2. SELF CAL button and LED indicator. Press to start
self-calibration routine.
LED Status
Green (Solid) Self-calibration completed successfully
Optional accessories
50XMMCXsensortip
TIVPMX50X
TIVP IsoVu™ Me asurement System
Quick Ref
*P0713
071-3733-03
Operating information
The Tektronix TIVP IsoVu™ Measurement System offers a
completely galvanically isolated (optical isolation) system.
Isolation is used to greatly increase the accuracy of differential
measurements with rapidly slewing reference voltages and to
isolate hazardous voltages from the oscilloscope.
The TIVP is designed for use with Tektronix 4/5/6 Series MSO
Oscilloscopes running software version 1.28 or higher.
to the TIVP User Manual (071-3692-xx) for complete
Refer
operating information and product specifications. The manual
ilable for download at www.tek.com/downloads.
is ava
The following illustration shows how to connect the probe to
your instrument and the controls on the compensation box.
erence Guide
73303*
Amber
(Intermittent flashing)
Red (Solid) Self-calibration failed
Amber (Solid) Self-calibration recommended
ration currently in progress
Self-calib
Internal probe temperatures or other settings
have invalidated the previous Self Cal.
3. INPUT Termination button and LED indicator. With no tip
attached, press to toggle the sensor head between 50 Ω and 1
MΩ termination.
LED Status
50 Sensor head SMA connector is terminated in 50 ±5 V
input range
1M Sensor head SMA connector is terminated in 1 M
±5 V input range
Neither
Sensor he
termination disabled)
ad has tip attached (switching input
4. MENU button. Press to open the Probe Vertical menu on
the oscil
loscope.
5. ALT Mode indicator.
LED Status
Magenta
(Flashing)
The probe compensation box is near its maximum
operating temperature. Additional cooling, such as
an external fan, may be required.
Compensating the probe
The TIVP contains a self-calibration routine that corrects
Gain Accuracy, DC offset, and frequency response. These
parameters change as the probe warms up to its operating
temperature and remain constant once the temperature reaches
steady-state. An indicator LED on the probe compensation
box and the SELFCAL:STATE? PI command both indicate
whether a self-calibration is RECOMMENDED, RUNNING,
or PASSED.
Before making critical measurements, run SELF CAL to
ensure the TIVP is compensated. The sensor tip does not need
to be removed from the test point to successfully complete
a self-calibration.
1. Connect TIVP to an oscilloscope channel.
2. Press the SELF CAL button on the probe compensation
box. Alternatively, send the CH<x>:PROBE:SELFCAL
EXECUTE PI command to start the self-calibration
programmatically.
The SELF CAL LED will blink while the
self-calibration is in process.
The SELF CAL LED is amber when there is a change
and a self-calibration is recommended. The LED is
green if the last self-calibration is still valid.
Cable handlin g
Tektronix TIVP probes operate using several optical fibers
within a single cable between the compensation box and probe
head. Therefore, standard optical fiber handling practices
are required. Avoid tight radius bends, crushing, crimping,
twisting, or otherwise stressing the fiber.
Standard accessories
MMCX probe tip
The 10X MMCX tip is included in every TIVP probe. The MMCX tip is
recommended for the best bandwidth and CMRR performance. 0.100"
e Pin and 0.200" Wide Square Pin tips are available as optional
Squar
accessories. Reorder part number: TIVPMX10X
100X sensor tip cable with 0.100" spaced square pin connectors
TIVPSQ100X
500X sensor tip cable with 0.200" spaced wide square pin connector
TIVPWS500X
1X MMCX sensor tip
TIVPMX1X
Square Pin to MMCX Adapter,
0.062" Spacing
131-9677-xx
Probe Tip Tripod Support
352-1170-xx
Lead, MMCX to IC Grabber
196-3546-xx
Lead, Square Pin to IC Grabber
196-3547-xx
Lead, Wide Square Pin to Banana
Jack
020-3189-xx
Square Pin Y-lead
196-3434-xx
MicroCKT grabbers
206-0569–xx
Spare Pins for 0.062" Spaced Test
s
Point
020-3169-xx
Solder Aid for 0.062" Spaced
Square Pins
003-1946-xx
Clearance r equirements
The unique common mode voltage range
of the measurement system allows it to be
n the presence of high frequency/high
used i
voltage common mode signals. It is
important to observe all precautions while
using this product.
