Teledyne 400 User Manual

Page 1
TELEDYNE HASTINGS
INSTRUMENTS
INSTRUCTION MANUAL
Power
Power Supply/Totalizer
Pod
400
ISO 9001
CERTIFIED
Page 2
Manual Print History
The print history shown below lists the printing dates of all revisions and addenda created for this manual. The revision level letter increases alphabetically as the manual undergoes subsequent updates. Addenda, which are released between revisions, contain important change information that the user should incorporate immediately into the manual. Addenda are numbered sequentially. When a new revision is created, all addenda associated with the previous revision of the manual are incorporated into the new revision of the manual. Each new revision includes a revised copy of this print history page.
Revision D (Document Number 164-082005) ................................................................ August 2005
Revision E (Document Number 164-102005) ............................................................... October 2005
Revision F (Document Number 164-012006) ................................................................ January 2006
Revision G (Document Number 164-092006) ........................................................... September 2006
Revision H (Document Number 164-112006) ........................................................... November 2006
Revision J (Document Number 164-022009) ............................................................... February 2009
Revision K (Document Number 164-062009) .................................................................... June 2009
Revision L (Document Number 164-082010) ................................................................. August 2010
Revision M (Document Number 164-102012) ............................................................. October 2012
Revision N (Document Number 164-122012) ........................................................... December 2012
Visit www.teledyne-hi.com for WEEE disposal guidance.
Hastings Instruments reserves the right to change or modify the design of its equipment
without any obligation to provide notification of change or intent to change.
164-122012 Power
POD
400 Page 2 of 42
Page 3
Table of Contents
1.0 QUICK START INSTRUCTIONS ..................................................................................................................................... 5
2.0 SAFETY ................................................................................................................................................................................ 6
3.0 FEATURES .......................................................................................................................................................................... 7
4.0 SPECIFICATIONS ............................................................................................................................................................... 9
5.0 FRONT PANEL ................................................................................................................................................................. 10
6.0 REAR PANEL..................................................................................................................................................................... 11
7.0 WIRING.............................................................................................................................................................................. 12
7.1. POWER ........................................................................................................................................................ 12
7.2. COMMUNICATIONS ............................................................................................................................... 12
7.3. TRANSDUCER CONNECTIONS ............................................................................................................ 13
7.4. ALARMS...................................................................................................................................................... 13
7.5. ANALOG SIGNAL FOLLOWERS ............................................................................................................ 14
8.0 MANUAL OPERATION ................................................................................................................................................... 15
8.1. POWER ON/OFF ........................................................................................................................................ 15
8.2. CHANGING THE COMMAND SET POINT .......................................................................................... 15
8.3. OVERRIDE OPEN ...................................................................................................................................... 15
8.4. OVERRIDE CLOSED ................................................................................................................................. 15
8.5. SETTING A CHANNEL TO AUTO CONTROL .................................................................................... 16
8.6. SETTING A CHANNEL TO DIRECTLY METER INCOMING SIGNALS ......................................... 16
8.7. SETTING A CHANNEL TO DISPLAY THE TOTALIZER FUNCTION ............................................ 16
9.0 EXTERNAL/REMOTE OPERATION .............................................................................................................................. 18
9.1. SELECTING EXTERNAL/REMOTE OPERATION (FRONT PANEL ONLY) ........................................ 18
9.2. CHOOSING RS-232 COMMUNICATION .............................................................................................. 18
9.3. CHOOSING RS-485 COMMUNICATION .............................................................................................. 18
10.0 SETUP/CAL MODE .......................................................................................................................................................... 19
10.1. ENTERING THE SETUP/CAL MODE .................................................................................................... 19
10.2. DISPLAY SETUP ........................................................................................................................................ 19
10.2.1. BLANKING A DISPLAY LINE/CHANNEL ........................................................................................................ 19
10.2.2. SELECTING DISPLAY BRIGHTNESS ............................................................................................................... 19
10.2.3. SELECTING UNIT-OF-MEASURE ................................................................................................................... 20
10.2.4. SELECTING GAS ID ...................................................................................................................................... 20
10.2.5. SETTING A/D CONVERSION RATE (FILTER) ................................................................................................ 20
10.2.6. FRONT PANEL LOCK OUT ............................................................................................................................ 21
10.3. TOTALIZER SETUP ................................................................................................................................. 21
10.3.1. COUNT UP TO A SET POINT ......................................................................................................................... 21
10.3.2. COUNT DOWN FROM A SET POINT .............................................................................................................. 22
10.3.3. COUNT CONTINUOUSLY .............................................................................................................................. 22
10.4. EXTERNAL COMMUNICATION SETUP ............................................................................................. 22
10.4.1. RS-232 SETTINGS ........................................................................................................................................ 22
10.4.2. RS-485 SETTINGS ........................................................................................................................................ 22
10.5. SELECTING THE ANALOG SIGNAL LEVEL ....................................................................................... 23
10.5.1. ZERO TO FIVE VOLT OPERATION ................................................................................................................. 23
10.5.2. ZERO TO TEN VOLT OPERATION ................................................................................................................. 23
164-122012 Power
POD
400 Page 3 of 42
Page 4
10.5.3. FOUR TO TWENTY MILLIAMP OPERATION .................................................................................................. 23
10.6. SETTING LIMIT ALARMS ...................................................................................................................... 24
10.6.1. SETTING A SINGLE CHANNELS HIGH LIMIT ............................................................................................... 24
10.6.2. SETTING A SINGLE CHANNELS LOW LIMIT ................................................................................................ 24
10.6.3. SETTING A SINGLE CHANNELS HYSTERESIS ............................................................................................... 24
10.7. SETTING RATIO CONTROL PARAMETERS ...................................................................................... 25
ENABLING RATIO CONTROL ..................................................................................................................................... 25
10.8. CALIBRATING A CHANNEL TO ITS INCOMING SIGNALS ............................................................ 26
10.8.1. RESETTING THE ZERO ONLY ....................................................................................................................... 26
10.8.2. RESETTING THE SPAN ONLY ........................................................................................................................ 26
10.8.3. RESETTING ZERO AND SPAN ........................................................................................................................ 27
10.8.4. SETTING A MULTIPLICATION FACTOR ........................................................................................................ 27
10.9. SERIAL COMMUNICATION ................................................................................................................... 28
10.9.1. COMMANDS .................................................................................................................................................. 28
10.9.2. SET POINT QUERIES ..................................................................................................................................... 28
10.9.3. ALARM/FLAG QUERIES ................................................................................................................................. 28
11.0 APPENDIX A ..................................................................................................................................................................... 30
12.0 APPENDIX B ..................................................................................................................................................................... 33
13.0 APPENDIX C ..................................................................................................................................................................... 33
14.0 APPENDIX D ..................................................................................................................................................................... 36
15.0 DRAWINGS ....................................................................................................................................................................... 41
16.0 WARRANTY ...................................................................................................................................................................... 42
16.1. WARRANTY REPAIR POLICY ......................................................................................................................... 42
16.2. NON-WARRANTY REPAIR POLICY ................................................................................................................ 42
164-122012 Power
POD
400 Page 4 of 42
Page 5
1.0 Quick Start Instructions
Important – The Power
Pod
-400 comes calibrated from the factory according to your specifications.
No set up is necessary unless you need to change the specs
1ea Power
1ea AC power cord
1ea 15-pin, Hi Density, D-style connector
1ea 9-pin, D-style connector
1ea user’s manual
When unpacking the Power
contain everything you need.
Pod
-400
Pod
, the package will
J-1 Ch1
J-2 Ch2
J-3 Ch3
J-4 Ch4
J-5
Analog
out
J-6 RS-232 Serial Com Port
J-7 RS-485
J-8 Alarms Open-collector Alarms
J-9 RS-485
Transducer Power,
Command and Analog
Signals.
Transducer Signal, Volts Only
Multi-drop com port (Daisy Chained)
Multi-drop com port (Daisy Chained)
Pin-outs (see section 7)
Connect a “known good” cable to the desired channel
(See next frame for pin-outs)
1) Check power setting switch on the back panel and make sure it is set appropriately.
2) Ensure the power switch is off, and then connect the power cord.
OPEN
1
CLOSE
OPEN
2
CLOSE
OPEN
3
CLOSE
OPEN
4
CLOSE
P
ower
1) Press the desired Channel # Key. Press the Auto key.
Turn power switch to on position. If you have just
turned on the mass flow instrument and the
power, some settling is necessary.
164-122012 Power
POD
400 Page 5 of 42
2) Press the desired Channel # Key. Enter desired flow. Press the enter key.
Pod
(For Controllers)
HASTINGS
3
2
1
5
6
4
8
7
9
.
0
Page 6
2.0 Safety
Read this manual in its entirety before operating the POWER POWER
and meters. Read all wiring and power hookup instructions and understand the requirements prior to using another manufacturer’s products with the POWER interfaced with the POWER standards before operating.
POD
-400 is designed to operate with most Teledyne Hastings Instruments (THI) flow controllers
POD
-400 is wired according to prevailing local safety and operational
The following symbols and terms may be found on THI products and/or in THI manuals and indicate important information.
When found on the device, this symbol indicates that the operator should refer to the manual for important instructions on the proper use of this device. When found in the manual, this symbol indicates that the reader should understand the implications contained in the text before operating the device.
POD
-400 Power Supply/Totalizer. The
POD
-400. Insure that any product being
This symbol indicates that a shock hazard may be present. Read the instruction manual carefully and insure that the device is wired properly and that all settings have been checked prior to applying power to the device.
