Teledyne 7600 User Manual

Page 1
OPERATING INSTRUCTIONS FOR
Model 7600
NDIR Infrared Gas Analyzer
P/N M7600
ECO:
DANGER
Toxic gases and or flammable liquids may be present in this monitoring system. Personal protective equipment may be required when servicing this instrument. Hazardous voltages exist on certain components internally which may persist
for a time even after the power is turned off and disconnected. Only authorized personnel should conduct maintenance and/or servicing.
Before conducting any maintenance or servicing, consult with authorized supervisor/manager.
Teledyne Analytical Instruments
Page 2
Model 7600
Copyright © 2005 Teledyne Instruments/ Analytical Instruments
All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in part, in any form or by any means, whether it be electronic, mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of Teledyne Instruments/ Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91749-
1580.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us free from defects of material and of construction, and that our liability shall be limited to replacing or repairing at our factory (without charge, except for transportation), or at customer plant at our option, any material or construction in which defects become apparent within one year from the date of shipment, except in cases where quotations or acknowledgements provide for a shorter period. Components manufactured by others bear the warranty of their manufacturer. This warranty does not cover defects caused by wear, accident, misuse, neglect or repairs other than those performed by TI/AI or an authorized service center. We assume no liability for direct or indirect damages of any kind and the purchaser by the acceptance of the equipment will assume all liability for any damage which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our apparatus, and to make any alterations in the dimensions, shape or weight of any parts, in so far as such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its user, and serves as a tool by which valuable data can be gathered. The information provided by the instrument may assist the user in eliminating potential hazards caused by his process; however, it is essential that all personnel involved in the use of the instrument or its interface, with the process being measured, be properly trained in the process itself, as well as all instrumentation related to it.
The safety of personnel is ultimately the responsibility of those who control process conditions. While this instrument may be able to provide early warning of imminent danger, it has no control over process conditions, and it can be misused. In particular, any alarm or control systems installed must be tested and understood, both as to how they operate and as to how they can be defeated. Any safeguards required such as locks, labels, or redundancy, must be provided by the user or specifically requested of TI/AI at the time the order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The purchaser is responsible for the training of personnel, for providing hazard warning methods and instrumentation per the appropriate standards, and for ensuring that hazard warning devices and instrumentation are maintained and operated properly.
Teledyne Instruments/ Analytical Instruments, the manufacturer of this instrument, cannot accept responsibility for conditions beyond its knowledge and control. No statement expressed or implied by this document or any information disseminated by the manufacturer or its agents, is to be construed as a warranty of adequate safety control under the user’s process conditions.
Teledyne Analytical Instruments ii
Page 3
Infrared Gas Analyzer
Specific Model Information
Instrument Serial Number: _______________________
Instrument Range: _______________
Calibrated for: _______________
Background Gas: _______________
Zero Gas: _______________
Span Gas: _______________
Teledyne Analytical Instruments iii
Page 4
Model 7600
Safety Messages
Your safety and the safety of others is very important. We have provided many important safety messages in this manual. Please read these messages carefully.
A safety message alerts you to potential hazards that could hurt you or others. Each safety message is associated with a safety alert symbol. These symbols are found in the manual and inside the instrument. The definition of these symbols is described below:
GENERAL WARNING/CAUTION: Refer to the instructions
for details on the specific danger. These cautions warn of specific procedures which if not followed could cause bodily Injury and/or damage the instrument.
CAUTION: HOT SURFACE WARNING: This warning is
specific to heated components within the instrument. Failure to heed the warning could result in serious burns to skin and underlying tissue.
WARNING: ELECTRICAL SHOCK HAZARD: Dangerous
voltages appear within this instrument. This warning is specific to an electrical hazard existing at or nearby the component or procedure under discussion. Failure to heed this warning could result in injury and/or death from electrocution.
Technician Symbol: All operations marked with this symbol
are to be performed by qualified maintenance personnel only.
No
Symbol
CAUTION: THE ANALYZER SHOULD ONLY BE USED FOR THE
NOTE: Additional information and comments regarding a
specific component or procedure are highlighted in the form of a note.
PURPOSE AND IN THE MANNER DESCRIBED IN THIS MANUAL.
Teledyne Analytical Instruments iv
Page 5
Infrared Gas Analyzer
IF YOU USE THE ANALYZER IN A MANNER OTHER THAN THAT FOR WHICH IT WAS INTENDED, UNPREDICTABLE BEHAVIOR COULD RESULT POSSIBLY ACCOMPANIED WITH HAZARDOUS CONSEQUENCES.
This manual provides information designed to guide you through the installation, calibration operation and maintenance of your new analyzer. Please read this manual and keep it available.
Occasionally, some instruments are customized for a particular application or features and/or options added per customer requests. Please check the front of this manual for any additional information in the form of an Addendum which discusses specific information, procedures, cautions and warnings that may be peculiar to your instrument.
Manuals do get lost. Additional manuals can be obtained from TI/AI at the address given in the Appendix. Some of our manuals are available in electronic form via the internet. Please visit our website at: www.teledyne-ai.com.
Teledyne Analytical Instruments v
Page 6
Model 7600
Table of Contents
List of Figures ...............................................................................x
List of Tables...............................................................................xii
Introduction.................................................................................15
1.1 Overview 15
1.2 Main Features of the Analyzer 15
1.3 Options Available 17
1.4 Applications 17
1.5 Description of the Main Unit 18
Installation................................................................................... 21
2.1 Unpacking the Analyzer 21
2.2 Choosing a Location 22
2.3 Mounting the Analyzer 23
2.4 Mounting the Input/Output Terminal Module 23
2.5 Gas Connections 25
2.5.1 Internal Piping Diagram 25
2.5.2 External Piping Diagram 27
2.5.3 Gas Conditioning 28
2.5.4 Flowrate 29
2.5.5 Preparation of Calibration Gas 29
2.5.6 Purging the Analyzer 30
2.5.7 Sample Gas Pressure 30
2.5.8 Example configuration of gas sampling system 30
2.6 Electrical Connections 32
2.6.1 Power Inlet 32
2.6.2 Input/Output Terminal Module 33
2.7 Testing the System 34
Teledyne Analytical Instruments vi
Page 7
Infrared Gas Analyzer
Operation .....................................................................................35
3.1 General Information 35
3.2 Display and Available Menus 36
3.3 The Display Screen 38
3.3.1 Measurement Mode 38
3.3.2 Setting/Selection Screen 40
3.4 Basic Operation 41
3.5 Setting Up the Analyzer in the User Mode 42
3.5.1 Switch Ranges 42
3.5.1.1 Manual Range Switching 43
3.5.1.2 Range identification contact operation 44
3.5.1.3 Remote Range Selection 45
3.5.1.4 Autoranging 46
3.5.2 Calibration Parameters 46
3.5.2.1 Calibration Value 46
3.5.2.2 Setting of Manual Zero Calibration 48
3.5.2.3 Setting of calibration range 50
3.5.2.4 Setting of auto calibration component/range 51
3.5.3 Alarm Setting 53
3.5.3.1 Configuring the Alarms 55
3.5.3.2 Hysteresis Setting 56
3.5.4 Setting of Auto Calibration 58
3.5.4.1 Auto Calibration 58
3.5.4.2 Gas Flow Time Setting 59
3.5.4.3 Cycle Setting Range 60
3.5.4.4 Remote Start 62
3.5.4.5 Forced Run/Stop of Auto Calibration 62
3.5.5 Setting of Auto Zero Calibration 65
3.5.5.1 Auto Zero Calibration 65
3.5.5.2 Forced run/stop of auto zero calibration 68
3.5.6 Peak Alarm Setting 70
3.5.7 Parameter Setting 73
Teledyne Analytical Instruments vii
Page 8
Model 7600
3.5.7.1 Output Hold 75
3.5.7.2 Average Value Reset: 78
3.5.7.3 Response Time 79
3.5.7.4 Average Period 79
3.5.7.5 Backlight Timer 80
3.5.7.6 Maintenance mode 81
3.6 Maintenance Mode 82
3.6.1 Sensor Input Value 83
3.6.2 Error Log 84
3.6.3 Calibration Log 84
3.6.4 Optical Adjustment 85
3.6.5 Moisture Interference Adjustment 86
3.6.6 Output Adjustment Screen 86
3.6.7 Other Parameters 87
3.7 Calibration 90
3.7.1 Zero Calibration 90
3.7.2 Span Calibration 91
Maintenance ................................................................................93
4.1 Routine Maintenance 93
4.2 Daily Check and Maintenance Procedures 93
4.3 Analyzer Maintenance 94
4.3.1 Sample Cell Cleaning 94
4.3.2 Cleaning the Cell Block 97
4.4 Optical Zero Adjustment 99
4.5 Moisture Interference Compensation Adjustment 102
4.6 Error Messages 104
4.6.1 Error Log File 106
4.6.2 Error Log Screen 106
4.6.3 Deleting Error Hstory 107
Appendix.................................................................................... 109
A.1 SPECIFICATIONS 109 A.2 Standard Functions 111
Teledyne Analytical Instruments viii
Page 9
Infrared Gas Analyzer
A.3 Optional Functions 115 A.4 Performance: 116 A.5 Standard Requirements for Sample Gas 117 A.6 Installation Requirements 118 A.7 EC Directive Compliance 118 A.8 Terminal Block Connections 119 A.9 Description on Terminal Block 120
Index...........................................................................................125
Teledyne Analytical Instruments ix
Page 10
Model 7600
List of Figures
Figure 1-1: Model 7600A Infrared Gas Analyzer...........................16
Figure 1-2: Model 7600 Description..............................................19
Figure 2-1: Slide Rail Mounting Dimensions .................................23
Figure 2-2: Input/Output Terminal Module.....................................24
Figure 2-3: Mounting the I/O Module............................................. 24
Figure 2-4: Internal Piping with Single or Dual Measuring Units ... 26
Figure 2-5: External Piping with Single Inlet and Outlet................27
Figure 2-6: External Piping with Two Pair of Inlet/Outlet (1).......... 28
Figure 2-7: External Piping with Two Pair of Inlet/Outlet (2).......... 28
Figure 2-8: Five Component Analysis System .............................. 31
Figure 2-9: Model 7600 Electrical Connections.............................32
Figure 2-10: Noise Suppression.................................................... 33
Figure 2-11: I/O Cable Connection................................................ 33
Figure 3-1: Front Panel of the Model 7600.................................... 35
Figure 3-2: Interface Keys on the Front Panel...............................35
Figure 3-3: Display Modes and Menu Hierarchy...........................37
Figure 3-4: Example Screen—5 Component Analysis 12
Channels................................................................................38
Figure 3-5: Setting/Selection Screen Areas .................................. 41
Figure 3-6: Hysteresis Example for a High Limit Alarm................. 57
Figure 3-7: Example Auto Calibration ........................................... 61
Figure 3-8: Example Auto Zero Calibration................................... 67
Figure 3-9: Peak Alarm Example ..................................................72
Figure 3-10: Output Hold for Manual Calibration........................... 75
Teledyne Analytical Instruments x
Page 11
Infrared Gas Analyzer
Figure 3-11: Output Hold for Auto Calibration ...............................76
Figure 3-12: Output Hold for External Hold ...................................76
Figure 3-13: Average Value Reset................................................79
Figure 3-14: Example Average Period ..........................................80
Figure 4-1: Top Cover Removal....................................................95
Figure 4-2: Sample Cell Removal..................................................95
Figure 4-3: Sample Cell.................................................................96
Figure 4-4: IR Window Assembly..................................................97
Figure 4-5: Detector Secure Nut....................................................98
Figure 4-6: Cell Block Assembly....................................................99
Figure 4-7: Optical Adjustment Display for Dual Optical System
Option...................................................................................100
Figure 4-8: Optical Zero Adjustment............................................101
Figure 4-9: Beam Adjustment Plate.............................................102
Figure 4-10: Bubbler Apparatus ..................................................103
Figure 4-11: The Error Log Screen..............................................107
Teledyne Analytical Instruments xi
Page 12
Model 7600
List of Tables
Table 2-1: Correspondence of Measured Components and
Measuring Units..................................................................... 27
Table 2-2: Zero and Span Gas......................................................29
Table 4.1 Troubleshooting and Maintenance................................ 93
Table 4-2: Components of Optical Adjustment Screen ............... 100
Table 4-3: Error Messages .......................................................... 104
Teledyne Analytical Instruments xii
Page 13
Infrared Gas Analyzer
DANGER
COMBUSTIBLE GAS USAGE
This is a general purpose instrument designed for use in a non-hazardous area. It is the customer's responsibility to ensure safety especially when combustible gases are being analyzed since the potential of gas leaks always exist.
