TDS3000C Series Oscilloscope Compliance and Safety
Instruct
1
*P0713
071-3578-00
Contacting Tektronix
Tektronix, Inc., 14150 SW Karl Braun Drive, P.O. Box 500, Beaverton, OR 97077, USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200. Worldwide, visit www.tektronix.com to nd contacts in your area.
This manual contains information and warnings that must be
wed by the user for safe operation and to keep the product
follo in a safe condition. See t he TDS3000C Series User Manual
ronix part number 071-2308-XX), which is available on
(Tekt the CD provided with your product and on the Tektronix Web
www.tek.com).
site (
To safely perform service on this product, additional information is provided at the end of this section.
Gener
Use the product only as specied. Review the following safety
utions to avoid injury and prevent damage to this product
preca or any products connected to it. Carefully read all instructions.
n these instructions for future reference.
Retai
Comply with local and national safety codes.
For correct and safe operation of the product, it is essential that you follow generally accepted safety procedures in addition to the safety precautions specied in this manual.
The product is designed to be used by trained personnel only.
Only qualied personnel who are aware of the hazards involved should remove the cover for repair, maintenance, or adjustment.
Before use, always check the product with a known source to be sure it is operating correctly.
product is not intended for detection of hazardous voltages.
This
Use personal protective equipment to prevent shock and arc blast injury where hazardous live conductors are exposed.
While using this product, you may need to access other parts of a larger system. Read the safety sections of the other component manuals for warnings and cautions related to operating the system.
When incorporating this equipment into a system, the safety of
system is the responsibility of the assembler of the system.
that
ions
57800*
tant safety information
al safety summary
To av oid re or personal injury
Use proper power cord. Use only the power cord specied for
this product and certied for the country of use.
Do not use the provided power cord for other products.
Ground the product. This product is grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, make sure that the product is properly grounded.
Power disconnect. The power cord disconnects the product from the power source. See instructions for the location. Do not position the equipment so that it is difcult to operate the power cord; it must remain accessible to the user at all times to allow for quick disconnection if needed.
Connect and disconnect properly. Do not connect or disconnect probes or test leads while they are connected to a voltage source.
ly insulated voltage probes, test leads, and adapters
Use on supplied with the product, or indicated by Tektronix to be
ble for the product.
suita
Do not disable the power cord grounding connection.
Observe all terminal ratings. To avoid re or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product. Do not exceed the Measurement Category (CAT) rating and voltage or current rating of the lowest rated individual component of a product, probe, or accessory. Use caution when using 1:1 test leads because the probe tip voltage is directly transmitted to the product.
Do not apply a potential to any terminal, including the c ommon terminal, that exceeds the maximum rating of that terminal.
Do not oat the common terminal above the rated voltage for that terminal.
The measuring terminals on this product are not rated for connection t o mains Category III or IV circuits, or to mains Category II circuits greater than 100 V.
When connected to circuits other than mains, the maximum transient overvoltage rating is 400 V peak.
Do not operate without covers. Do not operate this product with covers or panels removed, or with the case open. H azardous voltage exposure is possible.
Avoid exposed circuitry. Do not touch exposed connections and components when power is present.
Do not operate with suspected failures. If you suspect that there is damage to this product, have it inspected by qualied service personnel.
Disable the product if it is damaged. Do not use the product if it is damaged or operates incorrectly. If in doubt about safety of the product, turn it off and disconnect the power cord. Clearly mark the product to prevent its further operation.
Before use, inspect voltage probes, test leads, and accessories for mechanical damage and replace when damaged. Do not use probes or test leads if they are damaged, if there is exposed metal, or if a wear indicator shows.
Examine the exterior of the product before you use it. Look for cracks or missing pieces.
Use only specied replacement parts.
Do not operate in wet/damp conditions. Be aware that condensation may occur if a unit is moved from a cold to a warm environment.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry. Remove the input signals
before you clean the product.
Provide proper ventilation. Refer to the installation instructions in the TDS3000C Series User Manual for details on installing the product so it has proper ventilation. Slots and openings are provided for ventilation and should never be covered or otherwise obstructed. Do not push objects into any of the openings.
Provide a safe working environment. Always place the product in a location convenient for viewing the display and indicators.
Avoid improper or prolonged use of keyboards, pointers, and button pads. Improper or prolonged keyboard or pointer use may result in serious injury.
Be sure your work area meets applicable ergon Consult with an ergonomics professional to avoid stress injuries.
Probes and test leads
See the TDS3000C Series User Manual for Important Safety Information on the use, ratings, and selection of probes recommended for use with this product.
Before connecting probes or test leads, connect the power cord from the power connector to a properly grounded power outlet.
Keep ngers behind the nger guards on the probes. Do not
h metallic portion of the probe head while it is connected
touc to a voltage source and e nsure the reference lead or spring are
y mated before connecting the probe to the circuit under test.
full
omic standards.
Remove all probes, test leads and accessories that are not in use.
Use only correct Measurement Category (CAT), voltage, temperature and a dapters for any measurement.
