All Rights Reserved. No part of this manual may be reproduced, transmitted,
transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in part, in any form or by any means, whether it be electronic, mechanical, magnetic, optical, manual, or otherwise, without the prior written
consent of Teledyne Brown Engineering Analytical Instruments, 16830 Chestnut Street,
City of Industry, CA 91749-1580.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us
free from defects of material and of construction, and that our liability shall be limited
to replacing or repairing at our factory (without charge, except for transportation), or at
customer plant at our option, any material or construction in which defects become
apparent within one year from the date of shipment, except in cases where quotations or
acknowledgements provide for a shorter period. Components manufactured by others
bear the warranty of their manufacturer. This warranty does not cover defects caused by
wear, accident, misuse, neglect or repairs other than those performed by Teledyne or an
authorized service center. We assume no liability for direct or indirect damages of any
kind and the purchaser by the acceptance of the equipment will assume all liability for
any damage which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our
apparatus, and to make any alterations in the dimensions, shape or weight of any parts,
in so far as such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its user, and serves as a tool by
which valuable data can be gathered. The information provided by the instrument may
assist the user in eliminating potential hazards caused by his process; however, it is
essential that all personnel involved in the use of the instrument or its interface, with the
process being measured, be properly trained in the process itself, as well as all instrumentation related to it.
The safety of personnel is ultimately the responsibility of those who control
process conditions. While this instrument may be able to provide early warning of
imminent danger, it has no control over process conditions, and it can be misused. In
particular, any alarm or control systems installed must be tested and understood, both as
to how they operate and as to how they can be defeated. Any safeguards required such
as locks, labels, or redundancy, must be provided by the user or specifically requested of
Teledyne at the time the order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions.
The purchaser is responsible for the training of personnel, for providing hazard warning
methods and instrumentation per the appropriate standards, and for ensuring that hazard
warning devices and instrumentation are maintained and operated properly.
Teledyne Electronic Technologies/Analytical Instruments (TET/AI), the
manufacturer of this instrument, cannot accept responsibility for conditions beyond its
knowledge and control. No statement expressed or implied by this document or any
information disseminated by the manufacturer or its agents, is to be construed as a
warranty of adequate safety control under the user’s process conditions.
Material Safety Data Sheets............................................... MSDS-1
vi
Teledyne Electronic Technologies
Analytical Instruments
IntroductionIntroduction
Introduction
IntroductionIntroduction
Overview
The Teledyne Electronic Technologies/Analytical Instruments (TET/AI) Model
3060E Ultra Trace Oxygen Analyzer is a highly sophisticated microprocessorbased instrument designed to measure the oxygen concentration in a gas
mixture in a range as low as 0-50 parts per billion (PPB). The 3060E accurately
measures and analyzes the oxygen content in inert gases such as helium,
nitrogen and argon, and in flammable gases such as hydrogen and ethylene. The
analyzer is easy to use through LCD menu displays with five touch switches for
operator interface.
Ultra Ultra
Ultra
Ultra Ultra
Introduction
483 mm
TT
race Oxygrace Oxyg
T
race Oxyg
TT
race Oxygrace Oxyg
en Analen Anal
en Anal
en Analen Anal
yzyz
yz
yzyz
erer
er
erer
SAMPLE
OXYGEN
SCRUBBER
13
CALIBRATOR
6
7
FLOWSET
SPARGER
BYPASS
1
10
2
MV1
3
CALIBRATOR
SAMPLE
SPAN
VENT1 VENT2
BYPASS
5
MV2
HO
2
8
N
2
AR
4
H
2
HE
SELECTOR
9
MV3
CELL
MFC
GAS
TELEDYNE BROWN ENGINEERING
Analytical Instruments
Model 3060E
ULTRA TRACE OXYGEN ANALYZER
SENSOR COMPARTMENT
35
83
51
310
82
35
Figure 1-1: Model 3060E Front Panel
Teledyne Electronic Technologies
Analytical Instruments
1–1
Model 3060EModel 3060E
Model 3060E
Model 3060EModel 3060E
Components
The 3060E is designed to mount in a standard 19-inch rack. All of the
operational controls are easily accessed through the well-marked front panel.
Each of the following components is contained within the analyzer:
•Analysis
Electrochemical Ultra Trace Sensor
Electrochemical Calibrator
Sample System
•Power Supply Module
•Control Module
•LED Display
•Valve Control Board
•Temperature Controller
Chapter 1Chapter 1
Chapter 1
Chapter 1Chapter 1
The external design consists of easily accessible controls and ports. The front
panel consists of:
•LED Display
•LCD Display
•Flow Schematic Display
•Ultra Trace Sensor Access Panel
The back panel ports are:
•Gas Inlet/Outlet Ports
•Electrical Connections
•Water Reservoir Inlet and Drain
Main Features of the Analyzer
The 3060E is highly sophisticated yet simple to use. Useful features help you
to operate the analyzer and also provide you with high quality analysis data.
Below is a list of the main features.
1. Low range (0–50 PPB): This range provides highest resolution and
accuracy for ultra-pure gases. Three other ranges (0–100 PPB, 0–1
PPM, and 0–10 PPM) are also provided to meet your analysis needs.
