Teledyne 450 User Manual

INSTRUCTION MANUAL
MODEL 450
OZONE MONITOR
© Teledyne Instruments
Advanced Pollution Instrumentation Division
(T-API)
San Diego, CA 92121-5201
TOLL-FREE: 800-324-5190
TEL: 858-657-9800
FAX: 858-657-9816
EMAIL:
WEB SITE: www.teledyne-api.com
Rev D1
api-sales@teledyne.com
02395
11/22/05

SAFETY MESSAGES

Your safety and the safety of others is very important. We have provided many important safety messages in this manual. Please read these messages carefully.
A safety message alerts you to potential hazards that could hurt you or others. Each safety message is associated with a safety alert symbol. These symbols are found in the manual and inside the instrument. The definition of these symbols is described below:
General Warning/Caution danger.
Caution:
Caution:
Technician Symbol performed by qualified maintenance personnel only.
The monitor should only be used for the purpose and in the manner
If you use the monitor in a manner other than that for which it was
intended, unpredictable behavior could ensue with possibly hazardous
Hot Surface
Electrical Shock Hazard
: Refer to the instructions for details on the specific
: All operations marked with this symbol are to be
CAUTION
described in this manual.
consequences.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page ii

TABLE OF CONTENTS

SAFETY MESSAGES....................................................................................................................................II
TABLE
FIGURES........................................................................................................................................................
TABLES........................................................................................................................................................
1.0 INTRODUCTION...................................................................................................................................7
1.1
1.2
1.3
1.4
1.5
1.6
1.7
1.7
1.8
2.0 OPERATION........................................................................................................................................31
OF CONTENTS ...............................................................................................................................III
PREFACE..................................................................................................................................................... 7
WARRANTY POLICY.............................................................................................................................8
PRINCIPLE OF OPERATION ..........................................................................................................................9
SPECIFICATIONS .......................................................................................................................................11
INSTALLATION (19” RACK MOUNT ENCLOSURE).....................................................................................12
INSTALLATION (NEMA 4X, WALL MOUNT ENCLOSURE)........................................................................14
ELECTRICAL AND PNEUMATIC CONNECTIONS...........................................................................................20
1.7.1 Analog Output .................................................................................................................................20
1.7.2 Relay Outputs..................................................................................................................................20
1.7.3 Digital Outputs................................................................................................................................20
1.7.4 RS232 Output ..................................................................................................................................22
1.7.5 Pneumatic system ............................................................................................................................22
1.7.6 Sample gas connection....................................................................................................................23
1.7.7 Exhaust connections........................................................................................................................24
OPERATION VERIFICATION .......................................................................................................................27
OPTIONS...................................................................................................................................................29
V
VI
2.1
KEY FEATURES.........................................................................................................................................31
2.1.1 O3 readout.......................................................................................................................................31
2.1.2 Stream Selector (Optional)..............................................................................................................31
2.1.3 Concentration Alarms .....................................................................................................................31
2.1.4 O analog output.............................................................................................................................31
3
2.1.5 E2 ROM backup of software configuration.....................................................................................31
2.1.6 RS-232 interface..............................................................................................................................32
FRONT PANEL DISPLAY ...........................................................................................................................32
2.2
2.2.1 Front panel display fields................................................................................................................32
2.2.2 Status LED’s....................................................................................................................................34
SOFTWARE OPERATION............................................................................................................................35
2.3
2.3.1 Main Menu ......................................................................................................................................35
2.3.2 Multi-Stream Operation ..................................................................................................................36
2.3.3 Alarm Status Menu..........................................................................................................................37
3.0 SETUP MODE .....................................................................................................................................39
CONFIGURING AUTO SEQUENCE (SEQ)....................................................................................................39
3.1
STREAM CONFIGURATION (STRM)..........................................................................................................41
3.2
SETTING THE CONCENTRATION UNITS (UNIT) ........................................................................................42
3.3
SETTING THE TIME-OF-DAY AND DATE (CLK)..........................................................................................43
3.4
SETTING THE RS-232 BAUD RATE (COMM)............................................................................................43
3.5
SETUP VARIABLES (VARS)......................................................................................................................43
3.6
4.0 MAINTENANCE..................................................................................................................................45
4.1
REPLACING THE SAMPLE FILTER ELEMENT ...............................................................................................45
CLEANING EXTERIOR SURFACES OF THE M450........................................................................................46
4.2
5.0 ADJUSTMENTS..................................................................................................................................47
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page iii
5.1
CALIBRATION ...........................................................................................................................................47
ADJUSTING UV LAMP AND DETECTOR ......................................................................................................47
5.2
CHANGING THE ANALOG OUTPUT RANGE..................................................................................................49
5.3
5.3.1 Voltage output..................................................................................................................................49
5.3.2 Current Output (Optional)...............................................................................................................50
A/D - D/A CALIBRATION PROCEDURE......................................................................................................50
5.4
CURRENT LOOP CALIBRATION..................................................................................................................51
5.5
6.0 MODEL 450 SPARE PARTS...............................................................................................................53
APPENDIX A - ELECTRICAL SCHEMATIC INDEX .........................................................................55
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page iv
FIGURES
FIGURE 1.1 - REAR PANEL PNEUMATIC CONNECTIONS............................................................13
FIGURE 1.2 – NEMA 4X ENCLOSURE DIMENSIONS.......................................................................16
FIGURE 1.3 – NEMA 4X AC LINE POWER CONNECTION.............................................................17
FIGURE 1.4 – NEMA 4X FRONT PANEL..............................................................................................18
FIGURE 1.5 – NEMA 4X PNEUMATIC CONNECTIONS...................................................................19
FIGURE 1.6 - ELECTRICAL SIGNAL CONNECTIONS.....................................................................20
FIGURE 1.7 - CONNECTING DIGITAL OUTPUTS TO AN EXTERNAL DEVICE .......................22
FIGURE 1.8 - FLOW DIAGRAM.............................................................................................................23
FIGURE 1.9 - M450 ASSEMBLY LAYOUT...........................................................................................25
FIGURE 1.10 - M450 NEMA 4X LAYOUT.............................................................................................26
FIGURE 2.1 - MODEL 450 FRONT PANEL ..........................................................................................32
FIGURE 2.2 - MAIN MENU......................................................................................................................35
FIGURE 2.3 - MAIN MENU, MANUAL MODE ....................................................................................36
FIGURE 2.4 - ALARM STATUS MENU.................................................................................................37
FIGURE 3.1 - SETUP MENU....................................................................................................................39
FIGURE 3.2 - SEQUENCE EDITING......................................................................................................40
FIGURE 3.3 - STREAM CONFIGURATION MENU............................................................................41
FIGURE 4.1 - DETECTOR PREAMP ADJUSTMENT.........................................................................48
FIGURE 4.2- ANALOG OUTPUT JUMPERS........................................................................................50
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page v

TABLES

TABLE 1.1 DIGITAL OUTPUTS..............................................................................................................21
TABLE 1.2 FINAL TEST AND CALIBRATION VALUES...................................................................28
TABLE 2.1 SYSTEM MODES...................................................................................................................33
TABLE 2.2 TEST MEASUREMENTS......................................................................................................33
TABLE 2.3 WARNING MESSAGES........................................................................................................34
TABLE 2.4 STATUS LED'S.......................................................................................................................35
TABLE 2.5 ALARM CODES.....................................................................................................................37
TABLE 3.1 CONCENTRATION UNITS..................................................................................................42
TABLE 4.1 UV SOURCE LAMP AND DETECTOR DIAGNOSTICS.................................................48
TABLE 4.2 V/F BOARD DIP SWITCH - RANGES FOR ANALOG OUTPUT...................................50
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page vi

