Teledyne 8800T User Manual

instruction manual
Model 8800T
TRACE MOISTURE ANALYZER
Teledyne Analytical Instruments A Business Unit of Teledyne Electronic Technologies
i
8800T User’s Manual Table of Contents
1.0 Overview of the 8800T .................................................1
2.1 Precautions using the sensor .........................................3
2.2 Sensor Technical Specifications ...................................4
2.3 Sensor Installation & Sampling Techniques .................4
2.3.1 In-situ Installation ......................................................5
2.3.2 Extractive Installation ................................................7
2.4 Troubleshooting unexpected readings ..........................9
3.1 Precautions using the 8800T .......................................13
3.1.1 Electromagnetic Compatibility Considerations .......13
3.2 Instrument Technical Specifications ...........................14
3.3 Installation ..................................................................15
3.3.1 Mechanical Installation ............................................15
3.3.2 Electrical Installation ...............................................16
3.4 Operating the Instrument ............................................17
3.4.1 Starting up ................................................................17
3.4.2 Display Conventions ................................................17
3.4.3 Push Buttons ............................................................19
3.4.4 Operating State ........................................................19
3.4.4.1 Viewing Dewpoint Mode ......................................20
3.4.4.2 Viewing Temperature at the Sensor ......................20
3.4.4.3 Start Calibration (SpanCheck™) Mode ................21
3.4.4.4 Viewing Serial Number Mode ..............................24
3.4.5 SetUp State ..............................................................24
3.5 Troubleshooting the Instrument ..................................28
Appendix A: Operating State User Interface flowchart ....31
Appendix B: Set-Up State User Interface flowchart .........32
Appendix C: 8800T Mechanical Drawing ........................34
Appendix D: 8800T Electrical Connections .....................35
Appendix E: Sensor/SpanCheck™ Theory of Operation .36
Appendix F: Pressure Correction ......................................39
Appendix G: Current vs. Dewpoint ..................................40
ii
Appendix H: Uncertainty in LBS & ppmV calculations ..43
Appendix I: Sensor Response Time ..................................44
Appendix J: Return Authorization Request ......................45
iii
Section 1: Introduction
1.0 Overview of the 8800T
The 8800T is a microprocessor based 4-20mA loop powered (2 wire) hygrometer, for measuring moisture content in gases in the range from -100°C to +20°C. The measurement is displayed on the instrument’s custom LCD, and is trans­mitted by varying the current drawn (4-20mA) from the power supply. The current varies linearly proportional to the selected measurement units. An optional digital output is available which modulates/demodulates the 4-20mA loop line without interfering with its operation. With this option the 8800T is capable of communicating with properly equipped Personal Computers or other RS-232 capable con­trollers. Three front panel buttons provide the user with a wide variety of features. The 8800T’s advanced design allows it to be housed in a small stainless steel enclosure behind the sensor probe, thus the instrument and sensor are a single integrated unit. The 8800T uses the Teledyne HTF™ sensor which is encap­sulated in sintered stainless steel, thus it is capable of com­ing into contact with a wide variety of environments. However one should keep in mind that the sensor is a sensi­tive device and it should be handled accordingly.
page 1
page 2
Section 2: Sensor and Sampling Techniques

2.1 Precautions using the sensor

The Teledyne HTF™ Al2O3 sensor is designed and field proven to be highly reliable, rugged and maintenance free.
However the user should consider the following precau­tions:
To avoid the need for prolonged dry-down (when expecting to measure dewpoints dryer than -65ºC), do not expose the sensor to room air longer than necessary (1 - 2 minutes). Thus, do not open the sensor container before you are ready to install the sensor.
The sensor container has desiccant to keep the sensor dry during shipping and to avoid damage due to conden­sation. Close the container immediately after removing the sensor to avoid degradation of the desiccant.
Do not throw away the sensor container, you may use it again to transport the sensor between locations, to store it between uses or to ship it back to the factory for certi­fication. The container can be attached to the loop cable, by trapping the cable with the lid strap.
Do not expose the sensor to corrosive gases such as gases containing chlorine, ammonia or HCl. (SO2 can be monitored when the moisture content is low).
Except for the XTR65W sensor:
1. Do not expose the sensor to liquid water, as it may get damaged.
2. Do not breathe directly onto the sensor, as condensa­tion may form which could damage the sensor element.
Do not install the sensor near heat sources such as radia­tors or air ducts.
Do not install the sensor in places subject to extreme mechanical vibration or shock. If this is not avoidable, use resilient mounting. If in doubt, call your representa-
page 3
tive.
Do not disassemble the porous metal filter encapsula­tion, as this will damage the sensor and void your fac­tory warranty.
Prior to installation of the probe, ensure that no contam­inants are present in the system (e.g. oil, liquid water).

2.2 Sensor Technical Specifications

Type: .................................Hyper Thin Film high capacitance Al2O3.
Dewpoint range:
XTR-100........................-148°F to +68°F (-100°C to +20°C)
XTR-65..........................-85°F to +68°F (-65°C to +20°C).
Capacitance:......................15nF to 200nF.
Accuracy:..........................±5.5°F (±3°C).
Repeatability:....................±0.9°F (±0.5°C).
Response time:..................see graph in Appendix I.
Temperature range:...........-10°C to +70°C.
Sample Flow range:
(linear velocity @ 1ATM):Static to 100m/s.
Storage temperature:. ........-40°F to+176°F (-40°C to +80°C).
Mechanical:.......................encapsulated in 100µ sintered stainless steel.
Calibration method: ..........SpanCheck™, sensor saturates at dewpoint above
+68°F (+20°C). NIST/NPL traceable multi-point factory calibration available optionally.

2.3 Sensor Installation & Sampling Techniques

Keep in mind that the moisture content at the sensor is not only due to the moisture of the gas being measured, but also due to desorption of water from tubing, trapped moisture (at the interconnection points, valves, filters and other hygro­scopic materials in the system), leaks in the system, and oth­ers. Thus the measurement may vary from the expectation, and therefore care should be taken in choosing the sampling technique utilized in the measurement. Factors such as gas pressure, flow rate, materials of construction, length and
page 4
Section 2: Sensor and Sampling Techniques
diameter of tubing, number of interconnecting fittings, dead space in tubing and manifolds; will influence the measure­ment value and response time. The high capacitance HTF™ sensors can be installed either directly in the line to be sampled (in-situ), or in a slip stream of a sample system (extractive). To assure a long and accurate performance of the sensor, it should be protected from contaminants such as liquids (water, oil etc.), and particulates. The sintered stainless steel sensor encapsulation protects from particulates larger than 100 microns, finer particulates (e.g. from degraded desic­cant or rust) should be filtered with a particulate filter with suitable capability, do not use hygroscopic filter materials.

