Teledyne 465L User Manual

Page 1
OPERATION MANUAL
MODEL 465L
OZONE MONITOR
Teledyne Advanced Pollution Instrumentation (TAPI)
San Diego, CA 92121-5201
Phone: 800-324-5190 (toll free in North America)
858-657-9800 (direct)
FAX: 858-657-9816
EMAIL: api-sales@teledyne.com
WEB SITE: http://www.teledyne-api.com/
© 2007-2012 TAPI 05509F DCN6409 Teledyne Advanced Pollution Instrumentation 09 May 2012
Page 2
Page 3
ABOUT TELEDYNE ADVANCED POLLUTION INSTRUMENTATION (TAPI)
Teledyne Advanced Pollution Instrumentation, Inc. (TAPI) is a worldwide market leader in the design and manufacture of precision analytical instrumentation used for air quality monitoring, continuous emissions monitoring, and specialty process monitoring applications. Founded in San Diego, California, in 1988, TAPI introduced a complete line of Air Quality Monitoring (AQM) instrumentation, which comply with the United States Environmental Protection Administration (EPA) and international requirements for the measurement of criteria pollutants, including CO, SO2, NOX and Ozone.
Since 1988 TAPI has combined state-of-the-art technology, proven measuring principles, stringent quality assurance systems and world class after-sales support to deliver the best products and customer satisfaction in the business.
For further information on our company, our complete range of products, and the applications that they serve , please visit www.teledyne-api.com or contact sales@teledyne-api.com.
Notice of Copyright
© 2007-2012 Teledyne Advanced Pollution Instrumentation. Inc. All rights reserved.
Trademarks
All trademarks, registered trademarks, brand names or product names appearing in this document are the property of their respective owners and are used herein for identification purposes only.
05509F DCN6409 i
Page 4
This page intentionally left blank.
ii 05509F DCN6409
Page 5
IMPORTANT SAFETY INFORMATION
Important safety messages are provided throughout this manual for the purpose of avoiding personal injury or instrument damage. Please read these messages carefully. Each safety message is associated with a safety alert symbol and placed throughout this manual and inside the instrument. The symbols with messages are defined as follows:
WARNING: Electrical Shock Hazard
HAZARD: Strong oxidizer
GENERAL WARNING/CAUTION: Read the accompanying message
for specific information.
CAUTION: Hot Surface Warning
Do Not Touch: Touching some parts of the instrument without
protection or proper tools could result in damage to the part(s) and/or the instrument.
Technician Symbol: All operations marked with this symbol are to be performed by qualified maintenance personnel only.
Electrical Ground: This symbol inside the instrument marks the central safety grounding point for the instrument.
CAUTION
GENERAL SAFETY HAZARD
This instrument should only be used for the purpose and in the manner described in this manual. If you use it in a manner other than that for which it was intended, unpredictable behavior could ensue with possible hazardous consequences.
NEVER use any gas analyzer to sample combustible gas(es).
05509F DCN6409 iii
Page 6
Teledyne API Model 465L O3 Monitor Operation Manual
CONSIGNES DE SÉCURITÉ
Des consignes de sécurité importantes sont fournies tout au long du présent manuel dans le but d’éviter des blessures corporelles ou d’endommager les instruments. Veuillez lire attentivement ces consignes. Chaque consigne de sécurité est représentée par un pictogramme d’alerte de sécurité; ces pictogrammes se retrouvent dans ce manuel et à l’intérieur des instruments. Les symboles correspondent aux consignes suivantes :
AVERTISSEMENT : Risque de choc électrique
DANGER : Oxydant puissant
AVERTISSEMENT GÉNÉRAL / MISE EN GARDE : Lire la consigne
complémentaire pour des renseignements spécifiques
MISE EN GARDE : Surface chaude Ne pas toucher : Toucher à certaines parties de l’instrument sans
protection ou sans les outils appropriés pourrait entraîner des dommages aux pièces ou à l’instrument.
Pictogramme « technicien » : Toutes les opérations portant ce symbole doivent être effectuées uniquement par du personnel de maintenance qualifié.
Mise à la terre : Ce symbole à l’intérieur de l’instrument détermine le point central de la mise à la terre sécuritaire de l’instrument.
MISE EN GARDE
Cet instrument doit être utilisé aux fins décrites et de la manière décrite dans ce manuel. Si vous utilisez cet instrument d’une autre manière que celle pour laquelle il a été prévu, l’instrument pourrait se comporter de façon imprévisible et entraîner des conséquences dangereuses.
NE JAMAIS utiliser un analyseur de gaz pour échantillonner des gaz combustibles!
iv 05509F DCN6409
Page 7
Teledyne API Model 465L O3 Monitor Operation Manual
ABOUT THIS MANUAL
This manual is comprised of multiple documents as listed below.
Part
Number
05590 M465L Operation Manual (this manual) 05919 Rack Mount Spare Parts List (integrated into this manual as Section 12.1) 05910 NEMA Spare Parts List (integrated into this manual as Section 12.2)
Name/Description
NOTE
We recommend that this manual be read in its entirety before any attempt is made to operate the
instrument.
05509F DCN6409 v
Page 8
Teledyne API Model 465L O3 Monitor Operation Manual
REVISION HISTORY
M465L Operation Manual, PN05509
2012 May 09, Rev F, DCN 6409: technical and administrative updates 2011 August 2, Rev E, DCN 6198: technical and administrative updates 2009 October, Rev D, DCN 5522: technical and administrative updates 2009 October, Rev D, DCN 5522: technical and administrative updates 2009 June, Rev C, DCN 5456: administrative udpates
2009 February, Rev B, DCN 5371: administrative updates 2007, Rev A, Initial Release
vi 05509F DCN6409
Page 9
TABLE OF CONTENTS
ABOUT THIS MANUAL.........................................................................................i
SAFETY MESSAGES...........................................................................................ii
ABOUT TELEDYNE ADVANCED POLLUTION INSTRUMENTATION (TAPI).....I
1. INTRODUCTION..........................................................................................13
1.1. PREFACE................................................................................................13
2. SPECIFICATIONS AND AGENCY APPROVALS.......................................15
2.1. S
2.2. A
2.2.1. Safety..............................................................................................................16
2.2.2. EMC...............................................................................................................16
2.2.3. Other Type Certifications..............................................................................17
3. PRINCIPLE OF OPERATION......................................................................19
4. GETTING STARTED ...................................................................................21
4.1. PNEUMATIC CONNECTIONS.......................................................................23
4.1.1. Sample Gas Connection.................................................................................24
4.1.2. Exhaust Connection.......................................................................................28
4.2. ELECTRICAL I/O CONNECTIONS................................................................29
4.2.1. Location of I/O Connectors in the NEMA Configuration..............................29
4.2.2. Analog Output................................................................................................29
4.2.3. Status Outputs................................................................................................29
4.2.4. Hi-Current Relay Outputs .............................................................................31
4.2.5. Multi-Stream Relay Outputs (Optional) ........................................................32
4.2.6. Digital Serial Interfaces ................................................................................33
PECIFICATIONS......................................................................................15
PPROVALS ............................................................................................16
5. FREQUENTLY ASKED QUESTIONS .........................................................35
6. OPTIONS.....................................................................................................37
6.1. S
TREAM SELECTOR (3 OR 6 STREAM).......................................................37
6.2. DISPOSABLE SAMPLE INLET FILTERS.........................................................37
6.3. RACK MOUNT SLIDES ..............................................................................37
6.4. ELECTRONIC FLOWMETER........................................................................37
6.5. S
AMPLE CONDITIONER.............................................................................37
6.6. PERMEATION GAS DRYER........................................................................37
7. OPERATION ................................................................................................39
7.1. F
RONT PANEL OVERVIEW.........................................................................39
7.1.1. Ozone Concentration Display .......................................................................39
7.1.2. Menu Navigation Keypad..............................................................................39
7.1.3. Stream Selector Status/Control (Optional)....................................................40
05509F DCN6409 vii
Page 10
Teledyne API Model 465L O3 Monitor Operation Manual
7.1.4. Concentration Alarm Status/Reset.................................................................40
7.1.5. Status LEDs ...................................................................................................41
7.2. FRONT PANEL MENUS..............................................................................41
7.2.1. Menu Navigation Basics................................................................................41
7.2.2. Editing Numeric Values.................................................................................42
7.2.3. DIAG Menu....................................................................................................45
7.2.4. VIEW Menu....................................................................................................51
7.2.5. CONFIG Menu ..............................................................................................51
7.2.6. SETUP-VARS Menu.......................................................................................52
7.2.7. SETUP-STREAM Menu.................................................................................53
7.3. MULTI-STREAM OPERATION .....................................................................54
8. DIGITAL COMMUNICATIONS....................................................................55
8.1. D
8.2. C
IGITAL COMMUNICATIONS OVERVIEW......................................................55
HANGING COM PORT HARDWARE CONFIGURATION ..................................55
8.3. RS-232 CONFIGURATION.........................................................................57
8.4. RS-485 CONFIGURATION.........................................................................58
8.5. ETHERNET CONFIGURATION.....................................................................59
8.5.1. Hardware Setup for Configuring the Ethernet Module.................................59
8.5.2. Verifying Network Hardware Connection.....................................................59
8.5.3. Determining a Dynamically Assigned IP Address (DHCP Mode)................60
8.5.4. Manually Assigning a Static IP Address .......................................................60
8.6. MODBUS REGISTER MAPS.....................................................................61
8.6.1. MODBUS Commands Supported ..................................................................61
8.6.2. Notes on MODBUS registers.........................................................................61
8.6.3. Real-Time Concentration and Status Registers (Read Only)........................62
8.6.4. Instrument Setup and Configuration Registers (Read / Write)......................63
9. MAINTENANCE AND ADJUSTMENTS......................................................66
9.1. MAINTENANCE SCHEDULE........................................................................66
9.2. I
NSTRUMENT LAYOUT...............................................................................67
9.3. REPLACING INTERNAL PARTICULATE FILTER ..............................................68
9.4. UV LAMP ADJUSTMENT............................................................................69
9.5. UV
9.6. S
LAMP REPLACEMENT..........................................................................70
AMPLE PUMP REPLACEMENT..................................................................70
9.7. SENSOR VALVE REPLACEMENT.................................................................72
9.8. WATER DROP-OUT FILTER (OPTION) MAINTENANCE .................................73
9.9. CONFIGURING THE ANALOG OUTPUT.........................................................73
9.10. C
LEANING EXTERIOR SURFACES OF THE M465L....................................74
9.11. BOOT LOADER OPERATION...................................................................74
9.11.1. Boot Loader – Hardware Setup ..................................................................75
9.11.2. Boot Loader – Software Setup.....................................................................75
10. TROUBLESHOOTING.............................................................................78
10.1. R
EFERENCE DRAWINGS .......................................................................78
10.1.1. Pneumatic Diagram....................................................................................78
viii 05509F DCN6409
Page 11
Teledyne API Model 465L O3 Monitor Operation Manual
10.1.2. Interconnect Diagram.................................................................................79
10.2. TROUBLESHOOTING USING FRONT PANEL STATUS LED’S OR STATUS OUTPUTS 80
10.2.1. Sensor OK ...................................................................................................81
10.2.2. Invalid Reading...........................................................................................81
10.2.3. Check Lamp.................................................................................................81
10.2.4. Pneumatic Error..........................................................................................81
11. CALIBRATION.........................................................................................84
11.1. FACTORY CALIBRATION ........................................................................ 84
11.2. ZERO CALIBRATION..............................................................................84
12. SPARE PARTS........................................................................................86
12.1. M465L
RACK MOUNT SPARE PARTS LIST..............................................86
12.2. M465L NEMA SPARE PARTS LIST........................................................87
FIGURES
Figure 4-1: NEMA AC Power Connection to Monitor ....................................22
Figure 4-2: Pneumatic Connections, Rack Mount Configuration.................23
Figure 4-3: Pneumatic Connections, NEMA Configuration, Single Stream 23 Figure 4-4: Pneumatic Connections, NEMA Configuration, Multi-Stream...24 Figure 4-5. M465L Pneumatic Diagram with Sample Conditioner and Permeation
Gas Dryer Options (in single-stream configuration).......................26
Figure 4-6: Rear Panel Electrical I/O Connections........................................28
Figure 4-7: NEMA Electrical I/O Connections................................................28
Figure 4-8: Status Output Schematic .............................................................30
Figure 7-1: Front Panel Display.......................................................................39
Figure 7-2: Front Panel Menu Diagram (Page 1)............................................43
Figure 7-3: Front Panel Menu Diagram (Page 2)............................................44
Figure 7-4: Span Cal Menu.............................................................................. 49
Figure 7-5: Flow Cal Menu...............................................................................50
Figure 8-1: Com Configuration Jumper Settings (SW2) ...............................56
Figure 9-1: Instrument Layout, Rack Mount Configuration.......................... 67
Figure 9-2: Instrument Layout, NEMA Configuration....................................68
05509F DCN6409 ix
Page 12
Teledyne API Model 465L O3 Monitor Operation Manual
Figure 9-3: Sensor Detail................................................................................. 72
Figure 9-4: Mainboard – Analog Output Configuration.................................74
Figure 10-1: Pneumatic Diagram, 3 Stream Configuration...........................78
Figure 10-2: Interconnect Diagram.................................................................79
x 05509F DCN6409
Page 13
Teledyne API Model 465L O3 Monitor Operation Manual
TABLES
Table 4-1: Status Output Definitions ..............................................................30
Table 4-2: Relay Functions for Single Stream Config...................................31
Table 4-3: Relay Functions for Multi-Stream Configs...................................31
Table 4-4: Multi-Stream Relays.......................................................................32
Table 7-1: Alarm LED’s.....................................................................................40
Table 7-2: Status LED’s ...................................................................................41
Table 7-3: Analog Step Test Values................................................................45
Table 7-4: Signal I/O List..................................................................................46
Table 7-5: VIEW Menu Parameters..................................................................51
Table 7-6: CONFIG Menu Parameters .............................................................51
Table 7-7: VARS List........................................................................................52
Table 7-8: STREAM Properties........................................................................53
Table 8-1: RS-232 Connector Details..............................................................57
Table 8-2: RS-232 Port Setup ..........................................................................57
Table 8-3: RS-485 Connector Details..............................................................58
Table 8-4: RS-485 Port Setup ..........................................................................58
Table 8-5: Ethernet LED Definitions ...............................................................59
Table 8-6: Supported MODBUS Commands ..................................................61
Table 9-1: Maintenance Schedule...................................................................66
Table 10-1: Status LED/Output Definitions....................................................80
05509F DCN6409 xi
Page 14
Teledyne API Model 465L O3 Monitor Operation Manual
This page intentionally left blank.
xii 05509F DCN6409
Page 15
1. INTRODUCTION
1.1. Preface
The Teledyne API Model 465L is a UV photometric ozone monitor designed and manufactured for single/multi-channel industrial hygiene ozone analysis. It is a very stable analyzer, ideal for critical low-level PPM safety and process applications. Whether you have selected the rack mount or wall mount unit in NEMA 4X enclosure, the M465L is designed to be highly reliable and simple to operate with easy setup and user-friendly menu driven software that can also be set and monitored remotely.
