Your safety and the safety of others is very important. We have provided many
important safety messages in this manual. Please read these messages carefully.
A safety message alerts you to potential hazards that could hurt you or others. Each
safety message is associated with a safety alert symbol. These symbols are found in the
manual and inside the instrument. The definition of these symbols is described below:
General Warning/Caution
danger.
Caution:
Caution:
Technician Symbol
performed by qualified maintenance personnel only.
The monitor should only be used for the purpose and in the manner
If you use the monitor in a manner other than that for which it was
intended, unpredictable behavior could ensue with possibly hazardous
Hot Surface
Electrical Shock Hazard
: Refer to the instructions for details on the specific
: All operations marked with this symbol are to be
CAUTION
described in this manual.
consequences.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page ii
INSTALLATION (19”RACK MOUNT ENCLOSURE).....................................................................................12
INSTALLATION (NEMA4X,WALL MOUNT ENCLOSURE)........................................................................14
ELECTRICAL AND PNEUMATIC CONNECTIONS...........................................................................................20
1.7.1 Analog Output .................................................................................................................................20
1.7.3 Digital Outputs................................................................................................................................20
FRONT PANEL DISPLAY ...........................................................................................................................32
2.2
2.2.1 Front panel display fields................................................................................................................32
2.2.2 Status LED’s....................................................................................................................................34
2.3.1 Main Menu ......................................................................................................................................35
ADJUSTING UV LAMP AND DETECTOR ......................................................................................................47
5.2
CHANGING THE ANALOG OUTPUT RANGE..................................................................................................49
5.3
5.3.1 Voltage output..................................................................................................................................49
5.3.2 Current Output (Optional)...............................................................................................................50
TABLE 4.1 UV SOURCE LAMP AND DETECTOR DIAGNOSTICS.................................................48
TABLE 4.2 V/F BOARD DIP SWITCH - RANGES FOR ANALOG OUTPUT...................................50
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page vi
1.0 INTRODUCTION
1.1 Preface
Teledyne API is pleased that you have purchased the Model 450. We offer a full one
year warranty (see Section 1.2) and we at Teledyne API will be pleased to provide you
with any support required so that you may utilize our equipment to the fullest extent.
The Teledyne API Model 450 keyboard/operator interface with its "talking keys" makes
the Teledyne API a very user-friendly system. We hope you will not experience any
problems with the Teledyne API Model 450 but if you do, the built-in tests and
diagnostics should allow you to quickly and easily find the problem. In addition, our full
time customer service department is always available to answer your questions.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 7
Prior to shipment, Teledyne API equipment is thoroughly inspected and tested. Should equipment
failure occur,
COVERAGE
After the warranty period and throughout the equipment lifetime, Teledyne API stands ready to
provide on-site or in-plant service at reasonable rates similar to those of other manufacturers in the
industry. All maintenance and the first level of field troubleshooting are to be performed by the
customer.
NON-TELEDYNE API MANUFACTURED EQUIPMENT
Equipment provided but not manufactured by Teledyne API is warranted and will be repaired to the
extent and according to the current terms and conditions of the respective equipment manufacturers
warranty.
GENERAL
Teledyne API warrants each Product manufactured by Teledyne API to be free from defects in
material and workmanship under normal use and service for a period of one year from the date of
delivery. All replacement parts and repairs are warranted for 90 days after the purchase.
If a Product fails to conform to its specifications within the warranty period,
correct such defect by, in
refunding the purchase price of such Product.
The warranties set forth in this section shall be of no force or effect with respect to any Product:
(i) that has been altered or subjected to misuse, negligence or accident, or (ii) that has been used in
any manner other than in accordance with the instruction provided by
properly maintained.
THE WARRANTIES SET FORTH IN THIS SECTION AND THE REMEDIES THEREFORE
ARE EXCLUSIVE AND IN LIEU OF ANY IMPLIED WARRANTIES OF
MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE OR OTHER WARRANTY OF
QUALITY, WHETHER EXPRESSED OR IMPLIED. THE REMEDIES SET FORTH IN THIS
SECTION ARE THE EXCLUSIVE REMEDIES FOR BREACH OF ANY WARRANTY
CONTAINED HEREIN. TELEDYNE API SHALL NOT BE LIABLE FOR ANY INCIDENTAL
OR CONSEQUENTIAL DAMAGES ARISING OUT OF OR RELATED TO THIS AGREEMENT
OF TELEDYNE API'S PERFORMANCE HEREUNDER, WHETHER FOR BREACH OF
WARRANTY OR OTHERWISE.