1–2
2. Linearity on all ranges: This feature allows the user to calibrate the
analyzer on any of the four ranges and then analyze the sample gases
for all ranges.
Teledyne Electronic Technologies
Analytical Instruments
IntroductionIntroduction
Introduction
IntroductionIntroduction
3. AutoRanging with manual override: This feature allows the analyzer
4. Built-in oxygen scrubber: The built-in oxygen scrubber can be used
5. Faradaic calibrator: The built-in Faradaic calibrator generates oxy-
Ultra Ultra
Ultra
Ultra Ultra
to automatically select the appropriate range for a given measurement. A manual override allows the user to “lock in” a specific
range of interest.
to:
•Produce the zero gas needed to set the zero point.
•Feed the zero gas to sparge the electrolyte in the Ultra Trace
sensor.
•Feed the zero gas to the Faradaic calibrator to generate a
span gas for adjusting the span level.
Should very high oxygen levels be present in the sample gas, the
automatic scrubber inlet shut-off feature prolongs the life of the
scrubber.
gen from PPB levels up to 8 PPM, and allows the user to calibrate
the analyzer. A calibrator gauge logs the amount of time the calibrator is used and notifies you through the LCD display if the calibrator
electrolyte needs refilling.
TT
race Oxygrace Oxyg
T
race Oxyg
TT
race Oxygrace Oxyg
en Analen Anal
en Anal
en Analen Anal
yzyz
yz
yzyz
erer
er
erer
6. Autozero and Autospan: The analyzer can be set to calibrate itself
(zero and span) at pre-programmed intervals.
7. Span gas port: The analyzer has separate sample and span gas ports,
which allow the installation of an external source of span gas for
calibration without interfering with the sample gas line.
8. A built-in water/electrolyte level detector in the Ultra Trace sensor
signals an automatic water feed to the sensor when the electrolyte
falls below a minimum level.
9. Insensitivity to minor flow changes and mechanical vibrations: The
analyzer output is not affected by minor changes in the flow rate.
10. Built-in correction factors for various gases: Built-in correction
factors for various gases allow you to select nitrogen, argon, helium
or hydrogen as background gases without the need to recalibrate the
mass flow controller.
11. Flow schematic display: This feature allows you to visually verify
the exact status of each of the pneumatic valves in the sampling
system.
Teledyne Electronic Technologies
Analytical Instruments
1–3
Model 3060EModel 3060E
Model 3060E
Model 3060EModel 3060E
12. Meter readouts (the analyzer includes two meter readouts):
13. Minimal maintenance: The electrochemical Ultra Trace sensor in the
analyzer requires virtually no maintenance. Water is automatically
replenished in the Ultra Trace sensor. Replacing the electrolyte is
easily done and does not require removing the Ultra Trace sensor
from the analyzer.
14. Remote capability: The 3060E can be remotely located and controlled from a microcomputer linked to the analyzer by a phone line.
Software provided with the analyzer allows you to control analyzer
functions through menus displayed on the computer.
15. Five direct-reading alarms: Five programmable setpoints and Form
C SPDT relays can be configured to practically any requirement.
Chapter 1Chapter 1
Chapter 1
Chapter 1Chapter 1
·A red LED readout with large, easy-to-read numerals
displaying oxygen concentration.
·An LCD graphic display which includes alphanumeric
information such as alarms, output, help menus, etc.
16. Isolated 4-20 mADC and 0-1 VDC (negative ground) outputs: Four
signal outputs provide for both oxygen measurement and range
identification.
Applications
The analyzer is an invaluable tool in the following applications and industries:
•Analysis of blanketing gases in semiconductor and electronics
industries.
•Measuring the purity of various gases in air separation plants.
•Controlling oxygen for cracking and heating furnaces in petrochemical industries.
•Prevention of oxidation by measuring the purity of blanketing gases
in fiber and glass industries.
•Monitoring and controlling gas atmospheres in the heat treatment of
metals in steel and other metal industries.
•Gas analysis and research in laboratories and research and development areas.
1–4
Teledyne Electronic Technologies
Analytical Instruments
Operational Operational
Operational
Operational Operational
Overview
There are five main analyzer functional groups:
TheorTheor
Theor
TheorTheor
yy
y
yy
Operational Theory
1. Ultra Trace Sensor
2. Faradaic Calibrator
3. Sample System
Ultra Ultra
Ultra
Ultra Ultra
TT
race Oxygrace Oxyg
T
race Oxyg
TT
race Oxygrace Oxyg
en Analen Anal
en Anal
en Analen Anal
yzyz
yz
yzyz
erer
er
erer
4. Electronic Signal Processing
5. Temperature Controller.
The analyzer uses a sophisticated Ultra Trace oxygen (O2) sensor, which is an
electrochemical galvanic device with current output.
An electrochemical calibrator based on Faraday’s law is provided to facilitate
the calibration of the analyzer.
The sample system is designed to optimize the performance of the analyzer.
The components and the methods used for the fabrication of the sample system
assure leak-free transport of gases through the analyzer.