1.0 INTRODUCTION

1.1 Preface

Teledyne API is pleased that you have purchased the Model 450. We offer a full one year warranty (see Section 1.2) and we at Teledyne API will be pleased to provide you with any support required so that you may utilize our equipment to the fullest extent.
The Teledyne API Model 450 keyboard/operator interface with its "talking keys" makes the Teledyne API a very user-friendly system. We hope you will not experience any problems with the Teledyne API Model 450 but if you do, the built-in tests and diagnostics should allow you to quickly and easily find the problem. In addition, our full time customer service department is always available to answer your questions.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 7

1.2 WARRANTY POLICY

ADVANCED POLLUTION INSTRUMENTATION DIVISION (T-API)

02024b
Prior to shipment, Teledyne API equipment is thoroughly inspected and tested. Should equipment failure occur,
COVERAGE
After the warranty period and throughout the equipment lifetime, Teledyne API stands ready to provide on-site or in-plant service at reasonable rates similar to those of other manufacturers in the industry. All maintenance and the first level of field troubleshooting are to be performed by the customer.
NON-TELEDYNE API MANUFACTURED EQUIPMENT
Equipment provided but not manufactured by Teledyne API is warranted and will be repaired to the extent and according to the current terms and conditions of the respective equipment manufacturers warranty.
GENERAL
Teledyne API warrants each Product manufactured by Teledyne API to be free from defects in
material and workmanship under normal use and service for a period of one year from the date of delivery. All replacement parts and repairs are warranted for 90 days after the purchase.
If a Product fails to conform to its specifications within the warranty period, correct such defect by, in
refunding the purchase price of such Product.
The warranties set forth in this section shall be of no force or effect with respect to any Product: (i) that has been altered or subjected to misuse, negligence or accident, or (ii) that has been used in
any manner other than in accordance with the instruction provided by properly maintained.
THE WARRANTIES SET FORTH IN THIS SECTION AND THE REMEDIES THEREFORE ARE EXCLUSIVE AND IN LIEU OF ANY IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESSED OR IMPLIED. THE REMEDIES SET FORTH IN THIS SECTION ARE THE EXCLUSIVE REMEDIES FOR BREACH OF ANY WARRANTY CONTAINED HEREIN. TELEDYNE API SHALL NOT BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF OR RELATED TO THIS AGREEMENT OF TELEDYNE API'S PERFORMANCE HEREUNDER, WHETHER FOR BREACH OF WARRANTY OR OTHERWISE.
Teledyne API assures its customers that prompt service and support will be available.
Teledyne API shall
Teledyne API's discretion, repairing or replacing such defective Product or
Teledyne API or (iii) not
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 8
TERMS AND CONDITIONS
C
T
P
All units or components returned to Teledyne API should be properly packed for handling and returned freight prepaid to the nearest designated Service Center. After the repair, the equipment will be return e d , f r eight pr e p a i d .

1.3 Principle of Operation

The detection of ozone molecules is based on absorption of 254 nm UV light due to an internal electronic resonance of the O3 molecule. The Model 450 uses a mercury lamp constructed so that a large majority of the light emitted is at the 254nm wavelength. Light from the lamp shines down a hollow quartz tube that is alternately filled with sample gas, then filled with gas scrubbed to remove ozone. The ratio of the intensity of light passing through the scrubbed gas to that of the sample forms a ratio I/Io. This ratio forms the basis for the calculation of the ozone concentration.
The Beer-Lambert equation, shown below, calculates the concentration of ozone from the ratio of light intensities.
C

O
3
Where:
I = Intensity of light passed through the sample I
= Intensity of light through sample free of ozone
o
= absorption coefficient
l = path length
= concentration of ozone in ppb
O
3
= sample temperature in degrees Kelvin
= pressure in inches of mercury
As can be seen the concentration of ozone depends on more than the intensity ratio. Temperature and pressure influence the density of the sample. The density changes the number of ozone molecules in the absorption tube which impacts the amount of light removed from the light beam. These effects are addressed by directly measuring temperature and pressure and including their actual values in the calculation. The absorption coefficient is a number that reflects the inherent ability of ozone to absorb 254 nm light. Most current measurements place this value at 308 cm-1 atm-1 at STP. The value of this number reflects the fact that ozone is a very efficient absorber of UV radiation which is why stratospheric ozone protects the life forms lower in the atmosphere from the harmful effects from solar UV radiation. Lastly, the absorption path length determines how many molecules are present in the column of gas in the absorption tube.
9
10
l

o
273

2992
.
inHg
n
l
o
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 9
The intensity of light is converted into a voltage by the detector/preamp module. The voltage is converted into a number by a voltage-to-frequency (V/F) converter capable of 80,000 count resolution. The digitized signal, along with the other variables, are used by the CPU to compute the concentration using the above formula.
Every 6 seconds the M450 completes a measurement cycle consisting of a 2 second wait period for the sample tube to flush, followed by a 1 second measurement of the UV light intensity to obtain I. The sample valve is switched to admit scrubbed sample gas for 2 seconds, followed by a 1 second measurement of the UV light intensity to obtain Io. Measurement of the Io every 6 seconds minimizes instrument drift due to changing intensity of the lamp due to aging and dirt.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 10

1.4 Specifications

Ranges User selectable to any full scale range from
Measurement Units ppbv, ppmv, ppmw, pphm, µg/m3, mg/ m3 Zero Noise < .0015 ppm (rms) Span Noise < .5% of reading (rms) (above 0.1 ppm) Lower Detectable Limit < .003 ppm (rms) Linearity Better than 1% full scale Response Time (95%) <30 sec Minimum Stream Duration (Cycle Time)* Sample Flow Rate 1-2.5 L/min Temperature Range 5-45o C Humidity Range 10-90% RH, Non-Condensing Dimensions (H x W x D) 7" x 17" x 18.2"
Weight 28 lb. (12.7 kg) - Single Channel Monitor, 19”
Power 110V/60 Hz, 220V/50 Hz, 240V/50Hz 250 watts
Environmental Conditions Installation Category (Overvoltage Category) II
Maximum Operating Altitude 2000 meters Analog Output Voltage Mode 100mV, 1V, 5V, 10V, (User Selectable) Isolated Analog Output 0-20mA
Mode** *Multi-Channel Monitor only **Optional
1ppm to 1000 ppm
1.0 min
(178 mm x 432 mm x 463 mm)
Rack Mount 30 lb. (13.6 kg) - Multi-Channel Monitor, 19” Rack Mount 38 lb (17.3 kg) – NEMA 4X Enclosure
230V~, 50Hz, 2.5A
Pollution Degree 2
Maximum voltage between outputs and ground 60V peak
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 11

1.5 Installation (19” Rack Mount Enclosure)

The Model 450 is shipped with the following standard equipment:
1. Power cord.
2. Instruction manual.
CAUTION
Upon receiving the Model 450 please do the following:
1. Verify no apparent shipping damage. (If damage has occurred please advise
2. Remove all red colored shipping screws from the inside and bottom of the
3. When installing the Model 450, allow a minimum of 4 inches of clearance at
4. Connect sample inlet line(s) to the sample port on rear panel.
To avoid personal injury, always use two persons to lift and
carry the Model 450.
shipper first, then Teledyne API.)
instrument. There are four screws on the optical bench mounts that are accessed from the inside of the instrument and two on the pump bracket that are accessed from the bottom of the instrument. Note: Save these shipping
screws and re-install them whenever the unit is shipped to another location.
the back of the instrument and 1 inch of clearance on each side for proper ventilation.
Connect the exhaust fitting on the rear panel (See Fig. 1.1) to a
suitable vent outside the monitor area.
CAUTION
NOTE
See Figure 1.1 for rear panel pneumatic connections. Sample
lines made from an inert material such as Teflon should be used
to minimize sample degradation.
5. Connect the power cord to an appropriate power outlet (see the serial number
tag for correct voltage and frequency).
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 12
CAUTION
Verify that the instrument is set up for proper line voltage
and frequency.
6. Turn on the M450 by switching the switch on the lower right corner of the front panel. The front panel display should light with a sequence of messages,
-API - M450 - software version number, then a normal display as shown in Figure 2.2.
7. Allow about 20 minutes for the UV lamp temperature to come up to it’s setpoint then press the TST> button on the front keyboard to scroll through the TEST values. Compare these values to those noted during the final factory checkout listed in Table 1.2. The values observed should closely match the Table 1.2 values.
NOTE
Words in all caps are messages on the monitor front
panel.
FIGURE 1.1 - REAR PANEL PNEUMATIC CONNECTIONS
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 13

