The print history shown below lists the printing dates of all revisions and addenda created for this manual.
The revision level letter increases alphabetically as the manual undergoes subsequent updates. Addenda,
which are released between revisions, contain important change information that the user should incorporate
immediately into the manual. Addenda are numbered sequentially. When a new revision is created, all
addenda associated with the previous revision of the manual are incorporated into the new revision of the
manual. Each new revision includes a revised copy of this print history page.
Revision D (Document Number 164-082005) ................................................................ August 2005
Revision E (Document Number 164-102005) ............................................................... October 2005
Revision F (Document Number 164-012006) ................................................................ January 2006
Revision G (Document Number 164-092006) ........................................................... September 2006
Revision H (Document Number 164-112006) ........................................................... November 2006
Revision J (Document Number 164-022009) ............................................................... February 2009
Revision K (Document Number 164-062009) .................................................................... June 2009
Revision L (Document Number 164-082010) ................................................................. August 2010
Revision M (Document Number 164-102012) ............................................................. October 2012
Revision N (Document Number 164-122012) ........................................................... December 2012
Visit www.teledyne-hi.com for WEEE disposal guidance.
Hastings Instruments reserves the right to change or modify the design of its equipment
without any obligation to provide notification of change or intent to change.
3.0 FEATURES .......................................................................................................................................................................... 7
5.0 FRONT PANEL ................................................................................................................................................................. 10
7.1. POWER ........................................................................................................................................................ 12
10.2.4. SELECTING GAS ID ...................................................................................................................................... 20
10.2.6. FRONT PANEL LOCK OUT ............................................................................................................................ 21
10.3.1. COUNT UP TO A SET POINT ......................................................................................................................... 21
10.3.2. COUNT DOWN FROM A SET POINT .............................................................................................................. 22
10.5.1. ZERO TO FIVE VOLT OPERATION ................................................................................................................. 23
10.5.2. ZERO TO TEN VOLT OPERATION ................................................................................................................. 23
164-122012 Power
POD
400 Page 3 of 42
Page 4
10.5.3. FOUR TO TWENTY MILLIAMP OPERATION .................................................................................................. 23
ENABLING RATIO CONTROL ..................................................................................................................................... 25
10.8.1. RESETTING THE ZERO ONLY ....................................................................................................................... 26
10.8.2. RESETTING THE SPAN ONLY ........................................................................................................................ 26
10.8.3. RESETTING ZERO AND SPAN ........................................................................................................................ 27
10.8.4. SETTING A MULTIPLICATION FACTOR ........................................................................................................ 27
10.9.2. SET POINT QUERIES ..................................................................................................................................... 28
11.0 APPENDIX A ..................................................................................................................................................................... 30
12.0 APPENDIX B ..................................................................................................................................................................... 33
13.0 APPENDIX C ..................................................................................................................................................................... 33
14.0 APPENDIX D ..................................................................................................................................................................... 36
-400 comes calibrated from the factory according to your specifications.
No set up is necessary unless you need to change the specs
1ea Power
1ea AC power cord
1ea 15-pin, Hi Density, D-style connector
1ea 9-pin, D-style connector
1ea user’s manual
When unpacking the Power
contain everything you need.
Pod
-400
Pod
, the package will
J-1 Ch1
J-2 Ch2
J-3 Ch3
J-4 Ch4
J-5
Analog
out
J-6 RS-232 Serial Com Port
J-7 RS-485
J-8 Alarms Open-collector Alarms
J-9 RS-485
Transducer Power,
Command and Analog
Signals.
Transducer Signal, Volts
Only
Multi-drop com port
(Daisy Chained)
Multi-drop com port
(Daisy Chained)
Pin-outs (see section 7)
Connect a “known good” cable to the desired channel
(See next frame for pin-outs)
1) Check power setting switch on the back panel and make
sure it is set appropriately.
2) Ensure the power switch is off, and then connect the power
cord.
OPEN
1
CLOSE
OPEN
2
CLOSE
OPEN
3
CLOSE
OPEN
4
CLOSE
P
ower
1) Press the desired Channel # Key. Press the Auto key.
Turn power switch to on position. If you have just
turned on the mass flow instrument and the
power, some settling is necessary.
164-122012 Power
POD
400 Page 5 of 42
2) Press the desired Channel # Key. Enter desired flow.
Press the enter key.
Pod
(For Controllers)
HASTINGS
3
2
1
5
6
4
8
7
9
.
0
Page 6
2.0 Safety
Read this manual in its entirety before operating the POWER
POWER
and meters. Read all wiring and power hookup instructions and understand the requirements prior to
using another manufacturer’s products with the POWER
interfaced with the POWER
standards before operating.
POD
-400 is designed to operate with most Teledyne Hastings Instruments (THI) flow controllers
POD
-400 is wired according to prevailing local safety and operational
The following symbols and terms may be found on THI products and/or in THI manuals and indicate
important information.
When found on the device, this symbol indicates that the operator should refer to
the manual for important instructions on the proper use of this device. When found
in the manual, this symbol indicates that the reader should understand the
implications contained in the text before operating the device.
POD
-400 Power Supply/Totalizer. The
POD
-400. Insure that any product being
This symbol indicates that a shock hazard may be present. Read the instruction
manual carefully and insure that the device is wired properly and that all settings
have been checked prior to applying power to the device.
The WARNING label indicates important information that should be heeded for safe and proper
performance of the device.
The label, CAUTION, is used to indicate that damage to the power supply or equipment
connected to it could occur if directions are not followed. Warranty could be invalidated if the
instructions in this manual are not followed.
164-122012 Power
POD
400 Page 6 of 42
Page 7
3.0 Features
The POWER
rack hardware or can be used as a bench top unit. The POWER
character, vacuum fluorescent display (VFD). The display emulates a liquid crystal display in its
command structure but the VFD gives the unit a greater viewing angle and better visibility than available
with most conventional LED or LCD displays. The display can be set to four different brightness levels.
Use a lower brightness setting to extend the already long expected life time of the display. Use brighter
settings for viewing areas where ambient light may be too bright or cause glare, or where greater viewing
distances are required.
Most features are accessible via the membrane keys on the front panel. Consult the section on each
function to check its availability. Operators are guided through the many features and options by selecting
their choices from an intuitive menu structure.
FRONT PANEL LOCKOUT
ANALOG RANGE SELECTION
Analog signal and control ranges are operator selectable. The operator can choose between three different
DC ranges:
POD
-400 serves as a convenient control center that can be rack-mounted using standard half-
The Front Panel Lockout function is only available through serial communication. Manual
Overrides remain available during Lockout via a minimum number of keystrokes using
dedicated keys for this purpose and allow any command setting to be overridden in either the
high (open) or low (closed) state.
POD
-400 is equipped with a 4 X 20
A fifteen (15) pin, high density, sub-miniature, D-type connector is provided for separate monitoring of
each channel’s analog transducer signal.
CAUTION: Consult the appropriate section for limits to the loading of these signals.
SERIAL COMMUNICATION
The POWERPOD-400 comes equipped with standard RS-232 and RS-485, serial communication. Most
functions, features, signals and alarms are accessible and modifiable via any remote computer.
0 -5 VDC, 0-10 VDC or 4 -20 mA.