As the M465L was designed for stability, reliability and simplicity, it includes built-in tests and diagnostics to facilitate maximum uptime. In addition to a full one year warranty (see Section
1.2), Teledyne API is pleased to provide you with any support required so that you may utilize our equipment to the fullest extent. Our full time customer service department is available to answer your questions.
05509F DCN6409 13
Page 16
Teledyne API Model 465L O3 Monitor Operation Manual
This page intentionally left blank.
14 05509F DCN6409
Page 17
2. SPECIFICATIONS AND AGENCY
APPROVALS
2.1. Specifications
User selectable to any full scale range from: 1 PPM to 500 PPM
Ranges
Measurement Units Zero Noise < .0015 PPM (rms) Span Noise < .5% of reading (rms) (above 0.1 PPM)
1000 PPB to 10,000 PPB 2000 µg/m3 to 10,000 µg/m 100 PPHM to 10,000 PPHM (optional
configuration) PPM, PPB, µg/m3, PPHM (optional
configuration)
3
Lower Detectable Limit < .003 PPM (rms) Accuracy +/- 1% of full scale range
<30 sec
Response Time (95%) Sample Flow Rate 0.8 L/min Temperature Range 5-45o C Humidity Range 10-90% RH, Non-Condensing
Dimensions (H x W x D)
Weight
Power 100-240 VAC, 47-63 Hz, 74W max Environmental Conditions
(<60 sec with the Sample Conditioner option)
5.22” x 19.0" x 15.3" (133 mm x 483 mm x 388 mm) – Rack Mount Version
16.01” x 15.6” x 6.90” (407mm x 396mm x 175mm) – with NEMA 4X enclosure
12.6 lb. (5.73 kg) - Rack Mount Version
15.2 lb. (6.91 kg) – with NEMA 4X enclosure
Installation Category (Overvoltage Category) II Pollution Degree 2
Maximum Operating Altitude 2000 meters
05509F DCN6409 15
Page 18
Teledyne API Model 465L O3 Monitor Operation Manual
Analog Output, Voltage Mode 0 - 5V Isolated Analog Output, 4-
20mA Mode System OK Status Relay SPDT (Form C) Dry Contact, 250 VAC, 5A
Global Alarm Relays
Instrument Status Outputs
Multi-Stream Specifications:
Minimum Stream Duration (Cycle Time) 1.0 min
Stream Alarm Relays (HI and HI-HI)
Stream ID Relays
Maximum voltage between outputs and ground 60V peak
SPDT (Form C) Dry Contact, 250 VAC, 5A (x2, HI and HI-HI)
6 - SPST Dry Contact, 50VDC, 250mA Resistive Load
SPST Dry Contact, 50VDC, 250mA Resistive Load (2 per stream)
SPST Dry Contact, 50VDC, 250mA Resistive Load (1 per stream)
2.2. Approvals
This section presents Safety and Electromagnetic Compatibility (EMC) compliance approvals for the Model 465L monitor.
2.2.1. Safety
IEC 61010-1:2001, Safety requirements for electrical equipment for measurement, control, and laboratory use.
CE: 2006/95/EC, Low-Voltage Directive North American:
cNEMKO (Canada): CAN/CSA-C22.2 No. 61010-1-04 NEMKO-CCL (US): UL No. 61010-1 (2nd Edition)
2.2.2. EMC
EN 61326-1 (IEC 61326-1), Class A Emissions/Industrial Immunity EN 55011 (CISPR 11), Group 1, Class A Emissions FCC 47 CFR Part 15B, Class A Emissions CE: 2004/108/EC, Electromagnetic Compatibility Directive
16 05509F DCN6409
Page 19
Teledyne API Model 465L O3 Monitor Operation Manual
2.2.3. Other Type Certifications
For additional certifications, please contact Customer Service:
Toll-free
Phone: Phone:
Fax:
Email:
800-324-5190
858-657-9800 858-657-9816
api-customerservice@teledyne.com
05509F DCN6409 17
Page 20
Teledyne API Model 465L O3 Monitor Operation Manual
This page intentionally left blank.
18 05509F DCN6409
Page 21
3. PRINCIPLE OF OPERATION
The detection of ozone molecules is based on absorption of 254 nm UV light due to an internal electronic resonance of the O3 molecule. The Model 465 uses a mercury lamp constructed so that a large majority of the light emitted is at the 254nm wavelength. Light from the lamp shines down a hollow quartz tube that is alternately filled with sample gas, then filled with gas scrubbed to remove ozone. The ratio of the intensity of light passing through the scrubbed gas to that of the sample forms a ratio I/Io. This ratio forms the basis for the calculation of the ozone concentration.
The Beer-Lambert equation, shown below, calculates the concentration of ozone from the ratio of light intensities.
Where:
C
9
10
ln
O
3
273
o
92.29
inHg
o
I = Intensity of light passed through the sample Io = Intensity of light through sample free of ozone
= absorption coefficient
= path length
C = concentration of ozone in ppb
O
3
T = sample temperature in degrees Kelvin P = pressure in inches of mercury
As can be seen the concentration of ozone depends on more than the intensity ratio. Temperature and pressure influence the density of the sample. The density changes the number of ozone molecules in the absorption tube which impacts the amount of light removed from the light beam. These effects are addressed by directly measuring temperature and pressure and including their actual values in the calculation. The absorption coefficient is a number that reflects the inherent ability of ozone to absorb 254 nm light. Most current measurements place this value at 308 cm-1 atm-1 at STP. The value of this number reflects the fact that ozone is a very efficient absorber of UV radiation which is why stratospheric ozone protects the life forms lower in the atmosphere from the harmful effects from solar UV radiation. Lastly, the absorption path length determines how many molecules are present in the column of gas in the absorption tube.
05509F DCN6409 19
Page 22
Teledyne API Model 465L O3 Monitor Operation Manual
The intensity of light is converted into a voltage by a high resolution A/D (analog-to-digital) converter. The digitized signal and other variables are used by the CPU to compute the concentration using the above formula.
About every 2.5 seconds the M465 completes a measurement cycle consisting of a 1 second wait period for the sample tube to flush, followed by a 150 ms measurement of the UV light intensity to obtain I. The sample valve is switched to admit scrubbed sample gas for 1 second, followed by a 150 ms measurement of the UV light intensity to obtain Io. Measurement of the Io every 2.5 seconds eliminates instrument drift due to changing intensity of the lamp caused by aging and dirt.
20 05509F DCN6409
Page 23
4. GETTING STARTED
The Model 465 is shipped with the following standard equipment:
Power cord (Rack mount version only) Instruction manual.
Upon receiving the Model 465 please do the following:
1. Verify that there is no apparent shipping damage. (If damage has occurred please advise
shipper first, then Teledyne API.)
2. When installing the Model 465, allow a minimum of 4 inches for clearance at the back of
the instrument (rack mount only) and 1 inch of clearance on each side for proper ventilation.
3. Connect sample inlet line(s) to the sample port on rear panel or bottom panel of
instrument. See Figure 4-2 or Figure 4-3 for rear panel pneumatic connections.
CAUTION
Connect the exhaust fitting on the rear panel (See Figure 4-2) to a suitable vent outside the monitor area.
4. Ensure that there are no communication devices connected until after start-up is
complete.
5. For the NEMA configuration, the electrical connection must be hard-wired to the 3-pin
connector on the mainboard labeled J23. AC power connection to the monitor should be made with 12-14 AWG stranded copper wire, connected to the monitor as follows:
NOTE
Sample tubing made from an inert material such as Teflon should
be used to minimize sample degradation.
Earth Ground (green): Connect the earth ground wire to the screw terminal lug on the
chassis as shown in the Figure below. Do not connect the earth ground wire directly to the PCB connector on the mainboard PCB.
Line (Black): Connect the Line wire to the connector on the mainboard PCB
labeled “Line”.
Neutral (White): Connect the Neutral wire to the connector on the mainboard PCB
labeled “Neutral”.
05509F DCN6409 21
Page 24
Teledyne API Model 465L O3 Monitor Operation Manual
Figure 4-1: NEMA AC Power Connection to Monitor
6. Connect the power cord to an appropriate power outlet (see the serial number tag for
correct voltage and frequency).
CAUTION
Verify that the instrument is set up for proper line voltage and frequency. Observe local electrical codes when connecting power to the monitor.
7. Turn on the M465L by switching the switch on the left side of the rear panel, or applying
power to the monitor (NEMA configuration.) The front panel display should light with a sequence of messages, including the instrument serial number and software revision. (When the instrument first powers up, the display will show ‘xxxx’, indicating that the instrument is in start-up mode, waiting for the UV detector readings to stabilize)..
8. After the start-up period, the instrument will begin displaying the ozone concentration
and the optional stream selector will begin operating, if installed.
22 05509F DCN6409
Page 25
Teledyne API Model 465L O3 Monitor Operation Manual
4.1. Pneumatic Connections
Figure 4-2: Pneumatic Connections, Rack Mount Configuration
Figure 4-3: Pneumatic Connections, NEMA Configuration, Single Stream
05509F DCN6409 23
Page 26
Teledyne API Model 465L O3 Monitor Operation Manual
Figure 4-4: Pneumatic Connections, NEMA Configuration, Multi-Stream
4.1.1. Sample Gas Connection
Sample gas connections are made to the ¼” tube fittings on the rear panel. For single stream monitors, use the fitting labeled “Sample Inlet.” For multi-stream monitors, the sample tubing is connected to the fittings labeled Stream 1, Stream 2, etc (See Figure 4-2, Figure 4-3, Figure 4-4)
1/4" O.D. FEP (Teflon) tubing is recommended to connect the sample source to the monitor. Any fittings used in the sample lines should be constructed of stainless steel or Teflon.
Teledyne API recommends that the length of tubing connecting the sample points to the monitor should be kept to 50 feet or less.
NOTE
For the sampling lines, use only ¼” O.D. FEP tubing. For best analytical performance, the
tubing should be pre-conditioned to ozone prior to installation to minimize ozone loss in the
sampling lines. Pre-conditioned ¼” FEP tubing is available from Teledyne API (Part number
02639)
24 05509F DCN6409
Page 27
Teledyne API Model 465L O3 Monitor Operation Manual
4.1.1.1. Sample Filtration
The most important factor in maintaining an accurate and stable monitoring system for ozone is the cleanliness of entire pneumatic system leading up to (and including) the sensor assembly inside the instrument. Provision must be made for keeping dust and other particulate matter out of the monitor and upstream sample delivery tubing. The monitor is equipped with an internal particulate filter down-stream of the selector manifold.
In addition, it is recommended that inlet filters be installed at the inlet of each sample line. These filters should be constructed of PTFE or other inert material to avoid degradation of the ozone concentration. Appropriate disposable inlet filters can be purchased from TAPI. See the Spare Parts list in Appendix A of this manual for ordering information.
4.1.1.2. Sample System Design in Humid Environments
Care must be taken in the design of the sampling system to ensure that water vapor does not condense in the sampling lines or in the instrument itself. The sample lines should be routed to avoid large temperature gradients along the lines. If sample is brought in from a hot, humid area into a much cooler area, then low power heating tape should be wrapped around the lines to keep the temperature of the sample lines up and avoid condensation. The lines should also not have loops or low points where water could collect.
For applications that require sampling of very humid gas, a Sample Conditioner and/or a Permeation Gas Dryer may be required to lower sample gas humidity. A pneumatic block diagram with the locations of these M465L options is shown in Figure 4-5, followed by a description of each. Please contact the Teledyne API Sales Department for additional information.
