Teledyne T3-16 User Manual

Page 1
Summit T3-16 PCI Express Protocol Analyzer
Quick Start
1
Introduction
Use this document for quick installation and setup.
If you experience problems or need more information, see the Summit T3-16 User Manual on the Installation CD or at the Teledyne LeCroy web site.
2
Components
Summit T3-16 analyzer box
AC Power Cable
USB and Ethernet cables
•PETrac e r Software program CD-ROM
Depending on configuration, the package may also include:
One (for x8 recording) or two (for x16 recording) probe cables
Slot interposer
MidBus Probe kit
Please see the Summit T3-16 User Manual on the installation CD for component specifications and further details.
Summit T3-16 Protocol Analyzer
3
Unpacking the Analyzer
Inspect the received shipping container for any damage. Unpack the container and account for each of the system components listed on the accompanying packing list. Visually inspect each component for absence of damage.
In the event of damage, notify the shipper and Teledyne LeCroy. Retain all shipping materials for shipper’s inspection.
Page 2
4
Front Panel Description
The Analyzer has the following features on the front:
[1] Power Switch (black): 1 = On and 0 = Off. [2] LCD Menus: Allows you to configure Summit T3-16 and view
status.
[3] Navigation Buttons: Navigate through the LCD menu. [4] Speed LEDs: Indicate speed setting the analyzer is using for
each traffic direction.
[5] UPSTREAM [15:8] and UPSTREAM [7:0] connectors:
Connection to the probe for the capture of upstream direction of the link.
[6] DOWNSTREAM [15:8] and DOWNSTREAM [7:0]
connectors: Connection to the probe for the capture of
downstream direction of the link.
[7] Trigger IN and Out: Provide external trigger capabilities. You
can configure Summit T3 to trigger external equipment using Trigger Out. You can use Trigger In to trigger the Analyzer from another device.
[8] USB Type B Host Machine Connector: To connect the
Analyzer to the host machine using a USB connection.
[9] Ethernet Port: 1 GIGE Connectivity allows connection to the
analyzer through an Ethernet network.
[10] Expansion Slot: CATC Sync Expansion Card to use the
Teledyne LeCroy CrossSync application.
LCD Menus
The Summit T3-16 has a front LCD panel that displays the computer name that is currently connected to the system, the unit serial number, and the current network configuration. If no computer is connected to the unit, the LCD panel displays
Available.
The LCD panel also allows you to configure network settings for the analyzer. Use the Up and Down keypad buttons to cycle through the menu. Press the Center button to select.
Setup Static or Dynamic IP Mode
1. Use the Up and Down buttons to navigate to the Set IP Configuration menu.
2. Press the Center button to start the setup.
3. Use the Up and Down buttons to select Static or Dynamic IP configuration.
4. If you select the Static IP network configuration, you must specify the IP address, Subnet mask, and Gateway. Use the Up and Down buttons to navigate to the IP address, Subnet mask, or Gateway.
5. Press the Center button to enter edit mode.
6. In edit mode, use the Left and Right buttons to move the blinking cursor to any digit.
7. Use the Up and Down buttons to modify the digit.
8. After finishing modifications to the IP address, Subnet mask, or Gateway, press the Center button to return to the menu.
9. After you specify the IP address, Subnet mask, and Gateway, use the Up and Down buttons to navigate to the Accept and Reboot menu item. Press the Center button to apply the new settings and restart the analyzer.
Note: To cancel your modifications, select the Cancel Changes menu item.
Warning Do not open the enclosure. No operator serviceable
parts are inside.
5
Rear Panel Description
The rear panel of the Analyzer has:
Wide-range AC Connector Module
Power socket
Enclosed 5.0 A 250 V fuse
6
Interposers and Probes
Gen3 Slot Interposer
The Gen3 Slot Interposer is designed for use with the Summit T3-16 Analyzer and supports lane widths from x1 to x16 at data rates of 2.5 GT/s (Gen1),
5.0 GT/s (Gen2), or 8.0 GT/s (Gen3). A separate interposer is used for each lane width that you would like to probe. Lane reducers should NOT be used with the Gen3 Slot Interposer. For lane widths up to x8, one Analyzer Y-cable is required. For x16 applications, two Analyzer Y-cables are required.
Expansion Slot: Future use
Warning! For continued protection against fire, replace fuse
only with the type and rating specified above.