Beware of high voltages. Understand the voltage ratings for the probe yo ratings are important to know and understand:
These two vo application. Refer to the Specications section of the manual for more inf
Connect and disconnect properly. Connect the probe output to the measu to the circuit under test. Connect the probe reference lead to the circu Disconnect the probe input and the probe reference lead from the circuit measurement product.
Connect and disconnect properly. De-energize the circuit under test befor
Connect the probe reference lead to earth ground only.
Do not connect a current probe to any wire that carries voltages above the current probe voltage rating.
Inspect the probe and accessories. Before each use, inspect probe and accessories for damage (cuts, tears, or defects in the probe body, accessories, or cable jacket). Do not use if damaged.
Ground-referenced oscilloscope use. Do not oat the reference lead of this probe when using with ground-referenced oscilloscopes. The reference lead must be connected to e arth potential (0 V).
Floating measurement use. Do not oat the reference lead of this probe above the rated oat voltage.
, altitude, and amperage rated probes, test leads,
u are using and do not exceed those ratings. Two
The maximum measurement voltage from the probe tip to the probe reference lead
The maximum oating voltage from the probe reference lead to earth ground
ltage ratings depend on the probe and your
ormation.
WARNING. To prevent electrical shock, do not exceed the
maximum mea the oscilloscope input BNC connector, probe tip, or probe reference l
e connecting or disconnecting the current probe.
surement or maximum oating voltage for
ead.
rement product before connecting the probe
it under test before connecting the probe input.
under test before disconnecting the probe from the
Termsinthismanual
These terms may appear in this manual:
WARNING. Warning statements identify conditions or
practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or
practices that could result in damage to this product or other property.
Symbols and terms on the product
These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WAR N ING i accessible as you read the marking.
CAUTION indicates a hazard to property including the product.
The following symbol(s) may appear on the product:
ndicates an injury hazard not immediately
When this symbol is marked on the product, be sure to consult the manual to nd out the nature of the potential hazards and any actions which have
en to avoid them. (This symbol may also
to be tak be used to refer the user to ratings in the manual.)
x
Compliance information
This section lists the EMC (electromagnetic compliance), safety, and environmental standards with which the instrument complies.
Questions about the compliance information below may be directed to the following address:
Tektronix, Inc. PO Box 500, MS 19045 Beaverton, OR 97077, USA www.tek.com
EMC compliance
EU EMC Directive
Meets intent of Directive 2014/30/EU for Electromagnetic Compatibility. Compliance was demonstrated to the following specications as listed in the Ofcial Journal of the European Communities:
EN 61326-1, EN 61326-2-1. EMC requirements for electrical equipment for measurement, control, and laboratory use.
CISPR 11. Radiated and conducted emissions, Group 1, Class A
IEC 61000-4-2. Electrostatic discharge immunity
IEC 61000-4-3. RF electromagnetic eld immunity
IEC 61000-4-4. Electrical fast transient/burst immunity
IEC 61000-4-5. Power line surge immunity
IEC 61000-4-6. Conducted RF immunity
6
IEC 61000-4-11. Voltage dips and interruptions immunity
EN 61000-3-2. AC power line harmonic emissions
EN 61000-3-3. Voltage changes, uctuations, and icker
1
This product is intended for use in nonresidential areas only. Use in residential areas may cause electromagnetic interference.
2
Emissions which exceed the levels required by this standard may occur when this equipment is connected to a test object.
3
Equipment may not meet the immunity requirements of applicable listed standards when test leads and/or test probes are connected due to coupling of electromagnetic interference onto those leads/probes. To minimize the inuence of electromagnetic interference, minimize the loop area between the unshielded portions of signal and associated return leads, and keep leads as far away as possible from electromagnetic disturbance sources. Twisting unshielded test leads together is an effective way to reduce loop area. For probes, keep the ground return lead as short as possible and close to the probe body. Some probes have accessory probe tip adapters to accomplish this most effectively. In all cases, observe all safety instructions for the probes or leads used.
4
For compliance with the EMC standards listed here, high quality shielded interface cables that incorporate low impedance connection between the cable shield and the connector shell should be used.
1234
5
5
3.0 division waveform displacement or 6.0 division increase in peak-to-peak noise is allowed when the instrument is subjected to elds and signals as dened in the IEC 61000-4-3 test.
6
1.5 division waveform displacement or 2.0 division increase in peak-to-peak noise is allowed when the instrument is subjected to elds and signals as dened in the IEC 61000-4-6 test.
Australia / New Zealand Declaration of Conformity – EMC
Complies with the EMC provision of the Radiocommunications Act per the following standard, in accordance with ACM A:
CISPR 11. Radiated and Conducted Emissions, Group 1, Class A, in accordance with EN 61326-1 and EN 61326-2-1.
Safety compliance
This section lists the safety compliance information.
Pollution degree rating
Pollution degree 2 (as dened in IEC 61010-1). Rated for indoor, dry location use only.
Measurement and overvoltage category description
Measurement terminals on this product may be rated for measuring mains voltages from the following category (see specic ratings marked on the product and in the manual).
Category II. Circuits directly connected to the building wiring at utilization points (socket outlets and similar points).
