The Atrato range represents a new generation of through bore time of flight ultrasonic flowmeters that uses breakthrough technology to offer a wide ranging yet accurate meter. It is ideal for many process control, instrumentation and laboratory applications. The user must make sure that the flowmeter selected is suitable for the application and that the chemical compatibility, temperature and pressure requirements are within the Atrato’s operating range. Please check the model number before proceeding. All meters can be programmed and monitored via the USB connection.
1 General
4
First 3 digits Flow range
710 = 2 - 500 mL/min
720 = 0.01 - 1.7 L/min
740 = 0.02 - 5 L/min
760 = 0.1 - 20 L/min
Fourth digit seal material
V = Viton
®
N = Nitrile
E = EPDM
S = Silicon
Fifth digit end fittings
0 = 3/8” John Guest 10 bar
1 = 1/2” BSP PEEK 10 bar
2 = 1/2” NPT 316 St St 30 bar
3 = 1/2” BSP 316 St St 30 bar
Sixth digit Wetted materials
0 = PEEK/316 St St
1 = PEEK / Borosilicate glass
Seventh/eighth digit Electronics package
A = Analogue output
D = Display & analogue output
RA = 110°C Sensor remote electronics analogue output
RD = 110°C Sensor remote electronics display & analogue output
For example
760 V10 D
This is a 20 L/Min flowmeter with Viton seals, half inch BSP fittings and a 316 stainless steel flow tube with PEEK end fittings, fitted with a local digital display and an analogue output.
• Locate the flowmeter in a sheltered position away from falling water.
• Care must be taken to ensure that the end fittings on the meter are not stressed during use. Ideally, flexible tubes should be used.
• Ideally the meter should be installed with straight lengths of tube either side for a distance of 10 pipe diameters upstream and 5 downstream.
• Install the device well away from valves, regulators bends and other components that could cause excessive turbulence on the fluid entering or leaving the meter.
• If necessary use spacer blocks and mounting clips to raise the pipe work centre line 58mm above the surface.
• It is good practice to use upstream and downstream isolating full bore ball valves to facilitate easy meter installation or removal.
• If push-in 3/8” John Guest fittings are used, clip the pipes to the mounting surface 300mm upstream and 150 downstream.
• The Atrato must be installed in a positive pressure system. Ensure that there is sufficient back pressure on the flowmeter to keep any gas in solution. We recommend 500mbar plus two times the fluid vapour pressure.
• If there is any chance of air passing through the system mount the Atrato in a vertical pipe with the flow in an upward direction otherwise air can remain trapped in the meter and affect it’s performance.
2 Installation
5
Using suitable fixings, mount the connection box onto a rigid surface with the cable entry preferably at the bottom. The mounting screw centres are shown in the diagram on the right. The flowmeter should be clipped into place with the terminal screws positioned at the lower edge. Note that the rubber strip covering the terminals on the meter will not seal if water is persistently present in this area. Ensure the meter is not pressure washed. The Atrato can easily be dismounted by inserting a screwdriver under the mounting clip and gently unclipping it to release the main body. The PG9 thread can be either connected to suitable flexible conduit or may be fitted with the supplied cable gland.
End fitting ‘B’ Order code Dimension ‘A’ mm
3/8” John Guest 0 145
1/2” BSP PEEK 1 142
1/2” NPT 2 180
1/2” BSP stainless 3 180
75
‘A’
PG9
‘B’
58
82
14
60
22
The Atrato can be operated entirely from a computer using the USB connection but this may not be acceptable in many situations where remote operation or further functions are required. The unit will work with systems from Windows XP onwards. The maximum connector cable size is 2.5mm (22-14AWG), for ease of assembly we recommend 1mm maximum. Care should be taken when terminating the wires as the conductors should be stripped to 4 to 5mm maximum and the wire ends must be pushed fully into the connector before tightening. These cables pass through a slot cut in the aluminium housing please ensure that no bare conductors are clear of the surface of the connector strip prior to closing the housing.
