Transport Stream Compliance Analyzer (Demo) User Manual
Preface
Manual Conventions
This manual supports the Transport Stream Compliance Analyser demo.
Other documents are available on the Tektronix Web site (www.tektronix.com/
manuals):
HMTS400 Series MPEG Test System Getting Started Manual (071-1505-xx).
HMTS400 Series MPEG Test System Programmer Manual (071-1725-xx).
This manual specifies the remote control and status monitoring interfaces
available to a management application.
The following formatting conventions apply to this manual:
HBold text refers to specific interface elements that you are instructed to
select, click, or clear.
Example: Select Settings from the Configuration menu.
HMono-spaced text can indicate the following:
HText you enter from a keyboard.
Example: Enter the network identity (
HCharacters you press on your keyboard.
Example: Press CTRL+C to copy the selected text.
HPaths to components on your hard drive.
Example: The program files are installed at the following location:
C:\Program Files\Tektronix\.
http://TSMonitor01).
Transport Stream Compliance Analyzer (Demo) User Manual
vii
Preface
viii
Transport Stream Compliance Analyzer (Demo) User Manual
Introduction to the Demo Version
The MTS400 MPEG Transport Stream Compliance Analyzer (TSCA Demo
version) enables you to monitor and interpret the contents of real-time, previously recorded or synthesized transport streams using the latest MPEG, DVB,
ATSC, ISDB-T, and ISDB-S standards.
The demo analyzes only built-in TS and PCAP stream files.
Installing the Program
You can download the TSCADemoInstall.exe file from
www.tektronix.com/software and follow these steps to install the demo version
of TSCA.
1. Double-click TSCADemoInstall.exe to start the InstallShield Wizard that
installs the TSCA demo version on your computer. Click Next to continue.
Figure 1: InstallShield Wizard
Transport Stream Compliance Analyzer (Demo) User Manual
1
Introduction to the Demo Version
2. Click Next to install the desktop shortcut icons.
3. Click Install to begin the installation.
The InstallShield Wizard installs the application displaying the progress.
2
Transport Stream Compliance Analyzer (Demo) User Manual
Figure 2: Setup status
Introduction to the Demo Version
4. Click Next to install the WinPcap 3.1 Setup.
Figure 3: WinPcap 3.1 Setup
Transport Stream Compliance Analyzer (Demo) User Manual
3
Introduction to the Demo Version
5. Click I Agree to continue with WinPcap installation.
6. Click Finish to complete the WinPcap 3.1 Setup.
4
Transport Stream Compliance Analyzer (Demo) User Manual
Introduction to the Demo Version
If WinPcap 3.1 is already installed in your system, a message is displayed as
follows. Click OK to continue.
7. Click Finish to exit the Wizard and the demo version is successfully
installed.
Transport Stream Compliance Analyzer (Demo) User Manual
5
Introduction to the Demo Version
Starting the Program
Start the program by selecting the TS Compliance Analyser option from the
Start > Programs > Tektronix (Demo versions) > Analysers menu or by
double-clicking the TS Compliance Analyser Demo shortcut on the desktop.
Figure 4: TSCA desktop shortcut
Opening a File
When you start the TS Compliance Analyzer Demo version, the TSCA Demo
window with the overlaid Open Transport Stream dialog box is displayed.
The Open Transport Stream dialog box allows you to use either of the built-in
streams, namely the MPEG-2 Transport Steam file or the IP capture file. You do
not have the option to browse and select the stream of your choice.
6
Transport Stream Compliance Analyzer (Demo) User Manual
Introduction to the Demo Version
PCAP File
Figure 5: Initial appearance of the main window
The IP capture file (PCAP) supports two TS sessions, either of which may be
selected for file analysis.
Figure 6: Example PCAP file containing two TS sessions
Transport Stream Compliance Analyzer (Demo) User Manual
7
Introduction to the Demo Version
Stream Interpretation
The example TS file and the real-time TS input contain DVB and ATSC SI. You
can use the Stream Interpretation dialog box to select the standard of your
choice.
Figure 7: Stream Interpretation dialog box
Real−Time Analysis
Real-time analysis is enabled and stream inputs on the IP, ASI, DVB Parallel,
and SMPTE interfaces are simulated. The following steps describe how to select
a TS session for analysis when using the IP interface:
1. Select Real-time Analysis from the Open Transport Stream dialog box to
display the real-time analysis options.
8
Transport Stream Compliance Analyzer (Demo) User Manual
Figure 8: Real−time analysis dialog box
Introduction to the Demo Version
2. Select IP from the Interface drop down list to display the interface settings.
Click the + button to add a Configuration.
Transport Stream Compliance Analyzer (Demo) User Manual
9
Introduction to the Demo Version
3. Click Browse... to select the UDP flow.
A message that informs you to apply the configuration changes is displayed.
Click Apply to display the Browse for UDP flow... dialog box and select one
of the listed TS sessions.
Figure 9: Browse for UDP flow dialog box
10
Transport Stream Compliance Analyzer (Demo) User Manual
Introduction to the Demo Version
TSCA Main Window
Once the built-in stream is selected from the Open Transport Stream dialog box,
the main window is displayed as follows. The title bar of the main window
shows TSCA to be a demo version.
Menu Options
Figure 10: TSCA main demo window
The demo version of the application supports all the menus options available in
the full version except for the Record and File menu options.
Transport Stream Compliance Analyzer (Demo) User Manual
11
Introduction to the Demo Version
File Menu Options
The File menu options available in the demo version are listed in Table 1:
Figure 11: File menu options
Table 1: File menu options for demo version
CommandFunction
Open Transport
Stream...
Opens the Open Transport Stream dialog box from which the built−in
source stream file (file or real−time) and the interpretation standard
can be selected.
Restart AnalysisRestarts the current analysis.
Save Captured
Stream...
Save TS PacketsDisabled for the TS Compliance Analyser Demo version.
Recent FilesLists the most recently analyzed files (built−in files). Select a file
Close AnalysisStops and closes the current anlaysis, leaving the TS Compliance
ExitExits the program.
Disabled for the TS Compliance Analyser Demo version.
name to reanalyze the file.
Analyser Window open.
When you click the Save TS Packets option of the File menu, the file is not
saved. Instead a message informs you that the save option is disabled.
12
Transport Stream Compliance Analyzer (Demo) User Manual
Introduction to the Demo Version
Record Menu Option
The Record feature is not supported by the demo version and thus the Record
menu options are disabled.
Figure 12: Record menu options
Transport Stream Compliance Analyzer (Demo) User Manual
13
Introduction to the Demo Version
14
Transport Stream Compliance Analyzer (Demo) User Manual
Transport Stream Compliance Analyzer − Getting Started
The MTS400 MPEG Transport Stream Compliance Analyzer (TSCA) enables
you to monitor and interpret the contents of real-time, previously recorded or
synthesized transport streams using the latest MPEG, DVB, ATSC, ISDB-T, and
ISDB-S standards.
The TSCA is specifically designed to enable you to quickly locate and identify
problems within a transport stream using a minimum number of mouse clicks.
By quickly identifying the problem areas, the TSCA software helps you save
time during the development and test of equipment, networks, and services. You
can configure the TSCA software to display stream information in user-selected
fonts. This feature enables you to view stream information in your local language
or to use custom fonts.
The TSCA software will run stand-alone on computers with Microsoft Win dows 2000 or Windows XP operating systems.
Figure 13: TSCA Program view − Deferred analysis
Transport Stream Compliance Analyzer (Demo) User Manual
15
TSCA - Getting Started
TSCA Features
HEasy “program centric” user interface quickly isolates information of interest
HCaptureVut technology captures and analyzes system events in real or
deferred time
HIn-depth analysis of stored transport streams including support for MPEG,
ATSC, DVB, ISDB-T and ISDB-S table types
HData summaries and automated filters simplify the analysis of complex
transport streams
HTR 101 290 Priority 1, 2 and 3 tests
(European Telecommunications Standards Institute (ETSI) TR 101 290:
Digital Video Broadcasting (DVB); Measurement Guidelines for DVB
Systems.)
HSyntax analysis and display supported for TMCC and IIP data
HConsistency checks performed between SI, TMCC, and IIP data
HProprietary PSI/SI syntax section rate error testing
HError logging
HSuspension of all tests at program changeover based on PMT version change
HDII and DDB error checking
Deferred and Real−Time Modes
The TSCA can be run in Deferred or Real-Time mode.
The Deferred mode is available on any recommended platform (including the
user’s own PC). In Deferred mode, a static stream can be analyzed and displayed; the content can then be inspected at leisure.
The Real-time mode, with which a live stream can be analyzed, is available
when the TSCA is installed on a Tektronix MTS400 series MPEG Test System
or an MTX100B with an Option 7 interface card. Real-time streams can be
paused for detailed offline analysis and then when the issue has been found,
real-time analysis can be resumed. Real-time monitoring of Video over IP
(internet protocol) is also available as a standalone application on a user’s own
PC.
