Teledyne 2002 User Manual

TELEDYNE HASTINGS
INSTRUCTION MANUAL
INSTRUMENTS
MODEL 2002MODEL 2002
MODEL 2002
MODEL 2002MODEL 2002
VV
AA
V
VV
CUUM GCUUM G
A
AA
CUUM GCUUM G
AA
UGEUGE
A
UGE
AA
UGEUGE
Model 2002 page 1
Manual Print HistoryManual Print History
Manual Print History
Manual Print HistoryManual Print History
The print history shown below lists the printing dates of all revisions and addenda created for this manual. The revision level letter increases alphabetically as the manual undergoes subse­quent updates. Addenda, which are released between revisions, contain important change information that the user should incorporate immediately into the manual. Addenda are num­bered sequentially . When a new revision is created, all addenda associated with the previous revision of the manual are incorporated into the new revision of the manual. Each new revision includes a revised copy of this print history page.
Revision A (Document Number 149-0798) .......................................................September 1998
Revision B (Document Number 149-032000) ......................................................... March 2000
Revision C (Document Number 149-052002) .............................................................May 2002
Revision D (Document Number 149-102002)....................................................... October 2002
Revision E (Document Number 149-062004).............................................................. June 2004
Revision F (Document Number 149-082005)..........................................................August 2005
Revision F (Document Number 149-102005)........................................................ October 2005
Visit www.teledyne-hi.com for WEEE disposal guidance.
CAUTION
!
Refer to accompanying documents
CAUTION
!
If equipment is not used in the manner specified by this manual, protection provided by the equipment may be impaired
page 2 Model 2002
Hastings Instruments reserves the right to change or modify the design of its equipment without any obligation to provide notification of change or intent to change.
TT
able of Contentsable of Contents
T
able of Contents
TT
able of Contentsable of Contents
1.01.0
GENERAL INFORMAGENERAL INFORMA
1.0
GENERAL INFORMA
1.01.0
GENERAL INFORMAGENERAL INFORMA
1.1 Features................................................................................................................................5
1.2 Model 2002 Sensors ............................................................................................................5
1.3 Model 2002 Control Unit ....................................................................................................6
1.4 Specifications ....................................................................................................................... 6
2.02.0
INSTINST
2.0
INST
2.02.0
INSTINST
2.1 Receiving Inspection............................................................................................................ 7
2.2 Quick Start ...........................................................................................................................7
2.3 Transducer Installation ........................................................................................................8
2.4 Control Unit Installation......................................................................................................8
2.5 Initial Operation...................................................................................................................8
3.03.0
FRFR
3.0
3.03.0
3.1 Overall Functional Description............................................................................................9
ONT PONT P
FR
ONT P
FRFR
ONT PONT P
ALLAALLA
ALLA
ALLAALLA
TIONTION
TION
TIONTION
ANEL OPERAANEL OPERA
ANEL OPERA
ANEL OPERAANEL OPERA
TIONTION
TION
TIONTION
..............................................................................................................................................................................
.......................................................................................
..............................................................................................................................................................................
......................................................................................................................................
...................................................................
......................................................................................................................................
TIONTION
TION
TIONTION
................................................................................................................................
................................................................
................................................................................................................................
55
5
55
77
7
77
99
9
99
3.2 RUN Mode ........................................................................................................................10
3.3 HI and LOW Set Point Modes ..........................................................................................11
3.4 CAL Mode.........................................................................................................................11
3.5 GAS Mode.........................................................................................................................14
3.6 UNITS Mode ....................................................................................................................15
4.04.0
REAR PREAR P
4.0
REAR P
4.04.0
REAR PREAR P
4.1 Remote Zero Input ...........................................................................................................17
4.2 Analog Output ...................................................................................................................17
4.3 TTL Outputs.....................................................................................................................19
4.4 Power Entry Module..........................................................................................................20
5.05.0
THEORTHEOR
5.0
THEOR
5.05.0
THEORTHEOR
5.1 Piezoresistive Sensor..........................................................................................................22
5.2 Pirani Sensor......................................................................................................................24
5.3 Dual Sensor Operation ......................................................................................................26
ANEL DESCRIPTIONANEL DESCRIPTION
ANEL DESCRIPTION
ANEL DESCRIPTIONANEL DESCRIPTION
Y OF OPERAY OF OPERA
Y OF OPERA
Y OF OPERAY OF OPERA
TIONTION
TION
TIONTION
......................................................................................................................................
...................................................................
......................................................................................................................................
............................................................................................................................
..............................................................
............................................................................................................................
1717
17
1717
2121
21
2121
6.06.0
COMMUNICACOMMUNICA
6.0
COMMUNICA
6.06.0
COMMUNICACOMMUNICA
6.1 RS-232-E Interface Specifications.....................................................................................29
6.2 RS-232 Interface Connector Pin Assignments ..................................................................29
6.2.1 RS-485 Interface Connector Pin Assignments .........................................................30
6.3 Operation of the Serial Interface........................................................................................30
TIONS OPTION BOTIONS OPTION BO
TIONS OPTION BO
TIONS OPTION BOTIONS OPTION BO
ARDARD
ARD
ARDARD
........................................................................................
............................................
........................................................................................
