Teledyne 6600 User Manual

Oil in Water Analyzer
OPERATING INSTRUCTIONS
Model 6600
Oil in Water Analyzer
DANGER
HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM. PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM. HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST FOR A
TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED. ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING. BEFORE
CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/MANAGER.
Teledyne Analytical Instruments
P/N M71055
12/22/99
i
Model 6600
Copyright © 1999 Teledyne Analytical Instruments
All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in part, in any form or by any means, whether it be electronic, mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of Teledyne Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91749-
1580.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us free from defects of material and of construction, and that our liability shall be limited to replacing or repairing at our factory (without charge, except for transportation), or at customer plant at our option, any material or construction in which defects become apparent within one year from the date of shipment, except in cases where quotations or acknowledgements provide for a shorter period. Components manufactured by others bear the warranty of their manufacturer. This warranty does not cover defects caused by wear, accident, misuse, neglect or repairs other than those performed by Teledyne or an autho­rized service center. We assume no liability for direct or indirect damages of any kind and the purchaser by the acceptance of the equipment will assume all liability for any damage which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our apparatus, and to make any alterations in the dimensions, shape or weight of any parts, in so far as such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its user, and serves as a tool by which valuable data can be gathered. The information provided by the instrument may assist the user in eliminating potential hazards caused by his process; however, it is essential that all personnel involved in the use of the instrument or its interface, with the process being measured, be properly trained in the process itself, as well as all instrumen­tation related to it.
The safety of personnel is ultimately the responsibility of those who control process conditions. While this instrument may be able to provide early warning of imminent danger, it has no control over process conditions, and it can be misused. In particular, any alarm or control systems installed must be tested and understood, both as to how they operate and as to how they can be defeated. Any safeguards required such as locks, labels, or redundancy, must be provided by the user or specifically requested of Teledyne at the time the order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The purchaser is responsible for the training of personnel, for providing hazard warning methods and instrumentation per the appropriate standards, and for ensuring that hazard warning devices and instrumentation are maintained and operated properly.
Teledyne Analytical Instruments, the manufacturer of this instrument, cannot accept responsibility for conditions beyond its knowledge and control. No statement expressed or implied by this document or any information disseminated by the manufac­turer or its agents, is to be construed as a warranty of adequate safety control under the
user’s process conditions.
ii
Teledyne Analytical Instruments
Oil in Water Analyzer
Table of Contents
Part I: Control Section .................................Part I
Part II: Analysis Section .............................Part II
Part III: Oil in Water Sample System......... Part III
Appendix, Generic Info...................................A-1
Teledyne Analytical Instruments
iii
Model 6600
iv
Teledyne Analytical Instruments
Part I: Control Section
OPERATING INSTRUCTIONS
Model 6600
Oil in Water Analyzer
Part I: Control Section
of the Control/Analysis Unit
Z-PURGED
CLASS I, DIVISION II, GROUPS B, C, and D
Teledyne Analytical Instruments
Part I: i
Model 6600 Oil in Water Analyzer
Table of Contents
1 Introduction
1.1 Overview........................................................................ 1-1
1.2 Typical Applications ....................................................... 1-1
1.3 Main Features of the Analyzer ....................................... 1-1
1.4 Operator Interface .......................................................... 1-2
1.4.1 UP/DOWN Switch ................................................ 1-3
1.4.2 ESCAPE/ENTER Switch ..................................... 1-3
1.5 Control Unit Interface Panel ........................................... 1-6
2 Installation
2.1 Unpacking the Control Unit............................................ 2-1
2.2 Electrical Connections ................................................... 2-1
2.3 Testing the System......................................................... 2-9
3 Operation
3.1 Introduction .................................................................... 3-1