16
Transport Stream Compliance Analyzer (Demo) User Manual
Technical Background
TSCA - Getting Started
The analyzer has a client-server architecture. Every time the analyzer opens, it
starts two processes: tsca.exe, which is the client, and MpegCore.exe, which is
the MPEG analysis engine. Additionally, when the analyzer opens for the first
time, another single process, CoreManager.exe, is started. CoreManager.exe
forms an intermediary between TSCA clients and MPEG cores.
The tsca.exe, MpegCore.exe, and CoreManager.exe processes use SNMP
(Simple Network Management Protocol) and HTTP (Hypertext Transfer
Protocol) to communicate. The port numbers used for communication are read
from a configuration file, CoreInfo.xml. In a default installation, this file can be
found in c:\program files\tektronix\mpeg\mts400\apps\tsca\xml\. The default
port range is 15700-15800. These values can be changed if they conflict with
other applications.
On the MTS400 platform, another process, mint.exe, runs from system startup.
This process controls shared access to the MTS400 Series hardware by the
analyzer and player.
User Interface
Program View
You must have local administrator rights to be able to use IP analysis. If you
need to change your rights , please consult your system administrator.
The TSCA software uses a single main program summary window with different
context sensitive views contained within tabbed frames. This provides the
maximum amount of useful information while keeping the screen from appearing
cluttered. From the main window, you can access the following views: Program,
Tests, Tables,PID, Packets, and Interface.
The Program view provides a fast overview of the transport stream contents in
terms of program content, bit rate use by each program, and TR 101 290 test
results. Red, amber and green LEDs highlight errors associated with each
program, or element of a program, within the transport stream. Red LEDs
indicate that there are current errors in the stream, amber LEDs indicate that
errors have occurred but are now clear, and green LEDs indicate that no errors
have been detected.
Errors that are detected at lower levels in the program stream hierarchy propagate
up to the highest level. This allows you to monitor all of the programs in the
stream at a high level and then quickly go to lower levels as necessary to locate a
problem.
Transport Stream Compliance Analyzer (Demo) User Manual
17
TSCA - Getting Started
Tests View
SI/PSI (Tables) View
PID View
The Tests view enables you to isolate errors to the specific tests that have been
applied to the transport stream. The error log is automatically filtered by the
selected test, and can also be filtered by Packet Identifier (PID). In addition to
the standard first, second, and third priority tests included in TR 101 290
standard, tests are available for Program Clock Reference (PCR) jitter and
program/PID bit rate. A variability test enables you to test the changes in the bit
rate of a specific PID, and there are many tests that are specific to ISDB-T and
ISDB-S streams.
The SI/PSI (Tables) view displays the service information tables that have
occurred in the analyzed stream which comply with the selected digital video
standard. This includes MPEG-specific program information, DVB service
information, and ATSC, ISDB program, and system protocol information. The
tables are grouped together by function and hyperlinks enable you to quickly
access related information within other tables.
The PID view displays information about all of the PIDs found in the transport
stream. When you select the transport stream, the associated summary view
provides a PID-oriented overview of the transport stream, displaying the relative
data rates of all of the PIDs contained within the stream. The information can be
displayed as either a bar chart or as a pie chart. Pop up menus enable fast limit
selection and links.
Packets View
Interface View
Each PID is associated with a number of tests. When one or more tests fail, each
failed test will be listed under the relevant PID. Select a specific PID to display a
summary of all the associated tests. Select a specific test to display the Event
Log and Parameters for only that test.
The Packets view displays information about all of the packets found in the
transport stream grouped according to content. These groups include PID value,
SFN mega-initialization packets (MIPs) (DVB only), and ISDB-T information
packets (IIPs). When you select a specific PID or MIP, only packets carrying that
particular PID or MIP are displayed. A section view is also available. Deferred
analysis of PCAP files allows you to see the ethernet packets.
The Packets view is not available in real-time mode. However, packets in a
real-time stream can be examined using the CaptureVu feature.
The Interface view provides a central location for information related to the RF
interface card (where installed).
HTest results
HGraphical representation of specific RF card readings
HRF card readings and control settings
18
Transport Stream Compliance Analyzer (Demo) User Manual
CaptureVu Feature
Triggered Recording
TSCA - Getting Started
The graphical views display a selection of graphs of instantaneous measurements
and measurement trends. Also available are an Event Log and a Parameter view.
The Interface tab is displayed only when an interface is installed.
The CaptureVu feature captures a snapshot of system events in real and deferred
time, and allows you to debug the intermittent and complex problems that
traditional analyzers miss. You can initiate CaptureVu either manually or
automatically by associating a CaptureVu Breakpoint with any test event. When
a breakpoint is detected, either manual or automatic, the stream is paused and an
in-depth deferred time analysis can be performed on stored data. This powerful
debug mode enables fast debugging of troublesome intermittent problems.
During real-time analysis, stream information can only be held for a finite time eventually it will be overwritten by more up-to-date information. Triggered
recording allows a portion of a stream to be recorded and subsequently analyzed
in detail.
The following trigger sources and conditions apply:
HAny test event
HExternal input; (TTL) edge specifiable
HDate and time
HRecorded file size can be specified
HPretrigger buffer size can be specified as a percentage of the overall file size
range from 0 – 100%
Transport Stream Compliance Analyzer (Demo) User Manual
19
TSCA - Getting Started
Starting the TSCA Software
The TSCA can analyze transport streams in either real-time or off-line (deferred)
mode.
Opening the TSCA
Click File Analysis to select the trans−
port stream file to analyze and to se−
lect the method for calculating the
rate of the stream.
Click Real−time Analysis to select an
input source (IP or hardware inter−
face) for real−time analysis.
From the Windows Desktop, select Start > Programs > Tektronix MTS400 >
Analyzers > TS Compliance Analyzer.
The TSCA is opened, initially overlaid with the Open Transport Stream...
dialog box.
20
Figure 14: TSCA startup
The Open Transport Stream... dialog box allows you to:HSelect the interpretation standard to be used for analysis.
(The stream interpretation should be checked and, if necessary, set first.)
HIdentify a stream file for deferred analysis. Files can be selected either
directly from the disk drive or using previously set up shortcuts.
HIdentify an input (IP, ASI or SMPTE or DVB Parallel) receiving a stream
and start real-time analysis.
Transport Stream Compliance Analyzer (Demo) User Manual
TSCA - Getting Started
Interpreting a Stream
You can set up the stream interpretation standard in the following way:
1. Click Change... in the Open Transport Stream dialog box to open the Stream
Interpretation dialog box.
NOTE. The TSCA software analyzes streams in strict accordance with the
standards defined by the digital television authorities: DVB (Digital Video
Broadcasting), ATSC (Advanced Television Systems Committee), and ISDB
(Integrated Services Digital Broadcasting).
The TSCA software uses preloaded scripts which establish the expected content
and syntax of the analyzed stream. The analyzer is supplied with a comprehensive range of scripts, which, in different combinations, can interpret the full
range of digital television standards.
2. Select the Base Standards and then (if available) select the Region. The
analysis scripts that will be used are displayed in the Script Files pane.
Figure 15: Stream Interpretation dialog box
3. Select OK.
Transport Stream Compliance Analyzer (Demo) User Manual
21
TSCA - Getting Started
4. The stream interpretation is now set. Choose one of the analysis options to
select a file or real-time analysis.
(For details of the Stream Interpretation management, see Script Files on page
149.)
NOTE. The selected stream interpretation standard is not implemented until
stream analysis is started. If the Open Transport Stream dialog box is closed
without starting analysis, the stream interpretation settings are lost; they revert
to the settings that were in effect when the application was last closed.
Off−Line (Deferred)
Analysis
Perform the following steps to start analyzing a transport stream file:
1. In the Open Transport Stream dialog box, with File Analysis selected,
browse to a transport stream file.
If necessary, select a file type from the Files of type drop-down list.
Currently, two types are available: MPEG Transport Stream files and PCAP
files.
HMPEG transport stream files contain interlinked tables and coded
identifiers, which separate the programs and the elementary streams
within the programs.
HPCAP files contain raw packets of network data. A PCAP may contain
more than one session. You will need to select a session using the
Browse dialog box.
2. When you select a stream, the stream name appears in the field at the bottom
of the window.
22
Transport Stream Compliance Analyzer (Demo) User Manual
3. Calculating the Transport Rate. To analyze a transport stream correctly, you
must do one of the following:
HCalculate the bit rate from PCRs (Program Clock Reference) carried in
the stream
NOTE. When the TSCA software analyzes a transport stream, it compares the
bit rate of the stream to the expected rate. Errors are reported based on the
expected rate of the stream. To analyze a stream correctly, it is necessary to
know the original bit rate of the stream or to calculate the bit rate from
PCRs (Program Clock Reference) carried in the stream. PCRs are generated
as a stable clock reference and are added to the stream during its generation. During analysis, the PCRs are extracted and synchronized with a local
clock reference.
HKnow the original bit rate and enter it manually
HUse timestamps included with the stream
4. After you select the desired stream, select one of the following methods for
calculating the rate of the transport stream:
HPrescan the first nnn PCRs. When you select this method, the TSCA
software calculates the rate of the stream by prescanning the stream for
embedded PCRs. The entry box lists the number of PCRs that the TSCA
Transport Stream Compliance Analyzer (Demo) User Manual
23
TSCA - Getting Started
software recommends are necessary to determine the stream rate. You
can use the entry box to enter a different number.