Model 2002 page 3
2929
29
2929
7.07.0
RELARELA
7.0
RELA
7.07.0
RELARELA
7.1 Relay Board Specifications .............................................................................................. 35
7.2 Relay Connector Pin Assignment .................................................................................... 35
7.3 Operation ......................................................................................................................... 36
8.08.0
4-20mA INTERF4-20mA INTERF
8.0
4-20mA INTERF
8.08.0
4-20mA INTERF4-20mA INTERF
8.1 4-20 mA Interface Connector Pin Assignments .............................................................. 37
8.2 Operation ......................................................................................................................... 38
9.09.0
0-10 0-10
9.0
0-10
9.09.0
0-10 0-10
9.1 0-10 Volt Interface Connector Pin Assignments .............................................................. 39
10.010.0
10.0
10.010.0
11.011.0
11.0
W
11.011.0
11.1 Warranty Repair Policy .................................................................................................... 43
Y INTERFY INTERF
Y INTERF
Y INTERFY INTERF
VV
olt INTERFolt INTERF
V
olt INTERF
VV
olt INTERFolt INTERF
TRTR
OUBLESHOOOUBLESHOO
TR
OUBLESHOO
TRTR
OUBLESHOOOUBLESHOO
WW
ARRANTY ARRANTY
ARRANTY
WW
ARRANTY ARRANTY
AA
CE OPTION BOCE OPTION BO
A
CE OPTION BO
AA
CE OPTION BOCE OPTION BO
AA
CE OPTION BOCE OPTION BO
A
CE OPTION BO
AA
CE OPTION BOCE OPTION BO
AA
CECE
......................................................................................................................................................
A
CE
...........................................................................
AA
CECE
......................................................................................................................................................
TING GUIDETING GUIDE
TING GUIDE
TING GUIDETING GUIDE
AND REPAND REP
AND REP
AND REPAND REP
AIRAIR
AIR
AIRAIR
ARDARD
............................................................................................
ARD
..............................................
ARDARD
............................................................................................
ARDARD
..........................................................................................
ARD
.............................................
ARDARD
..........................................................................................
....................................................................................................................
..........................................................
....................................................................................................................
..................................................................................................................................
.................................................................
..................................................................................................................................
3535
35
3535
3737
37
3737
3939
39
3939
4141
41
4141
4343
43
4343
11.2 Non-Warranty Repair Policy ........................................................................................... 43
11
2.02.0
DIA DIA
1
11
2.0
2.02.0
DIA
DIA DIA
GRAMS GRAMS
GRAMS
GRAMS GRAMS
AND DRAAND DRA
AND DRA
AND DRAAND DRA
WINGSWINGS
WINGS
WINGSWINGS
....................................................................................................................
..........................................................
....................................................................................................................
4545
45
4545
page 4 Model 2002
The Model 2002 is a wide range vacuum measuring instrument consisting of a digital microproces­sor-based display unit, a miniature, rugged, thin film based sensing element that measures from below 1x10 EEPROM (Electrically Erasable/Programmable Read Only Memory), and an interconnecting cable. Together, these components provide accurate vacuum measurement over 7 decades of vacuum from 1x10-4 to 1000 Torr. This instrument is a new design approach based on over 50 years of experience by Hastings Instruments with well-known and widely used thermal conduc­tion vacuum measurement techniques. The Model 2002 is designed for quick, easy installation and will provide the user with long lasting, trouble free, accura te vacuum measurement.
1.1 Features
SECTION 1
General Information
-4
up to 100 Torr, a piezoresistiv e sensor that measures from 1 to 1000 Torr,
Wide dynamic range: 1x10
Compact rugged design
Attitude insensitive
Fast dynamic response
Withstands 150 psig positive pressure
Linear per decade
Bakeable to 250
TTL output process control
Interchangeable transducer
Corrosion resistant
1.2 Model 2002 Sensors
The Model 2002 transducer is comprised of an ion implanted, piezoresistive, direct force sensor and a thin film Pirani type sensor. The Pirani sensing element is a Pt thin film serpentine element (see Figure 5.3a) deposited on a 1 micron thick Si3N4 membrane. The membrane is peripherally supported by a Si box shaped die and is covered by a thick Si lid parallel to the membrane and open on two ends (see Figure 5.3b). The piezoresistive unit is an ion implanted Wheatstone bridge in a 50 micron thick Si membrane peripherally supported by a Si box shaped die which has been anodically bonded to a Pyrex substrate.
-4
to 1000 Torr
0
C
The dual sensor assembly is encased in a corrosion resistant 316 stainless steel tube shell. The durable tube design withstands high pressure (150 psig/10.2 bar) and high pressure surges. Since the Pirani sensor is miniaturized and employs a Pt thin film on a Si3N4 membrane (instead of a conventional long fragile wire), the transducer can withstand high lev els of mechanical shock.
The Model 2002 is designed for fast response. The micromachined sensing elements have a very small mass and operate in a constant temperature (Pirani) and a constant current (piezo) feed­back mode. This makes response time very fast as compared to other commercially available sensors which have to change the temperature of a significant mass to reflect pressure changes and have a large internal volume which must equalize in pressure with the system before the sensor can reach its final value. The transducer’s small internal volume (<1.5cc) permits rapid pneumatic response to system pressure changes. Further, the small geometry of the transducer prevents thermal convection currents which allows the sensor to be mounted in any orientation without calibration shifts.