05509F DCN6409 25
Page 28
Teledyne API Model 465L O3 Monitor Instruction Manual
Figure 4-5. M465L Pneumatic Diagram with Sample Conditioner and Permeation Gas Dryer Options (in single-
stream configuration)
26 05509F DCN6409
Page 29
Teledyne API Model 465L O3 Monitor Instruction Manual
Sample Conditioner Option
The Sample Conditioner option is a water drop-out filter and a permeation gas dryer, designed for high-humidity environments where the sampling gas has been saturated with water vapor and may contain small amounts of condensed water. This coalescing filter removes not only liquid, but also particulates with a Teflon filter element. Refer to Section 9.8 for maintenance procedures.
Due to the additional volume added by the water drop-out filter, there is an exception to the Response Time in the M465L specifications:
The Response Time of an M465L instrument equipped with the Sample Conditioner option, will be <60 seconds to 95%.
CAUTION
The Sample Conditioner will NOT protect the unit from damage by large amounts of liquid, such as that ingested through immersion of the sample line.
Furthermore, it is important to closely track the amount of water accumulating in the Water Drop-out Filter after the monitor has been installed and running: if excessive condensation is occurring in the sample lines leading to the monitor, then heat-traced lines should also be used to minimize condensation
The Sample Conditioner option includes a Permeation Gas Dryer, which is available as an alternate option (described next) if humidity is the only concern and there is no risk of condensation.
Permeation Gas Dryer Option
The Permeation Gas Dryer serves to remove water vapor from sampling gas in a humid environment. It is an alternative option when there is no risk of condensation. The permeation material is a Teflon Type compound and has only a minimal effect on the ozone concentration in the sample stream. When the M465L is equipped with this option, the unit has been calibrated at the factory to compensate for any ozone loss in the permeation dryer.
05509F DCN6409 27
Page 30
Teledyne API Model 465L O3 Monitor Operation Manual
4.1.2. Exhaust Connection
Connect a ¼” OD vent line to the “Exhaust” fitting on the rear panel. This line must be vented outdoors or to an appropriate discharge system capable of handling ozone. Local regulations regarding the discharge of ozone to the atmosphere must be observed. Note that the monitor may be equipped with an internal ozone scrubber for removal of ozone from the exhaust stream. For safety reasons, the exhaust port must still be vented appropriately as described above.
Figure 4-6: Rear Panel Electrical I/O Connections
Figure 4-7: NEMA Electrical I/O Connections
28 05509F DCN6409
Page 31
Teledyne API Model 465L O3 Monitor Operation Manual
4.2. Electrical I/O Connections
4.2.1. Location of I/O Connectors in the NEMA
Configuration
In the NEMA configuration of the M465L, the I/O connectors are located internally in the instrument. See Figure 4-7 for their location. The relative position of the connectors and the pinout information is the same as the Rack-Mount configuration shown in Figure 4-6. Connection to these terminals is usually made via a conduit connection to the NEMA enclosure. A standard conduit penetration may be added to one of the walls of the enclosure for this purpose, or the existing penetration may be used.
4.2.2. Analog Output
The analog output is located at the “General I/O” connector (See Figure 4-6) and is labeled “ANA.” The analog output can be configured for 0-5V voltage or 4-20 mA current output. See Section 9.8 for details on verifying or changing the analog output hardware configuration.
The analog output is automatically scaled to the range of the instrument, i.e. if the monitor range is 1000 PPB, then 5V would correspond to 1000 PPB and 0V would correspond to 0 PPB. See Section 7.2.7 for information on changing the range of the monitor.
The M465L has an analog output step-test mode that can be used for verifying the operation of the analog output and calibration of external measurement equipment. This mode is described in Section 7.2.3.1.
4.2.3. Status Outputs
There are four instrument status outputs located at the “General I/O” connector. These status outputs mirror the state of the four diagnostic LEDs on the front panel.
Electrically these outputs are SPST dry contact relay closures with a common contact, labeled ‘C’ on the rear panel. They are capable of switching up to 50 VDC, 250mA maximum current. There are no polarity requirements, so the outputs can be used to sink or source current from the ‘C’ pin, within the maximum voltage and current specified. Schematically, the outputs look as shown in Figure 4-8 below:
05509F DCN6409 29
Page 32
Teledyne API Model 465L O3 Monitor Operation Manual
Figure 4-8: Status Output Schematic
The outputs are defined as follows:
Table 4-1: Status Output Definitions
Status Output #
Name Normal Operating
State
1 Sensor OK On 2 Invalid Reading Off 3 Check Lamp Off 4 Pneumatic Error Off
30 05509F DCN6409
Page 33
Teledyne API Model 465L O3 Monitor Operation Manual
4.2.4. Hi-Current Relay Outputs
Three form C relay outputs are provided on the rear panel on the nine pin connector (See Figure 4-6). The relays are labeled “Relay 1,” “Relay 2,” and “Relay 3.”
The relays are dry contact type form C (SPDT) relays with Normally Open (N.O.), Normally Closed (N.C.) and Common (COM) contacts. The relays are capable of driving loads of up to 250VAC, 5A. For maximum contact life, the relays should only be used to drive resistive loads. Contact life may be dramatically shortened if inductive loads are driven without any provision for minimizing high voltage “inductive kick” that can occur.
These relays have slightly different functions depending on whether the monitor is a single stream or multi-stream configuration.
Table 4-2: Relay Functions for Single Stream Config
Relay # Function Normal
Operating State
1 System OK Diagnostic On 2 Hi Alarm Off 3 Hi-Hi Alarm Off
Table 4-3: Relay Functions for Multi-Stream Configs
Relay # Function Normal
Operating State
1 System OK Diagnostic On 2 Global Hi Alarm* Off 3 Global Hi-Hi Alarm* Off * The state of the global alarm represents all the Hi or
Hi-Hi alarm states OR’d together, i.e. if the Hi alarm for one or more of the sample streams is activated, then the Global Hi alarm will be activated.
05509F DCN6409 31
Page 34
Teledyne API Model 465L O3 Monitor Operation Manual
4.2.5. Multi-Stream Relay Outputs (Optional)
For instruments configured with the optional stream-selector, an additional set of output relays are provided on the rear panel and are labeled “Multi-Stream Outputs” (See Figure 4-6.)
Electrically these outputs are SPST dry contact relay closures with a common contact, labeled ‘C’ on the rear panel. They are capable of switching up to 50 VDC, 250mA maximum current. There are no polarity requirements, so the outputs can be used to sink or source current from the ‘C’ pin. One common terminal is provided for the six relays on each connector (See Figure 4-6.)
Each available stream has a set of 3 relays providing the following information:
Table 4-4: Multi-Stream Relays
Relay # Function Meaning
1 Stream ID Stream X is active 2 Hi Alarm Stream X Hi
Alarm is active
3 Hi-Hi Alarm Stream X Hi-Hi
Alarm is active
32 05509F DCN6409
Page 35
Teledyne API Model 465L O3 Monitor Operation Manual
4.2.6. Digital Serial Interfaces
There is one external communications port available on the Model 465 that can be configured as RS232, RS485 (Half Duplex,) or Ethernet (10Mbit.) This section provides only a brief overview of the digital serial connections. See Section 8 for more detailed information on configuring and using the digital serial interfaces.
4.2.6.1. RS232 Connection
With the instrument configured for RS232, connection can be made to the DB-9 connector, labeled “RS232/485”, or to the three “COM” pins on the “General I/O” connector. See Section
8.2 for additional details. When using RS232, a 9 pin serial cable can be used to connect the instrument to a standard PC
(personal computer) serial port. TAPI can provide this cable, please see the spare parts list in Appendix A.
4.2.6.2. RS485 Connection
With the instrument configured for RS485, connection can be made to the DB-9 connector, labeled “RS232/485”, or to the three “COM” pins on the “General I/O” connector. See Section
8.2 for additional details.
4.2.6.3. Ethernet Connection
With the instrument configured for Ethernet, connection should be made to the RJ-45 port labeled “Ethernet.” Connection can be made using a standard Category 5 or better network patch cable. TAPI can provide an appropriate cable, please see the spare parts list in Appendix A. See Section 8.2 for additional details.
05509F DCN6409 33
Page 36
Teledyne API Model 465L O3 Monitor Operation Manual
his page intentionally left blank
34 05509F DCN6409
Page 37
5. FREQUENTLY ASKED QUESTIONS
1. Q: How long does the UV Lamp last?
A: The typical lifetime is about 2-3 years.
05509F DCN6409 35
Page 38
Teledyne API Model 465L O3 Monitor Operation Manual
This page intentionally left blank.
36 05509F DCN6409
Page 39
6. OPTIONS
Following are descriptions of options available for the M465L.
6.1. Stream Selector (3 or 6 Stream)
This option allows the M465 to measure ozone at up to six different sampling points by using an internal stream selector. 3 or 6 channel stream selectors are available. This option includes digital outputs for HI and HI-HI alarms and a channel ID for each channel in the stream selector.
6.2. Disposable Sample Inlet Filters
This option consists of 50 disposable filters for installation at the inlet end of all sample lines.
6.3. Rack Mount Slides
Rack mount slides can be attached to the sides of the instrument, allowing it to slide completely out of a RETMA rack, facilitating maintenance activities.
6.4. Electronic Flowmeter
The M465 can be fitted with an internal electronic flowmeter that enables the instrument to display the flow digitally on the front panel and provides a Pneumatic Error if the flow is outside its normal limits.
6.5. Sample Conditioner
For humid environments where there is risk of water condensation in the sample lines, this option serves to filter out liquid and particulates and includes a permeation gas dryer as well. For details, see “Sample Conditioner” in Section 4.1.1.2.
6.6. Permeation Gas Dryer
This option removes water vapor from the sample stream in environments where humidity is a concern but where condensation is not a factor. For details, see “Permeation Gas Dryer” in Section 4.1.1.2.
05509F DCN6409 37
Page 40
Teledyne API Model 465L O3 Monitor Operation Manual
38 05509F DCN6409
Page 41
7. OPERATION
7.1. Front Panel Overview
The M465L front panel provides a number of functions in addition to displaying the current ozone concentration being measured. Figure 7-1 below shows the display and keypad portion of the front panel.
Figure 7-1: Front Panel Display
7.1.1. Ozone Concentration Display
The 4-character LED display in the center of the front panel normally shows the current ozone concentration being measured by the M465. While displaying concentration, one of the unit-of­measure indicators (i.e., PPM, PPB, or UGM) should be backlit, indicating the current measurement units. If none of these indicators are backlit, then the display is in Menu mode or Startup mode and is not displaying the concentration. If the display is in Menu mode, the display can be returned to Concentration mode by repeatedly pressing the CFG button until one of the unit-of-measure indicators is lit. Note that even when the display is in Menu mode, normal measurement and stream selector operation will continue in the background.
7.1.2. Menu Navigation Keypad
The menu navigation keypad (CFG, ENT and UP/DOWN buttons) is used to navigate the various menu items available on the M465L. The menus allow the user to view operational parameters of the instrument as well as view and modify the configuration of the instrument. See Section 7.2 for more detailed information.
05509F DCN6409 39
Page 42
Teledyne API Model 465L O3 Monitor Operation Manual
7.1.3. Stream Selector Status/Control (Optional)
The optional multi-channel stream selector can be controlled via the front panel. The stream selector operates in two modes, Auto and Manual. The Auto mode is normally the default mode. The ‘AUTO’ and ‘MAN’ LED’s show the current mode. To switch modes, press and hold the
SELECT button for at least 2 seconds. When in Manual mode, momentarily pressing the SELECT button will switch to the next stream.
The stream selector status is shown by the six LED’s labeled ‘SAMPLE POINT.’ One of the LED’s should always be lit, showing the current channel being measured. Note that for a three­channel instrument, only streams 1-3 can be measured. See Section 7.3 for additional details on the operation of the stream selector.
When the instrument is in its default configuration, there is a Manual mode timeout feature that will automatically return the stream selector to Auto mode when no keys are pressed for 10 minutes. This is the recommended setting for instruments used for critical safety applications. This feature prevents the instrument from being inadvertently left in Manual mode, and subsequent loss of monitoring data from the other, unmeasured channels. For applications where Manual mode is the preferred mode of operation, this timeout feature can be turned off. See Section 7.2.6 for details on changing this setting.
7.1.4. Concentration Alarm Status/Reset
The M465L front panel displays the concentration alarm status of all channels being sampled. The row of LED’s labeled ‘ALARM STATUS’ shows the alarm status for each channel as follows:
Table 7-1: Alarm LED’s
Alarm Status LED Color Alarm State
Green No alarm active Amber ‘HI’ alarm active Red ‘HI-HI’ alarm active
NOTE
If the alarms are configured for latching operation, then the RESET button can be used to attempt to reset all alarms. Note that alarms can only be reset if the concentration measured for a specific channel has subsequently dropped below the alarm limit. If the alarms are configured for non-latching operation, they will be automatically cleared when the concentration falls below the alarm threshold. The RESET button has no effect when the alarms are configured for non­latching operation.
See Sections 7.2.7, 4.2.4, and 4.2.5 for additional details on concentration alarms.
40 05509F DCN6409
Page 43
Teledyne API Model 465L O3 Monitor Operation Manual
7.1.5. Status LEDs
There are four Status LED’s on the front panel to the right of the concentration display. These LED’s provide a quick view of the instrument’s Status as well as an indication of any problems that may exist. The normal state for these LED’s after the initial 20 minute warm-up period is as follows:
Table 7-2: Status LED’s
Status LED Normal State
Sensor OK ON Invalid Reading OFF Check Lamp OFF Pneumatic Error OFF
If any of these LED’s are in a different state than shown above, then some adjustment or troubleshooting should be performed. See Section 10.1 for more detailed information.
