Teledyne 6000B User Manual

Photometric Analyzer
OPERATING INSTRUCTIONS
Model 6000B
Photometric Analyzer
DANGER
HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM.
FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED. ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING. BEFORE
CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/ MANAGER.
Teledyne Analytical Instruments
P/N M67986
11/24/04
ECO # 03-0126
i
Model 6000B
Copyright © 1999 Teledyne Analytical Instruments
All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in part, in any form or by any means, whether it be electronic, mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of Teledyne Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91749-
1580.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us free from defects of material and of construction, and that our liability shall be limited to replacing or repairing at our factory (without charge, except for transportation), or at customer plant at our option, any material or construction in which defects become apparent within one year from the date of shipment, except in cases where quotations or acknowledgements provide for a shorter period. Components manufactured by others bear the warranty of their manufacturer. This warranty does not cover defects caused by wear, accident, misuse, neglect or repairs other than those performed by Teledyne or an autho­rized service center. We assume no liability for direct or indirect damages of any kind and the purchaser by the acceptance of the equipment will assume all liability for any damage which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our apparatus, and to make any alterations in the dimensions, shape or weight of any parts, in so far as such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its user, and serves as a tool by which valuable data can be gathered. The information provided by the instrument may assist the user in eliminating potential hazards caused by his process; however, it is essential that all personnel involved in the use of the instrument or its interface, with the process being measured, be properly trained in the process itself, as well as all instrumen­tation related to it.
The safety of personnel is ultimately the responsibility of those who control process conditions. While this instrument may be able to provide early warning of imminent danger, it has no control over process conditions, and it can be misused. In particular, any alarm or control systems installed must be tested and understood, both as to how they operate and as to how they can be defeated. Any safeguards required such as locks, labels, or redundancy, must be provided by the user or specifically requested of Teledyne at the time the order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The purchaser is responsible for the training of personnel, for providing hazard warning methods and instrumentation per the appropriate standards, and for ensuring that hazard warning devices and instrumentation are maintained and operated properly.
Teledyne Analytical Instruments, the manufacturer of this instrument, cannot accept responsibility for conditions beyond its knowledge and control. No statement expressed or implied by this document or any information disseminated by the manufac­turer or its agents, is to be construed as a warranty of adequate safety control under the
user’s process conditions.
ii
Teledyne Analytical Instruments
Photometric Analyzer
Table of Contents
Specific Model Information................................. iv
Preface................................................................v
Part I: Control Unit.................................Part I: 1-1
Part II: Analysis Unit.............................Part II: 1-1
Appendix .........................................................A-1
Teledyne Analytical Instruments
iii
Model 6000B
iv
Teledyne Analytical Instruments
Part I: Control Unit
OPERA TING INSTRUCTIONS
Model 6000B
Photometric Analyzer
Part I: Control Unit
NEMA 4 Bulkhead Mount
Teledyne Analytical Instruments
Part I: i
Model 6000B Photometric Analyzer
Table of Contents
1 Introduction
1.1 Overview........................................................................ 1-1
1.2 Typical Applications ....................................................... 1-1
1.3 Main Features of the Analyzer ....................................... 1-1
1.4 Control Unit Inner Interface P anel .................................. 1-2
1.5 Control Unit Interface P anel........................................... 1-2
2 Installation
2.1 Unpacking the Control Unit............................................ 2-1
2.2 Mounting the Control Unit .............................................. 2-1
2.3 Electrical Connections ................................................... 2-3
2.4 Testing the System......................................................... 2-12
3 Operation
3.1 Introduction .................................................................... 3-1
3.2 Using the Data Entry and Function Buttons ................... 3-1
3.3 The System Function ..................................................... 3-4
3.3.1 Setting up an Auto-Cal........................................... 3-4
3.3.2 Pass w ord Protection.............................................. 3-6
