HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING
SYSTEM.
PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.
HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST
FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.
ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING. BEFORE
CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/
MANAGER.
All Rights Reserved. No part of this manual may be reproduced, transmitted,
transcribed, stored in a retrieval system, or translated into any other language or computer
language in whole or in part, in any form or by any means, whether it be electronic,
mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of
Teledyne Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91749-
1580.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us
free from defects of material and of construction, and that our liability shall be limited to
replacing or repairing at our factory (without charge, except for transportation), or at
customer plant at our option, any material or construction in which defects become
apparent within one year from the date of shipment, except in cases where quotations or
acknowledgements provide for a shorter period. Components manufactured by others bear
the warranty of their manufacturer. This warranty does not cover defects caused by wear,
accident, misuse, neglect or repairs other than those performed by Teledyne or an authorized service center. We assume no liability for direct or indirect damages of any kind and
the purchaser by the acceptance of the equipment will assume all liability for any damage
which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our
apparatus, and to make any alterations in the dimensions, shape or weight of any parts, in
so far as such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its user, and serves as a tool by
which valuable data can be gathered. The information provided by the instrument may
assist the user in eliminating potential hazards caused by his process; however, it is
essential that all personnel involved in the use of the instrument or its interface, with the
process being measured, be properly trained in the process itself, as well as all instrumentation related to it.
The safety of personnel is ultimately the responsibility of those who control process
conditions. While this instrument may be able to provide early warning of imminent
danger, it has no control over process conditions, and it can be misused. In particular, any
alarm or control systems installed must be tested and understood, both as to how they
operate and as to how they can be defeated. Any safeguards required such as locks, labels,
or redundancy, must be provided by the user or specifically requested of Teledyne at the
time the order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The
purchaser is responsible for the training of personnel, for providing hazard warning
methods and instrumentation per the appropriate standards, and for ensuring that hazard
warning devices and instrumentation are maintained and operated properly.
Teledyne Analytical Instruments, the manufacturer of this instrument, cannot
accept responsibility for conditions beyond its knowledge and control. No statement
expressed or implied by this document or any information disseminated by the manufacturer or its agents, is to be construed as a warranty of adequate safety control under the
user’s process conditions.
ii
Teledyne Analytical Instruments
Photometric Analyzer
Table of Contents
Specific Model Information ................................. iv
Preface ................................................................ v
Part I: Control Unit.................................Part I: 1-1
Part II: Analysis Unit ............................. Part II: 1-1
Model 6000A Specifications .................................................. A-3
Teledyne Analytical Instruments
Part I: iii
Model 6000A Photometric Analyzer
iv: Part I
Teledyne Analytical Instruments
Photometric AnalyzerPart I: Control Unit
Introduction
1.1Overview
The Teledyne Analytical Instruments Model 6000A Control Unit,
together with a 6000A Analysis Unit, is versatile microprocessor-based
instrument.
Part I, of this manual covers the Model 6000A General Purpose Panel
Mount Control Unit. (The Analysis Unit is covered in Part II of this
manual.) The Control Unit is for indoor use in a nonhazardous environment
only. The Analysis Unit (or Remote Section) can be designed for a variety
of hazardous environments.
1.2Typical Applications
A few typical applications of the Model 6000A are:
•Oil in refinery waste water condensates Streams
•CL2, HC, SO2, H2S in stack gases or Liquid Streams
•Chemical reaction monitoring
•Product Color monitoring liquids
•Petrochemical process control
•Quality assurance
•Phenol in water
•Hazardous waste incineration
•CLO2, Hypochlorite monitoring
•F2 monitoring
1.3Main Features of the Analyzer
The Model 6000A Photometric Analyzer is sophisticated yet simple to
use. The main features of the analyzer include:
Teledyne Analytical Instruments
Part I: 1-1
1 Introduction Model 6000A
•A 2-line alphanumeric display screen, driven by microprocessor
electronics, that continuously prompts and informs the operator.