The customer should ensure that the principles of operating this equipment are well understood by the user. Misuse of this product in any manner, tampering with its components, or unauthorized substitution of any component may adversely affect the safety of this instrument.
WARNING
Since the use of this instrument is beyond the control of Teledyne Instruments/ Analytical Instruments, referred as TI/AI, no responsibility by TI/AI, its affiliates, and agents for damage or injury from misuse or neglect of this equipment is implied or assumed.
Teledyne Analytical Instruments xiii
Page 14
Model 7600
Teledyne Analytical Instruments xiv
Page 15
Infrared Gas Analyzer Introduction
Introduction
1.1 Overview
The Model 7600 Infrared Gas Analyzer is a microprocessor based infrared gas analyzer used to measure the concentration of NO, SO CO
, CO and CH4 contained in sample gas. It is based on the principle
2
that different molecular species have unique absorption spectrum in the infrared, and the intensity of absorption is determined by the Lambert­Beer law.
In addition to the infrared analyzer, the Model 7600 can also be equipped with a built-in compact paramagnetic O2 sensor or employ an external zirconia oxygen sensor to increase to 5 the number of simultaneous component species the unit can analyze.
,
2
Teledyne's Series 7600 Infrared (IR) Gas Analyzer is conveniently packaged in either a 19" rack mount or NEMA-4 wall mount enclosure. The NEMA-4 enclosure can be X or Z-purged to satisfy hazardous area installation requirements.
A high-sensitivity mass flow type twin detector is used for infrared measurements. By utilizing a single beam, double path design in conjunction with a serial dual-layer transmission detector, the Series 7600 delivers long term, drift-free performance.
The concentration of the desired gases is displayed on a large, easy­to-read back-lit LCD. Figure 1-1 shows the standard rack mountable Model 7600. The user interface is very intuitive and the menu / mode selection buttons, which are readily accessible, provide the operator with dynamic control and extensive diagnostic capabilities.
1.2 Main Features of the Analyzer
The Model 7600 Infrared Gas Analyzer is designed for accurate and reliable gas analysis and is easy to operate. The following features are standard on the 7600 instrument:
Teledyne Analytical Instruments 15
Page 16
Introduction Model 7600
Figure 1-1: Model 7600A Infrared Gas Analyzer
Simultaneous measurement of up to five components
Excellent long-term stability
Large, easy to read LCD display showing all simultaneous
measurements and computations
Slide-out, chassis design to facilitate any optical or
maintenance adjustments required to fine tune analyzer performance (7600A)
Multiple, in-depth on-screen analyzer functions easily
accessible using the front-panel user interface buttons
Follow & Hold output signal control (during calibration)
Remote range change control
Low / Hi limit alarms
Range ID signals
Auto-calibration with user adjustable frequency and gas flow
time setting programming capabilities
Remote auto-calibration initiation
Auto-calibration status contacts
Teledyne Analytical Instruments 16
Page 17
Infrared Gas Analyzer Introduction
Instrument or calibration error contact outputs
Extra functions included such as average value computation,
O2 conversion
Pump ON/OFF contact
1.3 Options Available
Percent O
detector — Paramagnetic (built-in) or ZrO2
2
(externally installed), user preference
O
correction (the conversion of measured CO and SO2
2
readings into values at standard O2 concentration). Consult factory for more detail for this function.
Communication functions:
RS-232C (9 pins D-Sub connector)
Half-duplex bit serial
Modbus protocol
1.4 Applications
The Model 7600 Infrared Gas Analyzer is a versatile analytical instrument tool and is ideally suited for multi-parameter gas analysis requirements for applications such as:
Combustion control within the power, pulp and paper, steel,
and cement industries
Heat treating / Inert gas blanketing atmosphere control
Bulk-gas impurity analysis within the air separation industry
Anaerobic digester / Bio-gas / Land-fill gas analysis
Vent gas analysis of oxyhydrochlorination reactors (EDC)
Off-gas analysis on PTA and Maleic Anhydride reactors
Fluid Catalytic Cracker (FCC) regeneration gas analysis
Ammonia / Fertilizer process gas stream analysis
Continuous Emissions Monitoring Systems (CEMS)
Teledyne Analytical Instruments 17
Page 18
Introduction Model 7600
Biochemistry and fermentation,
Automotive emission analysis
Explosive and toxic gas analysis
Chemical analysis
Refinery operation
Research applications
1.5 Description of the Main Unit
Teledyne's Series 7600 Infrared Gas Analyzer measures the concentration of NO, SO2, CO2, CO, and CH4 in a gas mixture on a continuous basis. The Series 7600 can also be supplied with an oxygen sensor, which allows the simultaneous measurement of oxygen concentration as well. The NEMA-4 enclosure can be X or Z-purged to satisfy hazardous area installation requirements.
The system uses a high-sensitivity mass flow type twin detector for infrared measurements. A single beam, double path optical bench is installed in conjunction with a serial dual-layer transmission detector.
The concentration of the desired gases is displayed on a large, easy­to-read backlit LCD. The user interface is very intuitive and the menu / mode selection buttons, which are readily accessible, provide the operator with dynamic control and extensive diagnostic capabilities. See Figure 1-2.
The front panel mounted power switch turns the instrument on and off while the display switch controls power to the display.
Front panel mounted handles make sliding the instrument out of the panel rack or enclosure a simple task for easy maintenance.
Teledyne Analytical Instruments 18
Page 19
Infrared Gas Analyzer Introduction
Figure 1-2: Model 7600 Description
Gas connections for the 7600 instrument are made on the rear panel of the instrument using the ¼” NPT fittings installed. The standard instrument has two measuring units and a pair of inlet/outlet gas
connections exist for each unit. See Section 2.5 Gas Connections.
A power receptacle on the rear panel accepts the three-prong power cable supplied with the instrument. The Model 7600 operates on 100­240 VAC 50/60 Hz power. The power inlet conforms to EN60320 Protection Class 1 specifications.
Teledyne Analytical Instruments 19
Page 20
Introduction Model 7600
Teledyne Analytical Instruments 20
Page 21
Infrared Gas Analyzer Installation
Installation
Installation of the analyzer includes:
1. Unpacking the system.
2. Choosing a suitable location
3. Mounting the analyzer
4. Mounting the terminal module
5. Installing gas connections
6. Making electrical connections
7. Testing the installation.
CAUTION: READ THIS CHAPTER IN ITS ENTIRETY BEFORE
INSTALLING THE SYSTEM.
FOR INDOOR USE ONLY.
2.1 Unpacking the Analyzer
The Model 7600 Infrared Gas Analyzer is shipped with the following components:
7600 Analyzer
Input/Output terminal module set
Connection cable
Power cable
Fuse (2)
Cell window mounting tool
Slide rail (2)*
Relay board for auto calibration*
Relay board connection cable*
Instruction manual
Teledyne Analytical Instruments 21
Page 22
Installation Model 7600
* These optional items are included if specified at the time of purchase
As soon as you receive the instrument, carefully unpack and inspect the analyzer and components for damage. Immediately report any damage to the shipping agent. The analyzer is shipped with all the materials you need to install and prepare the system for operation.
2.2 Choosing a Location
CAUTION: THIS UNIT IS NOT EXPLOSION-PROOF. DO NOT
INSTALL IT IN A LOCATION WHERE EXPLOSIVE OR FLAMMABLE GASES ARE PRESENT.
Select an installation location that meets the following criteria:
This instrument should be rack mounted, or mounted in a
steel enclosure.
Indoor location.
Vibration-free.
Not exposed to direct sunlight.
Clean and non-cluttered space.
Depending on the options selected at the time of purchase,
the model 7600 requires an AC power source of 100V to 240V AC.
Operating voltage: 85V to 264V AC
Rated frequency: 50/60 Hz
Power consumption: 250 VA max.
Plug: Comformity to EN60320 class I type 3-pin inlet
Operation conditions:
Ambient temperature: -5° to 45°C
Ambient humidity: 90 % RH or less, no condensation
Teledyne Analytical Instruments 22
Page 23
Infrared Gas Analyzer Installation
2.3 Mounting the Analyzer
The 19” rack mountable 7600 analyzer is designed for either a guide rail or slide rail mounting within a rack or cabinet. The guide rail method supports the weight of the instrument from the bottom and can be used when there is adequate space for removing the top cover for maintenance. The slide rail mount supports the instrument from sliders mounted on the side of the instrument as shown in Figure 2-1. The instrument is then able to slide out of the rack or enclosure for required access during maintenance.
Figure 2-1: Slide Rail Mounting Dimensions
2.4 Mounting the Input/Output Terminal Module
The input/output module is the electronic interface that handles the various signals to and from the analyzer. It consists of up to five terminal blocks, a communications connector, a cable connector, and a solenoid drive output connector mounted together on a single mounting plate. See Figure 2-2.
Mount the input/output terminal module to the rack or cabinet panel using the six M4 screws supplied. For grounding, see Figure 2-3.
Note: To avoid noise generated from external units, mount the
I/O terminal module mounting plate to the panel making sure there is adequate metal contact for continuity at the
Teledyne Analytical Instruments 23
Page 24
Installation Model 7600
mounting surface. Connect the panel to the same ground as the analyzer main unit.
Figure 2-2: Input/Output Terminal Module
Figure 2-3: Mounting the I/O Module
Teledyne Analytical Instruments 24
Page 25
Infrared Gas Analyzer Installation
2.5 Gas Connections
Gas connections are made on the rear panel of the analyzer. Adhere to the following guidelines when making gas connections:
Use a corrosion resistant tube such as Teflon, stainless or
polyethylene to connect the instrument to a sampling system. Even if there is a danger of corrosion, refrain from using rubber or soft vinyl tubing. This would result in instrument inaccuracies due to gas absorption by the piping materials.
Pipe connection port is Rc1/4 female thread (or NPT1/4).
Piping should be cut as short as possible for quicker response. About 4 mm inner diameter is recommended.
Keep out dust and debris from the tubing and connections .
Always use clean tubing and fittings.
Connect the gas tube as follows:
Sample gas inlet: Attach the sample gas tube to the inlet
fitting. The sample gas should be filtered and dehumidified before passing into the analyzer. This port is also used to connect the zero and span calibration gases.
The gas flow should be constant within the range of 0.5
L/min ±0.2 L/min.
Sample gas outlet: Sample gas exits the analyzer through
the gas out port at atmospheric pressure. Exhaust gases must be vented safely.
