Teledyne t100, t108u, t100h, t108, t100u Operation Manual

when used in conjunction with:
T100U addendum, PN 06840
T100H addendum, PN 07265
T108 addendum, PN 07268
T108 addendum, PN 07268
Operation Manual
Model T100
UV Fluorescence SO2 Analyzer
Model T108U Analyzer
T100U addendum, PN 06840, and
© TELEDYNE API (TAPI)
9970 CARROLL CANYON ROAD
SAN DIEGO, CA 92131-1106
USA
Toll-free Phone: 800-324-5190
Phone: +1 858-657-9800
Fax: +1 858-657-9816
Email: api-sales@teledyne.com
Website: http://www.teledyne-api.com/
Copyright 2010-2016 Teledyne API 05 August 2016
06807F DCN7335
06807F DCN7335

NOTICE OF COPYRIGHT

© 2010-2016 Teledyne API Inc. All rights reserved.

TRADEMARKS

All trademarks, registered trademarks, brand names or product names appea ring in thi s document are the property of their respective owners and are used herein for identification purposes only.
i
Teledyne API - T100 UV Fluorescenc e SO2 Analyzer
and/or the instrument.
06807F DCN7335
IMPORTANT SAFETY INFORMATION
Important safety messages are provided throughout this manual for the purpose of avoiding personal injury or instrument damage. Please read these messages carefully. Each safety message is associated with a safety alert symbol and placed throughout this manual and inside the instrument. The symbols with messages are defined as follows:
WARNING: Electrical Shock Hazard
HAZARD: Strong oxidizer
GENERAL WARNING/CAUTION: Read the accompanying
message for s pecifi c information.
CAUTION: Hot Surface Warning
Note
Do Not Touch: Touching some parts of the instrument without
protection or proper tools could result in damage to the part(s)
Technician Symbol: All operations marked with this symbol are to be performed by qualified maintenance personnel only.
Electrical Ground: This symbol inside the instrument marks the central safety grounding point for the instrument.
CAUTION
GENER AL S AFETY HAZARD
The T100 Analyzer should only be used for the purpose and in the manner described in this manual. If you use the T100 in a manner other than that for which it was intended, unpredictable behavior could ensue with possible hazardous consequences.
NEVER use any gas analyzer to sample combustible gas(es).
Technical Assistance regarding the use and maintenance of the T100 or any other Teledyne API product can be obtained by contacting Teledyne API’s Technical Support Department:
Phone: 800-324-5190
Email: sda_techsupport@teledyne.com
or by accessing various service options on our website at
http://www.teledyne-api.com/
.
ii
AVERTISSEMENT : Risque de choc électrique
DANGER : Oxydant puissant
AVERTISSEMENT GÉNÉRAL / MISE EN GARDE : Lire la consigne
MISE EN GARDE : Surface chaude
Ne pas toucher
Mise à la terre : Ce symbole à l’intérieur de l’instrument détermine le point central
06807F DCN7335
CONSIGNES DE SÉCURITÉ
Des consignes de sécurité importantes sont fourn ies tout au long du présent m anuel dans le but d’éviter des blessures corporelles ou d’endommager les instruments. Veuillez lire attentivement ces consignes. Chaque consigne de sécuri té est représenté e par un pictogr amme d’aler te de sécurité; c es pictogram mes se retrouven t dans ce manuel et à l’intérieur des instruments. Les symboles correspondent aux consignes suivantes :
complémentaire pour des renseignements spécifiques
: Toucher à certaines parties de l’instrument sans protection ou sans les outils appropriés pourrait entraîner des dommages aux pièces ou à l’instrument.
Pictogramme « technicien » : Toutes les opérations portant ce symbole doivent être effectuées uniquement par du personnel de maintenance qualifié.
de la mise à la terre sécuritaire de l’instrument.
MISE EN GARDE
Cet instrument doit être utilisé aux fins décrites et de la manière décrite dans ce manuel. Si vous utilisez cet instrument d’une autre manière que celle pour laquelle il a été prévu, l’instrument pourrait se comporter de façon imprévisible et entraîner des conséquences dangereuses.
NE JAMAIS utiliser un analyseur de gaz pour échantillonner des gaz combustibles!
iii
Teledyne API - T100 UV Fluorescenc e SO2 Analyzer
arge (ESD) handling and packing
ESD handling and packing instructions
06807F DCN7335
WARRANTY
WARRANTY POLICY (02024J)
Teledyne API (TAPI), a business unit of Teledyne Instruments, Inc., provides that: Prior to shipment, TAPI equipment is thoroughly inspected and tested. Should equipment
failure occur, TAPI assures its customers that prompt service and support will be available. (For the instrument-specific warranty period, please refer to the “Limited Warranty” section in the Terms and Conditions of Sale on our website at the following link: http://www.teledyne-api.com/terms_and_conditions.asp).
COVERAGE
After the warranty period and throughout the equipment lifetime, TAPI stands ready to provide on-site or in-plant service at reasonable rates similar to those of other manufacturers in the industry. All maintenance and the first level of field troubleshooting are to be performed by the customer.
NON-TAPI MANUFACTURED EQUIPMENT
Equipment provided but not manufactured by TAPI is warranted and will be repaired to the extent and according to the current terms and conditions of the respective equipment manufacturer’s warranty.
PRODUCT RETURN
All units or components returned to Teledyne API should be properly packed for handling and returned freight prepaid to the nearest designated Service Center. After the repair, the equipment will be returned, freight prepaid.
The complete Terms and Conditions of Sale can be reviewed at
http://www.teledyne-api.com/terms_and_conditions.asp
Failure to comply with proper anti-Electro-Static Disch instructions and Return Merchandise Authorization (RMA) procedures when returning parts for repair or calibration may void your warranty. For anti­please refer to the manual, Fundamentals of ESD, PN 04786, in its “Packing Components for Return to Teledyne A PI’s Customer Service” section. The manual can be downloa ded from our website at
http://www.teledyne-api.com. RMA procedures can also be found on our website.
CAUTION – Avoid Warranty Invalidation
iv
06807F DCN7335
ABOUT THIS MANUAL
Presented here is information regarding the documents that are included with this manual ( content is organized ( in this manual (

STRUCTURE

This T100 manual, PN 06807, is comprised of multiple documents, assembled i n PDF format, as listed below.
Part No. Rev Name/Description
06807 F Operation Manual, T100 UV Fluorescence SO2 Analyzer 05036 F Appendix A, Menu Trees and related software documentation 06845 A Spare Parts List (in Appendix B of this manual) 04357 A AKIT, Expendables, basic (in Appendix B of this manual) 01475 A AKIT, Expendables, IZS (in Appendix B of this manual) 04728 A AKIT, Spares (in Appendix B of this manual) 04796 F Appendix C, Repair Form 06908 B Interconnect Diagram (in Appendix D of this manual)
Structure), its history of release and revisions (Revision History), how the
Organization), and the conventions used to present the information
Conventions Used).
Note We recommend that this manual be read in its entirety before any attempt
is made to operate the instrument.