The following commands are manual commands only:
OPEN,
The status of these settings can be read via serial communication but they cannot be changed
except manually, from the front panel.
AUTO and CLOSE.
164-122012 Power
POD
400 Page 7 of 42
Page 8
POWER SELECTION
Power input is switchable between 100 VAC, 115 VAC and 230VAC (50 or 60 Hz) via
the rear panel. For the safety of the operator as well as the device, the correct power
level should be selected prior to connecting to the power mains. See the table
“POWERPOD-400 Specifications” in section 3.0 for the proper fusing when changing
power settings.
RATIO CONTROL
Ratio control is possible between channels using a familiar master/slave configuration. Channel one
(1) must be enabled as the master channel. Any combination of the remaining channels is possible
for slave channel assignment.
TOTALIZER
A Totalizer function is present for each channel with the capability of counting down from a set
point, counting up to a set point or continuous count up. The maximum count is ±999999 units.
When the set points are reached, a memory flag for each set point is set to a digital ‘1’ indicating a
Boolean ‘true’ value. The Totalizer set point flags must be polled via digital communication to be
read. Each flow channel has one low-limit and one high-limit set point available. These alarms are
available via open-collector, opto-isolated outputs on the rear panel as well as serial
communication.
POWER OUTAGES AND THE OVERRIDE CONDITION
In the event of a power outage, even one of short duration, the POWER
POD
-400 is
designed to conduct a software reset. During the period of time in which the reset is
occurring, it will not accept or respond to any commands either manually or digitally
until the reset process is completed. After said reset, the POWER
POD
-400 will have
remembered all previously entered set-points but all channels are designed to come up
in the “Override-CLOSEd” condition. For meters, this should have no affect on their
behavior. All analog-only, THI, flow controllers with normally-closed (NC) valves will
close and remain closed until operator intervention manually returns selected channels’
Override condition back to the AUTO mode.
For all other controllers, this means that the control signal (pins 5 to 14) of J1 through
J4 will return to the previously set level. For controllers not taking advantage of the
Override function (pin 8), these controllers will return to their normal, preset operating
condition unless other intermediary steps are undertaken. Controllers that are making
use of the override function (Pin 8) and have normally-open valves will be driven to the
fully open condition.
164-122012 Power
POD
400 Page 8 of 42
Page 9
y
y
4.0 Specifications
Power
Specification Value Units Notes
Power Inputs
V
P 68 VA f 50 -60 Hz
Transducer
Number Channels 4
V
I
I/O
Display
Type
# Lines 4
# Characters 20
Brightness Levels 4
A/D Converter
Filtering Rate 4, 15, 30,
Alarms
3 per channel 1 High
1 Low
1 Total
Dimensions
Front Panel (h x w) 3.5 x 9.5 in
Case (h x w x d) 3 x 8 x 9.5 in
Hole Centers (h x
w)
Weight 5 1/4 lbs
Suppl
Suppl
Pod
-400 Specifications Table
100
115
VAC
230
±15
±250
0 - 5
0 - 10
VDC Bi-polar, per Channel
mA Bi-polar, per Channel
VDC
4 - 20 mADC
100
Hz
3 x 8.825 in
0.7 A, 250VAC, SB Fuse
0.6 A, 250VAC, SB Fuse
0.315 A, 250VAC, SB Fuse
Vacuum Fluorescent,
LCD Emulator
164-122012 Power
POD
400 Page 9 of 42
Page 10
(
A
A
(
(
p
A
e
5.0 Front Panel
1. CHANNEL NUMBER SELECT KEYS
Selects channel for editing. An asterisk (*) appears in the first column of the display to indicate that this is the channel to
be edited.
2. OVERRIDE INDICATORS
Indicates when a channel’s command signal is overridden high
3. OVERRIDE KEYS
Override the command signal on the
sets command to –15V. AUTO allows the user to set the command signal for normal operation. A channel must be
active before these keys can become operational.
4. DISPLAY AREA
Column 1: Reserved for displaying
Column 2: Reserved for Col’s 3 – 8: Signal monitor. Displays current input signal while in METER mode,
Column 9: SpacCol’s 10 – 14: UNITS OF MEASURE Column 15: Space
Col’s 16 – 20: GAS
5. KEYPAD
Use to enter
OPEN
1
CLOSE
OPEN
2
CLOSE
OPEN
3
CLOSE
OPEN
4
CLOSE
Pod
P
ower
HASTINGS
400
OPEN) or low (CLOSED).
CTIVE CHANNEL. OPEN sets control override (pin 8) to +15V. CLOSED
CTIVE CHANNEL
TOTAL (T).
olarity indicator.
to average readings or TOTAL while in TOTALIZER mode.
display.
ID.
SET POINTS or to modify the SETUP or CALIBRATION of control unit.
*), MASTER channel (M), SLAVE
3
2
1
5
4
7
0
6
8
9
.
VERAGE while set
S) or
164-122012 Power
POD
400 Page 10 of 42
Page 11
6.0 Rear Panel
1. POWER ON/OFF SWITCH
2. POWER INLET & FUSE
3. POWER SELECTOR SWITCH
4. FUSE Vs. POWER SETTING TABLE
5. RS-232 SERIAL PORT CONNECTOR (J6)
6. RS-485, DAISY CHAINED, SERIAL PORT CONNECTORS (J7, J9)
7. TRANSDUCER CONNECTORS (J1-J4)
8. ANALOG OUTPUT (J5)
9. ALARMS (J8)
164-122012 Power
POD
400 Page 11 of 42
Page 12
7.0 Wiring
7.1. POWER
Power is supplied through a fused, AC jack on the rear panel (item 2).Use the power cord
supplied with the unit (PN15-17-011 for 115 VAC, 60Hz). See the following table for
selecting the proper fuse rating. Use a metric, 5 x 20 mm sized, time-delayed fuse.
Cords without plugs are supplied with units shipped outside of the U.S. Consult and comply with any
local laws and/or codes when connecting to any AC main. The AC input is user selectable between 100,
115 or 230 VAC, 50 or 60 Hz, via an AC selector switch next to the AC jack (Item 3).
WARNING: Be sure to set the power select switch prior to connecting to mains. Re-fuse the
connector according to the table above.
7.2. COMMUNICATIONS
Connectors J6 (Item 4), J7 (Item 6) and J9 (Item 6) are for RS-232, RS-485 connections respectively.
Settings for serial communication are accessible via the front panel.
Connectors J1, 2, 3 and 4 (Item 7) are 15 pin D style connectors wired in the standard Hastings
Instruments pin-out (H pin-out).
J1, J2, J3, J4 1 NC
Transducer
Connectors
(DB-15)
(H-Pinout)
2 NC Valve Cntrl
3 NC mA Sig
4 NC mA Sig
5 Sig. Com.
6 Sig. In
7 Case Gnd.
8 Cntrl Over-ride 1.5mA
9 -15 VDC
10 NC
11 +15 VDC
12 Valve Return
13 NC Ext-In
14 Set Point Out
15 +5 VDC Ref. Not Used.
Voltage
7.4.ALARMS
Connector J8 (item 9) provides the user with open-collector, opto-isolated alarms for individual channels.
Each channel is provided with one user settable “High” and one “Low” alarm.
