DET-2 Theodolite
User Guide
•
CONTENTS
Cautions ............................................................................................. 3
Usage and applications of the instrument ........................................... 4
Specifications ...................................................................................... 5
Nomenclature ..................................................................................... 6
Functions of buttons ........................................................................... 8
Preparation before measurement ......................................................... 9
Battery installation .............................................................................. 9
Recharging of battery ........................................................................ 10
Eyepiece and telescope focusing ........................................................ 11
Parameter settings ............................................................................ 11
Centering and leveling with optical plummet ................................... 14
Operation ......................................................................................... 16
Inspection ......................................................................................... 22
Accessories ........................................................................................ 28
Error codes ........................................................................................ 29
CAUTIONS
Carefully read and understand this User Guide before use.
Avoid eye injury. Do not aim the telescope at the sun.
The instrument should be protected from hard shock.
Do not carry a tripod mounted instrument on your shoulder.
Avoid sudden changes in temperature. Sudden changes in temperature will influence the
precision of measurement, will effect the operation of the electronic system, and will cause
condensation on the objective lens. When the instrument is taken from the cold weather
outdoor to indoor, it must be placed in a warm dry place to allow condensation to
evaporate.
This instrument contains sensitive electronic components that are protected against dust
and moisture. Dust or moisture inside the instrument will cause damage. After using in a
humid environment, the instrument must be dried immediately and stored in a dry
instrument case.
The LCD will respond slowly in low temperatures.
To prevent battery leakage, please remove
the battery pack if the instrument will be
stored for an extended length of time.
Before placing the instrument into the
instrument case, align the dot marks on the
instrument and lightly tighten the clamp
knobs. Place the instrument in the case
with dots facing upwards. Loosen clamps
after the instrument is in the case and
insure instrument is properly seated.
Slightly tighten clamps after instrument is
fully seated.
USAGE AND APPLICATIONS OF THE INSTRUMENT
This electronic theodolite utilizes the photoelectric incremental angle measuring system.
The precision of angle measurement is 2”. It integrates optical, mechanical, electronic and
computer technologies for a variety of functions including angle measurement, display and
storage. It displays horizontal and vertical angles and makes conversions from vertical angle
to percent. The vertical angle measurement is also compensated.
The theodolite can be used in a variety of applications including measurement in railway,
highway, bridge, water projects, etc. It can be also used in various construction applications
including orientation of large equipment, topographic survey, and various kinds of
construction measurements.
SPECIFICATIONS
Handle
Screw
Handle
Objective Lens
LCD Display I
Optical Plummet
Horizontal Clamp /
Fine Motion Screw
Leveling Screw
Tribrach
Locking Lever
Focusing
Ring
Battery Case
Eyepiece
Collimator
Telescope Clamp /
Fine Motion Screw
Tubular Vial
LCD Display II
Circular Vial
Button Primary Function Other
ON/OFF
Turns the
instrument
ON/OFF
1. Function button for entering into instrument setup.
2. Function button for entering into index error setting.
3. Function button for entering into compensation setting.
Button for lighting of reticule and LCD
0SET
Reset of
horizontal angle
(zeroing)
1. Menu selection button in Instrument Setup.
2. Function button for entering into Compensator setting.
3. Function button for input while in Instrument Setup.
HOLD
Horizontal
angle Hold
Button
1. Menu selection button in Instrument Setup.
2. Function button for entering while in Instrument Setup.
3. Function button for entering into index error setting.
R/L
Increment of
left rotation and
right rotation of
horizontal angle.
1. Menu selection button in Instrument Setup.
2. Function button for input while in Instrument Setup.
V%
Vertical angle
and slope
convert
1. Function button for input while in Instrument Setup.
2. Button for confirmation after initial Setup
PREPARATION BEFORE MEASUREMENT
Power Options
The theodolite has two power options: Disposable and rechargeable batteries. The
rechargeable batteries are supplied in a sealed pack. The disposable batteries are inside a
similar shaped pack with a sliding door.
To remove the pack, turn the knob until the mark “▼” points to [UNLOCK], then
remove the battery pack.
To insert battery pack, place the raised bottom part of the battery case into the slot in the
theodolite.