3.2 Using the Controls ......................................................... 3-1
3.2.1 Mode/Function Selection ....................................... 3-2
3.3 The System Function ..................................................... 3-4
3.3.1 Setting up an Auto-Cal........................................... 3-5
3.3.2 Password Protection .............................................. 3-6
3.3.2.1 Entering the Password................................... 3-6
3.3.2.2 Installing or Changing the Password ............. 3-7
3.3.3 Logging Out ........................................................... 3-8
3.3.4 System Self-Diagnostic Test .................................. 3-9
3.3.5 The Model Screen ................................................. 3-10
3.3.6 Checking Linearity with Algorithm ......................... 3-10
3.3.7 Digital Flter Setup .................................................. 3-11
3.3.8 Homogenizer Function Setup ................................ 3-12
3.3.9 Hold/Track Setup ................................................... 3-13
3.3.10 Calibration/Hold Timer Setup................................. 3-14
ii: Part I
Teledyne Analytical Instruments
Part I: Control Section
3.3.11 Analog 4 to 20 mA Output Calibration.................... 3-14
3.3.12 Manual Control of Filter & Solenoids ..................... 3-15
3.4 The Zero and Span Functions ....................................... 3-16
3.4.1 Zero Cal ................................................................. 3-17
3.4.1.1 Auto Mode Zeroing ........................................ 3-17
3.4.1.2 Manual Mode Zeroing.................................... 3-18
3.4.1.3 Zero Offset Calibration................................... 3-19
3.4.1.4 Cell Failure .................................................... 3-19
3.4.2 Span Cal................................................................ 3-21
3.4.2.1 Auto Mode Spanning ..................................... 3-21
4.4.2.2 Manual Mode Spanning................................. 3-21
3.5 The Alarms Function...................................................... 3-22
3.6 The Range Select Function ........................................... 3-24
3.6.1 Manual (Select/Define Range) Screen .................. 3-25
3.6.2 Auto Screen ........................................................... 3-25
3.6.3 Precautions............................................................ 3-26
3.7 The Analyze Function .................................................... 3-27
3.8 Programming ................................................................. 3-28
3.8.1 The Set Range Screen .......................................... 3.29
3.8.2 The Curve Algorithm Screen ................................. 3-30
3.8.2.1 Manual Mode Linearization ........................... 3-30
3.8.2.2 Auto Mode Linearization................................ 3-30
4 Maintenance
4.1 Fuse Replacement ......................................................... 4-1
4.2 System Self Diagnostic Test........................................... 4-2
4.3 Major Internal Components............................................ 4-3
A Appendix
Model 6600 Specifications..................................................... A-3
Teledyne Analytical Instruments
Part I: iii
Model 6600 Oil in Water Analyzer
iv: Part I
Teledyne Analytical Instruments
Oil in Water Analyzer Part I: Control Section
Introduction
1.1 Overview
The Teledyne Analytical Instruments Model 6600 Control Section, together with a 6600 Analysis Section, is versatile microprocessor-based instrument.
Part I, of this manual covers the Model 6600 General Purpose, Bulk­head Mount Control Section. (The Analysis Section is covered in Part II of this manual, and Oil in Water application is covered in Part III) The Control Section is for indoor/outdoor use in hazardous environment only. The Analysis Section (or Remote Section) can be designed for a variety of hazardous environments. All Sections are mounted in a NEMA-4 enclosure (24”x20”x10”).
1.2 Typical Applications when Configured with the appropriate Sample System
A few typical applications of the Model 6600 are:
Offshore platforms, Produced Water, Sea Water
Waste Water, Tank Farms, Fuel Depots, Refinery Effluents
Oil Chemical Separators
On-Board Ship
Boiler Return Steam Condensate
Process Cooling Water
Bilge/Deballast Water Treatment
Water Soluble Oils
All Aromatic Hydrocarbons
Many Organic Hydrocarbons (Contact Factory)
Teledyne Analytical Instruments
Part I: 1-1
1 Introduction Model 6600
1.3 Main Features of the Analyzer
The Model 6600 Photometric Analyzer is sophisticated yet simple to
use. The main features of the analyzer include:
A 2-line alphanumeric display screen, driven by microprocessor electronics, that continuously prompts and informs the operator.
High resolution, accurate readings of concentration from low ppm levels through to 100%. Large, bright, meter readout.
Versatile analysis over a wide range of applications.
Microprocessor based electronics: 8-bit CMOS microprocessor with 32 kB RAM and 128 kB ROM.
Three user definable output ranges (from 0-1 ppm through 0-100 %) allow best match to users process and equipment.
