Teledyne 3000TA User Manual

OPERATING INSTRUCTIONS FOR
MODEL 3000TA
Trace Oxygen Analyzer
P/N M66316
3/09/11
DANGER
Toxic gases and or flammable liquids may be present in this monitoring
system.
instrument.
Hazardous voltages exist on certain components internally which may persist
for a time even after the power is turned off and disconnected.
Only authorized personnel should conduct maintenance and/or servicing. Before conducting any maintenance or servicing, consult with authorized supervisor/manager.
Teledyne Analytical Instruments
3000TA- EU
Copyright © 2011 Teledyne Analytical Instruments
All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in part, in any form or by any means, whether it be electronic, mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of Teledyne Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91749-1580.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us free from defects of material and of construction, and that our liability shall be limited to replacing or repairing at our factory (without charge, except for transportation), or at customer plant at our option, any material or construction in which defects become apparent within one year from the date of shipment, except in cases where quotations or acknowledgements provide for a shorter period. Components manufactured by others bear the warranty of their manufacturer. This warranty does not cover defects caused by wear, accident, misuse, neglect or repairs other than those performed by Teledyne or an authorized service center. We assume no liability for direct or indirect damages of any kind and the purchaser by the acceptance of the equipment will assume all liability for any damage which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our apparatus, and to make any alterations in the dimensions, shape or weight of any parts, in so far as such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its user, and serves as a tool by which valuable data can be gathered. The information provided by the instrument may assist the user in eliminating potential hazards caused by his process; however, it is essential that all personnel involved in the use of the instrument or its interface, with the process being measured, be properly trained in the process itself, as well as all instrumentation related to it.
The safety of personnel is ultimately the responsibility of those who control process conditions. While this instrument may be able to provide early warning of imminent danger, it has no control over process conditions, and it can be misused. In particular, any alarm or control systems installed must be tested and understood, both as to how they operate and as to how they can be defeated. Any safeguards required such as locks, labels, or redundancy, must be provided by the user or specifically requested of Teledyne at the time the order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The purchaser is responsible for the training of personnel, for providing hazard warning methods and instrumentation per the appropriate standards, and for ensuring that hazard warning devices and instrumentation are maintained and operated properly.
Teledyne Analytical Instruments, the manufacturer of this instrument, cannot accept responsibility for conditions beyond its knowledge and control. No statement expressed or implied by this document or any information disseminated by the manufacturer or its agents, is to be construed as a warranty of adequate safety control under the user’s process conditions.
Teledyne Analytical Instruments ii
Trace Oxygen Analyzer
Specific Model Information
The instrument for which this manual was supplied may incorporate one or more options not supplied in the standard instrument. Commonly available options are listed below, with check boxes. Any that are incorporated in the instrument for which this manual is supplied are indicated by a check mark in the box.
Instrument Serial Number: _______________________
Options Included in the Instrument with the Above Serial Number:
3000TA-C: In addition to all standard features, this model
also has separate ports for zero and span gases, and built-in control valves. The internal valves are entirely under the control of the 3000TA electronics to automatically switch between gases in synchronization with the analyzer’s operations
19" Rack Mnt: The 19" Relay Rack Mount units are available
with either one or two 3000 series analyzers installed in a standard 19" panel and ready to mount in a standard instrument rack.
Sensor Options Available for the Instrument with the Above Serial Number:
Insta-Trace B2C (Default) A2C L2C L2CL Insta-Trace A2C
Teledyne Analytical Instruments iii
3000TA- EU
Important Notice
Model 3000TA-EU complies with all of the requirements of the Commonwealth of Europe (CE) for Radio Frequency Interference, Electromagnetic Interference (RFI/EMI), and Low Voltage Directive (LVD).
The following International Symbols are used throughout the Instruction Manual. These symbols are visual indicators of important and immediate warnings and when you must exercise CAUTION while operating the instrument. See also the Safety Information on the next page.
STAND-BY: Instrument is on Stand-by, but circuit is
active
GROUND: Protective Earth
CAUTION: The operator needs to refer to the manual
for further information. Failure to do so may compromise the safe operation of the equipment.