The WARNING label indicates important information that should be heeded for safe and proper performance of the device.
The label, CAUTION, is used to indicate that damage to the power supply or equipment connected to it could occur if directions are not followed. Warranty could be invalidated if the instructions in this manual are not followed.
164-122012 Power
POD
400 Page 6 of 42
Page 7
3.0 Features
The POWER rack hardware or can be used as a bench top unit. The POWER character, vacuum fluorescent display (VFD). The display emulates a liquid crystal display in its command structure but the VFD gives the unit a greater viewing angle and better visibility than available with most conventional LED or LCD displays. The display can be set to four different brightness levels. Use a lower brightness setting to extend the already long expected life time of the display. Use brighter settings for viewing areas where ambient light may be too bright or cause glare, or where greater viewing distances are required.
Most features are accessible via the membrane keys on the front panel. Consult the section on each function to check its availability. Operators are guided through the many features and options by selecting their choices from an intuitive menu structure.
FRONT PANEL LOCKOUT
ANALOG RANGE SELECTION
Analog signal and control ranges are operator selectable. The operator can choose between three different DC ranges:
POD
-400 serves as a convenient control center that can be rack-mounted using standard half-
The Front Panel Lockout function is only available through serial communication. Manual Overrides remain available during Lockout via a minimum number of keystrokes using dedicated keys for this purpose and allow any command setting to be overridden in either the high (open) or low (closed) state.
POD
-400 is equipped with a 4 X 20
A fifteen (15) pin, high density, sub-miniature, D-type connector is provided for separate monitoring of each channel’s analog transducer signal.
CAUTION: Consult the appropriate section for limits to the loading of these signals.
SERIAL COMMUNICATION
The POWERPOD-400 comes equipped with standard RS-232 and RS-485, serial communication. Most functions, features, signals and alarms are accessible and modifiable via any remote computer.
0 -5 VDC, 0-10 VDC or 4 -20 mA.
The following commands are manual commands only:
OPEN,
The status of these settings can be read via serial communication but they cannot be changed except manually, from the front panel.
AUTO and CLOSE.
164-122012 Power
POD
400 Page 7 of 42
Page 8
POWER SELECTION
Power input is switchable between 100 VAC, 115 VAC and 230VAC (50 or 60 Hz) via the rear panel. For the safety of the operator as well as the device, the correct power level should be selected prior to connecting to the power mains. See the table “POWERPOD-400 Specifications” in section 3.0 for the proper fusing when changing power settings.
RATIO CONTROL Ratio control is possible between channels using a familiar master/slave configuration. Channel one (1) must be enabled as the master channel. Any combination of the remaining channels is possible for slave channel assignment.
TOTALIZER A Totalizer function is present for each channel with the capability of counting down from a set point, counting up to a set point or continuous count up. The maximum count is ±999999 units. When the set points are reached, a memory flag for each set point is set to a digital ‘1’ indicating a Boolean ‘true’ value. The Totalizer set point flags must be polled via digital communication to be read. Each flow channel has one low-limit and one high-limit set point available. These alarms are available via open-collector, opto-isolated outputs on the rear panel as well as serial communication.
POWER OUTAGES AND THE OVERRIDE CONDITION
In the event of a power outage, even one of short duration, the POWER
POD
-400 is
designed to conduct a software reset. During the period of time in which the reset is occurring, it will not accept or respond to any commands either manually or digitally until the reset process is completed. After said reset, the POWER
POD
-400 will have
remembered all previously entered set-points but all channels are designed to come up in the “Override-CLOSEd” condition. For meters, this should have no affect on their behavior. All analog-only, THI, flow controllers with normally-closed (NC) valves will close and remain closed until operator intervention manually returns selected channels’ Override condition back to the AUTO mode.
For all other controllers, this means that the control signal (pins 5 to 14) of J1 through J4 will return to the previously set level. For controllers not taking advantage of the Override function (pin 8), these controllers will return to their normal, preset operating condition unless other intermediary steps are undertaken. Controllers that are making use of the override function (Pin 8) and have normally-open valves will be driven to the fully open condition.
164-122012 Power
POD
400 Page 8 of 42
Page 9
y
y
4.0 Specifications
Power
Specification Value Units Notes
Power Inputs
V
P 68 VA f 50 -60 Hz
Transducer Number Channels 4 V I
I/O
Display
Type
# Lines 4 # Characters 20 Brightness Levels 4
A/D Converter Filtering Rate 4, 15, 30,
Alarms 3 per channel 1 High 1 Low 1 Total
Dimensions Front Panel (h x w) 3.5 x 9.5 in Case (h x w x d) 3 x 8 x 9.5 in Hole Centers (h x w)
Weight 5 1/4 lbs
Suppl
Suppl
Pod
-400 Specifications Table
100 115
VAC
230
±15
±250
0 - 5
0 - 10
VDC Bi-polar, per Channel
mA Bi-polar, per Channel
VDC
4 - 20 mADC
100
Hz
3 x 8.825 in
0.7 A, 250VAC, SB Fuse
0.6 A, 250VAC, SB Fuse
0.315 A, 250VAC, SB Fuse
Vacuum Fluorescent,
LCD Emulator
164-122012 Power
POD
400 Page 9 of 42
Page 10
(
A
A
(
(
p
A
e
5.0 Front Panel
1. CHANNEL NUMBER SELECT KEYS Selects channel for editing. An asterisk (*) appears in the first column of the display to indicate that this is the channel to
be edited.
2. OVERRIDE INDICATORS Indicates when a channel’s command signal is overridden high
3. OVERRIDE KEYS Override the command signal on the
sets command to –15V. AUTO allows the user to set the command signal for normal operation. A channel must be active before these keys can become operational.
4. DISPLAY AREA Column 1: Reserved for displaying
Column 2: Reserved for Col’s 3 – 8: Signal monitor. Displays current input signal while in METER mode,
Column 9: Spac Col’s 10 – 14: UNITS OF MEASURE Column 15: Space Col’s 16 – 20: GAS
5. KEYPAD Use to enter
OPEN
1
CLOSE
OPEN
2
CLOSE
OPEN
3
CLOSE
OPEN
4
CLOSE
Pod
P
ower
HASTINGS
400
OPEN) or low (CLOSED).
CTIVE CHANNEL. OPEN sets control override (pin 8) to +15V. CLOSED
CTIVE CHANNEL
TOTAL (T).
olarity indicator.
to average readings or TOTAL while in TOTALIZER mode.
display.
ID.
SET POINTS or to modify the SETUP or CALIBRATION of control unit.
*), MASTER channel (M), SLAVE
3
2
1
5
4
7
0
6
8
9
.
VERAGE while set
S) or
164-122012 Power
POD
400 Page 10 of 42
Page 11
6.0 Rear Panel
1. POWER ON/OFF SWITCH
2. POWER INLET & FUSE
3. POWER SELECTOR SWITCH
4. FUSE Vs. POWER SETTING TABLE
5. RS-232 SERIAL PORT CONNECTOR (J6)
6. RS-485, DAISY CHAINED, SERIAL PORT CONNECTORS (J7, J9)
7. TRANSDUCER CONNECTORS (J1-J4)
8. ANALOG OUTPUT (J5)
9. ALARMS (J8)
164-122012 Power
POD
400 Page 11 of 42
Page 12
7.0 Wiring
7.1. POWER
Power is supplied through a fused, AC jack on the rear panel (item 2).Use the power cord supplied with the unit (PN15-17-011 for 115 VAC, 60Hz). See the following table for selecting the proper fuse rating. Use a metric, 5 x 20 mm sized, time-delayed fuse.
Power Setting (50 – 60 Hz) Fuse Rating THI P/N
100 VAC 0.315 Amp/250 VAC 23-05-038 115 VAC 0.60 Amp/250 VAC 23-05-039 230 VAC 0.70 Amp/250 VAC 23-05-040
Cords without plugs are supplied with units shipped outside of the U.S. Consult and comply with any local laws and/or codes when connecting to any AC main. The AC input is user selectable between 100, 115 or 230 VAC, 50 or 60 Hz, via an AC selector switch next to the AC jack (Item 3).
WARNING: Be sure to set the power select switch prior to connecting to mains. Re-fuse the connector according to the table above.
7.2. COMMUNICATIONS
Connectors J6 (Item 4), J7 (Item 6) and J9 (Item 6) are for RS-232, RS-485 connections respectively. Settings for serial communication are accessible via the front panel.
J6 1 Unused
RS-232
(DB-9)
2 Tx 3 Rx 4 Unused 5 Gnd 6 Unused 7 RTS 8 CTS 9 Unused
J7, J9 1 Unused
RS-485
(DB-9)
2 Rx­3 Tx+ 4 Unused/Gnd (Gnd) 5 Unused 6 Gnd/VCC (VCC) 7 R+ 8 T­9 Unused
164-122012 Power
POD
400 Page 12 of 42
Page 13
7.3. TRANSDUCER CONNECTIONS
Connectors J1, 2, 3 and 4 (Item 7) are 15 pin D style connectors wired in the standard Hastings Instruments pin-out (H pin-out).
J1, J2, J3, J4 1 NC
Transducer
Connectors
(DB-15)
(H-Pinout)
2 NC Valve Cntrl
3 NC mA Sig 4 NC mA Sig 5 Sig. Com. 6 Sig. In 7 Case Gnd. 8 Cntrl Over-ride 1.5mA
9 -15 VDC 10 NC 11 +15 VDC 12 Valve Return 13 NC Ext-In 14 Set Point Out 15 +5 VDC Ref. Not Used.