Warning Interposer has static-sensitive components.
Handle only at static-safe work stations.
Gen3 MidBus Probe
The Gen3 MidBus Probe is designed for use with the Summit T3-16 Analyzer and supports lane widths from x1 to x16 at data rates of 2.5 GT/s (Gen1),
5.0 GT/s (Gen2), or 8.0 GT/s (Gen3).
Page 3
7
Installing the Software
The PETracer software operates all of Teledyne LeCroy’s PCI Express Protocol Analyzer and Exerciser products and should be installed on a Microsoft® Windows®-based host machine. You need to install the PETracer software on the host machine before attaching the Analyzer to the system. The minimum requirements for the host machine are:
Windows® XP, Windows 7 (x86, x64) Windows 8 (x86, x64), Windows Server 2003, Windows Server 2008 and Windows Server 2012 R2
Processor with clock speed of 1 GHz or higher
2 GB of RAM
Hard drive with at least 250 MB of free space for software installation and additional space for storing recorded data
Display with at least 1024 x 768 resolution , with 16-bit color
USB 2.0 port
In general, the higher the performance of the host machine, the better user experience you get. Please refer to the Readme notes
8
Setting Up and Connecting the Summit T3-16 Analyzer
You can connect the Analyzer to the host machine using USB or Ethernet.
and Summit T3-16 User Manual for recommended configurations and additional information. User manuals for your Teledyne LeCroy PCI Express products can be found in Start > All Programs > LeCroy > PETracer > PCIe Protocol Suite Documents. To install the software, follow the steps below:
1. Insert the CD into the CD-ROM drive of the host machine that will control the Analyzer. The installation window displays links to software installation, user manuals, application notes, and data sheets.
2. Select Install PCIe Protocol Suite and follow the on-screen instructions. The PETracer software installs automatically on the host machine’s hard disk. During installation, all necessary USB drivers will be installed.
3. To start the application, launch the PETracer program from the Start menu: Start > All Programs > LeCroy > PETracer
> LeCroy PETracer
Software installation can also be downloaded from the Teledyne LeCroy website.
Using an Ethernet Connection
To set up the Analyzer using an Ethernet connection:
1. Install PETracer software on the host machine.
2. Connect the Analyzer to a 100-volt to 240-volt, 50 Hz to 60 Hz, 120 W power outlet using the provided power cord.
3. Connect the Ethernet cable between the Ethernet port on the Analyzer and a Ethernet port in the local network.
4. Turn on the power switch on the front of the analyzer.
Note 1: No driver installation is needed for Summit T3-16 to
operate over a network.
Note 2: At power-on, the Analyzer initializes and performs a
self-diagnostic. The results are reflected by messages on the Summit T3-16 LCD display. If the LCD display indicates failure, call Teledyne LeCroy Customer Support for assistance.
Note 3: Summit T3 is configured at the factory to use dynamic IP
setting and will get all required network parameters from the DHCP server on your network. If the DHCP server is not available, or to connect Summit T3 directly to the host machine, you can reconfigure the network settings of Summit T3 using the menus in the LCD display on the front of the analyzer (see the “LCD Menus” part of section 4 of this document).
Connecting to Summit T3-16 in the software
Start the PETracer software and perform the following procedure to connect to a Summit T3-16 analyzer over the network.
1. Select the Setup > All connected devices… menu in the PETracer application to display the Analyzer Devices dialog.
2. Select your Summit T3 device in the list and press the Connect button to execute the connection procedure. After the connection is established, the application displays the Connection Properties dialog.
3. Select an option:
Automatically connect to the device: When the
Ask if I want to connect to the device: When the
Take no action: When you start the application or when
Note: When you close the application on this computer (or
4. Press OK in the Connection Properties dialog. After you finish the connect procedure, the Summit T3-16 to which you have connected is marked as Ready and you can use it for recording.
application is started or when the named device appears on the network while the PETracer application is running on this computer, the PETra c er application will try to connect to the named device.
application is started or when the named device appears on the network while the PETracer application is running on this computer, the PETra c er application will display a message box allowing you to connect to the named device.
the named device appears on the network while the PETracer application is running on this computer, you must connect manually to use the named device.
you perform manual disconnect), the application disconnects from the device.
The Summit devices in the list are marked:
Locked: Some other client on the network is already connected to that device
Ready to connect: Available for connection
Note: To disconnect from a device, display this dialog,
select the device, and click the Disconnect button.
