HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM.
PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.
HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST FOR A
TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.
ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING. BEFORE
CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/MANAGER.
All Rights Reserved. No part of this manual may be reproduced, transmitted,
transcribed, stored in a retrieval system, or translated into any other language or computer
language in whole or in part, in any form or by any means, whether it be electronic,
mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of
Teledyne Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91749-
1580.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us
free from defects of material and of construction, and that our liability shall be limited to
replacing or repairing at our factory (without charge, except for transportation), or at
customer plant at our option, any material or construction in which defects become
apparent within one year from the date of shipment, except in cases where quotations or
acknowledgements provide for a shorter period. Components manufactured by others bear
the warranty of their manufacturer. This warranty does not cover defects caused by wear,
accident, misuse, neglect or repairs other than those performed by Teledyne or an authorized service center. We assume no liability for direct or indirect damages of any kind and
the purchaser by the acceptance of the equipment will assume all liability for any damage
which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our
apparatus, and to make any alterations in the dimensions, shape or weight of any parts, in
so far as such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its user, and serves as a tool by
which valuable data can be gathered. The information provided by the instrument may
assist the user in eliminating potential hazards caused by his process; however, it is
essential that all personnel involved in the use of the instrument or its interface, with the
process being measured, be properly trained in the process itself, as well as all instrumentation related to it.
The safety of personnel is ultimately the responsibility of those who control process
conditions. While this instrument may be able to provide early warning of imminent
danger, it has no control over process conditions, and it can be misused. In particular, any
alarm or control systems installed must be tested and understood, both as to how they
operate and as to how they can be defeated. Any safeguards required such as locks, labels,
or redundancy, must be provided by the user or specifically requested of Teledyne at the
time the order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The
purchaser is responsible for the training of personnel, for providing hazard warning
methods and instrumentation per the appropriate standards, and for ensuring that hazard
warning devices and instrumentation are maintained and operated properly.
Teledyne Analytical Instruments, the manufacturer of this instrument, cannot
accept responsibility for conditions beyond its knowledge and control. No statement
expressed or implied by this document or any information disseminated by the manufacturer or its agents, is to be construed as a warranty of adequate safety control under the
user’s process conditions.
ii
Teledyne Analytical Instruments
Oil in Water Analyzer
Table of Contents
Part I: Control Section .................................Part I
Part II: Analysis Section .............................Part II
Part III: Oil in Water Sample System......... Part III
Model 6600 Specifications..................................................... A-3
Teledyne Analytical Instruments
Part I: iii
Model 6600 Oil in Water Analyzer
iv: Part I
Teledyne Analytical Instruments
Oil in Water AnalyzerPart I: Control Section
Introduction
1.1Overview
The Teledyne Analytical Instruments Model 6600 Control Section,
together with a 6600 Analysis Section, is versatile microprocessor-based
instrument.
Part I, of this manual covers the Model 6600 General Purpose, Bulkhead Mount Control Section. (The Analysis Section is covered in Part II of
this manual, and Oil in Water application is covered in Part III) The Control
Section is for indoor/outdoor use in hazardous environment only. The
Analysis Section (or Remote Section) can be designed for a variety of
hazardous environments. All Sections are mounted in a NEMA-4 enclosure
(24”x20”x10”).
1.2Typical Applications when Configured
with the appropriate Sample System
A few typical applications of the Model 6600 are:
•Offshore platforms, Produced Water, Sea Water
•Waste Water, Tank Farms, Fuel Depots, Refinery Effluents
•Oil Chemical Separators
•On-Board Ship
•Boiler Return Steam Condensate
•Process Cooling Water
•Bilge/Deballast Water Treatment
•Water Soluble Oils
•All Aromatic Hydrocarbons
•Many Organic Hydrocarbons (Contact Factory)
Teledyne Analytical Instruments
Part I: 1-1
1 Introduction Model 6600
1.3Main Features of the Analyzer
The Model 6600 Photometric Analyzer is sophisticated yet simple to
use. The main features of the analyzer include:
•A 2-line alphanumeric display screen, driven by microprocessor
electronics, that continuously prompts and informs the operator.
•High resolution, accurate readings of concentration from low
ppm levels through to 100%. Large, bright, meter readout.