NOTE. Only mains power supply circuits have an overvoltage
category rating. Only measurement circuits have a measurement category rating. Other circuits within the product do not have either rating.
Mains overvoltage category rating
Overvoltage category II (as dened in IEC 61010-1).
Environmental co nsiderations
Refer to the User Manual for information about the environmental impact of the product.
Operating overview
Refer to the User Manual for complete operating information and product specications.
Power source
Use a power source that delivers 100 VACto 240 VAC(±10%), with a frequency between 47 Hz and 440 Hz.
This Oscilloscope operates from a single-phase power source with the neutral conductor at or near earth ground. It is intended
for only ground-referenced measurements. A protective ground connection through the grounding conductor in the power cord is essential for safe operation.
The line fuse is internal and is not operator replaceable.
Table 1: Maximum voltage ratings
Rating Description
Maximum voltage at input BNC (1 M)
Overvoltage category
Non-mains environment
CAT II environment 100 VRMS (400 Vpk,
For steady-state sinusoidal waveforms, derate at
Maximum voltage
(400 VPK,Duty
150 V
RMS
Factor of 37.5%)
Duty Factor of 25%)
20 dB/decade above 200 kHz to 13 Vpk at 3 MHz and above.
Probe
0V ground reference lead
Table 2: Environmental ratings
Rating Description
Operating temperature
0°Cto+50°C(+32°Fto+122°F)
Nonoperating range (storage):
-40 °C to +71 °C (-40 °F to +160 °F)
Operating humidity
Operating
5% to 95% RH below 30 °C (+86 °F), derate upper limit to 45% RH at 50 °C (+122 °F)
Up to 3000 meters (10,000 feet)
altitude
Copyright © Tektronix, Inc. All rights reserved. www.tek.com
xx
TDS3000Cシリーズ・オシロスコープ
適合性と安全性
取扱説明書
2
*P0713
071-3578-00
当社へのお問合せ
Tektronix, Inc., 14150 SW Karl Braun Drive, P.O. Box 500, Beaverton, OR 97077, USA
製品情報、代理店、サービス、およびテクニカル・サポート:
北米:1-800-833-9200までお電話ください。 世界の他の地域では、www.tektronix.com にアクセスし、 お近くの代理店をお探しください。
安全性に関
このマニュアルには、操作を行うユーザの安全を確保し、製 品を安全な および警告が記載されています。『TDS3000Cシリーズ・ユー ザ・マニュ ださい。マニュアルは製品に付属するCDに収録されている ほか、当社 いただけます。
このセクションの最後には、製品を安全に保守するために必 要な追加情報が記載されています。
安全にご使用いただくために
製品は指定 傷を避け、本製品や本製品に接続されている製品の破損を 防止するた みください。すべての指示事項を注意深くお読みください。 必要なと しておいてください。
該当する地域および国の安全基準に従ってご使用ください。
本製品を正しく安全にご使用になるには、このマニュアルに 記載された注意事項に従うだけでなく、一般に認められてい る安全対策を徹底しておく必要があります。
本製品は 想定しています。
製品のカバーを取り外して修理や保守、または調整を実施 できるのは、あらゆる危険性を認識した専門的知識のある適 格者のみに限定する必要があります。
