Tektronix TekExpress USB2 User manual

TekExpress® USB2 Automated Test Solution
Printable Application Help
*P077102100*
077-1021-00
TekExpress® USB2 Automated Test Solution
Printable Application Help
www.tektronix.com
077-1021-00
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by national copyright laws and international treaty provisions. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specifications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc. 14150 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
Worldwide, visit www.tektronix.com to find contacts in your area.

Table of Contents

Welcome .............................................................................................................................................. 5
Getting help and support
Related documentation ................................................................................................................... 1
Conventions used in help ................................................................................................................ 2
Technical support ........................................................................................................................... 2
Getting started
Installing the software .................................................................................................................... 5
Minimum system requirements ................................................................................................. 5
Supported instruments ............................................................................................................... 6
Required windows 7 user account setting ................................................................................. 6
Install the software .................................................................................................................... 7
Verify application installation ................................................................................................... 8
Activate the license ................................................................................................................... 8
View software version ............................................................................................................... 8
Required my TekExpress folder settings ................................................................................... 9
Set the my TekExpress folder permissions ............................................................................... 9
Application directories and their contents ............................................................................... 10
File name extensions ............................................................................................................... 11
Where test files are stored ....................................................................................................... 11
Operating basics
Run the application ....................................................................................................................... 13
Exit the application ....................................................................................................................... 14
Application controls ..................................................................................................................... 15
Options menu ................................................................................................................................ 17
Options menu overview ........................................................................................................... 17
Instrument control settings ...................................................................................................... 18
View connected instruments ................................................................................................... 18
Email settings .......................................................................................................................... 20
Configure email settings .......................................................................................................... 20
Application Panels
Application panels overview ........................................................................................................ 23
TekExpress® USB2 Printable Application Help i
Table of Contents
Setup tab ....................................................................................................................................... 25
Setup controls overview .......................................................................................................... 25
Set DUT parameters ................................................................................................................ 26
Select tests ............................................................................................................................... 28
Set acquisition parameters ....................................................................................................... 29
Running tests on prerecorded (saved) waveforms .................................................................. 31
Configuration tab parameters .................................................................................................. 31
Configuration tab: global settings and measurement parameters ............................................ 32
Preferences tab ........................................................................................................................ 34
Status panel overview ................................................................................................................... 35
Results panel ................................................................................................................................. 37
Results panel overview ............................................................................................................ 37
Preferences menu .................................................................................................................... 38
View test-related files .............................................................................................................. 38
Plots panel ..................................................................................................................................... 39
Reports panel ................................................................................................................................ 40
Reports panel overview ........................................................................................................... 40
Select report options ................................................................................................................ 41
View a report ........................................................................................................................... 43
Report contents ........................................................................................................................ 43
Running tests
Test process flow .......................................................................................................................... 45
Instrument and DUT connection setup ......................................................................................... 45
Running tests ................................................................................................................................ 46
Prerun checklist ............................................................................................................................ 46
Saving and recalling test setup files
Test setup files overview .............................................................................................................. 47
Save a test setup file ..................................................................................................................... 47
Open (load) a saved test setup file ................................................................................................ 48
Run a saved test in prerecorded mode .......................................................................................... 49
Create a new test setup file based on an existing one ................................................................... 50
TekExpress USB2 programmatic interface
About the programmatic interface ................................................................................................ 51
To enable remote access ............................................................................................................... 52
Requirements for developing TekExpress USB2 client ............................................................... 54
ii TekExpress® USB2 Printable Application Help
Remote proxy object ..................................................................................................................... 55
Client proxy object ....................................................................................................................... 56
Client programmatic interface example ....................................................................................... 58
Program remote access code example .......................................................................................... 60
Command list ................................................................................................................................ 61
Application status() ................................................................................................................. 61
Check session saved() ............................................................................................................. 62
Connect() ................................................................................................................................. 62
Disconnect() ............................................................................................................................ 64
DUT automation() ................................................................................................................... 64
Enter controller PC IP address() .............................................................................................. 66
Get current status info() ........................................................................................................... 67
Get or set timeout value() ........................................................................................................ 68
Get pass fail status() ................................................................................................................ 70
Get report parameter() ............................................................................................................. 70
Get results value() ................................................................................................................... 72
Lock server() ........................................................................................................................... 73
Lock session() .......................................................................................................................... 73
Query status() ......................................................................................................................... 74
Register status change notification() ....................................................................................... 75
Run with set configurations or stop the run operation ............................................................ 76
Save, recall or query a saved session ...................................................................................... 78
Save session as() ...................................................................................................................... 80
Select device() ......................................................................................................................... 80
Select pre-recorded waveform files ......................................................................................... 81
Select port() ............................................................................................................................. 82
Select probe type() ................................................................................................................... 83
Get results value for sub measurement() ................................................................................. 85
Select power condition() ......................................................................................................... 86
Select qualifier() ...................................................................................................................... 87
Select single test() ................................................................................................................... 89
Select suite() ............................................................................................................................ 90
Select test method() ................................................................................................................. 90
Select test mode() .................................................................................................................... 92
Select test point:Near End() .................................................................................................... 93
Select test point:Far End .......................................................................................................... 95
Select Tier() ............................................................................................................................. 96
Send response() ....................................................................................................................... 98
Select versions() ...................................................................................................................... 99
Table of Contents
TekExpress® USB2 Printable Application Help iii
Table of Contents
Set or get the DUT ID ........................................................................................................... 100
Set instrument() ..................................................................................................................... 102
Set verbose mode() ................................................................................................................ 102
Status() ................................................................................................................................... 103
Transfer images() .................................................................................................................. 104
Transfer result() ..................................................................................................................... 105
Transfer waveforms() ............................................................................................................ 105
Unlock server() ...................................................................................................................... 106
Unlock session() .................................................................................................................... 106
Reference
Handle error codes ...................................................................................................................... 107
HSETT controller ....................................................................................................................... 108
Setting up controller PC for automated DUT test mode ....................................................... 108
Signal validation ......................................................................................................................... 108
iv TekExpress® USB2 Printable Application Help