1.6 Installation (NEMA 4X, Wall Mount Enclosure)

The Model 450 is shipped with the following standard equipment:
1. Instruction manual.
CAUTION
To avoid personal injury, always use two persons to lift and
carry the Model 450.
Upon receiving the Model 450 please do the following:
1. Verify no apparent shipping damage. (If damage has occurred please advise
shipper first, then Teledyne API.)
2. Mount the NEMA 4X enclosure to a vertical surface using four 5/16” bolts.
Mounting hole dimensions are shown in Figure 1.2.
3. Check the serial number tag on the front panel to ensure that the monitor has
been configured for the proper AC line voltage and frequency.
4. Route the line power cable through one of the conduit bulkheads on the
bottom surface of the enclosure.
5. Connect the line power to the power entry barrier strip shown in Figure 1.3.
CAUTION
Use a 3-wire grounded power cord
and connect as shown in Figure 1.3
6. Route signal wires into the enclosure through the conduit bulkheads. These
wires should be routed up to the front panel through the notch in the lower left-hand corner of the front panel as shown in Figure 1.4. Make sure to allow enough extra length in these wires so that the panel can still swing open without stressing the wires.
7. Connect ¼” sample inlet line(s) to the sample inlet fitting(s) shown in Figure
1.5.
8. Connect a ¼” exhaust line to the fitting shown in Figure 1.5.
CAUTION
Connect the exhaust fitting to a suitable vent outside the monitor
area.
9. The Model 450 can now be turned on using the power switch located on the
front panel.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 14
10. Allow about 20 minutes for the UV lamp temperature to come up to it’s setpoint then press the TST> button on the front keyboard to scroll through the TEST values. Compare these values to those noted during the final factory checkout listed in Table 1.2. The values observed should closely match the Table 1.2 values.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 15
FIGURE 1.2 – NEMA 4X ENCLOSURE DIMENSIONS
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 16
FIGURE 1.3 – NEMA 4X AC LINE POWER CONNECTION
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 17
FIGURE 1.4 – NEMA 4X FRONT PANEL
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 18
FIGURE 1.5 – NEMA 4X PNEUMATIC CONNECTIONS
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 19
FIGURE 1.6 - ELECTRICAL SIGNAL CONNECTIONS

1.7 Electrical and pneumatic connections

1.7.1 Analog Output

The analog output is a two pin connector on the rear panel (See Figure 1.6). The analog output can be configured for voltage or current output. The standard is a 0-5 volt output. 0-100mV, 0-1V, 0-10V, and 4-20 mA outputs are also available. See Section 5.3 for information on setting other output ranges.

1.7.2 Relay Outputs

Three form C relay outputs are provided on the rear panel on three pin connectors (See Figure 1.6). These outputs correspond to the ‘HI’ and ‘HI-HI’ alarms for Stream 1 and a System OK status output that is used to indicate a fault or error condition in the instrument. The relay contacts are rated to 3A at 240VAC. Do not exceed these ratings when connecting equipment to the instrument.

1.7.3 Digital Outputs

There are three sets of digital outputs on the rear panel. These outputs are optically isolated NPN transistors which can pass 50 mA of DC current. These outputs can be used to interface to devices that accept logic-level digital inputs, such as Programmable Logic Controllers(PLC’s).
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 20
The outputs labeled ‘STATUS OUT’ are used to indicate instrument operational status and fault conditions. The outputs labeled ‘CHANNEL ID’ are used to indicate which channel(stream), the instrument is currently monitoring. The outputs labeled ‘ALARM OUT’ are used to indicate the status of the HI and HI-HI alarms for each channel. Table
1.1 below summarizes the functions of all the digital outputs.
Output Description
STATUS OUT 1 Power OK 2 Diagnostics Mode(Instrument not monitoring) 3 Temperature Fault 4 Pressure Fault 5 UV Lamp Fault 6 Flow Fault
7-11 Not Used
12 System OK(no faults)
CHANNEL ID 1 Monitoring Stream #1 2 Monitoring Stream #2 3 Monitoring Stream #3 4 Monitoring Stream #4 5 Monitoring Stream #5 6 Monitoring Stream #6
ALARM OUT CH 1 HI and HI-HI alarms for Stream #1 CH 2 HI and HI-HI alarms for Stream #2 CH 3 HI and HI-HI alarms for Stream #3 CH 4 HI and HI-HI alarms for Stream #4 CH 5 HI and HI-HI alarms for Stream #5 CH 6 HI and HI-HI alarms for Stream #6
TABLE 1.1 DIGITAL OUTPUTS
Each digital output is configured as a Collector/Emitter pair. The rear panel labels are ‘C’ and ‘E’ for these contacts, respectively. Figure 1.7 below shows the most common way of connecting the digital outputs to an external device such as PLC. Note: Most devices, such as PLC’s, have internal provision for limiting the current that the input will draw from an external device. When connecting to a unit that does not have this feature, external dropping resistors must be used to limit the current through the transistor output to 50mA or less.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 21
FIGURE 1.7 - CONNECTING DIGITAL OUTPUTS TO AN EXTERNAL DEVICE

1.7.4 RS232 Output

The RS232 output is provided on the 9-pin D-Sub connector shown in Figure 1.6. The RS232 output can be connected to a computer or serial printer. The RS232 output can be used to record alarm events with a time and date stamp or can be used to control the instrument’s operation. Almost any function that can be done through the front panel interface can also be done remotely through the RS232 interface.
To use the RS232 for instrument control, all that is needed is a PC(personal computer) with an available serial communications port(COM port), a serial cable, and terminal emulation software. The serial cable must have a female DB-9 connector on one end and an appropriate connector on the other end to interface with the COM port on a PC.
For more details on the use of the RS232 interface, please contact Teledyne API and request document number 01350.

1.7.5 Pneumatic system

Figure 1.8 shows a block diagram of the M450 pneumatic system. The Model 450 is equipped with a vacuum pump that pulls the sample gas through the instrument. The sample flow is set using the needle valve located on the front panel flowmeter. The sample flow should be 1-2.5 L/min. A sample flow of less than 1 L/min will result in inaccurate ozone measurement.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 22
FIGURE 1.8 - FLOW DIAGRAM