Analog signals from each channel’s transducers are available for reading or for sending to another power
supply for ratio (Master/Slave) operation. The signal can be sent to a channel on another power supply
and programmed as a Master for that power supply, allowing the remaining three channels to operate as
slaves. When operating 4 – 20 mA instruments, the followers supply a 0.5 -2.5 volt signal.
J5 1 Channel 1 Signal
2 Channel 1 Return
3 Channel 2 Signal
4 Channel 2 Return
5
6
Analog
Out
(HD DB-15)
7
8
9
10
11 Channel 3 Signal
12 Channel 3 Return
13 Channel 4 Signal
14 Channel 4 Return
15
164-122012 Power
POD
400 Page 14 of 42
Page 15
8.0 Manual Operation
8.1. POWER ON/OFF
The Power On switch, item 1 in rear panel drawing, is located in the upper right corner of the
rear panel. Insure that the proper power setting is selected prior to turning the power on. See
the Power section of WIRING THE POWER
8.2. CHANGING THE COMMAND SET POINT
Channel two (2), example shown.
Press the desired CHANNEL # key. An asterisk appears in the first space on the line representing the
selected channel. The meter display is immediately replaced with the current set point. Pressing a
number key begins the editing process. The cursor lands on the channel set point to be edited, and the
new command is entered with the most significant bit (MSB) first. The set point is filled in from left to
right on the display. When entering a new set point, you must use the decimal key when it has been used in setting the SPAN. Pressing ENTER completes the editing process. The old command is not
changed until the ENTER key is actuated. Hitting the ESCape button at any time prior to the ENTER
key will return the display to its previous state without any changes being made. This command will not
be applied to the output until the channel is set for AUTO operation. See Setting a Channel to AUTO Control.
POD
-400, above.
8.3. OVERRIDE OPEN
Press the desired CHANNEL # key. An asterisk appears in the first space on the line representing the
selected channel. The meter display is immediately replaced with the current set point. Pressing the
OPEN key results in the application of +15 VDC to pin number eight (8) of the corresponding 15 pin
Sub-D connector and returns the previously programmed metering function to the display.
8.4. OVERRIDE CLOSED
Press the desired CHANNEL # key. An asterisk appears in the first space on the line representing the
selected channel. The meter display is immediately replaced with the current set point. Pressing the
CLOSE key results in the application of -15 VDC to pin number eight (8) of the corresponding 15 pin
Sub-D connector and returns the previously programmed metering function to the display.
164-122012 Power
POD
400 Page 15 of 42
Page 16
8.5. SETTING A CHANNEL TO AUTO CONTROL
Press the desired CHANNEL # key. An asterisk appears in the first space on the line representing the
selected channel. The meter display is immediately replaced with the current set point. Pressing the
AUTO key causes pin number eight (8) of the 15 pin Sub-D connector to float and returns the previously
programmed metering function to the display. Pin 14 signal levels are now available for control.
8.6. SETTING A CHANNEL TO DIRECTLY METER INCOMING SIGNALS
Press the MODE key. The MODE menu appears in the display. To select METER, press 1, and then
ENTER. The METER menu allows the selection of the desired channel. Press the number key that
corresponds with the desired channel followed by ENTER. The previously programmed display returns
with the selected channel reading the signal between pins 5 and 6 of the corresponding channel’s 15 pin
D-connector (J1 – J4).
8.7. SETTING A CHANNEL TO DISPLAY THE TOTALIZER FUNCTION
The POWERPOD-400 provides an integrated (Riemann Sum) value of the incoming signal for each
channel.
=
)**(
fCfT∑=, where:
IS
Each channel samples the incoming signal at a rate
corresponding to the preset A/D conversion rate in Hz. At
each sampling interval, the TOTALIZER function multiplies
the average signal, as a fraction of the full-scale value (5v, 10v
or 20 mA), times the span, or cal, value. This results in a rate
for that interval. Next, a time element factor is determined
according to the flow rate programmed for that channel as
follows.
For a time element of seconds, the factor is 1/10.
For minutes, the factor is 1/600.
Hours use a factor of 1/36000.
Each calculation is summed and stored for a TOTAL value.
TotalT
⎧
⎪
=
⎪
S
⎨
=
⎪
⎪
=
I
⎩
FactorSignalFractionalf
ValueSpanorvalueCALC
IntervalTimeFractionalf
⎫
⎪
⎪
⎬
⎪
⎪
⎭
164-122012 Power
POD
400 Page 16 of 42
Page 17
Example (See Graph):
A linear flow transducer with a DC output of 0 to 5 volts is calibrated for a maximum flow of 25 SLH.
The transducer is connected to one channel of a POWER
volts input and has been programmed to display rate in SLH. The transducer is sending a constant 1.000volt signal to the POWERPOD-400. What is the total flow after 40 minutes?
During one sampling interval, the TOTALIZER reads the one-volt signal and calculates the fraction of the
possible full-scale value.
1 volt/5 volts or 0.20
The fractional signal value is multiplied times the CAL value
1/5 * 25 = 5
Then, because the channel’s programmed rate is in SLH (standard liters per hour), a factor of 1/3600 is
applied to the result.
5 * 1/36000 = 1.389 x 10^-4 standard liters per sampling interval.
A constant signal received over a 40 minute time period would amount to 40 minutes * 60 seconds * 10
samples per second = 24000 samples. The TOTAL would be
2.400 E03 * 1.389 E-04 = 3 1/3 Std Liters.
If no rate is programmed on that channel, The TOTALIZER function is disabled for that channel and the
display returns to its regular metering configuration.
POD
-400 which is CAL’d to read 25.000 at 5
To enable the TOTALIZER, press the MODE key. The MODE menu appears in the display. To select
TOTAL, press 2, and then ENTER. The TOTALIZER menu allows the operator to select the desired channel. Press the number key that corresponds with the desired channel followed by ENTER. A
TOTALIZER menu is presented that allows the operator to go directly to displaying whatever the
Totalizer has currently stored as the total or to reset the counter. If DISPLAY is selected by pressing the
number 1 key followed by ENTER, the display will return to normal with the most recently stored total
being displayed on the appropriate line. If RESET is chosen by pressing the number 2 key followed by
ENTER, the display returns with the total value reset according to the pre-selected counting mode
presented in the table below with the Totalizer counting.
The POWER
The following text describes how to choose and activate the desired method. Further information and
instructions on how conditions may be changed can be found in section 10, SETUP/CAL mode.
POD
-400 allows the user to select different methods of serial communication and control.
9.2. CHOOSING RS-232 COMMUNICATION
Press the MODE button. The MODE menu is displayed. Press the number “4” key for XTRNL
CNTRL. The XTRNL CNTRL menu is displayed. Choose RS-232 by pressing the number “1” key.
The normal display will return to its previously programmed state.
9.3. CHOOSING RS-485 COMMUNICATION
Press the MODE button. The MODE menu is displayed. Press the number “4” key for XTRNL
CNTRL. The XTRNL CNTRL menu is displayed. Choose RS-485 by pressing the number “2” key.
The normal display will return to its previously programmed state.