7.2. Front Panel Menus
The M465L has an interactive menu structure that can be operated from the front panel. From within this menu structure, the operator can view real-time parameters such as temperature and pressure, view configuration information, and edit setup parameters such as alarm limits.
Figure 7-2 below shows the menu structure as well as the buttons that are used to navigate it. Starting with the display in Concentration mode, press CFG to enter Menu mode. At this point
the unit-of-measure LED should turn off and the first level of the Menu tree will be active. Pressing the UP/DOWN arrows will now select from the items available at this menu level. Press ENT to enter one of these sub-menus or press CFG to back out of this menu level back to Concentration mode.
7.2.1. Menu Navigation Basics
1. Press CFG from Concentration mode to enter Menu mode.
2. Once in Menu mode, CFG will move backwards one level in the menu structure and
ENT will move forward.
3. Each menu level contains an EXIT selection. Selecting EXIT and pressing ENT will exit
the Menu mode and return the display to Concentration mode.
4. The Menu mode will timeout if no buttons are pressed within 30 seconds, and the
instrument will automatically return to Concentration mode.
05509F DCN6409 41
Page 44
Teledyne API Model 465L O3 Monitor Operation Manual
7.2.2. Editing Numeric Values
At certain places in the menu structure the operator can edit numeric values. The menu buttons behave somewhat differently when in this mode.
Press CFG to select a digit to edit. The selected digit will blink. Next press the UP/DOWN arrows to increment/decrement the selected digit. Press CFG again to select another digit. Pressing CFG at the last digit will wrap back to the first digit.
Once all the digits have been edited, press ENT to accept the new value. To abort the editing process without changing the original value, press and hold the CFG button
until the display returns to the previous menu level. Note that limit checking is enforced while editing values and changes that would result in invalid
values are ignored.
42 05509F DCN6409
Page 45
Teledyne API Model 465L O3 Monitor Operation Manual
Concentration
Display
SETUP
CFG
ENTCFG
STREAM
CFG CFG
ENT
STR1
ENT
DURATION
DIAG
VIEW
ENT
ENT
VARS
EXIT
ANALOG
STEP TEST
SIGNAL I/O
RESET
MEMORY
ZERO CAL
SPAN CAL
FLOW CAL
STORE
VARS
RESTORE
VARS
EXIT
MEAS
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
SEE VARS
DETAIL, NEXT
PAGE
See Section
7.2.3.1
See Section
7.2.3.2
See Section
7.2.3.3
See Section
7.2.3.4
See Section
7.2.3.5
See Section
7.2.3.6
Currently Disabled
Currently Disabled
STRn
HI ALARM
HIHI ALARM
ON/OFF HI
ALARM
ON/OFF
HIHI ALARM
RANG
ENABLE
EXIT
REF
See Section
6.2.4
ENT
CONFIG
EXIT
ENT
PRESSURE
LAMP TEMP
SAMPLE
TEMP
FLOW
EXIT
S/N
COMM
ADDRESS
STREAM
COUNT
REV
BREV
BOOT
Figure 7-2: Front Panel Menu Diagram (Page 1)
05509F DCN6409 43
Page 46
Teledyne API Model 465L O3 Monitor Operation Manual
VARS
ENT
ALARM
MODE
STREAM
COUNT
DWEL 30.0
ALARM DELAY
MAN MODE
TIMEOUT
BC FILT
SIZE
FLOW
ENABLE
COMM
ADDRESS
UNIT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
LATCHING
NON-LATCHING
1,3,6
0005
ON
OFF
0010
ON
OFF
0001
PPM PPB
UGM
SCROLL
DELAY
USER
SLOPE
UNITS
TEMP
ENT
ENT
ENT
0025
1.000
NORMAL
EXIT
Figure 7-3: Front Panel Menu Diagram (Page 2)
44 05509F DCN6409
Page 47
Teledyne API Model 465L O3 Monitor Operation Manual
7.2.3. DIAG Menu
The DIAG menu (See Figure 7-2) contains functions that are useful for testing and configuring external equipment that
may be connected to the rear panel I/O of the instrument.
7.2.3.1. Analog Step-Test
When placed in this mode, the instrument will automatically step the analog output (See Section
4.2.2 for details on the analog output connection) through 5 points from 0 to 100 % and display the current value on the front panel. The operator can suspend the automatic cycling and manually control the stepping by pressing either one of the UP / DOWN arrow buttons. To resume automatic cycling, exit the function by pressing CFG and re-enter by pressing ENT.
The 5 points will correspond to the following analog output values depending on whether the output is configured for 0-5V or 4-20mA:
Table 7-3: Analog Step Test Values
Point 0-5V Output 4-20 mA Output
0% 0.00 V 4 mA 25% 1.25 V 8 mA 50% 2.50 V 12 mA 75% 3.75 V 16 mA 100% 5.00 V 20 mA
Press CFG to exit from the Analog Step-Test function.
05509F DCN6409 45
Page 48
Teledyne API Model 465L O3 Monitor Operation Manual
7.2.3.2. Signal I/O
The Signal I/O menu allows the operator to manually control the various relay outputs and control inputs available on the rear panel of the M465L. This function is useful for testing or debugging external control systems (i.e. dataloggers or PLC’s) to which the instrument may be connected.
See Section 4.2 for additional details on making connections to these I/O signals on the rear panel.
See Figure 7-2 for additional details on entering the Signal I/O menu.
Table 7-4: Signal I/O List
Signal Name Rear Panel Label Function
AUX 1 IN CTRL IN - 1 UNDEFINED AUX 2 IN CTRL IN - 1 UNDEFINED RELAY 1 RELAY 1 SENSOR OK RELAY 2 RELAY 2 GLOBAL HI ALARM RELAY 3 RELAY 3 GLOBAL HI-HI ALARM STAT OUT 1 STATUS OUT – 1 SENSOR OK STAT OUT 2 STATUS OUT – 2 Invalid Reading STAT OUT 3 STATUS OUT – 3 CHECK LAMP STAT OUT 4 STATUS OUT – 4 PNEUMATIC ERROR STAT AUX 1 AUX OUT – 1 UNDEFINED STAT AUX 2 AUX OUT – 2 UNDEFINED XRELAY 1 STRM 1 - 1 STREAM 1 ACTIVE XRELAY 2 STRM 1 – 2 STREAM 1 HI ALARM XRELAY 3 STRM 1 - 3 STREAM 1 HI-HI ALARM XRELAY 4 STRM 2 - 1 STREAM 2 ACTIVE XRELAY 5 STRM 2 – 2 STREAM 2 HI ALARM XRELAY 6 STRM 2 - 3 STREAM 2 HI-HI ALARM XRELAY 7 STRM 3 - 1 STREAM 3 ACTIVE XRELAY 8 STRM 3 – 2 STREAM 3 HI ALARM XRELAY 9 STRM 3 - 3 STREAM 3 HI-HI ALARM XRELAY 10 STRM 4 - 1 STREAM 4 ACTIVE
46 05509F DCN6409
Page 49
Teledyne API Model 465L O3 Monitor Operation Manual
Signal Name Rear Panel Label Function
XRELAY 11 STRM 4 – 2 STREAM 4 HI ALARM XRELAY 12 STRM 4 - 3 STREAM 4 HI-HI ALARM XRELAY 13 STRM 5 - 1 STREAM 5 ACTIVE XRELAY 14 STRM 5 – 2 STREAM 5 HI ALARM XRELAY 15 STRM 5 - 3 STREAM 5 HI-HI ALARM XRELAY 16 STRM 6 - 1 STREAM 6 ACTIVE XRELAY 17 STRM 6 – 2 STREAM 6 HI ALARM XRELAY 18 STRM 6 - 3 STREAM 6 HI-HI ALARM XVALVE 1 N/A STREAM 1 INPUT VALVE XVALVE 2 N/A STREAM 2 INPUT VALVE XVALVE 3 N/A STREAM 3 INPUT VALVE XVALVE 4 N/A STREAM 4 INPUT VALVE XVALVE 5 N/A STREAM 5 INPUT VALVE XVALVE 6 N/A STREAM 6 INPUT VALVE
7.2.3.3. Memory Reset
This function is currently disabled.
7.2.3.4. Zero Calibration
The zero calibration allows the instrument to calibrate its internal ozone offset factor. This should only be done with a source of zero air connected to the “Sample In” port of the M465L. Allow the instrument to stabilize on the zero air source before attempting to zero. This normally takes 10-15 minutes.
When entering the Zero Calibration menu, the prompt “ENT TO CAL” appears on the display. Simply confirm the calibration by pressing the ENT button to perform the calibration (to abort the calibration, press and hold the CFG button to return to the previous menu level.) After pressing ENT, the instrument will automatically exit the menu mode and return to Concentration mode. The concentration reading should quickly go to zero. Please note that while measuring zero air, a certain amount of noise or “dithering” of the concentration about the zero point will occur and is normal. This noise is typically 1-3 ppb (0.001 - 0.003 ppm) in magnitude.
05509F DCN6409 47
Page 50
Teledyne API Model 465L O3 Monitor Operation Manual
7.2.3.5. Span Calibration
The span calibration allows the instrument to calibrate its internal slope factor based on a known ozone concentration. This should be done with a source of ozone span gas connected to the “Sample In” port of the M465L. Allow the instrument to stabilize on the span gas source for a minimum of 1 hour before performing the span calibration.
NOTE
A Span Calibration should only be performed with a precision
source of ozone calibration gas, such as a Teledyne API Model
700 or Model 703 calibrator. Simple ozone generators without a
measurement feedback system should never be used for
performing span calibrations.
If you are unsure regarding the suitability of a particular source of
calibration gas, contact Customer Service at Teledyne API for
assistance.
Figure 7-4 below shows the Span Cal menu. After the instrument has stabilized on the source of
span gas, navigate to the Span Cal menu (See
Figure 7-2) and Press ENT. Next enter the Span Target concentration (the actual concentration of ozone being supplied to the monitor) and press ENT. The display will next show a confirmation menu, ENT TO SPAN. Press ENT to perform the Span Calibration, or CFG to abort back to the start. If the calibration is successful, the display will return to the concentration menu and the monitor reading should change adjust to read very close to the target value. If the calibration cannot be performed, an OUT OF RANGE error will be displayed. Press ENT to confirm and the display will return to the start of the Span Cal menu.
If the OUT OF RANGE error occurs, it means the Span Cal cannot be performed because it would result in an out of range slope value for the monitor. This means that either the sensor in the monitor is malfunctioning, causing improper readings, or the actual ozone concentration being supplied to the monitor is different than the target value being entered.
48 05509F DCN6409
Page 51
Teledyne API Model 465L O3 Monitor Operation Manual
Enter Span target
concentration (actual ozone
concentration being supplied
to monitor)
SPAN CAL 900
ENT
ENT
ENT TO
SPAN
CFG
ENT
SUCCESSFUL?
No
OUT OF
RANGE
ENT
Yes
RETURN TO
CONC MODE
Figure 7-4: Span Cal Menu
7.2.3.6. Flow Calibration (Optional Electronic Flowmeter only)
The flow calibration allows the instrument to calibrate its internal flowmeter reading. This should be done with a calibrated flowmeter connected to the “Sample In” port of the M465L. Figure 7-5 below shows the Flow Cal menu. After the instrument has stabilized on the source of span gas, navigate to the Flow Cal menu (See
Figure 7-2) and Press ENT. Next enter the Flow Cal target (the actual flow as measured at the sample inlet) and press ENT. The display will next show a confirmation menu, ENT TO CAL. Press ENT to perform the Flow Calibration, or CFG to abort back to the start. If the calibration is successful, the display will return to the concentration. If the calibration cannot be performed, an OUT OF RANGE error will be displayed. Press ENT to confirm and the display will return to the start of the Flow Cal menu.
If the OUT OF RANGE error occurs, it means the Flow Cal cannot be performed because it would result in an out of range flow slope value for the monitor. This means that either the flow sensor in the monitor is malfunctioning, causing improper readings, or the actual flow is different than the target value being entered.
05509F DCN6409 49
Page 52
Teledyne API Model 465L O3 Monitor Operation Manual
Enter Flow Cal target flow
(actual flow measured at
sample inlet)
FLOW CAL 0800
ENT
ENT
ENT TO CAL SUCCESSFUL?
CFG
ENT
Figure 7-5: Flow Cal Menu
No
OUT OF
RANGE
ENT
Yes
RETURN TO
CONC MODE
50 05509F DCN6409
Page 53
Teledyne API Model 465L O3 Monitor Operation Manual
7.2.4. VIEW Menu
The VIEW menu allows the operator to view various measurement parameters in real-time. This can be useful for diagnosing various instrument or system problems.
Note that these values are updated in real-time as they are repeatedly scrolled on the display.
Table 7-5: VIEW Menu Parameters
Parameter Meaning Units Normal Range*
MEAS UV detector reading, measure cycle mV 250 – 1230 REF UV detector reading, reference
mV 250 – 1230
cycle PRESSURE Sample pressure psia 9.0 – 14.9 LAMP
UV lamp temperature ºC 50.0 – 52.0 TEMP
SAMPLE
Sample temperature ºC 5.0 –50.0 TEMP
FLOW Sample Gas Flow Rate cc/min 640 – 960 *After initial 20 minute warm-up period.