WARNING. Electrical shock can occur while using this
urement system. The system is intended to isolate the
meas
operator from hazardous input voltages (common voltages);
lastic case of the sensor head and the s hield on the
the p
sensor tip cable do not supply safe isolation. Maintain the
clearance from the sensor head and sensor tip cable
safe
while the measurement system is connected to the energized
uit as recommended in this document. Do not access the
circ
RF Burn Hazard Zone while taking measurements on a live
uit. RF burn hazard zone is 1 meter (40 in) on all sides
circ
Controls and indicators
1. STATUS indicator.
LED Status Action
en (Solid)
Gre
Green
(Blinking)
Red (Solid)
Red
(Blinking)
Red
(Blinking • • – )
Normal operation
Bulk power failure
Probe application
failure
Probe application
failure and bulk
power failure
Self-Test failure
Power to probe
No
head
-
Try unplugging and plugging
back in. Inspect probe/scope
interface. Service of probe
may be required.
Try unplugging and plugging
back in. Service of probe
may be required.
Try unplugging and plugging
back in. Inspect probe/scope
interface. Service of probe
and/or scope may be
required.
Try unplugging and plugging
back in. Service of probe
may be required.
Probe bipod
Used to hold probe. TIVP can
rotate in holder to accommodate
square pin headers. Reorder part
number:
1179-xx
352-
Probe tip adapter
AdaptanMMCXIsoVutipto
dard 0.100" spaced, 0.025"
stan
square pins. Reorder part number:
131-9717-xx
SMA wrench/driver tool
5/16" wrench for use on SMA
connector. Reorder part number:
003-1947-xx
IsoVu c arrying case
Soft case (with foam insert)
protects the TIVP and enforces the
optical fiber minimum bend radius.
Reorder part number:
016-2147-xx (2 m cable models)
016-2149-xx (10 m cable models)
Specifications
e 1: Physical specifications
Tab l
Characteristic Description
Net weight
(Weight does not
include accessories
and packaging.)
sor tip cable length
Sen
Fiber cable length
Tip cable assembly: 25g (0.055 lb)
Probehead:0.363kg(0.8lbs)
Compensation box: 0.816kg (1.8lbs)
03 cm (7.886 inches)
20.
2 meter (6.56 ft)
10 meter (32.81 ft)
Table 2: Electrical specifications
Characteristic Description
Maximum
Non-Destructive
Input Voltage
ypical)
(T
DC Gain Accuracy (Typical) <3%
SMA Input
(50 mode)
SMA Input
(1 M mode)
TIVPMX10X 250 Vpk
IVPMX50X
T
TIVPSQ100X
TIVPWS500X
TIVPMX1X
5 Vrms, (DC + peak AC)
100 Vpk
00 Vpk
3
600 Vpk
3300 Vpk
5VRMS(50), 100 Vpk (1 M)
x

Characteristic Description
Linear Input
Voltage Range
(Typical)
Common Mode Voltage Range
(Typical)
System noise
(Typical)
SMA Input
(50 mode)
SMA Input
(1 M mode)
TIVPMX10X ±50 V
TIVPMX50X ±250 V
TIVPSQ100X
TIVPWS500X
TIVPMX1X ±5 V
SMA Input
(50 mode)
SMA Input
(1 M mode)
TIVPMX10X ±200 mV range: 4.3 mV rms
TIVPMX50X ±1 V range: 21.5 mV rms
TIVPSQ100X
TIVPWS500X
±5 V
±5 V
±500 V
±2.5 kV
60 kVpeak
±20 mV range: 0.43 mV rms
±320 mV range: 1.46 mV rms
±5 V range: 48 mV rms
±20 mV range: 0.43 mV rms
±320 mV range: 1.46 mV rms
±5 V range: 48 mV rms
±3.2 V range: 14.6 mV rms
±50 V range: 480 mV rms
±16 V range: 73 mV rms
±250 V range: 2.4 V rms
±2 V range: 43 mV rms
±32 V range: 146 mV rms
±500 V range: 4.8 V rms
±10 V range: 215 mV rms
±160 V range: 730 mV rms
±2500 V range: 24 V rms
This product is not rated for CAT II, III or IV
Table 3: Environmental specifications
Feature Description
Operating
temperature
Operating
humidity
Operating
altitude
Comp box: 0° C to 30° C
Probe head: 0° C to 50° C
Tip Cable/Adapters: -40° C to 85° C
Comp box: 5% to 85% RH (Relative Humidity) at
up to +30° C, non-condensing
Probe h ead: 5% to 85% RH (Relative Humidity)
o +40° C, 5% to 45% RH above +40° C up
at up t
to +50° C, non-condensing.