Push the top of the battery case in place. Turn the knob until the mark “▼” points to
[LOCK].
To Install Alkaline Batteries into the Disposable Battery Pack, open the battery pack cover,
insert four AA alkaline batteries into the battery case according to the (+) and (-).
Telescope
Image Erect
Magnification 30x
Aperture 45 mm (1.7 in)
Angle of view 1˚ 30’
Shortest distance 1.35 m (4.43 ft)
Stadia constant 100
Resolution 3”
Angle
Measuring
System
Angle measurement Incremental
Min. Reading 1” or 5”
Detection method H. Both sides, V. Single side
Precision 2”
Unit of angle Deg / mil / gon / V %
Display LCD both sides
Tilt
Sensor
Automatic Compensation Yes
Range of compensation +/-3’
Optical
Plummet
Image Erect
Magnification 3X
Angle of view 5˚
Focusing range 0.5 m ~ ∞ (1.6 ft ~ ∞)
Reticle type Crosshair
Vials
Tubular vial 30”/2 mm
Circular vial 8’/2 mm
Range Temperature -20˚C to +50˚C (-4˚F to +122˚F)
Power
Battery type 4xAA alkaline or NiMH Pack
Voltage 4.8
Battery Life 36 hours - alkaline
Dust / Water
Protection
IP54
Weight 4.5 kg (9.9 lbs)
Size
164x154x340 mm (6.4x6.1x13.4)
Warranty
1 Year
Recharging the battery
Connect the charger to the AC power source. The green light on the charger will light.
Connect the plug of the charger to the charging port of the battery pack. The green light
of the charger will turn red indicating the charging process has started. After 3-4 hours the
red light turns green indicating that the batteries are charged.
Caution: Do not charge Alkaline Batteries, severe damage can happen to the battery pack
or charger.
Attention: Turn off the instrument’s power before removing the battery pack.
EYEPIECE AND TELESCOPE FOCUSING
Adjustment of Eyepiece
Direct the telescope to a bright background. Turn the eyepiece ring so that the crosshairs
of the reticule are clearly seen.
Elimination of Optical Parallax
Adjust the focusing ring to clearly see an object on the reticule. Move your eyes up and
down to see if the image of the object moves relative to the graduation lines. If it does not
move, there is no optical parallax; otherwise turn the focusing ring to eliminate the optical
parallax.
Parameter Settings
Please enter all desired initial settings prior to the first measurement.
Items in bold indicate factory settings.
Initial Setting Selection
1. Unit of Angle
360
˚
/ 400G / 6400
Unit A / Unit B / Unit C
2. Zenith Angle ZEN = = 0 / ZEN = = 90
3. Auto Power Off Time 30 OFF / NO OFF
4. Min. Display dsp1 / dsp5
5. Tilt Sensor Switch TILT ON / TILT OFF
6. Indication of Horizontal
Angle Position
No Beep / 90
˚
Beep
Changing Parameter Settings
Press and hold [HOLD] button + [0SET] button and press [ON/OFF] button. Release
[ON/OFF) button when the full character display appears and release [HOLD] + [0SET]
buttons when four beeps are heard. The instrument enters into its initial setup mode and
the LCD displays:
• Press [►] button or [◄] button to change screens for selecting options.
• Press [▲] button for selecting specific content in the options.
• Finally, press [V / %] button to confirm and enter into angle measuring mode.
Unit of Angle
• UNIT A: 360˚ (Degree)
• UNIT B: 400 (GON)
• UNIT C: 6400 (Mil)
Zero Position of Vertical Angle
• ZEN == 0 : Zenith is 0°
• ZEN == 90 : Zenith is 90°
Auto Power Off Time
• NO OFF: Auto power off disabled
• 30 OFF: Turns power off if inactive
for 30 min
Minimum Display
• DSP 1: Minimum display is 1”
• DSP 5: Minimum display is 5”
Setting of Tilt Sensor
• V TILT ON: Turn on the tilt sensor
• V TILT OFF: Turn off the tilt sensor
Indication of Horizontal Angle
• NO BEEP: Horizontal angle indication
disabled
• 90 BEEP: Sounds beep when the
instrument is close to 0˚, 90°, 180˚ and 270°.