7.2.5. CONFIG Menu
The CONFIG menu displays specific information related to the instrument’s configuration. No changes are made within this menu.
Table 7-6: CONFIG Menu Parameters
CONFIG Meaning
S/N Serial number of M465L. (Should
match the serial number tag on the rear panel of the instrument).
COMM ADDRESS Displays the MODBUS slave ID for
the instrument.
STREAM COUNT Displays the number of streams
configured. (Always 1 for instruments
without the Multi-stream Option). REV Revision of instrument firmware BREV Revision of sensor firmware BOOT Revision of bootloader firmware
05509F DCN6409 51
Page 54
Teledyne API Model 465L O3 Monitor Operation Manual
L
N
7.2.6. SETUP-VARS Menu
The VARS menu allows viewing and editing of various global setup variables that effect how the M465L operates. These variables are stored in the instrument’s non-volatile memory. Table 7-7 lists these variables and their function.
Table 7-7: VARS List
VAR Meaning Default
S/N
Serial number of M465L. This number should match the serial number tag on the rear of the instrument.
ALARM MODE
STREAM COUNT*
DWEL Dwell value for stream selector operation. See Section 7.3
ALARM DELAY
MAN MODE TIMEOUT
BC FILT SIZE*
Controls the latching behavior of the concentration alarms. See Section 7.1.4 for additional details.
Specifies the number of streams available in the instrument. Note that the hardware in the instrument must be also configured to support the number of streams specified.
for additional details. Multi-stream units with the Sample Conditioner option
(Section 4.1.1.2), require additional dwell time to purge the filter between stream measurements.
Number of readings above alarm limit required to trigger alarms
Enables/disables the Manual mode timeout feature. See Section 7.1.3 for more detail.
Boxcar filter length for concentration filter. Increasing the size of the filter will smooth out the concentration readings but will also increase the response time.
Range
Value
N/A 1 – 9999
1 1,3,6
30.0 Seconds
45.0
Seconds
5 1 - 10
ON ON, OFF
10 1-30
ATCHING,
ON-LATCHING
0.0 – 30.0 Seconds
0.0 – 45.0 Seconds
FLOW ENABLE
COMM ADDRESS
Enables/disables operation of optional electronic flowmeter
MODBUS address for serial communications
OFF ON, OFF
1 1 - 20
UNIT Concentration unit-of-measure setting. PPM PPM, PPB, UGM SCROLL
DELAY
Inter-character delay for the front panel display, to increase speed (lower value) or decrease speed (higher value) the
20 6-100
display scroll.
USER SLOPE User-editable calibration slope value 1.000 0.8 – 1.0
UNITS TEMP
* Changing the value of these VARS forces the instrument to automatically reboot for the changes to take effect.
52 05509F DCN6409
Standard (20ºC) or Normal (0ºC) temperature selection for g/Nm3 units of measure.
NORMAL NORMAL
STANDARD
Page 55
Teledyne API Model 465L O3 Monitor Operation Manual
7.2.7. SETUP-STREAM Menu
The STREAM menu allows the operator to view and edit properties of each measurement stream available. For each stream the following properties are available:
Table 7-8: STREAM Properties
Property Meaning Default
Range
Value
DURATION Stream measurement duration for
AUTO stream-selector operation.
60.0 Seconds 60.0 – 360.0 Seconds
See Section 7.3 for additional details.
HI ALARM HI alarm trigger value 0.100 PPM .010 – 9.999
PPM*
HIHI ALARM HI-HI alarm trigger value 0.300 PPM .010 – 9.999
PPM*
ON/OFF HI ALARM
ON/OFF HI HI ALARM
RANG Concentration range for specified
HI alarm enable, enables HI alarm operation
HI-HI alarm enable, enables HI-HI alarm operation
stream. Used to scale the analog
ON ON/OFF
ON ON/OFF
1.0 PPM 1.0 – 500.0 PPM
output.
ENABLE Allows the choice of viewing or
ON ON-OFF skipping each stream in a multi­channel instrument when cycling through the streams in Auto or Manual modes. OFF = skip. At least one stream must be set to ON (enabled).
*
For low concentration instrument ranges. For certain range configurations, the
alarms may be set to higher values
05509F DCN6409 53
Page 56
Teledyne API Model 465L O3 Monitor Operation Manual
7.3. Multi-Stream Operation
This section describes the operation of the Multi-Stream M465 in Automatic mode (AUTO) and Manual mode (MAN.)
In AUTO mode the monitor is designed to continuously cycle through the enabled sampling streams (stream ENABLE set to ON), and skipping past the streams that are not enabled (stream ENABLE set to OFF). (The ENABLE parameter is accessed through the CFG-SETUP-STR[n] menu). In MAN mode the operator manually controls the cycling through enabled sampling streams by pressing the SELECT button. The stream cycling is accomplished by activating valves on the stream selector manifold on the rear panel to admit sample gas for a particular stream. Section 7.1.3 details the operation of the front panel stream selector interface.
When the M465L first switches to a new sample stream, the instrument goes through a 30­second dwell phase where sample gas is pulled from the new sample stream but no readings are taken. The purpose of this dwell phase is to purge out any “old” sample gas that has been sitting in the sample line while that stream was inactive. During the dwell phase the Sample Point LED on the display (See Figure 7-1) will blink and the concentration display will show the last concentration recorded for that stream. After the 30 seconds has elapsed, the Sample Point LED will stop blinking and the current concentration will be displayed. The current stream will now be monitored for the rest of the measurement Duration that has been set for that stream (see Section 7.2.7 for details on configuring the individual streams.) The actual time period that the stream is measured is the measurement Duration minus the 30 second dwell. So a 1 minute measurement Duration consists of 30 seconds of dwell followed by 30 seconds of measurement.
For systems that do not require a 30 second dwell between streams, this value can be changed using the DWEL Var. See Section 7.2.6 for additional details.
54 05509F DCN6409
Page 57
8. DIGITAL COMMUNICATIONS
8.1. Digital Communications Overview
The M465L comes equipped with a digital communications (Com) port that can be connected to a computer or digital data acquisition system. This Com port uses the standardized MODBUS® protocol and can be configured for RS232, RS485, or Ethernet (10 Mbit,) allowing for connections to a wide variety of devices and networks. See Section 8.2 for more information on configuring the Com port.
The Com port can be used for data acquisition, alarm triggering, and instrument configuration. All the functions that are available at the front panel of the instrument can also be performed over the Com port.
NOTE
MODBUS Interface Software for the M465L is available for
download from the TAPI website. Other TAPI software, such as
APICom, does not currently support communications with the
M465L.
8.2. Changing Com Port Hardware Configuration
WARNING - RISK OF ELECTRICAL SHOCK
HIGH VOLTAGE MAY BE PRESENT WHEN POWER IS CONNECTED TO THE INSTRUMENT!
THE OPERATIONS OUTLINED IN THIS SECTION ARE TO BE PERFORMED BY QUALIFIED MAINTENANCE PERSONNEL ONLY!
The Com port can be configured for RS232, RS485, or Ethernet. To change the Com port hardware configuration, perform the following steps:
1. Disconnect power from the M465L.
2. Remove the six screws and the top cover.
3. Locate the DIP switches labeled SW3 on the rear portion of the mainboard.
4. Turn ON the switches specified for the desired Com configuration (See Figure 8-1.) All
other switches on SW3 should be OFF.
5. Locate SW2 (DTE-DCE switch) on mainboard.
6. For RS-232 operation, set SW2 to either DTE or DCE (See Section 8.3.) For RS-485
operation, set SW2 to DCE.
05509F DCN6409 55
Page 58
Teledyne API Model 465L O3 Monitor Operation Manual
7. Re-install the instrument cover.
8. Re-connect power to the instrument.
9. The software portion of the Com port configuration will now automatically be completed
after the next instrument boot-up.
SW3
3
4
21
765
1098
RS232 Configuration
ETHERNET
SW3
RS485
Close Indicated Positions Only
3
4
21
765
RS232
1098
RS485 Configuration
ETHERNET
SW3
RS485
Close Indicated Positions Only
1234
567
RS232
8910
Ethernet Configuration
ETHERNET
RS485
Close Indicated Positions Only
RS232
Figure 8-1: Com Configuration Jumper Settings (SW2)
NOTE
To avoid start-up problems, ensure that there are no devices attempting communication with the instrument during start-up. Once the sign-on message is displayed on the front panel, it is safe to begin
communications.
56 05509F DCN6409
Page 59
Teledyne API Model 465L O3 Monitor Operation Manual
8.3. RS-232 Configuration
The RS-232 configuration is usually used when making a one-to-one connection between the instrument and a single computer or PLC. The communications protocol used for the RS-232 configuration is MODBUS RTU. For details on the MODBUS RTU specification, please see
http://www.modbus.org/.
With the instrument configured for RS-232, connection can be made to the DB-9 connector, labeled “RS232/485”, or to the three “COM” pins on the “General I/O” connector. Only one of these connections should be used, the other must be left un-connected. Table 8-1 below details the pinouts of these two connectors:
Table 8-1: RS-232 Connector Details
Function DB-9
Connector Pin (SW2 – DTE)
DB-9 Connector Pin (SW2 – DCE)
General I/O Connector
Transmit (Tx) 3 2 ‘T’ Receive (Rx) 2 3 ‘R’ Ground 5 5 ‘G’
Note that a DTE-DCE switch is provided so that the instrument can be connected to another device using either a straight-through or crossover (Null-Modem) cable. See Section 8.2 for details on changing this configuration.
The serial port setup for RS-232 configuration is shown in Table 8-2 below:
Table 8-2: RS-232 Port Setup
Property Value
Baud Rate 57600 Data Bits 8 Parity None Stop Bits 1 Flow Control None
05509F DCN6409 57
Page 60
Teledyne API Model 465L O3 Monitor Operation Manual
8.4. RS-485 Configuration
RS-485 is commonly used for factory automation busses with a computer or PLC acting as a master. The communications protocol used for the RS-485 configuration is MODBUS RTU. For details on the MODBUS RTU specification, please see http://www.modbus.org/.
With the instrument configured for RS-485, connection can be made to the DB-9 connector, labeled “RS232/485”, or to the three “COM” pins on the “General I/O” connector. Table 8-3 below details the pinouts of these two connectors:
Table 8-3: RS-485 Connector Details
Function DB-9
Connector Pin
General I/O Connector
(SW2 – DCE)
RS485 - A 2 ‘T’ RS485 - B 3 ‘R’ Ground 5 ‘G’
Note that for RS-485, the DTE-DCE switch should always be set to DCE. See Section 8.2 for details on changing this configuration.
The serial port setup for RS-485 configuration is shown in Table 8-4 below:
Table 8-4: RS-485 Port Setup
Property Value
Baud Rate 57600 Data Bits 8 Parity None Stop Bits 1 Flow Control None
58 05509F DCN6409
Page 61
Teledyne API Model 465L O3 Monitor Operation Manual
8.5. Ethernet Configuration
Ethernet is commonly used for factory automation busses with a computer or PLC acting as a master. The communications protocol used for the Ethernet configuration is MODBUS TCP/IP. For details on the MODBUS TCP/IP specification, please see http://www.modbus.org/.
A Windows software application, DeviceInstaller, is available for configuring the Ethernet module used in the M465L. This application is available for download here:
http://www.teledyne-api.com/software/
8.5.1. Hardware Setup for Configuring the Ethernet Module
To make the first-time configuration of the Ethernet module as easy as possible, the module comes from the factory with DHCP enabled. The M465L can simply be plugged into a network with a DHCP server and an IP Address should be automatically assigned to, and stored in, the device.
A simple network can easily be constructed for this purpose by using a small broadband router commonly used for home networks. The M465L and a PC can then be connected to the router and the router’s DHCP server will assign IP Addresses to both the PC and the M465L, enabling them to communicate.
8.5.2. Verifying Network Hardware Connection
With the M465L connected to a network and powered up, check the status of the two leds on the Ethernet connector on the rear panel. The left-hand LED should be lit; either amber or green indicating link status and the right-hand led should blink periodically, indicating activity on the network. See Table 8-5 below for more details on the status LED’s.
If neither of the LED’s are lit, then the Ethernet module may not be powered up or the instrument may not be connected to a valid network. Check the COM port configuration per Section 8.2 and make sure that it is set to “Ethernet.”
Table 8-5: Ethernet LED Definitions
Link LED (Left Side) Activity LED (Right Side) Color Meaning Color Meaning
Off No Link Off No Activity Amber 10 Mbps Connection Amber Half-Duplex Green 100 Mbps Connection Green Full-Duplex
05509F DCN6409 59
Page 62
Teledyne API Model 465L O3 Monitor Operation Manual
8.5.3. Determining a Dynamically Assigned IP Address (DHCP Mode)
The DeviceInstaller software application (See Section 8.5) can be used to search a network for instruments and determine the IP Address assigned by a DHCP server.
1. Click Start->Programs->Lantronix DeviceInstaller->DeviceInstaller. If your PC
has more than one network adapter, a message displays. Select an adapter and click OK.
2. Click on the Search icon. After a moment a list of instruments on the network will be
displayed. The M465L should be listed as an XPORT-IAP type device.