Welcome

Welcome to the TekExpress® USB2 Automated Test Solution application. TekExpress USB2 is a Signal Quality, Non-Signal Quality, Power Measurement and Receiver Sensitivity Measurement solution, which provides an automated, simple, and efficient way to test USB2 interfaces and devices consistent to the requirements of the USB2 specifications.
Key features and benefits
Comprehensive test coverage; select or deselect individual tests
Precise debugging and troubleshooting
USB-IF Signal Quality measurements integrated with TekExpress USB2
Automated Receiver Sensitivity measurement
User-friendly interface for plot inspection
Minimizes user intervention when performing time-consuming testing
Consolidated report for High Speed, Full Speed and Low Speed measurements
Complete programmatic interface enables automation scripts to call TekExpress USB2 functions
Flexible probe configuration
Comprehensive signal validation check
Automated DUT test mode control
USB-IF Signal Quality measurements integrated with TekExpress USB2

Getting help and support

Related documentation

The following documents are available as part of the TekExpress® USB2 Automated Test and Compliance Solution documentation set.
Table 1: Product documentation
Item Purpose Location
Application Help Application operation and User
PDF of the help Printable version of the
Interface help
Press F1 on the oscilloscope keyboard to open Application Help.
compiled help
See also
PDF file that ships with USB2 (USB2-Automated-Test-
Solution-Software-Printable­Help-EN-US.pdf).
Technical support
TekExpress® USB2 Printable Application Help 1
Getting help and support

Conventions used in help

Online Help uses the following conventions:
The term “DUT” is an abbreviation for Device Under Test.
The term “select” is a generic term that applies to the two methods of choosing a screen item (button, control, list item): using a mouse or using the touch screen.
Table 2: Icon descriptions
Icon Meaning
This icon identifies important information.
This icon identifies conditions or practices that could result in loss of data.
This icon identifies additional information that will help you use the application more efficiently.

Technical support

General information
Tektronix values your feedback on our products. To help us serve you better, please send us your suggestions, ideas, or comments on your application or oscilloscope. Contact Tektronix through mail, telephone, or the Web site. See
Contacting Tektronix for more information.
When you contact Tektronix Technical Support, please include the following information (be as specific as possible):
All instrument model numbers
Hardware options, if any
Probes used
Your name, company, mailing address, phone number, FAX number
Please indicate if you would like to be contacted by Tektronix about your suggestion or comments.
2 TekExpress® USB2 Printable Application Help
Getting help and support
Application specific
information
Software version number
Description of the problem such that technical support can duplicate the problem
If possible, save and send the setup files for all the instruments used and the application
If possible, save and send the TekExpress USB2 setup files, *.TekX (session files and folders), and status messages text file
If possible, save and send the waveform on which you are performing the measurement as a .wfm file
If possible, log files of Tektronix HSETT Controller from folder C:\Program Files\Tektronix\TekApplication\Tektronix.\HSETT Controller\
TekExpress® USB2 Printable Application Help 3
Getting help and support
4 TekExpress® USB2 Printable Application Help