CONVENTIONS USED

In addition to the safety symbols as presented in the Important Safety Information page, this manual provides special notices related to the safety and effective use of the analyzer and other pertinent information.
Special Notices appea r as follow s:
ATTENTION
COULD DAMAGE INSTRUMENT AND VOID WARRANTY
This special notice provides information to avoid damage to your instrument and possibly invalidate the warranty.
IMPORTANT IMPACT ON READINGS OR DATA
Could either affect accuracy of instrument readings or cause loss of data.
Note Pertinent information associated with the proper care, operation or
maintenance of the analyzer or its parts.
v
Teledyne API - T100 UV Fluorescenc e SO2 Analyzer
Date
Rev
DCN
Change Summary
06807F DCN7335
REVISION HISTORY
This section provides information regarding the history of changes to this manual.
T100 Manual, PN06807
2016 Aug 05 F 7335 Administrative updates. 2016 April 19 E 7230 Clarified Range setup when using dilution factor option; other administrative fixes. 2015 June 28 D 7088 Condensed content; implemented DCRs 2013 Apr 22 C 6650 Administrative corrections; technic al corre ctions 2011 Aug 22 B 6192 Administrative change: reorganized structure.
Technical Updates: added MODBUS Quick Setup (Section 6.6.1), update Appendices A and D with latest revisions.
2010 Sep 7 A 5834 Initial release
vi
06807F DCN7335
TABLE OF CONTENTS
Important Safety Information .............................................................................................................................. ii
CONSIGNES DE SÉCURITÉ ............................................................................................................................... iii
Warranty ...............................................................................................................................................................iv
About This Manual ............................................................................................................................................... v
Revision History ..................................................................................................................................................vi
TABLE OF CONTENTS .......................................................................................................... VII
List of Figures......................................................................................................................................................xi
List of Tables .....................................................................................................................................................xiv
1. INTRODUCTION, FEATURES AND OPTIONS ................................................................. 17
T100 Overview .............................................................................................................................................17 1.1.
Features .......................................................................................................................................................18
T100 Documentation ...................................................................................................................................18
O
ptions .........................................................................................................................................................18
2. SPECIFICATIONS, APPROVALS & COMPLIANCE ......................................................... 23
Specifications and Approvals ....................................................................................................................23 2.1.
E
PA Equivalency Designation ...................................................................................................................25
Approvals and Certifications .....................................................................................................................25
E
MC .......................................................................................................................................................25
S
afety .....................................................................................................................................................25
O
ther Type Certifications .......................................................................................................................25
3. GETTING STARTED .......................................................................................................... 27
Unpacking the T100 Analyzer ....................................................................................................................27 3.1.
V
entilation Clearance .............................................................................................................................28
Instrument Layout .......................................................................................................................................29
F
ront Panel.............................................................................................................................................29
R
ear Panel .............................................................................................................................................33
I
nternal Chassis Layout .........................................................................................................................35
onnections and Setup ..............................................................................................................................36
C
E
lectrical Connections ...........................................................................................................................36
P
neumatic Connections .........................................................................................................................52
Startup, Functional Checks, and Initial Calibration .................................................................................65
S
tartup ....................................................................................................................................................65
W
arning Messages ................................................................................................................................65
Functional Checks .................................................................................................................................67
I
nitial Calibr a tion ....................................................................................................................................69
4. OVERVIEW OF OPERATING MODES .............................................................................. 75
Sample Mode ...............................................................................................................................................77 4.1.
T
est Functions .......................................................................................................................................77
Warning Messages ................................................................................................................................80
Calibration Mode .........................................................................................................................................81
S
etup Mode ..................................................................................................................................................81
P
assword Security .................................................................................................................................81
P
rimary Setup Menu ..............................................................................................................................82
S
econdary Setup Menu (SETU P>MO RE) .............................................................................................82
5. SETUP MENU .................................................................................................................... 83
SETUP – CFG: Configuration Information ................................................................................................83 5.1.
SET
UP – ACAL: Automatic Calibration Option ........................................................................................83
SET
UP – DAS: Internal Data Acquisition System ....................................................................................84
SET
UP – RNGE: Analog Output Reporting Range Configuration..........................................................84
1.2.
1.3.
1.4.
2.2.
2.3.
2.3.1.
2.3.2.
2.3.3.
3.1.1.
3.2.
3.2.1.
3.2.2.
3.2.3.
3.3.
3.3.1.
3.3.2.
3.4.
3.4.1.
3.4.2.
3.4.3.
3.4.4.
4.1.1.
4.1.2.
4.2.
4.3.
4.3.1.
4.3.2.
4.3.3.
5.2.
5.3.
5.4.
vii
Teledyne API - T100 UV Fluorescence SO
06807F DCN7335
Available Analog Output Sig na ls ...........................................................................................................84 5.4.1.
Physical Range versus Analog Output Reporting Ranges ....................................................................85
R
eporting Range Modes: Single, Dual, Auto Ranges ...........................................................................86
R
ange Units ...........................................................................................................................................90
D
ilution Ratio (Option) ............................................................................................................................92
SETUP – PASS: Password Protection ......................................................................................................93
SETUP – CLK: Setting the Internal Time-of-Day Clock ...........................................................................96
SET
UP – COMM: Communications Ports .................................................................................................98
I
D (Instrument Identification) ..................................................................................................................98
I
NET (Ethernet) ......................................................................................................................................99
C
OM1 and COM2 (Mode, Baud Rate and Test Port) ............................................................................99
SETUP – VARS: Variables Setup and Definition ....................................................................................100
SETUP – DIAG: Diagnostics Functions ..................................................................................................102
S
ignal I/O .............................................................................................................................................104
A
nalog Output Step Test ......................................................................................................................105
A
nalog I/O Configuration ......................................................................................................................106
O
ptic Test .............................................................................................................................................119
E
lectrical Test ......................................................................................................................................120
Lamp Calibration ..................................................................................................................................121
P
ressure Calibration ............................................................................................................................122
F
low Calibration ...................................................................................................................................123
T
est Channel Output ............................................................................................................................124
Analyzer
2
6. COMMUNICATIONS SETUP AND OPERATION ............................................................ 127
Data Terminal / Communication Equipment (DTE DCE) .......................................................................127 6.1.
C
ommunication Modes, Baud Rate and Port testing ............................................................................127
C
ommunication Modes ........................................................................................................................128
C
OMM Port Baud Rate ........................................................................................................................130
C
OMM Port Testing .............................................................................................................................131
RS-232 ........................................................................................................................................................131
RS-485 (Option) .........................................................................................................................................132
E
thernet ......................................................................................................................................................132
C
onfiguring Ethernet Com munication Manually (Static IP Addr ess ) ...................................................133
C
onfiguring Ethernet Communication Using Dynamic Host Configuration Protocol (DHCP) .............135
U
SB Port for Remote access ...............................................................................................................137
ommunications Protocols .....................................................................................................................139
C
M
ODBUS .............................................................................................................................................139
H
ESSEN ..............................................................................................................................................141
7. DATA ACQUISITION SYSTEM (DAS) AND APICOM ..................................................... 147
DAS Structure ............................................................................................................................................148 7.1.
DA
S Channels .....................................................................................................................................148
D
AS Parameters ..................................................................................................................................149
D
AS Triggering Events ........................................................................................................................150
Default DAS Channels ..............................................................................................................................150
V
iewing DAS Data and Settings ..........................................................................................................153
7.
3Editing DAS Data Channels ............................................................................................................154
Trigger Events ......................................................................................................................................156
E
diting DAS Parameters ......................................................................................................................157
S
ample Period and Report Period .......................................................................................................159
N
umber of Records ..............................................................................................................................160
RS-232 C
Starting Date ........................................................................................................................................162
D H
APICOM Remote Control Program ..........................................................................................................165
R
emote DAS Configuration via APICOM ................................................................................................166
Report Function ......................................................................................................................162
ompact Report ...................................................................................................................................162
isabling/Enabling Data Channels ....................................................................................................163
OLDOFF Feature ............................................................................................................................164
5.4.2.
5.4.3.
5.4.4.
5.4.5.
5.5.
5.6.
5.7.
5.7.1.
5.7.2.
5.7.3.
5.8.
5.9.
5.9.1.
5.9.2.
5.9.3.
5.9.4.
5.9.5.
5.9.6.
5.9.7.
5.9.8.
5.9.9.
6.2.
6.2.1.
6.2.2.
6.2.3.
6.3.
6.4.
6.5.
6.5.1.
6.5.2.
6.5.3.
6.6.
6.6.1.
6.6.2.
7.1.1.
7.1.2.
7.1.3.
7.2.
7.2.1.
7.2.2.
7.2.3.
7.2.4.
7.2.5.
7.2.6.
7.2.7.
7.2.8.
7.2.9.
7.2.10.
7.2.11.
7.3.
7.4.
viii
Teledyne API - T100 UV Fluorescence SO2 Analyzer
06807F DCN7335
8. REMOTE OPERATION OF THE ANALYZER .................................................................. 169
Remote Operation Using the External Digital I/O ..................................................................................169 8.1.
Status Outputs .....................................................................................................................................169
C
ontrol Inputs .......................................................................................................................................170
Remote Operation Using the External Serial I/O ...................................................................................172
T
erminal Operating Modes ..................................................................................................................172
Help Commands in Terminal Mode .....................................................................................................172
C
ommand Syntax ................................................................................................................................173
D
ata Types ...........................................................................................................................................173
S
tatus Reporting ..................................................................................................................................174
Remote Access by Modem.......................................................................................................................175
COM Port Password Security ..................................................................................................................178
A
dditional Communications Documentation .........................................................................................178
9. CALIBRATION PROCEDURES ....................................................................................... 179
Calibration Preparations ..........................................................................................................................179 9.1.
R
equired Equipment, Supplies, and Expendables ..............................................................................179
D
ata Recording Devices ......................................................................................................................181
anual Calibration ....................................................................................................................................182
M M
anual Calibration Checks ......................................................................................................................185
M
anual Calibration with Zero/Span Valves.............................................................................................186
M
anual Calibration with IZS Option ........................................................................................................189
M
anual Calibration Checks with IZS or Zero/Span Valves ...................................................................189
Manual Calibration in DUAL or AUTO Reporting Range Modes ..........................................................192
C
alibration With Remote Contact Closures .........................................................................................192
Automatic Calibration (AutoCal) .............................................................................................................193
C
alibration Quality ....................................................................................................................................196
C
alibration of Optional Sensors ............................................................................................................197
Sensor Calibration ........................................................................................................................197 9.10.1.
O
2
Sensor Calibration ......................................................................................................................201 9.10.2.
CO
EPA Protocol Calibration .......................................................................................................................205
2
10. INSTRUMENT MAINTENANCE .................................................................................... 207
Maintenance Schedule ...........................................................................................................................209 10.1.
P
redictive Diagnostics............................................................................................................................211
Maintenance Procedures........................................................................................................................212
C
hanging the Sample Particulate Filter .............................................................................................212
C
hanging the IZS Permeation Tube ..................................................................................................213
C
hanging the External Zero Air Scrubber ..........................................................................................213
C
hanging the Critical Flow Orifice .....................................................................................................214
C
hecking for Light Leaks ...................................................................................................................215
Detailed Pressure Leak Check ..........................................................................................................216
P
erforming a Sample Flow Check .....................................................................................................217
Hydrocarbon Scrubber (Kicker) .........................................................................................................217
11. TROUBLESHOOTING & SERVICE ............................................................................... 219
General Troubleshooting .......................................................................................................................220 11.1.
F
ault Diagnostics with Warning Messages ........................................................................................220
F
ault Diagnosis with Test Functions ..................................................................................................223
U
sing the Diagnostic Signal I/O Functions ........................................................................................225
Status LEDs .............................................................................................................................................227
M
otherboard Status Indicator (Watchdog) .........................................................................................227
C
PU Status Indicators ........................................................................................................................227
Relay Board Status LEDs ..................................................................................................................228
Gas Flow Problems .................................................................................................................................228
Z
ero or Low Sample Flow ..................................................................................................................228
H
igh Flow ...........................................................................................................................................229
Calibration Problems ..............................................................................................................................229
N
egative Concentrations....................................................................................................................229
8.1.1.
8.1.2.
8.2.
8.2.1.
8.2.2.
8.2.3.
8.2.4.
8.2.5.
8.3.
8.4.
8.5.
9.1.1.
9.1.2.
9.2.
9.3.
9.4.
9.5.
9.6.
9.7.
9.7.1.
9.8.
9.9.
9.10.
9.11.
10.2.
10.3.
10.3.1.
10.3.2.
10.3.3.
10.3.4.
10.3.5.
10.3.6.
10.3.7.
10.3.8.
11.1.1.
11.1.2.
11.1.3.
11.2.
11.2.1.
11.2.2.
11.2.3.
11.3.
11.3.1.
11.3.2.
11.4.
11.4.1.
ix
Teledyne API - T100 UV Fluorescence SO
06807F DCN7335
No Response .....................................................................................................................................229 11.4.2.
Unstable Zero and Span ....................................................................................................................230
I
nability to Span - No SPAN Button ...................................................................................................230
I
nability to Zero - No ZERO Button ....................................................................................................231
N
on-Linear Response ........................................................................................................................231
D
iscrepancy Between Analog Output and Display ............................................................................232
ther Performance Problems ................................................................................................................232
O
E
xcessive noise .................................................................................................................................232
Slo
w Response ..................................................................................................................................232
T
he Analyzer Doesn’t Appear on the LAN or Internet .......................................................................232
Subsystem Checkout ..............................................................................................................................233
A
C Power Configuration ....................................................................................................................233
DC Power Supply ...............................................................................................................................234
2
C Bus ...............................................................................................................................................235 11.6.3.
I T
ouch-screen Interface ......................................................................................................................235
L
CD Display Module ..........................................................................................................................235
R
elay Board .......................................................................................................................................236
M
otherboard .......................................................................................................................................236
CPU ....................................................................................................................................................238
RS-23
Service Proced ures .................................................................................................................................243
Disk-on-Module Replacement ...........................................................................................................243 11.7.1.
S
Frequently Asked Questions (FAQs) ....................................................................................................260
T
echnical Assistance ..............................................................................................................................261
2 Communication .....................................................................................................................238
S
hutter System ................................................................................................................................239
P
MT Sensor .....................................................................................................................................240
P
MT Preamplifier Board ...................................................................................................................240
PMT Temperature Control PCA .......................................................................................................240
H
igh Voltage Power Supply .............................................................................................................240
P
neumatic Sensor Assembly ...........................................................................................................241
S
ample Pressure .............................................................................................................................242
I
ZS Option ........................................................................................................................................242
B
ox Temperature .............................................................................................................................242
PMT Temperature ............................................................................................................................242
ensor Module Repair & Cleaning ....................................................................................................245
Analyzer
2
12. PRINCIPLES OF OPERATION ...................................................................................... 263
Sulfur Dioxide (SO2) Sensor Principles of operation ..........................................................................263 12.1.
Ultraviolet Fluorescence Measurement Principle .......................................................................263 12.1.1.
SO
2
T
he UV Light Path ..............................................................................................................................266
U
V Source Lamp ................................................................................................................................267
The Reference Detector .....................................................................................................................268
T
he PMT ............................................................................................................................................268
U
V Lamp Shutter & PMT Offset .........................................................................................................268
O
ptical Filters .....................................................................................................................................269
O
ptical Lenses ...................................................................................................................................271
M
easurement Interferences ...............................................................................................................272
xygen (O
O
P O
Carbon Dioxide (CO
N CO E
Pneumatic Operation ..............................................................................................................................277
S F H
) Sensor Principles of Operation .......................................................................................273 12.2.
2
aramagnetic Measurement of O
Sensor Operation within the T100 Analyzer .................................................................................274 12.2.2.
2
DIR Measurement of CO
Operation within the T100 Analyzer ...........................................................................................276 12.3.2.
2
lectronic Operation of the CO
ample Gas Flow ...............................................................................................................................277
low Rate Control ..............................................................................................................................278
ydrocarbon Scrubber (Kicker) .........................................................................................................279
) Sensor Principles of Operation .......................................................................275 12.3.
2
2
.....................................................................................................273 12.2.1.
2
...............................................................................................................275 12.3.1.
Sensor ............................................................................................276 12.3.3.
2
11.4.3.
11.4.4.
11.4.5.
11.4.6.
11.4.7.
11.5.
11.5.1.
11.5.2.
11.5.3.
11.6.
11.6.1.
11.6.2.
11.6.4.
11.6.5.
11.6.6.
11.6.7.
11.6.8.
11.6.9.
11.6.10.
11.6.11.
11.6.12.
11.6.13.
11.6.14.
11.6.15.
11.6.16.
11.6.17.
11.6.18.
11.6.19.
11.7.
11.7.2.
11.8.
11.9.
12.1.2.
12.1.3.
12.1.4.
12.1.5.
12.1.6.
12.1.7.
12.1.8.
12.1.9.
12.4.
12.4.1.
12.4.2.
12.4.3.
x
Teledyne API - T100 UV Fluorescence SO2 Analyzer
06807F DCN7335
Pneumatic Sensors ............................................................................................................................280 12.4.4.
Electronic Operation ...............................................................................................................................281
C
PU ....................................................................................................................................................283
S
ensor Module ...................................................................................................................................284
P
hoto Multiplier Tube (PMT) ..............................................................................................................286
P
MT Cooling System .........................................................................................................................288
PMT Preamplifier ...............................................................................................................................289
P
neumatic Sensor Board ...................................................................................................................291
R
elay Board .......................................................................................................................................291
M
otherboard .......................................................................................................................................293
A
nalog Outputs ..................................................................................................................................294
E
xternal Digital I/O ...........................................................................................................................295
2
C Data Bus ....................................................................................................................................295 12.5.11.
I P
ower up Circuit ...............................................................................................................................295
P
ower Supply/ Circuit Breaker .........................................................................................................295
Front Panel/Display Interface ................................................................................................................297
L
VDS Transmitter Board ....................................................................................................................297
F
ront Panel Interface PCA .................................................................................................................297
oftware Operation .................................................................................................................................298
S
A
daptive Filter ....................................................................................................................................298
C
alibration - Slope and Offset ............................................................................................................299
T
emperature and Pressure Compensation (TPC) Feature ...............................................................299
I
nternal Data Acquisition System (DAS) ............................................................................................300
GLOSSARY ........................................................................................................................... 301
INDEX .................................................................................................................................... 305
12.5.
12.5.1.
12.5.2.
12.5.3.
12.5.4.
12.5.5.
12.5.6.
12.5.7.
12.5.8.
12.5.9.
12.5.10.
12.5.12.
12.5.13.
12.6.
12.6.1.
12.6.2.
12.7.
12.7.1.
12.7.2.
12.7.3.
12.7.4.
APPENDIX A - VERSION SPECIFIC SOFTWARE DOCUMENTATION APPENDIX B - SPARE PARTS APPENDIX C - REPAIR QUESTIONNAIRE APPENDIX D - INTERCONNECT DIAGRAM
LIST OF FIGURES
Figure 3-1: Front Panel Layout .......................................................................................................................29
Figure 3-2: Display Screen and Touch Control ..............................................................................................30
Figure 3-3: Display/Touch Control Screen Mapped to Menu Charts .............................................................32
Figure 3-4: Rear Panel Layout .......................................................................................................................33
Figure 3-5: Internal Layout, Basic (no Valve or Second Gas Options) ..........................................................35
Figure 3-6: Analog In Connector ....................................................................................................................37
Figure 3-7: Analog Output Connector ............................................................................................................38
Figure 3-8: Current Loop Option Installed on the Motherboard .....................................................................40
Figure 3-9: Status Output Connector .............................................................................................................41
Figure 3-10: Control Input Connector ...............................................................................................................43
Figure 3-11: Concentration Alarm Relay ..........................................................................................................44
Figure 3-12: Rear Panel Connector Pin-Outs for RS-232 Mode ......................................................................47
Figure 3-13: Default Pin Assignments for CPU Com Port Connector (RS-232) ..............................................48
Figure 3-14: Jumper and Cables for Multidrop Mode .......................................................................................50
Figure 3-15: RS-232-Multidrop PCA Host/Analyzer Interconnect Diagram .....................................................51
Figure 3-16: Pneumatic Connections–Basic Configuration–Using Bottled Span Gas .....................................55
Figure 3-17: Pneumatic Connections–Basic Configuration–Using Gas Dilution Calibrator .............................55
Figure 3-18: T100 Gas Flow, Basic Configuration ...........................................................................................56
Figure 3-19: Pneumatic Layout with Zero/Span Valves Option .......................................................................57
xi
Teledyne API - T100 UV Fluorescence SO
06807F DCN7335
Analyzer
2
Figure 3-20: Pneumatic Layout with Pressurized Span/Ambient Zero Option .................................................58
Figure 3-21: Pneumatic Layout with IZS Options .............................................................................................59
Figure 3-22: Pneumatic Layout with O2 Sensor ...............................................................................................62
Figure 3-23: Pneumatic Layout with CO2 Sensor.............................................................................................63
Figure 3-24: Warning Messages ......................................................................................................................65
Figure 3-25: Functional Check .........................................................................................................................68
Figure 3-26: Reporting Range Verification .......................................................................................................70
Figure 3-27: Dilution Ratio Setup .....................................................................................................................71
Figure 3-28: SO2 Span Gas Setting .................................................................................................................72
Figure 3-29: Zero/Span Calibration Procedure ................................................................................................73
Figure 4-1: Front Panel Display ......................................................................................................................76
Figure 4-2: Viewing T100 TEST Functions ....................................................................................................79
Figure 4-3: Viewing and Clearing T100 WARNING Messages ......................................................................80
Figure 5-1: SETUP – Configurat ion Inf or mation ............................................................................................83
Figure 5-2: SETUP – Analog Output Connector ............................................................................................84
Figure 5-3: SETUP RNGE – Reporting Range Mode ....................................................................................86
Figure 5-4: SETUP RNGE – Single Range Mode ..........................................................................................87
Figure 5-5: SETUP RNGE – Dual Range Mode ............................................................................................88
Figure 5-6: SETUP RNGE – Auto Range Mode ............................................................................................89
Figure 5-7: SETUP RNGE – Concentration Units Selection ..........................................................................90
Figure 5-8: SETUP RNGE – Dilution Ratio ....................................................................................................92
Figure 5-9: SETUP – Enable Password Security ...........................................................................................94
Figure 5-10: SETUP – Enter Calibration Mode Using Password .....................................................................95
Figure 5-11: SETUP – Clock ............................................................................................................................96
Figure 5-12: SETUP – Clock Speed Variable ..................................................................................................97
Figure 5-13: SETUP – COMM Menu ................................................................................................................98
Figure 5-14: COMM – Machine ID ..................................................................................................................99
Figure 5-15: SETUP – VARS Menu ...............................................................................................................101
Figure 5-16: DIAG Menu ................................................................................................................................103
Figure 5-17: DIAG – Signal I/O Menu ............................................................................................................104
Figure 5-18: DIAG – Analog Output Menu .....................................................................................................105
Figure 5-19: DIAG – Analog I/O Configuration Menu .....................................................................................108
Figure 5-20: DIAG – Analog Output Calibration Mode ...................................................................................109
Figure 5-21: DIAG – Analog Output Calibration Mode – Single Analog Channel ..........................................110
Figure 5-22: DIAG – Analog Output – Auto Cal or Manual Cal Selection for Channels ................................111
Figure 5-23: Setup for Calibrating Analog Outputs ........................................................................................112
Figure 5-24: Analog Output – Voltage Adjustment .........................................................................................113
Figure 5-25: Analog Output – Offset Adjustment ...........................................................................................114
Figure 5-26: Setup for Calibrating Current Outputs .......................................................................................115
Figure 5-27: Analog Output – Zero and Span Value Adjustment for Current Outputs ...................................116
Figure 5-28: DIAG – Analog Output – AIN Calibration ...................................................................................117
Figure 5-29. DIAG – Analog Inputs (Option) Configuration Menu .................................................................118
Figure 5-30: DIAG – Optic Test ......................................................................................................................119
Figure 5-31: DIAG – Electrical Test ................................................................................................................120
Figure 5-32: DIAG – Lamp Calibration ...........................................................................................................121
Figure 5-33: DIAG – Pressure Calibration......................................................................................................122
Figure 5-34: DIAG – Flow Calibration ............................................................................................................123
Figure 5-35: DIAG – Test Channel Output .....................................................................................................124
Figure 6-1: COMM – Communication Modes Set up ....................................................................................129
Figure 6-2: COMM – COMM Port Baud Rate ..............................................................................................130
Figure 6-3: COMM – COM1 Test Port..........................................................................................................131
Figure 6-4: COMM – LAN / Internet Manual Configuration ..........................................................................134
Figure 6-5: COMM – LAN / Internet Automatic Configuration ......................................................................135
Figure 6-6: COMM – Change Hostname ....................................................................................................136
Figure 6-7: COMM – Activating Hessen Protocol ........................................................................................142
Figure 6-8: COMM – Select Hessen Protocol Type .....................................................................................143
Figure 6-9: COMM – Select Hessen Protocol Response Mode ...................................................................144
xii
Teledyne API - T100 UV Fluorescence SO2 Analyzer
06807F DCN7335
Figure 6-10: COMM – Status Flag Bit Assignment ........................................................................................146
Figure 7-1: Default DAS Channels Setup ....................................................................................................152
Figure 7-2: DAS – Data Acquisition Me nu ...................................................................................................153
Figure 7-3: DAS – Editing DAS Data Channels ...........................................................................................154
Figure 7-4: DAS – Editing Data Channel Name ...........................................................................................155
Figure 7-5: DAS – Trigger Events ................................................................................................................156
Figure 7-6: DAS – Editing DAS Parameters ................................................................................................157
Figure 7-7: DAS – Configuring Parameters for a Specific Data Parameter .................................................158
Figure 7-8: DAS – Define the Report Period ................................................................................................160
Figure 7-9: DAS – Edit Number of Records .................................................................................................161
Figure 7-10: DAS – RS-232 Report Function .................................................................................................162
Figure 7-11: DAS – Disabling / Enabling Data Channels ...............................................................................163
Figure 7-12: DAS – Holdoff Feature ...............................................................................................................164
Figure 7-13: APICOM Remote Control Program Interface .............................................................................165
Figure 7-14: Sample APICOM User Interface for Configuring the DAS .........................................................166
Figure 7-15: DAS Configuration Through a Terminal Emulation Program .....................................................167
Figure 8-1: Status Output Connector ...........................................................................................................170
Figure 8-2: Control Inputs with Local 5 V Power Suppl y ..............................................................................171
Figure 8-3: Control Inputs with External 5 V Power Supply .........................................................................172
Figure 8-4: COMM – Remote Access by Modem ........................................................................................176
Figure 8-5: COMM – Initialize the Modem ...................................................................................................177
Figure 9-1: Setup for Manual Calibration without Z/S valve or IZS Option (Step 1) ....................................182
Figure 9-2: Setup for Manual Calibration without Z/S valve or IZS Option (Step 2) ....................................183
Figure 9-3: Setup for Manual Calibration without Z/S valve or IZS Option (Step 3) ....................................184
Figure 9-4: Setup for Manual Calibration Checks ........................................................................................185
Figure 9-5: Setup for Manual Calibration with Z/S Valve Option Installed (Step 1) .....................................186
Figure 9-6: Setup for Manual Calibration with Z/S Valve Option Installed (Step 2) .....................................187
Figure 9-7: Setup for Manual Calibration with Z/S Valve Option Installed (Step 3) .....................................188
Figure 9-8: Manual Calibration with IZS Optio n ...........................................................................................189
Figure 9-9: Setup for Manual Calibration Check with Z/S Valve or IZS Option (Step 1) .............................190
Figure 9-10: Setup for Manual Calibration Check with Z/S Valve or IZS Option (Step 2) .............................191
Figure 9-11: Manual Calibration in Dual/Auto Reporting Range Modes ........................................................192
Figure 9-12: AUTO CAL – User Defined Sequence .......................................................................................195
Figure 9-13: O2 Sensor Calibration Set Up ....................................................................................................197
Figure 9-14: O2 Span Gas Concentration Set Up ..........................................................................................198
Figure 9-15: Activate O2 Sensor Stability Function ........................................................................................199
Figure 9-16: O2 Zero/Span Calibration ...........................................................................................................200
Figure 9-17: CO2 Sensor Calibration Set Up ..................................................................................................201
Figure 9-18: CO2 Span Gas Concentration Setup .........................................................................................202
Figure 9-19: Activate CO2 Sensor Stability Function .....................................................................................203
Figure 9-20: CO2 Zero/Span Calibration ........................................................................................................204
Figure 10-1: Sample Particulate Filter Assembly ...........................................................................................212
Figure 10-2: Critical Flow Orifice Assembly ...................................................................................................214
Figure 10-3: Simple Leak Check Fixture ........................................................................................................217
Figure 10-4: Hydrocarbon Scrubber Leak Check Setup ................................................................................218
Figure 11-1: Viewing and Clearing Warning Messages .................................................................................221
Figure 11-2: Example of Signal I/O Function .................................................................................................226
Figure 11-3: CPU Status Indicator .................................................................................................................227
Figure 11-4: Location of Relay Board Power Configuration Jumper ..............................................................234
Figure 11-5: Manual Activation of the UV Light Shutter .................................................................................239
Figure 11-6: Sensor Module Wiring and Pneumatic Fittings ..........................................................................245
Figure 11-7: Sensor Module Mounting Screws ..............................................................................................246
Figure 11-8: Sample Chamber Mounting Bracket ..........................................................................................247
Figure 11-9: Hex Screw Between Lens Housing and Sample Chamber .......................................................248
Figure 11-10: UV Lens Housing / Filter Housing ..............................................................................................249
Figure 11-11: PMT UV Filter Housing Disassembled ......................................................................................249
Figure 11-12: Disassembling the Shutter As sembly ........................................................................................251
xiii
Teledyne API - T100 UV Fluorescence SO
06807F DCN7335
Analyzer
2
Figure 11-13: Shutter Assembly .......................................................................................................................252
Figure 11-14. UV Lamp Adjustment .................................................................................................................253
Figure 11-15: Location of UV Reference Detector Pot ent iometer ...................................................................254
Figure 11-16: PMT Assembly - Explod ed Vie w ................................................................................................256
Figure 11-17: Pre-Amplifier Board (Preamp PCA) Layout ...............................................................................258
Figure 12-1: UV Absorption ............................................................................................................................264
Figure 12-2: UV Light Path .............................................................................................................................267
Figure 12-3: Source UV Lamp Construction ..................................................................................................268
Figure 12-4: Excitation Lamp UV Spectrum Before/After Filtration ................................................................269
Figure 12-5: PMT Optical Filter Bandwidth ....................................................................................................270
Figure 12-6: Effects of Focusing Source UV in Sample Chamber .................................................................271
Figure 12-7: Oxygen Sensor - Principles of Operation ..................................................................................274
Figure 12-8: CO2 Sensor Principles of Operation ..........................................................................................275
Figure 12-9: CO2 Sensor Option PCA Layout and Electronic Connections ...................................................276
Figure 12-10: Gas Flow and Location of Critical Flow Orifice ..........................................................................277
Figure 12-11: Flow Control Assembly & Critical Flow Orifice...........................................................................278
Figure 12-12: T100 Hydrocarbon Scrubber (Kicker) ........................................................................................279
Figure 12-13: T100 Electronic Block Diagram .................................................................................................281
Figure 12-14: CPU Board Annotated ...............................................................................................................283
Figure 12-15: T100 Sensor Module ..................................................................................................................284
Figure 12-16: T100 Sample Chamber ..............................................................................................................285
Figure 12-17: PMT Housing Assembly .............................................................................................................286
Figure 12-18: Basic PMT Design .....................................................................................................................287
Figure 12-19: PMT Cooling System .................................................................................................................288
Figure 12-20: PMT Preamp Block Diagram .....................................................................................................290
Figure 12-21: Relay Board Status LED Locations ...........................................................................................292
Figure 12-22: Power Distribution Block Diagram .............................................................................................296
Figure 12-23: Front Panel and Display Interface Block Diagram .....................................................................297
Figure 12-24: Basic Software Operation ..........................................................................................................298
Figure 12-25: Calibration Slope and Offset ......................................................................................................299
LIST OF TABLES
Table 1-1: Analyzer Options ..........................................................................................................................19
Table 2-1 T100 Basic Unit Specifications ....................................................................................................23
Table 2-2: O2 Sensor Option Specifications ..................................................................................................24
Table 2-3: CO2 Sensor Option Specifications ...............................................................................................24
Table 3-1: Ventilation Clearance ...................................................................................................................28
Table 3-2: Display Screen and Touch Control Description ...........................................................................31
Table 3-3: Rear Panel Description ................................................................................................................34
Table 3-4: Electrical Connections References ..............................................................................................36
Table 3-5: Analog Input Pin Assignments .....................................................................................................38
Table 3-6: Analog Output Pin Assignments ..................................................................................................39
Table 3-7: Status Output Signals ..................................................................................................................42
Table 3-8: Control Input Signals ....................................................................................................................43
Table 3-9: Pneumatic Layout Reference ......................................................................................................54
Table 3-10: Zero/Span and Sample/Cal Valve Operating States ...................................................................57
Table 3-11: Zero/Span and Sample/Cal Valve Operating States ...................................................................58
Table 3-12: IZS Valve Operating States .........................................................................................................60
Table 3-13: NIST-SRM's Available for Traceability of SO2 Calibration Gases ...............................................64
Table 3-14: P ossible Startup Warning Messages – T100 Analyzers w/o Options .........................................66
Table 3-15: P ossible Startup Warning Messages – T100 Analyzers with Options .........................................67
Table 4-1: Analyzer Operating Modes ..........................................................................................................76
Table 4-2: Test Functions Defined ................................................................................................................78
Table 4-3: List of Warning Messages............................................................................................................80
xiv
Teledyne API - T100 UV Fluorescence SO2 Analyzer
06807F DCN7335
Table 4-4: Primary Setup Mode Features and Functions .............................................................................82
Table 4-5: Secondary Setup Mode Features and Functions ........................................................................82
Table 5-1: Password Levels ..........................................................................................................................93
Table 5-2: Variable Names (VARS) Revision 1.0.3 ....................................................................................100
Table 5-3: T100 Diagnostic (DIAG) Functions ............................................................................................102
Table 5-4: DIAG - Analog I/O Functions .....................................................................................................106
Table 5-5: Analog Output Voltage Ranges .................................................................................................106
Table 5-6: Analog Output Current Loop Range ..........................................................................................107
Table 5-7: Voltage Tolerances for Analog Output Calibration ....................................................................112
Table 5-8: Current Loop Output Calibration with Resistor ..........................................................................116
Table 5-9: Test Parameters Available for Analog Output A3 (standard configuration) ...............................125
Table 6-1: COMM Port Communication Modes ..........................................................................................128
Table 6-2: Ethernet Status Indicators..........................................................................................................132
Table 6-3: LAN/Internet Default Configuration Properties ..........................................................................133
Table 6-4: Hostname Editing Button Functions ..........................................................................................136
Table 6-5: RS-232 Communication Parameters for Hessen Protocol ........................................................141
Table 6-6: T100 Hessen Protocol Response Modes ..................................................................................144
Table 6-7: Default Hessen Status Bit Assignments ....................................................................................145
Table 7-1: Front Panel LED Status Indicators for DAS ...............................................................................148
Table 7-2: DAS Data Channel Properties ...................................................................................................149
Table 7-3: DAS Data Parameter Functions ................................................................................................150
Table 8-1: Status Output Pin Assignments .................................................................................................170
Table 8-2: Control Input Pin Assignments ..................................................................................................171
Table 8-3: Terminal Mode Software Commands ........................................................................................172
Table 8-4: Command Types ........................................................................................................................173
Table 8-5: Serial Interface Documents........................................................................................................178
Table 9-1: NIST-SRM's Available for Traceability of SO2 Calibration Gases .............................................180
Table 9-2: AutoCal Modes ..........................................................................................................................193
Table 9-3: AutoCal Attribute Setup Parameters ..........................................................................................193
Table 9-4: Example Auto-Cal Sequence .....................................................................................................194
Table 9-5: Calibration Data Quality Evaluation ...........................................................................................196
Table 10-1: T100 Preventive Maintenance Schedule ...................................................................................209
Table 10-2: Predictive Uses for Test Functions ............................................................................................211
Table 11-1: Warning Messages - Indicated Failures ....................................................................................222
Table 11-2: Test Functions - Possible Causes for Out-Of-Range Values ....................................................224
Table 11-3: Relay Board Status LEDs ..........................................................................................................228
Table 11-4: DC Power Test Point and Wiring Color Code ............................................................................234
Table 11-5: DC Power Supply Acceptable Levels ........................................................................................235
Table 11-6: Relay Board Control Devices .....................................................................................................236
Table 11-7: Analog Output Test Function - Nominal Values ........................................................................237
Table 11-8: Status Outputs Check Pin Out ...................................................................................................237
Table 11-9: Example of HVPS Power Supply Outputs .................................................................................241
Table 11-10: UV Lamp Signal Troubleshooting ..............................................................................................253
Table 12-1: Relay Board Status LED’s .........................................................................................................292
xv
Teledyne API - T100 UV Fluorescence SO
06807F DCN7335
Analyzer
2
This page intentionally left blank.
xvi
06807F DCN7335

1. INTRODUCTION, FEATURES AND OPTIONS

This section provides an overview of the Teledyne API Model T100 Analyzer, its features and its options.

