The servicing instructions are for
use by qualified personnel only.
To avoid personal injury, do not
perform any servicing unless you
are qualified to do so. Refer to all
safety summaries prior to
performing service.
Cleaning and decontamination.....................................................13
Returning your probe for recalibration or repair..........................13
TRCP0300, TRCP0600, TRCP3000 User Manual i
Table of Contents
ii TRCP0300, TRCP0600, TRCP3000 User Manual
Important safety information
This manual contains information and warnings that must be followed
by the user for safe operation and to keep the product in a safe
condition.
To safely perform service on this product, see the Service safetysummary that follows the General safety summary.
General safety summary
Use the product only as specified. Review the following safety
precautions to avoid injury and prevent damage to this product or any
products connected to it. Carefully read all instructions. Retain these
instructions for future reference.
This product shall be used in accordance with local and national codes.
For correct and safe operation of the product, it is essential that you
follow generally accepted safety procedures in addition to the safety
precautions specified in this manual.
The product is designed to be used by trained personnel only.
Only qualified personnel who are aware of the hazards involved should
remove the cover for repair, maintenance, or adjustment.
Before use, always check the product with a known source to be sure it
is operating correctly.
This product is not intended for detection of hazardous voltages.
Use personal protective equipment to prevent shock and arc blast injury
where hazardous live conductors are exposed.
While using this product, you may need to access other parts of a larger
system. Read the safety sections of the other component manuals for
warnings and cautions related to operating the system.
When incorporating this equipment into a system, the safety of that
system is the responsibility of the assembler of the system.
The intended use of this product is for the measurement of AC current.
TRCP0300, TRCP0600, TRCP3000 User Manual iii
Important safety information
To avoid fire or personal injury
Use proper power cord. Use only the power cord specified for this
product and certified for the country of use.
Use proper power cord. Use only the power cord specified for this
product and certified for the country of use. Do not use the provided
power cord for other products.
Ground the product.
Power disconnect.
Use proper AC adapter. Use only the AC adapter specified for this
product.
Connect and disconnect properly.
Observe all terminal ratings.
Observe all terminal ratings. To avoid fire or shock hazard, observe all
rating and markings on the product. Consult the product manual for
further ratings information before making connections to the product.
Do not apply a potential to any terminal, including the common
terminal, that exceeds the maximum rating of that terminal.
Do not operate without covers. Do not operate this product with covers or
panels removed, or with the case open. Hazardous voltage exposure is
possible.
Avoid exposed circuitry. Do not touch exposed connections and
components when power is present.
Do not operate with suspected failures. If you suspect that there is damage
to this product, have it inspected by qualified service personnel.
Disable the product if it is damaged. Do not use the product if it is
damaged or operates incorrectly. If in doubt about safety of the
product, turn it off and disconnect the power cord. Clearly mark the
product to prevent its further operation.
Before use, inspect probes, test leads, and accessories for mechanical
damage and replace when damaged. Do not use probes or test leads if
they are damaged, if there is exposed metal, or if a wear indicator
shows.
iv TRCP0300, TRCP0600, TRCP3000 User Manual
Important safety information
Examine the exterior of the product before you use it. Look for cracks
or missing pieces.
Use only specified replacement parts.
Replace batteries properly. Replace batteries only with the specified type
and rating.
Do not operate in wet/damp conditions. Be aware that condensation may
occur if a unit is moved from a cold to a warm environment.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry. Remove the input signals before
you clean the product.
Provide proper ventilation. Refer to the installation instructions in the
manual for details on installing the product so it has proper ventilation.
Slots and openings are provided for ventilation and should never be
covered or otherwise obstructed. Do not push objects into any of the
openings.
Provide a safe working environment. Always place the product in a
location convenient for viewing the display and indicators.
Be sure your work area meets applicable ergonomic standards. Consult
with an ergonomics professional to avoid stress injuries.
Probes, coil, and test leads
Remove all coils and BNC cables that are not in use
Use only correct Measurement Category (CAT), voltage, temperature,
altitude, and amperage rated coils and BNC cables for any
measurement.