3.3.2.1 Entering the Password...................... ... .......... 3-6
3.3.2.2 Installing or Changing the Password ............. 3-7
3.3.3 Logging Out ........................................................... 3-8
3.3.4 System Self-Diagnostic Test .................................. 3-9
3.3.5 The Model Screen ................................................. 3-10
3.3.6 Checking Linearity with Algorithm ......................... 3-10
3.3.7 Digital Flter Setup .................................................. 3-11
3.3.8 Filter or Solenoid Setup ......................................... 3-12
3.3.9 Hold/Track Setup ................................................... 3-13
3.3.11 Calibration/Hold Timer Setup................................. 3-14
3.3.12 Analog 4 to 20 mA Output Calibration.................... 3-15
3.3.12 Model..................................................................... 3-15
ii: Part I
Teledyne Analytical Instruments
Part I: Control Unit
3.3.13 Show Negative ...................................................... 3-15
3.4 The Zero and Span Functions ....................................... 3-16
3.4.1 Zero Cal ................................................................. 3-16
3.4.1.1 Auto Mode Zeroing ........................................ 3-16
3.4.1.2 Manual Mode Zeroing.................................... 3-17
3.4.1.3 Cell Failure .................................................... 3-18
3.4.2 Span Cal................................................................ 3-18
3.4.3 Offset Function....................................................... 3-20
3.4.2.1 Auto Mode Spanning ..................................... 3-19
3.4.2.2 Manual Mode Spanning................................. 3-19
3.5 The Alarms Function...................................................... 3-21
3.6 The Range Select Function ........................................... 3-23
3.6.1 Manual (Select/Define Range) Screen .................. 3-24
3.6.2 Auto Screen ........................................................... 3-24
3.6.3 Precautions............................................................ 3-26
3.7 The Analyze Function .................................................... 3-27
3.8 Programming ................................................................. 3-28
3.8.1 The Set Range Screen .......................................... 3.28
3.8.2 The Curve Algorithm Screen ................................. 3-30
3.8.2.1 Manual Mode Linearization ........................... 3-30
3.8.2.2 Auto Mode Linearization................................ 3-31
4 Maintenance
4.1 Fuse Replacement ......................................................... 4-1
4.2 System Self Diagnostic Test........................................... 4-2
4.3 Major Internal Components............................................ 4-3
A Appendix
Model 6000B Specifications .................................................. A-3
Teledyne Analytical Instruments
Part I: iii
Model 6000B Photometric Analyzer
iv: Part I
Teledyne Analytical Instruments
Photometric Analyzer Part I: Control Unit
Introduction
1.1 Overview
The Teledyne Analytical Instruments Model 6000B Control Unit, together with a 6000B Analysis Unit, is versatile microprocessor-based instrument.
Part I, of this manual covers the Model 6000B General Purpose NEMA 4 Bulkhead Mount Control Unit. (The Analysis Unit is covered in Part II of this manual.) The Control Unit is for indoor/outdoor use in a nonhazardous environment only. The Analysis Unit (or Remote Section) can be designed for a variety of hazardous environments.
1.2 Typical Applications
A few typical applications of the Model 6000B are:
Oil in refinery waste water condensates Streams
•CL2, HC, SO2, H2S in stack gases or Liquid Streams
Chemical reaction monitoring
Product Color monitoring liquids
Petrochemical process control
Quality assurance
Phenol in water
Hazardous waste incineration
CLO2, Hypochlorite monitoring
•F2 monitoring
1.3 Main Features of the Analyzer
The Model 6000B Photometric Analyzer is sophisticated yet simple to use. The main features of the analyzer include:
Teledyne Analytical Instruments
Part I: 1-1
1 Introduction Model 6000B
A 2-line alphanumeric display screen, driven by microprocessor electronics, that continuously prompts and informs the operator.
High resolution, accurate readings of concentration from low ppm levels through to 100%. Large, bright, meter readout.
Versatile analysis over a wide range of applications.
Microprocessor based electronics: 8-bit CMOS microprocessor with 32 kB RAM and 128 kB ROM.
Three user definable output ranges (from 0-1 ppm through 0-100 %) allow best match to users process and equipment.
Calibration range for convenient zeroing or spanning.
Auto Ranging allows analyzer to automatically select the proper preset range for a given measurement. Manual override allows the user to lock onto a specific range of interest.
Two adjustable concentration alarms and a system failure alarm.
Extensive self-diagnostic testing, at startup and on demand, with continuous power-supply monitoring.
RS-232 serial digital port for use with a computer or other digital communication device.
Analog outputs for concentration and range identification. (0-1 V dc standard, and isolated 4–20 mA dc)
Superior accuracy.
Internal calibration (optional).