T100 OVERVIEW 1.1.

The Model T100 (also referred to as T100) UV Fluorescence SO2 Analyzer is a microprocessor control led analyzer that determines the concentration of sulfur dioxide
), in a sample gas drawn through the instrument’s sample chamber where it is
(SO
2
exposed to ultraviolet light, which causes any S O measures the amount of fluorescence to determine the amount of SO sample gas.
present t o fluoresce. The instrument
2
present in the
2
The T100’s exceptional stability is achieved with the use of an optical shutter to compensate for sensor drift and a reference detector to correct for changes in UV lamp intensity. Additionally, an advanced optical design combined with a special scrubber, called a "kicker" that removes hydrocarbons (which fluoresce similarly to SO
), to
2
prevent inaccuracies caused by interferents. Calibration is performed in software that stores SO
when specific, known concentrations of SO
concentration measurements made
2
are supplied to the analyzer. The
2
microprocessor uses these calibration val ues along with other performance parameters such as the sensor offset, UV lamp intensity, amount of stray light present, and sample gas temperature and pressure measurements to compute the final SO
concentration.
2
Built-in data acquisition capability, using the analyzer's internal memory, allows the logging of multiple parameters including averaged or instantaneous concentration values, calibration data, and operating parameters such as pressure and flow rate. Stored data are easily retrieved through the serial port or Ethernet port via our APICOM software or from the front panel, allowing operators to perform predictive diagnostics and enhanced data analysis by tracking parameter trends. Multiple averaging periods of one minute to 365 days are available for over a period of one year.
17
Introduction, Features and Options Teledyne API - T100 UV Fluorescence SO
06807F DCN7335

FEATURES 1.2.

The features of your T100 UV Fluorescence Sulfur Dioxide Analyzer include:
LCD Graphical User Interface with capacitive touch screen
Ranges, 0-50 ppb to 0-20,000 ppb, user selectable
Dual ranges and auto ranging
Microprocessor control for v ersatil ity
Multi-tasking software to allow viewing test variables wh ile operat ing
Continuous self checking with alarms
Bi-directional USB, RS-232, and 10/100Base-T Ethernet ports for remote operation
(optional RS-485)
Front panel USB ports for peripheral devices
Digital status outputs to indicate instrument operating conditions
Adaptive signal filtering to optimize response time
Temperature and Pressure compensation
Analyzer
2
Internal Zero and Span check (optional)
Internal data logging with 1 min to 365 day multiple averages
Critical flow orifi ce s to provide flow stability

T100 DOCUMENTATION 1.3.

In addition to this operation manual (part number 06807), the APICOM and DAS manual (PN 07463) is available for download from Teledyne API’s website at http://www.teledyne-api.com/manuals/, to support the operation of this instrument.

OPTIONS 1.4.

The options available for your analyzer are presented in Table 1-1. To order these options or to learn more about them, please contact the Sales department of Teledyne API at:
TOLL-FREE: 800-324-5190 TEL: +1 858-657-9800 FAX: +1 858-657-9816 E-MAIL: Api-sales@teledyne.com WEB SITE: http://www.teledyne-api.com/
18
Teledyne API - T100 UV Fluorescence SO2 Analyzer Introduction, Features and Options
performance.
Carrying Strap/Handle
Side-mounted strap for hand-carrying analyzer
Parts Kits
Spare parts and expendables
Expendables Kit with IZS
06807F DCN7335
Table 1-1: Analyzer Options
OPTION
Pumps
Rack Mount Kits
29
OPTION
NUMBER
10A External Pump 100V - 120V @ 60 Hz N/A 10B External Pump 220V - 240V @ 50 Hz N/A 10C External Pump 220V - 240V @ 60 Hz N/A 10D External Pump 100V - 120V @ 50 Hz N/A 10E External Pump 100V @ 60 Hz N/A 11B Pumpless, internal or external Pump Pack N/A 13 High Voltage Internal Pump 240V @ 50Hz N/A
Options for mounting the analyzer in standard 19” racks
20A Rack mount brackets with 26 in. chassis slides N/A 20B Rack mount brackets with 24 in. chassis slides N/A 21 Rack mount brackets only (compatible with carrying strap, Option 29) N/A 23 Rack mount for external pump pack (no slides) N/A
DESCRIPTION/NOTES REFERENCE
Pumps meet all typical AC power supply standards while exhibiting same pneumatic
Extends from “flat” position to accommodate hand for carrying. Recesses to 9mm (3/8”) dimension for storage. Can be used with rack mount brackets, Option 21. Cannot be used with rack mount slides.
N/A
CAUTION
General Safety Hazar d
Analog Inputs
64
Current Loop Analog Outputs
41
NO Optical Filter Recommended for high NOX backgrounds.
47 Required for EN Certification. N/A
A FULLY LOADED T100 WITH VALVE OPTIONS WEIGHS ABOUT 18 KG (40 POUNDS).
To avoid personal injury we recommend that two persons lift and carry the analyzer. Disc on nect all cables and tubing from the analyzer before moving it.
Used for connecting external voltage signals from other instrumentation (such as meteorological instrument s).
42A
43 45
Also can be used for logging these signals in the analyzer’s interna l DAS
Adds isolated, voltage-to-current conversion circuitry to the analyzer’s analog outputs.
Can be configured for any output range between 0 and 20 mA. May be ordered separately for any of the analog outputs. Can be installed at the factory or retrofitted in the field.
Expendables Kit includes a recommended set of expendables for one year of operation of this instrument including replacement sample particulate filters.
includes the items needed to refurbish the
internal zero air scrubber (IZS) that is includ ed. Spare Parts Kit includes spares parts for one unit.
Sections 3.3.1.2,
5.3, 5.9.3.7,and 7
Sections 3.3.1.4,
5.4.1, 5.9.3, and
5.9.3.5
Appendix B
Appendix B Appendix B
19
Introduction, Features and Options Teledyne API - T100 UV Fluorescence SO
when immersed in a gas stream moving at the spec ified flow rate. This
Communication Cables
For remote serial, network and Internet communication with the analyzer.
Type
Description
Shielded, straight-through DB-9F to DB-25M cable, about activated switches with DB-25 serial connectors.
RS-232 Multidrop
Enables communications between host computer and up to eight analyzers.
06807F DCN7335
Analyzer
2
OPTION
Calibration Valves
50A
Internal Zero/Span (IZS) Gas Generator
OPTION
NUMBER
DESCRIPTION/NOTES REFERENCE
Used to control the flow of calibration gases generated from external sources, rather than manually switching the rear panel pneumatic connections.
Two Teflon® solenoid valve sets located inside the analyzer:
Zero/Span valve switches between zero air and span gas; Sample/Cal valve switches between sample gas and calibration gas.
Sections 3.3.2.3,
3.3.2.4, 9.4, 9.5 and 9.6
Generates internal zero air and span gas.
Includes heated enclosure for a permeation tube (tube not included –
51A
see SO producing zero air and a set of valves for switching between the sample gas inlet and the output of the zero/span subsystem, functionally very similar to the valves included in the zero/span valve
IZS Permeation Tubes options), an external scrubber for
2
Sections 3.3.2.4,
9.5, 10.3.2 and
11.6.17
option.
SO2 IZS Permeation Tubes Replacement tubes for the IZS option; identical size/shape; different effusion rates.
Effusion Rate (@ 50°C)
Approximate Concentration
Specified Flow Rate (of
indicated perm tube rate) 52C 796 ng/min 0.3-0.5 ppm 0.76 ± 5% lpm N/A 52H 1592 ng/min 0.8 ppm 0.76 ± 50% lpm N/A 52M 220 ng/min 150 ppb 0.56 ± 25% lpm N/A
Each tube comes with a calibration certificate, traceable to a NIST standard, specifying its actual effusion rate of that tube to within ± 5%
Sections 3.3.2.4 and 9.1.1.3
calibration is performed at a tube temperature of 50°C.
60A RS-232
60B RS-232
60C Ethernet
60D USB
Concentration Alarm Relay
Issues warning when gas concentration exceeds limits set by user.
Four (4) “dry contact” relays on the rear panel of the instrument. This
61
relay option is different from and in addition to the “Contact Closures” that come standard on all TAPI instruments.
Multidrop card seated on the analyzer’s CPU card.
62
Each instrument in the multidrop network requires this card and a communications cable (Option 60B).
1.8 m long. Used to interface with older computers or code
Shielded, straight-through DB-9F to DB-9F cable of about
1.8 m length. Patch cable, 2 meters long, used for Internet and LAN
communications. Cable for direct connection between instrument (rear panel
USB port) and personal computer.
Section 3.3.1.8 and 6.3
Sections 3.3.1.8, and 6.3, and 7.2.7
Sections 3.3.1.8 and 6.5
Sections3.3.1.8 and 6.5.1
Sections 3.3.1.7 and 3.4.4
Section 3.3.1.8
20
Teledyne API - T100 UV Fluorescence SO2 Analyzer Introduction, Features and Options
(specs)
(specs)
Special Features
Built in features, software act iv at ed
Maintenance Mode Switch
06807F DCN7335
OPTION
OPTION
NUMBER
DESCRIPTION/NOTES REFERENCE
Second Gas Sensors Choice of one additional gas sensor.
65A Oxygen (O2) Sensor
67A Carbon Dioxide (CO2) Sensor
, located inside the ins trument, places the
analyzer in maintenance mode where it can continue sampling, yet
N/A
ignore calibration, diagnostic, and reset instrument commands. This feature is of particular use for instruments connected to Multidrop or Hessen protocol networks .
Call Technical Support for activation.
Second Language Switch activ ates an alternate set of display
N/A
messages in a language other than the instrument’s default language.
Call Technical Support for a specially programmed Disk on Module containing the second language.
Dilution Ratio Option allows the user to compensate for diluted sample gas, such as in continuous emission monit orin g (CEM) where
N/A
the quality of gas in a s moke stack is being tested and the sampling method used to remove the gas from the stack dilutes the gas.
Call Technical Support for activation.
• Section 2.1
• Section 3.3.2.9, (pneumatic layout)
• Section 9.10.1 (calibration)
• Section 12.2 for principles of operation
• Section 2.1
• Section 3.3.2.10 (pneumatic layout)
• Section 9.10.2 (calibration)
• Section 12.3 (principles of operation)
N/A
N/A
Section 3.4.4.1,
5.4.5 and 12.1.9.3
21
Introduction, Features and Options Teledyne API - T100 UV Fluorescence SO
06807F DCN7335
Analyzer
2
This page intentionally left blank.
22
(Physical Analog Output)
Max: 0-20,000 ppb Full Scale (selectable, dual ranges and auto ranging supported)
Measurement Units
ppb, ppm, µg/m3, mg/m3 (selectable)
Zero Noise1
0.2 ppb (RMS)
Span Noise1
< 0.5% of reading, above 50 ppb
Lower Detectable Limit2
0.4 ppb
Zero Drift
0.5 ppb/24 hours
Span Drift
< 0.5% of full scale/24 hours
Lag Time1
20 seconds
Rise/Fall Time1
<100 sec to 95%
Linearity
1% of full scale
Precision1
0.5% of reading above 50 ppb
Sample Flow Rate
650 cc/min. ±10%
220-240 V~, 50/60 Hz 3.0A
140W
Analog Output Ranges
10 V, 5 V, 1 V, 0.1 V (selectable)
Recorder Offset
± 10 %
1 Ethernet: 10/100Base-T
4 analog outputs
3 4-20mA current outputs
06807F DCN7335

2. SPECIFICATIONS, APPROVALS & COMPLIANCE

This section presents specifications for the T100 analyzer and the O2 and CO2 sensor options, Agency approvals, EPA equivalency designation, and CE mark compliance.

SPECIFICATIONS AN D AP PR O VAL S 2.1.

Table 2-1 T100 Basic Unit Specifications
Parameter Description
Ranges
Power Requirements Power Rating
Standard I/O
Min: 0-50 ppb Full Scale
<
<
110-120 V~, 60 Hz 3.0A
2 RS-232 (300 – 115,200 baud) 2 USB device ports 8 opto-isolated digital outputs 6 opto-isolated digital inputs
Typical Power Consumption
165W
Optional I/O 1 USB com port
1 RS485 8 analog inputs (0-10V, 12-bit) 4 digital alarm outputs Multidrop RS232
23
Specifications, Approvals & Compliance Teledyne API - T100 UV Fluorescence SO
Intended for Indoor use only at altitudes ≤ 2000m
Operating Temperature
5 - 40 oC (with EPA Equivalency)
Humidity Range
0 - 95% RH, non-condensing
Dimensions HxWxD
7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)
Weight
31 lbs (14 kg); 35.7 lbs (16.2 kg) with internal pump
1
As defined by the USEPA.
1
1
Note: zero drift is typically <± 0.1% O
2
during the first 24 hrs of operation
Parameter
Description
1
1
Accuracy
<± (0.02% CO2 + 2% of reading)
Rise and Fall Time
<60 seconds to 95%
1
As defined by the USEPA
06807F DCN7335
Parameter Description
Environmental Installation category (over-voltage category) II; Pollution degree 2
2
Defined as twice the zero noise level by the USEPA.
Table 2-2: O2 Sensor Option Specifications
Parameter Description
Ranges 0-1% to 0-100% user selectable. Dual rang es and auto-ranging supported. Zero Noise Lower Detectable Limit2 <0.04% O2 Zero Drift (24 hours) 3 <± 0.02% O2 Zero Drift (7 days) <±- 0.05% O2 Span Noise Span Drift (7 days) <± 0.1% O2 Accuracy (intrinsic error) <± 0.1% O2 Linearity <± 0.1 % O2 Temp Coefficient <± 0.05% O2 /°C, Rise and Fall Time <60 seconds to 95%
1
As defined by the USEPA
2
Defined as twice the zero noise level by the USEPA
3
<0.02% O2
<± 0.05% O2
Analyzer
2
Table 2-3: CO2 Sensor Option Specifications
Ranges 0-1% to 0-20% user selectable. Dual ranges and auto-ranging supported. Zero Noise Zero Drift (24 hours) <± 0.02% CO2 Zero Drift (7 days) <± 0.05% CO2 Span Noise Span Drift (7 days) <± 0.1% CO2 Lower Detectable Limit2 <0.04% CO2
Linearity <± 0.1% CO2 Temperature Coefficient <± 0.01% CO2 /°C
2
Defined as twice the zero noise level by the USEPA
24
<0.02% CO
<± 0.1% CO2
2
Teledyne API - T100 UV Fluorescence SO2 Analyzer Specifications, Approvals & Compliance
Phone:
06807F DCN7335

EPA EQUIVALENCY DESIGNATION 2.2.

The Model T100 Fluorescence SO2 Analyzer is officially designated as US EPA Federal Equivalent Method Number EQSA-0495-100 for sulfur dioxide measurement. The official List of Designated Reference and Equivalent Methods is published in the U.S. Federal Register: http://www3.epa.gov/ttn/amtic/criteria.html
.

APPROVALS AND CERTIFICATIONS 2.3.

The Teledyne API Model T100 analyzer was tested and certified for Safety and Electromagnetic Compatibility (EMC). This section presents the compliance statements for those requirements and directives.
EMC 2.3.1.
EN 61326-1 (IEC 61326-1), Class A Emissions/Industrial Immunity EN 55011 (CISPR 11), Group 1, Class A Emissions FCC 47 CFR Part 15B, Class A Emissions CE: 2004/108/EC, Electromagnetic Compatibility Directive

SAFETY 2.3.2.

IEC 61010-1:2010 (3rd Edition), Safety requirements for electrical equipment for measurement, control, and laboratory use.
CE: 2006/95/EC, Low-Voltage Directive

OTHER TYPE CERTIFICATIONS 2.3.3.

MCERTS: Sira MC 050067/07 EN 15267 – Air Quality – Ambient Air Automated Measuring Systems EN 14212 – Ambient Air Measurement for SO
For additional certifications, please contact Technical Support:
Toll-free Phone:
Email:
800-324-5190 +1 858-657-9800
Fax:
+1 858-657-9816 sda_techsupport@teledyne.com
2
25
Specifications, Approvals & Compliance Teledyne API - T100 UV Fluorescence SO
06807F DCN7335
Analyzer
2
This page intentionally left blank.
26
06807F DCN7335

3. GETTING STARTED

This section addresses the procedures for unpac king the instrument and i nspecting for damage, presents clearance specifications for proper ventilation, introduces the instrument layout, then presents the procedures for getting started: making electrical and pneumatic connections, and conducting an initial calibration check.

UNPACKING THE T100 ANALYZER 3.1.

ATTENTION
CAUTION
GENER AL S AFETY HAZARD
Avoid personal injury: always use two persons to lift and carry the T100.
COULD DAMAGE INSTRUMENT AND VOID WARRANTY
Printed Circuit Assemblies (PCAs) are sensitive to electro-static discharges (ESD) too small to be felt by the human nervous system. Failure to use ESD protection when working with electronic assemblies will void the instrument warranty. Refer to the manual on Fundamentals of ESD, PN 04786, which can be downloaded from our website at
http://www.teledyne-api.com
the Special Manuals section.
Do not operate this instrument until you’ve removed dust plugs from SAMPLE and EXHAUST ports on the rear panel!
under Help Center > Product Manuals in
CAUTION
Note Teledyne API recommends that you store shipping containers/materials
for future use if/when the instrument should be returned to the factory for repair and/or calibration service. See Warranty section in this manual and shipping procedures on our Website at under Customer Support > Return Authorization.
27
http://www.teledyne-api.com
Getting Started Teledyne API - T100 UV Fluorescence SO
4 in.
1 in.
06807F DCN7335
Verify that there is no apparent external shipping damage. If damage has occurred, please advise the shipper first, then Teledyne API.
Included with your analyzer is a printed record of the final performance characterization performed on your instrument at the factory. It is titled Final Test and Vali dation Data Sheet (P/N 04551). This record is an important quality assurance and calibration record for this instrument. It should be placed in the quality records file for this instrument.
With no power to the unit, carefully remove the top cover of the analyzer and check for internal shipping damage by carrying out the following steps:
1. Remove the two flat head, Phillips screws on the sides of the instrument;
2. Slide the cover backwards until it clears the analyzer’s front bezel, and;
3. Lift the cover straight up.
4. Ins pect the interior of the inst rument to ensure that all circ uit boards and other components are in good shape and properly seated.
5. Chec k the connectors of the various internal wirin g harnesses and pneumatic hoses to ensure that they are firmly and properly seated.
Analyzer
2
6. Verif y that all of the option al hardware or dered with t he unit has been install ed. These are listed on the paperwork accompanying the analyzer.
WARNING
ELECTRICAL SHOCK HAZARD
Never disconnect PCAs, wiring harnesses or electronic subassemblies while under power.