Beware of high voltages.
Connect and disconnect properly.
Connect and disconnect properly. De-energize the circuit under test
before connecting or disconnecting the current probe.
Inspect the coil and accessories.
TRCP0300, TRCP0600, TRCP3000 User Manual v
Important safety information
Service safety summary
The Service safety summary section contains additional information
required to safely perform service on the product. Only qualified
personnel should perform service procedures. Read this Service safetysummary and the General safety summary before performing any
service procedures.
To avoid electric shock. Do not touch exposed connections.
Do not service alone. Do not perform internal service or adjustments of
this product unless another person capable of rendering first aid and
resuscitation is present.
Disconnect power. To avoid electric shock, switch off the product power
and disconnect the power cord from the mains power before removing
any covers or panels, or opening the case for servicing.
Use care when servicing with power on. Dangerous voltages or currents
may exist in this product. Disconnect power, remove battery (if
applicable), and disconnect test leads before removing protective
panels, soldering, or replacing components.
Verify safety after repair. Always recheck ground continuity and mains
dielectric strength after performing a repair.
Terms in the manual
These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that
could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that
could result in damage to this product or other property.
vi TRCP0300, TRCP0600, TRCP3000 User Manual
Important safety information
Terms on the product
These terms may appear on the product:
■
DANGER indicates an injury hazard immediately accessible as
you read the marking.
■
WARNING indicates an injury hazard not immediately accessible
as you read the marking.
■
CAUTION indicates a hazard to property including the product.
Symbols on the product
When this symbol is marked on the product, be sure to
consult the manual to find out the nature of the potential
hazards and any actions which have to be taken to avoid
them. (This symbol may also be used to refer the user to
ratings in the manual.)
The following symbols may appear on the product:
TRCP0300, TRCP0600, TRCP3000 User Manual vii
Important safety information
viii TRCP0300, TRCP0600, TRCP3000 User Manual
Compliance Information
This section lists the EMC (electromagnetic compliance), safety, and
environmental standards with which the instrument complies.
Safety compliance
This section lists the safety standards with which the product complies
and other safety compliance information.
EU declaration of conformity – low voltage
Compliance was demonstrated to the following specification as listed
in the Official Journal of the European Union:
Low Voltage Directive 2014/35/EU.
■
EN 61010-1. Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use – Part 1: General
Requirements.
■
EN 61010-2-032. Particular requirements for handheld current
clamps for electrical measurement and test equipment.
Additional compliances
■
IEC 61010-1. Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use – Part 1: General
Requirements.
■
IEC 61010-2-032. Particular requirements for handheld current
clamps for electrical measurement and test equipment.
Equipment type
Test and measuring equipment.
TRCP0300, TRCP0600, TRCP3000 User Manual ix
Compliance Information
Pollution degree description
A measure of the contaminants that could occur in the environment
around and within a product. Typically the internal environment inside
a product is considered to be the same as the external. Products should
be used only in the environment for which they are rated.
■
Pollution Degree 1. No pollution or only dry, nonconductive
pollution occurs. Products in this category are generally
encapsulated, hermetically sealed, or located in clean rooms.
■
Pollution Degree 2. Normally only dry, nonconductive pollution
occurs. Occasionally a temporary conductivity that is caused by
condensation must be expected. This location is a typical office/
home environment. Temporary condensation occurs only when the
product is out of service.
■
Pollution Degree 3. Conductive pollution, or dry, nonconductive
pollution that becomes conductive due to condensation. These are
sheltered locations where neither temperature nor humidity is
controlled. The area is protected from direct sunshine, rain, or
direct wind.
■
Pollution Degree 4. Pollution that generates persistent conductivity
through conductive dust, rain, or snow. Typical outdoor locations.
Pollution degree
Pollution Degree 2 (as defined in IEC 61010-1). Note: Rated for
indoor, dry location use only.
IP rating
IP20 (as defined in IEC 60529).
Measurement and overvoltage category descriptions
Measurement terminals on this product may be rated for measuring
mains voltages from one or more of the following categories (see
specific ratings marked on the product and in the manual).