1. 4 Control Unit Inner Control Panel
The standard 6000B Control Unit is housed in a rugged NEMA 4 metal case with all remote controls and displays accessible from the inner control panel. See Figure 1-1. The inner control panel has a digital meter, an alpha­numeric Vacuum Fluroscent Display (VFD), and thirteen buttons for operat­ing the analyzer.
1-2: Part I
Teledyne Analytical Instruments
Photometric Analyzer Part I: Control Unit
Figure 1-1: Front of Unmounted Control Unit
Function Keys: Six touch-sensitive membrane switches are used to change
the specific function performed by the analyzer:
Analyze Perform analysis for concentration content of a sample.
System Perform system-related tasks (described in detail in chapter
3, Operation.).
Span Span calibrate the analyzer.
Zero Zero calibrate the analyzer.
Alarms Set the alarm setpoints and attributes.
Range Set up the 3 user definable ranges for the instrument.
Teledyne Analytical Instruments
Part I: 1-3
1 Introduction Model 6000B
Data Entry Keys: Six touch-sensitive membrane switches are used to input data to the instrument via the alphanumeric VFD display:
Left & Right Arrows Select between functions currently
displayed on the VFD screen.
Up & Down Arrows Increment or decrement values of
functions currently displayed.
Enter Moves VFD display on to the next screen in a series. If
none remains, returns to the Analyze screen.
Escape Moves VFD display back to the previous screen in a
series. If none remains, returns to the Analyze screen.
Digital Meter Display: The meter display is a Light Emitting Diode LED device that produces large, bright, 7-segment numbers that are legible in any lighting. It is accurate across all analysis ranges. The 6000B models produce continuous readout from 0-10,000 ppm and then switch to continuous percent readout from 1-100 %.
Alphanumeric Interface Screen: The backlit VFD screen is an easy­to-use interface between operator and analyzer. It displays values, options, and messages for immediate feedback to the operator.
I/O Power Button: The red I/O button switches the instrument power between I (ON) and O (a Keep-Alive state). In the O state, the instrument’s circuitry is operating, but there are no displays or outputs.
CAUTION: The power must be disconnected to fully
disconnect power from the instrument. When chassis is exposed or when access door is open and power cable is connected, use extra care to avoid contact with live electrical circuits .
Access Door: For access to the electronics and interface panel, the front panel swings open when the latch in the panel is pressed all the way in with a narrow gauge tool. Accessing the main circuit board and other electronics requires unfastening the rear panel screws and sliding the unit out of the case.
1.5 Control Unit Interface Panel
The Control Unit interface panel, shown in Figure 1-2, contains the electrical terminal blocks for external inputs and outputs. The input/output functions are described briefly here and in detail in the Installation chapter of this manual.
1-4: Part I
Teledyne Analytical Instruments
Photometric Analyzer Part I: Control Unit
Figure 1-2: Model 6000B Rear Panel
Teledyne Analytical Instruments
Part I: 1-5
1 Introduction Model 6000B
Power Connection AC power source, 115VAC, 50/60 Hz
Analog Outputs 0-1 V dc concentration and 0-1 V dc
range ID. Isolated 4-20 mA dc and 4-20 mA dc range ID.
Alarm Connections 2 concentration alarms and 1 system
alarm.
RS-232 Port Serial digital concentration signal output
and control input.
Remote Bench Provides all electrical interconnect to the
Analysis Unit.
Remote Span/Zero Digital inputs allow external control of
analyzer calibration.
Calibration Contact To notify external equipment that
instrument is being calibrated and readings are not monitoring sample.
Range ID Contacts Four separate, dedicated, range relay
contacts. Low, Medium, High, Cal.
Network I/O Serial digital communications for local
network access. For future expansion. Not implemented at this printing.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the Model 6000B is accurate to 2-3 %. Accordingly, internally scheduled calibrations can vary 2-3 % per day.
1-6: Part I
Teledyne Analytical Instruments
Photometric Analyzer Part I: Control Unit
Installation
Installation of Model 6000B Analyzers includes:
1. Unpacking, mounting, and interconnecting the Control Unit and the Analysis Unit
2. Making gas connections to the system
3. Making electrical connections to the system
4. Testing the system.
This chapter covers installation of the Control Unit. (Installation of the
Analysis Unit is covered in Part II of this manual.)
2.1 Unpacking the Control Unit
The analyzer is shipped with all the materials you need to install and prepare the system for operation. Carefully unpack the Control Unit and inspect it for damage. Immediately report any damage to the shipping agent.