•High resolution, accurate readings of concentration from low
ppm levels through to 100%. Large, bright, meter readout.
•Versatile analysis over a wide range of applications.
•Microprocessor based electronics: 8-bit CMOS microprocessor
with 32 kB RAM and 128 kB ROM.
•Three user definable output ranges (from 0-1 ppm through
0-100 %) allow best match to users process and equipment.
•Calibration range for convenient zeroing or spanning.
•Auto Ranging allows analyzer to automatically select the proper
preset range for a given measurement. Manual override allows
the user to lock onto a specific range of interest.
•Two adjustable concentration alarms and a system failure alarm.
•Extensive self-diagnostic testing, at startup and on demand, with
continuous power-supply monitoring.
•RS-232 serial digital port for use with a computer or other digital
communication device.
•Analog outputs for concentration and range identification.
(0-1 V dc standard, and isolated 4–20 mA dc)
•Superior accuracy.
•Internal calibration (optional).
1.4Control Unit Inner Control Panel
The standard 6000A Control Unit is housed in a rugged panelmetal
case with all remote controls and displays accessible from the front control
panel. See Figure 1-1. The front control panel has a digital meter, an alphanumeric Vacuum Fluroscent Display (VFD), and buttons for operating the
analyzer.
1-2: Part I
Teledyne Analytical Instruments
Photometric AnalyzerPart I: Control Unit
Figure 1-1: Front of Panel Control Unit
Function Keys: ---- touch-sensitive membrane switches are used to change
the specific function performed by the analyzer:
•AnalyzePerform analysis for concentration content of a sample.
•SystemPerform system-related tasks (described in detail in chapter
3, Operation.).
•SpanSpan calibrate the analyzer.
•ZeroZero calibrate the analyzer.
•AlarmsSet the alarm setpoints and attributes.
•RangeSet up the 3 user definable ranges for the instrument.
Teledyne Analytical Instruments
Part I: 1-3
1 Introduction Model 6000A
Data Entry Keys: Six touch-sensitive membrane switches are used to
input data to the instrument via the alphanumeric VFD display:
•Left & Right ArrowsSelect between functions currently
displayed on the VFD screen.
•Up & Down ArrowsIncrement or decrement values of
functions currently displayed.
•EnterMoves VFD display on to the next screen in a series. If
none remains, returns to the
•Escape Moves VFD display back to the previous screen in a
series. If none remains, returns to the
Digital Meter Display: The meter display is a Light Emitting Diode
LED device that produces large, bright, 7-segment numbers that are legible
in any lighting. It is accurate across all analysis ranges. The 6000A models
produce continuous readout from 0-10,000 ppm and then switch to
continuous percent readout from 1-100 %.
Analyze
screen.
Analyze
screen.
Alphanumeric Interface Screen: The backlit VFD screen is an easyto-use interface between operator and analyzer. It displays values, options,
and messages for immediate feedback to the operator.
Power Button: The button switches the instrument power
between I (ON) and O (a Keep-Alive state). In the O state, the instrument’s
circuitry is operating, but there are no displays or outputs.
CAUTION: The power must be disconnected to fully
disconnect power from the instrument. When
chassis is exposed or when access door is open
and power cable is connected, use extra care to
avoid contact with live electrical circuits .
Access Door: For access to the electronics and interface panel, the front
panel swings open when the latch in the panel is pressed all the way in with
a narrow gauge tool. Accessing the main circuit board and other electronics
requires unfastening the rear panel screws and sliding the unit out of the
case.
1.5Control Unit Interface Panel
The Control Unit interface panel, shown in Figure 1-2, contains the
electrical terminal blocks for external inputs and outputs. The input/output
functions are described briefly here and in detail in the Installation chapter of
this manual.
1-4: Part I
Teledyne Analytical Instruments
Photometric AnalyzerPart I: Control Unit
Figure 1-2: Model 6000A Rear Panel
Teledyne Analytical Instruments
Part I: 1-5
1 Introduction Model 6000A
•Power ConnectionAC power source, 115VAC, 50/60 Hz
•Analog Outputs0-1 V dc concentration and 0-1 V dc
range ID. Isolated 4-20 mA dc and 4-20
mA dc range ID.