■
Measurement Category II. For measurements performed on
circuits directly connected to the low-voltage installation.
■
Measurement Category III. For measurements performed in the
building installation.
■
Measurement Category IV. For measurements performed at the
source of low-voltage installation.
x TRCP0300, TRCP0600, TRCP3000 User Manual
Compliance Information
Environmental compliance
This section provides information about the environmental impact of
the product.
Product end-of-life handling
Observe the following guidelines when recycling an instrument or
component:
Equipment recycling. Production of this equipment required the
extraction and use of natural resources. The equipment may contain
substances that could be harmful to the environment or human health if
improperly handled at the product’s end of life. To avoid release of
such substances into the environment and to reduce the use of natural
resources, we encourage you to recycle this product in an appropriate
system that will ensure that most of the materials are reused or recycled
appropriately.
This symbol indicates that this product complies with the
applicable European Union requirements according to
Directives 2012/19/EU and 2006/66/EC on waste
electrical and electronic equipment (WEEE) and
batteries. For information about recycling options, check
the Tektronix Web site (www.tek.com/productrecycling).
TRCP0300, TRCP0600, TRCP3000 User Manual xi
Compliance Information
xii TRCP0300, TRCP0600, TRCP3000 User Manual
Specifications
The Tektronix Rogowski Current Probes, the TRCP0300, TRCP0600,
and TRCP3000, combines high bandwidth performance with minimal
disruption to the circuit under test. Refer to these instructions when
using your probe. They will help you obtain the best and safest
performance from your product.
The Specifications in this section were valid at the time the product
was shipping. Go to www.tek.com for updated specifications for this
product.
TRCP0300TRCP0600TRCP3000
Frequency range9 Hz to 30 MHz12 Hz to 30 MHz1 Hz to 16 MHz
Peak current300 A600 A3000 A
Minimum current250 mA500 mA500 mA
Sensitivity20 mV/A10 mV/A2 mV/A
External probe attenuation
setting on scope
Accuracy (typical)+/- 2% of reading+/- 2% of reading+/- 2% of reading
Coil diameter1.7 mm4.5 mm8.5 mm
Termination1 MΩ BNC1 MΩ BNC1 MΩ BNC
50X100X500X
Peak output voltage±6 V (corresponding to ±Peak current)
Output cable0.5 m, BNC to BNC, 50 Ω
Absolute di/dt ratings
TRCP030070 kA/µs (peak) | 1.0k A/µs (rms)
TRCP0600100 kA/µs (peak) | 1.2 kA/µs (rms)
TRCP3000100 kA/µs (peak) | 1.2 kA/µs (rms)
TRCP0300, TRCP0600, TRCP3000 User Manual 1
Specifications
Operating temperature range
Integrator electronics0 °C to +40 °C
Coil and cable
(TRCP0300 and
TRCP0600)
Coil and cable
(TRCP3000)
Accuracy, typical5% to 100% peak current
CalibrationNominally ±0.2% with the conductor central in the Rogowski loop
Positional accuracy±2% variation of accuracy with conductor position in the loop
DC socket type3.5 mm OD and 1.3 mm ID jack socket, center positive
Only use the external
adaptor supplied with
product.
Adaptor mains input
voltage
Adaptor mains input
frequency
Adaptor supply
voltage/current
Mains input voltage
fluctuation
–40 °C to +125 °C
–20 °C to +100 °C
See the calibration certificate for further details.