The electronic signal processing unit (control module) is designed to simplify
the operation of the analyzer and accurately process the signal from various
components. The control module incorporates a microprocessor which allows
the operation of the analyzer with a minimum of operator interaction.
A temperature controller regulates the temperature of the Ultra Trace sensor
to minimize the effects of ambient temperature variations during analysis.
Figure 2-1 shows the locations of the major components of the analyzer.
Teledyne Electronic Technologies
Analytical Instruments
2-1
Model 3060EModel 3060E
Model 3060E
Model 3060EModel 3060E
Chapter 2Chapter 2
Chapter 2
Chapter 2Chapter 2
Figure 2-1. Main Components
Temperature
Controller
2-2
Teledyne Electronic Technologies
Analytical Instruments
Operational Operational
Operational
Operational Operational
Ultra Trace Sensor
The analyzer incorporates an electrochemical ultra trace O2 sensor exclusively
designed by TET/AI (U.S. Patent #5,085,760). The sensor shows exceptionally high sensitivity for O2 and remarkable long-term stability. The highly
accurate sensing element of the sensor enables it to detect as low as 0.5 PPB
of O
The components of the sensor are shown in Figure 2-2. The main body of the
sensor is made of clear acrylic. The sensor has a U-shaped profile with an open
top end. The left-hand wall of the sensor has a circular aperture for mounting
the O2-sensing cathode. A high surface area cadmium anode is inserted through
the top stainless steel plate. A fine porous frit disc is mounted at the bottom of
the sensor in order to remove any dissolved O2 by continuously sparging the
electrolyte with O2 free gas. A baffle plate is inserted between the sparger and
the sensing electrode to minimize the noise level caused by the sparger.
TheorTheor
Theor
TheorTheor
in a gas mixture.
2
yy
y
yy
Ultra Ultra
Ultra
Ultra Ultra
Ultra Trace Sensor Components
TT
race Oxygrace Oxyg
T
race Oxyg
TT
race Oxygrace Oxyg
en Analen Anal
en Anal
en Analen Anal
yzyz
yz
yzyz
erer
er
erer
The main body of the sensor contains 15% aqueous KOH electrolyte. A
thermistor is inserted into the sensor’s main body through an opening on the top
stainless steel mounting plate. The opening serves as a port for adding water/
electrolyte to the sensor. An electrolyte level detector is also mounted through
the top stainless steel plate.
WATER LEVEL
DETECTOR
CADMIUM ANODE
O-RING
CATHODE
SAMPLE IN
SAMPLE OUT
PURGE GAS OUT
(-)
THERMISTOR
PURGE GAS IN
TOP STAINLESS
STEEL PLATE
ELECTROLYTE
BAFFLE PLATE
ZERO GAS BUBBLES
(OXYGEN-FREE)
CELL BODY
SIDE STAINLESS
STEEL PLATE
Figure 2-2. Ultra Trace Sensor Components
(+)
CATHODE
Teledyne Electronic Technologies
Analytical Instruments
2-3
Model 3060EModel 3060E
Model 3060E
Model 3060EModel 3060E
The sensing cathode used in the sensor is a high surface-area metal-catalyzed
gas diffusion electrode, with a surface area of 150–180 m2/g. A 0.4 mg/cm
catalyst load provides an effective cathode surface area of up to 600 times the
geometric area of the cathode. This produces a very high signal output per unit
concentration of O2; hence, an excellent signal-to-noise ratio is achieved even
at the highest sensitivity.
The metal-catalyzed gas-diffusion electrode consists of a hydrophobic Tefloncarbon gas diffusion backing layer and a Teflon-carbon metal catalyst layer
bonded together. The thickness of the catalyst layer is approximately one-tenth
of the gas diffusion layer. The overall thickness of the cathode is approximately
0.5 mm.
The cathode is held against the sensor body by a polyethylene ring. The catalyst
layer of the cathode is exposed to the electrolyte phase; the hydrophobic
backing layer is exposed to the gas to be analyzed. The cathode assembly is
sealed by placing a polypropylene O-ring between the stainless steel plate,
containing stainless steel gas inlet and outlet tubes, and the sensor body.
Chapter 2Chapter 2
Chapter 2
Chapter 2Chapter 2
2
The electrolyte in the sensor is continuously purged by bubbling the zero gas
(generated by passing the sample gas through the built-in scrubber) through the
fine porous frit mounted in the inner floor of the sensor. Sparging of the
electrolyte removes most of the dissolved O
from the electrolyte, reducing the
2
background signal.
Ultra Trace Sensor Operation
The gas to be analyzed enters the cavity through an inlet tube between the
cathode and the stainless steel plate, and exits through the outlet tube. During
this process, the gas diffuses through the gas wicks of the hydrophobic backing
layer, and reaches the catalyst surface where O2 present in the gas mixture reacts
by the following mechanism:
O2 + 2H2O + 4e- → 4OH
Due to the high surface area of the catalyst, most of the O2 reacts at the cathode
surface and a very small amount of O2 dissolves in the bulk electrolyte. The
continuous sparging of the electrolyte with zero gas does not allow the O
concentration to build up in the bulk electrolyte. The amount of O2 that reaches
the catalyst surface is proportional to the partial pressure of O2 in the gas
mixture.