Teledyne API assures its customers that prompt service and support will be available.
Teledyne API shall
Teledyne API's discretion, repairing or replacing such defective Product or
Teledyne API or (iii) not
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 8
TERMS AND CONDITIONS
C
T
P
All units or components returned to Teledyne API should be properly packed for handling
and returned freight prepaid to the nearest designated Service Center. After the repair, the
equipment will be return e d , f r eight pr e p a i d .
1.3 Principle of Operation
The detection of ozone molecules is based on absorption of 254 nm UV light due to an
internal electronic resonance of the O3 molecule. The Model 450 uses a mercury lamp
constructed so that a large majority of the light emitted is at the 254nm wavelength. Light
from the lamp shines down a hollow quartz tube that is alternately filled with sample gas,
then filled with gas scrubbed to remove ozone. The ratio of the intensity of light passing
through the scrubbed gas to that of the sample forms a ratio I/Io. This ratio forms the
basis for the calculation of the ozone concentration.
The Beer-Lambert equation, shown below, calculates the concentration of ozone from the
ratio of light intensities.
C
O
3
Where:
I = Intensity of light passed through the sample
I
= Intensity of light through sample free of ozone
o
= absorption coefficient
l = path length
= concentration of ozone in ppb
O
3
= sample temperature in degrees Kelvin
= pressure in inches of mercury
As can be seen the concentration of ozone depends on more than the intensity ratio.
Temperature and pressure influence the density of the sample. The density changes the
number of ozone molecules in the absorption tube which impacts the amount of light
removed from the light beam. These effects are addressed by directly measuring
temperature and pressure and including their actual values in the calculation. The
absorption coefficient is a number that reflects the inherent ability of ozone to absorb 254
nm light. Most current measurements place this value at 308 cm-1 atm-1 at STP. The
value of this number reflects the fact that ozone is a very efficient absorber of UV
radiation which is why stratospheric ozone protects the life forms lower in the
atmosphere from the harmful effects from solar UV radiation. Lastly, the absorption path
length determines how many molecules are present in the column of gas in the absorption
tube.
9
10
l
o
273
2992
.
inHg
n
l
o
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 9
The intensity of light is converted into a voltage by the detector/preamp module. The
voltage is converted into a number by a voltage-to-frequency (V/F) converter capable of
80,000 count resolution. The digitized signal, along with the other variables, are used by
the CPU to compute the concentration using the above formula.
Every 6 seconds the M450 completes a measurement cycle consisting of a 2 second wait
period for the sample tube to flush, followed by a 1 second measurement of the UV light
intensity to obtain I. The sample valve is switched to admit scrubbed sample gas for 2
seconds, followed by a 1 second measurement of the UV light intensity to obtain Io.
Measurement of the Io every 6 seconds minimizes instrument drift due to changing
intensity of the lamp due to aging and dirt.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 10
1.4 Specifications
Ranges User selectable to any full scale range from
Measurement Units ppbv, ppmv, ppmw, pphm, µg/m3, mg/ m3
Zero Noise < .0015 ppm (rms)
Span Noise < .5% of reading (rms) (above 0.1 ppm)
Lower Detectable Limit < .003 ppm (rms)
Linearity Better than 1% full scale
Response Time (95%) <30 sec
Minimum Stream Duration
(Cycle Time)*
Sample Flow Rate 1-2.5 L/min
Temperature Range 5-45o C
Humidity Range 10-90% RH, Non-Condensing
Dimensions (H x W x D) 7" x 17" x 18.2"
Weight 28 lb. (12.7 kg) - Single Channel Monitor, 19”
Power 110V/60 Hz, 220V/50 Hz, 240V/50Hz 250 watts
Environmental Conditions Installation Category (Overvoltage Category) II
Maximum Operating Altitude 2000 meters
Analog Output Voltage Mode 100mV, 1V, 5V, 10V, (User Selectable)
Isolated Analog Output 0-20mA
Mode**
*Multi-Channel Monitor only
**Optional
1ppm to 1000 ppm
1.0 min
(178 mm x 432 mm x 463 mm)
Rack Mount
30 lb. (13.6 kg) - Multi-Channel Monitor, 19”
Rack Mount
38 lb (17.3 kg) – NEMA 4X Enclosure
230V~, 50Hz, 2.5A
Pollution Degree 2
Maximum voltage between outputs and ground
60V peak
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 11
1.5 Installation (19” Rack Mount Enclosure)
The Model 450 is shipped with the following standard equipment:
1. Power cord.
2. Instruction manual.
CAUTION
Upon receiving the Model 450 please do the following:
1. Verify no apparent shipping damage. (If damage has occurred please advise
2. Remove all red colored shipping screws from the inside and bottom of the
3. When installing the Model 450, allow a minimum of 4 inches of clearance at
4. Connect sample inlet line(s) to the sample port on rear panel.
To avoid personal injury, always use two persons to lift and
carry the Model 450.
shipper first, then Teledyne API.)
instrument. There are four screws on the optical bench mounts that are
accessed from the inside of the instrument and two on the pump bracket that
are accessed from the bottom of the instrument. Note: Save these shipping
screws and re-install them whenever the unit is shipped to another
location.
the back of the instrument and 1 inch of clearance on each side for proper
ventilation.
Connect the exhaust fitting on the rear panel (See Fig. 1.1) to a
suitable vent outside the monitor area.
CAUTION
NOTE
See Figure 1.1 for rear panel pneumatic connections. Sample
lines made from an inert material such as Teflon should be used
to minimize sample degradation.
5. Connect the power cord to an appropriate power outlet (see the serial number
tag for correct voltage and frequency).
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 12
CAUTION
Verify that the instrument is set up for proper line voltage
and frequency.
6. Turn on the M450 by switching the switch on the lower right corner of the
front panel. The front panel display should light with a sequence of messages,
-API - M450 - software version number, then a normal display as shown in
Figure 2.2.
7. Allow about 20 minutes for the UV lamp temperature to come up to it’s
setpoint then press the TST> button on the front keyboard to scroll through
the TEST values. Compare these values to those noted during the final
factory checkout listed in Table 1.2. The values observed should closely
match the Table 1.2 values.
NOTE
Words in all caps are messages on the monitor front
panel.
FIGURE 1.1 - REAR PANEL PNEUMATIC CONNECTIONS
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 13
1.6 Installation (NEMA 4X, Wall Mount Enclosure)
The Model 450 is shipped with the following standard equipment:
1. Instruction manual.
CAUTION
To avoid personal injury, always use two persons to lift and
carry the Model 450.
Upon receiving the Model 450 please do the following:
1. Verify no apparent shipping damage. (If damage has occurred please advise
shipper first, then Teledyne API.)
2. Mount the NEMA 4X enclosure to a vertical surface using four 5/16” bolts.
Mounting hole dimensions are shown in Figure 1.2.
3. Check the serial number tag on the front panel to ensure that the monitor has
been configured for the proper AC line voltage and frequency.
4. Route the line power cable through one of the conduit bulkheads on the
bottom surface of the enclosure.
5. Connect the line power to the power entry barrier strip shown in Figure 1.3.
CAUTION
Use a 3-wire grounded power cord
and connect as shown in Figure 1.3
6. Route signal wires into the enclosure through the conduit bulkheads. These
wires should be routed up to the front panel through the notch in the lower
left-hand corner of the front panel as shown in Figure 1.4. Make sure to allow
enough extra length in these wires so that the panel can still swing open
without stressing the wires.
7. Connect ¼” sample inlet line(s) to the sample inlet fitting(s) shown in Figure
1.5.
8. Connect a ¼” exhaust line to the fitting shown in Figure 1.5.
CAUTION
Connect the exhaust fitting to a suitable vent outside the monitor
area.
9. The Model 450 can now be turned on using the power switch located on the
front panel.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 14
10. Allow about 20 minutes for the UV lamp temperature to come up to it’s
setpoint then press the TST> button on the front keyboard to scroll through
the TEST values. Compare these values to those noted during the final
factory checkout listed in Table 1.2. The values observed should closely
match the Table 1.2 values.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 15
FIGURE 1.2 – NEMA 4X ENCLOSURE DIMENSIONS
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 16
FIGURE 1.3 – NEMA 4X AC LINE POWER CONNECTION
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 17
Loading...
+ 38 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.