Voltage
7.4. ALARMS
Connector J8 (item 9) provides the user with open-collector, opto-isolated alarms for individual channels. Each channel is provided with one user settable “High” and one “Low” alarm.
J8 1 Chnl 1, High Alarm
Open
Collector,
Opto-
Isolated, High
& Low
Alarms
(HD DB-26)
2 Chnl 1, Low Alarm 3 Chnl 1, Alarm Return 4 Chnl 2, High Alarm 5 Chnl 2, Low Alarm 6 Chnl 2, Alarm Return 7 NC 8 NC
9 NC 10 NC 11 NC 12 NC 13 NC 14 NC 15 NC 16 NC 17 NC 18 NC 19 Chnl 3, High Alarm 20 Chnl 3, Low Alarm 21 Chnl 3, Alarm Return 22 Chnl 4, High Alarm 23 Chnl 4, Low Alarm 24 Chnl 4, Alarm Return 25 NC 26 NC
164-122012 Power
POD
400 Page 13 of 42
Page 14
7.5. ANALOG SIGNAL FOLLOWERS
Analog signals from each channel’s transducers are available for reading or for sending to another power supply for ratio (Master/Slave) operation. The signal can be sent to a channel on another power supply and programmed as a Master for that power supply, allowing the remaining three channels to operate as slaves. When operating 4 – 20 mA instruments, the followers supply a 0.5 -2.5 volt signal.
J5 1 Channel 1 Signal
2 Channel 1 Return
3 Channel 2 Signal
4 Channel 2 Return
5
6
Analog
Out
(HD DB-15)
7
8
9 10 11 Channel 3 Signal 12 Channel 3 Return 13 Channel 4 Signal 14 Channel 4 Return 15
164-122012 Power
POD
400 Page 14 of 42
Page 15
8.0 Manual Operation
8.1. POWER ON/OFF
The Power On switch, item 1 in rear panel drawing, is located in the upper right corner of the rear panel. Insure that the proper power setting is selected prior to turning the power on. See the Power section of WIRING THE POWER
8.2. CHANGING THE COMMAND SET POINT
Channel two (2), example shown.
Press the desired CHANNEL # key. An asterisk appears in the first space on the line representing the selected channel. The meter display is immediately replaced with the current set point. Pressing a number key begins the editing process. The cursor lands on the channel set point to be edited, and the new command is entered with the most significant bit (MSB) first. The set point is filled in from left to right on the display. When entering a new set point, you must use the decimal key when it has been used in setting the SPAN. Pressing ENTER completes the editing process. The old command is not changed until the ENTER key is actuated. Hitting the ESCape button at any time prior to the ENTER key will return the display to its previous state without any changes being made. This command will not be applied to the output until the channel is set for AUTO operation. See Setting a Channel to AUTO Control.
POD
-400, above.
8.3. OVERRIDE OPEN
Press the desired CHANNEL # key. An asterisk appears in the first space on the line representing the selected channel. The meter display is immediately replaced with the current set point. Pressing the OPEN key results in the application of +15 VDC to pin number eight (8) of the corresponding 15 pin Sub-D connector and returns the previously programmed metering function to the display.
8.4. OVERRIDE CLOSED
Press the desired CHANNEL # key. An asterisk appears in the first space on the line representing the selected channel. The meter display is immediately replaced with the current set point. Pressing the CLOSE key results in the application of -15 VDC to pin number eight (8) of the corresponding 15 pin Sub-D connector and returns the previously programmed metering function to the display.
164-122012 Power
POD
400 Page 15 of 42
Page 16
8.5. SETTING A CHANNEL TO AUTO CONTROL
Press the desired CHANNEL # key. An asterisk appears in the first space on the line representing the selected channel. The meter display is immediately replaced with the current set point. Pressing the AUTO key causes pin number eight (8) of the 15 pin Sub-D connector to float and returns the previously programmed metering function to the display. Pin 14 signal levels are now available for control.
8.6. SETTING A CHANNEL TO DIRECTLY METER INCOMING SIGNALS
Press the MODE key. The MODE menu appears in the display. To select METER, press 1, and then ENTER. The METER menu allows the selection of the desired channel. Press the number key that
corresponds with the desired channel followed by ENTER. The previously programmed display returns with the selected channel reading the signal between pins 5 and 6 of the corresponding channel’s 15 pin D-connector (J1 – J4).
8.7. SETTING A CHANNEL TO DISPLAY THE TOTALIZER FUNCTION
The POWERPOD-400 provides an integrated (Riemann Sum) value of the incoming signal for each channel.
=
)**(
fCfT∑= , where:
IS
Each channel samples the incoming signal at a rate corresponding to the preset A/D conversion rate in Hz. At each sampling interval, the TOTALIZER function multiplies the average signal, as a fraction of the full-scale value (5v, 10v or 20 mA), times the span, or cal, value. This results in a rate for that interval. Next, a time element factor is determined according to the flow rate programmed for that channel as follows.
For a time element of seconds, the factor is 1/10. For minutes, the factor is 1/600. Hours use a factor of 1/36000.
Each calculation is summed and stored for a TOTAL value.
TotalT
⎧ ⎪
=
S
=
⎪ ⎪
=
I
FactorSignalFractionalf ValueSpanorvalueCALC
IntervalTimeFractionalf
⎫ ⎪
⎪ ⎬
⎪ ⎪
164-122012 Power
POD
400 Page 16 of 42
Page 17
Example (See Graph):
A linear flow transducer with a DC output of 0 to 5 volts is calibrated for a maximum flow of 25 SLH. The transducer is connected to one channel of a POWER volts input and has been programmed to display rate in SLH. The transducer is sending a constant 1.000­volt signal to the POWERPOD-400. What is the total flow after 40 minutes?
During one sampling interval, the TOTALIZER reads the one-volt signal and calculates the fraction of the possible full-scale value.
1 volt/5 volts or 0.20
The fractional signal value is multiplied times the CAL value
1/5 * 25 = 5
Then, because the channel’s programmed rate is in SLH (standard liters per hour), a factor of 1/3600 is applied to the result.
5 * 1/36000 = 1.389 x 10^-4 standard liters per sampling interval.
A constant signal received over a 40 minute time period would amount to 40 minutes * 60 seconds * 10 samples per second = 24000 samples. The TOTAL would be
2.400 E03 * 1.389 E-04 = 3 1/3 Std Liters.
If no rate is programmed on that channel, The TOTALIZER function is disabled for that channel and the display returns to its regular metering configuration.
POD
-400 which is CAL’d to read 25.000 at 5
To enable the TOTALIZER, press the MODE key. The MODE menu appears in the display. To select TOTAL, press 2, and then ENTER. The TOTALIZER menu allows the operator to select the desired channel. Press the number key that corresponds with the desired channel followed by ENTER. A TOTALIZER menu is presented that allows the operator to go directly to displaying whatever the
Totalizer has currently stored as the total or to reset the counter. If DISPLAY is selected by pressing the number 1 key followed by ENTER, the display will return to normal with the most recently stored total being displayed on the appropriate line. If RESET is chosen by pressing the number 2 key followed by ENTER, the display returns with the total value reset according to the pre-selected counting mode presented in the table below with the Totalizer counting.
164-122012 Power
Mode Reset Value
Count Down Set Point
Count Up Zero
Continuous Zero
POD
400 Page 17 of 42
Page 18
9.0 External/Remote Operation
9.1. SELECTING EXTERNAL/REMOTE OPERATION (Front Panel Only)
The POWER The following text describes how to choose and activate the desired method. Further information and instructions on how conditions may be changed can be found in section 10, SETUP/CAL mode.
POD
-400 allows the user to select different methods of serial communication and control.
9.2. CHOOSING RS-232 COMMUNICATION
Press the MODE button. The MODE menu is displayed. Press the number “4” key for XTRNL CNTRL. The XTRNL CNTRL menu is displayed. Choose RS-232 by pressing the number “1” key.
The normal display will return to its previously programmed state.
9.3. CHOOSING RS-485 COMMUNICATION
Press the MODE button. The MODE menu is displayed. Press the number “4” key for XTRNL CNTRL. The XTRNL CNTRL menu is displayed. Choose RS-485 by pressing the number “2” key.
The normal display will return to its previously programmed state.
164-122012 Power
POD
400 Page 18 of 42
Page 19
10.0 Setup/Cal Mode
10.1. ENTERING THE SETUP/CAL MODE
Pressing the MODE key causes the MODE menu to be displayed. Choose the SETUP/CAL mode by pressing the number “3” key then ENTER. A CHANNEL SELECT menu is presented on the display. It should be noted that, although the user must select a specific channel number, some choices in the following menus may affect the operation of all channels and/or the display (ex; Display Brightness, type of communication). See the following menu options for more details.
10.2. DISPLAY SETUP
After entering the SETUP/CAL mode and selecting a channel number as described in the previous section, several selections are available that will affect the display. Follow the steps below to gain the desired customized display.
Note: Front panel lockout and display line (channel) blanking are only available through serial communication.
10.2.1. Blanking a Display Line/Channel
Use the serial command “DndCr”, where
n = channel number, d = 1 for Totalizer Mode, 2 for Meter Mode, 3 for a blanked line and Cr = carriage return.