-
(cathode)
2
2-4
Teledyne Electronic Technologies
Analytical Instruments
Operational Operational
Operational
Operational Operational
When the O2 is reduced at the cathode, cadmium is simultaneously oxidized by
the following mechanism:
TheorTheor
Theor
TheorTheor
yy
y
yy
Ultra Ultra
Ultra
Ultra Ultra
TT
race Oxygrace Oxyg
T
race Oxyg
TT
race Oxygrace Oxyg
en Analen Anal
en Anal
en Analen Anal
yzyz
yz
yzyz
erer
er
erer
Cd → Cd+2 + 2e
The electrons released at the anode surface flow to the cathode surface via an
external circuit. This current is measured and used to determine the O
concentration in the gas mixture. The resulting current is directly proportional
to the O2 level in the gas mixture.
The current output of the sensor changes with temperature therefore, in order
to minimize the effects of ambient temperature variation on the current output,
the sensor is housed in a temperature-controlled oven maintained at 28°C
±2°C. Variations in temperature affecting the sensor are detected by a resistive
thermal device (RTD) attached to the sensor, which signals the temperature
controller to adjust the oven temperature accordingly.
A thermistor installed in the sensor monitors electrolyte temperature. Changes
in electrolyte temperature above 30°C are relayed to the microprocessor, which
adjusts the current output through a compensation algorithm.
NOTE: Since the gas to be analyzed does not flow through the bulk
electrolyte of the sensor, the probability of contaminating the
electrolyte by the particulates (if any) in the sample gas is very low.
This eliminates the requirements for frequent replacement of
sensor electrolyte. This is in contrast to Hercsh type galvanic cells
where gas is bubbled through the electrolyte and frequent replacement of electrolyte (3-6 month intervals) is essential to maintain
sensor performance.
-
(anode)
2
Faradaic Calibrator
The analyzer may be calibrated by using either the internal Faradaic calibrator
or an external span gas. The calibration using an external span gas is simple and
self-explanatory. The internal calibrator generates O2 at a precisely controlled
rate. The unique design of the calibrator allows repeated and reliable calibration
of the analyzer.
Faradaic Calibrator Components
Figure 2-3 illustrates major calibrator components. The calibrator is comprised
of a top electrode assembly and a bottom mounting block containing a cup for
electrolyte. An O-ring is used to make a leak-free seal between these two
components.
Teledyne Electronic Technologies
Analytical Instruments
2-5
Model 3060EModel 3060E
Model 3060E
Model 3060EModel 3060E
The two platinum/gold wire electrodes are attached to a feed-through header,
which is sealed to the stainless steel plate. The two electrodes are wrapped
around a porous polyethylene rod. The bottom end of the porous polyethylene
rod is submerged into the pool of electrolyte. Capillary action causes the
electrolyte to fill the pores, keeping the electrodes wet, which helps to maintain
the ionic conductivity between the two electrodes.
O
N
P
O
W
E
R
TE
LE
DYN
A
NA
E
LYTIC
AL
INS
TR
UM
EN
TS
PPBO XYGEN
P
P
M
O
X
Y
G
E
N
Chapter 2Chapter 2
Chapter 2
Chapter 2Chapter 2
C
O
N
T
R
O
L
M
O
D
U
L
E
2-6
Figure 2-3. Major Calibrator Components
Teledyne Electronic Technologies
Analytical Instruments
Operational Operational
Operational
Operational Operational
The calibrator is based on Faraday’s Law of Electrolysis. When applied to the
electrolysis of water, it states that the rate at which oxygen is generated is
directly proportional to the quantity of electric current flowing through the
electrodes. Based on this, the following formula is derived to determine the
current required to generate a mixture with the desired oxygen concentration
(PPM) and flow rate (SCCM).
The desired O2 concentration is entered through the calibration menu, and the
mass flow controller provides the flow rate of O2-free zero gas through the
calibrator. The required current for the electrolysis of water is calculated by the
microprocessor and then supplied to the calibrator electrodes.
A calibrator bypass uses zero gas to flush out any O2 that has seeped into the
calibration cavity path, eliminating excess O2 that might otherwise enter the
sampling system.
TheorTheor
Theor
TheorTheor
yy
y
yy
Ultra Ultra
Ultra
Ultra Ultra
Faradaic Calibrator Operation
µµ
I (
µA) =40/150 XO2 (PPM)XFlow Rate (SCCM)
µµ
TT
race Oxygrace Oxyg
T
race Oxyg
TT
race Oxygrace Oxyg
en Analen Anal
en Anal
en Analen Anal
yzyz
yz
yzyz
erer
er
erer
The accuracy of the calibration gas formed using the calibrator is best in the
0-10 PPM range. Since the sensor has linear output through all ranges, the
analyzer may be calibrated at one or two ranges above the normal range of
operation.
NOTE: For example, if the analyzer is to be used for analysis in the 0–50
PPB range, calibrate the analyzer in the 0–100 PPB or 0–1 PPM
range to minimize calibration error. At this level, errors due to
mass flow controller calibration, measurement of current from a
current source and current leakage are minimized.