164-122012 Power
POD
400 Page 18 of 42
Page 19
10.0 Setup/Cal Mode
10.1. ENTERING THE SETUP/CAL MODE
Pressing the MODE key causes the MODE menu to be displayed. Choose the SETUP/CAL mode by
pressing the number “3” key then ENTER. A CHANNEL SELECT menu is presented on the display.
It should be noted that, although the user must select a specific channel number, some choices in the
following menus may affect the operation of all channels and/or the display (ex; Display Brightness, type of
communication). See the following menu options for more details.
10.2. DISPLAY SETUP
After entering the SETUP/CAL mode and selecting a channel number as described in the previous
section, several selections are available that will affect the display. Follow the steps below to gain the
desired customized display.
Note: Front panel lockout and display line (channel) blanking are only available
through serial communication.
10.2.1. Blanking a Display Line/Channel
Use the serial command “DndCr”, where
n = channel number,
d = 1 for Totalizer Mode,
2 for Meter Mode,
3 for a blanked line and
Cr = carriage return.
10.2.2. Selecting Display Brightness
While the SETUP/CAL menu is displayed, choose DISPLAY by pressing the keys “1” + ENTER.
Choose BRIGHTNESS by pressing the number “1” key + ENTER. The BRIGHTNESS menu will be
visible. One of four different brightness levels can be chosen by pressing a number key. The number “1”
key selects the dimmest setting while the number “4” key selects the brightest. Any change in brightness
will affect the entire display after pressing the ENTER key.
164-122012 Power
POD
400 Page 19 of 42
Page 20
10.2.3. Selecting Unit-of-Measure
After entering the SETUP/CAL mode and after selecting a channel number, the SETUP/CAL menu
is displayed. Choose DISPLAY (1) + ENTER and then select UNITS by pressing the number “2” key + ENTER. The UNITS menu will be displayed allowing the selection of over 50 different units of measure
(UOM) simply by pressing the number key corresponding to the desired UOM. The number can be
found by scrolling up or down until the desired UOM is displayed or by consulting Appendix B, at the end
of this manual. Once selected, the UOM will be displayed in character spaces 10 to 14 on the appropriate
line. Selecting UOM = “0” results in the UOM characters being blanked.
10.2.4. Selecting Gas ID
While the SETUP/CAL menu is displayed, Select the DISPLAY menu by pressing 1 + ENTER. Then
choose GASID by pressing 3 + ENTER. The GASID menu will be displayed allowing the selection of
nearly one hundred different chemical symbols simply by pressing the number key corresponding to the
desired gas. The number can be found by scrolling up or down until the desired gas is displayed or by
consulting Appendix C at the end of this manual. Once selected, the formula will be displayed in
character spaces 16 through 20 on the appropriate line.
Note: Formulae too long to be completely displayed will display the Hastings
Instruments Gas ID number. Choosing “0” will blank the GASID elements on the
display.
10.2.5. Setting A/D Conversion Rate (Filter)
Press the MODE button. The MODE menu is displayed. Press the number “3” key followed by
ENTER. The CHANNEL SELECT menu is displayed. Choose the channel whose signal is to be filtered using the number keys. For example, “1” + ENTER. The SETUP/CAL menu is displayed.
164-122012 Power
POD
400 Page 20 of 42
Page 21
After SETUP/CAL is selected and a channel number has been chosen, the SETUP/CAL menu is
displayed. choose DISPLAY by pressing the number “1” key + ENTER. The FILTER option can
be reached by scrolling down once to view the selection number and then pressing the “4” key +
ENTER. A conversion rate of 4, 15, 30 or 100 Hertz may be selected by entering its menu number
+ ENTER. Enter the number of the desired A/D conversion rate and press ENTER. The normal
display will return. Changes may be noticed in the flickering of the LSB on the display. In most
situations, the flickering digits are an insignificant fraction of the display range and can be ignored.
Should the flickering become annoying or make reading the display difficult, choose a slower rate
of conversion.
NOTE: Slower rates of conversion may affect the accuracy of the TOTALIZER in
situations where flow is changing often. Steady flow rates will see little loss of
TOTALIZER accuracy.
10.2.6. Front Panel Lock Out
Front Panel Lockout is available only through serial communication. See the SERIAL
COMMUNICATIONS section of this manual and Appendix A.
CAUTION: While the front panel is locked out, all command functions, excluding
the override functions, are disabled. Commands can be overridden OPEN or
overridden CLOSED during front panel lockout. This was determined to be useful
both as a trouble shooting and as a safety tool. All other functions will be available
only through the use of serial communication.
10.3. TOTALIZER SETUP
The TOTALIZER can be set to count up to a set point, to count down from a set point or to count up
continuously to the maximum ability of the display (999999). In order for the TOTALIZER to work, a
unit of measure must be chosen that includes a time element. For example, standard liters per hour
(SLH) will totalize; percent (%) will not. Each mode is described in the following text and can be followed
in the diagram below.
10.3.1. Count Up to a Set Point
The TOTALIZER will be set to count from zero (0) to a set point. When the set point is reached, a
memory flag will be set to its logic high state. This flag is readable only through serial communications. The TOTALIZER will continue to count up to the maximum ability of the display.
While the SETUP/CAL menu is displayed, choose TOTAL by pressing the number “2” key followed by
ENTER. A TOTALIZER menu is displayed. Press the number “1” key to choose the CNT UP option.
The COUNT UP/DN window is displayed for the selected channel. Enter the desired set point using the
numbered key pads and press ENTER. The unit returns the previously programmed display with the
selected channel now displaying the Riemann Sum of the received transducer signal.
164-122012 Power
POD
400 Page 21 of 42
Page 22
10.3.2. Count Down from a Set Point
The TOTALIZER will be set to count from a preset value to zero (0). When zero is reached, a
memory flag will be set to its logic high state. This flag is readable only through serial
communications. The TOTALIZER will continue to count in the negative direction from zero to the
maximum ability of the display (-999999).
While the SETUP/CAL menu is displayed, choose TOTAL by pressing the number “2” key + ENTER.
A TOTALIZER menu is displayed. Press the number “2” key to choose the CNT DN option. The
COUNT UP/DN window is displayed. Enter the desired set point using the numbered key pads and
press ENTER. The unit returns the previously programmed display with the selected channel now
displaying the set point minus the Riemann Sum of the received transducer signal.
10.3.3. Count Continuously
The TOTALIZER will be set to count from zero (0) to the maximum ability of the display. No flags or
alarms will be set.
While the SETUP/CAL menu is displayed, choose TOTAL by pressing the number “2” key and
ENTER. A TOTALIZER menu is displayed. Press the number “3” key to choose the CONT option
followed by the ENTER key. The previously programmed display returns with the selected channel now
displaying the Riemann Sum of the received transducer signal.
10.4. EXTERNAL COMMUNICATION SETUP
The following text describes how the POWER
communication. Instruction structure and syntax is covered in the section, Serial Protocol.
POD
-400 can be setup to respond to serial and network
10.4.1. RS-232 Settings
From
SETUP/CAL
Menu
After selecting SETUP/CAL, select any channel number and hit ENTER, choose option number 3 via
the numbered key pad. The first menu to be displayed is the BAUD menu. To select the desired baud
rate press the number key 1 or 2 respectively. Eight bit bytes, no parity, one stop bit and no handshaking
are automatically set. The preprogrammed display is returned upon completion.