1.7.6 Sample gas connection

1/4" O.D. PTFE or FEP (Teflon) tubing is needed to connect the sample source to the monitor. Any fittings used in the sample lines should be constructed of stainless steel or PTFE.
Teledyne API recommends that the length of tubing connecting the sample points to the monitor should be kept to 100 feet or less. When running long sample lines, maintaining a high sample flow near 2 L/min will aid in reducing ozone loss in these lines. If more than 100 feet of sample line must be used, check to make sure that the pressure drop between the sampling point and the monitor be kept to 5 in-Hg or less. This can be measured by observing the Sample Pressure test function on the front panel of the instrument with the sample line connected at the rear panel and with the sample line disconnected. The difference between the two pressures is approximately equal to the pressure drop in the sample line.
NOTE
For the sampling lines, use only ¼” O.D. PTFE or FEP tubing. The tubing must be pre-
conditioned to ozone prior to installation to minimize ozone loss in the sampling lines.
Pre-conditioned ¼” FEP tubing is available from Teledyne API (Part number 02639)
Provision should be made for keeping dust and other particulate matter out of the monitor. As an option, the monitor may be configured with a PTFE particulate filter downstream of the stream selector manifold. In addition to this, in-line filters should be installed in the sample lines for each stream. To avoid dust build up in the sample lines, these filters should be placed at the inlet end of the sampling line. These filters should be
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 23
constructed of PTFE or other inert material to avoid degradation of the ozone concentration. Contact Teledyne API’s sales department for additional information on particulate filters.
Care must be taken in the design of the sampling system to ensure that water vapor does not condense in the sampling lines or in the instrument itself. The sample lines should be routed to avoid large temperature gradients along the lines. If sample is brought in from a hot, humid area into a much cooler area, then low power heating tape should be wrapped around the lines to keep the temperature of the sample lines up and avoid condensation. The lines should also not have loops or low points where water could collect.
For applications that require sampling of very humid gas, a Sample Conditioner may be required to lower sample gas humidity. Please contact the Teledyne API Sales Department for additional information.

1.7.7 Exhaust connections

A single 1/4" O.D. tube should be connected from the Monitor sample exhaust to an area outside of the room the monitor occupies. The maximum length of the exhaust line should not exceed 30 feet.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 24
FIGURE 1.9 - M450 ASSEMBLY LAYOUT
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 25
FIGURE 1.10 - M450 NEMA 4X LAYOUT
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 26

1.7 Operation verification

The Model 450 Monitor is now ready for operation.
1. Read Sections 1.3 and all of Section 4 of the manual to understand the
Monitor's operation.
2. Turn on the power by pressing the on/off switch on the front panel (see Figure
2.1). The display should turn on and green (sample) status LED should be energized. The yellow "fault" light may also be on until the flows, temperatures and voltages are within operating limits. Clear the fault messages.
3. Sample Flow - Adjust to 1-2.5 L/min using needle valve on flowmeter.
4. After a 20 minute warm-up, review the TEST function values in the front
panel display by pushing the keyboard button labeled TST TEST function is a diagnostic of correct monitor operation, therefore TEST functions not covered below can be ignored for now.
5. O3 REF, O3 MEAS - TEST function values should be between 4200 mV and
4700 mV.
6. Pressure - 29 to 30 Inches-Mercury-Absol ute at sea level. Other values will
be displayed depending on altitude of monitor.
7. Sample Temp - 20 -50oC
8. Box Temp - Ambient +10 oC
9. If the TEST functions are within the limits given above the instrument should
function correctly. If there is a problem please read the manual and check your set-up.
>. Not every
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 27
TEST Values Observed
Value
O3 MEAS mV 4200-4700
O3 REF mV 4200-4700
PRESS in-Hg-A 25 - 35
SAMPLE TEMP oC 20-50
FLOW L/MIN 1-2.5
Span Values
Noise at Zero(rms) PPB .5-1.5
Noise at
Span(rms)
Factory Installed Options Option Installed
Power Voltage/Frequency Stream Selector (3 channel) Stream Selector (6 channel)
Sample Filter
Flow Switch
All SS Fittings
Rack Mount, w/ Slides
Isolated 4-20mA Output
Voltage Range 0- V
PROM Rev # Date Technician
PPB .5% of reading
Units Nominal Range
Serial #
TABLE 1.2 FINAL TEST AND CALIBRATION VALUES
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 28

1.8 Options

1.8.1 Stream Selector (3 or 6 Channel)
This option allows the M450 to measure ozone at up to six different sampling points by using an internal stream selector. 3 or 6 channel stream selectors are available. This option includes digital outputs for HI and HI-HI alarms and a channel ID for each channel in the stream selector.
1.8.2 Sample Filter
This option consists of a 47mm diameter PTFE particulate filter inside the instrument. In instruments with the stream selector option, this filter is placed immediately downstream of the stream selector. The filter can be easily accessed through the front of the instrument by folding down the hinged front panel. No tools are necessary to change the filter element.
1.8.3 Flow Switch
This option is an on/off type flow switch that is used to send a signal to the CPU in case the flow drops below the minimum required for proper operation of the instrument. In case of flow failure, the instrument will issue a warning on the front panel and will turn off the System OK relay, indicating an instrument error.
1.8.4 Isolated Current Output(4-20mA)
This option converts the M450’s voltage output to an isolated 4-20mA driver with an integral loop power supply. When connecting to the current output, ensure that the maximum voltage between the outputs and ground is 60V peak.
1.8.5 Rack mount with slides
This option, including slides and rack mounting ears, permits the Monitor to be mounted in a standard 19" wide x 30" deep RETMA rack.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 29

2.0 OPERATION

2.1 Key features

The important features of the Teledyne API Model 450 O3 Monitor are listed below.

2.1.1 O3 readout

The Teledyne API Model 450 O3 Monitor constantly displays the current ozone reading (in units specified) in the upper right hand corner of the alphanumeric display.
If a multi-channel option has been installed, the reading represents the ozone concentration for the stream currently being monitored. The last concentration measurement for the other channels can be viewed by scrolling through the Test functions (See Section 2.2.)

2.1.2 Stream Selector (Optional)

As an option, the Model 450 can be equipped to sequentially monitor up to 6 separate sample streams. The monitoring duration for each stream as well as the sequence order can be configured by the user through the front panel software. See Sections 3.1 and 3.2 for stream and configuration sequencing.

2.1.3 Concentration Alarms

The Teledyne API O3 Monitor provides two concentration alarms, HI and HI-HI, per sample stream. The concentrations corresponding to each alarm levels can be independently configured for each sample stream. In addition, the alarms can operate in either latching or non-latching modes.

2.1.4 O3 analog output

The Teledyne API O3 Monitor provides an analog output of the current O3 reading on the rear panel (see Figure 1.6). The output can be configured for 0-100mV, 0-1V, 0-5V, 0-10V or 4-20mA non-isolated current output. Isolated 4-20mA or 2-20mA current output is available as an option. The voltage outputs are bi-polar and also provide for 20% over-range.

2.1.5 E2 ROM backup of software configuration

The Teledyne API O3 Monitor has few DIP switches or jumpers that need to be set by the operator. Configuration of the Monitor is done under software control and the configuration options are stored in electrically erasable (E options are saved even when the Monitor is powered off.
2
) ROM. Thus, configuration
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 31
There is one exception to this. The analog output voltage range is set by DIP Switches
A
on the A/D-I/O board as shown in Section 5.3.

2.1.6 RS-232 interface

The Teledyne API O3 Monitor features an RS-232 interface which can output the instantaneous and/or average O3 data to another computer. It can also be used as a command and status channel to allow a computer to control the Monitor.

2.2 Front Panel Display

This section describes the operator interface from the point of view of the front panel. The front panel consists of a 2-line by 40-character alphanumeric display, 8 pushbuttons, and 3 status LED’s. Each of these features is described below.