CAUTION: Risk of Electrical Shock
Teledyne Analytical Instruments iv
Trace Oxygen Analyzer
Safety Messages
Your safety and the safety of others is very important. We have provided many important safety messages in this manual. Please read these messages carefully.
A safety message alerts you to potential hazards that could hurt you or others. Each safety message is associated with a safety alert symbol. These symbols are found in the manual and inside the instrument. The definition of these symbols is described below:
GENERAL WARNING/CAUTION: Refer to the
instructions for details on the specific danger. These cautions warn of specific procedures which if not followed could cause bodily Injury and/or damage the instrument.
No
Symbol
CAUTION: HOT SURFACE WARNING: This warning is
specific to heated components within the instrument. Failure to heed the warning could result in serious burns to skin and underlying tissue.
WARNING: ELECTRICAL SHOCK HAZARD: Dangerous
voltages appear within this instrument. This warning is specific to an electrical hazard existing at or nearby the component or procedure under discussion. Failure to heed this warning could result in injury and/or death from electrocution.
Technician Symbol: All operations marked with this
symbol are to be performed by qualified maintenance personnel only.
NOTE: Additional information and comments regarding a
specific component or procedure are highlighted in the form of a note.
Teledyne Analytical Instruments v
3000TA- EU
CAUTION: THE ANALYZER SHOULD ONLY BE USED FOR THE
PURPOSE AND IN THE MANNER DESCRIBED IN THIS MANUAL.
IF YOU USE THE ANALYZER IN A MANNER OTHER
THAN THAT FOR WHICH IT WAS INTENDED, UNPREDICTABLE BEHAVIOR COULD RESULT POSSIBLY ACCOMPANIED WITH HAZARDOUS CONSEQUENCES.
This manual provides information designed to guide you through the installation, calibration and operation of your new analyzer. Please read this manual and keep it available.
Occasionally, some instruments are customized for a particular application or features and/or options added per customer requests. Please check the front of this manual for any additional information in the form of an Addendum which discusses specific information, procedures, cautions and warnings that may be peculiar to your instrument.
Manuals do get lost. Additional manuals can be obtained from Teledyne at the address given in the Appendix. Some of our manuals are available in electronic form via the internet. Please visit our website at: www.teledyne-ai.com.
Teledyne Analytical Instruments vi
Trace Oxygen Analyzer
This is a general purpose instrument designed for use in a nonhazardous area. It is the customer's responsibility to ensure safety especially when combustible gases are being analyzed since the potential of gas leaks always exist.
The customer should ensure that the principles of operation of this equipment are well understood by the user. Misuse of this product in any manner, tampering with its components, or unauthorized substitution of any component may adversely affect the safety of this instrument.
Since the use of this instrument is beyond the control of Teledyne, no responsibility by Teledyne, its affiliates, and agents for damage or injury from misuse or neglect of this equipment is implied or assumed.
Teledyne Analytical Instruments vii
3000TA- EU
Table of Contents
Safety Messages ........................................................................... v
Introduction ................................................................................... 1
1.1 Overview 1
1.2 Typical Applications 1
1.3 Main Features of the Analyzer 1
1.4 Model Designations 2
1.5 Front Panel (Operator Interface) 3
1.6 Recognizing Difference Between LCD & VFD 4
1.7 Rear Panel (Equipment Interface) 5
Operational Theory ....................................................................... 7
2.1 Introduction 7
2.2 Micro-Fuel Cell Sensor 7
2.2.1 Principles of Operation 7
2.2.2 Anatomy of a Micro-Fuel Cell 8
2.2.3 Electrochemical Reactions 9
2.2.4 The Effect of Pressure 10
2.2.5 Calibration Characteristics 10
2.3 Sample System 11
2.4 Electronics and Signal Processing 13
Installation ................................................................................... 17
3.1 Unpacking the Analyzer 17
3.2 Mounting the Analyzer 17
3.3 Rear Panel Connections 19
3.3.1 Gas Connections 19
3.3.2 Electrical Connections 21
3.3.2.1 Primary Input Power 21
3.3.2.2 50-Pin Equipment Interface Connector 22
3.3.2.3 RS-232 Port 27
Teledyne Analytical Instruments viii