•Versatile analysis over a wide range of applications.
•Microprocessor based electronics: 8-bit CMOS microprocessor
with 32 kB RAM and 128 kB ROM.
•Three user definable output ranges (from 0-1 ppm through
0-100 %) allow best match to users process and equipment.
•Calibration range for convenient zeroing or spanning.
•Auto Ranging allows analyzer to automatically select the proper
preset range for a given measurement. Manual override allows
the user to lock onto a specific range of interest.
•Two adjustable concentration alarms and a system failure alarm.
•Extensive self-diagnostic testing, at startup and on demand, with
continuous power-supply monitoring.
•RS-232 serial digital port for use with a computer or other digital
communication device.
•Analog outputs for concentration and range identification.
(0-1 V dc standard, and isolated 4–20 mA dc)
•Superior accuracy.
•Internal calibration-Manual or Automatic (optional).
1.4Operator Interface
All controls and displays on the standard 6600 are accessible from
outside the housing. The instrument has two simple operator controls. The
operator has constant feedback from the instrument through an alphanumeric
display, and a digital LED meter. The displays and controls are described
briefly here and in greater detail in chapter 3. See Figure 1-1.
1-2: Part I
Teledyne Analytical Instruments
Oil in Water AnalyzerPart I: Control Section
1.4.1UP/DOWN Switch
The UP/DOWN switch is used to select between any subfunctions displayed
on the VFD screen such as in the main menue, the system menue, the Alarm
menue, etc. When modifiable values are displayed on the VFD, the UP/DOWN
switch can be used to increment or decrement the values.
1.4.2ESCAPE/ENTER Switch
The ESCAPE/ENTER switch is used to input the data, to enter a function,
or to exit a function displayed in the alphanumeric display:
•EscapeMoves VFD display back to the previous screen in a
series. If none remains, returns to Analyze mode screen.
•EnterWithin a menue: the funtion selected is entered moving on
to the next screen in a series.
WARNING:
The power cable must be disconnected to fully remove power
from the instrument.
With Value selected: Enters the value into the analyzer as
data. Advances cursor on VFD to the next operation.
In the Analyze mode: it calls the main menue. Functions
called out by the main menue:
-SystemThis function is a menu that calls a number
of functions that regulate the analyzer
operation.
-SpanThis function spans the instrument.
-ZeroThis function zeros the instrument.
-AlarmsThis functions sets the alarm preferences.
-RangeThis function selects whether analyzer is
autoranging or locked on one range.
-Standby Places the analyzer in a sleep mode.
Teledyne Analytical Instruments
Part I: 1-3
1 Introduction Model 6600
Figure 1-1: Model 6600 Controls, Indicators, and Connectors
.
Digital Meter Display: The meter display is a Light Emitting Diode
LED device that produces large, bright, 7-segment numbers that are legible
in any lighting. It is accurate across all analysis ranges. The 6600 models
produce continuous readout from 0-10,000 ppm and then switch to
continuous percent readout from 1-100 %.
1-4: Part I
Teledyne Analytical Instruments
Oil in Water AnalyzerPart I: Control Section
Figure 1-2: Model 6600 Interface Panel
Teledyne Analytical Instruments
Part I: 1-5
1 Introduction Model 6600
Alphanumeric Interface Screen: The backlit VFD screen is an easy-
to-use interface between operator and analyzer. It displays values, options,
and messages for immediate feedback to the operator.
1.5Control Section Interface Panel
The Control Section interface panel, shown in Figure 1-2, contains the
electrical terminal blocks for external inputs and outputs. The input/output
functions are described briefly here and in detail in the Installation chapter of
this manual.
•Power ConnectionAC power source, 115VAC, 50/60 Hz
•Analog Outputs0-1 V dc concentration and 0-1 V dc
range ID. Isolated 4-20 mA dc and 4-20
mA dc range ID.
•Alarm Connections2 concentration alarms and 1 system
alarm.
•RS-232 PortSerial digital concentration signal output
and control input.
•Remote BenchProvides all electrical interconnect to the
Analysis Section.
Remote Span/ZeroDigital inputs allow external control of
analyzer calibration.
