Teledyne 3000TA User Manual 2

Teledyne Analytical Instruments
QUICKSTART GUIDE
MODEL 3000 TA
Trace Oxygen Analyzer
Quickstart Guide Trace Oxygen Analyzer
Teledyne Analytical Instruments
Model 3000TA Quickstart Guide
Teledyne Analytical Instruments 3
GETTING STARTED
This Quickstart Guide is designed to get you set up and operating your Teledyne Analytical Instruments Analyzer quickly. It shortcuts the details so you can install and use your new analyzer with a minimum of fuss. This Quickstart Guide should be used in conjunction with the Instruction Manual that shipped with your instrument. Only necessary features to get you operational are discussed in this guide. Many of the advanced features of this analyzer are not described here so you should refer to the Instruction Manual to get the most from your analyzer
Front Panel (Operator Interface)
Take a minute to familiarize yourself with the analyzer’s interface. The standard 3000TA is housed in a rugged metal case with all controls and displays accessible from the front panel. The operator interface consists of thirteen buttons for operating the analyzer, a digital meter, an alphanumeric display, and a window for viewing the sample flowmeter.
Function Keys:
Six touch-sensitive membrane switches
are used to change the specific function
performed by the analyzer:
Analyze Perform analysis for
oxygen content of a sample gas.
System Perform system-
related tasks (described in detail in chapter 4, Operation.).
Span Span calibrate the analyzer.
Zero Zero calibrate the analyzer.
Alarms Set the alarm setpoints and attributes.
Range Set up the 3 user definable ranges for the instrument.
Data Entry Keys: Six touch-sensitive membrane switches are used to input data to the instrument via the alphanumeric VFD display:
Quickstart Guide Trace Oxygen Analyzer
Teledyne Analytical Instruments 4
Left & Right Arrows Select between functions currently displayed on the VFD
screen.
Up & Down Arrows Increment or decrement values of functions currently displayed.
Enter Advances VFD display to the next screen in a series or returns to the Analyze
screen if none remain.
Escape Backs VFD display to the previous screen in a series or returns to the Analyze
screen if none remain.
Digital Meter Display: The meter display is a Light Emitting Diode (LED) device that produces large, bright, 7-segment numbers that are legible in any lighting. It produces a continuous readout from 0-10,000 ppm and then switches to a continuous percent readout from 1-25%.
Alphanumeric Interface Screen The VFD screen displays values, options, and messages for immediate feedback.
Set Flow: Needle valve used to adjust flow of gas sample
Flowmeter:
Monitors the flow of gas past the sensor. Readout is 0.2 to 2.4 SLPM of nitrogen.
Standby Button:
The Standby turns off the display and outputs, but circuitry is still operating.
Access Door: For access to the Micro-fuel Cell,the front panel swings open when the latch in the upper right corner of the panel is pressed all the way in.
Rear Panel (Equipment Interface)
At the rear panel you make the gas and electrical connections to the instrument.
Power Connection:
Universal AC power source.
Gas Inlet and Outlet:
Inlet, exhaust, zero and span.
Remote Probe
Connection: Used for controlling external solenoid valves only.
RS-232 Port: Serial digital concentration signal output and control input.
Model 3000TA Quickstart Guide
Teledyne Analytical Instruments 5
Analog Output Connection: Terminals for current and voltage analog output
signals.
Digital Input Terminals: Terminals for connecting digital input signals.
Alarm Connections: 2 concentration alarms and 1 system alarm.
Calibration Contact : To notify external equipment that instrument is being
calibrated and readings are not monitoring sample.
Range ID Contacts: Four separate, dedicated, range relay contacts. Low, Medium,
High, Cal.
Remote Span/Zero: Digital inputs for externally controlling calibration.
Network I/O: Serial digital communications for local network access.
Quickstart Guide Trace Oxygen Analyzer
Teledyne Analytical Instruments 6
SETUP AND INSTALL
Mount the Analyzer
The Model 3000TA is for indoor use in a general purpose area. It is NOT for hazardous environments of any type. It is designed for flush panel mounting. Use the four mounting holes— one in each corner of the rigid frame to panel mount the instrument.
Gas Connections
Note: Do not remove plastic caps on VCR fittings, unless you are ready to make
connections.
The analyzer is equipped with 1/4 inch tube fittings, and 6 mm adapters are supplied for metric system installations.
You will need:
Zero gas — regulated supply of oxygen free gas.
Span gas — regulated supply of a standard gas typically containing oxygen at 70-
90% of the full scale value on the range of interest.
Purge gas — oxygen free inert gas for purging if applicable.
Sample gas — regulated supply of sample gas available to the analyzer.
Make your gas connections to the analyzer as follows:
1. Install the flow restrictor that shipped with your instrument:
For positive pressure applications, install the restrictor with the blue dot to the
SAMPLE IN port. The small circular orifice should face away from the back of the unit. Remove the sticker after installing.
For low pressure applications (less than 5 psig), connect the restrictor without the
blue dot to the SAMPLE IN port.
For vacuum service (5-10 in Hg), install the restrictor without the blue dot to the
EXHAUST port.
2. Connect your sample gas to the SAMPLE IN port.
3. Connect a vent line to EXHAUST OUT port.
Loading...
+ 13 hidden pages