•Alarm Connections2 concentration alarms and 1 system
alarm.
•RS-232 PortSerial digital concentration signal output
and control input.
•Remote BenchProvides all electrical interconnect to the
Analysis Unit.
Remote Span/ZeroDigital inputs allow external control of
analyzer calibration.
•Calibration ContactTo notify external equipment that
instrument is being calibrated and
readings are not monitoring sample.
•Range ID ContactsFour separate, dedicated, range relay
contacts. Low, Medium, High, Cal.
•Network I/OSerial digital communications for local
network access. For future expansion.
Not implemented at this printing.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model 6000A is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.
1-6: Part I
Teledyne Analytical Instruments
Photometric AnalyzerPart I: Control Unit
Installation
Installation of Model 6000A Analyzers includes:
1. Unpacking, mounting, and interconnecting the Control Unit and
the Analysis Unit
2. Making gas connections to the system
3. Making electrical connections to the system
4. Testing the system.
This chapter covers installation of the Control Unit. (Installation of the
Analysis Unit is covered in Part II of this manual.)
2.1Unpacking the Control Unit
The analyzer is shipped with all the materials you need to install and
prepare the system for operation. Carefully unpack the Control Unit and
inspect it for damage. Immediately report any damage to the shipping agent.
2.2Mounting the Control Unit
The Model 6000A Control Unit is for indoor use in a general purpose
area. This Unit is NOT for any type of hazardous environments.
Figure 2-1 is an illustration of a Model 6000A standard Control Unit
front panel and mounting brackets as shown two mounting tabs are at the top
and two at the bottom of the units frame.
Teledyne Analytical Instruments
Part I: 2-1
2 InstallationModel 6000A
Figure 2-1: Front Panel of the Model 6000A Control Unit
All operator controls are mounted on the inner control panel "door",
which is hinged on the left edge and doubles as a door to provide access to
the internal components of the instrument. The door will swing open when
the button of the latch is pressed all the way in with a narrow gauge tool
(less than 0.18 inch wide), such as a small hex wrench or screwdriver
2-2: Part I
Teledyne Analytical Instruments
Photometric AnalyzerPart I: Control Unit
7.1
2
5 in
Figure 2-2: Required Front Door Clearance
Allow clearance for the door to open in a 90-degree arc of radius 11.75
inches. See Figure 2-2.
2.3Electrical Connections
Figure 2-3 shows the Control Unit interface panel. Connections for
power, communications, and both digital and analog signal outputs are
described in the following paragraphs. Wire size and maximum length data
appear in the Drawings at the back of this manual.
Figure 2-3: Interface Panel of the Model 6000A Control Unit
For safe connections, ensure that no uninsulated wire extends
outside of the terminal blocks. Stripped wire ends must insert completely
into terminal blocks. No uninsulated wiring should come in contact with
fingers, tools or clothing during normal operation.
Teledyne Analytical Instruments
Part I: 2-3
2 InstallationModel 6000A
Primary Input Power: The power supply requires a 115Vac, 50/60Hz
power source. The power cord receptacle and fuse block are located in the
same assembly. Insert the female plug end of the power cord into the power
cord receptacle.
CAUTION: Power is applied to the instrument's circuitry as
long as the instrument is connected to the power
source. The gray switch on the front panel is for
switching power on or off to the displays and out-
puts only.
Fuse Installation: The fuse block, at the right of the power cord
receptacle, accepts US or European size fuses. A jumper replaces the fuse in
whichever fuse receptacle is not used. Fuses are not installed at the factory.
Be sure to install the proper fuse as part of installation. (See Fuse Replace-ment in chapter 5, maintenance.)
Analog Outputs: There are four terminal blocks, one for each
analog output. There are two connectors per output with the polarity noted.
See Figure 2-5.