2.2 Mounting the Control Unit
The Model 6000B Control Unit is for indoor/outdoor use in a general purpose area. This Unit is NOT for any type of hazardous environ-
ments.
Figure 2-1 is an illustration of a Model 6000B standard Control Unit front panel and mounting brackets as shown two mounting tabs are at the top and two at the bottom of the units frame.
Teledyne Analytical Instruments
Part I: 2-1
2 Installation Model 6000B
NPT Fittings
supplied by
customer
Viewing W indow
0.0 % Anlz AL -1
AC POWER IN
50/60 HZ
115V
ALARM OUTPUTS
DIGITAL INPUT SPAN ZERO
CAL. CONTACT RANGE
ID CONTACTS RS-232
SOLENOID RETURN
ANALOG OUTPUTS
NET WORK
Figure 2-1: Front Panel of the Model 6000B Control Unit
3/4" NPT
Outer Door
Latch
3/4" NPT
1" NPT
1" NPT
All operator controls are mounted on the inner control panel "door", which is hinged on the left edge and doubles as a door to provide access to the internal components of the instrument. The door will swing open when the button of the latch is pressed all the way in with a narrow gauge tool (less than 0.18 inch wide), such as a small hex wrench or screwdriver
2-2: Part I
Teledyne Analytical Instruments
Photometric Analyzer Part I: Control Unit
11.75
Figure 2-2: Required Front Door Clearance
Allow clearance for the door to open in a 90-degree arc of radius 11.75
inches. See Figure 2-2.
2.3 Electrical Connections
Figure 2-3 shows the Control Unit interface panel. Connections for power, communications, and both digital and analog signal outputs are described in the following paragraphs. Wire size and maximum length data appear in the Drawings at the back of this manual.
Figure 2-3: Interface Panel of the Model 6000B Control Unit
For safe connections, ensure that no uninsulated wire extends outside of the terminal blocks. Stripped wire ends must insert completely
into terminal blocks. No uninsulated wiring should come in contact with fingers, tools or clothing during normal operation.
Teledyne Analytical Instruments
Part I: 2-3
2 Installation Model 6000B
Primary Input Power: The power supply in the Model 6000B will
accept a 115 Vac, 50/60 Hz power source. See Figure 2-4 for detailed connections.
DANGER: Power is applied to the instrument's circuitry as
long as the instrument is connected to the power source. The standby function switches power on or off to the displays and outputs only.
115VAC
Figure 2-4: Primary Input Power Connections
Fuse Installation: The fuse holders accept 5 x 20 mm, 4.0 A, T
type (slow blow) fuses. Fuses are not installed at the factory. Be sure to install the proper fuse as part of installation (See Fuse Replacement in chapter 4, maintenance.)
Analog Outputs: There are eight DC output signal connectors on
the ANALOG OUTPUTS terminal block. There are two connectors per output with the polarity noted. See Figure 2-5.
The outputs are:
0–1 V dc % of Range: Voltage rises linearly with increasing sample con-
centration, from 0 V at 0% to 1 V at 100%. (Full scale = 100% programmed range.)
0–1 V dc Range ID: 0.25 V = Low Range, 0.5 V = Medium Range,
0.75 V = High Range, 1 V = Cal Range.
2-4: Part I
Teledyne Analytical Instruments
Photometric Analyzer Part I: Control Unit
4–20 mA dc % Range: (-M Option) Current increases linearly with increas-
ing sample concentration, from 4 mA at 0% to 20 mA at full scale 100%. (Full scale = 100% of programmed range.)
4–20 mA dc Range ID: (-M Option) 8 mA = Low Range, 12 mA = Me-
dium Range, 16 mA = High Range, 20 mA = Cal Range.
Figure 2-5: Analog Output Connections
Examples:
The analog output signal has a voltage which depends on the sample concentration AND the currently activated analysis range. To relate the signal output to the actual concentration, it is necessary to know what range the instrument is currently on, especially when the analyzer is in the autoranging mode.
The signaloutput for concentration is linear over currently selected analysis range. For example, if the analyzer is set on a range that was defined as 0-10 %, then the output would be as shown in Table 2-1.