Model 2002 page 5
1.3 Model 2002 Control Unit
The Model 2002 control unit consists of the power supply, Pirani resistance bridge control, signal processing and display circuitry. The user interface consists of a flat panel display with smooth operat­ing tactile switches and large green LED digits (0.56"/14.2mm). The pressure is displayed in scientific format using a 3 digit mantissa and a 1 digit (plus the sign) exponent. The pressure can be displayed in Torr, mbar or Pascal. The instrument comes standard with 2 alarm setpoints that have TTL level digital outputs. An optional dual relay board can be internally mounted to trip on the alarm setpoints to turn on/off various equipment such as pumps, valves, heaters, bakeout ovens and safety equipment. An optional 4-20 mA or 0-10V analog output board is also available. The Model 2002 instrument accepts an external TTL level "Remote Zero" to zero the instrument when another instrument (e.g. an
-5
ion gauge) indicates that the pressure is less than 1x10
Torr. The standard analog output is linear per decade directly corresponding to the control unit’s display. The Model 2002 comes calibrated for nitrogen. Conversion factors for other gases are selectable by the user (see section 3.5). This direct readout display allows for flexibility in user system configurations which operate with various gases.
An optional serial interface board can be installed to provide RS232 or RS485 support. The Model 2002 control unit fits a standard DIN 43700 insert (1/4 DIN) and mounts into a 3.62 in. (92 mm) square hole. The case is constructed of NORYL rated UL-94-V1 (self-extinguishing plastic) and is
equipped with mounting brackets that are adjustable up to 3/16” (4.8mm).
1.4 Specifications
Measuring range................................................................................................ 1x10-4 to 10+3 Torr
1.3x10-4 to 1.3x10+3 mbar
Accuracy (N2, T=230C) ..................................................+ 15% of reading (5x10-4 to 3x10
+ 1.5% of reading (3x10+1 to 1x10
Ambient temperature operating range............................................................................. 00 to 500C
Bakeout temperature (with transducer electronics removed)................................................ 2500C
Analog output................................................................ 1.0-4.5 Volts (0.5 VDC linear per decade)
Process control ................................................................ 2 TTL outputs (1 TTL auto zero input)
Digital readout................................................................ Four green LED - resolution is three digit
!
(plus exponent)
Equipment operating range............................................................... 120/240 V~, 0.25A, 50/60 Hz
A 6 ft., 3-wire power cord is provided with each unit.
Fuse ..............................................................................................................................250 V , 1/4 A.
Transducer mounting................................................................ Any position without recalibration.
T ransducer internal volume .................................................................................................. <1.5cc
Wetted material ........................................................................................Au, Si3N4, Si, Pt, PyrexTM,
KovarTM and 316 stainless steel
Stability with voltage ....................................................................Variation in reading undetectable
as power fluctuates within power specification.
+1 +3
Tor r) Tor r)
page 6 Model 2002
Weight (display, cable, and sensor) ............................................... Net: 2.5 lbs., Shipping: 3.4 lbs.
T ransducer response time .................................................. Typically, less than 150 msec for a step
change from 10-3 Torr to 10+3 Tor r.
Positive pressure................................................................150 psig, Recalibration may be required
if unit is exposed to pressures greatly exceeding measuring range.
This section is designed to assist in getting a new pressure gauge into operation as quickly and easily as possible. Please read the following instructions thoroughly before installing the instru­ment.
2.1 Receiving Inspection
Carefully unpack the Hastings Model 2002 Instrument (part # HPM-2002), transducer (part #HPM-2002s) and cable (part # CB 2002). Inspect all items for any obvious signs of damage due to shipment. Immediately advise the carrier who delivered the shipment if any damage is suspected.
Compare each component shipped against the packing list. Ensure that all parts are present (i.e. transducer, power supply, cables, etc.). In addition to the components listed on the packing list, the shipment includes: AC power cord, panel mounting hardware, 9-pin female D connector. Optional equipment or accessories will be listed separately on the packing list.
2.2 Quick Start
SECTION 2
Installation
FF
olloollo
w this procedure to get yw this procedure to get y
F
ollo
w this procedure to get y
FF
olloollo
w this procedure to get yw this procedure to get y
1.1.
RemoRemo
vv
1.
Remo
1.1.
RemoRemo
2.2.
TT
2.
T
2.2.
TT
likely to occur,likely to occur,
likely to occur,
likely to occur,likely to occur,
3.3.
When installing 1/8”When installing 1/8”
3.
When installing 1/8”
3.3.
When installing 1/8”When installing 1/8”
the tube stem.the tube stem.
the tube stem.
the tube stem.the tube stem.
4.4.
Connect transducer cable (orange) to controller .Connect transducer cable (orange) to controller .
4.
Connect transducer cable (orange) to controller .
4.4.
Connect transducer cable (orange) to controller .Connect transducer cable (orange) to controller .
5.5.
Connect other end of transducer cable to transducerConnect other end of transducer cable to transducer
5.
Connect other end of transducer cable to transducer
5.5.
Connect other end of transducer cable to transducerConnect other end of transducer cable to transducer
is adequais adequa
is adequa
is adequais adequa
6.6.
If yIf y
6.
If y
6.6.
If yIf y
be abe a
ttached.ttached.
be a
ttached.
be abe a
ttached.ttached.
7.7.
I/O porI/O por
7.
I/O por
7.7.
I/O porI/O por
8.8.
Connect the Connect the
8.
Connect the
8.8.
Connect the Connect the
120/240120/240
120/240
120/240120/240
9.9.
TT
9.
T
9.9.
TT
“T“T
oror
r”r”
“T
or
r”
“T“T
oror
r”r”
10.10.
FF
10.