3. Double-click on the device in the right-hand section of the
of configuration parameters will be shown, including the IP Address. If multiple devices are shown, the correct one can be determined by matching the Hardware Address shown with the address printed on the label of the Ethernet module in the M465L. Note that the instrument cover must be removed to inspect this address in the instrument.
DeviceInstaller window; a list
8.5.4. Manually Assigning a Static IP Address
For applications where the IP Address must remain constant, a static IP Address can be manually assigned to the instrument and the DHCP function must be disabled.
1. Click Start->Programs->Lantronix DeviceInstaller->DeviceInstaller. If your PC
has more than one network adapter, a message displays. Select an adapter and click OK.
Note: If the unit already has an IP address (e.g., DHCP has assigned an IP address), click the Search icon and select the unit from the list of Lantronix device servers on the local network.
2. Click the Assign IP icon.
3. If prompted, enter the hardware address (on the product label) and click Next.
4. Select Assign a specific IP address and click Next.
5. Enter the IP address. The Subnet mask displays automatically based on the IP
address; if desired, you may change it. On a local network, you can leave the Default gateway blank (all zeros). Click Next.
6. Click the Assign button and wait several seconds until a confirmation message displays.
Click Finish.
7. Select the device from the main window list and select Ping from the Tools menu. The
Ping Device dialog box shows the IP address of the selected unit.
8. From the Tools menu, click the Ping button. The results display in the Status window.
Click the Clear Status button to clear the window so you can ping the device again.
Note: If you do not receive “Reply” messages, make sure the unit is properly attached to the network and that the IP address assigned is valid for the particular network segment you are working with. If you are not sure, check with your systems administrator.
9. Click the Close button to close the dialog box and return to the main window.
60 05509F DCN6409
Page 63
Teledyne API Model 465L O3 Monitor Operation Manual
8.6. MODBUS Register Maps
8.6.1. MODBUS Commands Supported
Table 8-6 lists the MODBUS commands that are supported by the M465L. Note that the “Write” commands will only work with registers that are configured as Read/Write, see Section 8.6.4. The “Read” commands can be used with any register.
Table 8-6: Supported MODBUS Commands
MODBUS Function Code
01 Read Coils 02 Read Discrete Inputs 03 Read Holding Registers 04 Read Input Registers 05 Write Single Coil 06 Write Single Register 15 Write Multiple Coils 16 Write Multiple Registers
Description
8.6.2. Notes on MODBUS registers
Concentration values in the MODBUS registers are always expressed in units of PPB (parts­per-billion,) regardless of the concentration units setting of the monitor.
05509F DCN6409 61
Page 64
Teledyne API Model 465L O3 Monitor Operation Manual
8.6.3. Real-Time Concentration and Status Registers (Read Only)
Decimal Address Description Register Type Start Bit # Bits
1 Stream 1 Concentration (PPB) Float Inverse 0 32 3 Stream 2 Concentration (PPB) Float Inverse 0 32 5 Stream 3 Concentration (PPB) Float Inverse 0 32 7 Stream 4 Concentration (PPB) Float Inverse 0 32 9 Stream 5 Concentration (PPB) Float Inverse 0 32 11 Stream 6 Concentration (PPB) Float Inverse 0 32 13 Stream 1 HI Alarm Status, 1= ON Byte 0 8 13 Stream 2 HI Alarm Status, 1= ON Byte 8 8 14 Stream 3 HI Alarm Status, 1= ON Byte 0 8 14 Stream 4 HI Alarm Status, 1= ON Byte 8 8 15 Stream 5 HI Alarm Status, 1= ON Byte 0 8 15 Stream 6 HI Alarm Status, 1= ON Byte 8 8 16 Stream 1 HI-HI Alarm Status, 1= ON Byte 0 8 16 Stream 2 HI-HI Alarm Status, 1= ON Byte 8 8 17 Stream 3 HI-HI Alarm Status, 1= ON Byte 0 8 17 Stream 4 HI-HI Alarm Status, 1= ON Byte 8 8 18 Stream 5 HI-HI Alarm Status, 1= ON Byte 0 8 18 Stream 6 HI-HI Alarm Status, 1= ON Byte 8 8 19 Stream 1 Active Status, 1= Sampling Byte 0 8 19 Stream 2 Active Status, 1= Sampling Byte 8 8 20 Stream 3 Active Status, 1= Sampling Byte 0 8 20 Stream 4 Active Status, 1= Sampling Byte 8 8 21 Stream 5 Active Status, 1= Sampling Byte 0 8 21 Stream 6 Active Status, 1= Sampling Byte 8 8 24 Gas Flow (cc/min) Float Inverse 0 32 39 Sensor Reference Value (mV) Float Inverse 0 32 41 Sensor Measure Value (mV) Float Inverse 0 32 43 Sensor Pressure Value (psia) Float Inverse 0 32 45 Sensor Gas Temperature (K) Float Inverse 0 32 47 Sensor UV Lamp Temp (°C) Float Inverse 0 32
62 05509F DCN6409
Page 65
Teledyne API Model 465L O3 Monitor Operation Manual
8.6.4. Instrument Setup and Configuration Registers (Read / Write)
Decimal Address Description Register Type Start Bit # Bits
1031 Stream 1 Monitoring Duration (s) Float Inverse 0 32 1033 Stream 2 Monitoring Duration (s) Float Inverse 0 32 1035 Stream 3 Monitoring Duration (s) Float Inverse 0 32 1037 Stream 4 Monitoring Duration (s) Float Inverse 0 32 1039 Stream 5 Monitoring Duration (s) Float Inverse 0 32 1041 Stream 6 Monitoring Duration (s) Float Inverse 0 32 1043 Stream 1 HI Alarm Limit (PPB) Float Inverse 0 32 1045 Stream 2 HI Alarm Limit (PPB) Float Inverse 0 32 1047 Stream 3 HI Alarm Limit (PPB) Float Inverse 0 32 1049 Stream 4 HI Alarm Limit (PPB) Float Inverse 0 32 1051 Stream 5 HI Alarm Limit (PPB) Float Inverse 0 32 1053 Stream 6 HI Alarm Limit (PPB) Float Inverse 0 32 1055 Stream 1 HI-HI Alarm Limit (PPB) Float Inverse 0 32 1057 Stream 2 HI-HI Alarm Limit (PPB) Float Inverse 0 32 1059 Stream 3 HI-HI Alarm Limit (PPB) Float Inverse 0 32 1061 Stream 4 HI-HI Alarm Limit (PPB) Float Inverse 0 32 1063 Stream 5 HI-HI Alarm Limit (PPB) Float Inverse 0 32 1065 Stream 6 HI-HI Alarm Limit (PPB) Float Inverse 0 32 1067 Stream 1 Analog Range (PPB) Float Inverse 0 32 1069 Stream 2 Analog Range (PPB) Float Inverse 0 32 1071 Stream 3 Analog Range (PPB) Float Inverse 0 32 1073 Stream 4 Analog Range (PPB) Float Inverse 0 32 1075 Stream 5 Analog Range (PPB) Float Inverse 0 32 1077 Stream 6 Analog Range (PPB) Float Inverse 0 32 1079 Stream 1 HI Alarm Enable, 1=ON Byte 0 8 1079 Stream 2 HI Alarm Enable, 1=ON Byte 8 8 1080 Stream 3 HI Alarm Enable, 1=ON Byte 0 8 1080 Stream 4 HI Alarm Enable, 1=ON Byte 8 8 1081 Stream 5 HI Alarm Enable, 1=ON Byte 0 8 1081 Stream 6 HI Alarm Enable, 1=ON Byte 8 8
05509F DCN6409 63
Page 66
Teledyne API Model 465L O3 Monitor Operation Manual
Decimal Address Description Register Type Start Bit # Bits
1082 Stream 1 HI-HI Alarm Enable, 1=ON Byte 0 8 1082 Stream 2 HI-HI Alarm Enable, 1=ON Byte 8 8 1083 Stream 3 HI-HI Alarm Enable, 1=ON Byte 0 8 1083 Stream 4 HI-HI Alarm Enable, 1=ON Byte 8 8 1084 Stream 5 HI-HI Alarm Enable, 1=ON Byte 0 8 1084 Stream 6 HI-HI Alarm Enable, 1=ON Byte 8 8
Alarm Latch Mode, 1= Latching, 0=
1085 1131 Stream Enable (bits 0-5) byte 0 16
Non-Latching Byte 0 8
64 05509F DCN6409
Page 67
Teledyne API Model 465L O3 Monitor Operation Manual
This page intentionally left blank.
05509F DCN6409 65
Page 68
9. MAINTENANCE AND ADJUSTMENTS
WARNING - RISK OF ELECTRICAL SHOCK
9.1. Maintenance Schedule
THE OPERATIONS OUTLINED IN THIS CHAPTER ARE TO BE
PERFORMED BY QUALIFIED MAINTENANCE PERSONNEL ONLY!
CAUTION
GENERAL SAFETY HAZARD
Do not look at the UV lamp while the unit is operating. UV light can cause eye damage. Always use safety glasses made from UV blocking material. (Generic plastic glasses are not adequate).
Table 9-1 below outlines the suggested maintenance procedures and intervals for ensuring the M465L continues to operate accurately and reliably. These intervals are based on continuous (24 hours a day – 7 days a week) operation. These intervals may be lengthened for intermittent operation.
Table 9-1: Maintenance Schedule
Maintenance Item Recommended Interval Section
Replace internal
6 months*1 9.3
particulate filter Adjust UV lamp As Indicated by ‘Check Lamp’
9.4
LED or status output
Replace lamp As required; when adjustment can
9.5
no longer be performed. Replace sample pump 2 years 9.6 Replace Sensor Module
2 years 9.7 Valve
*1
When external sample line pre-filters are used. If pre-filters are not
used, internal sample filter should be replaced every month.
05509F DCN6409 66
Page 69
Teledyne API Model 465L O3 Monitor Operation Manual
9.2. Instrument Layout
WARNING - RISK OF ELECTRICAL SHOCK
HIGH VOLTAGE MAY BE PRESENT WHEN POWER IS CONNECTED TO
Figure 9-1 and Figure 9-2 below shows the internal layout of the M465L. These figures will be referenced in the procedures that follow.
Note the caution areas where high voltage (line voltage) may be present when power is connected to the instrument.
THE INSTRUMENT!
Figure 9-1: Instrument Layout, Rack Mount Configuration
05509F DCN6409 67
Page 70
Teledyne API Model 465L O3 Monitor Operation Manual
Figure 9-2: Instrument Layout, NEMA Configuration
9.3. Replacing Internal Particulate Filter
1. Disconnect power from the M465L.
2. Remove the six screws and the top cover (Rack Mount Configuration) or open front
panel (NEMA Configuration.)
3. Locate the filter (see Figure 9-1 or Figure 9-2 above.) Un-strap the filter from the two
plastic hold-down clamps. Note the arrow on the filter showing flow direction.
4. Loosen the two nuts holding the filter into the two union fittings.
5. Install the nuts on the new filter. Hand-tighten until snug, do not use tools.
6. Perform steps 1 through 3 in reverse order to complete installation.
68 05509F DCN6409
Page 71
Teledyne API Model 465L O3 Monitor Operation Manual
9.4. UV Lamp Adjustment
CAUTION
GENERAL SAFETY HAZARD
Do not look at the UV lamp while the unit is operating. UV light can
1. Instrument should be running and warmed up for at least 20 minutes.
2. With instrument running, remove the six screws and the top cover.
cause eye damage. Always use safety glasses made from UV blocking material. (Generic plastic glasses are not adequate).
WARNING - RISK OF ELECTRICAL SHOCK
HIGH VOLTAGE MAY BE PRESENT WHEN POWER IS CONNECTED TO
THE INSTRUMENT!
3. Navigate the front panel menu to VIEW menu and scroll to REF display and press ENT.
At this point there should be a scrolling display similar to “REF = XXXX MV.” See Section 7.2 for details on menu navigation.
4. Locate the UV Detector adjustment pot on the forward end of the sensor module (see
Figure 9-1 or Figure 9-2 above.)
5. While observing the REF value on the display, slowly turn the pot to adjust the value.
The target adjustment range is as high as possible within the range of 800 – 1150 mV.
6. If the required adjustment cannot be achieved by adjusting the UV Detector pot alone,
then additional adjustment can be made by loosening the two UV lamp setscrews on the UV lamp housing (see Figure 9-1 or Figure 9-2) and rotating the lamp. Rotate the lamp very slowly while observing the REF value on the display. Make sure the lamp does not pull out and remains seated in the housing while it is being rotated. Re-tighten the two setscrews when a desired point has been reached.
7. If necessary, additional “fine tuning” can now be done with the UV Detector adjustment
pot per step 5.
8. Re-Install instrument cover and observe REF value on display for a couple minutes to
verify it does not drift out of the adjustment range.
05509F DCN6409 69
Page 72
Teledyne API Model 465L O3 Monitor Operation Manual
9.5. UV Lamp Replacement
CAUTION
GENERAL SAFETY HAZARD
Do not look at the UV lamp while the unit is operating. UV light can
1. Disconnect power from the M465L.
2. Remove the six screws and the top cover (Rack Mount Configuration) or open front
panel (NEMA Configuration.)
cause eye damage. Always use safety glasses made from UV blocking material. (Generic plastic glasses are not adequate).
WARNING - RISK OF ELECTRICAL SHOCK
HIGH VOLTAGE MAY BE PRESENT WHEN POWER IS CONNECTED TO
THE INSTRUMENT!