Calibration range for convenient zeroing or spanning.
Auto Ranging allows analyzer to automatically select the proper preset range for a given measurement. Manual override allows the user to lock onto a specific range of interest.
Two adjustable concentration alarms and a system failure alarm.
Extensive self-diagnostic testing, at startup and on demand, with continuous power-supply monitoring.
RS-232 serial digital port for use with a computer or other digital communication device.
Analog outputs for concentration and range identification. (0-1 V dc standard, and isolated 4–20 mA dc)
Superior accuracy.
Internal calibration-Manual or Automatic (optional).
1.4 Operator Interface
All controls and displays on the standard 6600 are accessible from outside the housing. The instrument has two simple operator controls. The operator has constant feedback from the instrument through an alphanumeric display, and a digital LED meter. The displays and controls are described briefly here and in greater detail in chapter 3. See Figure 1-1.
1-2: Part I
Teledyne Analytical Instruments
Oil in Water Analyzer Part I: Control Section
1.4.1 UP/DOWN Switch
The UP/DOWN switch is used to select between any subfunctions displayed on the VFD screen such as in the main menue, the system menue, the Alarm menue, etc. When modifiable values are displayed on the VFD, the UP/DOWN switch can be used to increment or decrement the values.
1.4.2 ESCAPE/ENTER Switch
The ESCAPE/ENTER switch is used to input the data, to enter a function, or to exit a function displayed in the alphanumeric display:
Escape Moves VFD display back to the previous screen in a
series. If none remains, returns to Analyze mode screen.
Enter Within a menue: the funtion selected is entered moving on
to the next screen in a series.
WARNING:
The power cable must be disconnected to fully remove power from the instrument.
With Value selected: Enters the value into the analyzer as data. Advances cursor on VFD to the next operation.
In the Analyze mode: it calls the main menue. Functions called out by the main menue:
-System This function is a menu that calls a number
of functions that regulate the analyzer operation.
-Span This function spans the instrument.
-Zero This function zeros the instrument.
-Alarms This functions sets the alarm preferences.
-Range This function selects whether analyzer is
autoranging or locked on one range.
-Standby Places the analyzer in a sleep mode.
Teledyne Analytical Instruments
Part I: 1-3
1 Introduction Model 6600
Figure 1-1: Model 6600 Controls, Indicators, and Connectors
.
Digital Meter Display: The meter display is a Light Emitting Diode
LED device that produces large, bright, 7-segment numbers that are legible in any lighting. It is accurate across all analysis ranges. The 6600 models produce continuous readout from 0-10,000 ppm and then switch to continuous percent readout from 1-100 %.
1-4: Part I
Teledyne Analytical Instruments
Oil in Water Analyzer Part I: Control Section
Figure 1-2: Model 6600 Interface Panel
Teledyne Analytical Instruments
Part I: 1-5
1 Introduction Model 6600
Alphanumeric Interface Screen: The backlit VFD screen is an easy-
to-use interface between operator and analyzer. It displays values, options, and messages for immediate feedback to the operator.
1.5 Control Section Interface Panel
The Control Section interface panel, shown in Figure 1-2, contains the electrical terminal blocks for external inputs and outputs. The input/output functions are described briefly here and in detail in the Installation chapter of this manual.
Power Connection AC power source, 115VAC, 50/60 Hz
Analog Outputs 0-1 V dc concentration and 0-1 V dc
range ID. Isolated 4-20 mA dc and 4-20 mA dc range ID.
Alarm Connections 2 concentration alarms and 1 system
alarm.
RS-232 Port Serial digital concentration signal output
and control input.
Remote Bench Provides all electrical interconnect to the
Analysis Section.
Remote Span/Zero Digital inputs allow external control of
analyzer calibration.
Calibration Contact To notify external equipment that
instrument is being calibrated and readings are not monitoring sample.
Range ID Contacts Four separate, dedicated, range relay
contacts.
Network I/O Serial digital communications for local
network access. For future expansion. Not implemented at this printing.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the Model 6600 is accurate to 2-3 %. Accordingly, internally sched­uled calibrations can vary 2-3 % per day.