Note 2: Summit T3 will not appear in the Analyzer Device dialog
if it is connected on a different Subnet on your local network. You can connect to it on the Subnet by specifying its IP address in the Add Device dialog.
Page 4
Using a USB Connection
To set up the Analyzer using a USB connection:
1. Install PETracer software on the host machine.
2. Connect the Analyzer to a 100-volt to 240-volt, 50 Hz to 60 Hz, 120 W power outlet using the provided power cord.
3. Connect the USB port to a USB port on the host machine using a USB cable.
4. Turn on the power switch on the front of the analyzer.
Note: At power-on, the Analyzer initializes and performs a
self-diagnostic. The results are reflected by messages on the Summit T3-16 LCD display. If the LCD display indicates failure, call Teledyne LeCroy Customer Support for assistance.
9
Making Your First Recording
Connect Summit T3 to the system under test using an interposer or probe, and then configure the Recording Options. You can then test the Analyzer by creating a short snapshot recording.
Connect Summit T3 to System under Test using an Interposer or Probe
For a slot interposer, do the following:
1. Insert the slot interposer into a PCI Express slot that you want to monitor.
2. Insert the PCI Express device into the PCIe connector on top of the interposer.
3. Connect Summit T3 to the interposer with Analyzer Y cables. Connect cable Side A to the Upstream port of the Analyzer, and connect cable Side B to the Downstream port. Note: If you are using x16 interposer, the lower connector carries lanes 0 to 7, and the upper connector carries lanes 8 to 15.
4. Connect power to the interposer using the provided 12-V power adapter.
For Mid Bus Probe or other types of probes and interposers, please see the Summit T3-16 User Manual.
Configure Recording Options
1. From the Setup menu, select Recording Options.
2. Select the General tab and then choose Summit T3 as the Analyzer type.
3. In the Link section, specify the lane width of the PCI Express link to be analyzed. The rest of the settings in this section can be left at the factory defaults for most PCI Express systems.
4. For multi-lane PCI Express links, the Analyzer needs to observe link training to record link traffic correctly.
5. Click OK at the bottom of the Recording Options dialog box to apply the Analyzer recording settings specified.
Start Recording
5. Summit T3 will be recognized by the host machine as a USB device.
Follow the Microsoft instructions for the automatic installation of the Analyzer as a USB device on your analyzing host machine. (The required USB drivers are included on the Installation CD and were installed on your system when you installed the software.) Click Finish when you see the message that says “Windows has finished installing the software that your new hardware requires” and the file has been installed in your host machine.
8. Start the PETracer software, so you can start working with Summit T3.
®
Windows® on-screen Plug-and-Play
1. Start a recording session by clicking the Record button in the toolbar. Recording is tracked and reported on an activity meter in the status bar.
2. After the recording buffer fills, uploading starts automatically. PCI Express traffic is saved to the hard drive as the file specified in the recording options, and this file is opened in the application for you to view the traffic.
Note: You can interrupt a session by pressing the Stop button. If Recording is finished and Upload has started but has not finished, a message box appears. You can:
Continue uploading.
Abort the upload and flush the data.
Abort the upload and preserve all the previously saved data.
Teledyne LeCroy Customer Support
Online Download Periodically check the Teledyne LeCroy Protocol Solutions Group web site for software updates and other support related to this product. Software updates are available to users with a current Maintenance Agreement.
Trademarks and Servicemarks
Teledyne LeCroy and Summit T3-16 are trademarks of Teledyne LeCroy. Microsoft and Windows are registered trademarks of Microsoft Inc. All other trademarks are property of their respective companies.
© 2012 Teledyne LeCroy, Inc. All rights reserved. Part Number: 923221-00 Rev A This document may be printed and reproduced without additional permission, but all copies should contain this copyright notice.
Mail: 3385 Scott Blvd., Santa Clara, CA 95054-3115 Web: teledynelecroy.com/Support/SoftwareDownload/ E-mail: psgsupport@teledynelecroy.com Tel: (800) 909-7112 (USA and Canada) Tel: (408) 653-1260 (worldwide) Fax: (408) 727-6622 (worldwide)
Changes
Product specifications are subject to change without notice. Teledyne LeCroy reserves the right to revise the information in this document without notice or penalty.