Trace Oxygen Analyzer
3.4 Installing the Micro-Fuel Cell 29
3.5 Testing the System 29
Operation ..................................................................................... 31
4.1 Introduction 31
4.2 Using the Data Entry and Function Buttons 31
4.3 The System Function 33
4.3.1 Tracking Oxygen Readings During Calibration and Alarm Delay 34
4.3.2 Setting up an Auto-Cal 36
4.3.3 Password Protection 37
4.3.3.1 Entering the Password 37
4.3.3.2 Installing or Changing the Password 38
4.3.4 Logout 40
4.3.5 System Self-Diagnostic Test 41
4.3.6 Version Screen 42
4.3.7 Showing Negative Oxygen Readings 42
4.4 The Zero and Span Functions 43
4.4.1 Zero Cal 44
4.4.1.1Auto Mode Zeroing 44
4.4.1.2 Manual Mode Zeroing 45
4.4.1.3Cell Failure 45
4.4.2 Span Cal 46
4.4.2.1 Auto Mode Spanning 46
4.4.2.2 Manual Mode Spanning 47
4.4.3 Span Failure 49
4.5 The Alarms Function 49
4.6 The Range Function 51
4.6.1 Setting the Analog Output Ranges 52
4.6.2 Fixed Range Analysis 53
4.7 The Analyze Function 54
4.8 Signal Output 54
Maintenance ................................................................................. 57
5.1 Routine Maintenance 57
5.2.1 Storing and Handling Replacement Cells 57
Teledyne Analytical Instruments ix
3000TA- EU
5.2.2 When to Replace a Cell 58
5.2.3 Removing the Micro-Fuel Cell 58
5.2.4 Installing a New Micro-Fuel Cell 59
5.2.5 Cell Warranty 61
5.3 Fuse Replacement 61
5.4 System Self Diagnostic Test 62
5.5 Major Internal Components 63
5.6 Cleaning 64
5.7 Troubleshooting 64
Appendix ...................................................................................... 67
A-1 Model 3000TA Specifications 67 A-2 Recommended 2-Year Spare Parts List 68 A-3 Drawing List 70 A-4 19-inch Relay Rack Panel Mount 70 A.5 Application notes 71 A-5 Material Safety Data Sheet 75
Teledyne Analytical Instruments x
Trace Oxygen Analyzer
List of Figures
Figure 1-1: Model 3000TA Front Panel ........................................... 3
Figure 1-2: Model 3000 TA Rear Panel ........................................... 5
Figure 2-1: Micro-Fuel Cell .............................................................. 8
Figure 2-2. Cross Section of a Micro-Fuel Cell (not to scale) .......... 8
Figure 2-3. Characteristic Input/Output Curve for a Micro-Fuel
Cell ............................................................................. 11
Figure 2-4: Piping Layout and Flow Diagram for Standard Model . 12
Figure 2-5: Flow Diagram .............................................................. 13
Figure 2-6: 3000TA Internal Electronic Component Location ........ 14
Figure 2-7: Block Diagram of the Model 3000TA-EU Electronics . 15
Figure 3-1: Front Panel of the Model 3000TA ............................... 18
Figure 3-2: Required Front Door Clearance .................................. 18
Figure 3-3: Rear Panel of the Model 3000TA ................................ 19
Figure 3-4: Equipment Interface Connector Pin Arrangement ....... 22
Figure 3-5: Remote Probe Connections ........................................ 27
Figure 3-6: FET Series Resistance ............................................... 27
Figure 5-1: Removing the Micro-Fuel ............................................ 59
Figure 5-2: Removing Fuse Block from Housing ........................... 61
Figure 5-3: Installing Fuses ........................................................... 62
Figure 5-4: Rear-Panel Screws ..................................................... 64
Figure A-1: Single and Dual 19" Rack Mounts .............................. 70
Teledyne Analytical Instruments xi
3000TA- EU
List of Tables
Table 3-1: Analog Output Connections Pin Function .................... 23
Table 3-2: Alarm Relay Contact Pins ............................................ 24
Table 3-3: Remote Calibration Connections .................................. 25
Table 3-4: Range ID Relay Connections ....................................... 26
Table 3-5: Commands via RS-232 Input ....................................... 28
Table 5-1: Self Test Failure Codes ................................................ 62
Teledyne Analytical Instruments xii
Trace Oxygen Analyzer Introduction
Introduction
1.1 Overview
The Teledyne Analytical Instruments Model 3000TA Trace Oxygen Analyzer is a versatile microprocessor-based instrument for detecting oxygen at the parts-per-million (ppm) level in a variety of gases. This manual covers the Model 3000TA General Purpose flush­panel and/or rack-mount units only. These units are for indoor use in a nonhazardous environment.