Purge gas inlet: This connection is used for purging the
inside of the analyzer. Purging is not always required. See
Section 2.5.6 Purging the Analyzer. When required, use dry
N
or instrumentation air for the purge gas. Use a flow rate of
2
1L/min or greater).
2.5.1 Internal Piping Diagram
Note: When the purge gas inlet is provided, an internal
connection to measuring unit 2 is installed.
Teledyne Analytical Instruments 25
Page 26
Installation Model 7600
An internal piping diagram for an instrument with one or two measuring units is shown in Figure 2-4. When there are two measuring units, if a built-in oxygen sensor is used, it must be installed in measuring unit 2. The diagram shows the combination of two cells used in measuring unit #2. This is possible by combining ranges. When a single measuring unit is used, there is only a single inlet and outlet gas connection.
Figure 2-4: Internal Piping with Single or Dual Measuring Units
Depending on the options chosen, the analyzer can be configured in several ways. Table 2-1 lists the possible configurations for single and dual measuring units with 1 to 4 component meters.
Teledyne Analytical Instruments 26
Page 27
Infrared Gas Analyzer Installation
/
/
Table 2-1: Correspondence of Measured Components and Measuring Units
Measuring components Measuring unit 1 Measuring unit 2
1-component meter for NO, SO
, CO2, CO and CH4
2
2-component meter for NO/SO
and CO
2
2
CO
2-component meter for
Each component None
2
CO
2
None
NO/SO CO
NO CO
NO/CO 3-component meter for
NO/SO
/CO
2
4-component meter for NO/SO
/CO2/CO
2
NO/SO
NO/SO
CO
2
CO2/CO
2
2.5.2 External Piping Diagram
There are several ways to bring sample gas to the analyzer depending on the number of inlet/outlet gas connections on your instrument Recommended piping diagrams are shown in Figures 2-5, 2­6 and 2-7. Figure 2-5 is a schematic for an instrument with a single pair of inlet/outlet gas connections. Figures 2-6 and 2-7 are used when there are two pair of inlet/outlet connections. Note that a NO2/NO converter is used when NO measurement is used for NOx analysis.
Figure 2-5: External Piping with Single Inlet and Outlet
Teledyne Analytical Instruments 27
Page 28
Installation Model 7600
Figure 2-6: External Piping with Two Pair of Inlet/Outlet (1)
Figure 2-7: External Piping with Two Pair of Inlet/Outlet (2)
2.5.3 Gas Conditioning
For optimum performance, the sample gas should be treated as
follows:
A filter should be installed to remove any dust or particles in
the sample gas. For a final stage filter, use a filter that can remove dust particles of 0.3 μm.
Teledyne Analytical Instruments 28
Page 29
Infrared Gas Analyzer Installation
The dew point of the sample gas must be lower than the
ambient temperature to avoid condensation in the analyzer. If vapor is contained in the sampling gas, the dew point should be lowered to 0°C by using a dehumidifier.
If SO
mist is contained in the sample gas, use a mist filter or
2
cooler to remove the SO2 mist. Other mists should be removed by using an appropriate mist filter or cooler.
Corrosive gases such as Cl
, F2 and HCl, if contained in the
2
sample gas in considerable amounts, will shorten the life of the instrument.
Sample gas temperature should be within 0 to 50°C. Hot gas
should not be fed directly into the instrument.
2.5.4 Flowrate
Use a flowrate of 0.5L/min ±0.2L/min. The sample system should be designed to avoid any flow fluctuation during measurement.
Use a flowmeter to observe the flow reading as shown in the external piping diagrams of Figures 2.5, 2.6 and 2.7.
2.5.5 Preparation of Calibration Gas
Routine calibration using calibration gases is required for optimizing the performance of this instrument. Once a week is a suggested calibration frequency. Table 2-2 indicates the zero and span gas required for calibration.
Table 2-2: Zero and Span Gas
Zero gas
Span gas other than for O
2
Span gas for
analysis
O
2
Teledyne Analytical Instruments 29
Analyzer without
O
2
N
gas N2 gas Dry air
2
Gas concentration of 90% or more of full scale
NA
Analyzer with built-in
O2
Gas concentration of 90% or more of full scale
Gas concentration of 90% or more of full scale
Gas concentration of 90% or more of full scale
1 to 2% O
Analyzer with external
zirconia O2 sensor
2
Page 30
Installation Model 7600
2.5.6 Purging the Analyzer
In general, purging the analyzer is not required unless one of the
following cases apply:
A combustible gas component is contained in sample gas.
Corrosive gas is contained in the atmospheric air at the
installation site.
The same gas as the sample gas component is contained in
the atmospheric air at the installation site.
In the above situations, the inside of analyzer should be purged
with instrument air or N2. Use a purge flow rate of about 1 L/min.
Use a filter to remove dust or mist from the purge gas.
2.5.7 Sample Gas Pressure
The sample gas pressure at the outlet should be atmospheric
pressure.
2.5.8 Example configuration of gas sampling system
A typical system configuration with five component gas analysis for monitoring combustion exhaust gas from boiler, refuse incinerator, etc. is shown in Figure 2-8. Contact Teledyne for specific application system configuration or further information.
The system shown is comprised of:
Model 7600—with dual measuring unit option and external
zirconia oxygen sensor.
Gas Extractor—with stainless steel filter of standard mesh 40
μm equipped with integral heater.
Mist Filter—to remove condensate, mist and dust from the
sample before it enters the analyzer.
Safety Drain Trap—with two compartments for positive and
negative pressure. It monitors and adjusts the sample gas pressure.
Gas Aspirator—to aspirate the sample gas.
Electronic Gas Cooler—used to dry the sample gas to a dew
point of approximately 2°C (35.6°F).
Teledyne Analytical Instruments 30
Page 31
Infrared Gas Analyzer Installation
Solenoid Valve—for introducing calibration gas and
OFF/ON flow of sample gas.
Membrane Filter—PTFE filter for rempval of fine dust
particles.
Flowmeter—adjusts and monitors the sample gas flowrate.
Standard Gas—a bank of calibration gases used for setting
zero and span of the analyzer for each species monitored.
Zirconia Sensor— an external zirconia sensor is used for
measuring the oxygen concentration in the sample gas. Some instruments use a built-in paramagnetic oxygen sensor instead aof an external zirconia sensor.
NO
/NO Converter—added to the NO analysis circuit which
2
uses an efficient catalyst for converting NO2 to NO for analysis.
Figure 2-8: Five-Component Analysis System
Teledyne Analytical Instruments 31
Page 32
Installation Model 7600
2.6 Electrical Connections
2.6.1 Power Inlet
The power inlet for the Model 7600 is located on the rear panel as shown in Figure 2-9. The proper 3-prong power cord has been supplied with your instrument. Connect the power cable to the power inlet.
Figure 2-9: Model 7600 Electrical Connections
The Model 7600 requires 100-240 VAC 50/60 Hz power.
Avoid installing this instrument near an electrical unit that generates electrical noise such as a high frequency furnace or electric welder. If using the instrument near such a noise-generating unit is unavoidable, use a separate power line to avoid noise.
Mount a noise suppressor such as varister or spark killer as shown in Figure 2-10 to the noise generating unit when noise is generated from relays or solenoid valves. The suppressor must be mounted near the noise generating source, or it will have no effect.
Teledyne Analytical Instruments 32
Page 33
Infrared Gas Analyzer Installation
Figure 2-10: Noise Suppression
2.6.2 Input/Output Terminal Module
Use the supplied cable to interface the I/O Module to the analyzer.
Plug the cable connector into the receptacle at the rear panel of the
analyzer and the receptacle on the PC board of the input/output module.
Make sure that the ferrite core attached to the I/O cable goes to the
analyzer. See Figures 2-9 and 2-11.
Figure 2-11: I/O Cable Connection
The I/O Module carries various input and output signals. A detailed
list of these signals and pinout information is given in the Appendix.
Teledyne Analytical Instruments 33
Page 34
Installation Model 7600
2.7 Testing the System
After the analyzer has been installed with gas and electrical connections but prior to powering up the unit make sure you have:
Installed the unit correctly
Checked the gas connections for leaks
Once the above checks have been made, you can connect the power source and turn the analyzer on using the power switch on the front panel. Allow the analyzer to warm up for four hours.
When the instrument is first turned on, you will see the measurement screen. While in the warm up stage, the readings are inaccurate. They may even be above the upper limit of range, however, this is not an error.
After the analyzer has come to equilibrium (approximately four hours after first powering up) the instrument is ready to be configured and calibrated for your process. This is described in Section 3.
Teledyne Analytical Instruments 34
Page 35
Infrared Gas Analyzer Operation
Operation
3.1 General Information
This section describes the front panel interface of the analyzer.
Figure 3-1 Shows the Model 7600 Infrared Gas Analyzer front panel. The user interface consists of 8 membrane switch buttons or keys, a power ON/OFF switch and a display ON/OFF switch. The 8 user keys are shown in Figure 3-2 and described below.
Figure 3-1: Front Panel of the Model 7600
Figure 3-2: Interface Keys on the Front Panel
Teledyne Analytical Instruments 35
Page 36
Operation Model 7600
MODE key: Used to switch the mode.
LEFT/RIGHT Arrow key:
UP key: Used to change the selected item by moving the
DOWN key: Used to change the selected item by moving the
ESC key: Used to return to a previous screen or cancel the
ENT key: Used for confirmation of selected item or value, and
ZERO key: Used for zero calibration.
SPAN key: Used for spancalibration.
Used to change the selected item by moving the cursor and numeral digit.
cursor and to increase a numeric value.
cursor and to decrease a numeric value.
setting midway.
for performing a calibration.
3.2 Display and Available Menus
The display on the Model 7600 has two modes: measurement mode
and user mode. To change between modes, press the MODE key.
In the measurement mode, the display indicates the composition of the sample gas. Up to 12 channels of output are available although the screen only shows 5 channels at a time. To see additional channels, scroll down or up using the UP or DOWN keys.
Calibration is performed from the measurement mode by pressing
ZERO or SPAN. See Section 3.7 Calibration.
Teledyne Analytical Instruments 36
Page 37
Infrared Gas Analyzer Operation
Figure 3-3: Display Modes and Menu Hierarchy
Teledyne Analytical Instruments 37
Page 38
Operation Model 7600
The user mode is where the operator configures the instrument and sets up various parameters for the intended application. From the user mode the following menus are available:
Switch Ranges
Calibration Parameters
Alarm Setting
Setting of Auto Calibration
Setting of Auto Zero Calibration
Setting of Peak Alarm
Parameter Setting
Figure 3-3 shows the overall structure of the Model 7600 display and menu hierarchy.
3.3 The Display Screen
3.3.1 Measurement Mode
The measurement mode screen is the default mode when the power is turned on. It displays component and concentration information about the process. The screen depends on the number of components. Figure 3-4 is an example of a measurement mode screen for an instrument configured for NO, SO2, CO2, CO and O2 analysis with a 12 channel output.
Figure 3-4: Example Screen—5 Component Analysis 12 Channels
Teledyne Analytical Instruments 38
Page 39
Infrared Gas Analyzer Operation
p
(Upp
)
p
play
The numbered call outs in Figure 3-4 refer to:
No. Name Function
(1) Component
display
(2) Concentration
display (3) Range display Displays range values.
(4) Unit display Displays unit with ppm and vol %.
(5) Peak alarm
com
onent
(6) Peak alarm
concentration (7) Peak alarm
times (8) Peak alarm
unit dis
Displays component of instantaneous value, corrected instantaneous value, corrected average value, etc.
Displays measured value of concentration.
Displays peak alarm component.