3 Electrical
Atrato Connector Assignments
Pin No Label Function
1 Relay NO Isolated relay contact; normally closed contact 24V 100mA
2 Relay Common Isolated relay contact; change over contact 24V 100mA
3 Relay NC Isolated relay contact; normally open contact 24V 100mA
4 4-20mA Analogue current (4-20mA) output (reference to 0V)
5 0-5/10V Analogue voltage (0-5/10V) output (reference to 0V)
6 PNP Output 1 open collector PNP OR Input 1 (5-24V dc) e.g. switch to pin 9
7 NPN Output 2 open collector NPN OR Input 2 (5-24V dc) e.g. switch to pin 8
8 0V External ground; common for PNP, NPN and Analogue outputs
9 +9 -24V dc External power; 9V ~24V (> 12V for Analogue output)
6
= Operates on External Power or USB = Operates on External Power only
Before use, download and install Atrato software from memory stick or website.
The Atrato low-flow ultrasonic flowmeter should be setup using the USB interface and a suitable computer. Below is a screen shot of the Configuration screen.
The panel below shows the functions of the panel on the top left of the screen. Control buttons to apply the settings and recover previously set parameters.
4 Setup
7
Send settings to system
Retrieve Settings from meter
Open settings file on the computer
Save settings to a file on the computer
Test outputs
Unlock/Lock User Settings
Diagnostic Tools
Software Information
To modify the configuration click “User Settings” tab at the bottom of the screen and ensure the system is unlocked for editing.
8
MULTIPLE WINDOW DISPLAY
To aid set up whilst operating the meter it is possible to detach the Flow Details and Advanced windows by dragging and dropping the bottom tabs away from the main window. This allows the user to view those tabs whilst adjusting the User Settings.
The Flow Details window may be further reduced or expanded using the minimise and maximise icons respectively in the top right hand of the window.
USER SETTINGS TAB: CALIBRATION SETTINGS:
Units: A drop down menu offers the choice of - Litres, cc, gms, kg, US gallon, Imperial gallon or Custom units i.e. blank.
Rate (units per): This is the time base for the flow rate and has the option of Second, Minute or Hour.
9
Scale Factor Adjustment: This is a fine tune adjustment on the signal to compensate for errors introduced by erratic flow or other system irregularities DENSITY: The flowmeter is fundamentally a volumetric device but a density figure can be entered here if one of the mass units are selected. Caution must be used however as there is no temperature/density correction.
Cut Off: Flow values below this level will be set to zero.
DISPLAY APPEARANCE:
Decrease Total for -ve flow: The meter will not register negative flow unless this box is checked. For reverse flow the rate will show a “-“ sign before the rate and the total will be reduced accordingly. One of the transistor outputs (PNP or NPN on Pin 6,7 or Pin 8) could be configured to give a logic level when reverse flow is detected. Once selected the totaliser will adjust with the negative flow and the NPN and PNP outputs will give a pulse output in proportion to the reverse flow seen. The analog pins will not give outputs.
Display Mode: The down arrow selects from the various display and Metraflow functions. The options are:-
Rate – The display will show flow rate only. Total – The display will show total flow only. Both – The display can be cycled from rate to total using the
left hand button on the Atrato or the left button in the “Real time window”. The TOTAL can be Reset at any time on the display model by pressing the Right Hand Button when the Total is being displayed. Alternatively, a remote input can be used if either pin 6 or pin 7 has been utilised.
Rate DP Position: Use the drop down menu to choose the required decimal point position. Total DP Position: Use the drop down menu to choose the required decimal point position. If the decimal point is set too low and the display value exceeds the setting capacity the display will show ----oF----.
Viewing Filter: This prevents the display, flow switches and analogue outputs from jittering with irregularities in the flow from say a peristaltic pump. The increments are arbitrary with the degree of damping approximately doubling with each level. Increments go up to 5, which may take up to a minute to stabilise.