1.2 Typical Applications
A few typical applications of the Model 3000TA are:
Monitoring inert gas blanketing
Air separation and liquefaction
Chemical reaction monitoring
Semiconductor manufacturing
Petrochemical process control
Quality assurance
Gas analysis certification.
1.3 Main Features of the Analyzer
The Model 3000TA Trace Oxygen Analyzer is sophisticated yet simple to use. The main features of the analyzer include:
A 2-line alphanumeric display screen, driven by
microprocessor electronics, that continuously prompts and informs the operator.
High resolution, accurate readings of oxygen content
from low ppm levels through 25%. Large, bright, meter readout.
Nylon cell block. (Stainless steel optional)
Teledyne Analytical Instruments 1
Introduction 3000TA- EU
Advanced Micro-Fuel Cell, designed for trace analysis,
has a one year warranty and an expected lifetime of two years.
Versatile analysis over a wide range of applications.
Microprocessor based electronics: 8-bit CMOS
microprocessor with 32 kB RAM and 128 kB ROM.
Three user definable output ranges (from 0-10 ppm
through 0- 250,000 ppm) allow best match to users process and equipment.
Air-calibration range for convenient spanning at 20.9 %.
Auto Ranging allows analyzer to automatically select the
proper preset range for a given measurement. Manual override allows the user to lock onto a specific range of interest.
Two adjustable concentration alarms and a system failure
alarm.
Extensive self-diagnostic testing, at startup and on
demand, with continuous power-supply monitoring.
CE Compliance.
RS-232 serial digital port for use with a computer or
other digital communication device.
Four analog outputs: two for measurement (0–1 VDC and
Isolated 4–20 mA DC) and two for range identification.
Convenient and versatile, steel, flush-panel or rack-
mountable case with slide-out electronics drawer.
1.4 Model Designations
3000TA: Standard model.
3000TA-C: In addition to all standard features, this model
also has separate ports for zero and span gases, and built-in control valves. The internal valves are entirely under the control of the 3000TA electronics, to automatically switch between gases in synchronization with the analyzer’s operations.
Teledyne Analytical Instruments 2
Trace Oxygen Analyzer Introduction
1.5 Front Panel (Operator Interface)
The standard 3000TA is housed in a rugged metal case with all controls and displays accessible from the front panel. See Figure 1-1. The front panel has thirteen buttons for operating the analyzer, a digital meter, an alphanumeric display, and a window for viewing the sample flowmeter.
Figure 1-1: Model 3000TA Front Panel
Function Keys: Six touch-sensitive membrane switches are used to change the specific function performed by the analyzer:
Analyze Perform analysis for oxygen content of a sample
gas.
System Perform system-related tasks (described in detail
in chapter 4, Operation.).
Span Span calibrate the analyzer.
Zero Zero calibrate the analyzer.
Alarms Set the alarm setpoints and attributes.
Range Set up the 3 user definable ranges for the
instrument.
Data Entry Keys: Six touch-sensitive membrane switches are used to input data to the instrument via the alphanumeric VFD display:
Teledyne Analytical Instruments 3
Introduction 3000TA- EU
Left & Right Arrows Select between functions currently
displayed on the VFD screen.
Up & Down Arrows Increment or decrement values of
functions currently displayed.
Enter Moves VFD display on to the next screen in a series.
If none remains, returns to the Analyze screen.
Escape Moves VFD display back to the previous screen in a
series. If none remains, returns to the Analyze screen.