Teledyne Analytical Instruments
Part I: 2-5
2 Installation Model 6000B
Table 2-1: Analog Concentration Output-Examples
Analyte Voltage Signal Current Signal
% Output (V dc) Output (mA dc)
0 0.0 4.0 1 0.1 5.6 2 0.2 7.2 3 0.3 8.8 4 0.4 10.4 5 0.5 12.0 6 0.6 13.6 7 0.7 15.2 8 0.8 16.8 9 0.9 18.4
10 1.0 20.0
To provide an indication of the range, a second pair of analog output terminals are used. They generate a steady preset voltage (or current when using the current outputs) to represent a particular range. Table 2-2 gives the range ID output for each analysis range.
Table 2-2: Analog Range ID Output - Example
Range Voltage (V) Current (mA)
LO 0.25 8 MED 0.50 12 HI 0.75 16 CAL (0-25%) 1.00 20
2-6: Part I
Teledyne Analytical Instruments
Photometric Analyzer Part I: Control Unit
N
Alarm Relays:
There are three alarm-circuit connectors on the alarm relays block (under RELAY OUTPUTS) for making connections to internal alarm relay contacts. Each provides a set of Form C contacts for each type of alarm. Each has both normally open and normally closed contact connections. The contact connections are indicated by diagrams on the rear panel. They are capable of switching up to 3 ampers at 250 V AC into a resistive load (Figure 2-6).
Normally closed Normally open Moving contact
Figure 2-6: Types of Relay Contacts
ormally open
Moving contact
The connectors are:
Threshold Alarm 1: • Can be configured as high (actuates when
concentration is above threshold), or low (actuates when concentration is below thresh old).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
Threshold Alarm 2: • Can be configured as high (actuates when concen-
tration is above threshold), or low (actuates when concentration is below threshold).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
System Alarm: Actuates when DC power supplied to circuits is
unacceptable in one or more parameters. Permanently configured as fail-safe and latching. Cannot be de­feated. Actuates if self test fails.
Teledyne Analytical Instruments
Part I: 2-7
2 Installation Model 6000B
To reset a System Alarm during installation, discon-
nect power to the instrument and then reconnect it Further detail can be found in chapter 3, section 4-5.
Digital Remote Cal Inputs Remote Zero and Span Inputs: The REMOTE SPAN and RE-
MOTE ZERO inputs are on the DIGITAL INPUT terminal block. They accept 0 V (OFF) or 24 V dc (ON) for remote control of calibration (See Remote Calibration Protocol below.)
Zero: Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the ZERO mode. Either side may be grounded at the source of the signal. 0 to 1 volt across the terminals allows ZERO mode to terminate when done. A synchronous signal must open and close the external zero valve appropriately. See Remote Probe Connector at end of section 3.3. (With the -C option, the internal valves automati­cally operate synchronously).
Span: Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the SPAN mode. Either side may be grounded at the source of the signal. 0 to 1 volt across the terminals allows SPAN mode to terminate when done. A synchronous signal must open and close the external span valve appropriately. See Remote Probe Connector at end of section 3.3. (With the -C option, the internal valves automati­cally operate synchronously.)
Cal Contact: This relay contact is closed while analyzer is spanning
and/or zeroing. (See Remote Calibration Protocol below.)
Remote Calibration Protocol: To properly time the Digital Remote
Cal Inputs to the Model 6000B Analyzer, the customer's controller must monitor the Cal Relay Contact.
When the contact is OPEN, the analyzer is analyzing, the Remote Cal
Inputs are being polled, and a zero or span command can be sent.
When the contact is CLOSED, the analyzer is already calibrating. It
will ignore your request to calibrate, and it will not remember that request.
Once a zero or span command is sent, and acknowledged (contact closes), release it. If the command is continued until after the zero or span is complete, the calibration will repeat and the Cal Relay Contact (CRC) will close again.
For example:
2-8: Part I
Teledyne Analytical Instruments
Photometric Analyzer Part I: Control Unit
1) Test the CRC. When the CRC is open, Send a zero command until the CRC closes (The CRC will quickly close.)
2) When the CRC closes, remove the zero command.
3) When CRC opens again, send a span command until the CRC closes. (The CRC will quickly close.)
4) When the CRC closes, remove the span command.
When CRC opens again, zero and span are done, and the sample is
being analyzed.
Note: The Remote Bench connector (paragraph 3.3) provides signals
to ensure that the zero and span gas valves will be controlled synchronously.