2.3.1 In-situ Installation

In-situ installation is recommended only for measurements where the gas pressure is expected to vary little, the gas is expected to be free of contaminants, the gas temperature is wit hin the oper a ting s pec if ic at ions of the se nsor , a nd th e re is no chance of liquids coalescing. Examples of applications suited for in-situ installations are: pure gases, output of des­iccant dryers (for instrument air), glove boxes, etc. For most other applications in-situ installation should be avoided for the following reasons:
Sample conditioning is almost always necessary to
avoid exposure of the sensor to liquid water and other contaminants, such as hydrocarbons, which may dam­age the sensor or affect accuracy over time.
Variations in line pressure affect the reading of the sen-
sor because dewpoint varies with pressure.
If the gas line is under pressure, it is more likely that
water condensation occurs which may damage the sen­sor.
page 5
Under a pressurized system removal of the sensor with-
out the installation of isolation valves can be dangerous.
If in-situ installation is required, bypass mounting is prefer­able; make sure to install the sensor at the upper surface of the gas line to minimize its exposure to liquid water, should condensation occur, the XTR65W sensor is best suited for these applications. Also consider the need to isolate (depres­surize) before installing or removing the sensor.
4-20mA loop cable
Main
Gas Line
8800T
oint Transmitter
Think
Read
Safety
www.xentaur.com
Mo
In-Line Installation, Sensor
Se
Manual
Measuring at Line Pressure
Ma
NOT RECOMMENDED
4-20mA loop cable
8800T
oint Transmitter
Think
Read
Safety
www.xentaur.com
Mo Se
Bypass Installation, Sensor
Manual
Ma
Measuring at Line Pressure
Safety shut-off Valve
Safety shut-off Valve
Sample Cell
Bypass
Main
Gas Line
Control
Valve *
* maintain differential pressure to provide adequate flow through sample cell
page 6
Section 2: Sensor and Sampling Techniques

2.3.2 Extractive Installation

For extractive installations we recommend our sample sys­tem ESS, which may be equipped with a variety of features, such as: isolation valve, coalescing or particulate filter, pressure regulator , calibration sample injection or ext raction port, pressure gauge, flow meter, weatherproof enclosure. Refer to the ESS literature for more information. If the resources to make your own sample system are avail­able, the following two diagrams may be used as a guideline to configure a simple system.
4-20mA loop cable
Safety shut-off Valve
oint Transmitter
Think Safety
8800T
Read
Manual
www.xentaur.com
Exhaust
Mo Se
Ma
Regulator or
Needle Valv e
Main
Gas Line
4-20mA loop cable
Regulator or
Needle Valv e
Main
Gas Line
Sample Cell
Extractive Installation, Sensor
Measuring at Line Pressure
8800T
oint Transmitter
Mo Se
Think
Read
Manual
Safety
Ma
www.xentaur.com
Exhaust
Sample Cell
Extractive Installation, Sensor
Measuring at Ambient Pressure
page 7
8800T User’s Manual
It is generally recommended to measure at ambient pressure for the following reasons:
The readings will not be affected by variations in line
pressure.
The risk of exposing the sensor to liquid water is signif-
icantly reduced.
ppm readings are computed for a pressure of one atmo-
sphere (1 bar); and have to be corrected using software in the instrument, or a pressure nomograph, or calcula-
tor if the sensor is measuring at different pressures. If readings at line pressure are necessary, it is recommended to measure at ambient pressure and to use the instrument’s pressure compensation feature to calculate the dewpoint at line pressure. See appendix F. Please make sure that:
The sample is taken from the upper surface of the main
gas line. This avoids problems with contamination. The
sample should be taken away from pipe line walls where
flow rates may be low, and dewpoint changes may lag.
For dewpoints dryer than -40°F, use stainless steel tub-
ing only. Copper tubing is acceptable for dewpoints wet-
ter than -40°F. Do not use plastic, rubber or tygon tubing
under any circumstances, as measurements would be
incorrect and/or response time slow due to water reten-
tion inside these materials.
Try to run pipes to the sensor upwards, so that contami-
nants tend to fall back into the main line.
Keep the length of the sample line to the sensor as short
as possible.
Use small diameter pipes (1/4” or 1/8” OD).
Use sufficient flow rates (e.g. 1 l/min with 6 feet of 1/8”
piping is adequate). The flow rate will influence the sys-
tems’ response time.
page 8
Section 2: Sensor and Sampling Techniques
Do not install any devices upstream of the sensor, such
as other measuring systems, flow meters etc., which are
not absolutely necessary as these are potential leak
sources.
Installation of a coalescing and / or particulate filter
ahead of the sensor is desirable to prevent any liquid or
particulate contamination of the sensor.
If filters are used upstream of the sensor, make sure
these contain non-hygroscopic filter materials only.
If pressure regulators, shut off valves etc. are used
upstream of the sensor, make sure these do not contain
rubber or other hygroscopic materials.