•Calibration ContactTo notify external equipment that
instrument is being calibrated and
readings are not monitoring sample.
•Range ID ContactsFour separate, dedicated, range relay
contacts.
•Network I/OSerial digital communications for local
network access. For future expansion.
Not implemented at this printing.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model 6600 is accurate to 2-3 %. Accordingly, internally scheduled calibrations can vary 2-3 % per day.
1-6: Part I
Teledyne Analytical Instruments
Oil in Water AnalyzerPart I: Control Section
Installation
Installation of Model 6600 Analyzers includes:
1. Unpacking, mounting, and interconnecting the Control/Analysis
Section
2. Making gas connections to the system
3. Making electrical connections to the system
4. Testing the system.
This chapter covers installation of the Control Section. (Installation of
the Analysis Section is covered in Part II of this manual.) The Oil in Water
application is covered in Part III.
2.1Unpacking the Control/Analysis Unit
The analyzer is shipped with all the materials you need to install and
prepare the system for operation. Carefully unpack the Control/Analysis
Unit and inspect it for damage. Immediately report any damage to the shipping agent. Figure 2-2: Required Front Door Clearance
Allow clearance for the door to open in a 90-degree arc of radius 15.5
inches. See Figure 2-2.
Figure 2-2: Required Front Door
Clearance
1
5
.
5
”
2.2Electrical Connections
Figure 2-3 shows the Control/Analysis Unit interface panel. Connections for power, communications, and both digital and analog signal outputs
are described in the following paragraphs. Wire size and maximum length
data appear in the Drawings at the back of this manual.
Teledyne Analytical Instruments
Part I: 2-1
2 InstallationModel 6600
Figure 2-3: Interface Panel of the Model 6600 Control Section
For safe connections, ensure that no uninsulated wire extends
outside of the terminal blocks. Stripped wire ends must insert completely
into terminal blocks. No uninsulated wiring should come in contact with
fingers, tools or clothing during normal operation.
Primary Input Power: The power supply in the Model 6600 will
accept a 115 Vac, 50/60 Hz power source. See Figure 2-4 for detailed
connections.
DANGER:Power is applied to the instrument's circuitry as
long as the instrument is connected to the power
source. The standby function switches power on or
off to the displays and outputs only.
115VAC
2-2: Part I
Figure 2-4: Primary Input Power Connections
Teledyne Analytical Instruments
Oil in Water AnalyzerPart I: Control Section
Fuse Installation: The fuse holders accept 5 x 20 mm, 4.0 A, T
type (slow blow) fuses. Fuses are not installed at the factory. Be sure to
install the proper fuse as part of installation (See Fuse Replacement in
chapter 4, maintenance.)
Analog Outputs: There are eight DC output signal connectors on
the ANALOG OUTPUTS terminal block. There are two connectors per
output with the polarity noted. See Figure 2-5.
The outputs are:
0–1 V dc % of Range:Voltage rises linearly with increasing sample con-
centration, from 0 V at 0% to 1 V at 100%. (Full
scale = 100% programmed range.)
0–1 V dc Range ID:0.25 V = Range 1, 0.5 V = Range 2, 0.75 V =
Range 3.
4–20 mA dc % Range: (-M Option) Current increases linearly with increas-
ing sample concentration, from 4 mA at 0% to 20
mA at full scale 100%. (Full scale = 100% of
programmed range.)
4–20 mA dc Range ID: (-M Option) 8 mA = Range 1, 12 mA = Range 2,
16 mA = Range 3.
Examples:
Figure 2-5: Analog Output Connections
Teledyne Analytical Instruments
Part I: 2-3
2 InstallationModel 6600
The analog output signal has a voltage which depends on the sample
concentration AND the currently activated analysis range. To relate the
signal output to the actual concentration, it is necessary to know what range
the instrument is currently on, especially when the analyzer is in the
autoranging mode.
The signaloutput for concentration is linear over currently selected
analysis range. For example, if the analyzer is set on a range that was
defined as 0-10 %, then the output would be as shown in Table 2-1.
To provide an indication of the range, a second pair of analog output
terminals are used. They generate a steady preset voltage (or current when
using the current outputs) to represent a particular range. Table 2-2 gives the
range ID output for each analysis range.