10
NOTE: the operation for Pin 6 and Pin 7 are identical with the exception of the type of transistor output. These instructions are the same for Pin 7 except the Pin 6 (PNP) would read Pin 7 (NPN).
PIN 6 (PNP): Pin 6 can be configured as either an
output or an input but not both simultaneously. If one input and one output is required the user must determine whether they require a NPN or PNP output and use the appropriate terminal for each action. The options are – Not used
Output Input .
PIN 7 (NPN): Pin 7 can be configured as either an
output or an input but not both simultaneously. If one input and one output is required the user must determine whether they require a NPN or PNP output and use the appropriate terminal for each action. The options are – Not used
Output Input .
EXTERNAL CONNECTIONS
PIN 6 (PNP)/PIN 7 (NPN)
The contents of this box will change depending on the selection made in the External
connection box.
If OUTPUT is selected the check box options are:­Pulse
Single Flow Switch Dual Flow Switch Reverses Flow
11
PULSE: Enter the number of pulses per unit volume
required. This figure can be adjusted to suit the application and the flow range required from the meter. The maximum output frequency is 1000 Hz. So care must be taken to ensure that this pulse rate is not exceeded.
E.g. 10 L/min at 1000 Hertz equals (1000x60)/10 = 6000 P/L
Therefore 6000 pulses per litre is the maximum resolution for 10 litres per minute assuming maximum frequency output. If the meter is being used up to 5 L/Min the pulse output and so resolution could be doubled to 12000 P/L.
SINGLE FLOW ALARM:
Checking this option opens up input boxes for THRESHOLD; LATCHING and INVERT
Set the Threshold to the flow value you require the transistor to operate. The logic of operation can be adjusted with the INVERT check box (see left).
The Latching box gives the user three options of None; High Threshold and Low Threshold.
• “None” will cause the alarm to be triggered when the threshold flow is met. The alarm state will then reset in accordance with the user set Hysteresis.
• “High Threshold” will cause an alarm signal to Latch on when flow is seen above the set value
• “Low Threshold” will cause an alarm signal to Latch on when flow is seen below the set value
Hysteresis (%): This is used when Latching is Set to NONE to prevent “fluttering” of the alarm signal when flow is around the threshold level.
LATCHED alarms will remain Active and the Local Display will show “LATCH” until they are reset. RESET LATCH ALARMS by either a signal Input on one of the other Input/ Output PINS, or by Pressing and Holding the Right Hand Button on the Local Display on the Flow meter for 3 seconds.
Note the alarm points are driven by the display value on the “flow reading” screen. Any changes to the filter value of the display will affect the response time of the alarm points so there are two adjustable parameters to ensure “fluttering” does not happen with small flow fluctuations - response time and hysteresis.
INVERT – This Check box allows the user to invert the operation of the signal
Checked – Alarm output logic HIGH
For details see the Alarm logic table or hover mouse over the set point to see the action of the output with the chosen settings.
12
Checking this will enable a range of flow to be entered. When the flow is outside/inside of this range the alarm will be triggered. The remaining settings in the input box are the same as for Single Flow alarms.
• When no Latching is chosen the alarm will automatically reset in proportion to the user set Hysteresis.
• The Latched Alarm can be reset by a signal Input on one of the other Input/Output PINS, or by Pressing and Holding the Right Hand Button on the Local Display on the Flow meter for 3 seconds.
• The INVERT checkbox will reverse the action of the alarm output. (see the Alarm Logic Table).
DUAL FLOW ALARM:
If INPUT is selected either the Reset Total or Reset Alarm can be operated remotely by using a switch and resistor to apply a suitable voltage (between 5 and 24 Vdc) to either Pin 6 or Pin 7 as described on page 13.
This option allows the user to send a signal from the assigned wire/PIN when reverse flow is seen. Selecting INVERT will reverse the action of the signal.