To see why, assume the absolute error during calibration on any range is 2 PPB.
If the analyzer is calibrated at 25 PPB, 2 PPB error constitutes an 8% error in
calibration. However, if the analyzer is calibrated at 80 PPB the calibration
error will be only 2.5%.
Teledyne Electronic Technologies
Analytical Instruments
2-7
Model 3060EModel 3060E
Model 3060E
Model 3060EModel 3060E
Sample System
The sample system delivers gases to the O2 sensor from the analyzer rear panel
inlet. Depending on the mode of operation (determined by which valves are
open or closed), either a sample, span or zero gas is delivered.
Chapter 2Chapter 2
Chapter 2
Chapter 2Chapter 2
The 3060E sample system is designed and fabricated to insure that the O
concentration of the gas is not altered (except in the O2 scrubber) as it travels
through the sample system.
•Electropolished 316L Stainless Steel Components
To eliminate O2 absorption and desorption from the internal wetted
surfaces of the sample system components, the sample system is
fabricated from electropolished 316L stainless steel.
•Welding/Metal Gasket-Type Fittings
All of the joints upstream of the O2 sensor are orbitally welded,
except for the metal gasket-type compression connections at the O
scrubber and mass flow controller. Orbital welding is used in the
sample system wherever feasible. Orbital welding fuses the electropolished 316L stainless steel components together, forming a
smooth, clean internal (wetted) weld junction and eliminating small
spaces around the weld junction where gases can get trapped or
absorbed. All of the weld junctions in the entire assembly are
purged using an inert gas during welding to ensure that there is no
O2 contamination.
Orbital welding is used where practical; otherwise, conventional
precision welding is used. For example, conventional precision
welding is used to fuse the tubes to the mounting plates. Metal
gasket-type compression connections are used at the O2 scrubber
and at the mass flow controller to facilitate replacement. The metal
gasket-type connection creates an airtight, metal-to-metal seal,
eliminating inboard and outboard gas leakage.
2
2
2-8
Teledyne Electronic Technologies
Analytical Instruments
Operational Operational
Operational
Operational Operational
TheorTheor
Theor
TheorTheor
•Valves
yy
y
yy
The analyzer sampling system utilizes three different types of valves.
Each valve is selected to prevent O2 contamination of the sample
depending on its position and purpose in the circuit.
Air-Actuated Bellows Valves
These valves are normally closed in the sample system. They are
used to control the delivery through the sample system of the
sample, span or zero gas. The valve bodies are orbitally welded
in the system and the valve bonnets make a metal-to-metal seal
to the body. This valve system eliminates inboard and outboard
gas leakage. The valves are activated (open/closed) by computer-controlled solenoid valves.
The valves have the following basic functions:
PV1:Sample gas to the Ultra Trace sensor.
PV2:Zero gas to the Ultra Trace sensor.
PV3:Zero gas to the calibrator.
PV4:Span gas out of the calibrator.
PV5:External span gas to the Ultra Trace sensor.
PV6:Sample gas into the scrubber.
PV7:Zero gas out of the scrubber.
PV8:Sample bypass vent.
PV9:Water to the Ultra Trace sensor.
PV10: Calibration bypass.
Ultra Ultra
Ultra
Ultra Ultra
TT
race Oxygrace Oxyg
T
race Oxyg
TT
race Oxygrace Oxyg
en Analen Anal
en Anal
en Analen Anal
yzyz
yz
yzyz
erer
er
erer
Metering Valve
The metering valve (labeled MV1 on the flow schematic) is used
to manually control the sparge rate to the sensor. The body of
the metering valve is orbitally welded and the bonnet is sealed to
the body with metal O-rings. The manual control knob is located in the sensor compartment.
Solenoid Valves
The solenoid valves control the air flow to the air-activated
bellows valves. The solenoid valves are controlled by the microprocessor. When de-energized, the valve outlet is open to
ambient air, allowing the air-activated bellows valve to close.
Teledyne Electronic Technologies
Analytical Instruments
2-9
Model 3060EModel 3060E
Model 3060E
Model 3060EModel 3060E
•Oxygen Scrubber
The O2 scrubber is used to remove O2 (down to less than 1 PPB,
typically 0.5 PPB) from the sample gas. The scrubbed gas is used
for:
Sparging
Electrolyte in the Ultra Trace oxygen sensor is sparged with
oxygen-free gas to reduce dissolved oxygen from the electrolyte.
Zero Calibration
The oxygen-free gas is used to electronically offset the background signal of the Ultra Trace sensor. This provides a zero
point for the system to measure the oxygen content of the
sample gas during analysis.
Span Calibration Using Internal Calibrator
Chapter 2Chapter 2
Chapter 2
Chapter 2Chapter 2
The oxygen-free gas is mixed with oxygen generated in the
electrochemical calibrator to produce a span gas with a known
oxygen concentration. This provides a span point for the system
to measure the oxygen content of the sample during analysis.
NOTE: The standard O2 scrubber has a capacity of 80 cc of pure O2 .