2-4: Part I
Teledyne Analytical Instruments
Oil in Water AnalyzerPart I: Control Section
Table 2-2: Analog Range ID Output - Example
RangeVoltage (V)Current (mA)
Range 10.258
Range 20.5012
Range 30.7516
Alarm Relays:
There are three alarm-circuit connectors on the alarm relays block
(under RELAY OUTPUTS) for making connections to internal alarm relay
contacts. Each provides a set of Form C contacts for each type of alarm.
Each has both normally open and normally closed contact connections. The
contact connections are indicated by diagrams on the rear panel. They are
capable of switching up to 3 ampers at 250 V AC into a resistive load
(Figure 2-6).
Normally closed
Normally open
Moving contact
Figure 2-6: Types of Relay Contacts
Normally open
Moving contact
The connectors are:
Threshold Alarm 1:• Can be configured as high (actuates when
concentration is above threshold), or low
(actuates when concentration is below thresh old).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
Teledyne Analytical Instruments
Part I: 2-5
2 InstallationModel 6600
Threshold Alarm 2:• Can be configured as high (actuates when concen-
tration is above threshold), or low (actuates when
concentration is below threshold).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
System Alarm:Actuates when DC power supplied to circuits is
unacceptable in one or more parameters. Permanently
configured as fail-safe and latching. Cannot be defeated. Actuates if self test fails.
To reset a System Alarm during installation, disconnect power to the instrument and then reconnect it
Further detail can be found in chapter 3, section 3-5.
Digital Remote Cal Inputs
Remote Zero and Span Inputs: The REMOTE SPAN and RE-
MOTE ZERO inputs are on the DIGITAL INPUT terminal block. They
accept 0 V (OFF) or 24 V dc (ON) for remote control of calibration (See
Remote Calibration Protocol below.)
Zero:Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the ZERO mode. Either side may be
grounded at the source of the signal. 0 to 1 volt across the
terminals allows ZERO mode to terminate when done. A
synchronous signal must open and close the external zero
valve appropriately. See Remote Probe Connector at end of
section 3.3. (With the -C option, the internal valves automatically operate synchronously).
Span:Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the SPAN mode. Either side may be
grounded at the source of the signal. 0 to 1 volt across the
terminals allows SPAN mode to terminate when done. A
synchronous signal must open and close the external span
valve appropriately. See Remote Probe Connector at end of
section 3.3. (With the -C option, the internal valves automatically operate synchronously.)
Cal Contact: This relay contact is closed while analyzer is spanning
and/or zeroing. (See Remote Calibration Protocol below.)
2-6: Part I
Teledyne Analytical Instruments
Oil in Water AnalyzerPart I: Control Section
Remote Calibration Protocol: To properly time the Digital Remote
Cal Inputs to the Model 6600 Analyzer, the customer's controller must
monitor the Cal Relay Contact.
When the contact is OPEN, the analyzer is analyzing, the Remote Cal
Inputs are being polled, and a zero or span command can be sent.
When the contact is CLOSED, the analyzer is already calibrating. It
will ignore your request to calibrate, and it will not remember that request.
Once a zero or span command is sent, and acknowledged (contact
closes), release it. If the command is continued until after the zero or span is
complete, the calibration will repeat and the Cal Relay Contact (CRC) will
close again.
For example:
1) Test the CRC. When the CRC is open, Send a zero command
until the CRC closes (The CRC will quickly close.)
2) When the CRC closes, remove the zero command.
3) When CRC opens again, send a span command until the CRC
closes. (The CRC will quickly close.)
4) When the CRC closes, remove the span command.
When CRC opens again, zero and span are done, and the sample is
being analyzed.
Note: The Remote Bench terminal strip (section 3.6 Part III) provides
signals to ensure that the zero and span gas valves will be
controlled synchronously.
Range ID Relays: Four dedicated RANGE ID CONTACT relays .
The first four ranges are assigned to relays in ascending order—Range 1 is
assigned to RANGE 1 ID, Range 2 is assigned to RANGE 2 ID, Range 3
is assigned to RANGE 3 ID, and Range 4 is assigned to RANGE 4 ID.
Network I/O: A serial digital input/output for local network protocol.