REVERSE FLOW:
ALARM LOGIC TABLE
Flow
Latching Invert
Alarm Action
Alarm Below Threshold Above Threshold
Single None Unticked OFF ON
Single None Ticked ON OFF
Single High Threshold Unticked OFF ON
Single High Threshold Ticked ON OFF
Single Low Threshold Unticked ON OFF
Single Low Threshold Ticked OFF ON
In Range Outside Range
Dual None Unticked OFF ON
Dual None Ticked ON OFF
Dual Out of Range Unticked OFF ON
Dual Out of Range Ticked ON OFF
Forward Flow Reverse Flow
Reverse Flow n/a Unticked OFF ON
Reverse Flow n/a Ticked ON OFF
13
Pin 7 : In input mode, the NPN open collector output is disabled, and the input pin responds to an external input voltage at TTL levels (VinH 2.4V; VinL 0.8V). This input must to be driven by an external voltage supply which can safely be in the range of zero to 24V. Because of the potential for conflict should the NPN output be activated accidentally, Titan recommend a resistor so that the source current cannot exceed 5mA. An example circuit is given below:
Example Input to Pin 7
Pin 6 : In input mode, the PNP open collector output is disabled, and the input pin
responds to an external input voltage at TTL levels (VinH 2.4V; VinL 0.8V). This input must to be driven by an external voltage supply which can safely be in the full range of zero to the external supply voltage. Because of the potential for conflict should the PNP output be activated accidentally, Titan Enterprises Ltd recommend a resistor so that the sink current cannot exceed 5mA. An example circuit is given below:
Example Input to Pin 6
14
RELAY
RELAY MODE: If the relay is required for a flow switch
check the flow switch box and a further set of options for set point and hysteresis will appear.
THRESHOLD: Set the required flow rate for the flow
switch. The Reset Offset (%) (Hysteresis) is either negative or positive depending on whether a falling or rising alarm is required. The relay has change over contacts but to reduce power consumption the relay is off if positive hysteresis is selected and only powers when the threshold is exceeded. If negative hysteresis is chosen the relay is on up to the set point. The increments are preset and can be cycled through using the + &- buttons. If more than one set point is required either both transistors must be used or one transistor and the relay.
ANALOGUE OUTPUTS
ANALOGUE MODE: There are four options on the
drop down menu: Not Used, 4-20mA, 0-5 Volt and 0-10 Volt.
Once the selection has been made you can enter the low and high flows for the analogue outputs as required. These are numeric entries in the units and time base selected in calibration settings.
15
FLOW DETAILS TAB
FLOW RATE: This is the instantaneous flow rate and is updated approximately every
100 milliseconds.
FILTERED
This is the filtered flow rate and is identical to the LCD display on the
FLOW
flowmeter (if fitted) and the speed of change will depend on the filter
RATE
setting. The filter will also affect the response time for the analogue
outputs and the flow switches.
TOTAL: The total liquid passed since the last reset.
PIN 6 & 7:
If no selection is made “Not used” will be in the Notes column. See
following text and chart for the display legends.
OR:- When Pin 6 or 7 is set to “Output” and set to “Pulse” this will
display the frequency of the running output and the number of pulses per litre selected.
OR:- If it is selected to “Output” and set to “Flow switch” the Value
column will show the transistor status either “low” or “high” and the notes will show the selected switch points.
OR:- If “Output” and “Reverse Flow” are selected the Value column will
show either “low” or “high” depending on forward or reverse flow.
OR:- If “Input” is selected the Value column will show “high” or “low” depending
on the switch condition and “Input Logic level” in the Notes column.
The large display reflects the meter readings showing on the Atrato LCD Display, if fitted and will mimic that display. The buttons either side have the functions selected on the ‘Configuration’ screen and also work in parallel with the buttons on the Atrato flowmeter unit itself. The small window below the rate and total display shows various relevant operational parameters whilst the unit is operating e.g. relay status and frequency output if these options are selected; and overfill or under fill if in batch mode.
16
The chart below shows the various display messages for various operating functions. During operation only one message will be shown for each parameter.