使用前に、既知の情報源と十分に照らし合わせて、製品が 正しく動作していることを常にチェックしてください。
本製品は危険電圧の検出用にはご利用になれません。
危険な通電導体が露出している部分では、感電やアーク・フ ラッシュによってけがをするおそれがありますので、保護具を 使用してください。
本製品をご使用の際に、より大きな他のシステムにアクセスし なければならない場合があります。他のシステムの操作に関 する警告や注意事項については、その製品コンポーネントの マニュアルにある安全に関するセクションをお読みください。
57800*
する重要な情報
状態に保つために順守しなければならない情報
アル』(当社部品番号077-1395-XX)を参照してく
のWebサイト (www.tek.com) よりダウンロー ドして
された方法でのみご使用ください。人体への損
めに、安全性に関する次の注意事項をよくお読
きに参照できるように、説明書を安全な場所に保管
訓練を受けた専門知識のあるユーザによる使用を
本機器をシステムの一部としてご使用になる場合には、その システムの構築者が安全性に関する責任を果たさなければ なりません。
火災や人体への損傷を避けるには
適切な電源コードを使用してください: 本製品用に指定さ
れ、使用される国で認定された電源コードのみを使用してく ださい。
他の製品の電源コードは使用しないでください。
本製品を接地してください: 本製品は、電源コードのグラン ド線を使用して接地します。感電を避けるため、グランド線を アースに接続する必要があります。本製品の入出力端子に 接続する前に、本製品が正しく接地されていることを確認し てください。
電源を切断してください: 電源コードの取り外しによって主 電源が遮断されます。スイッチの位置については、使用説明 書を参照してください。電源コードの取り扱いが困難な場所 には設置しないでください。必要に応じてすぐに電源を遮断 できるように、ユーザが常にアクセスできる状態にしておく必 要があります。
接続と切断の手順を守ってください: プローブとテスト・リー ドが電圧源に接続されている間は接続または切断しないで ください。
絶縁型の電圧プローブ、テスト・リード、およびアダプタは、 製品に付属する製品か、または当社により特別に指定され た製品のみを使用してください。
電源コードのグランド接続を無効にしないでください。
すべての端子の定格に従ってください: 火災や感電の危険 を避けるために、本製品のすべての定格とマーキングに従っ てください。本製品に電源を接続する前に、定格の詳細に ついて、製品マニュアルを参照してください。測定カテゴリ (CAT)の定格および電圧と電流の定格については、製品、 プローブ、またはアクセサリのうちで最も低い定格を超えな いように使用してください。1:1のテスト・リードを使用するとき は、プローブ・チップの電圧が直接製品に伝わるため注意が 必要です。
コモン端子を含むいかなる端子にも、その端子の最大定格 を超える電圧をかけないでください。
端子の定格電圧を超えてコモン端子をフローティングさせな いでください。
本製品の測定端子は、カテゴリIIIまたはIVの主電源回路、あ るいは100Vを超えるカテゴリIIの主電源回路には使用できま せん。
主電源以外の回路に接続するときの最大過渡過電圧定格 は、400Vピー クです。
カバーを外した状態では使用しないでください: カバーやパ ネルを外した状態やケースを開いたまま動作させないでくだ さい。危険性の高い電圧に接触してしまう可能性があります。
露出した回路への接触は避けてください: 電源が投入され ているときに、露出した接続部分やコンポーネントに触れな いでください。
故障の疑いがあるときは使用しないでください: 本製品に 故障の疑いがある場合には、資格のあるサービス担当者に 検査を依頼してください。
製品が故障している場合には、使用を停止してください。製 品が故障している場合や正常に動作していない場合には、 製品を使用しないでください。安全上の問題が疑われる場 合には、電源を切って電源コードを取り外してください。誤っ て使用されることがないように、問題のある製品を区別してお いてください。
使用前に、電圧プローブ、テスト・リード、およびアクセサリに 機械的損傷がないかを検査し、故障している場合には交換 してください。金属部が露出していたり、摩耗インジケータが 見えているなど、損傷が見られるプローブまたはテスト・リー ドは使用しないでください。
使用する前に、製品の外観に変化がないかよく注意してくだ さい。ひび割れや欠落した部品がないことを確認してくださ い。
指定された交換部品のみを使用するようにしてください。
湿気の多いところでは動作させないでください: 機器を寒い 場所から暖かい場所に移動する際には、結露にご注意くだ さい。
爆発性のガスがある場所では使用しないでください: 製品の表面を清潔で乾燥した状態に保ってください: 製品
の清掃を開始する前に、入力信号を取り外してください。
適切に通気してください: 適切な通気を確保できるように製 品を 設 置 する方法につい て は 、『TDS3000Cシリー ズ ・ユー ザ・マニュアル』にある設置手順を参照してください。製品に は通気用のスロットや開口部があります。その部分を覆った り、通気が妨げられたりすることがないようにしてください。開 口部には異物を入れないでください。
安全な作業環境を確保してください: 製品は常にディスプレ イやインジケータがよく見える場所に設置してください。