Getting started

Installing the software

Minimum system
requirements
The following table shows the minimum system requirements needed for an oscilloscope to run TekExpress USB2.
Table 3: TekExpress USB2 system requirements
Component Requirement
Oscilloscope See Supported instruments
Processor Same as the oscilloscope
Operating System Same as the oscilloscope:
Windows 7 (64-bit only) SP1 Windows
7 user account settings
Memory Same as the oscilloscope
Hard Disk Same as the oscilloscope
Display Super VGA resolution or higher video adapter
(800 x 600 minimum video resolution for small fonts or 1024 x 768 minimum video resolution
Firmware
Software
for large fonts). The application is best viewed at 96 dpi display settings
TekScope 6.8.1 and later (for Windows 7)
Microsoft Internet Explorer 7.0 SP1 or later, or other Web browser for viewing reports.
1
Adobe Reader software 7.0 or later for viewing portable document format (PDF) files.
1
If TekExpress USB2 is running on an instrument that has a video resolution less than 800x600, connect and configure a second monitor to the instrument.
TekExpress® USB2 Printable Application Help 5
Getting started

Supported instruments

Table 4: Required equipment
Resource Model supported
Real-time oscilloscope Tektronix DPO/DSA/MSO70000C, D, and DX
series oscilloscopes (Windows 7 OS 64 bits) Tektronix MSO/DPO5000 and MSO/DPO5000B Series
Probes Two TCA-SMA cables
P7313SMA differential probe TCP0030A/TCP202/A probe TDP1500/TDP2500 differential probe P7500 differential probe
Host test fixtures USB-IF fixture set (host/device SQ and drop/
Device test fixtures USB-IF fixtures (host/device SQ and drop/droop)
Tektronix AWG instruments AWG7102, AWG7122 Series
2
droop) TDSUSBR (Tek fixture)
TDSUSBR (Tek fixture)
All AWG7000, AWG5000 Series AWG5014B, AWG5014C,AWG5012C, AWG5002C, AWG7001A, AWG7002A
See also. Minimum system requirements
2
Required windows 7 user
account setting
Windows 7 instruments need to have the User Account Control Settings set to
Never Notify. To set User Account Control Settings:
1. Go to Control Panel > User Accounts > Change User Account Control settings.
2. Set the sliding control to Never Notify as shown in the image, and click OK.
2
Available through USB-IF.
6 TekExpress® USB2 Printable Application Help
Getting started
See also. Supported oscilloscopes

Install the software

Use the following steps to obtain the latest TekExpress USB2 software from the Tektronix Web site and install on any compatible instrument running Microsoft Windows 7 (64-bit). See Minimum System Requirements for details.
1. Close all applications (including the TekScope application).
2. Go to the www.tek.com Web site and locate the Downloads fields.
3. Enter tekexpress usb2 in the Model or Keyword field, select Software from the Select Download Type list, and click GO.
4. Select the latest version of software from the search result. Follow
instructions to download the software file.
5. Copy or download the TekExpress USB2 installer executable file to the oscilloscope.
6. Double-click the installer .exe file to extract the installation files and launch the InstallShield Wizard. Follow the on-screen instructions.
Software is installed at C:\Program Files (x86)\Tektronix
\TekExpress\TekExpress USB2
7. Verify application installation
See also.
Minimum system requirements
Supported instruments
Required \My TekExpress folder settings
TekExpress® USB2 Printable Application Help 7
Getting started
Verify application
installation

Activate the license

To verify the installation was successful:
1. Open the TekScope application.
2. Click the Analyze menu.
3. Verify that TekExpress USB2 is listed in the Analyze menu.
4. Click TekExpress USB2 to open the application. Verify that the application
opens successfully.
See also.
Activate the license
Required \My TekExpress folder settings
Activate the license using the Option Installation wizard in the TekScope application:
1. In the TekScope application menu bar, click Utilities > Option Installation.
TheTekScope Option Installation wizard opens.
2. Push the F1 key on the oscilloscope keyboard to open the Option Installation help topic.