At this printing, this port is not yet functional. It is to be used in future
versions of the instrument.
RS-232 Port: The digital signal output is a standard RS-232 serial
communications port used to connect the analyzer to a computer, terminal, or
other digital device. The pinouts are listed in Table 2-3.
Table 2-3: RS-232 Signals
RS-232 Sig RS-232 Pin Purpose
DCD1Data Carrier Detect
Teledyne Analytical Instruments
Part I: 2-7
2 InstallationModel 6600
RD2Received Data
TD3Transmitted Data
DTR4Data Terminal Ready
COM5Common
DSR6Data Set Ready
RTS7Request to Send
CTS8Clear to Send
RI9Ring Indicator
The data sent is status information, in digital form, updated every two
seconds. Status is reported in the following order:
•The concentration in percent
•The range is use (HI< MED< LO)
•The span of the range 0-100%, etc)
•Which alarm - if any - are disabled (AL-x DISABLED)
•Which alarms - if any - are tripped (AL-x ON)
Each status output is followed by a carriage return and line feed.
Three input functions using RS-232 have been implemented to date.
They are described in Table 2-4.
Table 2-4: Commands via RS-232 Input
CommandDescription
as<enter>Immediately starts an autospan.
az<enter>Immediately starts an autozero.
st<enter>Toggling input. Stops/Starts any status message output
from the RS-232, Until st<enter> is sent again.
The RS-232 protocol allows some flexibility in its implementation.
Table 2-5 lists certain RS-232 values that are required by the 6000B/6600.
Table 2-5: Required RS-232 Options
2-8: Part I
ParameterSetting
Baud2400
Byte8 bits
Paritynone
Stop Bits1
Message Interval2 seconds
Teledyne Analytical Instruments
Oil in Water AnalyzerPart I: Control Section
Remote Bench and Solenoid Valves: The 6600 is a single-chassis
instrument. However, the REMOTE BENCH and SOLENOID RETURN
connectors are provided on the interface PCB. The Remote Bench is wired
at the factory as well as any optional solenoid valves included in the system.
2.3Testing the System
After The Control/Analysis Unit is both installed and interconnected,
and the system gas and electrical connections are complete, the system is
ready to test. Before plugging the unit into its power sources:
•Check the integrity and accuracy of the gas connections. Make
sure there are no leaks.
•Check the integrity and accuracy of all electrical connections.
Make sure there are no exposed conductors
•Check that sample pressure typically between 0 and 30 psig,
according to the requirements of your process.
•Turn homogenizer power potentiometer fully counter-clockwise
(OFF), see section 3.3.8 for operation of homogenizer.
Warning:Do not operate the “ultrasonic homogenizer” in the
instrument for more than one (1) minute without a
liquid sample properly flowing through the homogenizer.
Power up the system, and test it by performing the following operation:
1. Repeat the Self-Diagnostic Test.
2. Zero the instrument.
3. Span the instrument.
For steps 2 and 3, refer to part II for gas calibration, and part III for Oil
in Water application.
Teledyne Analytical Instruments
Part I: 2-9
2 InstallationModel 6600
2-10: Part I
Teledyne Analytical Instruments
Oil in Water AnalyzerOperation 3
Operation
3.1Introduction
Although the Model 6600 is usually programmed to your application at the
factory, it can be further configured at the operator level, or even, cautiously,
reprogrammed. Depending on the specifics of the application, this might include
all or a subset of the following procedures:
•Setting system parameters:
•Establish a security password, if desired, requiring Operator
to log in (secure in safe file for referrence).
•Establish and start an automatic calibration cycle, if desired.
•Routine Operation:
•Calibrate the instrument.
•Choose autoranging or select a fixed range of analysis.
•Set alarm setpoints, and modes of alarm operation (latching,
fail-safe, etc).
•Program/Reprogram the analyzer:
•Define new applications.
•Linearize your ranges.
If you choose not to use password protection, the default password is
automatically displayed on the password screen when you start up, and you
simply press Enter for access to all functions of the analyzer.
3.2Using the Controls
To get the proper response from these controls, turn the control toward the
desired action (ESCAPE or ENTER—DOWN or UP), and then release it.