10.4.2. RS-485 Settings
From
SETUP/CAL
Menu
After selecting SETUP/CAL, select any channel number and hit ENTER, choose option number 3 via
the numbered key pad. The BAUD menu will be displayed. To select the desired baud rate press the
corresponding number key.
164-122012 Power
POD
400 Page 22 of 42
Page 23
The default address as programmed at the factory is 01. If a different address is required, it may be
changed using either RS-232 or 485. After making the proper serial connections, use the command *00X
to read the default address. Use *00Xdd, where dd = the new address. If the old address is known, use
DDXdd, where DD is the old address and dd is the new address. See the section on "Serial
Communication: Commands" for more on command structure.
All commands listed in Appendix B are available for RS-485 use by inserting an asterisk and the unit's
address in front of the command.
Ex) Set the units of measure on channel two (2) to %.
*10UM203Cr
The command, *00XCr can be used to poll the unit for its current address setting using either RS-232 or
RS-485.
10.5. SELECTING THE ANALOG SIGNAL LEVEL
The POWER
selectable. The following diagram applies no matter which level of operation is chosen.
From
SETUP/CAL
Menu
POD
-400 is capable of 0 to 5 V, 0 to 10 V or 4 to 20 mA operation. Each level is user
10.5.1. Zero to Five Volt Operation
While in the SETUP/CAL menu, select ANALOG by pressing the number 4 key and the ENTER key.
From the ANALOG SIGNAL menu, choose 0V – 5V by pressing the number 1 key and pressing the
ENTER key. The normal display returns with 0 volts intended to correspond to 0% signal and 5 volts
intended to correspond to the 100% signal. Calibration may be required.
10.5.2. Zero to Ten Volt Operation
While in the SETUP/CAL menu, select ANALOG by pressing the number 4 key and the ENTER key.
From the ANALOG SIGNAL menu, choose 0V – 10V by pressing the number 2 key and pressing the
ENTER key. The normal display returns with 0 volts intended to correspond to 0% signal and 10 volts
intended to correspond to the 100% signal. Calibration may be required.
10.5.3. Four to Twenty Milliamp Operation
While in the SETUP/CAL menu, select ANLG by pressing the number 4 key and the ENTER key.
From the ANALOG SIGNAL menu, choose 4mA – 20 mA operation by pressing the number 3
key and pressing ENTER. The normal display returns with 4 milliamps intended to correspond to
0% signal and 20 milliamps intended to correspond to the 100% signal. Calibration may be
required.
164-122012 Power
POD
400 Page 23 of 42
Page 24
10.6. SETTING LIMIT ALARMS
POD
Each channel of the POWER
400 has one high limit alarm and
one low limit alarm. Each alarm is
provided by an open-collector,
opto-isolated signal at a
corresponding pin on the rear
panel, connector J8. Use the
provided diagram when referring
to the text below.
From
SETUP/CAL
Menu
-
10.6.1. Setting a Single Channel’s High Limit
After selecting a channel in the SETUP/CAL menu, choose ALARMS by pressing the number 5 key
followed by the ENTER key. The ALARMS menu will be displayed. Pressing the number “1” key
selects the HIGH LIMIT display. Using the number keys, enter the desired high limit level and then
press ENTER. EXIT the loop by pressing the number “4” + ENTER key while the ALARMS menu is
displayed.
10.6.2. Setting a Single Channel’s Low Limit
After selecting a channel in the SETUP/CAL menu, choose ALARMS by pressing the number 5 key.
The ALARMS menu will be displayed. Pressing the number 2 key selects the LOW LIMIT display.
Using the number keys, enter the desired high limit level and then press ENTER. EXIT the loop by
pressing the number 4 key while the ALARMS menu is displayed, followed by ENTER.
10.6.3. Setting a Single Channel’s Hysteresis
After selecting a channel in the SETUP/CAL menu, choose ALARMS by pressing the number 5 key. The
ALARMS menu will be displayed. Pressing the number 3 key selects the HYSTERESIS display. Using
the number keys, enter the desired degree of Hysteresis followed by the ENTER key. EXIT the loop by
pressing the number 4 key while the ALARMS menu is displayed, followed by ENTER.
164-122012 Power
POD
400 Page 24 of 42
Page 25
10.7. SETTING RATIO CONTROL PARAMETERS
Enabling Ratio Control
Ratio control is achieved through the SETUP/CAL menu. Ratio Control is activated by selecting channel
number one (1) and enabling it as master. After enabling channel one, the operator may assign the
channels that will follow its signal by some factor.
Press the MODE key and then 3 + ENTER. Select channel one by pressing 1 + ENTER. Choose
RATIO by pressing the 6 key + ENTER. Channel one (1) is enabled as the master when you press 2 +
ENTER.
After enabling RATIO control, a series of menus labeled SLAVE A, SLAVE B and SLAVE C will be
displayed. These menus correspond to channels 2, 3 and 4 respectively. Each of the channels, 2 through
4, can be enabled or disabled individually. For example, pressing the 1 key + ENTER while SLAVE A is
being displayed will eliminate channel two (2) from ratio control and allow channel two to work normally
as an independent channel. Pressing the number 2 key + ENTER will make channel 2 the first SLAVE
and an S will appear in the first column of line two on the display. Note that, if EXIT is selected at any
time, the previous settings remain unchanged and the display is returned to its previously programmed
configuration.
164-122012 Power
POD
400 Page 25 of 42
Page 26
10.8. CALIBRATING A CHANNEL TO ITS INCOMING SIGNALS
The POWER
POD
-400 is designed to accept the input from almost any transducer that operates in the
ranges of zero to five volts, zero to ten volts or four to twenty milliamps. In most cases, transducers are
capable of having their minimum signal and maximum signal adjusted to correspond with the minimum
(ZERO) and maximum (SPAN) unit-of-measure that they are designed to reflect.
For example:
A transducer designed to output 0 volts while reading a temperature of 0 degrees centigrade and output 5
POD
volts while reading 100 degrees centigrade is connected to channel one of the POWER
-400.
In actuality, the transducer is sending -0.23 volts at 0 degrees and 5.11 volts at 100 degrees. By zeroing
POD
channel one of the POWER
-400 while reading the low signal from the transducer, the display will be
000.0 when the temperature is 0°C. Like wise, by setting the CAL value to 100.0 while reading the 5.11
volt signal causes the POWER
POD
-400 to display 100.0 when the temperature is actually 100°C.
This procedure will not account for any non-linearity in any transducer’s output. That is, the
accuracy/calibration of any transducer must be independently set, checked and verified before
connecting to the POWER
POD
-400, or, the transducer/ POWER
POD
-400 combination must be
calibrated together as a system, if the accuracy of the whole system is to be known.
10.8.1. Resetting the Zero Only
While the SETUP/CAL menu is being displayed and after making a selection from the CHANNEL
SELECT menu, select CALIBRATE by pressingthe number 7 keyfollowed by ENTER, then choose
ZERO ONLY by pressing the number 1 key and ENTER. The ZERO menu will be displayed which also
displays the selected channel number and the signal being received. If this signal is an acceptable signal
for zero, press the ENTER key to accept it. If the indicated signal is not an acceptable level for reading
zero, the operator has an opportunity at this stage to correct any problems while reading the incoming
signal before pressing the ENTER key and accepting it. Alternatively, the operator can press ESC to exit
the SETUP/CAL mode and return to it later. After accepting the displayed signal level as the zero point,
the display returns to its previously programmed settings. See Appendix D.