HPrescan all PCRs in a file. When you select this method, the TSCA
software calculates the rate of the stream by prescanning the stream for
all of the embedded PCRs.
HManually set nnnn Mbps. When you select this method, the TSCA
software uses the absolute value (in Mbps) that you enter as the expected
stream rate.
24
NOTE. Timestamps, which are added when the stream is generated, will be
used, if present, and if the Use Timestamps checkbox is enabled. Using
timestamps will override the PCR options (Prescan PCRs and Scan entire
file). The manual bit rate setting will override both PCRs and timestamps.
5. If the TSCA software cannot calculate the bit rate from the embedded PCRs,
or timestamps, a message box is displayed. You must enter an estimated bit
rate before the TSCA software can analyze the stream.
If the CaptureVu feature is to be used, the whole file can be captured or
analysis can be stopped either by a previously set CaptureVu breakpoint or at
a specific packet.
6. To stop analysis at a specific packet, enable Stop Analysis at Packet Index
and enter a packet number.
With a file selected and the transport rate calculation set up, you can start the
analysis (see page 26).
Transport Stream Compliance Analyzer (Demo) User Manual
TSCA - Getting Started
Real Time Analysis
Perform the following steps to start analyzing a transport stream received
through an IP or a hardware interface.
1. In the Open Transport Stream dialog box, with Real-time Analysis selected,
select an interface from the drop-down list.
NOTE. For stand-alone installations, only the IP interface is available.
Figure 17: TSCA Real−time Analysis dialog box
Real-time analysis will not be available if the appropriate software key
permissions are not found. A warning message is displayed when real-time
analysis is not available.
2. Make any necessary interface settings. During analysis, the interface selected
will be shown in the status bar. (The IP interface settings are described on
page 27.)
Transport Stream Compliance Analyzer (Demo) User Manual
25
TSCA - Getting Started
Start the Analysis
To start the analysis, follow these steps:
1. After you have selected the interpretation standard and the stream source,
click OK to continue.
The TSCA software starts analyzing the selected stream. During deferred
analysis, a progress bar is displayed. The TSCA starts reporting analysis
results (see Figure 18).
26
Figure 18: TSCA window showing analysis progress
2. When sufficient data has been gathered and the display has stabilized, you
can start using the TSCA to view the analysis.
3. To analyze another stream file or to reanalyze the current stream using
different standards, use the File > Open Transport Stream menu selection.
Transport Stream Compliance Analyzer (Demo) User Manual
Setting Up the IP Interface
Two Ethernet interfaces are standard on the MTS400 Series systems (one 10/100
Base T and one 10/100/1000 Base T GigE). A high performance Gigabit
Ethernet interface (1000 Base T GigE) can be acquired as an option; this
provides accurate hardware timestamping, hardware filtering and a variety of
connection options, including copper and optical. These allow for basic detection
and display of all UDP (User Datagram Protocol) traffic on a network. Extraction
of an MPEG over IP stream is possible (if the licensed option is enabled).
HIP multicast addresses are indicated. Note that the TSCA will not automati-
HUDP streams carrying MPEG TS traffic are indicated.
HA selected UDP stream carrying TS traffic can be selected for analysis and
TSCA - Getting Started
cally join a multicast group (see IGMP Group Membership, page 28).
recording, in a similar manner to any other physical interface.
Figure 19: Setting up the IP interface
In the Open Transport Stream… dialog box with IP selected in the Select
Interface drop-down menu, the current settings for the selected configuration are
shown. The individual settings are described next.
Select OK to confirm the IP setting and start analysis.
Transport Stream Compliance Analyzer (Demo) User Manual
27
TSCA - Getting Started
Interface Settings
Configurations
You can save individual IP settings configurations to disk; they can be recalled
when required. Up to 128 configurations can be specified.
The Open Transport Stream - Real-time Analysis dialog box shows the settings
associated with the currently selected configuration.
Different settings can be entered on this screen; the new settings will be used
when OK is selected. The new settings can also be saved as a new configuration
for future use if required.
IGMP Group Membership. Multicast broadcasts can be joined by entering the
multicast address in the IGMP Group Membership field in the Browse dialog
box. The sessions detected will be listed in the dialog box.
Create a New IP Configuration. In the Open Transport Stream - Real-time
Analysis dialog box, select the plus button to create a new configuration.
A default configuration name is allocated and displayed in the IP Configuration
Name field. You can enter a more meaningful name. New values can be entered
as required. This name and the settings associated with it will remain available
for recall.
Alternatively, the streams available can be viewed and a new one selected.
Select New to open the Edit IP Configuration dialog box. Select Browse and the
Browse for UDP flow... dialog box will be displayed (see Figure 20).
28
Transport Stream Compliance Analyzer (Demo) User Manual
TSCA - Getting Started
Figure 20: UDP flow selection dialog box
The names of any network devices installed in the instrument are displayed in
the Browse Device field. Select the network device required from the drop-down
list. The traffic present on the selected interface is displayed in the bottom half of
the dialog box. A summary of the UDP flows detected is displayed in the table
header. The bit rates given are UDP bit rates, not transport stream bit rates.
The Source and Destination addresses can be set up in two ways:
HIn Open Transport Stream > Real-time Analysis > Selected Configuration
area, enter the data into the Stream Settings fields. Multicast sessions can be
identified by entering the appropriate address in the IP Destination Address
field.
HIn the Browse for UDP Flow dialog box, highlight a session in the current
traffic area; the values are entered automatically in the Selected Configuration, Stream Settings fields. Traffic carrying transport streams will have a
check mark in the Has TS column.
The TSCA should detect whether traffic is transmitted over UDP or RTP; this
will be indicated in the Transport column. The Use RTP check box will be
updated to match the protocol indicated for the selected session; this setting can
be overridden by selecting or clearing the Use RTP setting if it is believed that
the protocol indicated is incorrect.
Transport Stream Compliance Analyzer (Demo) User Manual
29
TSCA - Getting Started
When the settings are correct, select OK to confirm and save the configuration.
The Edit IP Configuration dialog box closes and the new configuration is
immediately available in the Open Transport Stream dialog box.
Edit an Existing IP Configuration. An existing IP configuration can be modified as
required. Modify the individual fields in the Selected Configuration, Stream
Settings fields or select a new entry in the Browse for IP Flow dialog box.
Select an Alternative IP Connection. Select a new configuration from the Interface
Settings drop-down list. Note the change in the Current Settings section of the
window.
Delete an IP Configuration. To delete an IP configuration, ensure that the
configuration to be deleted is displayed in the Open Transport Stream dialog box
and select the minus button. The selected IP configuration is deleted and the next
available IP configuration is displayed.
30
Transport Stream Compliance Analyzer (Demo) User Manual
Setting the Stream Content Font
You can configure the TSCA software to display stream content information in a
user-defined font style or language, such as Japanese or Chinese.
NOTE. How the TSCA displays the text strings carried in the stream itself is
dictated by the TSCA font setting and by the interpretation standard used during
the analysis.
How the TSCA displays entries in the Event Log is dictated by the TSCA font
setting and fonts used by a local computer. For example, if the default language
of the TSCA computer is Japanese, the event log will be displayed in Japanese
characters. You must choose a suitable font for the characters to display
correctly.
TSCA - Getting Started
Selecting the Display Font
Perform the following steps to set the font that will be used by the TSCA
software to display stream content. You can configure the TSCA software to
display stream content using any font that is installed on your computer.
1. Select Settings > Preferences… to open the Preferences dialog box.
2. Select Font at the left of the Preferences dialog box to display the window
shown in Figure 21.
Figure 21: Selecting a display font from the Preferences dialog box
Transport Stream Compliance Analyzer (Demo) User Manual
31
TSCA - Getting Started
3. The dialog box lists the default font used by your computer for the Log
Entries and Stream Content displays on the TSCA software. Use the Font
Name and Font Size drop-down boxes to select a desired font style and size.
4. Click OK to accept and save any changes.
Installing Asian Language
Fonts
Loading a Custom
Extension Font
Computers that are using local versions of MS Windows 2000 (such as Japanese
or Chinese), or MS Windows XP, should have the Asian language fonts already
installed on their systems.
If your computer or MPEG test system does not have Asian fonts installed,
perform the following steps to install the Asian fonts:
1. Insert the MS Windows OS CD-ROM, which was supplied with your
computer or MPEG test system, into the CD-ROM drive.
2. Locate the LangPack directory on the CD-ROM. This directory contains a
number of *.inf files for various languages. For example, you can install
japanese.inf and tchinese.inf, for Japanese and Traditional Chinese language
fonts.
3. To install new fonts, right click on each desired font file in the LangPack
directory and select Install.
4. The new fonts you installed should now appear as possible font selections in
the Preferences dialog box. You must select the new font in the Preferences
dialog box before the new font will be used to display stream content.