The outputs are:
0–1 V dc % of Range:Voltage rises linearly with increasing sample con-
centration, from 0 V at 0% to 1 V at 100%. (Full
scale = 100% programmed range.)
0–1 V dc Range ID:0.25 V = Low Range, 0.5 V = Medium Range,
0.75 V = High Range, 1 V = Cal Range.
4–20 mA dc % Range: (-M Option) Current increases linearly with increas-
ing sample concentration, from 4 mA at 0% to 20
mA at full scale 100%. (Full scale = 100% of
programmed range.)
4–20 mA dc Range ID: (-M Option) 8 mA = Low Range, 12 mA = Me-
dium Range, 16 mA = High Range, 20 mA = Cal
Range.
Figure 2-5: Analog Output Connections
2-4: Part I
Examples:
Teledyne Analytical Instruments
Photometric AnalyzerPart I: Control Unit
The analog output signal has a voltage which depends on the sample
concentration AND the currently activated analysis range. To relate the
signal output to the actual concentration, it is necessary to know what range
the instrument is currently on, especially when the analyzer is in the
autoranging mode.
The signaloutput for concentration is linear over currently selected
analysis range. For example, if the analyzer is set on a range that was
defined as 0-10 %, then the output would be as shown in Table 2-1.
To provide an indication of the range, a second pair of analog output
terminals are used. They generate a steady preset voltage (or current when
using the current outputs) to represent a particular range. Table 2-2 gives the
range ID output for each analysis range.
Teledyne Analytical Instruments
Part I: 2-5
2 InstallationModel 6000A
Table 2-2: Analog Range ID Output - Example
RangeVoltage (V)Current (mA)
LO0.258
MED0.5012
HI0.7516
CAL (0-25%)1.0020
Alarm Relays:
There are three alarm-circuit connectors on the alarm relays block
(under RELAY OUTPUTS) for making connections to internal alarm relay
contacts. Each provides a set of Form C contacts for each type of alarm.
Each has both normally open and normally closed contact connections. The
contact connections are indicated by diagrams on the rear panel. They are
capable of switching up to 3 ampers at 250 V AC into a resistive load
(Figure 2-6).
Figure 2-6: Types of Relay Contacts
The connectors are:
Threshold Alarm 1:• Can be configured as high (actuates when
concentration is above threshold), or low
(actuates when concentration is below thresh old).
• Can be configured as fail-safe or non-fail-safe.
2-6: Part I
Teledyne Analytical Instruments
Photometric AnalyzerPart I: Control Unit
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
Threshold Alarm 2:• Can be configured as high (actuates when concen-
tration is above threshold), or low (actuates when
concentration is below threshold).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
System Alarm:Actuates when DC power supplied to circuits is
unacceptable in one or more parameters. Permanently
configured as fail-safe and latching. Cannot be defeated. Actuates if self test fails.
To reset a System Alarm during installation, disconnect power to the instrument and then reconnect it
Further detail can be found in chapter 3, section 4-5.
Digital Remote Cal Inputs
Remote Zero and Span Inputs: The REMOTE SPAN and RE-
MOTE ZERO inputs are on the DIGITAL INPUT terminal block. They
accept 0 V (OFF) or 24 V dc (ON) for remote control of calibration (See
Remote Calibration Protocol below.)
Zero:Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the ZERO mode. Either side may be
grounded at the source of the signal. 0 to 1 volt across the
terminals allows ZERO mode to terminate when done. A
synchronous signal must open and close the external zero
valve appropriately. See Remote Probe Connector at end of
section 3.3. (With the -C option, the internal valves automatically operate synchronously).
Span:Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the
grounded at the source of the signal. 0 to 1 volt across the
terminals allows SPAN mode to terminate when done. A
synchronous signal must open and close the external span
valve appropriately. See Remote Probe Connector at end of
section 3.3. (With the -C option, the internal valves automatically operate synchronously.)