Displays peak alarm concentration display.
er limit value
Displays the alarm times exceeding the
eak value.
Displays units of peak alarm with times/H.
Instantaneous value and concentration value:
The concentration display of Ch (component) where sampling
components such as “CO
”, “CO” or “O2 are displayed in the component
2
display, indicates current concentration values of the measured components contained in gas that is now under measurement.
O
correction concentration values:
2
Ch components where “cv**” is displayed as “cv CO” in the component display are calculated values. They are obtained from the following equation by setting sampling components, O2 instantaneous/ concentration values and O2 correction reference value (see item 6.8).
On-21
output Correction ×
=
⎜ ⎝
Cs
Os21
Teledyne Analytical Instruments 39
Page 40
Operation Model 7600
Where:
On: The value of the O
correction reference value (Value set by
2
application)
Os: Oxygen concentration (%)
Cs: Concentration of relevant measured component. Note that Os
does not exceed the O See Section 3.6
The corrected sampling components are NO
limit value set in “Other Parameter”
2
Maintenance Mode.
, SO2 and CO only.
X
O
2 correction concentration average value:
In the Ch (component) and O2 average value where “CV/AV **” is displayed as “ AV CO” in the component display, a value obtained by averaging O
correction concentration value or O2 average value in a
2
fixed time is output every 30 seconds.
Averaging time can be changed between 1 minute and 59 minutes or 1 hour and 4 hours according to the average time settings (See Section 3.5.7
Parameter Setting).
The set time is displayed as “1h” , for instance, in the range display.
* The measurement ranges of O
correction concentration value and O2
2
correction concentration average value are the same as that of the measuring components. Also, the measurement range of O value is the same as that of O
.
2
average
2
3.3.2 Setting/Selection Screen
The setting/selection screen is configured as shown in Figure 3-5:
In the status display area, the current status is displayed.
In the message display area, messages associated with
operation are displayed.
In the setting item and selection item display area, items or
values to be set are displayed, as required. To work in a particular area, move the cursor to any item by using the UP, DOWN and LEFT/RIGHT keys.
Teledyne Analytical Instruments 40
Page 41
Infrared Gas Analyzer Operation
Figure 3-5: Setting/Selection Screen Areas
3.4 Basic Operation
Measurement mode
The measurement mode is the default mode of the instrument and appears when the unit is first turned on. It displays information about the process being analyzed such as: channel number, component species analyzed, concentration or calculated value, etc. In this mode, the display can show up to 5 channels on a single screen. To view additional channels, scroll down by pressing the DOWN key or back up using the UP key. Each press scrolls the screen by one channel. You can also calibrate the instrument from this mode by pressing the appropriate ZERO or SPAN button. See Section 3.7
Calibration.
User mode displays:
In the user mode you can:
Switch ranges
Set calibration parameters
Adjust alarm settings
Teledyne Analytical Instruments 41
Page 42
Operation Model 7600
Set up the Auto Calibration feature
Setup an auto zero calibration
Adjust the peak alarm parameter Setting.
These settings are described in Section 3.5.
To enter the user mode from the measurement mode, press the MODE key. To return to the measurement mode, press the MODE key again or ESC.
3.5 Setting Up the Analyzer in the User Mode
3.5.1 Switch Ranges
From the Switch Ranges menu you can:
a.
Manually select a desired range for any channel.
b.
Use the remote range switch contacts to select a range.
c.
Select autoranging.
To enter the Switch Ranges menu:
1.
Press the MODE key to enter the user mode.
Teledyne Analytical Instruments 42
Page 43
Infrared Gas Analyzer Operation
2. Move the cursor to the “Switch Ranges” option and press
ENT. The “Channel Selection” screen appears.
Move the cursor by pressing the UP/DOWN keys to the
3. desired channel and press ENT.
There are 3 range switch modes in the second column to select from:
MR: Manual selection
RR: Remote range switch
AR: Autoranging
The selected range switch mode is highlighted for the channel you
selected.
4. Press the UP/DOWN keys to toggle between the
available switch modes.
5. Then press the ENT key to confirm the selection.
3.5.1.1 MANUAL RANGE SWITCHING
To manually switch between analysis ranges:
1.
Select the MR option from the Switch Range screen and
press the ENT key.
Teledyne Analytical Instruments 43
Page 44
Operation Model 7600
Move the highlight of the cursor to range selection and
2. select the desired range using the UP/DOWN keys. The highlighted arrow indicates the currently selected range.
Press ENT to accept the selection. Measurement is now
3. carried out using the selected range.
Note: If “RR” or “AR” is selected as range switch mode, manual
range selection as described above is not possible
The range for O2 correction value, O2 correction average
value, and O2 average value is automatically switched if the corresponding instantaneous value range is switched.
To abort the range selection, press the ESC key and the setting
operation is made invalid and the previous screen appears.
3.5.1.2 RANGE IDENTIFICATION CONTACT OPERATION
In all of the range switch modes (MR, RR, AR), the status of the range identification relay contact corresponding to each Ch (component) is closed when Range 1 is selected, and open when Range 2 is selected.
Note that even if the range is switched while a hold of measurement is in place, for instance, by a remote hold contact input or the hold of measurement value during calibration, the range identification contact maintains the contact state immediately before the hold. After stop of the hold, the contact state of the current range is resumed.
Teledyne Analytical Instruments 44
Page 45
Infrared Gas Analyzer Operation
3.5.1.3 REMOTE RANGE SELECTION
The range can be selected remotely if a remote switch has been installed. To use this feature, the range mode must be switched in the display to RR (Remote Range) .
To configure the instrument for remote range selection:
1.
Press the MODE key to enter the user mode.
Move the cursor to the “Switch Ranges” option and press
2. ENT. The “Channel Selection” screen appears.
Move the cursor by pressing the UP/DOWN keys to the
3. desired channel and press ENT.
The selected range switch mode is highlighted for the channel you
selected.
4. Press the UP/DOWN keys to toggle between the
available switch modes until RR is displayed.
5. Press the ENT key to accept the selection.
Teledyne Analytical Instruments 45
Page 46
Operation Model 7600
3.5.1.4 AUTORANGING
The Model 7600 can be set to autorange where the instrument will select the analysis range automatically. To use the autorange feature, AR must be selected for the switch mode. Note that each channel is independent so that autoranging can be used for some or all of the available channels. the remaining channels can be configured to manual or remote switching.
To setup autoranging use the same procedure as Section 3.5.1.1 except you must select AR for the switch mode on the channel or channels you want to autorange.
3.5.2 Calibration Parameters
The next menu choice from the user screen is the Calibration Parameters menu. Here calibration settings can be configured for your particular application. There are 4 submenus in the Calibration Parameters menu. They are:
Calibration Value—set the concentration of calibration gas About ZERO Calibration—set each or all components to be
zero calibrated during a manual zero calibration.
About Calibration Range—sets whether each component
during a zero or span calibration should be calibrated with a single or dual range.
Auto Calibration Components/Range— select the component
and range with which auto calibration is to be performed.
3.5.2.1 CALIBRATION VALUE
This submenu allows you to set the concentration of the standard gas (zero and span) for each channel used for calibration.
To enter the Calibration Value submenu:
1.
From measurement mode, press the MODE key to
display the user mode.
2.
Point the cursor to “Calibration Parameters” by pressing
the UP or DOWN key.
3. Press the ENT key to enter the Calibration Parameters
menu.
Teledyne Analytical Instruments 46
Page 47
Infrared Gas Analyzer Operation
4. In the “Calibration Parameters” screen that appears, point
the cursor to “Calibration Value” by pressing the UP or DOWN key. Press the ENT key.
To change the concentration of a calibration gas:
1.
Point the cursor to the channel you want to set by using
UP or DOWN key. Press the ENT key.
the
2.
Select the concentration item you want to set by pressing
the UP or DOWN or SIDE key. Press the ENT key, and the selected value is highlighted.
Enter the calibration gas concentration values (zero and
3. span). For any numerical value entry, press the UP or DOWN key to increase or decrease a single digit. Select the next digit by pressing the SIDE key.
4.
When the concentration is correct, save the entry by
pressing the ENT key. The saved value becomes valid from the next calibration process.
Teledyne Analytical Instruments 47
Page 48
Operation Model 7600
Note: Enter settings that correspond to each range. If a zirconia
type O2 sensor is used, select 21.00 for the field of Zero (when air is used), and select the concentration listed on the cylinder if the air contained in a cylinder is used.
To close the calibration concentration value setting process or
cancel this mode midway, press the ESC
key. The previous screen will
return.
Setting range of values:
For NOx, SO2, CO2, CO, CH\4, external O2 measurement and built-
in paramagnetic O
sensors, the setting range for span gas is 1 to 105%
2
of full scale. The instrument will not accept values outside of this range.
For external Zirconia O
sensors, the setting range for zero gas is 5
2
to 25 vol % and span from 0/01 to 5 vol %
3.5.2.2 SETTING OF MANUAL ZERO CALIBRATION
When zero calibration is performed manually, you can choose whether all components are to be calibrated simultaneously or each component calibrated while selecting one by one.
This is configured from the About Zero submenu of the Calibration Parameters menu. To get to the About Zero submenu:
1.
From the measurement mode, press the MODE key to
display the user mode.
2.
Point the cursor to “Calibration Parameters” by pressing
the UP or DOWN key and press the ENT key.
3.
In the “Calibration Parameters” screen that appears, point
the cursor to “About ZERO Calibration” by pressing the UP or DOWN key. Press the ENT key.
Teledyne Analytical Instruments 48
Page 49
Infrared Gas Analyzer Operation
This brings up the Cal. Settings ZERO Cal. screen where you can set whether all components or individually selected components are zeroed during a manual zero calibration. The choices are:
at once—these components will be zeroed simultaneously
during a manual zero.
each—these components will be zeroed individually during a
manual zero.
To change the manual zero setting:
1.
In the “Cal. Settings ZERO Cal.” screen, point the cursor
to the channel you want to set by using the UP or DOWN key. Press the ENT key.
2.
Select “at once” or “each” by pressing the UP or DOWN
key to toggle between the two options. Press the ENT key when the desired option is displayed.
To close the Cal. Settings ZERO Cal. screen or cancel this mode midway, press the ESC
Example:
key. The previous screen will return.
The options “each” or “at once” can be determined for each channel.
When the setting is “each”, select the channel on the manual
zero calibration screen (see Section 3.7) and perform a zero calibration. Only that component is zeroed.
When the setting is “at once”, performing a manual
calibration (see Section 3.7) will zero calibrate all components tagged “at once”. You will notice on the manual
Teledyne Analytical Instruments 49
Page 50
Operation Model 7600
zero calibration screen that all there are cursors on all components where “at once” is set.
Note: When the cylinder air or atmospheric air is used for the
zero gas, select “at once.”
3.5.2.3 SETTING OF CALIBRATION RANGE
Use the “About Calibration Range” submenu to configure the
instrument for calibration on one or two ranges for a channel.
To set calibration one or both ranges:
1.
From the measurement mode, press the MODE key to
display the user mode.
Point the cursor to “Calibration Parameters” by pressing
2. the UP or DOWN key and press the ENT key.
In the “Calibration Parameters” screen that appears, point
3. the cursor to “About Calibration Range” by pressing the UP or DOWN key. Press the ENT key.
In the next screen that appears, point the cursor to the
4. channel you want to set by pressing the UP or DOWN key then press the ENT key.
On the “calibration range selection” screen that appears,
5. select “both” or “current” by pressing the UP or DOWN key.
If “both” is selected, zero or span calibration is performed with Range 1 and Range 2 of the selected channel.