The ARIB specifications for the ISDB standards make use of private extensions
to the standard Unicode character set [0xE000-0xF8FF]. These extensions allow
you to add additional characters and nonstandard proprietary characters to an
existing font. The extension sets are provided as TrueType Extension files (*.tte).
Any extension sets should be installed using the Private Character Editor which
is supplied as part of the standard MS Windows installation.
32
To install a font extension set on your computer:
1. Open the MS Windows Private Character Editor.
a. Select Run from the Start menu.
b. Enter eudcedit in the Run dialog box and click OK.
2. After the Private Character Editor opens, close the Select Code dialog box.
3. Select Font Links0 from the File menu.
Transport Stream Compliance Analyzer (Demo) User Manual
TSCA - Getting Started
4. In the Font Links dialog box, select Link with Selected Fonts.
NOTE. If you select Link with All Fonts, the following actions will overwrite any
existing font extension sets.
5. In the Select Fonts box, find and highlight the font to which you are adding
the extension. This font should be the one you selected in the TSCA font
preferences dialog box.
6. After you select the font to which you are adding the extension, click on
Save As….
7. In the Modify Private Character Filename dialog box, browse to the
TrueType extension file you are adding (*.tte) and click Save.
8. In the Font Links dialog box, click OK.
9. Close the Private Character Editor window.
10. To verify that the font extension has been successfully installed, open the
Character Map application.
a. Select Run from the Start menu.
b. Enter charmap in the Run dialog box and click OK.
11. Browse through the font list for the name of the font to which you added the
extension. The term “Private Characters” should follow the font name. When
you select the font name in the list, you can view the extended character set.
12. Close the Character Map application.
The new font extension you installed should now appear as a possible font
selection in the Preferences dialog box. You must select the new font in the
Preferences dialog box before the new font will be used to display stream
content.
Transport Stream Compliance Analyzer (Demo) User Manual
33
TSCA - Getting Started
34
Transport Stream Compliance Analyzer (Demo) User Manual
Understanding the Analyzer Window
On start up, you can immediately open and analyze a transport stream
(see Starting the TSCA Software on page 20). This will result in a display similar
to that shown in Figure 22. This is the working display.
Menu bar
Toolbar
Navigation view
Detail views/panes
Status bar
Figure 22: Analyzer window
Transport Stream Compliance Analyzer (Demo) User Manual
35
TSCA - Understanding the Analyzer Window
TSCA Window Components
The TSCA window contains the following components:
Menu Bar
Toolbar
The Menu Bar provides access to a selection of system options using drop-down
menus (see Menu Bar and Options, page 144).
The buttons displayed on the toolbar provide shortcuts to often-used menu
options. Additional buttons displayed during real-time analysis are shown in
Table 2:
Table 2: Toolbar icons
Deferred and Real−time Analysis
Opens the Open Transport Stream dialog box.
Restarts analysis.
User actions are recorded during a session. The forward and back
buttons allow you to step through the actions in order to review or
repeat them.
Opens the Preferences dialog box from which you can set your
preferences.
CaptureVu
36
View CaptureVu status.
Resume file analysis.
Stop and capture analysis.
Real−time analysis
View record setting.
Arm record feature.
Start recording.
Stop recording.
Transport Stream Compliance Analyzer (Demo) User Manual
TSCA - Understanding the Analyzer Window
Progress Bar
Navigation View
Detail View/Panes(s)
The progress bar indicates the progress of processing during deferred analysis of
a stream.
To stop the analysis process, click the Stop button on the toolbar.
The left side pane in the main window is the navigation view and provides the
main analysis information for the transport stream in a number of forms:
Programs, Tests, SI/PSI, PID, Packets and Interface. Views are selected using the
tabs at the top of the pane. The background of each view is color-coded to aid
identification.
The right side pane in the main window is the detail view and provides details
related to the selection made in the current navigation view. The overall view
may be subdivided into one or more panes; you can access individual views in
multiple pane views using tabs.
Status Bar
The status bar, when not hidden, is displayed at the bottom of the main window
and shows a summary of stream statistics.
The fields, from left to right, are as follows:
TS Availability Indicates that either the stream bit rate has fallen below
the minimum value or that synchronization has been lost
(see also Sync below). The minimum stream bit rate is
set using the “Min stream bit rate for processing”
parameter, see Parameter Edit, page 118.
Sync Shows the status of the MPEG sync loss test. Loss of
sync during analysis of a stream may mean that analysis
of the complete stream was not possible.
Bit rate Total bit rate of the transport stream.
The packet value (188 or 204) is set using the Preferences feature (see page 147).
Transport Stream Compliance Analyzer (Demo) User Manual
37
TSCA - Understanding the Analyzer Window
Packet Size Packets in a stream can be either 188 or 204 bytes long.
Stream Interpretation, Base Standard
Stream Interpretation, Region
Physical Interface The current interface for real-time analysis.
Partial Reception Availability
This is detected automatically.
This is the set of tests that is being used to analyze the
stream.
This subset of tests modifies the base standard; it can
add tests or modify existing ones.
(Available only for ISDB-T streams.) This icon indicates
the presence of partial reception data in the stream. The
colors indicate the status as follows:
Gray:Partial reception availability is unknown.
Green:Partial reception data is available.
Gray +
red cross:
Processing Throttle Indicator
This icon consists of two adjacent LED indicators which
show the status of the TSCA processing engine. The left
indicator shows the “strained” status; that is whether
throttling is being applied to cut down on the information being processed. The right indicator shows the
“overwhelmed” status; that is whether the processing
engine is able to cope with the amount of information in
the stream even after throttling. The colors indicate the
status as follows (The general color coding scheme is
described in Error Status LEDs, see page 40):
Green:OK
Amber:Processing was strained/overwhelmed but is
Red:Processing is currently strained/over-
Partial reception data is not available.
currently OK
whelmed
38
White:Processor state is unknown
Gray:Test disabled
Transport Stream Compliance Analyzer (Demo) User Manual
TSCA - Understanding the Analyzer Window
System Log Status Shows the status of the system log which records
system-wide events, for example, scripting errors. The
color of the LED (in the body of the icon) indicates the
most serious event state in the log. When the top surface
of the icon is gray, it indicates that all entries in the log
have been viewed.
Double-click the icon to view the log.
Gray:The log is empty
Amber:There are entries in the log, but none since
the log was last viewed
Red:There are new entries in the log
Analyzer Window − No
Stream Available
If you choose not to select and analyze a stream, the screen will appear as shown
in Figure 23. The analysis structure is present, but with no stream content.
Figure 23: Analyzer window − no file
Transport Stream Compliance Analyzer (Demo) User Manual
39
TSCA - Understanding the Analyzer Window
Error Status LEDs
A major feature used throughout the TSCA interface is error status LEDs that
represent the status of the tests applied to associated items, for example,
programs, tests and PIDs. Colors used are as follows:
(Red) Error; test failed
(Yellow) Transient error (error not currently being detected but has
been seen since last reset)
(Green) No error; test passed
(Dark Blue) Warning (Interface tab only)
(Light Blue) Transient warning (warning not current, but has been
detected since last reset) (Interface tab only)
(Gray): Test disabled
(White) Test not applicable or unknown state
(Red) PID referenced but not found in the stream
(Green) PID referenced but not found in the stream.
Applicable test disabled.
When an LED represents a parent node in a tree (for example, a program node is
the parent of elementary stream nodes), the color represents the worst case of all
of its tests and the worst case of all of the tests represented by its child nodes.
Similarly, when an LED representing a parent node in a tree is disabled or
enabled, all subsidiary (or child) nodes are also disabled or enabled to reflect the
state of the parent node. When an LED representing a parent node is reset, all the
child nodes are also reset.
NOTE. The color scheme described here is used throughout the TSCA interface,
for example, the bit rate bars in the navigation views.
40
Transport Stream Compliance Analyzer (Demo) User Manual
TSCA - Understanding the Analyzer Window
Test Management
Shortcut Menu Options
Shortcut Menus
Tests can be disabled and enabled as required using the shortcut menu associated
with the LED icon. However, it is important to note that the extent of the action
depends on the hierarchical level of the icon chosen. For example, choosing
Disable from a program node shortcut menu will disable all tests in the program
group.
NOTE. In deferred time, if any disabled events are reenabled at the end of
analysis, the LEDs will become green. You will have to restart the analysis if you
wish to see the status of the test when the test is enabled.
Shortcut menus provide an important way of navigating around the TSCA
application. They are available on most displayed objects, for example, program
tree nodes. These menus provide options relevant to the selected object. For
example, after failure, all tests associated with a node can be reset, by selecting
the shortcut menu option Reset All Tests. The shortcut menu options provide
both actions (for example, enable and disable a function) and quick links to
associated objects in other views.
Navigation view tabs
Nodes
Shortcut menu (title bar)
Shortcut menu options
Figure 24: Shortcut menu (example)
To view shortcut menus, right-click an object. The menus are associated with
node types (for example, a PID node). Note that node types can be displayed in
more than one view; for example, PID nodes are displayed in the Program view
and the PID view. All shortcut menus have a title bar that indicates the node
selected.