キーボードやポインタ、ボタン・パッドを不適切に使用したり、 長く押しすぎたりしないでください。キーボードやポインタの 使用を誤ると、大けがにつながる可能性があります。
作業場が該当する人間工学規格を満たしていることを確認 してください。ストレスに由来するけががないように、人間工 学の専門家に助言を求めてください。
プローブとテスト・リード
本製品での使用が推奨されるプローブの使用、定格、および 選択に関しては、『TDS3000Cシリーズ・ユーザ・マニュアル』 の「安全性に関する重要な情報」を参照してください。
プローブやテスト・リードを接続する前に、電源コネクタから の電源コードを適切に接地されたコンセントに接続してくだ さい。
感電を避けるために、指ガードの先に指を出さないように注 意してください。プロー プローブ・ヘッドの金属部分に触れないでください。また、プ ローブを被測定回路に接 リングが確実に取り付けられていることを確認してください。
使用しないプローブ、テスト・リード、アクセサリはすべて取り 外してください。
測定に使用するプローブ、テスト・リード、アダプタは、測定 カテゴリ(CAT)、電圧、温度、高度、アンペア数の定格が適 切なもののみを使用してください。
高電圧に注意: 使用するプローブの電圧定格について理解 し、その定格を超えないようにしてください。特に次の2つの 定格についてはよく理解しておく必要があります。
プローブ・チップとプローブの基準リード間の最大測定 電圧
プローブ基準リードとアース間の最大フローティング電圧
上記の2つの電圧定格はプローブと用途によって異なります。 詳細については、プローブのマニュアルの仕様関連セクショ ンを参照してください。
警告: 感電を防止するために、オシロスコープの入力
BNCコネクタ、プローブ・チップ、またはプローブ基準リー ドの最大測定電圧や最大フローティング電圧を超えない ように注意してください。
接続と切断の手順を守ってください: プローブ出力を測定 器に接続してから、プローブを被測定回路に接続してくださ い。被測定回路にプローブの基準リードを接続してから、プ ローブ入力を接続してください。プローブ入力とプローブの 基準リードを被測定回路から切断した後で、プローブを測定 器から切断してください。
接続と切断の手順を守ってください: 被測定回路の電源を 切ってから、電流プローブの接続あるいは切断を行ってくだ さい。
プローブの基準リードは、グランドにのみ接続してください。
電流プローブを、その定格電圧を超える電圧がかかってい る電線に接続しないでください。
プローブとアクセサリを検査してください: 使用前には必ず プローブとアクセサリに損傷がないことを確認してください(プ ローブ本体、アクセサリ、ケーブル被覆などの断線、裂け目、 欠陥)。損傷がある場合には使用しないでください。
グランド基準のオシロスコープの使用: グランド基準のオシ ロスコープで使用する場合、プローブの基準リードはフロー ティングさせないでください。基準リードは接地電位(0V)に 接続しなければなりません。
フローティング測定の使用: 本プローブの基準リードは、定 格フローティング電圧を超えてフローティングさせないでくだ さい。
ブを電圧ソースに接続している間は、
続する前に、基準リードまたはスプ
本マニュアル内の用語
このマニュアルで
警告: 人体や生命に危害をおよぼすおそれのある状態
や行為を示します
注意: 本製品やその他の接続機器に損害を与えるおそ
れのある状態や行
本製品に使用さ
本製品では、次の用語を使用します。
危険: ただちに人体や生命に危険をおよぼす可能性が あることを示します。
警告: 人体や生命に危険をおよぼす可能性があること を示します。
注意: 本製品を あることを示します。
本製品では、次の記号を使用します。
は次の用語を使用します。
為を示します。
れる記号と用語
含む周辺機器に損傷を与える可能性が
製品にこの記号 ニュアルを参照して、想定される危険性とそれ らを回避するために必要な行動について確認 してください(マニュアルでは、この記号はユー ザに定格を示すために使用される場合があり ます。)
が表記されているときは、マ
x
適合性に関する情報
このセクションでは、本製品が適合しているEMC基準、安全 基準、および環境基準について説明します。
コンプライアンスに関するご質問は、以下の住所宛に、直接 お問い合わせいただくこともできます。
Tektronix, Inc.PO Box 500, MS 19‐045 Beaverton, OR 97077, USA www.tek.com
EMC適合性
欧州EMC指令
指令2014/30/EU電磁環境両立性に適合します。『Official Journal of the European Communities』に記載の以下の基 準 に準拠します。
EN 61326-1、EN 61326-2-1: 測定、制御、および実験用途 の電子機器を対象とするEMC基準。
CISPR 11: グループ1、クラスA、放射および伝導エミッ ション
IEC 61000-4-2: 静電気放電イミュニティ
IEC 61000-4-3: RF電磁界イミュニティ
IEC 61000-4-4: 電流高速トランゼント/バースト・イミュ ニティ
IEC 61000-4-5: 電力線サージ・イミュニティ
IEC 61000-4-6: 伝導RFイミュニティ
IEC 61000-4-11: 電圧低下と瞬時停電イミュニティ
EN 61000-3-2: AC電源ライン高調波エミッション