If “current” is selected, zero or span calibration is
Teledyne Analytical Instruments 50
Page 51
Infrared Gas Analyzer Operation
performed only for the range displayed when calibration of the selected channel is performed.
Press the ENT
6.
key after the selection, and the specified
calibration is set.
To close the calibration range selection screen or cancel this mode
midway, press the ESC key. The previous screen will return.
Note: To perform calibration for “both,” set the same calibration
gas concentration for both ranges.
After setting the calibration range to “both”, you will notice in the Manual Calibration screen (see Section 3.7), that there are cursors next to both ranges whereas there would only be a single cursor next to the range for calibration if “current” was set. For example, if channel 1 and channel 4 were set to “both” then cursors will appear in both ranges of Ch1 and Ch4 in the Manual Calibration screen shown below.
3.5.2.4 SETTING OF AUTO CALIBRATION COMPONENT/RANGE
This menu is used to select the channel and the range in which auto calibration is to be performed. If a channel has been set to autoranging
Teledyne Analytical Instruments 51
Page 52
Operation Model 7600
(AR), it will be calibrated in the range set in this menu when auto calibration is performed. From this menu you can enable or disable the auto calibration feature.
To navigate to the Auto Calibration Components/Range menu:
1.
From the measurement mode, press the MODE key to
display the user mode.
Point the cursor to “Calibration Parameters” by pressing
2. the UP or DOWN key and press the ENT key.
In the “Calibration Parameters” screen that appears, point
3. the cursor to “Auto Calibration Components/Range” by pressing the UP or DOWN key. Press the ENT key.
In the “Calibration Parameters” screen that appears, point
4. the cursor to “Auto Calibration Components / Range” by pressing the UP or DOWN key. Then press the ENT key.
To set the range used for auto calibration:
1.
In the “Auto Calibration Components / Range” selection
screen that appears, point the cursor to the channel whose
Teledyne Analytical Instruments 52
Page 53
Infrared Gas Analyzer Operation
range you want to set using the UP or DOWN key. Then press the ENT.
2.
The cursor next to the range of the selected channel is
highlighted. Select the range to be calibrated using the UP or DOWN key. Then press the ENT key. The auto calibration will be performed on the selected range.
Channels which have been set for autoranging (AR) will undergo the auto calibration and manual calibration on the range selected here. Once the calibration has started, the range will automatically switch and on completion of the calibration, the original range is resumed.
The range identification contact is interlocked with the range after the switch. However if the hold setting is set to “ON” the contact status before calibration is maintained.
To ENABLE/DISABLE the auto calibration feature:
1.
With the cursor next to the range of the selected channel
you want to enable or disable for autocalibration (see steps 1 and 2 above), use the SIDE key to highlight the “enable” or “disable” option.
Use the UP or DOWN keys to toggle between the two
2. options then select the desired status by pressing ENT.
To close the Auto Calibration Components/Range screen or cancel
this mode midway, press the ESC key. The previous screen will return.
Note: The order in which the Zero and Span calibrations are
performed in the auto calibration routine are different. For the zero calibration, the calibration is performed according to the order in which the channels were set to “enable”. For Span, the calibration begins with the lowest channel number first and proceeds sequentially to the highest channel number set to “enable”.
3.5.3 Alarm Setting
The Model 7600 is equipped with 6 alarms (5 or 6 concentration and a power failure alarm). The alarm configuration and settings can be assigned to any channel and adjusted at any time by entering the user mode and navigating to the Alarm Setting submenu. Arbitrary 6 alarm contact outputs are used which can be configured as:
Teledyne Analytical Instruments 53
Page 54
Operation Model 7600
High Limit: sets the upper value (concentration) above which
the alarm will trigger. Has an on-screen display.
Low Limit: sets a lower value (concentration) below which
the alarm will trigger. Has an on-screen display
High or Low Limit: triggers an alarm when the concentration
is above the high limit or below the low limit.
High-High: sets the upper value (concentration) above which
the alarm will trigger.
Low-Low: sets a lower value (concentration) below which
the alarm will trigger.
Power Failure: (Alarm 6 only) normally closed contacts open
when power is removed.
Configuring the alarms involves:
Selecting the alarm type
Assigning it to a channel
Establishing setpoints (concentration alarms only)
Enabling or disabling it
A channel can be configured for multiple alarms.
ON/OFF enables the alarm function if set at ON, or disables the alarm function if set at OFF. The H-limit value cannot be set below the L-limit value, and the L-limit value cannot be set above the H-limit value.
If it is desired to set the H-limit value below the L-limit value already stored in the memory, reduce the L-limit value beforehand, and vice versa.
A typical on-screen display for a High alarm is shown below.
When the H-limit alarm occurs, the “H-alarm” message comes on in the field of relevant channel. Similarly for L-alarm. There is no screen display for the HH-alarm”, and “LL-alarm” but otherwise, they function like the H and L limit alarms.
Teledyne Analytical Instruments 54
Page 55
Infrared Gas Analyzer Operation
CAUTION: ALARMS ARE INACTIVE FOR 10 MINUTES AFTER
TURNING ON POWER.
3.5.3.1
CONFIGURING THE ALARMS
Note: To change an alarm setting, first set the alarm ON/OFF
setting to OFF, and then change the value.
1. From the measurement mode, press the MODE key to
display the user mode.
Point the cursor to “Alarm Setting” by pressing the UP or
2. DOWN key and press the ENT key.
In the “Alarm Setting” screen that appears, point the
3. cursor to the alarm number you want to set (1 through 6) using the UP or DOWN key. Press the ENT key when the cursor is on the alarm number you desire.
Teledyne Analytical Instruments 55
Page 56
Operation Model 7600
4. The alarm item selection screen will appear where you
can select the alarm type, assign it to a channel, define setpoints, and enable or disable the alarm. Use the UP or DOWN keys until the cursor is aligned with the desired function and press the ENT key.
Note: Setpoints are adjustable form 0 to 100% of fullscale
however they must be defined so that H-limit value > L-limit value and that (H-limit value- L-limit value) > hysteresis.
After setting the last feature, store the values by pressing
5. ENT again.
To close the Alarm Setting screen or cancel this mode midway,
press the ESC key. The previous screen will return.
3.5.3.2 HYSTERESIS SETTING
The hysteresis setting is used prevent chattering of an alarm output near the alarm setpoint. The hysteresis value range is 0-20% of fullscale on any analysis range. Figure 3-6 shows the hysteresis width (in % fullscale) set for a high limit alarm.
Teledyne Analytical Instruments 56
Page 57
Infrared Gas Analyzer Operation
Figure 3-6: Hysteresis Example for a High Limit Alarm
An alarm output is turned ON if measurement value exceeds the upper limit value as shown in Figure 3-6. Once the alarm output has been turned ON, it is not turned OFF until the process value falls below the hysteresis width.
Note: The hysteresis setting is common to all alarms.
To set the hysteresis value:
1.
From the “Alarm No. Selection” screen, point the cursor
to “Hysteresis” by pressing the UP or DOWN key then press ENT.
In the “Hysteresis Value Setting” screen that appears,
2. enter hysteresis values using the UP or DOWN keys to increment or decrement a digit. Use the SIDE key to move to the next digit. When finished, press ENT.
Teledyne Analytical Instruments 57
Page 58
Operation Model 7600
To close the “Hysteresis Setting” or cancel the mode midway, press
the ESC key. A previous screen will return.
3.5.4 Setting of Auto Calibration
3.5.4.1 AUTO CALIBRATION
Auto calibration is automatically carried out when zero and span
calibrations are set.
In this menu you can set when and how frequent the auto calibration is performed. You also enable or disable the auto calibration feature here.
Description of setting items:
Start Time: Setting at the first calibration (day of the week, hour,
minute)
Cycle: A period between the start time of one calibration and
another (unit: hour/day)
Flow Time: The time required for replacement by calibration gas.
Time required for replacement of sample gas after the calibration is completed (Set by calibration gas.)
ON/OFF: Enable/disable the auto calibration feature.
Note: Before changing the settings of auto calibration, set the
ON/OFF to OFF.
To change the settings:
3.
From the measurement mode, press the MODE key to
display the user mode.
4.
Point the cursor to “Setting of Auto Calibration” by
pressing the UP or DOWN key and press the ENT key.
Teledyne Analytical Instruments 58
Page 59
Infrared Gas Analyzer Operation
5. In the “Setting of Auto Calibration” screen that appears,
point the cursor to any item you want to set by pressing
UP or DOWN key. Press the ENT key.
the
6. In the “Auto Calibration Parameter Setting” screen that
appears, perform the value entry or the setting. For the value entry or setting change, use the UP or DOWN key. To change the setting, use the SIDE key to move the cursor to the right.
After changing the setting, press the ENT key to accept
7. the change. Auto calibration will now be carried out using the entered setting value.
To close the “Setting of Auto Calibration” or cancel the mode
midway, press the ESC key. A previous screen will return.
3.5.4.2 GAS FLOW TIME SETTING
In this submenu you select the gas whose flow time is to be changed. The flow time range is from 60 to 900 seconds. The initial value is set for 300 seconds.
Teledyne Analytical Instruments 59
Page 60
Operation Model 7600
To set the flow time:
1.
Place the cursor next to “Flow Time,” then press the ENT
key.
On the flow time setting screen that appears, use the UP
2. and DOWN keys to move the cursor to the gas whose settings you want to change, then press ENT.
The highlighted value can be changed. Change the value
3. by pressing the UP or DOWN key, and then move the cursor to the right by pressing the SIDE key.
After changing the value, press the ENT key.
4.
5.
Press the ESC key to return to the automatic calibration
setting screen.
Note: Only the channels used are displayed on this screen. The
Ex. time is the output signal hold extension time after the completion of calibration. It is valid only when the hold setting is set to “ON.” The Ex. time set here is also used in a manual calibration.
The auto calibration contacts are closed during an auto calibration
event.
3.5.4.3 CYCLE SETTING RANGE
The cycle range is settable from 1 to 99 hours or 1 to 40 days. The
initial value is set at 7days.
Note: When an auto calibration starts, the measurement screen
automatically appears.
Teledyne Analytical Instruments 60
Page 61
Infrared Gas Analyzer Operation
Any operation other than “Stop Auto Calibration” (see
Section 3.5.4.5) is not permitted during auto calibration.
“Stop Auto Calibration” cannot be performed when the key
lock is ON.
To cancel auto calibration, set the key lock to OFF and
then execute “Stop Auto Calibration”.
Auto Calibration Example
Start Time SUN 12:00
Cycle 1 day
Flow Time Zero 350 sec
Ch1 Span 350 sec
Ch1 Span 350 sec
Ch1 Span 350 sec
Ch1 Span 300 sec
Ch1 Span 300 sec
EX. time 300 sec
ON/OFF ON
Figure 3-7 shows the auto calibration with the above settings.
Figure 3-7: Example Auto Calibration
Teledyne Analytical Instruments 61
Page 62
Operation Model 7600
When a channel enabled for auto calibration is in the calibration mode, a message indicating ZERO cal. or SPAN cal. will appear on the measurement screen. For example, if channels 1 and 2 have the auto calibration setting enabled, when the calibration event takes place, both channels 1 and 2 will indicate ZERO cal. when the zero calibration is being performed. Zero calibrations are performed first and all enabled channels will zero simultaneously.
When the zero calibrations are finished, channel 1 will enter the span calibration followed by channel 2.
3.5.4.4 REMOTE START
An auto calibration is always available by keeping the remote start input closed for at least 1.5 seconds. The remote start can initiate an auto calibration whether the parameter setting is ON or OFF.