2.2.1 Front panel display fields

The display is divided into 4 main "fields": the Mode field in the upper left, the Message/Test Function field in the top center, the ozone Concentration field consisting of the most recent instantaneous ozone value field in the upper right, and the Menu field which occupies the entire bottom line of the display. The Menu field is used to define the function of the 8 buttons directly below the display. The buttons are then used for selecting menu items and are also used for entering values such as alarm levels. A typical display is shown in Figure 2.1.
MODE
FIEL D
UTO #1 TIME=15:58:21 O3=0.005
<TST TST> MODE ALRMSTRM MSG CLR SETUP
MESSAGE/TEST
MEASUREMENT FIELD
MENU DEFINITIONS
CONCENTRA TION
FIEL D
FIGURE 2.1 - MODEL 450 FRONT PANEL
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 32
Mode Field
The mode field indicates the current mode of the Monitor. If it shows "AUTO #1", the instrument is in the automatic sequence mode(multi-stream instruments only) and that stream #1 is the active stream. Table 2.1 lists all the possible modes in the Monitor and their meanings.
Mode Meaning
AUTO #x Measuring stream #x in auto sequence mode(multi-stream
instruments only)
MANUAL #x Measuring stream #x in manual mode
SETUP xxx (1) Configuring monitor (monitoring continues)
DIAG CAL Calibration Menu
DIAG D/A Configure and Calibrate Digital to Analog converters
DIAG AOUT Test analog output
DIAG CFG Instrument Configuation List
(1) xxx = software revision (e.g. A.9)
TABLE 2.1 SYSTEM MODES
Message/Test Measurement Field
The message field shows warning messages or test measurements. Tables 2.2 and 2.3 summarize the test measurements and warning messages and their meanings. Refer to Section 2.3 for detailed information on viewing test measurements and warning messages and clearing warnings.
Test Message Meaning
TIME=xx:xx:xx Current time-of-day (HH:MM:SS)
STREAMx=x.xxx PPM Current O3 concentration for stream x
O3 MEAS=xxxxx MV Current UV reading, measure channel
O3 REF=xxxxx MV Current UV reading, reference channel
PRES=xxx IN-HG-A Absorption cell pressure - inches Hg
SAMPLE TEMP=xxx C Temperature of the sample (deg. C)
BOX TEMP=xxx C Internal box temperature (deg. C)
TABLE 2.2 TEST MEASUREMENTS
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 33
Warning Message Meaning
SYSTEM RESET Issued whenever Monitor is powered on
RAM INITIALIZED RAM was erased
ANA LAMP WARNING UV lamp < 2500mV OR => 5000mV
ANA LAMP SHUTDOWN UV lamp temp control not working
SAMPLE FLOW WARNING Sample flow < 700 cc/m
SAMPLE PRESSURE WARN Sample pressure < 15 or > 35 In-Hg-A
SAMPLE TEMP WARNING Sample temp < 10 or > 50 deg. C
BOX TEMP WARNING Box temp. < 12 deg. C or > 55 deg. C
ANA LAMP TEMP WARN UV lamp temp < 51 or > 61 deg. C
V/F NOT INSTALLED V/F card not installed or bad
TABLE 2.3 WARNING MESSAGES
Menu Field
The menu field changes depending on the mode of the Monitor and the buttons that have been pressed. It indicates the current function of each of the 8 pushbuttons below the display.
The 8 pushbuttons below the display are programmable by the CPU in that their functions change depending on the mode of the Monitor or the operations being performed. The legend above a button identifies its current function. If there is no legend above a button, it has no function and will be ignored if pressed.

2.2.2 Status LED’s

The three status LED’s to the right of the display indicate the general status of the Model 450 Monitor. The green SAMPLE LED indicates the sampling status. The yellow FAULT LED indicates the fault status. The red ALARM LED indicates the concentration alarm status. Table 2.4 below summarizes the meanings of the status LED’s.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 34
A
LED State Meaning
Green Off
Not monitoring(1)
On Monitoring normally Blinking Monitoring, Manual
Mode
Yellow Off No warnings exist
Blinking Warnings exist
Red Off No Alarms active
Blinking Alarms active
(1) This occurs during all diagnostics modes
TABLE 2.4 STATUS LED'S

2.3 Software Operation

This section describes the operation of the instrument software through the front panel interface described in the previous section. The instrument software has been designed to be easy to use, yet powerful enough to allow the user to customize the instrument for a particular application.

2.3.1 Main Menu

Figure 2.2 below shows the shows a typical main menu for a Model 450 with the multi­stream option.
UTO #1 TIME=15:58:21 O3=0.005
<TST TST> MODE ALRM MSG CLR SETUP
FIGURE 2.2 - MAIN MENU
If <TST or TST> are pushed, the upper center display cycles through list of available test functions (see Table 2.3).
When the MSG button is displayed, it indicates that one or more warnings have been issued. To view the warning messages, press the MSG button. Repeatedly pressing this button will cycle through all available warning messages. Pressing CLR will clear the warning message from the display. Note: If the conditions that caused the warning
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 35
messages are still in effect, the warning messages will re-appear after they have been cleared.
The ALRM button is displayed when concentration alarms have been triggered. See Section 2.3.3 for operation of the Alarm menu.
The MODE button will toggle the instrument between Auto and Manual modes. The current mode is shown in the upper left-hand corner of the display. In Auto mode, the instrument will continuously cycle through a user defined sequence of streams(See Section 3.2 for sequence configuration.) When the instrument is in manual mode, it will hold on a particular stream being measured until the user switches back to Auto mode or manually switches to a different stream. Figure 2.3 below shows the main menu when the instrument is in Manual mode.
MANUAL #1 TIME=15:58:21 O3=0.005
<TST TST> MODE ALRMSTRM MSG CLR SETUP
FIGURE 2.3 - MAIN MENU, MANUAL MODE
In Manual mode, the STRM button appears. Pressing this button will manually cycle through the available streams. Press STRM until the stream you wish to monitor appears in the mode display in the upper left-hand corner of the display. The instrument will hold and monitor this stream until the user places the instrument into Auto mode or manually switches to another stream by pressing the STRM button again.
Pressing MODE will toggle the instrument back to Auto mode and the automatic sequence will resume.

2.3.2 Multi-Stream Operation

This section describes the operation of the Multi-Stream M450 in automatic sequence mode (AUTO.) In this mode the monitor is designed to continuously cycle through a user-defined sequence of sampling points. This cycling is accomplished by activating valves on the stream selector manifold on the rear panel to admit sample gas for a particular stream.
When the M450 first switches to a new sample stream, the instrument goes through a 30 second dwell phase where sample gas is pulled from the new sample stream but no readings are taken. The purpose of this dwell phase is to purge out any “old” sample gas
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 36
that has been sitting in the sample line while that stream was inactive. During the dwell phase the Concentration field on the display (see Figure 2.1) will blink and the instrument will display the last concentration recorded for that stream. After the 30 seconds has elapsed, the concentration field will stop blinking and the current concentration will be displayed. The current stream will now be monitored for the rest of the Monitoring Duration that has been set for that stream (see Section 3.2 for details on configuring the individual streams.) The actual time period that the stream is measured is the Monitoring Duration minus the 30-second dwell. So a 1-minute Monitoring Duration consists of 30 seconds of dwell followed by 30 seconds of measurement.
For systems that do not require a 30 second dwell between streams, this value can be changed. Please consult Teledyne API Customer Service for instructions on how to do this.