Transport Stream Compliance Analyzer (Demo) User Manual
41
TSCA - Understanding the Analyzer Window
42
Transport Stream Compliance Analyzer (Demo) User Manual
Using the Program View
The Program view is displayed when the Program tab is selected in the
navigation tabbed pane. The navigation view initially displays the top node of
the program tree, which represents the analyzed transport stream from a program
perspective.
Figure 25: Program view
The detail view shows details of the node highlighted in the navigation view.
With the tree expanded, the navigation view shows the makeup of the stream as
shown in Figure 25; the stream display consists of the nodes shown in Figure 26.
Transport Stream Compliance Analyzer (Demo) User Manual
43
TSCA - Using the Program View
Transport stream node
Program node
Elementary stream node
Figure 26: Program view nodes
Table 3 shows which information is available for each node.
Table 3: Program view tabs
Summary /
Tabs /
Node
Transport
Stream
Programnn−−nn−nn
Elementary
Stream
Summary
nnnnn−−n−
−−−−nnnnn
Now
Playing
IP Graphs
MPE
Sessions
Associated
Tests
Graph −
Bit Rate
Graph −
Timing
Event
Log
Parameters
In DVB analysis mode, the MPE tab is always present. However, the MPE view
will be populated only when MPE data is present in the stream (see page 133).
Pressing the F3 key on the keyboard (or selecting Go to the next red LED incurrent tree from one of the shortcut menus) will highlight the next red LED in
the current view. Only the lowest level nodes will be considered and highlighted
in the search since the parent nodes simply reflect the state of the lowest level
nodes.
44
Transport Stream Compliance Analyzer (Demo) User Manual
Program Navigation − Transport Stream Node
The transport stream node represents the entire transport stream in terms of the
programs that it contains. The adjacent LED icon represents the status of the
entire transport stream; any status error in the stream will be represented by this
LED. When this node is selected, the detail view displays the transport stream
summary.
The associated detail view provides a program-orientated overview of the
transport stream, allowing you to see the relative data rates of all of the programs
contained within the transport. A summary of the overall stream is also
displayed. The information can be displayed as either a bar chart or as a pie
chart.
TSCA - Using the Program View
The data rate of a program is the cumulative data rate of all the PIDs that are
associated with that particular program, including PCR and ECM PIDs.
NOTE. Components can be shared between programs, so the overall data rate of
all the programs may appear to be greater than the total data rate of the
transport stream as shown in the status bar.
Transport Stream Compliance Analyzer (Demo) User Manual
45
TSCA - Using the Program View
Bar Chart
You can display the transport stream bar chart by selecting the bar chart icon.
Each program identified in the stream is allocated one row of the table. Related
information includes the overall program status (icon), the program name,
program number, and the maximum, minimum and average bit rates.
In addition to the basic display, you can associate a maximum and minimum
threshold value with each program in the bar chart (see page 125). For deferred
analysis, where these values are set after analysis, the stream must be reanalyzed
for the new values to be applied. In real-time analysis, the new values will take
effect immediately. The color of the bit rate bar indicates that the program has
stayed within the bit rate limits (green) or exceeded the limits (red).
46
Transport Stream Compliance Analyzer (Demo) User Manual
TSCA - Using the Program View
Pie Chart
You can display the transport stream pie chart by selecting the pie chart icon.
The pie chart view is divided into two panes. The left pane shows the pie chart,
and the legend on the right provides a list of the component programs. Each
program is color-coded. Related information given includes the overall program
status (icon), the program name and number, and the bit rate. The values given in
the pie chart labels are bit rates expressed as percentages of the overall bit rate.
The Null PID is shown in this view.
The color of the bit rate labels (and the bit rate values in the legend) reflects the
status of the bit rate measurement, for example, green indicates that the program
has stayed within the bit rate limits, and red indicates that the bit rate has
exceeded the limits.
Transport Stream Compliance Analyzer (Demo) User Manual
47
TSCA - Using the Program View
Bit Rate Value Display
You can toggle the bit rate displays between the mean bit rate and the current bit
rate by selecting or clearing the Show Mean Values option on the detail view
shortcut menus (see Figure 27). A check mark adjacent to the option indicates
that mean values are to be displayed.
Test Status Panes
Figure 27: Bit rate values display
NOTE. The bit rate value shown in the Status Bar is always the mean bit rate. It
is unaffected by the mean/current selection.
The lower view pane allows you to see a summary of the states of all the
associated tests as well as the event log for the entire transport stream (an
unfiltered event log). The test tabs available will depend on the interpretation
standard chosen. For example, with the DVB interpretation standard, the
available tabs are TR 101 209 and Other. Other interpretation standards will
generate different tabs and different sets of tests (see SI/PSI Nodes, page 171).
When any test in the transport stream has failed (and has not been reset), you can
identify which individual tests have failed by observing the red LEDs.
You can use the shortcut menus to reset and disable tests, to set alarms,
CaptureVu Breakpoints and record flags, and to jump to associated tests.
48
Transport Stream Compliance Analyzer (Demo) User Manual
TSCA - Using the Program View
Event Log
The Event Log (Figure 28) displays the log entries for the whole transport
stream. In deferred mode, the packet generating the error report can be inspected
by highlighting the error and selecting Go to packet from the shortcut menu.
Similarly, in real-time analysis mode, and where an event concerns a PID
carrying a PCR, you can use the shortcut menu to jump to the relevant PCR
graph. Deferred PCAP file analysis enables a shortcut that allows Ethernet
packets to be viewed.
Figure 28: Event log
Two buttons are available in the Error Log toolbar:
/ Pin log / Log is Pinned.
Export Log.
The buttons are described in detail in Event Log, page 121.
Transport Stream Compliance Analyzer (Demo) User Manual
49
TSCA - Using the Program View
IP Graphs
The IP graphs available are as follows:
HPacket Interarrival Time (PIT):
Shows the mean interarrival time of IP packets.
HPIT histogram:
Shows the distribution of packet arrival intervals. The X axis contains
accumulators where interval times are collected. The Y axis corresponds to
number of intervals that collect in each accumulator.
HEthernet flow rate:
Bit rate of the Ethernet flow being analyzed.
HDelay factor:
A measurement which characterizes IP packet cumulative jitter and delay.
(Delay factor is the amount of buffer, in milliseconds, that would be required
to subtract IP packet arrival deviations from the rate determined by the
media payload.)
HMedia loss rate:
Media packets lost per second.
The graphs actually available are dependent upon the IP interface options being
used. Unavailable graph names will be grayed out in the selection menu.
50
NOTE. Graph management and timing graphs are described in more detail in the
Common User Interface Concepts section of this manual (see page 99).
Transport Stream Compliance Analyzer (Demo) User Manual
Program Navigation − Program Nodes
Each of the program nodes in the transport stream represents a single program
referenced within the stream. When a program node is selected in the navigation
view, the detail pane displays the details of the program. The LED adjacent to
the program node represents the status of its child nodes (elementary streams)
and the related tests.
TSCA - Using the Program View
The program detail view provides you with an overview of the selected program,
allowing you to see the relative data rates and stream types of all of the PIDs that
are associated with the program (including PCR and ECM PIDs). If the program
contains an MPEG2 or H.264 video stream, it is decoded to produce a thumbnail
picture of the content; this is updated during real time analysis. Similarly, where
EPG information is associated with the program, a summary (Now and Next
program) is displayed. PID content is indicated using icons, which are described
on page 101. Bar chart and pie chart views are available.
PID related information is available under the Summary tab.
The Now Playing tab displays details of the streams and thumbnails for each of
the programs (where available). (See also Now Playing Views, page 116).
The Associated Test tab shows tests related to the program. Similarly, the Bit
rate Graphs tab displays the program bit rate in graph form.
Transport Stream Compliance Analyzer (Demo) User Manual
51
TSCA - Using the Program View
The Parameters view displays the parameters associated with the currently
selected test in the Associated Tests pane (if applicable). You can change the
parameter values, as well as display the current, minimum and maximum values
for each of the test parameters. For deferred analysis, where these values are set
after analysis, the stream must be reanalyzed for the new value to be applied. In
real-time analysis, the new values will take effect immediately.
The Event Log shows the errors for the highlighted test. When no tests are
selected, the event log will show the events related to all tests associated with the
elementary stream selected in the program tree.
VLC Media Player
The VLC Media Player is a portable multimedia player for various audio and
video formats. The player allows program content displayed in the TSCA
thumbnails to be played in a larger window. The media player is automatically
accessed from the TSCA when you want to view or listen to the video and audio
content of program streams.
NOTE. The VLC Media Player is not installed as part of the MTS400 Series
system. Installation of the player is described in the MTS400 Series GettingStarted manual (Tektronix part number: 071-1505-xx).
You can find full documentation of VLC on the http://www.videolan.com.
To start the media player, click the start button (below the thumbnail) in the
Program Summary view.
52
The VLC Media Player is opened automatically and the selected program
content (video and/or audio) will be played. You can close the player by clicking
the stop button (below the thumbnail). The player will close automatically.
Transport Stream Compliance Analyzer (Demo) User Manual
NOTE. The VLC Media Player can open “behind” the TSCA window. Use the
Windows controls to bring it forward.