10.2.2. Selecting Display Brightness
While the SETUP/CAL menu is displayed, choose DISPLAY by pressing the keys “1” + ENTER. Choose BRIGHTNESS by pressing the number “1” key + ENTER. The BRIGHTNESS menu will be visible. One of four different brightness levels can be chosen by pressing a number key. The number “1” key selects the dimmest setting while the number “4” key selects the brightest. Any change in brightness will affect the entire display after pressing the ENTER key.
164-122012 Power
POD
400 Page 19 of 42
Page 20
10.2.3. Selecting Unit-of-Measure
After entering the SETUP/CAL mode and after selecting a channel number, the SETUP/CAL menu is displayed. Choose DISPLAY (1) + ENTER and then select UNITS by pressing the number “2” key + ENTER. The UNITS menu will be displayed allowing the selection of over 50 different units of measure (UOM) simply by pressing the number key corresponding to the desired UOM. The number can be found by scrolling up or down until the desired UOM is displayed or by consulting Appendix B, at the end of this manual. Once selected, the UOM will be displayed in character spaces 10 to 14 on the appropriate line. Selecting UOM = “0” results in the UOM characters being blanked.
10.2.4. Selecting Gas ID
While the SETUP/CAL menu is displayed, Select the DISPLAY menu by pressing 1 + ENTER. Then choose GASID by pressing 3 + ENTER. The GASID menu will be displayed allowing the selection of nearly one hundred different chemical symbols simply by pressing the number key corresponding to the desired gas. The number can be found by scrolling up or down until the desired gas is displayed or by consulting Appendix C at the end of this manual. Once selected, the formula will be displayed in character spaces 16 through 20 on the appropriate line.
Note: Formulae too long to be completely displayed will display the Hastings Instruments Gas ID number. Choosing “0” will blank the GASID elements on the display.
10.2.5. Setting A/D Conversion Rate (Filter)
Press the MODE button. The MODE menu is displayed. Press the number “3” key followed by ENTER. The CHANNEL SELECT menu is displayed. Choose the channel whose signal is to be filtered using the number keys. For example, “1” + ENTER. The SETUP/CAL menu is displayed.
164-122012 Power
POD
400 Page 20 of 42
Page 21
After SETUP/CAL is selected and a channel number has been chosen, the SETUP/CAL menu is displayed. choose DISPLAY by pressing the number “1” key + ENTER. The FILTER option can be reached by scrolling down once to view the selection number and then pressing the “4” key + ENTER. A conversion rate of 4, 15, 30 or 100 Hertz may be selected by entering its menu number + ENTER. Enter the number of the desired A/D conversion rate and press ENTER. The normal display will return. Changes may be noticed in the flickering of the LSB on the display. In most situations, the flickering digits are an insignificant fraction of the display range and can be ignored. Should the flickering become annoying or make reading the display difficult, choose a slower rate of conversion.
NOTE: Slower rates of conversion may affect the accuracy of the TOTALIZER in situations where flow is changing often. Steady flow rates will see little loss of TOTALIZER accuracy.
10.2.6. Front Panel Lock Out
Front Panel Lockout is available only through serial communication. See the SERIAL COMMUNICATIONS section of this manual and Appendix A.
CAUTION: While the front panel is locked out, all command functions, excluding the override functions, are disabled. Commands can be overridden OPEN or overridden CLOSED during front panel lockout. This was determined to be useful both as a trouble shooting and as a safety tool. All other functions will be available only through the use of serial communication.
10.3. TOTALIZER SETUP
The TOTALIZER can be set to count up to a set point, to count down from a set point or to count up continuously to the maximum ability of the display (999999). In order for the TOTALIZER to work, a unit of measure must be chosen that includes a time element. For example, standard liters per hour (SLH) will totalize; percent (%) will not. Each mode is described in the following text and can be followed in the diagram below.
10.3.1. Count Up to a Set Point
The TOTALIZER will be set to count from zero (0) to a set point. When the set point is reached, a memory flag will be set to its logic high state. This flag is readable only through serial communications. The TOTALIZER will continue to count up to the maximum ability of the display.
While the SETUP/CAL menu is displayed, choose TOTAL by pressing the number “2” key followed by ENTER. A TOTALIZER menu is displayed. Press the number “1” key to choose the CNT UP option. The COUNT UP/DN window is displayed for the selected channel. Enter the desired set point using the numbered key pads and press ENTER. The unit returns the previously programmed display with the selected channel now displaying the Riemann Sum of the received transducer signal.
164-122012 Power
POD
400 Page 21 of 42
Page 22
10.3.2. Count Down from a Set Point
The TOTALIZER will be set to count from a preset value to zero (0). When zero is reached, a memory flag will be set to its logic high state. This flag is readable only through serial communications. The TOTALIZER will continue to count in the negative direction from zero to the
maximum ability of the display (-999999).
While the SETUP/CAL menu is displayed, choose TOTAL by pressing the number “2” key + ENTER. A TOTALIZER menu is displayed. Press the number “2” key to choose the CNT DN option. The COUNT UP/DN window is displayed. Enter the desired set point using the numbered key pads and press ENTER. The unit returns the previously programmed display with the selected channel now displaying the set point minus the Riemann Sum of the received transducer signal.
10.3.3. Count Continuously
The TOTALIZER will be set to count from zero (0) to the maximum ability of the display. No flags or alarms will be set.
While the SETUP/CAL menu is displayed, choose TOTAL by pressing the number “2” key and ENTER. A TOTALIZER menu is displayed. Press the number “3” key to choose the CONT option followed by the ENTER key. The previously programmed display returns with the selected channel now displaying the Riemann Sum of the received transducer signal.
10.4. EXTERNAL COMMUNICATION SETUP
The following text describes how the POWER communication. Instruction structure and syntax is covered in the section, Serial Protocol.
POD
-400 can be setup to respond to serial and network
10.4.1. RS-232 Settings
From SETUP/CAL Menu
After selecting SETUP/CAL, select any channel number and hit ENTER, choose option number 3 via the numbered key pad. The first menu to be displayed is the BAUD menu. To select the desired baud rate press the number key 1 or 2 respectively. Eight bit bytes, no parity, one stop bit and no handshaking are automatically set. The preprogrammed display is returned upon completion.
10.4.2. RS-485 Settings
From SETUP/CAL Menu
After selecting SETUP/CAL, select any channel number and hit ENTER, choose option number 3 via the numbered key pad. The BAUD menu will be displayed. To select the desired baud rate press the corresponding number key.
164-122012 Power
POD
400 Page 22 of 42
Page 23
The default address as programmed at the factory is 01. If a different address is required, it may be changed using either RS-232 or 485. After making the proper serial connections, use the command *00X to read the default address. Use *00Xdd, where dd = the new address. If the old address is known, use DDXdd, where DD is the old address and dd is the new address. See the section on "Serial Communication: Commands" for more on command structure.
All commands listed in Appendix B are available for RS-485 use by inserting an asterisk and the unit's address in front of the command.
Ex) Set the units of measure on channel two (2) to %.
*10UM203Cr
The command, *00XCr can be used to poll the unit for its current address setting using either RS-232 or RS-485.
10.5. SELECTING THE ANALOG SIGNAL LEVEL
The POWER selectable. The following diagram applies no matter which level of operation is chosen.
From SETUP/CAL Menu
POD
-400 is capable of 0 to 5 V, 0 to 10 V or 4 to 20 mA operation. Each level is user
10.5.1. Zero to Five Volt Operation
While in the SETUP/CAL menu, select ANALOG by pressing the number 4 key and the ENTER key. From the ANALOG SIGNAL menu, choose 0V – 5V by pressing the number 1 key and pressing the ENTER key. The normal display returns with 0 volts intended to correspond to 0% signal and 5 volts intended to correspond to the 100% signal. Calibration may be required.
10.5.2. Zero to Ten Volt Operation
While in the SETUP/CAL menu, select ANALOG by pressing the number 4 key and the ENTER key. From the ANALOG SIGNAL menu, choose 0V – 10V by pressing the number 2 key and pressing the ENTER key. The normal display returns with 0 volts intended to correspond to 0% signal and 10 volts intended to correspond to the 100% signal. Calibration may be required.
10.5.3. Four to Twenty Milliamp Operation
While in the SETUP/CAL menu, select ANLG by pressing the number 4 key and the ENTER key. From the ANALOG SIGNAL menu, choose 4mA – 20 mA operation by pressing the number 3 key and pressing ENTER. The normal display returns with 4 milliamps intended to correspond to 0% signal and 20 milliamps intended to correspond to the 100% signal. Calibration may be required.
164-122012 Power
POD
400 Page 23 of 42
Page 24
10.6. SETTING LIMIT ALARMS
POD
Each channel of the POWER 400 has one high limit alarm and
one low limit alarm. Each alarm is provided by an open-collector, opto-isolated signal at a corresponding pin on the rear panel, connector J8. Use the provided diagram when referring to the text below.
From SETUP/CAL Menu
-
10.6.1. Setting a Single Channel’s High Limit
After selecting a channel in the SETUP/CAL menu, choose ALARMS by pressing the number 5 key followed by the ENTER key. The ALARMS menu will be displayed. Pressing the number “1” key selects the HIGH LIMIT display. Using the number keys, enter the desired high limit level and then press ENTER. EXIT the loop by pressing the number “4” + ENTER key while the ALARMS menu is displayed.