EN 61000-3-3: 電圧の変化、変動、およびフリッカ
1
本製品は住居区域以外での使用を目的としたものです。住居区域 で使用すると、電磁干渉の原因となることがあります。
2
本製品をテスト対象に接続した状態では、この規格が要求するレ ベルを超えるエミッションが発生する可能性があります。
3
機器にテスト・リードまたはテスト・プローブが接続されているときに は、これらのリード/プローブに電磁干渉がカップリングされるた め、ここに記載されている標準により規定されたイミュニティ要件を 満たせるとは限りません。電磁干渉による影響を最小限に抑えるに は、信号の非シールド部分と対応するリターン・リードの間のループ 領域を最小にします。また、電波障害の発生源からできるだけ遠ざ けるようにします。ループ領域を少なくするための効率的な方法は、 非シールド部分のテスト・リードをツイストペアにすることです。プ ローブの場合、グランド・リターン・リードをできるだけ短くし、プロー ブ本体に近づけるようにします。そうした処置を効率的に行えるよう に、プローブによっては、アクセサリとしてプローブ・チップ・アダプタ が提供されている機種もあります。いずれの場合も、使用するプ ローブまたはリードの取扱説明書を十分に読むようにしてください。
4
ここに挙げた各種EMC規格に確実に準拠するには、ケーブル・シー ルドとコネクタ 質なシールドを持つインタフェース・ケーブルが必要です。
・シェルを低インピーダンスで接続できるように、高品
1234
5
6
5
IEC 61000-4-3試験に規定のRF電磁界/伝導性RFの干渉を受け た場合、本機器は3.0div以下の波形変位および6.0div以下のp-p ノイズ増加を生じます。
6
IEC 61000-4-6試験に規定のRF電磁界/伝導性RFの干渉を受け た場合、本機器は1.5div以下の波形変位および2.0div以下のp-p ノイズ増加を生じます。
オーストラリア/ニュージーランド適合宣言 - EMC
ACMAに従い、次の規格に準拠することで Radiocommunications ActのEMC条項に適合しています。
CISPR 11: グループ1、クラスA、放射および伝導エミッ ション(EN61326-1およびEN61326-2-1に準拠)
安全性に関する適合性
このセクションでは、安全規格への適合性に関する情報を示 します。
汚染度
汚染2(IEC 61010-1の定義による)。乾燥した屋内でのみ使 用できます。
測定および過電圧カテゴリについて
本製品の測定端子は、測定する電源電圧について次のカテ ゴリに評価されます(製品やマニュアルへの特定の評価を参 照)。
カテゴリII: 固定設備の屋内配線に直接接続される回路 (壁コンセントおよび類似する設備)。
注: 過電圧カテゴリ定格に該当するのは主電源回路のみで
す。測定カテゴリ定格に該当するのは測定回路のみです。製 品内部のその他の回路にはいずれの定格も該当しません。
主電源過電圧カテゴリ定格
過電圧カテゴリII(IEC 61010-1の定義による)。
環境条件
本製品が環境に及ぼす影響については、『ユーザ・マニュア ル』を参照してください。
操作方法の概要
操作方法および製品の仕様に関する詳細は、ユーザ・マニュ アルを参照してください。
電源
100VAC~240VAC(±10%)、47Hz~440Hzを供 給する電 源を 使用してください。
本製品は接地 します。接地基準測定を唯一の用途として想定しています。 安全な操作 護用のグランド接続が不可欠です。
電位近傍の中性線を有する単相電源で動作
のためには、電源コード内の接地線を通じた保
電源ヒューズはオシロスコープ内部に設置されているので、 ユーザによる交換はできません。
表 3: Maximum voltage ratings
定格 説明
最大 電圧 (BNC 入力 端子、 1MΩ)
過電圧カテゴリ
主電源 以外の環境 150V
最大電圧
RMS
( 400VPK、
デューティ・ファクタ:
37.5%)
CAT II環境 100VRMS ( 400Vpk 、
デューティ・ファクタ: 25%)
安定した正弦波では、200kHzを超えると 20dB/decade の 割 合で低下し、 3MHz を超えると 13Vpkになります。
プロー
0V
ブ・グラ ンド(基 準)リー ド
表 4: Environmental ratings
定格 説明
動作温度
0℃ ~+50
℃(+32°F~+122°F) 非動作時(保管): -40℃~+71℃(-40°F~+160°F)
動作湿度 30℃(+86゚F)以下で5%~95% RH、50℃(+
122゚F)では 上限 45% RH
動作高度
3,000m(10,000フィート)以下
Copyright © Tektronix, Inc.All rights reserved. www.tek.com
xx
TDS3000C 系列示波器
合规性和安全性
说明
将产品接地: 本产品通过电源线的接地导线接地。为避免
电击,必须将接地导线与大地相连。在对本产品的输入端 或输出端进行连接之前,请务必将本产品正确接地。
断开电源: 电源线可以使产品断开电源。请参阅有关位置 的说明。请勿将设备放在难以操作电源线的位置;必须保 证用户可以随时操作电源线,以便需要时快速断开连接。
正确连接并正确断开连接: 探头或测试导线连接到电压源 时请勿插拔。
仅使用产品附带的或 T 缘电压探头、测试导线和适配器。
不要切断电源线的接地连接。