Under normal operating conditions, the scrubber is expected to
last several years. The scrubber is installed in the system using
metal gasket and VCR type fittings.
The life of the scrubber is approximately 10 years when subjected
to an O2 concentration of 10 PPM at a flow rate of 150 cc/min.
•Mass Flow Controller
A mass flow controller is used to control the rate of gas flowing by
the sensing electrode of the sensor. The flow controller is located
upstream of the sensor. It provides signal input to the electronics in
the control module. The flow rate is displayed on the LCD screen
and is typically set at 150 cc/min.
•Fittings
The electropolished 316L stainless steel tees, elbows and crosses
used in the sampling system are orbitally welded for system integrity.
Small size fittings are used to make the sample system compact and
to minimize total system internal volume.
2-10
Teledyne Electronic Technologies
Analytical Instruments
Operational Operational
Operational
Operational Operational
TheorTheor
Theor
TheorTheor
yy
y
yy
Ultra Ultra
Ultra
Ultra Ultra
TT
race Oxygrace Oxyg
T
race Oxyg
TT
race Oxygrace Oxyg
en Analen Anal
en Anal
en Analen Anal
•Overall Sample System Design
The design of the sample system minimizes the volume of dead
space, which can retain residual gas from another route or previous
mode of analysis. Dead space contamination is minimized by the
structure of the gas flow routes (Figure 2-4).
yzyz
yz
yzyz
erer
er
erer
SAMPLE IN
OXYGEN SCRUBBER
SPAN GAS IN
PV
1
SHUT-OFF VALVEMETERING
Electronics
The analyzer has an embedded microcomputer, that controls all signal processing, input/output and display functions of the analyzer. System power is
supplied from a power supply module designed to be compatible with any
international power source.
PV
CALIBRATOR
BYPASS
6
PV
7
10
PV
3
CALIBRATOR
MV1
Figure 2-4. Flow Schematic
PV
PV
PV
PV
1
2
4
5
VALVE
MV2
8
MV1
PV
9
ULTRATRACE
SENSOR
H2O
MV3
BYPASSVENT (SAMPLE)
VENT #1 (SAMPLE/SPAN)
VENT #2 (SPARGER)
The microcomputer, a liquid crystal display (LCD) and all analog signal
processing electronics are located inside a replaceable control module. A light
emitting diode (LED) display and circuitry to actuate the valves in the sample
are located outside the control module.
Functional groups of analyzer electronics are:
1.Signal Processing
2.Temperature Control
3.Valve Control
4.Power Supply
Teledyne Electronic Technologies
Analytical Instruments
2-11
Model 3060EModel 3060E
Model 3060E
Model 3060EModel 3060E
Signal Processing Functions
Analog signal processing is accomplished in two plug-in circuit cards operating
under control of the microcomputer. One card processes analog input signals
and the other processes analog output signals. Digital signal processing is
accomplished directly by the microcomputer. All analog signals are converted
to digital early in the processing cycle to minimize analog processing and assure
maximum system accuracy, since digital processing is much more accurate than
analog and immune to many parameters such as drift and aging.
Analog Input Circuit Board
The analog input printed circuit board (PCB) is actually a plug-in module
consisting of a main board and a daughter board. Two analog signals are
connected to the inputs of this module for processing. They are:
•Ultra Trace Sensor Current Output
•Mass Flow Controller Output
Chapter 2Chapter 2
Chapter 2
Chapter 2Chapter 2
The circuitry on the daughter board converts the Ultra Trace sensor current to
a corresponding voltage, provides range control, isolates system grounds, and
filters high frequencies from the signal.
The circuitry on the main board processes the filtered output signal from the
daughter board and the signal outputs from the flow controller. All signals are
eventually connected through an analog multiplexer (electronic switch) to the
input of an analog-to-digital converter, where they are converted to digital
signals for use by the microcomputer.
This circuit board also contains a voltage measuring circuit to monitor the
voltage outputs of the system power supply, and a gas selector switch circuit
to adjust the mass flow controller for the background gas selected.
Analog Output Circuit Board
The analog output printed circuit board (PCB) generates the two 0-1 volt and
the two 4-20 mA analog signal outputs available on the rear panel of the
analyzer. These signals, generated in digital format by the microcomputer, are
converted into analog signals by the circuitry on this PCB. The output signals
represent the following:
•0-1V Signal (Oxygen Measurement)
2-12
This output goes from 0 to 1, representing 0 to 100% of the scale
that has been set; i.e., 0.6 volt is equal to 60% of the full scale, or 30
PPB when on the 50 PPB scale. It is possible that the signal may go
past zero into the negative range up to -0.25, especially if the ana-
Teledyne Electronic Technologies
Analytical Instruments
Operational Operational
Operational
Operational Operational
TheorTheor
Theor
TheorTheor
yy
y
yy
Ultra Ultra
Ultra
Ultra Ultra
TT
race Oxygrace Oxyg
T
race Oxyg
TT
race Oxygrace Oxyg
en Analen Anal
en Anal
en Analen Anal
lyzer has been zeroed with a gas that contains a significant concentration of O2 (Figure 2-3).
