Tektronix TIVPWS500X Quick Reference Guide

xx
2. SELF CAL button and LED indicator. Press to start self-calibration routine.
LED Status
Green (Solid) Self-calibration completed successfully
Optional accessories
50XMMCXsensortip
TIVPMX50X
TIVP IsoVu™ Me asurement System
Quick Ref
*P0713
071-3733-03
Operating information
The Tektronix TIVP IsoVu™ Measurement System offers a completely galvanically isolated (optical isolation) system. Isolation is used to greatly increase the accuracy of differential measurements with rapidly slewing reference voltages and to isolate hazardous voltages from the oscilloscope.
The TIVP is designed for use with Tektronix 4/5/6 Series MSO Oscilloscopes running software version 1.28 or higher.
to the TIVP User Manual (071-3692-xx) for complete
Refer operating information and product specications. The manual
ilable for download at www.tek.com/downloads.
is ava
The following illustration shows how to connect the probe to your instrument and the controls on the compensation box.
erence Guide
73303*
Amber (Intermittent ashing)
Red (Solid) Self-calibration failed
Amber (Solid) Self-calibration recommended
ration currently in progress
Self-calib
Internal probe temperatures or other settings have invalidated the previous Self Cal.
3. INPUT Termination button and LED indicator. With no tip attached, press to toggle the sensor head between 50 and 1 Mtermination.
LED Status
50 Sensor head SMA connector is terminated in 50 ±5 V
input range
1M Sensor head SMA connector is terminated in 1 M
±5 V input range
Neither
Sensor he termination disabled)
ad has tip attached (switching input
4. MENU button. Press to open the Probe Vertical menu on the oscil
loscope.
5. ALT Mode indicator.
LED Status
Magenta (Flashing)
The probe compensation box is near its maximum operating temperature. Additional cooling, such as an external fan, may be required.
Compensating the probe
The TIVP contains a self-calibration routine that corrects Gain Accuracy, DC offset, and frequency response. These parameters change as the probe warms up to its operating temperature and remain constant once the temperature reaches steady-state. An indicator LED on the probe compensation box and the SELFCAL:STATE? PI command both indicate whether a self-calibration is RECOMMENDED, RUNNING, or PASSED.
Before making critical measurements, run SELF CAL to ensure the TIVP is compensated. The sensor tip does not need to be removed from the test point to successfully complete a self-calibration.
1. Connect TIVP to an oscilloscope channel.
2. Press the SELF CAL button on the probe compensation
box. Alternatively, send the CH<x>:PROBE:SELFCAL EXECUTE PI command to start the self-calibration programmatically.
The SELF CAL LED will blink while the
self-calibration is in process.
The SELF CAL LED is amber when there is a change
and a self-calibration is recommended. The LED is green if the last self-calibration is still valid.
Cable handlin g
Tektronix TIVP probes operate using several optical bers within a single cable between the compensation box and probe head. Therefore, standard optical ber handling practices are required. Avoid tight radius bends, crushing, crimping, twisting, or otherwise stressing the ber.
Standard accessories
MMCX probe tip
The 10X MMCX tip is included in every TIVP probe. The MMCX tip is recommended for the best bandwidth and CMRR performance. 0.100"
e Pin and 0.200" Wide Square Pin tips are available as optional
Squar accessories. Reorder part number: TIVPMX10X
100X sensor tip cable with 0.100" spaced square pin connectors
TIVPSQ100X
500X sensor tip cable with 0.200" spaced wide square pin connector
TIVPWS500X
1X MMCX sensor tip
TIVPMX1X
Square Pin to MMCX Adapter,
0.062" Spacing
131-9677-xx
Probe Tip Tripod Support
352-1170-xx
Lead, MMCX to IC Grabber
196-3546-xx
Lead, Square Pin to IC Grabber
196-3547-xx
Lead, Wide Square Pin to Banana Jack
020-3189-xx
Square Pin Y-lead
196-3434-xx
MicroCKT grabbers
206-0569–xx
Spare Pins for 0.062" Spaced Test
s
Point
020-3169-xx
Solder Aid for 0.062" Spaced Square Pins
003-1946-xx
Clearance r equirements
The unique common mode voltage range of the measurement system allows it to be
n the presence of high frequency/high
used i voltage common mode signals. It is important to observe all precautions while using this product.
WARNING. Electrical shock can occur while using this
urement system. The system is intended to isolate the
meas operator from hazardous input voltages (common voltages);
lastic case of the sensor head and the s hield on the
the p sensor tip cable do not supply safe isolation. Maintain the
clearance from the sensor head and sensor tip cable
safe while the measurement system is connected to the energized
uit as recommended in this document. Do not access the
circ RF Burn Hazard Zone while taking measurements on a live
uit. RF burn hazard zone is 1 meter (40 in) on all sides
circ
Controls and indicators
1. STATUS indicator.
LED Status Action
en (Solid)
Gre
Green (Blinking)
Red (Solid)
Red (Blinking)
Red (Blinking • • – )
Normal operation
Bulk power failure
Probe application failure
Probe application failure and bulk power failure
Self-Test failure
Power to probe
No head
-
Try unplugging and plugging back in. Inspect probe/scope interface. Service of probe may be required.
