn the Meter power off before opening the Meter case.
l When servicing the Meter, use only the same model
number or identical electrical specifications replacement parts.
l The internal circuit of the Meter shall not be altered at
will to avoid damage of the Meter and any accident.
l Soft cloth and mild detergent should be used to clean the
surface of the Meter when servicing. No abrasive and solvent should be used to prevent the surface of the Meter from corrosion, damage and accident.
l l
l l
The Meter is suitable for indoor use. Under the environment with high (+/-4kV) electrostatic discharge, the Meter may not be operated as normal condition. The user may require resetting the Meter. Please take out the battery when it is not being used for a prolonged period to avoid damages to the Meter. Please constantly check the battery as it may leak when it has been using for some time, replace the battery as soon as leaking appears. A leaking battery will damage the Meter.
6
International Electrical Symbols
AC (Alternating Current)
Deficiency of Built-In
Battery DC (Direct Current) Continuity Test AC or DC Diode Grounding Capacitance Test
Double Insulated Fuse Warning. Refer to the
Operating Manual
Conforms to Standards
of European Union
The Meter Structure (see figure 1)
(figure 1)
1 LCD Display 2 Functional Buttons 3 Rotary Switch 4 HzV
Input Terminal: Input for voltage, frequency/duty cycle, resistance, diode, continuity and capacitance measurements.
5 COM Input Terminal:
Return terminal for all measurements.
Model UT60B
:
Input Terminal:
Input for 0.1 A to 400.0mA current measurements.
Input Terminal: Input for 0.1 A to 400.0mA current measurements and temperature testing.
7 10A Input Terminal:
Input for 0.01A to 10.00A current measurements.
7
Model UT60B/C/E: OPERATING MANUAL
Rotary Switch
Below table indicated for information about the rotary switch positions.
Rotary Switch
Position
Function
V DC voltage measurement range from 400.0mV to
1000V or AC voltage measurement range from 4.000V to
750.0V. Continuity test. Diode test. Resistance measurement range from 400.0
Temperature in celsius
Hz Frequency measurement range from 10.00Hz to
10.00MHz.
AC or DC current measurement range from
A.
mA AC or DC current measurement range from
40.00mA to 400.0mA.
A AC or DC current measurement range from
4.000A to 10.00A.
to 40.00M .
Model UT60B/C/E: OPERATING MANUAL
8
Below table indicated for information about the functional button operations.
Button Measuring
Function
Operation Performed
POWER Any rotary
switch
position
Turn the power on and off.
V
Switches between AC and DC voltage; the Meter beeps. DC is default. Model UT60B: Switches between continuity and diode and resistance measurements; the Meter beeps. Resistance is default.
Model UT60C/UT60E: Switches between continuity and diode and resistance and capacitance measurements; the Meter beeps. Resistance is default.
Switches between AC and DC current range from 400.0 A to 4000 A; the Meter beeps. DC is default.
mA
Switches between AC and DC current range from 40.00mA to
400.0mA; the Meter beeps. DC is default.
A
Switches between AC and DC current range from 4.000A to
10.00A; the Meter beeps. DC is default. Disable Sleep Mode feature.
(BLUE)
Functional Buttons (1)
9
Model UT60B/C/E: OPERATING MANUAL
RANGE
Any rotary
switch
position
exceptHz
and
l Press RANGE to enter the manual ranging mode; the Meter beeps. Manually selecting a range causes the Meter to exit the Hold and REL modes. l Press RANGE to step through the ranges available for the selected function; the Meter beeps. l Press and hold
RANGE
for 2
seconds to return to autoranging; the Meter beeps.
Hz
1. Press to start the frequency counter; the Meter beeps.
2. Press again to enter duty cycle mode; the Meter beeps.
3. Press again to return to the frequency counter mode; the Meter beeps.
Hz %
V , ,
mA or
A
1. Press to start the frequency counter; the Meter beeps.
2. Press again to enter duty cycle mode; the Meter beeps.
3. Press again to return to the previous measurement mode; the Meter beeps.
REL
Any rotary
switch
position
except Hz
Press to enter and exit the REL mode in any
measuring mode except in frequency/duty cycle; the Meter beeps.
Any rotary
switch
position
Press HOLD H
to enter and exit the Hold mode in any mode, the Meter beeps.
REL
Functional Buttons (2)
HOLD H
10
Model UT60B/C/E: OPERATING MANUAL
Display Symbols (1) (see figure 2)
Number Symbol Meaning
1 AC Indicator for AC voltage or current.
Model UT60B/UT60C: The displayed value
is the mean value. Model UT60E: The displayed value is the true rms value.
2 TRMS The Model UT60E: Indicator for true rms
value.
3
The Meter is in the auto range mode in which the Meter automatically selects the range with the best resolution.
4 RS232C Data output. It is always on the LCD, but
data output is only in progress when the Meter is connected to the computer via the
included RS232C Interface Cable. 5 % Percent: Used for duty cycle measurements. 6
Data hold is active. 7
The REL is on to display the present value
minus the stored value. 8
The battery is low.
Warning To avoid false readings, which could lead to possible electric shock or personal injury , replace the battery as
soon as the battery indicator appears.
9 Centigrade. The unit of temperature.
10
Test of diode
11
The continuity buzzer is on.
18
1
2
3
4
5
6
7 8 9
10 11
12 ~ 16
17
(figure 2)
11
Model UT60B/C/E: OPERATING MANUAL
, k ,M:
k
:
M
:
Ohm. The unit of resistance. kilohm. 1 x 10
or 1000 ohms.
Megaohm. 1 x 10
or 1,000,000 ohms.
nF
F: nF:
Farad. The unit of capacitance. Microfarad. 1 x 10
or 0.000001 farads. Nanofarad. 1 x 10
or 0.000000001
farads.
Hz,
kHz,
MHz
Hz: kHz:
MHz:
Hertz. The unit of frequency in cycles/second. Kilohertz. 1 x 10
or 1,000 hertz.
Megahertz. 1 x 10
or1,000,000
hertz.
V, mV V:
mV
Volts. The unit of voltage. Millivolt. 1 x 10
or 0.001 volts.
12~16
A, mA, A:
mA:
Amperes (amps). The unit of current. Milliamp. 1 x 10
or 0.001 amperes.
Microamp. 1x 10
or 0.000001
amperes.
17
Indicates negative reading.
18
The input value is too large for the selected range.
Measurement Ranges (1)
The Meter has both manual range and autorange options:
12
Display Symbols (1) (see figure 2)
A measurement range determines the highest value the Meter can measure. Most Meter functions have more than one range. See “Accuracy Specifications.”
A.Selecting a Measurement Range
Being in the right measurement range is important: l If the range is too low for the input, the Meter displays to
indicate an overload.
l If the range is too high, the Meter will not display the most
accurate measurement.
B. Manual Ranging and Autoranging
Model UT60B/C/E: OPERATING MANUAL