2.3.3 Alarm Status Menu

Pressing the ALRM button from the main menu when alarms have been triggered will bring up the alarm status menu as shown below in Figure 2.4.
1=H 2=HH 3=OK 4=OK 5=OK 6=OK
CLR CLR BEEP EXIT
FIGURE 2.4 - ALARM STATUS MENU
The Alarm menu displays the alarm status of all available streams along the top row of the display. The status of each stream is indicated by a one or two letter code shown in Table 2.5 below.
Alarm Code Meaning
OK No alarms triggered
H ‘HI’ alarm triggered
HH ‘HI-HI’ alarm triggered
TABLE 2.5 ALARM CODES
Clearing Alarms
If the monitor is configured for Latching type alarms, then a CLR button will appear below each stream with an active alarm. Pressing the CLR button will clear the alarm
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 37
condition, provided that the concentration for that stream has dropped below the alarm threshold.
If the monitor is configured for Non-Latching alarms, the CLR button will not appear, and all alarms will automatically clear when the concentration drops below the alarm threshold.
Audible Beeper
An audible beeper is sounded when any alarm is activated. A slow beep indicates that one ore more ‘HI’ alarms has been triggered, and a fast beep indicates that one or more ‘HI-HI’ alarms has been triggered. Pressing the BEEP button from the alarm menu will silence the audible beeper for 5 minutes. Note: Pressing the BEEP button does not clear any of the alarms, it simply silences the audible alarm. If the alarms are not cleared, the audible alarm will automatically resume in 5 minutes.
See Sections 3.2 and 3.6 for details on configuring concentration alarms.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 38

3.0 SETUP MODE

This section describes the setup variables which are used to configure the Monitor. All the setup variables are stored in the Monitor's EEPROM and are retained during power off and even when new software revisions are installed.
If a variable is modified, but ENTR is not pressed, the variable will
not be changed and the monitor will beep when exit is pressed.
The setup menus are accessed by pressing the SETUP button from the instrument’s main menu. The top level setup menu is shown in Figure 3.1 below.
SETUP A.8
NOTE
STRM UNIT COMMCLK VARS DIAG EXIT
SEQ
FIGURE 3.1 - SETUP MENU

3.1 Configuring Auto Sequence (SEQ)

The automatic sequence determines the monitoring order for the streams in a multi­stream instrument. The default setting for a six channel instrument is 1-2-3-4-5-6. With this setting the instrument will cycle through all six channels in numerical order and loop back to the beginning when the cycle is complete. This sequence can be modified, however, to cycle through the channels in any order desired. The sequence can also be modified to monitor certain streams more often than others, 1-2-1-3-1-4-1-5-1-6 for example.
To modify the auto sequence, press SEQ from the setup menu. The sequence editing screen is shown in Figure 3.2 below.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 39
SETUP A.8 SEQUENCE:[1]23456
<CH
CH> INS [1]DEL ENTR EXIT
FIGURE 3.2 - SEQUENCE EDITING
The current sequence is shown at the top center of the display as a string of digits. Each digit in the sequence represents a stream number. To edit a digit, press the <CH or CH> buttons until the digit you wish to edit is shown in brackets []. As you scroll through the steps in the sequence, the digit in brackets is displayed above one of the buttons on the display. To change this digit to another stream number, simply press the button below the digit to increment the stream number. Pressing INS will insert a digit in front of the bracketed digit and pressing DEL will delete the bracketed digit. A stream can be inserted in the sequence more than once. The sequence can consist of a maximum of 12 digits.
When you are finished editing the sequence, press ENTR to store the new sequence in memory. The new sequence remains in memory even when the instrument is powered down. Pressing EXIT will exit the sequence editing menu without saving any changes.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 40

3.2 Stream Configuration (STRM)

The stream configuration menu is used to set alarm thresholds, monitoring duration, and analog output ranges for each individual stream. This menu is accessed by pressing the STRM button from the setup menu. The stream configuration menu is shown in Figure
3.3 below.
SETUP A.8 STRM 1)H=0.10,HH=0.30 PPM
PREV
NEXT EDIT PRNT EXIT
FIGURE 3.3 - STREAM CONFIGURATION MENU
Pressing PREV or NEXT will cycle through the available streams. The top center of the display shows the stream number and a summary of the alarm thresholds for this stream. In the example shown in Figure 3.3 above, Stream 1 is configured for a HI alarm at 0.10 PPM and a HI-HI alarm at 0.30 PPM.
To edit the parameters for a particular stream, scroll using the PREV or NEXT buttons until the stream is displayed on the top line. Press the EDIT button. A series of prompts will now be displayed as follows:
HI ALARM:OFF
Pressing the leftmost button on the display will toggle the HI Alarm on and off. Setting the alarm to off disables the alarm. Press ENTR to proceed to the next prompt. If the HI Alarm is set to ON then the next prompt will show:
HI ALARM LIMIT:0.100 PPM
The default value for the HI alarm threshold is 0.100 PPM. To change this value, edit the number shown on the lower line of the display by pressing the button below the digit you want to change. When you are done, press ENTR to store the new value. The next set of prompts sets the HI-HI alarm as follows:
HI-HI ALARM:OFF
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 41
Pressing the leftmost button on the display will toggle the HI Alarm on and off. Press ENTR to proceed to the next prompt. If the HI-HI Alarm is set to ON then the next prompt will show:
HI-HI ALARM LIMIT:0.300 PPM
Edit the HI-HI alarm value and press ENTR to store the value. The next prompt is:
ANA. OUT RANGE: 1.000 PPM
This is the analog output range setting. This is the concentration level that will correspond to a full scale voltage or current(4-20mA) output at the instrument’s analog output on the rear panel. Edit this value and press ENTR. The next prompt is:
MONITOR DUR.: 10.00 Minutes
This prompt sets the monitoring duration for a particular stream. The monitoring duration includes a 30-second purge cycle at the beginning of each measurement. During the purge cycle, the monitor takes no readings. Therefore, for a monitoring duration of 10 minutes, the instrument will purge for 30 seconds and then take readings for the next 9 minutes, 30 seconds before switching to the next stream in the sequence.
Pressing ENTR takes you back to the main stream configuration menu shown in Figure
3.3. You may now scroll to another stream and repeat the steps shown above or press EXIT to exit back to the previous menu.

3.3 Setting the Concentration Units (UNIT)

To set the concentration units, press UNIT from the setup menu. The default concentration units are PPM(Parts per Million). Table 3.1 below lists the units available.
Software Abbreviation Unit
PPB Parts per Billion by Volume PPHM Parts per Hundred Million by Volume PPM Parts per Million by Volume PPMW Parts per Million by Weight UGM MGM Milligrams per Cubic Meter(mg/ m3)
TABLE 3.1 CONCENTRATION UNITS
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 42
Micrograms per Cubic Meter(g/m
3
)

3.4 Setting the time-of-day and date (CLK)

To set the current time-of-day, which is used for determining when to do an automatic calibration and for time-stamping the RS-232 reports, press SETUP-CLK-TIME. The CPU will display the current time-of-day as four digits in the format "HH: MM", where "HH" is the hour in 24-hour format (i.e. hours range from 00 to 23) and "MM" is the minute (00 - 59). The operator may change the time-of-day and then press ENTR to accept the new time, or press EXIT to leave the time unchanged.
To set the current date, which is used for time-stamping the RS-232 reports, press SETUP-CLK-DATE. The CPU will display the current date as "DD MMM YY". For example, April 1, 1990 would be displayed as "01 APR 90". Change the date by pressing the button under each field until the desired date is shown. Then press ENTR to accept the new date or press EXIT to leave the date unchanged.

3.5 Setting the RS-232 baud rate (COMM)

To set the baud rate for the RS-232 channel, press SETUP-COMM-BAUD. Press 300, 1200, 2450, 4800, 9600, or 19.2 followed by ENTR to accept the new baud rate, or EXIT to leave the baud rate unchanged.