1-6: Part I
Teledyne Analytical Instruments
Oil in Water Analyzer Part I: Control Section
Installation
Installation of Model 6600 Analyzers includes:
1. Unpacking, mounting, and interconnecting the Control/Analysis Section
2. Making gas connections to the system
3. Making electrical connections to the system
4. Testing the system.
This chapter covers installation of the Control Section. (Installation of the Analysis Section is covered in Part II of this manual.) The Oil in Water application is covered in Part III.
2.1 Unpacking the Control/Analysis Unit
The analyzer is shipped with all the materials you need to install and prepare the system for operation. Carefully unpack the Control/Analysis Unit and inspect it for damage. Immediately report any damage to the ship­ping agent. Figure 2-2: Required Front Door Clearance
Allow clearance for the door to open in a 90-degree arc of radius 15.5 inches. See Figure 2-2.
Figure 2-2: Required Front Door
Clearance
1
5
.
5
2.2 Electrical Connections
Figure 2-3 shows the Control/Analysis Unit interface panel. Connec­tions for power, communications, and both digital and analog signal outputs are described in the following paragraphs. Wire size and maximum length data appear in the Drawings at the back of this manual.
Teledyne Analytical Instruments
Part I: 2-1
2 Installation Model 6600
Figure 2-3: Interface Panel of the Model 6600 Control Section
For safe connections, ensure that no uninsulated wire extends
outside of the terminal blocks. Stripped wire ends must insert completely
into terminal blocks. No uninsulated wiring should come in contact with fingers, tools or clothing during normal operation.
Primary Input Power: The power supply in the Model 6600 will
accept a 115 Vac, 50/60 Hz power source. See Figure 2-4 for detailed connections.
DANGER: Power is applied to the instrument's circuitry as
long as the instrument is connected to the power source. The standby function switches power on or off to the displays and outputs only.
115VAC
2-2: Part I
Figure 2-4: Primary Input Power Connections
Teledyne Analytical Instruments
Oil in Water Analyzer Part I: Control Section
Fuse Installation: The fuse holders accept 5 x 20 mm, 4.0 A, T
type (slow blow) fuses. Fuses are not installed at the factory. Be sure to install the proper fuse as part of installation (See Fuse Replacement in chapter 4, maintenance.)
Analog Outputs: There are eight DC output signal connectors on
the ANALOG OUTPUTS terminal block. There are two connectors per output with the polarity noted. See Figure 2-5.
The outputs are: 0–1 V dc % of Range: Voltage rises linearly with increasing sample con-
centration, from 0 V at 0% to 1 V at 100%. (Full scale = 100% programmed range.)
0–1 V dc Range ID: 0.25 V = Range 1, 0.5 V = Range 2, 0.75 V =
Range 3.
4–20 mA dc % Range: (-M Option) Current increases linearly with increas-
ing sample concentration, from 4 mA at 0% to 20 mA at full scale 100%. (Full scale = 100% of programmed range.)
4–20 mA dc Range ID: (-M Option) 8 mA = Range 1, 12 mA = Range 2,
16 mA = Range 3.
Examples:
Figure 2-5: Analog Output Connections
Teledyne Analytical Instruments
Part I: 2-3
2 Installation Model 6600
The analog output signal has a voltage which depends on the sample concentration AND the currently activated analysis range. To relate the signal output to the actual concentration, it is necessary to know what range the instrument is currently on, especially when the analyzer is in the autoranging mode.
The signaloutput for concentration is linear over currently selected analysis range. For example, if the analyzer is set on a range that was defined as 0-10 %, then the output would be as shown in Table 2-1.
Table 2-1: Analog Concentration Output-Examples
Concentration Voltage Signal Current Signal
% Output (V dc) Output (mA dc)
0 0.0 4.0 1 0.1 5.6 2 0.2 7.2 3 0.3 8.8 4 0.4 10.4 5 0.5 12.0 6 0.6 13.6 7 0.7 15.2 8 0.8 16.8 9 0.9 18.4
10 1.0 20.0
To provide an indication of the range, a second pair of analog output terminals are used. They generate a steady preset voltage (or current when using the current outputs) to represent a particular range. Table 2-2 gives the range ID output for each analysis range.