Digital Meter Display: The meter display is a LED device that produces large, bright, 7-segment numbers that are legible in any lighting. It produces a continuous readout from 0-10,000 ppm and then switches to a continuous percent readout from 1-25%. It is accurate across all analysis ranges without the discontinuity inherent in analog range switching.
Alphanumeric Interface Screen: The VFD screen is an easy-to-use interface from operator to analyzer. It displays values, options, and messages that give the operator immediate feedback.
Flowmeter: Monitors the flow of gas past the sensor. Readout is 0.2 to
2.4 standard liters per minute (SLPM).
Standby Button: The Standby button turns off the display and outputs, but circuitry is still operating.
CAUTION: THE POWER CABLE MUST BE UNPLUGGED TO
FULLY DISCONNECT POWER FROM THE INSTRUMENT. WHEN CHASSIS IS EXPOSED OR WHEN ACCESS DOOR IS OPEN AND POWER CABLE IS CONNECTED, USE EXTRA CARE TO AVOID CONTACT WITH LIVE ELECTRICAL CIRCUITS.
Access Door: For access to the Micro-Fuel Cell, the front panel swings open when the latch in the upper right corner of the panel is pressed all the way in with a narrow gauge tool. Accessing the main circuit board requires unfastening rear panel screws and sliding the unit out of the case.
1.6 Recognizing Difference Between LCD & VFD
LCD has GREEN background with BLACK characters. VFD has DARK background with GREEN characters. In the case of VFD - NO CONTRAST ADJUSTMENT IS NEEDED.
Teledyne Analytical Instruments 4
Trace Oxygen Analyzer Introduction
1.7 Rear Panel (Equipment Interface)
The rear panel, shown in Figure 1-2, contains the gas and electrical connectors for external inlets and outlets. Those that are optional are shown shaded in the figure. The connectors are described briefly here
and in detail in the Installation chapter of this manual.
Figure 1-2: Model 3000 TA Rear Panel
Power Connection Universal AC power source.
Gas Inlet and Outlet One inlet (must be externally valved)
and one exhaust out. Three inlet when “C” option ordered.
RS-232 Port Serial digital concentration signal output and
control input.
Remote Valves Used in the 3000TA for controlling external
solenoid valves only.
50-Pin Equipment Interface Port:
Analog Outputs 0–1 VDC concentration plus 0-1 VDC
range ID, and isolated 4–20 mA DC plus 4-20 mA DC range ID.
Alarm Connections 2 concentration alarms and 1 system
alarm.
Teledyne Analytical Instruments 5
Introduction 3000TA- EU
Remote Span/Zero Digital inputs allow external control
of analyzer calibration.
Calibration Contact To notify external equipment that
instrument is being calibrated and readings are not monitoring sample.
Range ID Contacts Four separate, dedicated, range relay
contacts. Low, Medium, High, Cal.
Network I/O Serial digital communications for local
network access. For future expansion. Not implemented at this printing.
Optional:
Calibration Gas Ports (Auto Cal Option) Separate
fittings for zero, span and sample gas input, and internal valves for automatically switching the gases.
Note: If you require highly ac curate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the Model 3000TA is accurate to 2-3 %. Accordingly, internally scheduled calibrations can vary 2-3 % per day.
Teledyne Analytical Instruments 6
Trace Oxygen Analyzer Operational Theory
Operational Theory
2.1 Introduction
The analyzer is composed of three subsystems:
1. Micro-Fuel Cell Sensor
2. Sample System
3. Electronic Signal Processing, Display and Control
The sample system is designed to accept the sample gas and transport it through the analyzer without contaminating or altering the sample prior to analysis. The Micro-Fuel Cell is an electrochemical galvanic device that translates the amount of oxygen present in the sample into an electrical current. The electronic signal processing, display and control subsystem simplifies operation of the analyzer and accurately processes the sampled data. The microprocessor controls all signal processing, input/output and display functions for the analyzer.
2.2 Micro-Fuel Cell Sensor
2.2.1 Principles of Operation
The oxygen sensor used in the Model 3000T series is a Micro-Fuel Cell designed and manufactured by Analytical Instruments. It is a sealed plastic disposable electrochemical transducer.