Tip Cable/Adapters: 5% to 85% RH (Relative
Humidity) at up to +85° C, non-condensing
3,000 meters
Compliance information
This section lists the Safety and Environmental standards with
which the instrument complies. This product is intended for use
by professionals and trained personnel only; it is not designed
for use in households or by children.
Safety compliance
This section lists the safety standards with which the product
complies and other safety compliance information.
EU low v oltage directive
Compliance was demonstrated to the following specification as
listed in the Official Journal of the European Union:
Low Voltage Directive 2014/35/EU.
EN 61010-1. Safety Requirements for Electrical
Equipment for Measurement, Control, and Laboratory Use
– Part 1: General Requirements.
U.S. nationally recognized testing laboratory listing
UL 61010-1. Safety Requirements for Electrical
Equipment for Measurement, Control, and Laboratory Use
– Part 1: General Requirements.
Canadian certification
CAN/CSA-C22.2 No. 61010-1. Safety Requirements
for Electrical Equipment for Measurement, Control, and
Laboratory Use – Part 1: General Requirements.
Additional compliances
IEC 61010-1. Safety Requirements for Electrical
Equipment for Measurement, Control, and Laboratory Use
– Part 1: General Requirements.
EN 60825-1/IEC 60825-1. Safety of Laser Products-Part
1: Equipment Classification and Requirements - Edition
3 (2014)
US 21CFR PT1010 Performance S tandard for Electronic
Parts 2015.
US 21CFR PT1040 Performance Standards for Light
Emitting Products 2015.
Equipment type
Test and measuring equipment. CLASS 1 LASER PRODUCT.
Pollution degree d escriptions
A measure of the contaminants that could occur in the
environment around and within a product. Typically the
internal environment inside a product is considered to be the
same as the external. Products should be used only in the
environment for which they are rated.
Pollution degree 1. No pollution or only dry,
nonconductive pollution occurs. Products in this category
are generally encapsulated, hermetically sealed, or located
in clean rooms.
Pollution degree 2. Normally only dry, nonconductive
pollution occurs. Occasionally a temporary conductivity
that is caused by condensation must be expected. This
location is a typical office/home environment. Temporary
condensation occurs only when the product is out of
service.
Pollution degree 3. Conductive pollution, or dry,
nonconductive pollution that becomes conductive due
to condensation. These are sheltered locations where
neither temperature nor humidity is controlled. The area is
protected from direct sunshine, rain, or direct wind.
Pollution degree 4. Pollution that generates persistent
conductivity through conductive dust, rain, or snow.
Typical outdoor locations.
Pollution degree rating
Pollution degree 2 (as defined in IEC 61010-1). Rated for
indoor, dry location use only.
IP rating
IP20 (as defined in IEC 60529).
Measurement and overvoltage category descriptions
Measurement terminals on this product may be rated for
measuring mains voltages from one or more of the following
categories (see specific ratings marked on the product and in
the manual).
Category I. Circuits not directly connected to a mains
supply.
Category II. Circuits directly connected to the building
wiring at utilization points (socket outlets and similar
points).
Category III. In the building wiring and distribution
system.
Category IV. At the source of the electrical supply to the
building.
NOTE. Only measurement circuits have a measurement
category rating. Other circuits within the product do not have
either rating.
Environmental consideration s
This section provides information about the environmental
impact of the product.
Product end-of-life handling
Observe the following guidelines when recycling an instrument
or component:
Equipment recycling. Production of this equipment required
the extraction and use of natural resources. The e quipment may
contain substances that could be harmful to the environment or
human health if improperly handled at the product’s end of life.