3.5.4.5 FORCED RUN/STOP OF AUTO CALIBRATION
Auto calibration can be performed manually or forcibly stopped while the calibration is performed.
Teledyne Analytical Instruments 62
Page 63
Infrared Gas Analyzer Operation
3.5.4.5.1 MANUAL EXECUTION OF AUTO CALIBRATION
To perform the auto calibration cycle manually:
1.
From the measurement mode, press the MODE key to
display the user mode.
2.
Point the cursor to “Setting of Auto Calibration” by
pressing the UP or DOWN key and press the ENT key.
3.
In the “Setting of Auto Calibration” item selection screen
that appears, point the cursor to “Auto Calibration Run” by pressing the
UP or DOWN keys. Press the ENT
key.
“Run” is highlighted, displaying a message to confirm the
4. execution of auto calibration. Press the ENT key to execute the auto calibration, or press the ESC key to cancel.
3.5.4.5.2
FORCIBLY END AN AUTO CALIBRATION
To abort an auto calibration event:
Teledyne Analytical Instruments 63
Page 64
Operation Model 7600
5. From the measurement mode, press the MODE key to
display the user mode.
6.
Point the cursor to “Setting of Auto Calibration” by
pressing the UP or DOWN key and press the ENT key.
In the “Setting of Auto Calibration” item selection screen
7. that appears, point the cursor to “Auto Calibration Run”
by pressing the
UP or DOWN keys. Press the ENT
key.
“Stop” is highlighted, displaying a message to confirm
8. the stop of auto calibration. Press the ENT key to stop the auto calibration, or press the ESC key to cancel the abort.
Note: During auto calibration, the keys are inoperable other than
key lock ON/OFF and “Stop Auto Calibration.”
When the key lock is set to ON, even the “Auto Calibration
Stop” is locked out. To stop “Auto Calibration” forcedly, set the key lock to OFF and then execute “Auto Calibration Stop” as detailed above.
Teledyne Analytical Instruments 64
Page 65
Infrared Gas Analyzer Operation
3.5.5 Setting of Auto Zero Calibration
3.5.5.1 AUTO ZERO CALIBRATION
Auto zero calibration is automatically carried out when the zero calibration is set. Components for which a calibration is to be made are determined by setting of auto calibration component in Section 3.5.3.
Similar to the menu choices and operation of the auto calibration, in this menu you can set when and how frequent the auto zero calibration is performed. You also enable or disable the auto zero calibration feature here.
Description of setting items:
Start Time: Setting at the first calibration (day of the week, hour,
minute)
Cycle: A period between the start time of one calibration and
another (unit: hour/day)
Flow Time: The time required for replacement by calibration gas.
Time required for replacement of sample gas after the calibration is completed (Set by calibration gas.)
ON/OFF: Enable/disable the auto calibration feature.
Note: Before changing the settings of auto calibration, set the
ON/OFF to OFF.
To change the settings:
1.
From the measurement mode, press the MODE key to
display the user mode.
Point the cursor to “Setting of Auto Zero Calibration” by
2. pressing the UP or DOWN key and press the ENT key.
Teledyne Analytical Instruments 65
Page 66
Operation Model 7600
3. In the “Setting of Auto Zero Calibration” screen that
appears, point the cursor to any item you want to set by pressing the
UP or DOWN key. Press the ENT key.
4. In the “Auto Zero Calibration Parameter Setting” screen
that appears, perform the value entry or the setting. For the value entry or setting change, use the UP or DOWN key. To change the setting, use the SIDE key to move the cursor to the right.
After changing the setting, press the ENT key to accept
5. the change. Auto zero calibration will now be carried out using the entered setting value.
To close the “Setting of Auto Calibration” or cancel the mode
midway, press the ESC key. A previous screen will return.
The contacts for the auto zero calibration are closed during auto
zero calibration.
Auto Zero Calibration Example
Start Time SUN 12:00
Cycle 12 hour
Flow Time 300 sec
ON/OFF ON
Figure 3-8 shows the auto zero calibration with the above settings.
Teledyne Analytical Instruments 66
Page 67
Infrared Gas Analyzer Operation
Figure 3-8: Example Auto Zero Calibration
Note: When an auto zero calibration starts, the measurement
screen automatically appears.
Any operation other than “Stop Auto Zero Calibration” (see
Section 3.5.4.5 is not permitted during auto calibration.
“Stop Auto Zero Calibration” cannot be performed when
the key lock is ON.
To cancel auto zero calibration, set the key lock to OFF
and then execute “Stop Auto zero Calibration”. 0 If the auto calibration period and auto zero calibration period have overlapped, the auto calibration is retained, ignoring the auto zero calibration of that period.
If the auto calibration and the auto zero calibration periods
overlap, the auto calibration setting is retained. The instrument ignores the auto zero calibration period.
When the hold setting is set to ON, the hold time of auto
calibration contact and measurement value output signal is extended after calibration for gas replacement time.
Turn on the power again after it is turned off (including the
case of power failure) at the time set as the next start time in auto zero calibration, and then repeat it in the set cycle.
Teledyne Analytical Instruments 67
Page 68
Operation Model 7600
3.5.5.2 FORCED RUN/STOP OF AUTO ZERO CALIBRATION
Like the auto calibration, the auto zero calibration can be performed manually or forcibly stopped while the calibration is performed.
3.5.5.2.1 MANUAL EXECUTION OF AUTO ZERO CALIBRATION
To perform the auto calibration cycle manually:
1.
From the measurement mode, press the MODE key to
display the user mode.
2.
Point the cursor to “Setting of Auto Zero Calibration” by
pressing the UP or DOWN key and press the ENT key.
In the “Setting of Auto Zero Calibration” item selection
3. screen that appears, point the cursor to “Auto zero
Calibration Run” by pressing the
UP or DOWN keys.
Press the ENT key.
“Run” is highlighted, displaying a message to confirm the
4. execution of auto calibration. Press the ENT key to execute the auto calibration, or press the ESC key to cancel.
Teledyne Analytical Instruments 68
Page 69
Infrared Gas Analyzer Operation
3.5.5.2.2 FORCIBLY END AN AUTO ZERO CALIBRATION
To abort an auto zero calibration event:
1.
From the measurement mode, press the MODE key to
display the user mode.
Point the cursor to “Setting of Auto Calibration” by
2. pressing the UP or DOWN key and press the ENT key.
In the “Setting of Auto Calibration” item selection screen
3. that appears, point the cursor to “Auto Calibration Run”
by pressing the key.
“Stop” is highlighted, displaying a message to confirm
4. the stop of auto calibration. Press the ENT key to stop the auto calibration, or press the ESC key to cancel the abort.
UP or DOWN keys. Press the ENT
Teledyne Analytical Instruments 69
Page 70
Operation Model 7600
Note: During auto zero calibration, the keys are inoperable other
than key lock ON/OFF and “Stop Auto Zero Calibration.”
When the key lock is set to ON, even the “Auto Zero
Calibration Stop” is locked out. To stop “Auto Zero Calibration” forcedly, set the key lock to OFF and then execute “Auto Zero Calibration Stop” as detailed above.
3.5.6 Peak Alarm Setting
The optional peak alarm is triggered whenever the CO concentration exceeds the upper limit value for a preset number of times. The peak number is set from within the Peak Alarm Setting menu.
Note: Only instruments with the peak alarm option installed will
have this menu.
The following items are set within the peak alarm menu:
OFF/ON status of peak alarm
Alarm setpoint value
Alarm count (number of peaks)
Hysteresis
To adjust the peak alarm settings:
From the measurement mode, enter the user mode by
1. pressing the MODE key.
Point the cursor to “Setting of Peak Alarm” by pressing
2. the UP or DOWN key.
Teledyne Analytical Instruments 70
Page 71
Infrared Gas Analyzer Operation
Press the ENT key.
3.
4.
In the “Peak Alarm Setting” item selection screen that
appears, point the cursor to any item you want to set by pressing the UP or DOWN
key. Then press the ENT key.
5.
Enter numeric values by using the UP or DOWN keys.
Use the SIDE key to move the cursor to the right. Press the ENT key to save the change.
To close the “Peak Alarm Setting” or cancel the mode midway,
press the ESC key. A previous screen will return.
Description of setting items:
Peak Alarm: ON/OFF of peak alarm
Alarm Value: If measuring value exceeds the set alarm value in
ppm, a peak counter increments by 1.
Alarm Count: When a peak in excess of the alarm value occurs,
a peak count is incremented. The alarm count is
Teledyne Analytical Instruments 71
Page 72
Operation Model 7600
the numerical value which upon reaching will trigger the peak alarm.
Hysteresis: To prevent possible chattering when the
measuring value may exceed the set peak concentration by only 1 time, the peak count has an allowance in the hysteresis width.
Setting range:
Alarm value: 10 to 1000 ppm, 5 ppm step (initial value: 500
ppm)
Alarm count: 1 to 99 times (initial value: 5 times)
Hysteresis: 0 to 20 % of full scale (initial value: 0% of full
scale).
Action of peak alarm:
If CO concentration exceeds the alarm value, counting will begin. If the number of peaks is over the set times per hour, the peak alarm contact closes (ON). If it is less than the set times per hour, it is open (OFF).
Figure 3-9: Peak Alarm Example
Figure 3-9 depicts the behavior of the Peak alarm where the alarm count has been set to 5. At the point labeled (1), the peak count alarm is turned ON since the alarm count has reached the threshold of 5 per hour. Since peaks of more than 5 times within an hour occur between the
Teledyne Analytical Instruments 72
Page 73
Infrared Gas Analyzer Operation
interval (1) and (2), the peak count alarm remains ON. At (2), the peak count has fallen to 4 per hour so the peak alarm is turned OFF.
Like the hysteresis of the alarm setting , the hysteresis prevents possible chattering when measured gas is fluctuated near the alarm value.
CAUTION: FOR 10 MINUTES AFTER THE POWER IS TURNED
ON, A PEAK ALARM COUNTING IS NOT PERFORMED.
Releasing peak count alarm:
To release the peak count alarm, set the peak alarm to OFF. Turning the peak alarm back ON again resets the count to 0.
3.5.7 Parameter Setting
This menu allows you to adjust parameter settings such as:
Time
Key lock
Output hold
Reset average output
Response time
Average period
Backlight timer
Enter maintenance mode
Description of setting items:
Current Time: Current year, month, date, day of the week, hour, and minute setting in this order.
Note: The clock backup time is 2 days. If power is off for 2 days or
longer, the time setting will need to be set again.
Key Lock: Sets with ON/OFF. When set to ON, all key
operations except key lock OFF are disabled.
Output Hold: Sets whether calibration output is held or not, and
the holding value setting.
Teledyne Analytical Instruments 73
Page 74
Operation Model 7600
Reset Av. Output: Resets the average value.
Response time: Sets the response time of the electrical system.
Average Period: Sets the moving average time.
Backlight Timer: Sets an automatic OFF of the backlight of display
unit and the time until the backlight is turned OFF.
Maintenance mode: Enters passwords to switch to the Maintenance
mode.
For the maintenance mode, see Section 3.6.
To make parameter changes:
1.
From the measurement mode, enter the user mode by
pressing the MODE key.
Point the cursor to “Parameter Setting” by pressing the
2. UP and DOWN
keys. Then press ENT.
In the “Parameter Setting” screen that appears, point the
3. cursor to any item you want to set by pressing the UP or DOWN key. Then Press ENT.
Enter numeric values by using the UP or DOWN keys.