View software version

3. Follow the directions in the help topic to activate the license.
See also.
View version and license information
Required \My TekExpress folder settings
Use the following instructions to view version information for the application and for the application modules such as the Programmatic Interface and the Programmatic Interface Client.
To view version information for TekExpress USB2, click the Options button and select About TekExpress.
To view license and option key information in the TekScope applicaion:
1. In the TekScope application, select Help > About TekScope.
The Options dialog box is displayed.
2. Scroll through the Options list to locate USB2: TekExpress USB2.
3. To view the Option installation key value, look below the Options list.
See also.
Activate the license
Options menu
8 TekExpress® USB2 Printable Application Help
Getting started
Required my TekExpress
folder settings
Set the my TekExpress
folder permissions
Before you run tests for the first time, you need to Set the \My TekExpress folder
permissions.
See also.
Application directories and usage
File name extensions
Make sure that the My TekExpress folder has read and write access. Also verify that the folder is not set to be encrypted:
1. Right-click the folder and select Properties.
2. Select the General tab and then click Advanced.
3. In the Advanced Attributes dialog box, make sure that the option Encrypt contents to secure data is NOT selected.
4. Click the Security tab and verify that the correct read and write permissions
are set.
See also.
Application directories and usage
File name extensions
TekExpress® USB2 Printable Application Help 9
Getting started
Application directories
and their contents
The TekExpress USB2 application files are installed at the following location:
C:\Program Files (x86)\Tektronix\TekExpress\TekExpress USB2
The following table lists the application directory names and their purpose.
Table 5: Application directories and usage
Directory names Usage
Bin Contains TekExpress USB2 application libraries
Compliance Suites Contains compliance-specific files
Examples Contains various support files
ICP Contains instrument and TekExpress USB2
application-specific interface libraries
Lib Contains utility files specific to the TekExpress
USB2 application
Report Generator Contains style sheets for report generation
Tools Contains instrument and TekExpress USB2
application-specific files
AWG Files Contains files required for Receiver Sensitivity
measurement
See also.
View test-related files
File name extensions
10 TekExpress® USB2 Printable Application Help
Getting started

File name extensions

Where test files are stored

The TekExpress USB2 application uses the following file name extensions:
File name extension Description
.py Python sequence file
.xml Test-specific configuration information
(encrypted) files Application log files
.wfm Test waveform files
.mht Test result reports (default)
Test reports can also be saved in HTML format
.xslt Style sheet used to generate reports
See also.
View test-related files
Application directories and their contents
When you launch TekExpress USB2 for the first time, it creates the following folders on the oscilloscope:
\My Documents\My TekExpress\USB2
\My Documents\My TekExpress\USB2\Untitled Session
Every time you launch TekExpress USB2, an Untitled Session folder is created in the USB2 folder. The Untitled Session folder is automatically deleted when you exit the application. To preserve your test session files, save the test setup before exiting the application.
CAUTION.
Do not modify any of the session files or folders because this may result in loss of data or corrupted session files. Each session has multiple files associated with it. When you save a session, the application creates a .TekX file, and a folder named for the session that contains associated files, on the oscilloscope X: drive.
See also.
Set the \My TekExpress folder permissions
Application directories and usage
File name extensions
TekExpress® USB2 Printable Application Help 11
Getting started
12 TekExpress® USB2 Printable Application Help

Operating basics

Run the application

To start the TekExpress USB2 application, select Analyze > TekExpress USB2 from the TekScope menu.
The oscilloscope opens the TekExpress USB2 application:
When you first run the application after installation, the application checks for
Resources.xml located in the C:\Users\<username>\My TekExpress \USB2 folder. The Resources.xml file gets mapped to the X: drive when the
application launches. Session files are then stored inside the X:\USB2 folder. The
Resources.xml file contains information about available network-connected
instruments. If this file is not found, the application runs an instrument discovery program to detect connected instruments before launching TekExpress USB2.
NOTE. Do the steps in the Required\My TekExpress folder settings topic before running tests with the TekExpress USB2 application for the first time.
TekExpress® USB2 Printable Application Help 13
Operating basics
To keep the TekExpress USB2 application window on top, select Keep On Top from the Options menu. If the application goes behind the oscilloscope application, click Analyze > TekExpress USB2 to move the application to be in front.
See also

Exit the application

Required My TekExpress folder settings
Activate the license
Exit the application
Application controls
Application panel overview
To exit the application, click on the application title bar. Follow on-screen prompts to save any unsaved session, save test setup files, or exit the application.
NOTE. Using other methods to exit the application can result in abnormal termination of the application.
14 TekExpress® USB2 Printable Application Help
Operating basics