SPAN
mode. Either side may be
Cal Contact: This relay contact is closed while analyzer is spanning
and/or zeroing. (See Remote Calibration Protocol below.)
Teledyne Analytical Instruments
Part I: 2-7
2 InstallationModel 6000A
Remote Calibration Protocol: To properly time the Digital Remote
Cal Inputs to the Model 6000A Analyzer, the customer's controller must
monitor the Cal Relay Contact.
When the contact is OPEN, the analyzer is analyzing, the Remote Cal
Inputs are being polled, and a zero or span command can be sent.
When the contact is CLOSED, the analyzer is already calibrating. It
will ignore your request to calibrate, and it will not remember that request.
Once a zero or span command is sent, and acknowledged (contact
closes), release it. If the command is continued until after the zero or span is
complete, the calibration will repeat and the Cal Relay Contact (CRC) will
close again.
For example:
1) Test the CRC. When the CRC is open, Send a zero command
until the CRC closes (The CRC will quickly close.)
2) When the CRC closes, remove the zero command.
3) When CRC opens again, send a span command until the CRC
closes. (The CRC will quickly close.)
4) When the CRC closes, remove the span command.
When CRC opens again, zero and span are done, and the sample is
being analyzed.
Note: The Remote Bench connector (paragraph 3.3) provides signals
to ensure that the zero and span gas valves will be controlled
synchronously.
Range ID Relays: Four dedicated RANGE ID CONTACT relays .
The first three ranges are assigned to relays in ascending order—Low range
is assigned to RANGE 1 ID, Medium range is assigned to RANGE 2 ID,
and High range is assigned to RANGE 3 ID.
Network I/O: A serial digital input/output for local network protocol.
At this printing, this port is not yet functional. It is to be used in future
versions of the instrument.
RS-232 Port: The digital signal output is a standard RS-232 serial
communications port used to connect the analyzer to a computer, terminal, or
other digital device. The pinouts are listed in Table 2-3.
2-8: Part I
Table 2-3: RS-232 Signals
RS-232 Sig RS-232 Pin Purpose
DCD1Data Carrier Detect
Teledyne Analytical Instruments
Photometric AnalyzerPart I: Control Unit
RD2Received Data
TD3Transmitted Data
DTR4Data Terminal Ready
COM5Common
DSR6Data Set Ready
RTS7Request to Send
CTS8Clear to Send
RI9Ring Indicator
The data sent is status information, in digital form, updated every two
seconds. Status is reported in the following order:
•The concentration in percent
•The range is use (HI< MED< LO)
•The span of the range 0-100%, etc)
•Which alarm - if any - are disabled (AL-x DISABLED)
•Which alarms - if any - are tripped (AL-x ON)
Each status output is followed by a carriage return and line feed.
Three input functions using RS-232 have been implemented to date.
They are described in Table 2-4.
Table 2-4: Commands via RS-232 Input
CommandDescription
as<enter>Immediately starts an autospan.
az<enter>Immediately starts an autozero.
st<enter>Toggling input. Stops/Starts any status message output
from the RS-232, Until st<enter> is sent again.
The RS-232 protocol allows some flexibility in its implementation.
Table 2-5 lists certain RS-232 values that are required by the 6000A.
Teledyne Analytical Instruments
Part I: 2-9
2 InstallationModel 6000A
Table 2-5: Required RS-232 Options
ParameterSetting
Baud2400
Byte8 bits
Paritynone
Stop Bits1
Message Interval2 seconds
Remote Bench and Solenoid Valves: The 6000A is a single-chassis
instrument. However, the REMOTE BENCH and SOLENOID RETURN
connector is provided on the uper right corner of the backpanel as a 12 pin
connector. The Remote Bench is wired at the factory as well as any optional solenoid valves included in the system.
2.4Testing the System
After The Control Unit and the Analysis Unit are both installed and
interconnected, and the system gas and electrical connections are complete,
the system is ready to test. Before plugging either of the units into their
respective power sources:
•Check the integrity and accuracy of the gas connections. Make
sure there are no leaks.