VENTILATION CLEARANCE 3.1.1.

Whether the analyzer is set up on a bench or install ed into an instrument rack, be sure to leave sufficient ventilation clearance.
Table 3-1: Ventilation Clearance
AREA
Back of the instrument
Sides of the instrument
Above and below the
instrument
Various rack mount kits are available for this analyzer. Refer to Section 1.4 of this manual for more information.
MINIMUM REQUIRED
CLEARANCE
1 in.
28
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
06807F DCN7335

INSTRUMENT LAYOUT 3.2.

Instrument layout includes illustrations and descriptions of front panel and display, rear panel connectors, and internal chassis layout.
FRONT PANEL 3.2.1.
Figure 3-1 shows the analyzer’s front panel layout, followed by a close-up of the display screen in Figure 3-2, which is described in Table 3-2. The two USB ports on the front panel are provided for the connection of peripheral devices:
plug-in mouse (not included) to be used as an alternative to the touchscreen
interface
thumb drive (not included) to download updates to instruction software (contact
TAPI Technical Support for information).
Figure 3-1: Front Panel Layout
29
Getting Started Teledyne API - T100 UV Fluorescence SO
06807F DCN7335
Analyzer
2
ATTENTION
Figure 3-2: Display Screen and Touch Control
The front panel liquid crystal display screen includes touch control. Upon analyzer start­up, the interface shows a splash screen and other initialization indicators before the main display appears, similar to Figure 3-2 above (may or may not display a Fault alarm). The LEDs on the display screen indicate the Sample, Calibration and Fault states; also on the screen is the gas concentration field (Conc), which displays real-time readouts for the primary gas and for the secondary gas if installed. The display screen also shows what mode the analyzer is currently in, as well as messages and data (Param). Along the bottom of the screen is a row of touch control buttons; only those that are cu rrently appli cable will have a label. Table 3-2 provides detailed information for each component of the screen.
COULD DAMAGE INSTRUMENT AND VOID WARRANTY
Do not use hard-surfaced instruments such as pens to touch the control buttons.
30
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
Name
Color
State
Definition
06807F DCN7335
Table 3-2: Display Screen and Touch Control Description
Field Description/Function
Status LEDs indicating the states of Sample, Calibration and Fault, as follows:
Off On
SAMPLE Green
CAL Yellow
FAULT Red
Conc Mode Displays the name of the analyzer’s current operating mode Param Control Buttons Displays dynamic, context sensitive labels on each button, which is blank when inactive until applicable.
Displays the actual concentration of the sample gas currently being measured by the analyzer in the currently selected units of measure
Displays a variety of informational messages such as warning mes sages, operational data, test function values and response messages during interactive tasks.
Blinking
Off On Blinking
Off Blinking
Unit is not operating in sample mode, DAS is disabled. Sample Mode active; Front Panel Display being updated; DAS data
being stored. Unit is operating in sample mode, front panel display being updated,
DAS hold-off mode is ON, DAS disabled Auto Cal disabled
Auto Cal enabled Unit is in calibration mode
No warnings exist Warnings exist
Figure 3-3 shows how the front panel display is mapped to the menu charts illustrated in this manual. The Mode, Param (parameters), and Conc (gas concentration) fields in the display screen are represented across the top row of each menu chart. The eight touch control buttons along the bottom of the display screen are represented in the bottom row of each menu chart.
31
Getting Started Teledyne API - T100 UV Fluorescence SO
06807F DCN7335
Analyzer
2
Figure 3-3: Display/Touch Control Screen Mapped to Menu Charts
Note The menu charts in this manual cont ain condensed representations of the
analyzer’s display during the various operations being described. These menu charts are not intended to be exact visual representations of the actual display.
32
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
06807F DCN7335

REAR PANEL 3.2.2.

Figure 3-4: Rear Panel Layout
Table 3-3 provides a description of each component on the rear panel.
33
Getting Started Teledyne API - T100 UV Fluorescence SO
installed, connect a gas line to the source
RX TX
COM 2
RS-232
STATUS
CONTROL IN
ALARM
ETHERNET
06807F DCN7335
Table 3-3: Rear Panel Description
Component Function
cooling fan
AC power
connector
Pulls ambient air into chassis through side vents and exhausts through rear. Connector for three-prong cord to apply AC power to the analyzer.
CAUTION! The cord’s power specifications (specs) MUST comply with the power specs on the analyzer’s rear panel Model number label
Analyzer
2
Model/specs label
TO CONV
FROM CONV
SAMPLE
EXHAUST
SPAN 1
SPAN2/VENT
ZERO AIR
DCE DTE
ANALOG OUT
ANALOG IN
USB
Information Label
Identifies the analyzer model number and provides power specs
(not used in this model) (not used in this model)
Connect a gas line from the source of sample gas here. Calibration gases are also inlet here on units without zero/span/shutoff valve options
installed. Connect an exhaust gas l ine of not more than 10 meters long here that leads outside
the shelter or immediate area surrounding the instrument. On units with zero/span/shutoff valve options
of calibrated span gas here. Used as a second cal gas input line when instrument is configured with zero/span
valves and a dual gas option, or as a cal gas vent line when instrument is configured with a pressurized span option
Internal Zero Air: On units with zero/span/shutoff valve options installed but no internal zero air scrubber attach a gas line to the source of zero air here.
LEDs indicate receive (RX) and transmit (TX) activity on the when blinking. Serial communications port for RS-232 or RS-485. Serial communications port for RS-232 only. Switch to select either data terminal equipment or data communication equ ipm ent
during RS-232 communication. (Section 6.1). For outputs to devices such as Programmable Logic Controllers (PLCs).
For voltage or current loop outputs to a strip chart recorder and/or a data logger. For remotely activating the zero and span calibration modes. Option for concentration alarms and system warnings. Connector for network or Internet remote communication, using Ethernet cable Option for external voltage signals from other instrumentation and for logging these
signals Connector for direct connection to personal computer, using USB cable.
Includes voltage and frequency specifications
(Call factory for details).
34
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
06807F DCN7335
INTERNAL CHASSIS LAYOUT 3.2.3.
Figure 3-5 illustrates the internal layout of the chassis without options. Section 3.3.2 shows pneumatic diagrams for the basic configuration and for options.
Figure 3-5: Internal Layout, Basic (no Valve or Second Gas Options)
35
Getting Started Teledyne API - T100 UV Fluorescence SO
06807F DCN7335
Analyzer
2

CONNECTIONS AND SETUP 3.3.

This section presents the electrical (Section 3.3.1) and pneumatic (Section 3.3.2) connections for setup and preparing for instrument operation.

ELECTRICAL CONNECTIONS 3.3.1.