Centering and Leveling with Optical Plummet
• Extend the tripod legs to a suitable working height with equal length legs. Spread the
tripod legs and make the tripod head as level as possible while at the same time placing the
center of the tripod head directly over the ground point. Press the leg feet firmly into the
ground and make sure the tripod legs are locked.
• Set the instrument carefully on the center of the tripod head orientating the leveling
screws centered with each tripod leg. Attach the instrument to the tripod. Make sure the
center mark is visible on all three leveling screws (this ensures a complete leveling range).
• Adjust the Optical Plummet Eyepiece to focus the crosshairs. Adjust the Optical
Plummet telescope focus to see the ground clearly. If you can not see the ground point
while looking through the optical plummet, carefully lift two of the tripods legs, then
pivot on the third leg, carefully moving the tripod until the ground point is within one
inch of the reticule. Press the two tripod feet back into the ground and recheck the optical
plummet alignment. Repeat if necessary until the ground point can be seen in the optical
plummet field. Complete the alignment by turning the leveling screws (you will not be
level but you are pointed correctly).
Centering and Leveling with Optical Plummet
• Now center the instruments Circular Vial by carefully extending or shortening the tripod
leg closest to the bubble. Caution: Use only two legs. Repeat until alignment is within 6
mm
(¼ inch) or better.
• Next use the Tubular Vial to accurately level the instrument. Unlock and turn the
instrument so that the tubular vial is parallel to BC, any two leveling screws. Note the
direction to turn the leveling screws in the graphic. When turning the two screws, adjust
them equally. The bubble will move in the direction that your left thumb turns. To move
the bubble to the right, turn the B leveling screw in and simultaneously turn the C leveling
screw in. To move the bubble left, move both screws out simultaneously. Once centered
turn the instrument 90˚ over A leveling screw and turn screw A in or out until bubble is
centered. Go back to the first position BC, and repeat until the bubble is centered in both
positions. Then from position BC turn 180˚ to check the adjustment. If the bubble stays
centered or within ¼ division, you are leveled.
• Now, check the ground point centering. If you are not directly on the point, carefully
loosen the tripod fastener and move the instrument on the tripod head in an x - y
direction. Do not rotate the instrument. Recheck leveling and repeat until instruments is
level and over the ground point at the same time. With practice, this becomes easier.
![](/html/21/21a8/21a8042f8c4e29762f9f36f58e8a46fa986a14f073d932416b9f3e2236163b90/bg2.png)
– 18 –– 17 – – 19 – – 20 –
– 23 –– 22 – – 24 – – 25 –– 21 –
– 28 –– 27 – – 29 –– 26 –
– 16 –
Trimble Construction Division
5475 Kellenburger Road
Dayton, Ohio 45424-1099
U.S.A.
Phone 1.937.245.5600
Fax: 1.937.233-9004
www.trimble.com
© 2009, Trimble Navigation Limited. All rights reserved.
Reorder PN 1210-0110 (07/09)
Notice to Our European Union Customers
For product recycling instructions and more information,
please go to: www
.trimble.com/environment/summary.html
Recycling in Europe
To recycle Trimble WEEE, call: +31 497 53 2430, and ask for the ÒWEEE
associate,Ó or
mail a request for recycling instructions to:
Trimble Europe BV
c/o Menlo Worldwide Logistics
Meerheide 45
5521 DZ Eersel, NL
OPERATION
Start Up
• Press and hold [ON/OFF] button. Release [ON/OFF] button when the full character
display appears. The LCD displays:
• Move the telescope up and down when the instr ument is at the normal position. The
beeper beeps and the LCD displays the vertical angle. The instrument enters into
measuring mode.
• After the power is switched on and the instrument has entered into measuring mode, the
battery level is indicated by the batter y symbol in the lower right corner of LCD.
• If all of the three squares are displayed, the battery is fully charged.
• Decreasing squares indicates reduction of charge.
• If the battery symbol blinks, the battery is low and needs recharging or replacing.