The active components of the Micro-Fuel Cell are a cathode, an anode, and the 15% aqueous KOH electrolyte in which they are immersed. The cell converts the energy from a chemical reaction into an electrical current in an external electrical circuit. Its action is similar to that of a battery.
There is, however, an important difference in the operation of a battery as compared to the Micro-Fuel Cell: In the battery, all reactants are stored within the cell, whereas in the Micro-Fuel Cell, one of the reactants (oxygen) comes from outside the device as a constituent of the sample gas being analyzed. The Micro-Fuel Cell is therefore a hybrid between a battery and a true fuel cell. (All of the reactants are stored externally in a true fuel cell.)
Teledyne Analytical Instruments 7
Operational Theory 3000TA- EU
2.2.2 Anatomy of a Micro-Fuel Cell
The Micro-Fuel Cell is a cylinder only 11/4 inches in diameter and 11/4 inches thick. It is made of an extremely inert plastic, which can be placed confidently in practically any environment or sample stream. It is effectively sealed, although one end is permeable to oxygen in the sample gas. The other end of the cell is a contact plate consisting of two concentric foil rings. The rings mate with spring-loaded contacts in the sensor block assembly and provide the electrical connection to the rest of the analyzer. Figure 2-1 illustrates the external features.
Figure 2-1: Micro-Fuel Cell
Refer to Figure 2-2, Cross Section of a Micro-Fuel Cell, which
illustrates the following internal description.
Figure 2-2. Cross Section of a Micro-Fuel Cell (not to scale)
At the top end of the cell is a diffusion membrane of Teflon®, whose thickness is very accurately controlled. Beneath the diffusion membrane lies the oxygen sensing element—the cathode—with a surface area almost 4 cm2. The cathode has many perforations to ensure
Teledyne Analytical Instruments 8
Trace Oxygen Analyzer Operational Theory
sufficient wetting of the upper surface with electrolyte, and it is plated with an inert metal.
The anode structure is below the cathode. It is made of lead and has a proprietary design which is meant to maximize the amount of metal available for chemical reaction.
At the rear of the cell, just below the anode structure, is a flexible membrane designed to accommodate the internal volume changes that occur throughout the life of the cell. This flexibility assures that the sensing membrane remains in its proper position, keeping the electrical output constant.
The entire space between the diffusion membrane, above the cathode, and the flexible rear membrane, beneath the anode, is filled with electrolyte. Cathode and anode are submerged in this common pool. They each have a conductor connecting them to one of the external contact rings on the contact plate, which is on the bottom of the cell.
2.2.3 Electrochemical Reactions
The sample gas diffuses through the Teflon membrane. Any oxygen in the sample gas is reduced on the surface of the cathode by the following HALF REACTION:
O
+ 2H2O + 4e– 4OH– (cathode)
2
(Four electrons combine with one oxygen molecule—in the presence of water from the electrolyte—to produce four hydroxyl ions.)
When the oxygen is reduced at the cathode, lead is simultaneously oxidized at the anode by the following HALF REACTION:
Pb + 2OH
Pb+2 + H2O + 2e– (anode)
(Two electrons are transferred for each atom of lead that is oxidized. Therefore it takes two of the above anode reactions to balance one cathode reaction and transfer four electrons.)
The electrons released at the surface of the anode flow to the cathode surface when an external electrical path is provided. The current is proportional to the amount of oxygen reaching the cathode. It is
Teledyne Analytical Instruments 9
Operational Theory 3000TA- EU
measured and used to determine the oxygen concentration in the gas mixture.
The overall reaction for the fuel cell is the SUM of the half
reactions above, or:
2Pb + O
→2PbO
2
(These reactions will hold as long as no gaseous components capable of oxidizing lead—such as iodine, bromine, chlorine and fluorine—are present in the sample.)
The output of the fuel cell is limited by (1) the amount of oxygen in the cell at the time and (2) the amount of stored anode material.
In the absence of oxygen, no current is generated.
2.2.4 The Effect of Pressure
In order to state the amount of oxygen present in the sample in parts-per-million or a percentage of the gas mixture, it is necessary that the sample diffuse into the cell under constant pressure.