•Check the integrity and accuracy of all electrical connections.
Make sure there are no exposed conductors
•Check that sample pressure typically between 0 and 30 psig,
according to the requirements of your process.
Power up the system, and test it by performing the following operation:
1. Repeat the Self-Diagnostic Test as.
2-10: Part I
Teledyne Analytical Instruments
Photometric AnalyzerOperation 3
Operation
3.1Introduction
Although the Model 6000A is usually programmed to your application at
the factory, it can be further configured at the operator level, or even, cautious-ly, reprogrammed. Depending on the specifics of the application, this might
include all or a subset of the following procedures:
•Setting system parameters:
•Establish a security password, if desired, requiring Operator
to log in (secure in safe file for referrence).
•Establish and start an automatic calibration cycle, if desired.
•Routine Operation:
•Calibrate the instrument.
•Choose autoranging or select a fixed range of analysis.
•Set alarm setpoints, and modes of alarm operation (latching,
fail-safe, etc).
•Program/Reprogram the analyzer:
•Define new applications.
•Linearize your ranges.
If you choose not to use password protection, the default password is
automatically displayed on the password screen when you start up, and you
simply press
Enter
for access to all functions of the analyzer.
3.2Using the Data Entry and Function
Buttons
Data Entry Buttons: The < > buttons select options from the menu
currently being displayed on the VFD screen. The selected option blinks.
When the selected option includes a modifiable item, the
can be used to increment or decrement that modifiable item.
Teledyne Analytical Instruments
∆∆
∆∇ arrow buttons
∆∆
Part I 3-1
3 Operation Model 6000A
The
Enter
Escape
The
not yet accepted by use of the
Figure 4-1 shows the hierarchy of functions available to the operator via the
function buttons. The six function buttons on the analyzer are:
•
•
Four of these are for ordinary setup and operation:
button is used to accept any new entries on the VFD screen.
button is used to abort any new entries on the VFD screen that are
Enter
Analyze.
monitors the thermal conductivity of the sample, displays the
percent or parts-per-million of target gas or contamination, and
warns of any alarm conditions.
System.
•Setup an Auto-Cal
•Assign Passwords
•Log out to secure system
This is the normal operating mode. The analyzer
The system function consists of nine subfunctions.
button.
•Initiate a Self-Test
Three of the subfunctions do auxiliary tasks:
•Checking model and software version
•Adjust electronic filter of the signal
•Display more subfunctions
Two of these are for programming/reprogramming the analyzer:
•Define gas applications and ranges (Refer to programming
section, or contact factory.)
•Use the Curve Algorithm to linearize output. (Refer to
programming section, or contact factory.)
•
Zero
. Used to set up a zero calibration.
•
Span.
•
Alarms.
each alarm will be active or defeated, HI or LO acting, latching,
and/or fail-safe.
•
Range.
switched automatically with autoranging or used as individual
fixed ranges.
Used to set up a span calibration.
Used to set the alarm setpoints and determine whether
Used to set up three analysis ranges that can be
Any function can be selected at any time by pressing the appropriate button
(unless password restrictions apply). The order as presented in this manual is
appropriate for an initial setup.
Each of these functions is described in greater detail in the following procedures. The VFD screen text that accompanies each operation is reproduced, at
3-2 Part I
Teledyne Analytical Instruments
Photometric AnalyzerOperation 3
System
Dig_filt
SELF-TEST
PWD
LOGOUT
MORE
AUTOCAL
FILSOL
TRACK or
HOLD
CAL-HOLD
TIMER
Set Digital
Filter
Self-Test in
Progress
Enter
Password
Secure System
setup not allowed
Span/Zero status
and <>setup
Span/Zero
Solenoid or Filter
Set track or
hold output
Set cal. hold and
sample hold timer
Self-Test
Results
Change
Yes/No
Span/Zero timing
and on/off
Enter
Enter
Change
Password
Enter
Verify
Password
Enter
MORE
ALGORITHM
APPLICATION
MODEL
OUTPUT:
4 or 20 MA
Select range
Select range
Display
Model/Version
Set current
output
Display gas use
and range
Define
Application/Range
Enter
Enter
Select
Verify/Setup
Enter
Verify data
Points
Auto/Manual
linear Cal.