F
10.10.
FF
dodo
wn to lown to lo
do
wn to lo
dodo
wn to lown to lo
tion for one hourtion for one hour
tion for one hour
tion for one hourtion for one hour
e sensor from protective sensor from protectiv
v
e sensor from protectiv
vv
e sensor from protective sensor from protectiv
ransducer tube maransducer tube ma
ransducer tube ma
ransducer tube maransducer tube ma
then the por then the por
then the por
then the por then the por
te for proper operate for proper opera
te for proper opera
te for proper operate for proper opera ou haou ha
vv
e computer interface or process control options,e computer interface or process control options,
ou ha
v
e computer interface or process control options,
ou haou ha
vv
e computer interface or process control options,e computer interface or process control options,
t diagt diag
ram is gram is g
t diag
ram is g
t diagt diag
ram is gram is g
AA
C poC po
A
C po
AA
C poC po
VV
AA
C 50/60 HzC 50/60 Hz
V
A
C 50/60 Hz
VV
AA
C 50/60 HzC 50/60 Hz
urur
n on pon on po
ur
n on po
urur
n on pon on po
(default).(default).
(default).
(default).(default).
or best accuracor best accurac
or best accurac
or best accuracor best accurac
w 10w 10
w 10
w 10w 10
-6-6
-6
-6-6
..
.
..
ww
er ser s
witch.witch.
w
er s
witch.
ww
er ser s
witch.witch.
yy
,,
y
,
yy
,,
TT
oror
r (if possible).r (if possible).
T
or
r (if possible).
TT
oror
r (if possible).r (if possible).
..
.
..
our newour new
our new
our newour new
y be installed in any be installed in an
y be installed in an
y be installed in any be installed in an
t should be ort should be or
t should be or
t should be ort should be or
NPT style transducer tube, NPT style transducer tube,
NPT style transducer tube,
NPT style transducer tube, NPT style transducer tube,
tion.tion.
tion.
tion.tion.
iviv
en in the manual.en in the manual.
iv
en in the manual.
iviv
en in the manual.en in the manual.
ww
er cord.er cord.
w
er cord.
ww
er cord.er cord.
Gauge is no Gauge is no
Gauge is no
Gauge is no Gauge is no
the gauge should no the gauge should no
the gauge should no
the gauge should no the gauge should no
Model 2000 vModel 2000 v
Model 2000 v
Model 2000 vModel 2000 v
e package packag
e packag
e package packag
ientaienta
ienta
ientaienta
The Model 2002 automaThe Model 2002 automa
The Model 2002 automa
The Model 2002 automaThe Model 2002 automa
w reading pressure in units ofw reading pressure in units of
w reading pressure in units of
w reading pressure in units ofw reading pressure in units of
w be zeroed.w be zeroed.
w be zeroed.
w be zeroed.w be zeroed.
Ideally the gauge should be operaIdeally the gauge should be opera
Ideally the gauge should be opera
Ideally the gauge should be operaIdeally the gauge should be opera
ing.ing.
ing.
ing.ing. y ory or
y or
y ory or
ted doted do
ted do
ted doted do
acuum gauge up and racuum gauge up and r
acuum gauge up and r
acuum gauge up and racuum gauge up and r
ientaienta
tion.tion.
Ho Ho
ww
ienta
tion.
ientaienta
tion.tion.
wnwwnw
wnw
wnwwnw
use the 7/16” use the 7/16”
use the 7/16”
use the 7/16” use the 7/16”
evev
Ho
w
ev
Ho Ho
ww
evev
ard.ard.
ard.
ard.ard.
..
A finger tight connectionA finger tight connection
.
A finger tight connection
..
A finger tight connectionA finger tight connection
these cables can no these cables can no
these cables can no
these cables can no these cables can no
tically adjusts fortically adjusts for
tically adjusts for
tically adjusts fortically adjusts for
Pump the vPump the v
Pump the v
Pump the vPump the v
unning.unning.
unning.
unning.unning.
er if condensaer if condensa
er if condensa
er if condensaer if condensa
wrench fla wrench fla
wrench fla
wrench fla wrench fla
acuum systemacuum system
acuum system
acuum systemacuum system
ted in thisted in this
ted in this
ted in thisted in this
tion istion is
tion is
tion istion is
ts onts on
ts on
ts onts on
ww
w
ww
condi-condi-
condi-
condi-condi-
11.11.
Place the Model 2002 in the Place the Model 2002 in the
11.
Place the Model 2002 in the
11.11.
Place the Model 2002 in the Place the Model 2002 in the
up and doup and do
up and do
up and doup and do
12.12.
12.
12.12. mode smode s
mode s
mode smode s The gauge is now ready for normal operation and the display is a true indication of the system
pressure (P < 8 Torr is N2 equivalent).
wn arwn ar
wn ar
wn arwn ar
Press the Press the
Press the
Press the Press the
witch.witch.
witch.
witch.witch.
roro
ws simultaneously (“CALws simultaneously (“CAL
ro
ws simultaneously (“CAL
roro
ws simultaneously (“CALws simultaneously (“CAL
“ZER“ZER
O”O”
s s
“ZER
“ZER“ZER
witch once and returwitch once and retur
O”
s
witch once and retur
O”O”
s s
witch once and returwitch once and retur
“CAL“CAL
“CAL
“CAL“CAL
””
mode b mode b
mode b
””
mode b mode b
””
light is no light is no
light is no
””
light is no light is no
y using the mode sy using the mode s
y using the mode s
y using the mode sy using the mode s
w flashing).w flashing).
w flashing).
w flashing).w flashing).
n to the n to the
n to the
n to the n to the
“R“R
“R
“R“R
UN”UN”
UN”
UN”UN”
witch.witch.