2-4: Part I
Teledyne Analytical Instruments
Oil in Water Analyzer Part I: Control Section
Table 2-2: Analog Range ID Output - Example
Range Voltage (V) Current (mA)
Range 1 0.25 8 Range 2 0.50 12 Range 3 0.75 16
Alarm Relays:
There are three alarm-circuit connectors on the alarm relays block (under RELAY OUTPUTS) for making connections to internal alarm relay contacts. Each provides a set of Form C contacts for each type of alarm. Each has both normally open and normally closed contact connections. The contact connections are indicated by diagrams on the rear panel. They are capable of switching up to 3 ampers at 250 V AC into a resistive load (Figure 2-6).
Normally closed Normally open Moving contact
Figure 2-6: Types of Relay Contacts
Normally open
Moving contact
The connectors are:
Threshold Alarm 1: • Can be configured as high (actuates when
concentration is above threshold), or low (actuates when concentration is below thresh old).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
Teledyne Analytical Instruments
Part I: 2-5
2 Installation Model 6600
Threshold Alarm 2: • Can be configured as high (actuates when concen-
tration is above threshold), or low (actuates when concentration is below threshold).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
System Alarm: Actuates when DC power supplied to circuits is
unacceptable in one or more parameters. Permanently configured as fail-safe and latching. Cannot be de­feated. Actuates if self test fails.
To reset a System Alarm during installation, discon­nect power to the instrument and then reconnect it
Further detail can be found in chapter 3, section 3-5.
Digital Remote Cal Inputs Remote Zero and Span Inputs: The REMOTE SPAN and RE-
MOTE ZERO inputs are on the DIGITAL INPUT terminal block. They accept 0 V (OFF) or 24 V dc (ON) for remote control of calibration (See Remote Calibration Protocol below.)
Zero: Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the ZERO mode. Either side may be grounded at the source of the signal. 0 to 1 volt across the terminals allows ZERO mode to terminate when done. A synchronous signal must open and close the external zero valve appropriately. See Remote Probe Connector at end of section 3.3. (With the -C option, the internal valves automati­cally operate synchronously).
Span: Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the SPAN mode. Either side may be grounded at the source of the signal. 0 to 1 volt across the terminals allows SPAN mode to terminate when done. A synchronous signal must open and close the external span valve appropriately. See Remote Probe Connector at end of section 3.3. (With the -C option, the internal valves automati­cally operate synchronously.)
Cal Contact: This relay contact is closed while analyzer is spanning
and/or zeroing. (See Remote Calibration Protocol below.)
2-6: Part I
Teledyne Analytical Instruments
Oil in Water Analyzer Part I: Control Section
Remote Calibration Protocol: To properly time the Digital Remote Cal Inputs to the Model 6600 Analyzer, the customer's controller must monitor the Cal Relay Contact.
When the contact is OPEN, the analyzer is analyzing, the Remote Cal Inputs are being polled, and a zero or span command can be sent.
When the contact is CLOSED, the analyzer is already calibrating. It will ignore your request to calibrate, and it will not remember that request.
Once a zero or span command is sent, and acknowledged (contact closes), release it. If the command is continued until after the zero or span is complete, the calibration will repeat and the Cal Relay Contact (CRC) will close again.
For example:
1) Test the CRC. When the CRC is open, Send a zero command until the CRC closes (The CRC will quickly close.)
2) When the CRC closes, remove the zero command.
3) When CRC opens again, send a span command until the CRC closes. (The CRC will quickly close.)
4) When the CRC closes, remove the span command.
When CRC opens again, zero and span are done, and the sample is
being analyzed.
Note: The Remote Bench terminal strip (section 3.6 Part III) provides
signals to ensure that the zero and span gas valves will be controlled synchronously.
Range ID Relays: Four dedicated RANGE ID CONTACT relays .
The first four ranges are assigned to relays in ascending order—Range 1 is assigned to RANGE 1 ID, Range 2 is assigned to RANGE 2 ID, Range 3 is assigned to RANGE 3 ID, and Range 4 is assigned to RANGE 4 ID.