10.8.2. Resetting the Span Only
After selecting a channel in the CHANNEL SELECT portion of the SETUP/CAL menu, press 7 +
ENTER to select CALIBRATE. SPAN ONLY is selected by pressing number 3 + ENTER. The
SPAN window is displayed which contains the chosen channel and the incoming signal on that channel.
If the displayed signal is an acceptable full range value, press ENTER to accept it. The previously
programmed display returns with the selected channel now displaying the CAL’d value. If the incoming
164-122012 Power
POD
400 Page 26 of 42
Page 27
signal is not an acceptable full range value, the operator has an opportunity to correct any problems at this
stage prior to pressing ENTER. Alternatively, the operator can press ESC to exit SETUP/CAL and
return later. See Appendix D.
10.8.3. Resetting Zero and Span
Should it be required to zero and set a new span value for a given channel, The menu option is provided
that will present the programmer with the zero menu first, followed by the span, or cal, menu. After
selecting a channel in the CHANNEL SELECT portion of the SETUP/CAL menu, press the number 7
key to select CALIBRATE. ZERO & SPAN is selected by pressing the number 2 key. The ZERO
window is displayed first. It contains the chosen channel and the incoming signal on that channel. If the
signal displayed is within an acceptable range for zero, Press the ENTER key to accept this signal as
representing a display of zero while monitoring the transducer. If this signal is not appropriate, the user
may correct the problem while the window is displayed or press the ESC key and come back at a later
time to zero and span.
After setting the zero of the power supply, the SPAN menu window is displayed. If the displayed signal is
an acceptable full range value, press ENTER to accept it. The previously programmed display returns
with the selected channel now displaying the CAL’d value. If the incoming signal is not an acceptable full
range value, the operator try to correct any problems at this stage prior to pressing ENTER or, by pressing
the ESC key, they can exit SETUP/CAL and return to perform this procedure at a later date. See
Appendix D.
10.8.4. Setting a Multiplication Factor
For convenience, it is possible to leave a channel setup one way and merely multiply the display value by
some factor and replace the display with the new value. An example might be to set up a channel to read
from zero to one hundred percent and use a multiplication factor to make the display read from zero to
500 SCCM. An infinite number of possibilities are possible.
Once a channel in the CHANNEL SELECT portion of the SETUP/CAL menu has been selected, press
the number 7 key to select CALIBRATE. From the CALIBRATE menu, choose MULTIPLIER by
pressing the number 4 key. While the MULTIPLIER window is in the display, enter the multiplication
factor and press ENTER. The display returns to normal with the selected channel now displaying the
product of the normally displayed value and the entered factor.
164-122012 Power
POD
400 Page 27 of 42
Page 28
10.9. SERIAL COMMUNICATION
Follow the instructions in the section, WIRING THE POWER
serial communication with the unit. After insuring that the POWER
POD
-400 for cabling and proper pin out for
POD
-400 is wired properly, the unit
must be set up following the instructions in the section entitled, External Communications Setup.
POD
The instruction set for the POWER
-400 can be divided into two different types. They differ only in
that they
1. Write a command only or
2. Write a command and read a response.
Taking liberty with this vernacular, they may be differentiated using the terms COMMAND or QUERY.
The entire set of instructions can be found in Appendix B.
10.9.1. Commands
Typical RS-232 command structure is as follows: CMNDn<value>Cr
For example: To change channel three’s high alarm value to 75.00, send A3H075.00Cr.
RS-485 commands require only that an asterisk and the unit’s address precede the RS-232 Command.
Using the same example as above with a unit whose address is 10 (default), send *10A3H075.00Cr.
POD
When the POWER
shorted to add the required terminating resistors. Remove two machine screws from the rear of the top
cover. Slide the cover off. The two jumpers are near the serial communication connectors.
-400 is the terminating unit on a 485 bus, two jumpers, CJ1 and CJ2 can be
10.9.2. Set Point Queries
Queries are used when information from the unit is required. Each query is followed by a response from
the POWERPOD-400.
In this case, the query is asking for a programmed value. Specifically, the set point of channel three, which
happens to be set at 50.00 of 100.00.
Sent SP3Cr
Response SP3 050.00
10.9.3. Alarm/Flag Queries
The state of the high and low flow alarms for each channel can be monitored via the DB-15 connecter (J8)
on the rear panel for immediate action from an alarm situation but the Totalizer set points can only be monitored by reading a memory location’s Boolean setting via serial communication.
As shown in the section, Setting Limit Alarms, to simply QUERY channel three’s high alarm set point,
send A3HCr. Expect a response similar to A3H 075.00. The actual state of the alarm can be polled using
the STATUS query.
STCr
Which results in a response similar to
STATUS
OCA : CH1 AUTO CH2 CLOSED CH3 OPEN CH4 AUTO
HI/LO: 0/0 0/1 1/0 0/0
OCA is short for OPEN/CLOSED/AUTO.
HI/LO stands for the high flow limit alarm or the low flow limit alarm. One must parse the states if digital
tests are to be conducted.
164-122012 Power
POD
400 Page 28 of 42
Page 29
To query the status of the totalizer flag on channel four (4), use the following command.
TF4Cr
The response will be either TF4 0 corresponding to a Boolean ‘FALSE’ indication that the totalizer flag
has not been set, or TF4 1, a Boolean ‘TRUE’, indicating that the total is equal to or beyond the totalizer
set point..
D41Cr command tells channel four (4) to display the TOTAL. It is not necessary that the front panel
read the total for the TOTALIZER to work. The Totalizer works in the background no matter what mode
the display is in.
NOTE: It must be emphasized that, if the total reaches the maximum of the
display (±999999), the unit effectively stops counting.