Enter
Enter
Input/Output
Enter Span
gas value
Figure 3-1: Hierarchy of System Functions and Subfunctions
Enter
Enter
Teledyne Analytical Instruments
Part I 3-3
3 Operation Model 6000A
the appropriate point in the procedure, in a Monospaced type style. Push-button
names are printed in
Oblique
type.
3.3The
The subfuctions of the
procedures for their use follow the descriptions:
•Dig_Filt: Adjust how much digital filtering should be on the
signal
•SELF-TEST: Performs a self-diagnostic test to check the integrity
of the power supplies, outputs, detector signal and preamplifier.
•PWD: Login security system for accessing to the setup functions.
•LOGOUT: Prevents an unauthorized tampering with analyzer
settings.
•AUTOCAL: Set the automatic calibrated timer schedule for Zero
and Span cycling.
•FILSOL: Select Span/Zero flag (filter) or Span/Zero solenoid
valve for calibration method.
•TRACK: Set the system reading to be held or followed by the
concentration “gas or filter” during calibration.
•CAL-HOLD-TIMER: Set the timing for calibration holding and
timing for the sample reading after return to analyze mode.
System
System
Function
function are described below. Specific
•ALGORITHM: Linearize the output for nonlinear characteristic.
•APPLICATION: Used to define the analysis ranges and application
(gas used).
•MODEL: Displays model number and software version.
•OUTPUT: 4-20 MA: Adjust 4 and 20 mA output.
3.3.1Setting up an AUTO-CAL
When proper automatic valving is connected, the Analyzer can cycle itself
through a sequence of steps that automatically zero and span the instrument.
Note: Before setting up an AUTO-CAL, be sure you understand the
Zero
and
Span
functions as described in section 4.4, and
follow the precautions given there.
3-4 Part I
Teledyne Analytical Instruments
Photometric AnalyzerOperation 3
Note: If you require highly accurate AUTO-CAL timing, use external
AUTO-CAL control where possible. The internal clock in the
Model 6000A is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.
To setup an AutoCal cycle:
Choose
System
from the Function buttons. TheVFD will display five
subfunctions.
DIG_FILT SELFTEST
PWD LOGOUT MORE
Select MORE and press the Enter Key
AUTOCALFILSOLHOLD
CAL-HOLD-TIMERMORE
Use < > arrows to blink AUTOCAL, and press
Enter
. A new screen
for ZERO/SPAN set appears.
ZERO in Ød Øh off
SPAN in Ød Øh off
Press < > arrows to blink ZERO (or SPAN), then press
Enter
again. (You
won’t be able to set OFF to ON if a zero interval is entered.) A Span Every
... (or Zero Every ...) screen appears.
Zero schedule: OFF
Day: Ød Hour: Øh
∆∆
Use
∆∇ arrows to set an interval value, then use < > arrows to move to the
∆∆
start-time value. Use
∆∆
∆∇ arrows to set a start-time value.
∆∆
To turn ON the SPAN and/or ZERO cycles (to activate AUTOCAL): Press
System
again, choose AUTOCAL, and press
Enter
again. When the ZERO/
SPAN values screen appears, use the < > arrows to blink the ZERO (or SPAN)
and press
∆∆
∆∇ arrows to set the OFF/ON field to ON. You can now turn these fields ON
∆∆
Enter
to go to the next screen. Use < > to select OFF/ON field. Use
because there is a nonzero span interval defined.
If instrument is turned off, the next time the instrument is powered, the
instrument will automatically perform a calibration cycle after 3 minutes of
entering the sample mode if AUTOCAL functions were on prior to shut down.
Teledyne Analytical Instruments
Part I 3-5
Loading...
+ 65 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.