Parameter Value Notes Function
Flow rate 2.5000 Litres per minute Units selected in configuration
Filtered flow rate 2.4981 Litres per minute This will lag flow rate
Total 16.95 Litres Total passed since last reset
Pin 6. Off Pin not being used
83.33Hz Output Pulses (2000 per Litre)
Frequency output for current flow rate @
2000 pulses/Litre
On Will turn off when rate below 2.7 Flow switch set to 3 L/Min 10% -ve flow.
Off Will turn on when rate above 3.3 Flow switch set to 3 L/Min 10% +ve flow.
Forward Direction of flow Reverse flow tab selected
Low Input logic level
Left push button selected will momentarily
show high when remotely operated
Low Input logic level
Right push button selected will momentarily
show high when remotely operated
Pin 7. Off Pin not being used
83.33Hz Output Pulses (2000 per Litre) Pulse output selected
On Will turn off when rate below 2.7 Flow switch set to 3 L/Min 10% -ve flow.
Off Will turn on when rate above 3.3 Flow switch set to 3 L/Min 10% +ve flow.
Forward Direction of flow Reverse flow tab selected
High Input logic level
Left push button selected will momentarily
show low when remotely operated
High Input logic level
Right push button selected will momentarily
show low when remotely operated
Relay Off Relay not being used
On Will turn off when rate below 2.8 Flow switch set to 3 L/Min 10% -ve flow.
Off Will turn on when rate above 3.4 Flow switch set to 3 L/Min 10% +ve flow.
17
PIN 7:
The display for “Pin 7” will function the same as “Pin 6” (see page 13).
RELAY:
If no function is set for the relay the Value column will show “Off” otherwise
it will display the function and its operating parameters.
BATCHING
If Batch is selected in the “Display mode” the relay is automatically
OPTION:
assigned to the total display for batching pre-set volumes of liquid. The
preset batch size can be entered in the relay box. This value will be seen on the LCD display and will count down when the left hand start button is pressed. The batch can be interrupted by pressing the right hand button, in the paused state a further press of the right button cancels the batch. If you wish to continue the batch from the paused state the left hand button continues the dispense from the point it was interrupted.
If an alternative batch size is required after the unit is disconnected from
the computer pressing both buttons together for 3 seconds will permit the batch size to be altered by using the right hand key to increment the digit that is flashing and the left hand button to advance to the next digit. Holding the left button for 3 seconds enters the selected numbers and the meter is ready to begin batching. Both buttons can be operated remotely if required e.g. for remote start/stop signals.
ATRATO ADVANCED USER TAB
Median filter.
This filter removes short term anomalies in the recorded e.g. an air bubble passing through. It should be used with caution as it is theoretically possible that over use of this feature could lead to incorrect results in certain circumstances. Under normal operating conditions the meter returns around 25 results per second and a regular occurrence at this frequency could be completely ignored thus causing problematic readings.
The filter is a moving window taking the middle number from an odd number of results selectable between 1 and 21. It is not an average or mean and it is designed to totally ignore one or more results.
18
PIN protection. There are 2 possibilities for the PIN protection:-
To use these features it is first necessary to create a PIN. This is a four digit number in the range 0001 to 9999. Setting a value of 0000 tells the software there is no PIN and this is how to remove the PIN. After allocating a PIN these check boxes can lock all of the settings or just the advanced user, including the accumulated total reset. If you forget your PIN please contact you supplier for the default value. Once entered, all the setting parameters remain visible. In the ‘Help’ menu, the user can ‘Log On using PIN...’ but the user will be prompted for the PIN prior to making any changes. If the display version of the meter is being used and the display is set to view totals or rate and total the left hand button will cycle through the various options.
Rate The display shows units and time base e.g. 1.0 Ltr/min.
Total The display is static with just the volume e.g. 10.3Ltr.