3. Loosen the two UV lamp setscrews on the UV lamp housing (see Figure 9-1 or Figure
9-2 above.)
4. Unplug the lamp power cord from the connector labeled P1 on the sensor module.
5. Loosen the four slotted captive screws attaching the sensor module to the chassis.
6. Carefully slide the lamp out of housing. The sensor module may need to be raised
slightly to completely remove the lamp.
7. Install the new lamp, seating it in the lamp housing until it stops.
8. Re-attach the sensor module captive screws to the chassis.
9. Re-tighten the two UV lamp setscrews.
10. Plug the lamp power cord into P1 on the sensor module.
11. Reconnect power to the instrument and turn on power switch. Let instrument warm up
for at least 20 minutes.
12. Perform UV lamp adjustment procedure per Section 9.4.
9.6. Sample Pump Replacement
1. Disconnect power from the M465L.
2. Remove the six screws and the top cover (Rack Mount Configuration) or open front
panel (NEMA Configuration.)
3. Locate sample pump assembly (See Figure 9-1.)
4. Cut off the clamps holding the tubing to the pump nipples and remove tubing. Note
which tubing is connected to inlet and outlet.
70 05509F DCN6409
Page 73
Teledyne API Model 465L O3 Monitor Operation Manual
5. Unplug the pump power connector from the mainboard PCA at connector J12.
6. Remove the four screws attaching the sample pump assembly to the chassis.
7. Turn over pump assembly and remove the two screws holding the sheet-metal base to the
pump.
8. Install the base on the new pump.
9. Install new pump assembly in chassis.
10. Reconnect tubing, note that outlet fitting of pump should be connected to tubing routed to
the ‘Exhaust’ fitting on rear panel.
11. Re-install new tubing clamps or cable-ties to secure tubing connections.
05509F DCN6409 71
Page 74
Teledyne API Model 465L O3 Monitor Operation Manual
9.7. Sensor Valve Replacement
1. Disconnect power from the M465L.
2. Remove the six screws and the top cover (Rack Mount Configuration) or open front
panel (NEMA Configuration.)
3. Locate sensor module assembly (See Figure 9-1.)
4. Unplug the two-pin valve connector from the sensor PCA.
5. Remove the silver retainer clip from the top of the sensor valve. A pair of pliers may be
used to slide off the retainer clip.
6. The valve coil can now be removed by sliding upwards.
7. Remove the two (2) mounting screws using a #2 Phillips screwdriver. See Figure 9-3.
Note that there are four screw heads visible on the top of the valve body, only two (2) of these should be removed.
8. Remove valve body from sensor manifold.
9. Clean any residue or dirt off the surface of the manifold using a lint-free cloth and
distilled or DI water.
10. Install the new valve by reversing steps 1-7. Note the proper orientation of the new valve
as shown in Figure 9-3.
Figure 9-3: Sensor Detail
72 05509F DCN6409
Page 75
Teledyne API Model 465L O3 Monitor Operation Manual
9.8. Water Drop-Out Filter (Option)
Maintenance
Located externally on the right side of the enclosure, the Water Drop-Out Filter (Section 4.1.1.2) should be checked periodically for water build-up and drained. To drain this filter:
1. Disconnect the ozone supply line from the instrument.
2. Shut off flow to the instrument using the needle valve on the front panel.
3. Remove cap from the fitting located on the bottom of the filter and allow it to drain.
4. Replace cap and adjust flow to instrument.
Replacement of the particulate filter, a Teflon coalescing membrane should be rarely required. However, if it does become fouled, possibly indicated by a significant reduction in the gas pressure measured in the O3 cell, which would trigger a warning in the instrument, replace it as follows:
1. Remove the orange cap from the top of the water drop-out filter.
2. Located the membrane on the underside of the cap assembly and replace it.
3. Tightly secure the orange cap back onto the water drop-out filter.
9.9. Configuring the Analog Output
The analog output can be configured for either 0-5V DC or 4-20mA operation. To change or verify the configuration of the analog output:
1. Disconnect power from the M465L.
2. Remove the six screws and the top cover (Rack Mount Configuration) or open front
panel (NEMA Configuration.)
3. Locate the Mainboard PCA (see Figure 9-1 or Figure 9-2.)
4. Set the desired operation as shown in Figure 9-4.
5. Re-Install the top cover or re-secure the front panel (NEMA Configuration.)
05509F DCN6409 73
Page 76
Teledyne API Model 465L O3 Monitor Operation Manual
CPU ALIVE
D14
X1
C10
74HC373
DISPLAY
U7
J4
C20
74HC373
C27
R24
R23
+
U11
R26
C29
+
PRESS
C33
U17
R30
C38
U20
J7
C39
VR2
R35
LM7805
C46
C48
VR4
C54
C56
C58
U27
C60
VR6
C64
C65
C66
R58
C67
C68
VR7
C70
PS2
SW1
R1
J1
R64
C1 C2
R67
C4
PIC18F8720
C5
U1
R66
C7
C14
74HC373
U6
C21
SRAM128KX8-SOP
TP8
U13
C41
+
C31
U16
C34
C35
C36
U19
C40
HCF4052B
U22
R36
R40R39
C47
TP12
U24
J8
C53
REF_DET
C57
TC1044
U26
C59
C61
R53
C63
MEAS_DET
TP21
R55
J17
J16
R56
R57 J19
K6
RS-485/ Analog Out
J3
Expansion
DAC
R8 R9
R7
FLOW
TP4
R11
R15
C11 C12
U8
ISO124
ISO_GND
C19
TP6
C23
C22
TP7
C30
ISO_V
U14
C28
5.0VREF
+
D4
U15
M41T0
U18
J24
VALVE
Y1
+
Analog In
C37
R32
Expansion
D5
C50
VR3
R33
VR5
C43
U23
R37
+
TP11
R38
C49
4.096VREF
PR3
R44
R42
C51
R43
TP14
SDA
R47
TP13
R48
REF
PR4
C55
TP19
R51
TP18
R52
+12VL
MEAS
J14
J12
HEATER
R54
C62
D15
TP22
Q2
J20
PUMP
MODULE
+9V
+15V
+12V
LAMP INVERTER
PS1
JP5
R59
Q3
3.3V
DGND
VCC
J2
TP1
TP3
TP2
+
R2
R4
R3
C3
VR1
R6
R5
C6
C15
TP5
R19
C26
C32
R31
TP15
TP16
+15V
U4
R10
U2
R13 R14
C9
C8
LT1181
D1
R12
C13
R17
U5
C16
C17
R18
R16
C18
JP1
PCF8575
U10
U9
R20
XTR110
C24
.
.
R27
.
U21
D8
R46
SCL
-9V
J10
J15
AUX
7407
C25
SPAN
U12
JP2
R25
JP3
Q1
PR1
ZERO
PR2
JP4
I2
DCP010515
TP9
D6
C42
C44
C45
R34
I1 TP10
D9
R41
R45
R50
U25
R49
C52
D11
+15V
J9
TP17
TP20
J13
+12V
R65
D13
VALVE +15V
J18
R61
R62
J21
AC
G1
C71
PCA 05072 REV:
SW3
SWITCH DIP 10
ETHERNET
Close Indicated Position s Only
D2
R21
R28
D7
RELAY 1
D10
RELAY 2
D12
RELAY 3
K4
U28
7407
C69
R60
U29
R63
G2
J23
C72
K5
K7
K8
K9
K10
DANGER!
HIGH VOLTAGE
RS232RS485
D3
R22
J5
K1
K2
K3
EARTH LINE NEUT
TELEDYNE API
(C)2004
U3
XPORT XE
SW2
JP3
Q1
R27
J6
R29
SW4
J11 N.C. COM N.O. N.C. COM N.O. N.C. COM N.O.
J22
16 15 14 13 12 11 10
9 8 7 6 5 4 3 2 1
JP4
0-5V OUTPUT
JP3
Q1
R27
JP4
4-20mA OUTPUT
Figure 9-4: Mainboard – Analog Output Configuration
9.10. Cleaning Exterior Surfaces of the M465L
If necessary, the front panel mask and keyboard of the M465L can be cleaned with a damp cloth. Do not attempt to clean any of the other surfaces of the instrument. Do not submerge any part of the instrument in water or cleaning solution.
9.11. Boot Loader Operation
The main function of a boot loader is to load or update the firmware to the microcontroller on the Printed Circuit Board (PCB) of an instrument. There are two types of boot loader setup: hardware and software, described in the following sections.
NOTE
When an external MODBUS host device is attempting to communicate with the instrument over the serial or Ethernet port while the instrument is starting up, the instrument may halt. In order to prevent the problem from occurring, ensure that there are no devices attempting communication with the instrument during start-up. Once the sign-on message is displayed on the front panel, it is safe to begin communications.
74 05509F DCN6409
Page 77
Teledyne API Model 465L O3 Monitor Operation Manual
9.11.1. Boot Loader – Hardware Setup
Verify that the instrument is configured for an RS-232 connection. See Section 8.2 for details on changing the COM port configuration. The instrument can be connected to the PC COM port using a 9-pin serial cable (TAPI P/N WR77).
9.11.2. Boot Loader – Software Setup
Table 9-2 lists the required firmware loader software and firmware options.
Table 9-2: Required Software
SOFTWARE TYPE FILE TYPE/VERSION
PC Software FirmwareLoader.EXE
Instrument Firmware
The firmware loader is available on Teledyne API’s website at http://www.teledyne-
api.com/software/.
1. Install the firmware loader software onto a PC.
2. Verify the cable connections. Do not power up the instrument.
3. Run the firmware loader application on the PC.
4. Select the COM port that the instrument is connected to.
5. Select the HEX file for the appropriate firmware.
6. Refer to Section 9.11.2.1 to upload the firmware or to Section 9.11.2.2 to verify the
firmware.
05498-00_XX.HEX (for ppb, ppm, ugm)
05498-01_XX.HEX (for ppb, ppm, pphm)
9.11.2.1. Upload Firmware
1. Press “Write”. The screen will display, “Wait for boot loader”.
2. When the screen displays, “Power cycle instrument now”, power up the instrument.
NOTE
Search for boot loader will time out in 20 seconds. If the instrument is not powered up in 20 seconds, the screen will display, “No boot loader detected.” Go back to step 1 to rewrite the firmware.
05509F DCN6409 75
Page 78
Teledyne API Model 465L O3 Monitor Operation Manual
The firmware loader software should display a message “Boot loader found, uploading…”. If this message is not displayed, power down the instrument and receck the COM port configuration and cable (see Section 9.11.1) and retry.
3. When the upload is complete, a message, “Write-flash completed successfully” should be
displayed. If it is an unsuccessful upload, a message, “Write-flash error” is displayed. If a Write-flash error occurs, ensure that the correct PC COM port is selected for the instrument (see Section 9.11.2, step 4).
NOTE
It is not recommended to stop the writing process half way through. This will cause the firmware to not be fully uploaded.
9.11.2.2. Verify Firmware
The verify function compares the instrument’s current firmware to the selected HEX file (see Section 9.11.2, step 5) and confirms whether or not the files are a match.
1. Press “Verify”. The screen will display, “Wait for boot loader”.
2. When the screen displays, “Power cycle instrument now”, power up the instrument.
NOTE
Search for boot loader will time out in 20 seconds. If the instrument is not powered up in 20 seconds, the screen will display, “No boot loader detected.” Go back to step 1 to verify the firmware.
The firmware loader software should display a message “Boot loader found, verifying…”. If this message is not displayed, power down the instrument and receck the COM port configuration and cable (see Section 9.11.1) and retry.
3. When the upload is complete, a message, “Verify-flash completed successfully” should
be displayed. If it is an unsuccessful upload, a message “Verify-flash error” is displayed. If a Verify-flash error occurs, ensure that the correct PC COM port is selected for the instrument (see Section 9.11.2, step 4).
NOTE
A verification process can be stopped at any point in time. This will not affect the firmware that is already uploaded on the instrument. The instrument will automatically go to the firmware in 10 seconds.
76 05509F DCN6409
Page 79
Teledyne API Model 465L O3 Monitor Operation Manual
This page intentionally left blank.
05509F DCN6409 77
Page 80
10. TROUBLESHOOTING
WARNING
RISK OF ELECTRICAL SHOCK. THE OPERATIONS OUTLINED IN THIS CHAPTER ARE TO BE PERFORMED BY QUALIFIED MAINTENANCE PERSONNEL ONLY!
10.1. Reference Drawings
The drawings contained in this section are for general reference and may be useful when performing certain troubleshooting activities.
10.1.1. Pneumatic Diagram
Figure 10-1 below is a pneumatic diagram that can be referenced when performing troubleshooting on the monitor. Note that certain items, such as the Stream Selector and Electronic Flowmeter are optional and will not be present in all M465L monitors.
Figure 10-1: Pneumatic Diagram, 3 Stream Configuration
05509F DCN6409 78
Page 81
Teledyne API Model 465L O3 Monitor Operation Manual
10.1.2. Interconnect Diagram
Figure 10-2 below details the electrical connections between the various electronic modules in the M465L.