See Obtaining the best measurement on page 12
100 VAC to 240 VAC
50 Hz to 60 Hz
12V DC/450mA Max @ 100VAC input
Up to 10%
2 TRCP0300, TRCP0600, TRCP3000 User Manual
Specifications
Operation with battery
Battery supply, B
version
Battery life, B version25 hours, typical
SafetyEN61010-1:2010
Voltage rating (TRCP0300)300 V CAT II
Voltage rating (TRCP0600)600 V CAT III
Voltage rating (TRCP3000)1000 V CAT III
EnvironmentalRated for indoor use
AltitudeUp to 2,000 m (6,562 ft)
Installation CategoryII
Pollution Degree2
Maximum relative
humidity
4 x 1.5 V AA alkali batteries
EN61010-2-032:2012
80% up to 31 °C Decreasing linearly to 50% at 40 °C
TRCP0300, TRCP0600, TRCP3000 User Manual 3
Specifications
4 TRCP0300, TRCP0600, TRCP3000 User Manual
Tektronix Rogowski Current Probe
Index number
1 Rogowski coil (probe)
2 Probe free end
3 Ferrule
4 Connecting cable
5 DC power socket
6 DC power status LED
7 Enclosure
TRCP0300, TRCP0600, TRCP3000 User Manual 5
Description
Tektronix Rogowski Current Probe
Index numberDescription
8 BNC output socket
9 4 x AA batteries
10 Status indicator LED
GREEN = ON
RED = Batter LOW
11 Power button
Handling instructions
Every effort has been made to ensure that the TRCP models are as
robust as possible. To achieve such an extremely thin Rogowski coil
the probe is necessarily delicate.
To prolong the life of the coil please observe the following handling
instructions.
The free end of the coil is unclipped in the direction shown. While the
illustration below is specifically related to the TRCP0300, Similar
handling instructions apply for the TRCP0600 and TRCP3000.
6 TRCP0300, TRCP0600, TRCP3000 User Manual
Tektronix Rogowski Current Probe
When clipping or unclipping the coil, hold the ferrule as shown in this
image.
WARNING. Never force the free end of the coil into the ferrule. This
could damage the coil insulation.
WARNING. Do not apply force to the cable attached to the coil. This
could damage the coil.
NOTE. When not in use, return the probe coil to the protective case it
was shipped in.
Safety and pre-use checks
WARNING. The transducer does not protect you from electric shock
while applying or removing the probe from hazardous live conductors.
Additional protective means are necessary to avoid electric shock from
hazardous live conductors which cannot be de-energized.
If hazardous live voltages are present and accessible in the
installation, appropriate protective equipment must be used.
WARNING. The Rogowski coil must not be used if the coil outer layer is
damaged or in any way compromised.
TRCP0300, TRCP0600, TRCP3000 User Manual 7
Tektronix Rogowski Current Probe
WARNING. The probe must only be used in conjunction with properly
grounded test equipment.
If the probe is used in a manner not specified by the manufacturer, the
protection provided by the equipment may be impaired.
The Rogowski coil
WARNING. The integrity of the insulation around the Rogowski coil
itself must be visually inspected before use, and the probe must not be
used if there are signs of damage.
WARNING. When bending the flexible coil around a conductor, avoid
tight bends and sharp edges that could damage the coil.
WARNING. The voltage rating (safe PEAK working voltage) is clearly
labeled on the cable adjacent to the coil. For the TRCP0300,
TRCP0600, and TRCP3000, see the coil insulation and voltage rating
in Specifications. Never use at voltages greater than this value.
WARNING. Voltage ratings are only valid if the free-end of the coil is
fully inserted into the socket, and remains fully inserted during use.
The coil has a friction fit, the coil is fully inserted when you feel that
you can ease the free end of the coil into the ferrule no further.
WARNING. The voltage ratings are appropriate for intermittent use of
the probe as a test instrument and not for continuous use in a
permanent installation.
For permanent installation the coil should be situated such that
corona, which will eventually damage the coil insulation, cannot occur.
For information regarding permanent installation of TRCP models on
high voltage equipment, please consult Tektronix.
8 TRCP0300, TRCP0600, TRCP3000 User Manual
Tektronix Rogowski Current Probe
WARNING. You must ensure that the absolute maximum rated di/dt is
not exceeded.
The integrator
WARNING. The probe must only be used with oscilloscopes or
monitoring equipment that have their BNC inputs properly grounded.
Power options
The probe can be powered by an external DC source or by a battery
option.
External supply
The probe can be powered by an external DC supply. The DC voltage
must be 12 V (±10%) using a center positive connector. When the DC
supply is present a RED indicating LED adjacent to the socket is
illuminated.