Network I/O: A serial digital input/output for local network protocol.
At this printing, this port is not yet functional. It is to be used in future versions of the instrument.
RS-232 Port: The digital signal output is a standard RS-232 serial
communications port used to connect the analyzer to a computer, terminal, or other digital device. The pinouts are listed in Table 2-3.
Table 2-3: RS-232 Signals
RS-232 Sig RS-232 Pin Purpose
DCD 1 Data Carrier Detect
Teledyne Analytical Instruments
Part I: 2-7
2 Installation Model 6600
RD 2 Received Data TD 3 Transmitted Data DTR 4 Data Terminal Ready COM 5 Common DSR 6 Data Set Ready RTS 7 Request to Send CTS 8 Clear to Send RI 9 Ring Indicator
The data sent is status information, in digital form, updated every two
seconds. Status is reported in the following order:
The concentration in percent
The range is use (HI< MED< LO)
The span of the range 0-100%, etc)
Which alarm - if any - are disabled (AL-x DISABLED)
Which alarms - if any - are tripped (AL-x ON)
Each status output is followed by a carriage return and line feed.
Three input functions using RS-232 have been implemented to date.
They are described in Table 2-4.
Table 2-4: Commands via RS-232 Input
Command Description as<enter> Immediately starts an autospan. az<enter> Immediately starts an autozero. st<enter> Toggling input. Stops/Starts any status message output
from the RS-232, Until st<enter> is sent again.
The RS-232 protocol allows some flexibility in its implementation.
Table 2-5 lists certain RS-232 values that are required by the 6000B/6600.
Table 2-5: Required RS-232 Options
2-8: Part I
Parameter Setting
Baud 2400
Byte 8 bits
Parity none
Stop Bits 1
Message Interval 2 seconds
Teledyne Analytical Instruments
Oil in Water Analyzer Part I: Control Section
Remote Bench and Solenoid Valves: The 6600 is a single-chassis
instrument. However, the REMOTE BENCH and SOLENOID RETURN connectors are provided on the interface PCB. The Remote Bench is wired at the factory as well as any optional solenoid valves included in the system.
2.3 Testing the System
After The Control/Analysis Unit is both installed and interconnected,
and the system gas and electrical connections are complete, the system is ready to test. Before plugging the unit into its power sources:
Check the integrity and accuracy of the gas connections. Make sure there are no leaks.
Check the integrity and accuracy of all electrical connections. Make sure there are no exposed conductors
Check that sample pressure typically between 0 and 30 psig, according to the requirements of your process.
Turn homogenizer power potentiometer fully counter-clockwise (OFF), see section 3.3.8 for operation of homogenizer.
Warning: Do not operate the “ultrasonic homogenizer” in the
instrument for more than one (1) minute without a liquid sample properly flowing through the homog­enizer.
Power up the system, and test it by performing the following operation:
1. Repeat the Self-Diagnostic Test.
2. Zero the instrument.
3. Span the instrument.
For steps 2 and 3, refer to part II for gas calibration, and part III for Oil
in Water application.
Teledyne Analytical Instruments
Part I: 2-9
2 Installation Model 6600
2-10: Part I
Teledyne Analytical Instruments
Oil in Water Analyzer Operation 3
Operation
3.1 Introduction
Although the Model 6600 is usually programmed to your application at the factory, it can be further configured at the operator level, or even, cautiously, reprogrammed. Depending on the specifics of the application, this might include all or a subset of the following procedures:
Setting system parameters:
Establish a security password, if desired, requiring Operator to log in (secure in safe file for referrence).
Establish and start an automatic calibration cycle, if desired.
Routine Operation:
Calibrate the instrument.
Choose autoranging or select a fixed range of analysis.
Set alarm setpoints, and modes of alarm operation (latching, fail-safe, etc).
Program/Reprogram the analyzer:
Define new applications.
Linearize your ranges.
If you choose not to use password protection, the default password is
automatically displayed on the password screen when you start up, and you
simply press Enter for access to all functions of the analyzer.
3.2 Using the Controls
To get the proper response from these controls, turn the control toward the desired action (ESCAPE or ENTER—DOWN or UP), and then release it. Turn-and-release once for each action. For example, turn-and-release twice toward UP to move the VFD screen two selections upwards on the list of options (menu).