19
SPEED OF SOUND COMPENSATION
The Atrato ultrasonic flowmeter is factory calibrated on clean water and therefore the window of calibration is based around the measured time of flight of the ultrasonic signal based on the speed of sound of water. When a fluid is used in the meter that has a significantly different speed of sound to that of water, the stable signal may move outside of the window causing the meter to read incorrect and unstable time of flight signals. In this scenario you can compensate by manually adjusting the measurement window of the meter in the advanced tab feature for Fluid Speed of Sound Adjustment. A bar is seen in the graph indicating whether a strong signal is being measured. If a fluid has a faster speed of sound the user can click the FASTER button to adjust the measurement window a stronger signal point. A RED graphic is a very poor signal, ORANGE indicates a weak signal and GREEN is a Good strong signal
The user then enters a simple scaling factor on the User Settings tab, to accurately adjust the water calibration to that of the fluid being measured.
Recorded flow Used flow
0 0
1 1
1 1
2 2
2 2
3 2
0 3
3 3
4 4
4 4
4 4
Recorded and numerical order
2 3 0 0 2 3 2 is used
3 0 3 0 3 3 3 is used
0 3 4 0 3 4 3 is used
The seventh flow in the table above is zero. This could have been a small bubble passing through the 710 meter which has a 1mm bore which could completely absorb the ultrasound. The median filter completely ignores this reading and for the zero reading returns the value of 3 from the results either side. This filter is particularly useful at low flows where a small dip in value could drop a result below the internal cut off levels.
20
DATALOGGING (not available on all meters)
For general diagnostic purposes it is possible to log the meter performance using the Data logging Tab function. The logging is not internal to the flowmeter, but stored to the connected PC and therefore only occurs whilst the PC is connected to the meter.
Simply tick the relevant parameter you wish to log, choose a sensible sampling period (i.e. for longer data-logging periods choose 10+ second intervals to limit the size of the file created), and then click the Icon under Storage File and enter a simple filename (no spaces) at a location you desire.
Once all is selected correctly the START LOGGING button will become enabled. Click it to start the datalogging file creation. In the bottom left corner the samples collected will be listed as the log file builds.
To finish the log file, simply Click STOP LOGGING.
5 Technical Specification
Linearity ±1.0% of reading over flow range
Repeatability ±0.1% from 25% to 100%
±0.5% from 0 to 25%
Housing IP54
Temperature range -10 to 60°C assembly with enclosed electronics
or -10 to 110°C sensor only (for use with remote electronics)
-10 to 60°C remote electronics
Fluid temperature range -10 to 60°C or -10 to 110°C with remote electronics
Storage temperature -20 to 110°C
Pressure rating 10 bar standard, 30 bar with stainless steel end fittings
Pulse output PNP or NPN maximum frequency 400 Hz
Relay 24 Vdc 500mA max non inductive
PIN 6 transistor output PNP 24 V @ 20mA maximum
Input Pull down resistor required (10K ohm)
PIN 7 transistor output NPN 24 V @ 20mA maximum
Input Pull up resistor required (10K ohm)
LCD display Reflective
6 x 8mm high main characters
2.5mm enunciators Gal. cc. Kg. gms. Ltr. /min /Hr /Sec
4 – 20mA output into 250 ohm maximum 14 bit resolution ±0.1% linearity (plus flowmeter accuracy)
0 – 10 Volt output 14 bit resolution (14 Vdc min supply voltage) ±0.1% linearity (plus flowmeter accuracy)
0 – 5 Volt output 12 bit resolution
USB TypeA connector Windows XP or later
Wiring terminals 1mm maximum
Power supply 10 – 24 Vdc (15 -24 Vdc for 4-20mA or 0-10 V)
Power consumption 110mA (plus analogue output current)
Connections 1/2” BSP male PEEK or 1/2” NPT or BSP 316 stainless steel. 3/8” John Guest push-in
MATERIALS OF CONSTRUCTION
End fittings
PEEK, food and medical grade or 316 Stainless steel
Flow tube 316 Stainless steel as standard Alternative – Borosilicate glass
Seals Viton as standard Alternative – Nitrile, EPDM, Silicon
Housing Aluminium extrusion
End caps ABS
Mounting bracket ABS
Ext. elastomeric seals PTE
21