Note To maintain compliance with EMC standards, it is required that the cable
length be no greater than 3 meters for all I/O connections, which include Analog In, Analog Out, Status Out, Control In, Ethernet/LAN, USB, RS-232, and RS-485.
This section provides instructions for basic connections and for options. Table 3-4 provides a direct link to the instructions for the subsections that apply t o your analyzer’s configuration.
Table 3-4: Electrical Connections References
Connection Section
Power 3.3.1.1 Analog Inputs (Option) 3.3.1.2 Analog Outputs 3.3.1.3 Current Loop Analog Outputs (Option),
and converting current to voltage output Status Outputs 3.3.1.5 Control Inputs 3.3.1.6 Concentration Alarm Relay (Option) 3.3.1.7 Communications (Ethernet, USB,
RS-232, Multidrop, RS-485)*
* USB is an option with exceptions. * RS-485 is an option and requires special setup (contact the Factory). Either USB or RS-485 can be used; not both.
3.3.1.4
3.3.1.8
36
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
06807F DCN7335
3.3.1.1. CONNECTING POWER
Attach the power cord to the analyzer and plug it into a power outlet capable of carrying at least 10 Amps of current at your AC voltage and that it is equipped with a functioning earth ground.
WARNING
ELECTRICAL SHOCK HAZARD
High Voltages are present inside the analyzer’s case. Power connection must have functioning ground connection. Do not defeat the ground wire on power plug. Power off analyzer before disconnecting or connecting electrical
subassemblies. Do not operate analyzer with the cover off.
CAUTION
GENER AL S AFETY HAZARD
To avoid damage to your analyzer, ensure that the AC power voltage matches the voltage indicated on the Analyzer’s model identification label on the rear panel before plugging the T100 into line power.
3.3.1.2. CONNECT IN G ANALOG INPUTS (OPTION)
The Analog In connector is used for connecting external voltage signals from other instrumentation (such as meteorological instruments) and for logging these signals in the analyzer’s internal Data Acquisition System (DAS). The input voltage range for each analog input is 1-10 VDC, and input impedance is nominally 20kΩ in parallel with
0.1µF.
Figure 3-6: Analog In Connector
Pin assignments for the Analog In connector are presented in Table 3-5 .
37
Getting Started Teledyne API - T100 UV Fluorescence SO
DAS
1
ANALOG OUT
A1 A2 A3 A4 + - + - + - + -
06807F DCN7335
Table 3-5: Analog Input Pin Assignments
Analyzer
2
PIN DESCRIPTION
1 Analog input # 1 AIN 1 2 Analog input # 2 AIN 2 3 Analog input # 3 AIN 3 4 Analog input # 4 AIN 4 5 Analog input # 5 AIN 5 6 Analog input # 6 AIN 6 7 Analog input # 7 AIN 7 8 Analog input # 8 AIN 8
GND Analog input Ground N/A
See Section 7 for details on setting up the DAS.
3.3.1.3. CONNECTING ANALOG OUTPUTS
The T100 is equipped with several analog output channels accessible through a connector on the rear panel of the instrument. The standard configuration for these outputs is mVDC. An optional current loop output is available for each (Section
3.3.1.4). In default configuration, Channels A1 and A2 out put a signal proportional to the SO
concentration of the sample gas. Either can be used for connecting the analog out put signal to a chart recorder or for interfacing with a datalogger.
PARAMETER1
2
Channel A3 is only used on the T100 if the optional O
or CO2 sensor is install e d.
2
Channel A4 is special. It can be set by the user (refer to Section 5.9.9) to output any one of the parameters accessible through the <TST TST> buttons of the uni t’s sample display.
To access these signals attach a strip chart recorder and/or data-logger to the appropriate analog output connections on the rear panel of the analyzer.
Figure 3-7: Analog Output Connector
38
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
PIN
ANALOG OUTPUT
VOLTAGE SIGNAL
CURRENT SIGNAL
06807F DCN7335
Table 3-6: Analog Output Pin Assignments
1 2 Ground I Out ­3 4 Ground I Out ­5 A3
6 Ground I Out -
7 8 Ground I Out -
optional O sensor is installed)
A1
A2
(Only used if an
or CO2
2
A4
V Out I Out +
V Out I Out +
V Out I Out +
V Out I Out +
3.3.1.4. CURRENT LOOP ANALOG OUTPUTS (OPTION 41) SETUP
If your analyzer had this option installed at the factory, there are no further connections to be made. The current loop option can be configured for any output range between 0 and 20 mA. Section 5.9.3.5 provides information on calibrating or adjusting these outputs.
This section provides instructions for setting up the analog outputs for voltage and/or current output. Figure 3-8 provides installation instructions and illustrates a sample combination of one current output and two voltage outputs configuration.
For current output install the Current Loop option PCA on J19, on J 21 or on J23 of
the motherboard.
For voltage output, install jumpers on J19, J21 and/or J23. Following Figure 3-8 are instructions for converting current loop analog outputs to
standard 0-to-5 VDC outputs.
CAUTION – AVOID INVALIDATING WARRANTY
Servicing or handling of circuit components requires electrostatic discharge protection (ESD), i.e. ESD grounding straps, mats and containers. Failure to use ESD protection when working with electronic assemblies will void the instrument warranty. Refer to the manual on Fundamentals of ESD, PN 04786, which can be downloaded from our website at http://www.teledyne-api.com under Help Center > Product Manuals in the Special Manuals section.
39
Getting Started Teledyne API - T100 UV Fluorescence SO
06807F DCN7335
Analyzer
2
Figure 3-8: Current Loop Option Installed on the Motherboard
CONVERTING CURRENT LOOP ANALOG OUTPUTS TO STANDARD VOLTAGE OUTPUTS
To convert an output configured for cur rent loop operation to the standard 0 to 5 VDC output operation:
1. Turn off power to the analyzer.
2. If a recording device was connected to the output being modified, disconnect it.
3. Remove the top cover
Remove the side-panel screw fastening the top cover to the unit.
Lift the cover straight up.
4. Disc onnect the current loop option PCA from the appropriat e connector on the motherboard (refer to Figure 3-8).
5. Each connector, J19 and J23, requires two shunts. Place one shunt on the two left most pins and the second shunt on the two pins next to it (refer to Figure 3-8).
6. Reattach the top case to the analyzer.
The analyzer is now ready to have a voltage-sensing, recording device attached to that output.
Calibrate the analog output as described in Section 5.9.3
40
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
STATUS
1 2 3 4 5 6 7
8 D
+
SYSTEM OK
HIGH RANGE
CO NC VALI D
ZE RO CAL SP AN CAL DIAG MODE
06807F DCN7335
3.3.1.5. CONNECTING THE STATUS O U TPUTS
The status outputs report analyzer conditions via optically isolated NPN transistors, which sink up to 50 mA of DC current. These outputs can be used interface with devices that accept logic-level digital inputs, such as Programmable Logic Controllers (PLCs). Each status bit is an open collector output that can withstand up to 40 VDC. All of the emitters of these transistors are tied together and available at the “D” connector pin.
ATTENTION
COULD DAMAGE INSTRUMENT AND VOID WARRANTY
Most PLCs have internal provisions for limiting the current that the input will draw from an external device. When connecting to a unit that does not have this feature, an external dropping resistor must be used to limit the current through the transistor output to less than 50 mA . At 50 mA, the transistor will drop approximately 1.2V from its collector to emitter.
The status outputs are accessed via a 12-pin connector on the analyzer’s rear panel labeled STATUS (Figure 3-9). Pin-outs for this connector are presented in Table 3-7.
Figure 3-9: Status Output Connector
41
Getting Started Teledyne API - T100 UV Fluorescence SO
4
D
06807F DCN7335
Table 3-7: Status Output Signals
Analyzer
2
REAR PANEL
LABEL
1
2
3
5 6
7&8
+
STATUS
DEFINITION
SYSTEM OK
CONC VALID
HIGH RANGE
ZERO CAL ON whenever the instr u ment ’s ZERO point is being calibrated. SPAN CAL ON whenever the instr u ment ’s SPAN point is being cal ibrated.
DIAG MODE ON whenever the instrument is in DIAGNOSTIC mode
EMITTER BUS
DC POWER
Digital Ground The ground level from the analyzer’s internal DC power supplies
ON if no faults are pres ent.
OFF any time the HOLD OFF feature is active, such as during calibration or when other faults exist possi bly invalidating the current concentration measurement (example: sample flow rate is outside of acceptable limits).
ON if concentration measurement is valid. ON if unit is in high range of either the DUAL or Auto range modes.
SPARE
The emitters of the transistors on pins 1-8 are bussed together.
SPARE
+ 5 VDC, 300 mA source (combined rating with Control Output, if used).
CONDITION
3.3.1.6. CONNECTING THE CONTROL INPUTS
If you wish to use the analyze r to remotely activate the zero and span calibration modes, several digital control inputs are provided through a 10-pin connector labeled CONTROL IN on the analyzer’s rear panel.
There are two methods for energizing the control inputs. The internal +5V available from the pin labeled “+” is the most convenient method (Figure 3-10, lef t) . However, if full isolation is required, an external 5 VDC power supply should be used (Figure 3-10, right).
42
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
CONTROL IN
A B C D E F U +
SPAN CAL ZERO CAL
CONTROL IN
A B C D E F U
+
SPAN CAL ZERO CAL
+
5 VDC Power
Supply
Local Power Connections
External Power Connections
Input #
Status Definition
ON Condition
06807F DCN7335
Figure 3-10: Control Input Connector
Table 3-8: Control Input Signals
A
B
C, D, E & F SPARE
U External Power input Input pin for +5 VDC is required to activate pins A – F.
+
REMOTE ZERO CAL
REMOTE SPAN CAL
Digital Ground
5 VDC output
The analyzer is placed in Zero Calibration mode. The mode field of the display will read ZERO CAL R.
The analyzer is placed in span calibration mode as part of perform ing a low span (midpoint) calibration. The mode field of the display will read LO CAL R.
The ground level from the analyzer’s internal DC power supplies (same as chassis ground)
Internally generated 5V DC power. To activate inputs A – F, place a jumper between this pin and the “U” pin. The maximum amperage through this port is 300 mA (combined with the analog output supply, if used).
43
Getting Started Teledyne API - T100 UV Fluorescence SO
06807F DCN7335
3.3.1.7. CONNECTING THE CONCENTRATION ALARM RELAY (OPTION 61)
The concentration alarm option is comprised of four (4) “dry contact” relays on the rear panel of the instrument. This relay option is different from and in addition to the “Contact Closures” that come standard on all Teledyne API instruments. Each relay has 3 pins: Normally Open (NO), Common (C) and Normally Closed (NC).
Figure 3-11: Concentration Alarm Relay
Alarm 1 “System OK 2” Alarm 2 “Conc 1” Alarm 3 “Conc 2”
Analyzer
2
Alarm 4 “Range Bit”
“ALARM 1” RELAY
Alar m 1, which is “System OK 2” (system OK 1 is the status bit), is in the energized state when the instrument is “OK” and there are no warnings. If there is a warning active or if the instrument is put into the “DIAG” mode, Alarm 1 will change states. This alarm has “reverse logic” meaning that if you put a meter across the Common and Normally Closed pins on the connector, you will find that it is OPEN when the instrument is OK. This is so that if the instrument should turn off or lose power, it will change states and you can record this with a data logger or other recording device.
“ALARM 2” RELAY & “ALARM 3” RELAY
Alarm 2 relay is associated with the “Concentration Alarm 1” set point in the software; Alarm 3 relay is associated with the “Concentration Alarm 2” set point in the software.
Alarm 2 Relay SO Alarm 3 Relay SO Alarm 2 Relay SO Alarm 3 Relay SO
Alar m 2 relay will be turned on any time the concentration value exceeds the set-point, and will return to its normal state when the concentration value returns below the concentration set-point.
Alarm 1 = xxx PPM
2
Alarm 2 = xxx PPM
2
Alarm 1 = xxx PPM
2
Alarm 2 = xxx PPM
2
Even though the relay on the rear panel is a non-latching alarm and resets when the concentration goes back below the alarm set point, the warning on the front panel of the instrument will remain latched until it is cleared. You can clear the warning on the front panel either manually by pressing the CLR button on the front panel touch-screen or remotely through the serial port.
44
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
06807F DCN7335
It is possible to configure the alarms to have two alarm levels for each concentration.
SO
Alarm 1 = 20 PPM
2
SO
Alarm 2 = 100 PPM
2
SO
Alarm 1 = 20 PPM
2
SO
Alarm 2 = 100 PPM
2
In this example, SO
Alarm 1 and SO2 Alarm 1 will both be associated with the “Alarm
2
2” relay on the rear panel. This allows you to have multiple alarm levels for individual concentrations.
A more likely configuration for t his would be to put one concentration on t he “Alarm 1” relay and the other concentration on the “Alarm 2” relay.
SO
Alarm 1 = 20 PPM
2
SO
Alarm 2 = Disabled
2
SO
Alarm 1 = Disabled
2
SO
Alarm 2 = 100 PPM
2
“ALARM 4” RELAY
This relay is connected to the “range bit”. If the instrument is configured for “Auto Range” and the reading goes up into the high range, it will turn this relay on.
3.3.1.8. CONNECTING THE COMMUNIC ATIONS INTERFACES
Available remote communications interfaces are: Ethernet, USB, RS-232, RS-232 Multidrop and RS-485. Use appropriate cables, and configure each type of
communication method through the SETUP>COMM menu, Section 6. Although Ethernet is DHCP-enabled by default, it can also be configured manually (Section 6.5.1) to set up a static IP address, which is the recommended setting when operating the instrument via Ethernet.
ETHERNET CONNECTION
For network or Internet communication with the analyzer, connect an Ethernet cable from the analyzer’s rear panel Ethernet interface connector to an Ethernet port. Please refer to Section 6.5 for a description of the default configuration and setup instructions.
Configuration: Section 6.5
manual configuration: Section 6.5.1
automatic configuration (default): Section 6.5.2
45
Getting Started Teledyne API - T100 UV Fluorescence SO
Cables that appear to be compatible because of matching connectors cables acquired from sources other than Teledyne API for pin
06807F DCN7335
USB CONNECTION
For direct communication between the analyzer and a PC, connect a US B c ab l e be t w een the analyzer rear panel and desktop or laptop USB ports, and ensure that their baud rates match (Section 6.2.2).
Analyzer
2
Note
If this optio n is installed, the COM2 port cannot be used for anything other than Multidrop communication.
Configuration: Section 6.5.3
RS-232 CONNECTION
For RS-232 communications with data terminal equipment (DTE) or with data communication equipment (DCE) connect either a DB9-female-to-DB9-female cable (Teledyne API part number WR000077) or a DB9-female-to-DB25-male cable (Option 60A, Section 1.4), as applicable, from the analyzer’s rear panel RS-232 port to the device. Adjust the DCE-DTE switch (Figure 3-4) to select DTE or DCE as appropriate.
Configuration: Sections 5.7 and 6.3
IMPORTANT IMPACT ON READINGS OR DATA
may incorporate internal wiring that makes the link inoperable. Check assignments (Figure 3-12) before using.
46
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
06807F DCN7335
RS-232 COM PORT CONNECTOR PIN-OUTS
Figure 3-12: Rear Panel Connector Pin-Outs for RS-232 Mode
The signals from these two connectors are routed from the motherboard via a wiring harness to two 10-pin connectors on the CPU card, J11 and J12 (Figure 3-13).
47
Getting Started Teledyne API - T100 UV Fluorescence SO
06807F DCN7335
Analyzer
2
Figure 3-13: Default Pin Assignments for CPU Com Port Connector (RS-232)
RS-232 COM PORT DEFAULT SETTINGS
As received from the factory, the analyzer is set up to emulate a DCE (Section 6.1) or modem, with Pin 3 of the DB-9 connector designated for receiving data and Pin 2 designated for sending data.
RS-232: RS-232 (fixed) DB-9 male connector
Baud rate: 115200 bits per second (baud)
Data Bits: 8 data bits with 1 stop bit
Parity: None
COM2: RS-232 (configurable to RS 485), DB-9 female connector.
Baud rate: 19200 bits per second (baud).
Data Bits: 8 data bits with 1 stop bit.
Parity: None.
Configuration: Section 6.3
48
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
06807F DCN7335
RS-232 MULTIDROP (OPTION 62) CONNECTION
When the RS-232 Multidrop option is installed, connection adjustments and configuration through the menu system are required. This section provides instructions for the internal connection adjustments, then for external connections, and ends with instructions for menu-driven configuration.
Note
ATTENTION
Because the RS-232 Multidrop option uses both the RS232 and COM2 DB9 connectors on the analyzer’s rear panel to connect the chain of instruments, COM2 port is no longer available for separate RS-232 or RS-485 operation.
COULD DAMAGE INSTRUMENT AND VOID WARRANTY
Printed Circuit Assemblies (PCAs) are sensitive to electro-static discharges (ESD) too small to be felt by the human nervous system. Failure to use ESD protection when working with electronic assemblies will void the instrument warranty. Refer to the manual on Fundamentals of ESD, PN 04786, which can be downloaded from our website at
http://www.teledyne-api.com
the Special Manuals section.
In each instrument with the Multidrop option there i s a shunt that jumpers two pins on the serial Multidrop and LVDS printed circuit assembly (PCA), as shown in Figure 3-14. This shunt must be removed from all instruments except that designated as last in the multidrop chain, which must remain terminated. This requires powering off and opening each instrument and making the following adjustments:
under Help Center > Product Manuals in
1. With NO pow er to the instrument, rem ove its top cover and lay the rear panel open for access to the multidrop/LVDS PCA, which is seated on the CPU.
2. On the multidrop PCA’s JP2 connector, remove the shunt that jumpers Pins 21 22 as indicated in Figure 3-14. (Do this for all but the last instrument in the chain where the shunt should remain at Pins 21 22).
3. Check that the following cable connections are made in all instruments (again refer to Figure 3-14):
J3 on the Multidrop/LVDS PCA to the CPU’s COM1 connector
(Note that the CPU’s COM2 connector is not used in Multidrop)
J4 on the Multidrop/LVDS PCA to J12 on the motherboard
J1 on the Multidrop/LVDS PCS to the front panel LCD
49
Getting Started Teledyne API - T100 UV Fluorescence SO
06807F DCN7335
Analyzer
2
Figure 3-14: Jumper and Cables for Multidrop Mode
Note: If you are adding an instrum ent to the end of a previously configured chain, remove the shunt b etween Pins 21 22 of J P2 on t he Multidrop/LVDS PCA in the instrument that was previously the last instrument in the chain.
4. Close the instrument.
5. Referring to Figure 3-15 use straight-through DB9 male  DB9 female cables to interconnect the host RS232 port to the first analyzer’s RS232 port; then from the first analyzer’s COM2 port to the second analyzer’s RS232 port; from the second analyzer’s COM2 port to the third analyzer’s RS232 port, etc., connecting in this fashion up to eight analyzers, subject to the distance limitations of the RS-232 standard.
6. On the r ear panel of each analyzer, adj ust the DCE DTE switch so that t he green and the red LEDs ( RX and T X ) of the COM1 connect o r (lab ele d R S2 32) are both lit. (Ensure you are using the c orr ec t RS-232 cables internall y wired s pecif ic al ly for RS­232 communication; see Section 3.3.1.8 under “RS-232 Connection”.
7. BEFOR E communicating from the host, power on t he instruments and check that the Machine ID code is unique for each (Section 5.7.1).
a. In the S ETUP Mode menu go to SETUP >MORE>COMM>ID. T he default ID is
typically the model number or “0”.
50
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
Analyzer
Analyzer
Analyzer
Last Analyzer
Female DB9
Male DB9
RS-232
COM2
RS-232
COM2
RS-232
COM2
RS-232
COM2
Host
RS-232 port
Ensure jumper is installed between
JP2 pins 21 ↔ 22 in last instrument
of multidrop chain.
06807F DCN7335
b. To change the identification number, press the button below the digit to be
changed.
c. Press/select ENTER to accept the new ID for that instrument.
8. Next, in the SETUP>MORE>COMM>COM1 menu (do not use the COM2 menu for multidrop), edit the COM1 MODE parameter as follows: press/select EDIT and set only QUIET MODE, COMPUTER MODE, and MULTIDROP MODE to ON. Do not change any other settings.
9. Press/select ENTER to accept the changed settings, and ensure that COM1 MODE now shows 35.
10. Press/select SET> to go to the COM1 BAUD RATE menu and ensure it reads the same for all instruments (edit as needed so that all instruments are set at the same baud rate).
Note The (communication) Host instrument can address only one instrument
at a time, each by its unique ID (see step 7 above).
Note Teledyne API recommends setting up the first link, between the Host and
the first analyzer, and testing it before setting up the rest of the chain.
Figure 3-15: RS-232-Multidrop PCA Host
/Analyzer Interconnect D iagram
51
Getting Started Teledyne API - T100 UV Fluorescence SO
06807F DCN7335
RS-485 CONNECTION (OPTION)
As delivered from the factor y, COM2 is configured for RS-232 communications. This port can be reconfigured for oper ation as a non-isolated, half-duplex RS-485 port. Usin g COM2 for RS -485 communication disables the USB port. To reconfigure this port for RS-485 communication, please contact the factory.
PNEUMATIC CONNECTIONS 3.3.2.
This section provides not only pneumatic connection information, but also important information about the gases required for accurate calibration (Section 3.3.2.11); it also illustrates the pneumatic layouts for the analyzer in its basic configuration and with options.
Before making the pneumatic connections, carefully note the following cautionary and additional messages:
CAUTION
GENER AL S AFETY HAZARD
SULFUR DIOXIDE (SO2) IS A TOXIC GAS.
Analyzer
2
DO NOT vent calibration gas and sample gas into enclosed areas. Obtain a Material Safety Data Sheet (MSDS) for this material. Read and rigorously follow the safety guidelines described there.
CAUTION
GENER AL S AFETY HAZARD
Sample and calibration gases should only come into contact with PTFE (Teflon) or glass tubes and fixtures.
They SHOULD NOT come in contact with brass or stainless steel fittings prior to the reaction cell.
The exhaust from the analyzer’s internal pump MUST be vented outside the immediate area or shelter surrounding the instrument.
It is important to conform to all safety requirements regarding exposure to SO2.
52
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
06807F DCN7335
ATTENTION
COULD DAMAGE INSTRUMENT AND VOID WARRANTY
Maximum Pressure:
Ideally the maximum pressure of any gas at the sample inlet should equal ambient atmospheric pressure and should NEVER exceed 1.5 in­hg above ambient pressure.
Venting Pressurized Gas:
In applications where any gas (span gas, zero air supply, sample gas is) received from a pressurized manifold, a vent must be provided to equalize the gas with ambient atmospheric pressure before it enters the analyzer to ensure that the gases input do not exceed t he maximum inlet pressure of the analyzer, as well as to prevent back diffusion and pressure effects. These vents should be:
• at least 0.2m long
• no more than 2m long
• vented outside the shelter or immediate area surrounding the
instrument.
Dust Plugs: Remove dust plugs from rear panel exhaust and supply line fittings
before powering on/operating instrument. These plugs should be kept for reuse in the event of future storage or shipping to prevent debris from entering the pneumatics.
IMPORTANT
IMPORTANT
EPA Requirements: US EPA requirements state that zero air and span gases must be
supplied at twice the instrument’s specified gas flow rate. Therefore, the T100 zero and span gases should be supplied to their respective inlets in excess of 1300 cc
3
/min (650 cc3/min. x 2).
Leak Check: Run a leak check once the appropriate pneumatic connections have
been made; check all pneumatic fittings for leaks using the procedures defined in Section 10.3.6
.
CAUTION – GENERAL SAFETY HAZARD
Gas flow though the analyzer must be maintained at all time for units with a permeation tube installed. Insufficient gas flow allows gas to build up to levels that will contaminate the instrument or present a safety hazard to personnel.
53
Getting Started Teledyne API - T100 UV Fluorescence SO
06807F DCN7335
Analyzer
2
Section 3.3.2.1 provides external pneumatic connection instructions, and Table 3-9 provides links to the location of various internal pneumatic layout illustrations.
Table 3-9: Pneumatic Layout Reference
Pneumatic Layout Section
Basic 3.3.2.2 Zero/Span Valves 3.3.2.3 Internal Zero/Span (IZS) 3.3.2.4 Basic with O2 Sensor 3.3.2.9 Basic with CO2 Sensor 3.3.2.10
3.3.2.1. BASIC CONNECTIONS INCLUDING W/SPAN GAS AND W/GAS DILUTIO N CALIBRATOR
Refer to Figure 3-4 and Table 3-3 while making the pneumatic connections as follows: SAMPLE inlet Connect ¼” gas line not more than 2 m long, from sample gas
source to this inlet.
When no zero/span/shutoff valve options, also c onnect line from
calibration gas source to this inlet, but only when a calibration operation is actually being performed.
EXHAUST outlet Connect exhaust line made of PTEF tubing; minimum O.D ¼”, to
this fitting. The exhaust line should be no longer than 10 meters, and should lead outside the shelter or immedia te area surrounding the instrument.
Figure 3-16 and Figure 3-17 illustrate pneumatic connections for two of the possible basic configurations.
54
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
Source of
SAMPLE GAS
(Remove during
calibration)
VENT
MODEL T701
Zero Gas
Generator
Calibrated
SO
2
Gas
at
span gas
concentration
VENT
here if input
is pressurized
SAMPLE
EXHAUST
Chassis
Source of
SAMPLE GAS
(Remove during
calibration)
VENT
MODEL T701
Zero Gas
Generator
Calibrated
SO
2
Gas
at span gas
concentration
VENT
here if input
is pressurized
Model T700 Gas
Dilution
Calibrator
SAMPLE
EXHAUST
Chassis
06807F DCN7335
Figure 3-16: Pneumatic Connections–Basic Configurati o n–Using Bottled Span Gas
Figure 3-17: Pneumatic Connections–Basic Configurati o n–Using Gas Dilution Calibrator
55
Getting Started Teledyne API - T100 UV Fluorescence SO
SAMPLE
PRESSURE
SENSOR
FLOW PRESSURE
SENSOR PCA
Chassis
EXHAUST
gas outlet
SAMPLE
gas inlet
UV
LAMP
PMT
REACTION
CELL
PUMP
Particulate
Filter
HYDROCARBON
SCRUBBER
(Kicker)
VACUUM MANIFOLD
FLOW
SENSOR
06807F DCN7335
3.3.2.2. PNEUMATIC LAYOUT FOR BASIC CONF IG URATI ON
Analyzer
2
Figure 3-18: T100 Gas Flow, Basic Configuration
56
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
FLOW
SENSOR
FLOW / PRESSURE
SENSOR PCA
SAMPLE
PRESSURE
SENSOR
VACUUM MANIFOLD
FLOW CONTROL ASSY
EXHAUST TO OUTER
LAYER OF KICKER
SAMPLE/CAL
VALVE
SAMPLE GAS
INLET
ZERO AIR INLET
SPAN 1 INLET
ZERO/SPAN VALVE
SAMPLE FILTER
Chassis
EXHAUST GAS
OUTLET
KICKER EXHAUST
TO PUMP
HYDROCARBON
SCRUBBER
(KICKER)
UV LAMP
PMT
SAMPLE
CHAMBER
PUMP
NC
NO
COM
NC
NO
COM
06807F DCN7335
3.3.2.3. PNEUMATIC LAYOUT FOR ZERO/SPAN VALVES OPTION
Figure 3-19 shows the internal, pneumatic connections for a T100 with the zero/span valve option ins ta ll e d.
Figure 3-19: Pneumatic Layout with Zero/Span Valves Option
Table 3-10 describes the state of each valve during the analyzer’s vari ous operational modes.
Table 3-10: Zero/Span and Sample/Cal Valve Operating States
MODE VALVE CONDITION
SAMPLE
ZERO CAL
SPAN CAL
Sample/Cal Open to SAMPLE inlet
Zero/Span Open to ZERO AIR inlet
Sample/Cal Open to zero/span inlet
Zero/Span Open to ZERO AIR inlet
Sample/Cal Open to zero/span inlet
Zero/Span Open to SPAN GAS inlet
The state of the zero/span valves can also be controlled by any of the following means:
manually from the analyzer’s front panel by using the SIGNAL I/O controls located within the DIAG Menu (refer to Section 5.9.1)
by activating the instrument’s AutoCal feature (refer to Section 9.8)
57
Getting Started Teledyne API - T100 UV Fluorescence SO
06807F DCN7335
remotely by using the external digital control inputs (refer to Section 8.1.2 and
Section 9.7.1)
remotely through the RS-232/485 serial I/O ports (refer to Appendix A-6 for the
appropriate commands)
Sources of zero and span gas must be capable of supplying at least 1.55 L/min. (maximum 2.5L/min). Both supply lines should be vented outside of the analyzer’s enclosure. In order to prevent back-diffusion and pressure effects, these vent lines should be between 2 and 10 meters in length.
3.3.2.4. PNEUMATIC LAYOUT FOR PRESSURIZED SPAN/AMBIENT ZERO OPTION
Figure 3-21 shows the internal, pneumatic connections for the analyzer with the pressurized span ambient zero option installed.
Analyzer
2
58
Figure 3-20: Pneumatic Layout with Pressurized Span/Ambient Zero Option
Table 3-11 describes the state of each valve during the analyzer’s vari ous operational modes.
Table 3-11: Zero/Span and Sample/Cal Valve Operating States
MODE VALVE CONDITION
SAMPLE
ZERO CAL
SPAN CAL
Sample/Cal Open to SAMPLE inlet
Zero/Span Open to ZERO AIR inlet
Sample/Cal Open to zero/span inlet
Zero/Span Open to ZERO AIR inlet
Sample/Cal Open to zero/span inlet
Zero/Span Open to SPAN GAS inlet
Press Span Open to SPAN GAS inlet
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
EXHAUST TO OUTER LAYER
OF KICKER
VACUUM MANIFOLD
CRITICAL FLOW ORIFICE
CRITICAL FLOW ORIFICE
ZERO AIR
SCRUBBER
SAMPLE/CAL
VALVE
Chassis
EXHAUST GAS
OUTLET
SAMPLE GAS
INLET
KICKER EXHAUST
TO PUMP
HYDROCARBON
SCRUBBER
(KICKER)
ZERO AIR INLET
UV LAMP
PMT
SAMPLE
CHAMBER
FLOW
SENSOR
FLOW / PRESSURE
SENSOR PCA
SAMPLE
PRESSURE
SENSOR
PUMP
IZS
Permeation
Tube
SO2 Source
ZERO/SPAN VALVE
SAMPLE FILTER
NC
NO COM
NC
NO COM
06807F DCN7335
The state of the zero/span valves can also be controlled by any of the following means:
manually from the analyzer’s front panel by using the SIGNAL I/O controls located within the DIAG Menu (refer to Section 5.9.1)
by activating the instrument’s AutoCal feature (refer to Section 9.8)
remotely by using the external digital control inputs (refer to Section 8.1.2 and
Section 9.7.1)
remotely through the RS-232/485 serial I/O ports (refer to Appendix A-6 for the appropriate commands)
Sources of zero and span gas must be capable of supplying at least 1.55 L/min. (maximum 2.5L/min). The supply line should be vented outside of the analyzer’s enclosure. In order to prevent back-diffusion and pressure effects, these vent lines should be between 2 and 10 meters in length.
3.3.2.5. PNEUMATIC LAYOUT FOR INTERNAL ZERO/SPAN (IZS) GAS GENERATOR OPTION
Figure 3-21 shows the internal, pneumatic connections for the analyzer with the IZS option installed.
Figure 3-21: Pneumatic Layout with IZS Options
59
Getting Started Teledyne API - T100 UV Fluorescence SO
06807F DCN7335
The inte rnal zero air and span gas generato r (IZS) option includes a heated enclosure (Section 3.3.2.6) for a permeation tube (permeation tube must be purchased separately; see Section 1.4, in SO
IZS Permeation Tubes option), an external scrubber (Section
2
3.3.2.8) for producing zero air and a set of valves for switching between the sample gas inlet and the output of the zero/span subsystem, functionally very similar to the valves included in the zero/span valve option.
Table 3-12 describes the operational state of the valves associated with the IZS option during the analyzer’s various operating modes.
Table 3-12: IZS Valve Operating States
MODE VALVE CONDITION
SAMPLE
ZERO CAL
SPAN CAL
Sample/Cal Open to SAMPLE inlet
Zero/Span Open to ZERO AIR inlet
Sample/Cal Open to zero/span valve
Zero/Span Open to ZERO AIR inlet
Sample/Cal Open to zero/span valve
Zero/Span Open to SPAN GAS inlet
The state of the IZS valves can also be controlled by any of the following means:
Analyzer
2
Manually from the analyzer’s front panel by using the SIGNAL I/O controls under
the DIAG Menu (refer to Section 5.9.1),
By activating the instrument’s AutoCal feature (refer to Section 9.8),
Remotely by using the external digital control inputs (refer to Section 8.1.2 and
Section 9.7.1),
Remotely through the RS-232/485 serial I/O ports (refer to Appendix A-6 for the
applicable commands), or
Remotely via Ethernet
Note
The permeation tube is not included in the IZS Option and m ust be ordered separately. Refer to Section 1.4 for permeation tube options.
3.3.2.6. PERMEATION TUBE HEATER
In order to keep the permeation rate constant, the IZS e nclosure is heated to a constant 50 C (10° above the maximum operating temperature of the instrument ). The IZS heater is controlled by a precise PID (Proportional/Integral/Derivative) temperature control loop. A thermistor measures the actual temperature and reports it to the CPU for control feedback.
60
The IZS option includes an external zero air scrubber assembly that removes all SO zero air source. The scrubber is filled with activated charcoal.
the
2
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
06807F DCN7335
3.3.2.7. SPAN G AS CONCENTRATION V ARIATION
Span gas is created when zero air passes over a permeation tube containing liquid SO under high pressure, which slowly permeates through a PTFE membrane into the surrounding air. The speed at which the SO effusion rate. The concentration of the span gas is determined by three factors:
Size of the membrane: The larger the area of the membrane, the more permeation occurs.
Temperature of the SO inside the tube and therefore increases the effusion rate.
Flow rate of the zero air: If the previous two variables are constant, the permeation rate of air into the zero air stream will be constant. Therefore, a lower flow rate of zero air produces higher concentrations of SO flow rate and a constant permeation rate; hence, variations in concentration can be achieved by changing the IZS temperature.
3.3.2.8. EXTERNAL ZERO AIR SCRUBBER
The IZS option includes an external zero air scrubber assembly that removes all SO from the zero air source. The scrubber is filled with activated charcoal.
permeates the membrane is called the
2
: Increasing the temperature of the increases the pressure
2
. The T100 usually has a constant
2
2
2
61
Getting Started Teledyne API - T100 UV Fluorescence SO
FLOW PRESSURE
SENSOR PCA
Chassis
with O
2
Sensor Option
EXHAUST
GAS OUTLET
UV
LAMP
PMT
REACTION
CELL
Particulate
Filter
HYDROCARBON
SCRUBBER
(Kicker)
VACUUM MANIFOLD
O
2
Sensor
Flow Control
O2
Sensor
SAMPLE
GAS INLET
PUMP
SAMPLE
PRESSURE
SENSOR
FLOW
SENSOR
06807F DCN7335
3.3.2.9. PNEUMATIC LAYOUT WITH O2 SENSOR OPTION
Figure 3-22 shows the internal, pneumatic connections for the analyzer with the oxygen (O
) sensor option installed. Pneumatically, the O2 sensor draws a flow of 80 cm³/min in
2
addition to the normal sample flow rate. It is separately controlled with its own critical flow orifice.
Analyzer
2
Figure 3-22: Pneumatic Layout with O
Sensor
2
62
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
FLOW PRESSURE
SENSOR PCA
Chassis
with CO
2
Sensor Option
EXHAUST
GAS OUTLET
UV
LAMP
PMT
REACTION
CELL
PUMP
Particulate
Filter
HYDROCARBON
SCRUBBER
(Kicker)
VACUUM MANIFOLD
CO
2
Probe
SAMPLE
GAS INLET
SAMPLE
PRESSURE
SENSOR
FLOW
SENSOR
06807F DCN7335
3.3.2.10. PNEUMATIC LAYOUT WITH CO2 SENSOR OP TION
Figure 3-23 shows the internal, pneumatic connections for the analyzer with the carbon dioxide (CO
) sensor option installed. Pneumatically, the CO2 sensor is placed in line
2
with the sample gas line between the particulate filter and the analyzer’s sample chamber. It does not alter the gas flow rate of the sample through the analyzer.
Figure 3-23: Pneumatic Layout with CO
3.3.2.11. ABOUT ZERO AIR AND CALIBRATION (SPAN) GASES
Zero air and span gas are required for accurate calibration.
ZERO AIR
Zero air is a gas that is similar in chemical composition to the earth’s atmosphere but without the gas being measured by the analyzer, in this case SO equipped with an Internal Zer o Span (IZS) or an external zero air scrubber option, it is capable of creating zero air.
For analyzers without an IZS or external zero air scrubber option, a zero air generator such as the Teledyne API Model 701 can be used (Figure 3-16).
Sensor
2
. If your analyzer is
2
63
Getting Started Teledyne API - T100 UV Fluorescence SO
1
06807F DCN7335
CALIBRATION (SPAN) GAS
Calibration gas is specifically mixed to match the chemical composition of the type of gas being measured at near full scale of the desired measurement range. In this case,
measurements made with the Teledyne API T100 UV Fluorescence SO2 Analyzer,
SO
2
it is recommended that you use a span gas with a SO measurement range for your application.
EXAMPLE: If the application is to measure between 0 ppm and 500 ppb, an appropriate span gas concentration would be 450 ppb SO
.
2
concentration equal to 80% of the
2
Analyzer
2
Cylinders of calibrated SO
gas traceable to NIST-Standard Reference Material
2
specifications (also referred to as SRM’s or EPA protocol calibration gases) are commercially available. Table 3-13 lists specific NIST-SRM reference numbers for various concentrations of SO
Table 3-13: NIST-SRM's Available for Traceability of SO2 Calibration Gases
NIST-SRM Type Nominal Concentration
1693a Sulfur dioxide in N2 50 ppm 1694a Sulfur dioxide in N2 100 pp 1661a Sulfur dioxide in N2 500 ppm
2659a1 O2 in N2 21% by weight
2626a CO2 in N2 4% by weight 27452 CO2 in N2 16% by weight
Used to calibrate optional O2 sensor.
2
Used to calibrate optional CO2 sensor.
.
2
SPAN GAS FOR MULTIPOINT CALIBRATION
Some applications, such as EPA monitoring, require a multipoint calibration procedure where span gases of different concentrations are needed. We recommend using a bottle of calibrated SO calibrator such as a Teledyne API Model T700 (Figure 3-17). This typ e of calibrator precisely mixes a high concentration gas with zero air (both supplied externally) to accurately produce span gas of the correct concentration. Linearity profiles can be automated with this model and run unattended over night.
gas of higher concentration in conjunction with a gas dilution
2
64
If a dynamic dilution system is used to dilute high concentration gas standards to low, ambient concentrations, ensure that the SO
concentration of the reference gas matches
2
the dilution range of the calibrator. Choose the SO
gas concentr ation so t hat the dynamic dilution system operates in its
2
mid-range and not at the extremes of its dilution capabilities.
EXAMPLE:
A dilution calibrator with 10-10000 dilution ratio will not be able to accurately
dilute a 5000 ppm SO
gas to a final concentration of 500 ppb, as this would operate
2
at the very extreme dilution setting.
A 100 ppm SO
gas in nitrogen is much more suitable to calibrate the T100 analyzer
2
(dilution ratio of 222, in the mid-range of the system’s capabilities).
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
Suppresses the