V
R
H
86º 28΄ 48˝
150º 36΄ 10˝
Measurement of Angle
Observing in the “Normal” and “Reverse” Positions of the Telescope
The normal, or direct, position of the telescope refers to observation with the vertical circle
being on the left. The reverse position refers to obser vation with the vertical circle being on
the right. The mechanical errors can be offset by averaging the values measured in the
normal and reverse positions.
Normal
Reverse
Measurement of Vertical Angle
• 0° angle position can be set as follows in the initial setting:
ZEN==9 0 ( Zen ith 90°)
0°
90°
180°
270°
ZEN==0 (Z eni th 0°)
0°
90°
180°
270°
Compensation of the Tilt Sensor to
Vertical Angle.
• If the instrument is inclined within ±3’,
the tilt sensor can compensate the vertical
angle. If the inclination is greater than ±3’,
the instrument will display [TILT] as
shown in the figure.
Display of slope
• Press [V/%] button, the vertical angle
display is changed to slope display; press
[V/%] button, the vertical angle display is
resumed.
Note: When vertical angle is converted to
slope, the precision of the slope reading is
two digits after the decimal. The value of
slope is displayed only within ± 45°
(100%). There will be no slope display
exceeding this range.
V
R
H
90º 00΄ 10˝
108º 36΄ 38˝
V
R
H
69º 11΄ 00˝
108º 36΄ 38˝
V
R
H
38.88%
108º 36΄ 38˝
V
R
H
69º 11΄ 00˝
108º 36΄ 38˝
Measurement of Horizontal Angle
• Reset of Horizontal Angle
Press [0SET] button, the horizontal angle
returns to zero
Selecting the Direction of Measurement of
Horizontal Angle.
• Press [R/L] button to change the
direction of measurement of the horizontal
angle.
When “HR’ is displayed, the angle
increases with clockwise turning.
When “HL” is displayed, the angle
increases with counterclockwise turning.
Holding Horizontal Angle.
• Press [HOLD] button, the horizontal
angle will be held. The reading of the
horizontal angle will remain unchanged
even if the direction of telescope is
changed. Press [HOLD] button again, the
hold of horizontal angle is released.
V
R
H
90º 00΄ 10˝
150º 36΄ 10˝
V
R
H
90º 00΄ 10˝
150º 36΄ 10˝
V
R
H
90º 00΄ 10˝
00º 00΄ 00˝
V
L
H
90º 00΄ 10˝
209º 23΄ 50˝
V
R
H
90º 00΄ 10˝
150º 36΄ 10˝
V
R
H
90º 00΄ 10˝
150º 36΄ 10˝
Turning Off.
Press - [ON/OFF] button and hold,
“OFF” will be displayed, after a beep;
release [ON/OFF] button, the instrument
is turned OFF.
Measuring Distance Using the Stadia Method
Read the leveling rod using the stadia hair on the reticule of the telescope. Multiply the
reading by 100, to obtain the actual distance L from the target to the measured point.
(100 is the multiplication constant error of the instrument, i.e., L = l x 100)
Installation and Removal of the Base
Base Removal
• Turn the screw on the Tribrach Locking Lever outward using a flat screw driver to its stop.
• Turn Tribrach Locking Lever 180° counterclockwise. Holding the base with one hand,
take the main body off the base.
Base Installation
• Turn the Tribrach Locking Knob counterclockwise until it reaches the position limit.
Align the positioning block on the main body of the instrument to the notch on the base.
Install the main body onto the base .
• Turn the Tribrach Locking Lever clockwise until it reaches the position limit so that the
‘v’ mark points downward. Turn the locking screw inward until it hits its stop.
INSPECTION
Tubular vial
• Attach the instrument to a tripod and rough level. Position the tubular vial parallel to a
line connecting any two of the three leveling screws on the base. Adjust the two leveling
screws so that the tubular bubble is centered.
• Turn the instrument 180° and check if the bubble remains at the center.
• If the bubble remains at the center, no adjustment is required. Otherwise, perform
adjustments as follows:
• Using the bubble adjustment screws, move the bubble towards the tube center for half of
the error
• Turn the leveling screw to correct the other half of the error so that the bubble is
centered.