Press the Press the
witch.
Press the
witch.witch.
Press the Press the
mode using the mode using the
mode using the
mode using the mode using the
Model 2002 page 7
2.3 Transducer Installation
The transducer may be installed in any orientation. Although the transducer is rugged and will perform well in many harsh environments, the tube should be installed in a clean and careful manner. The tube is configured with the vacuum fitting requested. If your vacuum environment is highly contaminated or has unique fitting requirements, a Hastings filter or special adapter may be needed. Please contact the Hastings Instruments Sales Department for assistance in your system configuration.
2.4 Control Unit Installation
EnEn
vironment:vironment:
En
vironment:
EnEn
vironment:vironment:
Indoor use Altitude up to 2000 meters Operating temperature range from 50 to 400C Maximum relative humidity: 80% for tempera tures up to 310C decreasing linearly to 50%
relative humidity at 400C Installation category II
PP
anel Mount Instranel Mount Instr
P
anel Mount Instr
PP
anel Mount Instranel Mount Instr
The control unit can be panel mounted. See detail on page 43. The hole dimensions on the panel are 3.62" x 3.62"(92.8mm x 92.8 mm). Slide the neoprene gasket that was shipped with the control unit onto the case from the back. Slide the controller through the panel cutout. Hold the hardware against the side and tighten the two screws.
uctions:uctions:
uctions:
uctions:uctions:
TT
ransducer Cable ransducer Cable
T
ransducer Cable
TT
ransducer Cable ransducer Cable
The threaded connector attaches to the transducer. A finger tight connection is adequate for proper operation. The 9-pin male “D” connector attaches to the back panel (See Figure 4.2). The transducer cable connects to the left hand connector when looking at the back.
I/O Cable I/O Cable
I/O Cable
I/O Cable I/O Cable
The mating male plug to the I/O connector is supplied with the unit. An I/O cable can be wired with the 9-pin female “D” connector to use the analog output, digital alarms or the remote zero functions. The connector will accept 20 gauge or smaller wire. The pinout is shown in Figure 4.2 (back panel figure). A detailed description of these pins is given in Section 4.
Attachment:Attachment:
Attachment:
Attachment:Attachment:
2.5 Initial Operation
Upon applying power to the control unit a pressure measurement will be given in Torr for nitrogen. However, it is recommended that the user follow the instructions for zeroing and adjusting the output at atmospheric pressure in Section 3.4 ("Cal Mode").
Attachment:Attachment:
Attachment:
Attachment:Attachment:
page 8 Model 2002
3.1 Overall Functional Description
The front panel of the control unit is shown in Figure 3.1. The four circular blue buttons are used for the selection of display readout and the input of data. The green data field displays the data as determined by mode selection. The MODE switch toggles the control unit in a clockwise fashion among the six modes of operation:
RR
UNUN
R
UN Normal operation, pressure is displayed in scientific notation. The analog to digital
RR
UNUN
converter speed can be adjusted and the factory calibration can also be restored in
this mode. See Section 3.2 for further information. HIGHHIGH
HIGH High set point is displayed in scientific notation, above this pressure the HIGH
HIGHHIGH TTL output will be +5V.
LOLO
WW
LO
W Low set point is displayed in scientific notation, below this pressure the LO W TTL
LOLO
WW
output will be +5V.
CALCAL
CAL Pressure is displayed and can be adjusted.
CALCAL GASGAS
GAS Gas number is displayed and selection may be changed.
GASGAS UNITSUNITS
UNITS Units used to display pressure are selected.
UNITSUNITS
SECTION 3
Front Panel Operation
In the RUN, HIGH, and LOW modes, it is possible to see the display indicate that the data is out of range. Overrange is indicated by . x10- . Underrange is indicated by 0.0 x10-. A mea­sured pressure below 1x10 sensor(s), it is possible to see “ .“, “. “ or “ . “. See trouble shooting section (section 8) for more detail.
UP SWITCH
DOWN SWITCH
-4
Torr is indicated by 0.0 x10+. In the case of an unconnected or faulty
HASTINGS INSTRUMENTS DUAL SENSOR V ACUUM GAUGE MODEL 2002
mbar Torr Pa
Model 2002 Control Unit Front Panel
MODE SWITCH ZERO SWITCH
Model 2002 Control Unit Front Panel
Figure 3.1
Model 2002 page 9
All six modes of the Model 2002 have features which can be accessed and modified after bypassing the front panel interlock. The interlock is in place to prevent the accidental corruption of the instrument’s configuration and calibration. The interlock is bypassed using the following steps:
1. Place the instrument in one of the six modes.
2. Simultaneously press the up and down switches.
The mode light will now flash indicating that the interlock has been bypassed.
Scientific NotaScientific Nota
Scientific Nota
Scientific NotaScientific Nota
The Model 2002 measures pressure that spans more than seven decades. In order to easily display these readings, Hastings Instruments has employed scientific notation. In scientific notation, the mantissa (the fixed point part) is multiplied by some power of 10 (given by the exponent).
103 = 1000 10-1 = 0.1
2
10 101 = 10 10-3 = 0.001 100 = 1 10-4 = 0.0001
As an example: 7.60 x 10
"7.60 multiplied by 100" = 760 Torr which is one atmosphere of pressure.