2.4 Troubleshooting unexpected readings

If erroneous readings are suspected on a newly acquired instrument, compare the serial number engraved on the sen­sor sintered filter, to the one stored in the instrument mem­ory. The two should be the same; if they are not, the instrument may not be calibrated with the installed sensor. To troubleshoot other problems, identify the unexpected reading category in the following table, and consider the possible causes and appropriate diagnostic action and rem­edy.
page 9
Troubleshooting unexpected readings (table spans 2 pages)
Symptom Possible Cause
Reading is not
changing
Slow Response
Dry Reading SpanCheck™ wrongly set, or faulty sensor.
We t Read ing
Display Shows
Condensation in sample system.
1. Water vapor in the system.
2. Flow rate too low.
3. Sample pipe too large and/or too long.
4. Unsuitable sample pipe material.
5. Leaks
6. Hygroscopic materials in sample system
Leak in system or use of unsuitable pipe.
Comparison of readings with manual cooled­mirror instrument.
Prolonged exposure to wet gas.
6$7
Display Shows
1. In st ru men t F a i lu re
6+5
2.Short circuited sensor.
Display Shows
231
1. In st ru men t fai lu re.
2. Open circuit on sensor.
For non-sensor related problems (e.g. no reading on instru­ment) refer to section 3.5
page 10
Section 2: Sensor and Sampling Techniques
Troubleshooting unexpected readings (continued from previous page)
Symptom Diagnostic/Remedy
Reading is not
changing
Slow Response
Dry Reading Veri fy SpanCheck™, or return sensor for full calibration to your representative.
We t Read ing
Display Shows
6$7
Display Shows
6+5
Display Shows
Condensation will occur if the temperature of the sample system, at any point is below (colder) the dewpoint temperature of the sample gas. Once having formed, the sample reaching the sensor will have a dewpoint equal to the temperature of the condensation, regardless of the dewpoint of the sample at the sample point.
It is usually more satisfactory to bleed a sample gas at atmospheric pressure through the sensor sampling chamber, and to use 1/8” (3mm) o.d. sample pipe.
See below re: sample pipe material, also see section 2.3
Cure the leak, or replace unsuitable pipe with copper or stainless steel. Flexible con­nections should be made with PTFE pipe. NEVER use rubber or plastic pipe.
This type of indicator reads about 10°C dry at about -50°C dewpoint due to tempera­ture gradi ent s wit hin the de vi ce. Th e error in cr ease s at drier levels.
Dry the sensor, install sensor in either a known dry gas stream i.e. instrument quality air or dry nitroge n, or place sensor in a dry can or bottle of desiccan t an d seal the con­tainer from outside air (the shipping container is designed for this purpose)
Remove (unscrew) sensor, if the instrument still reads instrum ent. Howeve r , if the instrument rea ds
Remove (unscrew) sensor and note that the meter reading returns to sensor, or apply approximately 20V DC, between the center pin & the filter body of the sensor, MOMEN T AR ILY with the s ensor in a known dry condition. Pola rit y is not important, but the contact MUST be very brief or the sensor may be damaged.
Remove (unscrew) the sensor. Short the center contact of the 8800T sensor connection to the case, if the ins trumen t read s
the instrument for service.
231
the problem is in the sensor, otherwise return
6+5
the problem is with the
6+5
then check possible cause #2
. Use a new
231
231
Check sensor connection or replace sensor.
page 11
page 12
Section 3: Instrument
3.1 Precautions using the 8800T
The 8800T uses state-of-the-art microelectronics to provide a miniature full functioning instrument. The user should consider the following precautions when using any sensitive electronic device.
Do not install the unit near heat sources such as radia-
tors or air ducts.
Do not install the unit in places subject to extreme
mechanical vibration or shock. If this is not avoidable,
use resilient mounting. If in doubt, call your representa-
tive.
Observe the appropriate electrical safety codes and reg-
ulations
If weather proofing is required consult your representa-
tive for an optional cover.
3.1.1 Electromagnetic Compatibility Considerations
The 8800T has been designed and verified by testing to meet the requirements of the EC Council EMC Directive 89/336/EEC, for Industrial, Scientific & Medical equip­ment. The sensor ground (8800T housing) is isolated from the 4-20mA loop, however they are also shunted with a 2000pF capacitor and a 33V Transient Voltage Suppressor; this prevents electrostatic buildup, noise pick-up, and in conjunction with the internal fuse protects the instrument from over-voltage inputs. Please consider the following electromagnetic interference issues during installation:
In order to provide an acceptable noise environment for
the 8800T or any other digital equipment in the proxim-
ity of inductive loads, it is recommended that there be
varistors placed across the inductors to keep down the
high voltage spikes during transitions.
page 13
Any circuitry which is activated by relay contacts
should account for the contact bounce, one simple
debouncing method is placing a capacitor across the
relay contacts.
AC power wiring should be routed as far away from the
8800T and its wiring as practical.

3.2 Instrument Technical Specifications

Enclosure:.........................Stainless Steel, optional weather proof cover is avail-
able.
Dimensions & Weight:......~2”Dia. x ~4.75” long including sensor & connector
(see appendix C) 0.5lbs.
Pressure operating range:..Standard:.....500 PSI (34 bar).
Optional:.....5,000 PSI (340 bar).
Mechanical connections:...14mm x 1.25mm threads, and ¾”-16 threads.
Electrical connections:......2.1mm x 5.5mm (12mm minimum barrel length)
power jack, with retainer.
Cable:................................Two conductor cable, 12’ (~4m) provided.
Power Requirements:........10 to 33 VDC polarity independent, the instrument
draws 4-20mA depending on measured dewpoint.
Input resolution:................0.1°C dewpoint.
Indicators:.........................3.5 digit LCD with custom legends.
Engineering units:. ............°C,°F, PPM, LBS H
Controls:............................3 push buttons, user’s selections are stored in
EEPROM.
Outputs:.............................Analog and digital outputs are available from the
8800T.
A. 4-20mA drawn by the instrument from the power
supply. The 4-20mA is linear to the selected engi-
neering units, the range is programmable. Output
resolution is 0.1°C dewpoint.
B. The instrument can supply digital output by
modulating the 4-20mA loop line. The timing and
format of the data conforms to RS-232, however to
interface to a PC or other RS-232 device an optional
adapter is required.
In the digital mode the 8800T can be remotely
operated and the dewpoint as well as temperature
O/mm scf,gm H2O/M3.
2
page 14
Section 3: Instrument
(and pressure if installed) can be read.
In the digital mode multiple units can operate on
the same loop cable as a multi-channel instrument.
Alarms:..............................The 4-20mA signal or the digital output may be used
by an external device to operate relays.
Isolation:...........................Sensor and case are isolated from the current loop
but are shunted with a 33V transorb and 2000pF
capacitor.

3.3 Installation

3.3.1 Mechanical Installation

The 8800T has two thread sizes for mounting to the sample cavity (consult the sensor installation section) where the dewpoint will be measured. Various adapters are available for direct connection into existing system openings. Ask your representative for a Sample Cell, if you do not have the ability to provide an appropriate sample cavity mounting. If the ¾”x16 thread is used then the 8800T will seal against the wall of the sample cell with the provided Viton A O-ring. If the 14mm x 1.25 spark plug thread is used then an addi­tional Viton gasket must be installed to provide the proper seal. Ask your representative for this gasket, it is available free of charge. The 8800T is light enough such that either thread will mechanically support the whole instrument. To prevent any leaks, tighten the 8800T into the sample cavity, with a
metal-to-metal contact. Since the tightened positioning of the 8800T is unpredict­able, its front panel may be rotated (360°) to allow easy reading of the display. Press down gently on the spring loaded face plate and turn it to the desired position. The sensor can be removed from the 8800T by unscrewing it. Make sure that the sensor is securely fastened to the
” wrench, 1/8 turn past finger-tight to assure
17/
8
page 15
8800T (the tension washer should be compressed), so that it
does not come loose during use.