164-122012 Power
POD
400 Page 29 of 42
Page 30
Pod
11.0 Appendix A
Power
Command Name Query Response
Channel 1 Display C1 CH1 ddd.dd U of M GasID
Channel 2 Display C2 CH2 ddd.dd U of M GasID
Channel 3 Display C3 CH3 ddd.dd U of M GasID
Channel 4 Display C4 CH4 ddd.dd U of M GasID
All Channels’ Display C5
SP1<dd.ddd> Set Point - Channel 1 SP1 SP1 ddd.dd notes 1, 2
SP2<dd.ddd> Set Point - Channel 2 SP2 SP2 ddd.dd
SP3<dd.ddd > Set Point - Channel 3 SP3 SP3 ddd.dd
SP4<dd.ddd > Set Point - Channel 4 SP4 SP4 ddd.dd
A1H<dd.ddd > Alarm Set Point, Ch1, High A1H A1H ddd.dd notes 1, 2
A1L<dd.ddd > Alarm Set Point , Ch1, Low A1L A1L ddd.dd
A2H<dd.ddd > Alarm Set Point , Ch2, High A2H A2H ddd.dd
A2L<dd.ddd > Alarm Set Point , Ch2, Low A2L A2L ddd.dd
A3H<dd.ddd > Alarm Set Point , Ch3, High A3H A3H ddd.dd
A3L<dd.ddd > Alarm Set Point , Ch3, Low A3L A3L ddd.dd
A4H<dd.ddd > Alarm Set Point , Ch4, High A4H A4H ddd.dd
A4L<dd.ddd > Alarm Set Point , Ch4, Low A4L A4L ddd.dd
FL1<d> Filter Setting, Ch1 FL1 FL1 d
FL2<d> Filter Setting, Ch2 FL2 FL2 d
FL3<d> Filter Setting, Ch3 FL3 FL3 d
FL4<d> Filter Setting, Ch4 FL4 FL4 d
ML1<d.dddd> Multiplier, CH1 ML 1 ML 1 dddd.d notes 1, 2
ML2<dd.ddd> Multiplier , CH2 ML 2 ML 2 ddd.dd
ML 3<ddd.dd> Multiplier , CH3 ML 3 ML 3 dd.ddd
ML 4<dddd.d> Multiplier , CH4 ML 4 ML 4 d.dddd
-400 Serial Commands
CH1 ddd.dd U of M GasID
CH2 ddd.dd U of M GasID
CH3 ddd.dd U of M GasID
CH4 ddd.dd U of M GasID
IN1 d
rz - r
IN2 d
rz - r
IN3 d
rz - r
rz - r
IN4 d
fs
fs
fs
fs
notes 6, 7
164-122012 Power
POD
400 Page 30 of 42
Page 31
APPENDIX A (Cont.)
Pod
Power
Command Name Query Response
D1<d > Ch1 Disp Mode D1 d note 5
D2<d> Ch2 Disp Mode D2 d
D3<d > Ch3 Disp Mode D3 d
D4<d > Ch4 Disp Mode D4 d
T1S<dddddd> Totalizer Set Point, Ch1 T1S T1S dddddd
T2S<dddddd> Totalizer Set Point, Ch2 T2S T2S dddddd
T3S<dddddd> Totalizer Set Point, Ch3 T3S T3S dddddd
T4S<dddddd> Totalizer Set Point, Ch4 T4S T4S dddddd
T1M<d> Totalizer Mode, Ch1 T1M T1M d note 8
T2M<d> Totalizer Mode, Ch2 T2M T2M d
T3M<d> Totalizer Mode, Ch3 T3M T3M d
T4M<d> Totalizer Mode, Ch4 T4M T4M d
Units-of-Measure for Meter Reading & Corresponding Totalizer Units
Standard Cubic Centimeters per MinuteSCCMSCC
Standard Liters per MinuteSLMSL
Percent%
VoltsV
MillivoltsMV
CountsCNT
Normal Liters per MinuteNLMNL
Standard Liters per SecondSLSSL
Normal Liters per SecondNLSNL
Standard Liters per HourSLHS L
Normal liters per HourNLHNL
Standard Mililiters per MinuteSMLMSML
Normal Mililiters per MinuteNMLMNML
Standard Mililiters per SecondSMLSSML
Normal Mililters per SecondNMLSNML
Standard Mililiters per HourSMLHSML
Normal Mililiters per HourNMLHNML
Normal Cubic Centimet ers per MinuteNCCMNCC
Standard Cubic Centimeters per SecondSCCSSCC
Normal Cubic Centimeters per SecondNCCSNCC
Standard Cubic Centimeters per HourSCCHSCC
Normal Cubic Centimet ers per HourNCCHNCC
Standard Cubic Feet per MinuteSCFMSCF
Normal Cubic Feet per MinuteNCFMNCF
Standard Cubic Feet per SecondSCFSSCF
Normal Cubic Feet per SecondNCFSNCF
Standard Cubic Feet per HourSCFHSCF
Normal Cubic Feet per HourNCFHNCF
Standard Cubic Meters per MinuteSCMMSCM
Normal Cubic Meters per MinuteNCMMNCM
Standard Cubic Meters per SecondSCMSSCM
Normal Cubic Meters per SecondNCMSNCM
Standard Cubic Meters per HourSCMHSCM
Normal Cubic Meters per HourNCMHNCM
O 166 Sulfur Hexafluoride SF6 185 Vinyl Chloride 185
3H8
Propyl Amine C3H9N
168 Sulfur Trifluoride SF3 187 Water Vapor H2O
3H6
Pyradine C5H5N
170 Tetrachloroethylene 170 189 Xylene, m- C8H10
2F2
R123 R123
R125 C2HF5
R134A R134A
R143A R143A
157
R218 C3F8
159
Radon Rn
161
Silane SiH4
163
Sulfur Dioxide SO2
165
Sulfur Tetrafluoride SF4
167
Sulfur Trioxide SO3
169
Tetrafluoroethylene C2F4
171
Tert-butanol 173
173
Transbutene C4H8
176
Trichloroethylene 178
178
Triethylamine 180
180
Tungsten Hexafluoride WF6
182
Vinyl Bromide 184
184
Vinyl Flouride C2H3F
186
Xenon Xe
188
Xylene, o- C8H10
190
O 191 Xylene, p- C8H10
4H8
Mixtures MIXT
191
S 193 Other ????
4H4
194
164-122012 Power
POD
400 Page 35 of 42
Page 36
14.0 Appendix D
Setting the Zero & Span
on the
Power
All procedures outlined in this document must be performed with the power supply turned on and
warmed up for at least one hour.
All procedures outlined in this document must be performed with NO transducers connected to the
channel being adjusted*.
You CANNOT successfully zero a channel without, first, supplying zero volts or four milliamps to
the meter input for that channel.
You CANNOT successfully span a channel without, first, supplying the meter input for that channel
with its appropriate span voltage or span current.
14.1 Zeroing Unit (Voltage Range).
The following instructions explain how to short a given channel’s input signal to ground (0 volts DC)
and set the display to read zero at this voltage.
Locate the 15-pin, ‘D’ type connector that corresponds to the channel to be zeroed.
Short pin 5 (signal common) to pin 6 (signal in).
Press MODE
Press 3
Press ENTER
Press 1, 2, 3 or 4, depending on which channel is to be zeroed.
Press ENTER
Press 7
Press ENTER
Press 1
Press ENTER
Press ENTER again
You have just told a channel to display zero when the input signal is zero volts
The following instructions explain how to use the POWER
POD
-400 command signal to supply
4mAmp signal to a channel’s input and set the display to read zero at this current level.
Disconnect all connectors from the channel to be calibrated.
Insure that the channel is set to read 4 to 20 mA signals by performing the following steps.
Press MODE
Press 3
Press ENTER
Press 1, 2, 3 or 4, depending on which channel is to be zeroed.
Press ENTER
Press 4
Press ENTER
Press 3
Press ENTER
The normal display returns and will respond in the 4 to 20 mA mode.
Connect a milliamp meter between pins 14 and 6 on the 15-pin ‘D’ connector corresponding to the
channel to be calibrated.
Set the command for the selected channel to zero.
Press the CHANNEL # button
Press 1, 2, 3 or 4, depending on which channel is to be zeroed.
Press 0.0 using the numbered key pad
Press ENTER
The milliamp meter should be reading between 3.996 and 4.004 mAmps.