Teledyne Analytical Instruments
Part I 3-1
3 Operation Model 6600
The item that is blinking on the screen is the item that is currently selectable by choosing ENTER (turn-and-release toward ENTER with the ESCAPE/ ENTER control).
In these instructions, to ENTER means to turn-and-release toward EN­TER, and To ESCAPE means to turn-and-release towards ESCAPE. To scroll UP (or scroll DOWN) means to turn-and-release toward UP (or DOWN) as many times as necessary to reach the required menu item.
3.2.1 Mode/Function Selection
After the instrument has been powered up, and its initilization routine performed, the instrument will settle in the Analyze mode. To call up the Main menu from the Analyze mode, toggle the Enter switch. To return to the Analyze mode, toggle the Escape switch. The Main menu screens looks as shown below:
SYSTEM SPAN ZERO ALARM RANGE STBY
The Main menue screen is the top level in a series of screens used to configure the analyzer. The DOWN/UP selects the different options displayed in the VFD screen. The selectable option blinks on the VFD screen when you reach the desired option, toggle the Enter switch.
The Escape switch takes you back up to hierarchy of screens until you return back to the Analyze screen mode. Here is a brief description of the Main Menu:
System. The system function consists of nine subfunctions. Four of these are for ordinary setup and operation:
Setup an Auto-Cal
Assign Passwords
Log out to secure system
Initiate a Self-Test
Three of the subfunctions do auxiliary tasks:
Two of these are for programming/reprogramming the analyzer:
3-2 Part I
Checking model and software version
Adjust electronic filter of the signal
Display more subfunctions
Define gas applications and ranges (Refer to programming section, or contact factory.)
Teledyne Analytical Instruments
Oil in Water Analyzer Operation 3
System
Dig_filt
SELF-TEST
PWD
LOGOUT
MORE
AUTOCAL
HMGNZR
TRACK or
HOLD
CAL-HOLD
TIMER
Set Digital
Filter
Self-Test in
Progress
Enter
Password
Secure System
setup not allowed
Span/Zero status
and <>setup
Set Ultrasonic Homogenizer
ON/OFF
Set track or
hold output
Set cal. hold
and sample
hold timer
Self-Test
Results
Change
Yes/No
Span/Zero timing
and on/off
Enter
Enter
Change
Password
Enter
Verify
Password
Enter
MORE
ALGORITHM
APPLICATION
MODEL
OUTPUT_CAL
Select range
Select range
Display
Model/Version
Zero Analog
Output
Display gas use
and range
Define
Application/
Range
Enter
Enter
Select
Verify/Setup
Enter
Calibrate
Analog
Output
Verify data
Points
Auto/
Manual
linear Cal.
Enter
Enter
Enter
Input/Output
Enter Span
gas value
Figure 3-1: Hierarchy of System Functions and Subfunctions
Enter
Enter
Teledyne Analytical Instruments
Part I 3-3
3 Operation Model 6600
Use the Curve Algorithm to linearize output. (Refer to programming section, or contact factory.)
Zero. Used to set up a zero calibration.
Span. Used to set up a span calibration.
Alarms. Used to set the alarm setpoints and determine whether
each alarm will be active or defeated, HI or LO acting, latching, and/or fail-safe.
Range. Used to set up four analysis ranges that can be switched
automatically with autoranging or used as individual fixed ranges.
Any function can be selected at any time in the analyze mode (unless password restrictions apply). The order as presented in this manual is appropriate for an initial setup.
Each of these functions is described in greater detail in the following proce­dures. The VFD screen text that accompanies each operation is reproduced, at
the appropriate point in the procedure, in a Monospaced type style.
3.3 The System Function
The subfunctions of the System function are described below. Specific
procedures for their use follow the descriptions:
Dig_Filt: Adjust how much digital filtering should be on the
signal
SELF-TEST: Performs a self-diagnostic test to check the integrity
of the power supplies, outputs, detector signal and preamplifier.
PWD: Login security system for accessing to the setup functions.