3.2 RUN Mode
NorNor
mal Operamal Opera
Nor
mal Opera
NorNor
mal Operamal Opera
tiontion
tion
tiontion
Powers of 10
= 100 10-2 = 0.01
+2
Tor r
{
mantissa exponent
tiontion
tion
tiontion
The Model 2002 will automatically enter RUN mode upon start-up. This is the mode for normal operation and the mode in which the instrument will typically spend most of its time. In the run mode, the Model 2002 unit will continuously monitor the pressure and update the alarm conditions at the speed of the A/D con verter and will update the display about four times per second.
A/D Speed A/D Speed
A/D Speed
A/D Speed A/D Speed
The Model 2002 utilizes a 24 bit sigma-delta analog to digital converter. This type of converter averages the input over time to reject interference. Longer averaging times result in more accurate readings. Increasing the A/D speed will respond to sudden pressure changes faster.
While in the RUN mode, the user can adjust the A/D speed of operation. At start-up, the instrument will be operating with an A/D frequency of about 60 Hz. The speed can be adjusted one step faster or slower. Each step will increase or decrease the A/D frequency by a factor of two.
To adjust the A/D speed, press the MODE switch until the RUN light is illuminated. Then simulta­neously press the UP and DOWN switches. The RUN light will now flash indicating tha t the interlock has been bypassed and the A/D speed can be adjusted. Push the UP switch to speed up the A/D conv erter one step and decrease the response time . This will result in a faster triggering of alarms or relays if a sudden catastrophic failure occurs. Pressing the switch after the limit is reached has no effect.
Pushing the down switch will slow down the A/D converter and increase its accuracy. This will improve the ability to resolve low pressure readings. If a certain speed is desired, press the UP or DO WN switch three times and then press the other switch to select the other speed.
On older units. There will be no indication of the operating speed other than the least significant digits of the display will tend to run at higher speeds. On units with software version 1.60 and
AdjustAdjust
Adjust
AdjustAdjust
page 10 Model 2002
above, press the ZERO switch to display the decimation ratio (filtering). The larger the ratio, the longer the response time.
EEPREEPR
EEPR
EEPREEPR
There are interlocks built into the CAL mode to minimize the chance of accidental alteration of the calibration coefficients, but if calibration alteration does occur, the EEPROM calibration data can be restored in the RUN mode by pressing the UP and DOWN switches simultaneously (RUN light will flash), then press the “ZERO” switch, UP and DOWN switches (all three) simultaneously.
This will not correct the calibration if the tube has become contaminated or damaged. This reset will not affect the gas selection, units or setpoints.
Also see units mode for default calibration restoration.
OM CalibraOM Calibra
OM Calibra
OM CalibraOM Calibra
tion Restoration Restora
tion Restora
tion Restoration Restora
tiontion
tion
tiontion
3.3 HIGH and LOW Set Point Modes
The Model 2002 provides TTL outputs for process control. The I/O cable is attached via a 9-pin sub-D connector to the rear panel of the control unit. The pinout is shown in Figure 4.2.
To view the high set point, place the Model 2002 in the HIGH mode by pressing the MODE switch to illuminate the HIGH light and no other mode light. The display will show the set point selected. During normal operation the alarm light will illuminate and the TTL output (pin # 3) will go high (+5 V) if the pressure
Similarly, to view the low set point access LOW mode by pressing the mode switch until the LOW light is lit and no other mode light is illuminated. During normal operation, the LOW light will illuminate and the LOW alarm TTL output (pin # 4) will go high (+5 V) if the pressure becomes less less
less than the set point.
less less
exceedsexceeds
exceeds the set point.
exceedsexceeds
The alarm lights cannot indicate an alarm condition while in the set point modes. Therefore, it is advisable not to leave the instrument in these modes for extended periods.
To adjust a set point, press the mode switch until the set point is viewed. Then simultaneously press the UP and DOWN switches. The active mode light will now flash indicating that the interlock has been bypassed and the set point can be adjusted. The display shows the present set point.
Enter the new set point by using the UP and DOWN switches. Press and hold the UP switch to increase the set point and the DOWN switch to decrease the setpoint. Allow a few seconds for the circuitry to respond. If the button is held down for and extended period of time the adjustment rate will start to increase. Release the button and press again to get finer control of the exact trigger point.
At this point, the new set point is in temporary memory. If the instrument were unplugged now, the Model 2002 would revert back to the original set point upon restarting. Return the 2002 to the RUN mode to store the set point in permanent memory.
Once the interlock has been bypassed in the HIGH mode, it will stay active until the CAL mode is entered. This will allow the user to set both setpoints without repeating the interlock bypass.
3.4 CAL Mode
Optimal performance of the Model 2002 is achieved by performing in situ adjustments to the calibration coefficient in the CAL mode. There are three calibration coefficients. These are the zero coefficient, the midrange coefficient and the atmosphere coefficient. Once a tube has been fully calibrated the midrange coefficient should never need further adjustment, but it may be helpful to adjust the zero coefficient or the atmosphere coefficient under certain situations. Even though the operator inputs may be identical for adjustment of all three coefficients, microprocessor will detect
Model 2002 page 11
Maintainance and Repair 4-3
the power level of the tube and adjust the proper coefficient for pressure level of the adjustment.