warning messages

Press CLR to clear the current
message.
(If more than one warning is
active, the next message will
take its place until all warning
messages have been cleared.)
Once the last warning has been cleared,
the RANGE function will be displayed in
the analyzer’s Param field.
NOTE:
If a warning message persists after
several attempts to clear it, the message
may indicate a real problem and not an
artifact of the warm-up period
SAMPLE SYSTEM RESET
TEST CAL CLR SETUP
SAMPLE RANGE=500.0 PPB
<TST TST> CAL MSG SETUP
SAMPLE SYSTEM RESET
TEST CAL CLR SETUP
MSG returns the active
warnings to the message
field.
SAMPLE RANGE=500.0 PPB
<TST TST> CAL SETUP
06807F DCN7335
STARTUP, FUNCTIONAL CHECKS, AND INITIAL 3.4.
CALIBRATION
If you are unfamiliar with the T100 principles of operation, we recommend that you read Section 12. For information on navigating the analyzer’s software me nus, refer to the menu trees provided in Appendix A.
CAUTION - GENERAL SAFETY HAZARD
Do not look at the UV lamp while the unit is operating. UV light can cause eye damage. Always use safety glasses made from UV blocking

STARTUP 3.4.1.

material whenever working with the UV Lamp. (Generic plastic glasses are not adequate).
After the electrical and pneumatic connections are made, run an initial functional check. Turn on the instrument. The pump and exhaust fan should start immediately. The dis play will show a momentary splash screen of the Teledyne API l ogo and other information during the initialization process while t he CPU loads the operating system, the firmware and the configuration data.
WARNING MESSAGES 3.4.2.
The analyzer should automatically switch to Sample Mode after completing the boot-up sequence and start monitoring the
gas. However, there is an approximately one hour warm-up period before reliable gas measurements can be taken. During the warm-up period, the front panel display may show messages in the Parameters field.
Because internal temperatures and other conditions may be outside the specified limits during the analyzer’s warm-up period, the software will suppress most warning conditions for 30 minutes after power up. If warning messages persist after the 60 minutes warm up period is over, investigate their cause using the troubleshooting guidelines in Section 11.1.1.
To view and clear warning messages, press:
Figure 3-24: Warning Messages
65
Getting Started Teledyne API - T100 UV Fluorescence SO
Dark offset above limit specified indicating that too much stray light is present in
1
06807F DCN7335
Table 3-14 lists brief descriptions of the warning messages that may occur during start up for T100 analyzers with no options installed.
Table 3-14: Possible Startup Warning Messages – T100 Analyzers w/o Options
Message Meaning
Analyzer
2
ANALOG CAL W ARNING
BOX TEMP WARNING
CANNOT DYN SPAN2
CANNOT DYN ZERO3
CONFIG INITIALIZED
DARK CAL WARNING
DATA INITIALIZED
HVPS WARNING
PMT DET WARNING
PMT TEMP WARNING
RCELL TEMP WARNING
REAR BOARD NOT DET
RELAY BOARD WARN
The instrument's A/D circuitry or one of its analog outputs is not calibrate d.
The temperature inside the T100 chassis is outs ide the specified limits. Remote span calibration failed while the dynamic span feature was set to
turned on. Remote zero calibration failed while the dynamic zero feature was set to turned
on. Configuration was reset to factory defaults or was erased.
the sample chamber. DAS data storage was erased. High voltage power supply for the PMT is outside of specified limits.
PMT detector output is outside of operational limits. PMT temperature is outside of specified limits. Sample chamber temperature is outside of specified limits.
CPU unable to communicate with motherboard. CPU is unable to communicate with the relay PCA.
SAMPLE FLOW WARN
SAMPLE PRESS WARN
SYSTEM RESET1
UV LAMP W ARNING
Cl ears 45 minutes after power up.
2
Clears the next time successful zero calibration is performed.
3
Cl ears the next time successful span calibration is performed.
66
The flow rate of the sample gas is outside the specified limits. Sample gas pressure outside of operational parameters. The computer was rebooted.
The UV lamp intensity measured by the reference detector reading too low or too high.
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
1
06807F DCN7335
Table 3-15 lists brief descriptions of the warning messages that may occur during start up for T100 analyzers with optional second gas options or alarms installed.
Table 3-15: Possible Startup Warning Messages – T100 Analyzers with Options
Message Meaning
O2 CELL TEMP WARN
1
O2 sensor cell temperature outside of warning limits.
IZS TEMP WARNING
O2 ALARM 1 WARN
O2 ALARM 2 WARN
CO2 ALARM 1 WARN
CO2 ALARM 2 WARN
SO2 ALARM1 WARN
SO2 ALARM2 WARN
Only appears when the optional O2 sensor is installed.
2
Only appears when the optional internal zero span (IZS) option is installed.
3
Only appears when the optional CO2 sensor is installe d.
4
Only appears when the optional gas concentration alarms are installed
2
1, 4
1, 4
3, 4
3, 4`
4
4
On units with IZS options installed: The permeation tube temperature is outside of specified limits.
O2 Alarm limit #1 has been triggered. O2 Alarm limit #2 has been triggered. CO2 Alarm limit #1 has been triggered. CO2 Alarm limit #2 has been triggered. SO2 Alarm limit #1 has been triggered. SO2 Alarm limit #2 has been triggered.

FUNCTIONAL CHECKS 3.4.3.

After the analyzer’s components have warmed up for at least 60 minutes, verify that the software properly supports any hardware options that were installed.
4
4
4
4
4
4
For information on navigating through the analyzer’s software menus, refer to the menu trees described in Appendix A.
Check to ensure that the analyzer is functioning within allowable operating parameters.
Appendix C includes a list of test functions viewable from the analyzer’s front panel
as well as their expected values.
These functions are also useful tools for diagnosing performance problems with
your analyzer (refer to Section 11.1.2).
The Final Test and Validation Dat a Sheet (P/N 04551) included in the shipment
lists these values before the instrument left the factory.
67
Getting Started Teledyne API - T100 UV Fluorescence SO
1
This will match the currently selected units of measure for the range being displayed.
2
Only appears if the CO2 sensor option is installed.
3
Only appears if the O2 sensor option is installed.
4
The STABIL function can be set to display data related to any of the gasses the analyzer measures, e.g. (if either the CO
2
option or the O
2
sensor option is installed).
5
Only appears if the IZS option is installed.
6
Only appears if analog output A4 is actively reporting a
TEST FUNCTION
SAMPLE RANGE=500.00 PPB SO2=XXXX
<TST TST> CAL SETUP
Toggle <TST TST> buttons
to scroll through list of
functions
RANGE=[Value]PPB
1
RANGE1=[Value]PPB
1
RANGE2=[Value]PPB
1
SO2 RNG=[Value]PPB
1, 2, 3
SO2 RN1=[Value]PPB
1, 2, 3
SO2 RN2=[Value]PPB
1, 2, 3
CO2 RNG=[Value]%
2
O2 RNG=[Value]%
3
STABIL=[Value]PPM
4
RSP=[Value] SEC
PRES=[Value]IN-HG-A
SAMP FL=[Value]CC/M
PMT =[Value]MV
NORM PMT=[Value]MV
UV LAMP=[Value]MV
LAMP RATIO=[Value]%
STR LGT=[Value]PPB
DRK PMT=[Value]MV
DRK LMP =[Value]MV
SLOPE=[Value]
OFFSET=[Value]MV
CO2 SLOPE=[Value]
2
CO2 OFFSET=[Value]MV
2
O2 SLOPE=[Value]
3
O2 OFFSET=[Value]MV
3
HVPS =[Value]VOLTS
RCELL ON=[Value]SEC
RCELL TEMP=[Value]ºC
O2 CELL TEMP=[Value]ºC
2
BOX TEMP=[Value]ºC
PMT TEMP=[Value]º
IZS TEMP=[Value]ºC
5
PHT DRIVE =[Value]MV
TEST=[Value]MV
6
TIME=[HH:MM:SS]
06807F DCN7335
To view the current values of these parameters press the following control button sequence on the analyzer’s front panel. Remember until the unit has completed its warm up these parameters may not have stabilized.
Analyzer
2
68
Figure 3-25: Functional Check
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
06807F DCN7335
INITIAL CALIBRATION 3.4.4.
To perform the following calibration you must have sources for zero air and span gas available for input into the sample port on the back of the anal yzer. Refer to Section
3.3.2 for instructions for connecting these gas sources. The initial calibration should be carried out using the same reporting range set-up as
used during the analyzer’s factory calibration. This will allow you to compare your calibration results to the factory calibration as listed on the Final Test and Validation Data Sheet, P/N 04551.
If both available DAS parameters for a specific gas type are being reported via the instrument’s analog outputs (e.g. CONC1 and CONC2 when the DUAL range mode is activated), separate calibrations should be carried out for each parameter.
Use the LOW button when calibrating for CONC1 (equivalent to RANGE1).
Use the HIGH button when calibrating for CONC2 (equivalent to RANGE2).
Refer to the Configurable Analog Output Addendum, P/N 06270 for more information on the configurable analog output reporting ranges.
Note
The following procedure assumes that the instrument does not have any of the available valve options installed. Refer to Section 9.4 for instructions for calibrating instruments possessing valve options
Note The T100 analyzer has been tested for its ability to reject interference for
most sources. See Section 12.1.9 for more information on this topic.
3.4.4.1. INITIAL CALIBRATION PROCEDURE FOR BASIC ANALYZERS (NO 2ND GAS OPTION)
The following procedure assumes that:
The instrument DOES NOT have any of the available calibrati on valve or gas inl et
options installed;
Cal gas will be supplied t hrough the SAMPLE gas inlet on the back of the analyzer
(refer to Figure 3-4);
The pneumatic setup matches that described in Section 3.3.2.
VERIFYING THE REPORTING RANGE SETTINGS
While it is possible to perform the following procedure with any range setting we recommend that you perform this initial checkout using following reporting range settings:
Unit of Measure: PPB
Analog Output Reporting Range: 500.0 ppb
Mode Setting: SNGL
69
Getting Started Teledyne API - T100 UV Fluorescence SO
SETUP RANGE MODE:SN GL
SNGL DUAL AUTO ENTR EXIT
SAMPLE RANGE=500.0 PPB SO2= XXXX
<TST TST> CAL SETUP
SETUP PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP RANGE CONTROL MENU
MODE SET UNIT EXIT
SETUP RANGE: 500.0 Conc
0 0 5 0 0 .0 ENTR EXIT
SETUP CONC UNITS:PPM
PPB PPM UGM MGM ENTR EXIT
SETUP RANGE CONTROL MENU MODE SET UNIT EXIT
SETUP RANGE CONTROL MENU MODE SET UNIT EXIT
Verify that the MODE
is set for SNGL. If it is n o t, press
SINGL th en ENTR
Press EXIT
3x’s to return
the uni t to the
SAMPLE mode
Verify that the RANGE is
set fo r 500.0
If it is n o t, toggle each
numeric button until the
proper range is set, then
press ENTR.
Verify that the UNITs
is set for PPM
If it is n o t, press
PPM then ENT R
06807F DCN7335
While these are the default settings for the T100 analyzer, it is recommended that you verify them before proceeding with the calibration procedure, by pressing:
Analyzer
2
Figure 3-26: Reporting Range Verification
70
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
EXAMPLE
SAMPLE RANGE=500.0 PPB SO2= XXXX
<TST TST> CAL SETUP
SETUP PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP RANGE CONTROL MENU
UNIT DIL EXIT
SETUP SO2 DIL FACTOR:1.0 Gain
0 0 0 0 1 .0 ENTR EXIT
SETUP SO2 DIL FACTOR 20.0 Gain
0 0 0 2 0 .0 ENTR EXIT
Toggle these buttons to
set the dilution factor. This is the number by
which the analyze r will
multiply the SO
2
concentration of the gas
passing through the
reaction cell.
EXIT ignores the
new sett ing.
ENTR accepts the
new sett ing.
SETUP CO2 DIL FAC TOR:1.0 Gain
0 0 0 0 1 .0 ENTR EXIT
SETUP O2 DIL FACTOR:1.0 Gain
0 0 0 0 1 .0 ENTR EXIT
Only appears if the
optiona l CO
2
sensor
is instal le d .
Only appears if the optiona l O
2
sensor
is instal le d .
06807F DCN7335
DILUTION RATIO SETUP
If the dilution ratio option is enabled on your T100 and your application involves diluting the sample gas before it enters the analyzer, set the dilution ratio as follows:
Figure 3-27: Dilution Ratio Setup
71
Getting Started Teledyne API - T100 UV Fluorescence SO
M-P CAL SO2 SPAN CONC:400.0 Conc
0 4 0 0 .0 0 ENTR EXIT
SAMPLE GAS TO CAL:CO
CO O2 ENTR EXIT
SAMPLE RANGE=500.0 PPB SO2= XXXX
<TST TST> CAL SETUP
M-P CAL RANGE=400.0 PPB SO2= XXXX
<TST TST> ZERO SPAN CONC EXIT
SAMPLE RANGE TO CAL:LOW
LOW HIGH ENTR EXIT
The SO
2
span concentration value
is automatically default to
400.0 PPB.
If this is not the the concentration of
the span gas being used, toggle
these buttons to set the correct
concentration of the SO
2
calibration
gas.
EXIT ignores the new setting and returns to
the previous display.
ENTR accepts the new
setting and returns to
the
CONCENTRATION
MENU.
Only appears if either
the optional O
2
or CO
2
sensors are installed.
Only appears if either the analyzer is set for
DUAL range mode.
06807F DCN7335
SET SO2 SPAN GAS CONCENTRATION
Set the expected SO2 span gas concentration. This should be 80% of the concentration range for which the analyzer’s analog output range is set.
Analyzer
2
72
Figure 3-28: SO
Span Gas Setting
2
Teledyne API - T100 UV Fluorescence SO2 Analyzer Getting Started
Press ENTR to changes
the OFFSET & SLOPE
values for the SO
2
measurements.
Press EXIT to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE STABIL= XXXX PPB SO2=XXX.X < TST TST > CAL SETUP
M-P CAL STABIL= XXXX PPB SO2=XXX.X <TST TST> ZERO CONC EXIT
EXIT at this point
returns to the
SAMPLE menu.
Press ENTR to changes
the OFFSET & SLOPE
values for the SO2
measurements.
Press EXIT to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE RANGE=500.0 PPB SO2= XXXX < TST TST > CAL SETUP
Set the Display to show
the STABIL test function.
This function calculates
the stability of the SO
2
measurement.
Toggle TST> button until ...
Allow zero gas to enter the sample port
at the rear of the analyzer.
Wait until STABIL
falls below 0.5 ppb.
This may take several
minutes.
SAMPLE STABIL= XXXX PPB SO2=XXX.X < TST TST > CAL SETUP
M-P CAL STABIL= XXXX PPB SO2=XXX.X
<TST TST> ENTR CONC EXIT
M-P CAL STABIL= XXXX PPB SO2=XXX.X
<TST TST> ZERO SPAN CONC EXIT
Allow span gas to enter the sample port
at the rear of the analyzer.
Wait until STABIL
falls below 0.5 ppb.
This may take several
minutes.
SAMPLE STABIL= XXXX PPB SO2=XXX.X < TST TST > CAL SETUP
M-P CAL STABIL= XXXX PPB SO2=XXX.X
<TST TST> ENTR CONC EXIT
M-P CAL STABIL= XXXX PPB SO2=XXX.X
<TST TST> ENTR CONC EXIT
The SPAN button now
appears during the transition
from zero to span.
You may see both buttons.
If either the ZERO or SPAN
buttons fail to appear see the
Troubleshooting section for
tips.
06807F DCN7335
ZERO/SPAN CALIBRATION
To perform the zero/span calibration procedure (see Section 9 for manual and automat ic calibration procedures, including with valve options), press:
Figure 3-29: Zero/Span Calibration Procedure
73
Getting Started Teledyne API - T100 UV Fluorescence SO
06807F DCN7335
3.4.4.2. CALIBRATION PROCEDURE FOR THE O2 OPTION
Analyzer
2
If your analyzer is equipped with the optional O during installation of the instrument. Refer to Section 9.10.1 for instructions.
3.4.4.3. CALIBRATION PROCEDURE FOR THE CO
If your analyzer is equipped with the optional CO calibrated during installation of the instrument. Refer to Section 9.10.2 for instructions.
Note
Once you have completed the above set-up procedures, please fill out the Quality Questionnaire that was shipped with your unit and return it to Teledyne API. This information is vital to our efforts in continuously improving our service and our products. THANK YOU.
OPTION
2
sensor, this sensor should be calibrated
2
sensor, this sensor should be
2
74
06807F DCN7335