NOTE. If the VLC Media Player is installed on the MTX100B instrument, the
audio feature in the media player must be disabled. Refer to the MTS400 SeriesGetting Started manual (Tektronix part number: 071-1505-xx).
Program Navigation − Elementary Stream Node
Each of the elementary stream nodes in the program tree represents a single
elementary stream referenced by the parent program. When this node is selected,
the view pane displays the Elementary Stream summary view.
TSCA - Using the Program View
In addition, when the cursor hovers over the elementary stream node, a tooltip is
displayed giving stream type and description.
Transport Stream Compliance Analyzer (Demo) User Manual
53
TSCA - Using the Program View
An LED associated with each elementary stream node indicates the status of the
tests relating to the elementary stream.
When the selected elementary stream PID contains timing information, PCR
trend analysis views are also available, allowing you to display graphs of PCR
accuracy and PCR interval data. When the transport stream contains
timestamped data, the PCR overall jitter, PCR frequency offset, PCR arrival
time, and PCR drift rate graphs are also available.
NOTE. Graph management and timing graphs are described in more detail in the
Common User Interface Concepts section of this manual (see page 99).
54
Figure 29: PCR Graphs view
Transport Stream Compliance Analyzer (Demo) User Manual
TSCA - Using the Program View
The Associated Tests tab shows tests related to the elementary stream node.
The Parameters view displays all of the parameters for the currently selected test
in the Associated Tests pane (if applicable). As well as displaying the current,
minimum and maximum values for each of the test parameters, you can change
the parameter value. For deferred analysis, when values are changed, the stream
must be reanalyzed for the new values to be applied. In real time, the changes
will take effect immediately.
Similarly, the Event Log shows the errors for the highlighted test. When no tests
are selected, the event log will show the events related to all tests associated with
the elementary stream selected in the program tree.
Transport Stream Compliance Analyzer (Demo) User Manual
55
TSCA - Using the Program View
56
Transport Stream Compliance Analyzer (Demo) User Manual
Using the Tests View
The Tests view displays test information relevant to the transport stream being
analyzed. The view is displayed when the Tests tab is selected in the Navigation
tabbed pane (see Figure 30).
Figure 30: Tests Navigation view
The view displays a test tree which represents all the tests that have been applied
to the analyzed transport stream. Figure 31 shows the nodes available in the
navigation view.
Transport Stream Compliance Analyzer (Demo) User Manual
57
TSCA - Using the Tests View
All Tests
Group
Priority
Test
Associated PIDs
Subset
Group Other"
Figure 31: Test view nodes
58
Transport Stream Compliance Analyzer (Demo) User Manual
TSCA - Using the Tests View
Test Structure
The test tree structure represents all of the tests that can be applied to a transport
stream complying with a particular digital video standard: MPEG, DVB, ATSC,
ISDB-S, and ISDB-T. The standard must be selected before beginning analysis.
The test tree contains nodes for each test that the analyzer can implement.
An example of a tree structure for a selected standard is the TR 101 290 group of
tests for DVB transport streams (shown in Figure 31). The Group node is split
into priority groups: First Priority, Second Priority, and Third Priority. Each of
these priority groups contains a number of individual tests.
Test grouping differs between digital video standards. However, all of the
currently implemented standards contain a subset of the relevant TR 101 290
tests. Note that in standards other than DVB, the tests are not numbered.
Where relevant, PIDs will be associated with a test in the test tree when the test
has either been disabled, or when it has failed.
The LED associated with each node indicates the status of the node and its child
node tests (see Error Status LEDs on page 40). Total entries in the Event Log for
each group of tests and for each test are are indicated in parenthesis; resetting a
group of tests or a test restarts the count in parenthesis.
Transport Stream Compliance Analyzer (Demo) User Manual
59
TSCA - Using the Tests View
Test Filtering
Interpreting the display of tests can be made easier using the three filter buttons
at the top of the Tests navigation view.
Gray
WhiteGreen
Figure 32: Test filtering buttons
Each button, when selected, will hide a category of tests in the display.
HWhen you select the gray button, tests that are currently disabled will not be
displayed in the test tree.
HWhen you select the white button, tests that currently have either unknown
or not applicable status will not be displayed in the test tree.
HWhen you select the green button, tests that currently have OK status will
not be displayed in the test tree.
60
Transport Stream Compliance Analyzer (Demo) User Manual
Tests Navigation − All Tests
The detail view when the All Tests node is selected is shown in Figure 30 on
page 57. In addition to a summary of all tests, the event log shows the most
recent 10000 events that have occurred during analysis of the stream.
Test Failures View
TSCA - Using the Tests View
Figure 33: Test Failures screen
The Test Failures view provides an overview of the tests that have failed during
stream analysis. The view is divided into two panes. The left panes shows those
tests that have failed in the last second. The right pane shows those tests that
have failed, not including those that have failed in the last second, and are
displayed on the left side.
If a test fails, it will be displayed on the left side. When it no longer fails, it will
be removed from the left side and listed on the right side as a test that has
previously failed. If the test fails again, it will be removed from the left side and
redisplayed on the right side until it no longer fails.
Additionally, tests that have been listed (in either pane) in the last second are
displayed at the top of the list and shaded.
Transport Stream Compliance Analyzer (Demo) User Manual
61
TSCA - Using the Tests View
Test Navigation − Test Nodes
When you select a test node, the Associated PIDs pane displays a summary of all
of the PIDs and an event log listing all events that are associated with the test.
Where parameters are applicable to the test, you can modify them under the
Parameters tab.
62
Transport Stream Compliance Analyzer (Demo) User Manual
Test Navigation − PID Nodes
Where tests have failed, the PIDs affected are listed under the test node in the
Tests navigation view. Selecting a PID will display the Event Log, Parameters
for the PID, and the test in the detail view.
TSCA - Using the Tests View
Transport Stream Compliance Analyzer (Demo) User Manual
63
TSCA - Using the Tests View
64
Transport Stream Compliance Analyzer (Demo) User Manual
Using the Tables View
The Tables view (SI/PSI tab) displays service information in tree form, as in the
other navigation views. The tree represents the service information tables that
have occurred in the analyzed stream and which comply with the selected digital
video standard. This includes MPEG program specific information, DVB service
information, ATSC and ISDB program information, and system information
protocol. The tree contains nodes for each table found; tables/nodes are grouped
together by functionality. Figure 34 shows an example of the Table navigation
view with DVB stream interpretation selected.
Figure 34: Tables view
Every node within the service information tree has a corresponding summary
view. Nodes that represent tables will also provide you with access to the table
structure and hex data (see Tables Detail View - Section View on page 67), and
graphical displays of data rates and interval data (see Tables Detail View - SIRepetition Graphs on page 69).
Transport Stream Compliance Analyzer (Demo) User Manual
65
TSCA - Using the Tables View
Table Summary Pane
Some of the nodes that represent less common tables or table groups may not
have specific summary views available in the analyzer. In these cases, a generic
summary view is used.
The table summary view provides a summary of the table related information for
the analyzed transport stream. There is a table summary view for every node of
the service information. The content is dictated by the selected node.
All summary views for nodes that represent an individual table or subtable will
display the table identity and version number of the table from which the
information is extracted. Where relevant, links (underlined text) are provided to
associated objects, for example, in the programs summary (Figure 35), links are
provided to the individual program summary views. Similarly, arrow icons
provide links to the electronic program guide (EPG) tables. Where the section is
carried on a PID, a link is also provided to the PID view.
Figure 35 shows an example of the summary pane with the PMT PIDs node
selected in the navigation view.
66
Figure 35: Tables view − Programs (example)
Transport Stream Compliance Analyzer (Demo) User Manual
TSCA - Using the Tables View
EPG Views
A specific departure from the Table Summary view described previously is the
EPG view that is displayed when a section of the EPG table is highlighted. For a
description of the general presentation of the EPG view, see page 129.
Tables Detail View − Section View
The detailed section view is available when the selected node in the service
information tree represents a table or subtable in the transport stream. During
real-time analysis the view is continually updated.
The subtable identity, version, and section are displayed at the top of the table
pane.
NOTE. In deferred mode, where two or more subtables, versions or sections are
found in the stream, the relevant field is active and an alternative selection can
be made from the drop-down list.
In real-time mode, only the latest version and section can be viewed.
Transport Stream Compliance Analyzer (Demo) User Manual
67
TSCA - Using the Tables View
The table pane shows all fields that make up a section (for example, PMT). The
fields are arranged in a hierarchical order. The presence of subordinate fields is
indicated using { } bracket symbols adjacent to the container field. The container
can be expanded or collapsed by clicking the + or – icon.
Below the table pane, the data source is displayed. The data pane shows the data
bytes (in both hexadecimal number format and ASCII character format) for the
selected table, version, and section. When a node is selected in the table pane,
the bytes that make up the selected node (including the subordinate nodes) are
highlighted in the hexadecimal data pane (refer to Figure 36).
68
Figure 36: Section payload
If the section is corrupt or empty, the hexadecimal view background is red
instead of blue.