Application controls

Table 6: Application controls descriptions
Item Description
Options menu
Test Panel buttons
Menu to display global application controls.
Controls that open panels for configuring test settings and options.
Start / Stop button
Use the Start button to start the test run of the measurements in the selected order. If prior acquired measurements have not been cleared, the new measurements are added to the existing set. The button toggles to the Stop mode while tests are running. Use the Stop button to abort the test.
TekExpress® USB2 Printable Application Help 15
Operating basics
Item Description
Pause / Continue button
Clear button
Application window move icon
Use the Pause button to temporarily interrupt the current acquisition. When a test is paused, the button name changes to “Continue.”
Use the Clear button to clear all existing measurement results. Adding or deleting a measurement, or changing a configuration parameter of an existing measurement, also clears measurements. This is to prevent the accumulation of measurement statistics or sets of statistics that are not coherent. This button is available only on the Results panel.
Place the cursor over the three-dot pattern in the upper left corner of the application window. When the cursor changes to a hand, drag the window to the desired location.
See also
Application panel overview
16 TekExpress® USB2 Printable Application Help
Operating basics

Options menu

Options menu overview

The Options menu is located in the upper right corner of the application.
The Options menu has the following selections:
Menu Function
Default Test Setup Opens an untitled test setup with defaults
selected
Open Test Setup Opens a saved test setup
Save Test Setup Saves the current test setup selections
Save Test Setup As Creates a new test setup based on an existing
one
Open Recent Displays a menu of recently opened test setups
to select from
Instrument Control Settings Detects, lists, and refreshes the connected
instruments found on specified connections (LAN, GPIB, USB, and so on)
Keep On Top Keeps the TekExpress USB2 application on top
of other open windows on the desktop
Email Settings Use to configure email options for test run and
results notifications
Help Displays the TekExpress USB2 help
About TekExpress
Displays application details such as software name, version number, and copyright
Options menu
Provides access to License information for your TekExpress USB2 installation
Provides a link to the Tektronix Web site
TekExpress® USB2 Printable Application Help 17
Operating basics
See also. Application controls
Instrument control
settings
Use the TekExpress Instrument Control Settings dialog box to search for and list the connected resources (instruments) detected on selected connections (LAN, GPIB, USB), and each instruments connection information.
To access, click Options > Instrument Control Settings.
View connected
instruments
Use the Instrument Control Settings feature to search for connected instruments and view instrument connection details. You can select listed connected instruments for use in the Global Settings tab in the test Configuration pane.
See also. Options menu overview

Use the Instrument Control Settings dialog box to view or search for connected instruments required for the tests. The application uses TekVISA to discover the connected instruments on all selected connection types.

NOTE. The correct instruments for the current test setup must be connected and recognized by the application before running tests.
To refresh the list of connected instruments:
1. From the Options menu, select Instrument Control Settings.
2. In the Search Criteria section of the Instrument Control Settings dialog
box, select the connection types of the instruments for which to search.
18 TekExpress® USB2 Printable Application Help
Operating basics
Instrument search is based on the VISA layer, but different connections determine the resource type, such as LAN, GPIB, and USB. For example, if you choose LAN, the search will include all the instruments supported by TekExpress that are communicating over the LAN.
3. Click Refresh. TekExpress searches for connected instruments.
4. After searching, the dialog box lists the instrument-related details based on
the search criteria you selected. For example, if you selected LAN and GPIB as the search criteria, the application checks for the availability of instruments over LAN, then GPIB, and then lists detected instruments on those connection types.
The Retrieved Instruments table lists instrument details. The time and date of the last time this table was updated is displayed in the Last Updated field.
See also. Equipment connection setup
TekExpress® USB2 Printable Application Help 19
Operating basics

Email settings

Use the Email Settings utility to configure email notifications to receive an email message when a test completes, produces an error, or fails. Select the type of test session information to include in the email, such as test reports and test logs, the email message format, and the email message size limit.
Select Options > Email Settings to open this dialog box.
NOTE. Recipient email address, sender’s address, and SMTP Server are mandatory fields.

Configure email settings

See also.
Configure email settings
Options menu
Select test notification preferences
Use the Email Settings dialog box to be notified by email when a test completes, fails, or produces an error:
1. Select Options > Email Settings to open the Email Settings dialog box.
2. (Required) For Recipient email Address(es), enter one or more email
addresses to which the test notification has to be sent. To include multiple addresses, separate the addresses with commas.
3. (Required) For Sender’s Address, enter the email address used by the instrument. This address consists of the instrument name followed by an underscore followed by the instrument serial number, then the @ symbol and the email server used. For example: DPO72016C_B130099@yourcompany.com.
20 TekExpress® USB2 Printable Application Help