• Rotate the instrument 180 degrees and check if the bubble remains centered. If the
bubble is centered, the adjustment is complete. If not, repeat the steps until the bubble is
centered when the instrument is at any position.
Circle vial
After making sure that the tubular vial is correctly adjusted, check the circular level. If the
bubble is centered, no adjustment is required. Otherwise, adjust the three adjustment
screws with a needle to center the bubble.
Optical Plummet
• Attach the theodolite to a tripod (no leveling is required).
• Place a target under the instrument.
• Focus the image of the target then adjust the leveling screws so that the target is centered
in the reticule.
• Turn the instrument 180°.
• If the target remains at the center of the reticule, no adjustment is required. Otherwise,
adjust as follows.
• Unscrew the optical plummet protective cover ring
• Using the adjustment screws for the reticule, reduce the off—center distance by half.
Note: use the needle (as shown below).
• Repeat the above steps until the target coincides with the center.
Perpendicularity of Vertical Crosshair of Reticule of Telescope
• Attach the instrument on a tripod and precisely level
• Place a target point, A, 50M away from instrument.
• Aim the telescope at point A. Move the telescope using the vertical fine movement. If
point A moves along the vertical hair of the reticule, no adjustment is required.
Perform the following adjustment if the point A strays from the vertical crosshair:
• Remove the protective cover of the reticule and slightly loosen the four adjusting screws.
Turn the assembly so that point A coincides with the vertical hair, retighten the four
adjusting screws.
Repeat the above steps until there is no error.
Collimation Error
• Attach the instrument on a tripod and precisely level.
• Aim at point A in the distance with the normal, direct position of telescope. Record the
reading of the horizontal angle - HR-DIRECT. Next using the reverse position of the
telescope, take the reading of the horizontal angle - HR-REVERSE, then:
Collimation Error C = (HR-Direct -— HR-Reverse ± 180°)/2
If C <10”, no adjustment is required. If C >10”, the following adjustment is required:
Adjust the horizontal fine motion in the reverse position of the telescope so that the reverse
reading HR-Rev = HR-Rev + C.
Remove the protective cover of the reticule of the telescope and adjust both the left and
right adjusting screws so that the vertical hair of the reticule coincides with object A.
Repeat the steps until acceptable condition is reached.
Index Error of Vertical Circle
• Attach the instrument on a tripod and precisely level.
• Aim at any object, point P, in the normal position and take the reading of vertical angle -
V-Direct.
• Turn the telescope to the reverse position and aim it at point P again. Take the reading of
other vertical angle - V-Rev.
• If (V-Direct + V-Rev) - 360°= ≤15”, no adjustment is required. Otherwise, perform
adjustment as follows:
Press and hold [R/L] + [HOLD] buttons
while pressing [ON/OFF] button.
Release [ON/OFF] button when full
character display appears and then release
[R/L] + [HOLD] buttons when four
beeps are heard.
Swing the telescope near horizontal with
the instrument in the normal position.
Allow the vertical angle to reset after it
crosses zero. Aim the telescope in the
normal position at object P and press
[0SET] to confirm.
Aim the telescope in the reverse position
at object P and press [0SET] to confirm.
With this, the compensation of index
error is completed.
V
R
H
272º 36΄ 06˝
FACE-2
V
R
H
90º 00΄ 10˝
150º 36΄ 10˝
ACCESSORIES
1 plumb bob
1 tool kit (containing a screw driver and 2 needles)
2 bags of desiccant
1 rain cover
1 user guide
1 charger
1 battery pack alkaline
1 battery pack rechargeable
Display Meaning and remedy
E01 Count error, if displayed repeatedly, repair is needed
TOO FAST
The telescope or collimation unit rotated too fast, press any key
except [on/off] and [
], the instrument returns to normal state
E04 Horizontal sensor I error, repair is needed.
E05 Horizontal sensor II error, repair is needed.
E06 Vertical sensor error, repair is needed.
TILT
The tilt sensor is out of range. Level the instrument again. If this
does not clear the error code, repair is needed. Note: For a
temporary solution, the tilt sensor can be turned off.
Due to our policy of continuous improvement, we reserve the right to alter product design
and specification without prior notification.
ERROR CODES