4. Use the SIDE key to move the cursor to the right. After setting, press the ENT key to save the parameter setting.
Teledyne Analytical Instruments 74
Page 75
Infrared Gas Analyzer Operation
To close the “Setting of Auto Calibration” or cancel the mode
midway, press the ESC key. A previous screen will return.
Settings range:
Hold setting: 0 to 100% FS
Response time: 1 to 60 sec. (Initial value: 15 sec)
Average period: 1 to 59 min or 1 to 4 hours (Initial value: 1 hour)
Backlight Timer: 1 to 60 min (Initial value: OFF)
Maintenance mode: 0000 to 9999 (Initial value: 0000)
3.5.7.1 OUTPUT HOLD
By setting an output hold to ON, the output signal of each channel is held during the calibration (manual calibration and auto calibration) plus the gas flow time as shown in Figures 3-10 through 3-12. Refer to Section
3.5.4,
Setting of Auto Calibration for a description of the gas flow time.
Regardless of Hold ON/OFF setting, an output signal can always be held via an external input.
Manual calibration
Figure 3-10: Output Hold for Manual Calibration
Teledyne Analytical Instruments 75
Page 76
Operation Model 7600
Auto calibration
Figure 3-11: Output Hold for Auto Calibration
External hold
Figure 3-12: Output Hold for External Hold
Screen display during hold
For Auto calibration and external holds, the “on Hold” message
will be shown blinking on the measurement screen.
In a manual calibration, the screen displays the process of calibration so the “on Hold” message is not displayed even if the output signal is held. However, “on Hold” message is displayed during the hold extending time shown in Figure 3-10.
For both manual and auto calibration, if the calibration is cancelled after the calibration gas has been directed to the analyzer, the hold extending time will be performed.
There are two types of hold values that you can use: “current” or “setting”. Current uses the existing value immediately before entering output hold, while “setting” uses an arbitrary value that you set before initiating the hold.
To change or set the value for a hold:
Teledyne Analytical Instruments 76
Page 77
Infrared Gas Analyzer Operation
1. From the measurement screen, press the MODE key to
enter the user mode then navigate to the Parameter Setting option. Press the ENT key to bring up the parameter setting screen.
Select the Output Hold option by moving the cursor to it
2. using the UP and Down keys. Select it by pressing ENT.
“ON” or “OFF” is highlighted. If you desire to change
3. the state, press the UP or DOWN
key to toggle between
ON and OFF. Press the ENT key to select the desired state.
4. To change the hold value, use the SIDE key to highlight
“Current” or “Setting”. Select “Setting” by pressing the UP or DOWN key until that option appears then press ENT. The screen changes to the hold setting screen where you can select the channel and hold value for that channel.
Teledyne Analytical Instruments 77
Page 78
Operation Model 7600
On the parameter hold screen that appears, move the
5. cursor next to the channel you want to make the setting by pressing the UP or DOWN key, and then press ENT.
6. The value is highlighted, indicating that the value can be
changed. Change the numerical value by pressing the
UP or DOWN key, and then move the cursor to the
right by pressing the SIDE key.
7.
After the value is changed, press the ENT key to save.
8. Press the ESC key to return to the parameter setting
screen.
The setting is expressed in % against the range for both ranges. For example, when 0 to 1000 ppm is selected as the range, and 10% FS is selected as hold setting, the output equivalent of 100 ppm is output and held irrespective of the measurement value at that time.
Description of setting:
Instantaneous value display of the measurement cannot be held. Only the output can be held.
If set value is selected for hold, instantaneous O
correction value is
2
calculated and held based on the set value.
Range identification contact output cannot be switched even if the range is switched during the hold.
3.5.7.2 AVERAGE VALUE RESET:
This mode is used to clear all average values O2 correction average and O
average, and restarts averaging. All average values are reset at
2
the same time. The indication value and output value is 0 ppm, vol % etc. at the time of the reset input (Refer to the average period). To
Teledyne Analytical Instruments 78
Page 79
Infrared Gas Analyzer Operation
perform a reset, the contacts must close for at least 1.5 seconds. See Figure 3-13. Resetting occurs as the contacts reopen.
Figure 3-13: Average Value Reset
3.5.7.3 RESPONSE TIME
The response time of the electrical system can be changed.
Individual settings are available for each component.
Note: Response time does not provide exact seconds for the
setting time, but rather it is a guide to the overall response.
3.5.7.4 AVERAGE PERIOD
This submenu allows you to set an averaging period over which the
average value of O
correction and O2 average are sampled.
2
Teledyne Analytical Instruments 79
Page 80
Operation Model 7600
To modify the average period:
1.
Choose the channel or channels you want to modify. You
can set an average time from 1 to 59 minutes in 1-minute steps or 1 to 4 hours in 1-hour steps.
The UP and DOWN keys will increment or decrement
2. the numerical values while the SIDE key will move the cursor to the right to the next digit. Press ENT when the desired value is displayed.
Changing this parameter resets the average value of O
and O
average value. Pressing the ENT key validates the resetting only
2
correction
2
for components whose setting was changed.
Example of average action:
In the example shown in Figure 3-14, the average period was set to
1 hour.
Figure 3-14: Example Average Period
Sampling occurs every 30 seconds so every 30 seconds the average
for the last 1 hour (time setting) is output.
At the instant of resetting, zero is assumed for all past values. This means that the average value will be incorrect for 1 hour (1 time period) after resetting.
3.5.7.5 BACKLIGHT TIMER
Automatic OFF setting of the backlight of the LCD unit can be adjusted from this menu.
Upon returning to the measurement screen, when the specified time elapses, the backlight is automatically turned off. During operation,
Teledyne Analytical Instruments 80
Page 81
Infrared Gas Analyzer Operation
pressing any key will turn the backlight back ON for the specified duration.
When ON is selected, the time until auto OFF is displayed. To
change the backlight automatic shut off time:
Select Parameter Setting from the user screen.
1.
2.
With the UP or DOWN key, select Backlight Timer. and
press ENT.
3.
If the timer is not set to ON, change its state by pressing
the ENT key when OFF is highlighted.
Press the SIDE key to move the cursor to the time value
4. and the time setting can be changed using the UP and DOWN keys. Press the ENT key to confirm the selection.
Note: When OFF is selected, the backlight remains ON and is
not turned off automatically.
3.5.7.6 MAINTENANCE MODE
The maintenance mode is covered in section 3.6. It is a detailed password protected set of screens that allows you to check sensor input values, review error logs, set passwords and more.
The default password setting for entering this mode is “0000.” Use this password to initially enter the maintenance mode and then change it to one you can easily remember.
To enter the maintenance mode:
From the measurement screen press the MODE key to
1. open the user mode screen.
Teledyne Analytical Instruments 81
Page 82
Operation Model 7600
2. Select Parameter Setting from the user screen.
3.
With the UP or DOWN key, select Maintenance Mode
and press ENT.
3.6 Maintenance Mode
This mode is used for checking sensor input values, error log file display, or setting of passwords, etc. A password is required to enter this mode.
To get to the maintenance mode screen, select Maintenance Mode from the Parameter Setting menu. See Section 3.5.7 Parameter Setting.” This brings up a screen for entering the password. Enter the 4 digit password using the UP and DOWN keys and the SIDE key to move to the next digit. Press ENT when complete. If the password is entered correctly, the maintenance mode screen will appear on the display offering a range of submenus.
Note: The default password installed in new instruments is
“0000”.
The following submenus are accessed from the maintenance mode screen:
Sensor Input
1.
2.
Error Log
3.
Calibration Log
4.
Optical Adjustment
5.
Interference Compensation Adjustment
6.
Output Adjustment
Teledyne Analytical Instruments 82
Page 83
Infrared Gas Analyzer Operation
7. Additional Parameters
a.
Password Set
b.
O
Reference Value
2
c.
Limit
d.
MODBUS Communication Station No.
e.
Range Setting
8.
Factory Mode
In the maintenance mode screen, point the cursor to the item you want to set by pressing the UP or DOWN key and press the ENT key. The appropriate screen will automatically display.
Note: “To Factory Mode” is used for factory service engineers.
Do not use this mode.
3.6.1 Sensor Input Value
Upon selecting Sensor Input Value, the following screen will appear. It lists the installed sensors along with the input value from the measuring sensor (M) and the compensating detector (C).
Where:
NOx M: NOx sensor input value
NOx C: NOx interference compensation sensor input value
SO
M: SO2 sensor input value
2
SO
C: SO2 interference compensation sensor input value
2
CO
M: CO2 sensor input value
2
CO
C: CO2 interference compensation sensor input value
2
Teledyne Analytical Instruments 83
Page 84
Operation Model 7600
CO M: CO sensor input value
CO C: CO interference compensation sensor input value
Temperature: temperature sensor input value O
: O2 sensor input
2
value
3.6.2 Error Log
This screen gives the error history for the last fourteen logged errors. It also shows the date and time of occurrence (year, month day, hour, minute) and the channel the error occurred on. For error number, date and time of occurrence, channel and other details of error, refer to Section 4.6
Error Messages.
To clear the error log, select Clear Error Log and press the ENT key.
3.6.3 Calibration Log
The next screens detail the past calibration history on the analyzer. The first screen allows you to choose the channel whose information you want displayed. The next screen displays logged information on sensor input value, concentration value, and the dates when zero/span calibration were performed for that channel. The last 10 calibration events are logged for each channel.
The first screen also gives you the opportunity to clear the log. To erase the log, move the cursor to Clear Calibration Log and press the ENT key. The calibration log is cleared completely.
Teledyne Analytical Instruments 84
Page 85
Infrared Gas Analyzer Operation
Z1: Zero calibration (Z) of Range 1
S1: Span calibration (S) of Range 1
M: Value of measuring detector at the time of calibration
C: Value of the interference compensation detector at the time of calibration
Con: Concentration value displayed before calibration
3.6.4 Optical Adjustment
CAUTION: IF THE OPTICAL ADJUSTMENT IS MADE
INCORRECTLY, ANALYSIS MAY BE ADVERSELY AFFECTED. CARRY OUT THE OPERATION WITH UTMOST ATTENTION.
For details of this item, refer to Section4.4 Optical Zero
Adjustment.
Press ENT key and then turn ON the solenoid valve signal for each
calibration gas by using the UP or DOWN keys.
Teledyne Analytical Instruments 85
Page 86
Operation Model 7600
3.6.5 Moisture Interference Adjustment
This item is described in Section 4.5 Moisture Interference
Adjustment.
For the values listed in the middle column of the screen, the moisture interference for each component is already offset. The values in the right column are the interference compensation coefficients used.
To change the coefficient:
1.
Move the
cursor to a desired channel by pressing the UP or DOWN key, and then press the ENT key. The selected interference compensation coefficient at the right is highlighted.
2.
Check that the gas for moisture interference
compensation is flowing. Change the moisture interference compensation coefficient using the UP or DOWN key. Adjust the value until the value in the left column is near zero. Then press the ENT key to log moisture interference compensation value.
Note: An interference compensation detector is not provided so if
the 1st range is beyond 0 to 10 vol %, no interference adjustment can be performed (no need).
3.6.6 Output Adjustment Screen
In this menu you can adjust the analog output. First, connect a digital multi meter to the output terminal corresponding to the output to be adjusted. See the Appendix for terminal identification on the terminal blocks. Adjust the value so that 4 mA or 0 V is output at zero and 20 mA or 1 V is output at span.
Teledyne Analytical Instruments 86
Page 87
Infrared Gas Analyzer Operation
1. From the output adjustment screen which accessible from
the maintenance mode menu, move the cursor with the UP or DOWN key to the output (OUT No. and zero/ span) to be adjusted, and then press the ENT key.