LOGOUT: Prevents an unauthorized tampering with analyzer
settings.
AUTOCAL: Set the automatic calibrated timer schedule for Zero
and Span cycling.
HMGNZR: Turn Ultrasonic homogenizer ON and OFF on the
analyze mode.
TRACK: Set the system reading to be held or followed by the
concentration “gas or filter” during calibration.
CAL-HOLD-TIMER: Set the timing for calibration holding and
timing for the sample reading after return to analyze mode.
ALGORITHM: Linearize the output for nonlinear characteristic.
3-4 Part I
Teledyne Analytical Instruments
Oil in Water Analyzer Operation 3
APPLICATION: Used to define the analysis ranges and application
(gas used).
MODEL: Displays model number and software version.
OUTPUT_CAL: 4-20 MA: Adjust 4 and 20 mA output.
The hierarchy of the system menu is shown in figure 3-1.
3.3.1 Setting up an AUTO-CAL
When proper automatic valving is connected, the Analyzer can cycle itself
through a sequence of steps that automatically zero and span the instrument.
Note: Before setting up an AUTO-CAL, be sure you understand the
Zero and Span functions as described in section 4.4, and follow the precautions given there.
Note: If you require highly accurate AUTO-CAL timing, use external
AUTO-CAL control where possible. The internal clock in the Model 6600 is accurate to 2-3 %. Accordingly, internally sched­uled calibrations can vary 2-3 % per day.
To setup an Auto–Cal cycle:
Choose System from the main manu. TheVFD will display five subfunc-
tions.
DIG_FILT SELF—TEST PWD LOGOUT MORE
Select MORE and Enter
AUTOCAL HMGNZR H OLD CAL-HOLD-TIMER MORE
Use UP/DOWN to blink AUTO—CAL, and Enter. A new screen for
ZERO/SPAN set appears.
ZERO in Ød Øh off SPAN in Ød Øh off
Use UP/DOWN to blink ZERO (or SPAN), then Enter again. (You won’t
be able to set OFF to ON if a zero interval is entered.) A Span Every ...
(or Zero Every ...) screen appears.
Zero schedule: OFF Day: Ød Hour: Øh
Teledyne Analytical Instruments
Part I 3-5
3 Operation Model 6600
UP/DOWNUP/DOWN
Use
UP/DOWN to set the day interval, hour interval, then Enter
UP/DOWNUP/DOWN
Enter to turn ON the SPAN and/or ZERO cycles (to activate AUTO–CAL). Use the UP/DOWN to toggle the field between ON and OFF. Press Enter to return to The AUTO-CAL menu. You should be able to see that the screen has been updated with your new input. Escape to return to the System menu.
For Oil & Water Samples only setting of the Zero schedule is need­ed. Instrument will automatically perform a Span at the end of a sched­uled Zero.
If instrument is turned off, the next time the instrument is powered, the instrument will automatically perform a calibration cycle after 3 minutes of entering the sample mode if AUTOCAL functions were on prior to shut down.
3.3.2 Password Protection
Before a unique password is assigned, the system assigns TAI by default. This password will be displayed automatically. The operator just uses the Enter switch to be allowed total access to the instrument’s features.
If a password is assigned, then setting the following system parameters can be done only after the password is entered: alarm setpoints, assigning a new password, range/application selections, and curve algorithm linearization. (APPLICATION and ALGORITHM are covered in the programming section.) However, the instrument can still be used for analysis or for initiating a self-test without entering the password. To defeat security the password must be changed back to TAI.
NOTE: If you use password security, it is advisable to keep a copy of
the password in a separate, safe location.
3.3.2.1 Entering the Password
To install a new password or change a previously installed password, you
must key in and ENTER the old password first. If the default password is in effect, using ENTER three times will enter the default TAI password for you.
Enter the System menu...
DIG_FILT AUTO—CAL PWD LOGOUT MORE
Use theUP/DOWN to scroll the blinking over to PWD, and Enter to select
the password function. Either the default TAI password or AAA place holders for an existing password will appear on screen depending on whether or not a password has been previously installed.
3-6 Part I
Teledyne Analytical Instruments
Loading...
+ 121 hidden pages