3.3.2 Electrical Installation

The 8800T will operate properly with 10VDC to 33VDC at its input, if this voltage is exceeded the internal fuse may blow. When selecting the power supply voltage do not neglect the drop across any current measurement resistor and wiring in the loop. The connector is a standard 2.1mm x
5.5mm (12mm minimum barrel length) DC power jack, an internal diode bridge accommodates either polarity. The 8800T may be powered using an off-the-shelf wall trans­former with 2.1mm connector, or from a power supply using the provided connectorized cable. The provided cable is terminated with a connector which has a retainer, the retainer may be used to secure the connector to the 8800T thus avoiding accidental disconnection. The 8800T will draw 4mA to 20mA from the power supply depending on the dewpoint being measured. The dewpoints corresponding to 4mA and to 20mA are user selectable, in between the current will vary linearly to the selected engi­neering units, see appendix G. Various strategies for interfacing with the 8800T are shown in Appendix D. Please observe good electrical safety and grounding prac­tices when connecting any electrical equipment; connecting one end (e.g. negative) of the power supply to earth ground is advisable. After the installation is complete, proper detection by the user’s equipment of the 4-20mA out put, may be tested using the Analog Output test feature of the 8800T, see section
3.4.5.6.
page 16
Section 3: Instrument

3.4 Operating the Instrument

3.4.1 Starting up

The instrument is ready for use as soon as the power cable is installed. When power is applied the instrument will initial­ize its program and for a moment display enter the Operating State. If the MODE button is held pressed while the instrument is performing its power-up ini­tialization, it will enter the Set-Up State, which allows the user to select operating variables of the instrument. To accommodate a variety of installation possibilities, the 8800T front panel may be rotated (360°) by gently pressing down on the spring loaded face plate and turning it to the desired position.

3.4.2 Display Conventions

1. To display characters with the 7 segment numeric dis-
play, the following pseudo-alphanumerics are used:
Numbers:
0 123456789
 
, then it will
;(1
Letters:
ABCDEFGH IJ LNOPQRSTUXYZ
$%&'( ) *+, -/12345678;<=
Symbols:
?-.
"
2. The instrument will indicate whether a particular mode
lets you change a parameter by showing the word “SET” in the upper left corner of the display. Be careful
page 17
+20°C
-100°C
+68°F
-148°F
23,612 ppmV
0.013,849 ppmV
2,000,000
0.000,000,1
1104.2
lbs H
2
O/mmSCF
0.000,921,15
lbs H
2
O/mmSCF
17.688
grams H
2
O/meters
3
0.000,014,75
grams H
2
O/meters
3
not to change any parameter inadvertently.
3. °C°F appear simultaneously, to indicate the sensors’ attenuation in decibels.
4. Values larger than ±1999 or smaller than ±0.01 are dis-
±3
played in powers of 10
-3
“10
” will appear above and to the right of the dis-
. As required, either a “10 3” or
played value, the value will be rounded off to 3½ digits. The display will show
(out of ranGe), if the num-
51*
ber to be displayed is larger than 1,999,000. Twelve examples follow; for each the number and units desired to be displayed are shown in
italics,
depicted immedi-
ately below them is the resultant 3½ digit LCD display:
3
DEWPOINT
x10
 ˚C 
DEWPOINT
 ˚F 
DEWPOINT
PPM  
DEWPOINT
  LBS
DEWPOINT G/M
DEWPOINT
3
-3
x10
PPM  
DEWPOINT
x10
  LBS
-3
page 18
DEWPOINT
DEWPOINT G/M
 ˚C 
x10
3
 ˚F
DEWPOINT
-3
Section 3: Instrument

3.4.3 Push Buttons

Three push buttons are located on the front panel.
Teledyne 8800T
SET % HILO TEMP DEWPOINT PSI G/M³ LBS
MODE button UP button DOWN button
x10¯³ PPM ˚C˚F
Each button is marked with an icon engraved above it. In general the MODE button navigates through the different user options “Modes”; the UP and DOWN buttons modify the units, values or choices in the selected mode. Refer to the flow diagrams in Appendix A for detailed overview of button functionality. A button may be held down for a pro­longed time, for accelerated incrementing or decrementing of numeric values.

3.4.4 Operating State

Upon power up, the unit performs certain initialization tests (see table in section 3.5), and enters the ‘Operating State’, in the Viewing Dewpoint mode. Depressing the ‘MODE’ but­ton will change modes (see appendix A) in the following order: nViewing Dewpoint ‹ oViewing Temperature ‹
Start Calibration ‹ qViewing Serial Number ‹ (back to)
p
Viewing Dewpoint. The unit will return to Viewing Dew-
n
point mode if no buttons are pressed for 30 seconds, unless it is performing a calibration.
page 19

3.4.4.1 Viewing Dewpoint Mode

In this mode the user can view the dewpoint, this is indi­cated by the presence of the ‘DEWPOINT’ legend on the lower left of the display. The available engineering units in which to view the moisture content are °C, °F, PPM, LBS
3
and G/M
; the UP and DOWN buttons scroll back and forth through these units in respective order. The °C and °F are dewpoint readings. The PPM is parts per million by volume computed at the sensor pressure (more about pressure in
3
appendix F). The LBS and G/M
are pounds of water per million standard cubic feet and grams of water per standard cubic meters, both in Natural Gas, they are computed according to data derived by IGT Research Bulletin 8, tak­ing into account sensor pressure. Note that the analog output is linear to the selected engi­neering units, therefore be mindful that while scrolling through various units the analog output may change even though the measured dewpoint is stable. The PSI legend flashes at the bottom of the display, when there is pressure correction in the computation of the dis­played values, refer to appendix F.

3.4.4.2 Viewing Temperature at the Sensor

In this mode the user can view the thermodynamic tempera­ture at the dewpoint sensor inside the sampling chamber, this is indicated by the flashing ‘TEMP’ legend on the lower left of the display. The available units in which to view the temperature are °C and °F. The UP and DOWN buttons switch back and forth between these units. Pressing the MODE button changes to the Start Calibration Mode.
page 20
Section 3: Instrument