4. OVERVIEW OF OPERATING MODES

To assist in navigating the analyzer’s software, a series of menu trees can be found in Appendix A of this manual.
Note
Some control buttons on the touch screen do not appear if they are not applicable to the menu that you’re in, the task that you are performing, the command you are attempting to send, or to incorrect settings input by the user. For example, the ENTR button may disappear if you input a setting that is invalid or out of the allowable range for that parameter, such as trying to set the 24-hour clock to 25:00:00. Once you adjust the setting to an allowable value, the ENTR button will re-appear.
The T100 software has a variety of operating modes. Most commonly, the analyzer will be operating in SAMPLE mode. In this mode, a continuous read-out of the SO
2
concentration can be viewed on the front panel and output as an analog voltage from rear panel terminals, calibrations can be performed, and TEST functions and WARNING messages can be examined.
The second most important operating mode is SETUP mode. This mode is used for performing certain configuration operations, such as for the DAS system, the reporting ranges, or the serial (RS-232 / RS-485 / Ethernet) communication channels. The SET UP mode is also used for performing various diagnostic tests during troubleshooting.
75
Overview of Operating Modes Teledyne API - T100 UV Fluorescence SO2 Analyzer
DIAG
M-P CAL1
SPAN CAL R2 ZERO CAL M
2
1
06807F DCN7335
Figure 4-1: Front Panel Display
The Mode field of the front panel display indicates to the user which operating mode the unit is currently running.
In addition to SAMPLE and SETUP, other modes available are presented in Table 4-1.
Table 4-1: Analyzer Operating Modes
MODE EXPLANATION
One of the analyzer’s diagnostic modes is active (refer to Section 5.9).
LO CAL A1
LO CAL R1
SAMPLE
SAMPLE A
SETUP
SPAN CAL A2
SPAN CAL M2
ZERO CAL A2
ZERO CAL R2
Other calibration procedures under CAL mode are described separately in Section 9.
2
Only Appears on units with Z/S valve or IZS options.
Unit is performing LOW SPAN (midpoint) calibration initiated automatically by the analyzer’s AUTOCAL feature
Unit is performing LOW SPAN (midpoint) calibration initiated remotely through the COM ports or digital control inputs.
This is the basic calibration mode of the instrument and is activated by pressing the CAL button. Sampling normally, flashing text indicates adaptive filter is on. Indicates that unit is in SAMPLE mode and AUTOCAL feature is activated. SETUP mode is being used to configure the analyzer. The gas measurement will continue during
this process. Unit is performing SPAN calibration initiated automatically by the analyzer’s AUTOCAL feature Unit is performing SPAN calibration initiated manually by the user. Unit is performing SPAN calibration initiated remotely through the COM ports or digital control
inputs. Unit is performing ZERO calibration procedure initiated automatically by the AUTOCAL feature Unit is performing ZERO calibration procedure initiated manually by the user. Unit is performing ZERO calibration procedure initiated remotely through the COM ports or digital
control inputs.
76
Teledyne API - T100 UV Fluorescence SO2 Analyzer Overview of Operating Modes
06807F DCN7335

SAMPLE MODE 4.1.

This is the analyzer’s standard operating mode. In this mode, the instrument is analyzing
and calculating concentrations.
SO
2

TEST FUNCTIONS 4.1.1.

A series of test functions is available at the front panel while the analyzer is in SAMPLE mode. These parameters provide information about the present operating status of the instrument and are useful during troubleshooting (refer to Section 11.1.2). They can also be recorded in one of the DAS channels (refer to Section 7) for data analysis. To view the test functions, press one of the <TST TST> buttons repeatedly in either direction.
77
Overview of Operating Modes Teledyne API - T100 UV Fluorescence SO2 Analyzer
DISPLAY
PARAMETER
UNITS
DESCRIPTION
reference detector divided by the reading
2
2
TIME
1
Only appears if analog output A3 is actively reporting a test function.
06807F DCN7335
Table 4-2: Test Functions Defined
RANGE
RANGE
- -
RANGE1 RANGE2
PPB, PPM,
UGM & MGM
The Full Scale limit at which the reporting range of the analyzer’s ANALOG OUTPUTS is currently set.
THIS IS NOT the Physical Range of the instrument. Refer to Section
5.4 for more information. If DUAL or AUTO Range modes have been selected, two RANGE
functions will appear, one for each range.
STABIL
STABILITY mV Standard deviation of SO
recorded every ten seconds. The calculation uses the last 25 data points.
PRES
SAMP FL
PMT
NORM PMT
UV LAMP
LAMP
RATIO
STR. LGT
DRK PMT
DRK LMP
SLOPE
OFFSET
HVPS
RCELL
TEMP
BOX TEMP
PMT TEMP
IZS TEMP
TEST
SAMPLE
PRESSURE
SAMPLE FLOW
PMT Signal mV The raw output voltage of the PMT.
NORMALIZED
PMT Signal
Source UV Lamp
Intensity
UV Source lamp
ratio
Stray Light ppb
Dark PMT mV
Dark UV Source
Lamp
SO
measurement
Slope
SO
measurement
Offset HVPS V The PMT high voltage power supply.
SAMPLE
CHAMBER TEMP
BOX
TEMPERATURE
PMT
TEMPERATURE
1
IZS
TEMPERATURE
TEST SIGNAL
in-Hg-A
cm³/min
(cc/m)
mV
mV The output voltage of the UV r ef eren ce dete ctor .
%
mV
2
2
- The sensitivity of the instrument as calculated during the last
mV The overall offset of the instrument as calculated during the last
°C The current temperature of the sample chamber.
°C The ambient temperature of the inside of the analyzer case.
°C The current temperature of the PMT.
1
°C
mV Signal of a user-defined test function on output channel A4.
The current pressure of the sample gas as it enters the sample chamber, measured between the SO
The flow rate of the sample gas through the sample chamber. This value is not measured but calculated from the sample pressure.
The output voltage of the PMT after normalization for offset and temperature/pressure c ompensation (if activated).
The current output of the UV stored in the CPU’s memory from the last time a UV Lamp calibration was performed.
The offset due to stray light recorded by the CPU during the last zero­point calibration performed.
The PMT output reading recorded the last time the UV source lamp shutter was closed.
The UV reference detector output reading recorded the last time the UV source lamp shutter was closed.
calibration activity. The slope parameter is used to set the span calibration point of the analyzer.
calibration activity. The offset parameter is used to set the zero point of the analyzer response.
The current temperature of the internal zero/span option. Only appears when IZS option is enabled.
CLOCK TIME hh:mm:ss The current day time for DAS records and calibrat ion ev ents.
Only appears if Internal Gas Span Generator option is ins ta lled.
Concentration readings . Data points are
2
and Auto-Zero valves.
2
2
78
Teledyne API - T100 UV Fluorescence SO2 Analyzer Overview of Operating Modes
1
Only appears if the IZS option is installed.
2
Only appears if analog output A4 is actively reporting a TEST FUNCTION
SAMPLE RANGE=500.00 PPB SO2=XXXX
<TST TST> CAL SETUP
Toggle <TST TST> buttons
to scroll through list of
functions
RANGE
STABIL
RSP
PRES
SAMP FL
PMT
NORM PMT
UV LAMP
LAMP RATIO
STR LGTB
DRK PMT
DRK LMP
SLOPE
OFFSET
HVPS
RCELL ON
RCELL TEMP
BOX TEMP
PMT TEMP
IZS TEMP
1
PHT DRIVE
TEST
2
TIME
A at sea level. It
06807F DCN7335
To view the TEST Functions press the following button sequence:
IMPORTANT IMPACT ON READINGS OR DATA
A value of “XXXX” displayed for any of the TEST functions indicates an out-of-range reading or the analyzer’s inability to calculate it. All pressure measurements are represented in terms of absolute pressure. Absolute, atmospheric pressure is 29.92 in-Hg­decreases about 1 in-Hg per 300 m gain in altitude. A variety of factors such as air conditioning and passing storms can cause changes in the absolute atmospheric pressure.
Figure 4-2: Viewing T100 TEST Functions
79
Overview of Operating Modes Teledyne API - T100 UV Fluorescence SO2 Analyzer
MESSAGE
MEANING
ANALOG CAL W ARNING
The instrument's A/D circuitry or one of its analog outpu ts is not calibrated.
BOX TEMP WARNING
The temperature inside the T100 chassis is outs ide the specified limits.
CANNOT DYN SPAN
Remote span calibration failed while the dynamic span feature was set to turned on
CANNOT DYN ZERO
Remote zero calibration failed while the dynamic zero feature was set to turned on
CONFIG INITIALIZED
Configuration was reset to factory defaults or was erased.
DARK CAL WARNING
Dark offset above limit specified indicating that too much stray light is present in the sample chamber.
DATA INI TIALIZED
DAS data storage was erased.
HVPS WARNING
High voltage power supply for the PMT is outside of specified limits.
IZS TEMP WARNING
On units with IZS options installed: The permeation tube temperature is outside of specified limits .
PMT DET WARNING
PMT detector output outside of operational limits.
PMT TEMP WARNING
PMT temperature is outside of specified limits.
RCELL TEMP WARNING
Sample chamber temperature is outside of specified limits.
REAR BOARD NOT DET
The CPU is unable to communicate with the motherboard.
RELAY BOARD WARN
The firmware is unable to communicate with the relay board.
SAMPLE FLOW WARN
The flow rate of the sample gas is outside the specified limits.
SAMPLE PRESS WARN
Sample pressure outside of operational parameters.
SYSTEM RESET
The computer was rebooted.
UV LAMP W ARNING
The UV lamp intensity measured by the reference detector reading too low or too high
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
TEST CAL
MSG CLR
SETUP
Make sure warning messages are
not due to real problems.
Press CLR to clear the current
message.
If more than one warning is active, the
next message will take its place Once the last warning has been
cleared, the analyzer returns to
SAMPLE mode.
SAMPLE RANGE=500.0 00 PPM SO2=XXX.X < TST TST > CAL MSG CLR SETUP
SAMPLE HVPS WARNING SO2=XXX.X TEST CAL MSG CLR SETUP
TEST ignores warning messages
MSG activates warning
messages.
<TST TST> buttons replaced with
TEST button
NOTE:
If the warning message persists after several attempts to clear it, the message may indicate a real
problem and not an artifact of the
warm-up period.
06807F DCN7335

WARNING MESSAGES 4.1.2.

The most common instrument failures will be reported as a warning on the analyzer’s front panel and through the COMM ports. Section 11.1.1 explains how to use these messages to troubleshoot problems. Section 11.1.3 shows how to view and clear warning messages. Table 4-3 lists the warning messages for the current version of software.
Table 4-3: List of Warning Messages
To view and clear warning messages, press:
Figure 4-3: Viewing and Clearing T100 WARNING Messages
80
Teledyne API - T100 UV Fluorescence SO2 Analyzer Overview of Operating Modes
PASS menu
06807F DCN7335

CALI BRAT IO N M ODE 4.2.

Pressing the CAL button switches the analyzer into calibration mode. In this mode, the user can calibrate the instrument with the use of calibrated zero or span gases.
If the instrument includes either the ze ro/span valve option or IZS option, the displ ay will also include CALZ and CALS buttons. Pressing either of these buttons also puts the instrumen t into multipoint calibration m od e.
The CALZ button is used to initiate a calibration of the zero point.
The CALS button is used to calibrate the span point of the analyzer. It is
recommended that this span calibration is performed at 80% of full scale of the analyzer’s currently selected reporting range.
Because of their critical importance and complexity, calibration operations are described in detail in other sections of the manual:
Section 9 details basic calibration and calibration check operations.
Section 0 provides references for performing an EPA proto col c a lib ration.
IMPORTANT IMPACT ON READINGS OR DATA
To avoid inadvertent adjustments to critical settings, activate calibration security by enabling password protection in the SETUP – (5.5).

SETUP MODE 4.3.

The SETUP mode is used to configure the analyzer’s hardware and software features, perform diagnostic procedures, gat her information on the instrument’s performance and configure or access data from the internal data acquisition system (DAS). For a visual representation of the software menu trees, refer to Appendix A. Setup Mode is divided between Primary and Secondary Setup menus and can be protected through password security.

PASSWORD SECURITY 4.3.1.

Setup Mode can be protected by password security through the SETUP>PASS menu (Section 5.5) to prevent unauthorized or inadverten t configuration adjustm ents .
81
Overview of Operating Modes Teledyne API - T100 UV Fluorescence SO2 Analyzer
CLK
EXIT button is pressed before the ENTR button, the analyzer will beep,
06807F DCN7335

PRIMARY SETUP MENU 4.3.2.

Table 4-4: Primary Setup Mode Features and Functions
MODE OR FEATURE
Analyzer Configuration
Auto Cal Feature
Internal Data Acquisition
(DAS)
Analog Output Reporting
Range Configuration
Calibration Password Security
Internal Clock Configuration
Advanced SETUP features MORE
CONTROL
BUTTON
CFG
ACAL
DAS
RNGE
PASS
Lists key hardware and software configuration information. 5.1 Used to set up and operate the AutoCal feature.
Only appears if the analyzer has one of the internal valve options installed.
Used to set up the DAS system and view recorded data. 5.3 & 7 Used to configure the output signals generated by the
instrument’s Analog outputs. Turns the calibration password protection feature ON/OFF. 5.5 Used to Set or adjust the instrument’s internal clock. 5.6
This button accesses the instruments secondary setup menu.

SECONDARY SETUP MENU (SETUP>MORE) 4.3.3.

Table 4-5: Secondary Setup Mode Features and Functions
MODE OR FEATURE
External Communication
Channel Configuration
System Status Variables
System Diagnostic Features
MENU
ITEM
COMM
VARS
DIAG
Used to set up and operate the analyzer’s various external I/O channels including RS-232; RS 485, modem communi cati on and/or Ethernet access.
Used to view various variables related to the instrument’s current operational status
Used to access a variety of functions that are used to configure, test or diagnose problems with a variety of the analyzer’s basic systems
DESCRIPTION
DESCRIPTION
MANUAL
SECTION
5.2 & 9.8
5.4
See
Table 4-5
MANUAL
SECTION
5.7 & 6
5.8
5.9
IMPORTANT IMPACT ON READINGS OR DATA
Any changes made to a variable during the SETUP procedures are not acknowledged by the instrument until the ENTR button is pressed. If t he
alerting the user that the newly entered value has not been accepted.
82
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL SETUP
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
SAMPLE
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SAMPLE
T100 SO2 A
NALYZER
NEXT PREV EXIT
Press EXIT at
any time to
return to
SETUP menu
Press EXIT at
any time to
return to the
SAMPLE display
Press NEXT of PREV to move back and forth through the following list
of Configuration information:
MODEL NAME
SERIAL NUMBER
• SOFTWARE REVISION
• LIBRARY REVISION
iCHIP SOFTWARE REVISION
1
HESSEN PROTOCOL REVISION
1
ACTIVE SPECIAL SOFTWARE
OPTIONS
1
• CPU TYPE
DATE FACTORY CONFIGURATION
SAVED
1
Only appears if relevant option of Feature is active.
06807F DCN7335

5. SETUP MENU

The SETUP menu is used to set instrument parameters for performing configuration, calibration, reporting and diagnostics operations according to user needs.
SETUP – CFG: CONFIGURATION INFORMATION 5.1.
Pressing the CFG button displays the instrument configuration information. This display lists the analyzer model, serial number, firmware revision, software library revision, CPU type and other information. Use this information to identify the software and hardware when contacting Technical Support. Special instrument or software features or installed options may also be listed here.
Figure 5-1: SETUP – Configuration Information

SETUP – ACAL: AUTOMATIC CALIBRATION OPTION 5.2.

83
The menu button for this option appears only when the instrument has the zero span and/or IZS options. See Section 9.8 for details.
SETUP Menu Teledyne API - T100 UV Fluorescence SO2 Analyzer
Test output
ANALOG OUT
A1 A2 A3 A4 + - + - + - + -
SO2 concentration
outputs
HIGH range when
DUAL mode is selected
(not used in
standard
configuration)
LOW range when
DUAL mode is selected
06807F DCN7335
SETUP – DAS: INTERNAL DATA ACQUISITION SYSTEM 5.3.
Use the SETUP>DAS menu to capture and record data. Refer to Section 7 for configuration and operation details.
SETUP – RNGE: ANALOG OUTPUT REPORTING RANGE 5.4.
CONFIGURATION
Use the SETUP>RNGE menu to configure output reporting ranges, including scaled reporting ranges to handle data resolution challenges. This section describes configuration for Single, Dual, and Auto Range modes.

AVAIL ABLE ANALOG O UTPUT SIGN ALS 5.4.1.

The analyzer has three active ana log output si gnals, accessible through a connector on the rear panel.
Figure 5-2: SETUP – Analog Output Connector
All three outputs can be configured either at the factory or by the user for full scale outputs of 0.1 VDC, 1VDC, 5VDC or 10VDC. Additionally A1 and A2 may be equipped with optional 0-20 mA DC current loop drivers and configured for any current output within that range (e.g. 0 -20, 2-20, 4-20, etc.). The user may also adj ust the signal level and scaling of the actual output voltage or current to match the input requirements of the recorder or data logger (Refer to Section 5.9.3.3 and 5.9.3.5).
In its basic configuration, the A1 and A2 channels of the T100 output a signal that is proportional to the SO
concentration of the sample gas. Several operating mo des are
2
available which allow:
Single range mode (SNGL Mode, refer to Section 5.4.3.1): Both outputs are slaved
together and will represent the same concentration span (e.g. 0-50 ppm); however their electronic signal levels may be configured for different ranges (e.g. 0-10 VDC vs. 0-.1 VDC – Refer to Section 5.9.3).
Dual range mode (DUAL mode, refer to Section 5.4.3.2): The two outputs can to
configured for separate and independent units of measure and measurement spans as well as separate electronic signal levels.
84
Teledyne API - T100 UV Fluorescence SO2 Analyzer SETUP Menu
06807F DCN7335
Auto range mode (AUTO mode, refer to Section 5.4.3.3) gives the analyzer the
ability to automatically switch the A1 and A2 analog outputs between two ranges (low and high) dynamically as the concentration value fluctuates.
EXAMPLE: A1 OUTPUT: Output Signal = 0-5 VDC representing 0-1000 ppm concentration values
A2 OUTPUT: Output Signal = 0 – 10 VDC representing 0-500 ppm concentration values. A3 OUTPUT: Test channel; e.g., PMT signal = 0-5V
Output A4 is not available on the T100 Analyzer in standard configuration.
PHYSICAL RANGE VERSUS ANALOG OUTPUT REPORTING 5.4.2.
RANGES
The entire measurement range of the T100 is 0 – 20,000 ppb, but many applications use only a small part of the analyzer’s full measurement range. This creates two performance challenges:
The width of the T100’s physical range can create data resolution problems for most analog recording devices. For example, in an application where the expected concentration of SO only 0.25% of the instrument’s full 20,000 ppb measurement range. Unmodified, the corresponding output signal would also be recorded across only 0.25% of the range of the recording device.
is typically less than 500 ppb, the full scale of expect ed values is
2
The T100 solves this problem by allowing the user to select a scaled reporting range for the analog outputs that only includes that portion of the physical range relevant to the specific application. Only the reporting range of the analog outputs is scaled, the physical range of the analyzer and the readings displayed on the front panel remain unaltered.
Applications where low concentrations of SO
are measured require greater sensitivity
2
and resolution than typically necessary for measurements of higher concentrations. The T100 solves this issue by using two hardware physical ranges that cover the
instrument’s entire 0 and 20,000 ppb measurement range: a 0 to 2,000 ppb physical range for increased sensitivity and resolution when measuring very low SO concentrations, and a 0 to 20,000 ppb physical range for measuring higher SO
2 2
concentrations. The anal yzer’s software automatically selects which physical range is in effect based on the analog output reporting range selected by the user.
If the high end of the selected reporting range is ≤ 2,000 ppb. The low physical range is
selected.
If the high end of the selected reporting range is 2,001 ppb. The high physical range
is selected.
Once properly calibrated, the analyzer’s front panel display will accurately report concentrations along the entire span of its 0 and 20,000 ppb physical range regardless of which reporting range has been selected for the analog outputs and which physical range is being used by the instrument’s software.
85
SETUP Menu Teledyne API - T100 UV Fluorescence SO2 Analyzer
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL
SETUP
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
SETUP X.X
CFG DAS RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT EXIT
SETUP X.X RANGE MODE: SNGL
SNGL
DUAL AUTO ENTR EXIT
EXIT Returns
to the Main
SAMPLE Display
See the section on Single Range Mode
See the section on
Dual Range Mode
See the section on
Auto Range Mode
Only one of the
range modes may be
active at any time.
06807F DCN7335

REPORTING RANGE MODES: SINGLE, DUAL, AUTO RANGES 5.4.3.