10.6.2. Setting a Single Channel’s Low Limit
After selecting a channel in the SETUP/CAL menu, choose ALARMS by pressing the number 5 key. The ALARMS menu will be displayed. Pressing the number 2 key selects the LOW LIMIT display. Using the number keys, enter the desired high limit level and then press ENTER. EXIT the loop by pressing the number 4 key while the ALARMS menu is displayed, followed by ENTER.
10.6.3. Setting a Single Channel’s Hysteresis
After selecting a channel in the SETUP/CAL menu, choose ALARMS by pressing the number 5 key. The ALARMS menu will be displayed. Pressing the number 3 key selects the HYSTERESIS display. Using the number keys, enter the desired degree of Hysteresis followed by the ENTER key. EXIT the loop by pressing the number 4 key while the ALARMS menu is displayed, followed by ENTER.
164-122012 Power
POD
400 Page 24 of 42
Page 25
10.7. SETTING RATIO CONTROL PARAMETERS
Enabling Ratio Control
Ratio control is achieved through the SETUP/CAL menu. Ratio Control is activated by selecting channel number one (1) and enabling it as master. After enabling channel one, the operator may assign the channels that will follow its signal by some factor.
Press the MODE key and then 3 + ENTER. Select channel one by pressing 1 + ENTER. Choose
RATIO by pressing the 6 key + ENTER. Channel one (1) is enabled as the master when you press 2 + ENTER.
After enabling RATIO control, a series of menus labeled SLAVE A, SLAVE B and SLAVE C will be displayed. These menus correspond to channels 2, 3 and 4 respectively. Each of the channels, 2 through 4, can be enabled or disabled individually. For example, pressing the 1 key + ENTER while SLAVE A is being displayed will eliminate channel two (2) from ratio control and allow channel two to work normally as an independent channel. Pressing the number 2 key + ENTER will make channel 2 the first SLAVE and an S will appear in the first column of line two on the display. Note that, if EXIT is selected at any time, the previous settings remain unchanged and the display is returned to its previously programmed configuration.
164-122012 Power
POD
400 Page 25 of 42
Page 26
10.8. CALIBRATING A CHANNEL TO ITS INCOMING SIGNALS
The POWER
POD
-400 is designed to accept the input from almost any transducer that operates in the
ranges of zero to five volts, zero to ten volts or four to twenty milliamps. In most cases, transducers are capable of having their minimum signal and maximum signal adjusted to correspond with the minimum (ZERO) and maximum (SPAN) unit-of-measure that they are designed to reflect.
For example:
A transducer designed to output 0 volts while reading a temperature of 0 degrees centigrade and output 5
POD
volts while reading 100 degrees centigrade is connected to channel one of the POWER
-400.
In actuality, the transducer is sending -0.23 volts at 0 degrees and 5.11 volts at 100 degrees. By zeroing
POD
channel one of the POWER
-400 while reading the low signal from the transducer, the display will be
000.0 when the temperature is 0°C. Like wise, by setting the CAL value to 100.0 while reading the 5.11 volt signal causes the POWER
POD
-400 to display 100.0 when the temperature is actually 100°C.
This procedure will not account for any non-linearity in any transducer’s output. That is, the accuracy/calibration of any transducer must be independently set, checked and verified before connecting to the POWER
POD
-400, or, the transducer/ POWER
POD
-400 combination must be
calibrated together as a system, if the accuracy of the whole system is to be known.
10.8.1. Resetting the Zero Only
While the SETUP/CAL menu is being displayed and after making a selection from the CHANNEL SELECT menu, select CALIBRATE by pressing the number 7 key followed by ENTER, then choose ZERO ONLY by pressing the number 1 key and ENTER. The ZERO menu will be displayed which also
displays the selected channel number and the signal being received. If this signal is an acceptable signal for zero, press the ENTER key to accept it. If the indicated signal is not an acceptable level for reading zero, the operator has an opportunity at this stage to correct any problems while reading the incoming signal before pressing the ENTER key and accepting it. Alternatively, the operator can press ESC to exit the SETUP/CAL mode and return to it later. After accepting the displayed signal level as the zero point, the display returns to its previously programmed settings. See Appendix D.
10.8.2. Resetting the Span Only
After selecting a channel in the CHANNEL SELECT portion of the SETUP/CAL menu, press 7 + ENTER to select CALIBRATE. SPAN ONLY is selected by pressing number 3 + ENTER. The SPAN window is displayed which contains the chosen channel and the incoming signal on that channel.
If the displayed signal is an acceptable full range value, press ENTER to accept it. The previously programmed display returns with the selected channel now displaying the CAL’d value. If the incoming
164-122012 Power
POD
400 Page 26 of 42
Page 27
signal is not an acceptable full range value, the operator has an opportunity to correct any problems at this stage prior to pressing ENTER. Alternatively, the operator can press ESC to exit SETUP/CAL and return later. See Appendix D.
10.8.3. Resetting Zero and Span
Should it be required to zero and set a new span value for a given channel, The menu option is provided that will present the programmer with the zero menu first, followed by the span, or cal, menu. After selecting a channel in the CHANNEL SELECT portion of the SETUP/CAL menu, press the number 7 key to select CALIBRATE. ZERO & SPAN is selected by pressing the number 2 key. The ZERO window is displayed first. It contains the chosen channel and the incoming signal on that channel. If the signal displayed is within an acceptable range for zero, Press the ENTER key to accept this signal as representing a display of zero while monitoring the transducer. If this signal is not appropriate, the user may correct the problem while the window is displayed or press the ESC key and come back at a later time to zero and span.
After setting the zero of the power supply, the SPAN menu window is displayed. If the displayed signal is an acceptable full range value, press ENTER to accept it. The previously programmed display returns with the selected channel now displaying the CAL’d value. If the incoming signal is not an acceptable full range value, the operator try to correct any problems at this stage prior to pressing ENTER or, by pressing the ESC key, they can exit SETUP/CAL and return to perform this procedure at a later date. See Appendix D.
10.8.4. Setting a Multiplication Factor
For convenience, it is possible to leave a channel setup one way and merely multiply the display value by some factor and replace the display with the new value. An example might be to set up a channel to read from zero to one hundred percent and use a multiplication factor to make the display read from zero to 500 SCCM. An infinite number of possibilities are possible.
Once a channel in the CHANNEL SELECT portion of the SETUP/CAL menu has been selected, press the number 7 key to select CALIBRATE. From the CALIBRATE menu, choose MULTIPLIER by pressing the number 4 key. While the MULTIPLIER window is in the display, enter the multiplication factor and press ENTER. The display returns to normal with the selected channel now displaying the product of the normally displayed value and the entered factor.
164-122012 Power
POD
400 Page 27 of 42
Page 28
10.9. SERIAL COMMUNICATION
Follow the instructions in the section, WIRING THE POWER serial communication with the unit. After insuring that the POWER
POD
-400 for cabling and proper pin out for
POD
-400 is wired properly, the unit
must be set up following the instructions in the section entitled, External Communications Setup.
POD
The instruction set for the POWER
-400 can be divided into two different types. They differ only in
that they
1. Write a command only or
2. Write a command and read a response.
Taking liberty with this vernacular, they may be differentiated using the terms COMMAND or QUERY. The entire set of instructions can be found in Appendix B.
10.9.1. Commands
Typical RS-232 command structure is as follows: CMNDn<value>Cr
For example: To change channel three’s high alarm value to 75.00, send A3H075.00Cr.
RS-485 commands require only that an asterisk and the unit’s address precede the RS-232 Command. Using the same example as above with a unit whose address is 10 (default), send *10A3H075.00Cr.
POD
When the POWER shorted to add the required terminating resistors. Remove two machine screws from the rear of the top cover. Slide the cover off. The two jumpers are near the serial communication connectors.
-400 is the terminating unit on a 485 bus, two jumpers, CJ1 and CJ2 can be
10.9.2. Set Point Queries
Queries are used when information from the unit is required. Each query is followed by a response from the POWERPOD-400.
In this case, the query is asking for a programmed value. Specifically, the set point of channel three, which happens to be set at 50.00 of 100.00.
Sent SP3Cr Response SP3 050.00
10.9.3. Alarm/Flag Queries
The state of the high and low flow alarms for each channel can be monitored via the DB-15 connecter (J8) on the rear panel for immediate action from an alarm situation but the Totalizer set points can only be monitored by reading a memory location’s Boolean setting via serial communication.
As shown in the section, Setting Limit Alarms, to simply QUERY channel three’s high alarm set point, send A3HCr. Expect a response similar to A3H 075.00. The actual state of the alarm can be polled using the STATUS query.
STCr
Which results in a response similar to
STATUS
OCA : CH1 AUTO CH2 CLOSED CH3 OPEN CH4 AUTO
HI/LO: 0/0 0/1 1/0 0/0
OCA is short for OPEN/CLOSED/AUTO.
HI/LO stands for the high flow limit alarm or the low flow limit alarm. One must parse the states if digital tests are to be conducted.
164-122012 Power
POD
400 Page 28 of 42
Page 29
To query the status of the totalizer flag on channel four (4), use the following command.
TF4Cr
The response will be either TF4 0 corresponding to a Boolean ‘FALSE’ indication that the totalizer flag has not been set, or TF4 1, a Boolean ‘TRUE’, indicating that the total is equal to or beyond the totalizer set point..
D41Cr command tells channel four (4) to display the TOTAL. It is not necessary that the front panel read the total for the TOTALIZER to work. The Totalizer works in the background no matter what mode the display is in.