Range ID Relays: Four dedicated RANGE ID CONTACT relays .
The first three ranges are assigned to relays in ascending order—Low range is assigned to RANGE 1 ID, Medium range is assigned to RANGE 2 ID, and High range is assigned to RANGE 3 ID.
Network I/O: A serial digital input/output for local network protocol.
At this printing, this port is not yet functional. It is to be used in future versions of the instrument.
RS-232 Port: The digital signal output is a standard RS-232 serial
communications port used to connect the analyzer to a computer, terminal, or other digital device. The pinouts are listed in Table 2-3.
Table 2-3: RS-232 Signals
RS-232 Sig RS-232 Pin Purpose
DCD 1 Data Carrier Detect RD 2 Received Data TD 3 Transmitted Data DTR 4 Data Terminal Ready COM 5 Common DSR 6 Data Set Ready RTS 7 Request to Send CTS 8 Clear to Send RI 9 Ring Indicator
Teledyne Analytical Instruments
Part I: 2-9
2 Installation Model 6000B
The data sent is status information, in digital form, updated every two
seconds. Status is reported in the following order:
The concentration in percent
The range is use (HI< MED< LO)
The span of the range 0-100%, etc)
Which alarm - if any - are disabled (AL-x DISABLED)
Which alarms - if any - are tripped (AL-x ON)
Each status output is followed by a carriage return and line feed.
Three input functions using RS-232 have been implemented to date.
They are described in Table 2-4.
Table 2-4: Commands via RS-232 Input
Command Description as<enter> Immediately starts an autospan. az<enter> Immediately starts an autozero. st<enter> Toggling input. Stops/Starts any status message output
from the RS-232, Until st<enter> is sent again.
The RS-232 protocol allows some flexibility in its implementation.
Table 2-5 lists certain RS-232 values that are required by the 6110B.
Table 2-5: Required RS-232 Options
Parameter Setting
Baud 2400
Byte 8 bits
Parity none
Stop Bits 1
Message Interval 2 seconds
Remote Bench and Solenoid Valves: The 6000B is a single-chassis
instrument. However, the REMOTE BENCH and SOLENOID RETURN connectors are provided on the interface PCB. The Remote Bench is wired at the factory as well as any optional solenoid valves included in the system.
2-10: Part I
Teledyne Analytical Instruments
Photometric Analyzer Part I: Control Unit
2.4 Testing the System
After The Control Unit and the Analysis Unit are both installed and
interconnected, and the system gas and electrical connections are complete, the system is ready to test. Before plugging either of the units into their respective power sources:
Check the integrity and accuracy of the gas connections. Make sure there are no leaks.
Check the integrity and accuracy of all electrical connections. Make sure there are no exposed conductors
Check that sample pressure typically between 0 and 30 psig, according to the requirements of your process.
Power up the system, and test it by performing the following operation:
1. Repeat the Self-Diagnostic Test as.
Teledyne Analytical Instruments
Part I: 2-11
2 Installation Model 6000B
2-12: Part I
Teledyne Analytical Instruments
Photometric Analyzer Operation 3
Operation
3.1 Introduction
Although the Model 6000B is usually programmed to your application at
the factory, it can be further configured at the operator level, or even, cautious- ly, reprogrammed. Depending on the specifics of the application, this might include all or a subset of the following procedures:
Setting system parameters:
Establish a security password, if desired, requiring Operator to log in (secure in safe file for referrence).
Establish and start an automatic calibration cycle, if desired.
Routine Operation:
Calibrate the instrument.
Choose autoranging or select a fixed range of analysis.
Set alarm setpoints, and modes of alarm operation (latching, fail-safe, etc).
Program/Reprogram the analyzer:
Define new applications.
Linearize your ranges.
If you choose not to use password protection, the default password is
automatically displayed on the password screen when you start up, and you simply press Enter for access to all functions of the analyzer.
3.2 Using the Data Entry and Function Buttons
Data Entry Buttons: The < > buttons select options from the menu
currently being displayed on the VFD screen. The selected option blinks.
When the selected option includes a modifiable item, the
can be used to increment or decrement that modifiable item.
Teledyne Analytical Instruments
ΔΔ
Δ∇ arrow buttons
ΔΔ
Part I 3-1
3 Operation Model 6000
The Enter button is used to accept any new entries on the VFD screen.