遵守所有终端额定值: 为避免火灾或电击危险,请遵守产 品上所有的额定值和标记说明。在连接产品之前,请先查 看产品手册,了解额定值的详细信息。不要超过本产品、 探头或附件中各组件的额定值最低的测量类别 (CAT) 额 定值和电压或电流额定值。在使用 1:1 测试导线时要小 心,因为探头端部电压会直接传输到产品上。
对任何终端(包括公共终端)施加的电压不要超过该终端 的最大额定值。
请勿将公共终端浮动到
本产品的测量终端额定值不适用于连接到市电 III 或 IV 类型电路,或高于 100 V 的市电 II 类型电路。
连接至市电以外电路时,最大瞬态过压额定值是 400 V 峰值。
切勿开盖操作: 请勿在外盖或面板拆除或机壳打开的状态 下操作本产品。可能有危险电压暴露。
远离外露电路: 电源接通后请勿接触外露的接头和元件。
怀疑产品出现故障时,请勿进行操作: 如果怀疑本产品已
损坏,请让合格的维修人员进行检查。
产品损坏时请勿使用。本产品损坏或运行错误时请勿使 用。如果怀疑产品存 线。并做清晰标记以防其再被使用。
在使用之前,请检查电压探头、测试导线和附件是否有机 械损坏,如损坏则予以更换。如果探头或测试导线损坏、 金属外露或出现磨损迹象,请勿使用。
在使用之前请先检查产品外表面。查看是否有裂纹或缺失 部件。
仅使用规定的替换部件。
请勿在潮湿环境下操作: 请注意,如果某个单元从冷处移 到暖处,则可能发
ektronix 指明适合产品使用的绝
该终端的额定电压以上。
在安全问题,请关闭产品并断开电源
生凝结情况。
正确连接并正确断开连接: 将探头连接到被测电路之前, 先将探头输出端连接到测量产品。在连接探头输入端之 前,请先将探头参考导线与被测电路连接。将探头与测量 产品断开之前,请先将探头输入端及探头参考导线与被测 电路断开。
正确连接并正确断开连接: 连接电流探头或断开电流探头 的连接之前请先将被测电路断电。
只能将探头参考导线连
不要将电流探头连接到电压超过电流探头额定电压的任何 导线。
检查探头和附件: 在每次使用之前,请检查探头和附件是 否损坏(探头本体、附件、电缆外壳等的割裂、破损、缺 陷)。如果损坏,请勿使用。
使用以地为参考的示波器: 在使用以地为参考的示波器 时,不要将此探头的参考引线浮地。参考引线必须连接到 大地电势 (0 V)。
浮动测量使用: 不要将此探头的参考引线浮动到额定浮动 电压之上。
接到大地。
本手册中的术语
本手册中可能出现以
警告: “警告”声明指出可能会造成人身伤害或危及
生命安全的情况或操
注意: “注意”声明指出可能对本产品或其他财产造
成损坏的情况或操作
下术语:
作。
产品上的符号和术语
产品上可能出现以下术语:
看到“危险”标记时表示可直接导致人身伤害的危 险。
看到“警告”标记时表示不会直接导致人身伤害的危 险。
看到“注意”标记时 损害的危险。
产品上标示此符号 了解潜在危险的类别以及避免这些危险需采 取的措施。(此符号还可能用于指引用户参 阅手册中的额定值信息。)
产品上可能出现以下符号:
表示会对本产品在内的财产造成
时,请确保查阅手册,以
3
*P0713
071-3578-00
57800*
Tektronix 联系信息
Tektronix, Inc., 14150 SW Karl Braun Drive, P.O. Box 500, Beaverton, OR 97077, USA
有关产品信息、销售、服务和技术支持:
在北美地区 其他地区用户请访问 www.tektronix.com 查找当地的 联系信息。
,请拨打 1-800-833-9200。
重要安全信息
本手册包含用户必须遵守的信息和警告,以确保安全操 作并保证产品安全。请参阅 TDS3000C 系列用户手册(泰 克部件号 077-1397-XX),可从产品随附光盘和泰克网站 (www.tek.com) 中获得。
为保证安全地对本产品进行维修,本部分结尾还提供其他 信息。
常规安全概要
请务必按照规定使用产品。详细阅读下列安全性预防措 施,以避免人身伤害,并防止损坏本产品或与本产品连接 的任何产品。认真阅读所有说明。保留这些说明以备将来 参考。
遵守当地和国家安全法令。 为了保证正确安全地操作产品,除本手册规定的安全性预
防措施外,您还必须遵守普遍公认的安全规程。 产品仅限经过培训的人员使用。 只有了解相关危险的合格人员才能进行开盖维修、保养或
调整。 使用前,请务必检查产品是否来自已知来源,以确保正确
操作。 本产品不适用于检测危险电压。 如果有危
击和强电弧伤害。 使用本产品时,您可能需要使用一套大型系统的其他部
件。有关操作这类系统的警告和注意事项,请阅读其他组 件手册的安全性部分。
将本设备集成到某系统时,该系统的安全性由系统的组装 者负责。
险的带电导体暴露,请使用个人保护装备以防电
请勿在易燃易爆的 请保持产品表面清洁干燥: 清洁本产品前,请移除输入信
号。
请适当通风: 有关 信息,请参阅 TDS3000C 系列用户手册中的安装说明。所 提供的狭槽和开口 体放进任何开口。
提供安全的工作环境: 始终将产品放在方便查看显示器和 指示器的地方。
避免对键盘、指针和按钮盘使用不当或长时间使用。键盘 或指针使用不当或长时间使用可能导致严重损伤。
请确保工作区符合适用的人体工程学标准。请咨询人体工 程学专家,以避免应激损伤。
探头和测试导线
请参阅 TDS3000C 系列用户手册,以了解有关推荐与本产 品配合使用的探头的使用、额定值和选择方面的重要安全 信息。
连接探头或测试导线之前,请将电源线从电源连接器连接 到正确接地的电源插座。
请将手指放在探头上手指防护装置的后面。请勿在探头头 部连接至电源时 电路前确保参考导线或弹簧完全匹配。
拔掉所有不用的探头、测试导线和附件。
仅使用正确的测量类别 (CAT)、电压、温度、海拔高度和 电流额定的探头、测试导线和适配器进行测量。