NOTE: It must be emphasized that, if the total reaches the maximum of the display (±999999), the unit effectively stops counting.
164-122012 Power
POD
400 Page 29 of 42
Page 30
Pod
11.0 Appendix A
Power
Command Name Query Response
Channel 1 Display C1 CH1 ddd.dd U of M GasID Channel 2 Display C2 CH2 ddd.dd U of M GasID Channel 3 Display C3 CH3 ddd.dd U of M GasID Channel 4 Display C4 CH4 ddd.dd U of M GasID
All Channels’ Display C5
SP1<dd.ddd> Set Point - Channel 1 SP1 SP1 ddd.dd notes 1, 2 SP2<dd.ddd> Set Point - Channel 2 SP2 SP2 ddd.dd SP3<dd.ddd > Set Point - Channel 3 SP3 SP3 ddd.dd SP4<dd.ddd > Set Point - Channel 4 SP4 SP4 ddd.dd
A1H<dd.ddd > Alarm Set Point, Ch1, High A1H A1H ddd.dd notes 1, 2 A1L<dd.ddd > Alarm Set Point , Ch1, Low A1L A1L ddd.dd A2H<dd.ddd > Alarm Set Point , Ch2, High A2H A2H ddd.dd A2L<dd.ddd > Alarm Set Point , Ch2, Low A2L A2L ddd.dd A3H<dd.ddd > Alarm Set Point , Ch3, High A3H A3H ddd.dd A3L<dd.ddd > Alarm Set Point , Ch3, Low A3L A3L ddd.dd A4H<dd.ddd > Alarm Set Point , Ch4, High A4H A4H ddd.dd A4L<dd.ddd > Alarm Set Point , Ch4, Low A4L A4L ddd.dd
HY1<ddd> Hysteresis, Ch1 HY1 HY1 ddd HY2<ddd> Hysteresis, Ch2 HY2 HY2 ddd HY3<ddd> Hysteresis, Ch3 HY3 HY3 ddd HY4<ddd> Hysteresis, Ch4 HY4 HY4 ddd
UM1<dd> Unit of Meas, Ch1 UM1 UM1 dd UM2<dd> Unit of Meas, Ch2 UM2 UM2 dd UM3<dd> Unit of Meas, Ch3 UM3 UM3 dd UM4<dd> Unit of Meas, Ch4 UM4 UM4 dd
GS1<ddd > Gas ID, CH1 GS1 GS1 ddd GS2<ddd > Gas ID, CH2 GS2 GS2 ddd GS3<ddd > Gas ID, CH3 GS3 GS3 ddd GS4<ddd > Gas ID, CH4 GS4 GS4 ddd
IN1<d> I/O Select, Ch1 IN1 IN2<d> I/O Select, Ch2 IN2 IN3<d> I/O Select, Ch3 IN3 IN4<d> I/O Select, Ch4 IN4
FL1<d> Filter Setting, Ch1 FL1 FL1 d FL2<d> Filter Setting, Ch2 FL2 FL2 d FL3<d> Filter Setting, Ch3 FL3 FL3 d FL4<d> Filter Setting, Ch4 FL4 FL4 d
ML1<d.dddd> Multiplier, CH1 ML 1 ML 1 dddd.d notes 1, 2 ML2<dd.ddd> Multiplier , CH2 ML 2 ML 2 ddd.dd ML 3<ddd.dd> Multiplier , CH3 ML 3 ML 3 dd.ddd ML 4<dddd.d> Multiplier , CH4 ML 4 ML 4 d.dddd
-400 Serial Commands
CH1 ddd.dd U of M GasID CH2 ddd.dd U of M GasID CH3 ddd.dd U of M GasID CH4 ddd.dd U of M GasID
IN1 d
rz - r
IN2 d
rz - r
IN3 d
rz - r rz - r
IN4 d
fs
fs fs fs
notes 6, 7
164-122012 Power
POD
400 Page 30 of 42
Page 31
APPENDIX A (Cont.)
Pod
Power
Command Name Query Response
D1<d > Ch1 Disp Mode D1 d note 5 D2<d> Ch2 Disp Mode D2 d D3<d > Ch3 Disp Mode D3 d D4<d > Ch4 Disp Mode D4 d
T1S<dddddd> Totalizer Set Point, Ch1 T1S T1S dddddd T2S<dddddd> Totalizer Set Point, Ch2 T2S T2S dddddd T3S<dddddd> Totalizer Set Point, Ch3 T3S T3S dddddd T4S<dddddd> Totalizer Set Point, Ch4 T4S T4S dddddd
T1M<d> Totalizer Mode, Ch1 T1M T1M d note 8 T2M<d> Totalizer Mode, Ch2 T2M T2M d T3M<d> Totalizer Mode, Ch3 T3M T3M d T4M<d> Totalizer Mode, Ch4 T4M T4M d
T1R Totalizer Reset, Ch1 note 8 T2R Totalizer Reset, Ch2 T3R Totalizer Reset, Ch3 T4R Totalizer Reset, Ch4
Totalizer Flag, Ch1 TF1 TF1 d note 9 Totalizer Flag, Ch2 TF2 TF2 d Totalizer Flag, Ch3 TF3 TF3 d Totalizer Flag, Ch4 TF4 TF4 d
SN1<d.dddd> Cal Value, CH1 SN1 SN1 d.dddd notes 1, 2 SN2<dd.ddd> Cal Value, CH2 SN2 SN2 d.dddd SN3<ddd.dd> Cal Value, CH3 SN3 SN3 d.dddd SN4<dddd.d> Cal Value, CH4 SN4 SN4 d.dddd
Z1 Zero Ch1 Z2 Zero Ch2 Z3 Zero Ch3 Z4 Zero Ch4
F1 Set Full Scale Value, Ch1 F2 Set Full Scale Value, Ch2 F3 Set Full Scale Value, Ch3 F4 Set Full Scale Value, Ch4
Status, All Channels ST
R1<d> Ratio Mode, Ch1 R1 R1 d note 11 R2<d> Ratio Mode, Ch2 R2 R2 d R3<d> Ratio Mode, Ch3 R3 R3 d R4<d> Ratio Mode, Ch4 R4 R4 d
BR<d> Baud Rate BR BR d note 12 RE<d> Front Panel Lock Out RE REn: REMOTE/LOCAL note 13 *00X<dd> Multi-drop Address *00X MULTIDROP ADDRESS: dd
-400 Serial Commands (Cont'd)
OCA: CH1 text CH2 text Note 10 CH3 text CH4 text HI/LO: CH1 d/d CH2 d/d CH3 d/d CH4 d/d
164-122012 Power
POD
400 Page 31 of 42
Page 32
APPENDIX A (Notes:)
1. All returned values will include decimal points wherever unit has been programmed to display them.
2. All Commands needing decimal points must include them wherever they are intended to be displayed.
3. d = decimal digit (ASCII)
4. n = Channel Number
5.
⎪ ⎪
⎨ ⎪ ⎪
= = =
,3
6. rz = range, zero; . r
#:
channelselectedBlanksd
= range, full scale
fs
MassUnitsTotalDisplaysd
RateUnitsFlowDisplaysd
⎫ ⎪
)(,1
⎪ ⎬
)(,2
⎪ ⎪
=
ChnlnDnd
=
ChnlnINnd
7.
8.
9.
⎧ ⎪
⎪ ⎨
⎪ ⎪
⎧ ⎪
⎪ ⎨
⎪ ⎪
=
⎪ ⎨
⎪ ⎩
#:
=
VDCd
50,1
=
VDCd
100,2
=
mADCd
204,3
=
ChnlnTnMd
#:
= =
=
ChnlnTFnd
#:
<=
=
⎫ ⎪
⎪ ⎬
⎪ ⎪
Re,,1
ZerotosetsUpCountsTotalizerd
PoSettosetsDownCountsTotalizerd
Re,,3
⎪ ⎪
intRe,,2
⎪ ⎪
ZerotosetslyContinuousCountsTotalizerd
⎫ ⎪
PoSetTotald
PoSetTotald
int,0
⎬ ⎪
int,1
10. text = “OPEN” or “CLOSE” d = 0 or 1
#:
ChnlnRnd
=
=
==19200,2
==REMOTEd
⎫ ⎪
1
Enabledd
==
,2
Disabledd
9600,1:ddBRd
⎬ ⎪
⎫ ⎬ ⎭
LOCALdREd
,1:
,2
⎫ ⎬ ⎭
11.
12.
13.