3.4.4.3 Start Calibration (SpanCheck™) Mode

The instrument is calibrated at the factory with the sensor it is shipped with and does not need to be re-calibrated prior to installation.
Instrument calibration is recommended in approximately 12 month intervals, and the XTR65W sensor should be re-cali­brated after prolonged exposure to liquid water. Simply fol­low steps 1 - 7 of the procedure below, removing the sensor from the sample gas stream. Make sure the gas stream is depressurized before removing the sensor to avoid injury. It is recommended to keep the sensor exposure to room air as short as possible to avoid super saturation of the sensor. While super saturation is not damaging to the sensor, it will prolong the initial dry-down time after you install the sensor in the sample stream. Therefore, remove the sensor from the packaging container only after you are ready to proceed with the calibration procedure and install the sensor in the sample stream immediately after the calibration procedure is completed. If you are not ready to use the sensor right away after calibration, put the sensor back in the shipping container for dry storage. The instrument must be calibrated with the sensor it will be used with. The calibration procedure takes advantage of the designed saturation level (DSL) feature of the sensor and is executed by the instrument computer, by performing the fol­lowing steps:
1. If the instrument is locked and a calibration is
attempted; it will display
and will not perform the
/2&
calibration. To unlock the instrument consult section
3.4.5.9.
page 21
2. Push the MODE key a few times until the display shows
.
&$/
3. Press the UP button. The display will show
, prompt-
&1)
ing you to confirm that you want to start the calibration procedure. You can abort the calibration procedure by pressing the MODE key.
4. Remove the sensor from its packaging container so that
the porous metal filter is visible and the sensor is exposed to in-hand micro-climate. Close the packaging container as soon as you have removed the sensor to avoid degradation of the desiccant inside of the con­tainer. You may want to re-use the container at a later date.
5. Push the UP button again to confirm that you want to
start the calibration procedure. The display will flash
$&
for 60 seconds, while the sensor is saturating. Make sure you keep the sensor exposed to in-hand micro-climate until the display shows
(1'
.
6. After 60 seconds, the display will flash the selected sen-
sor type (see “
Selecting a sensor type 3.4.5.2
”) and then the instrument will calculate the slope and offset of the sensor curve while displaying the calculations and then will display
for a few seconds, after which the
(1'
instrument will automatically go into measuring mode. The sensor calibration is completed. (The display may flash
, indicating that the sensor has super-saturated.
6$7
As soon as the sensor is exposed to an atmosphere with a dewpoint lower than the saturation dewpoint, the dis­play will indicate the dewpoint measured by the sensor.) The display may show alternating
6(1
and
/2
as an in dica tion t h at the mea sure d capa cit ance is too low to be from a saturated sensor, in this case make sure that the sensor is fully engaged into the 8800T, and repeat the
page 22
Section 3: Instrument
calibration procedure.
7. Install the sensor in the sample cell or adapter fitting or put it back into the packaging container for later use.
The instrument will retain the calibration even if the power is turned off.
Under certain conditions, an over (super) saturated sensor may need to be completely dried out before a calibration is performed. Symptoms of these conditions are a sensor that will not go through the SpanCheck™ function to the
(1'
display, or a sensor that will not dry down after calibration. To dry, install sensor in either a known dry gas stream i.e. instrument quality air or dry nitrogen, or place sensor in a dry can or bottle of desiccant and seal the container from outside air (the shipping container is designed for this pur­pose). After a minimum dry out period of 24 hours, proceed with the calibration procedure of your choice. SpanCheck™ calibrates sensors by using micro climates. Please contact your representative if you have any questions about how to create such micro climates. A perfectly acceptable and accurate for calibration micro­climate may be created for the XTR100 and XTR65 sensors by cupping the sensor in the palm of one’s hand during the SpanCheck™ procedure. Remember that the micro-climate does not have to be accurate, it just has to be higher than the designed saturation level. The dewpoint of the micro-cli­mate within the fingers is usually higher than the dewpoint of the room air and probably well above the +20°C upper range of the sensor. On a dry day one may need to exhale in the hand before cupping the sensor to guarantee a high dew­point. The sensor should be warmed up close to body tem­perature before performing this procedure. Care must be
page 23
12
345
XSN
taken that the sensor temperature is not below the tempera­ture of the hand, as in such case condensation could occur and super saturate the sensor. For the same reason, avoid exhaling directly onto the sensor.

3.4.4.4 Viewing Serial Number Mode

In this mode the user can view the serial number of the sen­sor calibrated and shipped with the instrument. This number is also engraved on the stainless steel sintered filter of the sensor element. The display shows the serial number by alternately displaying
and the number. If the number is
;61
larger than 1999 then it is displayed in 2 parts, first part is
3
the thousands signified by the x10
legend in the upper right corner of the display and the second part is the units. For example serial number 12345 will be shown as:
3
x10
Pressing the UP and DOWN buttons simultaneously resets the instrument, this is useful for restarting in the set-up state (by holding the MODE button), in installations where power can not be turned on and off easily.
Pressing the MODE button changes the unit to the Viewing Dewpoint Mode.

3.4.5 SetUp State

To enter the Setup State power-up the unit (or reset it from the serial number Mode), and hold the MODE key pressed while the unit is initializing. Refer to Appendix B for a flow
page 24
Section 3: Instrument
diagram of the SetUp State. The setup state provides nine capabilities, each one con­trolled from its own mode. Depressing the MODE button navigates through the modes in the following order: nSe­lect Alternate Display Units ‹ oSelect Sensor Type ‹ pSet Measured Attenuation of Sensor at a Low Dewpoint ‹ qSet the Low Dewpoint at which the Attenuation was measured ‹
Set the Calibration Mid-range Adjust ‹ sPerform a Test
r
and Verification of the Analog Output ‹ tSet dewpoint cor­responding to 4mA of Analog Output ‹ uSet dewpoint cor­responding to 20mA of Analog Output ‹ vLock/Unlock the instrument ‹ (back to) nSelect Alternate Display Units. These nine functions are explained in their respective sec­tions which follow.
unlocked instrument
Changes can be made only to an
, the changes are stored (when the MODE button is pressed) and retained even if the unit is turned off.
1.
Display of alternate units
: In this mode, a second engi­neering unit can be chosen to be displayed alternating with the engineering unit selected in the operating state, for example, a dewpoint can be displayed alternating with the sample temperature, or dewpoint can be alter­nately shown in °C and in PPM.
2.
Selecting the sensor type
: In this mode the user can select the software matching the type of sensor installed in the instrument;
3&4
Adjusting low end sensor attenuation and dewpoint
: XTR-100 (-100°C to +20°C);
;7
: XTR-65 (-65°C to +20°C).
;7
These modes are used to enter a data pair representing a low dewpoint and the sensor attenuation measured at this low dewpoint. (see Sensor Theory of Operation appendix E). This data pair should not be modified unless the sensor is replaced. The sensor shipping con-
:
page 25
tainer is labeled with the proper values.
5.
Calibration Adjustment
: In this mode the user can enter a sensor specific Adjustment Value to improve the instruments’ accuracy in the range of -50°C to -10°C, this adjustment has very little or no effect outside this range. The Calibration Adjustment Value is printed on the shipping container supplied with the sensor, or may be derived if the sensor can be exposed to a known ref­erence.
Deriving an Adjustment Value
: If the adjustment value for your sensor is not known, contact your repre­sentative and follow the instructions in paragraph B on the next page. If the value is not available, then it may be derived using a manual method, provided that a moisture calibration lab is available. The calibration must be done in the range of -25°C to
-20°C (attempting to calibrate outside this range may cause inaccuracies), the dewpoint must be kept stable during calibration and it must be measured accurately by a reference instrument such as a chilled mirror. One must also note the state of pressure compensation and turn it on or off as applicable to the reference. The necessary steps are described below (also refer to the Set-Up State flow diagram in Appendix B): a. Make sure that the low attenuation and low dewpoint are correctly entered for this sensor. b. SpanCheck™ the sensor. c. Dry-down the sensor for at least 12 hrs. d. Expose the sensor to the known dewpoint (-25°C to
-20°C), for a sufficient time (at least 30 min.) e. Go to t he
&$/$'-
mode and press the UP button, the display will show the current value, in the range of -1.99 to +1.99 followed by blinking horizontal lines.
page 26
Section 3: Instrument
f. Observe the display, if the buttons are not being pressed; the display will every few seconds show for a short duration the dewpoint as being computed at the moment (utilizing the current cal-adjust value in the computation). g. Use the UP or DOWN buttons to modify the cal­adjust value and observe the alternately displayed dew­point. Perform this adjustment until the dewpoint matches the value shown by the reference instrument. h. Press the MODE button to go to the next mode, this will save the new cal-adjust value, and it will be retained even if power is turned off.
Entering a known Adjustment Value
: If the sensor is being replaced an adjustment value should be supplied with the sensor, to enter it into the instrument: a. Go to t he
&$/$'-
mode and press the UP button, the display will show the current value, in the range of -1.99 to +1.99 followed by blinking horizontal lines, ignore the alternately displayed dewpoint. b. Use the UP or DOWN buttons to modify the cal­adjust value as necessary. c. Press the MODE button to go to the next mode, this will save the new cal-adjust value, and it will be retained even if power is turned off.
2), 3), 4) and 5) are set at the factory and need only be modified when a sensor is changed.
6.
Testing the analog output
: By pushing the UP or DOWN buttons, the user forces the analog output to its low and high values, respectively. This facilitates the hook-up and testing of the user’s monitoring equipment.
7&8
Output range setting
: These modes are used to set
dewpoints corresponding to the low and/or high end of
page 27
the current loop output, refer to appendix G.
9.
Lock/Unlock the instrument
: This mode is used to block access to parameter settings, protecting the instru­ment from unauthorized or inadvertent changes of parameters. Attempting to change settings while instru­ment is locked will display
/2&
To unlock the instrument press the UP button, to lock the instrument press the DOWN button. Pressing the MODE button will navigate back to the Select Alternate Display Units mode. Pressing and holding either the UP or DOWN button, will save the lock or unlock setting depending on the button, and reset the instrument. This feature is useful if it is difficult to power off the instrument. Note: It is imperative that one returns the instrument to the locked mode to avoid unauthorized changes.