Transport Stream Compliance Analyzer (Demo) User Manual
Tables Detail View − SI Repetition Graphs
The section graph view will be available when the selected node in the service
information tree represents a table or subtable in the transport stream. You can
view graphs of the following data:
HSection repetition interval:
This graph displays the interval between two sections of a table on a
particular PID.
HSubtable intersection gap:
This graph displays the interval between sections in a particular subtable.
HSubtable repetition interval:
This graph displays the time between receiving one complete subtable and
receiving the next complete subtable.
NOTE. Not all of the above graph types are relevant to all table types.
TSCA - Using the Tables View
Graph Management is described in more detail in the Common User Interface
Concepts section of this manual (see page 99).
Transport Stream Compliance Analyzer (Demo) User Manual
69
TSCA - Using the Tables View
70
Transport Stream Compliance Analyzer (Demo) User Manual
Using the PID View
The PID view displays information about all of the PIDs found in the transport
stream being analyzed. The navigation view contains the PID tree structure
representing all of the PIDs that have occurred within the transport stream.
In DVB analysis mode, the MPE tab is always present. However, the MPE view
will only be populated when MPE data is present in the stream (see page 133).
Transport Stream Compliance Analyzer (Demo) User Manual
71
TSCA - Using the PID View
PID View − All PIDs
In the navigation view, PIDs are grouped under the following subordinate nodes:
The number of PIDs in each group is shown adjacent to the node. The root node
label shows the total number of PIDs found and the Network Name.
When highlighted, the All PIDs node displays in the detail pane all of the PIDs
found in the analyzed stream.
Each of these subordinate nodes contains PIDs belonging to that group. PIDs are
allocated during analysis, and the number of PIDs is identified for each node.
You can also allocate PID numbers to the User PID node, although User PIDs
may not necessarily have been found in the stream. This feature allows you to
compare the properties of one or more PIDs more easily.
The associated summary view provides a PID-oriented overview of the transport
stream, allowing you to see the relative data rates of all of the PIDs contained
within the transport stream. The information can be displayed as either a bar
chart, or as a pie chart. Columns in the summary views can be sorted in
ascending or descending order by clicking on the column header. PID related
information is available under the PID Information tab.
Each PID is associated with a number of tests. If one or more tests fail, they will
be listed under the relevant PID node in the navigation view. Selecting the PID
node will display a summary, in the detail view, of all associated tests that have
been found. Selecting a specific test in the Associated Tests pane will display the
Event Log and Parameters for only that test.
72
Transport Stream Compliance Analyzer (Demo) User Manual
PID View − Detail View (All PID and PID Group)
The All PID and PID Group detail view provides you with a PID-oriented
overview of the transport stream. This view is displayed when any of the main
nodes is selected in the navigation pane: All PIDs, Referenced PIDs, Unreferenced PIDs, or User PIDs.
The view shows the relative data rates of all of the PIDs contained within the
selected node. You can display the information in either a bar chart or a pie
format. The associated test states are shown in all views.
TSCA - Using the PID View
Figure 37: PID Summary view
The PID Information view (see Figure 38) provides PID related information, for
example, PID type stream type, scrambled status and the presence of PCRs.
Layer information is also shown when analyzing ISDB streams. The icons
provide an indication of the PID content (see Icons, page 101).
Where a PID carries video or audio, thumbnails are displayed under the Now
Playing tab.
In deferred mode (or during CaptureVu analysis), you can view the packets
carrying a PID by highlighting the PID, and selecting the Go To Packets view
Transport Stream Compliance Analyzer (Demo) User Manual
73
TSCA - Using the PID View
from the shortcut menu. The packet will be added to the PIDs list (in the Packets
view).
Figure 38: PID Information view
PID View − Detail View (PIDs)
The individual PID detail view provides you with information about the selected
PID. This view will be visible when you select an individual PID node in the
PID tree. Where a PID has failed a test, one or more test nodes are displayed as
child nodes of the PID.
The combination and detail of the views vary depending on the selected PID type
and node.
During deferred or captured analysis, the packets carrying the PID can be
examined in more detail by highlighting a PID, and selecting the Go to Packets
view from the shortcut menu.
74
Transport Stream Compliance Analyzer (Demo) User Manual
TSCA - Using the PID View
Ordinary PID Selected
PCR PID Selected
When a PID does not contain PCRs, the main view in the PID pane is the
Associated Tests view. This view lists the tests associated with the selected PID.
The status of the individual tests is indicated with LED icons. Also available are
Bit rate Graphs which show the bit rate of the selected PID in graphical form.
Also on the Associated Tests tab is an event log and a parameters pane. The
event log shows all PID related events. If a test is selected in the PID pane, only
the events associated with that test on that PID are shown. Similarly, when a test
is selected, the Parameters pane shows the parameters associated with that test
(when applicable).
When the selected PID contains PCR information, PCR inaccuracy and PCR
interval data graphs will be displayed in addition to a bit rate graph. When the
transport stream contains time stamped data, the PCR overall jitter, the PCR
frequency offset, and the PCR drift rate graphs will also be available. For
real-time IP analysis, the Packet Interval Timing (PIT) graph is also available
(see page 115).
A Presentation Time Stamp (PTS) graph may also be available on PIDs carrying
audio and video information.
(See also Graph Management, page 105 and Timing Graphs, page 111.)
Transport Stream Compliance Analyzer (Demo) User Manual
75
TSCA - Using the PID View
76
Transport Stream Compliance Analyzer (Demo) User Manual
Using the Packets View
NOTE. The Packets view is available only during deferred analysis and
CaptureVu analysis.
Data in the stream is carried in 188- or 204-byte packets. The Packets view
allows you to examine the contents of each packet in the analyzed stream.
The packet navigation view shows a tree structure which represents the packets
within a stream grouped according to content, including PID value, SFN
mega-initialization packets (MIPs) (DVB only), and ISDB-T information packets
(IIPs).
The All Packets, MIP and IIP nodes do not have any child nodes. The All
Packets node represents all of the packets in the transport stream. The MIP node
represents all of the DVB SFN MIPs in the transport stream. The IIP node
represents all of the ISDB-T information packets in the transport stream.
The PIDs node represents all of the PIDs that are of interest. You can add PIDs
in the transport stream to this node, and to PIDs that are contained within the
user PID list.
Transport Stream Compliance Analyzer (Demo) User Manual
77
TSCA - Using the Packets View
Packet Navigation Bar
Selecting any leaf node in the packet tree will apply a packet filter to the
associated packets. For example, selecting a PID node will only allow packets
carrying that particular PID to be displayed in the associated summary view.
Similarly, selecting the MIP node will only allow MIP information to be
displayed in the associated summary view.
In the packet summary view, you can use the top bar to navigate the packets in
the stream.
The packet position bar shows the position of the selected packet in the stream.
Current packet position
The navigation bar functionality depends on the selection in the packet navigation view. If you select All Packets, the controls will operate on all packets. If
you select a single PID, a filter is in operation, and the controls will work only
for those packets carrying that PID.
Selects the first available packet (in the stream or carrying the selected PID)
Selects the previous available packet (in the stream or carrying the selected PID)
Selects the next available packet (in the stream or carrying the selected PID)
Selects the last available packet (in the stream or carrying the selected PID)
Opens the Go To Packet dialog box
You can also use the packet position bar to navigate through the stream. To do
so, use the cursor to point to and click the next packet to be displayed. Doubleclick the packet number in the bar to open the Go To Packet dialog box. Enter
the required packet number and select OK. For both methods, remember that the
destination packet is dependent on the node selected in the navigation view.
78
Transport Stream Compliance Analyzer (Demo) User Manual
Packet Navigation − Filter Nodes
The packet filters allow you to view packets carrying PID.
To inspect the sequence of packets carrying a particular PID, you must add the
PID to the PID Filters node in the Packet Navigation view. PIDs carrying
sections can also be viewed. You can add as many PIDs as are required. There
are a number of ways to add a PID to the node; as described in the following
paragraphs.
Adding a PID in the Packet
View
1. In the Packet view, expand the PIDs node by doing one of the following:
HDouble-click the child node (double-click to add a PID)
HSelect Add PID... from the PIDs node shortcut menu
2. In the Add PID... dialog box, expand the nodes, locate and select the
required PID.
TSCA - Using the Packets View
Removing PIDs
3. Click OK.
The selected PID is added to the PIDs node, and the first occurrence of a
packet carrying the selected PID is displayed in the summary view.
Added PIDs will be stored when the application is closed. They will be
reinstated when the application is reopened. However, if a different stream is
analyzed, the stored PIDs can not be used.
All PIDs previously added to the PID node can be removed by selecting Remove
All from the PIDs node shortcut menu.
Transport Stream Compliance Analyzer (Demo) User Manual
79
TSCA - Using the Packets View
Packet Detail View
The detail view shows the fields and data that are contained in a packet. The
fields are arranged in a hierarchical order. The presence of subordinate fields is
indicated using { } bracket symbols adjacent to the container field. To expand or
collapse the container, click the + or - icon.
Below the table pane, the data source is displayed. The data pane shows the data
bytes (in both hexadecimal number format and ASCII character format) for the
selected packet. When you select a node in the table pane, the bytes that make up
the selected node (including the subordinate nodes) are highlighted in the
hexadecimal data pane.