•0-1V Range Identifier
This 0 to 1 volt output represents each range with a particular
voltage as shown in Table 2-1.
•Isolated 4-20 mA Signal (Oxygen Measurement)
This is a 4 to 20 mA output representing 0 to 100% of the scale,
with 4 mA equal to 0%, and 20 mA equal to 100% of that range.
This output may also range lower than 4 mA, especially if the
analyzer has been zeroed with a gas that contains a significant
concentration of O2 (Figure 2-5).
•Isolated 4-20 mA Range Identifier
This 4 to 20 mA output identifies individual ranges with discrete
current output as shown in Table 2-1.
yzyz
yz
yzyz
erer
er
erer
1 V
- 1/4
Full Scale
Value
IdentifierIdentifier
RangeVoltage (V) Current (mA)
50 PPB0.04.0
100 PPB0.27.2
1 PPM0.410.4
10 PPM0.613.6
30 PPM0.816.8
Table 2-1. Range Identifier
Voltage (V)
20 mA
O2
(0,0)
-0.25
Full
Scale
- 1/4
Full
Scale
Value
Current (I)
O2
(0,0)
Full
Scale
Figure 2-5. Analog Signal Output Offset
Teledyne Electronic Technologies
Analytical Instruments
2-13
Model 3060EModel 3060E
Model 3060E
Model 3060EModel 3060E
Digital Circuit Board
The digital PCB is a general purpose microcomputer used to control all
functions of the analyzer. The analog input PCB and the analog output PCB
plug directly into connectors located on the digital PCB. In addition to
controlling these analog PCBs, the digital board performs the following
functions:
1. Provides valve control commands for the sampling system.
2. Processes input from the control panel pushbuttons.
3. Provides signals for the selectable alarms.
4. Processes serial I/O functions (RS-232 data). The following serial
interface default parameters are used:
Defaults
1200 BaudFixed
8 Bits
No Parity
1 Stop Bit
Chapter 2Chapter 2
Chapter 2
Chapter 2Chapter 2
5. Controls the LCD and the LED displays.
Display Screens
There are two display screens on the front panel of the analyzer:
LCD
This screen is a dot-matrix display located on the control module of the
analyzer. It displays all of the menus and commands for the user to
control the system and is the user interface for system operations.
LED
This screen is a 7-segment display located on the front panel of the
analyzer, above the control module. This screen displays only oxygen
concentration, but it is large and bright to allow the operator to read it
at a greater distance. A dimmer switch for this display is located on the
display PCB behind the front panel.
2-14
Teledyne Electronic Technologies
Analytical Instruments
Operational Operational
Operational
Operational Operational
Temperature control in the oven, which houses the sensor and Faradaic
calibrator, is processed by a stand-alone proportional integrative derivative
(PID) controller. Input from the resistive thermal device (RTD) sensor located
in the oven is used to regulate two 25-watt heaters. Oven temperature is set at
28 ±2°C and should be kept at this temperature during analyzer operation.
The valve control PCB is external to the control module, across the top of the
front panel of the analyzer. This PCB contains driver circuitry to activate the
solenoid valves. It also contains and controls dual-color LEDs that indicate the
actual status of the pneumatic valves (green for OPEN and red for CLOSED).
The valve control PCB operates under the control of the system microcomputer.
TheorTheor
Theor
TheorTheor
yy
y
yy
T emperature Controller
Valve Control Circuit Board
Ultra Ultra
Ultra
Ultra Ultra
TT
race Oxygrace Oxyg
T
race Oxyg
TT
race Oxygrace Oxyg
en Analen Anal
en Anal
en Analen Anal
yzyz
yz
yzyz
erer
er
erer
Power Supply Module
NOTE: This power supply contains an International Power Entry Module.
This feature allows operation on any of four international voltage
ranges: 100V, 120V, 220V or 240V (50Hz or 60Hz). It also
facilitates both North American and European fusing arrangements. Instructions for programming this module are described in
Chapter 3, Voltage Selection.
The analyzer power supply module is a replaceable assembly containing four
power supplies and five alarm relays. Electronic circuitry used to drive and
interface the alarm relays to the output of the microcomputer is also located
inside this module.
Teledyne Electronic Technologies
Analytical Instruments
2-15
Model 3060EModel 3060E
Model 3060E
Model 3060EModel 3060E
Chapter 2Chapter 2
Chapter 2
Chapter 2Chapter 2
2-16
Teledyne Electronic Technologies
Analytical Instruments
InstallationInstallation
Installation
InstallationInstallation
Overview
Installation of the analyzer includes:
Ultra Ultra
Ultra
Ultra Ultra
TT
race Oxygrace Oxyg
T
race Oxyg
TT
race Oxygrace Oxyg
en Analen Anal
en Anal
en Analen Anal
Installation
1. Unpacking the system.
2. Recognizing the necessary precautions when installing the system.
3. Adding electrolyte to the Ultra Trace sensor.
4. Adding electrolyte to the calibrator.
5. Hooking up electrical connections.
6. Hooking up the sample/span gas and instrument air supply to the
appropriate connections.