3.5 Troubleshooting the Instrument

This instrument performs diagnostic tests on power up as well as once every two minutes. The table that follows, depicts all possible error/unexpected messages that may be displayed on the instrument. For each message the table has explanations for the reason, and if necessary a suggested action to remedy it.
page 28
Section 3: Instrument
DISPLAY EXPLANATION REQUIRED ACTION
LO
(55
HI
(55
%
(55
&6)
(55$'
(555()
/2%$7
LO
51*
TEMP
HI
51*
TEMP
231
6$7
6+5

'%
6(1/2
(55((3
/2&
51*
°C(


w/flashing
DEWPOINT
flashing
PSI
when
dew
;(1
Legend:
PROM check sum failed.
RAM write/read te st failed. cycle power
Unidentified power-up failure. if problem persists,
EEPROM Check Sum Failed. return to your representative
power on tests
converter failure.
A/D
erence voltage for A/D out of spec.
Ref
Low supply
Instrument low
has been excee ded . make sure that th e unit is at a
Instrument high
has been excee ded .
Sensor circuit is
Sensor is
Sensor circuit is
Trying to calculate dewpoint for
voltage.
erature
temp
eratur e
temp
open
urated. sensor troubleshooting
sat
shor
undefined sensor.
e
rang
system tests once per 2 min.
e
rang
.
ted. section
dewpoint display tests
Trying to calibrate an undefined
fo r s erv ice.
check input voltage 10-33VDC
temperature of -10°C to +70°C.
see table in
select sensor and
SpanCheck.
see Calibration instructions.
sensor.
sor reading is ‘
Sen
rated sensor, for SpanCheck.
ROM write cycle not completed.
EEP
Attempting to modify a
number can not be displayed in 3.5 digits.
°F)
An XTR-65 sensor is measuring less
w to be a satu-
too’ lo
Calibration. tests
if this persists, return for service.
ked unit. unlock unit, see set-up mode.
loc
than -70°C or -94°F dewpoint.
miscellaneous
Pressure correction is enabled.
Turn on message,
denotes alternately flashing messages.
Xen
taur (
Greek Ξ
=X)
page 29
The instrument contains a fuse which may open if the instrument is subjected to voltages exceeding the specifica­tions. Under these circumstances the unit display will be blank and the instrument will not draw any current when powered up. The fuse is not user replaceable, therefore the 8800T must be sent to your representative for servicing.
page 30
Appendices
SEN/2LO
AC/_AC
CNF/AC?
CAN
/
Appendix A: Operating State User Interface flowchart
Power ON initialization and sign on message
30 seconds anywhere other than calibration without a key press
Viewing
Dewpoint
Viewing
Sensor
Temperature
Calibration
Menu
Viewing
Serial
Number
Teledyne 8800T
DEWPOINT
PSI
select
units
Teledyne 8800T
TEMP
select
units
Teledyne 8800T
SET
Teledyne 8800T
a flashing PSI legend means pressure correction is being applied, see appendix F
˚C
˚F
Press together to RESET the Instrument. Then quickly press and hold MODE to go to Set-Up State.
1sec.
Confirm
SpanCheck™
1 minute
Sensor
Reading
OK
Teledyne 8800T
SET
numbers calculating
calibration
Teledyne 8800T
Teledyne 8800T
SET
Teledyne 8800T
SET
Too Low
few
seconds
Teledyne 8800T
SET
Teledyne 8800T
SET
1 sec.
Cancel Calibration (calibration values are not changed)
2 sec.
Values Stored
New Calibration
LEGEND: 1. Arrows leading out from below a button depict the flow as a result of  pressing the button.  2. Arrows leading out of other locations depict flow as a result of  completing a function or a time-out.  3. A slash '
' in the display area, is used to depict two alternately
shown (flashing back and forth) messages.
page 31
LOC
ALT/DSP
SEN/XT1
DB/
CAL/ADJ
A.0./TST LO/BAT
A.O./RNG/
XT1 XT6 DB