If the total pressure increases, the rate that oxygen reaches the cathode through the diffusing membrane will also increase. The electron transfer, and therefore the external current, will increase, even though the oxygen concentration of the sample has not changed. It is therefore important that the sample pressure at the fuel cell (usually vent pressure) remain relatively constant between calibrations.
2.2.5 Calibration Characteristics
Given that the total pressure of the sample gas on the surface of the Micro-Fuel Cell input is constant, a convenient characteristic of the cell is that the current produced in an external circuit is directly proportional to the rate at which oxygen molecules reach the cathode, and this rate is directly proportional to the concentration of oxygen in the gaseous mix­ture. In other words it has a linear characteristic curve, as shown in Figure 2-3. Measuring circuits do not have to compensate for nonlinearities.
In addition, since there is zero output in the absence oxygen, the characteristic curve has close to an absolute zero (within ± 1 ppm oxygen). In practical application, zeroing may still used to compensate for the combined zero offsets of the cell and the electronics. (The
Teledyne Analytical Instruments 10
Trace Oxygen Analyzer Operational Theory
electronics is zeroed automatically when the instrument power is turned on.)
Figure 2-3. Characteristic Input/Output Curve for a Micro-Fuel
Cell
2.3 Sample System
The sample system delivers gases to the Micro-Fuel Cell sensor from the analyzer rear panel inlet. Depending on the mode of operation either sample or calibration gas is delivered.
The Model 3000TA sample system is designed and fabricated to ensure that the oxygen concentration of the gas is not altered as it travels through the sample system. The sample encounters almost no dead space. This minimizes residual gas pockets that can interfere with trace analysis.
The sample system for the standard instrument incorporates 1/4 inch tube fittings for sample inlet and outlet connections at the rear panel. For metric system installations, 6 mm adapters are supplied with each instrument to be used if needed. The sample or calibration gas flows through the system is monitored by a flowmeter downstream from
Teledyne Analytical Instruments 11
Operational Theory 3000TA- EU
the cell. Figure 2-4 shows the piping layout and flow diagram for the standard model.
Figure 2-4: Piping Layout and Flow Diagram for Standard Model
Figure 2-5 is the flow diagram for the sampling system. In the stan­dard instrument, calibration gases (zero and span) can be connected di­rectly to the Sample In port by teeing to the port with appropriate valves. The shaded portion of the diagram shows the components added when the –C option is ordered. The valving is installed inside the 3000TA-C enclosure and is regulated by the instruments internal electronics.
Teledyne Analytical Instruments 12
Trace Oxygen Analyzer Operational Theory
Figure 2-5: Flow Diagram
2.4 Electronics and Signal Processing
The Model 3000TA Trace Oxygen Analyzer uses an 8031 microcontroller with 32 kB of RAM and 128 kB of ROM to control all signal processing, input/output, and display functions for the analyzer. System power is supplied from a universal power supply module designed to be compatible with any international power source. Figure 2-6 shows the location of the power supply and the main electronic PC boards.
The signal processing electronics including the microprocessor, analog to digital, and digital to analog converters are located on the motherboard at the bottom of the case. The preamplifier board is mounted on top of the motherboard as shown in the figure. These boards are accessible after removing the back panel. Figure 2-7 is a block diagram of the Analyzer electronics.
Teledyne Analytical Instruments 13
Operational Theory 3000TA- EU
Figure 2-6: 3000TA Internal Electronic Component Location
In the presence of oxygen the cell generates a current. A current to voltage amplifier converts this current to a voltage, which is amplified in the second stage amplifier.
The second stage amplifier also supplies temperature compensation for the oxygen sensor output. This amplifier circuit incorporates a ther­mistor, which is physically located in the cell block. The thermistor is a temperature dependent resistance that changes the gain of the amplifier in proportion to the temperature changes in the block. This change is in­versely proportional to the change in the cell output due to the same temperature changes. The result is a signal that is temperature independent. The output from the second stage amplifier is sent to an 18 bit analog to digital converter controlled by the microprocessor.
Teledyne Analytical Instruments 14
Trace Oxygen Analyzer Operational Theory
Figure 2-7: Block Diagram of the Model 3000TA-EU Electronics
Teledyne Analytical Instruments 15
Loading...
+ 62 hidden pages