To Zero the meter with a signal of 4mA, perform the following steps.
Press MODE
Press 3
Press ENTER
Press 1, 2, 3 or 4, depending on which channel is to be zeroed.
Press ENTER
Press 7
Press ENTER
Press 1
Press ENTER
Press ENTER again
You have just told a channel to display zero when the input signal is 4mA
!
.
164-122012 Power
POD
400 Page 37 of 42
Page 38
14.3 Spanning Unit.
The following instructions explain how to use the POWER
POD
-400 command signal to supply a given
channel’s input with the proper span voltage and to set the display to read a transducer’s span value.
You must know four things prior to setting the display’s span value:
1. You must know the maximum signal level for the specific transducer connected to the
channel to be spanned (5 VDC, 10VDC or 20mA).
2. You must know the maximum value to be displayed when the maximum signal level for the
transducer is connected to the channel to be spanned.
3. You must insure that the channel’s multiplier is set to 1.0000.
4. You must know the channel’s ‘Span Value’. That is, the value that would NOW be
displayed (prior to any adjustment) if the maximum signal was fed into the input of the
channel to be adjusted.
1 and 2. If this information is not on the transducer itself, consult the manufacturer of the
transducer. Their literature or other documentation should specify the maximum output signal
from the transducer and the maximum value to be displayed at that signal level.
3. Setting the MULTIPLIER to 1.0000.
Press MODE
Press 3
Press ENTER
Press 1, 2, 3 or 4 depending on which channel is to be adjusted.
Press ENTER
Press 7
Press ENTER
Press 4
Press ENTER
Press 1.0000 using the numeric key pad.
Press ENTER
You have just set a channel’s multiplier to 1.0000.
4. If you do not know what the selected channel’s maximum display (Span) value is currently set
to, follow these steps.
Notice that the last step in this exercise is to press ESC, not ENTER.
Press MODE
Press 3
Press ENTER
Press 1
, 2, 3 or 4 depending on which channel is to be adjusted.
Press ENTER
Press 7
Press ENTER
Press 3
Press ENTER
Read the VALUE from the display.
Press ESC
164-122012 Power
POD
400 Page 38 of 42
Page 39
An example of one possible transducer/ POWER
0
TRANSDUCER POWER
Max signal out
5 VDC 250.00 1.000
Setting the THPS-400 Analog level to correspond with the transducer to be attached.
Max value to be
displayed
POD
-400 combination.
POD
-400
Multiplier Current Span Value
100.00
Press MODE
Press 3
Press ENTER
Press 1, 2, 3 or 4 depending on which channel is to be adjusted.
Press ENTER
Press 7
Press ENTER
Press 1 for 0 – 5 VDC, 2 for 0 – 10 VDC or 3 for 4 – 20 mADC
Press ENTER
In the example listed above, you would have selected option number 1 for the 0 – 5 volt analog
signal level since the maximum signal from the transducer is 5 VDC.
You have just set the analog operating level (range) for the selected channel.
Setting the display to read the maximum transducer value at the maximum transducer input.
Locate the 15-pin, ‘D’ type connector that corresponds to the channel to be spanned.
For units set to meter a voltage range, short pin 14 (command out) to pin 6 (signal in) of the
selected channel and place a calibrated volt meter to read from pin 5 (common) to pin 6.
For units set to read a 4 to 20 mA range, connect a milliamp meter between pins 14 and 6.
Press the CHANNEL # key corresponding to the selected channel. An asterisk (*) appears.
Use the number keys to enter THPS-400 current span value. Use the decimal!
Press ENTER
Adjust the command to achieve a value as close to the max analog signal level as possible (±0.01
volts for the 5 volt range, ±0.02 volts for the 10 volt range and ±0.004 ampsfor the milliamp range.
Do this using the CHANNEL# key, the numeric key pad and the ENTER key until the desired
#
signal is established
.
Press MODE
Press 3
Press ENTER
Press 1, 2, 3 or 4 depending on which channel is to be adjusted.
Press ENTER
Press 3
Press ENTER
Use the numeric keys to enter the new SPAN value (Max transducer value to be displayed).
Press ENTER
Press ENTER
164-122012 Power
POD
400 Page 39 of 42
Page 40
Congratulations!
You have just calibrated your Power
Pod
-400 to read zero at zero volts and the transducer’s max
display value at the transducer’s max signal input.
The only thing left to do is to insure that your transducer is calibrated and that it is wired correctly
to the Power
Pod
-400.
* Individual channels may be calibrated with calibrated transducers connected, breakout
connectors (not supplied) and precision multi-meters with NIST traceable calibrations (also not
supplied). Such procedures are not covered in this document.
! If the calibration for the channel just zeroed had been severely compromised prior to beginning
this procedure, zero may not be immediately displayed and you may have to repeat the procedure
after resetting the span. Check that the span has been set correctly!
# If a signal level within .1% of the maximum DC voltage level is not attainable, the power supply
may need to be repaired.
164-122012 Power
POD
400 Page 40 of 42
Page 41
15.0 Drawings
4
P
ower
3
CLOSE
CLOSE
OPEN
1
2
CLOSE
CLOSE
OPEN
OPEN
OPEN
Pod
400
HASTINGS
7
825
936
4
1
0
.
164-122012 Power
POD
400 Page 41 of 42
Page 42
16.0 Warranty
16.1. Warranty Repair Policy
Hastings Instruments warrants this product for a period of one year from the date of shipment to be free
from defects in material and workmanship. This warranty does not apply to defects or failures resulting
from unauthorized modification, misuse or mishandling of the product. This warranty does not apply to
batteries or other expendable parts, nor to damage caused by leaking batteries or any similar occurrence.
This warranty does not apply to any instrument which has had a tamper seal removed or broken.
This warranty is in lieu of all other warranties, expressed or implied, including any implied warranty as to
fitness for a particular use. Hastings Instruments shall not be liable for any indirect or consequential
damages.
Hastings Instruments, will, at its option, repair, replace or refund the selling price of the product if
Hastings Instruments determines, in good faith, that it is defective in materials or workmanship during the
warranty period. Defective instruments should be returned to Hastings Instruments, shipment prepaid,
together with a written statement of the problem and a Return Material Authorization (RMA) number.
Please consult the factory for your RMA number before returning any product for repair. Collect freight
will not be accepted.
16.2. Non-Warranty Repair Policy
Any product returned for a non-warranty repair must be accompanied by a purchase order, RMA form
and a written description of the problem with the instrument. If the repair cost is higher, you will be
contacted for authorization before we proceed with any repairs. If you then choose not to have the product
repaired, a minimum will be charged to cover the processing and inspection. Please consult the factory for
your RMA number before returning any product repair.
TELEDYNE HASTINGS INSTRUMENTS
804 NEWCOMBE AVENUE
HAMPTON, VIRGINIA 23669 U.S.A.
ATTENTION: REPAIR DEPARTMENT
TELEPHONE (757) 723-6531
1-800-950-2468
FAX (757) 723-3925
E MAIL mail to:hastings_instruments@teledyne.com
INTERNET ADDRESS http://www.teledyne-hi.com/
Repair Forms may be obtained from the “Information Request” section of the Hastings Instruments
164-122012 Power
POD
400 Page 42 of 42
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.