Appendix B: Set-Up State User Interface flowchart

CAUTION! These parameters are set properly at the factory, to correspond to the shipped sensor. Do not modify them
unless you are replacing the sensor.
appears when changing
settings of a locked instrument
XTR-100 XTR-65 attenuation
+/- 0.01 units
If no buttons pressed
for 3 seconds the dew
is shown for a moment.
SET
Power-Up with
MODE key pressed
blank ˚F ˚C ppm LBS G/M3 Temp˚F Temp˚C%PSI ˚C˚F
Choose alternate units to show when displaying dewpoint
SET
Factory default is blank (there is no alternate display).
Choose Sensor type
SET
page 32
Set measured attenuation at a known low dewpoint
˚C˚F
SET
LO
+/- 0.01dB
Set low dewpoint for the above set attenuation
˚C
SET
LO
DEWPOINT
+/- 0.1˚C
Calibration Midrange Adjust
SET
Appendices
/
A.0./TST LO/BAT
A.O./RNG/
A.O./RNG/
UN/LOC
User Interface Flow Diagram - Set-Up State (continued from previous page
Ensure minimum voltage at LPDT
to be 10VDC, when drawing 20mA
Too
Low
Test
Voltage
Current Loop
OK
Test Analog Output
4mA
20mA
HILO
+/- 0.1˚C
Set dewpoint corresponding to low end of analog output
Factory default is -100˚C
˚C
SET
LO
SET
+/- 0.1˚C
Lock or unlock the instrument
Set dewpoint corresponding to high end of analog output
Factory default is 20˚C
˚C
When locked instrument parameters can not be changed.
Prolonged press of either button will reset the unit
Factory default is locked.
Lock
Unlock
HI
SET
)
' in the display area, is used to depict two alternately
page 33
LEGEND: 1. Arrows leading out from below a button depict the flow as a result of
pressing the button.
2. A slash '
shown (flashing back and forth) messages.
m
Appendix C: 8800T Mechanical Drawing
˚
0
6
3
y
l
e
e
r
f
s
e
t
a
t
o
r
l
e
n
a
p
t
n
o
r
f
DC Power Jack
oint Transmitter
Think Safety
Power Jack screw-on retainer
1.750"
x10¯³
PPM
˚C˚F
SET %
HILO
TEMP
DEWPOINT PSI G/M³ LBS
Teledyne 8800T
Read
Manual
www.teledyne-ai.com
Mo Se
Ma
8
/
7
flats for 1 " wrench
5/16"-24
tension washer
3/4"-16
sintered stainless steel,
unscrew to remove.
Removable Sensor encapsulated
in 100
Viton A O-ring
14mm x 12.5mm
2.230" 1.850" 0.600"
page 34
Appendix D: 8800T Electrical Connections
W
Methods of Using and Interfacing the 8800T
Operation with Wall Transformer
Dewpoint viewed on Instrument Display
10-33VDC
120V
or
220V
Wall Transformer
Operation with DC Power Supply
Dewpoint viewed on Instrument Display
and available as 4-20mA output
maintain  10-33VDC here
VDC
4-20mA
oint Transmitter
Think
Read
Safety
www.xentaur.com
oint Transmitter
Think
Read
Safety
www.xentaur.com
Mo Se
Manual
Ma
Mo Se
Manual
Ma
Operation with Relay Option Board
Dewpoint viewed on Instrument Display
and available as 4-20mA output
Relay Option Board
Provides two
independent Relays 
with programmable
trip point and 
hysteresis
DC or AC supply
Operation with Remote Option Board
Dewpoint viewed on Instrument Display
and available on remote
Display
Buttons
Relays
Voltage or Current
Output
RS-232 Output
Power Supply
4-20mA
Appendices
oint Transmitter
Mo Se
Think
Read
Manual
Safety
Ma
www.xentaur.com
oint Transmitter
Mo Se
Think
Read
Manual
Safety
Ma
www.xentaur.com
Operation with DC Power Supply
Dewpoint viewed on Instrument Display
and available as Voltage output
12-33VDC
0.4 - 2V output
100
oint Transmitter
Think
Read
Safety
www.xentaur.com
85 - 265VAC
Operation with Multichannel Option Board
Dewpoint viewed on each Instrument Display
and available on remote interface
Multichannel
Option Board
Mo Se
Manual
Ma
multiple units connected on the same two wires
oint Transmitter
oint Transmitter
Think
Think
Safety
Safety
www.xentaur.com
Read
Read
www.xentaur.com
Manual
Manual
Mo
Se
Ma
Ma
page 35
RS-232C 
to Computer
DC or AC supply
Mo Se
oint Transmitter
Think
Safety
Read
www.xentaur.com
Manual
Mo
Se
Ma

Appendix E: Sensor/SpanCheck™ Theory of Operation

The Sensor is constructed as a capacitor whose dielectric consists of porous Aluminum Oxide as well as the gas that has entered in the pores of the Aluminum Oxide. The plates (electrodes) of this capacitor are an aluminum substrate and a porous gold layer deposited on top of the Aluminum Oxide, the porous gold electrode allows transfer of gases into or out of the Aluminum Oxide pores. The capacitance due to the Aluminum Oxide is always con­stant, while the capacitance due to the gas varies according to the gas content and pressure. Since the dielectric constant of water is orders of magnitude larger than that of any gases being measured, the quantity of water vapor present in the pores changes the capacitance of the sensor to a much greater extent than any other system variable. For the same reason (the extremely large dielectric constant of the water molecule), any capacitance variations arising from the Alu­minum Oxide, such as changes due to temperature, are insignificant in relation to the capacitance due to the water content. Thus the sensor capacitance varies greatly in pro­portion to the water content in the surrounding gas. The sen­sor is designed such that the relationship of the quantity of water and the resultant capacitance has an "S" shaped curve. At the extremely dry end the curve is asymptotic to the capacitance due to the Aluminum Oxide and the gas, while at the very wet end the curve is asymptotic to the capaci­tance due to water molecules packed extremely tightly (sat­urated) in the pores of the Aluminum Oxide. This upper end is an excellent indication of the total pore volume; while the lower end of the curve is an indication of the distance between, and area of the capacitor plates (electrodes), this is the intrinsic capacitance of the sensor. Therefore knowing these two points, a sensor can be calibrated with compensa-
page 36
Appendices
tion for small manufacturing deviations. As the sensor is used in real world applications, it is exposed to various elements and stress; like any other sen­sor it may over time drift from calibration. However one must note that the intrinsic capacitance will not change with use because it is based simply on the dielectric constant of Al
and the distance between, & area of the plates (elec-
2O3
trodes), thus the original factory measured values will always be valid. What may change is the pore volume, due to clogging with contaminants, residual oxidation, metal migration, etc. Thus to re-calibrate the sensor the instrument needs only to "know" the capacitance at the wet end of the curve. As discussed previously the capacitance at the wet end approaches asymptotically a saturated pore capacitance, therefore if the sensor is saturated (the exact water content will not be important since the curve is asymptotic), then the instrument can measure the capacitance and re-calibrate the sensor. This is refered to as SpanCheck™. The advantages of this unique capability of Teledyne sensors and instruments are obvious in time and cost savings for re-calibration, as well as ease of sensor replacement.
Teledyne HTF™ Sensor: Dewpoint vs. Capacitance Response Curve
200
Saturated Capacitance Asymtote of New Sensor Saturated Capacitance Asymtote of Aged Sensor
160
rCurve
nsor Curve
120
Aged Se
New Senso
80
Capacitance (nF)
40
0
Dry Capacitance Asymtote of New & Aged Sensor
-80
-60
-40
Dewpoint Reference ˚C
-20
0
20
page 37
page 38