⎧ ⎪
⎨ ⎪
⎧ ⎨ ⎩
⎧ ⎨ ⎩
164-122012 Power
POD
400 Page 32 of 42
Page 33
12.0 Appendix B 13.0 Appendix C
# Name Rate Total # Name Rate Total
1 2 3 4 5 6 7 8
9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34
Units-of-Measure for Meter Reading & Corresponding Totalizer Units
Standard Cubic Centimeters per Minute SCCM SCC Standard Liters per Minute SLM SL Percent % Volts V Millivolts MV Counts CNT Normal Liters per Minute NLM NL Standard Liters per Second SLS SL Normal Liters per Second NLS NL Standard Liters per Hour SLH S L Normal liters per Hour NLH NL Standard Mililiters per Minute SMLM SML Normal Mililiters per Minute NMLM NML Standard Mililiters per Second SMLS SML Normal Mililters per Second NMLS NML Standard Mililiters per Hour SMLH SML Normal Mililiters per Hour NMLH NML Normal Cubic Centimet ers per Minute NCCM NCC Standard Cubic Centimeters per Second SCCS SCC Normal Cubic Centimeters per Second NCCS NCC Standard Cubic Centimeters per Hour SCCH SCC Normal Cubic Centimet ers per Hour NCCH NCC Standard Cubic Feet per Minute SCFM SCF Normal Cubic Feet per Minute NCFM NCF Standard Cubic Feet per Second SCFS SCF Normal Cubic Feet per Second NCFS NCF Standard Cubic Feet per Hour SCFH SCF Normal Cubic Feet per Hour NCFH NCF Standard Cubic Meters per Minute SCMM SCM Normal Cubic Meters per Minute NCMM NCM Standard Cubic Meters per Second SCMS SCM Normal Cubic Meters per Second NCMS NCM Standard Cubic Meters per Hour SCMH SCM Normal Cubic Meters per Hour NCMH NCM
Standard Cubic Inches per Minute SCIM SCI
35
Normal Cubic Inches per Minute NCIM NCI
36
Standard Cubic Inches per Second SCIS SCI
37
Normal Cubic Inches per Second NCIS NCI
38
Standard Cubic Inches per Hour S CI H SCI
39
Normal Cubic Inches per Hour NCIH NCI
40
Pounds per Minute LBM LB
41
Pounds per Second LBS LB
42
Pounds per Hour LBH LB
43
Kilograms per Minute KgM Kg
44
Kilograms per Second KgS Kg
45
Kilograms per Hour KgH Kg
46
Grams per Minute GRM GR
47
Grams per Second GMS GR
48
Grams per Hour GRH GR
49
Moles per Minute MolM Mol
50
Moles per Second MolS Mol
51
Moles per Hou r MolH Mol
52
Kilomoles per Minute KMolM KMol
53
Kilomoles per Second KMolS KMol
54
Kilomoles per Hour KMolH KMol
55
Watts W
56
Bits per Second BPS Bits
57
Seconds Sec
58
Minutes Min
59
Hours Hrs
60
Watt * Hours WH W
61
Torr Torr
62
Bar Bar
63
Pascals Pa
64
Inches of Water inH2O
65
Pounds per Square Inch, Absolute PSIA
66
Pounds per Square Inch, Gage PSIG
67
164-122012 Power
POD
400 Page 33 of 42
Page 34
Gas ID Table (Page 1 of 2)
#
1 Acetic Acid 1 46 Diethyl Ether 46 91 Hydrogen Cyanide CHN
Acetic Acid, Anhydride 2
2
3 Acetone C
Acetonitryl C2H3N
4
5 Acetylene C
Air Air
6
7 Allene C
Ammonia NH3
8
9 Argon Ar 54 Ethane, 1-chloro-1,1,2,2-tetrafluoro- 54 99 Isopentane C
Arsine AsH3
10
11 Benzene C
Boron Trichloride BCl3
12
13 Boron Triflouride BF
Bromine Br2
14
15 Bromochlorodifluoromethane 15 60 Ethyl Bromide 60 105 Methanol CH
Bromodifluoromethane 16
16
17 Bromotrifluormethane CBrF
Butane C4H10
18
19 Butanol 19 64 Ethylene Dibromide 64 109 Methyl Bromide CH
Butene C4H8
20
21 Carbon Di oxide CO
Carbon Disulfide CS2
22
23 Carbon Monoxide CO 68 Ethylidene Dichloride 68 113 Methyl Ethyl Ether C
Carbon Tetrachloride CCl4
24
25 Carbonyl Sulfide COS 70 Fluorine F
Chlorine Cl2
26
27 Chlorine Trifluoride ClF
Chlorobenzene 28
28
29 Chlorodifluoroethane 29 74 Freon 13 CClF
Chloroform CHCl3
30
31 Chloropentafluoroethane 31 76 Freon 22 76 121 Neon Ne
Chloropropane 32
32
33 Cisbutene C
Cyanogen C2N2
34
35 Cyanogen Chloride ClCN 80 Helium He 125 Nitrogen Tetroxide N
Cyclobutane C4H8
36
37 Cyclopropane C
Deuterium H22
38
39 Diborane B2H6 84 Hexane C
Dibromodifluoromethane 40
40
41 R21 R21 86 Hexene C
Dichloromethane 42
42
43 Dichloropropane 43 88 Hydrogen H
Dichlorosilane 44
44
45 Diethyl Amine 45 90 Hydrogen Chloride HCl 135 Pentaborane B
GAS NAME Symbol # GAS NA ME Symbol # GAS NAME Symbol
Diethyl Sulfide 47
47
O 48 Difluoroethylene 48 93 Hydrogen Iodide HI
3H6
Dimethylamine C2H7N
49
50 Dimethyl Ether C2H6O 95 Hydrogen Sulfide H2S
2H2
Dimethyl Sulfide C2H6S
51
52 Divinyl C4H6 97 Isobutanol 97
3H4
Ethane C2H6
53
Ethane, 1-chloro-1,2,2,2-tetrafluoro- 55
55
56 Ethanol C2H6O 101 Isoxazole 101
6H6
Ethylacetylene C4H6
57
58 Ethyl Amine C2H7N 103 Krypton Kr
3
Ethylbenzene C8H10
59
Ethyl Chloride 61
61
62 Ethyl Fluoride C2H5F 107 Methyl Acetylene C3H4
3
Ethylene C2H4
63
Ethylene Dichloride 65
65
66 Ethylene Oxide C2H4O 111 Methylcyclohexane C7H14
2
Ethyleneimine C2H4N
67
Ethyl Mercaptan C2H6S
69
115 Methyl Fluoride CH3F
2
Formaldehyde CH2O
71
72 Freon 11 CCl3F 117 Methyl Iodide CH3I
3
Freon 12 73
73
119 Methylpentene C6H12
3
Freon 14 CF4
75
Freon 23 CHF3
77
78 Freon 114 78 123 Nitrogen N2
4H8
Furan C4H4O
79
Heptafluoropropane C3HF7
81
82 HMDS HMDS 127 Nitromethane 127
3H6
Hexamethyldisiloxane 83
83
129 Nitrous Oxide N2O
6H14
Hexafluorobenzene C6F6
85
131 Octane C8H18
6H12
Hydrazine N2H4
87
133 Oxygen Difluoride F2O
2
Hydrogen Bromide HBr
89
Hydrogen Fluoride HF
92
Hydrogen Selenide H2Se
94
Isobutane C4H10
96
Isobutene C4H8
98
Isopropyl Alcohol C3H8O
100
Ketene C2H2O
102
Methane CH4
104
Methyl Acetate 106
106
Methylamine CH5N
108
Methyl Chloride CH3Cl
110
Methyl Ethyl Amine C3H9N
112
Methyl Ethyl Sulfide C3H8S
114
Methyl Formate 116
116
Methyl Mercaptan CH4S
118
Methyl Vinyl Ether C3H6O
120
Nitric Oxide NO
122
Nitrogen Dioxide NO2
124
Nitrogen Trifluoride NF3
126
Nitrosyl Chloride NOCl
128
n-Pentane C5H12
130
Oxygen O2
132
Ozone O3
134
5H12
3H8
2O4
5H9
3
4
Br
O
O
164-122012 Power
POD
400 Page 34 of 42
Page 35
Gas ID Table (Page 2 of 2)
136 Pentane C
137
138 Perfluorocyclobutane C
139
140 Perfluoropropane C
141
142 Phosgene COCl
143
144 Phosphorus Trifluoride PF
145
146 Propyl Alcohol C
147
148 Propylene C
149
150 R32 CH
151
152 R123A R123A 172 Tetrahydrofuran C
153
154 R134 R134 174 Thiophene C
155
GAS NAME Symbol GAS NAME Symbol GAS NAME Symbol
156 R143 R143 175 Toluene C7H8
5H12
Perchloryl Fluoride ClFO3
158 R152A R152A 177 Trichloroethane 177
4F8
R116 C2F6
160 R1416 R1416 179 R113 R113
3F8
Phenol C6H6O
162 Sec-butanol 162 181 Trimethyl Amine C3H9N
2
Phosphine PH3
164 Silicone Tetrafluoride SiF4 183 Uranium Hexafluoride UF6
3
Propane C3H8
O 166 Sulfur Hexafluoride SF6 185 Vinyl Chloride 185
3H8
Propyl Amine C3H9N
168 Sulfur Trifluoride SF3 187 Water Vapor H2O
3H6
Pyradine C5H5N
170 Tetrachloroethylene 170 189 Xylene, m- C8H10
2F2
R123 R123
R125 C2HF5
R134A R134A
R143A R143A
157
R218 C3F8
159
Radon Rn
161
Silane SiH4
163
Sulfur Dioxide SO2
165
Sulfur Tetrafluoride SF4
167
Sulfur Trioxide SO3
169
Tetrafluoroethylene C2F4
171
Tert-butanol 173
173
Transbutene C4H8
176
Trichloroethylene 178
178
Triethylamine 180
180
Tungsten Hexafluoride WF6
182
Vinyl Bromide 184
184
Vinyl Flouride C2H3F
186
Xenon Xe
188
Xylene, o- C8H10
190
O 191 Xylene, p- C8H10
4H8
Mixtures MIXT
191
S 193 Other ????