Try unplugging and plugging back in. Service of probe may be required.
Try unplugging and plugging back in. Inspect probe/scope interface. Service of probe and/or scope may be required.
Try unplugging and plugging back in. Service of probe may be required.
Probe bipod
Used to hold probe. TIVP can rotate in holder to accommodate square pin headers. Reorder part number:
1179-xx
352-
Probe tip adapter
AdaptanMMCXIsoVutipto
dard 0.100" spaced, 0.025"
stan square pins. Reorder part number:
131-9717-xx
SMA wrench/driver tool
5/16" wrench for use on SMA connector. Reorder part number:
003-1947-xx
IsoVu c arrying case
Soft case (with foam insert) protects the TIVP and enforces the optical ber minimum bend radius. Reorder part number:
016-2147-xx (2 m cable models)
016-2149-xx (10 m cable models)
Specications
e 1: Physical specications
Tab l
Characteristic Description
Net weight
(Weight does not include accessories and packaging.)
sor tip cable length
Sen
Fiber cable length
Tip cable assembly: 25g (0.055 lb)
Probehead:0.363kg(0.8lbs)
Compensation box: 0.816kg (1.8lbs)
03 cm (7.886 inches)
20.
2 meter (6.56 ft)
10 meter (32.81 ft)
Table 2: Electrical specications
Characteristic Description
Maximum Non-Destructive Input Voltage
ypical)
(T
DC Gain Accuracy (Typical) <3%
SMA Input (50 mode)
SMA Input (1 Mmode)
TIVPMX10X 250 Vpk
IVPMX50X
T
TIVPSQ100X
TIVPWS500X
TIVPMX1X
5 Vrms, (DC + peak AC)
100 Vpk
00 Vpk
3
600 Vpk
3300 Vpk
5VRMS(50), 100 Vpk (1 M)
x
Characteristic Description
Linear Input Voltage Range (Typical)
Common Mode Voltage Range (Typical)
System noise (Typical)
SMA Input (50 mode)
SMA Input (1 Mmode)
TIVPMX10X ±50 V
TIVPMX50X ±250 V
TIVPSQ100X
TIVPWS500X
TIVPMX1X ±5 V
SMA Input (50 mode)
SMA Input (1 Mmode)
TIVPMX10X ±200 mV range: 4.3 mV rms
TIVPMX50X ±1 V range: 21.5 mV rms
TIVPSQ100X
TIVPWS500X
±5 V
±5 V
±500 V
±2.5 kV
60 kVpeak
±20 mV range: 0.43 mV rms
±320 mV range: 1.46 mV rms
±5 V range: 48 mV rms
±20 mV range: 0.43 mV rms
±320 mV range: 1.46 mV rms
±5 V range: 48 mV rms
±3.2 V range: 14.6 mV rms
±50 V range: 480 mV rms
±16 V range: 73 mV rms
±250 V range: 2.4 V rms
±2 V range: 43 mV rms
±32 V range: 146 mV rms
±500 V range: 4.8 V rms
±10 V range: 215 mV rms
±160 V range: 730 mV rms
±2500 V range: 24 V rms
This product is not rated for CAT II, III or IV
Table 3: Environmental specications
Feature Description
Operating temperature
Operating humidity
Operating altitude
Comp box: 0° C to 30° C
Probe head: 0° C to 50° C
Tip Cable/Adapters: -40° C to 85° C
Comp box: 5% to 85% RH (Relative Humidity) at up to +30° C, non-condensing
Probe h ead: 5% to 85% RH (Relative Humidity)
o +40° C, 5% to 45% RH above +40° C up
at up t to +50° C, non-condensing.
Tip Cable/Adapters: 5% to 85% RH (Relative Humidity) at up to +85° C, non-condensing
3,000 meters
Compliance information
This section lists the Safety and Environmental standards with which the instrument complies. This product is intended for use by professionals and trained personnel only; it is not designed for use in households or by children.
Safety compliance
This section lists the safety standards with which the product complies and other safety compliance information.
EU low v oltage directive
Compliance was demonstrated to the following specication as listed in the Ofcial Journal of the European Union:
Low Voltage Directive 2014/35/EU.
EN 61010-1. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements.
U.S. nationally recognized testing laboratory listing
UL 61010-1. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements.
Canadian certication
CAN/CSA-C22.2 No. 61010-1. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements.
Additional compliances
IEC 61010-1. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements.
EN 60825-1/IEC 60825-1. Safety of Laser Products-Part 1: Equipment Classication and Requirements - Edition 3 (2014)
US 21CFR PT1010 Performance S tandard for Electronic Parts 2015.
US 21CFR PT1040 Performance Standards for Light Emitting Products 2015.
Equipment type
Test and measuring equipment. CLASS 1 LASER PRODUCT.
Pollution degree d escriptions
A measure of the contaminants that could occur in the environment around and within a product. Typically the internal environment inside a product is considered to be the same as the external. Products should be used only in the environment for which they are rated.
Pollution degree 1. No pollution or only dry, nonconductive pollution occurs. Products in this category are generally encapsulated, hermetically sealed, or located in clean rooms.