Turn-and-release once for each action. For example, turn-and-release twice
toward UP to move the VFD screen two selections upwards on the list of
options (menu).
Teledyne Analytical Instruments
Part I 3-1
3 Operation Model 6600
The item that is blinking on the screen is the item that is currently selectable
by choosing ENTER (turn-and-release toward ENTER with the ESCAPE/
ENTER control).
In these instructions, to ENTER means to turn-and-release toward ENTER, and To ESCAPE means to turn-and-release towards ESCAPE. To scroll
UP (or scroll DOWN) means to turn-and-release toward UP (or DOWN) as
many times as necessary to reach the required menu item.
3.2.1Mode/Function Selection
After the instrument has been powered up, and its initilization routine
performed, the instrument will settle in the Analyze mode. To call up the Main
menu from the Analyze mode, toggle the Enter switch. To return to the Analyze
mode, toggle the Escape switch. The Main menu screens looks as shown
below:
SYSTEM SPAN ZERO
ALARM RANGE STBY
The Main menue screen is the top level in a series of screens used to
configure the analyzer. The DOWN/UP selects the different options displayed
in the VFD screen. The selectable option blinks on the VFD screen when you
reach the desired option, toggle the Enter switch.
The Escape switch takes you back up to hierarchy of screens until you
return back to the Analyze screen mode. Here is a brief description of the Main
Menu:
•System. The system function consists of nine subfunctions.
Four of these are for ordinary setup and operation:
•Setup an Auto-Cal
•Assign Passwords
•Log out to secure system
•Initiate a Self-Test
Three of the subfunctions do auxiliary tasks:
Two of these are for programming/reprogramming the analyzer:
3-2 Part I
•Checking model and software version
•Adjust electronic filter of the signal
•Display more subfunctions
•Define gas applications and ranges (Refer to programming
section, or contact factory.)
Teledyne Analytical Instruments
Oil in Water AnalyzerOperation 3
System
Dig_filt
SELF-TEST
PWD
LOGOUT
MORE
AUTOCAL
HMGNZR
TRACK or
HOLD
CAL-HOLD
TIMER
Set Digital
Filter
Self-Test in
Progress
Enter
Password
Secure System
setup not
allowed
Span/Zero status
and <>setup
Set Ultrasonic
Homogenizer
ON/OFF
Set track or
hold output
Set cal. hold
and sample
hold timer
Self-Test
Results
Change
Yes/No
Span/Zero timing
and on/off
Enter
Enter
Change
Password
Enter
Verify
Password
Enter
MORE
ALGORITHM
APPLICATION
MODEL
OUTPUT_CAL
Select range
Select range
Display
Model/Version
Zero Analog
Output
Display gas use
and range
Define
Application/
Range
Enter
Enter
Select
Verify/Setup
Enter
Calibrate
Analog
Output
Verify data
Points
Auto/
Manual
linear Cal.
Enter
Enter
Enter
Input/Output
Enter Span
gas value
Figure 3-1: Hierarchy of System Functions and Subfunctions
Enter
Enter
Teledyne Analytical Instruments
Part I 3-3
3 Operation Model 6600
•Use the Curve Algorithm to linearize output. (Refer to
programming section, or contact factory.)
•Zero. Used to set up a zero calibration.
•Span. Used to set up a span calibration.
•Alarms. Used to set the alarm setpoints and determine whether
each alarm will be active or defeated, HI or LO acting, latching,
and/or fail-safe.
•Range. Used to set up four analysis ranges that can be switched
automatically with autoranging or used as individual fixed
ranges.
Any function can be selected at any time in the analyze mode (unless
password restrictions apply). The order as presented in this manual is
appropriate for an initial setup.
Each of these functions is described in greater detail in the following procedures. The VFD screen text that accompanies each operation is reproduced, at
the appropriate point in the procedure, in a Monospaced type style.
3.3The System Function
The subfunctions of the System function are described below. Specific
procedures for their use follow the descriptions:
•Dig_Filt: Adjust how much digital filtering should be on the
signal
•SELF-TEST: Performs a self-diagnostic test to check the integrity
of the power supplies, outputs, detector signal and preamplifier.
•PWD: Login security system for accessing to the setup functions.