7. Testing the system.
yzyz
yz
yzyz
erer
er
erer
Unpacking the Analyzer
Carefully unpack the analyzer and inspect it for damage. Immediately report
any damage to the shipping agent. Remove the packing slip and verify that you
have received all of the correct materials.
Packing list:
1. Model 3060E Ultra Trace Oxygen Analyzer
2. 3060E Accessory Kit (P/N A59582)
The analyzer is shipped with all the materials and special items you need to
install and prepare the system for operation. The shipping carton contains the
following components:
1. Analyzer assembly
•Ultra Trace sensor
•Calibrator
•Two cadmium anodes (in holders)
•Electronics
Teledyne Electronic Technologies
Analytical Instruments
3-1
Model 3060EModel 3060E
Model 3060E
Model 3060EModel 3060E
2. Standard accessory kit for the 3060E (P/N A59582):
1W66Wrench, open end, ¼ " 1S915Screwdriver, #0 Phillips drive
1S916Screwdriver, slotted drive
1S914Socket, 7/16 " for ¼ " drive
1R1549Rachet, quick release, ¼ " drive
1W68Allen wrench, with handle, 7/64 "
1W69Allen wrench, long arm,
1A51934KOH electrolyte
1A64239Calibration electrolyte
1B51752Syringe 50 cc (Assembled)
1A51886Syringe 2 cc (Assembled)
1T1052Fritted glass tube
1Safety glasses
1 pr.Gloves, rubber
1MP-A56133Procedure for Preparation of Electrolyte
Chapter 3Chapter 3
Chapter 3
Chapter 3Chapter 3
3. Instruction Manual
4. Quick Reference: Start Up
Cautions and Warnings
1. Always use extreme care when handling the electrolyte bottles.
Electrolyte can cause skin irritation. If the electrolyte comes into
contact with skin, immediately wash the affected area with cold
water.
CAUTION:The sensor electrolyte is caustic. Protective equipment
including, but not limited to, gloves and safety glasses should
be worn while handling electrolyte. See the Material Safety
Data Sheets for potential hazards and corrective action in
case of accident.
2. Use the correct syringe for each type of electrolyte. The 50 cc
syringe is used with the sensor electrolyte. The 2 cc syringe is used
with the calibrator electrolyte.
3. Before adding electrolyte to the sensor, examine the electrical
connections on the sensor body to make sure that they are firmly
secured to the proper terminals.
3-2
Teledyne Electronic Technologies
Analytical Instruments
InstallationInstallation
Installation
InstallationInstallation
Ultra Trace Sensor Installation
The analyzer is usually shipped with the sensor installed. If you need to install
or replace the sensor assembly before using the analyzer, use the procedure
below.
Ultra Ultra
Ultra
Ultra Ultra
TT
race Oxygrace Oxyg
T
race Oxyg
TT
race Oxygrace Oxyg
en Analen Anal
en Anal
en Analen Anal
4. When adding electrolyte to the calibrator, do not overfill.
5. When hooking up electrical connections, do not connect the output
to the chassis.
6. Do not begin operating the analyzer until you have completed all of
the installation procedures.
1. Prepare the analyzer casing for the sensor. Place four of the short
bolts in the holes on the side of the sensor and four of the long bolts
in the holes at the bottom of the sensor and bring the sensor between
the top and side stainless steel plates (see Figure 3-1). Leave all
eight bolts loose until Step 3.
yzyz
yz
yzyz
erer
er
erer
O
N
P
O
W
E
R
PPM O
X
YG
PPB O
T
E
L
E
D
Y
N
E
A
N
A
L
Y
T
I
C
A
L
I
N
S
T
R
U
M
E
XYG
N
T
S
EN
C
O
N
T
R
O
L
M
O
D
U
L
E
EN
SIDE BOLTS
BASE BOLTS
Figure 3-1. Ultra Trace Sensor Bolt Installation
Teledyne Electronic Technologies
3-3
Analytical Instruments
Model 3060EModel 3060E
Model 3060E
Model 3060EModel 3060E
2. Slip the Teflon tubing (mounted inside the sensor) onto the 1/8 "
stainless steel tube at the bottom of the stainless steel top plate (see
Figure 3-2).
Chapter 3Chapter 3
Chapter 3
Chapter 3Chapter 3
Figure 3-2. Tube Alignment
3. Push the sensor all the way up against the top stainless steel plate.
4. Finger-tighten the side and then the bottom bolts with your hand.
5. Using the rachet (RL547) supplied, tighten each of the four bolts to
secure the sensor to the top plate.
6. Using the rachet (RL547) supplied, tighten each of the four bolts on
the side plate.
NOTE: Make sure that the O-rings are in their grooves on the top and side
of the sensor before tightening the bolts.
Adding Electrolyte to the Ultra Trace Sensor
CAUTION:Protective equipment including but not limited to gloves,
safety glasses, face shield and rubber apron must be worn
while handling electrolyte. The safety of the people in the
vicinity of the area in which the electrolyte is being handled
must be given similar consideration. Please refer to Material
Safety Data Sheets to learn about potential hazards and
corrective action in case of accident.
3-4
Teledyne Electronic Technologies
Analytical Instruments
Loading...
+ 137 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.