To avoid release of such substances into the environment and
to reduce the use of natural resources, we encourage you to
recycle this product in an appropriate system that will ensure
that most of the materials are reused or recycled appropriately.
This symbol indicates that this product complies
with the applicable European Union requirements
according to Directives 2012/19/EU and
2006/66/EC on waste electrical and electronic
equipment (WEEE) and batteries. For information
about recycling options, check the Tektronix Web
site (www.tek.com/productrecycling).
Important safety information
This manual contains information and warnings that must be
followed by the user for safe operation and to keep the product
in a safe condition.
Any servicing is not allowed, please contact Tektronix service.
Laser source is contained in the products.
General safety summary
Use the product only as specified. Review the following safety
precautions to avoid injury and prevent damage to this product
or any products connected to it. Carefully read all instructions.
Retain these instructions for future reference.
The product is designed to be used by trained personnel only.
Before use, always check the product with a known source to
be sure it is operating correctly.
To Avoid Fire or Personal Injury
Connect and disconnect properly. Do not connect or disconnect
sensor tip cables, test leads, or accessories while they are
connected to a voltage source. Use only test leads and
accessories supplied with the product, or indicated by Tektronix
to be suitable for the product.
Observe all terminal ratings. To avoid fire or shock hazard,
observe all ratings and markings on the product. Consult the
product manual for further ratings information before making
connections to the product. Do not exceed the Measurement
Category (CAT) rating and voltage or current rating of the
lowest rated individual component of a product or accessory.
Do not apply a potential that exceeds the maximum rating.
Do not operate without covers.
Avoid Exposed Circuitry. Do not touch exposed connections
and components when power is present.
Do not operate with suspected failures.. If you suspect that
there is damage to this product, have it inspected by qualified
service personnel.
Disable the product if it is damaged.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Clean with Dry Cloth Only.
Sensor tip cables
Maintain safe clearance from the sensor head and sensor tip
cable while connected to the energized circuit as recommended
in this manual.
Remove the sensor tip cable and adapters from the test circuit
when not in use.
Leave the sensor tip cable connected to the sensor head when
not in use.
Use only correct Measurement Category (CAT), voltage,
temperature, altitude, and amperage rate
d sensor tip cables and
accessories for any measurement.
Beware of high voltages. Understand the voltage ratings for the
product you are using and do not exceed tho
se ratings. It is
important to know and understand the maximum m easurement
voltage rating of the product. The voltag
e rating depends on the
measurement category, the instrument, and your application.
Refer to the Specifications section of the
manual for more
information.
WARNING. To prevent electrical shock, do
not exceed the
maximum measurement or maximum voltage category.
Connect and disconnect properly.
CAUTION. To avoid damage to the equipment, de-energize
the test circuit before connecting or di
sconnecting the
sensor tip cable.
Terms in This Manual.
These terms may appear in this manual:
WARNING. Warning statements identify conditions or
practices that could result in injury or
loss of life.
CAUTION. Caution statements identify conditions or
practices that could result in damage to
this product or
other property.
Isolated, electrically floating. The t
erms isolated, electrically
floating,andgalvanically isolated are used in this document to
indicate a measurement where there is n
o direct conduction
path to earth ground.
Symbols and terms on the product
These terms may appear on the product:
DANGER indicates an injury hazard immediately
accessible as you read the marking.
WARNING indicates an injury hazard not immediately
accessible as you read the marking.
CAUTION indicates a hazard to property including the
product.
When this symbol is marked on the product, be
sure to consult the manual to find out the nature of
the potential hazards and any actions which have
to be taken to avoid them. (This symbol m
be used to refer the user to ratings in the manual.)
ay also
The following symbol(s) may appear on the product:
Contacting Tektronix
Web site: www.tek.com
Phone: 1-800-833-9200
Address: Tektronix, Inc.
Email:
Department or name (if known)
14200 SW Karl Braun Drive
P.O. Box 5 0 0
Beaverton, OR 97077 USA
techsupport@tek.com
Warranty information
For warranty information, go to www.tek.com/warranty.
Copyright © Tektronix, Inc. All rights reserved. www.tek.com