The selected value is highlighted. Adjust the value using
2. the UP or DOWN key while watching the output on the digital multi-meter. Press the SIDE key to select the next digit.
On completion of the adjustment, press the ENT key.
3.
3.6.7 Other Parameters
The last submenu of the maintenance mode menu contains a group
of parameters that may be modified by the user. These include:
Password Set: Set the password used to enter the
maintenance mode. An arbitrary 4-digit number can be selected.
O
Ref. Value: Set the oxygen concentration reference
2
value at the time of oxygen correction calculation. Settable in the range from 00 to 19%.
Limit: Set the oxygen concentration limit at the
time of oxygen correction calculation. Settable in the range from 01 to 20%. See O
correction concentration value in
2
Section 3.3.1
Measurement Mode for
oxygen correction calculation procedure.
Teledyne Analytical Instruments 87
Page 88
Operation Model 7600
Station No.: Set the station No. for MODBUS
communication. Settable in the range from 00 to 32.
Range setting: Moves to the screen on which measuring
range is changed.
To adjust or make changes to these parameters:
1.
Press the UP or DOWN key to move the cursor to the
item whose setting is to be changed.
The values for password, oxygen correction, limit, and
2. station no. are highlighted.
Press the UP or DOWN key to change the value and then
3. press the ENT key.
Note: If you forget the password you will be locked out from the
maintenance mode. Keep a record of the password in a secure place. Contact the factory for instruction on reclaiming forgotten or lost passwords.
To set or change the range:
The measuring range can be arbitrarily selected in the minimum and the maximum range specified at the time of purchase. The range to be used can be selected as 1 or 2.
1.
Move the cursor to the Range Setting option using the UP
or DOWN key, and then press the ENT key.
This brings up another screen that lists the channels and
2. components. Move the cursor to the channel whose setting is to be changed by pressing the UP or DOWN key, and then press the ENT key.
Teledyne Analytical Instruments 88
Page 89
Infrared Gas Analyzer Operation
Another screen appears which displays the current MIN
3. and MAX ranges established for that component. To change the existing range move the cursor to the range you want to modify using the UP or DOWN key, and then press the ENT key.
Note: The value for range 1 and range 2 must fall within or be
equal to the MIN the MAX range and range 1 must be less than range 2.
The number of ranges (Range num.) is 1 or 2.
Press the UP or DOWN key to change the value. Press
4. the SIDE key to select the next digit. When the decimal point is highlighted, press the UP or DOWN key, and the decimal point position can be changed.
5.
When complete, press the ENT key to save the change.
Teledyne Analytical Instruments 89
Page 90
Operation Model 7600
3.7 Calibration
3.7.1 Zero Calibration
The zero calibration is used for zero point adjustment. A suitable
zero calibration gas is described in Section 2.5.5
Calibration Gas.
To zero calibrate the analyzer:
1.
Press the ZERO key on the Measurement screen to
display the Manual Zero Calibration screen.
Preparation of
Select the channel to be calibrated by pressing the UP or
2. DOWN key. After selection, press the ENT key, and zero gas will be supplied.
Note: For channels that are set to “both” in the “Zero Calibration”
of the Calibration Setting mode, the zero calibrations are carried out at the same time.
3. Wait until the display has stabilized with the zero gas
flowing. After stabilization, press the ENT key. At this point, the zero calibration on the range indicated by the cursor is carried out.
Note: If the selected channel was set for autoranging “AR” (see
Section 3.5.1 Switch Ranges. the cursor automatically moves to the range selected in “Setting of Auto Calibration Component/ Range” (Section 3.5.2.4), and calibration is carried out within that range.
Teledyne Analytical Instruments 90
Page 91
Infrared Gas Analyzer Operation
To close the “Zero Calibration” screen or cancel this mode
midway, press the ESC key. A previous screen will return.
3.7.2 Span Calibration
Span calibration is used to perform a span point adjustment. It requires a span gas with a known concentration, preferably 90% or more of the range value of the component being calibrated. For oxygen, when measuring with the built-in O
use a standard gas containing 90% or
2
more of the range value. Use a standard gas of about 2 vol % when measuring with an external zirconia O
sensor.
2
To span calibrate the analyzer:
1.
From the measurement screen press the SPAN key to
display the Manual Span Calibration screen.
Select the channel to be calibrated by pressing the UP or
2. DOWN key. Then press the ENT key and the calibration gas will be supplied.
Note: For channels that are set to “both” in the “Span Calibration”
of the Calibration Setting mode, the span calibrations will be carried out for each component.
3. Wait until the display has stabilized with the span gas
flowing. After stabilization, press the ENT key. Span calibration on the range selected by the cursor is performed.
Note: If the selected channel was set for autoranging “AR” (see
Section 3.5.1 Switch Ranges. the cursor automatically moves to the range selected in “Setting of Auto Calibration Component/ Range” (Section 3.5.2.4), and calibration is carried out within that range.
Teledyne Analytical Instruments 91
Page 92
Operation Model 7600
To close "Span Calibration"
To close the “Span Calibration” screen or cancel this mode
midway, press the ESC key. A previous screen will return.
Teledyne Analytical Instruments 92
Page 93
Infrared Gas Analyzer Maintenance
Maintenance
4.1 Routine Maintenance
Aside from normal cleaning of the instrument and making sure connections are tight and secure, the following maintenance items should be carried out on a daily basis.
1.
Zero and Span Calibration:
Perform a zero calibration. For the calibration procedures, refer to Section 3.7.1
Then, perform span calibration. For the calibration procedures, refer to Section 3.7.2
Zero Calibration.
Span Calibration.
Flowrate Check:
2. Sampling and purge gas flowrates are as follows:
Sampling gas flow: 0.5L/min 0.2 L/min Purge gas flow: About 1 L/min
Check and adjust if necessary every day.
4.2 Daily Check and Maintenance Procedures
Table 4.1 Troubleshooting and Maintenance
Item to be
checked
Display value Display values are
Phenomena Cause Remedy
low. Display values are
high.
(1) Dust is mixed in sampling cell.
(1) Clean the sampling cell.
Check sampling system especially gas filter.
Teledyne Analytical Instruments 93
Page 94
Maintenance Model 7600
Item to be
checked
(2) Air is
Purge gas flow is included when purge gas in sampling gas flow rate.
Zero point of gas analyzer
Span point of gas analyzer
Phenomena Cause Remedy
absorbed midway in the sampling pipe.
Standard flow is beyond the specified flow rate of 0.5 L/min,
0.3 to 0.7 L/min.
It is deflected. Adjust.
It is deflected. Adjust.
Adjust using
4.3 Analyzer Maintenance
4.3.1 Sample Cell Cleaning
(2) Find cause of leak and repair.
needle valve on flowmeter.
In rare circumstances will the sample cell require cleaning. The procedure for dismantling and cleaning the sample cell should be handled by qualified personnel only. In general, if it is absolutely required, contact the factory.
The recommended procedure for cleaning the sample cell is as
follows:
1.
Turn off the power switch, stop the sample gas, and
allow the zero gas to flow for several minutes to purge the cell interior.
Loosen the setscrew (2 pieces) from the top cover and
2. remove it as shown in Figure 4-1.
Remove the internal gas inlet tube.
3.
4.
Loosen both right and left screws for cell holding plate
and remove only the sample cell. See Figure 4-2.
Turn the sample cell window to the left and remove it
5. from the sample cell. See Figures 4-3 and 4-4.
Teledyne Analytical Instruments 94
Page 95
Infrared Gas Analyzer Maintenance
6. For cleaning the window and inside surface of the cell,
first eliminate coarse dust using a soft brush and then wipe with a soft rag.
CAUTION: THE WINDOW IS EASILY SCRATCHED. USE
EXTREME CARE WHEN PERFORMING THIS OPERATION SO AS NOT TO DAMAGE IT.
Figure 4-1: Top Cover Removal
Figure 4-2: Sample Cell Removal
Teledyne Analytical Instruments 95
Page 96
Maintenance Model 7600
Figure 4-3: Sample Cell
Teledyne Analytical Instruments 96
Page 97
Infrared Gas Analyzer Maintenance
Figure 4-4: IR Window Assembly
Note: If the window or the cell interior is very dirty, gently wipe it
using a soft cloth moistened with absolute alcohol. A slightly corroded infrared transmission window/sample cell can be remedied by gentle rubbing with chromium oxide powder on a cleani cloth but if excessively corroded, it must be replaced. Do not use excessive force when cleaning the window.
Reinstall the cell after cleaning. Be sure to perform an optical zero adjustment (Section 4.4) and moisture interference compensation adjustment (Section 4.5) prior to analysis.
4.3.2 Cleaning the Cell Block
1. Turn off the power switch, stop the sample gas, and allow
the zero gas to flow for several minutes to purge the cell interior.
2.
Loosen the 2 setscrews that secure the top cover and
remove it.
Remove the internal gas inlet tube.
3.
4.
Loosen the 2 detector set bolts. See Figure 4-5.
Note: The gas distribution block, cell block, and detector are
fastened by the same bolts.
Teledyne Analytical Instruments 97
Page 98
Maintenance Model 7600
Figure 4-5: Detector Secure Nut
5.
Using the furnished cell mounting tool, turn the window
fixture to the left and remove it from the cell. See Figure 4-6.
6.
For cleaning the infrared transmission window and cell
inside surface, first eliminate coarse dust using a soft brush and then wipe clean with a soft rag.
CAUTION: THE WINDOW IS EASILY SCRATCHED. USE
EXTREME CARE WHEN PERFORMING THIS OPERATION SO AS NOT TO DAMAGE IT.
7. After cleaning, reinstall the sample cell and perform an
optical zero adjustment (see Section 4.4) and moisture interference compensation adjustment (see Section 4.5).
Teledyne Analytical Instruments 98
Page 99
Infrared Gas Analyzer Maintenance
Figure 4-6: Cell Block Assembly
4.4 Optical Zero Adjustment
If the following operation is not performed correctly,
subsequent analysis may adversely be affected. This operation should be performed by a trained technician or the unit sent back to the factory for service.
The following adjustment is performed at reassembly after removing and cleaning the sample cell.
1.
Remove the top cover.
2.
Allow dry N
gas inlet until the reading stabilizes. The sample gas is
Teledyne Analytical Instruments 99
or air to flow through the analyzer sample
2
Page 100
Maintenance Model 7600
introduced directly to the inlet of analyzer unit through the gas cylinder.
3.
Proceed to an optical adjustment as described in the
maintenance mode. The display on the operation panel of the main unit is illustrated in Figure 4-7. Balance adjust­ment is not required if the display falls within ±100.
Figure 4-7: Optical Adjustment Display for Dual Optical System
Option
Table 4-2: Components of Optical Adjustment Screen
No. Of Components
to be measured
1-component meter Main Comp – – – – – –
NO/SO2
2-
component
meter
3-component meter
NO/SO2/CO
4-component meter NO/SO2/CO2/CO
CO2/CO
NO/CO
a b c d e f g h
NO
Main
CO2
Main
NO
Main
NO
Main
NO
Main
NO
Comp
CO2
Comp
NO
Comp
NO
Comp
NO
Comp
SO2
Main
CO
Main
SO2
Main
SO2
Main
SO2
Comp
CO
Comp
– –
SO2
Comp
SO2
Comp
– – – –
CO
Main
CO
Main
CO2
Main
CO
Comp
CO
Comp
– –
– –
CO
Main
CO
Comp
Teledyne Analytical Instruments 100
Loading...