小心高电压: 了解您正在使用的探头的额定电压,请不要 超出这些额定值。重要的是知道并理解两个额定值:
探头端部到探头
探头参考导线到大地的最大浮动电压
这两个额定电压取决于探头和您的应用。请参阅手册的 “技术规格”部分了解更多详情。
警告: 为防止电击,请不要超出示波器输入 BNC 连
接器、探头端部或探头参考导线的最大测量电压或最 大浮动电压。
环境下操作:
如何安装产品使其保持适当通风的详细
用于通风,不得遮盖或阻挡。请勿将物
触摸其金属部分,并在将探头连接至被测
参考导线的最大测量电压
避免火灾或人身伤害
使用合适的电源线: 只能使用本产品专用并经所在国家/
地区认证的电源线。 不要使用为其他产品提供的电源线。
x
合规性信息
此部分列出仪器遵循的 EMC(电磁兼容性)、安全和环境 标准。
有关合规性信息相关问题,可以联系以下地址:
Tektronix, Inc.PO Box 500, MS 19‐045 Beaverton, OR 97077, USA www.tek.com
EMC 合规性
EU EMC 指令
符合 Directive 2014/30/EU 有关电磁兼容性的要求。已 证明符合《欧洲共同体公报》中所列的以下技术规格:
EN 61326-1、EN 61326-2-1: 测量、控制和实验室用电 气设备 EMC 要求。
CISPR 11。 放射和传导辐射量,组 1,A 类
IEC 61000-4-2。 对静电放电的抗干扰能力
IEC 61000-4-3。 对射频电磁场的抗干扰能力
IEC 61000-4-4。 对电快速瞬态/突发性的抗干扰能 力
IEC 61000-4-5。 对电源线电涌的抗干扰能力
IEC 61000-4-6。 对传导射频的抗干扰能力
IEC 61000-4-11。 对电压骤降和中断的抗干扰能力
EN 61000-3-2: 交流电源线谐波辐射
EN 61000-3-3: 电压变化、波动和闪变
1
本产品仅在非居民区内使用。在居民区内使用可能造成电磁 干扰。
2
当该设备与测试对象连接时,可能产生超过此标准要求的辐 射级别。
3
测试导线和/或测试探头由于电子干扰耦合而发生连接时,设备 可能无法满足所列适用标准的抗干扰能力要求。为了将电磁干扰 的影响降到最低,需最小化信号无屏蔽部分与关联返回导线之 间的环路面积,同时尽量让导线远离电磁干扰源。将未屏蔽的 测试导线缠绕在一起是减小环路面积的有效方法。探头方面, 需要使接地回路导线的长度尽可能得短,并靠近探头主体。为 了最有效地达到这一目的,一些探头配备了附件探头端部适配 器。在一切情况下,都应遵守所用探头或导线的所有安全说明。
4
为确保符合上面列出的 EMC 标准,应使用在电缆护套和连接器 外壳间包含低阻抗连接的高质量屏蔽接口电缆。
5
当仪器受到 IEC 61000-4-3 测试所定义的电磁场和信号的影响 时,允许 ≤3.0 格波形位移或峰-峰值噪声增加 ≤6.0 格。
1234
5
6
6
当仪器受到 IEC 61000-4-6 测试所定义的电磁场和信号的影响 时,允许 ≤1.5 格波形位移或峰-峰值噪声增加 ≤2.0 格。
澳大利亚/新西兰符合性声明 – EMC
根据 ACMA,符合 Radiocommunications Act(《无线电通 信法》)有关 EMC 规定的以下标准:
CISPR 11。 放射和传导发射量,组 1,A 类,依照 EN 61326- 1 和 EN 61326-2-1。
安全合规性
该部分列出了安全合规性信息。
污染度评级
污染度 2(如 IEC 61010-1 中定义)。仅适合在室内的 干燥场所使用。
测量和过压类别说明
本产品上的测量端子可能适合测量以下类别的市电电压 (请参阅产品和手册中标示的具体额定值)。
类别 II。 电路使用点(插座和类似点处)直接连接 到建筑物布线。
说明: 仅市电电源电路具有过压类别额定值。仅测量电
路具有测量类别额定值。产品中的其他电路不具有其中任 何一种额定值。
主线过压类别额定值
过压类别 II(如 IEC 61010-1 中的定义)。
环境注意事项
有关产品的环境影响,请参阅用户手册。
操作概述
有关完整的操作信息和产品规格,请参阅用户手册。
电源
请使用 100 VAC至 240 VAC(±10%)、 频 率 在 47 Hz 至 440 Hz 之间的电源。
本示波器使用带接地或近地中性导线的单相电源。该仪器 的用途仅限于以地为参考的测量。通过电源线中的接地导 线提供保护性接地对于安全操作十分重要。
线路保险丝在内部,操作人员不能更换。
表 5: Maximum voltage ratings
额定值 说明
输入 BNC 处 最大电 压(1 MΩ)
过压类别 最大电压
非市电环境 150 V
RMS
空系数为 37.5%)
CAT II 环境
100 VRMS(400 Vpk,
(400 VPK,占
占空系数为 25%)
对于稳态正弦波形,从大于 200 kHz 开始以 20 dB/10 倍 频程下降 到 大于等于 3 MHz 时的 13 Vpk。
探头接
0V
地基准 导线
表 6: Environmental ratings
额定值 说明
工作温度
0 °C 至 +50 °C(+32 °F 至 +122 °F)
非工作范围(存储):
-40 °C 至 +71 °C ( -40 °F 至 +160 °F)
工作湿度 低于 30 °C (+86 °F) 时为 5% 至 95%
RH,50°C (+122 °F) 时上限下降至 45% RH
工作海拔 最高 3000 米(10,000 英尺)
所有 © Tektronix, Inc. 保留所有利。 www.tek.com