80
The packet transport header bytes are displayed in red. Similarly, bytes greater
than 188 are displayed in gray. The adaptation field, where present, is displayed
in blue.
If the section is corrupted or empty, the background of the hex view is red
instead of blue.
Transport Stream Compliance Analyzer (Demo) User Manual
TSCA - Using the Packets View
ISDB Streams
Figure 39 and Figure 40 show an example of the packet detail view with an
analyzed ISDB-T stream. Note the IIP tab and the additional TMCC and IIP
nodes.
Figure 39: Packet view − ISDB detail −IIP
Transport Stream Compliance Analyzer (Demo) User Manual
81
TSCA - Using the Packets View
Figure 40: Packet view − ISDB detail − transport packet
82
Transport Stream Compliance Analyzer (Demo) User Manual
Using the Interface View
If real time analysis using an interface card is invoked, the Interface tab is
displayed in the Navigation view of the TSCA window.
NOTE. The Interface view is available only if an interface card is selected for
real-time analysis.
Figure 41: Interface view − example
Two interface card test nodes are displayed in the navigation view: IF Error
Collection and IF Warning Collection. Each collection contains a range of tests
(see Figure 41).
The detail of the interface view is different for each interface type.
Transport Stream Compliance Analyzer (Demo) User Manual
83
TSCA - Using the Interface View
Detail View Tabs
The detail view contains four tabs: Instantaneous graphs, Trend graphs, Event
Log, and Parameters. This section describes the tabs in general terms. The
following sections describe in detail the tabs associated with each interface type.
Instantaneous Graphs
Trend Graphs
Event Log
Parameters
Readings and Controls
The instantaneous graphs display instantaneous measurements in graphical form.
The trend graphs display trends in measurements in graphical form.
The event log displays a history of the events related to the item selected in the
navigation view.
The parameters tab display the parameters associated with the test highlighted in
the navigation view. Parameter editing is described in the Common UserInterface Concepts section, page 118.
The right panel contains Readings and Controls. The Readings area displays
values received from the card itself.
The RF Lock LED conveys different types of information. If full lock onto the
channel is acquired and it is receiving the signal correctly, the LED will light up
green. When a signal is present but no lock is achieved, the indicator is red.
When no signal is found, the indicator is gray.
84
In the Readings area, the LED colors are as follows:
GreenOn
RedOut of range or failed test
GrayOff or not applicable
Transport Stream Compliance Analyzer (Demo) User Manual
The Readings background colors are as follows:
WhiteIn range
RedOut of range; for example, too much or too little power or out of order packets
YellowNot recently updated
GrayNot applicable; for example, out of order packets not being available when the
The Controls area contains user controls. Values entered here will be applied to
the card and will affect its performance. These controls correspond exactly with
those on the Open Transport Stream dialog box.
RF Interface Card Overview
TSCA - Using the Interface View
per second being greater than zero
UDP protocol is selected
This section describes the configuration and interpretation of results of the RF
interface cards that can be installed in MTS400 Series MPEG Test Systems.
The interface settings described are as follows:
HQAM (Quadrature Amplitude Modulation) (Annex B II) (see page 88)
HCOFDM (Coded Orthogonal Frequency Division Multiplexing)
(see page 91)
HQPSK/8PSK (Phase Shift Keying) (see page 96)
H8VSB (Vestigial Side Band) (see page 94)
(Refer to the MTS400 Series Technical Reference, 071-1724-xx for the interface
technical specifications.)
Only one RF interface card can be installed in the MTS400. All RF interface
measurements and graphs include a constellation diagram with MER rings,
EVM, SNR, and BER measurements. The installed interface can be used
simultaneously with an IP input. All interfaces have dual-level warning and
failure alarms plus RF drift tests for long term trend monitoring.
The input card screen and configuration of the card will depend on which
interface card is installed.
Transport Stream Compliance Analyzer (Demo) User Manual
85
TSCA - Using the Interface View
RF Interface Card Setup
The RF interface card configuration is set up when analysis is started using the
Open Transport Stream dialog box. The settings can subsequently be changed
during analysis from the Interface View.
When real-time analysis is selected and an RF interface is selected from the
Interfaces drop-down list. The Firmware Version and the current Interface
Settings of the card are displayed.
86
Figure 42: Open Transport Stream dialog box
If the firmware version is incorrect, the Update Firmware... button is enabled,
and you will not be able to proceed with the setup and analysis until the
firmware is updated (see page 87).
1. In the Open Transport Stream dialog box, select Real-time Analysis.
2. Select the RF interface from the Interfaces drop-down list.
3. Enter the interface settings required (see the following pages for descriptions
of the settings for each RF interface card).
4. Select OK to start the analysis.
The name of the interface selected is displayed in the TSCA status bar.
Transport Stream Compliance Analyzer (Demo) User Manual
TSCA - Using the Interface View
RF Interface Card
Firmware Update
If the interface card firmware needs updating, a message will be displayed when
real-time analysis is selected. The Update Firmware... button will also be
enabled.
Update the interface card firmware as follows:
1. Select Update Firmware....
2. In the Update Firmware dialog box, select Start.
Figure 43: RF Firmware update
3. The dialog box will display progress of the update. The box will close when
the update is complete.
CAUTION. Removing power from the MTS400 while the firmware upgrade is in
progress can cause fatal corruption of the firmware files.
Transport Stream Compliance Analyzer (Demo) User Manual
87
TSCA - Using the Interface View
RF Interface View Screens
This section describes the settings and displays associated with the RF interface
cards.
RF Interface −
QAM (Annex B)
The settings and displays of the QAM (Annex B) RF interface card are as
follows: QAM(B) Interface Settings, QAM(B) Instantaneous Graphs, and
QAM(B) Trend Graphs.
88
Figure 44: Interface view − RF − QAM (Annex B)
QAM(B) Interface Settings. The following QAM (Annex B) settings are available
on the Open Transport Stream - Real Time Analysis dialog box, and also in the
Interface tab - Controls area.
Tuner Frequency :This value, set in kHz units, is the center frequency of
the channel to be received. Any frequency can be entered. There is no need to follow a channel plan. The
frequency resolution is determined by the tuner step size
of 62.5 kHz.
Transport Stream Compliance Analyzer (Demo) User Manual
TSCA - Using the Interface View
Modulation Format: This drop down list allows 64 QAM or 256 QAM opera-
tion to be selected. Note that SCTE07 and ITU-Tj83
annex B specify exact symbol rates for the two modulation formats, so there is no dialog box to enter the symbol rate separately.
Interleaving Mode: Interleaving assists in the correction of burst noise in-
duced errors. All interleaving modes mentioned in
SCTE 07 are supported, including all Level II modes up
to I = 128 J = 6.
Carrier Lock:This allows the carrier lock range to be increased from
the normal ±150 kHz to approximately ±500 kHz. This
is useful if you are attempting to lock on to badly mistuned or wandering carriers.
Tuner Phase Noise: This control modifies the carrier loop bandwidth to al-
low the acquisition of noisy carriers. Normal mode is
recommended for general use.
Equalization Range: The channel equalizer can be configured to give a long
equalization range, at the expense of equalizer resolution, or, in normal mode, optimized for typical cable
environments. Normal mode is recommended for general use.
QAM(B) Instantaneous Graphs. The following instantaneous graphs are available
for the QAM (Annex B) RF interface card.
RF Constellation:I and Q data are collected from the QAM front end and
displayed as a conventional constellation of 256 points
by 256 points. The samples collected are not real time,
so there is some dwell (delay) before a complete constellation is displayed. The persistence of the display can be
varied using the Persist slider control at the bottom of
the graph.
MER (Modulation Error Ratio) and EVM (Error Vector
Magnitude) calculations are performed on the constellation data according to the formulas in ETR290-101.
SNR After Equalizer: This graph represents the signal-to-noise ratio over time,
typically over 256 seconds. It is useful for examining
signal-to-noise trends, for example, if an alarm condition
has been breached, the graph can be used to examine
trends in the signal-to-noise performance before the
alarm occurred.
Transport Stream Compliance Analyzer (Demo) User Manual
89
TSCA - Using the Interface View
MER:This graph gives a time record of MER over the last
255 seconds.
Equalizer:This graph shows the status of the equalizer taps, and
gives an estimate of the channel state. If there are echoes
or other time related distortions on the signal, the Equalizer display will display the amplitude of the corresponding tap correction values versus time.
NOTE. The Equalizer graph amplitude values are approximate.
QAM(B) Trend Graphs. The trend graphs extend the principle of the instantaneous
graphs by graphing measurements over longer periods. Each graph is associated
with a measurement that is also displayed in the Readings panel on the right side
of the display. The measurements have status LEDs associated with them.
Individual trend graphs can be opened by making a selection from the drop-down
menu of the Trend graph or from the context menu of the measurement in the
Readings panel.
The following trend graphs are available for the QAM B interface card: RF
Lock, MER, EVM, Pre-RS BER, Input Level, SNR after Equalizer, and Carrier
Offset.