Pollution degree 2. Normally only dry, nonconductive pollution occurs. Occasionally a temporary conductivity that is caused by condensation must be expected. This location is a typical ofce/home environment. Temporary condensation occurs only when the product is out of service.
Pollution degree 3. Conductive pollution, or dry, nonconductive pollution that becomes conductive due to condensation. These are sheltered locations where neither temperature nor humidity is controlled. The area is protected from direct sunshine, rain, or direct wind.
Pollution degree 4. Pollution that generates persistent conductivity through conductive dust, rain, or snow. Typical outdoor locations.
Pollution degree rating
Pollution degree 2 (as dened in IEC 61010-1). Rated for indoor, dry location use only.
IP rating
IP20 (as dened in IEC 60529).
Measurement and overvoltage category descriptions
Measurement terminals on this product may be rated for measuring mains voltages from one or more of the following categories (see specic ratings marked on the product and in the manual).
Category I. Circuits not directly connected to a mains supply.
Category II. Circuits directly connected to the building wiring at utilization points (socket outlets and similar points).
Category III. In the building wiring and distribution system.
Category IV. At the source of the electrical supply to the building.
NOTE. Only measurement circuits have a measurement
category rating. Other circuits within the product do not have either rating.
Environmental consideration s
This section provides information about the environmental impact of the product.
Product end-of-life handling
Observe the following guidelines when recycling an instrument or component:
Equipment recycling. Production of this equipment required the extraction and use of natural resources. The e quipment may contain substances that could be harmful to the environment or human health if improperly handled at the product’s end of life. To avoid release of such substances into the environment and to reduce the use of natural resources, we encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused or recycled appropriately.
This symbol indicates that this product complies with the applicable European Union requirements according to Directives 2012/19/EU and 2006/66/EC on waste electrical and electronic equipment (WEEE) and batteries. For information about recycling options, check the Tektronix Web site (www.tek.com/productrecycling).
Important safety information
This manual contains information and warnings that must be followed by the user for safe operation and to keep the product in a safe condition.
Any servicing is not allowed, please contact Tektronix service.
Laser source is contained in the products.
General safety summary
Use the product only as specied. Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. Carefully read all instructions. Retain these instructions for future reference.
The product is designed to be used by trained personnel only.
Before use, always check the product with a known source to be sure it is operating correctly.
To Avoid Fire or Personal Injury
Connect and disconnect properly. Do not connect or disconnect
sensor tip cables, test leads, or accessories while they are connected to a voltage source. Use only test leads and
accessories supplied with the product, or indicated by Tektronix to be suitable for the product.
Observe all terminal ratings. To avoid re or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product. Do not exceed the Measurement Category (CAT) rating and voltage or current rating of the lowest rated individual component of a product or accessory.
Do not apply a potential that exceeds the maximum rating.
Do not operate without covers.
Avoid Exposed Circuitry. Do not touch exposed connections
and components when power is present.
Do not operate with suspected failures.. If you suspect that there is damage to this product, have it inspected by qualied service personnel.
Disable the product if it is damaged.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
Clean with Dry Cloth Only.
Sensor tip cables
Maintain safe clearance from the sensor head and sensor tip cable while connected to the energized circuit as recommended in this manual.
Remove the sensor tip cable and adapters from the test circuit when not in use.
Leave the sensor tip cable connected to the sensor head when not in use.
Use only correct Measurement Category (CAT), voltage, temperature, altitude, and amperage rate
d sensor tip cables and
accessories for any measurement.
Beware of high voltages. Understand the voltage ratings for the product you are using and do not exceed tho
se ratings. It is important to know and understand the maximum m easurement voltage rating of the product. The voltag
e rating depends on the measurement category, the instrument, and your application. Refer to the Specications section of the
manual for more
information.
WARNING. To prevent electrical shock, do
not exceed the
maximum measurement or maximum voltage category.
Connect and disconnect properly.
CAUTION. To avoid damage to the equipment, de-energize
the test circuit before connecting or di
sconnecting the
sensor tip cable.
Terms in This Manual.
These terms may appear in this manual:
WARNING. Warning statements identify conditions or
practices that could result in injury or
loss of life.
CAUTION. Caution statements identify conditions or
practices that could result in damage to
this product or
other property.
Isolated, electrically oating. The t
erms isolated, electrically oating,andgalvanically isolated are used in this document to indicate a measurement where there is n
o direct conduction
path to earth ground.
Symbols and terms on the product
These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
When this symbol is marked on the product, be sure to consult the manual to nd out the nature of the potential hazards and any actions which have to be taken to avoid them. (This symbol m be used to refer the user to ratings in the manual.)
ay also
The following symbol(s) may appear on the product:
Contacting Tektronix
Web site: www.tek.com
Phone: 1-800-833-9200
Address: Tektronix, Inc.
Email:
Department or name (if known) 14200 SW Karl Braun Drive P.O. Box 5 0 0 Beaverton, OR 97077 USA
techsupport@tek.com
Warranty information
For warranty information, go to www.tek.com/warranty.
Copyright © Tektronix, Inc. All rights reserved. www.tek.com