•LOGOUT: Prevents an unauthorized tampering with analyzer
settings.
•AUTOCAL: Set the automatic calibrated timer schedule for Zero
and Span cycling.
•HMGNZR: Turn Ultrasonic homogenizer ON and OFF on the
analyze mode.
•TRACK: Set the system reading to be held or followed by the
concentration “gas or filter” during calibration.
•CAL-HOLD-TIMER: Set the timing for calibration holding and
timing for the sample reading after return to analyze mode.
•ALGORITHM: Linearize the output for nonlinear characteristic.
3-4 Part I
Teledyne Analytical Instruments
Oil in Water AnalyzerOperation 3
•APPLICATION: Used to define the analysis ranges and application
(gas used).
•MODEL: Displays model number and software version.
•OUTPUT_CAL: 4-20 MA: Adjust 4 and 20 mA output.
The hierarchy of the system menu is shown in figure 3-1.
3.3.1Setting up an AUTO-CAL
When proper automatic valving is connected, the Analyzer can cycle itself
through a sequence of steps that automatically zero and span the instrument.
Note: Before setting up an AUTO-CAL, be sure you understand the
Zero and Span functions as described in section 4.4, and
follow the precautions given there.
Note: If you require highly accurate AUTO-CAL timing, use external
AUTO-CAL control where possible. The internal clock in the
Model 6600 is accurate to 2-3 %. Accordingly, internally scheduled calibrations can vary 2-3 % per day.
To setup an Auto–Cal cycle:
Choose System from the main manu. TheVFD will display five subfunc-
tions.
DIG_FILT SELF—TEST
PWD LOGOUT MORE
Select MORE and Enter
AUTOCAL HMGNZRH OLD
CAL-HOLD-TIMERMORE
Use UP/DOWN to blink AUTO—CAL, and Enter. A new screen for
ZERO/SPAN set appears.
ZERO in Ød Øh off
SPAN in Ød Øh off
Use UP/DOWN to blink ZERO (or SPAN), then Enter again. (You won’t
be able to set OFF to ON if a zero interval is entered.) A Span Every ...
(or Zero Every ...) screen appears.
Zero schedule: OFF
Day: Ød Hour: Øh
Teledyne Analytical Instruments
Part I 3-5
3 Operation Model 6600
UP/DOWNUP/DOWN
Use
UP/DOWN to set the day interval, hour interval, then Enter
UP/DOWNUP/DOWN
Enter to turn ON the SPAN and/or ZERO cycles (to activate AUTO–CAL).
Use the UP/DOWN to toggle the field between ON and OFF. Press Enter to
return to The AUTO-CAL menu. You should be able to see that the screen has
been updated with your new input. Escape to return to the System menu.
For Oil & Water Samples only setting of the Zero schedule is needed. Instrument will automatically perform a Span at the end of a scheduled Zero.
If instrument is turned off, the next time the instrument is powered, the
instrument will automatically perform a calibration cycle after 3 minutes of
entering the sample mode if AUTOCAL functions were on prior to shut down.
3.3.2Password Protection
Before a unique password is assigned, the system assigns TAI by default.
This password will be displayed automatically. The operator just uses the Enter
switch to be allowed total access to the instrument’s features.
If a password is assigned, then setting the following system parameters can
be done only after the password is entered: alarm setpoints, assigning a new
password, range/application selections, and curve algorithm linearization.
(APPLICATION and ALGORITHM are covered in the programming section.)
However, the instrument can still be used for analysis or for initiating a self-test
without entering the password. To defeat security the password must be
changed back to TAI.
NOTE: If you use password security, it is advisable to keep a copy of
the password in a separate, safe location.
3.3.2.1Entering the Password
To install a new password or change a previously installed password, you
must key in and ENTER the old password first. If the default password is in
effect, using ENTER three times will enter the default TAI password for you.
Enter the System menu...
DIG_FILT AUTO—CAL
PWD LOGOUT MORE
Use theUP/DOWN to scroll the blinking over to PWD, and Enter to select
the password function. Either the default TAI password or AAA place holders for
an existing password will appear on screen depending on whether or not a
password has been previously installed.
3-6 Part I
Teledyne Analytical Instruments
Loading...
+ 121 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.