The T100 provides three analog output range modes to choose from:
Single range (SNGL) mode sets a single maximum r ange for the analog output. If
single range is selected (refer to Section 5.4.3.1) both outputs are slaved together and will represent the same measurement span (e.g. 0-50 ppm); however, their electronic signal levels may be configured for different ranges (e.g. 0-10 VDC vs. 0­.1 VDC – Refer to Section 5.9.3.1).
Dual range (DUAL) allows the A1 and A2 outputs to be configured with dif ferent
measurement spans (refer to Section 5.4.3.2).
Auto range (AUTO) mode gives the analyzer to ability to output data via a low
range and high range. When this mode is selected (refer to Section 5.4.3.3) the T100 will automatically switch between the two ranges dynamically as the concentration value fluctuates.
Also, in this mode the RANGE Test function displayed on the front panel during SAMPLE mode will be replaced by two separate functions, Range1 and Range2.
Range status is also output via the External Digital I/O Status Bits (refer to Section
8.1.1). To select the Analog Output Range Type press:
86
Figure 5-3: SETUP RNGE – Reporting Range Mode
Teledyne API - T100 UV Fluorescence SO2 Analyzer SETUP Menu
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X RANGE CO NT RO L MENU
MODE
SET UNIT EXIT
SETUP X.X
RANGE MODE: SNGL
SNGL
IND AUTO ENTR EXIT
EXIT x 2 returns
to the main
SAMPLE display
SETUP X.X RANGE MODE: SNGL
SNGL
IND AUTO ENTR EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SETUP X.X RANGE: 500.0 Conc
0 0 5 0 0 .0 ENTR EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT EXIT
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL
SETUP
06807F DCN7335
5.4.3.1. SINGLE RANGE MODE (SNGL)
The default range mode for the analyzer is single range, in which all analog concentration outputs are set to t he same reporting range. This reporting range can be set to any value between 0.1 ppb and 20,000 ppb.
While the two outputs always have the same reporting range, the span and scaling of their electronic signals may also be configured differently (e.g., A1 = 0-10 V; A2 = 0-
0.1 V). To select SNGLE range mode and to set the upper limit of the range, press:
Figure 5-4: SETUP RNGE – Single Range Mode
87
SETUP Menu Teledyne API - T100 UV Fluorescence SO2 Analyzer
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT EXIT
SETUP X.X
RANGE MODE: SNGL
SNGL DUAL AUTO ENTR EXIT
SETUP X.X RANGE MODE: DUAL
SNGL
DUAL AUTO
ENTR EXIT
SAMPLE
ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
SETUP X.X RANGE CONT RO L MENU
MODE
SET UNIT EXIT
SETUP X.X HIGH RANGE: 500.0 Conc
0 0 5 0 0 .0 ENTR EXIT
SETUP X.X LOW RANGE: 500. 0 Co nc
0 0 1 0 0
.0
ENTR EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT EXIT
EXIT Returns
to the Main
SAMPLE Display
Toggle the
numeral buttons
to set the upper
limit of each
range.
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL SETUP
ranges have separate slopes
06807F DCN7335
5.4.3.2. DUAL RANGE MODE (DUAL)
Selecting Dual Range mode allows the A1 and A2 outputs to be configured with different reporting ranges. The analyzer software calls these two ranges low and high. The Low range setting corresponds wit h the analog output labeled A1 on the rear panel of the instrument. The high range setting corresponds with the A2 output. While the software names these two ranges low and high, they do not have to be configured that way. For example: the low range can be set for a span of 0-150 ppb while the high range is set for 0-50 ppb.
In DUAL range mode the RANGE test function displayed on the front panel will be replaced by two separate functions:
RANGE1: The range setting for the A1 output.
RANGE2: The range setting for the A2 output.
To set the ranges, press the following control button sequence:
.
IMPORTANT IMPACT ON READINGS OR DATA
88
Figure 5-5: SETUP RNGE – Dual Range Mode
In DUAL range mode the LOW and HIGH and offsets for computing SO2 concentration. The two ranges must be independently calibrated.
Teledyne API - T100 UV Fluorescence SO2 Analyzer SETUP Menu
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL
SETUP
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X R ANGE CONTROL MENU
MODE
SET UNIT EXIT
SETUP X.X RANG E MODE: SNGL
SNGL IND AUTO ENTR EXIT
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
EXIT x 2 returns
to the main
SAMPLE display
SETUP X.X RANGE MODE: AUTO
SNGL
IND AUTO ENTR EXIT
SETUP X.X RANG E CONTROL MENU
MODE
SET UNIT EXIT
SETUP X.X HI GH RANGE: 500.0 Conc
0 0 5 0 0 .0 ENTR EXIT
SETUP X.X L OW RANGE: 500.0 Conc
0 0 5 0 0 .0 ENTR EXIT
Toggle the numeral
buttons to set the
LOW and HIGH
range value. ENTR accepts the new setting, EXIT
ignores the new
setting.
06807F DCN7335
5.4.3.3. AUTO RANGE MO DE (AUTO)
In AUTO range mode, the analyzer automatically switches the reporting range between two user-defined ranges (low and high). The unit will switch from low range to high range when the SO return from high range back to low range once both the SO
concentration exceeds 98% of the low range span. The unit will
2
concentration falls below
2
75% of the low range span. In AUTO Range mode the instrument reports the same data in the same range on both
the A1 and A2 outputs and automatically switches both outputs between ranges as described above. Also, the RANGE test function displayed on the front panel will be replaced by two separate functions:
RANGE1: The LOW range setting for all analog outputs.
RANGE2: The HIGH range setting for all analog outputs.
The high/low range status is also reported thr ough the external, digital status bits (refer to Section 8.1.1).
To set individual ranges press the following control button sequence.
Figure 5-6: SETUP RNGE – Auto Range Mode
IMPORTANT IMPACT ON READINGS OR DATA
In AUTO range mode, the LOW and HIGH ranges have separate slopes and offsets for computing SO
concentration. The two ranges must be
2
independently calibrated.
89
SETUP Menu Teledyne API - T100 UV Fluorescence SO2 Analyzer
SETU P X.X CONC UNITS: PPM
PPM PPB UGM MGM ENTER EXIT
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL
SETUP
SETU P X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE
EXIT
SETUP X.X RANGE CONTROL MENU
MODE
SET UNIT EXIT
SETU P X.X CONC UNITS: PPB
PPM PPB UGM MG M ENTER EXIT
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
Select the preferred
concentration unit.
EXIT returns
to the main menu.
ENTR accepts
the new unit, EXIT returns
to the SETUP
menu.
06807F DCN7335

RANGE UNITS 5.4.4.

The T100 c an display concentrations in parts per billion (109 mols per mol, PPB), parts
6
per million (10 milligrams per cubic meter (mg/m
mols per mol, PPM), micrograms per cubic meter (µg/m3, UGM) or
3
, MGM). Changing units affects all of the display, analog outputs, COM port and DAS values for all reporting ranges regardless of the analyzer’s range mode.
To change the concentration units:
Figure 5-7: SETUP RNGE – Concentration Units Selection
IMPORTANT IMPACT ON READINGS OR DATA
Concentrations displayed in mg/m3 and µg/m3 use 0°C and 760 Torr as standard temperature and pressure (STP). Consult your local regulati ons for the STP used by your agency. See Section 5.4.4.1 volumetric to mass units
90
for converting
Teledyne API - T100 UV Fluorescence SO2 Analyzer SETUP Menu
06807F DCN7335
5.4.4.1. CONVERTING MICROGRAMS PER CUBIC M ET ER TO PARTS PER MILLION
The conversion between micrograms per cubic meter and parts per million is based on standard conditions (0
22.414 L. Thus, converting the mass of the pollutant M volume V
in liters at standard temperature and pressure (STP) takes the following
p
o
C and 101.325 kPa) where one mole of an ideal gas occupies
in grams to its equivalent
p
equation:
Vp = [(Mp)/(MW)] x 22.414 L.mol-1
Where MW is the molecular weight of the pollutant in units of grams per mole. For readings made at temperatures and pressures other than standard conditions, the standard volume, 22.414 L
.
mol-1, must be corrected. The ideal gas law to make the correction can
be used:
(22.414 L.mol-1) x [(T2)/(273 K)] x [(101.325 kPa)/(P2)]
Where T2 and P2 are the absolute temperature (in Kelvin) and absolute pressure (in kilopascals) at which the readings were made. Because parts per million is a volume ratio, it can be written as:
ppm = (Vp)/(Va + Vp)
where Va is the volume of the air in cubic meters at the temperature and pressure at which the measurement was taken. Then combine equations to yield:
ppm = {[(Mp)/(MW)] x (22.414 L.mol-1) x [(T2)/(273 K)] x [(101.325 kPa)/(P2)]}
-------------------------------------------------------------------------------------------
(V
) x (1000 L.m-3)
a
where Mp is the mass of the pollutant of interest in micr ograms. The factors converting micrograms to grams and liters to millions of liters cancel one another. Unless otherwise stated, it is assumed that V
= 1.00 m3
a
Example:
A 1-m pressure were 25.0 SO
3
sample of air was found to contain 80 µg.m-3 of SO2. The temperature and
concentration in parts per million?
2
o
C and 103.193 kPa when the air sample was taken. What was t he
Solution:
First, determine the MW of SO
MW of SO2 = 32.06 + 2(15.9994) = 64.06 g.mol-1
:
2
Next, convert the temperature from Celsius to Kelvin:
25oC + 273 K = 298 K
Using the equation derived above, Concentration is:
{[(80 µg)/(64.06 g.mol-1)] x (22.414 L.mol-1) x [(298 K)/(273 K)] x [(101.325 kPa)/(103.193 kPa)]} / [(1 m3) x (1000 L.m3)]
= 0.030 ppm of SO2
91
SETUP Menu Teledyne API - T100 UV Fluorescence SO2 Analyzer
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL SETUP
SETUP PRIMARY SETUP MENU
CFG DAS
RNGE PASS CLK MORE
EXIT
SETUP RANGE CONTROL MENU
MODE
SET UNIT DIL
EXIT
DIL only appears
if the dilution ratio
option has been
installed
SETUP
DIL FACTOR: 1.0 GAIN
0 0 0 1 .0 ENTR EXIT
Toggle to set the dilution factor.
This is the number by which the
analyzer will multiply the SO2
concentrations of the gas passing
through the reaction cell.
SETUP DIL FACTOR: 20.0 GAIN
0 0 2 0 .0 ENTR EXIT
EXIT ignores the
new setting.
ENTR accepts the
new setting.
SAMPLE ENTER SETUP PASS : 818
8 1 8
ENTR EXIT
06807F DCN7335
DILUTION RATIO (OPTION) 5.4.5.
The dilution ratio is a software option that allows the user to compensate for any dilution of the sample gas before it enters the sample inlet. Once t he degree of dilution is known, add an appropriate scaling factor to the analyzer’s SO the measurement range and concentration values reflect the undiluted values when shown on the instrument’s front panel display screen and reported via the analog and serial outputs.
Using the Dilution Rat io option is a 4-step process:
1. Select reporting range units: Follow the procedure in Section 5.4.4
2. Selec t the r eporting range m ode and s et the r eportin g r ange up per limit (see Section
5.4).
Ensure that the upper span limit entered for the reporting range is the maximum
expected concentration of the undiluted gas.
3. Set the dilution factor as a ga in (e.g., a value of 20 m eans 20 parts diluent and 1
part of sample gas):
concentration calculation so that
2
92
Figure 5-8: SETUP RNGE – Dilution Ratio
4. Calibrate the analyzer.
Ensure that the calibration span gas is either supplied through the same dilution
system as the sample gas or has an appropriate ly lower actual concentration.
Teledyne API - T100 UV Fluorescence SO2 Analyzer SETUP Menu
PASSWORD
Maintenance
/Maintenance
06807F DCN7335
EXAMPLE: If the reporting range limit i s set for 100 ppm and the dilution ratio of the sample gas is 20 gain, either:
a span gas with the concentration of 100 ppm can be used if the span gas passes
through the same dilution steps as the sample gas, or;
a 5 ppm span gas must be used if the span gas IS NOT routed through the
dilution system.
The analyzer multiplies the measured gas concentrations with this dilution factor and displays the result.
IMPORTANT IMPACT ON READINGS OR DATA
Once the above settings have been entered, the instrument needs to be recalibrated using one of the methods discussed in Section 9
.

SETUP – PASS: PASSWORD PROTECTION 5.5.

The menu system provides password protection of the calibration and setup functions to prevent unauthorized adjustments. When the passwords have been enabled in the PASS menu, the system will prompt the user for a password anytime a password-protected function (e.g., SETUP) is selected. T his allows normal operation of the instrument, but requires the password (101) to access to the menus under SETUP. When PASSWORD is disabled (SETUP>OFF), any operator can enter the Primary Setup (SETUP) and Secondary Setup (SETUP>MORE) menus. Whether PASSWORD is enabled or disabled, a password (default 818) is required to ente r the VARS or DIAG menus in the SETUP>MORE menu.
Table 5-1: Password Levels
LEVEL MENU ACCESS ALLOWED
Null (000) Operation All functions of the main menu (top level (Primary) menu)
101 818
Configuration/ Configuration
Access to Primary and Secondary SETUP Menus when PASSWORD is enabled
Access to Secondary SETUP Submenus VARS and DIAG whether PASSWORD is enabled or disabled.
93
SETUP Menu Teledyne API - T100 UV Fluorescence SO2 Analyzer
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
SAMPLE RANGE = 500.0 PPB SO2 =XXX.X
< TST TST > CAL SETUP
SETUP PRIMARY SET U P MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP PASSWORD ENA BLE: OFF
OFF ENTR EXIT
SETUP PASSWORD ENABLE: ON
ON ENTR EXIT
SETUP
PASSWORD ENABLE: ON
ON ENTR EXIT
EXIT ignores the
change.
ENTR accepts the
change. Once Password is enabled, exit back
out to the main
menu for this feature
to take effect.
Toggles
password
status On/Off
ENTR accepts
displayed
password value
EXIT returns to
SAMPLE display
PASSWORD
default state is
OFF
06807F DCN7335
To enable or disable passwords, press:
If the password feature is enabled, the default password displayed will be 000 upon entering either Calibration or Setup Mode, and the new password must be input. Example follows for Calibration Mode:
94
Figure 5-9: SETUP – Enable Password Security
Teledyne API - T100 UV Fluorescence SO2 Analyzer SETUP Menu
SAMPLE ENTER SETUP PASS : 0
0 0 0
ENTR EXIT
SAMPLE RANGE = 500.0 PPB SO2 =XXX.X
< TST TST >
CAL C ALZ CALS SETUP
M-P CAL RANGE = 500.0 PPB SO2 =XXX.X
< TST TST > ZERO CONC EXIT
Prompts
password
number
SAMPLE ENTER SETUP PASS : 0
1 0 1 ENTR EXIT
Press
individual
buttons to set
101
Continue calibration process …
06807F DCN7335
Figure 5-10: SETUP – Enter Calibration Mode Using Password
95
SETUP Menu Teledyne API - T100 UV Fluorescence SO2 Analyzer
SETUP X.X TIME-OF-DAY CLOCK
TIME DATE EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS
RNGE PASS CLK MORE
EXIT
EXIT returns
to the main
SAMPLE display
Enter Current
Date-of-Year
SETUP X.X DATE: 01-JAN-02
0 1 JAN 0 2 ENTR EXIT
SETUP X.X DATE: 01-JAN-02
0 1 JAN 0 2 ENTR EXIT
SETUP X.X TIME-OF-DAY CLOCK
TIME DATE EXIT
Enter Current
Time-of-Day
SETU P X.X3 TIME: 12:00
1 2 : 0 0 ENTR EXIT
SETUP X.X
TIME: 12:00
1 2 : 0 0
ENTR EXIT
SETU P X.X PRIMARY SETUP M ENU
CFG DAS
RNGE PASS CLK MORE
EXIT
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL
SETUP
06807F DCN7335
SETUP – CLK: SETTING THE INTERNAL TIME-OF-DAY 5.6.
CLOCK
The T100 has a built-in clock for the AutoCal timer, Time TEST functions, and time stamps on COM port messages and DAS data entries. To set the time-of-day, press:
Figure 5-11: SETUP – Clock
In order to compensate for CPU c locks, which may run fast or slow, there is a variable (in the SETUP>MOR E>VARS menu) to speed up or slow down the clock by a fixed amount every day. To change this variable, press:
96
Teledyne API - T100 UV Fluorescence SO2 Analyzer SETUP Menu
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL SETUP
SAMPLE ENTER SETUP PASS: 818
8 1 8
ENTR EXIT
SETUP X.X PRIMARY SETUP MENU CFG DAS
RNGE PASS CLK MORE
EXIT
3x EXIT returns
to the main SAMPLE display
Enter sign and number of seconds per
day the clock gains (-) or loses (+).
SETUPX.X 0 ) DAS_HOLD_OFF=15.0 Minutes
PREV NEXT JUMP EDIT PRNT EXIT
SETUP X.X
7) CLOCK_ADJ=0 Sec/Day
PREV JUMP
EDIT PRNT EXIT
SETUP X.X CLOCK_ADJ:
0 Sec/Day
+ 0 0 ENTR EXIT
SETUP X.X 8) CLOCK_ADJ=0 Sec/Day
PREV NEXT JUMP EDIT PRNT EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG
EXIT
Continue to press NEXT until …
SAMPLE ENTER SETUP PASS: 818
8 1 8 ENTR EXIT
06807F DCN7335
Figure 5-12: SETUP – Clock Speed Variable
97
SETUP Menu Teledyne API - T100 UV Fluorescence SO2 Analyzer
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL SETUP
SAMPLE
ENTER SETUP PASS : 8
18
8 1 8
ENTR EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS
RNGE PASS CLK
MORE
EXIT
SETUP X.X SECOND A RY SE TU P ME NU
COMM VARS DIAG EXIT
SETUP X.X COMMUNICATIONS MENU
ID INET COM1 COM2 EXIT
If the default password 818 is replaced by 000, then Password Protection has been enabled. Refer to SETUP: PA SS.
06807F DCN7335
SETUP – COMM: COMMUNICATIONS PORTS 5.7.
This section introduces the communications setup menu; Section 6 provides the setup instructions and operation information. Press SETUP>ENTR>MORE>COMM to arrive at the communications menu.
Figure 5-13: SETUP – COMM Menu

ID (INSTRUMENT IDENTIFICATION) 5.7.1.

In the SETUP>MORE>COMM menu press ID to display and/or change the Machine ID, which must be changed to a unique identifier (number) when more than one instrument of the same model is used in a multidrop configuration (Section 3.3.1.8) or when applying MODBUS protocol (Section 6.6.1). The default ID is typically the same as the model number; for the Model T100, the ID is 0100 (but could be 0000). Press any button(s) in the MACHINE ID menu (Figure 5-14) until the Machine ID Parameter field displays the desired identifier.
98
Loading...