The CAL MODE presupposes that the operator is applying a known pressure of the correct gas composition (see GAS MODE). The factory calibration points are 800 Torr, 7 Torr, and <1e-6 Torr. The users calibration points are not required to be exactly those values, but should be somewhat close, and must be within the ranges shown in the following figure. The HPM-2002 detects the voltage signal within the sensor tube, which is converted and displayed as a pressure reading. The resulting pressure reading determines which of the three coefficients will be ad­justed.
page 12 Model 2002
To adjust a calibration coefficient press the MODE switch until the CAL light is illuminated. Then simultaneously press the UP and DOWN switch. The CAL light will now flash indicating that the interlock has been bypassed and the calibration can be adjusted. Press and hold the UP switch to increase the displayed v alue or the DOWN switch to decrease the displayed value. If the switch is held down for an extended period of time the adjustment rate will start to increase. Release the switch and then press and hold again to regain finer control of the displayed number.
An important point to be noted is the fact that after the adjustment, the new calibration coeffi­cient is only in temporary memory. If the instrument were unplugged at this point, it would rev ert
back to the original displayed value upon restarting. Place the Model 2002 back in the RUN mode to store all of the current calibration coefficients in permanent memory.
To perform a full calibration on the Model 2002 system, first use the Zero Coefficient Adjust­ment Procedure. Followed by the Midrange Coefficient Adjustment Procedure, and finally perform the Atmosphere Coefficient Adjustment Procedure.
Zero Coefficient Zero Coefficient
Zero Coefficient
Zero Coefficient Zero Coefficient
The zero coefficient corrects for the constant power level which is present ov er the entire pressure range. Typically, this adjustment corrects for low pressure errors. The instrument will need to be re-zeroed often if measurements are being made in the 10-4 Torr range, especially if the ambient temperature changes. The transducer may have a temperature coefficient of up to 2x10-4 Tor r/oC. The instrument remote zero input will allow an external gauge such as an ion gauge to automati­cally re-zero the Model 2002 whenever the pressure drops below the desired pressure level (if it has a TTL output).
NOTE: Do not attempt to zero the Model 2002 in pressures abov e 10-2 Torr; the micro­processor will not accept a zero abov e this pressure.
To manually adjust the zero use the following procedure:
1. If possible, evacuate the vacuum system into the low 10-6 T orr (1.33x10-6 mbar) range or as
low as possible below 10
2. Allow the sensor to operate in this condition for a minimum of 15 minutes.
3. Place the instrument in the CAL mode, then simultaneously press the UP and DOWN switches to bypass the interlock. The CAL light will start to flash indicating that the calibration mode has been activated.
4. Press the ZERO switch once if below 6x10-5 Torr otherwise use the UP and DOWN switch to adjust the display of the Model 2002 to match the pressure indicated by the reference. If using the UP and DOWN switch to adjust the display , wait 20 seconds between presses at very low pressure to allow the adjustment to take affect the data to settle.
AdjustmentAdjustment
Adjustment
AdjustmentAdjustment
-4
Torr.
5. The unit is now fully zeroed. Place the Model 2002 back in the RUN mode to store the zero in permanent memory.
Midrange Coefficient Midrange Coefficient
Midrange Coefficient
Midrange Coefficient Midrange Coefficient
The midrange coefficient corrects for errors in the slope of the power curve of the thin film Pirani. Typically, this is due to the geometry of a particular sensor and will only need to be performed once in the lifetime of the sensor. This adjustment might be needed if a full calibration is being performed in a gas other than nitrogen. before making this adjustmentbefore making this adjustment
before making this adjustment.
before making this adjustmentbefore making this adjustment To adjust the midrange coefficient use the following procedure:
1. Evacuate the vacuum chamber and refill with the desired gas to a pressure of 7 Torr, as
indicated by a reference vacuum gauge.
2. Press the MODE switch repeatedly until the CAL mode is illuminated.
3. Simultaneously, press both the UP and DOWN switches. The CAL light will start to flash,
signalling that the calibration can now be conducted.
4. Press, hold and release the UP and DOWN switches as necessary to adjust the display to
match the pressure indicated by the reference. Note: It may be necessary to hold the
desired switch down for a few seconds before any movement is seen.
AdjustmentAdjustment
Adjustment
AdjustmentAdjustment
The zero adjustment will need to be perforThe zero adjustment will need to be perfor
The zero adjustment will need to be perfor
The zero adjustment will need to be perforThe zero adjustment will need to be perfor
medmed
med
medmed
Model 2002 page 13
5. Press and release the MODE switch as necessary to return the 2002 to the RUN mode and
store the coefficients to permanent memory.
Atmosphere Coefficient Atmosphere Coefficient
Atmosphere Coefficient
Atmosphere Coefficient Atmosphere Coefficient
If a reference high pressure gauge is not available, the ambient barometric pressure acquired from the weather channel or other weather service can be used to adjust the proper reading.
To adjust the atmosphere coefficient, use the following procedure:
1. Press the MODE switch a sufficient number of times to enter the CAL mode.
2. Simultaneously, press both the UP and DOWN switches to bypass the interlock. The CAL light will start to flash, signaling that the calibration can now be conducted.
3. Press, hold and release either of the UP and DOWN switches as necessary to adjust the display to match the pressure indicated by the reference. Note: It may be neces­sary to hold the desired switch down for a few seconds before any movement is seen.
4. Press and release the MODE switch as necessary to return the Model 2002 to the RUN mode and store the coefficients to permanent memory.