Standard alkaline batteries
Healthy batteries are indicated by the Green Status Indicator LED. If
the LED is Red, the batteries are delivering less than 2V and must be
replaced.
When the DC voltage is present the batteries are inoperative.
TRCP0300, TRCP0600, TRCP3000 User Manual 9
Tektronix Rogowski Current Probe
10 TRCP0300, TRCP0600, TRCP3000 User Manual
Using your Rogowski Probe
Replacing the battery
Low battery is indicated by a red LED located near the power switch.
To replace the battery, remove the probe from the test circuit, switchoff the transducer, remove the battery cover, and replace the used
batteries with appropriate batteries for your transducer type (standard
or rechargeable).
Make sure the probe is removed from the test circuit before replacing
the batteries.
Switching on
CAUTION. Make sure you've read the following sections before
switching on the transducer: Safety and pre-use checks on page 7 and
Handling instructions on page 6.
WARNING. The probe may be used only with oscilloscopes or
monitoring equipment with properly grounded BNC inputs.
1.Connect the BNC output of the transducer to your grounded
oscilloscope or current monitoring equipment.
2.Visually inspect the Rogowski coil ensuring that the insulation is
intact. If you see an abrasion or tear, use a different probe.
3.If the coil passes the inspection in the previous step, un-clip the
coil and wrap it around the de-energized conductor under test.
4.Insert the free-end of the coil fully inside the ferrule.
5.Re-energize the conductor.
6.Use the Power button to switch the transducer On by pressing and
releasing the button. Turn the transducer Off by pressing and
releasing the button when the transducer is on. The transducer is
on when the Status Indicator LED is Green.
TRCP0300, TRCP0600, TRCP3000 User Manual 11
Using your Rogowski Probe
After using this procedure, the transducer may require a settling period
of up to 2 minutes before providing a correct current measurement. The
settling period is dependent on warm-up time and the low frequency
bandwidth.
Obtaining the best measurement
The Rogowski coil should be positioned so that the conductor under
test is encircled by the coil but is not adjacent to the cable attachment
(see picture below). The diagram shows the direction a positive current
should pass through the coil loop in order to obtain a positive output
voltage. The Rowgowski Current probe has been calibrated with the
conductor near the center (position A), and this is the ideal position for
accuracy.
For the best high frequency performance the center of the current
should lie on the line shown A to B, where B is half way around the
circumference of the coil.
Sensitivity to external currents and voltages
The sensitivity of the TRCP to currents that are outside the closed loop
of the Rogowski coil is very small, provided that the external currents
are less than the current rating of the probe or that such currents are
relatively distant from the coil. In the vicinity of a multi-turn conductor
the effects are far stronger than from a single conductor carrying the
same current, and such positions should be avoided.
If there is a surface with a high voltage very close to the coil and the
voltage is subject to high rates of change (e.g. 1kV/µs) or high
frequency oscillations in the MHz range, then measurement error can
arise due to capacitive coupling to the coil.
12 TRCP0300, TRCP0600, TRCP3000 User Manual
Using your Rogowski Probe
To check for any unwanted response to adjacent fields, it is
recommended to quantify the output of the probe when close to (but
not encircling) the conductor under test. This will reveal the magnitude
of any unwanted responses to currents close to, but outside, the coil.
Output cables
Your probe includes a 500 mm BNC to BNC output. It is possible to
use longer cables; these should be 50 ohm, single screened co-axial
cable.
Extension cables can be used if necessary. Consideration should be
given to routing very long extension cables.
Cleaning and decontamination
To prevent surface contamination the Rogowski coil and cable should
be inspected regularly.
To clean the coil and cable use a mild detergent and water to remove
any contamination, wipe with clean water to remove the detergent and
dry thoroughly with a clean cloth. The probe must be disconnected
from any external circuit and turned off prior to cleaning.
Returning your probe for recalibration or repair
To return your product for repair or recalibration, please contact
Tektronix in advance for shipping instructions (refer to Contacting
Tektronix for contact details).
TRCP0300, TRCP0600, TRCP3000 User Manual 13