3.6 Setup variables (VARS)

The setup variables are global settings that can be configured by the user. Note: Do not arbitrarily change these settings since these variables affect the fundamental operation of the instrument. The setup variables are accessed from the main menu by pressing SETUP-VARS. Use the PREV or NEXT buttons to scroll to the desired variable and press EDIT to change the variable. Set the desired value for the variable and press ENTR to save the value. Pressing EXIT aborts the edit screen without changing the value. The setup variables available to the user are listed below along with their description.
0) LATCH_ALARMS
This variable sets the operation of the concentration alarms to either latched or non­latched mode. In latched mode, when a concentration alarm is triggered, the alarm will stay on, or latch, until the user resets the alarm through the ALRM menu on the front panel. When the alarms are set to non-latching mode(LATCH_ALARMS = OFF) the alarms will automatically turn off when the concentration drops back below the alarm threshold value.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 43
1) ALARM_BEEPER
This variable enables or disables the audible beeper that sounds whenever one of the alarms has been triggered. The beeper sounds a slow beep if any of the ‘HI’ alarms are active and sounds a fast beep if any of the ‘HI-HI’ alarms are active.
2) ANA_OUTPUT_MODE
This variable sets the Analog Output Mode for the Model 550 relay box accessory, if this is installed. This variable has no effect on the analog output of the M450 itself. The possible selections are: CURR (0-20mA or 4-20mA), 0.1V, 1.0V, 5V, 10V.
3) TPC_ENABLE
This variable enables temperature and pressure compensation for the calculation of the ozone concentration. This variable should always be set to ON unless an external correction is being applied to compensate for temperature and pressure changes.
4) STD_TEMP, 5) STD_PRESS
These variables are the standard temperature and pressure that are used to calculate the concentration when the units are set to Micrograms per Cubic Meter(g/m Milligrams per Cubic Meter(mg/ m
3
). These variables do not affect the ozone
3
) or
measurement in PPM or PPB units.
6) RS232_MODE
This variable sets the mode of operation for the RS232 interface. For more details on the use of the RS232 interface, please contact Teledyne API and request document number
01350.
7) CLOCK_ADJ
This variable is used to make slight adjustments to the instrument’s internal clock so that it keeps proper time. Setting this variable to a positive value will add that number of seconds each day to the clock while a negative value will subtract that number of seconds per day.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 44

4.0 Maintenance

4.1 Replacing the sample filter element

If your Model 450 has been equipped with an internal or external sample filter, the Teflon filter elements must be replaced on a regular basis to ensure accurate readings from the M450. The frequency of this operation will be determined by the particulate loading from the air being sampled and will vary greatly from location to location. When the instrument is first installed, the sample filters should be checked at least once a week for dirt loading and replaced if necessary. Once the replacement frequency is determined, a regular schedule for filter replacement should be instituted.
For replacement 47mm Teflon filter elements, please contact Teledyne API’s sales department and request part number 00969.
Filter Replacement Procedure:
1. If the M450 has been equipped with an internal filter, access the filter by releasing the
two quick-release tabs near the top of the front panel. The front panel will now swing down and the filter element can be viewed through the glass cover on the filter assembly, which is located on the left hand side of the panel.
2. To replace the element, unscrew the black knurled ring over the glass cover and
carefully remove the glass cover and the white Teflon retaining ring over the element. The element can now be removed and replaced.
1. Replace the Teflon retaining ring and glass cover. Note: When re-installing the
Teflon retaining ring, make sure the notches in the ring are visible and facing up towards the glass cover.
NOTE
Take care when handling the filter element and the sample filter parts that they are not contaminated with oil from your fingers or other sources. Even a small amount of oil on the wetted surfaces can cause the instrument to respond very slowly to ozone. Clean gloves or tweezers should be used to handle the clean filter elements.
2. Screw the black ring onto the glass cover. The ring should be screwed to a snug
position by hand, no tool should ever be used for installing the knurled ring.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 45

4.2 Cleaning Exterior Surfaces of the M450

If necessary, the front panel mask and keyboard of the M450 can be cleaned with a damp cloth. Do not attempt to clean any of the other surfaces of the instrument. Do not submerge any part of the instrument in water or cleaning solution.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 46

5.0 ADJUSTMENTS

CAUTION
RISK OF ELECTRICAL SHOCK. THE OPERATIONS
OUTLINED IN THE FOLLOWING SECTIONS OF THIS
CHAPTER ARE TO BE PERFORMED BY QUALIFIED
MAINTENANCE PERSONNEL ONLY!

5.1 Calibration

The Model 450 is calibrated to NIST traceable standards on the 0-1 ppm range prior to shipment. A calibration certificate for your instrument can be purchased from Teledyne API if required. Teledyne API also recommends that the Model 450 be re-calibrated once a year. Teledyne API can provide NIST traceable calibration services at our factory or on-site. Please contact our Customer Service department for details on these services.

5.2 Adjusting UV lamp and detector

Basic operation of the source lamp and detector can be determined by observing the value of the O3 REF test function. After the analyzer is warmed-up (15 min. to 30 min. after power-on), this value will give a good indication of the state of Lamp and Detector operation as follows:
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 47
UV Source Lamp and Detector Diagnostics
O3 REF Value Meaning
4700 mV to 5000 mV The Source Lamp and Detector are
operating, but adjustment is required
4000 mV to 4700 mV The Source Lamp and Detector are
operating properly; no adjustment is
needed
2500 mV to 4000 mV The Source Lamp and Detector are
operating. Adjustment is useful but not
required
150 mV to 2500 mV The Source Lamp and Detector are
operating, but adjustment is required
90 mV to 150 mV Either the Source Lamp or the UV
detector is not functioning
Less than 90 mV The Detector Pre-Amp or V/F Board
has failed or is disconnected
TABLE 4.1 UV SOURCE LAMP AND DETECTOR DIAGNOSTICS
Adjustment required or Adjustment useful:
The first adjustment that should be made is to adjust the gain on the UV Detector Preamp. The adjustment for the preamp is located through the preamp cover at the front of the optical bench (See Figure 4.1). The preamp should be adjusted to 4500 mV +/­100mV.
If the preamp cannot be adjusted to achieve this value, additional adjustment can be made by rotating the lamp inside the lamp housing at the rear of the optical bench(See Figure
1.9.) To adjust the lamp, loosen the lamp set-screw on the top of the optical bench and slowly rotate the lamp while observing the O3 REF value. Re-tighten the lamp setscrew.
FIGURE 4.1 - DETECTOR PREAMP ADJUSTMENT
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 48
If it is not possible to achieve an acceptable O3 REF test Value by means of adjustment, it is possible that the lamp has deteriorated beyond its useful range and should be replaced.
Lamp or Detector Failure:
WARNING
UV LIGHT PRESENT
DO NOT LOOK DIRECTLY AT THE UV LAMP SINCE UV
LIGHT COULD CAUSE EYE DAMAGE. ALAWYS USE UV
FILTERING GLASSES OR VIEW THROUGH GLASS.
A O3 REF value of approximately 125 mV usually indicates a total failure of either the source lamp or the detector. To determine which component is at fault, remove the top cover of the optical bench and observe the "lamp end" of the glass absorption tube. If a blue-white light is visible, the lamp is operating and the detector is at fault and should be replaced. If no light is visible, the UV lamp should be replaced. Please contact Teledyne API for obtaining replacement parts.
NOTE
In cold ambient conditions, it may require 5 to 15 minutes of
warm-up before the source lamp initially fires. Be sure to wait for
this period before troubleshooting the lamp/detector.