The Escape button is used to abort any new entries on the VFD screen that are not yet accepted by use of the Enter button.
Figure 4-1 shows the hierarchy of functions available to the operator via the
function buttons. The six function buttons on the analyzer are:
Analyze. This is the normal operating mode. The analyzer
monitors the thermal conductivity of the sample, displays the percent or parts-per-million of target gas or contamination, and warns of any alarm conditions.
System. The system function consists of nine subfunctions. Four of these are for ordinary setup and operation:
Setup an Auto-Cal
Assign Passwords
Log out to secure system
Initiate a Self-Test
Three of the subfunctions do auxiliary tasks:
Checking model and software version
Adjust electronic filter of the signal
Display more subfunctions
Display negative readings
Two of these are for programming/reprogramming the analyzer:
Define gas applications and ranges (Refer to programming section, or contact factory.)
Use the Curve Algorithm to linearize output. (Refer to programming section, or contact factory.)
Zero. Used to set up a zero calibration.
Span. Used to set up a span calibration.
Alarms. Used to set the alarm setpoints and determine whether
each alarm will be active or defeated, HI or LO acting, latching, and/or fail-safe.
Range. Used to set up three analysis ranges that can be
switched automatically with autoranging or used as individual fixed ranges.
Any function can be selected at any time by pressing the appropriate button (unless password restrictions apply). The order as presented in this manual is appropriate for an initial setup.
Each of these functions is described in greater detail in the following proce­dures. The VFD screen text that accompanies each operation is reproduced, at
3-2 Part I
Teledyne Analytical Instruments
Photometric Analyzer Operation 3
System
Dig_filt
SELF-TEST
PWD
LOGOUT
MORE
AUTOCAL
FILSOL
TRACK or
HOLD
CAL-HOLD
TIMER
Set Digital
Filter
Self-Test in
Progress
Enter
Password
Secure System
setup not allowed
Span/Zero status
and <>setup
Span/Zero
Solenoid or Filter
Set track or
hold output
Set cal. hold and
sample hold
timer
Self-Test
Results
Change Yes/No
Span/Zero timing
and on/off
Enter
Enter
Change
Password
Enter
Verify
Password
Enter
MORE
ALGORITHM
APPLICATION
OUTPUT:
4 or 20 MA
MORE
MODEL
SHOW.NEG
Select range
Select range
Set current
output
Display
Model/Version
Show Negative Reading
Display gas use
and range
Define
Application/
Range
Enter
Enter
Select
Verify/Setup
Enter
Verify data
Points
Auto/Manual
linear Cal.
Enter
Enter
Input/Output
Enter Span
gas value
Figure 3-1: Hierarchy of System Functions and Subfunctions
Enter
Enter
Teledyne Analytical Instruments
Part I 3-3
3 Operation Model 6000
the appropriate point in the procedure, in a Monospaced type style. Push­button names are printed in Oblique type.
3.3 The System Function
The subfuctions of the System function are described below. Specific
procedures for their use follow the descriptions:
Dig_Filt: Adjust how much digital filtering should be on the
signal
SELF-TEST: Performs a self-diagnostic test to check the
integrity of the power supplies, outputs, detector signal and preamplifier.
PWD: Login security system for accessing to the setup functions.
LOGOUT: Prevents an unauthorized tampering with analyzer
settings.
AUTOCAL: Set the automatic calibrated timer schedule for Zero
and Span cycling.
FILSOL: Select Span/Zero flag (filter) or Span/Zero solenoid
valve for calibration method.
TRACK: Set the system reading to be held or followed by the
concentration “gas or filter” during calibration.
CAL-HOLD-TIMER: Set the timing for calibration holding and
timing for the sample reading after return to analyze mode.
ALGORITHM: Linearize the output for nonlinear characteristic.
APPLICATION: Used to define the analysis ranges and
application (gas used).
MODEL: Displays model number and software version.
OUTPUT: 4-20 MA: Adjust 4 and 20 mA output.
SHOW_NEG: Whether to display negative readings or not; affects
analog output too. No negative readings is the default.
3.3.1 Setting up an AUTO-CAL
When proper automatic valving is connected, the Analyzer can cycle itself
through a sequence of steps that automatically zero and span the instrument.
3-4 Part I
Teledyne Analytical Instruments
Loading...
+ 68 hidden pages