3.5 GAS Mode
The Model 2002 can provide true pressure measurements in many gas environments. At pressure levels above 32 Torr where the direct force piezoresistive sensor is operative, the instrument is gas composition insensitive and measures the true pressure regardless of gas composition. The Pirani is gas composition sensitive so below 32 Torr, the actual composition must be known and the Pirani calibrated in that gas. When the vacuum system’s gas composition is dominated by a single gas species (for example, during system venting with an inert gas), the user can enter a gas selection into the Model 2002. To view the gas selection, place the instrument in the GAS mode
AdjustmentAdjustment
Adjustment
AdjustmentAdjustment
page 14 Model 2002
TABLE 1
Gas Mode
Displayed Number
0.0............................... nitrogen
0.1................................... argon
0.2................................. helium
0.3..........................water vapor
0.4.................................custom
and reference Table 1. The Model 2002 is factory set to display N2 pressure readings. To select a different gas, refer to Table 1 to find the number code of the gas to be entered. Press the UP and DOWN switches simultaneously, the GAS light will now flash. Use the UP and DOWN switches to individually select the desired code. Return the Model 2002 to the RUN mode.
Gas
Note: Factory calibration is performed using nitrogen only. Displayed pressure mea­surements using other gases are based on established g as thermal conductivity data.
3.6 UNITS Mode
The Model 2002 can display the measured pressure in different pressure units, Torr, mbar or Pascal. To switch between these units press the MODE switch until the UNITS field is lit. Press the UP and DOWN switches simultaneously. The UNITS light will now flash. Use the UP and DO WN switches individually to select the desired unit.
The units mode is also used to display the RS-485 address, (see section 6.3). A commonly used unit is mTorr which is also known as the micron (short for micron of Hg).
One Torr is equal to 1000 mTorr. The table below illustrates how to interpret the display in mTorr.
Table 2 gives conversion factors for the most often used pressure units.
mTorr Conversions 1 mTorr = 1.00 x 10-3 Tor r 10 mTorr = 1.00 x 10-2 Tor r 100 mT orr = 1.00 x 10
-1
Tor r
TABLE 2
Torr
Torr (mm Hg) mTorr (micron Hg) atm (athmosphere) mbar (millibar) Pa
1 1.32x10
-3
10
760
0.75
7.5x10
-3
psi (lb/in ) 51.72
Default CalibraDefault Calibra
Default Calibra
Default CalibraDefault Calibra
The UNITS mode is also used to load default calibration data in rare cases where the transducer EEPROM has become corrupted. The default calibration data should be considered an approxi­mate calibration because the values entered in the controller’s software are based on averaged data collected over a large number of transducers. While this type of calibration download is not a true calibration, it does allow the user to return the unit to a reasonable status from which a full calibration can be easily performed.
tion Restoration Restora
tion Restora
tion Restoration Restora
tiontion
tion
tiontion
atm
-3
1.3332 133 .32
1.32x10
9.869x10
9.869x10
6.805x10 68.953
-6
0.00133
1
1013.23
-4 -2
-6
-2
1
0.01
.13332
1.013x10 100
1
6895.3
psiPambar
1.934x10
1.934x10
14.7
1.45x10
1.45x10 1
-2
-5
-4
Model 2002 page 15
This page was intentionally left blank
page 16 Model 2002
A schematic of the rear panel of the Model 2002 control unit is shown in Figure 4.2 ( Rear Panel Figure). The transducer is connected to the control unit via an female 9-pin “D” connector on the rear panel. An I/O cable can be connected via a male 9-pin “D” connector on the rear panel. The I/O cable provides an analog output signal, a remote zero input capability, high and low alarm TTL outputs, and a + 5 VDC Power Supply Output. The I/O cable also provides an analog ground and a digital common.
4.1 Remote Zero Input
The Model 2002 can be zeroed remotely via the I/O cable using the remote zero input line. Upon setting this line to TTL low level, the instrument will set its current pressure reading to zero. A common applica­tion of this feature is to utilize a process control output from a high vacuum gauge (for example an ion gauge controller) to automatically zero the Model 2002 whenever the pressure is below the low 10-5 Torr range. A Digital Common is provided on pin # 5 which can be used with relay contacts or switch contacts to remotely zero the instrument.
4.2 Analog Output
The Model 2002 provides an analog voltage output signal (0.5 VDC linear per decade) for process control and for pressure monitoring. A plot of the analog output signal as a function of pressure is shown in Figure 4.1 . Note that the output voltage is directly proportional to the pressure over each complete decade (see detailed region in Figure 4.1) and that each decade has a different linear slope.
SECTION 4
Rear Panel Description
Above an indicated pressure of 103 Torr, the analog output signal will be +5 VDC. Below the minimal
displayable pressure of 1 x 10-4 Torr, the analog output will be 1.0 VDC.
(EXP + 6) (MANT - 1)
V =
The analog output voltage is given by the following equation: Where EXP is equal to the exponent of the displayed pressure.
218
+
EXP = -2 MANT = 1.53
(1.53 - 1)
+
18
example: p = 1.53 x 10
(-2 + 6)
V =
-2
2
2 + 0.0294
=
2.0294 VDC
=
EXP = +2 MANT = 9.87
(+2 + 6)
V =
4 + 0.4928
=
4.4928 VDC
=
2
(9.87 - 1)
+
18
Model 2002 page 17
Loading...
+ 37 hidden pages