5.3 Changing the analog output range

5.3.1 Voltage output

Output voltage ranges are set by DIP switches on the V/F board. To change the range for the analog output:
1. Turn off instrument power near the front of the analyzer using Figure 1.9.
2. Locate switch S1, along the top edge of the card. Select the desired range per Table 4.2 below.
3. Recalibrate the ADC as described in Section 5.4.
. Remove instrument cover. Locate the V/F board
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 49
Full Scale Output Switch Settings
100 mV 1,6 = ON
1 V 1,5 = ON 5 V 1,4 = ON
10 V 1,3 = ON
TABLE 4.2 V/F BOARD DIP SWITCH - RANGES FOR ANALOG OUTPUT

5.3.2 Current Output (Optional)

To configure the M450 for 4-20mA current output, perform the following steps:
1. Set the analog voltage range to the 5 Volt scale as described in Section 5.3.1
above.
2. Install Jumper in J2 on the rear panel PCB assembly.
3. Set Jumpers JP3 and JP4 on the rear panel PCB assembly as shown in Figure
4.2 below.
FIGURE 4.2- ANALOG OUTPUT JUMPERS

5.4 A/D - D/A Calibration Procedure

Due to the stability of modern electronics, this procedure should not have to be performed more than once a year or whenever a major sub-assembly is exchanged or whenever analog output voltage range is changed .
To calibrate the ADC, do the following:
1. Press SETUP-DIAG.
2. Enter Diagnostic password and press NEXT until D/A CALIBRATION
appears in the display and press ENTR.
3. Press ADC to perform the A/D Cal.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 50
4. The M450 display will read "ADJUST ZERO:A/D=xx.x MV.” Put the probe of a voltmeter between TP3(AGND) and TP9(DAC #0) on the top of the V/F card.
5. The value displayed by the voltmeter should be close(+/- 20 mV) to the value on the M450 display. If they are not close then the V/F card has probably been configured improperly. Adjust the Zero pot(R27) on the6. V/F card until the value on the M450 display matches the v
adjusting R27, the value on the M450 display will change, the valu voltmeter will remain constant.
7.
Press ENTR.
8. The M450 disp
9. Adjust the Span pot(R31) on the V/F card until the value on the M450 display matches the value on the voltmeter to within +/-
10. Press ENTR.
11. The ADC is now calibrated and the M450 will automatically calibrate all the DAC's. This process takes only a fe
12. Press EXIT 3 times to return to the s
5.5

Current Loop Calibration

alue on the voltmeter to within +/- 2 mV. Note that when
lay will now read "ADJUST GAIN:A/D=xx.x MV.”
e on the
2 mV.
w seconds.
ample menu.
A current loop option can be ordered that will provide 0-20mA or 4-20mA analog output on the a an A/D - D/A calibra
an log output. This calibration must be repeated every time
tion is performed. To calibrate the current output, perform the following steps:
1.
Perform an A/D - D/A calibration as outlined in Section 5.4.
2.
Connect a Multimeter capable of measuring milliamperes to the analog output on the rear panel. Note: When measuring the current
multimeter or similar low-impedance current measuring device, a 400­450 ohm resistor must be placed in series with the meter to simulate a load. Failure to do this will result in erroneous readings.
3. l, press SETUP-DIAG. Press NEXT until D/A
From the front pane CALIBRATION appears and press ENTR. Press CFG and the properties for analog output Channel 0 will be displayed on the top line. The display should show something like:
DIAG D/A 0) O3_CONC_1,CURR, NOT CAL
NEXT EXIT
CAL SET
output with a
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 51
4. is setup for current output and has not been
This indicates that channel 0 calibrated. If the display shows VOLT instead of CURR then the channel must be
setup for current output. To do this, press SET, select CURR as the
output type and press ENTR.
5. Next press CAL to begin the calibration. At this point the display should
show:
DIAG D/A CAL O3_CONC_1,CURR, ZERO
UP10 EXIT
UP DOWNU100 DN10 D100 ENTR
The zero point for current output can now be set. Pressing UP, UP10 and
6.
U100 w
ill step the zero point up in increments of 1, 10 and 100 steps. The
zero point can be adjusted anywhere between 0 and 4 milliamps. Press ENTR when you have reached the desired zero point as measured by your test meter.
7.
The display will now prompt you to adjust the Gain, or full-scale output of the current loop. Using the UP and DOWN buttons as in step 4, adjust the full­scale (usually 20ma) and press ENTR. This completes the current loop calibration. Press EXIT several times until you are back at the sample menu.
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 52

6.0 MODEL 450 SPARE PARTS

002760450 ASSY, CPU, 256K PROM/SW, M450 (AMX) 003290000 THERMISTOR, BASIC 004204000 ASSY, O3 PROM W/SOFTWARE, M450 005140300 PCA, V/F CARD, W/ SCHEMATIC 005260100 ASSY, BENCH LAMP, M400/M401/M700 007040000 PCA, KEYBOARD DISPLAY 007280000 DISPLAY ASSY 009690000 AKIT, TFE FLTR ELEMENTS, 47MM, (100) 009690100 AKIT, TFE FLTR ELEMENTS, 47MM, (25) 019300000 PCA, KEYBOARD DISPLAY, CEMARK 020280000 PCA, UV DET PREAMP, M400A/M450 022920000 ABSORPTION TUBE, M450 023430000 ASSY, REF SCRUBBER, M450 023430100 OPTION, OZONE SCRUBBER, W/FLOW SWITCH 023430200 ASSY, OZONE SCRUBBER, W/O FLOW SWITCH 023480000 FILTER PAD, OZONE SCRUBBER, M450 023620000 PCA, REAR PANEL, M450 023620100 PCA, REAR PANEL, SINGLE CHANNEL, M450 023850000 ASSY, OPTICAL BENCH, M450, 1000PPM 023900000 ASSY, UV LAMP SUPPLY, M450 023930000 ASSY, PRESSURE XDUCER, M450 023940000 ASSY, SWITCHING VALVE, M450 023950000 MANUAL, INSTRUCTION, M450 024020000 OPTION, SAMPLE FILTER, M450 (47MM) 024040000 ASSY, POWER SUPPLY MODULE, M450 024140000 ASSY, SAMPLE THERMISTOR, M450 024190000 ASSY, HTR/THERM, OPTICAL BENCH, M450 024190000 ASSY, OPTICAL BENCH HEATER/THERMISTOR FL0000003 FILTER, DFU FM0000014 FLOWMETER W/NEEDLE VALVE, .2-2.5LPM FM0000015 FLOW SWITCH, ACRYLIC 560 CCM HW0000036 TFE TAPE, 1/4" (48 FT/ROLL) HW0000037 TIE, CABLE HW0000150 CLAMP, HOSE, NYLON, 1/4" OP0000012 PHOTOTUBE, ALSO SEE OP1 OR0000011 ORING, 2-109S OR0000014 ORING, 2-116V OR0000048 ORING, 2-112S OR0000061 ORING, 2-113S OR0000069 ORING, 2-129V PS0000011 SWITCHING PS, 40W, +5V, +/-15V PS0000012 SWITCHING PS, 20W, 12V PU0000020 PUMP, KNF, TEFLON PU0000022 DIAPHRAM KIT, KNF ,TEFLON, PU20
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 53
SW0000026 PRESSURE XDUCER, 0-15 PSIA TU0000009 TUBING, 1/8 X 1/4 TYGON TU0000018 FEP TEFLON TUBING 1/4 OD 5/32 ID VA0000034 VALVE, 2-WAY, SS, MANIFOLD MOUNT VA0000035 VALVE, 3-WAY, SS, 12VDC, VITON, M450
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 54

APPENDIX A - ELECTRICAL SCHEMATIC INDEX

Drawing Number Title
00514 V/F - I/O Card Assembly 00515 V/F - I/O Card Schematic 00704 Keyboard Assembly 00705 Keyboard Schematic 0147802 Motherboard Assembly 01479 Motherboard Schematic 01561 I2C/SubMux Assembly 01562 I2C/SubMux Schematic 01930 CE Keyboard Assembly 01931 CE Keyboard Schematic 02028 UV Detector Preamp Assembly 02029 UV Detector Preamp Schematic 02322 UV Lamp Driver Assembly 02323 UV Lamp Driver Schematic 0236200 M450 Rear Panel Assembly 02363 M450 Rear Panel Schematic 02421 Pressure Transducer Assembly 02422 Pressure Transducer PCB Schematic
P/N 02395D1 Teledyne API Model 450 O3 Monitor Instruction Manual - Page 55
Loading...