Appendix F: Pressure Correction

Appendices
Sensor Pressure
is used in the context that this is the pres­sure inside the sampling chamber when performing the measurement, i.e. it is the operating pressure of the sensor. Gas Pressure
is use d in th e c ontex t that this is the pr essur e at which the dewpoint is to be calculated. Pressure Correction
is used in the context that the values displayed signify the moisture content at some pressure (we refer to this as the ‘Gas Pressure’) different from the pres-
3
sure at the sensor. Note that PPM, LBS and G/M
readings are by definition unaffected by pressure correction because only the pressure at the sensor affects their value. While °C and °F are affected by pressure correction by reporting what the dewpoint would be at the Gas Pressure when the dew­point is what is measured at the pressure at the sensor. How­ever, this also implies that whether pressure correction is
3
applied or not the PPM, LBS and G/M
readings are affected by the setting of the sensor pressure. The factory default settings are: 14.7psi for both sensor and gas pressure and pressure correction disabled. When Pressure correction is disabled all dewpoints are computed by assuming that both Sensor and Gas Pressures are 14.7psi. As all Teledyne dewpoint meters, the 8800T is equipped with the necessary software to perform pressure correction. However, for the lack of space, the 8800T does not have a Pressure Correct button. Therefore the user can not set or enable Pressure Correction, this must be done at the factory . Consult with your representative.
page 39

Appendix G: Current vs. Dewpoint

The current being pulled by the 8800T, varies with the dew­point being measured by the 8800T. To use the current to calculate the value of the dewpoint measurement, one must know the settings of the low and high ends of the analog output range, then:
4
()
I
----------------------------------------L+=
D
where:I = current drawn by 8800T loop in mA.
= value of High end of Analog Output range
H
converted to sele cted engineering units
= value of Low end of Analog Output range
L
converted to sele cted engineering units
= dewpoint measured by instrument in selected
D
engineering units.
()×
HL
16
Consult section 3.4.5.7&8 and/or appendix B, to check and set the Analog Output low and high ranges; the factory default settings are -100°C and +20°C respectively. For example a unit with factory default settings, drawing 12mA is computed to be measuring a dewpoint of -40°C:
12 4
()
----------------------------------------------------------- -
20 100
()
()×
16
100
+
()
40
=
Note that the computation is such that the current is linear to the selected engineering units. Hence, selecting ppmV or LBS or G/M3 units, will cause the analog output to be lin­early proportional to those units (approximately logarithmi­cally proportional to dewpoint), refer to the graph that follows. Naturally selecting °C or °F will cause the analog output to be linearly proportional to dewpoint. When monitoring in ppmV or LBS or G/M3, the analog output low & high ranges may have to be adjusted to pro-
page 40
Appendices
vide a useful output. Consider an example where the area of interest to be monitored is 10 to 100 ppmV, and the analog output is set up with the factory defaults of -100°C to +20°C (which is 0.014 to 23612 ppmV); then the current loop out­put will vary only from ~4.1 to ~4.2 mA in the area of inter­est (consult with the graph on the following page). In most instances this would be an unacceptable output for proper monitoring of the measurement. In this example the user should adjust the analog output low & high ranges such that the output range is better suited to the measurement of 10 to 100 ppmV. It may be useful to select the low and high ranges to be 5 and 150 ppmV respectively, thus out of range conditions will be detected properly. Then the low range will be set to 5ppmV which is -65.5°C dewpoint, and the high range will be set to 150ppmV which is -38.5°C dew­point. Now the current loop output will be 4.55 to 14.48 mA in the range of 10 to 100 ppmV, the ~10mA variation is more than sufficient for a good measurement by the user’s equipment. One may carry out similar calculations for LBS or G/M3 and choose the appropriate settings. While making these computations it may be useful to obtain a copy of Teledyne’s dewpoint calculator, this is a Microsoft Win­dows™ program which simplifies the process of converting dewpoint measurement units. It is available at www.Tel- edyne-ai.com. If you are not certain how to carry out such calcu­lations send, by e-mail to Teledyne@Teledyne-ai.com or by fax to (626) 961-2538, your system specifics, and some one will get back to you with appropriate analog output settings. In general, if the dewpoint is monitored in °C or °F, there is no need to change the factory default -100°C to +20°C set­tings, because the 4-20mA provides sufficient resolution to measure the output better than the specified accuracy of the sensor.
page 41
Relationship of 4-20mA output and instrument reading in ppmV or LBS
289.1 LBS = 0˚C = 6063.8 ppmV
49.3 LBS = -20˚C = 1019.3 ppmV
Low AO range = -100˚C High AO range = +20˚C
Instrument reading in ppmV
0.56 LBS = -60˚C = 10.66 ppmV
10 100 1000 10000
8mA
12mA
20mA
4/20 mA output
1.97 LBS = -50˚C = 38.83 ppmV
High AO range = -38.5˚C
5mA
4.5mA
page 42
Instrument reading in LBS of H2O / million cubic feet
Low AO range = -65.5˚C
0.1 1 10 100 1000
4.1mA
Appendices
Appendix H: Uncertainty in LBS & ppmV calculations
48 44 38
34 29
24 19
15
900 800
700 600
500 400
300
1000
10
5
2.5
O / mmscf
2
1
Calculated LBS H
0.5
0.25
0.1
Uncertainty of LBS & ppmV calculations due to +/-3˚C measurment accuracy
0.05
-65
-60
-55
-50
-45
-40
-35
-30
-25
-20
Dewpoint +3˚C
Dewpoint -3˚C
-75
-70
200
100
90 80 70
60 50
40
30
20
10
9 8 7
6 5
4
3
2
1
0.9
0.8
0.7
0.6
0.5
-80
Measured Dewpoint ˚C
Calculated ppmV
page 43

Appendix I: Sensor Response Time

100 95
90 85
80
after 48hrs of equilibrium
(dp)
Sensors to a Step-change (DOWN)
3
O
2
at time 0, to -62.2 ˚C
(dp)
from -39.4 ˚C
Response Time of TAI Al
0
5
1015202530354045505560657075808590
65% of step change
% of step change
75 70 65 60 55 50 45
Time (minutes)
40 35 30 25 20 15 10
5
95% of step change
0
95
100
105
page 44
Teledyne, 16830 Chestnut St., City of Industry, CA 91748 Tel: (626) 934-1500
Loading...