4H4
194
164-122012 Power
POD
400 Page 35 of 42
Page 36
14.0 Appendix D
Setting the Zero & Span
on the
Power
All procedures outlined in this document must be performed with the power supply turned on and warmed up for at least one hour.
All procedures outlined in this document must be performed with NO transducers connected to the channel being adjusted*.
You CANNOT successfully zero a channel without, first, supplying zero volts or four milliamps to the meter input for that channel.
You CANNOT successfully span a channel without, first, supplying the meter input for that channel with its appropriate span voltage or span current.
14.1 Zeroing Unit (Voltage Range).
The following instructions explain how to short a given channel’s input signal to ground (0 volts DC) and set the display to read zero at this voltage.
Locate the 15-pin, ‘D’ type connector that corresponds to the channel to be zeroed.
Short pin 5 (signal common) to pin 6 (signal in).
Press MODE Press 3 Press ENTER Press 1, 2, 3 or 4, depending on which channel is to be zeroed. Press ENTER Press 7 Press ENTER Press 1 Press ENTER Press ENTER again
You have just told a channel to display zero when the input signal is zero volts
Pod
-400 Power Supply/Totalizer
J1 = Channel 1 J2 = Channel 2 J3 = Channel 3 J4 = Channel 4
!
.
164-122012 Power
POD
400 Page 36 of 42
Page 37
14.2 Zeroing Unit (4 to 20 mA Range)
The following instructions explain how to use the POWER
POD
-400 command signal to supply
4mAmp signal to a channel’s input and set the display to read zero at this current level.
Disconnect all connectors from the channel to be calibrated. Insure that the channel is set to read 4 to 20 mA signals by performing the following steps. Press MODE Press 3 Press ENTER Press 1, 2, 3 or 4, depending on which channel is to be zeroed. Press ENTER Press 4 Press ENTER Press 3 Press ENTER The normal display returns and will respond in the 4 to 20 mA mode.
Connect a milliamp meter between pins 14 and 6 on the 15-pin ‘D’ connector corresponding to the channel to be calibrated.
J1 = Channel 1 J2 = Channel 2 J3 = Channel 3 J4 = Channel 4
Set the command for the selected channel to zero. Press the CHANNEL # button Press 1, 2, 3 or 4, depending on which channel is to be zeroed. Press 0.0 using the numbered key pad Press ENTER The milliamp meter should be reading between 3.996 and 4.004 mAmps.
To Zero the meter with a signal of 4mA, perform the following steps. Press MODE Press 3 Press ENTER Press 1, 2, 3 or 4, depending on which channel is to be zeroed. Press ENTER Press 7 Press ENTER Press 1 Press ENTER Press ENTER again
You have just told a channel to display zero when the input signal is 4mA
!
.
164-122012 Power
POD
400 Page 37 of 42
Page 38
14.3 Spanning Unit.
The following instructions explain how to use the POWER
POD
-400 command signal to supply a given
channel’s input with the proper span voltage and to set the display to read a transducer’s span value.
You must know four things prior to setting the display’s span value:
1. You must know the maximum signal level for the specific transducer connected to the
channel to be spanned (5 VDC, 10VDC or 20mA).
2. You must know the maximum value to be displayed when the maximum signal level for the
transducer is connected to the channel to be spanned.
3. You must insure that the channel’s multiplier is set to 1.0000.
4. You must know the channel’s ‘Span Value’. That is, the value that would NOW be
displayed (prior to any adjustment) if the maximum signal was fed into the input of the channel to be adjusted.
1 and 2. If this information is not on the transducer itself, consult the manufacturer of the transducer. Their literature or other documentation should specify the maximum output signal from the transducer and the maximum value to be displayed at that signal level.
3. Setting the MULTIPLIER to 1.0000.
Press MODE Press 3 Press ENTER Press 1, 2, 3 or 4 depending on which channel is to be adjusted. Press ENTER Press 7 Press ENTER Press 4 Press ENTER Press 1.0000 using the numeric key pad. Press ENTER
You have just set a channel’s multiplier to 1.0000.
4. If you do not know what the selected channel’s maximum display (Span) value is currently set
to, follow these steps.
Notice that the last step in this exercise is to press ESC, not ENTER.
Press MODE Press 3 Press ENTER Press 1
, 2, 3 or 4 depending on which channel is to be adjusted. Press ENTER Press 7 Press ENTER Press 3 Press ENTER Read the VALUE from the display.
Press ESC
164-122012 Power
POD
400 Page 38 of 42
Page 39
An example of one possible transducer/ POWER
0
TRANSDUCER POWER
Max signal out
5 VDC 250.00 1.000
Setting the THPS-400 Analog level to correspond with the transducer to be attached.
Max value to be
displayed
POD
-400 combination.
POD
-400
Multiplier Current Span Value
100.00
Press MODE Press 3 Press ENTER Press 1, 2, 3 or 4 depending on which channel is to be adjusted. Press ENTER Press 7 Press ENTER Press 1 for 0 – 5 VDC, 2 for 0 – 10 VDC or 3 for 4 – 20 mADC Press ENTER
In the example listed above, you would have selected option number 1 for the 0 – 5 volt analog signal level since the maximum signal from the transducer is 5 VDC.
You have just set the analog operating level (range) for the selected channel.
Setting the display to read the maximum transducer value at the maximum transducer input.
Locate the 15-pin, ‘D’ type connector that corresponds to the channel to be spanned.
J1 = Channel 1 J2 = Channel 2 J3 = Channel 3 J4 = Channel 4
For units set to meter a voltage range, short pin 14 (command out) to pin 6 (signal in) of the selected channel and place a calibrated volt meter to read from pin 5 (common) to pin 6. For units set to read a 4 to 20 mA range, connect a milliamp meter between pins 14 and 6. Press the CHANNEL # key corresponding to the selected channel. An asterisk (*) appears. Use the number keys to enter THPS-400 current span value. Use the decimal! Press ENTER
Adjust the command to achieve a value as close to the max analog signal level as possible (±0.01 volts for the 5 volt range, ±0.02 volts for the 10 volt range and ±0.004 ampsfor the milliamp range.
Do this using the CHANNEL# key, the numeric key pad and the ENTER key until the desired
#
signal is established
. Press MODE Press 3 Press ENTER Press 1, 2, 3 or 4 depending on which channel is to be adjusted. Press ENTER Press 3 Press ENTER Use the numeric keys to enter the new SPAN value (Max transducer value to be displayed). Press ENTER Press ENTER
164-122012 Power
POD
400 Page 39 of 42
Page 40
Congratulations! You have just calibrated your Power
Pod
-400 to read zero at zero volts and the transducer’s max
display value at the transducer’s max signal input.
The only thing left to do is to insure that your transducer is calibrated and that it is wired correctly to the Power
Pod
-400.
* Individual channels may be calibrated with calibrated transducers connected, breakout
connectors (not supplied) and precision multi-meters with NIST traceable calibrations (also not supplied). Such procedures are not covered in this document.
! If the calibration for the channel just zeroed had been severely compromised prior to beginning
this procedure, zero may not be immediately displayed and you may have to repeat the procedure after resetting the span. Check that the span has been set correctly!
# If a signal level within .1% of the maximum DC voltage level is not attainable, the power supply
may need to be repaired.
164-122012 Power
POD
400 Page 40 of 42
Page 41
15.0 Drawings
4
P
ower
3
CLOSE
CLOSE
OPEN
1
2
CLOSE
CLOSE
OPEN
OPEN
OPEN
Pod
400
HASTINGS
7
825
936
4
1
0
.
164-122012 Power
POD
400 Page 41 of 42
Page 42
16.0 Warranty
16.1. Warranty Repair Policy
Hastings Instruments warrants this product for a period of one year from the date of shipment to be free from defects in material and workmanship. This warranty does not apply to defects or failures resulting from unauthorized modification, misuse or mishandling of the product. This warranty does not apply to batteries or other expendable parts, nor to damage caused by leaking batteries or any similar occurrence. This warranty does not apply to any instrument which has had a tamper seal removed or broken.
This warranty is in lieu of all other warranties, expressed or implied, including any implied warranty as to fitness for a particular use. Hastings Instruments shall not be liable for any indirect or consequential damages.
Hastings Instruments, will, at its option, repair, replace or refund the selling price of the product if Hastings Instruments determines, in good faith, that it is defective in materials or workmanship during the warranty period. Defective instruments should be returned to Hastings Instruments, shipment prepaid, together with a written statement of the problem and a Return Material Authorization (RMA) number. Please consult the factory for your RMA number before returning any product for repair. Collect freight will not be accepted.
16.2. Non-Warranty Repair Policy
Any product returned for a non-warranty repair must be accompanied by a purchase order, RMA form and a written description of the problem with the instrument. If the repair cost is higher, you will be contacted for authorization before we proceed with any repairs. If you then choose not to have the product repaired, a minimum will be charged to cover the processing and inspection. Please consult the factory for your RMA number before returning any product repair.
TELEDYNE HASTINGS INSTRUMENTS
804 NEWCOMBE AVENUE
HAMPTON, VIRGINIA 23669 U.S.A.
ATTENTION: REPAIR DEPARTMENT
TELEPHONE (757) 723-6531
1-800-950-2468
FAX (757) 723-3925
E MAIL mail to:hastings_instruments@teledyne.com
INTERNET ADDRESS http://www.teledyne-hi.com/
Repair Forms may be obtained from the “Information Request” section of the Hastings Instruments
164-122012 Power
POD
400 Page 42 of 42
Loading...