04968-XXXX SENSOR ASSY
02571-0100 DETECTOR
PCA
VA0000054
VALVE
03742 LAMP ASSY P1
04003-1100 FLOWMETER PCA (OPTIONAL)
J1
04849-0100 SENSOR PCA
J17
J8
RACK-MOUNT
VERSION ONLY
05072-0100 MAINBOARD PCA
J19
J24 “ANALOG IN
EXPANSION”
J1
J6
RS485
+15V
J10 MODULE RS485
J15 “AUX”
AC/DC Power
CN0000073 IEC TYPE POWER ENTRY MODULE
SW0000051 POWER SWITCH/CIRCUIT BREAKER
Supply
J23 AC IN
05313-0100 DISPLAY PCA
04702 PUMP ASSY
J4 +15V
J2
12V
I2C
J14 “MODULE +15”
J4
J3 “FLOW”
N/C
J2
J12 “PUMP +12V”
J5
i2C
05264-0000 EXPANSION RELAY PCA
J5
Figure 10-2: Interconnect Diagram
05509F DCN6409 79
Page 82
Teledyne API Model 465L O3 Monitor Operation Manual
10.2. Troubleshooting Using Front Panel Status
LED’s or Status Outputs
The M465L has 4 front-panel status LED’s that reflect the current operating status of the monitor, and indicate fault conditions. There are also four relay Status Outputs on the rear panel (or internally in the NEMA configuration) that also reflect the state of these status LED’s. See Section 4.2.3 for more information on the Status Outputs.
The Status LED’s are categorized into Critical and Non-Critical warnings. Critical warnings are those that would normally require immediately removing the monitor from service and repairing it. Non-critical warnings are those that indicate some maintenance would be useful, but not immediately required.
Table 10-1: Status LED/Output Definitions
Status Output #
1 Sensor OK On
2 Invalid Reading Off
3 Check Lamp Off 4 Pneumatic
Status LED Label
Error
Normal Operatin g State
Off
Triggers Critical
Warning?
No bench communications Reference < 125.0 mV Reference > 1230.0 mV O3 Concentration < -10.0 PPB  O3 Concentration > F.S. Range  Reference < 250.0 mV Pressure < 9.0 psia Pressure > 14.9 psia Flow < 640 cc/min Flow > 960 cc/min
Yes
No
No No
80 05509F DCN6409
Page 83
Teledyne API Model 465L O3 Monitor Operation Manual
10.2.1. Sensor OK
The Sensor OK LED indicates the status of the O3 sensor module in the monitor. The normal state of this LED (or Status Output) is On. If this LED remains off after the normal 20 minute warm-up period, then a failure has occurred and the monitor should be removed from service and repaired.
The most common cause of this warning is a failure of the UV Lamp. A UV Lamp Adjustment (See Section 9.4) should be attempted. If the UV Lamp cannot be adjusted to meet the specified values, then a UV Lamp Replacement should be performed (See Section
9.5.) This warning can also be caused by a communications error with the sensor module. Inspect
the two cables between the Mainboard PCA and Sensor Module for loose or intermittent connections (See Section 9.2 for monitor layout.) If no cable problem can be found, then the Sensor Module should be replaced.
10.2.2. Invalid Reading
The Invalid Reading LED indicates that the instrument is reading a value that cannot be represented properly on the analog output. Since the analog output is limited to 0-5V (or 4­20mA,) it cannot properly represent negative values, or values in excess of the full-scale range.
If the monitor is consistently reading negative values, then a zero calibration should be performed, see Section 7.2.3.4.
If the monitor is consistently reading values in excess of the full-scale range, then the range value should be adjusted higher. See Section 7.2.7 for details on setting the range value.
10.2.3. Check Lamp
The Check Lamp LED indicates that the UV Lamp intensity has dropped below 250mV, a level where UV Lamp Adjustment (See Section 9.4) should be made at the next convenient opportunity. Note that this is a non-critical warning and immediate service is not required. However if the UV Lamp intensity drops below 125mV, then the Sensor OK LED will also turn off, indicating that the monitor must be immediately serviced or taken off-line.
10.2.4. Pneumatic Error
The Pneumatic Error LED indicates that one of the pneumatic parameters, flow or pressure, has gone outside of normal ranges. Note that this is a non-critical warning and immediate service is not required. However if the Sensor OK LED turns off then the monitor must be immediately serviced or taken off line.
The first step in troubleshooting a Pneumatic Error is determining which parameter has caused the warning. At the monitor front panel, navigate to the VIEW menu (See Section
05509F DCN6409 81
Page 84
Teledyne API Model 465L O3 Monitor Operation Manual
7.2.4.) Examine the Flow and Pressure values and compare them to the limits described in Table 10-1 and take appropriate action as described below.
10.2.4.1. Pressure Too High
The monitor inlet is being pressurized. The monitor inlet should be allowed to sample gas at ambient pressure. Disconnect all sample lines from the monitor and read the pressure again. If the value then drops to within the acceptable range, then one of the sample lines is at elevated pressure.
If the inlet is being pressurized, then the design of the sample system should be reviewed and corrected. Contact Teledyne API customer service for assistance.
10.2.4.2. Pressure Too Low
Something may be restricting the flow through the monitor, causing the sample pump to pull a vacuum on the sensor module. First, unplug the sample pump power plug on the Mainboard PCA and observe the pressure reading. If the pressure reading is still too low, then likely there is a problem with the pressure sensor on the Sensor Module and the Sensor Module should be replaced. Next reconnect the pump and disconnect all sample lines from the monitor and see if the pressure increases to normal levels. If it does, then the problem is somewhere in the sampling system external to the monitor. If the pressure is still below the limit with the sample lines removed, then there is a restriction in the pneumatics of the monitor upstream of the sensor assembly. The most likely cause is a plugged particulate filter. Disconnect the outlet fitting on the sample filter and observe the pressure. If it returns to normal levels then the restriction is in the particulate filter or inlet manifold assembly (multi-stream configurations only.) Replace particulate filter per Section 9.3.
82 05509F DCN6409
Page 85
Teledyne API Model 465L O3 Monitor Operation Manual
This page intentionally left blank.
05509F DCN6409 83
Page 86
11. CALIBRATION
11.1. Factory Calibration
The Model 465 is calibrated to NIST traceable standards on the 0-1 ppm range prior to shipment. A calibration certificate for your instrument can be purchased from Teledyne API if required. Teledyne API also recommends that the Model 465 be re-calibrated once a year. Teledyne API can provide NIST traceable calibration services at our factory or on-site. Please contact our Customer Service department for details on these services.
11.2. Zero Calibration
A zero calibration can be easily performed in the field, and Teledyne API recommends that this be done at least annually. See Section 7.2.3.4 for details on performing a zero calibration.
05509F DCN6409 84
Page 87
Teledyne API Model 465L O3 Monitor Operation Manual
This page intentionally left blank.
05509F DCN6409 85
Page 88
12. SPARE PARTS
This section provides the names and part numbers of the spare parts for the M465L rack mount units and for the NEMA units.
12.1. M465L Rack Mount Spare Parts List
PARTNUMBER DESCRIPTION
025710100 PCA, UV DETECTOR PREAMP, M454/M460 036750000 AKIT, ELEMENTS, 47MM, 30 MICRON (10 PCS) 037420000 ASSY, UV LAMP (BIR) M452/454/460X/465L * 040031200 PCA, FLOW SENSOR, M465 046170000 ABSORPTION TUBE, M460L 046740000 ASSY, PUMP, 12VDC, M460M/M700E 048490100 PCA, O3 BENCH, M460L/M465L 050720300 PCA, MAINBOARD, M465L, CE 052640000 PCA, RELAY/VALVE EXPANSION, M460/M465L 053010000 AKIT, EXP KIT, CARULITE 200 053130300 PCA, FRONT, MULTI CHL, M465L, CE 053130400 PCA, FRONT PANEL, SNGL CHL, M465, CE 053200000 PCA, FRONT PANEL DAUGHTER BOARD, M465L 055090000 MANUAL, OPERATORS, M465L 059350000 ASSY, FLOWMETER W/O VALVE, M465L 061760100 ASSY, FLOWMETER, M465L/M 062310000 AKIT, OPTION, PRE-FILTERS, 1 STREAM, M46 062680000 ASSY, ZERO SCRUBBER, OZONE DR0000006 DRYER, 24", 1/4" SS FITTINGS FL0000033 DFU, KYNAR, 1/4" TUBE FM0000004 FLOWMETER (KB) HW0000120 SHOCKMOUNT, GROMMET ISOLATOR KIT000267 AKIT, M46XL PRE-FILTERS (FL37-25 pcs) OP0000031 WINDOW, QUARTZ, 1/2"DIA, .063" THICK (KB OR0000039 ORING, 2-012V OR0000050 ORING, 2-014V
05509F DCN6409 86
Page 89
Teledyne API Model 465L O3 Monitor Operation Manual
PARTNUMBER DESCRIPTION
OR0000098 ORING, 2-108S OR0000099 ORING, 2-129, CV-75 COMPOUND OR0000104 ORING, 5-125V PS0000042 PS, 60W SWITCHING, 2X4, 15V SW0000051 SWITCH, POWER CIRC BREAK VDE/CE, w/RG(KB VA0000054 MANIFOLD, 3-WAY VALVE, PEEK/VITON WR0000008 POWER CORD, 10A
05919G – M465L, SPL, (DCN 5123) 10/01/08
12.2. M465L NEMA Spare Parts List
PART NUMBER DESCRIPTION
025710100 PCA, UV DETECTOR PREAMP 036280000 ASSY, WATER DROP OUT FILTER, PFA 036750000 AKIT, ELEMENTS, (FL26) 47MM, 30 MICRON 040031200 PCA, FLOW SENSOR, 046170000 ABSORPTION TUBE 046740000 ASSY, PUMP, 12VDC (OP63) 048490100 PCA, O3 BENCH 048990300 ASSY, HI-CON O3 SCRUBBER, CH71 049680000 ASSY, SENSOR, HI-CONC 049680100 ASSY, SENSOR, LO-CONC 050300000 ASSY, REF SCRUBBER, HI-CONC 050720300 PCA, MAINBOARD, CE 052640000 PCA, RELAY/VALVE EXPANSION 053010000 AKIT, EXP KIT, CARULITE 200 053130300 PCA, FRONT, MULTI CHL, CE 053130400 PCA, FRONT PANEL, SNGL CHL, CE 053200000 PCA, FRONT PANEL DAUGHTER BD, M465XXX 055090000 MANUAL, OPERATORS, M465L 058400001 ASSY, REF SCRUBBER, LO-CONC,*(OBS) 061760100 ASSY, FLOWMETER 062310000 AKIT, PRE-FILTERS, 1 STREAM 062680000 ASSY, ZERO SCRUBBER, OZONE 065660000 ASSY UV LAMP UVP OP39 GEN III
05509F DCN6409 87
Page 90
Teledyne API Model 465L O3 Monitor Operation Manual
PART NUMBER DESCRIPTION
CN0000350 CONNECTOR, 16 PIN, W/SCREW FLANGE DR0000006 DRYER, 24", 1/4" SS FITTINGS FL0000033 DFU, KYNAR, 1/4" TUBE HW0000120 SHOCKMOUNT, GROMMET ISOLATOR KIT000267 AKIT, M46XL PRE-FILTERS (FL37-25 pcs) OP0000031 WINDOW, QUARTZ, 1/2"DIA, .063" THICK (KB OR0000039 ORING, 2-012V OR0000050 ORING, 2-014V OR0000098 ORING, 2-108S OR0000099 ORING, 2-129, CV-75 COMPOUND OR0000104 ORING, 5-125V PS0000042 PS, 60W SWITCHING, 2X4, 15V VA0000054 MANIFOLD, 3-WAY VALVE, PEEK/VITON
05910J – M465L NEMA SPL, (DCN 5796) 08/06/2010
88 05509F DCN6409
Page 91
Teledyne API Model 465L O3 Monitor Operation Manual
This page intentionally left blank.
05509F DCN6409 89
Page 92
INDEX
Accuracy
sample filtratiom, 25 sampling lines, 25 Specifications, 15
Alarm Relays
Specifications, 16
Analog Output
Specifications, 16
asurement Units
Specifications, 15
Cleanliness in pneumatic system
importance of, 25
Connection
Ethernet, 33 RS232, 33
RS485, 33 Diagnostic LEDs, 29 Dimensions
Specifications, 15 Environmental Conditions
Specifications, 15 Humid Environments, 25 Humidity Range
Specifications, 15 Isolated Analog Output
Specifications, 16 Lower Detectable Limit
Specifications, 15 Maximum Operating Altitude
Specifications, 15 Multi-Stream Specifications:, 16 Noise
Specifications, 15 Option
Flowmeter, 37
Permeation Gas Dryer, 27, 37
Rack Mount Slides, 37
Sample Conditioner, 27, 37
Sample Inlet Filters, 37
Stream Selector, 37
Permeation Gas Dryer, 27 Power
Specifications, 15
Ranges
Specifications, 15
Response Time
Specifications, 15
rument Status Outputs
Specifications, 16
Safety Messages
General, iii, 65, 68, 69
Sample Conditioner, 27 Sample Filtration Accuracy, 25 Sample Flow Rate
Specifications, 15
Sample gas connection
multi-stream, 24 single stream, 24
Spare Parts
for NEMA units, 86 for rack mount units, 85
Startup
initial behavior, 22
Status Relay
Specifications, 16
Teledyne Contact Information
Email Address, 17 Fax, 17 Phone, 17
Temperature Range
Specifications, 15 UV Lamp, 65, 68, 69 uv Light, 65, 68, 69 UV light absorption, 19 Weight
Specifications, 15
05509F DCN6409 90
Loading...